CARDIOVASCULAR Continuous cardiac output during off-pump coronary artery bypass surgery: pulse-contour analyses vs pulmonary artery thermodilution

Size: px
Start display at page:

Download "CARDIOVASCULAR Continuous cardiac output during off-pump coronary artery bypass surgery: pulse-contour analyses vs pulmonary artery thermodilution"

Transcription

1 British Journal of Anaesthesia 99 (4): (2007) doi: /bja/aem199 Advance Access publication July 24, 2007 CARDIOVASCULAR Continuous cardiac output during off-pump coronary artery bypass surgery: pulse-contour analyses vs pulmonary artery thermodilution P. S. Halvorsen 1 *, A. Sokolov 13, M. Cvancarova 2,P.K.Hol 1, R. Lundblad 4 and T. I. Tønnessen The Interventional Centre, 2 Department of Biostatistics, 3 Department of Anaesthesiology and 4 Department of Cardio-Thoracic Surgery, Rikshospitalet-Radiumhospitalet Medical Centre, Oslo, Norway. 5 Medical Faculty, University of Oslo, Oslo, Norway *Corresponding author: The Interventional Centre, Rikshospitalet-Radiumhospitalet Medical Centre, NO-0027 Oslo, Norway. per.steinar.halvorsen@rikshospitalet.no Background. No gold standard method exists for monitoring continuous cardiac output (CO). In this study, the agreement between the two most frequently used methods, PiCCO pulse-contour analysis (PCCO) and STAT pulmonary artery thermodilution (STAT-CO), was assessed during multiple-vessel off-pump coronary artery bypass (OPCAB) surgery. Methods. Thirty patients were enrolled in the study. Two time periods were defined during surgery; Period 1 included positioning of the heart and stabilizer device and Period 2 included the coronary occlusion. Measurements were obtained every minute during both periods. The agreement for the continuous CO and the change in CO (DCO) was estimated using the Bland Altman method. Results. Significant changes in mean arterial pressure (DMAP), central venous saturation, PCCO and STAT-CO were seen only during Period 1. DMAP correlated only with changes in PCCO, (P,0.001, r¼0.60). The mean difference (2SD) between PCCO and STAT-CO ranged from (1.82) to (2.57) litre min 21, and the percentage error varied from 32 to 50%. For the CO measurements, the limits of agreements did not differ between Period 1 and Period 2. In contrast, for the DCO measurements, the limits of agreements were wider in Period 1 than in the more haemodynamically stable Period 2. Conclusions. PCCO and STAT-CO show large discrepancies in CO during OPCAB surgery. Clinically acceptable agreement was seen only for trends in CO during haemodynamically stable periods. Br J Anaesth 2007; 99: Keywords: anaesthesia, cardiovascular; heart, cardiac output; measurement techniques, pulsecontour analysis; measurement techniques, thermodilution; surgery, coronary artery bypass Accepted for publication: June 3, 2007 During off-pump coronary artery by-pass (OPCAB) surgery, haemodynamic instability is common 1 3 and continuous monitoring of cardiac output (CO) is advisable. 4 There is no gold standard method for continuous CO monitoring. The most commonly used methods are the STAT pulmonary artery catheter (STAT-CO), which gives semi-continuous thermodilution measurements, and PiCCO pulse-contour analysis (PCCO) enabling almost real-time CO. Both methods have drawbacks. Because of a time delay of the CO calculations, the STAT method may miss rapid and transient haemodynamic changes. 5 7 With PCCO, flow is estimated from the femoral artery blood pressure curve. The algorithm uses properties of the vascular system which are non-linear and differs from patient to patient. 8 Thus, the reliability of PCCO has been questioned during haemodynamic instability. 910 Continuous CO is used to monitor cardiac performance and to guide medical interventions. Therefore, these # The Board of Management and Trustees of the British Journal of Anaesthesia All rights reserved. For Permissions, please journals.permissions@oxfordjournals.org

2 Cardiac output monitoring during OPCAB surgery methods should be reliable and in agreement with each other. In this paper, we address the agreement between PCCO and STAT-CO during multiple-vessel OPCAB surgery. We hypothesize that the agreement depends on the patients haemodynamic situation, that is, during haemodynamic instability, the agreement may be poorer than during more stable periods. For each coronary artery, the agreement was estimated in two periods; Period 1 included the positioning of the heart and the stabilizer, and Period 2 included the coronary occlusion and suturing of the anastomosis. Methods Patients The Regional Ethics Committee approved the study and informed consent was obtained from all patients. Exclusion criteria were unstable angina pectoris, left ventricle ejection fraction,0.30, arrhythmias, valvular heart disease, intracardiac shunts, and severe peripheral arterial occluding diseases. The patients were premedicated with diazepam 5 15 mg orally. An arterial catheter was placed in the radial artery and anaesthesia was induced with fentanyl 2 5 mg kg 21, diazepam 0.1 mg kg 21, thiopentone 2 5 mg kg 21, and cisatracurium 0.15 mg kg 21. Anaesthesia was maintained with sevoflurane. Ephedrine or phenylephrine was administered when necessary in order to keep the mean arterial pressure (MAP) above 60 mm Hg, and atropine was given to maintain heart rate (HR) above 45 beats min 21.All patients received an infusion of glycerylnitrate ( mg kg 21 except when MAP was below 60 mm Hg. The haemoglobin (Hb) concentration was kept above 9.0 g dl 21. A 7.5 French pulmonary artery catheter (REF: 774HF75, Edwards Lifesciences, USA) was inserted via a 9.0 French triple lumen introducer (REF: m3l9fhsi, Edwards Lifesciences, USA) in the right internal jugular vein and advanced to the wedge position guided by waveform and pressure analysis. The pulmonary artery catheter was connected to a Vigilance monitor (version 6.2 Edwards Lifesciences, USA) with an inline temperature sensor (model ) for intermittent pulmonary artery thermodilution CO. The mixed venous oxygenation (Sv O2 ) was calibrated in vitro before the insertion of the pulmonary artery catheter. A 5 French arterial thermodilution catheter (PV 2025L20, PULSIOCATH, Pulsion Medical Systems, Munich, Germany) was inserted in the right femoral artery and connected to a pressure transducer (PV8115, PULSION Medical Systems, Munich, Germany), and the signals were computed in a PiCCO monitor (Pulsion Plus, software version 5.1, PULSION Medical Systems, Munich, Germany). A PiCCO inline injectate temperature sensor was fitted in the central venous line for transpulmonary thermodilution CO measurements. Three boluses of 10 ml of ice-cold glucose 5% were injected into the proximal port of the pulmonary artery catheter, and thermodilution CO for both the PiCCO and the pulmonary artery catheter were then obtained simultaneously. The injections were carried out by the same person to minimize variation. The mean value of CO obtained by transthoracic thermodilution was used to calibrate the pulse-contour analysis. Measurements For surgery on the anterior, lateral, and posterior coronary arteries, three time points were defined: T1 (baseline), T2 (immediately before coronary occlusion), and T3 (immediately before coronary reperfusion). Measurements of HR, MAP, Sv O2 and PCCO were collected at all three time points. Because of the interruption of the thermodilution measurements, STAT-CO values could be obtained only at T2 and T3. Instead, at T1, the mean CO of the three pulmonary artery thermodilutions was used. All measurements were obtained with the table in the zero position and the heart untouched in the pericardial sack. For each coronary artery, T1 T2 defined Period 1 and included positioning of the heart and the placement of the stabilizer device. Period 2 was defined from T2 to T3 and included the coronary occlusion and suturing of the anastomosis. PCCO was recalibrated before surgery on each coronary artery. The recalibrated pulse-contour value was used for baseline measurements (T1). Each minute, during the two time periods, both PCCO and STAT-CO were measured simultaneously. Since the STAT-CO value updates only every minute, a measurement of 1 min was used. The software program ICUPilot (CMA, Sweden) was used for data collection. Statistical analysis The number of patients was based on a calculation of power. A clinically relevant mean difference was set to 0.20 litre min 21, with an SD of 0.20 litre min 21. Thirty patients were included to ensure power of at least 80%. For the T1, T2, and T3 measurements, the variables were not normally distributed, so the values are presented as median and inter-quartile range (25th and 75th percentiles). Differences between the time points were analysed with the Friedman test, and a P-value,0.05 was considered statistically significant. The Wilcoxon rank sum tests with the Bonferroni correction were used for variables for which the Friedman test showed significant differences. For the correlation analysis, the Pearson correlation was used and a P-value,0.01 was considered statistically significant. For the CCO and DCO measurements, the agreement was assessed using the Bland Altman method. 11 The Bland Altman method could be applied since these variables were normally distributed. Percentage error was calculated as follows: twice the SD for the difference between the two methods being compared was divided by the mean value 485

3 Halvorsen et al. obtained from both methods {2 SD (difference between method 1 and method 2)/[(mean of method 1þmean of method 2)/2]100g. The statistical analysis was computed with SPSS (Version 13, SPSS Inc., Chicago, IL, USA). Results Thirty patients were enrolled (26 men and 4 women). Median (range) age was 63 yr (39 86), body surface area 2.02 ( ) m 2, New York Heart Association functional class of angina pectoris 2.0 (1 4), and left ventricle ejection fraction 0.72 ( ). Sixteen patients (53%) had a previous myocardial infarction and 27 (90%) of the patients received preoperative beta-blockers. The number of distal coronary anastomoses was 3.0 (1 4). End-tidal volume concentration of sevoflurane was 1.70 ( )%, fentanyl 1.70 ( ) mg kg 21 h 21, ephedrine 10.0 ( ) mg, phenylephrine 0.05 ( ) mg, and atropine 0.50 ( ) mg. One patient was converted from OPCAB to on-pump surgery before grafting because of severe myocardial hypertrophy and fragile coronary arteries. Surgery on the anterior wall was performed in 29 patients, on the posterior wall in 15 patients, and on the lateral wall in 24 patients. Ventricular fibrillation occurred during reperfusion of the circumflex artery in one patient. He was successfully defibrillated and recovered without sequelae. Two patients had atrio-ventricular block during occlusion of the right coronary artery. Both patients spontaneously recovered to sinus rhythm after reperfusion. There were no deaths or myocardial or cerebral infarctions. Erythrocyte concentrates were infused in seven patients (24%), with a median (range) volume of 400 ml ( ). Twenty-two patients (76%) received autotransfusion; the median (range) volume given was 500 ml (0 2000). A small but significant (P¼0.005) change in the Hb was seen from the start to the end of surgery, mean (SD) 11.5 (1.3) and 10.4 (1.3) g dl 21, respectively. Blood gases were measured on average three times during surgery. The lowest mean Hb observed was 10.0 (1.1) g dl 21 (P,0.001 towards the start, and P¼0.519 towards the end of surgery). All patients were externally warmed, and the mean (SD) rectal temperature was 36.2 (0.5)8C at the start and 36.1 (0.4)8C at the end of the surgery (P¼0.497). Figure 1 shows an example of haemodynamic data during surgery on the right coronary artery. There is a large discrepancy between PCCO and STAT-CO. The curves for Sv O2, PCCO, and MAP decrease until the end of the period, then Sv O2 increases, whereas PCCO and MAP continue to decrease. The MAP and PCCO curves look remarkably similar. Significant changes in MAP, Sv O2, PCCO, and STAT-CO were observed from baseline (T1) to coronary occlusion (T2), but not from the start of occlusion (T2) to reperfusion (T3) (Table 1). Thus, in each subgroup of patients, Period 1 (T1 T2) is the most haemodynamically unstable period. For surgery on the anterior wall, median times (range) for Period 1 and Period 2 were 14 (6 25) and 12 (7 26) min, respectively. The corresponding times for lateral wall surgery were 12 (7 23) and 14 (8 20) min, and for posterior wall surgery 11 (4 20) and 14 (6 22) min, respectively. The mean difference and limits of agreements between PCCO and STAT-CO for the 1 min measurements during Period 1 and Period 2 are shown in Table 2 and Figure 2. For surgery on all heart regions, a significant discrepancy in the mean difference between the methods was observed for both periods. As shown in Table 2, wide limits of agreement were observed for both periods during surgery on the different coronary arteries. The limits of agreements in Period 1 and Period 2 were and litre min 21, respectively. Trends in CO, that is, changes in CO from the previous value (DCO), were also calculated using both methods, and their agreements are shown in Table 3 and Figure 3. For all regions, no significant discrepancy in the mean difference for DCO was observed. However, the limits of agreement were markedly reduced in Period 2 compared with Period 1. Thus, when assessing trends in CO, the best agreement was seen during stable haemodynamics. In Table 4, the correlations between changes in MAP (DMAP), changes in Sv O2 (DSv O2 ), changes in PCCO (DPCCO), and changes in STAT-CO (DSTAT-CO) for the 1 min measurements in Period 1 and Period 2 are presented. As seen, DMAP was correlated only with DPCCO (P, 0.001, mean r¼0.60) and not with DSv O2 and DSTAT-CO. DSv O2 was correlated to neither DPCCO nor DSTAT-CO. Discussion In this study, the agreement between the two most frequently used methods for monitoring CCO during OPCAB surgery was assessed. We found that the agreement between PCCO and STAT-CO was poor, with wide limits of agreements seen during surgery on every coronary artery. There were no differences in the limits of agreements between the two time periods. When measuring trends in CO, the methods agreed more closely, particularly during stable haemodynamics. In the present study, the largest haemodynamic changes appeared during the positioning of the heart and the stabilizer. This is in accordance with other studies. 13 No significant changes were observed during the coronary occlusion period. Thus, in our study, Period 1 is the most haemodynamically unstable. Studies focusing on the agreement between different methods for assessing CO (both continuous and intermittent) usually compare measurements at clearly defined time points Typically, the new methods are 486

4 Cardiac output monitoring during OPCAB surgery Fig 1 Haemodynamic changes during surgery on the right coronary artery. Event 1, positioning of the heart; Event 2, application of the stabilizer device. compared with intermittent thermodilution pulmonary artery CO, which is considered the most accurate. Since the thermodilution measurements take a few minutes, comparisons are generally performed in haemodynamically stable phases. This improves the precision and the agreement between the methods. In our view, it is more important to know how the different methods agree during haemodynamic instability. Ideally, clinically acceptable mean difference and limits of agreement should be defined before starting the study. Instead, these are commonly based on a subjective judgement. 14 In order to define objective criteria, Critchley and Critchley 15 recommended accepting a new method where limits of agreements are up to +30%. In OPCAB surgery, trends in CO, that is, information on the direction and relative magnitude of changes in CO, may give more valuable clinical information than accurate intermittent measurements. Thus, one might accept reduced accuracy when using a continuous method. However, there is no consensus on the level of agreement, and therefore clinical judgement has to be used. In our study, there was a poor agreement between PCCO and STAT-CO. For all heart regions, the methods differed significantly in Period 1 and Period 2, and the mean difference between STAT-CO and PCCO varied from 0.71 to 0.78 litre min 21. For both time periods and for every heart region, wide limits of agreements were observed (range to litre. When 487

5 Halvorsen et al. Table 1 Haemodynamic variables at baseline (T1), start occlusion (T2) and end of coronary occlusion (T3) Surgery T1 T2 T3 P-value Median 25th 75th percentile Median 25th 75th percentile Median 25th 75th percentile Intragroup T1 T2 T2 T3 T1 T3 Anterior (n¼29) Lateral (n¼24) Posterior (n¼15) HR (beats MAP (mm Hg) , Sv O2 % PCCO (litre , STAT-CO (litre HR (beats MAP (mm Hg) Sv O2 % ,0.001, ,0.001 PCCO (litre , STAT-CO (litre HR (beats MAP (mm Hg) Sv O2 % PCCO (litre STAT-CO (litre applying the method described by Critchley and Critchley, our percentage error was in the range of 32 50%. The two methods agree better when trends in CO were measured. For the DCO values, no significant difference between the methods was observed. The best agreement was seen during stable haemodynamics (Period 2), with limits of agreements litre min 21. For Period 1, the limits of agreements were litre min 21. This is in contrast to the CO measurements, where no difference between periods was observed. Given these results, clinically acceptable agreement between PCCO and STAT-CO can be seen only for trends in CO during haemodynamically stable phases. Our study allows no conclusion on which method is the most accurate. One of the main reasons for the lack of agreement is that STAT-CO is a semi-continuous method with a slow response time, whereas PCCO reflects CO in almost real time. Thus, rapid haemodynamic changes may be missed with the STAT method. This may explain why the mean difference changed from Period 1 to Period 2 for the absolute CO measurements. In Period 1, the mean differences were negative, showing that the STAT pulmonary artery was measuring higher values compared with PCCO. Since STAT-CO responds slowly, it continues to decrease into Period 2. In contrast, PCCO fell early in Period 1, but was stabilized during Period 2 (Table 1). Table 2 Agreement in CO (litre between PCCO and STAT-CO Surgery Period n Mean difference 95% Limits of agreements Lower 95% Upper 95% %Error Anterior (n¼29) Lateral (n¼24) Posterior (n¼15) to to to to to to to to to to to to to to to to to to

6 Cardiac output monitoring during OPCAB surgery Fig 2 Bland Altman plots, showing the agreements between PCCO and STAT-CO for the CO values. Solid lines depict the mean difference and limit of agreements, and dotted lines depict the corresponding 95% confidence s. (A) Anterior wall surgery, Period 1; (B) anterior wall surgery, Period 2. Table 3 Agreement in DCO (litre between PCCO and STAT-CO Surgery Period n Mean difference 95% Limits of agreements Lower 95% Upper 95% Anterior (n¼29) Lateral (n¼24) Posterior (n¼15) to to to to to to to to to to to to

7 Halvorsen et al. Fig 3 Bland Altman plots, showing the agreements between PCCO and STAT-CO for the DCO values. Solid lines depict the mean difference and limit of agreements, and dotted lines depict the corresponding 95% confidence s. (A) Anterior wall surgery, Period 1; (B) anterior wall surgery, Period 2. PCCO estimates flow from the femoral artery blood pressure curve and the algorithm is based on properties of the vascular system which are non-linear and differ from patient to patient. 8 Vascular compliance is altered by myogenic intrinsic local regulation, sympathetic vasoconstriction, manipulation of the aorta, and changes in inotropic, vasoactive, or anaesthetic drugs during the time course of surgery. In a previous study, during stable haemodynamics, we found a percentage error of % between PCCO and intermittent pulmonary artery and transthoracic thermodilution. 16 In a study including postoperative cardiac surgery patients, Østergaard and colleagues 17 reported a percentage error of 49.5% between PCCO and intermittent pulmonary artery thermodilution. Reduced accuracy of PCCO during haemodynamic instability has been reported by others Thus, in our study, since significant haemodynamic changes were seen and inotropes and vasoactive drugs were given in order to maintain acceptable MAP and HR, one should expect a high percentage error between PCCO and true CO. In the present study, minor bleeding occurred and the patients were adequately oxygenated. Under such circumstances, Sv O2 has been shown to correlate with CO, but small changes in MAP and CO may not alter Sv O2 and 490

8 Cardiac output monitoring during OPCAB surgery Table 4 Correlation between D-values for the haemodynamic variables at the different time periods during surgery Anterior surgery Lateral surgery Posterior surgery Period 1 Period 2 Period 1 Period 2 Period 1 Period 2 P-value r P-value r P-value r P-value r P-value r P-value r DMAP (mm Hg) vs DPCCO (litre DMAP (mm Hg) vs DSTAT-CO (litre DMAP (mm Hg) vs DSv O2 % DPCCO (litre vsdstat-co (litre DPCCO (litre vs DSv O2 % DSTAT-CO (litre vs DSv O2 %, , , , , , , , the correlation between CO and Sv O2 is better during haemodynamic instability. Thus, one should expect DPCCO and DSTAT-CO to be correlated with DSv O2 in Period 1 when the circulatory changes were the most prominent. However, this was not found, and the DPCCO and DMAP were the only haemodynamic variables that correlated. This may indicate a too strong mathematical coupling between MAP and PCCO, implying that PCCO does not reflect true CO precisely. Another reason for the poor correlation found between PCCO, DSv O2 and DSTAT-CO might be our short sampling. Although Sv O2 can detect rapid and large haemodynamic changes, the time response is delayed compared with the blood pressure. 7 In the present study, a major portion of the D values for Sv O2 and STAT-CO were small or even zero, and, since no linearity can be found on data clustered around zero, this may account for the poor correlations observed. In our opinion, the reliability of PCCO during haemodynamic instability is uncertain, and if PCCO needs recalibration every time there are changes in vasoactive drugs or after manipulation of the heart and great vessels, we argue that the use of PCCO is limited during OPCAB surgery. For low-risk patients, we consider ECG, invasive arterial blood pressure, peripheral oxygen saturation, end-tidal carbon dioxide tension, and sufficient monitoring of central venous pressure. From the central venous line, intermittent central venous oxygenation can be measured, and a close correlation with Sv O2 has been shown. 21 In high-risk patients, we add a pulmonary artery catheter with continuous Sv O2 monitoring and transoesophageal echocardiography (TOE). Sv O2 provides valuable information on tissue oxygenation and, although slower than PCCO, rapid changes in CO can be detected. Sv O2 adds no direct information about the different determinates of CO, such as preload, contractility, and afterload, and for that purpose, intermittent bolus thermodilution and TOE can be applied. In OPCAB surgery, the TOE also permits the monitoring of valvular function and regional myocardial ischaemia. 22 A limitation in our study is the lack of a gold standard method for continuous CO measurements. Since no such method exists, we cannot conclude which method reflects CO most accurately. In our study, we compared the CO values each minute during surgery. This is long enough to obtain a new STAT-CO value, but is too short to detect large changes because of the time delay of the calculations. Since CO might change considerably during 1 min in OPCAB surgery, we think that a measurement of 1 min is clinically relevant. The duration of Period 1 and Period 2 was different for each patient. Thus, different numbers of measurements were obtained, implying unequal weighting for each patient. Our results show that there is less agreement between DPCCO and DSTAT-CO during haemodynamic instability. Thus, if haemodynamic stability is associated with the time of surgery, this might have affected our results on the agreement between the methods. However, the measurements at different time points were compared for the same patients (each patients being their own control), and this may compensate for weighting. In conclusion, we found a poor agreement between PCCO and STAT-CO during OPCAB surgery. Clinically acceptable agreement was seen only for trends in CO during haemodynamically stable periods. We could not conclude which method was the most accurate. In our opinion, intermittent bolus thermodilution is still the reference method for measuring CO, whereas PCCO and STAT-CO may be used only for monitoring trends. Particularly during haemodynamic instability, these trends have to be interpreted with caution. References 1 Shinn HK, Oh YJ, Kim SH, Lee JH, Lee CS, Kwak YL. Evaluation of serial haemodynamic changes during coronary artery anastomoses in patients undergoing off-pump coronary artery bypass 491

9 Halvorsen et al. graft surgery: initial experiences using two deep pericardial stay sutures and octopus tissue stabilizer. Eur J Cardiothorac Surg 2004; 25: Do QB, Goyer C, Chavanon O, Couture P, Denault A, Cartier R. Hemodynamic changes during off-pump CABG surgery. Eur J Cardiothorac Surg 2002; 21: Mishra M, Malhotra R, Mishra A, Meharwal ZS, Trehan N. Hemodynamic changes during displacement of the beating heart using epicardial stabilization for off-pump coronary artery bypass graft surgery. J Cardiothorac Vasc Anesth 2002; 16: Buhre W, Weyland A, Kazmaier S, et al. Comparison of cardiac output assessed by pulse-contour analysis thermodilution in patients undergoing minimally invasive direct coronary artery bypass grafting. J Cardiothorac Vasc Anesth 1999; 13: Aranda M, Mihm FG, Garrett S, Mihm MN, Pearl RG. Continuous cardiac output catheters: delay in in vitro response time after controlled flow changes. Anesthesiology 1998; 89: Siegel LC, Hennessy MM, Pearl RG. Delayed time response of the continuous cardiac output pulmonary artery catheter. Anesth Analg 1996; 83: Lazor MA, Pierce ET, Stanley GD, Cass JL, Halpern EF, Bode RH. Evaluation of the accuracy and response time of STAT-mode continuous cardiac output. J Cardiothorac Vasc Anesth 1997; 11: Wesseling KH, Jansen JR, Settels JJ, Schreuder JJ. Computation of aortic flow from pressure in humans using a nonlinear, three-element model. J Appl Physiol 1993; 74: Tannenbaum GA, Mathews D, Weissman C. Pulse contour cardiac-output in surgical intensive-care unit patients. J Clin Anesth 1993; 5: Rodig G, Prasser C, Keyl C, Liebold A, Hobbhahn J. Continuous cardiac output measurement: pulse contour analysis vs thermodilution technique in cardiac surgical patients. Br J Anaesth 1999; 82: Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 1: Della RG, Costa MG, Pompei L, Coccia C, Pietropaoli P. Continuous and intermittent cardiac output measurement: pulmonary artery catheter versus aortic transpulmonary technique. Br J Anaesth 2002; 88: Goedje O, Hoeke K, Lichtwarck-Aschoff M, Faltchauser A, Lamm P, Reichart B. Continuous cardiac output by femoral arterial thermodilution calibrated pulse contour analysis: comparison with pulmonary arterial thermodilution. Crit Care Med 1999; 27: Mantha S, Roizen MF, Fleisher LA, Thisted R, Foss J. Comparing methods of clinical measurement: reporting standards for Bland and Altman analysis. Anesth Analg 2000; 90: Critchley LA, Critchley JA. A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput 1999; 15: Halvorsen PS, Espinoza A, Lundblad R, et al. Agreement between PiCCO pulse-contour analysis, pulmonal artery thermodilution and transthoracic thermodilution during off-pump coronary artery by-pass surgery. Acta Anaesthesiol Scand 2006; 50: Østergaard M, Nielsen J, Rasmussen JP, Berthelsen PG. Cardiac output-pulse contour analysis vs. pulmonary artery thermodilution. Acta Anaesthesiol Scand 2006; 50: Tannenbaum GA, Mathews D, Weissman C. Pulse contour cardiac-output in surgical intensive-care unit patients. J Clin Anesth 1993; 5: Inomata S, Nishikawa T, Taguchi M. Continuous monitoring of mixed venous oxygen-saturation for detecting alterations in cardiac-output after discontinuation of cardiopulmonary bypass. Br J Anaesth 1994; 72: Reinhart K, Kuhn HJ, Hartog C, Bredle DL. Continuous central venous and pulmonary artery oxygen saturation monitoring in the critically ill. Intensive Care Med 2004; 30: Reinhart K, Rudolph T, Bredle DL, Hannemann L, Cain SM. Comparison of central-venous to mixed-venous oxygensaturation during changes in oxygen-supply demand. Chest 1989; 95: Skulstad H, Andersen K, Edvardsen T, et al. Detection of ischemia and new insight into left ventricular physiology by strain Doppler and tissue velocity imaging: assessment during coronary bypass operation of the beating heart. J Am Soc Echocardiogr 2004; 17:

Less Invasive, Continuous Hemodynamic Monitoring During Minimally Invasive Coronary Surgery

Less Invasive, Continuous Hemodynamic Monitoring During Minimally Invasive Coronary Surgery Less Invasive, Continuous Hemodynamic Monitoring During Minimally Invasive Coronary Surgery Oliver Gödje, MD, Christian Thiel, MS, Peter Lamm, MD, Hermann Reichenspurner, MD, PhD, Christof Schmitz, MD,

More information

Button et al. using a thermistor-tipped catheter. The CO is calculated using an algorithm based on the area under the systolic part of the pressure wa

Button et al. using a thermistor-tipped catheter. The CO is calculated using an algorithm based on the area under the systolic part of the pressure wa British Journal of Anaesthesia 99 (3): 329 36 (2007) doi:10.1093/bja/aem188 Advance Access publication on July 12, 2007 CARDIOVASCULAR Clinical evaluation of the FloTrac/Vigileo TM system and two established

More information

Edwards FloTrac Sensor & Performance Assessments of the FloTrac Sensor and Vigileo Monitor

Edwards FloTrac Sensor & Performance Assessments of the FloTrac Sensor and Vigileo Monitor Edwards FloTrac Sensor & Edwards Vigileo Monitor Performance Assessments of the FloTrac Sensor and Vigileo Monitor 1 Topics System Configuration Performance and Validation Dr. William T. McGee, Validation

More information

Continuous cardiac output measurement: pulse contour analysis vs thermodilution technique in cardiac surgical patients

Continuous cardiac output measurement: pulse contour analysis vs thermodilution technique in cardiac surgical patients British Journal of Anaesthesia 82 (4): 525 30 (1999) Continuous cardiac output measurement: pulse contour analysis vs thermodilution technique in cardiac surgical patients G. Rödig*, C. Prasser, C. Keyl,

More information

The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for

The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for 1 2 The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for continuous central venous oximetry (ScvO2) 3 The Vigileo

More information

THE MEASUREMENT of cardiac output (CO) is a parameter

THE MEASUREMENT of cardiac output (CO) is a parameter Comparison of Esophageal Doppler, Pulse Contour Analysis, and Real-Time Pulmonary Artery Thermodilution for the Continuous Measurement of Cardiac Output Berthold Bein, MD,* Frank Worthmann, MD,* Peter

More information

Measurement of the cardiac output (CO) and

Measurement of the cardiac output (CO) and Original Article Annals of Cardiac Anaesthesia 2007; 10: 121 126 Chakravarthy et al. Cardiac Output Measurement by 4 Techniques 121 Comparison of Simultaneous Estimation of Cardiac Output by Four Techniques

More information

Perioperative monitoring of cardiac

Perioperative monitoring of cardiac Continuous cardiac output monitoring with pulse contour analysis: A comparison with lithium indicator dilution cardiac output measurement James Pittman, MBBS, FRCA; Shahar Bar-Yosef, MD; John SumPing,

More information

Minimally invasive cardiac output monitors

Minimally invasive cardiac output monitors Continuing Education in Anaesthesia, Critical Care & Pain Advance Access published October 3, 2011 Minimally invasive cardiac output monitors Kate E Drummond MBBS FANZCA Edward Murphy MBBS FANZCA Matrix

More information

Hemodynamic Monitoring and Circulatory Assist Devices

Hemodynamic Monitoring and Circulatory Assist Devices Hemodynamic Monitoring and Circulatory Assist Devices Speaker: Jana Ogden Learning Unit 2: Hemodynamic Monitoring and Circulatory Assist Devices Hemodynamic monitoring refers to the measurement of pressure,

More information

This PDF is available for free download from a site hosted by Medknow Publications

This PDF is available for free download from a site hosted by Medknow Publications Indian J Crit Care Med Oct-Dec 006 Vol 10 Issue 4 IJCCM October-December 003 Vol 7 Issue 4 Research Article Correlation of mixed venous and central venous oxygen saturation and its relation to cardiac

More information

A pilot assessment of the FloTrac TM cardiac output monitoring system

A pilot assessment of the FloTrac TM cardiac output monitoring system Intensive Care Med (2007) 33:344 349 DOI 10.1007/s00134-006-0410-4 TECHNICAL NOTE Helen Ingrid Opdam Li Wan Rinaldo Bellomo A pilot assessment of the FloTrac TM cardiac output monitoring system Received:

More information

Minimally invasive cardiac output monitors

Minimally invasive cardiac output monitors Kate E Drummond MBBS FANZCA Edward Murphy MBBS FANZCA Matrix reference 2A04 Key points Minimally invasive cardiac output monitors have varying degrees of invasiveness with some being totally non-invasive

More information

Volumetric preload measurement by thermodilution: a comparison with transoesophageal echocardiography

Volumetric preload measurement by thermodilution: a comparison with transoesophageal echocardiography BJA Advance Access published March 24, 25 British Journal of Anaesthesia Page 1 of 8 doi:1.193/bja/aei123 Volumetric preload measurement by thermodilution: a comparison with transoesophageal echocardiography

More information

PulsioFlex Patient focused flexibility

PulsioFlex Patient focused flexibility PulsioFlex Patient focused flexibility Modular platform with intelligent visualisation for advanced patient Minimally invasive perioperative cardiac output trend with ProAQT Enables calibrated cardiac

More information

Pulse Pressure Analysis

Pulse Pressure Analysis Pulse Pressure Analysis M. Cecconi, J. Wilson, and A. Rhodes z Introduction Cardiac output monitoring is part of routine practice in the critically ill patient. Recently, there has been increasing interest

More information

Received: 7 Jun 2006 Revisions requested: 28 Jun 2006 Revisions received: 30 Aug 2006 Accepted: 21 Nov 2006 Published: 21 Nov 2006

Received: 7 Jun 2006 Revisions requested: 28 Jun 2006 Revisions received: 30 Aug 2006 Accepted: 21 Nov 2006 Published: 21 Nov 2006 Available online http://ccforum.com/content/10/6/r164 Vol 10 No 6 Research Open Access Comparison of uncalibrated arterial waveform analysis in cardiac surgery patients with thermodilution cardiac output

More information

Impedance Cardiography (ICG) Method, Technology and Validity

Impedance Cardiography (ICG) Method, Technology and Validity Method, Technology and Validity Hemodynamic Basics Cardiovascular System Cardiac Output (CO) Mean arterial pressure (MAP) Variable resistance (SVR) Aortic valve Left ventricle Elastic arteries / Aorta

More information

At present, cardiac output is usually determined

At present, cardiac output is usually determined Original Article / Transplantation Continuous versus bolus cardiac output monitoring during orthotopic liver transplantation Fang-Ping Bao and Jian Wu Hangzhou, China BACKGROUND: Cardiac output monitoring

More information

University of Florida Department of Surgery. CardioThoracic Surgery VA Learning Objectives

University of Florida Department of Surgery. CardioThoracic Surgery VA Learning Objectives University of Florida Department of Surgery CardioThoracic Surgery VA Learning Objectives This service performs coronary revascularization, valve replacement and lung cancer resections. There are 2 faculty

More information

Chapter 2. Performance of three minimally invasive cardiac output monitoring systems

Chapter 2. Performance of three minimally invasive cardiac output monitoring systems Chapter 2 Performance of three minimally invasive cardiac output monitoring systems Rob de Wilde, Bart Geerts, Jisheng Cui, Paul van den Berg and Jos Jansen Anaesthesia 2009; 64: 762 769 (45 46) Ideally

More information

Shock, Monitoring Invasive Vs. Non Invasive

Shock, Monitoring Invasive Vs. Non Invasive Shock, Monitoring Invasive Vs. Non Invasive Paula Ferrada MD Assistant Professor Trauma, Critical Care and Emergency Surgery Virginia Commonwealth University Shock Fluid Pressors Ionotrope Intervention

More information

Radboud University Nijmegen Medical Centre Why measure cardiac output in critically ill children?

Radboud University Nijmegen Medical Centre Why measure cardiac output in critically ill children? Radboud University Nijmegen Medical Centre Why measure cardiac output in critically ill children? J. Lemson Anesthesiologist/(pediatric)intensivist Case; Girl 2 years, 12 kg, severe meningococcal septic

More information

ESTIMATION OF STROKE VOLUME BY NON-INVASIVE BLOOD PRESSURE MONITORING IN HUMAN CENTRIFUGE

ESTIMATION OF STROKE VOLUME BY NON-INVASIVE BLOOD PRESSURE MONITORING IN HUMAN CENTRIFUGE 84TH ASMA ANNUAL SCIENTIFIC MEETING CHICAGO - MAY 12-16, 213 ESTIMATION OF STROKE VOLUME BY NON-INVASIVE BLOOD PRESSURE MONITORING IN HUMAN CENTRIFUGE PROGRAM ID NUMBER 445 MANEN O, DUSSAULT C, SAUVET

More information

Posted: 11/27/2011 on Medscape; Published Br J Anaesth. 2011;107(2): Oxford University Press

Posted: 11/27/2011 on Medscape; Published Br J Anaesth. 2011;107(2): Oxford University Press Posted: 11/27/2011 on Medscape; Published Br J Anaesth. 2011;107(2):209-217. 2011 Oxford University Press Effect of Phenylephrine and Ephedrine Bolus Treatment on Cerebral Oxygenation in Anaesthetized

More information

Monitoring of cardiac index (CI) after surgery for congenital heart

Monitoring of cardiac index (CI) after surgery for congenital heart Cardiac index monitoring by pulse contour analysis and thermodilution after pediatric cardiac surgery U. Fakler, MD, a Ch. Pauli, MD, a G. Balling, MD, a H. P. Lorenz, MD, a A. Eicken, MD, a M. Hennig,

More information

Cardiac Output Monitoring - 6

Cardiac Output Monitoring - 6 Cardiac Output Monitoring - 6 How to use Wrexham s Cardiac Output Monitors. Wrexham Maelor Critical Care Version 02.05.16 Introduction Types of Devices: NICOM - Cheetah Oesophageal Doppler +/- Pulse Contour

More information

Nothing to Disclose. Severe Pulmonary Hypertension

Nothing to Disclose. Severe Pulmonary Hypertension Severe Ronald Pearl, MD, PhD Professor and Chair Department of Anesthesiology Stanford University Rpearl@stanford.edu Nothing to Disclose 65 year old female Elective knee surgery NYHA Class 3 Aortic stenosis

More information

Introduction. Invasive Hemodynamic Monitoring. Determinants of Cardiovascular Function. Cardiovascular System. Hemodynamic Monitoring

Introduction. Invasive Hemodynamic Monitoring. Determinants of Cardiovascular Function. Cardiovascular System. Hemodynamic Monitoring Introduction Invasive Hemodynamic Monitoring Audis Bethea, Pharm.D. Assistant Professor Therapeutics IV January 21, 2004 Hemodynamic monitoring is necessary to assess and manage shock Information obtained

More information

Cytoreductive Surgery and Hyperthermic Intraperitoneal Chemotherapy

Cytoreductive Surgery and Hyperthermic Intraperitoneal Chemotherapy Cytoreductive Surgery and Hyperthermic Intraperitoneal Chemotherapy A 44 year old female undergoing 10 hour Cytoreductive (CRS) procedure followed by Hyperthermic Intraperitoneal Chemotherapy (HIPEC).

More information

Transpulmonary versus continuous thermodilution cardiac output after valvular and coronary artery surgery

Transpulmonary versus continuous thermodilution cardiac output after valvular and coronary artery surgery doi:10.1510/icvts.2009.204545 Interactive CardioVascular and Thoracic Surgery 9 (2009) 4 8 www.icvts.org Work in progress report - Cardiac general Transpulmonary versus continuous thermodilution cardiac

More information

Hemodynamic Monitoring

Hemodynamic Monitoring Perform Procedure And Interpret Results Hemodynamic Monitoring Tracheal Tube Cuff Pressure Dean R. Hess PhD RRT FAARC Hemodynamic Monitoring Cardiac Rate and Rhythm Arterial Blood Pressure Central Venous

More information

Emergency Intraoperative Echocardiography

Emergency Intraoperative Echocardiography Emergency Intraoperative Echocardiography Justiaan Swanevelder Department of Anaesthesia, Glenfield Hospital University Hospitals of Leicester NHS Trust, UK Carl Gustav Jung (1875-1961) Your vision will

More information

Chapter 3. An evaluation of cardiac output by five arterial pulse contour techniques during cardiac surgery

Chapter 3. An evaluation of cardiac output by five arterial pulse contour techniques during cardiac surgery Chapter 3 An evaluation of cardiac output by five arterial pulse contour techniques during cardiac surgery R.B.P. de Wilde 1, J.J. Schreuder 2, P.C.M. van den Berg 1 and J.R.C. Jansen 1 1 Department of

More information

Cardiac Output MCQ. Professor of Cardiovascular Physiology. Cairo University 2007

Cardiac Output MCQ. Professor of Cardiovascular Physiology. Cairo University 2007 Cardiac Output MCQ Abdel Moniem Ibrahim Ahmed, MD Professor of Cardiovascular Physiology Cairo University 2007 90- Guided by Ohm's law when : a- Cardiac output = 5.6 L/min. b- Systolic and diastolic BP

More information

Perioperative Goal- Protocol Summary

Perioperative Goal- Protocol Summary Perioperative Goal- Directed Therapy Protocol Summary Evidence-based, perioperative Goal-Directed Therapy (GDT) protocols. Several single centre randomized controlled trials, meta-analysis and quality

More information

Assessing Preload Responsiveness Using Arterial Pressure Based Technologies. Patricia A. Meehan, RN, MS Education Consultant Edwards Lifesciences, LLC

Assessing Preload Responsiveness Using Arterial Pressure Based Technologies. Patricia A. Meehan, RN, MS Education Consultant Edwards Lifesciences, LLC Assessing Preload Responsiveness Using Arterial Pressure Based Technologies Patricia A. Meehan, RN, MS Education Consultant Edwards Lifesciences, LLC Content Description : Fluid administration is a first

More information

Perioperative Goal- Protocol Summary

Perioperative Goal- Protocol Summary Perioperative Goal- Directed Therapy Protocol Summary Evidence-based, perioperative Goal-Directed Therapy (GDT) protocols. Several single centre randomized controlled trials, meta-analysis and quality

More information

Intra-operative Echocardiography: When to Go Back on Pump

Intra-operative Echocardiography: When to Go Back on Pump Intra-operative Echocardiography: When to Go Back on Pump GREGORIO G. ROGELIO, MD., F.P.C.C. OUTLINE A. Indications for Intraoperative Echocardiography B. Role of Intraoperative Echocardiography C. Criteria

More information

Assessing fluid responses after coronary surgery: role of mathematical coupling of global end-diastolic volume to

Assessing fluid responses after coronary surgery: role of mathematical coupling of global end-diastolic volume to 954 Original article Assessing fluid responses after coronary surgery: role of mathematical coupling of global end-diastolic volume to cardiac output measured by transpulmonary thermodilution Rose-Marieke

More information

Interesting Cases - A Case Report: Renal Cell Carcinoma With Tumor Mass In IVC And Heart. O Wenker, L Chaloupka, R Joswiak, D Thakar, C Wood, G Walsh

Interesting Cases - A Case Report: Renal Cell Carcinoma With Tumor Mass In IVC And Heart. O Wenker, L Chaloupka, R Joswiak, D Thakar, C Wood, G Walsh ISPUB.COM The Internet Journal of Thoracic and Cardiovascular Surgery Volume 3 Number 2 Interesting Cases - A Case Report: Renal Cell Carcinoma With Tumor Mass In IVC And Heart O Wenker, L Chaloupka, R

More information

Continuous monitoring of cardiac output: why and how

Continuous monitoring of cardiac output: why and how 33rd International Symposium of Intensive Care and Emergency Medicine, Brussels 19.03.-22.03.2013 UNIVERSITÄTSKLINIKUM Schleswig-Holstein Continuous monitoring of cardiac output: why and how Berthold Bein,

More information

Emergency surgery in acute coronary syndrome

Emergency surgery in acute coronary syndrome Emergency surgery in acute coronary syndrome Teerawoot Jantarawan Division of Cardiothoracic Surgery, Department of Surgery, Faculty of Medicine Siriraj Hospital, Mahidol University, Bangkok, Thailand

More information

Capture every aspect of hemodynamic status

Capture every aspect of hemodynamic status Capture every aspect of hemodynamic status Edwards Advanced Technology Swan-Ganz Catheters Hemodynamic Monitoring Systems with Continuous CO, SvO 2, EDV, RVEF, SVR, and SV for Optimal Patient Care The

More information

The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for

The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for 1 2 The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for continuous central venous oximetry (ScvO 2 ) 3 The Vigileo

More information

escco, estimated Continuous Cardiac Output device DESEBBE OLIVIER SAUVEGARDE CLINIC, LYON, FRANCE

escco, estimated Continuous Cardiac Output device DESEBBE OLIVIER SAUVEGARDE CLINIC, LYON, FRANCE escco, estimated Continuous Cardiac Output device DESEBBE OLIVIER SAUVEGARDE CLINIC, LYON, FRANCE No conflict of interest u Technology, based on Pulse Wave Transit Time u Limits of the technology u Precision

More information

Topics to be Covered. Cardiac Measurements. Distribution of Blood Volume. Distribution of Pulmonary Ventilation & Blood Flow

Topics to be Covered. Cardiac Measurements. Distribution of Blood Volume. Distribution of Pulmonary Ventilation & Blood Flow Topics to be Covered MODULE F HEMODYNAMIC MONITORING Cardiac Output Determinants of Stroke Volume Hemodynamic Measurements Pulmonary Artery Catheterization Control of Blood Pressure Heart Failure Cardiac

More information

Hemodynamic Monitoring in Critically ill Patients in Arthur Simonnet, interne Tuteur : Pr. Raphaël Favory

Hemodynamic Monitoring in Critically ill Patients in Arthur Simonnet, interne Tuteur : Pr. Raphaël Favory Hemodynamic Monitoring in Critically ill Patients in 2017 Arthur Simonnet, interne Tuteur : Pr. Raphaël Favory Rationale for Hemodynamic Monitoring Identify the presence of hemodynamic instability Identify

More information

The effects of vasoactive drugs on pulse pressure and stroke volume variation in postoperative ventilated patients,

The effects of vasoactive drugs on pulse pressure and stroke volume variation in postoperative ventilated patients, Journal of Critical Care (211) 26, 328.e1 328.e8 The effects of vasoactive drugs on pulse pressure and stroke volume variation in postoperative ventilated patients, Mehrnaz Hadian MD a, Donald A. Severyn

More information

Coronary Artery Bypass Graft: Monitoring Patients and Detecting Complications

Coronary Artery Bypass Graft: Monitoring Patients and Detecting Complications Coronary Artery Bypass Graft: Monitoring Patients and Detecting Complications Madhav Swaminathan, MD, FASE Professor of Anesthesiology Division of Cardiothoracic Anesthesia & Critical Care Duke University

More information

Mechanical Ventilation & Cardiopulmonary Interactions: Clinical Application in Non- Conventional Circulations. Eric M. Graham, MD

Mechanical Ventilation & Cardiopulmonary Interactions: Clinical Application in Non- Conventional Circulations. Eric M. Graham, MD Mechanical Ventilation & Cardiopulmonary Interactions: Clinical Application in Non- Conventional Circulations Eric M. Graham, MD Background Heart & lungs work to meet oxygen demands Imbalance between supply

More information

Hemodynamic monitoring beyond cardiac output

Hemodynamic monitoring beyond cardiac output Hemodynamic monitoring beyond cardiac output Prof Xavier MONNET Medical Intensive Care Unit Bicêtre Hospital Assistance publique Hôpitaux de Paris FRANCE Conflicts of interest Lilly GlaxoSmithKline Pulsion

More information

Cardiothoracic Fellow Expectations Division of Cardiac Anesthesia, Beth Israel Deaconess Medical Center

Cardiothoracic Fellow Expectations Division of Cardiac Anesthesia, Beth Israel Deaconess Medical Center The fellowship in Cardiothoracic Anesthesia at the Beth Israel Deaconess Medical Center is intended to provide the foundation for a career as either an academic cardiothoracic anesthesiologist or clinical

More information

Heart may be rotated but not compressed

Heart may be rotated but not compressed Tips And Techniques For Multivessel OPCAB John D. Puskas, MD, Emory University, Atlanta AATS Adult Cardiac Skills April 28, 2012 San Francisco, CA Beating Heart Surgery vs Beat The Heart Surgery OPCAB

More information

W. J. RUSSELL*, M. F. JAMES

W. J. RUSSELL*, M. F. JAMES Anaesth Intensive Care 2004; 32: 644-648 The Effects on Arterial Haemoglobin Oxygen Saturation and on Shunt of Increasing Cardiac Output with Dopamine or Dobutamine During One-lung Ventilation W. J. RUSSELL*,

More information

Importance of the third arterial graft in multiple arterial grafting strategies

Importance of the third arterial graft in multiple arterial grafting strategies Research Highlight Importance of the third arterial graft in multiple arterial grafting strategies David Glineur Department of Cardiovascular Surgery, Cliniques St Luc, Bouge and the Department of Cardiovascular

More information

Sepsis Wave II Webinar Series. Sepsis Reassessment

Sepsis Wave II Webinar Series. Sepsis Reassessment Sepsis Wave II Webinar Series Sepsis Reassessment Presenters Nova Panebianco, MD Todd Slesinger, MD Fluid Reassessment in Sepsis Todd L. Slesinger, MD, FACEP, FCCM, FCCP, FAAEM Residency Program Director

More information

Cardiovascular Nursing Practice: A Comprehensive Resource Manual and Study Guide for Clinical Nurses 2 nd Edition

Cardiovascular Nursing Practice: A Comprehensive Resource Manual and Study Guide for Clinical Nurses 2 nd Edition Cardiovascular Nursing Practice: A Comprehensive Resource Manual and Study Guide for Clinical Nurses 2 nd Edition Table of Contents Volume 1 Chapter 1: Cardiovascular Anatomy and Physiology Basic Cardiac

More information

Das recht Ventrikel ist auch noch da! RV function The RV operates as. Physiology Not very sensitive to preload Good compliance of the free wall

Das recht Ventrikel ist auch noch da! RV function The RV operates as. Physiology Not very sensitive to preload Good compliance of the free wall Das recht Ventrikel ist auch noch da! I.Michaux Intensive Care Medicine University Hospital CHU UCL Namur Mont-Godinne Belgium RV function The RV operates as a low pressure, volume pump, moving the blood

More information

Frequently asked questions and answers:

Frequently asked questions and answers: Frequently asked questions and answers: General 1) What are the indications and contraindications for PiCCO-Technology? Indications: Patients in whom cardiovascular and circulatory volume status monitoring

More information

Assessing Cardiac Risk in Noncardiac Surgery. Murali Sivarajan, M.D. Professor University of Washington Seattle, Washington

Assessing Cardiac Risk in Noncardiac Surgery. Murali Sivarajan, M.D. Professor University of Washington Seattle, Washington Assessing Cardiac Risk in Noncardiac Surgery Murali Sivarajan, M.D. Professor University of Washington Seattle, Washington Disclosure None. I have no conflicts of interest, financial or otherwise. CME

More information

Disclaimer. Improving MET-based patient care using treatment algorithms. Michael R. Pinsky, MD, Dr hc. Different Environments Demand Different Rules

Disclaimer. Improving MET-based patient care using treatment algorithms. Michael R. Pinsky, MD, Dr hc. Different Environments Demand Different Rules Michael R. Pinsky, MD - June 29, 26 Improving MET-based patient care using Michael R. Pinsky, MD, Dr hc Department of Critical Care Medicine University of Pittsburgh Disclaimer Michael R. Pinsky, MD is

More information

ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE

ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE Vet Times The website for the veterinary profession https://www.vettimes.co.uk ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE Author : CARL BRADBROOK Categories : Vets Date : October 7, 2013

More information

Adult Echocardiography Examination Content Outline

Adult Echocardiography Examination Content Outline Adult Echocardiography Examination Content Outline (Outline Summary) # Domain Subdomain Percentage 1 2 3 4 5 Anatomy and Physiology Pathology Clinical Care and Safety Measurement Techniques, Maneuvers,

More information

FloTrac Sensor and Edwards PreSep Central Venous Oximetry Catheter Case Presentations

FloTrac Sensor and Edwards PreSep Central Venous Oximetry Catheter Case Presentations Edwards FloTrac Sensor & Edwards Vigileo Monitor FloTrac Sensor and Edwards PreSep Central Venous Oximetry Catheter Case Presentations 1 Topics System Configuration FloTrac Sensor and PreSep Catheter Thoracotomy

More information

PERIOPERATIVE MYOCARDIAL INFARCTION THE ANAESTHESIOLOGIST'S VIEW

PERIOPERATIVE MYOCARDIAL INFARCTION THE ANAESTHESIOLOGIST'S VIEW PERIOPERATIVE MYOCARDIAL INFARCTION THE ANAESTHESIOLOGIST'S VIEW Bruce Biccard Perioperative Research Group, Department of Anaesthetics 18 June 2015 Disclosure Research funding received Medical Research

More information

Cardiovascular Practice Quiz

Cardiovascular Practice Quiz Cardiovascular Practice Quiz 1. When caring for a patient following a cardiac catheterization with coronary angiography, which of the following findings would be of most concern to the nurse? a. Swelling

More information

Special circulations, Coronary, Pulmonary. Faisal I. Mohammed, MD,PhD

Special circulations, Coronary, Pulmonary. Faisal I. Mohammed, MD,PhD Special circulations, Coronary, Pulmonary Faisal I. Mohammed, MD,PhD 1 Objectives Describe the control of blood flow to different circulations (Skeletal muscles, pulmonary and coronary) Point out special

More information

Images have been removed from the PowerPoint slides in this handout due to copyright restrictions.

Images have been removed from the PowerPoint slides in this handout due to copyright restrictions. Percutaneous Coronary Intervention https://www.youtube.com/watch?v=bssqnhylvma Types of PCI Procedures Balloon Angioplasty Rotational Atherectomy Coronary Stent Balloon Inflation Rotational Atherectomy

More information

The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the Edwards PreSep oximetry catheter

The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the Edwards PreSep oximetry catheter 1 2 The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the Edwards PreSep oximetry catheter for continuous central venous oximetry (ScvO2) 3

More information

Transpulmonary thermodilution in septic shock & ARDS

Transpulmonary thermodilution in septic shock & ARDS Transpulmonary thermodilution in septic shock & ARDS F Michard, MD, PhD Lea Michard Photography Disclosure Intensivist Pulsion, UPMED & Edwards Patents Transpulmonary thermodilution CVC, A line (femoral)

More information

Impedance Cardiography (ICG) Application of ICG in Intensive Care and Emergency

Impedance Cardiography (ICG) Application of ICG in Intensive Care and Emergency Impedance Cardiography (ICG) Application of ICG in Intensive Care and Emergency Aim of haemodynamic monitoring in ICU and ED Detection and therapy of insufficient organ perfusion Answers to common cardiovascular

More information

Useful? Definition of High-risk? Pre-OP/Intra-OP/Post-OP? Complication vs Benefit? Mortality? Morbidity?

Useful? Definition of High-risk? Pre-OP/Intra-OP/Post-OP? Complication vs Benefit? Mortality? Morbidity? Preoperative intraaortic balloon counterpulsation in high-risk CABG Stefan Klotz, M.D. Preoperative IABP in high-risk CABG Questions?? Useful? Definition of High-risk? Pre-OP/Intra-OP/Post-OP? Complication

More information

Tailored Volume Resuscitation in the Critically Ill is Achievable. Objectives. Clinical Case 2/16/2018

Tailored Volume Resuscitation in the Critically Ill is Achievable. Objectives. Clinical Case 2/16/2018 Tailored Volume Resuscitation in the Critically Ill is Achievable Heath E Latham, MD Associate Professor Fellowship Program Director Pulmonary and Critical Care Objectives Describe the goal of resuscitation

More information

COMPARISON OF SUFENTANIL-OXYGEN AND FENTANYL-OXYGEN ANAESTHESIA FOR CORONARY ARTERY BYPASS GRAFTING

COMPARISON OF SUFENTANIL-OXYGEN AND FENTANYL-OXYGEN ANAESTHESIA FOR CORONARY ARTERY BYPASS GRAFTING Br. J. Anaesth. (1988), 60, 530-535 COMPARISON OF SUFENTANIL-OXYGEN AND FENTANYL-OXYGEN ANAESTHESIA FOR CORONARY ARTERY BYPASS GRAFTING H. M. L. MATHEWS, G. FURNESS, I. W. CARSON, I. A. ORR, S. M. LYONS

More information

Dr. F Javier Belda Dept. Anesthesiology and Critical Care Hospital Clinico Universitario Valencia (Spain) Pulsion MAB

Dr. F Javier Belda Dept. Anesthesiology and Critical Care Hospital Clinico Universitario Valencia (Spain) Pulsion MAB State of the Art Hemodynamic Monitoring III CO, preload, lung water and ScvO2 The winning combination! Dr. F Javier Belda Dept. Anesthesiology and Critical Care Hospital Clinico Universitario Valencia

More information

Pre-operative usage of IABP for patients for by pass surgery

Pre-operative usage of IABP for patients for by pass surgery Pre-operative usage of IABP for patients for by pass surgery Mitrev Z, Anguseva T, Hristov N Special hospital for surgery Filip Vtori Skopje - Macedonija Oktomvri, 2008 IABP Background Preload Afterload

More information

PCV and PAOP Old habits die hard!

PCV and PAOP Old habits die hard! PCV and PAOP Old habits die hard! F Javier Belda MD, PhD Head of Department Associate Professor Anaesthesia and Critical Care Hospital Clínico Universitario Valencia (SPAIN) An old example TOBACO SMOKING

More information

PHYSIOLOGY MeQ'S (Morgan) All the following statements related to blood volume are correct except for: 5 A. Blood volume is about 5 litres. B.

PHYSIOLOGY MeQ'S (Morgan) All the following statements related to blood volume are correct except for: 5 A. Blood volume is about 5 litres. B. PHYSIOLOGY MeQ'S (Morgan) Chapter 5 All the following statements related to capillary Starling's forces are correct except for: 1 A. Hydrostatic pressure at arterial end is greater than at venous end.

More information

Rationale for Prophylactic Support During Percutaneous Coronary Intervention

Rationale for Prophylactic Support During Percutaneous Coronary Intervention Rationale for Prophylactic Support During Percutaneous Coronary Intervention Navin K. Kapur, MD, FACC, FSCAI Assistant Director, Interventional Cardiology Director, Interventional Research Laboratories

More information

FUNDAMENTALS OF HEMODYNAMICS, VASOACTIVE DRUGS AND IABP IN THE FAILING HEART

FUNDAMENTALS OF HEMODYNAMICS, VASOACTIVE DRUGS AND IABP IN THE FAILING HEART FUNDAMENTALS OF HEMODYNAMICS, VASOACTIVE DRUGS AND IABP IN THE FAILING HEART CINDY BITHER, MSN, ANP, ANP, AACC, CHFN CHIEF NP, ADV HF PROGRAM MEDSTAR WASHINGTON HOSPITAL CENTER CONFLICTS OF INTEREST NONE

More information

Goal-directed vs Flow-guidedresponsive

Goal-directed vs Flow-guidedresponsive Goal-directed vs Flow-guidedresponsive therapy S Magder Department of Critical Care, McGill University Health Centre Flow-directed vs goal directed strategy for management of hemodynamics S Magder Curr

More information

Cardiology. Objectives. Chapter

Cardiology. Objectives. Chapter 1:44 M age 1121 Chapter Cardiology Objectives art 1: Cardiovascular natomy and hysiology, ECG Monitoring, and Dysrhythmia nalysis (begins on p. 1127) fter reading art 1 of this chapter, you should be able

More information

Evaluation of Stroke Volume Variation Obtained by the FloTrac /Vigileo System to Guide Preoperative Fluid Therapy in Patients Undergoing Brain Surgery

Evaluation of Stroke Volume Variation Obtained by the FloTrac /Vigileo System to Guide Preoperative Fluid Therapy in Patients Undergoing Brain Surgery The Journal of International Medical Research 2012; 40: 1175 1181 [first published online as 40(3) 11] Evaluation of Stroke Volume Variation Obtained by the FloTrac /Vigileo System to Guide Preoperative

More information

Section 6 Intra Aortic Balloon Pump

Section 6 Intra Aortic Balloon Pump Section 6 Intra Aortic Balloon Pump The Intra Aortic Balloon Pump (IABP) The balloon is synthetic and is made for single use only. It is threaded into the aorta, usually via a femoral approach. The balloon

More information

Fluid bolus of 20% Albumin in post-cardiac surgical patient: a prospective observational study of effect duration

Fluid bolus of 20% Albumin in post-cardiac surgical patient: a prospective observational study of effect duration Fluid bolus of 20% Albumin in post-cardiac surgical patient: a prospective observational study of effect duration Investigators: Salvatore Cutuli, Eduardo Osawa, Rinaldo Bellomo Affiliations: 1. Department

More information

Alfa Ferry FRCS Cardiac Surgeon OPERATIVE MANAGEMENT IN CORONARY ARTERY DISEASE

Alfa Ferry FRCS Cardiac Surgeon OPERATIVE MANAGEMENT IN CORONARY ARTERY DISEASE Alfa Ferry FRCS Cardiac Surgeon OPERATIVE MANAGEMENT IN CORONARY ARTERY DISEASE Management in CHD Medical (medikamentosa) Intervensi 1. Percutaneous ( PTCA & stenting ) 2. Surgical ( CABG, CABG & mitral

More information

FAILURE IN PATIENTS WITH MYOCARDIAL INFARCTION

FAILURE IN PATIENTS WITH MYOCARDIAL INFARCTION Br. J. clin. Pharmac. (1982), 14, 187S-19lS BENEFICIAL EFFECTS OF CAPTOPRIL IN LEFT VENTRICULAR FAILURE IN PATIENTS WITH MYOCARDIAL INFARCTION J.P. BOUNHOURE, J.G. KAYANAKIS, J.M. FAUVEL & J. PUEL Departments

More information

CARDIAC OUTPUT Monitoring ANDY CAMPBELL JOURNAL CLUB NOV 2011

CARDIAC OUTPUT Monitoring ANDY CAMPBELL JOURNAL CLUB NOV 2011 CARDIAC OUTPUT Monitoring ANDY CAMPBELL JOURNAL CLUB NOV 2011 Is keeping up the pressure enough? It is a source of regret that the measurement of flow is so much more difficult than the measurement of

More information

Cigna - Prior Authorization Procedure List Cardiology

Cigna - Prior Authorization Procedure List Cardiology Cigna - Prior Authorization Procedure List Cardiology Category CPT Code CPT Code Description 33206 Insertion of new or replacement of permanent pacemaker with transvenous electrode(s); atrial 33207 Insertion

More information

Propofol or etomidate: Does it genuinely matter for induction in cardiac surgical procedures?

Propofol or etomidate: Does it genuinely matter for induction in cardiac surgical procedures? Original Research Article DOI: 10.18231/2394-4994.2016.0012 Propofol or etomidate: Does it genuinely matter for induction in cardiac surgical procedures? Manjunath Ratnakara Kamath 1,*, Suchitha Kamath

More information

The Cardiovascular System

The Cardiovascular System The Cardiovascular System The Cardiovascular System A closed system of the heart and blood vessels The heart pumps blood Blood vessels allow blood to circulate to all parts of the body The function of

More information

British Journal of Anaesthesia 101 (6): (2008) doi: /bja/aen277 Advance Access publication October 12, 2008

British Journal of Anaesthesia 101 (6): (2008) doi: /bja/aen277 Advance Access publication October 12, 2008 British Journal of Anaesthesia 11 (6): 761 8 (28) doi:1.193/bja/aen277 Advance Access publication October 12, 28 CARDIOVASCULAR Uncalibrated pulse contour-derived stroke volume variation predicts fluid

More information

Heart disease remains the leading cause of morbidity and mortality in industrialized nations. It accounts for nearly 40% of all deaths in the United

Heart disease remains the leading cause of morbidity and mortality in industrialized nations. It accounts for nearly 40% of all deaths in the United Heart disease remains the leading cause of morbidity and mortality in industrialized nations. It accounts for nearly 40% of all deaths in the United States, totaling about 750,000 individuals annually

More information

Adult Cardiac Surgery

Adult Cardiac Surgery Adult Cardiac Surgery Mahmoud ABU-ABEELEH Associate Professor Department of Surgery Division of Cardiothoracic Surgery School of Medicine University Of Jordan Adult Cardiac Surgery: Ischemic Heart Disease

More information

International Journal of Medical and Health Sciences

International Journal of Medical and Health Sciences International Journal of Medical and Health Sciences Journal Home Page: http://www.ijmhs.net ISSN:2277-4505 Original article A Comparative Study Of Etomidate And Midazolam Induction In Patients Undergoing

More information

Management of new-onset AF: Initial rate control treatment

Management of new-onset AF: Initial rate control treatment Geneva Acute Crdiovascular Care Congress 2014 - October 18-20, 2014 Management of new-onset AF: Initial rate control treatment Antonio Raviele, MD, FESC, FHRS ALFA Alliance to Fight Atrial fibrillation,

More information

PERIPARTUM CARDIOMYOPATHY

PERIPARTUM CARDIOMYOPATHY PERIPARTUM CARDIOMYOPATHY Dr.T.Venkatachalam. Professor of Anaesthesiology Madras Medical College, Chennai Peripartum cardiomyopathy is defined as the onset of acute heart failure without demonstrable

More information

Central haemodynamics during spontaneous angina pectoris

Central haemodynamics during spontaneous angina pectoris British Heart Journal, I974, 36, I0-I09I Central haemodynamics during spontaneous angina pectoris From the Department of Clinical Physiology, Malmo General Hospital, S-214 OI Malmo, Sweden. Central pressures

More information

CATCH A WAVE.. INTRODUCTION NONINVASIVE HEMODYNAMIC MONITORING 4/12/2018

CATCH A WAVE.. INTRODUCTION NONINVASIVE HEMODYNAMIC MONITORING 4/12/2018 WAVES CATCH A WAVE.. W I S C O N S I N P A R A M E D I C S E M I N A R A P R I L 2 0 1 8 K E R I W Y D N E R K R A U S E R N, C C R N, E M T - P Have you considered that if you don't make waves, nobody

More information