Dosimetry and QA of proton and heavier ion beams
|
|
- Damian Webster
- 6 years ago
- Views:
Transcription
1 Dosimetry and QA of proton and heavier ion beams Stanislav Vatnitskiy EBG MedAustron GmbH Wiener Neustadt, Austria
2 Content Introduction Reference dosimetry Methods, detectors, protocols Dosimetry in non-reference conditions and QA measurements Methods, detectors Recommendations
3 about 25 treatment facilities are established another 20 proton and light-ion beam centers are planned to be open in next 5 years
4 Dosimetry tasks in proton and heavier ion beam radiotherapy Acceptance testing and commissioning of delivery beam lines Reference calibration of clinical beams Commissioning of treatment planning systems Periodic QA checks Verification of dose delivery
5 Ensure exact delivery of prescribed dose Consistent and harmonized dosimetry guidelines Accurate beam calibration Perform planning of high-precision conformal therapy Provide interchange of clinical experience and treatment protocols between facilities Provide standardization of dosimetry in radiobiology experiments
6 Reference dosimetry Basic output calibration of a clinical beam, is a direct determination of absorbed dose per MU in water under specific reference conditions Reference dosimetry techniques Faraday Cup Calorimetry Ionization chamber
7 Absorbed dose determination in reference conditions for proton and heavier ion beams Faraday Cup Calorimeter Lack of national and international dosimetry standards Thimble air-filled ionization chamber
8 FC-based calibration of the ionization chamber D w = (N/A) (S/ρ) w * x N : number of protons per MU collected in the FC A : effective area of the beam (S/ρ) w : stopping power for water at the incident proton energy
9 N Dw( Q, FC) = ( S/ ) A FC-based calibration of the ionization chamber 10 D ρ ( w QFC, ) w NDwFC,, = corr M FC IC
10 1995 PTCOG Dosimetry intercomparison Institution s statement of proton dose ratio to the mean Institution FC calibration
11 Calorimetry-based calibration of the ionization chamber Main characteristics of calorimetry dosimetry: Energy imparted to matter by radiation causes an increase in temperature T. Dose absorbed in the sensitive volume is proportional to T. T is measured with thermocouples or thermistors. Calorimetric dosimetry is the most precise of all absolute dosimetry techniques.
12 Calorimetry-based calibration of the ionization chamber Two types of absorbed dose calorimeter are used in charged particle beams: In graphite calorimeters the average temperature rise is measured in a graphite body that is thermally insulated from surrounding bodies (jackets) by evacuated vacuum gaps.
13 Calorimetry-based calibration of the ionization chamber In sealed water calorimeters low thermal diffusivity of water enables the temperature rise to be measured directly at a point in continuous water.
14 Calorimetry-based calibration of the ionization chamber D ( Q, cal) = c Θ V (1 + D ) w T
15 Calorimetry-based calibration of the ionization chamber Dw( Qcal, ) = c Θ V (1 + DT),, N Dwc = D (, ) w Qcal corr M
16 Water calorimetery in a 250 MeV proton beam at LLUMC Water Calorimeter
17 Ion chamber measurements (dummy calorimeter) in a 250 MeV proton beam Dummy calorimeter with IC
18 Graphite calorimetry protons at CCO carbon ions at NIRS Sakama et al 2008 Palmans et al 2004
19 Absorbed dose determination in reference conditions for light ion beams Faraday Cup Calorimeter Lack of national and international dosimetry standards Thimble air-filled ionization chamber
20 Protocols/Code of Practice for proton and heavier ion beam dosimetry Only a Protocol based on standards of absorbed dose to water is being recommended by ICRU/IAEA Reports for protons and heavier ions
21 N D,w - based formalism - IAEA TRS-398 D w (z ref ) at any user quality Q (photons, electrons, protons, heavier ions) D w,q = M Q N D,w,Q o k Q,Qo corrected instrument reading at Q calibration coefficient at Q o beam quality factor
22 Ionization chambers Both cylindrical and parallel plate chambers are recommended for reference dosimetry Parallel plate chambers yield higher u c in absolute D w, though are better suited for relative dosimetry cylindrical chambers recommended for SOBP width>2 cm ICRU/IAEA Report 78: dosimetry equipment For SOBP widths < 2 cm parallel plate chambers must be used Versatile electrometer with cable and connectors fitting to the electrometer and all chambers thermometer, barometer
23 Water phantoms for reference dosimetry Water is the only standard and most universal phantom material for proton and heavier ions dosimetry measurements
24 Reference calibration of proton and heavier ion beams: reference conditions Passive Scattering protons, carbon ions Calibration at SOBP Goitein, Lomax and Pedroni, 2002 Protons spot scanning Carbon ions spot scanning Calibration at SOBP Calibration at plateau Plateau versus SOBP: superposition of beams with different intensities not continuous and reproducible mix of particles with high and low LET fluence corrections are small
25 Reference geometry passive beam delivery relative dose relative dose off-axis distance, cm depth, cm phantom protons ion chamber
26 Proton spot scanning calibration at PSI dose model predicts the number of protons/gy 10x10x10 cm 3 box is filled in with homogeneous dose distribution to a dose of 1Gy IC 10x10x10 ( Pedroni et al) ion chamber is placed at a residual range 5g/cm 2
27 A solid phantom rather than a water phantom is used in scanned ion beam A field of 5 cm x 5 cm rather than 10 cm x 10 cm is used The reference depth is in the plateau region of a monoenergetic Bragg peak (depth of 7 mm) instead of the center of a SOBP The calibration is dependent on the initial particles energy and several energy points are selected
28 ( ) ( s ) Lack of standards => Q o = 60 Co This approach would ultimately lead to a dosimetry system, where the dose applied to a patient is traceable to the dosimetry standards of the national PSDL. k Q = s ( ) wair, W p wair, 60 Q air Q Q Co W p = p p p p ( ) air Q dis wall cav cel 1 for protons 1 for heavier ions 1 for 60 Co Co p Co
29 Stopping powers for proton beams Basic proton stopping powers from ICRU 49 Calculation using MC code PETRA following Spencer- Attix cavity theory Transport included secondary electrons and nuclear inelastic process
30 1.140 water/air stopping-power ratio alpha particles carbon ions protons C (Salamon) Ne (Salamon) Ar (Salamon) He (Salamon) p (ICRU-49) He (ICRU-49) C (H and B) residual range (g cm -2 ) Ratio of stopping powers water/air for heavy ions calculated using the computer codes developed by Salamon (for C, Ne, Ar and He) and by Hiraoka and Bichsel (for C). Data for protons and He from ICRU 49. A constant value of s w,air = 1.13 adopted in TRS 398 (ignores fragments)
31 Geitner et al 2006
32 Values of w/e for protons and carbon ions deduced from comparison of ionization chamber and calorimeter measurements (w air/e )p / (J C -1 ) Proton beam 34.2 J C -1 Delacroix et al., 1997 Palmans et al., 2004 Palmans et al., 1996 Hashemian et al., 2003 Siebers et al., 1995 Schulz et al., 1992 Brede et al., 2006 Medin et al., J C J C -1 TRS Proton energy / MeV ICRU 78 Sakama et al 2008
33 Proton perturbation factors Palmans et al. 2000
34 intra-track one single track inter-tracks multiple tracks dose or dose-rate independent dose or dose-rate dependent Courtesy of P. Andreo The user should verify recombination corrections against independent method
35 Initial recombination in pp chamber in different LET beams Kanai t. et al., Phys. Med. Biol. (1998) 43,
36 Issues to be resolved in upcoming ICRU report on light-ion beams The proposed approach in upcoming ICRU Report on light-ion beams would ultimately lead to a dosimetry system, where the dose applied to a patient is traceable to the dosimetry standards of the national PSDL. TRS 398 may be adopted for light-ion beam dosimetry with beamline specific adjustments The currently recommended values of s w,air (and W air ) for absolute dosimetry should be re-considered Uncertainties in stopping powers, including those of the I-values for different tissues (5-10%), must be taken into account to reestimate what precision is really achievable in clinical practice
37 Standard uncertainties in D w (TRS 398, ICRU 78) u(n D,w SSDL ) = 0.6 k Q calc Co-60 gamma-rays 0.9 High-energy photons 1.5 High-energy electrons Proton beams Heavier ions
38 Dosimetry in non-reference conditions Relative dose measurements require no detector calibration other than verification of linearity of response within assumed dynamic range of measurement conditions Dosimetry tasks Routine daily clinical physics activity Beam line commissioning Collecting data for TPS Periodic QA Beam characteristics Depth dose Lateral profiles Output factors
39 Detectors for measurements in non-reference conditions Active detectors: Ion chambers, diodes, diamond detector, scintillators (single and multiple) Direct display of the current dose rate or the accumulating dose Passive detectors: Destructive TLD Non-destructive Films, alanine Probe accumulates the dose during irradiation. The value of dose is obtained after irradiation with read-out device
40 Detectors for measurements in non-reference conditions Passive detectors Active detectors
41 Diodes for characterization of small beams Because of their small size silicon diodes are convenient for profile measurements and characterization of small proton beams. Ionization chamber Diode The response of diodes must always be checked against ionometric measurements before use.
42 Diodes for characterization of small proton beams Newhauser et al, 2003
43 1.1 Characterization of small proton beams 126 MeV protons, collimator 30 mm, no modulation Markus chamber Diode DEB 50 Diamond detector PinPoint chamber Relative dose Depth in water, mm
44 Characterization of scanned proton pencil beams Depth dose distribution Beam monitor calibration Gillin et al, 2010
45 Multi-detector systems for characterization and QA of proton and carbon beams courtesy by PTW Cirio et al, 2004 Nichiporov et al courtesy by IBA
46 Alanine ESR proton dosimetry Advantages: Signal is preserved after readout Exhibits long term stability Near tissue equivalent Linear dose response Disadvantages: Low sensitivity Reduction of Bragg Peak ESR technique is quite elaborous Palmans, 2003
47 2-D dosimetry and QA with fluorescent screen and CCD camera Boon et al, 2000
48 2-D dosimetry and QA with fluorescent screen and CCD camera Courtesy of J. Heese E. Pedroni et al 2005
49 2-D dosimetry and QA with fluorescent screen and CCD camera Rosenthal et al, 2004
50 TLD film for 2-D characterization of clinical proton beams Olko et al, 2004
51 Radiochromic film for characterization of clinical proton beams 1.0 MD- 55 film Kodak XV-2 film 0.8 A. Entrance B. SOBP region relative dose off-axis data, mm
52 Radiochromic film for characterization of clinical proton beams 155 MeV, modulation 3 cm, normalized at 11 cm parallel plate ionization chamber MD-55 film, 70 Gy relative dose water equivalent depth, cm
53 3-D Gel dosimetry for characterization of clinical proton beams Magic Gel Heufelder et al, 2003
54 3-D Gel dosimetry for characterization of clinical proton beams Heufelder et al, 2003
55 Current status of proton and heavier ion beam dosimetry Implementation of ICRU Report 78 IAEA TRS 398 harmonize clinical dosimetry at proton and heavier ion beam facilities provide a level of accuracy comparable to that in calibration of photon and electron beams
56 Current status of proton and light-ion beam dosimetry: PROTONS Passive beam delivery Active beam delivery ICRU Report 78 IAEA TRS 398 (center of SOBP) CARBON IONS Active beam delivery IAEA TRS 398 (Plateau or center SOPB) PROTONS Stereotactic radiosurgery MC
57 Acknowledgements Members of ICRU/IAEA Committee Report 78 O. Jäkel and colleagues at DKFZ E. Pedroni and colleagues at PSI H. Palmans
Recent advances in dosimetry in reference conditions for proton and light-ion beams
Recent advances in dosimetry in reference conditions for proton and light-ion beams S. Vatnitskiy a), P. Andreo b) and D.T.L. Jones c) a) MedAustron, Wiener Neustadt, Austria b) Medical Radiation Physics,
More informationTopics covered 7/21/2014. Radiation Dosimetry for Proton Therapy
Radiation Dosimetry for Proton Therapy Narayan Sahoo Department of Radiation Physics University of Texas MD Anderson Cancer Center Proton Therapy Center Houston, USA Topics covered Detectors used for to
More informationActive Scanning Beam 3 Checking Delivery/Dosimetry
Active Scanning Beam 3 Checking Delivery/Dosimetry C Algranati, J Salk, A Coray, A Lomax, E Pedroni, T Boeringer, S Lin, E Hug Center for Proton Radiation Therapy Checking Delivery/Dosimetry Contents Absolute
More informationStandard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties
Standard calibration of ionization chambers used in radiation therapy dosimetry and evaluation of uncertainties A. Solimanian and M. Ghafoori * Iran. J. Radiat. Res., 2010; 8 (3): 195-199 Radiation Dosimetry
More informationPractical Reference Dosimetry Course April 2015 PRDC Program, at a glance. Version 1.0. Day 1 Day 2 Day 3 Day 4
Practical Reference Dosimetry Course 21-24 April 2015 PRDC 2015 Program, at a glance Version 1.0 Day 1 Day 2 Day 3 Day 4 Quantities and Units Free air chambers Uncertainties Brachytherapy traceability
More informationTwo-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance
Two-Dimensional Thermoluminescence Dosimetry System for Proton Beam Quality Assurance Jan Gajewski Institute of Nuclear Physics, Kraków, Poland German Cancer Research Center, Heidelberg, Germany Existing
More informationABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING. 5 th June 2007
ABSTRACTS FOR RADIOTHERAPY STANDARDS USERS MEETING 5 th June 2007 An Overview of Radiotherapy Dosimetry at the NPL Hugo Palmans In relation to radiotherapy applications, The National Physical Laboratory
More informationPMP. Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers in Low-Energy Electron Dosimetry. Original Article.
Original Article PMP Progress in Medical Physics 29(1), March 218 https://doi.org/1.14316/pmp.218.29.1.16 pissn 258-4445, eissn 258-4453 Use of Cylindrical Chambers as Substitutes for Parallel- Plate Chambers
More informationPeak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam
Peak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam University of Chicago CDH Proton Center LET study C. Reft 1, H. Ramirez 2 and M. Pankuch
More informationCOMPARISON OF RADIOBIOLOGICAL EFFECTS OF CARBON IONS TO PROTONS ON A RESISTANT HUMAN MELANOMA CELL LINE
COMPARISON OF RADIOBIOLOGICAL EFFECTS OF CARBON IONS TO PROTONS ON A RESISTANT HUMAN MELANOMA CELL LINE I. Petrovi a, A. Risti -Fira a, L. Kori anac a, J. Požega a, F. Di Rosa b, P. Cirrone b and G. Cuttone
More informationTreatment Planning (Protons vs. Photons)
Treatment Planning Treatment Planning (Protons vs. Photons) Acquisition of imaging data Delineation of regions of interest Selection of beam directions Dose calculation Optimization of the plan Hounsfield
More informationDosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy
Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti a*, M. Bucciolini a, L. Marrazzo a, D. Menichelli a. a) Department
More informationCalibration of dosimeters for small mega voltage photon fields at ARPANSA
Calibration of dosimeters for small mega voltage photon fields at ARPANSA G Ramanathan 1, C.Oliver 1, D J Butler 1, P D Harty 1, Viliami Takau 1 Tracy Wright 1 and Tom Kupfer 2 1 Australian Radiation Protection
More informationMania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane
Mania Aspradakis John Byrne Hugo Palmans John Conway Jim Warrington Karen Rosser Simon Duane Why are we concerned with small MV photon fields? SRS dosimetry Total scatter factor with various detectors
More informationMeasurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients
Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Moyed Miften, PhD Professor and Chief Physicist University of Colorado Chester Reft, PhD Associate Professor University of Chicago
More informationProtons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry J. Phys.: Conf. Ser.
Protons Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry T Gorjiara, Z Kuncic, J Adamovics and C Baldock 2013 J. Phys.: Conf. Ser. 444 012090 PRESAGE is a radiochromic
More informationDOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS
DOSIMETRIC COMPARISION FOR RADIATION QUALITY IN HIGH ENERGY PHOTON BEAMS EUGENIA BADITA 1, CATALIN VANCEA 1,3, ION CALINA 1,3, DANIELA STROE 2, MIHAELA DUMITRACHE 2,3, MIRABELA DUMITRACHE 1,3 1 National
More informationPREDICTION OF ABSORBED DOSE DISTRIBUTIONS AND NEUTRON DOSE EQUIVALENT VALUES IN PROTON BEAM RADIATION THERAPY
PREDICTION OF ABSORBED DOSE DISTRIBUTIONS AND NEUTRON DOSE EQUIVALENT VALUES IN PROTON BEAM RADIATION THERAPY IDENTIFICATION NUMBER: ANS-RT-PROTON-01 BENCHMARK CLASSIFICATION: Radiation Therapy BENCHMARK
More informationDosimetry. Dr. Manfred Sassowsky Cantonal Hospital Lucerne (KSL) Institute for Radio-Oncology
Kantonsspital Luzern Dosimetry Dr. Manfred Sassowsky Cantonal Hospital Lucerne (KSL) Institute for Radio-Oncology 3.9.2007 Introduction Metrological traceability; the dosimetry chain Absolute / relative
More informationCharacterization and implementation of Pencil Beam Scanning proton therapy techniques: from spot scanning to continuous scanning
Characterization and implementation of Pencil Beam Scanning proton therapy techniques: from spot scanning to continuous scanning Supervisors Prof. V. Patera PhD R. Van Roermund Candidate Annalisa Patriarca
More informationHow accurately can we measure dose clinically? Accurate clinical dosimetry. Dosimetric accuracy requirement
Accurate clinical dosimetry Part I: Fundamentals and New developments in reference dosimetry Jan Seuntjens McGill University Montreal, Canada, H3G 1A4 jseuntjens@medphys.mcgill.ca Dosimetric accuracy requirement
More informationEDITORIAL NOTE. factor, which has been used instead of the chamber factor N D
IAEA/WHO NETWORK OF SECONDARY STANDARD DOSIMETRY LABORATORIES Prepared by the Joint IAEA/WHO Secretariat of the SSDL Network Published semi-annually by the International Atomic Energy Agency Vienna No.
More informationRadiation qualities in carbon-ion radiotherapy at NIRS/HIMAC
Radiation qualities in carbon-ion radiotherapy at NIRS/ Shunsuke YONAI Radiological Protection Section Research Center for Charged Particle Therapy National Institute of Radiological Sciences (NIRS) E-mail:
More informationDisclosure. Outline. Machine Overview. I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past.
Clinical Implementation of the CyberKnife Disclosure I have received honoraria from Accuray in the past. I have had travel expenses paid by Accuray in the past. Mary Ellen Masterson-McGary McGary CyberKnife
More informationApplications of Monte Carlo simulations to radiation dosimetry
Applications of Monte Carlo simulations to radiation dosimetry D.W.O. Rogers Carleton Laboratory for Radiotherapy Physics. Physics Dept, Carleton University, Ottawa http://www.physics.carleton.ca/~drogers
More informationSmall field diode dosimetry
Small field diode dosimetry Parham Alaei, Ph.D. Department of Radiation Oncology University of Minnesota NCCAAPM Symposium-October 10, 2013 1 Diodes as beam data collection detectors Diodes as in vivo
More informationDetector comparison for small field output factor measurements with flattening filter free photon beams
Detector comparison for small field output factor measurements with flattening filter free photon beams 1 Wolfgang Lechner, 1 Lukas Sölkner, 2 Paulina Grochowska, 1 Dietmar Georg 1 Christian Doppler Laboratory
More informationReport to the 18 th Meeting of the CCRI(I), May 2007 Recent Activities in Measurement Standards and Dosimetry at METAS,
Report to the 18 th Meeting of the CCRI(I), May 2007 Recent Activities in Measurement Standards and Dosimetry at, 2005-2007 1. Introduction Gerhard Stucki, Solange Gagnebin, Damian Twerenbold, Sándor Vörös
More informationAbsolute Dosimetry. Versatile solid and water phantoms. Introduction Introduction Introduction Introtro Intro Intro
In Phantoms Vivo treatment for Absolute Dosimetry verification Versatile solid and water phantoms Introduction Introduction Introduction Introtro Intro Intro The rapid development of advanced treatment
More informationD DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology
Journal of Life Sciences 9 (2015) 120-126 doi: 10.17265/1934-7391/2015.03.005 D DAVID PUBLISHING Uncertainties of in vivo Dosimetry Using Semiconductors Zeina Al Kattar, Hanna El Balaa and Saeed Zahran
More informationMuch of what we will say
New Dosimetry Measurements - New Developments J. Seuntjens J. Gibbons I. Das WA-A-211A Continuing Education Session Much o what we will say Clinical Dosimetry Measurements in Radiotherapy (AAPM 2009 Summer
More informationDevelopment of Graphite Calorimetry at the NPL for Proton Beam Therapy
Development of Graphite Calorimetry at the NPL for Proton Beam Therapy Russell A S Thomas Senior Research & Clinical Scientist Radiation Dosimetry Group 1 Background Current protocols based on chambers
More informationLimits of Precision and Accuracy of Radiation Delivery Systems
Limits of Precision and Accuracy of Radiation Delivery Systems Jean M. Moran, Ph.D. 1 and Timothy Ritter, Ph.D. 2 1 University of Michigan, Ann Arbor, Michigan 2 Ann Arbor Veterans Affairs Hospital, Ann
More informationDosimetric Uncertainties in Reference and Relative Dosimetry of Small Fields
Dosimetric Uncertainties in Reerence and Relative Dosimetry o Small Fields Jan Seuntjens, Ph.D., FAAPM, FCCPM McGill University Health Centre Canada Outline o Presentation Uncertainty concepts and requirements
More informationIon recombination corrections of ionization chambers in flattening filter-free photon radiation
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 13, NUMBER 5, 2012 Ion recombination corrections of ionization chambers in flattening filter-free photon radiation Yuenan Wang, a Stephen B. Easterling,
More informationTraceability and absorbed dose standards for small fields, IMRT and helical tomotherapy
Traceability and absorbed dose standards for small fields, IMRT and helical tomotherapy Simon Duane, Hugo Palmans, Peter Sharpe NPL, UK Stefaan Vynckier UCL, Brussels, Belgium LNE-LNHB / BIPM workshop,
More informationDosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy
Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti M. Bruzzi,M. Bucciolini, L. Marrazzo, D. Menichelli University of
More informationCalibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia
Abstract Calibration of Radiation Instruments Used in Radiation Protection and Radiotherapy in Malaysia Taiman Bin Kadni (taiman@mint.gov.my) Secondary Standard Dosimetry Laboratory (SSDL) Malaysian Institute
More informationLearning Objectives. Clinically operating proton therapy facilities. Overview of Quality Assurance in Proton Therapy. Omar Zeidan
Overview of Quality Assurance in Proton Therapy Omar Zeidan AAPM 2012 Charlotte, NC July 30 st, 2012 Learning Objectives Understand proton beam dosimetry characteristics and compare them to photon beams
More informationUniversità degli Studi di Napoli Federico II. Dipartimento di Fisica "Ettore Pancini"
Università degli Studi di Napoli Federico II Scuola Politecnica e delle Scienze di Base Collegio di Scienze Dipartimento di Fisica "Ettore Pancini" Corso di Laurea Magistrale in Fisica TESI DI LAUREA SPERIMENTALE
More informationIntroduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.
Measurement of Secondary Radiation for Electron and Proton Accelerators D. Followill, Ph.D. Radiological Physics Center U. T. M. D. Anderson Cancer Center Introduction Patients undergoing radiation therapy
More informationIntensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies
Intensity Modulated Radiation Therapy: Dosimetric Aspects & Commissioning Strategies ICPT School on Medical Physics for Radiation Therapy Justus Adamson PhD Assistant Professor Department of Radiation
More informationApplication(s) of Alanine
Application(s) of Alanine Simon Duane Radiotherapy Standards User Group, 5 June 2007 Outline Alanine/EPR dosimetry characteristics, usage (dis)advantages for reference dosimetry Traceable dosimetry for
More informationSTEREOTACTIC DOSE VERIFICATION PHANTOM VERSATILE STEREOTACTIC QA PHANTOMS
PHANTOMS VERSATILE STEREOTACTIC QA For fast and accurate commissioning of Accuray CyberKnife treatment systems and patient specific dose verification plans STEREOTACTIC DOSE VERIFICATION PHANTOM Stereotactic
More informationA preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF
Nuclear Science and Techniques 24 (2013) 060102 A preliminary clinic dosimetry study for synchrotron radiation therapy at SSRF LI Zhaobin SHI Zeliang ZHANG Qing WANG Yong FU Shen * The 6 th People s Hospital
More informationMEDICAL PHYSICS ASPECTS OF PARTICLE THERAPY Oliver Jäkel 1,2, *
Radiation Protection Dosimetry (2009), Vol. 137, No. 1 2, pp. 156 166 Advance Access publication 14 October 2009 doi:10.1093/rpd/ncp192 MEDICAL PHYSICS ASPECTS OF PARTICLE THERAPY Oliver Jäkel 1,2, * 1
More informationDosimetry Standards and Dissemination Systems for Radiation Processing in China
Dosimetry Standards and Dissemination Systems for Radiation Processing in China Zhang Yanli National Institute of Metrology, China April 2017 in Vienna, Austria Organization Founded in 1955, National Institute
More informationM. J. Maryanski, Three Dimensional BANG Polymer Gel Dosimeters AAPM'99, CE Course
Three Dimensional BANG Polymer Gel Dosimeters Marek J. Maryanski MGS Research, Inc. Guilford, CT Educational objectives: Describe the need for high-resolution 3D dosimetry in 3D CRT. Explain the physics
More informationSemiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry
DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector
More informationDependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose
Submitted to Radiation Measurements Dependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose P. Bilski 1, M. Sadel 1, J. Swakon 1, A. Weber 2 1
More informationProcess Control Methods in Radiation Technologies
Process Control Methods in Radiation Technologies András Kovács Nuclear Security Department Centre for Energy Research Hungarian Academy of Sciences Radiation Processing Applications Radiation sterilization:
More informationAssessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine
2017 IJSRST Volume 3 Issue 3 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine Samuel Nii Adu Tagoe
More informationSCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z540-1-1994 UNIVERSITY OF WISCONSIN RADIATION CALIBRATION LABORATORY Room B1002, WIMR 1111 Highland Avenue Madison, WI 53705-2275 Larry A. DeWerd,
More informationSmall Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy)
Small Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) IJDas (1) Indra J. Das, PhD, FAAPM, FACR, FASTRO Department of Radiation Oncology Indiana University School of
More informationKaren E. Christaki and Ahmed Meghzifene Dosimetry and Medical Radiation Physics Section, International Atomic Energy Agency, A-1400 Vienna, Austria
Dosimetry of small static fields used in external photon beam radiotherapy: Summary of TRS-483, the IAEA AAPM international Code of Practice for reference and relative dose determination Hugo Palmans a)
More informationDevelopment and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba.
Development and prospects on dosimetry at radiotherapy levels in the Secondary Standard Dosimetry Laboratory of Cuba. Gonzalo Walwyn Salas, Stefan Gutiérrez Lores Center for Radiation Protection and Hygiene,
More informationS. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned
S. Derreumaux (IRSN) Accidents in radiation therapy in France: causes, consequences and lessons learned MEDICAL LINEAR ACCELERATORS Electron beam (MeV) Photon beam (MV) PRECISION REQUIRED IN RADIOTHERAPY
More informationDOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS
Romanian Reports in Physics, Vol. 63, No. 3, P. 700 706, 2011 DOSE MEASUREMENTS IN TELETHERAPY USING THERMOLUMINESCENT DOSIMETERS ZOE GHITULESCU 1, ANA STOCHIOIU 2, MIHAI DUMITRACHE 3, 1 CNCAN- National
More informationA comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV
A comparison of dose distributions measured with two types of radiochromic film dosimeter MD55 and EBT for proton beam of energy 175 MeV M. Mumot, G. V. Mytsin, Y. I. Luchin and A. G. Molokanov Medico-Technical
More informationSmall Field Dosimetry: Overview of AAPM TG-155 and the IAEA-AAPM Code of Practice (Therapy) What is a Small Field? Indra J. Das/ Small Fields (Page-1)
Indra J. Das/ Small Fields (Page-1) Small Field Dosimetry: Overview of AAP TG-155 and the IAEA-AAP Code of Practice (Therapy) Indra J. Das, PhD, FAAP, FACR, FASTRO Department of Radiation Oncology Indiana
More informationph fax
This product is available through: JRT Associates www.standardimaging.com 800-261-4446. ph 608-831-0025. fax 608-831-2202 5 Nepperhan Avenue, Suite 2B 3120 Deming Way Middleton WIElmsford, 53562-1461 NY
More informationHampton University Proton Therapy Institute
Hampton University Proton Therapy Institute Brief introduction to proton therapy technology, its advances and Hampton University Proton Therapy Institute Vahagn Nazaryan, Ph.D. Executive Director, HUPTI
More informationSemiflex 3D. Always perfectly oriented. 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry
DETECTORS Always perfectly oriented. Semiflex 3D 3D Thimble Ionization Chamber for Relative and Absolute Dosimetry The New Reference Class Full 3D Geometry For FFF and FF Beams Semiflex 3D 3D Detector
More informationBasic Press Information
Basic Press Information Contact MedAustron EBG MedAustron GmbH Marie Curie-Strasse 5 A-2700 Wiener Neustadt Austria T +43 2622 26 100-0 e-mail: office@medaustron.at Internet: www.medaustron.at Press contact:
More information8/2/2018. Disclosures. In ICRU91: SRT = {SBRT/SABR, SRS}
High Dose, Small Field Radiation Therapy: Lessons from the HyTEC Project and the ICRU 91 Report Part 1: Small Field Dosimetry Jan Seuntjens, Ph.D, FCCPM, FAAPM, FCOMP Director and Professor, Medical Physics
More informationOut-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients
EURAMED,MELODI and EURADOS Workshop, Helsinki 19.-20.4.2017 Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients Jan Seppälä, Chief Physicist Tuomas Virén, Medical
More informationPHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)
PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) Course Description: This course is an introduction to physics in medicine and is intended to introduce
More informationFuture upcoming technologies and what audit needs to address
Future upcoming technologies and what audit needs to address Dr R.I MacKay History of audit Absolute dose - Simple phantom standard dose measurement Point doses in beams - Phantoms of relatively simple
More informationImportance of Radiation Dosimetry standards in preclinical radiobiology studies
Importance of Radiation Dosimetry standards in preclinical radiobiology studies Ceferino Obcemea Radiation Research Program National Cancer Institute, Bethesda, MD, USA CIRMS 2018 No financial conflict
More informationEDITORIAL NOTE A short CV
IAEA/WHO NETWORK OF SECONDARY STANDARD DOSIMETRY LABORATORIES Prepared by the Joint IAEA/WHO Secretariat of the SSDL Network Published semi-annually by the International Atomic Energy Agency Vienna No.
More informationCan we hit the target? Can we put the dose where we want it? Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy
Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern
More informationA fast Monte Carlo toolkit on GPU for treatment plan dose recalculation in proton therapy
A fast Monte Carlo toolkit on GPU for treatment plan dose recalculation in proton therapy Martina Senzacqua PhD student in Accelerator Physics at University of Rome La Sapienza Department of Basic and
More informationSUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES
A Personal Dosimetry Intercomparison Study in Asian and Pacific Region Hiroyuki MURAKAMI and Fumiaki TAKAHASHI Department of Health Physics, Japan Atomic Energy Research Institute and Richard V. Griffith
More informationIAEA/AAPM code of practice for the dosimetry of small static fields used in external beam radiotherapy
IAEA/AAPM code of practice for the dosimetry of small static fields used in external beam radiotherapy Hugo Palmans MedAustron, Wiener Neustadt, Austria and National Physical Laboratory, Teddington, UK
More informationDosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 11, NUMBER 2, SPRING 2010 Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification Sathiyan Saminathan, a Ravikumar
More informationProton and helium beams: the present and the future of light ion beam therapy
Proton and helium beams: the present and the future of light ion beam therapy Dr. Andrea Mairani Group Leader Biophysics in Particle Therapy Heidelberg Ion Beam Therapy Center HIT Department of Radiation
More informationTG-191 The clinical use of luminescent dosimeters
TG-191 The clinical use of luminescent dosimeters Task Group Members Stephen F. Kry, chair MD Anderson Cancer Center Paola Alvarez, MD Anderson Cancer Center Joanna Cygler, Ottawa Hospital Cancer Center
More informationHot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties. Sonja Dieterich, PhD University of California Davis
Hot Topics in SRS: Small Field Dosimetry & Other Treatment Uncertainties Sonja Dieterich, PhD University of California Davis Lecture in Two Parts SRS Dosimetry Other Uncertainties SRS DOSIMETRY Outline
More information8/1/2016. TG-191 The clinical use of luminescent dosimeters. Task Group Members. Scope of Report
TG-191 The clinical use of luminescent dosimeters Task Group Members Stephen F. Kry, chair MD Anderson Cancer Center Paola Alvarez, MD Anderson Cancer Center Joanna Cygler, Ottawa Hospital Cancer Center
More informationAccuracy Requirements and Uncertainty Considerations in Radiation Therapy
Departments of Oncology and Medical Biophysics Accuracy Requirements and Uncertainty Considerations in Radiation Therapy Introduction and Overview 6 August 2013 Jacob (Jake) Van Dyk Conformality 18 16
More informationEfficient Dosimetry for Proton Therapy
Efficient Dosimetry for Proton Therapy Why IBA Dosimetry? IBA Dosimetry offers the full product range tailored to fit any Proton Therapy QA needs. PT treatment safety, as well as most efficient dosimetry
More informationRadiotherapy Standards Users Meeting, 1 Dec 2008 Abstracts. New calorimeters Simon Duane, NPL
Radiotherapy Standards Users Meeting, 1 Dec 2008 Abstracts New calorimeters Simon Duane, NPL Calorimetry remains the most fundamental method for the measurement of absorbed dose and is the basis of almost
More informationCatharine Clark NPL, Royal Surrey County Hospital and RTTQA
Catharine Clark NPL, Royal Surrey County Hospital and RTTQA Background Historically UK has been relatively independent of the rest of the world in audit NPL PSDL International intercomparison for reference
More informationIII. Proton-therapytherapy. Rome SB - 5/5 1
Outline Introduction: an historical review I Applications in medical diagnostics Particle accelerators for medicine Applications in conventional radiation therapy II III IV Hadrontherapy, the frontier
More informationMonte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry
Monte Carlo water-equivalence study of two PRESAGE formulations for proton beam dosimetry T Gorjiara 1, Z Kuncic 1, J Adamovics 2 and C Baldock 1,3 1 Institute of Medical Physics, School of Physics, University
More informationActivity report from JCOG physics group
2013.5.9 Global Harmonization Group meeting ICCR2013 @ Melbourne Activity report from JCOG physics group 1 Hokkaido University, Graduate School of Medicine, 2 National Cancer Center, Center for Cancer
More informationPhantom Positioning Variation in the Gamma Knife Perfexion Dosimetry
Journal of Physics: Conference Series PAPER OPEN ACCESS Phantom Positioning Variation in the Gamma Knife Perfexion Dosimetry To cite this article: N A Costa et al 2016 J. Phys.: Conf. Ser. 733 012098 View
More informationCorrection of Prompt Gamma Distribution for Improving Accuracy of Beam Range Determination in Inhomogeneous Phantom
Original Article PMP Progress in Medical Physics 28(4), December 217 https://doi.org/1.14316/pmp.217.28.4.27 pissn 258-4445, eissn 258-4453 Correction of Prompt Gamma Distribution for Improving Accuracy
More informationDiode calibration for dose determination in total body irradiation
Iran. J. Radiat. Res., 2008; 6 (1): 43-50 Diode calibration for dose determination in total body irradiation M.Allahverdi 1*, Gh. Geraily 1, M. Esfehani 3,A. Sharafi 2,A. Shirazi 1 1 Department of Medical
More informationIAEA/AAPM Code of Practice for the Dosimetry of Static Small Photon Fields. Jan Seuntjens McGill University Montréal, Canada
IAEA/AAPM Code o Practice or the Dosimetry o Static Small Photon Fields Jan Seuntjens McGill University Montréal, Canada Acknowledgements IAEA/AAPM small and composite ield working group: Hugo Palmans
More informationHigh-Dose Standardization Study For γ-rays radiation processing at NIM
High-Dose Standardization Study For γ-rays radiation processing at NIM ZHANG Yanli, GAO Junchen, YANG Yuandi. Division of Metrology in Ionizing Radiation and medicine, National Institute of Metrology,
More informationCalculated LET spectrum from antiproton beams stopping in water
Acta Oncologica, 2009; 48: 223226 ORIGINAL ARTICLE Calculated LET spectrum from antiproton beams stopping in water NIELS BASSLER 1,2 & MICHAEL HOLZSCHEITER 3 1 Department of Experimental Clinical Oncology,
More informationNon-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.
Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact Stephen F. Kry, Ph.D., D.ABR. Goals Compare out-of-field doses from various techniques Methods to reduce out-of-field doses Impact of
More informationLarge-area Two-Dimensional Thermoluminescence Dosimetry System in Ion Beam Quality Assurance
Large-area Two-Dimensional Thermoluminescence Dosimetry System in Ion Beam Quality Assurance J. Gajewski1, M. Kłosowski1, S. Greilich2, P. Olko1 Institute of Nuclear Physics, Kraków, Poland German Cancer
More informationNeutrons. ρ σ. where. Neutrons act like photons in the sense that they are attenuated as. Unlike photons, neutrons interact via the strong interaction
Neutrons Neutrons act like photons in the sense that they are attenuated as I = I 0 e μx where Unlike photons, neutrons interact via the strong interaction μ = The cross sections are much smaller than
More informationAmendment No. 2. Item No. 2 (Rfx/ Event number )
Amendment 2 Sub: Amendment to the Document 06.12.2018 Ref.: Notice Inviting Bid ref. HITES/PCD/NCI-AIIMS/36/18-19 dated 26.09.2018 read with its Amendment no. 1 dated 19.11.18 The following changes have
More informationProgress Report on Radiation Dosimetry at NPL
Progress Report on Radiation Dosimetry at NPL Peter Sharpe and Simon Duane National Physical Laboratory Teddington, UK April 2009 1 Introduction This report gives a brief overview of radiation dosimetry
More informationJoint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009
233-3 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 29 Dosimetry Framework, Formalism including Uncertainties Costas Hourdakis Greek Atomic
More informationThe determination of timer error and its role in the administration of specified doses
JURNAL FIZIK MALAYSIA VOLUME 29, NUMBER 1&2 2008 The determination of timer error and its role in the administration of specified doses S. B. Samat a *, C. J. Evans b, T. Kadni c and M. T. Dolah c a School
More informationRadiation Monitoring Instruments
Radiation Monitoring Instruments This set of slides is based on Chapter 4 authored byg. Rajan, J. Izewska of the IAEA publication (ISBN 92-0-107304-6): Radiation Oncology Physics: A Handbook for Teachers
More information