APPLICATION OF HAND PHANTOMS IN SIMULATIONS TO DETERMINE THE RADIATION EXPOSURE OF MEDICAL STAFF
|
|
- Stanley Rice
- 5 years ago
- Views:
Transcription
1 APPLICATION OF HAND PHANTOMS IN SIMULATIONS TO DETERMINE THE RADIATION EXPOSURE OF MEDICAL STAFF Frank Becker 1, Christoph Blunck 1, Catarina Figueira 2, Fabian Göpfert 1 1 Karlsruhe Institute of Technology, KIT, Germany 2 Radiological Protection and Safety Unit, IST/ITN, Portugal INSTITUTE FOR NUCLEAR WASTE DISPOSAL - RADIATION PROTECTION RESEARCH KIT Universität des Landes Baden-Württemberg und nationales Forschungszentrum in der Helmholtz-Gemeinschaft
2 CT-Fluoroscopy staff dosimetry December 2009 Thousands of New Cancers Predicted Due to Increased Use of CT Doses Higher and More Variability Women Face Greater Risk Thousands of Future Cancers? CMPWG Nara, Japan - Frank Becker et al.
3 CT-Fluoroscopy - staff dosimetry Typical situation in CT fluoroscopy (Siemens) CMPWG Nara, Japan - Frank Becker et al.
4 CT-Fluoroscopy - staff dosimetry CTF biopsy showing biopsy needle and lesion (Katada, Fujita Health University, JP) CMPWG Nara, Japan - Frank Becker et al.
5 CT-Fluoroscopy - staff dosimetry * Scattered dose rates to equipment operator from CT (Exposure parameters: 120 kv, 50 ma) * New recommendation limit 20 msv : 50 procedures CMPWG Nara, Japan - Frank Becker et al.
6 CT-Measurements and Simulations => EURADOS Measurements and Simulations => Handphantoms CMPWG Nara, Japan - Frank Becker et al.
7 Measurements with Handphantoms phantom from LPS Berlin CMPWG Nara, Japan - Frank Becker et al.
8 Typical Dose Distribution H p (0.07) on Hands Simulations useful to reveal the overall/maximum dose distribution? CMPWG Nara, Japan - Frank Becker et al.
9 CT-Fluoroscopy - staff dosimetry CT devices in Karlsruhe and Coimbra CMPWG Nara, Japan - Frank Becker et al.
10 CT-Fluoroscopy - staff dosimetry Measurements in Karlsruhe and Coimbra (cont.) CMPWG Nara, Japan - Frank Becker et al.
11 CT-Fluoroscopy - staff dosimetry Simulations in of the scenarios validation of two programs C. Figueira: MCNPX Voxel Phantoms from DICOM-Files Rotational movement simulated by 36 source positions F. Göpfert: GMctdospp developed by Ralph Schmidt, Giessen GUI to handle EGSnrc Rotation mode in EGSnrc available DICOM-Files directly accepted in DICOM-RT- STRUCT format CMPWG Nara, Japan - Frank Becker et al.
12 CT-Fluoroscopy - staff dosimetry DICOM-Files CMPWG Nara, Japan - Frank Becker et al.
13 CT-Fluoroscopy - staff dosimetry MCNPX CMPWG Nara, Japan - Frank Becker et al.
14 CT-Fluoroscopy - staff dosimetry GMctdospp CMPWG Nara, Japan - Frank Becker et al.
15 CT-Fluoroscopy - staff dosimetry Placing TLDs in DICOM-RT-STRUCT files with the program Prosoma CMPWG Nara, Japan - Frank Becker et al.
16 CT-Fluoroscopy - staff dosimetry Simulations Measurements CMPWG Nara, Japan - Frank Becker et al.
17 CT-Fluoroscopy - staff dosimetry Strong beam spot influence (simulations) CMPWG Nara, Japan - Frank Becker et al.
18 CT-Fluoroscopy - staff dosimetry Beamspot Dosimeters Simulations CMPWG Nara, Japan - Frank Becker et al.
19 Methods Ch. Blunck, PhD Thesis Reales Szenario Handmodell Modell der radioaktiven Quelle 3-D Informationen Modellierung Übereinstimmung Simulation CMPWG Nara, Japan - Frank Becker et al.
20 Markers and Position Tracking CMPWG Nara, Japan - Frank Becker et al.
21 Dose Rate during Injection CMPWG Nara, Japan - Frank Becker et al.
22 Dose Rate during Injection 2.7 s CMPWG Nara, Japan - Frank Becker et al.
23 Dose Rate during Injection 17.3 s CMPWG Nara, Japan - Frank Becker et al.
24 Dose Rate during Injection 24.0 s CMPWG Nara, Japan - Frank Becker et al.
25 Dose Rate during Injection 28.0 s CMPWG Nara, Japan - Frank Becker et al.
26 Dose Rate during Handling under Laboratory Conditions CMPWG Nara, Japan - Frank Becker et al.
27 Video + Hand Simulation Ch. Blunck, PhD Thesis CMPWG Nara, Japan - Frank Becker et al.
28 Near Future: KIT and BOOSTER Gamma Camera BiO-dOSimetric Tools for triage to Responders Contribution to WP 300 Fast Evaluation Include sensors in existing tracing systems for dosimetry analysis - superimposition with visible images Laboratory tests will be carried out defining the performance of the gamma camera in running projects Comparison with Monte Carlo simulations and traditional dose distribution measurements Possibilities in nuclear medicine centres: Detecting traces of radioactive substances Detecting patients undergoing nuclear medicine therapy within a group of people Campus-Blog 2009 Nürnberger Zeitung CMPWG Nara, Japan - Frank Becker et al.
29 FIN F. Becker 1, Ch. Blunck 1, C. Figueira 2, F. Göpfert 1, B. Esteves 3, S. Di Maria 2, G. Paulo 4, J. Santos 4, P. Teles 2, P. Vaz 2, E. Martini 5 and J. Engelhardt 5 1 KIT 2 ITN, 3 Radiology Department of Hospital dos Covões 4 College of Health Technology of Coimbra 5 LPS Berlin CMPWG Nara, Japan - Frank Becker et al.
EURADOS: Overview of actions and initiatives in radiation protection in medicine
EURADOS: Overview of actions and initiatives in radiation protection in medicine Filip Vanhavere Oslo, June 8 th, 2011 EURADOS: Carries out projects and network activities to Advance the scientific understanding
More informationQuality and Safety of Patient Healthcare and Personalized Medicine Using Ionizing Radiation
Quality and Safety of Patient Healthcare and Personalized Medicine Using Ionizing Radiation Ana Catarina Antunes, Mariana Baptista, Ana Belchior, Jorge Borbinha, Salvatore di Maria, Célia Fernandes, Joana
More informationHighlights and Conclusions Thematic Area 4: Radiation of Patients
Highlights and Conclusions Thematic Area 4: Radiation of Patients by Robert H. Corbett and Eliseo Vañó Main topics: Diagnostic Radiology Interventional Radiology Nuclear Medicine Radiotherapy New Techniques
More informationPROSTATE BRACHYTHERAPY DOSIMETRY USING VOXEL PHANTOMS
PROSTATE BRACHYTHERAPY DOSIMETRY USING VOXEL PHANTOMS Sílvia Barros 1, Pedro Teles 1, Simone Cardoso 2, Alessandro Facure 3, Luiz da Rosa 4, Maíra Santos 4, Pedro Paulo Pereira Jr. 5, Isabel Gonçalves
More informationHow to assess doses from internal emitters
How to assess doses from internal emitters in Radiation Protection and Medicine PD Dr. Bastian Breustedt, Safety and Environment (SUM) KIT Die Forschungsuniversität in der Helmholtz-Gemeinschaft www.kit.edu
More informationMonte Carlo Modelling: a reliable and efficient tool in radiation dosimetry
Monte Carlo Modelling: a reliable and efficient tool in radiation dosimetry G. Gualdrini, P. Ferrari ENEA Radiation Protection Institute, Bologna (Italy) Contribution to the Italy in Japan 2011 initiative
More informationA. DeWerd. Michael Kissick. Larry. Editors. The Phantoms of Medical. and Health Physics. Devices for Research and Development.
Larry Editors A. DeWerd Michael Kissick The Phantoms of Medical and Health Physics Devices for Research and Development ^ Springer Contents 1 Introduction to Phantoms of Medical and Health Physics 1 1.1
More informationInternal dosimetry in emergency situations: challenges and recent developments
10th EURADOS Winter School (AM2017) Internal Dosimetry for Radiation Protection and Medicine KIT Karlsruhe (Germany), 1 March 2017 Internal dosimetry in emergency situations: challenges and recent developments
More informationPerformance of several active personal dosemeters in interventional radiology and cardiology
Performance of several active personal dosemeters in interventional radiology and cardiology S. Chiriotti 1,*, M. Ginjaume 1, E. Vano 2,3, R. Sanchez 3, J.M. Fernandez 2,3, M.A. Duch 1, J. Sempau 1 1.
More informationThe need for standardization of dosimetry in experimental radiation biology
The need for standardization of dosimetry in experimental radiation biology Kurt Pedersen University of Wisconsin Medial Radiation Research Center Under the direction of Larry DeWerd, PhD. NCCAAPM Fall
More informationUse of Monte Carlo Simulation Software for Calculating Effective dose in Cone Beam Computed Tomography
Use of Monte Carlo Simulation Software for Calculating Effective dose in Cone Beam Computed Tomography Wilson Otto Gomes Batista Instituto Federal da Bahia Rua Emidio dos Santos s/n. Barbalho 40301-015
More informationEffect of non- homogeneous breast 3ssue on mean glandular dose assessment in digital breast tomosynthesis
Effect of nonhomogeneous breast 3ssue on mean glandular dose assessment in digital breast tomosynthesis M. Bap3sta, S. Di Maria, C. Figueira, L. Orvalho, A. Silva, P. Vaz, M. Zankl o Mo3va3on Mammography
More informationGuidelines to optimize extremity monitoring and to reduce skin doses in nuclear medicine. Results of the ORAMED Project
Guidelines to optimize etremity monitoring and to reduce skin doses in nuclear medicine. Results of the ORAMED Project M. Ginjaume 1, M. Sans Merce 2*, S. Baechler 2, I. Barth 3, A. Carnicer 1, L. Donadille
More informationEURADOS Work on Internal Dosimetry
B. Breustedt (KIT, Germany), E. Blanchardon (IRSN, France), C.-M. Castellani (ENEA, Italy), G. Etherington (PHE (retired), UK), D. Franck (IRSN, France), A. Giussani (BfS, Germany), W. Hofmann (U Salzburg,
More informationGATE MONTE CARLO DOSIMETRY SIMULATION OF MARS SPECTRAL CT
GATE MONTE CARLO DOSIMETRY SIMULATION OF MARS SPECTRAL CT R Aamir, C Lowe, J Damet, P Carbonez, A P H Butler, N Schleich, N G Anderson MARFO, Emmanuel Geant4 User Workshop 2017, Wollongong Overview MARS
More informationA more accurate method to estimate patient dose during body CT examinations with tube current modulation
A more accurate method to estimate patient dose during body CT examinations with tube current modulation Poster No.: C-0738 Congress: ECR 2014 Type: Scientific Exhibit Authors: A. Kawaguchi 1, Y. Matsunaga
More informationOut-of-field dosimetry in radiotherapy for input to epidemiological studies. Roger Harrison
MELODI 7th Workshop, 9 11 November 2015 Helmholtz Zentrum München Next Generation Radiation Protection Research Out-of-field dosimetry in radiotherapy for input to epidemiological studies Roger Harrison
More informationMEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS
RADIOPROTECTION AND DOSIMETRY MEASUREMENT OF THE EQUIVALENT INDIVIDUAL DOSES FOR PATIENTS IN ANGIOGRAPHY PROCEDURE AND INTERVENTIONAL RADIOLOGY WITH THERMOLUMINESCENT SYSTEMS DANIELA ADAM 1, ANA STOCHIOIU
More informationMODELLING A GAMMA IRRADIATION PROCESS USING THE MONTE CARLO METHOD
2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte, MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 MODELLING A GAMMA IRRADIATION
More informationA comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography:
A comparison of radiation doses from modern multi-slice Computed Tomography angiography and conventional diagnostic Angiography: Rob Loader Oliver Gosling Introduction Approached by Dr Oliver Gosling (Research
More informationStrategic Research Agenda of EURAMED, highlighting synergies
Strategic Research Agenda of EURAMED, highlighting synergies Christoph Hoeschen Otto-von-Guericke University Magdeburg on behalf of the steering committee The Strategic Research Agenda Common Strategic
More informationMultislice CT versus cone beam CT: dosimetric and image quality comparision.
Multislice CT versus cone beam CT: dosimetric and image quality comparision. Poster No.: C-2589 Congress: ECR 2015 Type: Authors: Scientific Exhibit C. Dionisi 1, A. Crispo 2, E. Stefani 3, A. Lovaglio
More informationNational Cancer Institute
National Cancer Institute A Dosimetry Summary of CT Participants in the National Lung Screening Trial (NLST) U.S. DEPARTMENT OF HEALTH AND HUMAN SERVICES National Institutes of Health AAPM 2015 Anaheim,
More informationEURADOS WG12. Dosimetry in medical imaging
EURADOS WG12 EURADOS WORKING GROUP 12 Dosimetry in medical imaging (Main actions from February 2017) Željka Knežević RBI WG 12 group status WG12 is focused on patient and staff dosimetry in the medical
More informationNon-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.
Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact Stephen F. Kry, Ph.D., D.ABR. Goals Compare out-of-field doses from various techniques Methods to reduce out-of-field doses Impact of
More informationTowards a model of DTPA decorporation therapy
Towards a model of DTPA decorporation therapy B. Breustedt, KIT, Germany E. Blanchardon, IRSN, France P. Berard, CEA, France P. Fritsch, CEA, France O. Gremy, CEA, France A. Giussani, BfS, Germany M. Kastl,
More informationMeasurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method
Iran. J. Radiat. Res., 2004; 1(4): 187-194 Measurement of organ dose in abdomen-pelvis CT exam as a function of ma, KV and scanner type by Monte Carlo method M.R. Ay 1, M. Shahriari 2, S. Sarkar 3, P.
More informationPatient Dose Estimates. from CT Examinations. Patient Dose Estimates
Patient Dose Estimates from CT Examinations Patient Dose Estimates from CT Examinations John M. Boone, Ph.D. Professor and Vice Chairman of Radiology Professor of Biomedical Engineering University of California,
More informationEric E. Klein, Ph.D. Chair of TG-142
Eric E. Klein, Ph.D. Chair of TG-142 Professor of Radiation Oncology Washington University St. Louis, MO 2010 AAPM Annual Meeting Med. Phys. 21(4) 1994 Performance-based, comprehensive guidelines for preventing
More informationDosimetry benchmark for MBDCA
Clinical implementation for advanced brachytherapy dose calculation algorithms beyond the TG-43 formalism: Dosimetry benchmark for MBDCA F. Ballester PhD University of Valencia, Spain Disclosure Research
More informationIEC Subcommittee 62C (Equipment for Radiotherapy, Nuclear Medicine and Radiation Dosimetry): Recent and Active Projects
IEC Subcommittee 62C (Equipment for Radiotherapy, Nuclear Medicine and Radiation Dosimetry): Recent and Active Projects 2015 AAPM Annual Meeting Anaheim, CA Wes Culberson, PhD, DABR University of Wisconsin
More informationAssessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom
Assessing Heterogeneity Correction Algorithms Using the Radiological Physics Center Anthropomorphic Thorax Phantom David Followill, Ph.D. Associate Director Radiological Physics Center RPC History Lesson
More informationMONITORING OF OCCUPATIONAL EXPOSURE AT NUCLEAR FACILITIES
GUIDE YVL 7.10 / 29 JANUARY 2002 MONITORING OF OCCUPATIONAL EXPOSURE AT NUCLEAR FACILITIES 1 GENERAL 3 2 PROVISIONS OF THE RADIATION ACT AND DECREE 3 3 MONITORING OF RADIATION EXPOSURE 4 3.1 General requirements
More informationOut-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients
EURAMED,MELODI and EURADOS Workshop, Helsinki 19.-20.4.2017 Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients Jan Seppälä, Chief Physicist Tuomas Virén, Medical
More informationCALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE
CALCULATION OF BACKSCATTER FACTORS FOR LOW ENERGY X-RAYS USING THE TOPAS MONTE CARLO CODE Emily Hewson 1 Martin Butson 1,2 Robin Hill 1,2 1 Institute of Medical Physics, School of Physics, University of
More informationImproving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations
Improving personal dosimetry of medical staff wearing radioprotective garments: Design of a new whole-body dosimeter using Monte Carlo simulations Clarita Saldarriaga Vargas, Corinne Amalberto, Lara Struelens,
More informationMAESTRO: Methods and Advanced Equipment for Simulation and Treatment in Radiation Oncology
What is MAESTRO? Sixth Framework Program FP6 (2002-2006) European Commission Integrated Project Duration: 5 years MAESTRO: Methods and Advanced Equipment for Simulation and Treatment in Radiation Oncology
More informationValidation of the MC platform GATE for radiation therapy applications Evaluation of dose deposited by electrons. Yann Perrot, LPC, CNRS/IN2P3
Validation of the MC platform GATE for radiation therapy applications Evaluation of dose deposited by electrons Yann Perrot, LPC, CNRS/IN2P3 1 Projects @ PCSV 2 Projects @ PCSV Computation for life science
More informationX-RAYS INDIVIDUAL DOSE ASSESSMENT USING TLD DOSIMETERS
X-RAYS INDIVIDUAL DOSE ASSESSMENT USING TLD DOSIMETERS Carlos Salas Nucleoeléctrica Argentina SA - Embalse NPP Environmental Control and Dosimetry Division INTRODUCTION: This paper describes the methodology
More informationThoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters
Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters Poster No.: C-2584 Congress: ECR 2015 Type: Scientific
More informationBorges C 1, Zarza- Moreno M 2, Teixeira N 2, Vaz P 3 1
Borges C 1, Zarza- Moreno M 2, Teixeira N 2, Vaz P 3 1 Medicalconsult SA, Lisboa, Portugal; 2 Escola Superior de Tecnologias da Saúde, Lisboa, Portugal; 3 InsEtuto Tecnológico e Nuclear, InsEtuto Superior
More informationMetrology Laboratory of Ionizing Radiation
Metrology Laboratory of Ionizing Radiation Carlos Oliveira This year it was signed a new collaborative protocol between the ITN and the IPQ, replacing the former, dated from 1989, and stepping up forms
More informationPHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)
PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015) Course Description: This course is an introduction to physics in medicine and is intended to introduce
More informationDependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose
Submitted to Radiation Measurements Dependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose P. Bilski 1, M. Sadel 1, J. Swakon 1, A. Weber 2 1
More informationImaging of Radiation Dose Using Cherenkov Light
Imaging of Radiation Dose Using Cherenkov Light Eric Brost 1, Yoichi Watanabe 1, Fadil Santosa 2, Adam Green 3 1 Department of Radiation Oncology, University of Minnesota 2 Institute for Mathematics and
More informationMetrology Laboratory of Ionizing Radiation
Metrology Laboratory of Ionizing Radiation Carlos Oliveira The Metrology Laboratory of Ionizing Radiation (LMRI) has been involved in activities related with scientific, technical and legal metrology.
More informationComparison of absorbed fraction of Gamma and Beta rays of I-124 and I-131radio-isotopes in thyroid gland with Monte Carlo Simulation
Available online at http://www.ijabbr.com International journal of Advanced Biological and Biomedical Research Volume 1, Issue 9, 2013: 993-998 Comparison of absorbed fraction of Gamma and Beta rays of
More informationDosimetric Analysis Report
RT-safe 48, Artotinis str 116 33, Athens Greece RT-safe +30 2107563691 info@rt-safe.com Dosimetric Analysis Report SAMPLE, for demonstration purposes only Date of report: ------ Irradiation system: ------
More informationAsk EuroSafe Imaging. Tips & Tricks. Paediatric Imaging Working Group. Shielding in pediatric CT
Ask EuroSafe Imaging Tips & Tricks Paediatric Imaging Working Group Shielding in pediatric CT Claudio Granata (IRCCS Istituto Giannina Gaslini, IT) Joana Santos (ESTeSC-Coimbra Health School, PT) Elina
More informationThe x-rays produced penetrate the body which absorbs, refracts, or reflects the x-ray beam energy depending on the tissue. Bone
Authors Sari Cohen, Poh Yan Lim, Merng Koon Wong, Siew Hong Lau, Donna Russell-Larson 1.6.2 Image intensifier Poh Yan Lim, Merng Koon Wong The discovery of x-rays had a profound impact on the diagnosis
More informationPatient dose assessment of CT perfusion scanning at the RSCH
Patient dose assessment of CT perfusion scanning at the RSCH Lesley Leavesley, Emma Whitehead, Matthew Pryor, Debbie Peet Regional Radiation Protection Service Royal Surrey County Hospital, Guildford Overview
More informationPractice and Risk at Medical Facilities in Agency Operations
Practice and Risk at Medical Facilities in Agency Operations Igor Gusev Radiation Protection Unit IAEA International Atomic Energy Agency Outline What is medical radiation exposure? Radiation sources and
More informationA COMPARISON BETWEEN GEANT4 AND MCNPX ON THE DOSIMETRY OF THE 192IR MICROSELECTRON V2 HDR BRACHYTHERAPY
2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 A COMPARISON BETWEEN GEANT4
More informationPersonal Dosimetry in Sweden - Challenges and Developments
Personal Dosimetry in Sweden - Challenges and Developments - Present status - Unresolved issues - Developments and trends Swedish Facilities Nuclear facilities 10 operating and 3 closed reactors, fuel
More informationComputed tomography Acceptance testing and dose measurements
Computed tomography Acceptance testing and dose measurements Jonas Andersson Medical Physicist, Ph.D. Department of Radiation Sciences University Hospital of Norrland, Umeå Sweden Contents The Computed
More information45 Hr PET Registry Review Course
45 HR PET/CT REGISTRY REVIEW COURSE Course Control Document Timothy K. Marshel, MBA, R.T. (R), (N)(CT)(MR)(NCT)(PET)(CNMT) The PET/CT Training Institute, Inc. SNMMI-TS 028600-028632 45hr CEH s Voice Credits
More informationJoint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009
2033-4 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for General Radiology and Clinical Uncertainty Peter Homolka EFOMP Training
More informationIN VIVO IMAGING Proton Beam Range Verification With PET/CT
IN VIVO IMAGING Proton Beam Range Verification With PET/CT Antje-Christin Knopf 1/3 K Parodi 2, H Paganetti 1, T Bortfeld 1 Siemens Medical Solutions Supports This Project 1 Department of Radiation Oncology,
More informationIon radiography as a tool for patient set-up & image guided particle therapy: a Monte Carlo study
University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2014 Ion radiography as a tool for patient set-up
More informationIMAGE-GUIDED RADIATION THERAPY
IMAGE-GUIDED RADIATION THERAPY Your Single Source Oncology Solutions Provider Plan. Target. Treat. At Best NOMOS, we design products and solutions that help medical professionals treat a variety of cancers.
More informationCOMPARING THE RESPONSES OF TLD 100, TLD 600, TLD 700 AND TLD 400 IN MIXED NEUTRON-GAMMA FIELDS
2015 International Nuclear Atlantic Conference - INAC 2015 São Paulo, SP, Brazil, October 4-9, 2015 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-06-9 COMPARING THE RESPONSES OF TLD
More informationPreliminary study of MAGAT polymer gel dosimetry for boron-neutron capture therapy
Journal of Physics: Conference Series OPEN ACCESS Preliminary study of MAGAT polymer gel dosimetry for boron-neutron capture therapy To cite this article: Shin-ichiro Hayashi et al 2015 J. Phys.: Conf.
More informationImaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT
Imaging of Scattered Radiation for Real Time Tracking of Tumor Motion During Lung SBRT April 25 nd, 2015 Lung Cancer Lung cancer is the most lethal cancer: Over 224,000 new diagnoses in the U.S. predicted
More informationToshiba Aquillion 64 CT Scanner. Phantom Center Periphery Center Periphery Center Periphery
Comparison of radiation dose and imaging performance for the standard Varian x-ray tube and the Richardson Healthcare ALTA750 replacement tube for the Toshiba Aquillion CT scanners. by Robert L. Dixon,
More informationRadiation Exposure and Image Quality in X-Ray Diagnostic Radiology
Horst Aichinger Joachim Dierker Sigrid Joite-Barfuß Manfred Säbel Radiation Exposure and Image Quality in X-Ray Diagnostic Radiology Physical Principles and Clinical Applications Second Edition Springer
More informationFEE RULES RADIATION ONCOLOGY FEE SCHEDULE CONTENTS
Tel: +27-21-9494060 Fax: +27-21-9494112 E-mail: leon.gouws@cancercare.co.za FEE RULES RADIATION ONCOLOGY FEE SCHEDULE CONTENTS 1. EXTERNAL BEAM RADIATION... 2 2. PLANNING OF TREATMENT... 2 3. DELIVERY
More informationMonte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation.
Monte Carlo simulation of 192 Ir radioactive source in a phantom designed for brachytherapy dosimetry and source position evaluation Samuel Chiquita 1 1 University of Porto, Porto, Portugal Abstract In
More informationSmall field dosimetry, an example of what a Medical Physicist does (& some more examples) A/PROF SCOTT CROWE MEDICAL PHYSICIST
Small field dosimetry, an example of what a Medical Physicist does (& some more examples) A/PROF SCOTT CROWE MEDICAL PHYSICIST What is a Medical Physicist? Medical physics application of physics concepts
More information6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS
6/29/2012 MANAGEMENT OF RADIOTHERAPY PATIENTS WITH IMPLANTED CARDIAC DEVICES Dimitris Mihailidis, PhD., Charleston Radiation Therapy Consultants Charleston, WV 25304 WHAT IS IN THIS PRESENTATION? Types
More informationCalculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer
Calculation of Effective Doses for Radiotherapy Cone-Beam CT and Nuclear Medicine Hawkeye CT Laura Sawyer Department of Medical Physics and Bioengineering, Royal United Hospital, Bath Overview Varian Acuity
More informationRadiation protection in proton therapy
Radiation protection in proton therapy Pieternel van der Tol Medical Physicist - HollandPTC pvandertol@hollandptc.nl Marjan Dwarswaard René Bolt Marc-Jan van Goethem Lars Murrer Outline Introduction Interactions
More informationArticle. Reference. Comparison of organ doses and image quality between CT and flat panel XperCT scans in wrist and inner ear examinations
Article Comparison of organ doses and image quality between CT and flat panel XperCT scans in wrist and inner ear examinations DAMET, J, et al. Abstract The aim of this study was to evaluate and compare
More informationFLUKA Monte Carlo Simulation for the Leksell Gamma Knife Perfexion TM radiosurgery system
FLUKA Monte Carlo Simulation for the Leksell Gamma Knife Perfexion TM radiosurgery system Collaboration Project: Θ Θ Θ INFN, Milan; Medical Physics Department, Niguarda Ca'Granda Hospital, Milan; Elekta
More informationPrepublication Requirements
Issued Prepublication Requirements Standards Revisions for Organizations Providing Fluoroscopy Services The Joint Commission has approved the following revisions for prepublication. While revised requirements
More informationRadiologic Units: What You Need to Know
Radiologic Units: What You Need to Know TODD VAN AUKEN M.ED. RT (R)(MR) Agenda Greys, Sieverts, Coulombs per kg, & Becquerel's Conventional Units Other Concepts (LET, Q-Factor, Effective Dose, NCRP Report
More informationAgenda: Dental Cone Beam Imaging
Cone Beam Imaging Agenda: Dental Cone Beam Imaging *Definition and Functionality *Usage and diagnostics benefits *Comparative radiation information *Federal regulatory responsibilities: manufacturing *State
More informationCOMPLEMENTARY TRAINING MATERIALS DUE TO NEW REGULATION. Ringhals AB
COMPLEMENTARY TRAINING MATERIALS DUE TO NEW REGULATION Ringhals AB 1 NEW REGULATION IN RADIATION PROTECTION Since 1 June 2018, Sweden has a new Radiation Protection Act (SFS 2018: 396), a new Radiation
More informationTesting of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution
Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Ferenc Giczi a*, Sándor Pellet b, Ian Donald McLean c and Ahmed Meghzifene c a Széchenyi
More informationWN MEDICAL IMAGING. RADIOTHERAPY. MEDICAL PHYSICS. NUCLEAR MEDICINE. RADIOACTIVITY
WN MEDICAL IMAGING. RADIOTHERAPY. MEDICAL PHYSICS. NUCLEAR MEDICINE. RADIOACTIVITY Definitions : (from various online encyclopaedias/dictionaries) Medical Imaging/Radiology/ Diagnostic imaging: the use
More informationVariation of Occupational Doses among Subspecialties in Diagnostic Radiology. A.N. Al-Haj, C.S. Lagarde, A.M. Lobriguito
Variation of Occupational Doses among Subspecialties in Diagnostic Radiology A.N. Al-Haj, C.S. Lagarde, A.M. Lobriguito Biomedical Physics Department, MBC 03 King Faisal Specialist Hospital and Research
More informationRadiation physics and radiation protection. University of Szeged Department of Nuclear Medicine
Radiation physics and radiation protection University of Szeged Department of Nuclear Medicine Radiation doses to the population 1 Radiation doses to the population 2 Sources of radiation 1 Radiation we
More informationRecent Achievements in External Radiation Dosimetry A. Delgado CIEMAT Radiation Dosimetry Unit Avda Complutense 22, Madrid, Spain
Topical Session TS 1.1.1 External Exposure to Ionizing Radiation Keynote Lecture: Recent Achievements in External Radiation Dosimetry A. Delgado CIEMAT Radiation Dosimetry Unit Avda Complutense 22, 28040
More informationThe Management of Imaging Procedure Dose Nuclear Medicine Dose Indices
The Management of Imaging Procedure Dose Nuclear Medicine Dose Indices Wesley E. Bolch, PhD, PE, DABHP, FHPS, FAAPM Director, Advanced Laboratory for Radiation Dosimetry Studies Department of Biomedical
More informationTracking Doses in the Pediatric Population
Tracking Doses in the Pediatric Population Frederic H. Fahey DSc Boston Children s Hospital Harvard Medical School frederic.fahey@childrens.harvard.edu Disclosures Sadly, none that pay me any money! SNMMI
More informationUncertainties on internal dosimetry
Uncertainties on internal dosimetry Augusto Giussani 2 March 2017 agiussani@bfs.de Internal dosimetry Internal dose is evaluated with mathematical models Intake Biokinetic Model Time-activity curves in
More informationSunCHECK Patient Comprehensive Patient QA
The SunCHECK Platform SunCHECK Patient Comprehensive Patient QA SunCHECK Patient // Comprehensive Patient QA Automate Your Patient QA SunCHECK Patient automates all your Patient QA needs from Secondary
More informationEORTC Member Facility Questionnaire
Page 1 of 9 EORTC Member Facility Questionnaire I. Administrative Data Name of person submitting this questionnaire Email address Function Phone Institution Address City Post code Country EORTC No Enter
More informationOrgan Dose Reconstruction for Wilms Tumor Patients Treated with Radiation Therapy
Organ Dose Reconstruction for Wilms Tumor Patients Treated with Radiation Therapy Rasha Makkia Biomedical Physics Ph.D. student East Carolina University March 3 rd, 2016 Outlines The Purpose Wilms Tumor,
More informationInvivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters
Original Article PROGRESS in MEDICAL PHYSICS Vol. 26, No. 2, June, 2015 http://dx.doi.org/10.14316/pmp.2015.26.2.93 Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters
More informationDosimetric Consideration in Diagnostic Radiology
Dosimetric Consideration in Diagnostic Radiology Prof. Ng Kwan-Hoong Department of Biomedical Imaging University of Malaya ngkh@um.edu.my Radiation Dosimetry Workshop, 28-29 March 2014 2 Why do we measure
More informationQuality Assurance beam tracking Nami Saito GSI, Darmstadt, Germany biophysics department
Quality Assurance beam tracking Nami Saito GSI, Darmstadt, Germany biophysics department 1 Oct 2009 N. Saito PTCOG 48 Heidelberg 1 Introduction Beam tracking GSI beam tracking system Beam tracking QA Function
More informationCalculation of Normalised Organ and Effective Doses to Adult Reference Computational Phantoms from Contemporary Computed Tomography Scanners
Progress in NUCLEAR SCIENCE and TECHNOLOGY, Vol. 2, pp.165-171 (2011) ARTICLE Calculation of Normalised Organ and Effective Doses to Adult Reference Computational Phantoms from Contemporary Computed Tomography
More informationDOSE ASSESSMENT WITH PASSIVE PERSONAL DOSIMETERS EXPOSED TO X-RAY GENERATOR USING THE 241 Am CALIBRATION CURVE
DOSE ASSESSMENT WITH PASSIVE PERSONAL DOSIMETERS EXPOSED TO X-RAY GENERATOR USING THE 241 Am CALIBRATION CURVE F. MIHAI, A. STOCHIOIU, C. TUCA National Institute for Nuclear Physics and Engineering, P.O.Box
More informationA Navigation System for Minimally Invasive CT-Guided Interventions
A Navigation System for Minimally Invasive CT-Guided Interventions Markus Nagel 1,GerdSchmidt 2, Ralf Petzold 3, and Willi A. Kalender 1 1 Institute of Medical Physics, University of Erlangen-Nürnberg,
More informationRADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY. L19: Optimization of Protection in Mammography
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L19: Optimization of Protection in Mammography
More informationRecent proceedings in Brachytherapy Physics
Recent proceedings in Brachytherapy Physics Frank-André Siebert UKSH, Campus Kiel, Germany Clinic of Radiotherapy Dept. Medical Physics Physical characteristics of brachytherapy (Courtesy Luc Beaulieu,
More informationAccreditation of Dosimetry Services
Accreditation of Dosimetry Services J. Van Cauteren, E. De Geest, M. Bricoult AV-Controlatom Content History Regulatory requirements Discussion of reference documents/standards Practical implementation
More informationRadiation Protection Dosimetry (2008), Vol. 131, No. 1, pp Advance Access publication 23 August 2008
Radiation Protection Dosimetry (2008), Vol. 131, No. 1, pp. 62 66 Advance Access publication 23 August 2008 doi:10.1093/rpd/ncn229 AN OVERVIEW OF THE USE OF EXTREMITY DOSEMETERS IN SOME EUROPEAN COUNTRIES
More informationQuanitative and qualitative analysis of Image Quality and Imaging Dose of KV CBCT from XVI Elekta Linac.
IOSR Journal of Applied Physics (IOSR-JAP) e-issn: 2278-4861.Volume 9, Issue 5 Ver. IV (Sep. - Oct. 2017), PP 46-52 www.iosrjournals.org Quanitative and qualitative analysis of Image Quality and Imaging
More informationEXTREMITY EXPOSURE IN NUCLEAR MEDICAL THERAPY WITH Y-90 LABELLED SUBSTANCES
EXTREMITY EXPOSURE IN NUCLEAR MEDICAL THERAPY WITH Y-9 LABELLED SUBSTANCES A. Rimpler 1, I. Barth 1, P. Ferrari 2, S. Baechler 6, A. Carnicer 3, L. Donadille 4, M. Fulop 5, M. Ginjaume 3, M. Mariotti 2,
More information