Measurement Error in Grip and Pinch Force Measurements in Patients With Hand Injuries

Size: px
Start display at page:

Download "Measurement Error in Grip and Pinch Force Measurements in Patients With Hand Injuries"

Transcription

1 Research Report Measurement Error in Grip and Pinch Force Measurements in Patients With Hand Injuries Background and Purpose. There is limited documentation of measurement error of grip and pinch force evaluation methods. The purposes of this study were (1) to determine indexes of measurement error for intraexaminer and interexaminer measurements of grip and pinch force in patients with hand injuries and (2) to investigate whether the measurement error differs between measurements of the injured and noninjured hands and between experienced and inexperienced examiners. Subjects. The subjects were a consecutive sample of 33 patients with hand injuries who were seen in the Department of Rehabilitation Medicine of Erasmus MC University Medical Center Rotterdam in the Netherlands. Methods. Repeated measurements were taken of grip and pinch force, with a short break of 2 to 3 minutes between sessions. For the grip force in 2 handle positions (distance between handles of 4.6 and 7.2 cm, respectively), tip pinch (with the index finger on top and the thumb below, with the other fingers flexed) and key pinch force (with the thumb on top and the radial side of the index finger below) data were obtained on both hands of the subjects by an experienced examiner and an inexperienced examiner. Intraclass correlation coefficients (ICCs), standard errors of measurement (SEMs), and associated smallest detectable differences (SDDs) were calculated and compared with data from previous studies. Results. The reliability of the measurements was expressed by ICCs between.82 and.97. For grip force measurements (in the second handle position) by the experienced examiner, an SDD of 61 N was found. For tip pinch and key pinch, these values were 12 N and 11 N, respectively. For measurements by the inexperienced examiner, SDDs of 56 N for grip force and 13 N and 18 N for tip pinch and key pinch were found. Discussion and Conclusion. Based on the SEMs and SDDs, in individual patients only relatively large differences in grip and pinch force measurements can be adequately detected between consecutive measurements. Measurement error did not differ between injured and noninjured hands or between experienced and inexperienced examiners. Criteria for judging whether the measurement error does allow application of the measurements in individual patients are discussed. [Schreuders TAR, Roebroeck ME, Goumans J, et al. Measurement error in grip and pinch force measurements in patients with hand injuries. Phys Ther. 2003;83: ] Key Words: Hand, Hand force, Hand injuries, Measurement error, Reliability of results. Ton AR Schreuders, Marij E Roebroeck, Janine Goumans, Johan F van Nieuwenhuijzen, Theo H Stijnen, Henk J Stam 806 Physical Therapy. Volume 83. Number 9. September 2003

2 Only data that have acceptable reliability and validity are valuable in the clinical decisionmaking process for determination of impaired function and comparison of surgical repair techniques to document progress during rehabilitation and to evaluate disability after injury. Such data will decrease the need for use of unsubstantiated opinions and will increase the physical therapist s ability to obtain reproducible findings and meaningful results. 1 Better reliability implies better precision of single measurements, which is a prerequisite for better tracking of changes in measurements in research and clinical practice. 2 Generally accepted instruments in the evaluation of hand function measure grip and pinch force. Grip and pinch force measurements have been promoted as an important measure of outcome in, for example, the evaluation of peripheral nerve function. 3 5 For a correct interpretation of these force measurements, we contend information on the reliability (measurement error) of these measurements is required. The reliability of grip force measurements has been studied in patients with hand injuries 6,7 and in people without impairments of the hand. 8,9 It also has been studied in several other categories of patients, including women with nonspecific regional pain (NSRP) 10 and patients with reflex sympathetic dystrophy (RSD) 11 or epicondylitis, 12 and in people after a stroke. 13 The reliability of pinch force measurements has been studied in patients with RSD 11 and repetitive strain injury (RSI). 14 The studies of Spijkerman et al 6 and Brown et al 7 on the reliability of force measurements in patients with hand injuries have important drawbacks. The study by Spijkerman et al included only 8 patients, and Brown et al, in their study of grip and pinch force measurements in patients with hand injuries, did not specify the measurement error. As in many studies, the intraclass correlation coefficient (ICC) was used as a measure of reliability. For TAR Schreuders, PT, is Physical Therapist, Department of Rehabilitation Medicine, Hand Therapy Unit, Erasmus MC University Medical Center Rotterdam, PO Box 2040, 3000 CA Rotterdam, the Netherlands (a.schreuders@erasmusmc.nl). Address all correspondence to Mr Schreuders. ME Roebroeck, PhD, is Assistant Professor, Department of Rehabilitation Medicine, Erasmus MC University Medical Center Rotterdam. J Goumans, MD, is Medical Student, Erasmus MC University Medical Center Rotterdam. JF van Nieuwenhuijzen, PT, is Research Therapist, Department of Rehabilitation Medicine, Erasmus MC University Medical Center Rotterdam. TH Stijnen, PhD, is Professor of Biostatistics, Department of Epidemiology and Biostatistics, Erasmus MC University Medical Center Rotterdam. HJ Stam, MD, PhD, is Professor of Rehabilitation Medicine and Chairman, Department of Rehabilitation Medicine, Erasmus MC University Medical Center Rotterdam. Mr Schreuders, Dr Roebroeck, and Dr Stam provided concept/idea/research design. Mr Schreuders and Dr Roebroeck provided writing. Dr Goumans and Mr van Nieuwenhuijzen provided data collection, and Mr Schreuders, Dr Roebroeck, Dr Goumans, and Dr Stijnen provided data analysis. Mr Schreuders and Dr Goumans provided project management. Dr Roebroeck and Dr Stam provided fund procurement and facilities/equipment. Dr Stijnen and Dr Stam provided consultation (including review of manuscript before submission). The Ethical Committee of the Erasmus Medical Center, the Netherlands, approved this study. This article was received February 24, 2003, and was accepted May 19, Physical Therapy. Volume 83. Number 9. September 2003 Schreuders et al. 807

3 force). A third aim was to judge whether the measurement error is small enough to justify use of these measurements to discern real changes in grip and pinch force in individual patients. Because no clear criteria to judge the SDD are available for grip and pinch force measurements, we compared our findings with the results of studies in which grip and pinch force measurements were examined in different patient categories (ie, patients with hand injuries, pain, RSD, or epicondylitis and people after stroke). Method Figure 1. Measurements were performed with the participant seated at a table on which the dynamometers were positioned. clinical use, however, we believe the most relevant information is the magnitude of change between tests that is required to detect a real change, preferably quantified in the same units as the force measurement. 10,15 This magnitude of change is specified in the smallest detectable difference (SDD), which is calculated from the standard error of measurement (SEM), the absolute error component of the measurements. 16,17 The smaller the measurement error, the better the measure. 18 Various factors relating to the patient s condition may influence the measurement error, such as the amount of pain and loss of normal function of the fingers or thumb after injury. The experience of the examiner also might influence the measurement error. However, studies on these aspects of grip and pinch force measurements for patients with hand injuries are scarce. One aim of our study was to assess the SEM and related indexes of reliability for intraexaminer and interexaminer applications of grip (2 handle positions, with distances between the handles of 4.6 and 7.2 cm) and pinch (tip pinch and key pinch) force measurements for patients with hand injuries obtained with Lode handgrip and pinch-grip dynamometers (Fig. 1).* A second aim was to investigate whether the measurement error of grip and pinch force measurements differs between specific applications, such as measurements of injured and noninjured hands, measurements obtained by experienced and inexperienced examiners, and measurements between 2 handle positions for grip force and 2 different types of pinch force (ie, tip and key pinch Subjects A consecutive sample of 33 patients (20 male, 13 female) who were seen in the Department of Rehabilitation Medicine of Erasmus MC University Medical Center Rotterdam in the Netherlands participated in the study. They had a mean age of 36 years (SD 13.7, range 17 67). All patients had an injury on only one hand (6 with tendon lacerations, 9 with nerve lesions, 4 with fractures, and 14 with a combination of lesions or surgeries [eg, crush injury, finger amputation, arthrodesis]). During a period of 5 months, patients who were capable of being evaluated for grip and pinch force were asked to participate in this study. Patients who were not permitted to grip or pinch with maximum force because of injured tissue and patients who complained of pain were excluded. We presumed that pain would inhibit the patients from exerting full maximum force and that force measurements may reflect the amount of pain rather than the force-generating capacity of the patients. The participants were informed about the purpose of the study and gave informed consent. The participants were in different phases of their rehabilitation process. Study entry was, on average, 22 months after injury or surgery when including one patient who had a 16-year period between the date of injury and time of examination. Excluding this patient, the average period was 9 months (SD 7.0, range 1 28) since injury or surgery. Force measurements of the hand are routinely obtained in our department during different phases of the rehabilitation process. Such measurements can be used to determine progression of muscle force recovery, the ability to perform in a vocation, and indications for continuing therapy and to assess longterm outcome after hand injuries. Instrumentation We used Lode handgrip and pinch-grip dynamometers (Fig. 1). The distances between the handles of the Lode handgrip dynamometer were adapted to create spaces between the handles comparable to those of the Jamar * Lode Medical Technology, Zernikepark 16, 9747 AN Groningen, the Netherlands. Distributed in the United States by ElectraMed Corp, G-5332 Hill-23 Dr, Flint, MI Schreuders et al Physical Therapy. Volume 83. Number 9. September 2003

4 hand dynamometer. For our study, the grip handle positions 2 (distance between handles 4.6 cm) and 4 (distance between handles 7.2 cm), which we designated grip 2 and grip 4, respectively, were used with the handgrip dynamometer. With the Lode pinch-grip dynamometer, which is similar in design to the Preston pinch dynamometer,,19 2 types of pinch were measured: tip pinch and key pinch. Tip pinch force was measured with only the index finger on top and the other fingers flexed with the thumb below, and key pinch force was measured with the thumb on top and the radial side of the index finger below. The grip handle is connected to an amplifier from which the values give a digital readout on a display. According to the manufacturer, the measuring range is 0 to 1,000 N, with an accuracy of 1% deviation for 2 to 500 N and 2% for 500 to 900 N. Calibration was done with suspended weights according to the method described by Mathiowetz et al. 8 We believe the use of strain gauge technology is preferable to dynamometers, which are spring-based or use a hydraulic pressure system, because such dynamometers might produce erroneous data due to the wear and tear of metal, slow leaks, and hysteresis. 20 Procedure Measurements were taken according to the recommendations of the American Society of Hand Therapists (ASHT). 21 Participants were seated at a table on which the dynamometers were positioned (Fig. 1). The subjects were told to keep their elbow flexed without resting their arm or the grip handle of the dynamometer on the table. The digital display was not visible to the subjects. Corresponding to the ASHT recommendations for each measurement, the mean of 3 repetitions was recorded. Measurements were obtained of the left and right hands alternately, which is the method of testing used at our clinic. Testing both hands, we contend, enables comparisons of the injured and noninjured hands, as proposed by Gaul. 22 The side (injured or noninjured) at which measurements were started and the order of tests (grip 2, grip 4, tip pinch, or key pinch) were randomly selected. Both examiners obtained all the measurements twice for each subject; the order of examiners was randomly selected. During each measurement session, the examiners obtained 12 measurements (2 grip and 2 pinch measurements 3 repetitions from one hand. Then the examiners were changed while the subject took a short break (2 3 minutes) and moved the arm and hand out of the testing position. In this way, we attempted to reflect practice with examiners independently testing subjects, except there was a short time interval between examiners. Both examiners participated in 2 sessions per patient, which resulted in 48 measurements per hand. The total time required for testing, including the breaks, was 35 to 40 minutes per patient. Examiners Because we were interested in the effect of experience of the examiner on measurement error, we selected 2 examiners with widely differing experience. The first examiner was a 54-year-old male physical therapist ( JFvN) with 20 years of experience in testing hand function. The second examiner ( JG) was a 22-year-old female medical student with no previous experience with these tests. During a 2-week period, the inexperienced examiner had 5 hours of training by an experienced physical therapist to become familiarized with the equipment and testing protocol. During the training, measurements were practiced on several subjects without hand injuries and some patients with hand injuries. We believe that the results obtained by this student are reasonably representative of those that could be obtained by an inexperienced examiner, regardless of whether this is a starting physician or physical therapist, because at this stage they have comparable (ie, limited) knowledge, attitudes, and skills concerning force measurements. Data Analysis Statistical analyses were done to determine intraexaminer reliability of data obtained in the 2 sessions performed by the same examiner both for the experienced examiner (Texp1 and Texp2) and the inexperienced examiner (Tinx 1 and Tinx 2) separately. Interexaminer reliability was calculated between the first sessions of the experienced examiner (Texp1) and the inexperienced examiner (Tinx1). Analyses of the measurements of the injured hands (n 33) and noninjured hands (n 33) were conducted separately. An analysis of variance (ANOVA) was conducted with an SPSS/PC program to determine the multiple sources of measurement error. The variance attributed to differences among participants was designated Var (P). The variance attributed to differences between sessions by the same examiner (Texp1 and Texp2, Tinx1 and Tinx2) was designated Var (S). The variance ascribed to the different examiners (Texp1 and Tinx1) was designated Var (T). Variance due to interaction among participants for intraexaminer sessions was designated Var (P S), and variance due to interaction among participants for interexaminer sessions was designated Var (P T). The interaction components (P S) and (P T) were confounded by the residual Sammons Preston/Rolyan, 4 Sammons Ct, Bolingbrook, IL SPSS Inc, 233 S Wacker Dr, Chicago, IL Physical Therapy. Volume 83. Number 9. September 2003 Schreuders et al. 809

5 Table 1. Grip and Pinch Force Measurements (in Newtons) Obtained by Experienced and Inexperienced Examiners for Both Hands of Participants With Hand Injury of One Hand (N 33) Hand Test a Experienced Examiner Inexperienced Examiner Session 1 Session 2 Session 1 Session 2 X SD X SD X SD X SD Injured Grip Grip Tip pinch Key pinch Noninjured Grip Grip Tip pinch Key pinch a Grip 2 measurements of grip force in the second handle position, grip 4 measurements of grip force in the fourth handle position. error. In the intraexaminer analyses, Var (S) and Var (P S) (ie, the variance within participants) constituted the error variance. Correspondingly, for the interexaminer analyses, the variance components Var (T) and Var (P T) constituted the error variance. From the error variance, the SEM was computed as its square root. 17 Based on the SEM, the SDD with 95% confidence was calculated as SEM. 23 This SDD can be applied in such a way that only differences between 2 consecutive measurements greater than the SDD can be interpreted with 95% certainty as real change in grip or pinch force. Intraclass correlation coefficients were computed as the ratio of variances among participants (ie, Var [P]) and the total variance. 15 Comparing SEMs In order to test whether differences existed in error variances of different applications of grip and pinch force measurements, a test of equality of variances for paired samples was applied. In this analysis, scores between sessions were compared by calculating the correlation coefficient (Pearson r) between the sum and difference of the difference scores and testing whether this correlation differed from zero. 16,24(pp ) In this way, we tested whether differences existed in error variance, and thus SEM and SDD, between the experienced and inexperienced examiners, between measurements of the injured and noninjured hands, and between the 2 handle positions of the grip force measurements and between the 2 pinch force measurements. Commonly accepted criteria to judge whether the SEM and SDD of measurements are adequate for application of the measurements to individual patients do not exist. Therefore, we compared the SEMs found in our study with those from other studies in which grip and pinch force measurements were investigated. For an adequate comparison, We selected studies from the literature in Figure 2. Scatterplot of intraexaminer grip force measurements with the second handle position obtained in 2 sessions by an experienced examiner (Texp1 and Texp2). represents measurements of the injured hands (N 33); F represents measurements of the noninjured hands (N 33), which are the higher values. which values for SEM or SDD were reported and studies from which SEM and SDD can be estimated by using the formulas: SEM SD (1 ICC) 15(p119),23 or SEM SD difference / 2, 15(p120) where SD standard deviation. These studies included similar groups of patients as in our study (ie, patients with hand injuries or patients from different groups such as patients with pain, RSD, epicondylitis, and stroke). Results Descriptive values (mean and standard deviations) were calculated for all measurements (2 grip and 2 pinch measurements) for the injured and noninjured hands and for the 4 sessions conducted by the experienced and inexperienced examiners (Tab. 1). Evidently noninjured 810. Schreuders et al Physical Therapy. Volume 83. Number 9. September 2003

6 inexperienced examiners. Similarly, no differences were found in the error variances of the measurements between the 2 handle positions of grip force or between the 2 pinch techniques. Tables 4 and 5 present comparisons among studies investigating grip and pinch force measurements, including the type of dynamometer used, characteristics of the subjects, means and standard deviations of force measurements, and reliability indexes (ie, SEM, SDD, and ICC). The measurement errors of the SEMs of grip force measurements are comparable in most of the studies, except for those of Nitschke et al 10 and Smidt et al, 12 in which relatively low values were found (Tab. 4). For pinch force measurements, the SEMs were 4 or 5 N, with the exception of the higher values of 11 N found by Geertzen et al 11 (Tab. 5). Figure 3. Scatterplot of interexaminer tip pinch force measurements by an experienced examiner (Texp1) and an inexperienced examiner (Tinx1). represents measurements of the injured hands (N 33); F represents measurements of the noninjured hands (N 33). hands have stronger grip and pinch forces. The mean grip force in position 4 was less than in position 2. Key pinch had a considerably greater mean force than tip pinch. Two scatterplots of measurements obtained for both injured and noninjured hands (n 66) are presented in Figures 2 and 3. Figure 2 shows the measurements of grip force with the second handle position for the 2 sessions conducted by the experienced examiner (Texp1 and Texp2). Figure 3 shows the measurements of the tip pinch force obtained by the 2 different examiners (Texp1 and Tinx1). Both scatterplots demonstrate that there was a high level of correspondence between the 2 sessions and the 2 examiners. Table 2 contains the results of the ANOVA: the sum of squares, mean of squares, and estimated variance components attributed to different sources. Reliability indexes (ie, SEM, SDD, and ICC) for the intraexaminer grip and pinch measurements are presented, ordered by examiner (experienced and inexperienced) and by hand (injured and noninjured). The SEMs ranged from 4 to 29 N, and the corresponding SDDs ranged from 11 to 80 N. The lowest ICC was.82 and the highest ICCs were.97 which, according to the scale suggested by Shrout and Fleiss, 25 should be classified as excellent. The interexaminer data are presented in Table 3 and show comparable values. No differences were found in the error variances of grip and pinch force measurements between injured and noninjured hands or performed by experienced and Discussion In our study, high ICCs were found for the reliability of grip and pinch force measurements in subjects with hand injuries. In our view, however, a high ICC should not be interpreted as a small measurement error because, in an ICC, measurement error and real variability between subjects are expressed in relative terms. Expressing the reliability as a dimensionless ratio of variances does not allow us to interpret the reliability in terms of an individual score. To decide whether a person s grip force has changed after a period of rehabilitation, a physical therapist must know which part of the change measured was real and which part by is due to measurement error. To accomplish this, we calculated the SEM and SDD. In our study, the SDD of the grip force measurements (second handle position) of the injured hands was 61 N for measurements obtained by the experienced examiner (Tab. 2). This finding means that for grip force measurements taken by the experienced examiner (Texp), a change in force of at least 61 N is needed between 2 sessions to be 95% confident that a real change has occurred. The 2 different types of pinch force measurements (tip pinch and key pinch) showed a much smaller SDD; for the experienced examiner, the SDDs for the injured hands were 11 to 12 N (Tab. 2). In the case of our 2 examiners with widely differing experience, comparison of the SEMs showed no differences between them. Thus, for grip and pinch force measurements, it appears that a brief training period was sufficient for an inexperienced examiner to reach a level of reliability comparable to that of an experienced examiner. Given the small number of examiners in our study, however, this cannot be stated with certainty. Physical Therapy. Volume 83. Number 9. September 2003 Schreuders et al. 811

7 Table 2. Intraexaminer Measurements of Grip and Pinch Force (in Newtons) of Injured Hands of Participants (N 33) Obtained by Experienced and Inexperienced Examiners a Examiner Hand Test Source SS MS Estimated Variance SEM SDD ICC Experienced Injured Grip 2 Between patients Within patients Grip 4 Between patients Within patients Tip pinch Between patients Within patients Key pinch Between patients Within patients Inexperienced Injured Grip 2 Between patients Within patients Grip 4 Between patients Within patients Tip pinch Between patients Within patients Key pinch Between patients (n 32) Within patients Experienced Noninjured Grip 2 Between patients Within patients Grip 4 Between patients Within patients Tip pinch Between patients Within patients Key pinch Between patients Within patients Inexperienced Noninjured Grip 2 Between patients (n 32) Within patients Grip 4 Between patients Within patients Tip pinch Between patients Within patients Key pinch Between patients Within patients a Grip 2 measurements of grip force in the second handle position, grip 4 measurements of grip force in the fourth handle position. SS sum of squares, MS mean squares, SEM standard error of measurement, SDD smallest detectable difference, ICC intraclass correlation coefficient. More examiners, however, would entail a longer measurement time, which might fatigue the patient or decrease his or her concentration, both of which may influence the measurement error. We attempted to compensate for the small number of examiners by selecting 2 examiners with what we considered maximally different experience. Given that no differences in measurement error between the 2 examiners were found, we contend that having more examiners (eg, with intermediate levels of experience) obtain the measurements would not have affected our conclusions. Although we do not have data to support this view, we believe it is a reasonable interpretation. Other factors also may influence the measurement error. For example, we also examined the difference in SEMs between grip and pinch force measurements of the injured and noninjured hands. The measurement error, however, did not differ between the injured and noninjured hands. No differences were found between the SEMs of grip force measurements in the 2 handle positions or between both types of pinch force measurements. Therefore, for applications in clinical practice, we contend one SEM value can be applied for grip force measurements for both handle positions and one SEM value can be applied for pinch force measurements, irrespective of injury and experience of the examiner. Accepted criteria to judge the SDD of grip and pinch force measurements do not exist. Therefore, we compared our findings with those of 6 studies in which grip and pinch force measurements were investigated in similar and different patient categories (ie, patients with hand injuries, 6,7 women with NSRP, 10 patients with RSD, 11 patients with epicondylitis, 12 and patients after a stroke 23 ) (Tab. 4). The SEMs of grip force measurements in these different studies are comparable. However, the differences in means and ranges of the mea Schreuders et al Physical Therapy. Volume 83. Number 9. September 2003

8 Table 3. Interexaminer Measurements of Grip and Pinch Force of Injured and Noninjured Hands of Participants (N 33) a Hand Test Source SS MS Estimated Variance SEM (N) SDD (N) ICC Injured Grip 2 Between patients Within patients Grip 4 Between patients Within patients Tip pinch Between patients Within patients Key pinch Between patients (n 32) Within patients Noninjured Grip 2 Between patients Within patients Grip 4 Between patients Within patients Tip pinch Between patients Within patients Key pinch Between patients Within patients a Grip 2 measurements of grip force in the second handle position, grip 4 measurements of grip force in the fourth handle position. SS sum of squares, MS mean squares, SEM standard error of measurement, SDD smallest detectable difference, ICC intraclass correlation coefficient, N newtons. Table 4. Comparison of Intratester Measurements From 6 Grip Force Studies With Data From Our Study a Study Handheld Dynamometer Subjects n X SD SEM SDD ICC Spijkerman et al 6 Prototype Lode No injuries Hand injuries Brown et al 7 Dexter grip b Hand injuries Nitschke et al 10 Jamar Women with no injuries Women with NSRP Geertzen et al 11 CITEC grip c RSD Unaffected hand Affected hand Smidt et al 12 Jamar Epicondylitis Uninvolved hand e.98 Involved hand e.98 Boissy et al 13 Lafayette grip d Not affected Stroke Our study Lode grip handle Hand injuries Noninjured hand position 2 Injured hand Lode grip handle position 4 Noninjured hand Injured hand a NSRP nonspecific regional pain syndrome, RSD reflex sympathetic dystrophy, SEM standard error of measurement, SDD smallest detectable difference, ICC intraclass correlation coefficient (except for study by Nitschke et al, in which Pearson correlation coefficients were calculated). All data converted to newtons (data in studies by Nitschke et al, Geertzen et al, and Smidt et al were in kilograms of force and were converted to newtons by multiplying by 10; Brown and colleagues within-patient data were in pounds and were multiplied by 4.448). Data are for intratester measurements, except for Geertzen et al and Smidt et al, who reported intertester data. Data from Brown et al are the mean of 3 testers. Figures in italics are estimated from ICC values presented in the studies. b Dexter Extremity Evaluation & Therapy Systems, Cedaron Medical Inc, PO Box 2100, Davis, CA c CITEC hand-held dynamometer, CIT Technics, Rijksstraatweg 384, 9752 CR Haren, the Netherlands. d Lafayette Instrument Co, 3700 Sagamore Pkwy N, PO Box 5729, Lafayette, IN e Figures from study by Smidt et al are estimated from the SD diff values they reported. These values of the SDD differ from the values reported by Smidt et al, who submitted an erratum on this point. Physical Therapy. Volume 83. Number 9. September 2003 Schreuders et al. 813

9 Table 5. Comparison of Data From Earlier Pinch Force Measurement Studies With Data From Our Study (All Data in Newtons) a Study Handheld Dynamometer Subjects n Tester X SD SEM SDD ICC Geertzen et al 11 CITEC RSD Unaffected side Key pinch b Affected side Intertester b Brown et al 7 Dexter Hand injuries Our study Lode Hand injuries Tip pinch 30 Intratester (mean of Jaw pinch 3 testers) Noninjured hand Tip pinch 33 Intratester Injured hand Lode Noninjured hand Key pinch Injured hand a SEM standard error of measurement, SDD smallest detectable difference, ICC intraclass correlation coefficient, RSD reflex sympathetic dystrophy. b No data available. sured force levels in these studies are remarkable (eg, the mean grip force levels in the affected hands ranged from 84 to 300 N). To determine the value of a particular measurement in a population during clinical practice, we believe it is essential to know in which range the changes of muscle force take place. For example, in people with hand injury, the grip force will be lower shortly after the trauma. During later phases of rehabilitation, the increased muscle force will generally reach a plateau after many months. If the difference in the grip force between the start and end of the rehabilitation is, for example, 100 N, an SDD of 80 N will mean the measurement is inadequate to detect changes. If the range in which the forces change is as wide as 300 N, an SDD of 30 N is certainly sufficient to detect changes during the rehabilitation process. In this latter example, a virtual scale of 10 steps is achieved. To determine whether the calculated measurement error is small enough to make a test valuable in clinical practice, different methods have been applied in the studies we cite. Geertzen et al 11 concluded that the SDD is too large in relation to the mean force measured to be useful in patients with RSD. Smidt et al 12 suggested that an SDD less than 10% of the total range of the measurements would be acceptable. The range, however, is determined by the extreme values of the measurement, whereas the standard deviation is less influenced by the extreme values. In our study, therefore, we re-examined the SDD in relation to the standard deviation of the measurements by calculating the SDD/standard deviation ratios. The SDD/standard deviation ratios were 0.5 to 0.6 for grip force measurements using the second handle position and 0.6 to 0.7 when using the fourth handle position. These findings mean that 6 to 7 steps of changes in force level can be detected within the 95% distribution ( virtual scale ) of impaired to normal levels of grip force. In our opinion, this SDD/standard deviation ratio is useful in order to gain better insight as to whether measurements can be used to detect clinical meaningful changes. A more definitive conclusion concerning the usefulness of the measurements can be made by relating the SDD relative to the magnitude of changes in force measurements in a prospective study of patients throughout the entire rehabilitation process. Conclusions According to the ICC values obtained in our study, the reliability of grip and pinch force measurements is excellent. However, assessment of the measurement error and detectable change in muscle force between 2 consecutive measurements demonstrated that, in grip and pinch force measurements in individual patients, only relatively large changes can be adequately detected. Measurement error did not differ between the experienced and inexperienced examiners or the injured and noninjured hands. Similarly, no differences were found in the grip force measurements for the 2 handle positions or in the tip pinch and key pinch force measurements. The limited number of examiners and the use of subjects who were pain-free should be considered when applying our results. Further study of the SDD/standard deviation ratio is needed to develop clear criteria to judge whether measurement error is acceptable to detect changes in individual patients. References 1 Bear-Lehman J, Abreu BC. Evaluating the hand: issues in reliability and validity. Phys Ther. 1989;69: Beckerman H, Roebroeck ME, Lankhorst GJ, et al. Smallest real difference: a link between reproducibility and responsiveness. Qual Life Res. 2001;10: Dellon AL. A numerical grading scale for peripheral nerve function. J Hand Ther. 1993;6: Rosen B, Dahlin LB, Lundborg G. Assessment of functional outcome after nerve repair in a longitudinal cohort. Scand J Plast Reconstr Surg Hand Surg. 2000;34: Schreuders et al Physical Therapy. Volume 83. Number 9. September 2003

10 5 Strickland JW, Idler, RS, DelSignore JL. Ulnar nerve repair. In: Gelberman RH, ed. Operative Nerve Repair and Reconstruction. Vol 1. Philadelphia, Pa: JB Lippincott Co; 1991; Spijkerman DC, Snijders CJ, Stijnen T, Lankhorst GJ. Standardization of grip strength measurements: effects on repeatability and peak force. Scand J Rehabil Med. 1991;23: Brown A, Cramer LD, Eckhaus D, et al. Validity and reliability of the Dexter hand evaluation and therapy system in hand-injured patients. J Hand Ther. 2000;13: Mathiowetz V, Weber K, Volland G, Kashman N. Reliability and validity of grip and pinch strength evaluations. J Hand Surg [Am]. 1984;9: Hamilton A, Balnave R, Adams R. Grip strength testing reliability. J Hand Ther. 1994;7: Nitschke JE, McMeeken JM, Burry HC, Matyas TA. When is a change a genuine change? A clinically meaningful interpretation of grip strength measurements in healthy and disabled women. J Hand Ther. 1999;12: Geertzen JH, Dijkstra PU, Stewart RE, et al. Variation in measurements of grip strength: a study in reflex sympathetic dystrophy patients. Acta Orthop Scand Suppl. 1998;279: Smidt N, van der Windt DA, Assendelft WJ, et al. Interobserver reproducibility of the assessment of severity of complaints, grip strength, and pressure pain threshold in patients with lateral epicondylitis. Arch Phys Med Rehabil. 2002;83: Boissy P, Bourbonnais D, Carlotti MM, et al. Maximal grip force in chronic stroke subjects and its relationship to global upper extremity function. Clin Rehabil. 1999;13: MacDermid JC, Kramer JF, Woodbury MG, et al. Interrater reliability of pinch and grip strength measurements in patients with cumulative trauma disorders. J Hand Ther. 1994;7: Streiner DL, Norman GR. Health Measurement Scales: A Practical Guide to Their Development and Use. 2nd ed. Oxford, England: Oxford University Press; Stratford PW, Goldsmith CH. Use of the standard error as a reliability index of interest: an applied example using elbow flexor strength data. Phys Ther. 1997;77: Roebroeck ME, Harlaar J, Lankhorst GJ. The application of generalizability theory to reliability assessment: an illustration using isometric force measurements. Phys Ther. 1993;73: ; discussion Hopkins WG. Measures of reliability in sports medicine and science. Sports Med. 2000;30: Young VL, Pin P, Kraemer BA, et al. Fluctuation in grip and pinch strength among normal subjects. J Hand Surg [Am]. 1989;14: Pronk CNA. Technical note: measuring hand-grip force, using a new application of strain gauges. Med Biol Eng Comput. 1981;19: Clinical Assessment Recommendations. 2nd ed. Garner, NC: American Society of Hand Therapists; Gaul JS Jr. Intrinsic motor recovery: a long-term study of ulnar nerve repair. J Hand Surg [Am]. 1982;7: Roebroeck ME, Harlaar J, Lankhorst GJ. Reliability assessment of isometric knee extension measurements with a computer-assisted hand-held dynamometer. Arch Phys Med Rehabil. 1998;79: Armitage P, Berry G. Statistical Methods in Medical Research. 3rd ed. Boston: Mass: Blackwell Scientific Publications; Shrout P, Fleiss J. Intraclass correlation: uses in assessing rater reliability. Psychol Bull. 1979;86: Physical Therapy. Volume 83. Number 9. September 2003 Schreuders et al. 815

Ton A. R. Schreuders, Marij E. Roebroeck, PhD, Jean-Bart Jaquet, MD, Steven E. R. Hovius, MD, PhD, Henk J. Stam, MD, PhD, Rotterdam, the Netherlands

Ton A. R. Schreuders, Marij E. Roebroeck, PhD, Jean-Bart Jaquet, MD, Steven E. R. Hovius, MD, PhD, Henk J. Stam, MD, PhD, Rotterdam, the Netherlands Measuring the Strength of the Intrinsic Muscles of the Hand in Patients with Ulnar and Median Nerve Injuries: Reliability of the Rotterdam Intrinsic Hand Myometer (RIHM) Ton A. R. Schreuders, Marij E.

More information

Ton A. R. Schreuders, 1 Marij E. Roebroeck, 1 Jean-Bart Jaquet, 2 Steven E. R. Hovius 2 and Henk J. Stam 1 INTRODUCTION

Ton A. R. Schreuders, 1 Marij E. Roebroeck, 1 Jean-Bart Jaquet, 2 Steven E. R. Hovius 2 and Henk J. Stam 1 INTRODUCTION J Rehabil Med 2004; 36: 273 278 LONG-TERM OUTCOME OF MUSCLE STRENGTH IN ULNAR AND MEDIAN NERVE INJURY: COMPARING MANUAL MUSCLE STRENGTH TESTING, GRIP AND PINCH STRENGTH DYNAMOMETERS AND A NEW INTRINSIC

More information

Dynamometry of intrinsic hand muscles in patients with Charcot Marie Tooth disease

Dynamometry of intrinsic hand muscles in patients with Charcot Marie Tooth disease Dynamometry of intrinsic hand muscles in patients with Charcot Marie Tooth disease R.W. Selles, PhD; B.T.J. van Ginneken, MSc; T.A.R. Schreuders, PT, PhD; W.G.M. Janssen, MD; and H.J. Stam, MD, PhD, FRCP

More information

Instructions for administering the Rotterdam Intrinsic Hand Myometer (RIHM) test July 2008

Instructions for administering the Rotterdam Intrinsic Hand Myometer (RIHM) test July 2008 Instructions for administering the Rotterdam Intrinsic Hand Myometer (RIHM) test July 2008 RIHM HAPE EXTURE DENTIFICATION Instructions for administering the Rotterdam Intrinsic Hand Myometer (RIHM) A dynamometer

More information

The Reliability of Four Different Methods. of Calculating Quadriceps Peak Torque Angle- Specific Torques at 30, 60, and 75

The Reliability of Four Different Methods. of Calculating Quadriceps Peak Torque Angle- Specific Torques at 30, 60, and 75 The Reliability of Four Different Methods. of Calculating Quadriceps Peak Torque Angle- Specific Torques at 30, 60, and 75 By: Brent L. Arnold and David H. Perrin * Arnold, B.A., & Perrin, D.H. (1993).

More information

Department of Rehabilitation Medicine, Gyeongsang National University Hospital, 1

Department of Rehabilitation Medicine, Gyeongsang National University Hospital, 1 Original Article Ann Rehabil Med 2012; 36: 394-399 pissn: 2234-0645 eissn: 2234-0653 http://dx.doi.org/10.5535/arm.2012.36.3.394 Annals of Rehabilitation Medicine Reliability of the Pinch Strength with

More information

INTRODUCTION SUBJECTS AND METHODS

INTRODUCTION SUBJECTS AND METHODS J Korean Med Sci 23; 18: 2-8 ISSN 111-8934 Copyright The Korean Academy of Medical Sciences Normative Data and Developmental Characteristics of Hand Function for Elementary School Children in Suwon Area

More information

Muscle strength in patients with chronic pain

Muscle strength in patients with chronic pain Clinical Rehabilitation 2003; 17: 885 889 Muscle strength in patients with chronic pain CP van Wilgen Painexpertise Centre, Department of Rehabilitation, Department of Oral and Maxillofacial Surgery University

More information

Assessment protocol of limb muscle strength in critically ill. patients admitted to the ICU: Dynamometry

Assessment protocol of limb muscle strength in critically ill. patients admitted to the ICU: Dynamometry Assessment protocol of limb muscle strength in critically ill patients admitted to the ICU: Dynamometry To proceed to voluntary muscle strength assessment, the neurologic en hemodynamic stability of the

More information

Hand grip strength is a measure of. Reliability and validity of an electronic dynamometer for measuring grip strength

Hand grip strength is a measure of. Reliability and validity of an electronic dynamometer for measuring grip strength Research Reliability and validity of an electronic dynamometer for measuring grip strength Deborah Allen, Fiona Barnett Aims: This article describes a study aiming to establish the concurrent validity

More information

TEST-RETEST INTRA-RATER RELIABILITY OF GRIP FORCE IN PATIENTS WITH STROKE

TEST-RETEST INTRA-RATER RELIABILITY OF GRIP FORCE IN PATIENTS WITH STROKE J Rehabil Med 2003; 35: 189 194 TEST-RETEST INTRA-RATER RELIABILITY OF GRIP FORCE IN PATIENTS WITH STROKE Ann Hammer 1 and Birgitta Lindmark 2 From the 1 Department of Rehabilitation Medicine, Örebro University

More information

Downloaded from umj.umsu.ac.ir at 22: on Friday March 22nd 2019

Downloaded from umj.umsu.ac.ir at 22: on Friday March 22nd 2019 * 91/06/07 : 91/04/02 : :.. -. :. ICC. ICC.. :. // // :. : - - : :.(-) - (SLR) Email: eterafoskouei@tbzmed.ac.ir.(). (ULTT).().. ( ) Straight Leg Raise Upper Limb Tension Tests .().(-) Hough.().() ULTT1.()

More information

Citation for published version (APA): Soer, R. (2009). Functional capacity evaluation: measurement qualities and normative values s.n.

Citation for published version (APA): Soer, R. (2009). Functional capacity evaluation: measurement qualities and normative values s.n. University of Groningen Functional capacity evaluation Soer, Remko IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite from it. Please check the document

More information

Relationship of the Penn Shoulder Score with Measures of Range of Motion and Strength in Patients with Shoulder Disorders: A Preliminary Report

Relationship of the Penn Shoulder Score with Measures of Range of Motion and Strength in Patients with Shoulder Disorders: A Preliminary Report The University of Pennsylvania Orthopaedic Journal 16: 39 44, 2003 2003 The University of Pennsylvania Orthopaedic Journal Relationship of the Penn Shoulder Score with Measures of Range of Motion and Strength

More information

Intertester Reliability of Assessing Postural Sway Using the Chattecx Balance System

Intertester Reliability of Assessing Postural Sway Using the Chattecx Balance System Intertester Reliability of Assessing Postural Sway Using the Chattecx Balance System By: Carl G. Mattacola, MEd, ATC; Denise A. Lebsack, PhD, ATC; David H. Perrin, PhD, ATC * Mattacola, C.G., Lebsack,

More information

The Patient and Observer Scar Assessment Scale: A Reliable and Feasible Tool for Scar Evaluation

The Patient and Observer Scar Assessment Scale: A Reliable and Feasible Tool for Scar Evaluation The Patient and Observer Scar Assessment Scale: A Reliable and Feasible Tool for Scar Evaluation Lieneke J. Draaijers, M.D., Fenike R. H. Tempelman, M.D., Yvonne A. M. Botman, M.D., Wim E. Tuinebreijer,

More information

lntertester and lntratester Reliability of a Dynamic Balance Protocol Using the Biodex Stability System

lntertester and lntratester Reliability of a Dynamic Balance Protocol Using the Biodex Stability System Journal of Sport Rehabilitation, 1998, 7, 95-101 0 1998 Human Kinetics Publishers, Inc. lntertester and lntratester Reliability of a Dynamic Balance Protocol Using the Biodex Stability System Randy Schmitz

More information

Journal of Sport Rehabilitation. The reliability of strength tests performed in elevated shoulder positions using a hand-held dynamometer

Journal of Sport Rehabilitation. The reliability of strength tests performed in elevated shoulder positions using a hand-held dynamometer The reliability of strength tests performed in elevated shoulder positions using a hand-held dynamometer Journal: Manuscript ID: JSR.2015-0034.R2 Manuscript Type: Technical Report Keywords: dynamometry,

More information

Nutrition-Focused Physical Examination for Adults, Part 2: Grip Strength in Patient Functional Assessment

Nutrition-Focused Physical Examination for Adults, Part 2: Grip Strength in Patient Functional Assessment Idaho Health Care Association July 18, 2018 Boise Centre Boise, Idaho Handgrip strength (HGS) is one of the clinical characteristics of protein-calorie malnutrition as described by the Academy of Nutrition

More information

Goniometric Reliability in a Clinical Setting

Goniometric Reliability in a Clinical Setting Goniometric Reliability in a Clinical Setting Elbow and Knee Measurements JULES M. ROTHSTEIN, PETER J. MILLER, and RICHARD F. ROETTGER Reliability of goniometric measurements has been examined only under

More information

Electronic Supplementary Materials

Electronic Supplementary Materials Electronic Supplementary Materials Study 1 Body Measurements. To find proxies of lifting strength that could later be used in laboratory settings without cumbersome weight-lifting equipment, a variety

More information

Reliability of Lower Extremity Functional Performance Tests

Reliability of Lower Extremity Functional Performance Tests Reliability of Lower Extremity Functional Performance Tests Lori A. Bolgla, PT' Douglas R. Keskula, Ph D, PT, ATC2 ports medicine clinicians commonly see athletes who have sustained a lower extremity injury

More information

A new computer mouse called Horse

A new computer mouse called Horse A new computer mouse called Horse Anatomy and technology; The computer mouse as extension of the human body C.J. Snijders and P.C. Helder Department of Biomedical Physics and Technology, Erasmus MC, University

More information

The Effect of Inter-Trial Rest Interval on the Assessment of lsokinetic Thigh Muscle Torque

The Effect of Inter-Trial Rest Interval on the Assessment of lsokinetic Thigh Muscle Torque The Effect of Inter-Trial Rest Interval on the Assessment of lsokinetic Thigh Muscle Torque PAUL W. STRATFORD, MSc,' ANNETTE BRUULSEMA,2 BILL MAXWELL,= TIM BLACK, BSc,' BEVERLY HARDING, BSc4 Journal of

More information

Therapy Manual DO NOT PRINT

Therapy Manual DO NOT PRINT Therapy Manual Contents 1. Shoulder 2. Shoulder and elbow a. Protraction: 1 DoF 1 b. Flexion: 1 DoF 1-6 c. Extension: 1 DoF 1-2 d. Abduction: 1 DoF 1-4 e. External rotation: 1 DoF 1-14 a. Combined shoulder

More information

PROCEDURAL OPTIONS FOR MEASURING MUSCLE STRENGTH

PROCEDURAL OPTIONS FOR MEASURING MUSCLE STRENGTH 10.1515/AMB-2016-0020 PROCEDURAL OPTIONS FOR MEASURING MUSCLE STRENGTH S. Mindova, I. Karaganova and I. Stefanova Faculty of Public Health and Health Care, University of Ruse Angel Kanchev, Ruse, Bulgaria

More information

Spinal cord injury (SCI) and lower-extremity amputation

Spinal cord injury (SCI) and lower-extremity amputation Using the BTE Primus to Measure Grip and Wrist Flexion Strength in Physically Active Wheelchair Users: An Exploratory Study Orit Shechtman, Leanne MacKinnon, Catherine Locklear Key Words: hand strength

More information

The following pages constitute the final, accepted and revised manuscript of the article:

The following pages constitute the final, accepted and revised manuscript of the article: The following pages constitute the final, accepted and revised manuscript of the article: Flansbjer, Ulla-Britt and Holmbäck, Anna Maria and Downham, David and Lexell, Jan What change in isokinetic knee

More information

Title: Rotational strength, range of motion, and function in people with unaffected shoulders from various stages of life

Title: Rotational strength, range of motion, and function in people with unaffected shoulders from various stages of life Author's response to reviews Title: Rotational strength, range of motion, and function in people with unaffected shoulders from various stages of life Authors: Jean-Sébastien Roy (jean-sebastien.roy.1@ulaval.ca)

More information

Validity of Data Extraction Techniques on the Kinetic Communicator (KinCom) Isokinetic Device

Validity of Data Extraction Techniques on the Kinetic Communicator (KinCom) Isokinetic Device Validity of Data Extraction Techniques on the Kinetic Communicator (KinCom) Isokinetic Device By: Laurie L. Tis, PhD, AT,C * and David H. Perrin, PhD, AT,C Tis, L.L., & Perrin, D.H. (1993). Validity of

More information

REPRODUCIBILITY AND RESPONSIVENESS OF EVALUATIVE OUTCOME MEASURES

REPRODUCIBILITY AND RESPONSIVENESS OF EVALUATIVE OUTCOME MEASURES International Journal of Technology Assessment in Health Care, 17:4 (2001), 479 487. Copyright c 2001 Cambridge University Press. Printed in the U.S.A. REPRODUCIBILITY AND RESPONSIVENESS OF EVALUATIVE

More information

Three Different Methods of Calculating Vertical Jump Height from Force Platform Data in Men and Women

Three Different Methods of Calculating Vertical Jump Height from Force Platform Data in Men and Women Measurement in Physical Education and Exercise Science ISSN: 1091-367X (Print) 1532-7841 (Online) Journal homepage: http://www.tandfonline.com/loi/hmpe20 Three Different Methods of Calculating Vertical

More information

INFLUENCE OF DIFFERENT TESTING POSTURES ON HAND GRIP STRENGTH

INFLUENCE OF DIFFERENT TESTING POSTURES ON HAND GRIP STRENGTH INFLUENCE OF DIFFERENT TESTING POSTURES ON HAND GRIP STRENGTH Walaa M. El-Sais, MSc, PT College of Applied Medical Sciences, Majmaah University, KSA Walaa S. Mohammad, PhD, PT Faculty of Physical Therapy,

More information

Test-retest reliability of the Shape/Texture Identification testtm in people with chronic stroke

Test-retest reliability of the Shape/Texture Identification testtm in people with chronic stroke Test-retest reliability of the Shape/Texture Identification testtm in people with chronic stroke Ekstrand, Elisabeth; Lexell, Jan; Brogårdh, Christina Published in: Clinical Rehabilitation DOI: 10.1177/0269215515608512

More information

Use of the Standard Error as a Reliability Index of Interest:

Use of the Standard Error as a Reliability Index of Interest: Use of the Standard Error as a Reliability Index of Interest: An Applied Example Using Elbow Flexor Strength Data Key Words: Measurement error, Reliability, components. The intraclass correlatio~l coefficient

More information

Serial Grip Strength Testing- Its Role In Assessment Of Wrist And Hand Disability

Serial Grip Strength Testing- Its Role In Assessment Of Wrist And Hand Disability ISPUB.COM The Internet Journal of Surgery Volume 5 Number 2 Serial Grip Strength Testing- Its Role In Assessment Of Wrist And Hand Disability L Ashton, S Myers Citation L Ashton, S Myers.. The Internet

More information

Tung-Wu Lu, DPhil 1, Horng-Chaung Hsu, MD 2, Ling-Ying Chang, MS, PT 3 and Hao-Ling Chen, PhD 4 ORIGINAL REPORT. J Rehabil Med 2007; 39:

Tung-Wu Lu, DPhil 1, Horng-Chaung Hsu, MD 2, Ling-Ying Chang, MS, PT 3 and Hao-Ling Chen, PhD 4 ORIGINAL REPORT. J Rehabil Med 2007; 39: J Rehabil Med 2007; 39: 679 684 ORIGINAL REPORT ENHANCING THE EXAMINER S RESISTING FORCE IMPROVES THE RELIABILITY OF MANUAL MUSCLE STRENGTH MEASUREMENTS: COMPARISON OF A NEW DEVICE WITH HAND-HELD DYNAMOMETRY

More information

CRITICALLY APPRAISED PAPER (CAP)

CRITICALLY APPRAISED PAPER (CAP) CRITICALLY APPRAISED PAPER (CAP) Svens, B., Ames, E., Burford, K., & Caplash, Y. (2015). Relative active motion programs following extensor tendon repair: A pilot study using a prospective cohort and evaluating

More information

Reliability and Minimum Detectable Change for Common Clinical Physical Function Tests in Sarcopenic Men and Women

Reliability and Minimum Detectable Change for Common Clinical Physical Function Tests in Sarcopenic Men and Women University of Nebraska - Lincoln DigitalCommons@University of Nebraska - Lincoln Nutrition and Health Sciences -- Faculty Publications Nutrition and Health Sciences, Department of 2017 Reliability and

More information

Measuring equipment for professionals. Manual. Handleiding MicroFET 2

Measuring equipment for professionals. Manual. Handleiding MicroFET 2 Measuring equipment for professionals Manual Handleiding MicroFET 2 Contents 1 Introduction... 2 2 Testing Method... 3 3 Reproducibility... 4 4 MicroFET2 Features... 5 5 Units of measurement... 7 6 Batteries...

More information

ACGIH TLV for Hand Activity Level (HAL)

ACGIH TLV for Hand Activity Level (HAL) ACGIH TLV for Hand Activity Level (HAL) 1(6) ACGIH TLV for Hand Activity Level (HAL) General description and development of the method The ACGIH HAL TLV uses HAL (Hand activity level) and peak hand forces

More information

NATIONAL ASSESSMENT MANUAL. For assessment of hand function after nerve repair

NATIONAL ASSESSMENT MANUAL. For assessment of hand function after nerve repair NATIONAL ASSESSMENT MANUAL For assessment of hand function after nerve repair Manual for assessment of hand function after nerve repair according to Rosen score 1 Version 1, 2018 Documentation of hand

More information

Validity and Interobserver Reliability of Visual Observation to Assess Partial Weight-Bearing

Validity and Interobserver Reliability of Visual Observation to Assess Partial Weight-Bearing ORIGINAL ARTICLE Validity and Interobserver Reliability of Visual Observation to Assess Partial Weight-Bearing Henri L. Hurkmans, PhD, Johannes B. Bussmann, PhD, Eric Benda, PT 309 ABSTRACT. Hurkmans HL,

More information

Advanced Ergonomic Benefits of Pneumatic Scissors

Advanced Ergonomic Benefits of Pneumatic Scissors Advanced Ergonomic Benefits of Pneumatic Scissors Abstract While a certain of automation within the poultry industry has been reached, there will always be an interaction between the product, individual,

More information

Research Report. A Comparison of Five Low Back Disability Questionnaires: Reliability and Responsiveness

Research Report. A Comparison of Five Low Back Disability Questionnaires: Reliability and Responsiveness Research Report A Comparison of Five Low Back Disability Questionnaires: Reliability and Responsiveness APTA is a sponsor of the Decade, an international, multidisciplinary initiative to improve health-related

More information

CRITICALLY APPRAISED PAPER (CAP)

CRITICALLY APPRAISED PAPER (CAP) CRITICALLY APPRAISED PAPER (CAP) Che Daud, A. Z., Yau, M. K., Barnett, F., Judd, J., Jones, R. E., & Muhammad Nawawi, R. F. (2016). Integration of occupation based intervention in hand injury rehabilitation:

More information

PREDICTORS OF MUSCLE STRENGTH USING QUANTITATIVE MUSCLE TESTING FOR HAND MUSCLES IN YOUNG INDIAN ADULTS

PREDICTORS OF MUSCLE STRENGTH USING QUANTITATIVE MUSCLE TESTING FOR HAND MUSCLES IN YOUNG INDIAN ADULTS The Indian Journal of Occupational Therapy : Vol. XLII : No. 1 () PREDICTORS OF MUSCLE STRENGTH USING QUANTITATIVE MUSCLE TESTING FOR HAND MUSCLES IN YOUNG INDIAN ADULTS * Rasna Ratn, B.O.T., Co-Authors

More information

Adult grip strength norms for the Baseline digital dynamometer

Adult grip strength norms for the Baseline digital dynamometer The University of Toledo The University of Toledo Digital epository Master s and Doctoral Projects Adult grip strength norms for the Baseline digital dynamometer Chelsey. Meek Follow this and additional

More information

Effects of Grip Power Training on Elbow Flexion Strength after Oberlin Transfer in Upper Arm Type Brachial Plexus Injuries

Effects of Grip Power Training on Elbow Flexion Strength after Oberlin Transfer in Upper Arm Type Brachial Plexus Injuries Effects of Grip Power Training on Elbow Flexion Strength after Oberlin Transfer in Upper Arm Type Brachial Plexus Injuries Yi-Jung Tsai 1, Chih-Kun Hsiao 2, Yuan-Kun Tu 2, Chin-Hsien Wu 2, Fong-Chin Su

More information

An Investigation of Hand Force Distribution, Hand Posture and Surface Orientation

An Investigation of Hand Force Distribution, Hand Posture and Surface Orientation An Investigation of Hand Force Distribution, Hand Posture and Surface Orientation R. FIGUEROA and T. ARMSTRONG Department of Industrial and Operations Engineering, University of Michigan, Ann Arbor, MI

More information

Test-retest reliability and measurement errors of six mobility tests in the community-dwelling elderly

Test-retest reliability and measurement errors of six mobility tests in the community-dwelling elderly Asian J Gerontol Geriatr 2009; 4: 8 13 Test-retest reliability and measurement errors of six mobility tests in the community-dwelling elderly ORIGINAL ARTICLE CY Wang 1 PhD, CF Sheu 2 PhD, EJ Protas 3

More information

Influence of Lever Arm and Stabilization on Measures of Hip Abduction and Adduction Torque Obtained by Hand-Held Dynamometry

Influence of Lever Arm and Stabilization on Measures of Hip Abduction and Adduction Torque Obtained by Hand-Held Dynamometry 37 ORIGINAL ARTICLE Influence of Lever Arm and Stabilization on Measures of Hip Abduction and Adduction Torque Obtained by Hand-Held Dynamometry David A. Krause, PT, MBA, DScPT, OCS, Susan J. Schlagel,

More information

OBJECTIVE AND ACCURATE measurements of lumbar

OBJECTIVE AND ACCURATE measurements of lumbar 99 Strapped Versus Unstrapped Technique of the Prone Press-Up for Measurement of Lumbar Extension Using a Tape Measure: Differences in Magnitude and Reliability of Measurements William D. Bandy, PhD, PT,

More information

Specialist in motion. Handleiding. User Manual MicroFET 2

Specialist in motion. Handleiding. User Manual MicroFET 2 Specialist in motion Handleiding User Manual MicroFET 2 Contents 1 Introduction... 2 2 Testing Method... 3 3 Reproducibility... 4 4 MicroFET2 Features... 5 5 Units of measurement... 7 6 Battery... 8 7

More information

Power grip is a commonly used index to assess impairment

Power grip is a commonly used index to assess impairment Brief Report Effect of Wrist Positioning on the Repeatability and Strength of Power Grip Pauline W. K. Fong, Gabriel Y. F. Ng Key Words: hand strength testing posture and positioning evaluation wrist joint

More information

Hands on Sports Therapy KNOWLEDGE REVIEW QUESTIONS 2004 Thomson Learning It can help to shape a basic fitness training programme

Hands on Sports Therapy KNOWLEDGE REVIEW QUESTIONS 2004 Thomson Learning It can help to shape a basic fitness training programme Hands on Sports Therapy KNOWLEDGE REVIEW QUESTIONS 2004 Thomson Learning 1 CHAPTER 13 Knowledge Review Q1: Why is fitness testing useful? A1: Fitness testing is useful for various reasons: 1. It can help

More information

Grip strength is a representative measure of muscle weakness in the upper extremity after stroke

Grip strength is a representative measure of muscle weakness in the upper extremity after stroke Grip strength is a representative measure of muscle weakness in the upper extremity after stroke Ekstrand, Elisabeth; Lexell, Jan; Brogårdh, Christina Published in: Topics in Stroke Rehabilitation DOI:

More information

Interrater and Intrarater Reliability of the Assisting Hand Assessment

Interrater and Intrarater Reliability of the Assisting Hand Assessment Interrater and Intrarater Reliability of the Assisting Hand Assessment Marie Holmefur, Lena Krumlinde-Sundholm, Ann-Christin Eliasson KEY WORDS hand pediatric reliability OBJECTIVE. The aim of this study

More information

Test-retest relative and absolute reliability of knee extensor strength measures and minimal detectable change

Test-retest relative and absolute reliability of knee extensor strength measures and minimal detectable change Isokinetics and Exercise Science 22 (2014) 17 26 17 DOI 10.3233/IES-130522 IOS Press Test-retest relative and absolute reliability of knee extensor strength measures and minimal detectable change Asunción

More information

Statistical Methods For Assessing Measurement Error (Reliability) in Variables Relevant to Sports Medicine

Statistical Methods For Assessing Measurement Error (Reliability) in Variables Relevant to Sports Medicine REVIEW ARTICLE Sports Med 1998 Oct; 26 (4): 217-238 0112-1642/98/0010-0217/$11.00/0 Adis International Limited. All rights reserved. Statistical Methods For Assessing Measurement Error (Reliability) in

More information

A Comparison of Two Stretching Protocols on Hip Range of Motion: Implications for Total Daily Stretch Duration

A Comparison of Two Stretching Protocols on Hip Range of Motion: Implications for Total Daily Stretch Duration Journal of Strength and Conditioning Research, 2003, 17(2), 274 278 2003 National Strength & Conditioning Association A Comparison of Two Stretching Protocols on Hip Range of Motion: Implications for Total

More information

Intramachine and intermachine reproducibility of concentric performance: A study of the Con-Trex MJ and the Cybex Norm dynamometers

Intramachine and intermachine reproducibility of concentric performance: A study of the Con-Trex MJ and the Cybex Norm dynamometers Isokinetics and Exercise Science 12 (4) 91 97 91 IOS Press Intramachine and intermachine reproducibility of concentric performance: A study of the Con-Trex MJ and the Cybex Norm dynamometers C. Bardis

More information

CRITICALLY APPRAISED PAPER (CAP)

CRITICALLY APPRAISED PAPER (CAP) CRITICALLY APPRAISED PAPER (CAP) Horng, Y. S., Hsieh, S. F., Tu, Y. K., Lin, M. C., Horng, Y. S., & Wang, J. D. (2011). The comparative effectiveness of tendon and nerve gliding exercises in patients with

More information

Original Article. MUNENORI KATOH, PT, PhD 1), YUKINOBU HIIRAGI, PT, PhD 2), MANABU UCHIDA, PT, PhD 3)

Original Article. MUNENORI KATOH, PT, PhD 1), YUKINOBU HIIRAGI, PT, PhD 2), MANABU UCHIDA, PT, PhD 3) Original Article Validity of Isometric Muscle Strength Measurements of the Lower Limbs Using a Handheld Dynamometer and Belt: a Comparison with an Isokinetic Dynamometer J. Phys. Ther. Sci. 23: 553 557,

More information

Chapter 20: Test Administration and Interpretation

Chapter 20: Test Administration and Interpretation Chapter 20: Test Administration and Interpretation Thought Questions Why should a needs analysis consider both the individual and the demands of the sport? Should test scores be shared with a team, or

More information

Model 1: Subject As Single Factor

Model 1: Subject As Single Factor 3 Model 1: Subject As Single Factor 3.1 The Model In this section, I am considering reliability experiments where each subject is scored by a different group of r raters 1. The experimenter controls which

More information

The collection of reliable and valid evaluation

The collection of reliable and valid evaluation Grip and -Pinch Strength: Norms for 6- to 19-Year 1ds Virgil Mathiowetz, Diana M. Wiemer, Susan M. Federman Key Words: hand. hand strength testing. manual strength. motor skills. tests The purpose ofthis

More information

UTILITY OF THE POWERBALL IN THE INVIGORATION OF THE MUSCULATURE OF THE FOREARM

UTILITY OF THE POWERBALL IN THE INVIGORATION OF THE MUSCULATURE OF THE FOREARM Hand Surgery, Vol. 13, No. 2 (2008) 79 83 World Scientific Publishing Company UTILITY OF THE POWERBALL IN THE INVIGORATION OF THE MUSCULATURE OF THE FOREARM Sebastián Axel Balan and Marc Garcia-Elias Institut

More information

TO MAINTAIN THE INDEPENDENCE of older adults as

TO MAINTAIN THE INDEPENDENCE of older adults as 1747 ORIGINAL ARTICLE Grip Strength in Older Adults: Test-Retest Reliability and Cutoff for Subjective Weakness of Using the Hands in Heavy Tasks Ching-Yi Wang, PhD, Li-Yuan Chen, MS ABSTRACT. Wang C-Y,

More information

4 Smallest detectable difference of maximal mouth opening in patients with painfully restricted temporomandibular joint function.

4 Smallest detectable difference of maximal mouth opening in patients with painfully restricted temporomandibular joint function. 4 Smallest detectable difference of maximal mouth opening in patients with painfully restricted temporomandibular joint function. Kropmans ThJB, Dijkstra PU, Stegenga B, Stewart R, de Bont LGM European

More information

The Reliability of Measuring Neck Muscle Strength with a Neck Muscle Force Measurement Device

The Reliability of Measuring Neck Muscle Strength with a Neck Muscle Force Measurement Device Original Article The Reliability of Measuring Neck Muscle Strength with a Neck Muscle Force Measurement Device J. Phys. Ther. Sci. 15: 7 12, 2003 ASGHAR REZASOLTANI, Ph D, PT 1, 2), AMIR AHMADI, B Sc,

More information

Effect of Preload and Range of Motion on Isokinetic Torque in Women

Effect of Preload and Range of Motion on Isokinetic Torque in Women Effect of Preload and Range of Motion on Isokinetic Torque in Women By: Laurie L. Tis, David H. Perrin, Arthur Weltman, Donald W. Ball, and Joe H. Gieck * Tis, L.L., Perrin, D.H., Weltman, A., Ball, D.W.,

More information

Resources: Ergonomic/Musculoskeletal Hazards in Patient Handling. Sustained and/or Awkward Postures. Contact Stresses or Pressures

Resources: Ergonomic/Musculoskeletal Hazards in Patient Handling. Sustained and/or Awkward Postures. Contact Stresses or Pressures Resources: Ergonomic/Musculoskeletal Hazards in Patient Handling Forceful Exertions Sustained and/or Awkward Postures Repetition Gripping Contact Stresses or Pressures High Risk Patient Handling Tasks

More information

Reliability and validity of the Biodex system 3 pro isokinetic dynamometer velocity, torque and position measurements

Reliability and validity of the Biodex system 3 pro isokinetic dynamometer velocity, torque and position measurements Reliability and validity of the Biodex system 3 pro isokinetic dynamometer velocity, torque and position measurements By: Joshua M. Drouin 1, Tamara C. Valovich-McLeod 2, Sandra J. Shultz 3, Bruce M. Gansneder

More information

ISOMETRIC AND ISOKINETIC MUSCLE STRENGTH IN THE UPPER EXTREMITY CAN BE RELIABLY MEASURED IN PERSONS WITH CHRONIC STROKE

ISOMETRIC AND ISOKINETIC MUSCLE STRENGTH IN THE UPPER EXTREMITY CAN BE RELIABLY MEASURED IN PERSONS WITH CHRONIC STROKE J Rehabil Med 2015; 47: 706 713 ORIGINAL REPORT ISOMETRIC AND ISOKINETIC MUSCLE STRENGTH IN THE UPPER EXTREMITY CAN BE RELIABLY MEASURED IN PERSONS WITH CHRONIC STROKE Elisabeth Ekstrand, RPT, MSc 1,2,

More information

Design & manufacturing a computerized Multi-Channel Isometric Dynamometer

Design & manufacturing a computerized Multi-Channel Isometric Dynamometer Journal of Advanced Sport Technology 1(1): 15-21 Technical Research Design & manufacturing a computerized Multi-Channel Isometric Dynamometer Abbas Meamarbashi 1 1. Professor in Sport Biomechanics, University

More information

The Patient-Rated Tennis Elbow Evaluation (PRTEE) User Manual. December 2007

The Patient-Rated Tennis Elbow Evaluation (PRTEE) User Manual. December 2007 The Patient-Rated Tennis Elbow Evaluation (PRTEE) User Manual December 2007 Joy C. MacDermid, BScPT, MSc, PhD School of Rehabilitation Science, McMaster University, Hamilton, Ontario, Canada Clinical Research

More information

Original Article A simple way to improve the relative and absolute reliability of handheld dynamometer measurements using learners

Original Article A simple way to improve the relative and absolute reliability of handheld dynamometer measurements using learners Int J Clin Exp Med 2016;9(1):199-208 www.ijcem.com /ISSN:1940-5901/IJCEM0015678 Original Article A simple way to improve the relative and absolute reliability of handheld dynamometer measurements using

More information

Chapter 4B: Reliability Trial 2. Chapter 4B. Reliability. General issues. Inter-rater. Intra - rater. Test - Re-test

Chapter 4B: Reliability Trial 2. Chapter 4B. Reliability. General issues. Inter-rater. Intra - rater. Test - Re-test Chapter 4B: Reliability Trial 2 Chapter 4B Reliability General issues Inter-rater Intra - rater Test - Re-test Trial 2 The second clinical trial was conducted in the spring of 1992 using SPCM Draft 6,

More information

The Patient-Rated Wrist Evaluation (PRWE) User Manual. December 2007

The Patient-Rated Wrist Evaluation (PRWE) User Manual. December 2007 The Patient-Rated Wrist Evaluation (PRWE) User Manual December 2007 Joy C. MacDermid, BScPT, MSc, PhD School of Rehabilitation Science, McMaster University, Hamilton, Ontario, Canada Clinical Research

More information

Evaluating concomitant lateral epicondylitis and cervical radiculopathy

Evaluating concomitant lateral epicondylitis and cervical radiculopathy Evaluating concomitant lateral epicondylitis and cervical radiculopathy March 06, 2010 This article describes a study of the prevalence of lateral epicondylitis or tennis elbow among patients with neck

More information

Reliability of Measuring Trunk Motions in Centimeters

Reliability of Measuring Trunk Motions in Centimeters Reliability of Measuring Trunk Motions in Centimeters MARGARET ROST, SANDRA STUCKEY, LEE ANNE SMALLEY, and GLENDA DORMAN A method of measuring trunk motion and two related motions using a tape measure

More information

MIE Medical Research Ltd Digital Analyser - Myometer Instruction Manual

MIE Medical Research Ltd Digital Analyser - Myometer Instruction Manual Doc 117-04 Copyright 1998-2004 MIE Medical Research Ltd Getting Started Unpack your Digital Analyser and check that the following components are present: 1 x Digital Analyser Figure 1 1 x Myometer Transducer

More information

Regulation of Human Heart Rate

Regulation of Human Heart Rate Name: Date: Period: Regulation of Human Heart Rate Pre-Lab 1. List some activities or stimuli that you think may increase a person s heart rate. An activity is something a person does, and a stimulus is

More information

Power grip is an isometric and static condition defined

Power grip is an isometric and static condition defined An Original Study Biomechanical Analysis of Flexor Digitorum Profundus and Superficialis in Grip-Strength Generation Robert A. Kaufmann, MD, Scott H. Kozin, MD, Adam Mirarchi, BS, Burt Holland, PhD, and

More information

Chapter 14 Training Muscles to Become Stronger

Chapter 14 Training Muscles to Become Stronger Chapter 14 Training Muscles to Become Stronger Slide Show developed by: Richard C. Krejci, Ph.D. Professor of Public Health Columbia College 11.22.11 Objectives 1. Describe the following four methods to

More information

Electromyography II Laboratory (Hand Dynamometer Transducer)

Electromyography II Laboratory (Hand Dynamometer Transducer) (Hand Dynamometer Transducer) Introduction As described in the Electromyography I laboratory session, electromyography (EMG) is an electrical signal that can be recorded with electrodes placed on the surface

More information

The amount of active range of motion (AROM) for forearm

The amount of active range of motion (AROM) for forearm Reliability of 2 Functional Goniometric Methods for Measuring Forearm Pronation And Supination Active Range of Motion Christos Karagiannopoulos, MPT, MEd, ATC 1 Michael Sitler, EdD, ATC 2 Susan Michlovitz,

More information

Testing Protocol: Grip and Pinch

Testing Protocol: Grip and Pinch Testing Protocol: Grip and Pinch OF POWER GRIP or upright - test arm at side with elbow flexed 90 - palm facing inward Adjust handle to appropriate rung, where grip is comfortable and the thumb overlaps

More information

ANTICIPATING DYNAMIC LOADS IN HANDLING OBJECTS.

ANTICIPATING DYNAMIC LOADS IN HANDLING OBJECTS. ANTICIPATING DYNAMIC LOADS IN HANDLING OBJECTS. Alan M. Wing Sensory Motor Neuroscience Centre, The University of Birmingham, Edgbaston, Birmingham, B15 2TT, UK. a.m.wing@bham.ac.uk J. Randall Flanagan

More information

Many upper extremity motor function outcome measures do

Many upper extremity motor function outcome measures do Assessing Wolf Motor Function Test as Outcome Measure for Research in Patients After Stroke Steven L. Wolf, PhD, PT; Pamela A. Catlin, EdD, PT; Michael Ellis, MPT; Audrey Link Archer, MPT; Bryn Morgan,

More information

A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability Research

A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability Research Journal of Chiropractic Medicine (2016) 15, 155 163 www.journalchiromed.com Cracking the Code: Providing Insight Into the Fundamentals of Research and Evidence-Based Practice A Guideline of Selecting and

More information

KEEPING IRONWORKERS HEALTHY: ERGONOMICS AND WMSDs

KEEPING IRONWORKERS HEALTHY: ERGONOMICS AND WMSDs KEEPING IRONWORKERS HEALTHY: ERGONOMICS AND WMSDs Funded by: With assistance from: WMSDs are preventable! 1 This training is dedicated to the memory of 2 GOALS OF THIS CLASS You will understand: What WMSDs

More information

PDF hosted at the Radboud Repository of the Radboud University Nijmegen

PDF hosted at the Radboud Repository of the Radboud University Nijmegen PDF hosted at the Radboud Repository of the Radboud University Nijmegen The following full text is a publisher's version. For additional information about this publication click this link. http://hdl.handle.net/2066/23532

More information

Augmented reflection technology for stroke rehabilitation a clinical feasibility study

Augmented reflection technology for stroke rehabilitation a clinical feasibility study Augmented reflection technology for stroke rehabilitation a clinical feasibility study S Hoermann 1, L Hale 2, S J Winser 2, H Regenbrecht 1 1 Department of Information Science, 2 School of Physiotherapy,

More information

EVALUATION AND MEASUREMENTS. I. Devreux

EVALUATION AND MEASUREMENTS. I. Devreux EVALUATION AND MEASUREMENTS I. Devreux To determine the extent and degree of muscular weakness resulting from disease, injury or disuse. The records obtained from these tests provide a base for planning

More information

Humerus. Ulna. Radius. Carpals

Humerus. Ulna. Radius. Carpals Posture Analysis Exercise T. Armstrong M. Ebersole 1.0 Objectives: 1. Improve skill for rating over all job and identifying specific force and posture problems 2. Learn how to characterize posture 3. Learn

More information

Reliability of Stationary Dynamometer Muscle Strength Testing in Community-Dwelling Older Adults

Reliability of Stationary Dynamometer Muscle Strength Testing in Community-Dwelling Older Adults 1128 Reliability of Stationary Dynamometer Muscle Strength Testing in Community-Dwelling Older Adults Cheryl D. Ford-Smith, PT, MS, NCS, Jean F. Wyman, PhD, RN, R.K. Elswick Jr, PhD, Theresa Fernandez,

More information

Cubital Tunnel Syndrome

Cubital Tunnel Syndrome Cubital Tunnel 2 3333 N CALVERT ST, SUITE 370, BALTIMORE, MD 21218 T410 467 5400 F410 366 9826 delloninstitutes.com your complaints are Numbness or tingling in the little/ring side of the front and back

More information