The Last Generation of Pericardial Valves in the Aortic Position: Ten-Year Follow-up in 589 Patients

Size: px
Start display at page:

Download "The Last Generation of Pericardial Valves in the Aortic Position: Ten-Year Follow-up in 589 Patients"

Transcription

1 The Last Generation of Pericardial Valves in the Aortic Position: Ten-Year Follow-up in 589 Patients Michel R. Aupart, MD, Agnes L. Sirinelli, MD, Frank F. Diemont, MD, Yvon A. Meurisse, MD, Xavier B. Dreyfus, MD, and Michel A. Marchand, MD Department of Cardiac Surgery, Trousseau University Hospital, Tours, France Background. The first generation of pericardial valves has been withdrawn from the market because of excessively high rates of premature failure. With its original design, the Carpentier-Edwards pericardial valve has promised improved results. Methods. In our institution, 589 patients underwent an isolated aortic valve replacement with a Carpentier- Edwards pericardial bioprosthesis between July 1984 and December The patients' mean age was years, and 49% of the patients were in New York Heart Association clinical class III or IV. The operative mortality rate was 2.3% (14 of 595). All patients but 4 were followed up for an average of 4.1 years after their operation, and total follow-up was 2,408 patient-years. Results. At the time of the study, more than 85% of the patients were in New York Heart Association class I or II. There were 79 late deaths. After 10 years, the actuarial survival rate was 71% + 7%. Nineteen patients died of valve-related causes (3 endocarditis, 7 thromboembolic complications, 1 structural failure, and 8 sudden deaths). The actuarial rate of freedom from valve-related death was 94% + 3% at 10 years. Valve-related complications included 23 thromboembolic episodes (0.9% per patient-year), 14 endocarditis (0.5% per patient-year), 9 reoperations (0.4% per patient-year), and 4 structural valve failures with calcification and stenosis (0.2% per patient-year). After 10 years, freedom from valve-related complications was 84% --. 6%, from reoperation 97% - 2%, and from valve failure 96% + 4%. Conclusions. Because of its low rate of valve-related events at 10 years and low rate of structural deterioration with no leaflet tears, this prosthesis is an outstanding choice for patients who need tissue valves and for patients aged 60 years or older. (Ann Thorac Surg 1996;61:615-20) he first generation of pericardial valves has been T abandoned by physicians because of poor clinical results and excessively high rates of deterioration, characterized by leaflet tears. With its original design, the Carpentier-Edwards valve has promised improved results. Intermediate follow-up has shown low rates of valve-related events, in particular a low rate of deterioration with no leaflet tears [1-3]. This study reports our experience at 10 years with 595 Carpentier-Edwards pericardial aortic valves (model 2900; Baxter Healthcare Corp., Santa Ana, CA) in 589 patients. Material and Methods The Valve The Carpentier-Edwards pericardial valve is a threeleaflet bioprosthesis made of glutaraldehyde-preserved bovine pericardium. This valve presents two conceptual improvements: Complete strut flexibility is achieved with an Elgiloy wire, and the design of the stent allows the pericardial cusp to be mounted by drawing it into the stent from below rather than suturing. Accepted for publication Sep 21, Address reprint requests to Dr Aupart, Unite de Chirurgie Cardiaque, Hopital Trousseau, ~fours Cedex, France. Patients From July 1984 to December 1993, 595 Carpentier- Edwards pericardial bioprostheses were used in 589 consecutive patients for isolated aortic valve replacement in our hospital. Patients undergoing multiple valve replacement were excluded from this study, but there were no exclusions for other concomitant operations. There were 426 men (71.5%) and 169 women (28.5%). The age ranged from 16 to 88 years, with a mean of 67.5 _ years; 46 patients were younger than 50 years old (Fig 1). The main indications for valve replacement were idiopathic calcific stenosis in 361 patients, dystrophic insufficiency in 157, rheumatic fever in 23, prosthetic valve dysfunction in 21, endocarditis in 16, congenital anomaly in 12, and acute aortic dissection in 5. Clinical presentation was aortic stenosis in 308 patients, aortic insufficiency in 217, and mixed disease in 70. The mean New York Heart Association clinical status was 2.4. Ten percent of patients had chronic atrial fibrillation, and 12% had coronary artery bypass grafting. There were 125 associated procedures (21%). These included aortocoronary bypass in 72 patients, mitral valvuloplasty in 21, tricuspid annuloplasty in 11, ascending aortic operations in 8, myomectomy in 10, and aneurysm repair in 3. The patients were operated on through a median sternotomy with standard cardiopulmonary bypass, hemodilution, and general body hypothermia. Myocardial 1996 by The Society of Thoracic Surgeons /96]$15.00 Published by Elsevier Science Inc SSDI (95)

2 616 AUPART ET AL Ann Thorac Surg EDWARDS AORTIC PERICARD1AL VALVES 1996;61: Fig 1. Age distribution of patients having aortic valve replacement AGE protection was obtained with crystalloid or blood cardioplegia and topical cooling. Sizes of the valves were 19 mm (90 patients), 21 mm (172), 23 mm (160), 25 mm (155), 27 mm (14), and 29 mm (4). The postoperative anticoagulant protocol included heparin for 2 days, followed by 1 month of calcium heparin (activated partial thromboplastin time > 1.5 N) or Acenocoumarol (international normalized ratio 2.5 to 3) (Ciba-Geigy, Rueil-Malmaison, France). After 1 month, anticoagulation treatment was discontinued at the cardiologist's discretion. Follow-up information was obtained during a 3-month interval (June 1994 to August 1994) through questionnaire and telephone contact with cardiologists, family physicians, and patients. All patients but 4 were followed up for an average of 4.1 years after their operation, and total follow-up was 2,408 patient-years (0.6% lost during follow-up). More than 95% of patients underwent an echographic study at follow-up; among them, 50 patients underwent echocardiographic Doppler studies at our institution. These were performed by the same physician using the same procedure. All data were input in an IBM computer and analyzed with Sedistat (SEDIA SA, Paris, France). Standard actuarial and linearized statistical techniques were used to describe survival and the incidence of valve-related complications. Only the first event for each patient during the study was considered in the actuarial analysis, whereas all events for each patient were considered in linearized rates. Continuous data are presented as mean + standard deviation; actuarial probability and linearized rates are given as mean _+ 95% confidence limits of the mean. The guidelines for reporting mortality and morbidity after cardiac valvular operations were observed in the preparation of this report [4]. Results Patient Survival Fourteen patients died within 30 days of the operation, for a mortality rate of 2.3%. The major cause of early death was cardiac failure in 11 patients. One patient died of neurologic embolism. There were no deaths for New York Heart Association class I or II. There were 79 late deaths. The causes of these deaths were cardiac related in 17 (21%), valve related in 19 (24%), and noncardiac in 43 (54%). The actuarial survival rate including operative deaths was 71% _+ 7% after 10 years (Fig 2). At the time of the study, more than 85% of the 492 followed patients were in New York Heart Association class I or II (Fig 3), 9% received anticoagulation treatment, and 8% were in atrial fibrillation. Valve-Related Complications Nineteen patients died of valve-related causes (3 endocarditis, 7 thromboembolic complications, 1 structural failure, and 8 sudden deaths). The actuarial rate of freedom from valve-related death was 94% +_ 3% at 10 Fig 2. Actuarial survival including operative deaths. 100 i Years

3 Ann Thorac Surg AUPART ET AL ;61: EDWARDS AORTIC PERICARDIAL VALVES Fig 3. Preoperative and postoperative New York Heart Association (NYHA) class. 200 z I II III IV NYHA Class years. No postmortem examination was available for the sudden deaths. Twenty-three patients presented with a thromboembolic accident, for an actuarial rate of 93% -+ 3% of patients free of thromboembolism at 10 years. Seventeen events were neurologic, with 7 deaths, 3 permanent neurologic deficits, and 7 without sequelae. Sixteen patients were in sinus rhythm, 5 in atrial fibrillation, and 2 with a pacemaker. Six of these had anticoagulation treatment (3 in sinus rhythm, 2 in atrial fibrillation, and I with a pacemaker). Two of them underwent mitral repair at the time of operation. The linearized rate of thromboembolism was 0.9% per patient-year. No thrombosis of the bioprosthesis was observed. There were seven nonfatal cases of hemorrhagic complications meeting the definition of the guidelines. The indication for anticoagulation treatment was atrial fibrillation or cardiologist decision. Endocarditis was reported in 14 patients (1 in the first 3 months, 2 in the first year); 4 required reoperation in an elective situation, without mortality. We used another bioprosthesis for this procedure in 3. Three patients died without reoperation after being diagnosed in another institution, and 7 had successful medical treatment without recurrence. The linearized rate of endocarditis was 0.5% per pafient-year, with an actuarial rate of 96% + 2% of patients free of endocarditis at 10 years. No hemolysis was noted in the followed patients. Nine patients required reoperation, for an actuarial rate of 97% _+ 2% of patients free of reoperation at 10 years and a linearized rate of 0.4% per patient-year. Causes of reoperation were endocardifis in 4, perivalvular leak in 3, and structural failure in 2. We used another pericardial bioprosthesis in 6 patients and a mechanical prosthesis in 3 (1 structural failure, 1 perivalvular leak, 1 endocarditis). No deaths occurred in these elective reoperations. There were only 4 patients with structural deterioration of the valve, for an actuarial rate of 96% + 4% of patients free of structural valve failure at 10 years and a linearized rate of 0.2% per patient-year. All required reoperation. One patient died of sudden death just before reoperation, whereas the other 3 had reoperations without problem (1 after the end of the study). The structural change was stenosis without regurgitation, caused by valve calcification. The explanted valve showed diffuse micromineralization of the three leaflets. It is notable that no leaflet tear was observed. The 4 patients' ages were 42, 58, 58, and 61 years, and the interval until calcification was 8, 7, 4, and 4 years, respectively. No structural change was noted in the other followed patients, but echographic studies were not performed in all patients. No deterioration was noted in patients less than 40 years of age, and only one case occurred in a patient older than 60 years. Age One hundred ten patients were less than 60 years of age. Among the 12 associated procedures, six were coronary artery bypass grafting and four were aortic arch operations. In this group, with a mean follow-up of 4.8 years and total follow-up of 454 patient-years, we observed one valve-related death (0.2%) and eight valve-related events (Table 1). Actuarial survival was 91% +_ 6% at 10 years. Actuarial freedom from valve-related death was 99% 1%, from endocarditis 96% + 4%, from thromboembolism 98% + 2%, from deterioration 92% _+ 6%, and from reoperation 96% _+ 4%. Of the total, 479 patients were older than 60 years. The major cause of the 113 associated procedures was coronary artery bypass grafting in 66. In this group, with a mean follow-up of 4 years and a total follow-up of 1,954 patient-years, we observed 18 valve-related deaths (0.9%) Table 1. Valve-Related Morbidity by Age Grouy Event <60 Years e60 Years Thromboembolism 1 (0.2%) 22 (1.1%) Hemorrhage 0 7 (0.3%) Endocarditis 3 (0.6%) 11 (0.6%) Reoperations 2 (0.4%) 6 (0.3%) Structural failure 3 (0.6%) 1 (0.05%) Valve-related death 1 (0.2%) 18 (0.9%) "~ Values are number with linearized rate in parentheses.

4 618 AUPART ET AL Ann Thorac Surg EDWARDS AORTIC PERICARDIAL VALVES 1996;61: Table 2. Hemodynamic Data Valve Size Mean Gradient Effective Orifice (ram) N (mrn Hg) Area (cn'l 2) and 47 valve-related events (see Table 1). Actuarial survival was 64% _+ 9% at 10 years. Actuarial freedom from valve-related death was 92% _+ 4%, from endocarditis 97% +_ 2%, from thromboembolism 92% -- 4%, from deterioration 99% +_ 1%, and from reoperation 98% _+ 2%. Statistical analysis showed no significant difference between the age groups in terms of valve-related morbidity. Hemodynamic Data Despite the short coaptation area of the leaflet, no major central insufficiency or structural changes were noted except in patients with structural failure. Among the 50 patients assessed by echocardiography in our institution, 8 had an aortic valve replacement with a 19-mm bioprosthesis, 14 with a 21-mm, 13 with a 23-mm, 12 with a 25-mm, and 3 with a 27-ram bioprosthesis. Hemodynamic data are presented in Table 2. Comment Using pericardium to manufacture heart valve prostheses remains controversial. Currently, valve durability is the major problem with all bioprostheses. Pericardial bioprostheses have been used clinically in humans since The first generation of this type of valve has been withdrawn from the market because of poor clinical results [5, 6]. The mode of deterioration was documented not as pericardium failure, but as design failure [7, 8] or preparation failure [9]. The Carpentier-Edwards pericardial valve is designed with an original leaflet clamping, which eliminates the need for retention suture as well as the risk of abrasion. Procurement and selection of pericardium are strictly monitored. Preservation is done with glutaraldehyde. With these specifications, this pericardial bioprosthesis has shown satisfactory intermediate results [10, 11] in both positions for rehabilitating pericardium as a valve substitute. These results, especially in the aortic position and with small sizes, promise very interesting long-term follow-up. However, few reports are available with this type of pericardial valve. The Carpentier-Edwards pericardial valve was first used in Europe and received Food and Drug Administration approval in We report here a large study with this prosthesis. We have used these pericardial valves in the aortic position since 1984 in 589 patients, representing more than half of all valve replacements during the same period. It is the only bioprosthesis used for valve replacement in our institution. Major indications were aortic stenosis and age greater than 60 years. We also used this pericardial valve in patients with a small aortic annulus, with poor ventricular function, and, of course, in cases of contraindication to anticoagulation therapy and in young women who want pregnancy. Finally, after informed consent, the patient's choice is taken into account. In this study, the overall rate of survival was 71%. The rate and causes of death were similar to those in other series of aortic valve replacement [12]. There were 19 valve-related deaths, among them eight sudden deaths. No postmortem examination was possible because these deaths occurred in another institution or in the patient's home. Despite this fact, the actuarial freedom from valve-related death is comparable to that in other studies. Morbidity with this prosthesis was low. The most important complication was thromboembolism in 23 patients. This complication was significantly more common in patients over 60 years of age. In patients younger than 60, only one thromboembolic event was noted, and freedom from thromboembolism was 98% _+ 2% at 10 years. We think that the definition of the guidelines is too large and that some thromboembolic events were not in fact valve related, especially in the elderly population, but were instead related to the vascular status of the patients [13]. We observed only four structural deteriorations, with no leaflet tears. Deterioration of the tissue was always the same: calcification of the leaflet leading to valve stenosis. Explanted valves showed intrinsic calcification of the three leaflets. The cause of this calcification is poorly understood, but we expect better results with improvement in preparation of the tissue. The hemodynamic data, as in other publications on pericardial valves [14], were satisfactory. In this study, 90 patients underwent replacement with a 19-mm valve. The mean gradient was 18 mm Hg, the peak gradient was 32 mm Hg, and the effective orifice area was 1.08 cm 2. Unlike other investigators [15], we observed no increased rate of sudden death. The hemodynamic performance is one of the major reasons to continue developing the pericardial device. Among the 589 patients, 110 were less than 60 years of age at the time of the operation. The results in this particular group were very interesting. The 10-year performance appears comparable to that obtained with mechanical valves, classically used in these patients [12, 16]. Freedom from deterioration was 96% _+ 4% at 10 years, and better quality of life (no valve-related noise and no anticoagulation therapy) was especially reported by younger patients. The behavior of this pericardial valve in young patients appears better than that obtained with porcine valves [17]. But because of the history, of pericardial valves, it is necessary to be very careful, and more follow-up is mandated in young patients. Comparison of the long-term results between different valves is difficult, because many factors vary among cardiac surgical teams. Our results at 10 years appear

5 Ann Thorac Surg AUPART ET AL ;61: EDWARDS AORTIC PERICARD1AL VALVES better than those obtained with previous pericardial bioprostheses. There is a great difference in the structural deterioration rate between the Carpentier-Edwards and the first generation of pericardial valves (Ionescu-Shiley, Shiley, Inc, Irvine, CA; Hancock, Medtronic Cardiopulmonary Division, Anaheim, CA). The reported rate is approximately 80% at 5 years for the Ionescu-Shiley [18] and 75% for the Hancock pericardial bioprosthesis [19]. Too few data are available about the Pericarbon pericardial valve (Pericarbon; Sorin Biomedica, Saluggia, Italy) to do a long-term comparison [20]. Comparison between porcine and pericardial valves was described previously [21, 22]. Cosgrove and associates [21] showed that pericardial valves are less obstructive than porcine, and Pelletier and colleagues [22] found a higher rate of structural failure with pericardial devices over 6 years. However, the prosthesis used in the latter study was the Ionescu-Shiley valve, which presents a high rate of premature failure. At 10 years, our low deterioration rate with Carpentier-Edwards valves compares favorably with the best published results on porcine prostheses [23, 24]. The difference with our study may be due to the better stress behavior of pericardium than of porcine tissue over time. The comparison between mechanical and biologic valves was established by Bloomfield and co-workers [25] and Hammermeister and associates [26]. These studies involved the first generation of pericardial valves or porcine valves. The better results of the last generation of pericardial valves, especially in younger patients, will certainly change the results of the comparison. In many studies, homograft is the procedure of choice in terms of valve-related morbidity. The problems with homograft involve technical procedures and procurement. Structural deterioration rates were variable and sometimes included substantial regurgitation. Nevertheless, at 10 years the reported rates of deterioration and reoperation with cryopreserved [27] and 4 C antibioticstored valves [28] are identical to those in our study, although the patients are younger. The coming years will determine the long-term behavior of this bioprosthesis compared with homografts, especially in young patients. Despite their hemodynamic results and their interesting intermediate morbidity rates, no comparison can be made with stentless valves, which have had a shorter follow-up [29]. We conclude that this aortic pericardial valve appears as good and perhaps better than available porcine valves, and is at least as good as the results with homografts at 10 years. The results with the stentless valves and the new porcine valves in the coming years must be compared with this very good tissue valve. For us, this is the prosthesis of choice when a tissue valve is needed, especially in patients older than 60 years and in those with a small aortic annulus. Nevertheless, calcification appears to be unavoidable with time, and longer follow-up and further studies are necessary to provide definitive conclusions. References 1. Frater RW, Salomon NW, Rainer WG, Cosgrove DM, Wickman E. The Carpentier Edwards pericardial valve: intermediate results. Ann Thorac Surg 1992;53: Perier P, Mihaileanu S, Fabiani JN, et al. Long term evaluation of the Carpentier-Edwards pericardial valve in the aortic position. J Cardiac Surg 1991;6: Aupart M, Neville P, Dreyfus X, Meurisse Y, Sirinelli A, Marchand M. The Carpentier Edwards pericardial aortic valve: intermediate results in 420 patients. Eur J Cardiothorac Surg 1994;8: Edmunds LH Jr, Clark RE, Cohn LH, Miller DC, Weisel RD. Guidelines for reporting morbidity and mortality after cardiac valvular operations. Ann Thorac Surg 1988;46: Gallo I, Nistal F, Revuelta JM, Garcia-Statue E, Artinano E, Duran CG. Incidence of primary, valve failure with the Ionescu-Shiley pericardial valve. J Thorac Cardiovasc Surg 1985;90: Bortolotti U, Milano A, Guerra F, et al. Failure of Hancock pericardial xenografts: is prophylactic bioprosthetic replacement justified? Ann Thorac Surg 1991;51: Walley VM, Keon WJ. Patterns of failure in lonescu-shiley bovine pericardial valves. J Thorac Cardiovasc Surg 1987;93: Walley VM, Rubens FD, Campagna M, Pipe AL, Keon WJ. Patterns of failure in Hancock pericardial bioprostheses. J Thorac Cardiovasc Surg 1991;102: Leandri J, Bertrand P, Mazzucotelli JP, Loisance D. Mode of failure of the Mitroflow pericardial valve. J Heart Valve Dis 1992;1: Meyns B, Szecsi J, Flameng W, Daenen W. Aortic and mitral valve replacement with the Carpentier-Edwards pericardial bioprosthesis: mid-term clinical results. J Heart Valve Dis 1993;2: Aupart M, Neville P, Meurisse Y, Dreyfus X, Sirinelli A, Marchand M. The Carpentier-Edwards pericardial mitral bioprosthesis: intermediate follow up in 121 patients. In: Gabbay S, Frater R, eds. New horizon and the future of heart valve bioprostheses. Proceedings of the third meeting of the International Association for Cardiac Biological Implants. Austin: Silent Partners, 1994: Czer LSC, Chaux A, Matloff JM, et al. Ten year experience with the St Jude Medical valve for prima~ valve replacement. J Thorac Cardiovasc Surg 1990;100: Butchard EG, Moreno de la Santa P, Rooney SJ, Lewis PA. Arterial risk factors and ischemic cerebrovascular events after aortic replacement. J Heart Valve Dis 1995;4: Cosgrove DM, Lytle BW, Williams W. Hemodynamic performance of the Carpentier Edwards pericardial valve in the aortic position in vivo. Circulation 1985;72(Suppl 2): Kratz JM, Sade RM, Crawford FA, Crumbley BAJ, Stroud MR. The risk of small St Jude aortic valve prostheses. Ann Thorac Surg 1994;57: Nitter-Hauger S, Abdelnoor M. Ten year experience with the Medtronic Hall valvular prosthesis. Circulation 1989;80 (Suppl 1): Al-Khaja N, Belboul A, Rashid M, et al. The influence of age on the durability of Carpentier-Edwards biological valves. Eur J Cardiothorac Surg 1991;5: Daenen W, Noyez L, Lesaffre E. The Ionescu-Shiley pericardial valve: results in 473 patients. Ann Thorac Surg 1988;46: Scully H, Goldman B, Fulop J, et al. Five year follow up of Hancock pericardial valves: management of premature failure. J Cardiac Surg 1988;3: Wallberger E, Jault F, Fontanel M, et al. Bioprostheses p6ricardiques de Sorin-Pericarbon. Arch Mal Coeur 1994;87: Cosgrove DM, Lytle BW, Gill CC, et al. In vivo hemodynamic comparison of porcine and pericardial valves. J Thorac Cardiovasc Surg 1985;89: Pelletier CL, Carrier M, Leclerc Y, Lepage G, DeGuise P,

6 620 AUPART ET AL Ann Thorac Surg EDWARDS AORTIC PERICARDIAL VALVES 1996;61: Dyrda I. Porcine versus pericardial bioprostheses: a comparison of late results in 1593 patients. Ann Thorac Surg 1989;47: Burdon TA, Miller DC, Oyer PE, et al. Durability of porcine valve at fifteen years in a representative North American patient population. J Thorac Cardiovasc Surg 1992;103: Barratt-Boyes BG, Jaffe WM, Hong Ko P, Whitlock RML. The zero pressure fixed Medtronic Intact porcine valve: an 8.5 year review. J Heart Valve Dis 1993;2: Bloomfield P, Wheatley DJ, Prescott RJ, Miller HC. Twelve year comparison of a Bj6rk-Shiley mechanical valve with porcine bioprosthesis. N Engl J Med 1991;324: Hammermeister KE, Sethi GK, Henderson WG, et al. A comparison of outcomes in men 11 years after heart valve replacement with a mechanical valve or bioprosthesis. N Engl J Med 1993;328: Kirklin JK, Smith D, Novik W, et al. Long term function of cryopreserved aortic homografts. J Thorac Cardiovasc Surg 1993;106: Barratt-Boyes BG, Roche AHG, Subramanyan R, Pemberton JR, Whitlock RML. Long term follow up with the antibiotic sterilized aortic homograft valve inserted freehand in the aortic position. Circulation 1987;75: David TE, Feindel CM, Bos J, Sun Z, Scully HE, Rakowski H. Aortic valve replacement with a stentless porcine aortic valve. J Thorac Cardiovasc Surg 1994;108: INVITED COMMENTARY Aupart and colleagues have served an important purpose by putting second-generation pericardial valves into proper perspective. Although the Ionescu-Shiley pericardial valve showed excellent hemodynamics [1], its ultimate failure to receive widespread acceptance was based on (1) a design problem using a rigid stent and a fixation suture at the top of each commissure, creating a "stress point" that was responsible for the starting point of most leaflet disruptions, and (2) implantation of the valve into patients of nearly all ages. The resultant difficulties at reoperation were magnified by rigid fixation of a Dacron patch on the back of each strut, which formed a firm fibrous bond to the adjacent aortic wall making removal frustrating, difficult, and associated with higher risk. The ingenious and innovative design by an Edwards Laboratory engineer whereby the tissue leaflets are firmly attached behind the struts without the need for anchoring sutures has nearly eliminated the problem of leaflet disruption. This achievement, along with the use of a flexible stent, highlights the importance of close collaboration between the clinical cardiac surgeon and the biomedical engineer. Contemporaneously, based on several clinical studies, it is now apparent that tissue valve longevity and performance are far better in patients in the older age groups (whether this be 60, 65, or 70 years and older). Aupart and colleagues have added further evidence to the thesis that pericardial valves in the aortic position function very well and over an extended period of time if placed in patients selected on the basis of age and if the valve has the improvements incorporated in the second-generation design. W. Gerald Rainer, MD Department of Surgery University of Colorado Health Sciences Center 2005 Franklin St, Suite 380 Denver, CO Reference 1. Rainer WG, Christopher RA, Sadler TR Jr, Hilgenberg AD. Dynamic behavior of prosthetic aortic tissue valves as viewed by high-speed cinematography. Ann Thorac Surg 1979;28:

CARPENTIER-EDWARDS PERICARDIAL VALVES IN THE MITRAL POSITION: TEN-YEAR FOLLOW-UP

CARPENTIER-EDWARDS PERICARDIAL VALVES IN THE MITRAL POSITION: TEN-YEAR FOLLOW-UP CARPENTIER-EDWARDS PERICARDIAL VALVES IN THE MITRAL POSITION: TEN-YEAR FOLLOW-UP M. R. Aupart, MD P. H. Neville, MD S. Hammami, MD A. L. Sirineili, MD Y. A. Meurisse, MD M. A. Marchand, MD Objective: The

More information

15-Year Comparison of Supra-Annular Porcine and PERIMOUNT Aortic Bioprostheses

15-Year Comparison of Supra-Annular Porcine and PERIMOUNT Aortic Bioprostheses ORIGINAL CONTRIBUTION 15-Year Comparison of Supra-Annular Porcine and PERIMOUNT Aortic Bioprostheses WR Eric Jamieson, MD, Eva Germann, MSc, Michel R Aupart, MD 1, Paul H Neville, MD 1, Michel A Marchand,

More information

16 YEAR RESULTS Carpentier-Edwards PERIMOUNT Mitral Pericardial Bioprosthesis, Model 6900

16 YEAR RESULTS Carpentier-Edwards PERIMOUNT Mitral Pericardial Bioprosthesis, Model 6900 CLINICAL COMMUNIQUé 6 YEAR RESULTS Carpentier-Edwards PERIMOUNT Mitral Pericardial Bioprosthesis, Model 69 The Carpentier-Edwards PERIMOUNT Mitral Pericardial Valve, Model 69, was introduced into clinical

More information

Durability of Pericardial Versus Porcine Aortic Valves

Durability of Pericardial Versus Porcine Aortic Valves Journal of the American College of Cardiology Vol. 44, No. 2, 2004 2004 by the American College of Cardiology Foundation ISSN 0735-1097/04/$30.00 Published by Elsevier Inc. doi:10.1016/j.jacc.2004.01.053

More information

Read at the Twenty-fourth Annual Meeting of The Western Thoracic Surgical Association, Whistler, British Columbia, June 24-27, 1998.

Read at the Twenty-fourth Annual Meeting of The Western Thoracic Surgical Association, Whistler, British Columbia, June 24-27, 1998. STRUCTURAL VALVE DETERIORATION IN MITRAL REPLACEMENT SURGERY: COMPARISON OF CARPENTIER-EDWARDS SUPRA-ANNULAR PORCINE AND PERIMOUNT PERICARDIAL BIOPROSTHESES W. R. Eric Jamieson, MD a Michel A. Marchand,

More information

CLINICAL COMMUNIQUE 16 YEAR RESULTS

CLINICAL COMMUNIQUE 16 YEAR RESULTS CLINICAL COMMUNIQUE 6 YEAR RESULTS Carpentier-Edwards PERIMOUNT Mitral Pericardial Bioprosthesis, Model 6900 Introduction The Carpentier-Edwards PERIMOUNT Mitral Pericardial Valve, Model 6900, was introduced

More information

Carpentier-Edwards Pericardial Valve in the Aortic Position: 25-Years Experience

Carpentier-Edwards Pericardial Valve in the Aortic Position: 25-Years Experience SURGERY: The Annals of Thoracic Surgery CME Program is located online at http://www.annalsthoracicsurgery.org/cme/ home. To take the CME activity related to this article, you must have either an STS member

More information

A 20-year experience of 1712 patients with the Biocor porcine bioprosthesis

A 20-year experience of 1712 patients with the Biocor porcine bioprosthesis Acquired Cardiovascular Disease Mykén and Bech-Hansen A 2-year experience of 1712 patients with the Biocor porcine bioprosthesis Pia S. U. Mykén, MD, PhD, a and Odd Bech-Hansen, MD, PhD b Objective: The

More information

P have been used for mitral and aortic valve replacement

P have been used for mitral and aortic valve replacement A -Year Comparison of Mitral Valve Replacement With Carpentier-Edwards and Hancock Porcine Bioprostheses P. Perier, MD, A. Deloche, MD, S. Chauvaud, MD, J. C. Chachques, MD, J. Relland, MD, J. N. Fabiani,

More information

Carpentier-Edwards supra-annular aortic porcine bioprosthesis: Clinical performance over 20 years

Carpentier-Edwards supra-annular aortic porcine bioprosthesis: Clinical performance over 20 years Surgery for Acquired Cardiovascular Disease Carpentier-Edwards supra-annular aortic porcine bioprosthesis: Clinical performance over 20 years W. R. Eric Jamieson, MD, Lawrence H. Burr, MD, Robert T. Miyagishima,

More information

Reoperation for Bioprosthetic Mitral Structural Failure: Risk Assessment

Reoperation for Bioprosthetic Mitral Structural Failure: Risk Assessment Reoperation for Bioprosthetic Mitral Structural Failure: Risk Assessment W.R.E. Jamieson, MD; L.H. Burr, MD; R.T. Miyagishima, MD; M.T. Janusz, MD; G.J. Fradet, MD; S.V. Lichtenstein, MD; H. Ling, MD Background

More information

P substitutes since the introduction of the Ionescu-

P substitutes since the introduction of the Ionescu- Mitroflow Pericardial Valve: Long-Term Durability Daniel Y. Loisance, MD, Jean-Philippe Mazzucotelli, MD, Patrick C. Bertrand, MD, Philippe H. Deleuze, MD, and Jean-Paul Cachera, MD Department of Surgical

More information

A Surgeon s Perspective Guidelines for the Management of Patients with Valvular Heart Disease Adapted from the 2006 ACC/AHA Guideline Revision

A Surgeon s Perspective Guidelines for the Management of Patients with Valvular Heart Disease Adapted from the 2006 ACC/AHA Guideline Revision A Surgeon s Perspective Guidelines for the Management of Patients with Valvular Heart Disease Adapted from the 2006 ACC/AHA Guideline Revision Prof. Pino Fundarò, MD Niguarda Hospital Milan, Italy Introduction

More information

Primary Tissue Valve Degeneration in Glutaraldehvde-Preserved Porcine Biomostheses: Hancock I Vekus Carpentier-Edwards at 4- to 7-Years Follow-up

Primary Tissue Valve Degeneration in Glutaraldehvde-Preserved Porcine Biomostheses: Hancock I Vekus Carpentier-Edwards at 4- to 7-Years Follow-up Primary Tissue Valve Degeneration in Glutaraldehvde-Preserved Porcine Biomostheses: A Hancock I Vekus Edwards at 4- to 7-Years Follow-up Francisco Nistal, M.D., Edurne Artifiano, M.D., and Ignacio Gallo,

More information

Clinical material and methods. Copyright by ICR Publishers 2003

Clinical material and methods. Copyright by ICR Publishers 2003 Fourteen Years Experience with the CarboMedics Valve in Young Adults with Aortic Valve Disease Jan Aagaard 1, Jens Tingleff 2, Per V. Andersen 1, Christel N. Hansen 2 1 Department of Cardio-Thoracic and

More information

T sors in the following aspects: the porcine aortic valve

T sors in the following aspects: the porcine aortic valve Clinical and Hemodynamic Assessment of the Hancock I1 Bioprosthesis Tirone E. David, MD, Susan Armstrong, MSc, and Zhao Sun, MA Division of Cardiovascular Surgery, The Toronto Hospital and University of

More information

The St. Jude Medical Biocor Bioprosthesis

The St. Jude Medical Biocor Bioprosthesis The St. Jude Medical Biocor Bioprosthesis Clinical Evidence of Long-term Durability Long-term Biocor Experience A Review and Comparative Assessment Long-term Biocor Stented Tissue Valve Studies Twenty-year

More information

Department of Cardiothoracic Surgery, Heart and Lung Center, Lund University Hospital, Lund, Sweden

Department of Cardiothoracic Surgery, Heart and Lung Center, Lund University Hospital, Lund, Sweden Long-Term Outcome of the Mitroflow Pericardial Bioprosthesis in the Elderly after Aortic Valve Replacement Johan Sjögren, Tomas Gudbjartsson, Lars I. Thulin Department of Cardiothoracic Surgery, Heart

More information

Aortic valve replacement with the Sorin Pericarbon Freedom stentless prosthesis: 7 years experience in 130 patients

Aortic valve replacement with the Sorin Pericarbon Freedom stentless prosthesis: 7 years experience in 130 patients Aortic valve replacement with the Sorin Pericarbon Freedom stentless prosthesis: 7 years experience in 130 patients Augusto D Onofrio, MD, Stefano Auriemma, MD, Paolo Magagna, MD, Alessandro Favaro, MD,

More information

The CarboMedics bileaflet prosthetic heart was introduced

The CarboMedics bileaflet prosthetic heart was introduced The CarboMedics Valve: Experience With 1,049 Implants José M. Bernal, MD, José M. Rabasa, MD, Francisco Gutierrez-Garcia, MD, Carlos Morales, MD, J. Francisco Nistal, MD, and José M. Revuelta, MD Department

More information

Outcomes of Mitral Valve Repair for Mitral Regurgitation Due to Degenerative Disease

Outcomes of Mitral Valve Repair for Mitral Regurgitation Due to Degenerative Disease Outcomes of Mitral Valve Repair for Mitral Regurgitation Due to Degenerative Disease TIRONE E. DAVID, MD ; SEMIN THORAC CARDIOVASC SURG 19:116-120c 2007 ELSEVIER INC. PRESENTED BY INTERN 許士盟 Mitral valve

More information

AORTIC VALVE REPLACEMENT WITH FREEHAND AUTOLOGOUS PERICARDIUM

AORTIC VALVE REPLACEMENT WITH FREEHAND AUTOLOGOUS PERICARDIUM AORTIC VALVE REPLACEMENT WITH FREEHAND AUTOLOGOUS PERICARDIUM Fifty-one patients with a mean age of 31.2 years underwent aortic valve replacement with glutaraldehyde-treated autologous pericardium. Pure

More information

W e have previously reported the results of a randomised

W e have previously reported the results of a randomised 715 CARDIOVASCULAR MEDICINE Twenty year comparison of a mechanical heart valve with porcine bioprostheses H Oxenham, P Bloomfield, D J Wheatley, R J Lee, J Cunningham, R J Prescott, H C Miller... See end

More information

P found suitable valve substitutes for clinical use.

P found suitable valve substitutes for clinical use. Porcine Versus Pericardial Bioprostheses: A Comparison of Late Results in 1,593 Patients L. Conrad Pelletier, MD, Michel Carrier, MD, Yves Leclerc, MD, Gilles Lepage, MD, Pierre deguise, MD, and Ihor Dyrda,

More information

Does Patient-Prosthesis Mismatch Affect Long-term Results after Mitral Valve Replacement?

Does Patient-Prosthesis Mismatch Affect Long-term Results after Mitral Valve Replacement? Original Article Does Patient-Prosthesis Mismatch Affect Long-term Results after Mitral Valve Replacement? Hiroaki Sakamoto, MD, PhD, and Yasunori Watanabe, MD, PhD Background: Recently, some articles

More information

TSDA Boot Camp September 13-16, Introduction to Aortic Valve Surgery. George L. Hicks, Jr., MD

TSDA Boot Camp September 13-16, Introduction to Aortic Valve Surgery. George L. Hicks, Jr., MD TSDA Boot Camp September 13-16, 2018 Introduction to Aortic Valve Surgery George L. Hicks, Jr., MD Aortic Valve Pathology and Treatment Valvular Aortic Stenosis in Adults Average Course (Post mortem data)

More information

Late failure of transcatheter heart valves: An open question

Late failure of transcatheter heart valves: An open question Late failure of transcatheter heart valves: An open question A comparison with surgically implanted bioprosthetic heart valves. A. Rashid The Cardiothoracic Centre Liverpool, UK. Conflict of Interest Statement

More information

Reconstruction of the intervalvular fibrous body during aortic and

Reconstruction of the intervalvular fibrous body during aortic and Aortic and mitral valve replacement with reconstruction of the intervalvular fibrous body: An analysis of clinical outcomes Nilto C. De Oliveira, MD Tirone E. David, MD Susan Armstrong, MSc Joan Ivanov,

More information

LONG-TERM OUTCOME AFTER BIOLOGIC VERSUS MECHANICAL AORTIC VALVE REPLACEMENT IN 841 PATIENTS

LONG-TERM OUTCOME AFTER BIOLOGIC VERSUS MECHANICAL AORTIC VALVE REPLACEMENT IN 841 PATIENTS LONG-TERM OUTCOME AFTER BIOLOGIC VERSUS MECHANICAL AORTIC VALVE REPLACEMENT IN 841 PATIENTS David S. Peterseim, MD Ye-Ying Cen, MA Srinivas Cheruvu, MHS Kevin Landolfo, MD Thomas M. Bashore, MD James E.

More information

Aspirin or Coumadin as the Drug of Choice

Aspirin or Coumadin as the Drug of Choice Aspirin or Coumadin as the Drug of Choice for Valve Replacement with Porcine Bioprosthesis L. Nufiez, M.D., M. Gil Aguado, M.D., D. Celemin, M.D., A. Iglesias, M.D., and J. L. Larrea, M.D. ABSTRACT Eight

More information

Hani K. Najm MD, Msc, FRCSC FACC, FESC President Saudi Society for Cardiac Surgeons Associate Professor of Cardiothoracic Surgery King Abdulaziz

Hani K. Najm MD, Msc, FRCSC FACC, FESC President Saudi Society for Cardiac Surgeons Associate Professor of Cardiothoracic Surgery King Abdulaziz Hani K. Najm MD, Msc, FRCSC FACC, FESC President Saudi Society for Cardiac Surgeons Associate Professor of Cardiothoracic Surgery King Abdulaziz Cardiac Centre Riyadh, Saudi Arabia Decision process for

More information

Hani K. Najm MD, Msc, FRCSC, FRCS (Glasgow), FACC, FESC President of Saudi Heart Association King Abdulaziz Cardiac Centre Riyadh, Saudi Arabia.

Hani K. Najm MD, Msc, FRCSC, FRCS (Glasgow), FACC, FESC President of Saudi Heart Association King Abdulaziz Cardiac Centre Riyadh, Saudi Arabia. Hani K. Najm MD, Msc, FRCSC, FRCS (Glasgow), FACC, FESC President of Saudi Heart Association King Abdulaziz Cardiac Centre Riyadh, Saudi Arabia. Decision process for Management of any valve Timing Feasibility

More information

Hemodynamics Benefit of Supra-Annular Design in Failed Bio-Prosthetic Valves

Hemodynamics Benefit of Supra-Annular Design in Failed Bio-Prosthetic Valves Hemodynamics Benefit of Supra-Annular Design in Failed Bio-Prosthetic Valves Speaker's name: I have the following potential conflicts of interest to report: Proctorship for Medtronic Agenda Failure modes

More information

CoreValve in a Degenerative Surgical Valve

CoreValve in a Degenerative Surgical Valve CoreValve in a Degenerative Surgical Valve Ran Kornowski, MD, FESC, FACC Chairman Department of Cardiology Rabin Medical Center, Petach Tikva, Israel Disclosure Statement of Financial Interest I, Ran Kornowski,

More information

ORIGINAL PAPER. The long-term results and changing patterns of biological valves at the mitral position in contemporary practice in Japan

ORIGINAL PAPER. The long-term results and changing patterns of biological valves at the mitral position in contemporary practice in Japan Nagoya J. Med. Sci. 78. 369 ~ 376, 2016 doi:10.18999/nagjms.78.4.369 ORIGINAL PAPER The long-term results and changing patterns of biological valves at the mitral position in contemporary practice in Japan

More information

Presenter Disclosure. Patrick O. Myers, M.D. No Relationships to Disclose

Presenter Disclosure. Patrick O. Myers, M.D. No Relationships to Disclose Presenter Disclosure Patrick O. Myers, M.D. No Relationships to Disclose Aortic Valve Repair by Cusp Extension for Rheumatic Aortic Insufficiency in Children Long term Results and Impact of Extension Material

More information

Mitroflow Synergy Prostheses for Aortic Valve Replacement: 19 Years Experience With 1,516 Patients

Mitroflow Synergy Prostheses for Aortic Valve Replacement: 19 Years Experience With 1,516 Patients Mitroflow Synergy Prostheses for Aortic Valve Replacement: 19 Years Experience With 1,516 Patients Kazutomo Minami, MD, Armin Zittermann, PhD, Sebastian Schulte-Eistrup, MD, Heinrich Koertke, MD, and Reiner

More information

Nineteen-Millimeter Aortic St. Jude Medical Heart Valve Prosthesis: Up to Sixteen Years Follow-up

Nineteen-Millimeter Aortic St. Jude Medical Heart Valve Prosthesis: Up to Sixteen Years Follow-up Nineteen-Millimeter Aortic St. Jude Medical Heart Valve Prosthesis: Up to Sixteen Years Follow-up Dilip Sawant, FRCS, Arun K. Singh, MD, William C. Feng, MD, Arthur A. Bert, MD, and Fred Rotenberg, MD

More information

Bioprostheses are prone to continuous degeneration

Bioprostheses are prone to continuous degeneration Twenty-Year Experience With the St. Jude Medical Biocor Bioprosthesis in the Aortic Position Walter B. Eichinger, MD, Ina M. Hettich, MD, Daniel J. Ruzicka, MD, Klaus Holper, MD, Carolin Schricker, Sabine

More information

Ball Valve (Smeloff-Cutter) Aortic Valve Replacement Without Anticoagulation

Ball Valve (Smeloff-Cutter) Aortic Valve Replacement Without Anticoagulation Ball Valve (Smeloff-Cutter) Aortic Valve Replacement Without Anticoagulation Begonia Gometza, MD, and Carlos M. G. Duran, MD, PhD Department of Cardiovascular Diseases, King Faisal Specialist Hospital

More information

Controversy exists regarding which valve type is best

Controversy exists regarding which valve type is best Treatment of Endocarditis With Valve Replacement: The Question of Tissue Versus Mechanical Prosthesis Marc R. Moon, MD, D. Craig Miller, MD, Kathleen A. Moore, BS, Phillip E. Oyer, MD, PhD, R. Scott Mitchell,

More information

Long-term results (22 years) of the Ross Operation a single institutional experience

Long-term results (22 years) of the Ross Operation a single institutional experience Long-term results (22 years) of the Ross Operation a single institutional experience Authors: Costa FDA, Schnorr GM, Veloso M,Calixto A, Colatusso D, Balbi EM, Torres R, Ferreira ADA, Colatusso C Department

More information

CONTRIBUTION. Aortic valve replacement in young patients: long-term follow-up

CONTRIBUTION. Aortic valve replacement in young patients: long-term follow-up CONTRIBUTION Aortic valve replacement in young patients: long-term follow-up DOUGLAS S. MOODIE, MD; USAMA HANHAN, MD; RICHARD STERBA, MD; DANIEL J. MURPHY, Jr, MD; ELIOT R. ROSENKRANZ, MD; ANDREA M. KOVACS,

More information

by Age Groups W. R. E. Jamieson, M.D., L. J. Rosado, M.D., A. I. Munro, M.D., A. N. Gerein, M.D., L. H. Burr, M.D., R. T. Miyagishima, M.D.

by Age Groups W. R. E. Jamieson, M.D., L. J. Rosado, M.D., A. I. Munro, M.D., A. N. Gerein, M.D., L. H. Burr, M.D., R. T. Miyagishima, M.D. Carpentier-Edwards Standard Porcine Bioprosthesis: Primary Tissue Failure (Structural Valve Deterioration) by Age Groups W. R. E. Jamieson, M.D., L. J. Rosado, M.D., A. I. Munro, M.D., A. N. Gerein, M.D.,

More information

Durability and Outcome of Aortic Valve Replacement With Mitral Valve Repair Versus Double Valve Replacement

Durability and Outcome of Aortic Valve Replacement With Mitral Valve Repair Versus Double Valve Replacement Durability and Outcome of Aortic Valve Replacement With Mitral Valve Repair Versus Double Valve Replacement Masaki Hamamoto, MD, Ko Bando, MD, Junjiro Kobayashi, MD, Toshihiko Satoh, MD, MPH, Yoshikado

More information

The operative mortality rate after redo valvular operations

The operative mortality rate after redo valvular operations Clinical Outcomes of Redo Valvular Operations: A 20-Year Experience Naoto Fukunaga, MD, Yukikatsu Okada, MD, Yasunobu Konishi, MD, Takashi Murashita, MD, Mitsuru Yuzaki, MD, Yu Shomura, MD, Hiroshi Fujiwara,

More information

THE IMPACT OF AGE, CORONARY ARTERY DISEASE, AND CARDIAC COMORBIDITY ON LATE SURVIVAL AFTER BIOPROSTHETIC AORTIC VALVE REPLACEMENT

THE IMPACT OF AGE, CORONARY ARTERY DISEASE, AND CARDIAC COMORBIDITY ON LATE SURVIVAL AFTER BIOPROSTHETIC AORTIC VALVE REPLACEMENT THE IMPACT OF AGE, CORONARY ARTERY DISEASE, AND CARDIAC COMORBIDITY ON LATE SURVIVAL AFTER BIOPROSTHETIC AORTIC VALVE REPLACEMENT Gideon Cohen, MD Tirone E. David, MD Joan Ivanov, MSc Sue Armstrong, MSc

More information

Iatrogenic pathology of the heart:

Iatrogenic pathology of the heart: Iatrogenic pathology of the heart: Complications of mitral valve plasty and replacement Patrick Bruneval D pt of Pathology Hôpital Européen Georges Pompidou Enterprise Interest None Mitral valve surgery

More information

A valve was initiated at the Medical University of

A valve was initiated at the Medical University of St. Jude Prosthesis for Aortic and Mitral Valve Replacement: A Ten-Year Experience John M. Kratz, MD, Fred A. Crawford, Jr, MD, Robert M. Sade, MD, Arthur J. Crumbley, MD, and Martha R. Stroud, MS Division

More information

Porcine bioprosthesis use for surgical treatment of

Porcine bioprosthesis use for surgical treatment of Fifteen-Year Clinical Experience With the Biocor Porcine Bioprostheses in the Mitral Position Kaan Kırali, MD, Mustafa Güler, MD, Altuğ Tuncer, MD, Bahadır Dağlar, MD, Gökhan İpek, MD, Ömer Işık, MD, and

More information

42yr Old Male with Severe AR Mild LV dysfunction s/p TOF -AV Replacement(tissue valve) or AoV plasty- Kyung-Hwan Kim

42yr Old Male with Severe AR Mild LV dysfunction s/p TOF -AV Replacement(tissue valve) or AoV plasty- Kyung-Hwan Kim 42yr Old Male with Severe AR Mild LV dysfunction s/p TOF -AV Replacement(tissue valve) or AoV plasty- Kyung-Hwan Kim Current Guideline for AR s/p TOF Surgery is reasonable in adults with prior repair of

More information

The diameter of the aortic valve is in direct proportion

The diameter of the aortic valve is in direct proportion The CarboMedics Top-Hat Supraannular Prosthesis José M. Bernal, MD, Rafael Martin-Duran, MD, José M. Rabasa, MD, and José M. Revuelta, MD Departments of Cardiovascular Surgery and Echocardiography, Hospital

More information

Medtronic Mosaic porcine bioprosthesis: Assessment of 12-year performance

Medtronic Mosaic porcine bioprosthesis: Assessment of 12-year performance Medtronic Mosaic porcine bioprosthesis: Assessment of 12-year performance W. R. Eric Jamieson, MD, a Friedrich-Christian Riess, MD, b Peter J. Raudkivi, MD, c Jacques Metras, MD, d Edward F. G. Busse,

More information

The clinical experience reported in recent Western series has provided

The clinical experience reported in recent Western series has provided Surgery for Acquired Cardiovascular Disease Yu et al Long-term evaluation of Carpentier-Edwards porcine bioprosthesis for rheumatic heart disease Hsi-Yu Yu, MD a Yi-Lwun Ho, MD b Shu-Hsun Chu, MD c Yih-Sharng

More information

Long-Term Results With the Medtronic-Hall Valvular Prosthesis

Long-Term Results With the Medtronic-Hall Valvular Prosthesis Long-Term Results With the Medtronic-Hall Valvular Prosthesis Cary W. Akins, MD Cardiac Surgical Unit, Massachusetts General Hospital, Boston, Massachusetts Background. Although more than 170,000 Medtronic-

More information

Surgery for Acquired Cardiovascular Disease

Surgery for Acquired Cardiovascular Disease Performance of bioprostheses and mechanical prostheses assessed by composites of valve-related complications to 15 years after mitral valve replacement W. R. E. Jamieson, MD, O. von Lipinski, MD, R. T.

More information

RESEARCH AND REVIEWS: JOURNAL OF PHARMACOLOGY AND TOXICOLOGICAL STUDIES

RESEARCH AND REVIEWS: JOURNAL OF PHARMACOLOGY AND TOXICOLOGICAL STUDIES e-issn:2322-0139 RESEARCH AND REVIEWS: JOURNAL OF PHARMACOLOGY AND TOXICOLOGICAL STUDIES Comparative Evaluation of Safety Outcomes of Different Prosthetic Valves in Indian Subjects. Kama Raval 1 *, Reena

More information

Long-Term Survival After Bovine Pericardial Versus Porcine Stented Bioprosthetic Aortic Valve Replacement: Does Valve Choice Matter?

Long-Term Survival After Bovine Pericardial Versus Porcine Stented Bioprosthetic Aortic Valve Replacement: Does Valve Choice Matter? Long-Term Survival After Bovine Pericardial Versus Stented Bioprosthetic Aortic Valve Replacement: Does Valve Choice Matter? Asvin M. Ganapathi, MD, Brian R. Englum, MD, Jeffrey E. Keenan, MD, Matthew

More information

Intensity of oral anticoagulation after implantation of St. Jude Medical mitral or multiple valve replacement: lessons learned from GELIA (GELIA 5)

Intensity of oral anticoagulation after implantation of St. Jude Medical mitral or multiple valve replacement: lessons learned from GELIA (GELIA 5) European Heart Journal Supplements () 3 (Supplement Q), Q39 Q43 Intensity of oral anticoagulation after implantation of St. Jude Medical mitral or multiple valve replacement: lessons learned from GELIA

More information

The risk-benefit ratio of mitral valve operation is

The risk-benefit ratio of mitral valve operation is Degenerative Mitral Regurgitation: When Should We Operate? Malcolm J. R. Dalrymple-Hay, PhD, Mark Bryant, Richard A. Jones, MRCP, Stephen M. Langley, FRCS, Steven A. Livesey, FRCS, and James L. Monro,

More information

Aortic valve replacement: is porcine or bovine valve better?

Aortic valve replacement: is porcine or bovine valve better? Interactive CardioVascular and Thoracic Surgery Advance Access published December 4, 2012 Interactive CardioVascular and Thoracic Surgery (2012) 1 13 doi:10.1093/icvts/ivs447 BEST EVIDENCE TOPIC Aortic

More information

Results of Aortic Valve Preservation and Repair

Results of Aortic Valve Preservation and Repair Results of Aortic Valve Preservation and Repair Department of Cardiothoracic and Vascular Surgery Cliniques Universitaires St. Luc Brussels, Belgium Gebrine Elkhoury Institutional experience in AV preservation

More information

Isolated Mitral Valve Replacement with the Hancock Bioprosthesis: A 13-Year Appraisal

Isolated Mitral Valve Replacement with the Hancock Bioprosthesis: A 13-Year Appraisal Isolated Mitral Valve Replacement with the Hancock ioprosthesis: A 13-Year Appraisal Vincenzo Gallucci, M.D., berto ortolotti, M.D., Aldo Milano, M.D., Carlo Valfrk, M.D., Alessandro Mazzucco, M.D., and

More information

Eight-Year Results of Aortic Root Replacement With the Freestyle Stentless Porcine Aortic Root Bioprosthesis

Eight-Year Results of Aortic Root Replacement With the Freestyle Stentless Porcine Aortic Root Bioprosthesis Eight-Year Results of Aortic Root Replacement With the Freestyle Stentless Porcine Aortic Root Bioprosthesis Neal D. Kon, MD,* Robert D. Riley, MD, Sandy M. Adair, RN, Dalane W. Kitzman, MD, and A. Robert

More information

Long-term relative survival after primary heart valve replacement 1

Long-term relative survival after primary heart valve replacement 1 European Journal of Cardio-thoracic Surgery 11 (1997) 81 91 Long-term relative survival after primary heart valve replacement 1 Abstract E. Ståhle a, *, P. Kvidal b, S.O. Nyström a, R. Bergström c a Department

More information

Standarized definition of bioprosthetic valve deterioration and failure

Standarized definition of bioprosthetic valve deterioration and failure Translational aortic valve research. From biology to treatment Standarized definition of bioprosthetic valve deterioration and failure Anna Sonia Petronio, MD, FESC Head of Cardiac Catheterization Lab

More information

Expanding Relevance of Aortic Valve Repair Is Earlier Operation Indicated?

Expanding Relevance of Aortic Valve Repair Is Earlier Operation Indicated? Expanding Relevance of Aortic Valve Repair Is Earlier Operation Indicated? RM Suri, V Sharma, JA Dearani, HM Burkhart, RC Daly, LD Joyce, HV Schaff Division of Cardiovascular Surgery, Mayo Clinic, Rochester,

More information

Hemodynamic Performance of the Medtronic Mosaic Porcine Bioprosthesis Up to Ten Years

Hemodynamic Performance of the Medtronic Mosaic Porcine Bioprosthesis Up to Ten Years Hemodynamic Performance of the Medtronic Mosaic Porcine Bioprosthesis Up to Ten Years Friedrich-Christian Riess, MD, Ralf Bader, MD, Eva Cramer, MD, Lorenz Hansen, MD, Bèr Kleijnen, MS, Gunther Wahl, MD,

More information

Indication, Timing, Assessment and Update on TAVI

Indication, Timing, Assessment and Update on TAVI Indication, Timing, Assessment and Update on TAVI Swedish Heart and Vascular Institute Ming Zhang MD PhD Interventional Cardiology Structure Heart Disease Conflict of Interest None Starr- Edwards Mechanical

More information

Emergency Intraoperative Echocardiography

Emergency Intraoperative Echocardiography Emergency Intraoperative Echocardiography Justiaan Swanevelder Department of Anaesthesia, Glenfield Hospital University Hospitals of Leicester NHS Trust, UK Carl Gustav Jung (1875-1961) Your vision will

More information

The Edge-to-Edge Technique f For Barlow's Disease

The Edge-to-Edge Technique f For Barlow's Disease The Edge-to-Edge Technique f For Barlow's Disease Ottavio Alfieri, Michele De Bonis, Elisabetta Lapenna, Francesco Maisano, Lucia Torracca, Giovanni La Canna. Department of Cardiac Surgery, San Raffaele

More information

Aortic Root Replacement With the Carboseal Composite Graft: 7-Year Experience With the First 100 Implants

Aortic Root Replacement With the Carboseal Composite Graft: 7-Year Experience With the First 100 Implants Aortic Root Replacement With the Carboseal Composite Graft: 7-Year Experience With the First 100 Implants Giovanni Battista Luciani, MD, Gianluca Casali, MD, Luca Barozzi, MD, and Alessandro Mazzucco,

More information

Dvnamic Behavior of Prosthetic Aortic Tissue Vdves as Viewed by High-speed Cinematography

Dvnamic Behavior of Prosthetic Aortic Tissue Vdves as Viewed by High-speed Cinematography Dvnamic Behavior of Prosthetic Aortic Tissue Vdves as Viewed by High-speed Cinematography W. Gerald Rainer, M.D., Robert A. Christopher, Ph.D., Theodore R. Sadler, Jr., M.D., and Alan D. Hilgenberg, M.D.

More information

The CarboMedics prosthetic valve (Sulzer CarboMedics,

The CarboMedics prosthetic valve (Sulzer CarboMedics, Long-Term Result of 1144 CarboMedics Mechanical Valve Implantations Chang Hyun Kang, MD, Hyuk Ahn, MD, Kyung Hwan Kim, MD, and Ki-Bong Kim, MD Department of Thoracic and Cardiovascular Surgery, Seoul National

More information

Twenty-year experience with the St Jude Medical mechanical valve prosthesis

Twenty-year experience with the St Jude Medical mechanical valve prosthesis Surgery for Acquired Cardiovascular Disease Ikonomidis et al Twenty-year experience with the St Jude Medical mechanical valve prosthesis John S. Ikonomidis, MD, PhD John M. Kratz, MD Arthur J. Crumbley

More information

PATIENT BOOKLET MEDTRONIC SURGICAL VALVE REPLACEMENT. Tissue Valve for Aortic and Mitral Valve Replacement

PATIENT BOOKLET MEDTRONIC SURGICAL VALVE REPLACEMENT. Tissue Valve for Aortic and Mitral Valve Replacement PATIENT BOOKLET MEDTRONIC SURGICAL VALVE REPLACEMENT Tissue Valve for Aortic and Mitral Valve Replacement ARE MEDTRONIC SURGICAL TISSUE HEART VALVES RIGHT FOR YOU? Medtronic surgical heart valves are for

More information

The pulmonary valve is the most common heart valve

The pulmonary valve is the most common heart valve Biologic versus Mechanical Valve Replacement of the Pulmonary Valve After Multiple Reconstructions of the RVOT Tract S. Adil Husain, MD, and John Brown, MD Indiana University School of Medicine, Department

More information

Long-term Experience with the Bjork-Shiley Monostrut Tilting Disc Valve

Long-term Experience with the Bjork-Shiley Monostrut Tilting Disc Valve J Korean Med Sci 2007; 22: 1060-4 ISSN 1011-8934 DOI: 10.3346/jkms.2007.22.6.1060 Copyright The Korean Academy of Medical Sciences Long-term Experience with the Bjork-Shiley Monostrut Tilting Disc Valve

More information

Eleven years experience with the Biocor stentless aortic bioprosthesis: clinical and hemodynamic follow-up with long-term relative survival rate

Eleven years experience with the Biocor stentless aortic bioprosthesis: clinical and hemodynamic follow-up with long-term relative survival rate European Journal of Cardio-thoracic Surgery 22 (2002) 912 921 www.elsevier.com/locate/ejcts Eleven years experience with the Biocor stentless aortic bioprosthesis: clinical and hemodynamic follow-up with

More information

Aortic valve replacement in predominant aortic stenosis: What is an appropriate size valve?

Aortic valve replacement in predominant aortic stenosis: What is an appropriate size valve? IJTCVS Joshi et al 141 Aortic valve replacement in predominant aortic stenosis: What is an appropriate size valve? Kishore Joshi, M.Ch., Sachin Talwar, M.Ch., Devagourou Velayoudham, M.Ch., Arkalgud Sampath

More information

VALVE REPAIR VERSUS REPLACEMENT FOR MITRAL INSUFFICIENCY: WHEN IS A MECHANICAL VALVE STILL INDICATED?

VALVE REPAIR VERSUS REPLACEMENT FOR MITRAL INSUFFICIENCY: WHEN IS A MECHANICAL VALVE STILL INDICATED? VALVE REPAIR VERSUS REPLACEMENT FOR MITRAL INSUFFICIENCY: WHEN IS A MECHANICAL VALVE STILL INDICATED? Eugene A. Grossi, MD Aubrey C. Galloway, MD Jeffrey S. Miller, MD Greg H. Ribakove, MD Alfred T. Culliford,

More information

Long-term results of a strategy of aortic valve repair in the paediatric population: Should we avoid cusp extension?

Long-term results of a strategy of aortic valve repair in the paediatric population: Should we avoid cusp extension? Long-term results of a strategy of aortic valve repair in the paediatric population: Should we avoid cusp extension? Y d Udekem, J Siddiqui, C Seaman, I Konstantinov, J Galati, M Cheung, C Brizard Royal

More information

Indications and Late Results of Aortic Valve Repair

Indications and Late Results of Aortic Valve Repair Indications and Late Results of Aortic Valve Repair Prof. Gebrine El Khoury Department of Cardiovascular and Thoracic Surgery Cliniques St. Luc Brussels, Belgium Aortic Valve Repair Question # 1 Can the

More information

Management of Difficult Aortic Root, Old and New solutions

Management of Difficult Aortic Root, Old and New solutions Management of Difficult Aortic Root, Old and New solutions Hani K. Najm MD, Msc, FRCSC,, FACC, FESC Chairman, Pediatric and Congenital Heart Surgery Cleveland Clinic Conflict of Interest None Difficult

More information

The Tricuspid Valve: The Not So Forgotten Valve. Manuel J Antunes Cardiothoracic Surgery Coimbra, Portugal

The Tricuspid Valve: The Not So Forgotten Valve. Manuel J Antunes Cardiothoracic Surgery Coimbra, Portugal The Tricuspid Valve: The Not So Forgotten Valve Manuel J Antunes Cardiothoracic Surgery Coimbra, Portugal No Conflicts of Interest to declare with regards to this subject 2 INCIDENCE OF TRICUSPID REGURGITATION

More information

Echocardiographic Evaluation of Mitral Valve Prostheses

Echocardiographic Evaluation of Mitral Valve Prostheses Echocardiographic Evaluation of Mitral Valve Prostheses Dennis A. Tighe, M.D., FACC, FACP, FASE Cardiovascular Medicine University of Massachusetts Medical School Worcester, MA www.asecho.org 1 Nishimura

More information

Mitroflow Aortic Bioprosthesis 5-Year Follow-Up: North American Prospective Multicenter Study

Mitroflow Aortic Bioprosthesis 5-Year Follow-Up: North American Prospective Multicenter Study Mitroflow Aortic Bioprosthesis 5-Year Follow-Up: North American Prospective Multicenter Study Federico M. Asch, MD, David Heimansohn, MD, Daniel Doyle, MD, Walter Dembitsky, MD, Francis D. Ferdinand, MD,

More information

TAVI- Is Stroke Risk the Achilles Heel of Percutaneous Aortic Valve Repair?

TAVI- Is Stroke Risk the Achilles Heel of Percutaneous Aortic Valve Repair? TAVI- Is Stroke Risk the Achilles Heel of Percutaneous Aortic Valve Repair? Elaine E. Tseng, MD and Marlene Grenon, MD Department of Surgery Divisions of Adult Cardiothoracic and Vascular and Endovascular

More information

Stentless aortic valves. Current aspects

Stentless aortic valves. Current aspects Endorsed by proceedings in Intensive Care Cardiovascular Anesthesia EXPERT OPINION HSR Proceedings in Intensive Care and Cardiovascular Anesthesia 2012; 4(2): 77-82 Stentless aortic valves. Current aspects

More information

14 Valvular Stenosis

14 Valvular Stenosis 14 Valvular Stenosis 14-1. Valvular Stenosis unicuspid valve FIGUE 14-1. This photograph shows severe valvular stenosis as it occurs in a newborn. There is a unicuspid, horseshoe-shaped leaflet with a

More information

Incidence of prosthesis-patient mismatch in patients receiving mitral Biocor porcine prosthetic valves

Incidence of prosthesis-patient mismatch in patients receiving mitral Biocor porcine prosthetic valves INTERVENTION/VALVULAR HEART DISEASE ORIGINAL ARTICLE Cardiology Journal 2016, Vol. 23, No. 2, 178 183 DOI: 10.5603/CJ.a2016.0011 Copyright 2016 Via Medica ISSN 1897 5593 Incidence of prosthesis-patient

More information

Replacement of the mitral valve in the presence of

Replacement of the mitral valve in the presence of Mitral Valve Replacement in Patients with Mitral Annulus Abscess Christopher M. Feindel Replacement of the mitral valve in the presence of an abscess of the mitral annulus presents a major challenge to

More information

Mechanical vs. Bioprosthetic Aortic Valve Replacement: Time to Reconsider? Christian Shults, MD Cardiac Surgeon, Medstar Heart and Vascular Institute

Mechanical vs. Bioprosthetic Aortic Valve Replacement: Time to Reconsider? Christian Shults, MD Cardiac Surgeon, Medstar Heart and Vascular Institute Mechanical vs. Bioprosthetic Aortic Valve Replacement: Time to Reconsider? Christian Shults, MD Cardiac Surgeon, Medstar Heart and Vascular Institute Assistant Professor, Georgetown School of Medicine

More information

Heart Valves: Before and after surgery

Heart Valves: Before and after surgery Heart Valves: Before and after surgery Tim Sutton, Consultant Cardiologist Middlemore Hospital, Auckland Auckland Heart Group Indications for intervention in Valvular disease To prevent sudden death and

More information

Late Results with Autogenous

Late Results with Autogenous Late Results with Autogenous Tissue Heart Valves W. Sterling Edwards, M.D. ABSTRACT Five years of experience using unsupported and mounted fascia lata and pericardial heart valves in human beings is summarized.

More information

T annulus (521 mm in diameter) is tempered by concerns

T annulus (521 mm in diameter) is tempered by concerns Clinical and Hemodynamic Performance of the 19-mm Carpentier-Edwards Porcine Bioprosthesis Robert M. Bojar, MD, Hassan Rastegar, MD, Douglas D. Payne, MD, Charles A. Mack, MD, and Steven L. Schwartz, MD

More information

Repeated mitral valve replacement in a patient with extensive annular calcification

Repeated mitral valve replacement in a patient with extensive annular calcification CASE REPORT Open Access Repeated mitral valve replacement in a patient with extensive annular calcification Tadashi Kitamura 1,2*, Sachito Fukuda 1, Takahiro Sawada 1, Sumio Miura 1, Ikutaro Kigawa 1,3

More information

Trifecta Valve. Clinical Compendium. Five Year Data

Trifecta Valve. Clinical Compendium. Five Year Data Trifecta Valve Clinical Compendium Five Year Data Trifecta Valve Compendium INTRODUCTION The Trifecta valve is a tri-leaflet stented pericardial valve designed for supra-annular placement in the aortic

More information

and Coronary Artery Surgery George M. Callard, M.D., John B. Flege, Jr., M.D., and Joseph C. Todd, M.D.

and Coronary Artery Surgery George M. Callard, M.D., John B. Flege, Jr., M.D., and Joseph C. Todd, M.D. Combined Valvular and Coronary Artery Surgery George M. Callard, M.D., John B. Flege, Jr., M.D., and Joseph C. Todd, M.D. ABSTRACT Between July, 97, and March, 975,45 patients underwent combined valvular

More information