Hardware-in-the-loop-simulation of the cardiovascular system, with assist device testing application

Size: px
Start display at page:

Download "Hardware-in-the-loop-simulation of the cardiovascular system, with assist device testing application"

Transcription

1 Medical Engineering & Physics 29 (2007) Hardware-in-the-loop-simulation of the cardiovascular system, with assist device testing application B.M. Hanson a,, M.C. Levesley a, K. Watterson b, P.G. Walker a a School of Mechanical Engineering, University of Leeds, Leeds LS2 9JT, UK b Yorkshire Heart Centre, Leeds General Infirmary, Leeds, UK Received 29 November 2005; received in revised form 13 March 2006; accepted 9 May 2006 Abstract This paper presents a technique for evaluating the performance of biomedical devices by combining physical (mechanical) testing with a numerical, computerised model of a biological system. This technique is developed for evaluation of a cardiac assist device prior to in vivo trials. This device will wrap around a failing heart and provide physical beating assistance (dynamic cardiac compression). In vitro, the device to be tested is placed around a simulator comprising a mechanical simulation of the beating ventricles. This hardware model interfaces with a computerised (software) model of the cardiovascular system. In real time the software model calculates the effect of the assistance on the cardiovascular system and controls the beating motion of the hardware heart simulator appropriately. The software model of the cardiovascular system can represent ventricles in various stages of heart failure, and/or hardened or congested blood vessels as required. The software displays physiological traces showing the cardiac output, depending on the natural function of the modelled heart together with the physical assist power provided. This system was used to evaluate the effectiveness of control techniques applied to the assist device. Experimental results are presented showing the efficacy of prototype assist on healthy and weakened hearts, and the effect of asynchronous assist IPEM. Published by Elsevier Ltd. All rights reserved. Keywords: Hardware-in-the-loop (HIL); Cardiac assist device; Modelling; Simulation; Cardiovascular system; LVAD 1. Introduction Cardiac assist devices are currently being developed with the aim of providing physical pumping assistance to a weakened or failing heart. Implantable impeller pump-based left ventricular assist devices (LVADs) are emerging. Alternatively, dynamic cardiac compression (DCC) can assist by providing compression to the surface of the ventricle(s) [1] thereby avoiding some problems of immune-system rejection and thromboses [2]. In the early development of LVADs, numerical simulations of circulatory systems have been valuable tools when used to simulate the effect of assist devices on the cardiovascular Correspondence to: Department of Mechanical Engineering, University College London, Torrington Place, London WC1E 7JE, UK. Tel.: address: ben@benhanson.com (B.M. Hanson). system (CVS) [3,4]. These models have a long history of use and some are highly detailed (e.g. [5]). However, when working prototypes have been constructed purely numerical techniques become less attractive; it can be inconvenient and inaccurate to create numerical models of prototype devices, whose physical behaviour may not be fully understood yet. Physical testing is therefore required. The actual hydraulic performance of prototype LVAD systems has been tested on electro-hydraulic servo-systems [6 8]. Investigators have used simple models of the circulatory system to present a realistic hydraulic load to the LVAD, however these testing models have not shown how the mechanical performance of an LVAD directly affects the complete circulatory system. With a DCC assist device, the interaction between the assist device and the surface of the heart is crucial. This interaction is likely to depend on physical features which are particularly difficult to model, such as non-linear friction and backlash. Physical testing of DCC devices requires /$ see front matter 2006 IPEM. Published by Elsevier Ltd. All rights reserved. doi: /j.medengphy

2 368 B.M. Hanson et al. / Medical Engineering & Physics 29 (2007) a physical heart (or model) on which to apply compression, and a means of measuring the compressive effort applied. In vitro studies on dead hearts are unfortunately not feasible due to non-function of papillary muscles, collapse of the ventricular outflow tracts and increased myocardial stiffness (unpublished results). However, excised hearts have been sustained for in vitro tests using a blood supply from a support animal [1,9,10]. In these studies, a hydraulic servo pump was used to present a realistic outflow impedance using a windkessel model, as with the LVAD tests. Again, the effects on the closed-loop circulatory system were not studied. These animal studies have produced invaluable results, however they are costly in terms of time, resources and animal lives. 2. Hardware-in-the-loop concept This paper describes the use of hardware-in-the-loop (HIL) simulation to test a DCC assist device in vitro. This HIL simulation combines a realistic numerical model of the heart and cardiovascular system with a controllable physical heart simulator, and this interacts in real time with a prototype DCC assist device. The nature of these interactions is shown in Fig. 1. Hardware-in-the-loop-simulation has been developing in industrial control for testing of systems comprising some physical and some simulated components [11 13]. Simulation is used to represent processes that are physically unavailable, or whose use would be too costly, dangerous, or time-consuming. Proven benefits of HIL include: Reproducibility of experiments. The ability to perform tests which would otherwise be impossible, impractical or unsafe: testing a component under extreme or dangerous operating or environmental conditions (e.g. extremes of temperature, pressure, acceleration); testing effects of sensor and/or actuator faults; long-term durability testing until failure. In this investigation, the cardiovascular system is the simulated component, and the benefits described above apply equally to the biomedical field. Using HIL simulation for biological systems could provide: The possibility to test on a wide range of simulated patient geometries and pathologies. Repeated testing on a consistent model. Facilitated numerical quantification of performance by recording simulated physiological parameters. Replacement of human and/or animal subjects: sterile, clinical environment is not required; ethical issues are removed; cost and development time are reduced. 3. Hardware-in-the-loop simulation Fig. 1 shows the overall structure of the HIL simulation for the assist device application. The system involves a position control loop (indicated) whereby the diameter of the heart simulator is controlled by computer, such that it is a real-time physical display of the diameter of the simulated heart. The assist device contracts around this simulator and a sensor records the assist force at the physical interface between assist device and simulator. The HIL aspect is that this physical force signal forms part of the control loop: the force signal is fed into the CVS model, which calculates its effect on pressure within the heart and therefore blood Fig. 1. Structure of the hardware-in-the-loop-simulation of the cardiovascular system.

3 B.M. Hanson et al. / Medical Engineering & Physics 29 (2007) flow into and out of the heart, therefore the diameter of the simulated heart. Thus, the motion of the hardware simulator depends on the effect that hardware interaction with the assist device has on the software CVS model. From the point of view of the assist band, the heart s physical motion and response to assistance appear realistic, as governed by the software CVS model. The components of the HIL cardiovascular simulation will now be described, with reference to Fig Numerical model of cardiovascular system The human cardiovascular system has been modelled many times at various levels of complexity (for a review see, e.g. [14], and the state-of-the-art [15,16]). This particular model is based on some elements of previously-reported models, selected as appropriate to the requirements of the DCC assist device application. The structure of the model is shown in Fig. 2; it is an important feature that the model is haemodynamically closed-loop in order to assess the effects of applying assistance. The function of the software model is described in detail in [17], however an overview of the model is provided here. In a HIL simulation, the interface between hardware and software is crucial in this case that is the heart. The model is therefore biased with more detail being used to describe the heart than the rest of the circulatory system. The four heart chambers are modelled separately, allowing assessment of the effect of disease or incompetence in any or all of the chambers. Two further passive, compliant compartments represent the aorta and pulmonary artery. The equation relating flow and pressure in each compartment is of the form: P x = Z x φ x + P x additional (1) where P additional is applied to the heart chambers only and represents the sum of pressures generated by passive stretching of the pericardium, natural systolic function, and assist pressure, where appropriate. For the atria we apply a pressure time curve; this is not dependent on atrial volume. For the ventricles we use a function, f(t), to generate a time-varying myocardial wall stress; the instantaneous active systolic pressure within the ventricles is then calculated from the wall stress and ventricle dimensions. This stress, σ, depends on the volume, V, togive a representation of the Frank Starling relationship, and is also rate-dependent [5]: [ ( )] dv σ(t) = σ max f (t)k V (V (t) V 0 ) 1 D V (2) dt Since research has concentrated on modelling the primary mechanical effects of assistance, the model does not include hormonal effects, vasomotor control, orthostatic stress or breathing, although these could of course be added at a later stage. The model has been implemented using LabVIEW TM to produce not only the software/hardware interface but also the numerical circulatory model itself, in LabVIEW TM s visual programming language. The use of a graphical programming environment makes it easy to incorporate physiological data into the model, and to manipulate it into a suitable form. Non-linear circulatory elements and functions are shown graphically and can therefore be consulted, verified, and manipulated more easily than tabulated data. As an example Fig. 3 shows the shape of a typical activation function for myocardial stress generation. This is scaled in the X and Y directions and used in the heart chamber models to create the function f(t) which is a stress, Y, acting over a time, X. The graphical code is compiled efficiently to take advantage of the hardware computation processes available on a Fig. 2. Representation of the circulatory system model using an electrical equivalent, indicating six compartments in which blood can be stored (vertical branches). Fig. 3. A normalised activation function curve: used to generate myocardial pressures within the CVS model.

4 370 B.M. Hanson et al. / Medical Engineering & Physics 29 (2007) typical modern processor, thus in practice may be computed as fast as code written in, e.g. C Heart simulator An interface is required to communicate between the physical (hardware) assist device and the software cardiovascular model. The requirements of the physical simulator of the heart for this assist device application are: it must be possible to wrap a contractile band-type heart assist device around the simulator, and sense the assist pressure produced, the device must represent the motion of one slice through the ventricles the volume of the heart encircled by one assist belt, the device must be able to simulate normal and pathological heart motion at rates of up to 150 bpm. In use, when combined with the software model and controller, the device is required to beat in a real-time display of the changing volume of the ventricles. The combined system must respond to assistance compression in a physically realistic manner. This interface takes the form of a heart simulator, as illustrated in Fig. 4. The simulator was constructed using electromagnetic swing-arm actuators that can be controlled by computer easily and accurately. These are arranged in a circular array as shown in Fig. 4 the swing arms are shown, but the actuation method is omitted for clarity. On the end of each of the six arms is mounted a vertical post; these form a hexagon around which to wrap the assist device. With the configuration used, external diameters from 24 to 96 mm (vertex to vertex) can be simulated. Though independently actuated, the six actuators are currently all linked with coupling rods, giving the sim- ulator just one degree of freedom: the diameter. Therefore, it is just the gross change in ventricle volume that is represented; this is sufficient since the assist band to be tested also has just one degree of freedom (circumferential contraction) Assist device The prototype DCC assist device being tested consists of several contractile belts which are to be placed around both ventricles of the heart to form a contractile blanket. One band was tested in isolation on the heart simulator as indicated in Fig. 4. The circumference of these flexible belts is controlled using direct-drive, miniature dc motors. Compared to pneumatic actuation used in other DCC devices, the torque and position of these motors are easily controlled by computer, and are suitable for use with an implantable battery power supply. Long-term device life is being determined by endurance trials. If necessary, brushless commutation could be used to increase motor life; brushless motors are currently in use within implanted LVADs. Further details of the form and control of the device can be found in [18] Interfacing between hardware and software The key factor in an HIL simulation is the interfacing; this must be designed to suit both the hardware and software systems. A custom-made force sensor was used to record the force produced by the assist device. This comprises a thin aluminium cantilever beam structure (dimensions: 5.5 mm 12 mm) with strain gauges on both beam surfaces. The force signal from this sensor is read into the software via analogue-to-digital conversion using an interfacing card (National Instruments PCI-MIO-16E). The increase in ventricular pressure created by this force was calculated using a simple model, Eq. (3), and added to the ventricular pressures within the software model. P assist = T rh (3) Fig. 4. Plan view diagram of the heart simulator, constructed from six swingarm actuators. Key:, 6 pivot points of swing-arm actuators;, 6 posts, around which is wrapped;, flexible belt of assist device;, force sensor;, monitor unit of assist device. where T is the circumferential tension in the assist belt (= force recorded, with the current geometry), r the external radius of the ventricles, and h is the effective width of the assist band. The same assist pressure was added to both ventricles. A servo-potentiometer is used to measure the position of one swing-arm actuator, and from this, the diameter of the heart simulator is calculated. This is the feedback used to control the instantaneous diameter of the simulator. The model is paced using an ex-planted pacemaker interfaced to the computerised model. This allows synchronous in vitro/in vivo comparison to be performed in future, where

5 B.M. Hanson et al. / Medical Engineering & Physics 29 (2007) Fig. 5. Position tracking performance of the heart simulator at 50, 100, and 150 beats per minute (bpm). a pacemaker would relay real physiological pacing signals to the computer model. For the data presented herein, the pacemaker was set to a constant rate and the model could equally have been paced numerically Position control loop The physical simulator s task is to display the exact timevarying dimensions of the modelled slice through the heart, as shown in Fig. 4. The instantaneous position of the simulator is controlled in a feedback loop operating concurrently with the circulatory system model data is passed along each arrow in Fig. 1 at a loop rate of 500 Hz. This rate provides a high resolution of the CVS simulation that allows detailed identification of the effects that assistance might have. A fast rate is also desirable to reduce the delay associated with digital filters that are used to remove high-frequency electrical noise from analogue input signals. The upper limit on loop rate is in practice governed by the time required for analogue interfacing rather than model computation. A non-linear PID control algorithm is used for feedback position control of the hexagonal array of swing-arm actuators. This gives good positional accuracy in the presence of unpredictable disturbance forces from the assist device. The tracking performance of the heart simulator is shown in Fig. 5, where the simulator replicated the motion of a heart beating at 50, 100, and 150 bpm. The actual position of the simulator followed the desired position to within 0.5 mm diameter, and this was deemed sufficient accuracy for the application. 4. Experimental methods The HIL simulator allows quantification of the circulatory effects of real, physical assist. In this paper we present results that demonstrate the efficacy of the HIL testing environment, and its specific benefits. The HIL simulator was used to evaluate the performance of a prototype assist band, demonstrating the direct effect of its assistance on a model of a weakened cardiovascular system. A study of assist synchronisation was performed to attempt to determine the effect of assistance that is poorly synchronised with the heart s own efforts: when used in vivo, the device will use the heart s natural pacing signal, if available, sensed from a pacemaker. This will be exhibit beat-to-beat timing variations, and without careful control it is possible that an assist device may lose synchronisation with the natural heart. To assess this effect, a delay of up to 200 ms was imposed between the pacing signal at the start of natural systole and the onset of assist compression. The assist band was mounted to the heart simulator, as shown in Fig. 4. A weakened heart model was used, as described below. With the model operating in a steady haemodynamic state, the assist device was switched on, applying compression every beat. The CVS reached a new steady operating state after approximately 8 10 beats. Simulated physiological traces were recorded from the model over this period; the pressure within ventricles and main arteries was studied in each case, along with the ventricle volumes and cardiac output. The energy efficiency of the assist device was measured: instantaneous electrical power in to the assist device was calculated as the product of voltage and current, and mechanical power applied to the simulated ventricle was measured by multiplying the applied belt tension by the rate of change of circumference. The energy efficiency of the assist device was measured in each delayed case as the ratio of total mechanical energy (out) to electrical energy (in). This was averaged over a period of three cardiac cycles, once the CVS had reached a steady state. To simulate acute ischaemic heart disease, the contractility of both left and right ventricles was scaled down to 50% of their nominal healthy values [4]. Although autonomous nervous system (ANS) control of peripheral resistance is not included in this current model, the model s values of vascular resistances were increased manually to maintain blood pressure in the weakened condition (values in Appendix A). Other parameters, including heart rate, were unchanged; use of a software model ensures that the experimental conditions are identical for each repeated test something that would be impossible on a biological model. 5. Results Fig. 6 shows some traces from the numerical CVS simulation while undergoing testing. The assistance in this example was synchronised with the natural systolic effort. The closed-loop CVS model has shown that when one compression band is applied around both ventricles, assistance affects the systemic (left heart) and pulmonary (right heart) circulation in different ways. These effects are dis-

6 372 B.M. Hanson et al. / Medical Engineering & Physics 29 (2007) Fig. 6. Simulated physiological traces from HIL simulation. Healthy CVS shown, and weakened CVS with effect of direct compression from a prototype assist device: LV, left ventricle; RV, right ventricle; PA, pulmonary artery. cussed in more detail in [17]. The model suggests that the assistance acts to empty the right ventricle, and the increased pulmonary pressure would then tend to increase the operating volume of the left ventricle. These effects would be reduced in vivo by the body s ANS applying compensatory mechanisms. Therefore, this model is valuable in showing the direct, mechanical effects, as this allows the development of a control scheme to maintain both ventricle volumes without relying on ANS control. Assistance is seen to immediately increase the modelled blood pressure within ventricles and major arteries. Again, in vivo, ANS control would act to reduce vascular resistance to decrease the aortic blood pressure and increase the cardiac output. This model only shows the direct, mechanical effects of compression, which is sufficient for evaluating prototype assist devices. The increase in cardiac output (C.O.) as a result of assistance is shown in Table 1. The C.O. of the simulated weakened state is dramatically reduced in comparison to the healthy state, however the C.O. was then increased with assistance. The beneficial increase in C.O. is highest for synchronous assist, however a delay of up to 50 ms in assist action did not indicate a significant effect on performance. Fig. 7 shows a comparison between synchronous and asynchronous assist. In Fig. 7(a), assist is applied at beginning of systole, and a positive assist force is recorded over the period of ventricle contraction. In Fig. 7(b), the assist is delayed (by 150 ms), and a force is only recorded over part of the contraction period. The assist force continues into the isovolumetric relaxation period, which is extended as a result diastole only begins once the assist pressure has been removed. Fig. 7(b) also shows a significant peak in force during the isovolumetric period, which could be clinically important the increased contact force in diastole could restrict blood flow over the surface of the myocardium and increase the risk of further ischaemia. The efficiency of the device in converting electrical power into mechanical power is indicated by the relative magnitudes Table 1 Effects of prototype assist device with delayed onset on HIL simulation of the CVS Condition Assist Delay (ms) L.V. (E.D.V.) C.O. (l/min) Healthy None Weak None Weak Assisted Weak Assisted Weak Assisted Weak Assisted Weak Assisted Assist efficiency (%)

7 B.M. Hanson et al. / Medical Engineering & Physics 29 (2007) Fig. 7. Force and power traces recorded from heart simulator, shown with ventricle diameter: (a) synchronous assist and (b) asynchronous assist. of the electrical and mechanical traces in Fig. 7. With no delay, the efficiency recorded was approximately 9%. This is below the theoretical maximum efficiency of dc electric motors (up to 70%), however that occurs at much higher rotational speeds than used in this application. Friction in the belt and the motor s pulley system will have reduced the potential efficiency of the device. The electrical power used by the device does not change very significantly between Fig. 7(a and b), however the useful mechanical power out from the device is reduced if the assistance is applied to ventricles that have finished contracting. Table 1 shows how the efficiency reduces with increasingly asynchronous assist. Further investigation of the conversion of mechanical assist power into fluid power is recommended, taking into account the work done by the myocardium (in simulation). Preliminary investigations, as yet unpublished, have indicated that when assisted through systolic contraction, the myocardium generates a lower active component of stress. It is hoped that this could promote remodelling of the muscle. 6. Discussion and conclusions The results of HIL testing have demonstrated that applying mechanical assistance in the form of direct cardiac compression can increase blood pressure and cardiac output from a weakened heart. The experimental testing described in Sections 4 and 5 has demonstrated several of the benefits of HIL simulation identified in Section 2: compared to testing on an animal model, it would not have been possible to perform these repeated experiments, all on an identical patient model, in a short space of time, in a non-clinical setting. The HIL environment also facilitated numerical evaluation of the experiments and assessment of the device s efficiency. Compared with a purely numerical simulation, HIL simulation enabled evaluation of the effect of the real, physical performance of the prototype assist device. This included the electromechanical properties of the motor and the mechanics of tension transmission via the flexible band to the heart surface. The efficacy of the prototype assist device was assessed as the dimensions of the ventricles changed over the cardiac cycle. The numerical CVS model used in HIL simulation can be further developed as required by future investigations. The short computational time required for the current model did not suggest that future models will have to be greatly simplified, especially given the increasingly available computing power. Ferrari et al. [4] found that LabVIEW running under Microsoft Windows provided limited time for computation, and suggested LabVIEW Real Time. For this apparatus we are also investigating LabVIEW Real Time, installed on a conventional PC. The assist device bands each have one degree of freedom to contract circumferentially however, this form of assist has been shown to produce differing effects on the two ventricles. Future work will consider ventricle-specific assist. The simulator can be enhanced by removing the mechanical links between the swing-arms, and providing separate position controllers for each of the six actuators. Then the different compliances of the right and left ventricles can be represented, as can regional wall motion abnormalities. Given the proven benefits of this technique, it is likely that this hardware-in-the-loop technique will be suitable for evaluation of prostheses and interaction with other biological systems. In such applications, it is the interface between hardware and software that will require the most attention. In this instance that interface took the form of a heart simulator; taking the example of an LVAD, the interface would necessarily involve fluid and may take the form of a precision servo-controlled displacement pump, with pressure transducers to measure the instantaneous pressure rise over the LVAD. This pressure would be fed into a circulation model similar to that of Fig. 2, with the addition of a branch through the LVAD. That flow would then be presented physically to the LVAD via the servo-controlled displacement pump. Arterial

8 374 B.M. Hanson et al. / Medical Engineering & Physics 29 (2007) grafts, stents, and valves could be evaluated in a very similar manner, and the HIL technique may also be applied to musculoskeletal prostheses using software models of muscle function. Appendix A Some cardiovascular system (CVS) parameters used: Parameter Healthy condition Ischaemic heart disease Peripheral venous resistance (mmhg/ml/s) Pulmonary resistance (mmhg/ml/s) LV contractility scaling (dimensionless) RV contractility scaling (dimensionless) 4 2 Heart rate (bpm) References [1] Oz MC, Artrip JH, Burkhoff D. Direct cardiac compression devices. J Heart Lung Transplant 2002;21(10): [2] Macnair R, Underwood MJ, Angelini GD. Biomaterials and cardiovascular devices. Proc Inst Mech Eng 1998;212(Part II): [3] Wu Y, Allaire P, Tao G, Wood H, Olsen D, Tribble C. An advanced physiological controller design for a left ventricular assist device to prevent left ventricular collapse. Artif Organs 2003;27: [4] Ferrari G, Kozarski M, De Lazzari C, Gorczynska K, Mimmo R, Guaragno M, et al. Modelling of cardiovascular system: development of a hybrid (numerical-physical) model. Int J Artif Organs 2003;26(12): [5] Urbaszek A, Schaldach M. A numerical heart and circulation model to simulate hemodynamics for rate-responsive pacing. In: Power H, editor. Bio-fluid mechanics. Southampton/Boston: Computational Mechanics Publications; p [6] Pillon M, Duffour H, Jufer M. In vitro experiments: circulatory assist device interaction with a virtual cardiovascular system. In: Proceedings of Annual International Conference IEEE-EMBS, vol p [7] Wu Y, Allaire P, Tao G, Liu Y. In-vitro test of an adaptive flow controller for a continuous flow LVAD. In: Proceedings of American control conference p [8] Kozarski M, Ferrari G, Clemente F, Gorczynska K, De Lazzari C, Darowski M, et al. A hybrid mock circulatory system: development and testing of an electro-hydraulic impedance simulator. Int J Artif Organs 2003;26(1): [9] Kawaguchi O, Goto Y, Ohgoshi Y, Yaku H, Murase M, Suga H. Dynamic cardiac compression improves contractile efficiency of the heart. J Thorac Cardiovasc Surg 1997;113(5): [10] Sunagawa K, Burkhoff D, Lim KO, Sagawa K. Impedance loading servo pump system for excised canine ventricle. Am J Physiol 1982;243(2):H [11] Isermann R, Schaffnit J, Sinsel S. Hardware-in-the-loop simulation for the design and testing of engine-control systems. Control Eng Pract 1999;7: [12] Ferreira JA, Gomes Almeida F, Quintas MR, Estima de Oliveira JP. Hybrid models for hardware-in-the-loop simulation of hydraulic systems. Proc Instn Mech Engrs Part 1: J Syst Control Eng 2004;218: [13] Hanselmann H. Hardware-in-the-loop simulation testing and its integration into a CACSD toolset. In: IEEE international symposium of computer-aided control system design [14] Melchior FM, Srinivasan RS, Charles JB. Mathematical modelling of human cardiovascular system for simulation of orthostatic response. Am J Physiol (Heart Circ Physiol) 1992;262(31): [15] Ha R, Qian J, Wang D, Zwischenberger JB, Bidani A, Clark JW. A closed-loop model of the ovine cardiovascular system. In: Proceedings of 26th annual international conference of the IEEE EMBS p [16] Olansen JB, Clark JW, Khoury D, Ghorbel F, Bidani A. A closed-loop model of the canine cardiovascular system that includes ventricular interaction. Comput Biomed Res 2000;33: [17] Hanson BM, Levesley MC, Watterson K, Walker PG. Simulation of the human cardiovascular system for real-time interaction with an assist device. In: Proceedings of 27th annual international conference of the IEEE EMBS, (EMBC 05) [18] Hanson BM, Richardson RC, Davies GLJ, Watterson K, Levesley MC, Walker PG. Control of a non-blood contacting cardiac assist device. In: Proceedings of IASTED international conference on biomedical engineering BioMED p

CONTROL OF A NON-BLOOD CONTACTING CARDIAC ASSIST DEVICE

CONTROL OF A NON-BLOOD CONTACTING CARDIAC ASSIST DEVICE CONTROL OF A NON-BLOOD CONTACTING CARDIAC ASSIST DEVICE B.M.Hanson 1, R.C.Richardson 2, G.L.J.Davies 1, K. Watterson 3, M.C.Levesley 1, P.G.Walker 1. 1 School of Mechanical Engineering, University of Leeds,

More information

Application of Electroactive Polymers to Cardiovascular Flows

Application of Electroactive Polymers to Cardiovascular Flows Application of Electroactive Polymers to Cardiovascular Flows Dave Morgan Department of Mechanical Engineering, Concordia University, Montreal, Canada Abstract The ability of electroactive polymers (EAPs)

More information

Physiological Control of Left Ventricular Assist Devices Based on Gradient of Flow (1)

Physiological Control of Left Ventricular Assist Devices Based on Gradient of Flow (1) 2005 American Control Conference June 8-10, 2005. Portland, OR, USA FrA13.4 Physiological Control of Left Ventricular Assist Devices Based on Gradient of Flow (1) Shaohui Chen (2), James F. Antaki (3),

More information

ENT 318/3 Artificial Organs

ENT 318/3 Artificial Organs ENT 318/3 Artificial Organs Modeling of cardiovascular system and VAD Lecturer Ahmad Nasrul bin Norali 1 What is modeling and why we need it? In designing product, sometimes we have to make sure that the

More information

PROBLEM SET 2. Assigned: February 10, 2004 Due: February 19, 2004

PROBLEM SET 2. Assigned: February 10, 2004 Due: February 19, 2004 Harvard-MIT Division of Health Sciences and Technology HST.542J: Quantitative Physiology: Organ Transport Systems Instructors: Roger Mark and Jose Venegas MASSACHUSETTS INSTITUTE OF TECHNOLOGY Departments

More information

EI2311 BIOMEDICAL INSTRUMENTATION

EI2311 BIOMEDICAL INSTRUMENTATION 66 EI2311 BIOMEDICAL INSTRUMENTATION 1. What is meant by cell? UNIT I PHYSIOLOGY AND TRANSDUCERS The basic living unit of the body is cell. The function of organs and other structure of the body is understood

More information

Application of a PExSim for modeling a POLVAD artificial heart and the human circulatory system with left ventricle assistance

Application of a PExSim for modeling a POLVAD artificial heart and the human circulatory system with left ventricle assistance Pol J Med Phys Eng 2010;16(2):107-124. PL ISSN 1425-4689 doi: 10.2478/v10013-010-0010-z website: http://www.pjmpe.waw.pl Alicja Siewnicka, Bartlomiej Fajdek, Krzysztof Janiszowski Application of a PExSim

More information

Principles of Biomedical Systems & Devices. Lecture 8: Cardiovascular Dynamics Dr. Maria Tahamont

Principles of Biomedical Systems & Devices. Lecture 8: Cardiovascular Dynamics Dr. Maria Tahamont Principles of Biomedical Systems & Devices Lecture 8: Cardiovascular Dynamics Dr. Maria Tahamont Review of Cardiac Anatomy Four chambers Two atria-receive blood from the vena cave and pulmonary veins Two

More information

BME 5742 Bio-Systems Modeling and Control. Lecture 41 Heart & Blood Circulation Heart Function Basics

BME 5742 Bio-Systems Modeling and Control. Lecture 41 Heart & Blood Circulation Heart Function Basics BME 5742 Bio-Systems Modeling and Control Lecture 41 Heart & Blood Circulation Heart Function Basics Dr. Zvi Roth (FAU) 1 Pumps A pump is a device that accepts fluid at a low pressure P 1 and outputs the

More information

SymBioSys Exercise 2 Cardiac Function Revised and reformatted by C. S. Tritt, Ph.D. Last updated March 20, 2006

SymBioSys Exercise 2 Cardiac Function Revised and reformatted by C. S. Tritt, Ph.D. Last updated March 20, 2006 SymBioSys Exercise 2 Cardiac Function Revised and reformatted by C. S. Tritt, Ph.D. Last updated March 20, 2006 The goal of this exercise to explore the behavior of the heart as a mechanical pump. For

More information

Chapter 9, Part 2. Cardiocirculatory Adjustments to Exercise

Chapter 9, Part 2. Cardiocirculatory Adjustments to Exercise Chapter 9, Part 2 Cardiocirculatory Adjustments to Exercise Electrical Activity of the Heart Contraction of the heart depends on electrical stimulation of the myocardium Impulse is initiated in the right

More information

Cardiovascular system

Cardiovascular system BIO 301 Human Physiology Cardiovascular system The Cardiovascular System: consists of the heart plus all the blood vessels transports blood to all parts of the body in two 'circulations': pulmonary (lungs)

More information

2.6 Cardiovascular Computer Simulation

2.6 Cardiovascular Computer Simulation 2.6 Cardiovascular Computer Simulation ROOM 23G22 Contents 1. INTRODUCTION... 4 1.1. GENERAL REMARKS... 4 1.2. LEARNING GOALS... 4 1.3. PHYSIOLOGICAL PARAMETERS... 5 1.4. GLOSSARY... 5 2. USING THE COMPUTER

More information

AS Level OCR Cardiovascular System

AS Level OCR Cardiovascular System AS Level OCR Cardiovascular System Learning Objectives The link between the Cardiac Cycle and the Conduction system of the heart. The relationship between Stroke volume, Heart rate and Cardiac Output.

More information

IP: Regulation of Cardiac Output

IP: Regulation of Cardiac Output ANP 1105D Winter 2013 Assignment 9: The Heart, part 2: Chap... Assignment 9: The Heart, part 2: Chapter 18 Signed in as Alex Sokolowski Help Close Resources Due: 11:59pm on Monday, March 25, 2013 Note:

More information

QUIZ 1. Tuesday, March 2, 2004

QUIZ 1. Tuesday, March 2, 2004 Harvard-MIT Division of Health Sciences and Technology HST.542J: Quantitative Physiology: Organ Transport Systems Instructors: Roger Mark and Jose Venegas MASSACHUSETTS INSTITUTE OF TECHNOLOGY Departments

More information

Effect of physiological heart rate changes on left ventricular dimensions and mitral blood flow velocities in the normal fetus

Effect of physiological heart rate changes on left ventricular dimensions and mitral blood flow velocities in the normal fetus ELSEVIER Early Human Development 40 (1995) 109-114 Effect of physiological heart rate changes on left ventricular dimensions and mitral blood flow velocities in the normal fetus P.B. Tsyvian a, K.V. Malkin

More information

Introducing a Hardware-in-the-Loop Simulation of the Cardiovascular System

Introducing a Hardware-in-the-Loop Simulation of the Cardiovascular System The Fourth IEEE RAS/EMBS International Conference on Biomedical Robotics and Biomechatronics Roma, Italy. June 24-27, 2012 Introducing a Hardware-in-the-Loop Simulation of the Cardiovascular System Ali

More information

THE CARDIOVASCULAR SYSTEM

THE CARDIOVASCULAR SYSTEM THE CARDIOVASCULAR SYSTEM AND RESPONSES TO EXERCISE Mr. S. Kelly PSK 4U North Grenville DHS THE HEART: A REVIEW Cardiac muscle = myocardium Heart divided into two sides, 4 chambers (L & R) RS: pulmonary

More information

Translating In-vitro and Computational Multi-scale Models for the Vascular Surgeon

Translating In-vitro and Computational Multi-scale Models for the Vascular Surgeon Translating In-vitro and Computational Multi-scale Models for the Vascular Surgeon Benjamin A. Eslahpazir MS, BSME 1, Kyle W. Beggs 2, Zoe M. Lauters 2 Alain J. Kassab PhD 2 William M. DeCampli MD, PhD

More information

A MINIMAL CARDIOVASCULAR SYSTEM HAEMODYNAMIC MODEL FOR RAPID DIAGNOSTIC ASSITANCE

A MINIMAL CARDIOVASCULAR SYSTEM HAEMODYNAMIC MODEL FOR RAPID DIAGNOSTIC ASSITANCE A MINIMAL CARDIOASCULAR SYSTEM HAEMODYNAMIC MODEL FOR RAID DIAGNOSTIC ASSITANCE Bram W. Smith 1, J. Geoffrey Chase 2, Geoffry M. Shaw 3 and Roger Nokes 2 Department of Mechanical Engineering, University

More information

Cardiovascular System

Cardiovascular System Cardiovascular System The Heart Cardiovascular System The Heart Overview What does the heart do? By timed muscular contractions creates pressure gradients blood moves then from high pressure to low pressure

More information

Cardiac Physiology an Overview

Cardiac Physiology an Overview Cardiac Physiology an Overview Dr L J Solomon Department of Paediatrics and Child Health School of Medicine Faculty of Health Sciences University of the Free State and PICU Universitas Academic Hospital

More information

Electrical Conduction

Electrical Conduction Sinoatrial (SA) node Electrical Conduction Sets the pace of the heartbeat at 70 bpm AV node (50 bpm) and Purkinje fibers (25 40 bpm) can act as pacemakers under some conditions Internodal pathway from

More information

Patient-specific modeling of heart and circulation

Patient-specific modeling of heart and circulation Patient-specific modeling of heart and circulation Theo Arts t.arts@bf.unimaas.nl Maastricht University, *Maastricht University Hospital **University of Tecnology, Eindhoven The Netherlands *Tammo Delhaas,

More information

Chapter 13 The Cardiovascular System: Cardiac Function

Chapter 13 The Cardiovascular System: Cardiac Function Chapter 13 The Cardiovascular System: Cardiac Function Overview of the Cardiovascular System The Path of Blood Flow through the Heart and Vasculature Anatomy of the Heart Electrical Activity of the Heart

More information

A PARADIGM FOR QUANTIFYING VENTRICULAR CONTRACTION USA

A PARADIGM FOR QUANTIFYING VENTRICULAR CONTRACTION USA CELLULAR & MOLECULAR BIOLOGY LETTERS 331 A PARADIGM FOR QUANTIFYING VENTRICULAR CONTRACTION JOSEPH L. PALLADINO 1 and ABRAHAM NOORDERGRAAF 2 1 Department of Engineering, Trinity College, Hartford, CT,

More information

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model Instrumentation and Control Department, JSS Academy of Technical Education, Noida (U.P.), India Abstract- The non-invasive

More information

Cardiovascular Physiology. Heart Physiology. Introduction. The heart. Electrophysiology of the heart

Cardiovascular Physiology. Heart Physiology. Introduction. The heart. Electrophysiology of the heart Cardiovascular Physiology Heart Physiology Introduction The cardiovascular system consists of the heart and two vascular systems, the systemic and pulmonary circulations. The heart pumps blood through

More information

CIRCULATORY PHYSIOLOGY SECTION 3: CARDIAC MECHANICS *

CIRCULATORY PHYSIOLOGY SECTION 3: CARDIAC MECHANICS * CIRCULATORY PHYSIOLOGY SECTION 3: CARDIAC MECHANICS * Summary: In this we will look at the mechanical action of the heart as a pump and how it is able to adjust its stroke volume as a result of the Frank-

More information

MASSACHUSETTS INSTITUTE OF TECHNOLOGY

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Harvard-MIT Division of Health Sciences and Technology HST.542J: Quantitative Physiology: Organ Transport Systems Instructors: Roger Mark and Jose Venegas MASSACHUSETTS INSTITUTE OF TECHNOLOGY Departments

More information

Left atrial function. Aliakbar Arvandi MD

Left atrial function. Aliakbar Arvandi MD In the clinic Left atrial function Abstract The left atrium (LA) is a left posterior cardiac chamber which is located adjacent to the esophagus. It is separated from the right atrium by the inter-atrial

More information

From PV loop to Starling curve. S Magder Division of Critical Care, McGill University Health Centre

From PV loop to Starling curve. S Magder Division of Critical Care, McGill University Health Centre From PV loop to Starling curve S Magder Division of Critical Care, McGill University Health Centre Otto Frank 1890 s Frank-Starling Relationship ( The Law of the Heart ) The greater the initial stretch

More information

Practice Exercises for the Cardiovascular System

Practice Exercises for the Cardiovascular System Practice Exercises for the Cardiovascular System On the diagram below, color the oxygen-rich blood red and the oxygen-poor blood blue. Label the parts: Continued on the next page... Label the parts on

More information

The relation between stroke work and end-diastolic volume in the ventricles

The relation between stroke work and end-diastolic volume in the ventricles Modelling in Medicine and Biology VI 123 The relation between stroke work and end-diastolic volume in the ventricles R. M. Shoucri Department of Mathematics and Computer Science, Royal Military College

More information

Dynamic Hollow Cylinder System Proposal

Dynamic Hollow Cylinder System Proposal Dynamic Hollow Cylinder System Proposal from Wykeham Farrance Limited Weston Road Slough SL1 4HW England Tel:+44 (0)1753 571241 Fax: +44 (0)1753 811313 E-mail: sales@wfi.co.uk Advanced Soils Testing 2

More information

Circulation: Chapter 25. Cardiac Output. The Mammalian Heart Fig Right side of the heart

Circulation: Chapter 25. Cardiac Output. The Mammalian Heart Fig Right side of the heart Circulation: Chapter 25 1. Limits of Diffusion A. Small organisms use diffusion B. rapid over small distances 2. Most animals have circulatory systems A. Blood B. Pump (Heart) or propulsive structures

More information

Circulatory system of mammals

Circulatory system of mammals Circulatory system of mammals Explain the cardiac cycle and its initiation Discuss the internal factors that control heart action Blood flows through the heart as a result of pressure differences Blood

More information

Hemodynamic Assessment. Assessment of Systolic Function Doppler Hemodynamics

Hemodynamic Assessment. Assessment of Systolic Function Doppler Hemodynamics Hemodynamic Assessment Matt M. Umland, RDCS, FASE Aurora Medical Group Milwaukee, WI Assessment of Systolic Function Doppler Hemodynamics Stroke Volume Cardiac Output Cardiac Index Tei Index/Index of myocardial

More information

The Cardiovascular System

The Cardiovascular System The Cardiovascular System The Cardiovascular System A closed system of the heart and blood vessels The heart pumps blood Blood vessels allow blood to circulate to all parts of the body The function of

More information

Evaluation of Native Left Ventricular Function During Mechanical Circulatory Support.: Theoretical Basis and Clinical Limitations

Evaluation of Native Left Ventricular Function During Mechanical Circulatory Support.: Theoretical Basis and Clinical Limitations Review Evaluation of Native Left Ventricular Function During Mechanical Circulatory Support.: Theoretical Basis and Clinical Limitations Tohru Sakamoto, MD, PhD Left ventricular function on patients with

More information

Outline. Electrical Activity of the Human Heart. What is the Heart? The Heart as a Pump. Anatomy of the Heart. The Hard Work

Outline. Electrical Activity of the Human Heart. What is the Heart? The Heart as a Pump. Anatomy of the Heart. The Hard Work Electrical Activity of the Human Heart Oguz Poroy, PhD Assistant Professor Department of Biomedical Engineering The University of Iowa Outline Basic Facts about the Heart Heart Chambers and Heart s The

More information

The Cardiovascular System

The Cardiovascular System Essentials of Human Anatomy & Physiology Elaine N. Marieb Slides 11.1 11.19 Seventh Edition Chapter 11 The Cardiovascular System Functions of the Cardiovascular system Function of the heart: to pump blood

More information

CONSIDERATIONS ABOUT THE LUMPED PARAMETER WINDKESSEL MODEL APPLICATIVITY IN THE CARDIOVASCULAR SYSTEM STRUCTURE

CONSIDERATIONS ABOUT THE LUMPED PARAMETER WINDKESSEL MODEL APPLICATIVITY IN THE CARDIOVASCULAR SYSTEM STRUCTURE CONSIDERATIONS ABOUT THE LUMPED PARAMETER WINDKESSEL MODEL APPLICATIVITY IN THE CARDIOVASCULAR SYSTEM STRUCTURE VASILE MANOLIU Electrical Engineering Faculty, POLITEHNICA University of Bucharest, Splaiul

More information

The ADL Force 5 configured for knee testing

The ADL Force 5 configured for knee testing The ADL Force 5 configured for knee testing Description The AMTI Force 5 is a versatile simulator that can be configured for a variety of single and multiaxis tests, such as joint simulation, product testing,

More information

GIGA - In Silico Medicine, University of Liege, Belgium, 2

GIGA - In Silico Medicine, University of Liege, Belgium, 2 S. Kosta 1,*, A. Pironet 1, J.A. Negroni 2, E.C. Lascano 2, P.C. Dauby 1 1 GIGA - In Silico Medicine, University of Liege, Belgium, 2 Department of Comparative Cellular and Molecular Biology, Favaloro

More information

How does the heart pump? From sarcomere to ejection volume

How does the heart pump? From sarcomere to ejection volume How does the heart pump? From sarcomere to ejection volume Piet Claus Cardiovascular Imaging and Dynamics Department of Cardiovascular Diseases University Leuven, Leuven, Belgium Course on deformation

More information

*Generating blood pressure *Routing blood: separates. *Ensuring one-way blood. *Regulating blood supply *Changes in contraction

*Generating blood pressure *Routing blood: separates. *Ensuring one-way blood. *Regulating blood supply *Changes in contraction *Generating blood pressure *Routing blood: separates pulmonary and systemic circulations *Ensuring one-way blood flow: valves *Regulating blood supply *Changes in contraction rate and force match blood

More information

Heart. Structure Physiology of blood pressure and heartbeat

Heart. Structure Physiology of blood pressure and heartbeat Heart Structure Physiology of blood pressure and heartbeat Location and Anatomy Location and Anatomy Pericardial cavity: surrounds, isolates, and anchors heart Parietal pericardium lined with serous membrane

More information

Lab 16. The Cardiovascular System Heart and Blood Vessels. Laboratory Objectives

Lab 16. The Cardiovascular System Heart and Blood Vessels. Laboratory Objectives Lab 16 The Cardiovascular System Heart and Blood Vessels Laboratory Objectives Describe the anatomical structures of the heart to include the pericardium, chambers, valves, and major vessels. Describe

More information

Section 5.1 The heart and heart disease

Section 5.1 The heart and heart disease Section 5.1 The heart and heart disease Mammals are too large to rely on diffusion. They need a circulatory system to move substances around the body. Blood moves down pressure gradients, from high to

More information

Eindhoven University of Technology. Exam Modeling Cardiac Function (8W160)

Eindhoven University of Technology. Exam Modeling Cardiac Function (8W160) Eindhoven University of Technology department of Biomedical Engineering group Cardiovascular Biomechanics Exam Modeling Cardiac Function (8W160) January 21, 2011, 14.00 17.00 h This exam consists of 6

More information

During exercise the heart rate is 190 bpm and the stroke volume is 115 ml/beat. What is the cardiac output?

During exercise the heart rate is 190 bpm and the stroke volume is 115 ml/beat. What is the cardiac output? The Cardiovascular System Part III: Heart Outline of class lecture After studying part I of this chapter you should be able to: 1. Be able to calculate cardiac output (CO) be able to define heart rate

More information

Suction Detection And Feedback Control For The Rotary Left Ventricular Assist Device

Suction Detection And Feedback Control For The Rotary Left Ventricular Assist Device University of Central Florida Electronic Theses and Dissertations Doctoral Dissertation (Open Access) Suction Detection And Feedback Control For The Rotary Left Ventricular Assist Device 2013 Yu Wang University

More information

McHenry Western Lake County EMS System Paramedic, EMT-B and PHRN Optional Continuing Education 2018 #12 Understanding Preload and Afterload

McHenry Western Lake County EMS System Paramedic, EMT-B and PHRN Optional Continuing Education 2018 #12 Understanding Preload and Afterload McHenry Western Lake County EMS System Paramedic, EMT-B and PHRN Optional Continuing Education 2018 #12 Understanding Preload and Afterload Cardiac output (CO) represents the volume of blood that is delivered

More information

The circulatory system

The circulatory system Introduction to Physiology (Course # 72336) 1 הלב עקרונות בסיסיים (הכנה למעבדת לב) Adi Mizrahi mizrahia@cc.huji.ac.il Textbook Chapter 12 2 The circulatory system To the heart Away from the heart 3 L 2.5

More information

Introduction. Cardiac Imaging Modalities MRI. Overview. MRI (Continued) MRI (Continued) Arnaud Bistoquet 12/19/03

Introduction. Cardiac Imaging Modalities MRI. Overview. MRI (Continued) MRI (Continued) Arnaud Bistoquet 12/19/03 Introduction Cardiac Imaging Modalities Arnaud Bistoquet 12/19/03 Coronary heart disease: the vessels that supply oxygen-carrying blood to the heart, become narrowed and unable to carry a normal amount

More information

COMPUTER PLAY IN EDUCATIONAL THERAPY FOR CHILDREN WITH STUTTERING PROBLEM: HARDWARE SETUP AND INTERVENTION

COMPUTER PLAY IN EDUCATIONAL THERAPY FOR CHILDREN WITH STUTTERING PROBLEM: HARDWARE SETUP AND INTERVENTION 034 - Proceeding of the Global Summit on Education (GSE2013) COMPUTER PLAY IN EDUCATIONAL THERAPY FOR CHILDREN WITH STUTTERING PROBLEM: HARDWARE SETUP AND INTERVENTION ABSTRACT Nur Azah Hamzaid, Ammar

More information

The Arterial and Venous Systems Roland Pittman, Ph.D.

The Arterial and Venous Systems Roland Pittman, Ph.D. The Arterial and Venous Systems Roland Pittman, Ph.D. OBJECTIVES: 1. State the primary characteristics of the arterial and venous systems. 2. Describe the elastic properties of arteries in terms of pressure,

More information

Blood Pressure Laboratory

Blood Pressure Laboratory Introduction The blood that circulates throughout the body maintains a flow and pressure. The nervous system can change the flow and pressure based on the particular needs at a given time. For example,

More information

On the feasibility of speckle reduction in echocardiography using strain compounding

On the feasibility of speckle reduction in echocardiography using strain compounding Title On the feasibility of speckle reduction in echocardiography using strain compounding Author(s) Guo, Y; Lee, W Citation The 2014 IEEE International Ultrasonics Symposium (IUS 2014), Chicago, IL.,

More information

Heart Pump and Cardiac Cycle. Faisal I. Mohammed, MD, PhD

Heart Pump and Cardiac Cycle. Faisal I. Mohammed, MD, PhD Heart Pump and Cardiac Cycle Faisal I. Mohammed, MD, PhD 1 Objectives To understand the volume, mechanical, pressure and electrical changes during the cardiac cycle To understand the inter-relationship

More information

CHAPTER 4 Basic Physiological Principles

CHAPTER 4 Basic Physiological Principles 4-1 CHAPTER 4 Basic Physiological Principles Now that we have a working anatomical knowledge of the heart and circulatory system, we will next develop a functional and quantitative knowledge of the cardiovascular

More information

Achieving Physiologic Perfusion with Ventricular Assist Devices: Comparison of Control Strategies

Achieving Physiologic Perfusion with Ventricular Assist Devices: Comparison of Control Strategies 25 American Control Conference June 8-1, 25. Portland, OR, USA FrA13.3 Achieving Physiologic Perfusion with Ventricular Assist Devices: Comparison of Control Strategies Guruprasad Giridharan, George Pantalos,

More information

SUPPLEMENTAL MATERIAL

SUPPLEMENTAL MATERIAL SUPPLEMENTAL MATERIAL Supplemental methods Pericardium In several studies, it has been shown that the pericardium significantly modulates ventricular interaction. 1-4 Since ventricular interaction has

More information

Lab #3: Electrocardiogram (ECG / EKG)

Lab #3: Electrocardiogram (ECG / EKG) Lab #3: Electrocardiogram (ECG / EKG) An introduction to the recording and analysis of cardiac activity Introduction The beating of the heart is triggered by an electrical signal from the pacemaker. The

More information

Introduction to Physiology (Course # 72336) 1. Adi Mizrahi Textbook Chapter 12

Introduction to Physiology (Course # 72336) 1. Adi Mizrahi Textbook Chapter 12 Introduction to Physiology (Course # 72336) 1 עקרונות בסיסיים (הכנה למעבדת לב) הלב Adi Mizrahi mizrahia@cc.huji.ac.il Textbook Chapter 12 2 The circulatory system To the heart Away from the heart 3 L 2.5

More information

Pro 01: Development and In Vitro Evaluation of a Hydraulic System for a Mini axial Flow Blood Pump

Pro 01: Development and In Vitro Evaluation of a Hydraulic System for a Mini axial Flow Blood Pump Pro 01: Development and In Vitro Evaluation of a Hydraulic System for a Mini axial Flow Blood Pump Background: Fully implantable rotary blood pumps are used for mechanical circulatory support of heart

More information

The Cardiac Cycle Clive M. Baumgarten, Ph.D.

The Cardiac Cycle Clive M. Baumgarten, Ph.D. The Cardiac Cycle Clive M. Baumgarten, Ph.D. OBJECTIVES: 1. Describe periods comprising cardiac cycle and events within each period 2. Describe the temporal relationships between pressure, blood flow,

More information

Material characterization of HeartPrint models and comparison with arterial tissue properties

Material characterization of HeartPrint models and comparison with arterial tissue properties Material characterization of HeartPrint models and comparison with arterial tissue properties Over the years, catheter-based interventions have gained popularity for the treatment of cardiovascular diseases

More information

Ventricular Assisting Devices in the Cathlab. Unrestricted

Ventricular Assisting Devices in the Cathlab. Unrestricted Ventricular Assisting Devices in the Cathlab Unrestricted What is a VAD? A single system device that is surgically attached to the left ventricle of the heart and to the aorta for left ventricular support

More information

SIKLUS JANTUNG. Rahmatina B. Herman

SIKLUS JANTUNG. Rahmatina B. Herman SIKLUS JANTUNG Rahmatina B. Herman The Cardiac Cycle Definition: The cardiac events that occur from the beginning of one heartbeat to the beginning of the next The cardiac cycle consists of: - Diastole

More information

Ventricular Assist Device. Lauren Bartlett 10/5/16 BME 281, section 1

Ventricular Assist Device. Lauren Bartlett 10/5/16 BME 281, section 1 Ventricular Assist Device Lauren Bartlett 10/5/16 BME 281, section 1 What is a Ventricular Assist Device (VAD)? Electromechanical device for assisting cardiac circulation Used to partially or completely

More information

PD233: Design of Biomedical Devices and Systems

PD233: Design of Biomedical Devices and Systems PD233: Design of Biomedical Devices and Systems (Lecture-7 Biopotentials- 2) Dr. Manish Arora CPDM, IISc Course Website: http://cpdm.iisc.ac.in/utsaah/courses/ Electromyogram (EMG) Skeletal muscles are

More information

Lab 4: Introduction to Physiological Measurements - Cardiovascular

Lab 4: Introduction to Physiological Measurements - Cardiovascular Lab 4: Introduction to Physiological Measurements - Cardiovascular INTRODUCTION: This lab will demonstrate cardiovascular measurements by creating an ECG with instruments used in previous labs. Students

More information

THE CARDIOVASCULAR SYSTEM. Heart 2

THE CARDIOVASCULAR SYSTEM. Heart 2 THE CARDIOVASCULAR SYSTEM Heart 2 PROPERTIES OF CARDIAC MUSCLE Cardiac muscle Striated Short Wide Branched Interconnected Skeletal muscle Striated Long Narrow Cylindrical PROPERTIES OF CARDIAC MUSCLE Intercalated

More information

Anatomy Review: The Heart Graphics are used with permission of A.D.A.M. Software, Inc. and Benjamin/Cummings Publishing Co.

Anatomy Review: The Heart Graphics are used with permission of A.D.A.M. Software, Inc. and Benjamin/Cummings Publishing Co. Anatomy Review: The Heart Graphics are used with permission of A.D.A.M. Software, Inc. and Benjamin/Cummings Publishing Co. Anatomy Views Label the diagrams of the heart below: Interactive Physiology Study

More information

DETECTION OF HEART ABNORMALITIES USING LABVIEW

DETECTION OF HEART ABNORMALITIES USING LABVIEW IASET: International Journal of Electronics and Communication Engineering (IJECE) ISSN (P): 2278-9901; ISSN (E): 2278-991X Vol. 5, Issue 4, Jun Jul 2016; 15-22 IASET DETECTION OF HEART ABNORMALITIES USING

More information

Evolutionary origins of the right ventricle. S Magder Department of Critical Care, McGill University Health Centre

Evolutionary origins of the right ventricle. S Magder Department of Critical Care, McGill University Health Centre Evolutionary origins of the right ventricle S Magder Department of Critical Care, McGill University Health Centre Fully separated four chamber heart only evolved in birds and mammals What are the evolutionary

More information

Modeling Unloading of the Left Ventricle by the Levitronix CentriMag LVAS using a Cardiovascular Simulator

Modeling Unloading of the Left Ventricle by the Levitronix CentriMag LVAS using a Cardiovascular Simulator Modeling Unloading of the Left Ventricle by the Levitronix CentriMag LVAS using a Cardiovascular Simulator BMTE 06.17 Lieke Cox Sandra Loerakker Internship at the Texas Heart Institute, Houston, USA Supervisors

More information

The Cardiovascular System

The Cardiovascular System Essentials of Human Anatomy & Physiology Elaine N. Marieb Seventh Edition Chapter 11 The Cardiovascular System Slides 11.1 11.19 Lecture Slides in PowerPoint by Jerry L. Cook The Cardiovascular System

More information

Heart. Heart 2-Tunica media: middle layer (media ='middle') muscle fibers (smooth or cardiac).

Heart. Heart 2-Tunica media: middle layer (media ='middle') muscle fibers (smooth or cardiac). t. innermost lumenal General Circulatory system heart and blood vessels walls have 3 layers (inside to outside) 1-Tunica interna: aka tunica intima layer--lumenal layer epithelium--endothelium simple squamous

More information

Mechanics of Cath Lab Support Devices

Mechanics of Cath Lab Support Devices Mechanics of Cath Lab Support Devices Issam D. Moussa, MD Chief Medical Officer First Coast Cardiovascular Institute, Jacksonville, FL Professor of Medicine, UCF, Orlando, FL None DISCLOSURE Percutaneous

More information

Health Science 20 Circulatory System Notes

Health Science 20 Circulatory System Notes Health Science 20 Circulatory System Notes Functions of the Circulatory System The circulatory system functions mainly as the body s transport system. It transports: o Oxygen o Nutrients o Cell waste o

More information

Impedance Cardiography (ICG) Method, Technology and Validity

Impedance Cardiography (ICG) Method, Technology and Validity Method, Technology and Validity Hemodynamic Basics Cardiovascular System Cardiac Output (CO) Mean arterial pressure (MAP) Variable resistance (SVR) Aortic valve Left ventricle Elastic arteries / Aorta

More information

Age-related changes in cardiovascular system. Dr. Rehab Gwada

Age-related changes in cardiovascular system. Dr. Rehab Gwada Age-related changes in cardiovascular system Dr. Rehab Gwada Objectives explain the main structural and functional changes in cardiovascular system associated with normal aging Introduction aging results

More information

The Cardiovascular System

The Cardiovascular System 11 PART A The Cardiovascular System PowerPoint Lecture Slide Presentation by Jerry L. Cook, Sam Houston University ESSENTIALS OF HUMAN ANATOMY & PHYSIOLOGY EIGHTH EDITION ELAINE N. MARIEB The Cardiovascular

More information

The Heart 1 of 34 Boardworks Ltd 2012

The Heart 1 of 34 Boardworks Ltd 2012 The Heart 1 of 34 Boardworks Ltd 2012 2 of 34 Boardworks Ltd 2012 What does the heart do? 3 of 34 Boardworks Ltd 2012 The heart is a muscular organ located in the thorax. It pumps blood continuously around

More information

The cardiovascular system is composed of a pump the heart and blood

The cardiovascular system is composed of a pump the heart and blood 5 E X E R C I S E Cardiovascular Dynamics O B J E C T I V E S 1. To understand the relationships among blood flow, pressure gradient, and resistance 2. To define resistance and describe the main factors

More information

Advanced imaging of the left atrium - strain, CT, 3D, MRI -

Advanced imaging of the left atrium - strain, CT, 3D, MRI - Advanced imaging of the left atrium - strain, CT, 3D, MRI - Monica Rosca, MD Carol Davila University of Medicine and Pharmacy, Bucharest, Romania Declaration of interest: I have nothing to declare Case

More information

Stretching Cardiac Myocytes: A Finite Element Model of Cardiac Tissue

Stretching Cardiac Myocytes: A Finite Element Model of Cardiac Tissue Megan McCain ES240 FEM Final Project December 19, 2006 Stretching Cardiac Myocytes: A Finite Element Model of Cardiac Tissue Cardiac myocytes are the cells that constitute the working muscle of the heart.

More information

CIRCULATORY SYSTEM BLOOD VESSELS

CIRCULATORY SYSTEM BLOOD VESSELS Name: Block: CIRCULATORY SYSTEM Multicellular organisms (above the level of roundworms) rely on a circulatory system to bring nutrients to, and take wastes away from, cells. In higher organisms such as

More information

A Cardiovascular Model for the Analysis of Pacing Configurations in Cardiac Resynchronization Therapy

A Cardiovascular Model for the Analysis of Pacing Configurations in Cardiac Resynchronization Therapy A Cardiovascular Model for the Analysis of Pacing Configurations in Cardiac Resynchronization Therapy Kevin Tse Ve Koon, Virginie Le Rolle, Guy Carrault, Alfredo Hernandez To cite this version: Kevin Tse

More information

Cardiovascular Physiology

Cardiovascular Physiology Cardiovascular Physiology Introduction The cardiovascular system consists of the heart and two vascular systems, the systemic and pulmonary circulations. The heart pumps blood through two vascular systems

More information

A NONINVASIVE METHOD FOR CHARACTERIZING VENTRICULAR DIASTOLIC FILLING DYNAMICS

A NONINVASIVE METHOD FOR CHARACTERIZING VENTRICULAR DIASTOLIC FILLING DYNAMICS A NONINVASIVE METHOD FOR CHARACTERIZING VENTRICULAR DIASTOLIC FILLING DYNAMICS R. Mukkamala, R. G. Mark, R. J. Cohen Haard-MIT Division of Health Sciences and Technology, Cambridge, MA, USA Abstract We

More information

Project 1: Circulation

Project 1: Circulation Project 1: Circulation This project refers to the matlab files located at: http://www.math.nyu.edu/faculty/peskin/modsimprograms/ch1/. Model of the systemic arteries. The first thing to do is adjust the

More information

Biomedical Instrumentation E. Blood Pressure

Biomedical Instrumentation E. Blood Pressure Biomedical Instrumentation E. Blood Pressure Dr Gari Clifford Adapted from slides by Prof. Lionel Tarassenko Blood pressure Blood is pumped around the body by the heart. It makes its way around the body

More information

Determination of Cardiac Output By Equating Venous Return Curves With Cardiac Response Curves1

Determination of Cardiac Output By Equating Venous Return Curves With Cardiac Response Curves1 Determination of Cardiac Output By Equating Venous Return Curves With Cardiac Response Curves1 ARTHUR C. GUYTQN From the Department of Physiology and Biophysics, School of Medicine, University of Mississippi,

More information