At present, cardiac output is usually determined
|
|
- Leonard Grant
- 6 years ago
- Views:
Transcription
1 Original Article / Transplantation Continuous versus bolus cardiac output monitoring during orthotopic liver transplantation Fang-Ping Bao and Jian Wu Hangzhou, China BACKGROUND: Cardiac output monitoring is important for critical patients. This study aimed to determine the delayed response of continuous cardiac output (CCO) thermodilution measurement, whether CCO and bolus cardiac output (BCO) thermodilution agree sufficiently to be used interchangeably, and whether CCO monitoring is reliable for patients undergoing liver transplantation. METHODS: Thirteen patients undergoing liver transplantation without veno-venous bypass were studied (37-66 years old, weight kg). Continuous and bolus thermodilution measurements were performed at predefined time points using an "Opti-Q" SvO 2 /CCO monitor (Abbott Laboratories, North Chicago, IL, USA). Bias and 95% limits of agreement were calculated according to Bland and Altman analysis. The limits of agreement by which two methods are judged to be interchangeable were defined in advance as ±(13% BCO mean ) L/min. The repeatability and relative error of CCO, and the differences between CCO and the mean of the two measurements were calculated. RESULTS: Cardiac output measurements yielded 196 data pairs with ranges of 1.9 to 17.9 L/min for CCO and 2.1 to 18.3 L/min for BCO. The response time of CCO was delayed in the early phases after caval clamping and after reperfusion. At most of the measurement points, bias and 95% limits of agreement were -0.18±1.91 L/min. 95% limits of agreement did not fall within the predetermined limits of agreement of ±1.14 L/min. The repeatability coefficient of CCO was 0.36 L/min and the relative error was 4.6±4.7%. The mean difference between CCO and the average of the two methods was L/min (0.49 L/min). Author Affiliations: Department of Anesthesiology, First Affiliated Hospital, Zhejiang University School of Medicine, Hangzhou , China (Bao FP and Wu J) Corresponding Author: Jian Wu, MD, Department of Anesthesiology, First Affiliated Hospital, Zhejiang University School of Medicine, Hangzhou , China (Tel: ; wujian888888@ 163.com) 2008, Hepatobiliary Pancreat Dis Int. All rights reserved. CONCLUSIONS: In patients undergoing liver transplantation, the delayed response of CCO limits its application during the early phases after caval clamping and after reperfusion of the graft. The two methods are not interchangeable even in hemodynamic stability. Continuous thermodilution monitoring, however, is reliable or acceptable for clinical purposes. (Hepatobiliary Pancreat Dis Int 2008; 7: ) KEY WORDS: cardiac output; thermodilution; anesthesia; liver transplantation Introduction At present, cardiac output is usually determined by the intermittent bolus thermodilution technique. However, the accuracy of the bolus cardiac output (BCO) measurement can be influenced by various factors, and some shortcomings limit its clinical application. [1, 2] Acute hemodynamic instability may occur in critically ill patients, so continuous measurement of cardiac output is preferable to intermittent measurement. Among the methods of continuous measurement, thermodilution is userindependent and semi-continuous and is widely accepted. Two continuous cardiac output (CCO) monitors are on the market: the "Vigilance" CCO/ oximetry system (Baxter Edwards Critical-Care, Irvine, CA, USA) and the "Opti-Q" SvO 2 /CCO system (Abbott Laboratories, North Chicago, IL, USA). They measure cardiac output using the same basic thermodilution principles and use a filament to generate heat. However, they have different signal generation and signal processing technologies. Both systems have a response time of 3-6 minutes, [3] which is too long to monitor acute changes in cardiac output. [4-7] This delayed response of CCO measurement limits its clinical application during acute hemodynamic instability. [4-9] 138 Hepatobiliary Pancreat Dis Int,Vol 7,No 2 April 15,2008
2 Continuous versus bolus cardiac output monitoring during OLT Several studies have analyzed the agreement between CCO and BCO measurements under [3, 4, 10-16] stable conditions, and the performance of the "vigilance" monitor in liver transplantation patients has also been evaluated. [8, 16, 17] However, the performance and equivalence of CCO compared with BCO has not been tested at the level of scrutiny expected to qualify as "evidence-based" monitoring options. [1, 18] It may be that the two methods are not interchangeable [19, 20] according to Bland and Altman [21, 22] analysis. In this study, we used the "Opti-Q" SvO 2 /CCO system in patients undergoing orthotopic liver transplantation without veno-venous bypass to 1) measure the delayed response of CCO during the early periods after caval clamping and after reperfusion of the liver graft; 2) assess the agreement between CCO and BCO measurements, that is, to determine whether we can use the two interchangeably; and 3) evaluate whether CCO monitoring is reliable and acceptable for clinical purposes. Methods After approval by our clinical research committee and written informed consent had been obtained, 13 adult patients (11 male and 2 female, years old, weighing kg) were enrolled in the study. They were suffering from end-stage liver disease or hepatocellular carcinoma, and were about to undergo orthotopic liver transplantation, without veno-venous bypass. Patients with valvular heart disease and intracardiac shunts were excluded. All patients were not premedicated. On the patient's arrival at the operating room, 3 forearm venous and left radial arterial lines were established, and a pulmonary artery flotation catheter (Opti-Q SvO 2 /CCO, 8F) was inserted through a 8.5-French introducer via the right internal jugular vein under local anesthesia before induction of anesthesia. The catheter was connected to a Q-Vue computer operated in the "normal" mode (cardiac output values were averaged over a period of 5 minutes in this mode). The proper position of the catheter was confirmed by pressure monitoring from its proximal and distal ports. Anesthesia was induced with etomidate (0.3 mg/kg), midazolam (0.05 mg/kg), fentanyl (5 μg/kg), and vecuronium (0.1 mg/kg), and maintained with midazolam, fentanyl, isoflurane (0.5%-1.5%) and vecuronium. Ventilation was adjusted to maintain end-tidal pco 2 at mmhg. Heart rate, SpO 2, invasive blood pressure, central venous pressure and pulmonary artery mean pressure were monitored. The patients were placed on a warming blanket (Gaymar Meditherm, Orchard Park, NY, USA). Blood products and colloid solutions were given through a warming infusion device (HOT LINE, SIMS Medical System, Graseby Ltd., UK) and other room temperature crystalloid solutions infusions were maintained at a rate of about 500 ml/h via the peripheral veins. Through another peripheral vein, dopamine (3 μg/kg/min) was given throughout the operation, and epinephrine was administered to stabilize the circulation after upper inferior vena cava clamping and, if needed, was continued to the end of the operation. CCO and BCO measurements were obtained at the following predefined time during the study period: 20, 40, 60, and 90 minutes after induction of anesthesia, 5, 15, 25, 35, and 45 minutes after caval clamping, and 5, 15, 25, 60, 90, and 120 minutes after reperfusion of the liver graft. At each measurement point, CCO was measured immediately before and after BCO measurement, and the means of these CCO data pairs were recorded. BCO was measured by manual injection of 10 ml iced saline solution through the injectate port of the CCO catheter using a closed injectate system. The average of three measurements, measured within a 10% range distributed randomly over the respiratory cycle, was calculated. All bolus thermodilution measurements were performed by the same person to avoid inter-individual variations. Intravenous infusions through the central venous catheter were stopped and the setting of the ventilator remained the same during the measurement. The peripheral volume infusion rates and change of epinephrine administration were independent of the cardiac output measurements and were adjusted to maintain hemodynamic stability according to clinical considerations (i.e. blood loss, mean blood pressure, and operation stage). In addition, to measure the delayed response phenomenon of CCO, its values were recorded before caval clamping and at 1, 2, 3, and 5 minutes, then at 5-minute intervals after caval clamping until before reperfusion of the liver graft, and at 1, 2, 3, and 5 minutes, then at 5-minute intervals until 30 minutes after the reperfusion. Rectal temperature and blood temperature in the pulmonary artery were recorded continuously. Statistical analysis Statistical analysis was performed using linear regression for the data pairs, Pearson's correlation coefficient, and the method described by Bland and Hepatobiliary Pancreat Dis Int,Vol 7,No 2 April 15,
3 Hepatobiliary & Pancreatic Diseases International Table 1. Mean values of cardiac output and correlation, bias and 95% limits of agreement between continuous and bolus cardiac output measurements at predefined times Measurement time points (min) CCO (L/min) BCO (L/min) Correlation (r)/p value Bias (L/min) 95% limits of agreement (L/min) After induction of anesthesia (3.5) 9.0 (3.4) 0.93/< to (3.2) 9.8 (3.2) 0.93/< to (2.7) 9.9 (2.3) 0.95/< to (1.9) 9.1 (2.2) 0.88/ to (2.3) 8.6 (1.8) 0.88/ to 2.85 After caval clamping (2.4) 5.5 (1.8) 0.72/ to (2.1) 5.7 (2.1) 0.94/< to (1.8) 6.3 (1.9) 0.89/< to (2.1) 6.0 (2.2) 0.96/< to (2.7) 6.1 (2.6) 0.96/< to 1.26 After reperfusion of liver graft (2.8) 11.7 (3.8) 0.73/ to (3.5) 11.3 (2.9) 0.93/< to (3.5) 10.9 (3.5) 0.98/< to (3.6) 10.2 (3.7) 0.98/< to (3.6) 10.4 (4.1) 0.99/< to (1.8) 8.8 (1.8) 0.88/ to 1.84 Altman. [21, 22] Bias between the methods was calculated as the mean difference between CCO and BCO, and tests for normality of distribution were made. The upper and lower limits of agreement were calculated as bias ±1.96SD, and defined the range in which 95% of the differences between the methods were expected to lie. The differences between each pair of values were plotted over the average of each pair. The precision of the bias analysis and the limits of agreement were assessed using 95% confidence intervals (CI). The repeatability coefficient of each method was calculated using one-way analysis of variance. [22] We defined the limits within which the two methods were judged to be interchangeable as ±(13% BCO mean ) L/min. [23] In addition, the mean difference between CCO and the assumed true value (average value of two methods) was calculated, and these differences were also plotted over the assumed true values accordingly. The relative error of CCO measurement, defined as 100 [CCO- (BCO+CCO)/2]/[(BCO+CCO)/2], was calculated. P<0.05 was considered statistically significant. Results The mean time in the three phases (paleohepatic, anhepatic, and neohepatic) during the operation was 133.8±22.8, 53.9±12.3 and 148.8±34.7 minutes, respectively. Boluses of epinephrine (10-80 μg) were required after upper inferior vena cava clamping and after reperfusion of the graft; meanwhile, infusions of epinephrine were given to stabilize the circulation, and mean arterial pressure was maintained above 55 mmhg throughout most of the procedure in all patients. All patients had successful transplants and survived at the end of the operation. No adverse effects of the "Opti-Q" SvO 2 /CCO catheter were noted during the study period. Since it was impossible to make all measurement as scheduled without interfering with treatment, a total of 196 data pairs were collected from 13 patients. The ranges of measured cardiac output values using bolus and continuous cardiac output were and L/min, respectively. At 5 minutes after caval clamping and 5 minutes after reperfusion of the graft, the correlation markedly decreased and the bias increased (Table 1). CCO decreased slowly in the early phase after caval clamping and was close to BCO at 15 minutes of clamping. In contrast, CCO increased slowly in the early phase after reperfusion and was close to BCO at 15 minutes of reperfusion. In the early phases after caval clamping and reperfusion, the delayed response of CCO when acute changes in cardiac output occurred was evident. So, the data pairs of two time points (5 minutes after caval clamping and 5 minutes after reperfusion) were excluded from the ensuing data analysis. The results of the analysis of agreement, 140 Hepatobiliary Pancreat Dis Int,Vol 7,No 2 April 15,2008
4 Continuous versus bolus cardiac output monitoring during OLT Fig. 1. Bland and Altman diagram for CCO and BCO. This analysis included 170 data pairs in the stable phases during liver transplantation, and a bias of L/min with 95% limits of agreement of to 1.73 L/min was determined. Table 2. Mean values of pulmonary arterial blood and rectal temperatures and difference between these measurements at each predefined time Measurement time points (min) After induction of anesthesia Pulmonary arterial blood temperature ( ) Rectal temperature ( ) Temperature difference ( ) (1.1) 36.9 (1.0) 0.2 (0.1) (1.0) 36.8 (1.0) 0.2 (0.1) (0.9) 36.8 (1.0) 0.3 (0.1) (0.9) 36.8 (1.0) 0.2 (0.1) (1.0) 37.0 (1.1) 0.3 (0.2) After caval clamping (0.8) 36.9 (0.9) 0.3 (0.2) (0.9) 36.8 (1.0) 0.4 (0.2) (1.0) 36.7 (1.0) 0.5 (0.2) (0.7) 36.9 (0.8) 0.6 (0.2) (0.7) 36.9 (0.8) 0.8 (0.2) After reperfusion of liver graft (0.9) 36.3 (0.9) 0.6 (0.1) (0.9) 36.2 (0.9) 0.3 (0.2) (0.9) 36.3 (0.9) 0.3 (0.2) (0.8) 36.5 (0.9) 0.2 (0.1) (0.9) 36.7 (0.9) 0.2 (0.1) (0.8) 36.8 (0.9) 0.2 (0.1) Fig. 2. Difference between continuous cardiac output and the mean of two measurements plotted against the average of these two values (n=170). A mean difference of L/min (SD 0.49 L/min) was determined and 95% of these differences between CCO and the average were expected to lie in the range of to 0.87 L/min (±1.96SD). comprising 170 data pairs from the other time points in 13 patients, revealed that the bias was L/min (95% CI: to L/min), and lower and upper limits of agreement (bias ±1.96SD) were L/min (95% CI: to L/min) and 1.73 L/min (95% CI: 1.48 to 1.99 L/min), respectively. The ±1.96SD values of the bias did not fall within the predetermined limits of agreement of ±1.14 L/min [13% BCO mean (8.75 L/min)]. The differences between CCO and BCO are plotted against the mean of the two methods (Fig. 1). The repeatability coefficient of CCO was 0.36 L/min and that of BCO was 0.86 L/min. The mean difference between CCO and the average of the two methods was L/min (SD 0.49 L/min) and the differences of these data pairs are plotted against the assumed true value (Fig. 2). The relative error of CCO measurement was 4.6±4.7%. Pulmonary artery temperature and rectal temperature were maintained well in all patients after induction of anesthesia. Pulmonary artery temperature, however, decreased while rectal temperature remained constant after caval clamping. The temperature difference increased to a maximum at 45 minutes after clamping. After reperfusion of the graft, both pulmonary artery temperature and rectal temperature increased gradually to basal levels (Table 2). Discussion The standard method of statistical evaluation of [21, 22] comparison studies is that of Bland and Altman. The primary aim of a comparison study is to determine whether two methods agree sufficiently to be used interchangeably. In a review, Mantha et al [18] pointed out several inadequacies and inconsistencies in the statistical results of comparison studies with regard to interchangeability of measurement methods, although 95% of the studies used Bland and Altman methodology for analysis. In our study, we compared the "Opti-Q" CCO with the conventional intermittent Hepatobiliary Pancreat Dis Int,Vol 7,No 2 April 15,
5 Hepatobiliary & Pancreatic Diseases International bolus thermodilution technique by strictly applying Bland and Altman analysis, [21, 22] and found that the two methods were not interchangeable. Ideally, one should define satisfactory agreement in advance in the comparison of two methods of measurement because how far apart measurements can be performed without leading to problems is a matter of clinical judgement, and statistical methods cannot make such a judgement. [22] In addition, a literature review revealed that acceptable, clearly defined criteria as to whether a newer technique can replace an older and more established one are lacking. [18] In the present study, we used the approach recommended by LaMantia et al, [23] who suggested using the results of Stetz et al, [24] i.e., 13% based on triplicate measurements as the criterion for comparing a newer method with thermodilution. In our study, the mean value of BCO from 170 data pairs was 8.75 L/min. Hence, ±1.96SD of the bias did not fall within the predetermined limits of agreement of ±1.14 L/min, indicating a lack of agreement between CCO and BCO. Factors associated with the lack of agreement between CCO and BCO included the repeatability of each method, the errors of thermodilution cardiac output measurement, rapid changes in cardiac output, and the delayed response of the CCO catheter. Under ideal circumstances, intermittent thermodilution [24, 25] has a 10%-15% error. Potential errors of 15%-50% can be caused by respiratory variations in pulmonary arterial blood temperature >0.05. [26] Therefore, a variation of ±10% within quadruplicate measurements is generally defined to be acceptable during intermittent thermodilution cardiac output measurements in order to reduce the influence of within-technique variability. In this study, the 95% limits of agreement were considerably wider than the repeatability. Therefore, the lack of agreement between the two methods cannot be explained by lack of repeatability, implying that other factors lowered the agreement. [22] Intermittent BCO is not the true value of cardiac output, although it is currently used as the standard method for its determination. Several factors including the volume and temperature of the injectate, the injection technique, and the timing of the injection within the respiratory cycle can influence the accuracy of intermittent bolus thermodilution. Because of methodological differences, the CCO method is not influenced by these factors, and may be more accurate than intermittent BCO under conditions with stable hemodynamics. Better accuracy with CCO than with intermittent BCO has been demonstrated in both in vitro [13] and in vivo [19] studies. The present study also showed better repeatability with CCO method with intermittent BCO. Several previous studies demonstrated the delayed response of CCO catheters to acute changes in cardiac output. [4-6] Significant delays in the response of CCO are induced by clinically relevant maneuvers, including rapid fluid infusion, hemorrhage, and drug therapy. [5] In the "normal" mode of the "Q-Vue" monitor, cardiac output values are averaged over 5 minutes. So, the monitor may ensure a response time of about 5 minutes under ideal conditions according to the averaging algorithms. However, response time may be significantly longer in the perioperative period. Aranda and colleagues [6] demonstrated in vitro that the response time for an 80% change in the "Opti-Q" CCO catheter is 5-11 minutes after controlled flow changes. Marked changes in cardiac output often occur after caval clamping and reperfusion during orthotopic liver transplantation, and they result in more thermal noise. Therefore, the monitor needs a longer analysis period, and the response time of CCO is delayed. In the present study, BCO and CCO cardiac output determinations could not be made simultaneously because of signal interference. Therefore, we could not calculate exactly the response time of the CCO catheter when cardiac output abruptly decreased after caval clamping and increased after reperfusion. However, the CCO values at 15 minutes after caval clamping as well as at 15 minutes after reperfusion were close to the BCO values, indicating that the response time of the CCO catheter was similar to that reported by Aranda et al. [6] In addition, this delay in response time of the CCO catheter is important for appropriate clinical use and must be considered when rapid changes in cardiac output occur. The rapid infusion of cold fluids may interfere with thermodilution cardiac output measurements by increasing thermal noise. Mihaljevic et al [13] found that the various rates of room-temperature fluid infusions (100 to 1000 ml/h) influences bolus measurements at flow rates of both 2 and 4 L/min, and affects the values from continuous measurement only at 2 L/min. Greim and colleagues [17] compared the two methods during liver transplantation and found that the bias increased with rising intravenous infusion rates. A small bias between the methods was achieved at low rates (<1000 ml/h) and during high cardiac output, whereas a large bias was most evident at high rates (>2000 ml/h) and low cardiac output. [17] Although our results of bias and 142 Hepatobiliary Pancreat Dis Int,Vol 7,No 2 April 15,2008
6 Continuous versus bolus cardiac output monitoring during OLT 95% limits of agreement were similar to those reported [3, 4, 8-12, 14-17, 19, 20] by other researchers, fluid infusion may still have contributed to the lack of agreement between the two methods found in this study. Bottiger et al [8] reported a bias (SD) of L/min (1.789 L/min) between the two methods during liver transplantation. The bias increased while the differences between pulmonary artery blood temperature and rectal temperature showed dynamic changes in the early phases after caval clamping and after reperfusion. [8] These findings are in agreement with our results. Both reflected hemodynamic and thermal instability during surgery and may also have contributed to the lack of agreement between the two methods. In this study, however, the difference between pulmonary artery blood temperature and rectal temperature reached a maximum at 45 minutes after caval clamping, indicating that the significant differences between the two methods, at 5 minutes after caval clamping and 5 minutes after reperfusion, were mainly induced by marked cardiac output changes and deficiency of thermodilution measurement. Another question in the current literature concerns what degree of accuracy is needed in CCO determinations for adequate clinical care [27] and correspondingly, if its accuracy is acceptable for clinical purposes. That the two methods are not interchangeable does not mean that continuous measurement is unacceptable for clinical purposes. They involve two different concepts because the bolus measurement does not give the true value of cardiac output. We did not know the true values, so the mean of two measurements was the best estimate we had. [22] Therefore, we compared continuous measurement with the average to decide its acceptability. Both our results and previous [13, 19] studies showed that continuous measurement had excellent repeatability in stable conditions. During the stable conditions in our study, the mean difference between continuous measurement and the average and its standard deviation were small (-0.09 L/min and 0.49 L/min, respectively). They were not clinically important, giving a mean cardiac output of 8.66 L/min. Combined with the small relative error of the continuous measurement, we considered the continuous measurement reliable and acceptable for clinical purposes. In conclusion, the delayed response of continuous thermodilution measurement limits its application during the early phases after caval clamping and after reperfusion of the graft. The two methods are not interchangeable even during hemodynamic stability in patients undergoing liver transplantation. Continuous thermodilution monitoring, however, is reliable and acceptable for clinical purposes. Funding: This work was supported by a grant from the Health Bureau of Zhejiang Province (No. 2006QN011). Ethical approval: Not needed. Contributors: WJ proposed the study and wrote the first draft. BFP analyzed the data. All authors contributed to the design and interpretation of the study and to further drafts. WJ is the guarantor. Competing interest: No benefits in any form have been received or will be received from a commercial party related directly or indirectly to the subject of this article. References 1 Salmenpera M, Aittomaki J. Cardiac output monitoring: need for improvement? Acta Anaesthesiol Scand 2003;47: Bottiger BW. Continuous cardiac output monitoring-- further applications of the thermodilution principle. Intensive Care Med 1999;25: Mihm FG, Gettinger A, Hanson CW 3rd, Gilbert HC, Stover EP, Vender JS, et al. A multicenter evaluation of a new continuous cardiac output pulmonary artery catheter system. Crit Care Med 1998;26: Haller M, Zollner C, Briegel J, Forst H. Evaluation of a new continuous thermodilution cardiac output monitor in critically ill patients: a prospective criterion standard study. Crit Care Med 1995;23: Siegel LC, Hennessy MM, Pearl RG. Delayed time response of the continuous cardiac output pulmonary artery catheter. Anesth Analg 1996;83: Aranda M, Mihm FG, Garrett S, Mihm MN, Pearl RG. Continuous cardiac output catheters: delay in in vitro response time after controlled flow changes. Anesthesiology 1998;89: Poli de Figueiredo LF, Malbouisson LM, Varicoda EY, Carmona MJ, Auler JO Jr, Rochae Silva M. Thermal filament continuous thermodilution cardiac output delayed response limits its value during acute hemodynamic instability. J Trauma 1999;47: Bottiger BW, Sinner B, Motsch J, Bach A, Bauer H, Martin E. Continuous versus intermittent thermodilution cardiac output measurement during orthotopic liver transplantation. Anaesthesia 1997;52: Lazor MA, Pierce ET, Stanley GD, Cass JL, Halpern EF, Bode RH Jr. Evaluation of the accuracy and response time of STAT-mode continuous cardiac output. J Cardiothorac Vasc Anesth 1997;11: Yelderman ML, Ramsay MA, Quinn MD, Paulsen AW, McKown RC, Gillman PH. Continuous thermodilution cardiac output measurement in intensive care unit patients. J Cardiothorac Vasc Anesth 1992;6: Boldt J, Menges T, Wollbrück M, Hammermann H, Hempelmann G. Is continuous cardiac output measurement using thermodilution reliable in the Hepatobiliary Pancreat Dis Int,Vol 7,No 2 April 15,
7 Hepatobiliary & Pancreatic Diseases International critically ill patient? Crit Care Med 1994;22: Jakobsen CJ, Melsen NC, Andresen EB. Continuous cardiac output measurements in the perioperative period. Acta Anaesthesiol Scand 1995;39: Mihaljevic T, von Segesser LK, Tonz M, Leskosek B, Seifert B, Jenni R, et al. Continuous versus bolus thermodilution cardiac output measurements--a comparative study. Crit Care Med 1995;23: Bottiger BW, Soder M, Rauch H, Bohrer H, Motsch J, Bauer H, et al. Semi-continuous versus injectate cardiac output measurement in intensive care patients after cardiac surgery. Intensive Care Med 1996;22: Zollner C, Polasek J, Kilger E, Pichler B, Jaenicke U, Briegel J, et al. Evaluation of a new continuous thermodilution cardiac output monitor in cardiac surgical patients: a prospective criterion standard study. Crit Care Med 1999; 27: Della Rocca G, Costa MG, Pompei L, Coccia C, Pietropaoli P. Continuous and intermittent cardiac output measurement: pulmonary artery catheter versus aortic transpulmonary technique. Br J Anaesth 2002;88: Greim CA, Roewer N, Thiel H, Laux G, Schulte am Esch J. Continuous cardiac output monitoring during adult liver transplantation: thermal filament technique versus bolus thermodilution. Anesth Analg 1997;85: Mantha S, Roizen MF, Fleisher LA, Thisted R, Foss J. Comparing methods of clinical measurement: reporting standards for bland and altman analysis. Anesth Analg 2000;90: Seguin P, Colcanap O, Le Rouzo A, Tanguy M, Guillou YM, Mallédant Y. Evaluation of a new semi-continuous cardiac output system in the intensive care unit. Can J Anaesth 1998;45: Zollner C, Goetz AE, Weis M, Morstedt K, Pichler B, Lamm P, Kilger E, Haller M. Continuous cardiac output measurements do not agree with conventional bolus thermodilution cardiac output determination. Can J Anaesth 2001;48: Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1: Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res 1999;8: LaMantia KR, O'Connor T, Barash PG. Comparing methods of measurement: an alternative approach. Anesthesiology 1990;72: Stetz CW, Miller RG, Kelly GE, Raffin TA. Reliability of the thermodilution method in the determination of cardiac output in clinical practice. Am Rev Respir Dis 1982;126: Moore FA, Haenel JB, Moore EE. Alternatives to Swan- Ganz cardiac output monitoring. Surg Clin North Am 1991;71: Latson TW, Whitten CW, O'Flaherty D. Ventilation, thermal noise, and errors in cardiac output measurements after cardiopulmonary bypass. Anesthesiology 1993;79: Gardner RM. Continuous cardiac output: how accurate and how timely? Crit Care Med 1998;26: Received November 5, 2007 Accepted after revision January 29, Hepatobiliary Pancreat Dis Int,Vol 7,No 2 April 15,2008
Continuous cardiac output measurement: pulse contour analysis vs thermodilution technique in cardiac surgical patients
British Journal of Anaesthesia 82 (4): 525 30 (1999) Continuous cardiac output measurement: pulse contour analysis vs thermodilution technique in cardiac surgical patients G. Rödig*, C. Prasser, C. Keyl,
More informationA pilot assessment of the FloTrac TM cardiac output monitoring system
Intensive Care Med (2007) 33:344 349 DOI 10.1007/s00134-006-0410-4 TECHNICAL NOTE Helen Ingrid Opdam Li Wan Rinaldo Bellomo A pilot assessment of the FloTrac TM cardiac output monitoring system Received:
More informationEdwards FloTrac Sensor & Performance Assessments of the FloTrac Sensor and Vigileo Monitor
Edwards FloTrac Sensor & Edwards Vigileo Monitor Performance Assessments of the FloTrac Sensor and Vigileo Monitor 1 Topics System Configuration Performance and Validation Dr. William T. McGee, Validation
More informationMeasurement of the cardiac output (CO) and
Original Article Annals of Cardiac Anaesthesia 2007; 10: 121 126 Chakravarthy et al. Cardiac Output Measurement by 4 Techniques 121 Comparison of Simultaneous Estimation of Cardiac Output by Four Techniques
More informationThe Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for
1 2 The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for continuous central venous oximetry (ScvO2) 3 The Vigileo
More informationTHE MEASUREMENT of cardiac output (CO) is a parameter
Comparison of Esophageal Doppler, Pulse Contour Analysis, and Real-Time Pulmonary Artery Thermodilution for the Continuous Measurement of Cardiac Output Berthold Bein, MD,* Frank Worthmann, MD,* Peter
More informationCARDIOVASCULAR Continuous cardiac output during off-pump coronary artery bypass surgery: pulse-contour analyses vs pulmonary artery thermodilution
British Journal of Anaesthesia 99 (4): 484 92 (2007) doi:10.1093/bja/aem199 Advance Access publication July 24, 2007 CARDIOVASCULAR Continuous cardiac output during off-pump coronary artery bypass surgery:
More informationFluid bolus of 20% Albumin in post-cardiac surgical patient: a prospective observational study of effect duration
Fluid bolus of 20% Albumin in post-cardiac surgical patient: a prospective observational study of effect duration Investigators: Salvatore Cutuli, Eduardo Osawa, Rinaldo Bellomo Affiliations: 1. Department
More informationHoiyin Cheung Quan Dong Rong Dong Buwei Yu. Keywords Measurement techniques Cardiac output Thermodilution Non-invasiveness Bioreactance
DOI 10.1007/s00540-014-1938-z ORIGINAL ARTICLE Correlation of cardiac output measured by non invasive continuous cardiac output monitoring (NICOM) and thermodilution in patients undergoing off pump coronary
More informationReceived: 7 Jun 2006 Revisions requested: 28 Jun 2006 Revisions received: 30 Aug 2006 Accepted: 21 Nov 2006 Published: 21 Nov 2006
Available online http://ccforum.com/content/10/6/r164 Vol 10 No 6 Research Open Access Comparison of uncalibrated arterial waveform analysis in cardiac surgery patients with thermodilution cardiac output
More informationButton et al. using a thermistor-tipped catheter. The CO is calculated using an algorithm based on the area under the systolic part of the pressure wa
British Journal of Anaesthesia 99 (3): 329 36 (2007) doi:10.1093/bja/aem188 Advance Access publication on July 12, 2007 CARDIOVASCULAR Clinical evaluation of the FloTrac/Vigileo TM system and two established
More informationCardiac output (CO) and invasive hemodynamic. Not for External Distribution
Technology, Computing, and Simulation Section Editor: Jeffrey M. Feldman Not for External Distribution Cardiac Output Measurement in Patients Undergoing Liver Transplantation: Pulmonary Artery Catheter
More informationThis PDF is available for free download from a site hosted by Medknow Publications
Indian J Crit Care Med Oct-Dec 006 Vol 10 Issue 4 IJCCM October-December 003 Vol 7 Issue 4 Research Article Correlation of mixed venous and central venous oxygen saturation and its relation to cardiac
More informationComparison of central venous oxygen saturation and mixed venous oxygen saturation during liver transplantation
Anaesthesia, 9,, pages 37 3 doi:1.1111/j.135-..5793.x Comparison of central venous oxygen saturation and mixed venous oxygen saturation during liver transplantation A. El Masry, 1 A. M. Mukhtar, 1 A. M.
More informationTechniques of Cardiac Output Measurement During Liver Transplantation: Arterial Pulse Wave Versus Thermodilution
LIVER TRANSPLANTATION 15:287-291, 2009 ORIGINAL ARTICLE Techniques of Cardiac Output Measurement During Liver Transplantation: Arterial Pulse Wave Versus Thermodilution Peter Nissen, 1 Johannes J. Van
More informationAssessing Agreement Between Methods Of Clinical Measurement
University of York Department of Health Sciences Measuring Health and Disease Assessing Agreement Between Methods Of Clinical Measurement Based on Bland JM, Altman DG. (1986). Statistical methods for assessing
More informationCapture every aspect of hemodynamic status
Capture every aspect of hemodynamic status Edwards Advanced Technology Swan-Ganz Catheters Hemodynamic Monitoring Systems with Continuous CO, SvO 2, EDV, RVEF, SVR, and SV for Optimal Patient Care The
More informationLess Invasive, Continuous Hemodynamic Monitoring During Minimally Invasive Coronary Surgery
Less Invasive, Continuous Hemodynamic Monitoring During Minimally Invasive Coronary Surgery Oliver Gödje, MD, Christian Thiel, MS, Peter Lamm, MD, Hermann Reichenspurner, MD, PhD, Christof Schmitz, MD,
More informationHemodynamic Monitoring
Perform Procedure And Interpret Results Hemodynamic Monitoring Tracheal Tube Cuff Pressure Dean R. Hess PhD RRT FAARC Hemodynamic Monitoring Cardiac Rate and Rhythm Arterial Blood Pressure Central Venous
More informationW. J. RUSSELL*, M. F. JAMES
Anaesth Intensive Care 2004; 32: 644-648 The Effects on Arterial Haemoglobin Oxygen Saturation and on Shunt of Increasing Cardiac Output with Dopamine or Dobutamine During One-lung Ventilation W. J. RUSSELL*,
More informationObjectifs : REcemment, un nouveau systeme de mesure semi-continue du debit cardiaque (CCO) a EtE commercialise
578 Equipment Evaluation of a new semi-continuous cardiac output system in the intensive care unit Philippe Seguin MD, Olivier Colcanap MD, Anne Le Rouzo MD, Mich61e Tanguy MD, Yves-Marie Guillou MD, Yannick
More informationThe Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the Edwards PreSep oximetry catheter
1 2 The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the Edwards PreSep oximetry catheter for continuous central venous oximetry (ScvO2) 3
More informationThe effects of vasoactive drugs on pulse pressure and stroke volume variation in postoperative ventilated patients,
Journal of Critical Care (211) 26, 328.e1 328.e8 The effects of vasoactive drugs on pulse pressure and stroke volume variation in postoperative ventilated patients, Mehrnaz Hadian MD a, Donald A. Severyn
More informationPerioperative monitoring of cardiac
Continuous cardiac output monitoring with pulse contour analysis: A comparison with lithium indicator dilution cardiac output measurement James Pittman, MBBS, FRCA; Shahar Bar-Yosef, MD; John SumPing,
More informationInteresting Cases - A Case Report: Renal Cell Carcinoma With Tumor Mass In IVC And Heart. O Wenker, L Chaloupka, R Joswiak, D Thakar, C Wood, G Walsh
ISPUB.COM The Internet Journal of Thoracic and Cardiovascular Surgery Volume 3 Number 2 Interesting Cases - A Case Report: Renal Cell Carcinoma With Tumor Mass In IVC And Heart O Wenker, L Chaloupka, R
More informationComparison of Bier's Block and Systemic Analgesia for Upper Extremity Procedures: A Randomized Clinical Trial
J Arch Mil Med. 1 August; (3): e1977. Published online 1 August 3. DOI: 1.81/jamm.1977 Research Article Comparison of Bier's Block and Systemic Analgesia for Upper Extremity Procedures: A Randomized Clinical
More informationThe validity of trans-esophageal Doppler ultrasonography as a measure of cardiac output in critically ill adults
Intensive Care Med (2004) 30:2060 2066 DOI 10.1007/s00134-004-2430-2 O R I G I N A L Paul M. Dark Mervyn Singer The validity of trans-esophageal Doppler ultrasonography as a measure of cardiac output in
More informationComparison between oxygen consumption calculated by Fick s principle using a continuous thermodilution technique and measured by indirect calorimetry
British Journal of Anaesthesia 1995; 75: 719 73 Comparison between oxygen consumption calculated by Fick s principle using a continuous thermodilution technique and measured by indirect calorimetry P.
More informationChapter 3. An evaluation of cardiac output by five arterial pulse contour techniques during cardiac surgery
Chapter 3 An evaluation of cardiac output by five arterial pulse contour techniques during cardiac surgery R.B.P. de Wilde 1, J.J. Schreuder 2, P.C.M. van den Berg 1 and J.R.C. Jansen 1 1 Department of
More informationIMPROVE PATIENT OUTCOMES AND SAFETY IN ADULT CARDIAC SURGERY.
Clinical Evidence Guide IMPROVE PATIENT OUTCOMES AND SAFETY IN ADULT CARDIAC SURGERY. With the INVOS cerebral/somatic oximeter An examination of controlled studies reveals that responding to cerebral desaturation
More informationMinimally invasive cardiac output monitors
Continuing Education in Anaesthesia, Critical Care & Pain Advance Access published October 3, 2011 Minimally invasive cardiac output monitors Kate E Drummond MBBS FANZCA Edward Murphy MBBS FANZCA Matrix
More informationTranspulmonary versus continuous thermodilution cardiac output after valvular and coronary artery surgery
doi:10.1510/icvts.2009.204545 Interactive CardioVascular and Thoracic Surgery 9 (2009) 4 8 www.icvts.org Work in progress report - Cardiac general Transpulmonary versus continuous thermodilution cardiac
More informationThe cornerstone of treating patients with hypotension,
Does the Central Venous Pressure Predict Fluid Responsiveness? An Updated Meta-Analysis and a Plea for Some Common Sense* Paul E. Marik, MD, FCCM 1 ; Rodrigo Cavallazzi, MD 2 Background: Aim: Data Sources:
More informationPropofol or etomidate: Does it genuinely matter for induction in cardiac surgical procedures?
Original Research Article DOI: 10.18231/2394-4994.2016.0012 Propofol or etomidate: Does it genuinely matter for induction in cardiac surgical procedures? Manjunath Ratnakara Kamath 1,*, Suchitha Kamath
More informationEvaluation of Stroke Volume Variation Obtained by the FloTrac /Vigileo System to Guide Preoperative Fluid Therapy in Patients Undergoing Brain Surgery
The Journal of International Medical Research 2012; 40: 1175 1181 [first published online as 40(3) 11] Evaluation of Stroke Volume Variation Obtained by the FloTrac /Vigileo System to Guide Preoperative
More informationORIGINAL ARTICLE. Abstract. Introduction. Materials and Methods
ORIGINAL ARTICLE Effects of Cardiac Hemodynamics on Agreement in the ph, HCO3- and Lactate Levels between Arterial and Venous Blood Samples in Patients with Known or Suspected Chronic Heart Failure Satoshi
More informationPulsioFlex Patient focused flexibility
PulsioFlex Patient focused flexibility Modular platform with intelligent visualisation for advanced patient Minimally invasive perioperative cardiac output trend with ProAQT Enables calibrated cardiac
More informationMinimally invasive cardiac output monitors
Kate E Drummond MBBS FANZCA Edward Murphy MBBS FANZCA Matrix reference 2A04 Key points Minimally invasive cardiac output monitors have varying degrees of invasiveness with some being totally non-invasive
More informationClinical evaluation of USCOM ultrasonic cardiac output monitor in cardiac surgical patients in intensive care unit
British Journal of Anaesthesia 94 (3): 287 91 (2005) doi:10.1093/bja/aei054 Advance Access publication January 14, 2005 Clinical evaluation of USCOM ultrasonic cardiac output monitor in cardiac surgical
More informationPermanent City Research Online URL:
Kyriacou, P. A., Pal, S. K., Langford, R. & Jones, DP (2006). Electro-optical techniques for the investigation of oesophageal photoplethysmographic signals and blood oxygen saturation in burns. Measurement
More informationMagnesium sulfate; Laryngoscopy; Sternotomy; Hemodynamic response; Coronary artery bypass graft
Acad J Surg, 1 10 Dec. 2015 10 Aug. 2016 4 Mar. 2016 This study was designed to evaluate the effectiveness of low and high dose magnesium sulfate in reducing pressure responses to laryngoscopy and sternotomy.
More informationSWISS SOCIETY OF NEONATOLOGY. Prolonged arterial hypotension due to propofol used for endotracheal intubation in a newborn infant
SWISS SOCIETY OF NEONATOLOGY Prolonged arterial hypotension due to propofol used for endotracheal intubation in a newborn infant July 2001 2 Wagner B, Intensive Care Unit, University Children s Hospital
More informationThe Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for
1 2 The Vigileo monitor by Edwards Lifesciences supports both the FloTrac Sensor for continuous cardiac output and the PreSep oximetry catheter for continuous central venous oximetry (ScvO2) 3 The Vigileo
More informationClinical Research Article. Jung-Yeon Yun 1, So-Hee Park 2, Dae-Soon Cho 1, Hae-Jeung Jeung 1, Soon-Ae Lee 1, and So Jin Seo 3.
Clinical Research Article Korean J Anesthesiol 2011 October 61(4): 281-287 http://dx.doi.org/10.4097/kjae.2011.61.4.281 Comparison of the central venous pressure from internal jugular vein and the pressure
More informationRevision of 10/27/2017 Form #280 Page 1 of 12 PVDOMICS STUDY Clinical Center Right Heart Catheterization (RHC) Results Form #280
Revision of 10/27/2017 Form #280 Page 1 of 12 PVDOMICS STUDY Clinical Center Right Heart Catheterization (RHC) Results Form #280 Instructions: Review PVDOMICS MOP Chapter 100 prior to completing right
More informationHemodynamic Monitoring and Circulatory Assist Devices
Hemodynamic Monitoring and Circulatory Assist Devices Speaker: Jana Ogden Learning Unit 2: Hemodynamic Monitoring and Circulatory Assist Devices Hemodynamic monitoring refers to the measurement of pressure,
More informationResistive Heating during Off-Pump Coronary Bypass Surgery
(Acta Anaesth. Belg., 2007, 58, 27-31) Resistive Heating during Off-Pump Coronary Bypass Surgery S. ENGELEN, J.BERGHMANS, S.BORMS, M.SUY-VERBURG and D. HIMPE Summary : Background : Maintaining normothermia
More informationescco, estimated Continuous Cardiac Output device DESEBBE OLIVIER SAUVEGARDE CLINIC, LYON, FRANCE
escco, estimated Continuous Cardiac Output device DESEBBE OLIVIER SAUVEGARDE CLINIC, LYON, FRANCE No conflict of interest u Technology, based on Pulse Wave Transit Time u Limits of the technology u Precision
More informationOriginal Article. * Received for Publication: September 2, 2006 * Revision Received: December 27, 2006 * Revision Accepted: April 4, 2007
Original Article COMPARATIVE EVALUATION OF THE CEREBRAL STATE INDEX TM AND BISPECTRAL INDEX TM MONITORING DURING PROPOFOL -REMIFENTANIL ANESTHESIA FOR OPEN HEART SURGERY Shahrbanoo Shahbazi 1, Farid Zand
More informationMONITORIA EN CUIDADO INTENSIVO DEL PACIENTE CON ENFERMEDAD HEPATICA DIEGO FERNANDO BAUTISTA R. INTERNISTA INTENSIVISTA
MONITORIA EN CUIDADO INTENSIVO DEL PACIENTE CON ENFERMEDAD HEPATICA DIEGO FERNANDO BAUTISTA R. INTERNISTA INTENSIVISTA COMO MONITOREAMOS HOY COMO LO HARIAMOS EN EL FUTURO Frederic Michard, Michael R. Pinsky
More informationNon-Invasive PCO 2 Monitoring in Infants Hospitalized with Viral Bronchiolitis
Non-Invasive PCO 2 Monitoring in Infants Hospitalized with Viral Bronchiolitis Gal S, Riskin A, Chistyakov I, Shifman N, Srugo I, and Kugelman A Pediatric Department and Pediatric Pulmonary Unit Bnai Zion
More informationDepth of Central Venous Catheterization by Intracardiac ECG in Paediatric Patients
Original Article Elmer Press Depth of Central Venous Catheterization by Intracardiac ECG in Paediatric Patients Prerana N. Shah a, b, Jithesh Appukutty a, Deepa Kane a Abstract Background: Central venous
More informationThe effect of vasopressin on the hemodynamics in CABG patients
Yimin et al. Journal of Cardiothoracic Surgery 2013, 8:49 RESEARCH ARTICLE Open Access The effect of vasopressin on the hemodynamics in CABG patients Hu Yimin 1, Liu Xiaoyu 1, Hu Yuping 1, Li Weiyan 1
More informationInternational Journal of Medical and Health Sciences
International Journal of Medical and Health Sciences Journal Home Page: http://www.ijmhs.net ISSN:2277-4505 Original article A Comparative Study Of Etomidate And Midazolam Induction In Patients Undergoing
More informationDepartment of Anaesthesiology, Erasmus University Medical Center, Rotterdam, The Netherlands
Intravenous device feasible for controlled cooling and rewarming of individuals with abnormal body core temperature A. Struijs 1, F. De Ruiter 1, A. Weijerse 1, J. Klein 2, A.J.J.C. Bogers 1 1 Department
More informationSwan Ganz catheter: Does it still have a role? Daniel De Backer Department of Intensive Care Erasme University Hospital Brussels, Belgium
Swan Ganz catheter: Does it still have a role? Daniel De Backer Department of Intensive Care Erasme University Hospital Brussels, Belgium How can cardiac output be measured? Thermodilution Arterial waveform
More informationComparison of cardiac output derived from FloTrac TM /Vigileo TM and impedance cardiography during major abdominal surgery
Clinical Note Comparison of cardiac output derived from FloTrac TM /Vigileo TM and impedance cardiography during major abdominal surgery Journal of International Medical Research 41(4) 1342 1349! The Author(s)
More informationChapter 2. Performance of three minimally invasive cardiac output monitoring systems
Chapter 2 Performance of three minimally invasive cardiac output monitoring systems Rob de Wilde, Bart Geerts, Jisheng Cui, Paul van den Berg and Jos Jansen Anaesthesia 2009; 64: 762 769 (45 46) Ideally
More informationCan peripheral venous pressure be interchangeable with central venous pressure in patients undergoing cardiac surgery?
Intensive Care Med (2004) 30:627 632 DOI 10.1007/s00134-003-2052-0 ORIGINAL Roger Desjardins André Y. Denault Sylvain Bélisle Michel Carrier Denis Babin Sylvie Lévesque Raymond Martineau Can peripheral
More informationFloTrac Sensor and Edwards PreSep Central Venous Oximetry Catheter Case Presentations
Edwards FloTrac Sensor & Edwards Vigileo Monitor FloTrac Sensor and Edwards PreSep Central Venous Oximetry Catheter Case Presentations 1 Topics System Configuration FloTrac Sensor and PreSep Catheter Thoracotomy
More informationESTIMATION OF STROKE VOLUME BY NON-INVASIVE BLOOD PRESSURE MONITORING IN HUMAN CENTRIFUGE
84TH ASMA ANNUAL SCIENTIFIC MEETING CHICAGO - MAY 12-16, 213 ESTIMATION OF STROKE VOLUME BY NON-INVASIVE BLOOD PRESSURE MONITORING IN HUMAN CENTRIFUGE PROGRAM ID NUMBER 445 MANEN O, DUSSAULT C, SAUVET
More informationRESEARCH ARTICLE. Helmut Hager & Goutham Mandadi & Debra Pulley & J. Chris Eagon & Edward Mascha & Benjamin Nutter & Andrea Kurz
OBES SURG (2009) 19:717 724 DOI 10.1007/s11695-008-9607-7 RESEARCH ARTICLE A Comparison of Noninvasive Blood Pressure Measurement on the Wrist with Invasive Arterial Blood Pressure Monitoring in Patients
More informationBMC Anesthesiology. Open Access. Abstract
BMC Anesthesiology BioMed Central Research article Evaluation of a new arterial pressure-based cardiac output device requiring no external calibration Christopher Prasser 1, Sylvia Bele 2, Cornelius Keyl
More informationIndex. Note: Page numbers of article titles are in boldface type.
Index Note: Page numbers of article titles are in boldface type. A Ablation, radiofrequency, anesthetic considerations for, 479 489 Acute aortic syndrome, thoracic endovascular repair of, 457 462 aortic
More informationMonitoring of cardiac index (CI) after surgery for congenital heart
Cardiac index monitoring by pulse contour analysis and thermodilution after pediatric cardiac surgery U. Fakler, MD, a Ch. Pauli, MD, a G. Balling, MD, a H. P. Lorenz, MD, a A. Eicken, MD, a M. Hennig,
More informationHemodynamic Monitoring in Critically ill Patients in Arthur Simonnet, interne Tuteur : Pr. Raphaël Favory
Hemodynamic Monitoring in Critically ill Patients in 2017 Arthur Simonnet, interne Tuteur : Pr. Raphaël Favory Rationale for Hemodynamic Monitoring Identify the presence of hemodynamic instability Identify
More informationThe Hemodynamics of PH Interpreting the numbers
The Hemodynamics of PH Interpreting the numbers Todd M Bull MD Associate Professor of Medicine Division of Pulmonary Sciences and Critical Care Medicine Pulmonary Hypertension Center University of Colorado
More informationMin Hur, Eun-Hee Kim, In-Kyung Song, Ji-Hyun Lee, Hee-Soo Kim, and Jin Tae Kim INTRODUCTION. Clinical Research
Anesth Pain Med 2016; 11: 375-379 https://doi.org/10.17085/apm.2016.11.4.375 Clinical Research http://crossmark.crossref.org/dialog/?doi=10.17085/apm.2016.11.4.375&domain=pdf&date_stamp=2016-10-25 pissn
More informationContinuous monitoring of cardiac output: why and how
33rd International Symposium of Intensive Care and Emergency Medicine, Brussels 19.03.-22.03.2013 UNIVERSITÄTSKLINIKUM Schleswig-Holstein Continuous monitoring of cardiac output: why and how Berthold Bein,
More informationInfluence of Intrapleural Infusion of Marcaine on Post Thoracotomy Pain
ORIGINAL ARTICLE Tanaffos (2007) 6(1), 47-51 2007 NRITLD, National Research Institute of Tuberculosis and Lung Disease, Iran Influence of Intrapleural Infusion of Marcaine on Post Thoracotomy Pain Hamid
More informationA comparison of cardiac output derived from the arterial pressure wave against thermodilution in cardiac surgery patients ²
British Journal of Anaesthesia 87 (2): 212±22 (2001) A comparison of cardiac output derived from the arterial pressure wave against thermodilution in cardiac surgery patients ² J. R. C. Jansen 1 *, J.
More information福島県立医科大学学術成果リポジトリ. Title laparoscopic adrenalectomy in patie pheochromocytoma. Midori; Iida, Hiroshi; Murakawa, Ma
福島県立医科大学学術成果リポジトリ Examination of the usefulness of no Title variation monitoring for adjusting laparoscopic adrenalectomy in patie pheochromocytoma Isosu, Tsuyoshi; Obara, Shinju; Oha Author(s) Atsuyuki;
More informationPerioperative Fluid Management in ERPs
Perioperative Fluid Management in ERPs Robert H. Thiele, M.D. Assistant Professor University of Virginia First Do No Harm Intravenous fluids should be considered a pharmacotherapeutic agent Just like all
More informationStandard Operating Procedure (SOP) Management of intervention group patients SOP 001
` Standard Operating Procedure (SOP) Management of intervention group patients SOP 001 Authors: Mark Edwards & Rupert Pearse Authorisation: Rupert Pearse (Chief Investigator) Scope To provide guidance
More informationA Cardiac Surgery ICU Discharge Model; for Research and Logistic Purpose
12 The Open Cardiovascular and Thoracic Surgery Journal, 2009, 2, 12-17 Open Access A Cardiac Surgery ICU Discharge Model; for Research and Logistic Purpose Carl-Johan Jakobsen *, Alice Lundbøl Vestergaard,
More informationComparison of Calibrated and Uncalibrated Arterial Pressure Based Cardiac Output Monitors During Orthotopic Liver Transplantation
LIVER TRANSPLANTATION 16:773-782, 2010 ORIGINAL ARTICLE Comparison of Calibrated and Uncalibrated Arterial Pressure Based Cardiac Output Monitors During Orthotopic Liver Transplantation Vladimir Krejci,
More informationThe Comparison of Stroke Volume Variation and Arterial Pressure Based Cardiac Output with Standard Hemodynamic Measurements during Cardiac Surgery
ISPUB.COM The Internet Journal of Anesthesiology Volume 22 Number 2 The Comparison of Stroke Volume Variation and Arterial Pressure Based Cardiac Output with Standard H Liu, M Konia, Z Li, N Fleming Citation
More informationActualités sur le remplissage peropératoire. Philippe Van der Linden MD, PhD
Actualités sur le remplissage peropératoire Philippe Van der Linden MD, PhD Fees for lectures, advisory board and consultancy: Fresenius Kabi GmbH B Braun Medical SA Perioperative Fluid Volume Administration
More informationDESIGNER RESUSCITATION: TITRATING TO TISSUE NEEDS
DESIGNER RESUSCITATION: TITRATING TO TISSUE NEEDS R. Phillip Dellinger MD, MSc, MCCM Professor and Chair of Medicine Cooper Medical School of Rowan University Chief of Medicine Cooper University Hospital
More informationAnimals. Male C57Bl/6 mice (n=27) were obtained from Charles River (Sulzfeld, Germany) and
Supplemental Methods Animals. Male C57Bl/6 mice (n=27) were obtained from Charles River (Sulzfeld, Germany) and housed in groups of 5-10 in individually ventilated BCU cages within a temperature controlled
More informationRadboud University Nijmegen Medical Centre Why measure cardiac output in critically ill children?
Radboud University Nijmegen Medical Centre Why measure cardiac output in critically ill children? J. Lemson Anesthesiologist/(pediatric)intensivist Case; Girl 2 years, 12 kg, severe meningococcal septic
More information( 12 17mLO 2 /dl) 1.39 Hb S v O P v O2
32 1970 Harold James Swan William Ganz N Engl J Med 1) 40 (mixed venous oxygen saturation S v O2 ) (central venous oxygen saturation Scv O2 ) (Hb) S v O2 Scv O2 1971 Ganz 2) 20 Forrester Swan Forrester
More informationPCV and PAOP Old habits die hard!
PCV and PAOP Old habits die hard! F Javier Belda MD, PhD Head of Department Associate Professor Anaesthesia and Critical Care Hospital Clínico Universitario Valencia (SPAIN) An old example TOBACO SMOKING
More informationPREOPERATIVE CARDIOPULMONARY ASSESSMENT FOR LIVER TRANSPLANTATION James Y. Findlay Mayo Clinic College of Medicine, Rochester, MN, USA.
PREOPERATIVE CARDIOPULMONARY ASSESSMENT FOR LIVER TRANSPLANTATION James Y. Findlay Mayo Clinic College of Medicine, Rochester, MN, USA Introduction Liver transplantation (LT) has gone from being a high-risk
More informationDynamic left ventricular outflow tract obstruction in living donor liver transplantation recipients -A report of two cases-
Case Report Korean J Anesthesiol 2010 December 9(Suppl): S128-S132 DOI: 10.4097/kjae.2010.9.S.S128 Dynamic left ventricular outflow tract obstruction in living donor liver transplantation recipients -A
More informationComparison of dexmedetomidine and midazolam in prevention of myoclonus occurring due to etomidate in coronary bypass surgery
Research Article 52 myoclonus occurring due to etomidate in coronary bypass surgery Nevriye Salman 1, Hasan Alper Gurbuz 2, Huseyin Oguzalp 3, Halil Ibrahim Ucar 2, Cem Yorgancioglu 2, Sumru Sekerci 1
More informationPerioperative Goal- Protocol Summary
Perioperative Goal- Directed Therapy Protocol Summary Evidence-based, perioperative Goal-Directed Therapy (GDT) protocols. Several single centre randomized controlled trials, meta-analysis and quality
More informationOriginal Article Influence of butorphanol on the postoperative remifentanil hyperalgesia
Int J Clin Exp Med 2016;9(2):4685-4689 www.ijcem.com /ISSN:1940-5901/IJCEM0017991 Original Article Influence of butorphanol on the postoperative remifentanil hyperalgesia Chen Lv, Hui Zheng, Mingyang Wei,
More informationCOMPARISON OF SUFENTANIL-OXYGEN AND FENTANYL-OXYGEN ANAESTHESIA FOR CORONARY ARTERY BYPASS GRAFTING
Br. J. Anaesth. (1988), 60, 530-535 COMPARISON OF SUFENTANIL-OXYGEN AND FENTANYL-OXYGEN ANAESTHESIA FOR CORONARY ARTERY BYPASS GRAFTING H. M. L. MATHEWS, G. FURNESS, I. W. CARSON, I. A. ORR, S. M. LYONS
More informationRegional Anaesthesia for Caesarean Section
Regional Anaesthesia for Caesarean Section "The Best Recipe" Warwick D. Ngan Kee Dept of Anaesthesia & Intensive Care The Chinese University of Hong Kong What I will not do. Magic recipes One shoe to fit
More informationPilot Of Spontaneous Breathing Vs. Ventilated Model For Hemorrhage And Resuscitation In The Rabbit
Article ID: WMC001137 Pilot Of Spontaneous Breathing Vs. Ventilated Model For Hemorrhage And Resuscitation In The Rabbit Author(s):Dr. Jonathan S. Jahr, MD, Dr. Robert A. Gunther, PhD, Dr. Bernd Driessen,
More informationThe Utilization Of Venovenous Bypass And Transfusion In Orthotopic Liver Transplantation
ISPUB.COM The Internet Journal of Surgery Volume 15 Number 1 The Utilization Of Venovenous Bypass And Transfusion In Orthotopic Liver Transplantation O Bamgbade, S Pelletier, A Tait, O Nafiu, P Dorje Citation
More informationCardiac Output by Thermodilution and Arterial Pulse Contour Techniques
Cardiac Output by Thermodilution and Arterial Pulse Contour Techniques J. R. C. Jansen and P. C. M. van den Berg Introduction A reliable method of cardiac output monitoring is particularly desirable in
More information61% SvO 2. Clarity. Simplified.
61% SvO 2 Clarity Simplified. HemoSphere Advanced Monitoring Platform See advanced hemodynamic parameters with a new level* of clarity. The HemoSphere advanced monitoring platform from Edwards Lifesciences
More informationDe-Jie Wang 1, I-Shan Lee 1, An-Hsun Chou 1,2, Chun-Yu Chen 1,2, Pei-Chi Ting 1, Yun-Hui Teng 1, Jr-Rung Lin 1,2,3 and Hsin-I Tsai 1,2*
Wang et al. BMC Anesthesiology (2018) 18:138 https://doi.org/10.1186/s12871-018-0600-y RESEARCH ARTICLE Open Access Non-invasive cardiac output measurement with electrical velocimetry in patients undergoing
More informationPressure recording analytical method (PRAM) for measurement of cardiac output during various haemodynamic states
British Journal of Anaesthesia 95 (): 59 65 (5) doi:.93/bja/aei54 Advance Access publication May 3, 5 Pressure recording analytical method (PRAM) for measurement of cardiac output during various haemodynamic
More informationConflicts of Interest
Anesthesia for Major Abdominal Cancer Resection John E. Ellis MD Adjunct Professor University of Pennsylvania johnellis1700@gmail.com Conflicts of Interest 1 Upper Abdominal Surgery Focus on oncologic
More informationPerioperative Goal- Protocol Summary
Perioperative Goal- Directed Therapy Protocol Summary Evidence-based, perioperative Goal-Directed Therapy (GDT) protocols. Several single centre randomized controlled trials, meta-analysis and quality
More informationAjay Yadlapati, MD;* Tristan Grogan, MS; David Elashoff, PhD; Robert B. Kelly, MD
The Journal of ExtraCorporeal Technology Correlation of a Novel Noninvasive Tissue Oxygen Saturation Monitor to Serum Central Venous Oxygen Saturation in Pediatric Patients with Postoperative Congenital
More informationCERVICAL PLEXUS BLOCK FOR CAROTID ENDARTERECTOMY FOLLOWED BY GENERAL ANESTHESIA FOR ABDOMINAL AORTIC SURGERY
CERVICAL PLEXUS BLOCK FOR CAROTID ENDARTERECTOMY FOLLOWED BY GENERAL ANESTHESIA FOR ABDOMINAL AORTIC SURGERY - A Case Report - ALEXANDRE YAZIGI *, FADIA HADDAD *, SAMIA MADI-JEBARA *, GEMMA HAYECK * AND
More information