Assessment of aortic distensibility with ECG-gated 64-row multi-detector computed tomography in abdominal aortic aneurysms
|
|
- Andrew Miles
- 5 years ago
- Views:
Transcription
1 Assessment of aortic distensibility with ECG-gated 64-row multi-detector computed tomography in abdominal aortic aneurysms Poster No.: C-3106 Congress: ECR 2010 Type: Scientific Exhibit Topic: Vascular - CT Authors: H. Yokoyama, S. Sanada ; Sapporo/JP, Kanazawa/JP Keywords: Multi-detector CT, Imaging,three-Dimensional, Abdominal Aortic Aneurysm Rupture Keywords: Vascular, CT DOI: /ecr2010/C Any information contained in this pdf file is automatically generated from digital material submitted to EPOS by third parties in the form of scientific presentations. References to any names, marks, products, or services of third parties or hypertext links to thirdparty sites or information are provided solely as a convenience to you and do not in any way constitute or imply ECR's endorsement, sponsorship or recommendation of the third party, information, product or service. ECR is not responsible for the content of these pages and does not make any representations regarding the content or accuracy of material in this file. As per copyright regulations, any unauthorised use of the material or parts thereof as well as commercial reproduction or multiple distribution by any traditional or electronically based reproduction/publication method ist strictly prohibited. You agree to defend, indemnify, and hold ECR harmless from and against any and all claims, damages, costs, and expenses, including attorneys' fees, arising from or related to your use of these pages. Please note: Links to movies, ppt slideshows and any other multimedia files are not available in the pdf version of presentations. Page 1 of 25
2 Purpose Most abdominal aortic aneurysms (AAAs) are found accidentally, sometimes a calcified aneurysm can be recognized on plain abdominal x-rays. Surgery is indicated when the diameter of aneurysm exceeds 5.5cm,and AAAs are frequently fatal vascular condition. The aim of this study is that to detect the aortic distensibility can predict the rupture of AAAs using ECG-gated 64-row MDCT? Page 2 of 25
3 Methods and Materials Measurements of the aortic distensibility were performed in 80 (including 3 ruptured cases) patients with AAA, which all with thrombus. Equipment & Acquisition parameter Aquilion 64 SuperHeart MDCT(Toshiba medical Co.) Zio Station Workstation(ZIO Soft Co.) Dual shot GX Injector #Nemoto kyorindo Co.) Iomeron 350(Eisai pharmaceutical Co.) 100kV,350mA,0.35sec,1mm slice,fc 43,hp 5.6 Iomeprol 350mgI/ml 3.0ml/sec total 80ml +saline3.0m/se total40ml CTDIvol 47.0±2mGy DLP 1.99±0.8Gy.cm Radiation Dose First 32 cases were using 135 kilovoltage, rest 48 were using 100 kilovoltage. 135kV CTDIvol 129.6±3.3mGy DLP 5.37±0.1Gy.cm 100kV CTDIvol 47.0±2mGy DLP 1.99±0.8Gy.cm ECG-assisted MDCT is higher radiation dose than non ECG-assisted,but is acceptable since we only have vessel areas. Patients'clinical characteristics Subjects(n=80) Ø>5cm Ø<5cm Gender ratio male:female (52:28) 24:12 28:16 Mean age(years) 72±6 74±3 Systolic Blood Pressure(mmHg) 144±7 138±2 Diastolic Blood Pressure(mmHg) 79±6 74±3 Page 3 of 25
4 Mean of maximal aortic diameter (cm) 6.3± ±0.4 Aortic cross-sectional area of positions(fig.1) Above the renal artery Normal vessel size of artery Maximal diameter (aneurysm) Below the aneurysm Methods ( analysis of data aquisition fig.2-fig.4) 1.Reconstructed at 0%,10%,~90% of the RR interval 2.Calculated the vessel area using workstation 3.area-versus-time curve Elasticity quantification d = dsys # ddia d: dilation(mm) D = #A / Ao##p D:distensibility(Pa-1) #A:difference between maximum and minimum area Ao:minimum vessel area #p:change in blood pressure Page 4 of 25
5 Images for this section: Fig. 0 Page 5 of 25
6 Fig. 0 Page 6 of 25
7 Fig. 0 Page 7 of 25
8 Fig. 0 Page 8 of 25
9 Results Resulting area-versus-time curves were shown that the inner wall areas were larger than the outer wall areas, which mean the outer wall were stiffer. Distensibility of aneurysms exceed 5cm (36/80) were 0.2±0.5(10-4Pa-1), under 5cm (44/80) were 1.3±0.2 (10-4Pa-1).All in normal size were 3.2± 0.4 (10-4Pa-1). Three aneurysms of impending rupture indicated 0.19±0.1 (10-4Pa-1). The cine images also were seen the difference of movement clearly. (fig.1-fig.2) Changes of dilation d(mm) d(mm) Ø>5cm Ø<5cm Normal vessel size p. 2.1± ±1 Above the renal artery 1.8±1 1.7±0.9 At aneurysm 1.3± ±1 Below the aneurysm 1.4± ±0.8 Changes of cross-sectional areas sys/dia sys/dia Ø>5cm Ø<5cm Normal vessel size p 1.14± ±0.5 Above the renal artery 1.08± ±0.8 At aneurysm 1.07± ±0.5 Below the aneurysm 1.14± ±0.5 Distensibility D D[10-4 Pa-1] D[10-4 Pa-1] Ø>5cm Ø<5cm Normal vessel size p 3.2± ±0.6 Above the renal artery 4.1± ±0.5 At aneurysm 0.2± ±0.2 Below the aneurysm 2.1± ±0.6 Page 9 of 25
10 1mmHg=1.33Pa p<0.05 Case Report (fig.3) Clinical Summary: A 73-year-old man was referred to the Cardio-Vascular Center Hokkaido Ohno Hospital with abdominal pain.on the patient's admission,ct examination was performed.ct images revealed a 78.7x 64.8mm abdominal aneurysm in diameter. Patient condition was stable, therefore operation was scheduled electively. Two days after admission,patient complaint of backpain,then underwent CT immediately. Rupture of aneurysm was suspected and then emergent operation was carried out. Labo Data on admission onset(ruptured) 3 WBC(10 /µl) # 4 RBC(10 /µl) # Hb (g/dl) # Ht (%) # MCV (fl) MCH (pg) PLT (10 /µl) # Blood pressure 151/80 mmhg 163/88 mmhg Risk factor Hypertension Operative Impression: Impending rupture (Defect of Intima) (fig.4) Page 10 of 25
11 Discussion (fig.5-fig.6) We had experienced three cases which all with rupture of aneurysms, after underwent MDCT examinations. #1 case; time from routine study to onset 2days #2 case; time from routine study to onset 7days #3 case; time from routine study to onset 31days Distensibility D of ruptured cases were significantly lower in the aneurysms Distensibility #D) (10-4Pa-1) two days before onset (rupture) Normal vessel size p Above the renal artery At aneurysm Below the aneurysm Region of rupture 0.19 Factors influencing the rupture of AAAs Size of Aneurysm #5#5.5cm Aneurysm Formation saccular # fusiform Epidemic Factors gender,age,hypertention, smoking,copd Page 11 of 25
12 Images for this section: Fig. 0: VR(Volume Rendering Image) of AAA(abdominal aortic aneurysm Page 12 of 25
13 Fig. 4: 4D Image of AAA(abdominal aortic aneurysm). Page 13 of 25
14 Fig. 0: 4D images are seen the difference of movement clearly. Page 14 of 25
15 Fig. 2: 4D(Cine Mode) #1 Page 15 of 25
16 Fig. 3: 4D(Cine Mode)#2 Page 16 of 25
17 Fig. 1: 4D(Cine Mode) #3 Page 17 of 25
18 Fig. 0: Aneurysm VR(Volume Rendering image),two days after rupture was occured Page 18 of 25
19 Fig. 0 Page 19 of 25
20 Fig. 0 Brewster DC,et al:guidelines for the treatment of AAA J Vasc Surg,37; ,2003 Page 20 of 25
21 Fig. 0 Vardulaki KA,et al:growth rates and risk of rupture of AAABr J Surg,85; ,1998 Page 21 of 25
22 Conclusion The resulting area-versus-time curves were shown that the inner wall areas larger than the outer wall areas.the 4-D images were seen the difference of movement clearly. To detect distensibility changes that might predict an increased risk of rupture,regardless of AAA diameter. The authors acknowledge to all the scholars who are interested in Distensibility of the aorta,particularly to Dr.Ganten and her research papers. Page 22 of 25
23 Images for this section: Fig. 0: Wired frame method was useful to evaluate the aortic distensibility Page 23 of 25
24 References Ganten M-K.et al.motion characterization of aortic wall and intimal flap by ECG-gated CT in patients with chronic B-dissection.Eur J Radiol(2008) Ganten M-K,Krautter U,von Tengg-Koligk H,et al.quantification of aortic distensibility in abdominal aortic aneurysm using ECG-gated multi-detector computed tomography.eur Radiol 2008(Jan 12,Epub ahead of print) Ganten M,Boese JM,Leitermann D,et al.quantification aortic elasticity: development and experimental validation of a method using computed tomography.eur Radiol 2005;15(12) Ekaterinaris JA,Ioannou CV,Katsamouris AN.Flow dynamics in expantions characterizing abdominal aorta aneurysms.ann Vasc Surg2006;20(3):351-9 Page 24 of 25
25 Personal Information Hiroichi Yokoyama RT,MSc.,1) Shigeru Sanada ##.# 2) Dept.Diagnostic Imagings, Cardio-Vascular Center Hokkaido Ohno Hospital1) Kanazawa University Graduate School of Medical Science2) Nishino Nishiku Sapporo, Japan mail Page 25 of 25
Scientific Exhibit. Authors: D. Takenaka, Y. Ohno, Y. Onishi, K. Matsumoto, T.
The feasibility of biphasic contrast-media-injection-protocol for chest imaging on 320-slice volume MDCT: Direct comparison of biphasic and bolus contrast-media injection protocols on 320-slice volume
More information64-MDCT imaging of the pancreas: Scan protocol optimisation by different scan delay regimes
64-MDCT imaging of the pancreas: Scan protocol optimisation by different scan delay regimes Poster No.: C-051 Congress: ECR 2009 Type: Scientific Exhibit Topic: Abdominal and Gastrointestinal Authors:
More informationHow to plan a Zenith AAA stent-graft from a CTA: Basic measurements and concepts explained
How to plan a Zenith AAA stent-graft from a CTA: Basic measurements and concepts explained Poster No.: C-3077 Congress: ECR 2010 Type: Educational Exhibit Topic: Vascular Authors: D. V. Thomas; Northampton/UK
More informationScientific Exhibit Authors:
Clinical Audit on Optimization of Radiation Dose from MDCT: Effect on Diagnostic Reference Levels for Brain, Sinus, Cervical Spine, Chest, Abdomen-Pelvis, and Lumbar Spine Examinations and on Nationwide
More informationAFib is the most common cardiac arrhythmia and its prevalence and incidence increases with age (Fuster V. et al. Circulation 2006).
Feasibility, image quality and radiation dose of coronary CT angiography (CCTA) in patients with atrial fibrillation using a new generation 256 multi-detector CT (MDCT) Poster No.: C-2378 Congress: ECR
More informationLow-dose computed tomography (CT) protocol in the screening of patients with social exposure to asbestos
Low-dose computed tomography (CT) protocol in the screening of patients with social exposure to asbestos Poster No.: C-3032 Congress: ECR 2010 Type: Scientific Exhibit Topic: Radiographers Authors: P.
More informationTreatment options for endoleaks: stents, embolizations and conversions
Treatment options for endoleaks: stents, embolizations and conversions Poster No.: C-0861 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Exhibit G. Lombardi; napoli/it Arteries / Aorta, Abdomen,
More information3D ultrasound applied to abdominal aortic aneurysm: preliminary evaluation of diameter measurement accuracy
3D ultrasound applied to abdominal aortic aneurysm: preliminary evaluation of diameter measurement accuracy Poster No.: C-0493 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Paper A. LONG
More informationMonophasic versus biphasic contrast application in CT of patients with head and neck tumour
Monophasic versus biphasic contrast application in CT of patients with head and neck tumour Poster No.: C-3331 Congress: ECR 2010 Type: Topic: Authors: Keywords: DOI: Scientific Exhibit Head and Neck G.
More information3D cine PCA enables rapid and comprehensive hemodynamic assessment of the abdominal aorta
3D cine PCA enables rapid and comprehensive hemodynamic assessment of the abdominal aorta Poster No.: C-1138 Congress: ECR 2013 Type: Scientific Exhibit Authors: M. Takahashi 1, Y. Takehara 2, H. Isoda
More informationIdentification and numbering of lumbar vertebrae using various anatomical landmarks on MRI of lumbosacral spine
Identification and numbering of lumbar vertebrae using various anatomical landmarks on MRI of lumbosacral spine Poster No.: C-2125 Congress: ECR 2015 Type: Authors: Scientific Exhibit S. patil 1, A. M.
More informationComputed tomographic dacryocystography as compared with X-ray dacryocystography in patients with dacryostenosis
Computed tomographic dacryocystography as compared with X-ray dacryocystography in patients with dacryostenosis Poster No.: C-1887 Congress: ECR 2016 Type: Authors: Keywords: DOI: Educational Exhibit M.
More informationQuantitative imaging of hepatic cirrhosis on abdominal CT images
Quantitative imaging of hepatic cirrhosis on abdominal CT images Poster No.: C-0556 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit S. Kido, A. Nakamura, Y. Hirano; Ube/JP Cirrhosis,
More informationSingle cold nodule in Graves' disease: benign vs malignant
Single cold nodule in Graves' disease: benign vs malignant Poster No.: C-0073 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Paper L. I. Sonoda, M. Halim, K. Balan; Cambridge/UK Head and neck,
More informationContrast enhancement of the right ventricle during coronary CTA: is it necessary?
Contrast enhancement of the right ventricle during coronary CTA: is it necessary? Poster No.: C-1545 Congress: ECR 2014 Type: Authors: Keywords: DOI: Scientific Exhibit M. Kok, C. Mihl, B. Kietselaer,
More informationSlowly growing malignant nodules and rapidly growing benign nodules: Evaluation of the value of volume doubling time
Slowly growing malignant nodules and rapidly growing benign nodules: Evaluation of the value of volume doubling time Poster No.: C-208 Congress: ECR 2009 Type: Educational Exhibit Topic: Chest Authors:
More information128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients
128-slice dual-source CT coronary angiography using highpitch scan protocols in 102 patients Poster No.: C-0634 Congress: ECR 2010 Type: Scientific Exhibit Topic: Cardiac Authors: Y. H. Choe, J. W. Lee,
More informationCryoplasty versus conventional angioplasty in peripheral arterial disease: 3 year analysis of reintervention free survival by treatment received.
Cryoplasty versus conventional angioplasty in peripheral arterial disease: 3 year analysis of reintervention free survival by treatment received. Poster No.: C-0343 Congress: ECR 2011 Type: Scientific
More informationHigh density thrombi of pulmonary embolism on precontrast CT scan: Is it dangerous?
High density thrombi of pulmonary embolism on precontrast CT scan: Is it dangerous? Poster No.: C-1753 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit B. Y. Lee, H. R. KIM, J. I. Jung,
More informationScientific Exhibit Authors: V. Moustakas, E. Karallas, K. Koutsopoulos ; Rodos/GR, 2
Diagnosis of Acute Appendicitis: the role of Color Doppler Ultrasound as first-line imaging method and evaluation of the higher diagnostic performances of CT against its disadvantages. Poster No.: C-0708
More informationScientific Exhibit Authors: M. Sugiyama, Y. Takehara, T. Saito, N. Ooishi, M. Alley,
Abnormal flow dynamics within the ascending aorta of the patients with aortic valve stenosis. Assessments with phase resolved three dimensional phase contrast MR image (4DFlow). Poster No.: C-2504 Congress:
More informationAetiologies of normal CT main pulmonary arterial (PA) measurements in patients with right heart catheter (RHC) confirmed pulmonary hypertension (PH)
Aetiologies of normal CT main pulmonary arterial (PA) measurements in patients with right heart catheter (RHC) confirmed pulmonary hypertension (PH) Poster No.: C-0964 Congress: ECR 2010 Type: Scientific
More informationContrast agents, Abdomen, CT, Contrast agent-intravenous, Cancer /ecr2015/C-1760
Can Weight-Adapted IV Contrast Media Protocols Reduce Iodine Dose and Still Produce a Diagnostic Contrast Enhancement Level in Abdomino-Pelvic CT Scans? Poster No.: C-1760 Congress: ECR 2015 Type: Authors:
More informationA pictorial review of normal anatomical appearences of Pericardial recesses on multislice Computed Tomography.
A pictorial review of normal anatomical appearences of Pericardial recesses on multislice Computed Tomography. Poster No.: C-1787 Congress: ECR 2012 Type: Educational Exhibit Authors: N. Ahmed 1, G. Avery
More informationReliability of the pronator quadratus fat pad sign to predict the severity of distal radius fractures
Reliability of the pronator quadratus fat pad sign to predict the severity of distal radius fractures Poster No.: C-0669 Congress: ECR 2014 Type: Scientific Exhibit Authors: J. Tonak, I. Wobbe, R. L. Duschka,
More informationSingle ventricle on cardiac MRI
Single ventricle on cardiac MRI Poster No.: C-0414 Congress: ECR 2011 Type: Scientific Exhibit Authors: F. Secchi, V. G. Nardella, A. Giardino, G. Di Leo, F. 1 2 1 1 1 1 2 Sardanelli ; Milano/IT, Milan/IT
More informationThe Role of Radionuclide Lymphoscintigraphy in the Diagnosis of Lymphedema of the Extremities
The Role of Radionuclide Lymphoscintigraphy in the Diagnosis of Lymphedema of the Extremities Poster No.: C-1229 Congress: ECR 2014 Type: Authors: Keywords: DOI: Scientific Exhibit M. Osher 1, A. Pallas
More informationComparison of Cardiac MDCT with MRI and Echocardiography in the Assessement of Left Ventricular Function
Comparison of Cardiac MDCT with MRI and Echocardiography in the Assessement of Left Ventricular Function Poster No.: C-0969 Congress: ECR 2012 Type: Scientific Exhibit Authors: B. Kara, Y. Paksoy, C. Erol,
More informationBolus administration of esmolol allows for safe and effective heart rate control during coronary computed tomography angiography
Bolus administration of esmolol allows for safe and effective heart rate control during coronary computed tomography angiography Poster No.: C-1342 Congress: ECR 2013 Type: Scientific Exhibit Authors:
More informationDiffusion-weighted MRI (DWI) "claw sign" is useful in differentiation of infectious from degenerative Modic I signal changes of the spine
Diffusion-weighted MRI (DWI) "claw sign" is useful in differentiation of infectious from degenerative Modic I signal changes of the spine Poster No.: C-0894 Congress: ECR 2012 Type: Scientific Exhibit
More informationQuantifying Dual-energy computed tomography (DECT) in patients with renal calculi using a Toshiba Aquilion One Scanner.
Quantifying Dual-energy computed tomography (DECT) in patients with renal calculi using a Toshiba Aquilion One Scanner. Poster No.: C-2613 Congress: ECR 2015 Type: Authors: Keywords: DOI: Scientific Exhibit
More informationA Randomized Controlled Study to Compare Image Quality between Fenestrated and Non-Fenestrated Intravenous Catheters for Cardiac MDCT
A Randomized Controlled Study to Compare Image Quality between Fenestrated and Non-Fenestrated Intravenous Catheters for Cardiac MDCT Poster No.: C-0623 Congress: ECR 2017 Type: Authors: Keywords: DOI:
More informationInfluence of pulsed fluoroscopy and special radiation risk training on the radiation dose in pneumatic reduction of ileocoecal intussusceptions.
Influence of pulsed fluoroscopy and special radiation risk training on the radiation dose in pneumatic reduction of ileocoecal intussusceptions. Poster No.: C-0599 Congress: ECR 2013 Type: Authors: Keywords:
More informationAssessment of renal cell carcinoma by two PET tracer : dual-time-point C-11 methionine and F-18 fluorodeoxyglucose
Assessment of renal cell carcinoma by two PET tracer : dual-time-point C-11 methionine and F-18 fluorodeoxyglucose Poster No.: C-0805 Congress: ECR 2015 Type: Scientific Exhibit Authors: S. Ito, K. Kato,
More informationComputed tomography and Modified RECIST criteria for assessment of response in malignant pleural mesothelioma
Computed tomography and Modified RECIST criteria for assessment of response in malignant pleural mesothelioma Poster No.: C-0729 Congress: ECR 2013 Type: Scientific Exhibit Authors: A. Marin, I. Pozek,
More informationSeemingly isolated greater trochanter fractures do not exist
Seemingly isolated greater trochanter fractures do not exist Poster No.: B-0950 Congress: ECR 2012 Type: Scientific Paper Authors: D. Dunker, J. H. Göthlin, M. Geijer ; Gothenburg/SE, Lund/SE Keywords:
More informationCT evaluation of small bowel carcinoid tumors
CT evaluation of small bowel carcinoid tumors Poster No.: C-0060 Congress: ECR 2015 Type: Educational Exhibit Authors: N. V. V. P. Costa, L. Nascimento, T. Bilhim ; Estoril/PT, PT, 1 2 3 1 2 3 Lisbon/PT
More informationIdiopathic dilatation of the pulmonary artery : radiographic and MDCT features in 6 cases
Idiopathic dilatation of the pulmonary artery : radiographic and MDCT features in 6 cases Poster No.: P-0075 Congress: ESTI 2014 Type: Authors: Educational Poster J. J. Woo 1, K. Y. Lee 2, Y. Cho 1, J.
More informationUltra-low dose CT of the acute abdomen: Spectrum of imaging findings
Ultra-low dose CT of the acute abdomen: Spectrum of imaging findings Poster No.: C-1452 Congress: ECR 2010 Type: Educational Exhibit Topic: GI Tract Authors: P. A. Vlachou, C. Kloeters, S. Kandel, P. Hein,
More informationPurpose. Methods and Materials
Thin-section CT findings in peripheral lung cancer of 3 cm or smaller: are there any characteristic features for predicting tumor histology or do they depend only on tumor size? Poster No.: C-1893 Congress:
More informationCierny-Mader classification of chronic osteomyelitis: Preoperative evaluation with cross-sectional imaging
Cierny-Mader classification of chronic osteomyelitis: Preoperative evaluation with cross-sectional imaging Poster No.: C-590 Congress: ECR 2009 Type: Topic: Educational Exhibit Musculoskeletal Authors:
More informationComputed tomography for pulmonary embolism: scan assessment of a one-year cohort and estimated cancer risk associated with diagnostic irradiation.
Computed tomography for pulmonary embolism: scan assessment of a one-year cohort and estimated cancer risk associated with diagnostic irradiation. Award: Poster No.: E-0004 Certificate of Merit Congress:
More informationPostmortem Computed Tomography Finding of Lungs in Sudden Infant Death.
Postmortem Computed Tomography Finding of Lungs in Sudden Infant Death. Poster No.: C-1147 Congress: ECR 2013 Type: Educational Exhibit Authors: Y. Kawasumi, A. Usui, Y. Hosokai, M. Sato, A. Nakajima,
More informationShear Wave Elastography in diagnostics of supraspinatus tendon.
Shear Wave Elastography in diagnostics of supraspinatus tendon. Poster No.: C-2168 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit V. Saltykova; Moscow/RU Musculoskeletal joint, Musculoskeletal
More informationPurpose. Methods and Materials. Results
Prevalence and significance of hypoattenuating hepatic lesions deemed too small to characterise: How are we following up these lesions and what are the outcomes? Poster No.: C-014 Congress: ECR 2009 Type:
More informationAbdominal fat distribution (subcutaneous vs. visceral abdominal fat compartments): correlation with gender, age, BMI and waist circumference
Abdominal fat distribution (subcutaneous vs. visceral abdominal fat compartments): correlation with gender, age, BMI and waist circumference Poster No.: C-934 Congress: ECR 202 Type: Scientific Exhibit
More informationRadiological Investigation of Renal Colic in an Emergency Department of a Teaching Hospital
Radiological Investigation of Renal Colic in an Emergency Department of a Teaching Hospital Poster No.: C-0892 Congress: ECR 2014 Type: Authors: Keywords: DOI: Scientific Exhibit A. Koo; Leeds, West Yorkshire/UK
More informationMultislice CT versus cone beam CT: dosimetric and image quality comparision.
Multislice CT versus cone beam CT: dosimetric and image quality comparision. Poster No.: C-2589 Congress: ECR 2015 Type: Authors: Scientific Exhibit C. Dionisi 1, A. Crispo 2, E. Stefani 3, A. Lovaglio
More informationValsalva-manoeuvre or prone belly position for computed tomography (CT) scan when an orbita varix is suspected: a single-case study.
Valsalva-manoeuvre or prone belly position for computed tomography (CT) scan when an orbita varix is suspected: a single-case study. Poster No.: C-0512 Congress: ECR 2012 Type: Authors: Keywords: DOI:
More informationOur experience in the endovascular treatment of female varicocele
Our experience in the endovascular treatment of female varicocele Poster No.: C-0347 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit A. Sáez de Ocáriz García, M. M. Mendigana Ramos,
More informationComputational simulation of 4D blood flow dynamics of the thoraco-abdominal aorta: prediction of long-term changes in aneurysm morphology
Computational simulation of 4D blood flow dynamics of the thoraco-abdominal aorta: prediction of long-term changes in aneurysm morphology Poster No.: B-044 Congress: ECR 2011 Type: Scientific Paper Topic:
More informationRadiological features of Legionella Pneumophila Pneumonia
Radiological features of Legionella Pneumophila Pneumonia Poster No.: E-0048 Congress: ESTI 2012 Type: Scientific Exhibit Authors: M. Vinciguerra, L. Stefanetti, E. Teti, G. Argentieri, L. G. 1 1 1 1 1
More informationDiffuse high-attenuation within mediastinal lymph nodes on non-enhanced CT scan: Usefulness in the prediction of benignancy
Diffuse high-attenuation within mediastinal lymph nodes on non-enhanced CT scan: Usefulness in the prediction of benignancy Poster No.: C-1785 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific
More informationFDG-18 PET/CT - radiation dose and dose-reduction strategy
FDG-18 PET/CT - radiation dose and dose-reduction strategy Poster No.: C-1856 Congress: ECR 2014 Type: Authors: Keywords: DOI: Scientific Exhibit P. Nicholson, S. McSweeney, K. O'Regan; Cork/IE Radiation
More informationComparison of multidetector-row computed tomography findings of IgG4-related sclerosing cholangitis and cholangiocarcinoma
Comparison of multidetector-row computed tomography findings of IgG4-related sclerosing cholangitis and cholangiocarcinoma Poster No.: C-0245 Congress: ECR 2014 Type: Scientific Exhibit Authors: M. Yata,
More informationIntraluminal gas in non-perforated acute appendicitis: a CT sign of gangrenous appendicitis
Intraluminal gas in non-perforated acute appendicitis: a CT sign of gangrenous appendicitis Poster No.: C-978 Congress: ECR 202 Type: Scientific Exhibit Authors: D. Plata Ariza, E. MARTINEZ CHAMORRO, J.
More informationAudit of CT Pulmonary Angiogram in suspected pulmonary embolism patients
Audit of CT Pulmonary Angiogram in suspected pulmonary embolism patients Poster No.: C-2511 Congress: ECR 2012 Type: Scientific Exhibit Authors: N. D. Gupta, M. K. Heir, P. Bradding; Leicester/UK Keywords:
More informationCan aortic aneurysm growth rate be predicted in clinical practice using 18-fluorodeoxyglucose positron emission tomography (18-FDG PET)?
Can aortic aneurysm growth rate be predicted in clinical practice using 18-fluorodeoxyglucose positron emission tomography (18-FDG PET)? Poster No.: C-1743 Congress: ECR 2011 Type: Authors: Keywords: DOI:
More informationFeasibility of contrast agent volume reduction on 640-slice CT coronary angiography in patients with low heart rate
Feasibility of contrast agent volume reduction on 640-slice CT coronary angiography in patients with low heart rate Poster No.: B-0742 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Paper
More informationRadiation induced DNA double strand breaks undergoing PET/CT examinations.
Radiation induced DNA double strand breaks undergoing PET/CT examinations. Poster No.: C-1685 Congress: ECR 2011 Type: Scientific Exhibit Authors: M. S. May, M. Brand, T. Löwe, M. Sommer, C. Engert, U.
More informationPneumo-esophageal 64-MDCT technique for gastric cancer evaluation
Pneumo-esophageal 64-MDCT technique for gastric cancer evaluation Poster No.: C-1627 Congress: ECR 2010 Type: Scientific Exhibit Topic: GI Tract Authors: M. Ulla, E. Gentile, E. Levy, D. Cavadas, J. Ithurralde
More informationArtifact in Head CT Images Due to Air Bubbles in X-Ray Tube Oil
Artifact in Head CT Images Due to Air Bubbles in X-Ray Tube Oil Poster No.: C-0671 Congress: ECR 2016 Type: Educational Exhibit Authors: H. Patel 1, W. Liu 2, J. DeSanto 2, S. Meagher 2, M. Zagardo 2,
More informationExtravasation of Contrast Medium during CT Scanning Tracking and Reduction of Rate of Extravasation
Extravasation of Contrast Medium during CT Scanning Tracking and Reduction of Rate of Extravasation Poster No.: C-0072 Congress: ECR 2015 Type: Scientific Exhibit Authors: M. S. K. Teo, C. Ong, A. Ying,
More informationCognitive target MRI-TRUS fusion biopsies of MRI detected PIRADS 4 and 5 lesions
Cognitive target MRI-TRUS fusion biopsies of MRI detected PIRADS 4 and 5 lesions Poster No.: B-0704 Congress: ECR 2015 Type: Scientific Paper Authors: P. P. van Westerveld, J. Vriesema, J. H. W. van den
More informationDeath due to Hypothermia: Postmortem Forensic Computed Tomography.
Death due to Hypothermia: Postmortem Forensic Computed Tomography. Poster No.: C-1128 Congress: ECR 2012 Type: Educational Exhibit Authors: Y. Kawasumi, A. Usui, Y. Hosokai, M. Sato, H. Saito, T. Ishibashi,
More informationBasic low - field MR imaging of meniscal injuries in children.
Basic low - field MR imaging of meniscal injuries in children. Poster No.: C-2365 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Exhibit A. Yakimov, M. Nikonova, E. Prokhorova, D. Vybornov,
More informationBI-RADS 3, 4 and 5 lesions on US: Five categories and their diagnostic efficacy and pitfalls in interpretation
BI-RADS 3, 4 and 5 lesions on US: Five categories and their diagnostic efficacy and pitfalls in interpretation e-poster: C-118 Congress: ECR 2008 Type: Educational Exhibit Topic: Breast / Ultrasound Authors:
More informationAudit of CT Pulmonary Angiogram in suspected pulmonary embolism patients
Audit of CT Pulmonary Angiogram in suspected pulmonary embolism patients Poster No.: C-2511 Congress: ECR 2012 Type: Scientific Exhibit Authors: N. D. Gupta, M. K. Heir, P. Bradding; Leicester/UK Keywords:
More informationThe "whirl sign". Diagnostic accuracy for intestinal volvulus.
The "whirl sign". Diagnostic accuracy for intestinal volvulus. Poster No.: C-0670 Congress: ECR 2014 Type: Scientific Exhibit Authors: M. Pire, M. Marti, A. Borobia, A. Verón; Madrid/ES Keywords: Abdomen,
More informationPopliteal pterygium syndrome
Popliteal pterygium syndrome Poster No.: C-1816 Congress: ECR 2011 Type: Educational Exhibit Authors: L. B. S. Santos, J. L. D. O. Schiavon, O. O. Guimaraes Neto, 1 1 2 3 1 1 C. A. P. Braga, R. S. LEMOS,
More informationCT Fluoroscopy-Guided vs Multislice CT Biopsy ModeGuided Lung Biopies:a preliminary experience
CT Fluoroscopy-Guided vs Multislice CT Biopsy ModeGuided Lung Biopies:a preliminary experience Poster No.: C-0097 Congress: ECR 2016 Type: Scientific Exhibit Authors: A. Casarin, G. Rech, C. Cicero, A.
More informationLong bones manifestations of congenital syphilis
Long bones manifestations of congenital syphilis Poster No.: C-0139 Congress: ECR 2011 Type: Educational Exhibit Authors: T. F. de Souza 1, P. P. Collier 1, E. J. M. Bronzatto 1, G. L. P. Keywords: DOI:
More informationUser Guide for Dental and Maxillofacial Cone Beam Computed Tomography (CBCT)
User Guide for Dental and Maxillofacial Cone Beam Computed Tomography (CBCT) Poster No.: C-0756 Congress: ECR 2014 Type: Educational Exhibit Authors: J. Ukkonen, J. Asp; Helsinki/FI Keywords: Education
More informationCarotid intima media thickness as an usefull tool in predicting cerebrovaskular events
Carotid intima media thickness as an usefull tool in predicting cerebrovaskular events Poster No.: C-0005 Congress: ECR 2015 Type: Authors: Keywords: DOI: Scientific Exhibit A. Rahimic - Catic; Sarajevo/BA
More informationLusory Artery - random finding without further implications or important challenge which conditions treatment strategy.
Lusory Artery - random finding without further implications or important challenge which conditions treatment strategy. Poster No.: C-2118 Congress: ECR 2014 Type: Authors: Keywords: DOI: Scientific Exhibit
More informationECG Gated CT Aorta in Transcatheter Aortic Valve Implantation
ECG Gated CT Aorta in Transcatheter Aortic Valve Implantation Poster No.: C-2014 Congress: ECR 2014 Type: Educational Exhibit Authors: M. A. Ottesen; Oslo/NO Keywords: Cardiac, Arteries / Aorta, CT, CT-Angiography,
More informationPage 1 of 8. 4 Charleston, SC/US
Comparison of contrast-induced discomfort between Visipaque (Iodixanol) injection 320 mg I/mL and Isovue (Iopamidol) 370 mg I/mL in patients undergoing peripheral arteriography Poster No.: C-1976 Congress:
More informationAbdominal Vascular Emergencies in MDCT Imaging
Abdominal Vascular Emergencies in MDCT Imaging Poster No.: C-0913 Congress: ECR 2016 Type: Educational Exhibit Authors: K. SHIRODKAR, D. N. Dasappa, S. L. DEVARU, D. S. 1 2 2 2 2 2 1 Nandikoor, A. R. Patil,
More informationStandardizing mammographic breast compression: Pressure rather than force?
Standardizing mammographic breast compression: Pressure rather than force? Poster No.: C-2383 Congress: ECR 2012 Type: Authors: Scientific Paper J. E. de Groot 1, W. Branderhorst 1, M. Broeders 2, G. J.
More informationComparison of Image quality in temporal bone MRI at 3T using 2D selective RF excitation versus a routine SPACE sequence
Comparison of Image quality in temporal bone MRI at 3T using 2D selective RF excitation versus a routine SPACE sequence Poster No.: C-1065 Congress: ECR 2015 Type: Authors: Keywords: DOI: Scientific Exhibit
More informationCoarctation of aorta in an adult-a case report
Coarctation of aorta in an adult-a case report Poster No.: P-0057 Congress: ESTI 2014 Type: Educational Poster Authors: R. Challa, R. Ahmed; Bolton/UK Keywords: Imaging sequences, CT, Thorax, Congenital
More informationGastrointestinal Angiodysplasia: CT Findings
Gastrointestinal Angiodysplasia: CT Findings Poster No.: C-1792 Congress: ECR 2012 Type: Authors: Keywords: DOI: Educational Exhibit G. Anguita Martinez, A. Fernandez Alfonso, D. C. Olivares Morello, J.
More informationImaging characterization of renal clear cell carcinoma
Imaging characterization of renal clear cell carcinoma Poster No.: C-0327 Congress: ECR 2011 Type: Educational Exhibit Authors: S. Ballester 1, A. Gaser 2, M. Dotta 1, M. F. CAPPA 1, F. Hammar 1 ; 1 2
More informationEndovascular Repair Of Traumatic, Degenerative And Mycotic Aortic Aneurysms: A Single Center Experience
Endovascular Repair Of Traumatic, Degenerative And Mycotic Aortic Aneurysms: A Single Center Experience Poster No.: C-2349 Congress: ECR 2014 Type: Scientific Exhibit Authors: Y. M. H. Al Bulushi, R. ALSukaiti;
More informationChest CT with Ultra-High Resolution Collimator for Submillimeter Fat Plane Detection: A Phantom Study
Chest CT with Ultra-High Resolution Collimator for Submillimeter Fat Plane Detection: A Phantom Study Poster No.: C-1207 Congress: ECR 2013 Type: Scientific Exhibit Authors: Y. Shimomiya, M. Kondo, M.
More informationImaging Gorham's disease (vanishing bone)
Imaging Gorham's disease (vanishing bone) Poster No.: C-2201 Congress: ECR 2010 Type: Scientific Exhibit Topic: Musculoskeletal Authors: D. Vanel, P. Ruggieri, M. Alberghini; Bologna/IT Keywords: Gorham,
More informationInfluence of X-ray contrast media on blood viscoelasticity
Influence of X-ray contrast media on blood viscoelasticity Poster No.: B-630 Congress: ECR 2009 Type: Topic: Authors: Keywords: DOI: Scientific Paper Contrast Media V. Ribitsch, C. Well; Graz/AT Blood
More informationUsing diffusion-tensor imaging and tractography (DTT) to study biological characteristics of glyoma in brain stem for neurosurgical planning
Using diffusion-tensor imaging and tractography (DTT) to study biological characteristics of glyoma in brain stem for neurosurgical planning Poster No.: C-0016 Congress: ECR 2014 Type: Authors: Keywords:
More informationPI-RADS classification: prognostic value for prostate cancer grading
PI-RADS classification: prognostic value for prostate cancer grading Poster No.: C-1622 Congress: ECR 2014 Type: Scientific Exhibit Authors: I. Platzek, A. Borkowetz, T. Paulus, T. Brauer, M. Wirth, M.
More informationDEXA Scores and Bone Density Measured on Routine CT Scans
DEXA Scores and Bone Density Measured on Routine CT Scans Poster No.: R-0072 Congress: 2015 ASM Type: Scientific Exhibit Authors: C. Annabattula, P. Phadke; Sydney/AU Keywords: Screening, Nuclear medicine
More informationThe role of abdominal CT and MRI in detection of complications after transplantations of liver, kidney and pancreas.
The role of abdominal CT and MRI in detection of complications after transplantations of liver, kidney and pancreas. Poster No.: C-1319 Congress: ECR 2015 Type: Educational Exhibit Authors: R. Muslimov,
More informationUreteropelvic Junction Obstruction (UPJO) syndrome: imaging with Multidetector CT (MDCT) prior to minimally invasive treatment
Ureteropelvic Junction Obstruction (UPJO) syndrome: imaging with Multidetector CT (MDCT) prior to minimally invasive treatment Poster No.: C-1753 Congress: ECR 2011 Type: Scientific Exhibit Authors: E.
More information"Ultrasound measurements of the lateral ventricles in neonates: A comparison of multiple measurements methods."
"Ultrasound measurements of the lateral ventricles in neonates: A comparison of multiple measurements methods." Poster No.: C-1557 Congress: ECR 2014 Type: Authors: Keywords: DOI: Educational Exhibit I.
More informationHyperechoic breast lesions can be malignant.
Hyperechoic breast lesions can be malignant. Poster No.: C-0041 Congress: ECR 2015 Type: Educational Exhibit Authors: G. Babu, R. bradley; Edinburgh/UK Keywords: Breast, Ultrasound, Biopsy, Cancer DOI:
More informationValue of shear wave ultrasound elastography in the myofascial pain syndrome diagnosis in sportsmen
Value of shear wave ultrasound elastography in the myofascial pain syndrome diagnosis in sportsmen Poster No.: C-1202 Congress: ECR 2014 Type: Authors: Keywords: DOI: Scientific Exhibit V. Saltykova; Moscow/RU
More informationContrast-enhanced ultrasound (CEUS) in the evaluation and characterization of complex renal cysts
Contrast-enhanced ultrasound (CEUS) in the evaluation and characterization of complex renal cysts Poster No.: C-2812 Congress: ECR 2018 Type: Educational Exhibit Authors: J. A. Torres de Abreu Macedo,
More informationIntegrating CT in minimally invasive treatment of the coronary arteries
Integrating CT in minimally invasive treatment of the coronary arteries e-poster: C-200 Congress: ECR 2008 Type: Scientific Exhibit Topic: Cardiac / Miscellaneous MeSH: Keywords: Coronary Vessels [A07.231.114.269]
More informationUse of IV-contrast versus IV-and oral-contrast in the evaluation of abdominal pain on CT in the emergency department
Use of IV-contrast versus IV-and oral-contrast in the evaluation of abdominal pain on CT in the emergency department Poster No.: B-0693 Congress: ECR 2016 Type: Authors: Scientific Paper M. Wasserman 1,
More informationCavitary lung lesion: Two different diagnosis with similar appearence
Cavitary lung lesion: Two different diagnosis with similar appearence Poster No.: P-0043 Congress: ESTI 2015 Type: Educational Poster Authors: M. Yesildag, H. Kalkan, K. Ödev; Konya/TR Keywords: Infection,
More information