Continuous cardiac output measurement: pulse contour analysis vs thermodilution technique in cardiac surgical patients
|
|
- Polly Morris
- 5 years ago
- Views:
Transcription
1 British Journal of Anaesthesia 82 (4): (1999) Continuous cardiac output measurement: pulse contour analysis vs thermodilution technique in cardiac surgical patients G. Rödig*, C. Prasser, C. Keyl, A. Liebold and J. Hobbhahn Department of Anaesthesia, University Hospital, Franz-Josef-Strauss Allee 11, D Regensburg, Germany *To whom correspondence should be addressed We have analysed the clinical agreement between two methods of continuous cardiac output measurement: pulse contour analysis (PCCO) and a continuous thermodilution technique (CCO), were both compared with the intermittent bolus thermodilution technique (BCO). Measurements were performed in 26 cardiac surgical patients (groups 1 and 2, 13 patients each, with an ejection fraction 45% and 45%, respectively) at 12 selected times. During operation, mean differences (bias) between PCCO BCO and CCO BCO did not differ in either group. However, phenylephrine-induced increases in systemic vascular resistance (SVR) by approximately 60% resulted in significant differences. Significantly higher absolute bias values of PCCO BCO compared with CCO BCO were also found early after operation in the ICU. Thus PCCO and CCO provided comparable measurements during coronary bypass surgery. After marked changes in SVR, further calibration of the PCCO device is necessary. Br J Anaesth 1999; 82: Keywords: measurement techniques, cardiac output; measurement techniques, pulse contour analysis; measurement techniques, thermodilution; surgery, cardiovascular Accepted for publication: October 19, 1998 Continuous measurement of cardiac output (CO) is preferable to intermittent measurement, especially during rapid changes in cardiovascular function in critically ill patients. A nearly continuous measurement of CO, using a thermodilution technique generating thermal pulses by a heating filament attached to a pulmonary artery catheter, produces a clinically acceptable level of accuracy compared with the intermittent bolus technique. 1 8 A new system of pulse contour analysis is available which is less invasive than continuous CO monitoring and may cost less. The method of calculating stroke volume from the contour of the arterial pressure curve dates from 1899 and was resurrected by advancements in computer technology in the 1980s. 9 Arterial pulse pressure waveform analysis involves measuring the area under the systolic portion of the arterial pulse wave from the end of diastole to the end of the ejection phase, together with an individual calibration factor to account for individual vascular impedance. Cardiac output, determined by arterial thermodilution, is used for calibration of this device which uses an enhanced version of the algorithm of Wesseling and colleagues for analysis of the arterial pulse pressure curve We have compared two methods of continuous CO measurement: pulse contour analysis (PiCCO, Pulsion Medical Systems, Munich, Germany) and a continuous thermodilution technique (Opti-Q SvO 2 / CCO, Abbott Critical Care Systems, Moutain View, CA, USA), with conventional intermittent thermodilution measurement in cardiac surgical patients with both good and poor ventricular performance. Patients and methods After obtaining approval from the Ethics Committee and written informed consent, we studied 26 patients (22 males), aged yr (mean 62 yr), undergoing elective coronary bypass surgery. Thirteen patients with an ejection fraction of 45% were allocated to group 1 and 13 patients with an ejection fraction 45% to group 2. Patients with valvular heart disease, intracardiac shunts or peripheral vascular disease were excluded. Preoperative ECG showed sinus rhythm without arrhythmia. Patients were anaesthetized with fentanyl, etomidate and pancuronium; fentanyl and isoflurane were used for maintenance of anaesthesia. Mechanical ventilation was set (FI O2 0.5, ventilatory frequency 10 bpm) to give an endtidal PCO 2 of kpa throughout the study. In all patients, a 4-French gauge arterial thermodilution catheter (PiCCO, Pulsion Medical Systems, Munich, Germany) was inserted via the femoral artery for clinical monitoring of arterial pressure, continuous CO (PCCO) measurements derived from arterial pressure and intermittent arterial thermodilution CO measurements. A pulmonary artery British Journal of Anaesthesia
2 Rödig et al. catheter (Opti-Q SvO 2 /CCO, 8F) was also inserted via an 8.5-French gauge introducer, into the right internal jugular vein for intermittent conventional pulmonary artery thermodilution CO and continuous CO (CCO) measurement. The arterial catheter was connected to a computer for pulse contour analysis (Pulsion Medical Systems, Munich, Germany). To calibrate this system, individual arterial input impedance to arterial pressure is calculated by simultaneously determining the area under the systolic portion of the arterial pulse wave (A sys ) and measuring arterial thermodilution CO. The mean of three subsequent measurements, randomized within the respiratory cycle, was used. These were performed by injection of 10 ml of iced saline solution via a central venous catheter with subsequent detection by the thermistor embedded into the wall of the arterial catheter. PCCO is computed by multiplying stroke volume by heart rate. A moving average of the preceding 30 s is calculated and displayed on the monitor. The passage of the PAC to the pulmonary artery, with its thermal filament connected to the cardiac computer (Q-vue CCO), was accomplished by monitoring the pressure waveform from the distal port of the catheter. This catheter has a lumen (DTPP lumen, distal thermal coil positioning port) at a distance of 1 cm from the tip of the thermal filament to check if its position is correct within the right ventricle by waveform analysis. Intracardiac and pulmonary artery pressures were monitored continuously to ensure correct placement of the catheter. The displayed CCO reflected either the average flow of the previous 3 min ( fast mode ) or of the previous s ( urgent mode ). Intermittent pulmonary artery thermodilution CO measurements (BCO) were performed by manual injection of iced saline solution. The average of three measurements, measured within a 10% range, distributed randomly over the respiratory cycle, was calculated. Furthermore, mixed venous blood (PAC tip) and rectal (Sirecust 1281 monitor, Siemens, Erlangen, Germany) temperatures were measured continuously and recorded at each time during cardiac surgery. A standard cardiopulmonary bypass (CPB) technique was used in all patients. Non-pulsatile perfusion at a pump flow rate of 2.4 litre min 1 m 2 was used during mild hypothermia (arterial temperatures C). Before disconnection from CPB, patients were rewarmed to a minimum rectal temperature of 36 C. PCCO, CCO and BCO measurements were compared after induction of anaesthesia, before surgical incision (T1), after the start of operation when dissection of the saphenous vein was performed (T2), after sternotomy during preparation of the internal mammary artery (T3), after pericardiotomy (T4), and 5 min (T5), 15 min (T6), 30 min (T7) and 45 min (T8) after weaning from CPB. Measurements were performed during stable surgical stimulation where the surgeons paused if necessary. After admission to the ICU, further sets of measurements were obtained during the subsequent 2 h (T9 T12). Sedation was not given in the postoperative period, according to the standard procedures of this institution. At each time, simultaneous readings of CO values, displayed on the pulse contour computer and on the Q-vue CCO-computer, were compared with BCO measurements obtained immediately after switching the Q-vue CCO-monitor from the continuous fast mode to the bolus mode. If PCCO changed by more than 15% during the pulmonary thermodilution period, CO values were rejected and comparison of measurements was repeated. Initial calibration of the pulse contour analysis system was performed before initiation of the test sequence at T1. This system was recalibrated immediately after weaning from CPB (before T5) and after admission to the ICU (before T9). In patients requiring intervention because of low systemic vascular resistance (MAP 55 mm Hg, CI 2.5 litre min 1 m 2 ), at the end of operation and before admission to the ICU, an i.v. infusion of phenylephrine was given in a standardized manner to restore arterial pressure to MAP values 80 mm Hg within approximately 20 min. PCCO (recalibrated), CCO (urgent mode) and BCO measurements were obtained before starting administration of phenylephrine, and after 20 min of administration, when an increase in arterial pressure to 80 mm Hg had been achieved. Statistical analysis was performed using the method described by Bland and Altman. 12 Bias between the methods was calculated as the mean difference between PCCO and BCO, and between CCO and BCO. The upper and lower limits of agreement were calculated as bias 2SD, and defined the range in which 95% of the differences between the methods were expected to lie. The precision of the bias analysis and limits of agreement were assessed using 95% confidence intervals (CI). The relative error, defined as 100 (BCO PCCO)/[(BCO PCCO)/2] and 100 (BCO CCO)/[(BCO CCO)/2], was calculated. To compensate for the variability of the samples, absolute values of these differences were also calculated. Student s t test for paired samples was used to compare bias, relative error at each time for groups 1 and 2, and inter-group comparisons. The Wilcoxon paired sample test was used to compare absolute differences which were not normally distributed. The level of significance was corrected according to Bonferroni to compensate for the effect of multiple comparisons. Bias between PCCO and BCO, and CCO and BCO, respectively, before and after 20 min of administration of phenylephrine, was analysed using the paired Student s t test. P 0.05 was considered significant. Results There were 308 comparative measurements performed between PCCO and BCO, and 312 measurements between CCO and BCO in 26 patients. In one patient, PCCO was not available from T5 to T8 because of prominent diastolic waves recorded in the iliac artery which the pulse contour analysis monitor was unable to differentiate from the systolic peak. The range of PCCO values was litre min 1 526
3 Pulse contour vs thermodilution cardiac output Table 1 Pulse contour cardiac output (PCCO) and continuous cardiac output (CCO) measurements (mean (SD)) at sample times T1 T12 in group 1 (ejection fraction 45%) and group 2 (ejection fraction 45%) Group 1 Group 2 PCCO CCO PCCO CCO (litre min 1 ) (litre min 1 ) (litre min 1 ) (litre min 1 ) T1 4.5 (1.8) 4.1 (1.6) 3.9 (1.4) 3.4 (1.1) T2 4.4 (1.6) 4.7 (1.3) 4.3 (1.4) 3.8 (1.1) T3 4.2 (1.9) 4.4 (1.5) 4.3 (1.0) 4.2 (1.1) T4 4.3 (1.7) 4.6 (1.5) 4.2 (0.9) 3.9 (0.8) T5 6.8 (2.4) 6.5 (2.2) 6.7 (1.5) 6.6 (1.6) T6 6.2 (2.1) 6.4 (1.9) 6.6 (1.3) 6.9 (1.6) T7 6.1 (2.4) 6.6 (1.6) 6.5 (1.0) 6.9 (1.7) T8 6.0 (2.0) 6.5 (2.0) 6.6 (1.0) 6.4 (1.3) T9 6.7 (1.5) 6.6 (1.8) 6.4 (1.2 ) 6.8 (1.4) T (1.5) 6.6 (1.6) 7.0 (2.2) 6.9 (1.4) T (2.3) 7.0 (1.4) 7.0 (1.2) 6.9 (1.4) T (2.1) 6.8 (1.4) 7.0 (1.3) 7.4 (1.4) in group 1 and litre min 1 in group 2. CCO values were litre min 1 in group 1 and litre min 1 in group 2. Mean CO values, measured by PCCO and CCO for the different sample times in both groups, are presented in Table 1. There were no significant differences between CO values, assessed by PCCO and CCO, at any time in either group, nor in PCCO and CCO measurements, respectively, between groups. Differences between PCCO and BCO, and between CCO and BCO, were plotted against the mean of the two methods in both groups (Fig. 1). The results of the analysis of agreement, assessed by bias, and the distribution of the observed differences, indicated by bias 2SD as upper and lower limits of agreement, including 95% confidence intervals at each time, are presented in Tables 2 and 3. Differences between PCCO and BCO values were not significantly different from differences between CCO and BCO at any time in either group. Analysis of the absolute values of the differences showed significantly higher biases of PCCO BCO compared with CCO BCO at T10 and T12 in group 1 and at T12 in group 2. The relative error was within 15% for 64% of comparisons between PCCO and BCO in group 1, and for 76% in group 2, compared with 86% in group 1 and 80% in group 2 of the differences between CCO and BCO. The absolute values of the relative error were significantly higher for PCCO and BCO compared with CCO and BCO measurements at T10 and T12 in group 1 and at T12 in group 2. Ten patients received phenylephrine to restore arterial pressure as a result of low systemic vascular resistance (SVR). Phenylephrine increased SVR by 62 (SD 29)% (from 732 (190) to 1153 (264) dyn s cm 5 m 2 ) after 20 min of administration. CO, assessed by the bolus technique, did not change significantly 20 min after administration of phenylephrine (5.7 (1.1) vs 6.1 (0.9) litre min 1 ). PCCO BCO bias did not differ from CCO BCO bias before the start of administration of phenylephrine. Phenylephrineinduced increases in SVR resulted in significantly higher PCCO BCO bias compared with CCO BCO after 20 min (see Fig. 2). From T1 to T4, mean pulmonary artery blood temperature (35.5 (SD 0.6) to 34.9 (0.5) C in group 1 and 35.7 (0.4) to 35.0 (0.6) C in group 2) and rectal temperature (35.8 (0.8) to 35.3 (0.7) C in group 1 and 35.7 (0.9) to 35.1 (1.0) C in group 2) decreased in both groups. After CPB, mean blood temperature (37.1 (0.4) to 36.6 (0.4) C in group 1 and 37.1 (0.7) to 36.6 (0.5) C in group 2) and rectal temperature (37.0 (0.9) to 36.4 (0.9) C in group 1 and 36.8 (0.7) to 36.3 (0.9) C in group 2) decreased until the end of operation. After admission to the ICU, only pulmonary blood temperatures were recorded. They increased during the first 2 h (from T9 to T12) in the ICU (36.4 (0.7) to 36.9 (0.8) C in group 1 and 36.5 (0.7) to 36.9 (1.0) C in group 2). The 26 patients were operated on in a uniform manner. Mean duration of CPB was 105 (SD 32) min in group 1 and 105 (20) min in group 2. Aortic cross-clamp time was 65 (18) min and 64 (15) min, respectively. A mean of 3.5 bypasses were constructed. In all patients the left internal mammary artery was dissected in order for it to be used as one of the bypass vessels. Weaning from CPB was achieved in all patients by administration of dopamine 3 7 g kg 1 min 1. Seven patients in group 2 required additional infusions of epinephrine to increase inotropic support. All patients had sinus rhythm after weaning from CPB. Volume infusions were administered in accordance with individual patient needs via a peripheral venous cannula. No adverse effects of either the PiCCO catheter or the Opti-Q SvO 2 /CCO catheter were observed. Discussion Previous studies have reported a small mean difference (bias) and limits of agreement (bias 2SD), considered clinically acceptable, between CCO, based on the pulsed warm thermodilution technique, and BCO measurements. 1 8 The BCO technique was used as the reference because it is a widespread clinical means of measuring CO despite its poor reproducibility and a variability of 15% under clinical situations Our results comparing these two methods in cardiac surgical patients with good and reduced ventricular performance were similar to others, 1 7 including early after mild hypothermic bypass. 8 However, factors such as uneven rewarming after CPB, indicated by opposite changes in blood and rectal temperatures (not the case in our patients, who were rewarmed to a minimum rectal temperature of 36 C at the end of CPB) or delivery of large volumes of cold fluids may affect the reliability of the continuous thermodilution CO measurement Arterial pulse contour analysis may be a less invasive alternative requiring an arterial thermodilution catheter and a central venous line for indicator injection. 20 The Wesseling algorithm, 10 which involves measuring the area under the systolic portion of the arterial pulse wave (A sys ) from the end of diastole to the end of the ejection phase divided by 527
4 Rödig et al. Fig 1 Bland and Altman plot for comparison between PCCO and BCO, and between CCO and BCO in groups 1 and 2 for all measurements. Solid lines represent bias 2SD. Table 2 Mean difference between PCCO BCO and between CCO BCO (bias), and lower and upper limits of agreement (bias 2SD), together with 95% confidence intervals (in parentheses) at sample times T1 T12 in group 1 (ejection fraction 45%) PCCO BCO (litre min 1 ) CCO BCO (litre min 1 ) Bias Lower limit Upper limit Bias Lower limit Upper limit T ( ) 2.27 ( 2.60 to 1.90) 2.89 ( ) 0.02 ( ) 1.59 ( 1.83 to 1.35) 1.55 ( ) T ( ) 1.81 ( 2.05 to 1.57) 1.45 ( ) 0.29 ( ) 0.96 ( 1.13 to 0.79) 1.54 ( ) T ( ) 1.86 ( 2.10 to 1.62) 1.46 ( ) 0.01 ( ) 1.65 ( 1.89 to 1.41) 1.63 ( ) T ( ) 1.62 ( 1.84 to 1.40) 1.24 ( ) 0.18 ( ) 0.70 ( 0.85 to 0.55) 1.06 ( ) T ( ) 2.06 ( 2.34 to 1.78) 1.98 ( ) 0.30 ( ) 1.80 ( 2.00 to 1.58) 1.20 ( ) T ( ) 1.85 ( 2.13 to 1.57) 2.23 ( ) 0.07 ( ) 2.03 ( 2.34 to 1.72) 2.20 ( ) T ( ) 3.25 ( 3.69 to 2.81) 2.63 ( ) 0.13 ( ) 1.85 ( 2.13 to 1.57) 2.11 ( ) T ( ) 3.21 ( 3.67 to 2.75) 3.11 ( ) 0.29 ( ) 1.12 ( 1.32 to 0.92) 1.70 ( ) T ( ) 1.31 ( 1.57 to 1.05) 2.15 ( ) 0.31 ( ) 1.26 ( 1.50 to 1.02) 1.88 ( ) T ( ) 2.37 ( 2.89 to 1.85) 4.77 ( ) 0.02 ( ) 0.78 ( 0.89 to 0.67) 0.82 ( ) T ( ) 2.94 ( 3.46 to 2.42) 4.42 ( ) 0.14 ( ) 1.37 ( 1.59 to 1.15) 1.32 ( ) T ( ) 2.94 ( 3.42 to 2.46) 3.52 ( ) 0.28 ( ) 0.88 ( 0.99 to ( ) aortic impedance (Z ao ) to provide a measure of stroke volume, has proved superior to other approaches in calculating CO from the contour of the arterial pulse wave. 21 In an enhanced version, Wesseling and colleagues 11 based the pressure flow relationship on a three-element model of arterial input impedance to arterial pressure to compute aortic flow from pressure. The algorithm uses the three major properties of the aorta and arterial system (i.e. aortic characteristic impedance, windkessel compliance of the arterial system and peripheral vascular resistance) to com- pute a flow pulsation from an arterial pressure pulsation. Wesseling and colleagues reported a bias of 0.09 litre min 1 and precision (SD) of 0.36 in eight cardiac surgical patients. 11 Others, using the above approach, found biases of 0.06 (0.93) litre min 1, (1.56) litre min 1 21 and 0.09 (0.85) litre min 1 23 in intensive care patients. The results of PCCO measurement improved to a bias of litre min 1 and a precision of 0.56 litre min 1 if additional calibration was used. 23 A radial artery was used for pressure monitoring in all patients, except for five patients where a femoral 528
5 Pulse contour vs thermodilution cardiac output Table 3 Mean difference between PCCO BCO and between CCO BCO (bias), and lower and upper limits of agreement (bias 2SD), together with 95% confidence intervals (in parentheses) at sample times T1 T12 in group 2 (ejection fraction 45%) PCCO BCO (litre min 1 ) CCO BCO (litre min 1 ) Bias Lower limit Upper limit Bias Lower limit Upper limit T ( ) 1.16 ( 1.40 to 0.92) 2.04 ( ) 0.09 ( ) 1.49 ( 1.69 to 1.29) 1.31 ( ) T ( ) 1.62 ( 1.93 to 1.31) 2.58 ( ) 0.10 ( ) 1.35 ( 1.52 to 1.18) 1.15 ( ) T ( ) 0.99 ( 1.19 to 0.79) 1.67 ( ) 0.20 ( ) 0.96 ( 1.13 to 0.79) 1.36 ( ) T ( ) 0.85 ( 1.00 to 0.70) 1.31 ( ) 0.05 ( ) 1.27 ( 1.44 to 1.10) 1.17 ( ) T ( ) 2.34 ( 2.69 to 1.99) 2.52 ( ) 0.02 ( ) 1.41 ( 1.63 to 1.20) 1.45 ( ) T ( ) 2.50 ( 2.85 to 2.15) 2.18 ( ) 0.05 ( ) 1.75 ( 2.01 to 1.49) 1.85 ( ) T ( ) 2.01 ( 2.29 to 1.73) 1.83 ( ) 0.29 ( ) 1.95 ( 2.28 to 1.62) 2.53 ( ) T ( ) 1.87 ( 3.67 to 2.75) 3.11 ( ) 0.10 ( ) 2.22 ( 2.53 to 1.91) 2.02 ( ) T ( ) 1.86 ( 2.12 to 1.60) 1.82 ( ) 0.42 ( ) 0.73 ( 0.90 to 0.56) 1.57 ( ) T ( ) 2.66 ( 3.10 to 2.22) 3.32 ( ) 0.22 ( ) 1.29 ( 1.51 to 1.07) 1.73 ( ) T ( ) 2.88 ( 3.29 to 2.47) 2.68 ( ) 0.29 ( ) 2.19 ( 2.45 to 1.93) 1.61 ( ) T ( ) 3.54 ( 4.04 to 3.04) 3.26 ( ) 0.19 ( ) 0.76 ( 0.89 to 0.63) 1.14 ( ) Fig 2 PCCO BCO and CCO BCO bias ( SD) before and 20 min after the start of administration of phenylephrine in 10 patients. *P Broken line represents an ideal bias of zero. artery was used. 23 In all of the studies, concern was expressed as to whether greater degrees of hyper- or hypotension ( 25% 22 or 50% 23 ) would result in significant alterations in arterial input impedance to arterial pressure, requiring recalibration of the system. Aortic impedance and arterial compliance are non-linear, pressure-dependent properties of the arterial system. The third element, total systemic peripheral resistance (SVR), according to Wesseling s model of arterial impedance, is a time-varying property of the vascular bed. As actual aortic impedance, arterial compliance and peripheral vascular resistance for an individual patient are unknown, CO, determined by a separate method, is used as a calibration factor. Measurement of CO by arterial thermodilution, used for calibration of the device we used, is a reliable alternative to conventional CO measurement via a pulmonary artery cather. 24 An enhanced version of the Wesseling algorithm, which has not yet been published by the manufacturer, was applied to the pulse contour analysis system we used with a refined correction factor to reduce the effects of changing mean arterial pressure on arterial impedance. Comparison of PCCO, provided by this new device, using femoral arterial access, to conventional BCO measurement showed a bias of litre min 1 and precision of 1.10 in experimental animals. 25 In cardiac surgical patients, Rauch and colleagues found a bias of (1.15) litre min 1 between PCCO and BCO compared with a value of 0.4 (1.25) litre min 1 between CCO and BCO. 26 Our results on the agreement, assessed by bias and upper and lower limits of agreement between PCCO and conventional BCO determination, were comparable with others When PCCO, recalibrated after CPB, and the CCO measurement were compared with standard BCO determination, they did not differ significantly during coronary bypass surgery, either in patients with good ventricular performance or in those requiring an additional infusion of epinephrine after weaning from CPB. Phenylephrine-induced increases in systemic vascular resistance by approximately 60%, however, resulted in a significantly greater PCCO BCO bias compared with CCO BCO. After operation, the absolute differences and absolute values of the relative error between PCCO and BCO were significantly higher than between CCO and BCO measurements at T10 and T12 in group 1 and at T12 in group 2. As sedation was not given in the postoperative period, short-term episodes such as increases in arterial pressure occurred during early recovery of anaesthesia, where patients underwent continuous mechanical ventilation. Our data suggest that rapid haemodynamic changes during the first 2 h in the ICU, such as increases in arterial pressure caused by increases in systemic vascular resistance, may account for the observed differences in the degree of agreement between PCCO and CCO, compared with BCO. The peripheral pulse wave reflects the interplay between left ventricular output and capacitance of the vascular tree. 27 The ability to trend changes in CO, despite changes in peripheral resistance, is ascribed to factors in the pulse contour algorithms that compensate for changes in arterial characteristics. We found that although changes in vascular tone of approximately 20% had no effect on the pulse contour method, greater changes in arterial pressure of 529
6 Rödig et al. 50% may affect PCCO measurement. When marked changes in systemic vascular resistance are expected after administration of vasoactive drugs or during recovery from anaesthesia, then our data indicate that the device should be recalibrated to compensate for changes in vascular resistance. In summary, assessment of continuous CO by pulse wave form analysis, and by a thermodilution technique, provided comparable measurements during coronary bypass surgery as long as significant changes in systemic vascular resistance did not occur. Recalibration of the PCCO device is necessary after marked changes in systemic vascular resistance. References 1 Böttiger BW, Soder M, Rauch H, et al. Semi-continuous injectate cardiac output measurement in intensive care patients after cardiac surgery. Intensive Care Med 1996; 22: Boldt J, Menges T, Wollbruck M, Hammermann H, Hempelmann G. Is continuous cardiac output measurement using thermodilution reliable in the critically ill patient? Crit Care Med 1994; 22: Burchell SA, Yu M, Takiguchi SA, Ohta RM, Myers SA. Evaluation of a continuous cardiac output and mixed venous oxygen saturation catheter in critically ill surgical patients. Crit Care Med 1997; 25: Jakobsen CJ, Melsen NC, Andresen EB. Continuous cardiac output measurements in the perioperative period. Acta Anaesthesiol Scand 1995; 39: Jacquet L, Hanique G, Glorieux D, Matte P, Goenen M. Analysis of the accuracy of continuous thermodilution cardiac output measurement. Intensive Care Med 1996; 22: Lefrant JY, Bruelle P, Ripart J, et al. Cardiac ouput measurement in critically ill patients: comparison of continuous and conventional thermodilution techniques. Can J Anaesth 1995; 42: Yelderman ML, Ramsya MA, Quinn MD, Paulsen AW, McKnown RC, Gillman PH. Continuous thermodilution cardiac output measurement in intensive care unit patients. J Cardiothorac Vasc Anesth 1992; 6: Rödig G, Keyl C, Liebold A, Hobbhahn J. Intra-operative evaluation of a continuous versus intermittent bolus thermodilution technique of cardiac output measurement in cardiac surgical patients. Eur J Anaesthesiol 1998; 15: Nelson LD. The new pulmonary arterial catheters. Crit Care Clin 1996; 12: Wesseling KH, de Wit B, Weber JAP, Ty Smith N. A simple device for the continuous measurement of cardiac output. Adv Cardiovasc Physiol 1983; 5: Wesseling KH, Jansen JRC, Settels JJ, Schreuder JJ. Computation of aortic flow from pressure in humans using a nonlinear, threeelement model. J Appl Physiol 1993; 74: Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 1: LeTulzo Y, Belgith M, Seguin P, et al. Reproducibility of thermodilution cardiac output in critically ill patients: comparison between bolus and continuous method. J Clin Monit 1996; 12: Taylor SH, Silke B. Is the measurement of cardiac output useful in clinical practice? Br J Anaesth 1988; 60: 90 8S 15 Stetz CW, Miller RG, Kelly GE, Raffin TA. Reliability of the thermodilution method in the determination of cardiac output in clinical practice. Am Rev Respir Dis 1982; 126: Latson TW, Whitten CW, O Flaherty D. Ventilation, thermal noise, and errors in cardiac output measurements after cardiopulmonary bypass. Anesthesiology 1993; 79: Nishikawa T, Dohi S. Errors in the measurement of cardiac output by thermodilution. Can J Anaesth 1993; 40: Böttiger BW, Rauch H, Böhrer H, et al. Continuous versus intermittent cardiac output measurement in cardiac surgical patients undergoing hypothermic cardiopulmonary bypass. J Cardiothorac Vasc Anesth 1995; 9: Böttiger BW, Sinner B, Motsch J, Bach A, Bauer H, Martin E. Continuous versus intermittent thermodilution cardiac output measurement during orthotopic liver transplantation. Anaesthesia 1997; 52: Gödje O, Höke K, Fischlein T, Vetter H, Reichart B. Less invasive, continuous cardiac-output measurement through pulse contour analysis versus conventional thermal dilution. Intensive Care Med 1996; 22: 58S 21 Apenburg C, Specht M. Continuous cardiac-output measurement by pulse contour analysis. Intensive Care Med 1996; 22 (Suppl. 3): Weissman C, Ornstein EJ, Young WL. Arterial pulse contour analysis trending of cardiac output. Hemodynamic manipulations during arteriovenous malformation resection. J Clin Monit 1993; 9: Irlbeck M, Forst H, Briegel J, Haller M, Peter K. Continuous measurement of cardiac output using pulse contour analysis. Anaesthesist 1995; 44: Spiegel T, Wietasch G, Bürsch J, Hoeft A. Cardiac output measurement by transpulmonary indicator dilution technique. Anaesthesist 1996; 45: Faltlhauser A, Erhardt W, Gödje O, Reichart B, Pfeiffer U. Accuracy of a less invasive device for rapid beat-to beat continuous cardiac output monitoring. Intensive Care Med 1996; 22: (Suppl. 3): Rauch H, Böttiger BW, Motsch J, Müller M, Fleischer F, Martin E. Pulse contour cardiac output measurement corresponds better with standard dilution technique than continuous thermodilution technique after hypothermic bypass. Br J Anaesth 1997; 78 (Suppl. 2): O Rourke MF, Yaginuma T. Wave reflections and the arterial pulse. Arch Intern Med 1984; 144:
Less Invasive, Continuous Hemodynamic Monitoring During Minimally Invasive Coronary Surgery
Less Invasive, Continuous Hemodynamic Monitoring During Minimally Invasive Coronary Surgery Oliver Gödje, MD, Christian Thiel, MS, Peter Lamm, MD, Hermann Reichenspurner, MD, PhD, Christof Schmitz, MD,
More informationAt present, cardiac output is usually determined
Original Article / Transplantation Continuous versus bolus cardiac output monitoring during orthotopic liver transplantation Fang-Ping Bao and Jian Wu Hangzhou, China BACKGROUND: Cardiac output monitoring
More informationButton et al. using a thermistor-tipped catheter. The CO is calculated using an algorithm based on the area under the systolic part of the pressure wa
British Journal of Anaesthesia 99 (3): 329 36 (2007) doi:10.1093/bja/aem188 Advance Access publication on July 12, 2007 CARDIOVASCULAR Clinical evaluation of the FloTrac/Vigileo TM system and two established
More informationA pilot assessment of the FloTrac TM cardiac output monitoring system
Intensive Care Med (2007) 33:344 349 DOI 10.1007/s00134-006-0410-4 TECHNICAL NOTE Helen Ingrid Opdam Li Wan Rinaldo Bellomo A pilot assessment of the FloTrac TM cardiac output monitoring system Received:
More informationTHE MEASUREMENT of cardiac output (CO) is a parameter
Comparison of Esophageal Doppler, Pulse Contour Analysis, and Real-Time Pulmonary Artery Thermodilution for the Continuous Measurement of Cardiac Output Berthold Bein, MD,* Frank Worthmann, MD,* Peter
More informationThe Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for
1 2 The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for continuous central venous oximetry (ScvO2) 3 The Vigileo
More informationEdwards FloTrac Sensor & Performance Assessments of the FloTrac Sensor and Vigileo Monitor
Edwards FloTrac Sensor & Edwards Vigileo Monitor Performance Assessments of the FloTrac Sensor and Vigileo Monitor 1 Topics System Configuration Performance and Validation Dr. William T. McGee, Validation
More informationPerioperative monitoring of cardiac
Continuous cardiac output monitoring with pulse contour analysis: A comparison with lithium indicator dilution cardiac output measurement James Pittman, MBBS, FRCA; Shahar Bar-Yosef, MD; John SumPing,
More informationImpedance Cardiography (ICG) Method, Technology and Validity
Method, Technology and Validity Hemodynamic Basics Cardiovascular System Cardiac Output (CO) Mean arterial pressure (MAP) Variable resistance (SVR) Aortic valve Left ventricle Elastic arteries / Aorta
More informationChapter 3. An evaluation of cardiac output by five arterial pulse contour techniques during cardiac surgery
Chapter 3 An evaluation of cardiac output by five arterial pulse contour techniques during cardiac surgery R.B.P. de Wilde 1, J.J. Schreuder 2, P.C.M. van den Berg 1 and J.R.C. Jansen 1 1 Department of
More informationHemodynamic Monitoring and Circulatory Assist Devices
Hemodynamic Monitoring and Circulatory Assist Devices Speaker: Jana Ogden Learning Unit 2: Hemodynamic Monitoring and Circulatory Assist Devices Hemodynamic monitoring refers to the measurement of pressure,
More informationCARDIOVASCULAR Continuous cardiac output during off-pump coronary artery bypass surgery: pulse-contour analyses vs pulmonary artery thermodilution
British Journal of Anaesthesia 99 (4): 484 92 (2007) doi:10.1093/bja/aem199 Advance Access publication July 24, 2007 CARDIOVASCULAR Continuous cardiac output during off-pump coronary artery bypass surgery:
More informationCardiac Output Monitoring - 6
Cardiac Output Monitoring - 6 How to use Wrexham s Cardiac Output Monitors. Wrexham Maelor Critical Care Version 02.05.16 Introduction Types of Devices: NICOM - Cheetah Oesophageal Doppler +/- Pulse Contour
More informationMonitoring of cardiac index (CI) after surgery for congenital heart
Cardiac index monitoring by pulse contour analysis and thermodilution after pediatric cardiac surgery U. Fakler, MD, a Ch. Pauli, MD, a G. Balling, MD, a H. P. Lorenz, MD, a A. Eicken, MD, a M. Hennig,
More informationESTIMATION OF STROKE VOLUME BY NON-INVASIVE BLOOD PRESSURE MONITORING IN HUMAN CENTRIFUGE
84TH ASMA ANNUAL SCIENTIFIC MEETING CHICAGO - MAY 12-16, 213 ESTIMATION OF STROKE VOLUME BY NON-INVASIVE BLOOD PRESSURE MONITORING IN HUMAN CENTRIFUGE PROGRAM ID NUMBER 445 MANEN O, DUSSAULT C, SAUVET
More informationPulse Pressure Analysis
Pulse Pressure Analysis M. Cecconi, J. Wilson, and A. Rhodes z Introduction Cardiac output monitoring is part of routine practice in the critically ill patient. Recently, there has been increasing interest
More informationReceived: 7 Jun 2006 Revisions requested: 28 Jun 2006 Revisions received: 30 Aug 2006 Accepted: 21 Nov 2006 Published: 21 Nov 2006
Available online http://ccforum.com/content/10/6/r164 Vol 10 No 6 Research Open Access Comparison of uncalibrated arterial waveform analysis in cardiac surgery patients with thermodilution cardiac output
More informationHemodynamic Monitoring in Critically ill Patients in Arthur Simonnet, interne Tuteur : Pr. Raphaël Favory
Hemodynamic Monitoring in Critically ill Patients in 2017 Arthur Simonnet, interne Tuteur : Pr. Raphaël Favory Rationale for Hemodynamic Monitoring Identify the presence of hemodynamic instability Identify
More informationPulsioFlex Patient focused flexibility
PulsioFlex Patient focused flexibility Modular platform with intelligent visualisation for advanced patient Minimally invasive perioperative cardiac output trend with ProAQT Enables calibrated cardiac
More informationInteresting Cases - A Case Report: Renal Cell Carcinoma With Tumor Mass In IVC And Heart. O Wenker, L Chaloupka, R Joswiak, D Thakar, C Wood, G Walsh
ISPUB.COM The Internet Journal of Thoracic and Cardiovascular Surgery Volume 3 Number 2 Interesting Cases - A Case Report: Renal Cell Carcinoma With Tumor Mass In IVC And Heart O Wenker, L Chaloupka, R
More informationCytoreductive Surgery and Hyperthermic Intraperitoneal Chemotherapy
Cytoreductive Surgery and Hyperthermic Intraperitoneal Chemotherapy A 44 year old female undergoing 10 hour Cytoreductive (CRS) procedure followed by Hyperthermic Intraperitoneal Chemotherapy (HIPEC).
More informationNothing to Disclose. Severe Pulmonary Hypertension
Severe Ronald Pearl, MD, PhD Professor and Chair Department of Anesthesiology Stanford University Rpearl@stanford.edu Nothing to Disclose 65 year old female Elective knee surgery NYHA Class 3 Aortic stenosis
More informationHemodynamic Monitoring
Perform Procedure And Interpret Results Hemodynamic Monitoring Tracheal Tube Cuff Pressure Dean R. Hess PhD RRT FAARC Hemodynamic Monitoring Cardiac Rate and Rhythm Arterial Blood Pressure Central Venous
More informationEstimation of changes in cardiac output from the arterial blood pressure waveform in the upper limb ²,²²
British Journal of Anaesthesia 86 (4): 486±96 (2001) Estimation of changes in cardiac output from the arterial blood pressure waveform in the upper limb ²,²² N. W. F. Linton* and R. A. F. Linton The Rayne
More informationFrequently asked questions and answers:
Frequently asked questions and answers: General 1) What are the indications and contraindications for PiCCO-Technology? Indications: Patients in whom cardiovascular and circulatory volume status monitoring
More informationThe Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for
1 2 The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for continuous central venous oximetry (ScvO 2 ) 3 The Vigileo
More informationMinimally invasive cardiac output monitors
Continuing Education in Anaesthesia, Critical Care & Pain Advance Access published October 3, 2011 Minimally invasive cardiac output monitors Kate E Drummond MBBS FANZCA Edward Murphy MBBS FANZCA Matrix
More informationIntroduction. Invasive Hemodynamic Monitoring. Determinants of Cardiovascular Function. Cardiovascular System. Hemodynamic Monitoring
Introduction Invasive Hemodynamic Monitoring Audis Bethea, Pharm.D. Assistant Professor Therapeutics IV January 21, 2004 Hemodynamic monitoring is necessary to assess and manage shock Information obtained
More informationCardiothoracic Fellow Expectations Division of Cardiac Anesthesia, Beth Israel Deaconess Medical Center
The fellowship in Cardiothoracic Anesthesia at the Beth Israel Deaconess Medical Center is intended to provide the foundation for a career as either an academic cardiothoracic anesthesiologist or clinical
More informationAdmission of patient CVICU and hemodynamic monitoring
Admission of patient CVICU and hemodynamic monitoring Prepared by: Rami AL-Khatib King Fahad Medical City Pi Prince Salman Heart tcentre CVICU-RN Admission patient to CVICU Introduction All the patients
More informationMinimally invasive cardiac output monitors
Kate E Drummond MBBS FANZCA Edward Murphy MBBS FANZCA Matrix reference 2A04 Key points Minimally invasive cardiac output monitors have varying degrees of invasiveness with some being totally non-invasive
More informationDetermination of left ventricular stroke volume from the arterial pressure wave
Integrative Human Cardiovascular Control The Panum Institute/ Rigshospitalet, University of Copenhagen May 2018 Determination of left ventricular stroke volume from the arterial pressure wave Jo(Han)nes
More informationA comparison of cardiac output derived from the arterial pressure wave against thermodilution in cardiac surgery patients ²
British Journal of Anaesthesia 87 (2): 212±22 (2001) A comparison of cardiac output derived from the arterial pressure wave against thermodilution in cardiac surgery patients ² J. R. C. Jansen 1 *, J.
More informationCardiac Output by Thermodilution and Arterial Pulse Contour Techniques
Cardiac Output by Thermodilution and Arterial Pulse Contour Techniques J. R. C. Jansen and P. C. M. van den Berg Introduction A reliable method of cardiac output monitoring is particularly desirable in
More informationCOMPARISON OF SUFENTANIL-OXYGEN AND FENTANYL-OXYGEN ANAESTHESIA FOR CORONARY ARTERY BYPASS GRAFTING
Br. J. Anaesth. (1988), 60, 530-535 COMPARISON OF SUFENTANIL-OXYGEN AND FENTANYL-OXYGEN ANAESTHESIA FOR CORONARY ARTERY BYPASS GRAFTING H. M. L. MATHEWS, G. FURNESS, I. W. CARSON, I. A. ORR, S. M. LYONS
More informationThe Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the Edwards PreSep oximetry catheter
1 2 The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the Edwards PreSep oximetry catheter for continuous central venous oximetry (ScvO2) 3
More informationMeasurement of the cardiac output (CO) and
Original Article Annals of Cardiac Anaesthesia 2007; 10: 121 126 Chakravarthy et al. Cardiac Output Measurement by 4 Techniques 121 Comparison of Simultaneous Estimation of Cardiac Output by Four Techniques
More informationTranspulmonary versus continuous thermodilution cardiac output after valvular and coronary artery surgery
doi:10.1510/icvts.2009.204545 Interactive CardioVascular and Thoracic Surgery 9 (2009) 4 8 www.icvts.org Work in progress report - Cardiac general Transpulmonary versus continuous thermodilution cardiac
More informationShock, Monitoring Invasive Vs. Non Invasive
Shock, Monitoring Invasive Vs. Non Invasive Paula Ferrada MD Assistant Professor Trauma, Critical Care and Emergency Surgery Virginia Commonwealth University Shock Fluid Pressors Ionotrope Intervention
More informationPosted: 11/27/2011 on Medscape; Published Br J Anaesth. 2011;107(2): Oxford University Press
Posted: 11/27/2011 on Medscape; Published Br J Anaesth. 2011;107(2):209-217. 2011 Oxford University Press Effect of Phenylephrine and Ephedrine Bolus Treatment on Cerebral Oxygenation in Anaesthetized
More informationThe effects of vasoactive drugs on pulse pressure and stroke volume variation in postoperative ventilated patients,
Journal of Critical Care (211) 26, 328.e1 328.e8 The effects of vasoactive drugs on pulse pressure and stroke volume variation in postoperative ventilated patients, Mehrnaz Hadian MD a, Donald A. Severyn
More informationFloTrac Sensor and Edwards PreSep Central Venous Oximetry Catheter Case Presentations
Edwards FloTrac Sensor & Edwards Vigileo Monitor FloTrac Sensor and Edwards PreSep Central Venous Oximetry Catheter Case Presentations 1 Topics System Configuration FloTrac Sensor and PreSep Catheter Thoracotomy
More informationDefinition- study of blood flow Haemodynamic monitoring refers to monitoring of blood in the cardiovascular system Uses Is NB in the critically ill
By Craig Definition- study of blood flow Haemodynamic monitoring refers to monitoring of blood in the cardiovascular system Uses Is NB in the critically ill pt Can assist diagnosis and decision making
More informationBMC Anesthesiology. Open Access. Abstract
BMC Anesthesiology BioMed Central Research article Evaluation of a new arterial pressure-based cardiac output device requiring no external calibration Christopher Prasser 1, Sylvia Bele 2, Cornelius Keyl
More informationAssessing Preload Responsiveness Using Arterial Pressure Based Technologies. Patricia A. Meehan, RN, MS Education Consultant Edwards Lifesciences, LLC
Assessing Preload Responsiveness Using Arterial Pressure Based Technologies Patricia A. Meehan, RN, MS Education Consultant Edwards Lifesciences, LLC Content Description : Fluid administration is a first
More informationPressure recording analytical method (PRAM) for measurement of cardiac output during various haemodynamic states
British Journal of Anaesthesia 95 (): 59 65 (5) doi:.93/bja/aei54 Advance Access publication May 3, 5 Pressure recording analytical method (PRAM) for measurement of cardiac output during various haemodynamic
More informationCardiac anaesthesia. Simon May
Cardiac anaesthesia Simon May Contents Cardiac: Principles of peri-operative management for cardiac surgery Cardiopulmonary bypass, cardioplegia and off pump cardiac surgery Cardiac disease and its implications
More informationSection 6 Intra Aortic Balloon Pump
Section 6 Intra Aortic Balloon Pump The Intra Aortic Balloon Pump (IABP) The balloon is synthetic and is made for single use only. It is threaded into the aorta, usually via a femoral approach. The balloon
More informationThis PDF is available for free download from a site hosted by Medknow Publications
Indian J Crit Care Med Oct-Dec 006 Vol 10 Issue 4 IJCCM October-December 003 Vol 7 Issue 4 Research Article Correlation of mixed venous and central venous oxygen saturation and its relation to cardiac
More informationChapter 2. Performance of three minimally invasive cardiac output monitoring systems
Chapter 2 Performance of three minimally invasive cardiac output monitoring systems Rob de Wilde, Bart Geerts, Jisheng Cui, Paul van den Berg and Jos Jansen Anaesthesia 2009; 64: 762 769 (45 46) Ideally
More informationCase scenario V AV ECMO. Dr Pranay Oza
Case scenario V AV ECMO Dr Pranay Oza Case Summary 53 y/m, k/c/o MVP with myxomatous mitral valve with severe Mitral regurgitation underwent Mitral valve replacement with mini thoracotomy Pump time nearly
More informationAn algorithmic approach to the very high risk surgical patient
An algorithmic approach to the very high risk surgical patient Daniel A. Reuter Center of Anesthesiology and Intensive Care Medicine Hamburg-Eppendorf Universiy Medical Center Disclosures: Member of the
More informationWhere there s flow, there s life. Measuring flow and pressure together, for even greater control
Where there s flow, there s life Measuring flow and pressure together, for even greater control Introducing the CardioQ-ODM+ The CardioQ-ODM+ is the world s first fluid management and cardiac output monitoring
More informationIMPROVE PATIENT OUTCOMES AND SAFETY IN ADULT CARDIAC SURGERY.
Clinical Evidence Guide IMPROVE PATIENT OUTCOMES AND SAFETY IN ADULT CARDIAC SURGERY. With the INVOS cerebral/somatic oximeter An examination of controlled studies reveals that responding to cerebral desaturation
More informationUniversity of Florida Department of Surgery. CardioThoracic Surgery VA Learning Objectives
University of Florida Department of Surgery CardioThoracic Surgery VA Learning Objectives This service performs coronary revascularization, valve replacement and lung cancer resections. There are 2 faculty
More informationCARDIOVASCULAR MONITORING. Prof. Yasser Mostafa Kadah
CARDIOVASCULAR MONITORING Prof. Yasser Mostafa Kadah Introduction Cardiovascular monitoring covers monitoring of heart and circulatory functions It makes it possible to commence interventions quickly in
More informationHemodynamic monitoring beyond cardiac output
Hemodynamic monitoring beyond cardiac output Prof Xavier MONNET Medical Intensive Care Unit Bicêtre Hospital Assistance publique Hôpitaux de Paris FRANCE Conflicts of interest Lilly GlaxoSmithKline Pulsion
More informationCapture every aspect of hemodynamic status
Capture every aspect of hemodynamic status Edwards Advanced Technology Swan-Ganz Catheters Hemodynamic Monitoring Systems with Continuous CO, SvO 2, EDV, RVEF, SVR, and SV for Optimal Patient Care The
More informationRadboud University Nijmegen Medical Centre Why measure cardiac output in critically ill children?
Radboud University Nijmegen Medical Centre Why measure cardiac output in critically ill children? J. Lemson Anesthesiologist/(pediatric)intensivist Case; Girl 2 years, 12 kg, severe meningococcal septic
More informationHeart transplantation is the gold standard treatment for
Organ Care System for Heart Procurement and Strategies to Reduce Primary Graft Failure After Heart Transplant Masaki Tsukashita, MD, PhD, and Yoshifumi Naka, MD, PhD Primary graft failure is a rare, but
More informationHoiyin Cheung Quan Dong Rong Dong Buwei Yu. Keywords Measurement techniques Cardiac output Thermodilution Non-invasiveness Bioreactance
DOI 10.1007/s00540-014-1938-z ORIGINAL ARTICLE Correlation of cardiac output measured by non invasive continuous cardiac output monitoring (NICOM) and thermodilution in patients undergoing off pump coronary
More informationCogent 2-in-1 Hemodynamic Monitoring System
Cogent 2-in-1 Hemodynamic Monitoring System Minimally invasive and invasive hemodynamic monitoring technologies in a single, lightweight system with wireless communication The flexibility you ve been asking
More informationTailored Volume Resuscitation in the Critically Ill is Achievable. Objectives. Clinical Case 2/16/2018
Tailored Volume Resuscitation in the Critically Ill is Achievable Heath E Latham, MD Associate Professor Fellowship Program Director Pulmonary and Critical Care Objectives Describe the goal of resuscitation
More informationescco, estimated Continuous Cardiac Output device DESEBBE OLIVIER SAUVEGARDE CLINIC, LYON, FRANCE
escco, estimated Continuous Cardiac Output device DESEBBE OLIVIER SAUVEGARDE CLINIC, LYON, FRANCE No conflict of interest u Technology, based on Pulse Wave Transit Time u Limits of the technology u Precision
More informationIntra-operative Echocardiography: When to Go Back on Pump
Intra-operative Echocardiography: When to Go Back on Pump GREGORIO G. ROGELIO, MD., F.P.C.C. OUTLINE A. Indications for Intraoperative Echocardiography B. Role of Intraoperative Echocardiography C. Criteria
More informationThe Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for
1 2 The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for continuous central venous oximetry (ScvO2) 3 The Vigileo
More informationW. J. RUSSELL*, M. F. JAMES
Anaesth Intensive Care 2004; 32: 644-648 The Effects on Arterial Haemoglobin Oxygen Saturation and on Shunt of Increasing Cardiac Output with Dopamine or Dobutamine During One-lung Ventilation W. J. RUSSELL*,
More informationGoal-directed vs Flow-guidedresponsive
Goal-directed vs Flow-guidedresponsive therapy S Magder Department of Critical Care, McGill University Health Centre Flow-directed vs goal directed strategy for management of hemodynamics S Magder Curr
More informationAssessing fluid responses after coronary surgery: role of mathematical coupling of global end-diastolic volume to
954 Original article Assessing fluid responses after coronary surgery: role of mathematical coupling of global end-diastolic volume to cardiac output measured by transpulmonary thermodilution Rose-Marieke
More informationSepsis Wave II Webinar Series. Sepsis Reassessment
Sepsis Wave II Webinar Series Sepsis Reassessment Presenters Nova Panebianco, MD Todd Slesinger, MD Fluid Reassessment in Sepsis Todd L. Slesinger, MD, FACEP, FCCM, FCCP, FAAEM Residency Program Director
More informationTechnique. Technique. Technique. Monitoring 1. Local anesthetic? Aseptic technique Hyper-extend (if radial)
Critical Care Monitoring Hemodynamic Monitoring Arterial Blood Pressure Cannulate artery Uses 2 Technique Sites Locate artery, prep 3 1 Technique Local anesthetic? Aseptic technique Hyper-extend (if radial)
More informationAdult Echocardiography Examination Content Outline
Adult Echocardiography Examination Content Outline (Outline Summary) # Domain Subdomain Percentage 1 2 3 4 5 Anatomy and Physiology Pathology Clinical Care and Safety Measurement Techniques, Maneuvers,
More informationFluid Resuscitation and Monitoring in Sepsis. Deepa Gotur, MD, FCCP Anne Rain T. Brown, PharmD, BCPS
Fluid Resuscitation and Monitoring in Sepsis Deepa Gotur, MD, FCCP Anne Rain T. Brown, PharmD, BCPS Learning Objectives Compare and contrast fluid resuscitation strategies in septic shock Discuss available
More informationFluid bolus of 20% Albumin in post-cardiac surgical patient: a prospective observational study of effect duration
Fluid bolus of 20% Albumin in post-cardiac surgical patient: a prospective observational study of effect duration Investigators: Salvatore Cutuli, Eduardo Osawa, Rinaldo Bellomo Affiliations: 1. Department
More informationNon-invasive continuous blood pressure monitoring based on radial artery tonometry (T-Line TL-200pro device) in the intensive care unit
Non-invasive continuous blood pressure monitoring based on radial artery tonometry (T-Line TL-200pro device) in the intensive care unit Bernd Saugel; Agnes S Meidert; Alexander Hapfelmeier; Florian Eyer;
More informationCath Lab Essentials: Basic Hemodynamics for the Cath Lab and ICU
Cath Lab Essentials: Basic Hemodynamics for the Cath Lab and ICU Ailin Barseghian El-Farra, MD, FACC Assistant Professor, Interventional Cardiology University of California, Irvine Department of Cardiology
More informationADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE
Vet Times The website for the veterinary profession https://www.vettimes.co.uk ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE Author : CARL BRADBROOK Categories : Vets Date : October 7, 2013
More informationCitation for published version (APA): Truijen, J. (2018). Withstanding the flow: Human cardiovascular control during postural challenges
UvA-DARE (Digital Academic Repository) Withstanding the flow Truijen, J. Link to publication Citation for published version (APA): Truijen, J. (2018). Withstanding the flow: Human cardiovascular control
More informationDiastolic Augmentation with an External Pulsating Device To Treat Cardiogenic Shock
Diastolic Augmentation with an External Pulsating Device To Treat Cardiogenic Shock Nickolas Trubov, B.Sc., C.C.P. and Steven J. Phillips, M.D. Mercy Hospital, Des Moines, Iowa 50314 ABSTRACT An 80 cc
More informationCardiac output by Portapres
Clinical Science (2004) 106, 407 412 (Printed in Great Britain) 407 Cardiac output by Portapres Marjorie S. PITT, Paul MARSHALL, Jonathan P. DIESCH and Roger HAINSWORTH Institute for Cardiovascular Research,
More informationPerioperative Management of DORV Case
Perioperative Management of DORV Case James P. Spaeth, MD Department of Anesthesia Cincinnati Children s Hospital Medical Center University of Cincinnati Objectives: 1. Discuss considerations regarding
More informationSaphenous Vein Autograft Replacement
Saphenous Vein Autograft Replacement of Severe Segmental Coronary Artery Occlusion Operative Technique Rene G. Favaloro, M.D. D irect operation on the coronary artery has been performed in 180 patients
More informationMechanics of Cath Lab Support Devices
Mechanics of Cath Lab Support Devices Issam D. Moussa, MD Chief Medical Officer First Coast Cardiovascular Institute, Jacksonville, FL Professor of Medicine, UCF, Orlando, FL None DISCLOSURE Percutaneous
More informationMechanical Ventilation & Cardiopulmonary Interactions: Clinical Application in Non- Conventional Circulations. Eric M. Graham, MD
Mechanical Ventilation & Cardiopulmonary Interactions: Clinical Application in Non- Conventional Circulations Eric M. Graham, MD Background Heart & lungs work to meet oxygen demands Imbalance between supply
More informationComparison between oxygen consumption calculated by Fick s principle using a continuous thermodilution technique and measured by indirect calorimetry
British Journal of Anaesthesia 1995; 75: 719 73 Comparison between oxygen consumption calculated by Fick s principle using a continuous thermodilution technique and measured by indirect calorimetry P.
More informationGeorgios C. Bompotis Cardiologist, Director of Cardiological Department, Papageorgiou Hospital,
Georgios C. Bompotis Cardiologist, Director of Cardiological Department, Papageorgiou Hospital, Disclosure Statement of Financial Interest I, Georgios Bompotis DO NOT have a financial interest/arrangement
More informationA New Technique for Repeated Measurement of Cardiac Output During Cardiopulmonary Resuscitation
Purdue University Purdue e-pubs Weldon School of Biomedical Engineering Faculty Publications Weldon School of Biomedical Engineering 1980 A New Technique for Repeated Measurement of Cardiac Output During
More informationPulse Contour Analysis and Transpulmonary Thermodilution
Pulse Contour Analysis and Transpulmonary Thermodilution MARK LENNON, FCARCSI, FANZCA Dr Lennon is a Senior Registrar in Anaesthesia at Sir Charles Gairdner Hospital in Western Australia. He is also completing
More informationEVOLUCIÓN DE LA MONITORIZACIÓN CARDIOVASCULAR EN LA UCI
EVOLUCIÓN DE LA MONITORIZACIÓN CARDIOVASCULAR EN LA UCI Antonio Artigas Critical Care Center Sabadell Hospital CIBER Enfermedades Respiratorias Autonomos University of Barcelona Spain aartigas@tauli.cat
More informationhaemodynamic monitoring
haemodynamic monitoring Mostcare Up is the only monitor able to follow, in real time and from beat to beat, even the slightest haemodynamic variations in the patient. The patented algorithm based on the
More informationHow and why I give IV fluid Disclosures SCA Fluids and public health 4/1/15. Andrew Shaw MB FRCA FCCM FFICM
How and why I give IV fluid Andrew Shaw MB FRCA FCCM FFICM Professor and Chief Cardiothoracic Anesthesiology Vanderbilt University Medical Center 2015 Disclosures Consultant for Grifols manufacturer of
More informationCardiac output (CO) and invasive hemodynamic. Not for External Distribution
Technology, Computing, and Simulation Section Editor: Jeffrey M. Feldman Not for External Distribution Cardiac Output Measurement in Patients Undergoing Liver Transplantation: Pulmonary Artery Catheter
More informationWhat is. InSpectra StO 2?
What is InSpectra StO 2? www.htibiomeasurement.com What is InSpectra StO 2? Hemoglobin O 2 saturation is measured in three areas: 1) Arterial (SaO 2, SpO 2 ) Assesses how well oxygen is loading onto hemoglobin
More informationCardiac Output Technique For Small Animals
Cardiac Output Technique For Small Introduction Cardiac output (CO) is a measure of the quantity of blood pumped by the heart each minute and is the product of stroke volume (ie. volume of blood ejected
More informationClinical evaluation of USCOM ultrasonic cardiac output monitor in cardiac surgical patients in intensive care unit
British Journal of Anaesthesia 94 (3): 287 91 (2005) doi:10.1093/bja/aei054 Advance Access publication January 14, 2005 Clinical evaluation of USCOM ultrasonic cardiac output monitor in cardiac surgical
More informationDetermination of Stroke Volume from Left Ventricular Isovolumetric Contraction and Ejection Times
Determination of Stroke Volume from Left Ventricular Isovolumetric Contraction and Ejection Times Clarence M. AGRESS, M.D. and Stanley WEGNER SUMMARY Examination was made of the relationship of left ventricular
More informationManagement of Cardiogenic Shock. Dr Stephen Pettit, Consultant Cardiologist
Dr Stephen Pettit, Consultant Cardiologist Cardiogenic shock Management of Cardiogenic Shock Outline Definition, INTERMACS classification Medical management of cardiogenic shock PA catheters and haemodynamic
More informationPROBLEM SET 2. Assigned: February 10, 2004 Due: February 19, 2004
Harvard-MIT Division of Health Sciences and Technology HST.542J: Quantitative Physiology: Organ Transport Systems Instructors: Roger Mark and Jose Venegas MASSACHUSETTS INSTITUTE OF TECHNOLOGY Departments
More informationDESIGNER RESUSCITATION: TITRATING TO TISSUE NEEDS
DESIGNER RESUSCITATION: TITRATING TO TISSUE NEEDS R. Phillip Dellinger MD, MSc, MCCM Professor and Chair of Medicine Cooper Medical School of Rowan University Chief of Medicine Cooper University Hospital
More informationNATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE
NATIONAL INSTITUTE FOR HEALTH AND CLINICAL EXCELLENCE Medical technologies guidance SCOPE CardioQ-ODM oesophageal Doppler monitor for patients undergoing major or high-risk surgery and patients in critical
More information