Implementation and analysis of international standard for electroacoustic performance evaluation of hearing aids
|
|
- Bertram Lang
- 5 years ago
- Views:
Transcription
1 Acoustical Measurements and Instrumentation: Paper ICA Implementation and analysis of international standard for electroacoustic performance evaluation of hearing aids Zargos Neves Masson (a), Eduardo Bresciani (b), Stephan Paul (c), Júlio A. Cordioli (d) (a) Federal University of Santa Catarina, Brazil, (b) Federal University of Santa Catarina, Brazil, (c) Federal University of Santa Catarina, Brazil, (d) Federal University of Santa Catarina, Brazil, Abstract Hearing aids are the most common device used by hearing impaired people. The hearing aids electroacoustic performance is of utmost importance for hearing care professionals to properly choose and adapt the device for a particular individual. Harmonized methods for determining the electroacoustic characteristics are fundamental to allow comparison between different hearing aids. Therefore, the International Electrotechnical Commission has published a series of standards giving recommendations for hearing aids electroacoustic measurement procedures. This article presents an analysis of the implementation of the standard IEC : Measurement of the performance characteristics of hearing aids. This standard is used to obtain results for manufacturer data sheets. First, a test setup validation was conducted by comparing the results with those measured by an accredited laboratory for a reference hearing aid. After that, investigations were made to analyze the influence of different measurement aspects, like hearing aid positioning, presence of a control microphone at the test space and differences between an ear simulator and a 2CC coupler. Comparisons with the reference showed good agreement between the results considering the respective uncertainties. The use of a 2CC coupler in IEC was considered the better approach from the perspective of the standard purpose. The influence of the control microphone position and presence was found to be minimum in all investigations made. Keywords: Hearing aids; Electroacoustic; Measurements.
2 Implementation and analysis of international standard for electroacoustic performance evaluation of hearing aids 1 Introduction Hearing aids (HA) are the most common device used by hearing impaired people. Generally, an audiologist will adapt a hearing aid in order to amplify sounds so that the sound pressure level (SPL) produced remain inside the range that a particular individual is still capable of hearing. The electroacoustic performance plays an important role in this process, and hearing care professionals need to know how the hearing aid performs to properly choose and adapt the device to the patients hearing loss. Harmonized methods for determining the electroacoustic characteristics are fundamental to allow comparison between different hearing aids. Therefore, the International Electrotechnical Commission has published a series of standards giving recommendation for hearing aids electroacoustic measurement procedures under the number IEC The first part of the series is entitled Measurement of the performance characteristics of hearing aids [1] and standardizes the procedures using a free-field technique to obtain results for manufacturer data sheets, which are baseline information for clinicians. The measurement procedure is defined in such a way as to be practical and reproducible, and is consequently based on fixed parameters. Recently, the IEC standard went through a revision and significant technical changes with respect to the previous edition were made. One of the changes was the substitution of the acoustic coupler used. In previous editions the Occluded Ear Simulator 1 was used [2], while the current version of the IEC standard recommends the 2CC Coupler 2 [1]. The 2CC coupler is intended for loading a hearing aid with a specified acoustic impedance, but not to model the sound pressure in a subject s ear canal. It is therefore important to note that the performance measured under test conditions may deviate substantially from the performance of the hearing aid in actual conditions of use. This paper will focus only on presenting results and analysis for the acoustic frequency response 3 procedures obtained by means of the hearing aid microphone as the inlet method provided by the standard aforementioned. First, a measurement setup validation, that used a reference results issued by an accredited laboratory, will be discussed. An investigation of different aspects that influence the measurements results, like test arrangement layout and test space, is carried out subsequently. 1 Described by IEC [3], referred only as ear simulator throughout the paper. 2 Described by IEC [4]. 3 Every mention of frequency response in this paper should be interpreted as sound pressure level frequency spectrum. Nomenclature used by IEC [1]. 2
3 2 IEC Standard 2.1 Measurement Setup IEC requires the measurement to be carried out in a test enclosure with low background noise level (10 db below the lowest SPL generated during tests) that allows the simulation of free-field conditions by maintaining a constant sound pressure level at a reference point using a control microphone. In the present work, measurements were carried out in a hemi anechoic chamber with working dimensions of 7 m x 7 m x 4 m with a test assembly as shown in Figure 1. The chamber s floor was covered by acoustic edges in order to minimize reflections. Overall background noise measured at the test point was below 30 db, complying with the requirements established by the standard. (a) Assembly of the test in the chamber (b) Omnidirectional HA. (c) Directional HA. Figure 1: IEC test arrangements As seen in Figure 1, the axis of the control microphone shall be orthogonal to the speaker axis and shall intersect it at the reference point of the hearing aid. The reference point is located at the center of the inlet microphone port, or midpoint of the microphone port array, marked in red in Figure 1b and 1b, respectively. A laser leveling system was used to aid positioning. 2.2 Uncertainty Analysis Sound pressure level measured in the acoustic coupler, L p,coup, is calculated by L p, coup = L V + L s + δ T + δ P + δ H + δ coup + δ cal + δ Tub + δ Ass + δ Lp,ref, (1) being L V the voltage level db re.1v acquired by the acquisition board, L s is the microphone sensitivity db re.1v/pa, δ T is the correction associated with variations in temperature, δ P is the correction associated with variations in ambient pressure, δ H is the correction associated with variations in humidity, δ coup is an input quantity to allow any errors caused by the coupler, δ cal is an input quantity to allow any errors caused by the calibration check procedure, δ Tub is an input quantity to allow any errors caused by the tube connecting hearing aid and coupler, δ Ass is an input quantity to allow any errors from the assembly, δ Ass is an input quantity to allow any errors from the sound pressure level at the reference point. All values are given in decibels. The combination of uncertainties on logarithmic scale, is realized in ISO GUM [5], and it is 3
4 considered a conservative approach [6]. Since equation 1 is a sum of logarithmic terms, combined uncertainty µ Lp, coup through is obtained µ 2 L p, coup = µ 2 L V + µ 2 L s + δ 2 T + δ 2 P + δ 2 H + δ 2 coup + δ 2 cal + δ 2 Tub + δ 2 Ass + δ 2 L p,ref. (2) The expanded uncertainty is then calculated by U Lp, coup = kµ Lp, coup, (3) where k is a coverage factor equal to 2 for a confidence level of 95,45%. Corrections due to ambient conditions were not applied, and instead it was considered in the combined uncertainty as a maximum error assuming the quantity was within a determined range. During measurements, ambient conditions were monitored to be always within the range defined by IEC Uncertainty analysis was conducted for both couplers and results are presented in Tables 1 and 2. Table 1: Uncertainty analysis table for Sound pressure level with the 2CC Coupler Source Random effects Symbol Description of Uncertainty Values [db] Distribution µ [db] µ LV Voltage measurement normal 0.16 µ Ls Microphone sensibility normal 0.06 δ T Temperature correction rectangular 0.03 δ P Ambient pressure correction rectangular 0.01 δ H Humidity correction rectangular 0.01 δ coup,2cc 2CC coupler maximum error rectangular 0.09 δ cal Pistonphone calibration maximum error rectangular 0.06 δ Tub Tubing maximum error rectangular 0.29 δ Ass Assembly maximum error normal 0.30 δ Lp,ref Sound pressure level at the reference point 0.42 normal 0.21 µ Lp, coup:2cc Combined uncertainty SPL 2CC coupler - normal 0.51 U Lp, coup:2cc Expanded uncertainty (95.45%) SPL 2CC coupler normal 1.02 The difference in the combined uncertainty between the two couplers is mainly due the lack of calibration for the ear simulator s microphone by an accredited laboratory, requiring the uncertainty to be extracted from the data sheet. Ear simulator sound pressure level uncertainty declared by the accredited laboratory was 1.5 db for frequencies up to 4kHz and 1.9 db from 4kHz up to 8kHz. The maximum permitted expanded uncertainty allowed by the standard for sound pressure level are: 2 db for frequencies up to 4kHz and 2.5 db for frequencies from 4kHz up to 8kHz. 4
5 Table 2: Uncertainty analysis table for the sound pressure level with the Ear Simulator Coupler Source Random effects Symbol Description of Uncertainty Values [db] Distribution µ [db] µ LV Voltage acquisition normal 0.16 µ Ls Microphone sensibility rectangular 0.58 δ T Temperature correction rectangular 0.03 δ P Ambient Pressure correction rectangular 0.01 δ H Humidity correction rectangular 0.01 δ coup:es,up4 Ear Simulator up to 4kHz maximum error rectangular 0.23 δ coup:es,4-8 Ear Simulator 4kHz - 8kHz maximum error rectangular 0.46 δ cal Pistonphone maximum calibration error rectangular 0.29 δ Tub Tubing maximum error rectangular 0.06 δ Ass Assembly maximum error normal 0.30 δ Lp,ref Sound pressure level at the reference point normal 0.21 µ Lp, coup:es,up4 Combined Uncertainty up tp 4kHz 0.80 µ Lp, coup:es,4-8 Combined Uncertainty 4 khz - 8kHz 0.89 U Lp, coup:es,up4 Expanded Unc. (95.45%) up to 4 khz SPL ES 1.59 U Lp, coup:es,4-8 Expanded Unc. (95.45%) 4kHz - 8 khz SPL ES Measurements All measurements to obtain the frequency response were performed using pure tone signals over the range of 200 Hz Hz, using the ISO R40 preferred frequencies (approximately 1/24th octave frequencies). The signal was processed using a Hanning window, the autospectrum was calculated and the SPL in the frequency of excitation was extracted Sound pressure level adjustment Sound pressure level adjustment at the reference point is critical for carrying out the tests since it is generally assumed that the SPL at the reference point is constant. An adjustment was carried out by determining the sound source sensitivity, which can be calculated by S( f ) = L V ( f ) L p,ref ( f ), (4) since the relationship between the logarithmic voltage level L V ( f ) in db and the sound pressure level at the reference point L p,ref ( f ) in db can be assumed to be linear. Also, the calculation of sensitivity must be performed for each frequency, since the speaker s response is not flat OSPL90 The OSPL90 curve is obtained by setting the input SPL to 90 db, varying the frequency of the sound source and recording the coupler SPL versus frequency. This input SPL is more than 5
6 enough to produce the highest possible output level produced by the hearing aid at each frequency and is also known as saturation sound pressure level (SSPL). The gain control settings of the hearing aid are set to be full-on, and all adaptive functions should be turned off. From that measurement two parameters are determined: the maximum value in the curve known as Max-OSPL90, and an arithmetic mean over the OSPL at three frequencies, typically 1, 1.6 and 2.5 khz, is referred as HFA-OSPL Full-on gain response curve To measure the full-on gain response curve the input sound pressure level is set to 50 db and the gain is recorded as the difference obtained by subtracting 50 db from the acoustic coupler SPL. The full-on gain response curve (FOGRC) reflects the maximum gain that the hearing aid is capable. A 50 db input sound pressure level is generally low enough to ensure that the hearing aid does not saturate. From this response the HFA-FOG can be calculated as the arithmetic mean at the same three frequencies used in HFA-OSPL Basic frequency Response curve The basic frequency response (BFR) is measured by setting the hearing aid gain to the reference test settings (RTS). The RTS is defined as the setting of the gain control required to produce a HFA-gain within ±1.5 db of the HFA-OSPL90 minus 77 db, or, if the full-on HFA gain for an input SPL of 60 db is less than the HFA-OSPL90 minus 77 db, RTS will be the the full-on setting of the gain control. After the adjustment, input SPL is set to 60 db and the Basic Frequency Response curve is obtained stepping through all pure tone frequencies of interest and recording the response from the acoustic coupler. The frequency range is limited by f min e f max, being these the frequencies at which the BFR corresponds to the HFA output level minus 20 db. 3 Results An analog hearing aid previously tested by an European accredited laboratory using the previous version of the standard 5 was selected for the validation of the test setup and procedures. It is important to note that the only difference found between the versions that affected the results was the change of couplers. So, in order to properly compare the results and comply with the new version of the standard 6, measurements were performed with an ear simulator and a 2CC coupler and will be referred as LVA ES and LVA 2CC, respectively. Results obtained by using a commercial FONIX 8000 hearing aid analyzer were also included in the comparison, referred to as FONIX 2CC. Figure 2 presents the frequency response curves obtained, whereas Table 3 shows the parameters calculated from them. 4 HFA stands for High Frequency Average. 5 IEC :1983/amd1994 [2]. 6 IEC :2015 [1]. 6
7 Table 3: Parameters of the reference hearing aid measured according to IEC [1]. Reference ES LVA ES LVA 2CC FONIX 2CC Max OSPL90 [db] 123,94 123,61 119,21 119,21 HFA-OSPL90 [db] 120,57 120,65 113,84 114,16 Max-FOG [db] 52,67 52,35 46,43 45,86 HFA-FOG [db] 47,79 48,64 41,49 41,50 f min [Hz] 221,8 212,8 <200 <200 f max [Hz] Analyzing the results from a metrological point of view, all LVA ES curves can be considered consistent with the reference since all lie inside the respective uncertainty limits. Small variations between the different test systems are expected since tolerances are allowed by the standard for various elements of the measurement chain. In the analysis with the 2CC coupler good agreement between LVA 2CC and FONIX 2CC results were also found. Parameters calculated from the tests, presented in Table 3, also confirm the consistency among the results. A small misalignment is observed at the first and fourth peaks ot the curves and may be caused by differences between the length of the tubes used, altering the transfer function and influencing the curve shape. It is also noteworthy that the OSPL90 curve shows smaller discrepancies when compared to the reference curve than the other curves. This is expected since this test is less prone to variations of the input SPL as the FOG measurements. That also means hearing aid positioning errors in the test space will not considerably influence the result. The 2CC coupler has a volume that approximates the adult residual ear canal when a hearing aid is worn [8]. However, it does not provide a good approximation to the ear s acoustic impedance. An ear simulator, on the other hand, mimics the ear s variation of impedance with frequency, therefore gives a better approximation of the ear s response [7]. While the resonances are almost the same for the ear simulator and the 2CC coupler, the later underestimates SPL at high frequencies. The simple geometry of the 2CC coupler makes it cheaper and less susceptible to variation in geometry compared to the ear simulator. The internal configuration of the ear simulator consists of cavities connected by small openings, which can be easily blocked. Transfer functions and correction factors can be found in the literature to transform the 2CC coupler response in a ear simulator response and vice versa [9]. The results obtained by the IEC measurement procedures are a good starting point for clinicians, but to verify the actual hearing aid acoustical characteristics, by performing measurements in each individual ear 7 adapted, is indispensable. Therefore, results obtained with the ear simulator have no clinical advantage compared to the 2CC coupler. Those reasons make the 2CC coupler more suitable to the IEC purposes. 7 These measurements generally use a probe tube microphone to measure the acoustic response inside the ear canal and are known as real-ear measurements. 7
8 SPL [db ref 20µPa] SPL [db ref 20µPa] SPL [db ref 20µPa] OSPL ,000 4,000 8, Frequency [Hz] FOGRC ,000 4,000 8, Frequency [Hz] Basic Frequency Response Curve ,000 4,000 8,000 Frequency [Hz] Reference ES LVA ES LVA 2CC FONIX 2CC Reference Uncertainty Figure 2: Frequency response results. Reference result provided by accredited laboratory using an occluded ear simulator ( ) with respective uncertainty ( ); ( ) results obtained at LVA using an occluded ear simulator; ( ) results obtained at LVA with 2CC coupler; ( ) results obtained with FONIX 8000 and 2CC coupler. 8
9 In order to evaluate relative position sensibility between the HA and the control microphone, three measurements were carried out for three different distances between them: 5 mm, 12 mm and 20 mm. Furthermore, an analysis of how the control microphone presence affected the result was realized by taking the control microphone out of the test after adjusting the sound pressure level at the reference point. After that, a different configuration was evaluated. A free field microphone was used to perform the adjustment of sound pressure level at the reference point, and then was substituted by the HA in the same position. This configuration is referenced as FFM in the graph. The results from these measurements are presented in Figure 3 for the BFR. Basic Frequency Response Curve SPL [db ref 20µPa] ,000 4,000 8,000 Frequency [Hz] 5mm 5mm - Without Control Mic 12mm 12mm Without Control Mic 20mm FFM Figure 3: Basic Frequency Response Investigation varying parameters. From the results presented in Figure 3 it is possible to notice a small deviation between the different control microphone position results. Therefore, it is expected that small errors in the microphone positioning would not have significant impact on the results. Similarly, taking the control microphone out of position after adjusting the input sound pressure level does not alters the acoustic field around the reference point in a way that influences the final result. It is important to note that adjusting the reference input sound pressure level with a free field microphone, and then replacing it with the hearing aid, resulted in slightly different results if compared to the other curves. So, for the FFM case, the acoustic field is significantly altered by the geometry differences and positioning uncertainty ends up being greater than in the other test arrangements. 4 Conclusions The authors evaluated different test set-ups for the eletroacoustic evaluation of hearing aids according to different revisions of the IEC standard, and validated the approaches using reference results issued by an accredited laboratory and a commercial test-box system used for this purpose. Good agreement comparing the results from different measurement 9
10 systems was found. It was observed that the results obtained by the authors are within the uncertainty bounds given by the accredited laboratory. Although the ear simulator gives a better approximation of the SPL at the ear canal, the 2CC coupler delivers a better cost benefit for the IEC measurement purposes. Investigations regarding the relative positioning of the control microphone and hearing aid, showed that this parameters did not represent substantial influence in the results of the hearing aid tested. This is noteworthy because it credits confidence to single microphone systems, that are used both for reference SPL adjustment and hearing aid testing. Moreover, the results showed that taking the control microphone out of position after adjusting the sound pressure level at the reference point, does not affect the result. However further investigation with directional hearing aids is necessary. The test using a free field microphone to adjust the sound pressure level at the reference point without the HA in place presented a larger discrepancy due to a lack of correspondence between the geometries of the hearing aid geometry and the microphone used to adjust the sound pressure level at the reference point. Acknowledgements The Authors would like to Acknowledge the National Council for Scientific and Technological Development (CNPq), the National Council for the Improvement of Higher Education (CAPES) for the financial support under this project and DELTA laboratory for the hearing aid and results provided. References [1] IEC, International Standard IEC :2015 Electroacoustics - Hearing aids - Measurement of the performance characteristics of hearing aids, [2] IEC, International Standard IEC :1983/amd1:1994 Electroacoustics - Hearing aids. Part 0: Measurement of electroacoustical characteristics, [3] IEC, International Standard IEC :2010 Electroacoustics - Simulators of human head and ear - Part 4: Occluded-ear simulator for the measurement of earphones coupled to the ear by means of ear inserts, [4] IEC, International Standard IEC :2006 Electroacoustics - Simulators of human head and ear - Part 5: 2 cm3 coupler for the measurement of hearing aids and earphones coupled to the ear by means of ear inserts, [5] JCGM, Evaluation of measurement data Guide to the expression of uncertainty in measurement, JCGM, [6] Pedroso, M. A.; Aspectos metrológicos da calibração de audiômetros. Master s thesis, Universidade Federal de Santa Catarina, [7] Dillon, H.; Hearing Aids, Boomerang Press, Sydney (Australia), second edition, [8] Romanow, FF.; Methods of measuring the performance of hearing aid, Journal Acoustic Society of America, volume 13(1), pp , [9] Bentler, R. A.; Pavlovic, C. V., Transfer functions and correction factors used in hearing aid evaluation and research. Ear and Hearing, 10(1), 1989, pp
Testing FM Systems on the 7000 Hearing Aid Test System
Testing FM Systems on the 7000 Hearing Aid Test System Introduction Testing FM Systems on the 7000 Hearing Aid Test System This workbook describes how to test FM systems with the 7000 Hearing Aid Test
More informationFM SYSTEMS. with the FONIX 6500-CX Hearing Aid Analyzer. (Requires software version 4.20 or above) FRYE ELECTRONICS, INC.
T E S T I N G FM SYSTEMS with the FONIX 6500-CX Hearing Aid Analyzer (Requires software version 4.20 or above) FRYE FRYE ELECTRONICS, INC. P.O. Box 23391 Tigard, OR 97281-3391 (503) 620-2722 (800) 547-8209
More informationOIML R 122 Annex C RECOMMENDATION. Edition 1999 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION
INTERNATIONAL RECOMMENDATION OIML R 122 Annex C Edition 1999 (E) Equipment for speech audiometry Annex C Test report format Appareils pour l audiométrie vocale Annexe C Format du rapport d essai OIML R
More information보청기의전기음향적성능분석. Kyoung Won Lee, Ph.D Hallym University of Graduate Studies
보청기의전기음향적성능분석 Kyoung Won Lee, Ph.D Hallym University of Graduate Studies 차례 I. 성능분석이필요한이유 II. 성능분석에필요한장비의구성 III. 성능분석시구성품의위치 IV. 성능분석의기준 V. 성능분석의내용 VI. 성능분석의절차 I. 성능분석이필요한이유 보청기의품질관리보청기의적합 - frequency
More informationGeneral about Calibration and Service on Audiometers and Impedance Instruments
General about Calibration and Service on Audiometers and Impedance Instruments Calibration and Service of Audiometers - page 1 of 14 Rev.4 2008-05-21/JF General about Calibration and Service on Audiometers
More informationTwo Modified IEC Ear Simulators for Extended Dynamic Range
Two Modified IEC 60318-4 Ear Simulators for Extended Dynamic Range Peter Wulf-Andersen & Morten Wille The international standard IEC 60318-4 specifies an occluded ear simulator, often referred to as a
More informationFeng Xiujuan National Institute of Metrology (NIM),China
The acoustic calibration service in transportation at NIM Feng Xiujuan National Institute of Metrology (NIM),China 1. Calibration requirements 2. Calibration service at NIM 2.1 Microphone 2.2 Ultrasonic
More informationWelcome to the LISTEN G.R.A.S. Headphone and Headset Measurement Seminar The challenge of testing today s headphones USA
Welcome to the LISTEN G.R.A.S. Headphone and Headset Measurement Seminar The challenge of testing today s headphones USA 2017-10 Presenter Peter Wulf-Andersen Engineering degree in Acoustics Co-founder
More informationTesting FM Systems with FONIX FP40 Analyzers
Testing FM Systems with FONIX FP40 Analyzers FRYE FRYE ELECTRONICS, INC. P.O. Box 23391 Tigard, OR 97281-3391 USA (503) 620-2722 (800) 547-8209 Fax: (503) 639-0128 http://www.frye.com e-mail: general@frye.com
More informationINTRODUCTION TO PURE (AUDIOMETER & TESTING ENVIRONMENT) TONE AUDIOMETERY. By Mrs. Wedad Alhudaib with many thanks to Mrs.
INTRODUCTION TO PURE TONE AUDIOMETERY (AUDIOMETER & TESTING ENVIRONMENT) By Mrs. Wedad Alhudaib with many thanks to Mrs. Tahani Alothman Topics : This lecture will incorporate both theoretical information
More informationCOPYRIGHT Danish Standards. NOT FOR COMMERCIAL USE OR REPRODUCTION. DS/EN ISO :1998
INTERNATIONAL STANDARD IS0 8253-l First edition 1989-11-15 COPYRIGHT Danish Standards. NOT FOR COMMERCIAL USE OR REPRODUCTION. DS/EN ISO 8253-1:1998 Acoustics - Audiometric test methods - Part I: Basic
More informationPercentiles Study Group. Elk test for percentile analysis
Percentiles Study Group Elk test for percentile analysis Percentiles Study Group: Beate Gromke (Chairwoman), EUHA, Leipzig Martin Blecker, EUHA, Hanover Harald Bonsel, EUHA, Reinheim Dr.-Ing. Josef Chalupper,
More informationNOTICE. (Formulated under the cognizance of the CTA R6 Portable, Handheld and In-Vehicle Electronics Committee.)
ANSI/CTA Standard Personal Sound Amplification Performance Criteria ANSI/CTA-2051 January 2017 NOTICE Consumer Technology Association (CTA) Standards, Bulletins and other technical publications are designed
More informationThe Effect of Analysis Methods and Input Signal Characteristics on Hearing Aid Measurements
The Effect of Analysis Methods and Input Signal Characteristics on Hearing Aid Measurements By: Kristina Frye Section 1: Common Source Types FONIX analyzers contain two main signal types: Puretone and
More informationImportance of a Good Start. Topics. Rationale. In-Situ Audiometry. In-Situ Audiometry. Enhancing the Initial
Enhancing the Initial Hearing Aid Fitting Michael Block, Ph.D. Session 0228 Importance of a Good Start The consequences of doing a poor job of selling the hearing-impaired client on a positive treatment
More informationRECD-Measurements with Connexx 6.3 and Unity 2
RECD-Measurements with Connexx 6.3 and Unity 2 Vera Spitzlei, Corporate Audiology Siemens Audiologische Technik GmbH For internal use only / Copyright Siemens AG 2009. All rights reserved. Overview Why
More informationQuick Guide Binaural REM
Quick Guide Binaural REM The purpose of this document is to provide a quick guide for the Binaural REM feature found in the REM440 Real-Ear Measurement module in the Affinity 2.0 and Callisto Suites. This
More informationISO INTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 11904-1 First edition 2002-10-01 Acoustics Determination of sound immission from sound sources placed close to the ear Part 1: Technique using a microphone in a real ear (MIRE
More informationbinax fit: How it s simulated in Connexx, how to verify it, and how to match-to-target the easy way
www.bestsound-technology.com binax fit: How it s simulated in Connexx, how to verify it, and how to match-to-target the easy way Richard Schultz-Amling Sivantos, 2015. 1 Introduction Since the release
More informationHEARING CONSERVATION PROGRAM
CALIFORNIA STATE UNIVERSITY, CHICO HEARING CONSERVATION PROGRAM PREPARED BY THE OFFICE OF ENVIRONMENTAL HEALTH AND SAFETY REVISED June 2008 TABLE OF CONTENTS Section Page 1.0 Introduction... 1-1 2.0 Exposure
More informationISO INTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 389-8 First edition 2004-05-15 Acoustics Reference zero for the calibration of audiometric equipment Part 8: Reference equivalent threshold sound pressure levels for pure tones
More informationStep-by-Step RECD Guide
Precision Audiometric Instruments www.medrx-usa.com Step-by-Step RECD Guide The RECD task involves 4 steps: 1 - Complete Calibration of the Speakers and Probe Tube 2 - Measure an Ear Response 3 - Perform
More information3M Center for Hearing Conservation
3M Center for Hearing Conservation Key Terms in Occupational Hearing Conservation Absorption A noise control method featuring sound-absorbing materials that are placed in an area to reduce the reflection
More informationHearing Solutions Catalog Winter Supplement January 2009
Hearing Solutions Catalog Winter Supplement January 2009 LX Virtue The following information is a supplement containing information for products released January 1, 2009. It should be used along with
More informationProceedings of Meetings on Acoustics
Proceedings of Meetings on Acoustics Volume 19, 2013 http://acousticalsociety.org/ ICA 2013 Montreal Montreal, Canada 2-7 June 2013 Noise Session 1aNS: Advanced Hearing Protection and Methods of Measurement
More informationACOUSTIC INFRASOUND AND LOW-FREQUENCY SOUND
116 Acoustic ACOUSTIC INFRASOUND AND LOW-FREQUENCY SOUND These limits represent sound exposures to which it is believed nearly all workers may be repeatedly exposed without adverse effects that do not
More informationOrganisation Internationale de Métrologie Légale
Organisation Internationale de Métrologie Légale OIML INTERNATIONAL RECOMMENDATION Equipment for speech audiometry Appareils pour l audiométrie vocale OIML R 122 Edition 1996 (E) CONTENTS Foreword... 3
More informationNOAH Sound Equipment Guideline
NOAH Sound Equipment Guideline Version 1.0 Disclaimer The information contained in this document is subject to change without notice. HIMSA A/S MAKES NO WARRANTY OF ANY KIND WITH REGARD TO THIS MATERIAL,
More informationDigital hearing aids are still
Testing Digital Hearing Instruments: The Basics Tips and advice for testing and fitting DSP hearing instruments Unfortunately, the conception that DSP instruments cannot be properly tested has been projected
More informationCITY OF FORT BRAGG HEARING CONSERVATION PROGRAM
CITY OF FORT BRAGG HEARING CONSERVATION PROGRAM A. PURPOSE It is the City s policy to protect the health and safety of employees through the establishment and enforcement of this Hearing Conservation Program
More informationProcedure Number 310 TVA Safety Procedure Page 1 of 6 Hearing Conservation Revision 0 January 6, 2003
Procedure Number 310 TVA Safety Procedure Page 1 of 6 Hearing Conservation Revision 0 January 6, 2003 1. Purpose 1.1. The purpose of this procedure is to establish a TVA Hearing Conservation Program (HCP)
More informationISO INTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 389-7 Second edition 2005-11-01 Acoustics Reference zero for the calibration of audiometric equipment Part 7: Reference threshold of hearing under free-field and diffuse-field
More informationDigital. hearing instruments have burst on the
Testing Digital and Analog Hearing Instruments: Processing Time Delays and Phase Measurements A look at potential side effects and ways of measuring them by George J. Frye Digital. hearing instruments
More informationSterilization of health care products Radiation. Part 3: Guidance on dosimetric aspects of development, validation and routine control
Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 11137-3 Second edition 2017-06 Sterilization of health care products Radiation Part 3: Guidance on dosimetric aspects of development, validation
More informationTitle: Evaluation of the Kuduwave 5000 audiometer for compliance with standards for hearing conservation purposes
Title: Evaluation of the Kuduwave 5000 audiometer for compliance with standards for hearing conservation purposes Author: AL Edwards Affiliation: Council for Scientific and Industrial Research, Centre
More informationAURICAL Plus with DSL v. 5.0b Quick Guide. Doc no /04
AURICAL Plus with DSL v. 5.0b Quick Guide 0459 Doc no. 7-50-0900/04 Copyright notice No part of this Manual or program may be reproduced, stored in a retrieval system, or transmitted, in any form or by
More informationTRAINING MANUAL AVANT HIT+ Hearing Instrument Test Chamber.
TRAINING MANUAL AVANT Hearing Instrument Test Chamber HIT+ www.medrx-int.com Contents Getting to know your HIT+... 3 AVANT HIT Software Overview.. 4 AVANT HIT Software Options.... 5 Preparing for Testing...
More informationSimulating cartilage conduction sound to estimate the sound pressure level in
*Manuscript Click here to download Manuscript: JSV_Cartilage conduction simulation_resubmission_text5.doc Click here to view linked References Original paper Simulating cartilage conduction sound to estimate
More informationSanta Clarita Community College District HEARING CONSERVATION PROGRAM. Revised
Santa Clarita Community College District HEARING CONSERVATION PROGRAM Revised March 2018 TABLE OF CONTENTS INTRODUCTION...1 HEARING CONSERVATION PROGRAM...2 1.1 District Policy...2 1.2 Plan Review...2
More informationIssues faced by people with a Sensorineural Hearing Loss
Issues faced by people with a Sensorineural Hearing Loss Issues faced by people with a Sensorineural Hearing Loss 1. Decreased Audibility 2. Decreased Dynamic Range 3. Decreased Frequency Resolution 4.
More informationJapan Suggestions for AVAS sound requirements JASIC
Japan Suggestions for AVAS sound requirements JASIC 1 1. Outline Japan suggestions for the sound requirements for AVAS: 1.At least two bands with adequate sound pressure levels in 1/3 octave band, 2.Audible
More informationHow to use AutoFit (IMC2) How to use AutoFit (IMC2)
How to use AutoFit (IMC2) 1 AutoFit is a beneficial feature in the Connexx Fitting Application that automatically provides the Hearing Care Professional (HCP) with an optimized real-ear insertion gain
More informationJacob Sulkers M.Cl.Sc (AUD) Candidate University of Western Ontario: School of Communication Sciences and Disorders
Critical Review: The (Sound) Wave of the Future: Is Forward Pressure Level More Accurate than Sound Pressure Level in Defining In Situ Sound Levels for Hearing Aid Fitting? Jacob Sulkers M.Cl.Sc (AUD)
More informationInternational Journal of Health Sciences and Research ISSN:
International Journal of Health Sciences and Research www.ijhsr.org ISSN: 2249-9571 Original Research Article Effect of Compression Parameters on the Gain for Kannada Sentence, ISTS and Non-Speech Signals
More informationELECTROACOUSTIC EVALUATION OF THE RESOUND UNITE MINI MICROPHONE WITH OTOMETRICS AURICAL HIT
ELECTROACOUSTIC EVALUATION OF THE RESOUND UNITE MINI MICROPHONE WITH OTOMETRICS AURICAL HIT Astrid Haastrup, GN ReSound Mona Dworsack-Dodge, AuD, GN Otometrics Abstract With ReSound s 2.4 GHz wireless
More informationAUDL GS08 and GAV1: 2013 Final exam page 1/13. You must complete all sections. Label all graphs. Show your work!
AUDL GS08 and GAV1: 2013 Final exam page 1/13 You must complete all sections. Label all graphs. Show your work! Section A: Short questions concerning Signals & Systems A1. Give the sound pressure levels
More informationFitting Decisions and their Impact on Hearing Aid User Benefit. Mallory Maine, AuD Audiologist, GN ReSound
Fitting Decisions and their Impact on Hearing Aid User Benefit Mallory Maine, AuD Audiologist, GN ReSound Agenda Common Fitting Oversights #1 Setting the coupler type in fitting software To set or not
More informationThe Use of a High Frequency Emphasis Microphone for Musicians Published on Monday, 09 February :50
The Use of a High Frequency Emphasis Microphone for Musicians Published on Monday, 09 February 2009 09:50 The HF microphone as a low-tech solution for performing musicians and "ultra-audiophiles" Of the
More informationHearing Conservation Program
Hearing Conservation Program Table of Contents I. Program Goals and Objectives... 2 II. Scope and Application... 2 III. Regulatory Authority and Related Information... 2 IV. Definitions... 2 V. Responsibilities...
More informationUser Guide. For Jacoti Hearing Center Version 1.1. Manufacture Year 2016
User Guide For Jacoti Hearing Center Version 1.1 Manufacture Year 2016 Revision 11/ 07 / 2016 Table of contents 1_ Intended Use... 3 1.1_ USA... 3 1.2_ Requirements... 3 1.3_ Headphones... 3 1.4_ Startup...
More informationTesting Digital Hearing Aids
Testing Digital Hearing Aids with the FONIX 6500-CX Hearing Aid Analyzer Frye Electronics, Inc. Introduction The following is a quick guide for testing digital hearing aids using the FONIX 6500-CX. All
More informationWeek 2 Systems (& a bit more about db)
AUDL Signals & Systems for Speech & Hearing Reminder: signals as waveforms A graph of the instantaneousvalue of amplitude over time x-axis is always time (s, ms, µs) y-axis always a linear instantaneousamplitude
More informationHearing Protection Systems
March 2018 ANR, AHNR, In-Ear How to Choose? Simon Smith Chief Engineer Gentex Europe ssmith@gentexcorp.eu.com +44 (0)1462 478000 Michelle Collier Product Manager Aircrew Hearing Protection & Communications
More informationVerification of soft speech amplification in hearing aid fitting: A comparison of methods
Verification of soft speech amplification in hearing aid fitting: A comparison of methods Sarah E. Dawkins, B.A. AuD Research Project April 5, 2007 University of Memphis Project Advisor Robyn M. Cox, PhD.
More informationThe registration testing for audiometric device in CFDA
The registration testing for audiometric device in CFDA Beijing Center for Medical Devicee Quality Supervision and Testing of China Food and Drug Administration Beijing Institute of Medical Device Testing
More informationDiagnostic Audiometer
AD629 Technical Specificatons Page 0 Technical Specifications AD629 Diagnostic Audiometer Item No.8107429 3/2015 AD629 Technical Specificatons Page 1 License Overview International configuration AD629
More informationPhysikalisch-Technische Bundesanstalt
Physikalisch-Technische Bundesanstalt Short Report for CCAUV, September2017 1. Organisation PTB is the National Metrology Institute of Germany and the highest technical authority for the field of metrology
More informationC HAPTER FOUR. Audiometric Configurations in Children. Andrea L. Pittman. Introduction. Methods
C HAPTER FOUR Audiometric Configurations in Children Andrea L. Pittman Introduction Recent studies suggest that the amplification needs of children and adults differ due to differences in perceptual ability.
More informationFONIX FP35. Ease and Accuracy with the FP35 Touch Hearing Aid Analyzer! Visible Speech
FONIX FP35 Ease and Accuracy with the FP35 Touch Hearing Aid Analyzer! Move your hearing aid fittings into the world of the touch screen! The new FP35 Touch adds a new interactive dimension to hearing
More informationTympanometry and Reflectance in the Hearing Clinic. Presenters: Dr. Robert Withnell Dr. Sheena Tatem
Tympanometry and Reflectance in the Hearing Clinic Presenters: Dr. Robert Withnell Dr. Sheena Tatem Abstract Accurate assessment of middle ear function is important for appropriate management of hearing
More informationElements of Effective Hearing Aid Performance (2004) Edgar Villchur Feb 2004 HearingOnline
Elements of Effective Hearing Aid Performance (2004) Edgar Villchur Feb 2004 HearingOnline To the hearing-impaired listener the fidelity of a hearing aid is not fidelity to the input sound but fidelity
More informationSignals, systems, acoustics and the ear. Week 1. Laboratory session: Measuring thresholds
Signals, systems, acoustics and the ear Week 1 Laboratory session: Measuring thresholds What s the most commonly used piece of electronic equipment in the audiological clinic? The Audiometer And what is
More informationipod Noise Exposure Assessment in Simulated Environmental Conditions
ipod Noise Exposure Assessment in Simulated Environmental Conditions Kyle N. Acker Advised by: Robert Novak Michael Heinz Background Since the 80s and the invention of the personal audio player, there
More informationINSERT EARPHONES. User Guide
TM INSERT EARPHONES User Guide INTRODUCTION AND INTENDED USE ER-3C earphones are the next generation of insert earphones by Etymotic Research, designed to replace Etymotic s ER-3A Tubephone earphones,
More informationUNITY 2. a state-of-the-art audiological integrated diagnostic and fitting system
The information in this document contains general descriptions of the technical options available, which do not always have to be present in individual cases and are subject to change without prior notice.
More informationPercentiles Study Group. The practice of calibrating test boxes
Percentiles Study Group The practice of calibrating test boxes Percentiles Study Group: Beate Gromke (Chairwoman), EUHA, Leipzig Martin Blecker, EUHA, Hanover Harald Bonsel, EUHA, Reinheim Dr.-Ing. Josef
More informationAND9020/D. Adaptive Feedback Cancellation 3 from ON Semiconductor APPLICATION NOTE INTRODUCTION
Adaptive Feedback Cancellation 3 from ON Semiconductor APPLICATION NOTE INTRODUCTION This information note describes the feedback cancellation feature provided in ON Semiconductor s latest digital hearing
More informationTechnical Data Insio micon TM
Technical Data Insio micon TM 7mi 5mi CIC ITC - HS ITE 113 / 40 113 / 40 118 / 55 51 db / 124 (ear simulator) 40 db / 113 (2 ccm coupler) 50 db / 124 (ear simulator) 40 db / 113 (2 ccm coupler) 66 db /
More information#0086.
USER MANUAL Contents Introduction... 3 AVANT POLAR HIT Software Overview.. 4 AVANT POLAR HIT Software Options. 5 Preparing for Testing.. 7 Rotational Table Control.. 10 Making HIT Measurements 11 Sample
More informationEEL 6586, Project - Hearing Aids algorithms
EEL 6586, Project - Hearing Aids algorithms 1 Yan Yang, Jiang Lu, and Ming Xue I. PROBLEM STATEMENT We studied hearing loss algorithms in this project. As the conductive hearing loss is due to sound conducting
More informationImpact of the ambient sound level on the system's measurements CAPA
Impact of the ambient sound level on the system's measurements CAPA Jean Sébastien Niel December 212 CAPA is software used for the monitoring of the Attenuation of hearing protectors. This study will investigate
More informationTechnical Data Nitro micon TM
Technical Data Nitro micon TM 7mi 3mi Earhook 85 db / 144 (ear simulator) 82 db / (2 ccm coupler) Key eatures Super power Super connectivity micon BestSound TM Technology Data Sheet www.siemens.com/hearing
More informationTechnical Data. Orion RIC. usa.siemens.com/hearing. Life sounds brilliant. Description. Standard Features
usa.siemens.com/hearing Technical Data Orion RIC S-Receiver M-Receiver P-Receiver HP-Receiver 2 2 2 2 4 4 4 4 6 6 6 6 Description Very small Receiver-in Canal (RIC) hearing system appropriate or mild,
More informationpep Step up with pep Pep Specification Guide _Pep_Specification_Guide_UK.indd 1
pep Step up with pep Pep Specification Guide 124843_Pep_Specification_Guide_UK.indd 1 20.06.12 14:14 What is pep? It s that energetic feeling you get when life is great and you are ready to take on the
More informationINTERNATIONAL STANDARD
INTERNATIONAL STANDARD IEC 60601-2-37 2001 AMENDMENT 1 2004-08 Amendment 1 Medical electrical equipment Part 2-37: Particular requirements for the safety of ultrasonic medical diagnostic and monitoring
More informationinter.noise 2000 The 29th International Congress and Exhibition on Noise Control Engineering August 2000, Nice, FRANCE
Copyright SFA - InterNoise 2000 1 inter.noise 2000 The 29th International Congress and Exhibition on Noise Control Engineering 27-30 August 2000, Nice, FRANCE I-INCE Classification: 6.3 PSYCHOLOGICAL EVALUATION
More informationSOLUTIONS Homework #3. Introduction to Engineering in Medicine and Biology ECEN 1001 Due Tues. 9/30/03
SOLUTIONS Homework #3 Introduction to Engineering in Medicine and Biology ECEN 1001 Due Tues. 9/30/03 Problem 1: a) Where in the cochlea would you say the process of "fourier decomposition" of the incoming
More informationTesting Digital Hearing Aids
Testing Digital Hearing Aids with the FONIX FP40 Hearing Aid Analyzer FRYE 7 Introduction The following is a quick guide for testing digital hearing aids. All digital aids can be tested, but some of the
More informationIEC/TC 29 Electroacoustics. Status: September 2010
IEC/TC 29 Electroacoustics Status: September 2010 Secretary: Leif Nielsen Danish Standards Association Kollegievej 6 DK-2920 Charlottenlund Denmark Email address: len@ds.dk Chairman: Susan Dowson National
More informationPhysikalisch-Technische Bundesanstalt
CCAUV/15-05 Status Report PTB Physikalisch-Technische Bundesanstalt Short Report for CCAUV, October 2015 1. Organisation PTB is the National Metrology Institute of Germany and the highest technical authority
More informationDifferences in Sensation Level between the Widex SoundTracker and Two Real-Ear Analyzers DOI: /jaaa
J Am Acad Audiol 24:660 670 (2013) Differences in Sensation Level between the Widex SoundTracker and Two Real-Ear Analyzers DOI: 10.3766/jaaa.24.8.3 Kristi Oeding* Michael Valente* Abstract Background:
More informationTop 10 ideer til en god høreapparat tilpasning. Astrid Haastrup, Audiologist GN ReSound
HVEM er HVEM Top 10 ideer til en god høreapparat tilpasning Astrid Haastrup, Audiologist GN ReSound HVEM er HVEM 314363 WRITE DOWN YOUR TOP THREE NUMBER 1 Performing appropriate hearing assessment Thorough
More informationImplants for surgery Active implantable medical devices. Part 7: Particular requirements for cochlear implant systems
Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 14708-7 First edition 2013-01-15 Implants for surgery Active implantable medical devices Part 7: Particular requirements for cochlear implant systems
More informationISO INTERNATIONAL STANDARD. Acoustics Audiometric test methods Part 2: Sound field audiometry with pure-tone and narrow-band test signals
INTERNATIONAL STANDARD ISO 8253-2 Second edition 2009-12-15 Acoustics Audiometric test methods Part 2: Sound field audiometry with pure-tone and narrow-band test signals Acoustique Méthodes d'essais audiométriques
More informationNoise reduction in modern hearing aids long-term average gain measurements using speech
Noise reduction in modern hearing aids long-term average gain measurements using speech Karolina Smeds, Niklas Bergman, Sofia Hertzman, and Torbjörn Nyman Widex A/S, ORCA Europe, Maria Bangata 4, SE-118
More informationEfficiency of attenuating high-level acoustic impulses with double protection (earplugs and earmuffs used simultaneously)
Efficiency of attenuating high-level acoustic impulses with double protection (earplugs and earmuffs used simultaneously) J. ŻERA 1,2, R. MŁYŃSKI 1 1 Central Institute for Labour Protection - National
More informationMethods of validation of occupational noise exposure measurement with multi aspect personal sound exposure meter
Proceedings of ACOUSTICS 2016 Methods of validation of occupational noise exposure measurement with multi aspect personal sound exposure meter Marek Podgórski Svantek Sp. z o.o., Warsaw, Poland ABSTRACT
More informationProceedings of Meetings on Acoustics
Proceedings of Meetings on Acoustics Volume 19, 213 http://acousticalsociety.org/ ICA 213 Montreal Montreal, Canada 2-7 June 213 Engineering Acoustics Session 4pEAa: Sound Field Control in the Ear Canal
More informationIS INTERNATIONAL STANDARD
INTERNATIONAL STANDARD IS0 4869-2 First edition 1994-09-O 1 Acoustics - Hearing protectors - Part 2: Estimation of effective A-weighted sound pressure levels when hearing protectors are worn Acoustique
More informationAcoustics Estimation of noiseinduced
Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO 1999 Third edition 2013-10-01 Acoustics Estimation of noiseinduced hearing loss Acoustique Estimation de la perte auditive induite par le bruit
More informationHearing threshold measurements of infrasound combined with audio frequency sound
12th ICBEN Congress on Noise as a Public Health Problem Hearing threshold measurements of infrasound combined with audio frequency sound Elisa Burke 1, Johannes Hensel 1, Thomas Fedtke 1 1 Physikalisch-Technische
More informationinter.noise 2000 The 29th International Congress and Exhibition on Noise Control Engineering August 2000, Nice, FRANCE
1 inter.noise 2000 The 29th International Congress and Exhibition on Noise Control Engineering 27-30 August 2000, Nice, FRANCE I-INCE Classification: 3.6 FINITE ELEMENT METHOD (FEM) MODEL FOR THE HEARING
More informationComparing Speech Perception Abilities of Children with Cochlear Implants and Digital Hearing Aids
Comparing Speech Perception Abilities of Children with Cochlear Implants and Digital Hearing Aids Lisa S. Davidson, PhD CID at Washington University St.Louis, Missouri Acknowledgements Support for this
More informationNoise at Work Regulations. Mick Gray MRSC, LFOH, ROH. MWG Associates Ltd
Noise at Work Regulations Mick Gray MRSC, LFOH, ROH. MWG Associates Ltd The Issue NIHL is a significant occupational disease 170,000 people in the UK suffer deafness, tinnitus or other ear conditions as
More informationEUHA Guideline Wireless remote microphone systems configuration, verification and measurement of individual benefit
EUHA Guideline Wireless remote microphone systems configuration, verification and measurement of individual benefit Guideline 04-06 - v1.0 - issued 9 May 2017 Audiology Expert Group Beate Gromke (Chairwoman),
More informationUsing Multi-Curve in Real-Ear Measurements on the 6500-CX Hearing Aid Analyzer
Using Multi-Curve in Real-Ear Measurements on the 6500-CX Hearing Aid Analyzer FRYE FRYE ELECTRONICS, INC. P.O. Box 23391 Tigard, OR 97281-3391 (503) 620-2722 (800) 547-8209 Fax: (503) 639-0128 www.frye.com
More informationTechnical Topics. Where did my gain go? Thin Tube Open Fit BTE Verification. History of Thin Tubes: Summary of presentation by C. Staples and S.
Technical Topics Where did my gain go? Thin Tube Open Fit BTE Verification Summary of presentation by C. Staples and S. Aiken History of Thin Tubes: Growth of new BTE market: First quarter of 2006 market:
More information