Comparison of Exercise Blood Pressure Measured by Technician and an Automated System

Size: px
Start display at page:

Download "Comparison of Exercise Blood Pressure Measured by Technician and an Automated System"

Transcription

1 Clin. Cardiol. 7, (1984) Clinical Cardiology Publishing Co., nc. Practitioner's Corner Comparison of Exercise Blood Pressure Measured by Technician and an Automated System J. A. GARCA-GREGORY, M.D., A. S. JACKSON, P.E.D., J. SrUDEVLLE, M.D., W. G. SQURES, Ph.D., C. A. OWEN, M.D. Section of Cardiology, Kelsey-Seybold Clinic; Department of Health, Physical Education and Recreation, University of Houston; The Cardiopulmonary Laboratory, NASA/Johnson Space Center, Kelsey-Seybold Clinic, Houston, Texas, USA Summary: We evaluated the automated system Blood Pressure Measuring System (BPMS) developed by NASA on 277 adult males who elected to have a treadmill test as part of their annual physical. The BPMS uses acoustic transduction with a computer-assisted ECG gating to detect nonsynchronous noise. The BPMS readings were compared to pressures simultaneously measured by trained technicians. For all stages of work, BPMS readings were higher for systolic and lower for diastolic than technician readings. At peak stages of work, BPMS systolic pressures were about 20 mmhg higher than technician readings. Within each 3-min workstage, BPMS readings were found to be more inconsistent than technician readings. The standard errors of measurement for BPMS were from two to three times higher than technician values. These data showed automated blood pressure readings were significantly different than technician values and subject to more random fluctuations. These findings demonstrate the need to view exercise blood pressure measured by automated systems with caution. Address for reprints: Dr. J. A. Garcia-Gregory Section of Cardiology Kelsey-Seybold Clinic 6624 Fannin Houston, Texas 77030, USA Received: February, 1984 Accepted: February 10, 1984 Key words: exercise blood pressure, automated blood pressure measurement, exercise hypertension ntroduction Hemodynamic changes induced by exercise are of diagnostic value to practicing physicians. Walking and/or jogging on a treadmill introduces both arm movement and noise which makes it more difficult to measure blood pressure. For diagnostic and safety reasons, interest in measuring exercise blood pressure accurately is high and several electronic systems have been developed. Glasser and Ramsey (1981) have provided data to suggest that an automated system which used electrocardiographic gating and microcomputer processing of nonsynchronous noise to isolate acoustically the Korotkoff sounds measures blood pressure accurately. The correlations between exercise blood pressure measured by technicians and the automated system were high, but standard deviations for systolic pressure differences varied 7-9 mmhg, which showed that the readings varied substantially. Therefore, this study was initiated to evaluate a similar automated system developed by the National Aeronautics and Space Administration (NASA). The study was designed to determine: (1) if there were systematic differences in exercise blood pressure measured by a trained technician and the NASA automated system, and (2) if the automated readings were less subject to random fluctuations than readings by a trained technician.

2 316 Clio. Cardio!' Vo!' 7, May 1984 Methods The Blood Pressure Measuring System (BPMS) was designed to measure an astronaut's blood pressure during rest and at exercise on a bicycle ergometer. The BPMS measured blood pressure by the noninvasive Korotkoff sound technique (Nolte, 1977). To measure blood pressure, the occlusive cuff was automatically filled to one of three pressures: 160,200, or 250 mmhg. The occlusive cuff pressure was decreased at a systematic rate of 7 mmhg/s. A microphone with the necessary electrical modules was used to monitor the appearance and disappearance of the Korotkoff sounds and register systolic and diastolic pressures. The system was designed to coordinate Korotkoff sounds with the electrocardiogram. At each QRS complex of the heart, the electrocardiograph supplied a switch closure used to generate an "operating window," which was the time interval during which the system would accept the sounds from the microphone. A feature of BPMS was the capacity to record pressure electronically and simultaneously on a mercury manometer which was part of the BPMS. The mercury manometer was attached by a t connector to the gas line used to increase and decrease cuff pressure. The BPMS was used to measure exercise blood pressure during graded exercise stress tests administered by the staff of the Cardiopulmonary Laboratory, Kelsey Seybold Clinic, National Aeronautics and Space Administration/Johnson Space Center, Houston, Texas. The subjects of this study were 277 males who elected to have a graded exercise stress test as part of their annual health examination. Data describing the subjects at rest and exercise are provided in Table. All tests were administered on a treadmill (Quinton Model 18-60) utiliz- ing the Bruce protocol. The stress test was terminated when the subject either reached voluntary exhaustion or was terminated early by the physician in charge. During the stress test, blood pressure was simultaneously measured by BPMS and a trained technician at: (1) rest, (2) each minute of exercise, and (3) each minute during an 8-min recovery period. The procedures followed by the technician were: (1) the BPMS inflated the cuff to the target pressure; (2) using a stethoscope, the technician listened for Korotkoff sounds; and (3) at the appearance and disappearance of the sounds, read the systolic and diastolic (phase V) pressures directly from the mercury manometer. The BPMS provided pressures which were automatically stored in the memory of an LS 11 microcomputer. The technician was "blind" to the blood pressure readings of BPMS. The data analysis sought to compare the mean differences of blood pressures measured simultaneously by the two methods, and intraclass analysis was used to evaluate the consistency of exercise blood pressure measured by each method within each stage of the Bruce protocol. The data were analyzed by two work intensities: first, percentage of age predicted heart rate and second, minutes of the test. Results Provided in Table are the descriptive statistics of the subjects studied. The subjects varied considerably in terms of age and physical working capacity. All blood pressure values in Table were technician-determined. The resting values were taken after lying quietly for five minutes. The exercise values were recorded during the last minute of work. TABLE Characteristics of patients studied Variable Mean Standard deviation Range Physicial characteristics Age (yr) 42.9 Height (cm) Weight (kg) 78.5 Resting characteristics Heart rate (beats/min) 63.5 Diastolic blood pressure " 74.8 Systolic blood pressure " Peak exercise characteristics Treadmill time (min) 9.6 Heart rate (beats/min) Diastolic blood pressure " 81.3 Systolic blood pressure " a Blood pressures are auscultatory with resting pressures taken in a supine position and exercise values recorded at the final minute of exercise.

3 J. A. Garcia-Gregory et al.,' Automated blood pressure measurement 317 The resting and recovery blood pressure readings were simultaneously measured by technician and B})MS. At rest; the mean systolic pressures were not statistically different, with a difference of only 1 mmhg. The mean resting diastolic pressure of BPMS was significantly (p<0.01) lower than the pressure obtained by technician, but the difference was small, only 3 mmhg At recovery, the mean systolic reading of BPMS was significantly (p <0.01) higher (6 mmhg) than the technician reading and the recovery diastolic means were nearly identical to the differences found at rest. Furnished in Fig. 1 is exercise blood pressure response as a function of treadmill time and percentage of agedetermined maximum heart rate (HRmax). These data are consistent with expected exercise blood pressure variation: diastolic pressures remained relatively stable and systolic pressures steadily rose with work intensity. For all stages of work, the systolic readings of BPMS were significantly (p<o.o) higher than technician values. As work intensity increased, the differences in systolic pressure became more pronounced. At peak workloads, the difference in systolic pressure was about 20 mmhg. For all stages of work the diastolic pressures of BPMS were significantly (p<o.o) lower than technician readings. The difference range was 5-10 mmhg. The intraclass analysis is furnished in Fig. 2. This analysis was designed to examine the degree to which blood pressure measured by each method could be expected to vary due to random fluctuations within a 3-min workstage. The analysis strategy was to use each of the three blood pressure readings within a workstage for the analysis. Provided are the standard errors of measurement for a blood pressure reading for each workstage. For all levels of work, the standard errors were lower for technician-determined values than for BPMS. At peak workloads, the standard errors for exercise systolic pressure of BPMS were over three times higher than the technician-determined values. The standard error is an estimate of the random variation, and these data showed that BPMS was subject to more random fluctuations than a trained technician. The final analysis conducted was to examine the correlations of exercise blood pressure measured by both methods for various stages of work, and correlations were computed for the middle minute of each stage of work. The coefficients for exercise blood pressure measured by BPMS and technician were for systolic pressures and for diastolic pressures. n general, the lowest correlations were found at the more intensive workloads when a patient was more likely to be jogging rather than walking. Discussion The results of this study showed that exercise blood pressure measured with BPMS varied systematically from readings obtained by trained technicians. As the workload increased, the differences became more pronounced. Unexpectedly, the technician-determined readings were more consistent than the automated readings. Glasser and Ramsey (1981) found that systolic pressures obtained with an automated system were higher than values obtained by trained technicians. The differences in exercise systolic readings ranged from 5 to 12 mmhg over the workloads studied. These systematic differences led Glasser and Ramsey to conclude that at a systolic pressure of 200 mmhg, the automated system could be expected to read 9-15 mmhg higher than simultaneously recorded manual readings. The average exercise diastolic differences between automated and manual measured blood pressure reported by Glasser and Ramsey were slight; the differences were less than 3 mmhg, which are smaller than the values we found. However, the differences were in the same direction; lower diastolic pressures were obtained with the automated system. t is suspected that the smaller diastolic differences found by Glasser and Ramsey can be traced to equipment. For both studies, phase V Korotkoff sounds were used by the technician. The BPMS used by NASA followed a set criterion for registering diastolic pressure, whereas the system evaluated by Glasser and Ramsey recorded phase V if the pressure was greater than 80 mmhg, but phase V if the pressure was less than 80 mmhg. A major problem with exercise diastolic pressure is that Korotkoff sounds can sometimes persist at very low' pressures. The BPMS tended to register some low values which would account for the lower diastolic pressures recorded by the NASA system. One group of investigators has reported that it is difficult to obtain accurate exercise diastolic readings (rving et al. 1977). Automated systems are believed to be more sensitive in determining the onset of Korotkoff sounds than the human ear (Glasser and Ramsey, 1981). This would be one explanation for the higher systolic readings recorded by automated systems. A second reason, however, is that if a noise artifact was present, it would tend to "trigger" the automated reading above the "true" systolic pressure. This possibility can be seen by examining a hardcopy tracing of blood pressure obtained by BPMS (Fig. 3). The top reading clearly delineates the systolic and diastolic pressures, whereas the Korotkoff sound patterns of the bottom reading are not clearly apparent. We suspect that some of the recorded sounds are exercise artifacts which would tend to produce higher systolic and lower diastolic pressures. With the rapid advances in microcomputers and electronic technology, an attractive option is to automate procedures in an effort to reduce human error. This is one suggested advantage of automated blood pressure systems. The intraclass analysis (Fig. 2) showed that the technician-determined readings were, surprisingly, more consistent than the automated readings. An important consideration for these results is that the technician was

4 318 Clio. Cardiol. Vol. 7, May 1984! ::: A C ,'l Treadmill Time (min) ! ::: 160!! t:-- ff,. A &- 60 1: < >85 B % Maximum Heart Rate FG. Exercise systolic and diastolic blood pressure trends for given levels of work. Plots are by (A) minutes of Bruce protocol and (B) percentage of HRmax. Both show systematic differences at each stage of work. (6-6, Technician-determined readings;.-., BPMS readings.)

5 1. A. Garcia-Gregol)' et al.: Automated blood pressure measurement 319 Systolic pressure Diastolic Pressure 15 5 o A Treadmill Time (min) Systolic Pressure Diastolic Pressure 15 5 o -L- < >84 < >84 B % Maximum Heart Rate FG. 2 Standard emus of measurement for exercise blood pressure measured by technician (0) and BPMS (.) for (A) Bruce protocol and (B) HRmax. The figures show that BPMS was subject to more random variance than technician-measured blood pressure.

6 320 Clin. Cardio!. Vol. 7, May 1984 r-r-r-, r "T' r " ' / ;!...l : --. i-',.. ; " ; - c-. -' -1-,-,--1-, Tt! i - :.,,-' t- J i.. i i L -t-:.,,,:'-- 1 '(-! + '1- -., L J. -L ' =!:-v-=f "o-..."=-t"...:..., -=t-"---.,;"=' """"'/""'"'"=r-"-=! T-R' i i " f/f)t i. i 'i... : ll' 'b,' P! i"--, '" r'\ '" '" '" j f! j._., t+ r. n. b - f-- T, j! '-,.. r-..4 L flo t. b t. ft, ft. "'- 1l ft,. '" "'...' - Fro. 3 Hardcopy of blood pressure tracings from BPMS. Top tracing shows the clear location of the Korotkoff sounds and the lower tracing isolates " noise" at the extremes of the blood pressure ramp. ft.. highly trained and experienced. The intraclass analysis used the three blood pressure readings within each 3-min workstage with the assumption that each patient was internally consistent over this constant workload. The standard error of measurement is a statistic that represents the degree to which blood pressure could be expected to vary owing to chance. These analyses showed that at each workstage, the technician was more consistent than BPMS. We suspect this may be partly due to extreme scores caused by arti facts recorded by the BPMS. The technician used in this study was consistent; a reading was not recorded until it was certain a Korotkoff sound was heard. f there was doubt, the pressure was immediately retaken by the technician. These data showed that the technician was more consistent than the BPMS, but the intraclass analysis does not show that the technician was more accurate. Steinfeld et ai. (1981) compared resting blood pressures measured by electronic and stethoscopic sphygmomanometers with intra-arterial pressures. The correlations between auscultatory and intra-arterial pressures were somewhat higher than the coefficients found between intra-arterial and electronic pressures. Bergman et al. (1972) reported that the correlations between the intra-arterial exerci se blood pressure and the readings obtained by BPMS and auscultatory methods were similar. Exercise blood pressure was measured while riding a bicycle ergometer, and these readings were less subject to the noise introduced by walking or jogging on a treadmill. The results of these studies do not support the conclusion that automated methods are more accurate than manual readings while walking or jogging on a treadmill. n our experience, a trained technician can be a safeguard for screening obviously erroneous readings which would be recorded by automated methods. Glasser and Ramsey (l981) reported correlations of 0.96 and 0.85, respectively, between concurrent manual and automated systolic and diastolic pressures. These correlations are higher than the coefficients we found, but the findings are not directly comparable. Glasser and Ramsey calcul ated a si ngle coefficient for systolic and diastolic pressure. These high correlations were computed on the patient's blood pressure measured at rest, several stages of work, and recovery. By using all stages of work, recovery, and rest, the correlations were inflated because more variability was introduced into the bivariate relationship. This was especially true for systolic pre sure which changes su bstantially with exercise. The coeffi cients reported for our data represent the relation between BPMS and manual determinations for each defined stage of work. The results of this study

7 J. A. Garcia-Gregory et al.: Automated blood pressure measurement 321 showed that the correlation between simultaneously measured BPMS and technician-detennined blood pressure was lower at a defined workload and that the correlations were somewhat lower at more intense workloads when the subject was jogging. There is a growing trend to define exercise hypertension for men at 230 mmhg systolic (Sannerstedt, 1981) and an exercise hypotensive response :s 140 mmhg. systolic (Bruce et al., 1977; rving et al., 1977), and exercise hypertensive responses have been shown (Jackson et al., 1983; Wilson and Meyer, 1981) to be predictive of the development of resting hypertension. These criteria were developed with nonnative data obtained by the auscultatory method. Using the technician readings, only 2.9% of the NASA employees were found to be hypertensive during the exercise test and none were hypotensive. The difference between systolic pressure measured by BPMS and technician at maximum work levels tended to be about 20 mmhg. f BPMS was used to measure exercise blood pressure, about 15% of the NASA executives would be judged to be hypertensive during exercise. The data presented in Fig. 1 shows the difference is present at the lower levels of work. f the concern of exercise blood pressure response was a hypotensive response because of left ventricular disfunction and a criterion of a systolic pressure of :s 140 mmhg was used, the automated system would identify few hypotensive patients. f automated systems are to be used, the data need to be compared with technician readings so that meaningful interpretations can be made. The results of this study showed that exercise blood pressures measured by an automated system were substantially different than pressures measured by a trained technician. The differences become more pronounced with exercise intensity, and the automated pressure readings were more prone to random fluctuations than readings by trained technicians. References Bergman SA, Blomquist G, Mitchell J, Hoffler GS: Skylab blood pressure measuring system: Correlative study with intm-arterial pressures. Unpublished study, Southwestern Medical School, Dallas (1972) Bruce RA, DeRouen RA, Peterson D, rving JB, Chinn N, Blake B, Haffey, V: Noninvasive predictors of sudden cardiac death in men with coronary heart diesase: Predictive value of maximal stress testing. Am J Cardiol 39, 833 (1977) Glasser SP, Ramsey MR, : An automated system for blood pressure detennination during exercise. Circulation 63, 348 (1981) rving JB, Bruce RA, DeRouen TA: Variations in and significance of systolic pressure during maximal exercise (treadmill) testing. Am J Cardiol 39, 841 (1977) Jackson AW, Squires WG, Grimes G, Beard EF: Prediction of future resting hypertension from exercise blood pressure. J Cardiac Rehab 3, 263 (1983) Nolte RW: Automated blood pressure measuring system (M092). n Biomedical Results from Skylab. RS Johnson and LF Dietlein (Eds.). NASA, Washington, D.C. (1977) Sannerstedt R: Exercise in the patient with arterial hypertension. Prac Cardiol 7, 89 (1981) Steinfeld L, Cole A, Nazarian H, Richard A: A comparison of electronic and stethoscopic sphygomomanometers. Circulation (Abstr.) 64, (No.4) V-213 (1981) Wilson NV, Meyer BM: Early prediction of hypertension using exercise blood pressure. Prevent Med 10, 62 (1981)

An Automated System for Blood Pressure Determination During Exercise

An Automated System for Blood Pressure Determination During Exercise An Automated System for Blood Pressure Determination During Exercise SEPHEN P. GLASSER, M.D., AND MAYNARD R. RAMSEY III, M.D., PH.D. with the technical assistance ofblanche Price SUMMARY Automation of

More information

CAMOSUN COLLEGE BIOLOGY 144 (2010) LABS

CAMOSUN COLLEGE BIOLOGY 144 (2010) LABS LAB 8: CARDIOVASCULAR PHYSIOLOGY PART 1. HEART SOUNDS AND PULSE DETERMINATIONS Introduction Two distinct sounds can be heard during each cardiac cycle. These sounds are commonly described as lub and dup

More information

Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard

Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard EOIN O'BRIENt, NEIL ATKINS, JAMES SHERIDAN, COLM KILLEEN, RICHARD COSTELLOE, SHANE LYDON and

More information

Taking and recording blood pressure and pulse. City Gate Training Centre all rights reserved

Taking and recording blood pressure and pulse. City Gate Training Centre all rights reserved Taking and recording blood pressure and pulse City Gate Training Centre all rights reserved Course Content: Definition of Blood Pressure Normal Blood Pressure Readings Systolic and Diastolic Blood Pressure

More information

In looking for other diagnostic variables we. might on exercise suffer a sufficient fall in cardiac

In looking for other diagnostic variables we. might on exercise suffer a sufficient fall in cardiac Br Heart j 1985; 53: 598-62 Increased diastolic blood pressure response to exercise testing when coronary artery disease is suspected An indication ofseverity FAWAZ AKHRAS, JAMES UPWARD, GRAHAM JACKSON

More information

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE. MEASURING BLOOD PRESSURE - MANUAL (equ04)

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE. MEASURING BLOOD PRESSURE - MANUAL (equ04) SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE TITLE: ISSUED FOR: MEASURING BLOOD PRESSURE - MANUAL Nursing DATE: REVIEWED: PAGES: 2/80 7/17 1 of 5 RESPONSIBILITY: RN, LPN, Patient Care Technician Multi-skilled

More information

2. Measure a subject's blood pressure and heart rate both at rest and during exercise.

2. Measure a subject's blood pressure and heart rate both at rest and during exercise. Lab Activity 11 The Cardiovascular System Student Learning Objectives After completing this lab, you should be able to: 1. Define, explain and correctly use the key terms. 2. Measure a subject's blood

More information

Protocol 4: Measuring Blood Pressure

Protocol 4: Measuring Blood Pressure Zool 430L Protocols Page 8 of 12 Protocol 4: Measuring Blood Pressure 4.1 Via Auscultation (listening through stethoscope) The estimate of blood pressure is obtained by cutting off blood flow with a known

More information

Characterization of the Oscillometric Method for Measuring Indirect Blood Pressure

Characterization of the Oscillometric Method for Measuring Indirect Blood Pressure Purdue University Purdue e-pubs Weldon School of Biomedical Engineering Faculty Publications Weldon School of Biomedical Engineering 1982 Characterization of the Oscillometric Method for Measuring Indirect

More information

Biomedical Instrumentation E. Blood Pressure

Biomedical Instrumentation E. Blood Pressure Biomedical Instrumentation E. Blood Pressure Dr Gari Clifford Adapted from slides by Prof. Lionel Tarassenko Blood pressure Blood is pumped around the body by the heart. It makes its way around the body

More information

External Oscillatory Blood Pressure - EOBPTM

External Oscillatory Blood Pressure - EOBPTM External Oscillatory Blood Pressure - EOBPTM Development of Novel Principle To Measure Blood Pressure Mindaugas Pranevicius, M.D., Osvaldas Pranevicius, M.D., Ph.D. Pranevicius Biotech Inc., Forest Hills,

More information

Data Collection Worksheet

Data Collection Worksheet Data Collection Worksheet 1. Has a doctor or nurse ever said that you have: High blood pressure or hypertension? [ ] No [ ] Yes [ ] Not Sure 1.a. If Yes, then at what age were you first told this? Age

More information

Blood Pressure Laboratory

Blood Pressure Laboratory Introduction The blood that circulates throughout the body maintains a flow and pressure. The nervous system can change the flow and pressure based on the particular needs at a given time. For example,

More information

BP-600 Noninvasive Blood Pressure Sensor

BP-600 Noninvasive Blood Pressure Sensor Technical Note BP-600 Overview A person's cardiac output, peripheral vascular resistance, blood pressure, and other cardiovascular parameters change in response to the activities and events taking place

More information

1 Non-invasive measurement of arterial pressure

1 Non-invasive measurement of arterial pressure Non-invasive measurement of arterial pressure I. Background A. Circulatory systems Human circulation takes place in a closed system that consists of two subsystems, pulmonary circulation and systemic circulation,

More information

Automated Blood Pressure Monitor

Automated Blood Pressure Monitor Automated Blood Pressure Monitor For Cardiac Stress and Exercise Testing SunTech Tango M2 Our cardiac stress blood pressure (BP) monitoring products give you the confidence to know that patient movement,

More information

HUMAN ANATOMY AND PHYSIOLOGY

HUMAN ANATOMY AND PHYSIOLOGY HUMAN ANATOMY AND PHYSIOLOGY NAME Detection of heart sounds. Clean the ear pieces of the stethoscope before using. The ear pieces should be pointing slightly forward when inserted into the ears because

More information

IB TOPIC 6.2 THE BLOOD SYSTEM

IB TOPIC 6.2 THE BLOOD SYSTEM IB TOPIC 6.2 THE BLOOD SYSTEM THE BLOOD SYSTEM TERMS TO KNOW circulation ventricle artery vein 6.2.U1 - Arteries convey blood at high pressure from the ventricles to the tissues of the body Circulation

More information

AFFORDABLE TECHNOLOGY

AFFORDABLE TECHNOLOGY World Health Organization AFFORDABLE TECHNOLOGY BLOOD PRESSURE MEASURING DEVICES FOR LOW RESOURCE SETTINGS CARDIOVASCULAR DISEASES Blood Pressure Measurement in Low Resource Settings Annex 1: Blood Pressure

More information

Hands on Sports Therapy KNOWLEDGE REVIEW QUESTIONS 2004 Thomson Learning It can help to shape a basic fitness training programme

Hands on Sports Therapy KNOWLEDGE REVIEW QUESTIONS 2004 Thomson Learning It can help to shape a basic fitness training programme Hands on Sports Therapy KNOWLEDGE REVIEW QUESTIONS 2004 Thomson Learning 1 CHAPTER 13 Knowledge Review Q1: Why is fitness testing useful? A1: Fitness testing is useful for various reasons: 1. It can help

More information

Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol

Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol Abstract Objectives Existing concerns over the accuracy of automated blood

More information

Vital Signs. Vital Signs. Pulse. Temperature. Respiration. Blood Pressure

Vital Signs. Vital Signs. Pulse. Temperature. Respiration. Blood Pressure Vital Signs Jarvis, Chapter 9 Vital Signs Classic Vital Signs TPR/BP Temperature Pulse Respirations Blood Pressure Additional Vital Signs Height Weight BMI (Kg/m2) or (702Xlbs/in2) Supine, orthostatic

More information

CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME

CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME 64 CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME 4.1 GENERAL This chapter presents the methodologies that are usually adopted for the measurement of blood pressure, heart rate and pulse

More information

Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods

Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods CLINICAL STUDIES Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods Linda A. Green, RN, MS, Robin D. Froman, RN, PhD, FAAN Objective: To determine the equivalence of auscultatory

More information

The Cardiac Cycle Clive M. Baumgarten, Ph.D.

The Cardiac Cycle Clive M. Baumgarten, Ph.D. The Cardiac Cycle Clive M. Baumgarten, Ph.D. OBJECTIVES: 1. Describe periods comprising cardiac cycle and events within each period 2. Describe the temporal relationships between pressure, blood flow,

More information

The measurement of blood pressure and hypertension. Handout Fenyvesi Tamás III.Department of Medicine

The measurement of blood pressure and hypertension. Handout Fenyvesi Tamás III.Department of Medicine The measurement of blood pressure and hypertension Handout Fenyvesi Tamás III.Department of Medicine 1 2 History of the blood pressure measurement direct Stephen HALES 1726 horse carotid indirect Riva-Rocci

More information

Lab #10 Physiology of the Circulatory System

Lab #10 Physiology of the Circulatory System Lab #10 Physiology of the Circulatory System Introduction The circulatory system functions to deliver oxygen an nutrients to tissues for growth and metabolism, and to remove metabolic wastes. The heart

More information

BRIEF COMMUNICATIONS. KEY WORDS: Ambulatory blood pressure monitoring, placebo effect, antihypertensive drug trials.

BRIEF COMMUNICATIONS. KEY WORDS: Ambulatory blood pressure monitoring, placebo effect, antihypertensive drug trials. AJH 1995; 8:311-315 BRIEF COMMUNICATIONS Lack of Placebo Effect on Ambulatory Blood Pressure Giuseppe Mancia, Stefano Omboni, Gianfranco Parati, Antonella Ravogli, Alessandra Villani, and Alberto Zanchetti

More information

IB TOPIC 6.2 THE BLOOD SYSTEM

IB TOPIC 6.2 THE BLOOD SYSTEM IB TOPIC 6.2 THE BLOOD SYSTEM TERMS TO KNOW circulation ventricle artery vein THE BLOOD SYSTEM 6.2.U1 - Arteries convey blood at high pressure from the ventricles to the tissues of the body Circulation

More information

Physiology of the Circulatory System modified from

Physiology of the Circulatory System modified from Physiology of the Circulatory System modified from http://www.ekcsk12.org/science/aplabreview/aplab10.htm Introduction The circulatory system functions to deliver oxygen and nutrients to tissues for growth

More information

Cardiovascular Physiology

Cardiovascular Physiology Cardiovascular Physiology The mammalian heart is a pump that pushes blood around the body and is made of four chambers: right and left atria and right and left ventricles. The two atria act as collecting

More information

and bias, which are known to be present in self-home and in professional office BP measurements taken using the auscultatory technique [7].

and bias, which are known to be present in self-home and in professional office BP measurements taken using the auscultatory technique [7]. Devices and technology 37 Validation of the A&D UM-11 professional hybrid device for office blood pressure measurement according to the International Protocol George S. Stergiou, Periklis P. Giovas, Charilaos

More information

MANUAL BLOOD PRESSURE MONITOR BPM 168B Contents Blood Pressure Monitor Intended Use What is blood pressure?

MANUAL BLOOD PRESSURE MONITOR BPM 168B Contents Blood Pressure Monitor Intended Use What is blood pressure? MANUAL BLOOD PRESSURE MONITOR BPM 168B Contents 01... Blood Pressure Monitor Intended Use 02... 1. What is blood pressure? 02... 2. Why is it useful measure blood pressure at home?... A. WHO blood pressure

More information

Blood Pressure Competency Training

Blood Pressure Competency Training Blood Pressure Competency Training WORKBOOK Table of Contents 3..... Introduction 4..... Background 6..... Equipment Accurate blood pressure What impacts a patient s blood pressure Strategies for proper

More information

The evaluation of cardiovascular response to exercise in healthy Turkish children

The evaluation of cardiovascular response to exercise in healthy Turkish children The Turkish Journal of Pediatrics 2009; 51: 472-477 Original The evaluation of cardiovascular response to exercise in healthy Turkish children Hülya Akdur 1, Ahmet Bilge Sözen 2, Zerrin Yiğit 3, Funda

More information

LAB 4. Human Cardiovascular Lab

LAB 4. Human Cardiovascular Lab 35 Assignments: LAB 4 Human Cardiovascular Lab Due before lab: Do IP exercises (Factors that Affect Blood Pressure and Arterial Baroreceptor Reflex (Pgs. 46-47). Quiz: Define the terms on pg 36 and IP

More information

Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination

Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination JIRI JILEK 1, MILAN STORK 2 1 Carditech, Culver City, California, USA jilekj@usa.net 2 Dept. of Applied Electronics and

More information

AS OCR PHYSICAL EDUCATION The Vascular System

AS OCR PHYSICAL EDUCATION The Vascular System AS OCR PHYSICAL EDUCATION The Vascular System Learning Objectives: - Distribution of Cardiac output at rest and during exercise - Role of the Vasomotor centre, arterioles and pre-capillary sphincters -

More information

Large Arteries of Heart

Large Arteries of Heart Cardiovascular System (Part A-2) Module 5 -Chapter 8 Overview Arteries Capillaries Veins Heart Anatomy Conduction System Blood pressure Fetal circulation Susie Turner, M.D. 1/5/13 Large Arteries of Heart

More information

The magnitude and duration of ambulatory blood pressure reduction following acute exercise

The magnitude and duration of ambulatory blood pressure reduction following acute exercise Journal of Human Hypertension (1999) 13, 361 366 1999 Stockton Press. All rights reserved 0950-9240/99 $12.00 http://www.stockton-press.co.uk/jhh ORIGINAL ARTICLE The magnitude and duration of ambulatory

More information

science-u.org What affects blood Blood Pressure Directions You Will Need BEST FOR GRADES 5-8 ESTIMATED TIME Minutes

science-u.org What affects blood Blood Pressure Directions You Will Need BEST FOR GRADES 5-8 ESTIMATED TIME Minutes What affects blood pressure? BEST FOR GRADES 5-8 ESTIMATED TIME 25-60 Minutes You Will Need Sphygmomanometer (blood pressure cuff) Stethoscope Blood pressure cuffs and stethoscopes can be found online

More information

Masters two step exercise test, also known as. Augmented workload in DMT exercise. Original Article. Wg Cdr VN Jha * ABSTRACT

Masters two step exercise test, also known as. Augmented workload in DMT exercise. Original Article. Wg Cdr VN Jha * ABSTRACT Original Article Augmented workload in DMT exercise: VN Jha Augmented workload in DMT exercise Wg Cdr VN Jha * ABSTRACT Masters two step exercise test (DMT) forms the mainstay in the field screening of

More information

How To Measure Vital Signs

How To Measure Vital Signs How To Measure Vital Signs How to Use This Tutorial This tutorial is intended for healthcare providers or students to teach basic vital signs skills Use the navigation buttons below to move through the

More information

Biology 13A Lab #10: Cardiovascular System II ECG & Heart Disease

Biology 13A Lab #10: Cardiovascular System II ECG & Heart Disease Biology 13A Lab #10: Cardiovascular System II ECG & Heart Disease Lab #10 Table of Contents: Expected Learning Outcomes...... 83 Introduction....... 84 Activity 1: Collecting ECG Data..... 85 Activity

More information

Blood Pressure and Exercise Lab

Blood Pressure and Exercise Lab Blood Pressure and Exercise Lab Rob MacLeod March 9, 2006 1 Purpose and Background Purpose: To learn about external means to measure blood pressure, observe features of venous circulation, and observe

More information

Importance Of Taking Manual Blood Pressure In Both Arms

Importance Of Taking Manual Blood Pressure In Both Arms Importance Of Taking Manual Blood Pressure In Both Arms Its importance to health is now well-recognised, in that high blood pressure can lead to Have a look at our slide show on how to take a blood pressure

More information

Lab #3: Electrocardiogram (ECG / EKG)

Lab #3: Electrocardiogram (ECG / EKG) Lab #3: Electrocardiogram (ECG / EKG) An introduction to the recording and analysis of cardiac activity Introduction The beating of the heart is triggered by an electrical signal from the pacemaker. The

More information

& Wilkins. a Division of Cardiology, Schulich Heart Centre, b Institute for Clinical and

& Wilkins. a Division of Cardiology, Schulich Heart Centre, b Institute for Clinical and Original article 333 Optimum frequency of office blood pressure measurement using an automated sphygmomanometer Martin G. Myers a, Miguel Valdivieso a and Alexander Kiss b,c Objective To determine the

More information

Blood Pressure Estimation Using Photoplethysmography (PPG)

Blood Pressure Estimation Using Photoplethysmography (PPG) Blood Pressure Estimation Using Photoplethysmography (PPG) 1 Siddhi Sham Karande, BE E&TC, VIIT Pune. 2 Kiran Rajendrasingh Thakur, BE E&TC, VIIT Pune. 3 Sneha Dattatraya Waghmare, BE E&TC, VIIT Pune 4

More information

12.2 Monitoring the Human Circulatory System

12.2 Monitoring the Human Circulatory System 12.2 Monitoring the Human Circulatory System Video 1: 3D Animation of Heart Pumping Blood blood flow through the heart... Video 2: Hank Reviews Everything on the Heart https://www.youtube.com/watch?v=x9zz6tcxari

More information

Managing Blood Pressure: It Takes a Team

Managing Blood Pressure: It Takes a Team Managing Blood Pressure: It Takes a Team Module Two How to Take Blood Pressure October 2011 MODULE TWO: How to Take Blood Pressure Module Two: HOW TO TAKE BLOOD PRESSURE Objectives Demonstrate proper technique

More information

Lectures on Medical Biophysics Department of Biophysics, Medical Faculty, Masaryk University in Brno. Biophysics of cardiovascular system

Lectures on Medical Biophysics Department of Biophysics, Medical Faculty, Masaryk University in Brno. Biophysics of cardiovascular system Lectures on Medical Biophysics Department of Biophysics, Medical Faculty, Masaryk University in Brno Biophysics of cardiovascular system 1 Lecture outline Mechanical properties of blood vessels Reynolds

More information

LAB 9: Metabolic Rates

LAB 9: Metabolic Rates LAB 9: Metabolic Rates Introduction: The cardiovascular (circulatory) system functions to deliver oxygen and nutrients to tissues for growth and metabolism, and to remove metabolic wastes. The heart pumps

More information

Chapter 29 1/8/2018. Vital Signs. Measuring and Reporting Vital Signs. Key Terms

Chapter 29 1/8/2018. Vital Signs. Measuring and Reporting Vital Signs. Key Terms Chapter 29 Vital Signs Key Terms Blood pressure Body temperature Diastolic pressure Hypertension Hypotension Pulse Pulse rate Respiration Sphygmomanometer Stethoscope Systolic pressure Vital signs Measuring

More information

PLANK 1 Direct Care Staff Trained in Accurate BP Measurement

PLANK 1 Direct Care Staff Trained in Accurate BP Measurement Direct Care Staff Trained in Accurate BP Measurement 1. 2. 3. 5. 22 to 26 cm Small adult (12X22 cm) 27 to 34 cm Adult (16X30 cm) 35 to 44 cm Large adult (16X36 cm) 45 to 52 cm Adult thigh (16X42 cm) widths

More information

The patient with coronary heart disease at altitude: observations during acute exposure to 3100 meters

The patient with coronary heart disease at altitude: observations during acute exposure to 3100 meters Journal of Wilderness Medicine 1, 147-153 (1990) The patient with coronary heart disease at altitude: observations during acute exposure to 3100 BJ. MORGAN!, J.K. ALEXANDER2*, S.A. NICOLI l and H.L. BRAMMELU

More information

Lab Six: Maximal Exercise. Stephanie Smith. University of Otago. PHSE 203: Exercise Physiology. Due 5pm Monday, 9 th May 2011.

Lab Six: Maximal Exercise. Stephanie Smith. University of Otago. PHSE 203: Exercise Physiology. Due 5pm Monday, 9 th May 2011. Lab 6 PP4, S Smith, 1 LAB REPORT SIX Lab Six: Maximal Exercise Stephanie Smith University of Otago PHSE 203: Exercise Physiology Due 5pm Monday, 9 th May 2011 Lab Stream: PP4 E-mail: smitsm31@suny.oneonta.edu

More information

Influence of Age, Hypertension or Myocardial Infarction on Cardiovascular Responses to Changes in Body Position

Influence of Age, Hypertension or Myocardial Infarction on Cardiovascular Responses to Changes in Body Position Influence of Age, Hypertension or Myocardial Infarction on Cardiovascular Responses to Changes in Body Position A population-based study in 30-, 50- and 60-year-old men BY ANNA HOFSTEN UPPSALA UNIVERSITY

More information

1. Department of Gynecology and Obstetrics, St. Joseph's Hospital Berlin Tempelhof, Germany

1. Department of Gynecology and Obstetrics, St. Joseph's Hospital Berlin Tempelhof, Germany Page 1 of 9 Validation of the TONOPORT VI ambulatory blood pressure monitor, according to the European Society of Hypertension International Protocol revision 2010 Michael Abou-Dakn 1, Cornelius Döhmen

More information

Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation

Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation ORIGINAL RESEARCH Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation Tyler S Lamb BSc Amar Thakrar MD, FRCPC Mahua Ghosh MD, PhD Merne P Wilson MSc Thomas W Wilson

More information

Blood Pressure in Exercise Prof. Dr. Alfred Hager

Blood Pressure in Exercise Prof. Dr. Alfred Hager Blood Pressure in Exercise Prof. Dr. Alfred Hager Department of Pediatric Cardiology and Congenital Heart Disease Head of Outpatient Section Head of Exercise Laboratory Member of DGPK Guideline Committee

More information

following: the readout..

following: the readout.. Read Me Vital Signs In-Lab Guide We will be studying 5 concepts in lab 1. Study the anatomy of the Intrinsic Conduction System using heart models. 2. Study the ECG tracings, including the following: -

More information

How Low Do We Go? Update on Hypertension

How Low Do We Go? Update on Hypertension How Low Do We Go? Update on Beth L. Abramson, MD, FRCPC, FACC As presented at the University of Toronto s Saturday at the University Session (September 2003) Arecent World Health Organization report states

More information

Webinar Logistics. Now What Do I Need To Do? Blood Pressure in Kids January NC Department of Health and Human Services Division of Public Health

Webinar Logistics. Now What Do I Need To Do? Blood Pressure in Kids January NC Department of Health and Human Services Division of Public Health NC Department of Health and Human Services Division of Public Health Now What Do I Need To Do? Blood Pressure in Kids January 2018 Gerri Mattson, MD, MSPH, FAAP Pediatric Medical Consultant Children and

More information

Training and Certification of Blood Pressure Observers

Training and Certification of Blood Pressure Observers Training and Certification of Blood Pressure Observers J. DAVID CURB, M.D., M.P.H., DARWIN R. LABARTHE, M.D., PH.D., SHARON POIZNER COOPER, PH.D., GARY R. CUTTER, PH.D., AND C. MORTON HAWKINS, SC.D. SUMMARY

More information

DEFIBRILLATORS. Prof. Yasser Mostafa Kadah

DEFIBRILLATORS. Prof. Yasser Mostafa Kadah DEFIBRILLATORS Prof. Yasser Mostafa Kadah Basics Defibrillation is definitive treatment for life-threatening cardiac arrhythmias such as ventricular fibrillation Defibrillation consists of delivering therapeutic

More information

Identification of Ischemic Lesions Based on Difference Integral Maps, Comparison of Several ECG Intervals

Identification of Ischemic Lesions Based on Difference Integral Maps, Comparison of Several ECG Intervals 1.2478/v148-9-21-7 MEASUREMENT SCIENCE REVIEW, Volume 9, No. 5, 29 Identification of Ischemic Lesions Based on Difference Integral Maps, Comparison of Several ECG Intervals J. Švehlíková, M. Kania 1, M.

More information

CARDIOVASCULAR MONITORING. Prof. Yasser Mostafa Kadah

CARDIOVASCULAR MONITORING. Prof. Yasser Mostafa Kadah CARDIOVASCULAR MONITORING Prof. Yasser Mostafa Kadah Introduction Cardiovascular monitoring covers monitoring of heart and circulatory functions It makes it possible to commence interventions quickly in

More information

iworx Sample Lab Experiment HC-5: Body Position, Exercise, and Cardiac Output

iworx Sample Lab Experiment HC-5: Body Position, Exercise, and Cardiac Output Experiment HC-5: Body Position, Exercise, and Cardiac Output Exercise 1: Cardiac Output While Reclining Aim: To determine the cardiac output of a subject through the measurement of blood pressures and

More information

To Correlate Ejection Fraction with 6 Minute Walked Distance and Quality of Life in Patients with Left Ventricular Heart Failure

To Correlate Ejection Fraction with 6 Minute Walked Distance and Quality of Life in Patients with Left Ventricular Heart Failure To Correlate Ejection Fraction with 6 Minute Walked Distance and Quality of Life in Patients with Left Ventricular Heart Failure Pramila S Kudtarkar*, Mariya P Jiandani*, Ashish Nabar** Abstract Purpose

More information

ECG, Blood Pressure, and Exercise Lab

ECG, Blood Pressure, and Exercise Lab ECG, Blood Pressure, and Exercise Lab Rob MacLeod March 31, 2006 1 Purpose and Background Purpose: The purpose of the lab is to learn about measuring the ECG and blood pressures and observing the effects

More information

Vital Signs. Vital Signs. Vital Signs

Vital Signs. Vital Signs. Vital Signs Vital Signs Vital Signs Why do vital signs? Determine relative status of vital organs Establish baseline Monitor response to Rx, meds Observe trends Determine need for further evaluation, Rx, intervention

More information

LEARNING OUTCOME The students will be able to elicit vital signs correctly on human volunteers/patients

LEARNING OUTCOME The students will be able to elicit vital signs correctly on human volunteers/patients Vital signs (pulse, blood pressure, temperature, respiratory rate, pain) are physiological parameters that a healthcare professional requires when dealing with patients. Accurate measurement of vital signs

More information

Exercise Test: Practice and Interpretation. Jidong Sung Division of Cardiology Samsung Medical Center Sungkyunkwan University School of Medicine

Exercise Test: Practice and Interpretation. Jidong Sung Division of Cardiology Samsung Medical Center Sungkyunkwan University School of Medicine Exercise Test: Practice and Interpretation Jidong Sung Division of Cardiology Samsung Medical Center Sungkyunkwan University School of Medicine 2 Aerobic capacity and survival Circulation 117:614, 2008

More information

#6 - Cardiovascular III Heart Sounds, Pulse Rate, Hemoglobin Saturation, and Blood Pressure

#6 - Cardiovascular III Heart Sounds, Pulse Rate, Hemoglobin Saturation, and Blood Pressure #6 - Cardiovascular III Heart Sounds, Pulse Rate, Hemoglobin Saturation, and Blood Pressure Objectives: Observe slide of artery and vein cross-section Auscultate heart sounds using a stethoscope Measure

More information

Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis

Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis Journal of Physics: Conference Series PAPER OPEN ACCESS Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis To cite this article: Valerie Tan et al 2018 J. Phys.: Conf. Ser. 1049 012062

More information

High Blood Pressure Blood Pressure Monitoring at Home -- F...

High Blood Pressure Blood Pressure Monitoring at Home -- F... Return to Web version High Blood Pressure Blood Pressure Monitoring at Home Why should I measure my blood pressure at home and keep a record of it? Measuring your blood pressure at home and keeping a record

More information

Bio& 242, Unit 3/ Lab 4 Blood Vessels, Lymphatic System and Blood Pressure G. Blevins/ G. Brady Summer 2009

Bio& 242, Unit 3/ Lab 4 Blood Vessels, Lymphatic System and Blood Pressure G. Blevins/ G. Brady Summer 2009 Bio& 242, Unit 3/ Lab 4 Blood Vessels, Lymphatic System and Blood Pressure G. Blevins/ G. Brady Summer 2009 Major Arteries and for arteries and veins with common names your answer must include either artery

More information

10 Graded Exercise Testing

10 Graded Exercise Testing 10 Graded Exercise Testing Clinton A. Brawner, MS, RCEP CONTENTS Background 111 Timing of the GXT 113 Repeating the GXT 114 Pre-Test Considerations 114 Protocol Selection 114 Preparing for the GXT 116

More information

Influence of 50-Hz electric and magnetic fields on human blood pressure

Influence of 50-Hz electric and magnetic fields on human blood pressure Radiat Environ Biophys (1996) 35: 199 204 Springer-Verlag 1996 ORIGINAL PAPER L. Korpinen J. Partanen Influence of 50-Hz electric and magnetic fields on human blood pressure Received: 6 June 1994 / Accepted

More information

Human Anatomy and Physiology II Laboratory Cardiovascular Physiology

Human Anatomy and Physiology II Laboratory Cardiovascular Physiology Human Anatomy and Physiology II Laboratory Cardiovascular Physiology 1 This lab involves two exercises: 1) Conduction System of the Heart and Electrocardiography and 2) Human Cardiovascular Physiology:

More information

UNIVERSITY OF UTAH HEALTH CARE HOSPITALS AND CLINICS

UNIVERSITY OF UTAH HEALTH CARE HOSPITALS AND CLINICS UNIVERSITY OF UTAH HEALTH CARE HOSPITALS AND CLINICS CARDIAC MECHANICAL SUPPORT PROGRAM GUIDELINES CARDIAC MECHANICAL SUPPORT: LVAD BASICS FREQUENT SCENARIOS AND TROUBLESHOOTING Review Date: July 2011

More information

Cardiovascular Physiology Laboratory Clive M. Baumgarten, Ph.D.

Cardiovascular Physiology Laboratory Clive M. Baumgarten, Ph.D. Cardiovascular Physiology Laboratory Clive M. Baumgarten, Ph.D. To pump blood, the heart has a rhythmical sequence of both electrical and mechanical events, the cardiac cycle. The electrical activity,

More information

BLOOD PRESSURE ASSESSMENT

BLOOD PRESSURE ASSESSMENT BLOOD PRESSURE ASSESSMENT Course Principles of Health Science Unit X Vital Signs Course Health Science Unit VI CPR / AED & First Aid Essential Question How do health care skills help to promote health

More information

Objectives of the Heart

Objectives of the Heart Objectives of the Heart Electrical activity of the heart Action potential EKG Cardiac cycle Heart sounds Heart Rate The heart s beat separated into 2 phases Relaxed phase diastole (filling of the chambers)

More information

AP Biology Lab 10 PHYSIOLOGY OF THE CIRCULATORY SYSTEM

AP Biology Lab 10 PHYSIOLOGY OF THE CIRCULATORY SYSTEM AP Biology Laboratory Date: Name and Period: AP Biology Lab 10 PHYSIOLOGY OF THE CIRCULATORY SYSTEM OVERVIEW In this lab you will: 1. in Exercise 10A you will learn how to measure blood pressure. 2. in

More information

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010 Short title: Validation

More information

STUDIES INVOLVING THE NON-INVASIVE MEASUREMENT OF BLOOD PRESSURE IN THE ARM

STUDIES INVOLVING THE NON-INVASIVE MEASUREMENT OF BLOOD PRESSURE IN THE ARM STUDIES INVOLVING THE NON-INVASIVE MEASUREMENT OF BLOOD PRESSURE IN THE ARM 1. SCOPE Blood pressure is a vital sign that is used in many clinical situations, and assesses the pressure of blood as it flows

More information

Welcome to the S410. Blood Pressure System. Instruction Manual

Welcome to the S410. Blood Pressure System. Instruction Manual Welcome to the S410 Blood Pressure System Instruction Manual The simulator is to be used as part of an approved educational program for healthcare. It is not a substitute for traditional learning methods

More information

CHRONIC HEART FAILURE : WHAT ELSE COULD WE OFFER TO OUR PATIENTS? Cardiac Rehabilitation Society of Thailand

CHRONIC HEART FAILURE : WHAT ELSE COULD WE OFFER TO OUR PATIENTS? Cardiac Rehabilitation Society of Thailand CHRONIC HEART FAILURE : WHAT ELSE COULD WE OFFER TO OUR PATIENTS? Cardiac Rehabilitation Society of Thailand ENHANCED EXTERNAL COUNTER PULSATION Piyanuj Ruckpanich, MD. Cardiac Rehabilitation Center Perfect

More information

5.2. MEASUREMENT OF BLOOD PRESSURE

5.2. MEASUREMENT OF BLOOD PRESSURE 5.2. MEASUREMENT OF BLOOD PRESSURE Purpose of experiment To examine a person s blood pressure. Tasks of experiment To record the pressure changes in the vascular system: using the traditional method measure

More information

Human Cardiovascular Physiology: Blood Pressure and Pulse Determinations

Human Cardiovascular Physiology: Blood Pressure and Pulse Determinations ighapmlre33apg269_274 5/12/04 3:10 PM Page 269 impos03 302:bjighapmL:ighapmLrevshts:layouts: NAME Human Cardiovascular Physiology: Blood Pressure and Pulse Determinations LAB TIME/DATE REVIEW SHEET exercise

More information

Assessment of techniques for measurement of blood pressure in infants and children

Assessment of techniques for measurement of blood pressure in infants and children Archives of Disease in Childhood, 1973, 48, 932. Assessment of techniques for measurement of blood pressure in infants and children A. M. ELSEED, E. A. SHINEBOURNE, and M. C. JOSEPH From the Paediatric

More information

Unit 1: Human Systems. The Circulatory System

Unit 1: Human Systems. The Circulatory System Unit 1: Human Systems The Circulatory System nourish all cells with oxygen, glucose, amino acids and other nutrients and carry away carbon dioxide, urea and other wastes Purposes Transport chemical messengers

More information

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model Instrumentation and Control Department, JSS Academy of Technical Education, Noida (U.P.), India Abstract- The non-invasive

More information

ASSESSMENT OF CARDIOVASCULAR RESPONSE TO TREADMILL EXERCISE IN NORMAL HEALTHY INDIAN ADOLESCENTS

ASSESSMENT OF CARDIOVASCULAR RESPONSE TO TREADMILL EXERCISE IN NORMAL HEALTHY INDIAN ADOLESCENTS Indian 36 Pande J Physiol et al Pharmacol 01; 56(1) : 36 41 Indian J Physiol Pharmacol 01; 56(1) ASSESSMENT OF CARDIOVASCULAR RESPONSE TO TREADMILL EXERCISE IN NORMAL HEALTHY INDIAN ADOLESCENTS SUSHMA

More information

Chapter 08. Health Screening and Risk Classification

Chapter 08. Health Screening and Risk Classification Chapter 08 Health Screening and Risk Classification Preliminary Health Screening and Risk Classification Protocol: 1) Conduct a Preliminary Health Evaluation 2) Determine Health /Disease Risks 3) Determine

More information

Each student should record the ECG of one of the members of the lab group and have their own ECG recorded.

Each student should record the ECG of one of the members of the lab group and have their own ECG recorded. EXPERIMENT 1 ELECTROCARDIOGRAPHY The purpose of this experiment is to introduce you to the techniques of electrocardiography and the interpretation of electrocardiograms. In Part A of the experiment, you

More information