CATHERINE FILLÉE, JEAN CUMPS, JEAN-MARIE KETELSLEGERS. Cliniques Universitaires St-Luc, Université Catholique de Louvain, Brussels, Belgium SUMMARY

Size: px
Start display at page:

Download "CATHERINE FILLÉE, JEAN CUMPS, JEAN-MARIE KETELSLEGERS. Cliniques Universitaires St-Luc, Université Catholique de Louvain, Brussels, Belgium SUMMARY"

Transcription

1 Clin. Lab. 2012;58: Copyright ORIGINAL ARTICLE Comparison of Three Free T4 (FT4) and Free T3 (FT3) Immunoassays in Healthy Subjects and Patients with Thyroid Diseases and Severe Non-Thyroidal Illnesses CATHERINE FILLÉE, JEAN CUMPS, JEAN-MARIE KETELSLEGERS Cliniques Universitaires St-Luc, Université Catholique de Louvain, Brussels, Belgium SUMMARY Background: Free T4 (FT4) and free T3 (FT3) immunoassays exhibit wide inter-assay variations. We therefore established control values with three different immunoassays and compared their clinical performances in thyroidal diseases and severe/acute non-thyroidal illnesses (NTI). Methods: The UniCel DxI 800, Architect i2000, and Elecsys 2010 assays were used. FT4 and FT3 reference ranges were established in 68 controls, without conditions interfering with thyroid function, with normal TSH ( mu/l) and negative anti-thyroid peroxidase antibodies. Free hormones were determined in 60 patients with thyroid diseases (TSH: < mu/l) and 45 NTI patients (TSH: mu/l). Control values were normalized as Z-scores; patients results were expressed as Z-scores, using the control values of each assay. Pairwise comparisons of the Z-scores were performed by Deming s regression. Classification of patients results based on 95% control values were evaluated by kappa agreement statistics. Results: Control values for FT4 (pmol/l; geometrical means; 95% confidence intervals) were: 11.1 ( ), 12.3 ( ), 15.6 ( ) and for FT3: 4.8 ( ), 4.0 ( ), 4.3 ( ), with DxI 800, Architect, and Elecsys, respectively. Pairs of control Z-scores correlated significantly, but with different strengths (FT4: r = 0.915, 0.740, 0.770; FT3: r = 0.615; 0.589; 0.790, for DxI 800 vs. Elecsys; DxI 800 vs. Architect; Elecsys vs. Architect; p<0.001); slopes and intercepts of paired controls were 1.00 and zero. In thyroid diseases, slopes of FT4 Z-scores among assays differed slightly from 1.00 (1.11, 0.88, 0.87 for DxI versus Elecsys, DxI 800 versus Architect, Elecsys versus Architect, respectively; p<0.05); slopes of FT3 Z-scores were consistent with unity, except for DxI 800 versus Elecsys (0.88; p<0.05). In NTI patients, regression slopes were consistent with unity (p<0.05). The agreement statistics showed moderate to very good inter-assay concordances for thyroid and NTI patients results. Conclusions: FT4 and FT3 assays show moderate to very good agreement, in patients with thyroid diseases or NTI when compared pairwise to their control values. Slight quantitative differences between some pairs of assays are observed in thyroid diseases, after normalization as Z-scores. (Clin. Lab. 2012;58: DOI: /Clin.Lab ) KEY WORDS Free thyroxin; free triiodothyronine; immunoassay; thyroid; non-thyroidal illness LIST OF ABBREVIATIONS FT3 - free triiodothyronine FT4 - free thyroxin IRQ - interquartile range LC-MS/MS - liquid chromatography-tandem mass spectrometry NTI - non-thyroidal illness TPO - thyroid peroxidase TSH - thyrotropin TTR - transthyretin Manuscript accepted November 27, 2011 Clin. Lab. 7+8/

2 C. FILLÉE et al. INTRODUCTION Thyrotropin (TSH) is the first line assay for the routine evaluation of thyroid function status. However, free thyroxin (FT4) and free triiodothyronine (FT3) are essential biomarkers, together with thyrotropin (TSH) for final diagnosis and management of thyroid diseases [1]. The majority of third generation TSH assays show excellent performances [2]. The measurement of free thyroid hormones remains a complex analytical issue [3]. The various immunoassay systems for FT4 and FT3 show wide inter-assay variations [4-7]. In particular, the normal ranges among manufacturers exhibit important differences, a feature causing much confusion in the comparison of results among different laboratories. This lack of consensus in the reference values of FT4 and FT3 among assays may also shed some doubt on the clinical value of the results generated by different assays. Thus, it could be hypothesized that the FT4 or FT3 results categorized as normal, elevated or low in patients with thyroid diseases or severe/acute non-thyroidal illnesses (NTI) are at variance among assays. Such inter-assay comparisons, based upon the analysis of clinically defined groups of patients is at present poorly documented. To obviate the problems related to the inter-assay variations of free thyroid hormone immunoassays, the use of methods based upon the physiccal separation of the free hormones from the hormonebinding complexes and the subsequent measurement of T4 or T3 by liquid chromatography-tandem mass spectrometry (LC-MS/MS) have been advocated. However, these methods will require further standardization, in particular, in the preanalytical phases and significant effort is still necessary to implement them in automated formats suitable for large scale routine volumes [8-12]. Since routine laboratories will continue for some time to use the available automated immunoassays, we addressed the question of the comparison of three different FT4 and FT3 immunoassays, in a group of patients with known thyroid diseases or NTI, compared to the reference ranges of each assay. To this end, we first determined FT4 and FT3 with the three immunoassays, in a control group. In this group, we also tested if FT4 and FT3 would be influenced by age, in all assays. The results of the control group were expressed as Z-scores. FT4 and FT3 were also determined in a cohort of patients with thyroid diseases or NTI and the data expressed as Z-scores based on the reference control values of each assay. The patients results obtained in each assay were compared pairwise by regression analysis. Concordance among assays, in relation to the reference control values was also computed. MATERIALS AND METHODS Patients The patient characteristics are presented in Table 1. A group of 68 control patients (ambulatory or hospitalized, 35 to 95 years old) were selected on the basis that their medical record mentioned only benign conditions, not prone to alter the thyroid status; their TSH levels were normal (median: 1.24 mu/l; range: ); all of them had undetectable anti thyroid peroxidase antibodies (atpo) levels. A group of 60 patients with established thyroid diseases (substituted primary hyperthyroidism, secondary hypothyroidism, Basedow diseases, euthyroid goiter) were included. For all of them, the diagnosis was based on clinical grounds, thyroid imaging, as well as several pre-treatment thyroid panels; these patients were either under early or long-term therapy and were either well controlled on the basis of TSH levels or were in the course of therapeutic adjustments; indeed, the TSH values in this subgroup ranged from undetectable to very high values, consistent with the wide range of thyroid dysfunctions encountered at the time of blood sampling. In the subgroup of 45 patients with NTI, the TSH values were normal (n = 38), low (n = 4; µu/ml) or slightly elevated (n = 3; µu/ml). Assays TSH and atpo were determined on the UniCel DxI 800 (Beckman Coulter); FT4 and FT3 were measured on the UniCel DxI 800, the Elecsys 2010 (Roche), and the Architect i2000 (Abbott). All assays were performed on the same day, all three instruments were maintained according to the manufacturers recommendations, and the whole procedure was directly supervised by the same biologist (CF). The procedure was in agreement with the internal rules of the hospital ethical committee. FT4 determination on both DxI 800 and Architect i2000 use a 2-step enzyme immunoassay, in which the FT4 of the sample reacts first with anti-t4 antibody coated microparticles; after washing, the antibodies remaining free on the solid phase are back-titrated by addition of labeled T3 and the signal is inversely related to the FT4 concentrations in the sample. The electrochemiluminescent Elecsys FT4 assay differs from the other two, as it is based upon a first incubation in which the FT4 of the sample reacts with ruthenium labeled anti-t4 antibodies; in a subsequent incubation, microparticle-bound T4 is added to the medium and binds to the remaining free labeled anti-t4 antibodies; the solid phase is then separated from the incubation medium and the amount of labeled anti-t4 antibodies captured by the T4 coated microparticles is inversely related to the FT4. The DxI 800 and Elecsys 2010 FT3 assays are based on the single labeled antibody technique. In contrast, the FT3 Architect i2000 assay makes use of the 2-step assay scheme. The assays imprecisions were in accordance with those stated in the data inserts (data not shown). Statistics Comparisons between FT4 and FT3 results obtained with the three assays from controls and all patients considered together were performed using the non-paramet- 726 Clin. Lab. 7+8/2012

3 COMPARISON OF FT4 AND FT3 IMMUNOASSAYS ric Passing-Bablok regression. Relations between age and FT4 and FT3 in the control group were computed by Pearson regression analyses after logarithmic transformation of the free hormone concentrations. Confidence limits of the FT4 and FT3 control values were calculated after logarithmic transformation of the data and were expressed as geometrical means with 95% confidence intervals; the 90% confidence of the upper and lower limits were also computed. The significance of differences between control values for FT4 and FT3 among the three assays was computed using paired t- test (after Bonferoni correction for multiple comparisons). For each assay, Z-scores of the 68 controls group samples were calculated on the basis of the means of the logarithmic transformed data and the standard deviations [(log of each control value - mean of the log of the control group)/sd of log of the control values)]. The Shapiro-Wilk test showed that the distribution pattern of the control Z-scores were consistent with a normal distribution (p>0.05) with the exception of the Architect i2000 FT3 for which the normality test was very close to normality (p = 0.041). The correlations among the control Z-scores were determined, as well as the 95% confidence ellipses for each pair of assays. Deming regression, assuming equal variances, showed a slope of 1.00 and intercept of zero for all pairs of FT4 and FT3 control Z-scores. FT4 and FT3 Z-scores for the groups with thyroid illnesses and NTI were also determined (using the corresponding control means and SD for each assay). Deming regression was also used to determine the slopes and intercepts, with their 95% confidence intervals, for each pair of FT4 and FT3 Z-scores among all three assays, in the patients with thyroid diseases and NTI. The observation of slopes different from unity would suggest differences of immunoassay behaviors for samples from diseased patients. We also computed the between-assay agreements (kappa statistic) for FT4 and FT3 in thyroid diseases and acute NTI after classification of their Z-scores, using the 2.5 SD limits of the control data of each assay. The strength of the agreement was evaluated following the guidelines: <0.20: poor; : fair; : moderate; : good; : very good [13]. Statistical analysis was performed with JMP, version 8.0.1, from SAS Institute, Analyse-it for Excel, version 2.05 and MedCalc for Windows, version (MedCalc Software, Mariakerke, Belgium). RESULTS Relations between FT4 values obtained with the three assays including all subjects (n = 173) are depicted in Figure 1 (panels A - C). The Elecsys FT4 values were systematically higher than those generated with DxI 800 or the Architect. In contrast, FT4 levels determined with the DxI 800 compared to the Architect were closer to each other. For FT3 (Figure 1; panels D - F), the larger between-assay differences were observed when Elecsys results were compared to those of the DxI 800. One patient included in the analysis had extremely high FT3 values (15.2, 18.2, and 18.9 pmol/l, with the DxI 800, Architect, and Elecsys assays, respectively). The corresponding FT4 values were also very high (42.60, 35.91, and pmol/l), and TSH levels were undetectable (<0.01 µu/ml). These results were from a 49 year old female who at the time of blood sampling presented clinically overt hyperthyroidism, due to acute recurrence of Basedow disease, caused by the fact that the patient had interrupted her anti thyroid drug therapy. FT4 levels in the control group were positively related to age with a high degree of significance, a 20 30% increase being observed over the age span of years. This relation was observed with all three assays (r = 0.473, 0.545, and 0.396; DxI 800, Architect, and Elecsys assays, respectively; p<0.001). In contrast, FT3 and age showed no significant correlation. Table 2 represents estimations of normal intervals for FT4 and FT3 assays derived from the control group. Mean FT4 values obtained with the DxI 800 were 10% and 29% lower than those measured with the Architect i2000 or the Elecsys 2010, respectively. The mean FT3 values with the DxI 800 were 20% and 12% higher than those measured with the Architect i2000 or the Elecsys 2010, respectively. These differences between groups were highly significant (p<0.001). For comparison, the estimated normal reference ranges stated by the manufacturers in the data inserts are summarized in Table 3. Figure 2 shows the distribution of the Z-scores for FT4 and FT3 in the control group, as well as their correlations among the three assays; the density ellipse, a surface containing 95% of all observations is also represented. The width of the density ellipse was related to the correlation coefficients between two variables. The coefficients of correlation among the different pairs of assays showed great variability. The strongest correlation of Z-scores was observed for the relation between the FT4 levels obtained with the Elecsys 2010 and the DxI 800 (despite the fact that the absolute values are very different). For FT3, the best correlation of Z-scores was obtained when the Elecsys 2010 was related to the Architect i2000. Deming regressions of the Z-scores between all FT4 and FT3 assays from controls had slopes of 1.00 (95% confidence intervals as on the Figure), with an intercept of zero. In the 60 patients with thyroid diseases, the FT4 Z- scores were very similar in the three assays, as they ranged from -4.6 to 7.7 (interquartile: ), -4.2 to 7.0 (interquartile: ) and -6.5 to 7.6 (interquartile: ), with the DxI 800, the Architect i2000, and the Elecsys 2010, respectively. For FT3, these values were -5.1 to 12.3 (interquartile: ), -6.6 to 10.8 (interquartile: ), and -6.1 to 8.9 (interquartile: ) (Figure 3). Intercepts of the Deming regressions were not significantly different from zero for any of the pairs of FT4 and FT3 assays. The slope of FT4 - DxI 800 versus FT4 - Elecsys 2010 Z-scores was slightly above unity (p<0.05); in contrast, the slopes of Clin. Lab. 7+8/

4 C. FILLÉE et al. Table 1. Patients characteristics. Pathology N Age a (years) Gender ratio (M/F) TSH (mu/l) a Controls Patients (outpatients or hospitalized) under stable conditions with pathologies unrelated to thyroid disorders, no severe clinical conditions and undetectable atpo (16-94) [35-66] b 29/ ( ) Basedow disease at various states of treatment by surgery, radioiodone or antithyroid agents (31-84) 6/ (< ) Thyroidectomy for papillary / follicular cancer (n = 11) or multinodular goiter (n = 2) (19-67) 3/ ( ) Thyroidal illnesses Primary hypothyroidism,t4 substituted (2-76) 2/ (< ) Secondary hypothyroidism,t4 substituted 3 25; 54; 76 c 2/1 0.02; 0.02; 0.21 Euthyroid goiter 6 48 (35-65) 1/ ( ) Acute cardiovascular condition / Arrhythmia (38-81) 18/ ( ) Severe / acute nonthyroidal clinical conditions Inflammatory / infectious states (38-96) 9/ ( ) Malignancies under treatment (17-72) 5/ ( ) a median (range) b [interquartile range], values determined with DxI 800 c individual values Table 2. FT4 and FT3 values in 68 control subjects using DxI 800, Architect, and Elecsys. FT4 (pmol/l) FT3 (pmol/l) DxI 800 Architect Elecsys DxI 800 Architect Elecsys Geometrical means % Reference limits % Confidence interval of lower limits % Confidence interval of upper limits Clin. Lab. 7+8/2012

5 COMPARISON OF FT4 AND FT3 IMMUNOASSAYS Table 3. Expected normal values for FT4 and FT3 in adults as stated in the manufacturer s data insert. Manufacturer N FT4 (pmol/l) N FT3 (pmol/l) Beckman Coulter UniCel DxI 800 Abbott Architect i2000 Roche Elecsys a b e c f d 5, g h a 95% confidence interval; USA population (North, Central, West); years old; TSH mu/l b FT4 values reported as pmol/l until December 2006, data insert of Access 2 c 99% central interval d % non parametric percentile e 95th percentile (median: 4.9 pmol/l) f central 95% interval g population from a private laboratory, near the German coast; TSH: mu/l; non-parametric 95% confidence limits h samples from apparently healthy blood donors; population from Central Germany; non-parametric 95% confidence limits Table 4. Agreement statistics for the classification of Z-scores obtained with the different FT4 and FT3 assays, in thyroid diseases and acute/critical illness conditions a, b. Thyroid diseases (n = 60) FT4 FT3 DxI 800 vs. Elecsys (0.104) (0.108) DxI 800 vs. Architect (0.096) (0.164) Elecsys vs Architect (0.138) (0.147) Acute/Critical illness (n = 45) FT4 FT3 DxI 800 vs. Elecsys (0.204) (0.195) DxI 800 vs. Architect (0.151) (0.164) Elecsys vs. Architect (0.229) (0.152) a Kappa (standard error) b Classifications are based upon the ±2.5 SD limits of the control group in each assay, as represented in Figures 4 and 5 the Deming regressions relating FT4 DxI 800 versus FT4 - Abbott and FT4 Elecsys 2010 versus FT4 Architect Z-scores were slightly below unity (p<0.05). Comparisons of the FT3 Z-scores showed a slope slightly below unity for the DxI 800 versus the Elecsys 2010 assays (p<0.05), while the slopes were within the limits of unity for DxI 800 versus Architect and Elecsys versus Architect. In 45 patients with acute NTI, the range of pathological FT4 and FT3 Z-scores was narrower than in the thyroid diseases (Figure 4). Moreover, FT4 Z-scores were systematically and significantly (p<0.001) higher than the Clin. Lab. 7+8/

6 C. FILLÉE et al. Figure 1. Passing-Bablok regressions of FT4 (left panel) and FT3 (right panel), using data from controls, thyroid diseases, and non thyroidal severe illnesses obtained with the DxI 800, Architect i2000, and Elecsys 2010, compared pairwise. The actual values of one patient with extremely high FT3 values are given in parentheses. 730 Clin. Lab. 7+8/2012

7 COMPARISON OF FT4 AND FT3 IMMUNOASSAYS Figure 2. Correlations between the Z-scores of FT4 (left panel) and FT3 (right panel) of the control group, determined with the DxI 800, Architect i2000, and Elecsys 2010, compared pairwise. The red lines represent the 95% confidence ellipses of the regression. The distributions of the Z-scores for each assay are represented at the upper and right side of the regression graphs. The dotted lines represent the Deming regressions; the intercepts are zero and the slopes are 1.00; the 95% confidence limits of the slopes are represented on the Figures. Clin. Lab. 7+8/

8 C. FILLÉE et al. Figure 3. Z-scores of FT4 (left panel) and FT3 (right panel) of patients with thyroid diseases determined with the DxI 800, Architect i2000, and Elecsys 2010, compared pairwise. The red lines represent the 95% confidence ellipses of the control groups. The dotted lines represent the Deming regressions; the slopes and intercepts with their 95% confidence intervals are represented on the figures. O: Basedow disease; Δ: hypothyroidism; : euthyroid goiter (see Table 1). 732 Clin. Lab. 7+8/2012

9 COMPARISON OF FT4 AND FT3 IMMUNOASSAYS Figure 4. Z-scores of FT4 (left panel) and FT3 (right panel) of patients with severe non thyroidal illnesses determined with the DxI 800, Architect i2000, and Elecsys 2010, compared pairwise. The red lines represent the 95% confidence ellipses of the control groups. The dotted lines represent the Deming regressions; the slopes and intercepts with their 95% confidence intervals are represented on the figures. О: inflammatory/infection; Δ: malignancies; ; acute cardiovascular condition (see Table 1). Clin. Lab. 7+8/

10 C. FILLÉE et al. FT3 Z-scores with all three assays. The intercepts and slopes of the Deming regressions were not significantly different from zero and unity, respectively. Table 4 shows the between assay agreements after classification of FT4 and FT3 Z-scores according to ±2.5 SD of their respective controls. For FT4 Z-scores, agreements in cases of thyroid diseases and acute NTI, the kappa statistic showed good agreement, except for Elecsys 2010 versus Architect i2000, which was moderate. For FT3 Z-scores, agreements were moderate to good in cases of thyroid diseases; the lowest kappa statistics (still moderate) were observed for FT3 Z-scores in acute NTI. DISCUSSION In this report, we have compared the FT4 and FT3 values obtained with three automated systems, DxI 800, Architect i2000, and Elecsys 2010, in a control group as well as in patients with various stages of thyroid diseases and patients with severe/acute non thyroidal illnesses (NTI). Significant inter-assay biases were observed. To allow inter-assay comparison, we normalized the control values on the basis of Z-scores and expressed the FT4 and FT3 from the patient groups as Z- scores, using the corresponding control references. All assay Z-score pairs were compared. There was considerable variation among the correlation coefficients indicating differences in random error among assays. In the control group, slopes and intercepts of the regressions of paired assays did not differ from 1.00 and zero; these results are in agreement with the validity of Z-score normalization in these cases. In the patients with thyroid disease, comparison of the slopes of paired FT4 and FT3 Z-scores revealed modest or no significant departures from unity suggesting some differences in assay behaviors in these cases. In NTI patients, regression analysis did not reveal such between-assay differences. Between-assay classification agreements in all cases were rated from moderate to very good. FT4 values generated with the Elecsys are considerably higher than those reported by DxI 800 or Architect i2000. Results of the DxI 800 and the Architect i2000 were closer to each other than those of the Elecsys The important divergence of the Elecsys FT4 from DxI and Architect i2000 may possibly result from assay design differences. Indeed, the Elecsys assay is based on a one step labeled anti-t4 antiserum, while the two others make use of the 2-step procedure FT4 extraction on solid phase anti-t4 antibodies followed by back-titration of the remaining free antibodies with labeled T3 analogue. Differences in the composition of the first buffers may also be of importance. Antibody affinities and their interactions with binding proteins (thyroxin binding protein (TBG) and transthyretin (TTR)) may also play a role in the inter-assay differences. FT3 values also exhibited significant differences among assays, the Architect i2000 and the Elecsys showing the least differences. Inter-assay variations observed in this study for free thyroid hormones are in line with many previous reports [7,14-15]. The FT4 limits derived from our control group obtained by DxI 800 ( pmol/l) are not in line with the manufacturer s data insert ( pmol/l) nor with the report of d Herbomez et al. ( pmol/l) [16]. Indeed, we observed an upper control limit ~13% higher. This may be due to an increase of FT4 with high age, together with the fact that our control group consisted of subjects from years old, while the cohorts used by the DxI 800 manufacturer and d Herbomez et al. were younger (20-60 years and years, respectively). The increase of FT4 with age was clearly evidenced in our control group, not only with the DxI 800 assay but also with the Architect i2000 and the Elecsys. The age-dependence of FT4 observed with all three immunoassays in this study may indirectly and partly be related to changes in the TBG and TTR, especially after 50 years. Our observation of a positive relation between age and FT4 levels determined by immunoassays is not unique and has been documented in other studies [17-20]. Thorough evaluations based on well defined populations are thus required to assess references applicable to specific clinical settings. FT4 values of our control group determined with the Architect i2000 or the Elecsys 2010 were close to the expected values of the data inserts. This was also observed by others with both assays [21,22], with the exception of one report with upper values with the Architect i2000 exceeding by ~40% the expected value [23]. FT3 values of our control group were also similar to the expected values described in the data inserts, as well as those from several other reports [15,21-23]. The recent report of the IFCC working group for standardization of thyroid tests (free thyroxin and free triodothyronine) [7] did highlight the important biases of most immunoassays for these free hormones. The report also shows that these biases may be largely corrected by an international conventional reference measurement procedure. This would not solve all problems, as the report also points to the fact that precision, within-run stability, and between-run consistency should be improved for some assays. Important information from this report is that mathematical recalibration of most of the results of the assays, based upon their relation to a reference LC-MS/MS method eliminates the biases. The authors, however, warn that after such recalibration, large sample-related effects remained. In the present report, we normalized the data of the three assays by expressing the results of control groups in terms of Z-sores. It appeared that all Z-scores describing the control values were, as expected, centered on a null value and that their distribution did not deviate from normality, except for the FT3 Architect i2000 for which the data were lightly skewed (Shapiro-Wilk = 0.041). The correlation coefficients of these Z-scores across assays were highly significant, even though the strength of these relations varied considerably, indicat- 734 Clin. Lab. 7+8/2012

11 COMPARISON OF FT4 AND FT3 IMMUNOASSAYS ing important variations in inter-assay scatter. For FT4, the strongest coefficient of correlation was observed between the DxI 800 and the Elecsys values. For FT3, there was less disparity, but the best relation was between the Architect i2000 and the Elecsys. These differences in the strength of correlation between Z-scores of the controls were visually reflected by the width of their 95% confidence ellipses. These observations are in line with the conclusions of the IFCC report, concerning the relative variations among assays as well as possible differences in precision. The Deming regression lines, assuming equal variance of both axes for each pair of FT4 and FT3 control Z-scores, were, in all cases, equal to unity with an intercept of zero, providing indirect evidence for the validity of the Z-score transformation. Of the 60 patients with thyroid diseases, most of the data were confined within the normality ellipses for all three FT4 and FT3 assays and consistent with therapeutic action leading to normalization or near normalization of thyroid function in this group. The slopes of the Deming regressions comparing FT4 Z-scores of assay pairs in these patients showed a slight but significant departure from unity, suggesting that quantitative extrapolation from one of these assays to the other must be taken with some caution. For the FT3 Z-score values in the thyroid cases, the regression slope did not depart from unity except for the relation of the DxI 800 compared to Elecsys For the group of patients with NTI, the mean FT3 Z- scores were lower than those for FT4. This is in accordance with alterations in peripheral conversion of FT4 to FT3. It is possible that the alternative pathway (conversion of T4 to inactive rt3) is activated, resulting in the production of lower amounts of bioactive T3. This was particularly marked for the patients with acute infections. Intercepts of the regression lines were not different from zero and the slopes of the regression lines were consistent with unity suggesting that differences in the assay behavior in this group of patients was not statistically rejected. We also used the Kappa statistics to evaluate the between-assay agreements of the data expressed as Z- scores. They were ranked as moderate to very good when the Z-scores for the patients with different pathological conditions were classified on the basis of the ±2.5 SD limits of the controls. These results indicate that all three assays are of clinical value, provided that appropriate reference values are used. It remains that for some pairs of assays, the concordance is only moderate, and caution should be exercised in extrapolating the clinical classification from one assay to another. This study has several limitations. First, the control group is relatively small. It represents patients attending a hospital physician, even though they were proven to be free of any conditions which could possibly interfere with the thyroid function, that their TSH values were normal, and atpo was undetectable. However, this is the kind of population, in which decisions have to be made on the presence or absence of significant thyroid dysfunction in a hospital based population, a situation quite different from general practice. Again, as already evidenced by several studies, locally normal values should always be checked by each laboratory, and data insert expected values can only be used for general guidance. The problem of a multicenter approach for derivation of reference intervals for thyroid hormones for laboratories, using identical analyzers has already been discussed by Andrew et al. [24]. All our control subjects had undetectable anti-tpo antibodies, measured with a highly sensitive method, but the anti-tg antibodies could not be measured due to practical constraints in all of them. However, a recent report concludes that anti-tg antibody alone, in the absence of anti-tpo is not significantly associated with thyroid diseases [25]. In conclusion, we confirmed the wide disparity in FT4 and FT3 concentrations derived from three immunoassays. Normalization of the results by transformation as Z-scores allows the elimination of between-assay differences. However, caution must be exercised when precise quantitative inter-assay comparisons are performed, as regression analysis comparing assays show that not all behave similarly, particularly in a group of patients with thyroid diseases. Classification of the results in thyroid diseases and NTI among pairs of assays showed moderate to very good agreement on the basis of their respective control Z-scores. It remains that any clinically or biologically questionable results should be referred to a reference laboratory, using physical separation of free hormones followed by their measurement by LC- MS/MS. In the future, it will be desirable to use international conventional reference measurands, or better, to implement routinely automated and standardized LC- MS/MS methods. Disclaimers: None. Declaration of Interest: None. Financial Support: Department of Clinical Biology and Pathology - Cliniques Universitaires St-Luc, Université Catholique de Louvain, Brussels, Belgium. References: 1. Baloch Z, Carayon P, Conte-Devolx B, Demers LM, Feldt-Rasmussen U, Henry JF, et al. Laboratory medicine practice guidelines. Laboratory support for the diagnosis and monitoring of thyroid disease. Thyroid 2003;13: Thienpont LM, Van UK, Beastall G, et al. Report of the IFCC Working Group for Standardization of Thyroid Function Tests; part 1: thyroid-stimulating hormone. Clin Chem 2010;56: Clin. Lab. 7+8/

12 C. FILLÉE et al. 3. Ekins R. The free hormone hypothesis and measurement of free hormones. Clin Chem 1992;38: Nelson JC, Wilcox RB. Analytical performance of free and total thyroxine assays. Clin Chem 1996;42: Christofides ND, Wilkinson E, Stoddart M, Ray DC, Beckett GJ. Assessment of serum thyroxine binding capacity-dependent biases in free thyroxine assays. Clin Chem 1999;45: Stockigt JR. Free thyroid hormone measurement. A critical appraisal. Endocrinol Metab Clin North Am 2001;30: Thienpont LM, Van UK, Beastall G, et al. Report of the IFCC Working Group for Standardization of Thyroid Function Tests; part 2: free thyroxine and free triiodothyronine. Clin Chem 2010; 56: Jonklaas J, Kahric-Janicic N, Soldin OP, Soldin SJ. Correlations of free thyroid hormones measured by tandem mass spectrometry and immunoassay with thyroid-stimulating hormone across 4 patient populations. Clin Chem 2009;55: Holm SS, Hansen SH, Faber J, Staun-Olsen P. Reference methods for the measurement of free thyroid hormones in blood: evaluation of potential reference methods for free thyroxine. Clin Biochem 2004;37: Van Uytfanghe K, Stockl D, Ross HA, Thienpont LM. Use of frozen sera for FT4 standardization: investigation by equilibrium dialysis combined with isotope dilution-mass spectrometry and immunoassay. Clin Chem 2006;52: Thienpont LM, Beastall G, Christofides ND, et al. Proposal of a candidate international conventional reference measurement procedure for free thyroxine in serum. Clin Chem Lab Med 2007;45: Yue B, Rockwood AL, Sandrock T, La'ulu SL, Kushnir MM, Meikle AW. Free thyroid hormones in serum by direct equilibrium dialysis and online solid-phase extraction-liquid chromatography/tandem mass spectrometry. Clin Chem 2008;54: Altman DG. Practical Statisitcs for Medical Research. Chapman & Hall/CRC, Steele BW, Wang E, Klee GG, et al. Analytic bias of thyroid function tests: analysis of a College of American Pathologists fresh frozen serum pool by 3900 clinical laboratories. Arch Pathol Lab Med 2005;129: Sapin R, D'Herbomez M. Free thyroxine measured by equilibrium dialysis and nine immunoassays in sera with various serum thyroxine-binding capacities. Clin Chem 2003;49: D'Herbomez M, Jarrige V, Darte C. Reference intervals for serum thyrotropin (TSH) and free thyroxine (FT4) in adults using the Access Immunoassay System. Clin Chem Lab Med 2005;43: Hoogendoorn EH, Hermus AR, de VF, et al. Thyroid function and prevalence of anti-thyroperoxidase antibodies in a population with borderline sufficient iodine intake: influences of age and sex. Clin Chem 2006;52: Ross HA, den HM, Hermus AR, Sweep FC. Composite reference interval for thyroid-stimulating hormone and free thyroxine, comparison with common cutoff values, and reconsideration of subclinical thyroid disease. Clin Chem 2009;55: Davey R. Thyroxine, thyrotropin, and age in a euthyroid hospital patient population. Clin Chem 1997;43: Hershman JM, Pekary AE, Berg L, Solomon DH, Sawin CT. Serum thyrotropin and thyroid hormone levels in elderly and middle-aged euthyroid persons. J Am Geriatr Soc 1993;41: Gonzalez-Sagrado M, Martin-Gil FJ. Population-specific reference values for thyroid hormones on the Abbott ARCHITECT i2000 analyzer. Clin Chem Lab Med 2004;42: Kratzsch J, Fiedler GM, Leichtle A, et al. New reference intervals for thyrotropin and thyroid hormones based on National Academy of Clinical Biochemistry criteria and regular ultrasonography of the thyroid. Clin Chem 2005;51: Hubl W, Schmieder J, Gladrow E, Demant T. Reference intervals for thyroid hormones on the Architect analyser. Clin Chem Lab Med 2002;40: Andrew CE, Hanning I, McBain AM, Moody D, Price A. A model for a multicentre approach to the derivation of reference intervals for thyroid hormones and testosterone for laboratories using identical analysers. Clin Chem Lab Med 2000;38: Hollowell JG, Staehling NW, Flanders WD, Hannon WH, Gunter EW, Spencer CA, Braverman LE, et al. Serum TSH, T(4), and thyroid antibodies in the United States population (1988 to 1994): National Health and Nutrition Examination Survey (NHANES III). J Clin Endocrinol Metab 2002;87: Correspondence: Catherine Fillée Cliniques Universitaires St-Luc Université Catholique de Louvain Avenue Hippocrate, Brussels, Belgium Tel.: Fax: catherine.fillee@uclouvain.be 736 Clin. Lab. 7+8/2012

PDF hosted at the Radboud Repository of the Radboud University Nijmegen

PDF hosted at the Radboud Repository of the Radboud University Nijmegen PDF hosted at the Radboud Repository of the Radboud University Nijmegen The following full text is a publisher's version. For additional information about this publication click this link. http://hdl.handle.net/2066/81732

More information

Reference Intervals for Children and Adults

Reference Intervals for Children and Adults for Children and Adults TSH, FT4, FT3, T4, T3, T-Uptake, FT4-index, Anti-TPO, Anti-Tg, Tg Elecsys systems /20 MODULAR ANALYTICS E70 cobas e 4 and cobas e 60 analysers Contents Page Introduction 4 2 Summary

More information

Evaluation of the analytical performances of six measurands for thyroid functions of Mindray CL-2000i system

Evaluation of the analytical performances of six measurands for thyroid functions of Mindray CL-2000i system Original Article Page 1 of 7 Evaluation of the analytical performances of six measurands for thyroid functions of Mindray CL-2i system Andrea Padoan, Chiara Cosma, Mario Plebani Department of Laboratory

More information

Standardization of Thyroid Function Tests

Standardization of Thyroid Function Tests Standardization of Thyroid Function Tests Saturday, Dec 3 rd 211 39th meeting Location: Nycomed Belgium 18 Brussels Linda Thienpont Linda.thienpont@ugent.be Thyroid dysfunction: clinical importance Relatively

More information

Free hormone estimates. Never ending story.

Free hormone estimates. Never ending story. Free hormone estimates Never ending story. Roger Ekins (1926-2016) - saturation analysis - comp. immunoassay - microarray - RIA (T4, B12) - Free T4 (never really succeeded) Free hormone hypothesis Law

More information

Research. Although thyrotropin (thyroidstimulated

Research. Although thyrotropin (thyroidstimulated Research www.ajog.org OBSTETRICS Free T4 immunoassays are flawed during pregnancy Richard H. Lee, MD; Carole A. Spencer, PhD; Jorge H. Mestman, MD; Erin A. Miller, BS; Ivana Petrovic, MS; Lewis E. Braverman,

More information

Table 1: Thyroid panel. Result (reference interval) TSH 89.5 miu/l ( ) Total T4 5.2 µg/dl ( ) T3 uptake 39% (22-35)

Table 1: Thyroid panel. Result (reference interval) TSH 89.5 miu/l ( ) Total T4 5.2 µg/dl ( ) T3 uptake 39% (22-35) Introduction Thyroid disease is the second most common endocrine disorder (behind diabetes), and its prevalence increases with increasing age. The incidence of newly diagnosed thyroid cancer is increasing

More information

Review Article Challenges in Interpretation of Thyroid Function Tests in Pregnant Women with Autoimmune Thyroid Disease

Review Article Challenges in Interpretation of Thyroid Function Tests in Pregnant Women with Autoimmune Thyroid Disease SAGE-Hindawi Access to Research Journal of Thyroid Research Volume 2011, Article ID 598712, 7 pages doi:10.4061/2011/598712 Review Article Challenges in Interpretation of Thyroid Function Tests in Pregnant

More information

Discontinuation of Smoking Increases the Risk for Developing Thyroid Peroxidase Antibodies and/or Thyroglobulin Antibodies: A Prospective Study

Discontinuation of Smoking Increases the Risk for Developing Thyroid Peroxidase Antibodies and/or Thyroglobulin Antibodies: A Prospective Study ORIGINAL ARTICLE Endocrine Care Brief Report Discontinuation of Smoking Increases the Risk for Developing Thyroid Peroxidase Antibodies and/or Thyroglobulin Antibodies: A Prospective Study Grigoris Effraimidis,

More information

Free Thyroid Hormones in Serum by Direct Equilibrium Dialysis and On-line Solid-Phase Extraction Liquid Chromatography/Tandem Mass Spectrometry

Free Thyroid Hormones in Serum by Direct Equilibrium Dialysis and On-line Solid-Phase Extraction Liquid Chromatography/Tandem Mass Spectrometry Papers in Press. Published February 7, 2008 as doi:10.1373/clinchem.2007.098293 The latest version is at http://www.clinchem.org/cgi/doi/10.1373/clinchem.2007.098293 Clinical Chemistry 54:4 000 000 (2008)

More information

IFCC Committee for Standardization of Thyroid Function Tests C-STFT- Update on the project

IFCC Committee for Standardization of Thyroid Function Tests C-STFT- Update on the project IFCC Committee for Standardization of Thyroid Function Tests C-STFT- Update on the project Maria-Magdalena Patru, MD, PhD Director, Medical and Scientific Affairs, Ortho Clinical Diagnostics, Member C-STFT

More information

Evaluation of maternal thyroid function during pregnancy: the importance of using gestational age-specific reference intervals

Evaluation of maternal thyroid function during pregnancy: the importance of using gestational age-specific reference intervals European Journal of Endocrinology (2007) 157 509 514 ISSN 0804-4643 CLINICAL STUDY Evaluation of maternal thyroid function during pregnancy: the importance of using gestational age-specific reference intervals

More information

Biomed Environ Sci, 2017; 30(6):

Biomed Environ Sci, 2017; 30(6): Biomed Environ Sci, 2017; 30(6): 455-459 455 Letter to the Editor Serum Fetuin-A Levels and Thyroid Function in Middle-aged and Elderly Chinese * DENG Xin Ru 1,2,&, DING Lin 1,2,&, WANG Tian Ge 1,2, XU

More information

Chapter 3. Autoimmunity and Hypothyroidism: Anti-TPO antibodies in Hypothyroid Patients in Gujarat Population

Chapter 3. Autoimmunity and Hypothyroidism: Anti-TPO antibodies in Hypothyroid Patients in Gujarat Population Chapter 3. Autoimmunity and Hypothyroidism: Anti-TPO antibodies in Hypothyroid Patients in Gujarat Population I. INTRODUCTION Hypothyroidism is an endocrine disorder characterized by decreased activity

More information

Original Article. 106

Original Article. 106 Original Article Endocrinol Metab 217;32:16-114 https://doi.org/1.383/enm.217.32.1.16 pissn 293-596X eissn 293-5978 Thyroid Stimulating Hormone Reference Range and Prevalence of Thyroid Dysfunction in

More information

Analytic Bias of Thyroid Function Tests. Analysis of a College of American Pathologists Fresh Frozen Serum Pool by 3900 Clinical Laboratories

Analytic Bias of Thyroid Function Tests. Analysis of a College of American Pathologists Fresh Frozen Serum Pool by 3900 Clinical Laboratories Analytic Bias of Thyroid Function Tests Analysis of a College of American Pathologists Fresh Frozen Serum Pool by 3900 Clinical Laboratories Bernard W. Steele, MD; Edward Wang, PhD; George G. Klee, MD,

More information

Decoding Your Thyroid Tests and Results

Decoding Your Thyroid Tests and Results Decoding Your Thyroid Tests and Results Wondering about your thyroid test results? Learn about each test and what low, optimal, and high results may mean so you can work with your doctor to choose appropriate

More information

Thyroid in the elderly. Akbar Soltani M.D. Endocrinology and Metabolism Research Center (EMRC) Shariati Hospital

Thyroid in the elderly. Akbar Soltani M.D. Endocrinology and Metabolism Research Center (EMRC) Shariati Hospital Thyroid in the elderly Akbar Soltani M.D. Endocrinology and Metabolism Research Center (EMRC) Shariati Hospital soltania@tuma.ac.ir Case 1 A 79 year old female is seen because of a 6 month history of fatigue,

More information

Performance Characteristics of Eight Estradiol Immunoassays

Performance Characteristics of Eight Estradiol Immunoassays Clinical Chemistry / EVALUATION OF EIGHT ESTRADIOL IMMUNOASSAYS Performance Characteristics of Eight Estradiol Immunoassays David T. Yang, MD, 1 William E. Owen, MT(ASCP), 2 Carol S. Ramsay, 3 Hui Xie,

More information

concentration in young people.

concentration in young people. NAOSITE: Nagasaki University's Ac Title Author(s) Citation Prevalence of antithyroid antibodie concentration in young people. Sekitani, Yui; Hayashida, Naomi; Ka Kozlovsky, Alexander; Yamashita, Sh Clinical

More information

Prevalence of thyroid autoimmunity and subclinical hypothyroidism in persons with chronic kidney disease not requiring chronic dialysis

Prevalence of thyroid autoimmunity and subclinical hypothyroidism in persons with chronic kidney disease not requiring chronic dialysis Clin Chem Lab Med 2009;47(11):1367 1371 2009 by Walter de Gruyter Berlin New York. DOI 10.1515/CCLM.2009.304 2009/284 Prevalence of thyroid autoimmunity and subclinical hypothyroidism in persons with chronic

More information

DBC s panel of thyroid hormone immunoassays is now more comprehensive with the new Reverse T3 ELISA kit. Test Your Hormones

DBC s panel of thyroid hormone immunoassays is now more comprehensive with the new Reverse T3 ELISA kit. Test Your Hormones DBC s panel of thyroid hormone immunoassays is now more comprehensive with the new Reverse T3 ELISA kit Test Your Hormones INTRODUCTION For years thyroid hormone testing has been concentrated on TSH and

More information

Subclinical Hyperthyroidism: Considerations in Defining the Lower Limit of the Thyrotropin Reference Interval

Subclinical Hyperthyroidism: Considerations in Defining the Lower Limit of the Thyrotropin Reference Interval Papers in Press. Published January 15, 2009 as doi:10.1373/clinchem.2008.110627 The latest version is at http://www.clinchem.org/cgi/doi/10.1373/clinchem.2008.110627 Clinical Chemistry 55:3 000 000 (2008)

More information

Harmonisation of TFTs

Harmonisation of TFTs Royal Prince Alfred Hospital and University of Sydney Harmonisation of TFTs Harmonisation Workshop May 2015 Paul Williams Harmonisation May 2015 Talk Outline IFCC initiatives C-STFT (Committee for standardisation

More information

Measurement of total thyroxine concentration in serum from dogs and cats by use of various methods Objective Sample Population Procedure Results

Measurement of total thyroxine concentration in serum from dogs and cats by use of various methods Objective Sample Population Procedure Results Measurement of total thyroxine concentration in serum from dogs and cats by use of various methods Robert J. Kemppainen, DVM, PhD, and Jeremy R. Birchfield, DVM Objective To compare results obtained from

More information

Sanjay B. Dixit, M.D. BHS Endocrinology Associates November 11, 2017

Sanjay B. Dixit, M.D. BHS Endocrinology Associates November 11, 2017 Sanjay B. Dixit, M.D. BHS Endocrinology Associates November 11, 2017 I will not be discussing this Outline of discussion Laboratory tests for thyroid function Diagnosis of hypothyroidism Treatment of

More information

Chapter I.A.1: Thyroid Evaluation Laboratory Testing

Chapter I.A.1: Thyroid Evaluation Laboratory Testing Chapter I.A.1: Thyroid Evaluation Laboratory Testing Jennifer L. Poehls, MD and Rebecca S. Sippel, MD, FACS THYROID FUNCTION TESTS Overview Thyroid-stimulating hormone (TSH) is produced by the anterior

More information

Multicenter Comparison of Four Automated Immunoassay Analyzers for Prostate Specific Antigen

Multicenter Comparison of Four Automated Immunoassay Analyzers for Prostate Specific Antigen Letter to the Editor Clinical Chemistry Ann Lab Med 19;39:6-41 https://doi.org/1.3343/alm.19.39.4.6 ISSN 2234-36 eissn 2234-3814 Multicenter Comparison of Four Automated Immunoassay Analyzers for Prostate

More information

TSH ELISA Kit Medical Device Licence No.: 16419

TSH ELISA Kit Medical Device Licence No.: 16419 TSH ELISA Kit Medical Device Licence No.: 16419 Enzyme immunoassay kit for the quantitative determination of TSH concentration in serum. Catalog Number: SL100305 Catalog Number: SL100306 Catalog Number:

More information

Early stages of thyroid autoimmunity: follow-up studies in the Amsterdam AITD cohort Effraimidis, G.

Early stages of thyroid autoimmunity: follow-up studies in the Amsterdam AITD cohort Effraimidis, G. UvA-DARE (Digital Academic Repository) Early stages of thyroid autoimmunity: follow-up studies in the Amsterdam AITD cohort Effraimidis, G. Link to publication Citation for published version (APA): Effraimidis,

More information

Thyroid profile in geriatric population

Thyroid profile in geriatric population Original article: Thyroid profile in geriatric population Dr. Abhijit Pratap, Dr. Mona A. Tilak, Dr. Pradnya Phalak Dept of Biochemistry, Dr. D. Y. Patil Medical College, Pimpri, Pune 18 Corresponding

More information

Comparison of Five Automated Serum and Whole Blood Folate Assays

Comparison of Five Automated Serum and Whole Blood Folate Assays Clinical Chemistry / FIVE AUTOMATED FOLATE ASSAYS Comparison of Five Automated Serum and Whole Blood Folate Assays William E. Owen, MT(ASCP), 1 and William L. Roberts, MD, PhD 2 Key Words: Hemolysate;

More information

Tandem Mass Spectrometry as a Novel Tool for Elucidating Pituitary Thyroid Relationships

Tandem Mass Spectrometry as a Novel Tool for Elucidating Pituitary Thyroid Relationships THYROID Volume 18, Number 12, 2008 ª Mary Ann Liebert, Inc. DOI: 10.1089=thy.2008.0155 Tandem Mass Spectrometry as a Novel Tool for Elucidating Pituitary Thyroid Relationships Jacqueline Jonklaas 1 and

More information

Endocrine Journal 2011, 58 (11),

Endocrine Journal 2011, 58 (11), Endocrine Journal 2011, 58 (11), 995-1002 Or i g i n a l An epidemiological study of the serum thyrotropin reference range and factors that influence serum thyrotropin levels in iodine sufficient areas

More information

Gentian Canine CRP Immunoassay Application Note for scil VitroVet*

Gentian Canine CRP Immunoassay Application Note for scil VitroVet* v02-jan2014 Gentian Canine CRP Immunoassay Application Note for scil VitroVet* Intended Use The canine CRP immunoassay on scil VitroVet is an in vitro diagnostic test for quantitative determination of

More information

Thyroid function after assisted reproductive technology in women free of thyroid disease

Thyroid function after assisted reproductive technology in women free of thyroid disease Thyroid function after assisted reproductive technology in women free of thyroid disease Kris Poppe, M.D., a Daniel Glinoer, M.D., Ph.D., b Herman Tournaye, M.D., Ph.D., c Johan Schiettecatte, c Patrick

More information

Natural History and Comorbidities of Subjects with Subclinical Hyperthyroidism: Analysis at a Tertiary Hospital Setting

Natural History and Comorbidities of Subjects with Subclinical Hyperthyroidism: Analysis at a Tertiary Hospital Setting 506 Original Article Natural History and Comorbidities of Subjects with Subclinical Hyperthyroidism: Analysis at a Tertiary Hospital Setting Shaikh Abdul Kader Kamaldeen Abdul Shakoor, 1 MBBS, MD, MRCP

More information

PREVALENCE OF THYROID DISORDERS IN A TERTIARY CARE CENTER

PREVALENCE OF THYROID DISORDERS IN A TERTIARY CARE CENTER IJCRR Section: Healthcare Sci. Journal Impact Factor 4.016 Research Article PREVALENCE OF THYROID DISORDERS IN A TERTIARY CARE CENTER Deokar PG 1, Nagdeote AN 2, Lanje MJ 3, Basutkar DG 4 1 Senior Resident

More information

Cortisol Assays. The Good, The Bad and The Indifferent. David Ducroq. Cardiff and Vale ulhb WEQAS. Unit 6, Parc Tŷ Glas. Llanishen.

Cortisol Assays. The Good, The Bad and The Indifferent. David Ducroq. Cardiff and Vale ulhb WEQAS. Unit 6, Parc Tŷ Glas. Llanishen. Cortisol Assays The Good, The Bad and The Indifferent David Ducroq Cardiff and Vale ulhb WEQAS Unit 6, Parc Tŷ Glas Llanishen Cardiff www.weqas.com Summary Brief overview of current methods and challenges

More information

Research Article Trimester- and Assay-Specific Thyroid Reference Intervals for Pregnant Women in China

Research Article Trimester- and Assay-Specific Thyroid Reference Intervals for Pregnant Women in China International Endocrinology Volume 2016, Article ID 3754213, 5 pages http://dx.doi.org/10.1155/2016/3754213 Research Article Trimester- and Assay-Specific Thyroid Reference Intervals for Pregnant Women

More information

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY INFORMATION BULLETIN USING THE ACCESS AMH ASSAY IN YOUR LABORATORY ///////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////

More information

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats Authors: Kate Cote, Ph.D., Graham Bilbrough, MA, VetMB, CertVA, MRCVS,

More information

Grave Clinicopathologic Correlation: A Case of Hyperthyroxinemia

Grave Clinicopathologic Correlation: A Case of Hyperthyroxinemia Grave Clinicopathologic Correlation: A Case of Hyperthyroxinemia Andre Mattman, 1 * Mari L. DeMarco, 1 Sophia Wong, 1 Daniel T. Holmes, 1 and Julie Lee 2 CASE DESCRIPTION A family physician ordered a set

More information

Human T4-HRP ELISA Kit Medical Device Licence No.: 21177

Human T4-HRP ELISA Kit Medical Device Licence No.: 21177 Human T4-HRP ELISA Kit Medical Device Licence No.: 21177 Enzyme immunoassay kit for the quantitative determination of T4 concentration in serum. Catalog Number: SL100303 96 tests For in vitro diagnostic

More information

ASSOCIATIONS BETWEEN THYROID DYSFUNCTION AND CHRONIC KIDNEY DISEASE

ASSOCIATIONS BETWEEN THYROID DYSFUNCTION AND CHRONIC KIDNEY DISEASE 2014 ILEX PUBLISHING HOUSE, Bucharest, Roumania http://www.jrdiabet.ro Rom J Diabetes Nutr Metab Dis. 21(1):37-42 doi: 10.2478/rjdnmd-2014-0006 ASSOCIATIONS BETWEEN THYROID DYSFUNCTION AND CHRONIC KIDNEY

More information

THYROID HORMONES & THYROID FUNCTION TESTS

THYROID HORMONES & THYROID FUNCTION TESTS THYROID HORMONES & THYROID FUNCTION TESTS SCHOOL OF MEDICINE AND HEALTH SCIENCES DIVISION OF BASIC MEDICAL SCIENCES DISCIPLINE OF BIOCHEMISTRY AND MOLECULAR BIOLOGY CLINICAL BIOCHEMISTRY LECTURE BMLS III

More information

Screening for thyroid dysfunction in adults

Screening for thyroid dysfunction in adults Screening for thyroid dysfunction in adults External review against programme appraisal criteria for the UK National Screening Committee (UK NSC) Version: Draft 2 Solutions for Public Health June 2017

More information

Complex relationship between free thyroxine and TSH in the regulation of thyroid function

Complex relationship between free thyroxine and TSH in the regulation of thyroid function European Journal of Endocrinology (20) 62 23 29 ISSN 0804-4643 CLINICAL STUDY Complex relationship between free thyroxine and TSH in the regulation of thyroid function Rudolf Hoermann, Walter Eckl, Christian

More information

Original. Jing Cai 1) *, Yujie Fang 1) *, Da Jing 2) *, Shaoyong Xu 1), Jie Ming 1), Bin Gao 1), Han Shen 3), Rong Zhang 1) and Qiuhe Ji 1)

Original. Jing Cai 1) *, Yujie Fang 1) *, Da Jing 2) *, Shaoyong Xu 1), Jie Ming 1), Bin Gao 1), Han Shen 3), Rong Zhang 1) and Qiuhe Ji 1) 2016, 63 (4), 381-388 Original Reference intervals of thyroid hormones in a previously iodine-deficient but presently more than adequate area of Western China a population-based survey Jing Cai 1) *, Yujie

More information

THYROID HORMONES: An Overview

THYROID HORMONES: An Overview 1 SCHOOL OF MEDICINE AND HEALTH SCIENCES DIVISION OF BASIC MEDICAL SCIENCES DISCIPLINE OF BIOCHEMISTRY AND MOLECULAR BIOLOGY PBL SEMINAR MBBS III; BMLS & BDS Year 3 What are the Thyroid Hormones? THYROID

More information

HHS Public Access Author manuscript J Pediatr Gastroenterol Nutr. Author manuscript; available in PMC 2016 August 16.

HHS Public Access Author manuscript J Pediatr Gastroenterol Nutr. Author manuscript; available in PMC 2016 August 16. Hypothyroidism is a rare cause of isolated constipation: 5-year review of all thyroid tests in a pediatric gastroenterology office William E. Bennett Jr., M.D. 1 and Robert O. Heuckeroth, M.D., Ph.D. 2,3

More information

Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000

Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000 v2-jan214 Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4 Intended Use The canine CRP immunoassay on Abbott's Architect c4 is an in vitro diagnostic test for quantitative determination

More information

Thyroid Stimulating Hormone (TSH) ELISA Catalog No. GWB , legacy id (96 Tests)

Thyroid Stimulating Hormone (TSH) ELISA Catalog No. GWB , legacy id (96 Tests) For Research Use Only. Not for use in Diagnostic Procedures. INTENDED USE The GenWay, Inc. TSH ELISA Kit is intended for the quantitative measurement of TSH in human serum or plasma. For research use only.

More information

Triiodothyronine (T3) ELISA

Triiodothyronine (T3) ELISA For Research Use Only. Not for use in Diagnostic Procedures. INTENDED USE The GenWay, Inc. Triiodothyronine (T3) ELISA Kit is intended for the detection of total T3 in human serum or plasma. For research

More information

International Journal of Research and Review E-ISSN: ; P-ISSN:

International Journal of Research and Review   E-ISSN: ; P-ISSN: International Journal of Research and Review www.ijrrjournal.com E-ISSN: 2349-9788; P-ISSN: 2454-2237 Original Research Article Free Thyroid Hormones in Subclinical Hypothyroidism and Its Preponderance

More information

PDF hosted at the Radboud Repository of the Radboud University Nijmegen

PDF hosted at the Radboud Repository of the Radboud University Nijmegen PDF hosted at the Radboud Repository of the Radboud University Nijmegen The following full text is a publisher's version. For additional information about this publication click this link. http://hdl.handle.net/2066/50214

More information

FREE THYROXINE (f-t4) CHEMILUMINESCENCE IMMUNOASSAY KIT

FREE THYROXINE (f-t4) CHEMILUMINESCENCE IMMUNOASSAY KIT FREE THYROXINE (f-t4) CHEMILUMINESCENCE IMMUNOASSAY KIT Catalog No. CL1005-2 INTENDED USE The Autobio f-t4 CLIA test kit is intended for the quantitative determination of free thyroxine (f-t4) concentration

More information

Week 17 and 21 Comparing two assays and Measurement of Uncertainty Explain tools used to compare the performance of two assays, including

Week 17 and 21 Comparing two assays and Measurement of Uncertainty Explain tools used to compare the performance of two assays, including Week 17 and 21 Comparing two assays and Measurement of Uncertainty 2.4.1.4. Explain tools used to compare the performance of two assays, including 2.4.1.4.1. Linear regression 2.4.1.4.2. Bland-Altman plots

More information

Thyroid Function. Thyroid Antibodies. Analyte Information

Thyroid Function. Thyroid Antibodies. Analyte Information Thyroid Function Thyroid Antibodies Analyte Information - 1-2013-04-30 Thyroid Antibodies Determination of thyroid autoantibodies are, besides TSH and FT4, one of the most important diagnostic parameters.

More information

Keywords: albuminuria; albumin/creatinine ratio (ACR); measurements. Introduction

Keywords: albuminuria; albumin/creatinine ratio (ACR); measurements. Introduction Clin Chem Lab Med 2015; 53(11): 1737 1743 Beryl E. Jacobson, David W. Seccombe*, Alex Katayev and Adeera Levin A study examining the bias of albumin and albumin/creatinine ratio measurements in urine DOI

More information

Toward Standardization of Insulin Immunoassays

Toward Standardization of Insulin Immunoassays Clinical Chemistry 55:5 1011 1018 (2009) General Clinical Chemistry Toward Standardization of Insulin Immunoassays W. Greg Miller, 1* Linda M. Thienpont, 2 Katleen Van Uytfanghe, 2 Penelope M. Clark, 3

More information

Using Hashimoto thyroiditis as gold standard to determine the upper limit value of thyroid stimulating hormone in a Chinese cohort

Using Hashimoto thyroiditis as gold standard to determine the upper limit value of thyroid stimulating hormone in a Chinese cohort Li et al. BMC Endocrine Disorders (2016) 16:57 DOI 10.1186/s12902-016-0137-3 RESEARCH ARTICLE Open Access Using Hashimoto thyroiditis as gold standard to determine the upper limit value of thyroid stimulating

More information

Index. Graves disease, 111 thyroid autoantigens, 110 Autoimmune thyroiditis, 11, 58, 180, 181. B Bamforth Lazarus syndrome, 27

Index. Graves disease, 111 thyroid autoantigens, 110 Autoimmune thyroiditis, 11, 58, 180, 181. B Bamforth Lazarus syndrome, 27 Index A Adrenergic activation, 77 Allan Herndon Dudley syndrome, 31 Ambulatory practice choice of test, 156, 157 screening general population, thyroid dysfunction, 163, 164 targeted population, 164 167

More information

A Clinical Study on Patients Presenting with Thyroid Swelling and Its Correlation with TFT, USG, FNAC and Anti TPO Antibodies

A Clinical Study on Patients Presenting with Thyroid Swelling and Its Correlation with TFT, USG, FNAC and Anti TPO Antibodies A Clinical Study on Patients Presenting with Thyroid Swelling and Its Correlation with TFT, USG, FNAC and Anti TPO Antibodies 1* Hanushraj. R, 2 Sudharsan.S, 3 Balasubramaniyan. S, 4 Pradeep Kumar. M 1,4,

More information

Thyroid Management. Evolving Controversy - Science, Dogma, Opinion. The Ogden Surgical Medical Society May 2016

Thyroid Management. Evolving Controversy - Science, Dogma, Opinion. The Ogden Surgical Medical Society May 2016 Thyroid Management Evolving Controversy - Science, Dogma, Opinion The Ogden Surgical Medical Society May 2016 Published Guidelines AACE and ATA - Clinical Practice Guidelines in 2012 Guidelines are neither

More information

A STUDY ON PREVALANCE OF HYPOTHYROIDISM IN DIABETICS. 1.Assistant Professor,Dept of Gen. Medicine, National Institute of Medical Sciences(NIMS),Jaipur

A STUDY ON PREVALANCE OF HYPOTHYROIDISM IN DIABETICS. 1.Assistant Professor,Dept of Gen. Medicine, National Institute of Medical Sciences(NIMS),Jaipur Original research article International Journal of Medical Science and Education pissn- 2348 4438 eissn-2349-3208 Dr Asrar Ahmed 1* A STUDY ON PREVALANCE OF HYPOTHYROIDISM IN DIABETICS 1.Assistant Professor,Dept

More information

This supplementary material has been provided by the authors to give readers

This supplementary material has been provided by the authors to give readers Supplementary Online Content LI D, Radulescu A, Shrestha RT, et al. Association of biotin ingestion with performance of hormone and nonhormone assays in healthy adults. JAMA. doi:1.11/jama.217.1375 efigure

More information

EliKine Free Thyroxine (ft4) ELISA Kit

EliKine Free Thyroxine (ft4) ELISA Kit EliKine Free Thyroxine (ft4) ELISA Kit Booklet Item NO. KET0005 Product Name EliKine Free Thyroxine (ft4) ELISA Kit ATTENTION For laboratory research use only. Not for clinical or diagnostic use TABLE

More information

Intermethod Differences in Results for Total PSA, Free PSA, and Percentage of Free PSA

Intermethod Differences in Results for Total PSA, Free PSA, and Percentage of Free PSA Clinical Chemistry / PSA and Free PSA Method Differences Intermethod Differences in Results for Total PSA, Free PSA, and Percentage of Free PSA Patricia R. Slev, PhD, Sonia L. La ulu, and William L. Roberts,

More information

ACCESS hstni SCIENTIFIC LITERATURE

ACCESS hstni SCIENTIFIC LITERATURE ACCESS hstni SCIENTIFIC LITERATURE 2017 2018 Table of contents Performance Evaluation of Access hstni A critical evaluation of the Beckman Coulter Access hstni: Analytical performance, reference interval

More information

DISORDERS OF THE THYROID GLAND SIGNS, SYMPTOMS, & TREATMENT ENDOCRINE SYSTEM AT A GLANCE OBJECTIVES ANATOMY OF THE THYROID

DISORDERS OF THE THYROID GLAND SIGNS, SYMPTOMS, & TREATMENT ENDOCRINE SYSTEM AT A GLANCE OBJECTIVES ANATOMY OF THE THYROID OBJECTIVES DISORDERS OF THE THYROID GLAND SIGNS, SYMPTOMS, & TREATMENT Stephanie Blackburn, MHS, MLS(ASCP) CM LSU Health Shreveport Clinical Laboratory Science Program Discuss the synthesis and action

More information

Guidance for Thyroid Function Testing in Primary Care in Lothian

Guidance for Thyroid Function Testing in Primary Care in Lothian Guidance for Thyroid Function Testing in Primary Care in Lothian In July 2006 following a lengthy consultation process, a joint working group comprising representatives from the Association of Clinical

More information

Design Verification IMTEC-TSH R -A C Intended Use... 2 Diagnostic Sensitivity and Diagnostic Specificity... 2 Interferences... 4 Imprecision...

Design Verification IMTEC-TSH R -A C Intended Use... 2 Diagnostic Sensitivity and Diagnostic Specificity... 2 Interferences... 4 Imprecision... Design Verification IMTEC-TSH RECEPTOR-ANTIBODIES CONTENTS 1 Intended Use... 2 2 Diagnostic Sensitivity and Diagnostic Specificity... 2 2.1 Determination of Diagnostic Sensitivity and Diagnostic Specificity...

More information

MANY REPORTS SUGGEST an increasing incidence of

MANY REPORTS SUGGEST an increasing incidence of 21-972X/7/$1./ The Journal of Clinical Endocrinology & Metabolism 92(12):47 482 Printed in U.S.A. Copyright 27 by The Endocrine Society doi: 1.121/jc.27-1499 -Specific Distribution of Serum Thyrotropin

More information

LABORATORY TESTS FOR EVALUATION OF THYROID DISORDERS

LABORATORY TESTS FOR EVALUATION OF THYROID DISORDERS LABORATORY TESTS FOR EVALUATION OF THYROID DISORDERS Maryam Tohidi Anatomical & clinical pathologist Research Institute for Endocrine Sciences THYROID GLAND (15-25 gr), (12-20 gr), 2 lobes connected by

More information

Thyroid disease is one of the most common endocrine disorders (1). The

Thyroid disease is one of the most common endocrine disorders (1). The Original Article Performance characteristics of three automated immunoassays for thyroid hormones Faranak Kazerouni (PhD) 1 Houshang Amirrasouli (PhD) *1 1- Department of Laboratory Medicine, Faculty of

More information

liquicolor (AMP Buffer, IFCC) Design Verification

liquicolor (AMP Buffer, IFCC) Design Verification Design Verification (AMP Buffer, IFCC) liquicolor 1 Introduction... 2 2 Imprecision... 2 3 arity and Detection Limit... 2 3.1 arity... 2 3.2 Detection Limit... 3 4 Comparison of Methods... 3 5 Stability...

More information

Thyroid Function. Thyroglobulin Analyte Information

Thyroid Function. Thyroglobulin Analyte Information Thyroid Function Thyroglobulin Analyte Information - 1-2011-01-11 Thyroglobulin Introduction Thyroglobulin (Tg) is a big dimeric protein consisting of two identical subunits. It has 2,748 amino acids in

More information

Comparison between two automated chemiluminescence immunoassays for quantifying 25 (OH) vitamin D

Comparison between two automated chemiluminescence immunoassays for quantifying 25 (OH) vitamin D 7 ORIGINALS / Rev Osteoporos Metab Miner. 216;8(2):7-74 Torrubia B 1, Alonso I 1, López-Ramiro E 1, Mahillo I 2, De la Piedra C 1 1 Laboratorio de Bioquímica 2 Servicio de Epidemiología Fundación Jiménez

More information

The Relationship Between TSH and Free T 4 in a Large Population Is Complex and Nonlinear and Differs by Age and Sex

The Relationship Between TSH and Free T 4 in a Large Population Is Complex and Nonlinear and Differs by Age and Sex ORIGINAL Endocrine ARTICLE Research The Relationship Between TSH and Free T 4 in a Large Population Is Complex and Nonlinear and Differs by Age and Sex Narelle C. Hadlow, Karen M. Rothacker, Robert Wardrop,

More information

FREE TRIIODOTHYRONINE (F-T3) CHEMILUMINESCENCE

FREE TRIIODOTHYRONINE (F-T3) CHEMILUMINESCENCE FREE TRIIODOTHYRONINE (F-T3) CHEMILUMINESCENCE IMMUNOASSAY KIT Catalog No. CL1004-2 INTENDED USE The Autobio f-t3 CLIA test kit is intended for the quantitative determination of free triiodothyronine (f-t3)

More information

DBC 25-Hydroxyvitamin D

DBC 25-Hydroxyvitamin D D I A G N O S T I C S B I O C H E M C A N A D A DBC 25-Hydroxyvitamin D ELISA OVERVIEW The worldwide aging and chronically ill population is increasing rapidly. It is forecasted that the Global Vitamin

More information

Stability of thyroxine and triiodothyronine in

Stability of thyroxine and triiodothyronine in Stability of thyroxine and triiodothyronine in biological fluids LYNN NYE, T. H. YEO, VIVIAN CHAN1, D. GOLDIE2, AND J. From the Department of Chemical Pathology, St Bartholomew's Hospital, London EC] J.

More information

Grave s disease (1 0 )

Grave s disease (1 0 ) THYROID DYSFUNCTION Grave s disease (1 0 ) Autoimmune - activating AB s to TSH receptor High concentrations of circulating thyroid hormones Weight loss, tachycardia, tiredness Diffuse goitre - TSH stimulating

More information

Immunoprotective Steroids and SHBG in Non-Treated Hypothyroidism and their Relationship to Autoimmune Thyroid Disorders

Immunoprotective Steroids and SHBG in Non-Treated Hypothyroidism and their Relationship to Autoimmune Thyroid Disorders Physiol. Res. 57 (Suppl. 1): S119-S125, 2008 Immunoprotective Steroids and SHBG in Non-Treated Hypothyroidism and their Relationship to Autoimmune Thyroid Disorders K. DRBALOVÁ, P. MATUCHA, M. MATĚJKOVÁ-BĚHANOVÁ,

More information

See external label 96 tests ULTRASENSITIVE THYROID STIMULATING HORMONE (u-tsh) TSH Ultra Sensitive

See external label 96 tests ULTRASENSITIVE THYROID STIMULATING HORMONE (u-tsh) TSH Ultra Sensitive DIAGNOSTIC AUTOMATION, INC. 21250 Califa Street, Suite 102 and 116, Woodland Hills, CA 91367 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com

More information

The Thr92Ala polymorphism in the type 2 deiodinase is not associated with thyroxin dose in athyroid patients or patients with Hashimoto thyroiditis

The Thr92Ala polymorphism in the type 2 deiodinase is not associated with thyroxin dose in athyroid patients or patients with Hashimoto thyroiditis 12 The Thr92Ala polymorphism in the type 2 deiodinase is not associated with thyroxin dose in athyroid patients or patients with Hashimoto thyroiditis K.A. Heemstra, H.C. Hoftijzer, W.M. van der Deure,

More information

Endogenous antibody interference in immunoassays. Ellen Anckaert, M.D., Ph.D. Dienst Klinische Chemie en Radio-immunologie UZ Brussel

Endogenous antibody interference in immunoassays. Ellen Anckaert, M.D., Ph.D. Dienst Klinische Chemie en Radio-immunologie UZ Brussel Endogenous antibody interference in immunoassays Ellen Anckaert, M.D., Ph.D. Dienst Klinische Chemie en Radio-immunologie UZ Brussel To help protect your privacy, PowerPoint has blocked automatic download

More information

THYROID FUNCTION EVALUATION IN PATIENTS WITH INCREASED OR DECREASED THYROXINE-BINDING PROTEIN

THYROID FUNCTION EVALUATION IN PATIENTS WITH INCREASED OR DECREASED THYROXINE-BINDING PROTEIN THE AMEBICAN JOURNAL OF CLINICAL PATHOLOGY Vol. 50, No. 3 Copyright 1968 by The Williams & Wilkins Co. Printed in U.S.A. THYROID FUNCTION EVALUATION IN PATIENTS WITH INCREASED OR DECREASED THYROXINE-BINDING

More information

Human Thyroid-Peroxidase Antibody, TPO-Ab ELISA Kit

Human Thyroid-Peroxidase Antibody, TPO-Ab ELISA Kit Human Thyroid-Peroxidase Antibody, TPO-Ab ELISA Kit Catalog No: E0442h 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH

More information

Instructions for Use. Tg Antibody ELISA

Instructions for Use. Tg Antibody ELISA Instructions for Use Tg Antibody ELISA Enzyme Immunoassay for the Quantitative Determination of Autoantibodies to Thyroglobulin (Tg) in Serum and Plasma I V D REF EA618/96 12 x 8 2 8 C DLD Gesellschaft

More information

25 OH Vitamin D Rapid Test

25 OH Vitamin D Rapid Test INSTRUCTION FOR USE REF:GDB 7120-25T 25 OH Vitamin D Rapid Test A Rapid Sandwich Immunochromatographic Test for Quantitative Detection of total 25-OH Vitamin D in human finger-prick blood For In Vitro

More information

Establishing Reference Intervals for Clinical Laboratory Test Results Is There a Better Way?

Establishing Reference Intervals for Clinical Laboratory Test Results Is There a Better Way? Clinical Chemistry / Estimating Reference Intervals Establishing Reference Intervals for Clinical Laboratory Test Results Is There a Better Way? Alex Katayev, MD, 1 Claudiu Balciza, 2 and David W. Seccombe,

More information

Thyroid Peroxidase IgG ELISA

Thyroid Peroxidase IgG ELISA For Research Use Only. Not for use in Diagnostic Procedures. INTENDED USE The GenWay, Inc. Thyroid Peroxidase (TPO) IgG Kit is intended for the detection of IgG antibody to Thyroid Peroxidase in human

More information

Standard Operating Procedure

Standard Operating Procedure Subject Free Thyroxine (FT4-II) cobas e411 Index Number Lab-1594 Section Regional/Affiliate Subsection Laboratory Category Departmental Contact Goplin, Darcy Last Revised 2/2/2017 References Required document

More information

Human Thyroid Stimulating Hormone CLIA kit

Human Thyroid Stimulating Hormone CLIA kit Human Thyroid Stimulating Hormone CLIA kit Cat. No.:DEEL0223 Pkg.Size:96 tests Intended use For the direct quantitative determination of Thyroid Stimulating Hormone in human serum by chemiluminescence

More information

A RARE CAUSE OF DISCORDANT TSH: MACRO-TSH

A RARE CAUSE OF DISCORDANT TSH: MACRO-TSH doi: 10.4183/aeb.2015.170 general endocrinology A RARE CAUSE OF DISCORDANT TSH: MACRO-TSH M.M. Yalcin 1,*, G. Ayvaz 1, O. Gulbahar 2, F. Toruner 1, C. Ozkan 1, A.E. Altinova 1, M. Akturk 1, M. Arslan 1

More information

SCIEX Vitamin D 200M Assay for the Topaz System

SCIEX Vitamin D 200M Assay for the Topaz System The First FDA-Cleared LC-MS/MS Assay for Vitamin D SCIEX Vitamin D 200M Assay for the Topaz System The first FDA-cleared LC-MS/MS assay for Vitamin D Vitamin D is an important building block for human

More information