Article. Reference. Light-curing time reduction with a new high-power halogen lamp. STAUDT, Christine Bettina, et al.

Size: px
Start display at page:

Download "Article. Reference. Light-curing time reduction with a new high-power halogen lamp. STAUDT, Christine Bettina, et al."

Transcription

1 Article Light-curing time reduction with a new high-power halogen lamp STAUDT, Christine Bettina, et al. Abstract Orthodontic brackets are routinely bonded with light-cured adhesives. Conventional halogen lights used in bonding have the disadvantage of a long curing time, and the available alternatives (laser and plasma lights) are expensive. Our aim was to investigate the minimum time necessary to bond brackets with a new, relatively low-priced, high-power halogen light. Reference STAUDT, Christine Bettina, et al. Light-curing time reduction with a new high-power halogen lamp. American Journal of Orthodontics and Dentofacial Orthopedics, 2005, vol. 128, no. 6, p DOI : /j.ajodo PMID : Available at: Disclaimer: layout of this document may differ from the published version.

2 ORIGINAL ARTICLE Light-curing time reduction with a new high-power halogen lamp Christine Bettina Staudt, a Anestis Mavropoulos, a Serge Bouillaguet, b Stavros Kiliaridis, c and Ivo Krejci d Geneva, Switzerland Introduction: Orthodontic brackets are routinely bonded with light-cured adhesives. Conventional halogen lights used in bonding have the disadvantage of a long curing time, and the available alternatives (laser and plasma lights) are expensive. Our aim was to investigate the minimum time necessary to bond brackets with a new, relatively low-priced, high-power halogen light. Methods: Five groups of 15 deciduous bovine incisors were bonded with stainless steel brackets (Mini Diamond Twin, Ormco, Orange, Calif) by using different lamps and curing times. Three of the groups were bonded by using a high-power halogen light (Swiss Master Light, Electro Medical Systems, Nyon, Switzerland) for 2, 3, and 6 seconds, respectively. The fourth group, bonded with a fast halogen light (Optilux 501, Sybron Dental Specialties, Danbury, Conn) for 40 seconds, served as the positive control group. The fifth group, the comparison group, was bonded with a plasma light (Remecure, Remedent, Deurle, Belgium) for 4 seconds. After storage for 24 hours in the dark at 37 C in water, shear bond strength was measured with a universal testing machine. Results: A curing time of 2 seconds with the high-power halogen light negatively affected the bond strength and the probability of bond survival. The adhesive remnant index scores were not significantly different among the groups. Most failures ( 60%) occurred at the bracket base/adhesive interface. Conclusions: The high-power halogen light seems to be a cost-effective solution to reducing curing time. The recommended curing times to bond stainless steel brackets are 6 seconds and, with caution, even 3 seconds. (Am J Orthod Dentofacial Orthop 2005;128:749-54) From the University of Geneva, Geneva, Switzerland. a Research associate, Department of Orthodontics. b Lecturer, Department of Cariology and Endodontology. c Professor and chair, Department of Orthodontics. d Professor and chair, Department of Cariology and Endodontology. Reprint requests to: Dr Christine Bettina Staudt, Faculté de Médecine, Section de Médecine Dentaire, Division d Orthodontie, 19, rue Barthélemy-Menn, CH-1205 Geneva, Switzerland; , christine.staudt@medecine.unige.ch. Submitted, June 2004; revised and accepted, August /$30.00 Copyright 2005 by the American Association of Orthodontists. doi: /j.ajodo Orthodontic brackets are routinely bonded by using chemical or light-curing adhesive systems. With light-curing materials, the clinician has sufficient time to position the brackets accurately and initiate polymerization when ready. The disadvantage, however, of conventional halogen lights is the long curing time required for each bracket (up to 40 seconds); this wastes valuable clinical time. The curing efficiency of conventional halogen lights is limited, because 98% of their radiation does not contribute to polymerization but is lost as heat. 1 In addition, only part of the halogen light spectrum is useful because the absorption spectrum of camphorquinone is comparatively narrow. 2 To reduce curing time, light sources with higher power densities have been introduced. Argon lasers can achieve bond strengths comparable to the 40 seconds of light-curing with conventional devices in only 10 seconds 3-6 or even 5 seconds. 7 Unfortunately, argon lasers are larger and more expensive than the conventional devices. Xenon plasma lights are less expensive than argon laser lights, but they are 3 times more expensive than halogen lights. 8 Conflicting information exists concerning curing time for xenon plasma lights. Some authors recommend short exposure times of 2 or 3 seconds Others suggest longer exposure times from 6 to 9 seconds 2,12-14 to avoid compromising the mechanical properties or eluting of monomers. 15 Light-emitting diode (LED) units vary widely in their power density. The LEDs evaluated until now for bonding orthodontic brackets have relatively low power densities with irradiation times of 20 to 40 seconds. 8,16-18 These irradiation times are similar to halogen lights. Only 1 study describes a curing time of 10 seconds as sufficient for a specific LED, 17 but this low curing time does not agree with findings that 20 seconds are necessary with this lamp. 18 LED units are relatively inexpensive. No information is available on the new generation of high intensity LEDs. The light-curing devices investigated in other studies have either a long curing time (conventional halogen lights, low power density LEDs) or a comparatively high price (laser, plasma-lights). 749

3 750 Staudt et al American Journal of Orthodontics and Dentofacial Orthopedics December 2005 Table I. Technical characteristics of 3 curing lights Curing device Type Power density Wavelength Tip diameter Swiss Master Light High-power halogen 3000 mw/cm nm 11 mm Optilux 501 Fast halogen 1000 mw/cm nm 8 mm Remecure Plasma 1600 mw/cm nm 5.5 mm The aim of this study was to evaluate a relatively low-cost, high-power halogen light (Swiss Master Light, no. M 00042, Electro Medical Systems SA, Nyon, Switzerland) to estimate the minimum exposure time necessary to achieve bond strength equivalent to the bond strength obtained with a fast halogen light. These results were also compared with a plasma light, which until now has been the most advanced device available for curing time and price. Our null hypothesis was that the shear bond strength obtained with the new high-power halogen light for short irradiation periods would be equivalent to a fast halogen light with a relatively long curing time. MATERIAL AND METHODS Seventy-five deciduous bovine incisors (extracted about age 18 months) were divided into 5 groups of 15 teeth. They were stored in 0.2% thymol for 2 to 3 months. The root of each tooth was cut off, and the crown was polished for 15 seconds with a handpiecebrush by using water and a fluoride-free pumice (Bimsstein Pulver, Prochimie, Avenches, Switzerland). Each tooth was then rinsed thoroughly under tap water for 20 seconds. The teeth were dried with oil-free air for 5 seconds and etched on the buccal side for 30 seconds with 35% phosphoric acid gel (Transbond XT etching gel, no , 3M Unitek, Monrovia, Calif). Each tooth was then rinsed under tap water for 30 seconds and dried again with oil-free air for 5 seconds, followed by application of the primer (Transbond XT Light Cure Adhesive Primer, no , 3M Unitek). Composite (Transbond XT Light Cure Paste, no , 3M Unitek) was applied to the bracket-base (stainless steel brackets: Mini Diamond Twin, for the maxillary right central incisor, no , Ormco, Orange Calif). The bracket was then firmly placed onto the flattest area of the crown with respect to the occlusal/gingival side. Excess composite was carefully removed. Light-curing was performed with the following devices and curing times: Three groups were bonded with a high-power halogen light (Swiss Master Light) in the fast-cure mode. In the first 2 groups, curing times of 2 and 3 seconds were used. The light tip was positioned on the middle of the bracket during each exposure. For the third group, the total curing time was 6 seconds. Because the manufacturer suggested not more than 4 seconds of curing time at once, the time of 6 seconds was divided into 3 seconds on the mesial side and 3 seconds on the distal side of the bracket. The fourth group was bonded with a fast halogen light (Optilux 501 with turbo light guide, serial no , Sybron Dental Specialities, Danbury, Conn) for 20 seconds on the mesial side and 20 seconds on the distal side of the bracket (40 seconds total). This group served as a positive control. The fifth group, the comparison group, was bonded with a xenon plasma arc curing light (Remecure, model CL-15E, serial no. CL15E-A0100, Remedent, Deurle, Belgium) for 2 seconds on the mesial side and 2 seconds on the distal side of the bracket. The total curing time of 4 seconds was according to the light s operation instructions. The technical characteristics of these curing lights are described in Table I. We made sure that the polymerization of the composite was not affected by the position of the lightguiding tip with respect to the bracket. The large tip of the high-power halogen lamp during the 1-shot light exposure covered the bracket completely. Throughout the mesial and distal light exposures, the tip overlapped the middle of the bracket (about 1 mm) to resemble the coverage of the high-power halogen lamp. Therefore, even with the smallest light-guiding tip diameter, the 2 radii of light it produced covered all areas around the bracket. Furthermore, the tip was held at an angle of 90 to the bracket to direct the light at the same angle on each occasion. The shear bond strength was measured with a universal testing machine (Model 1114, serial no. H-1377, Instron Corp, Canton, Mass) (Fig 1). A rectangular rod transmitted the shear force on the occlusal edge of the bracket base until the bracket failed. The following procedure of standardization was developed to ensure that the bracket base would be parallel to the force-exerting rod during bond strength testing. 9,19 Each bonded specimen was embedded in a rectangular acrylic block (Technovit 4071, Lot , Ch.-B , Heraeus Kulzer, Wehrheim/Taunus, Germany). For this purpose, a metallic cubic blade (2

4 American Journal of Orthodontics and Dentofacial Orthopedics Volume 128, Number 6 Staudt et al 751 tooth; 3, all adhesive left on the tooth with a distinct impression of the bracket mesh. Fig 1. Assembly for shear bond strength testing: acrylic block with bonded tooth secured onto lower jaw of testing machine. Force-exerting rod moved along axis parallel to tooth-bracket interface, and its extremity was parallel to occlusal border of bracket base. 2 mm) was fixed into the vertical slot of the bracket to ensure that the tooth-bracket interface was parallel to and projecting slightly above the acrylic surface. 10 When this blade was laid over the mold, the bracket was situated in the middle of the acrylic block, and the edges of the bracket base were parallel to the edges of the block. After storage for 24 hours in the dark at 37 C 2,12,20 in water, 21 the acrylic block with the bonded tooth was secured onto the lower jaw of the testing machine. The shear bond force was applied by the rod. The samples were stressed in an occlusogingival direction with a crosshead speed of 0.5 mm/minute. 22 Due to our special embedding procedure, the direction of the shear force exerted by the rod was parallel to the tooth-bracket interface, and the edge of the rod was parallel to the occlusal edge of the bracket base (Fig 1). The force necessary to debond the brackets was recorded in Newtons and then converted into megapascals (MPa) by dividing the Newtons by the surface area mm 2 of the bracket base. 9,19 The adhesive remnant index (ARI) was used to measure according to the following 4-point scale introduced by Årtun and Bergland 23 : 0, no adhesive left on the tooth; 1, less than half of the adhesive left on the tooth; 2, more than half of the adhesive left on the Statistical analyses For the statistical evaluation, SPSS software (version 11.5, SPSS, Chicago, Ill) was used. The level of significance was set at P.05. Because the distribution of the shear bond strength values was normal, parametric tests were applied. The mean, standard deviation, and 95% confidence interval of the mean were estimated for each group. Comparisons of shear bond strength between lights and curing times were performed by analysis of variance (ANOVA). If a statistically significant difference was found, the post-hoc Tukey test was used to identify which groups were different. Additionally, the groups were classified into Tukey homogeneous subsets. The Weibull survival analysis was also applied to compare the performance of each sample by calculating survival probability at given values of shear stress. 10,13 The nonparametric Kruskal-Wallis test was used to test for significant differences between groups regarding the ARI. RESULTS Descriptive statistics for shear bond strength of the different groups are shown in Table II. ANOVA indicated statistically significant differences between groups (P.001). Shear bond strength was significantly lower for the 2-second high-power halogen light than the 6-second high-power halogen light (P.001), the fast halogen (P.002), and the xenon plasma arc light (P.007) (Table II). The Tukey homogeneous subsets test showed a weaker subset, which included the 2- and 3-second high-power halogen groups (P.06). The stronger subset included the 3- and 6-second high-power halogen light, the fast halogen, and the plasma light (P.565). The bracket survival probability at given levels of shear stress, estimated with the Weibull survival distribution, is shown in Figure 2. The curve of the 2-second high-power halogen lamp differs significantly: the survival probability declines more rapidly at a lower stress level; eg, the vertical dotted lines in Figure 2 correspond to 6-8 MPa that is suggested as the minimum acceptable bracket bond strength for clinical use. 24 In the range of 6-8 MPa, the probability of survival would be approximately 80% for the 2-second high-power halogen light, whereas it would be above 90% for the other groups. A shear stress of approximately 10 MPa would cause 50% bond failures in the case of the 2-second high-power

5 752 Staudt et al American Journal of Orthodontics and Dentofacial Orthopedics December 2005 Table II. Descriptive statistics for shear bond strength (MPa) Curing device* Time (seconds) N Mean SD Lower 95% Upper Minimum Maximum Swiss Master Light b a Optilux a Remecure a Total *Swiss Master Light, high-power halogen light; Optilux 501, fast halogen light; Remecure, xenon plasma arc light. a, b significant difference (P.05) between groups with different symbols. Fig 2. Bracket survival probability plotted against shear stress [MPa] (Weibull survival analysis) for high-power halogen (Swiss Master Light, SML), fast halogen (Optilux 501), and xenon plasma arc light (Remecure) with different curing times (seconds). Vertical dotted lines represent stress range of 6-8 MPa. Horizontal dotted line corresponds to failure probability of 50%. halogen lamp, whereas, to reach the same bond failure rate, a stress of over 15 MPa would be required for the other groups. No statistically significant differences were found between the groups regarding the ARI (P.72) (Table III). More than half of the adhesive was left on the tooth in most cases ( 60%) in every group after debonding. DISCUSSION These results support our hypothesis that the highpower halogen light minimizes curing time dramatically without compromising the bond strength of orthodontic brackets. Exposure to the high-power halogen light for 6 seconds and even for 3 seconds led to shear bond strengths equivalent to those achieved after 40 seconds of exposure to the fast halogen light. An exposure time of only 2 seconds is, however, not recommended because it led to significantly inferior shear bond strength values. An exposure time of 3 seconds should be used with caution, because the 3-second group was not significantly different from the 2-second group. We, therefore, recommend a curing time of 6 seconds for bonding stainless steel brackets. For ceramic brackets, the curing time might be less because the light polymerizes the composite through transillumination of the bracket as well. Curing time is reportedly not influenced by the bandwidth of the emission spectrum of different lights as long as it corresponds approximately to the absorption spectrum of camphorquinone. It is shown that the double bond conversion rate of resin composites is not influenced by the emission spectrum, if the irradiated light energy is the same. 25 Therefore, the parameter that allowed relevant reduction in curing time in our study seems to be the increase in power density (3000 mw/cm 2 for the high-power halogen light compared with 1000 mw/cm 2 for the fast halogen light). However, such a high power density might not be necessary to reduce curing time in the order of magnitude we

6 American Journal of Orthodontics and Dentofacial Orthopedics Volume 128, Number 6 Staudt et al 753 Table III. Frequency distribution of ARI ARI scores Curing device* Time (seconds) Swiss Master Light 2 0 (0.0%) 1 (6.7%) 9 (70.3%) 5 (33.3%) 3 0 (0.0%) 0 (0.0%) 13 (87.1%) 2 (13.3%) 6 0 (0.0%) 1 (6.7%) 10 (60.7%) 4 (26.6%) Optilux (0.0%) 2 (13.3%) 11 (73.3%) 2 (13.3%) Remecure 4 0 (0.0%) 2 (13.3%) 9 (70.3%) 4 (26.6%) *Swiss Master, high-power halogen light; Optilux 501, fast halogen light; Remecure, xenon plasma arc light. 0, no adhesive left on tooth; 1, less than half of adhesive left on tooth; 2, more than half of adhesive left on tooth; 3, all adhesive left on tooth. demonstrated. The xenon plasma arc light had a power density of only 1600 mw/cm 2. Exposure to the xenon plasma arc light for 4 seconds was sufficient to obtain bond strength equivalent to that achieved after 3 or 6 seconds of exposure to the high-power halogen light and after 40 seconds to the fast halogen light. This observation leads to the assumption that, from a certain power density on, further increase in power density ceases to reduce curing time considerably. Such a threshold was previously detected, suggesting that there might be no benefit to polymerization of resin composite when power density is increased above approximately 1000 mw/cm The innovations of the high-power halogen light are its high power density and the unique size of its light-guiding tip (diameter, 11 mm). The light-guiding tip covers the entire bracket, and, in our study, a 1-shot application of 3 seconds was shown to be acceptable. However, the power increase of the high-power halogen light, made possible by an innovative watercooling system, is accompanied by a bulky ( mm) and heavy (7 kg) device. Nevertheless, the high-power halogen lamp is considerably less expensive than the laser or the plasma lamp. Only the LED lamps are less expensive, but no information is available on the new generation of high-intensity LEDs. The bond strength to bovine enamel is weaker than to human enamel. 27 Nevertheless, the use of bovine enamel from deciduous teeth (as in this study) instead of permanent bovine enamel resembles more closely the bond strength values of human enamel. 27 Even when human enamel is used in vitro and in vivo, there is a difference in bond strength: in-vivo values are significantly lower. 28 Thus, a direct extrapolation of in-vitro observations to the clinical situation is difficult, and clinical studies are necessary. Furthermore, clinical studies could determine the minimum bracket bond strength necessary to withstand the forces due to occlusal loading. Recent information on such a threshold is lacking in the literature, and the only indication for the minimum necessary bond strength is given by Reynolds, 24 who suggested 6-8 MPa. If we rely on this finding, the mean shear bond strength of every group in our study would be sufficient (Table II, Fig 2). Even the minimum in every group would be within Reynolds suggested limits, with the exception of the 2-second light-curing with the high-power halogen light. However, the bond strength values measured in this study are common in analogous studies. 2,9,12,20 One must also consider that in-vivo bond strength values are lower. 28 The site of bond failure was unaffected by light source and curing time. Most failures at debonding were located at the bracket-adhesive interface; this agrees with other shear-bond testing studies. 6,8,11,13,19 The relatively weak mechanical retention between the composite and the bracket base reduces the risk of enamel fractures. CONCLUSIONS The results of this study support the use of the high-power halogen light. It is a cost-effective solution to reduce light-curing time. The recommended curing time for bonding stainless steel brackets is 6 seconds or, with caution, even 3 seconds. Our results are based on in-vitro laboratory conditions, and the clinical relevance of our findings should be confirmed in vivo. We thank Dr M. Cattani for his help with the shear bond strength testing. REFERENCES 1. Althoff O, Hartung M. Advances in light curing. Am J Dent 2000;13(spec no):77d-81d. 2. Oesterle LJ, Newman SM, Shellhart WC. Rapid curing of bonding composite with a xenon plasma arc light. Am J Orthod Dentofacial Orthop 2001;119: Blankenau RJ, Kelsey WP, Powell GL, Shearer GO, Barkmeier WW, Cavel XT. Degree for composite resin polymerization with visible light and argon laser. Am J Dent 1991;4: Kurchak M, Desantos B, Powers J, Turner D. Argon laser for light-curing adhesives. J Clin Orthod 1997;31:371-4.

7 754 Staudt et al American Journal of Orthodontics and Dentofacial Orthopedics December Weinberger SJ, Foley TF, McConnell RJ, Wright GZ. Bond strengths of two ceramic brackets using argon laser, light, and chemically cured resin systems. Angle Orthod 1997;67: Talbot TQ, Blankenau RJ, Zobitz ME, Weaver AL, Lohse CM, Rebellato J. Effect of argon laser irradiation on shear bond strength of orthodontic brackets: an in vitro study. Am J Orthod Dentofacial Orthop 2000;118: Lalani N, Foley TF, Voth R, Banting D, Mamandras A. Polymerization with the argon laser: curing time and shear bond strength. Angle Orthod 2000;70: Dunn WJ, Taloumis LJ. Polymerization of orthodontic resin cement with light-emitting diode curing units. Am J Orthod Dentofacial Orthop 2002;122: Sfondrini MF, Cacciafesta V, Pistorio A, Sfondrini G. Effects of conventional and high-intensity light-curing on enamel shear bond strength of composite resin and resin-modified glassionomer. Am J Orthod Dentofacial Orthop 2001;119: Pettemerides AP, Ireland AJ, Sherriff M. An ex vivo investigation into the use of a plasma arc light when using a visible light-cured composite and a resin-modified glass poly(alkenoate) cement in orthodontic bonding. J Orthod 2001;28: Klocke A, Korbmacher HM, Luck LG, Ghosh J, Kahl-Nieke B. Plasma arc curing of ceramic brackets: an evaluation of shear bond strength and debonding characteristics. Am J Orthod Dentofacial Orthop 2003;124: Ishikawa H, Komori A, Kojima I, Ando F. Orthodontic bracket bonding with a plasma-arc light and resin-reinforced glass ionomer cement. Am J Orthod Dentofacial Orthop 2001;120: Klocke A, Korbmacher HM, Huck LG, Kahl-Nieke B. Plasma arc curing lights for orthodontic bonding. Am J Orthod Dentofacial Orthop 2002;122: Dietschi D, Marret N, Krecji I. Comparative efficiency of plasma and halogen light sources on composite micro-hardness in different curing conditions. Dent Mater 2003;19: Munksgaard EC, Peutzfeldt A, Asmussen E. Elution of TEG- DMA and BisGMA from a resin and a resin composite cured with halogen or plasma light. Eur J Oral Sci 2000;108: Bishara SE, Ajlouni R, Oonsombat C. Evaluation of a new curing light on the shear bond strength of orthodontic brackets. Angle Orthod 2003;73: Swanson T, Dunn WJ, Childers DE, Taloumis LJ. Shear bond strength of orthodontic brackets bonded with light-emitting diode curing units at various polymerization times. Am J Orthod Dentofacial Orthop 2004;125: Usumez S, Buyukyilmaz T, Karaman AI. Effect of light-emitting diode on bond strength of orthodontic brackets. Angle Orthod 2004;74: Sfondrini MF, Cacciafesta V, Klersy C. Halogen versus highintensity light-curing of uncoated and pre-coated brackets: a shear bond strength study. J Orthod 2002;29: Oesterle LJ, Newman SM, Shellhart WC. Comparative bond strength of brackets cured using a pulsed xenon curing light with 2 different light-guide sizes. Am J Orthod Dentofacial Orthop 2002;122: Fox NA, McCabe JF, Buckley JG. A critique of bond strength testing in orthodontics. Br J Orthod 1994;21: Eliades T, Brantley WA. The inappropriateness of conventional orthodontic bond strength assessment protocols. Eur J Orthod 2000;22: Årtun J, Bergland S. Clinical trials with crystal growth conditioning as an alternative to acid-etch enamel pretreatment. Am J Orthod 1984;85: Reynolds IR. A review of direct orthodontic bonding. Br J Orthod 1975;2: Yoon TH, Lee YK, Lim BS, Kim CW. Degree of polymerization of resin composites by different light sources. J Oral Rehabil 2002;29: Musanje L, Darvell BW. Polymerization of resin composite restorative materials: exposure reciprocity. Dent Mater 2003;19: Oesterle LJ, Shellhart WC, Bellanger GK. The use of bovine enamel in bonding studies. Am J Orthod Dentofacial Orthop 1998;113: Murray SD, Hobson RS. Comparison of in vivo and in vitro shear bond strength. Am J Orthod Dentofacial Orthop 2003;123: 2-9.

Bracket bond strength dependence on light power density. STAUDT, Christine Bettina, KREJCI, Ivo, MAVROPOULOS, Anestis. Abstract

Bracket bond strength dependence on light power density. STAUDT, Christine Bettina, KREJCI, Ivo, MAVROPOULOS, Anestis. Abstract Article Bracket bond strength dependence on light power density STAUDT, Christine Bettina, KREJCI, Ivo, MAVROPOULOS, Anestis Abstract In order to reduce curing time for bracket bonding with light-cured

More information

Bonding characteristics of a self-etching primer and precoated brackets: an in vitro study

Bonding characteristics of a self-etching primer and precoated brackets: an in vitro study The European Journal of Orthodontics Advance Access published February 22, 26 European Journal of Orthodontics of 5 doi:.9/ejo/cji7 The Author 26. Published by Oxford University Press on behalf of the

More information

The acid-etch technique has been widely used in

The acid-etch technique has been widely used in ORIGINAL ARTICLE Comparison of bond strength between orthodontic brackets bonded with halogen and plasma arc curing lights: An in-vitro and in-vivo study Michael D. Signorelli, a Elizabeth Kao, b Peter

More information

Immediate versus Delayed Force Application after Orthodontic Bonding; An In Vitro Study

Immediate versus Delayed Force Application after Orthodontic Bonding; An In Vitro Study Journal of Dentistry, Tehran University of Medical Sciences Original Article Immediate versus Delayed Force Application after Orthodontic Bonding; An In Vitro Study M. Basafa 1, F. Farzanegan 2 1 Professor,

More information

A Comparison of Bond Strength Between Directand Indirect-bonding Methods

A Comparison of Bond Strength Between Directand Indirect-bonding Methods Marquette University e-publications@marquette School of Dentistry Faculty Research and Publications Dentistry, School of 3-1-2006 A Comparison of Bond Strength Between Directand Indirect-bonding Methods

More information

Shear Bond Strength of Orthodontic Brackets Cured with Different Light Sources under Thermocycling

Shear Bond Strength of Orthodontic Brackets Cured with Different Light Sources under Thermocycling Shear Bond Strength of Orthodontic Brackets Cured with Different Light Sources under Thermocycling Oral Sokucu a Seyda Herguner Siso b Firat Ozturk c Ruhi Nalcaci d ABSTRACT Objectives: The aim of this

More information

The effect of orthodontic bracket pad shape on shear bond strength, an in vitro study on human enamel

The effect of orthodontic bracket pad shape on shear bond strength, an in vitro study on human enamel Journal section: Orthodontics Publication Types: Research doi:10.4317/jced.55121 http://dx.doi.org/10.4317/jced.55121 The effect of orthodontic bracket pad shape on shear bond strength, an in vitro study

More information

Comparison of shear bond strength and adhesive remnant index between precoated and conventionally bonded orthodontic brackets

Comparison of shear bond strength and adhesive remnant index between precoated and conventionally bonded orthodontic brackets Guzman et al. Progress in Orthodontics 2013, 14:39 RESEARCH Open Access Comparison of shear bond strength and adhesive remnant index between precoated and conventionally bonded orthodontic brackets Ulises

More information

A Comparative Study between Bond Strength of Rebonded and Recycled Orthodontic Brackets

A Comparative Study between Bond Strength of Rebonded and Recycled Orthodontic Brackets 1 A Comparative Study between Bond Strength of Rebonded and Recycled Orthodontic Brackets ABSTRACT F. Heravi DDS MS*, R. Naseh DDS MS** Introduction: A large number of orthodontists prefer to rebond the

More information

Adhesive Thickness Effects on the Bond Strength of a Light- Cured Resin-Modified Glass Ionomer Cement

Adhesive Thickness Effects on the Bond Strength of a Light- Cured Resin-Modified Glass Ionomer Cement Original Article Adhesive Thickness Effects on the Bond Strength of a Light- Cured Resin-Modified Glass Ionomer Cement Selim Arici, DDS, MMedSci, PhD a ; Cem Mustafa Caniklioglu, DDS, PhD b ; Nursel Arici,

More information

Original Research. Shear bond strength of new self etching primer (RSEP) Sorake A et al

Original Research. Shear bond strength of new self etching primer (RSEP) Sorake A et al Journal of International Oral Health 2015; 7(7):17-21 Received: 17 th January 2015 Accepted: 10 th April 2015 Conflicts of Interest: None Source of Support: Nil Original Research Comparison of Shear Bond

More information

Orthodontic In Vivo Bond Strength: Comparison with In Vitro Results

Orthodontic In Vivo Bond Strength: Comparison with In Vitro Results Original Article Orthodontic In Vivo Bond Strength: Comparison with In Vitro Results Kevin L. Pickett, DMD, MS a ; P. Lionel Sadowsky, DMD, BDS, Dip Ortho, M Dent b ; Alex Jacobson, DMD, MDS, MS, PhD c

More information

Comparison of Shear Bond Strength of Metal Brackets Bonded to Porcelain Surface using Different Surface Conditioning Methods: An in vitro Study

Comparison of Shear Bond Strength of Metal Brackets Bonded to Porcelain Surface using Different Surface Conditioning Methods: An in vitro Study 10.5005/jp-journals-10024-1174 ORIGINAL RESEARCH Comparison of Shear Bond Strength of Metal Brackets Bonded to Porcelain Surface using Different Surface Conditioning Methods: An in vitro Study PV Girish,

More information

Various whitening systems are currently being

Various whitening systems are currently being ORIGINAL ARTICLE Can previously bleached teeth be bonded safely? Tancan Uysal, DDS, a Faruk Ayhan Basciftci, DDS, MS, b Serdar Üşümez, DDS, PhD, b Zafer Sari, DDS, PhD, b and Ahmet Buyukerkmen, DDS a Konya,

More information

A new flash-free orthodontic adhesive system: A first clinical and stereomicroscopic study

A new flash-free orthodontic adhesive system: A first clinical and stereomicroscopic study Original Article A new flash-free orthodontic adhesive system: A first clinical and stereomicroscopic study Moritz Foersch a ; Christian Schuster b ; Roman K. Rahimi c ; Heinrich Wehrbein d ; Collin Jacobs

More information

COMPARISON OF SHEAR BOND STRENGTH OF NEW AND RECYCLED BRACKETS WITH DIFFERENT MESH DESIGNS

COMPARISON OF SHEAR BOND STRENGTH OF NEW AND RECYCLED BRACKETS WITH DIFFERENT MESH DESIGNS COMPARISON OF SHEAR BOND STRENGTH OF NEW AND RECYCLED BRACKETS WITH DIFFERENT MESH DESIGNS Author: Dr.Pragati, M.D.S. 3 rd year student Department of Orthodontics, Co-Authors: Prof. Praveen Mehrotra, M.D.S.

More information

Effect of light-curing units in shear bond strength of metallic brackets: an in vitro study

Effect of light-curing units in shear bond strength of metallic brackets: an in vitro study www.scielo.br/jaos Effect of light-curing units in shear bond strength of metallic brackets: an in vitro study Luciana Borges RETAMOSO 1, Niége Michelle Lazzari ONOFRE 2, Luciane HANN 3, Ernani Menezes

More information

Assessment of the Confidence of the Adhesive Remnant Index Score With Different Methods

Assessment of the Confidence of the Adhesive Remnant Index Score With Different Methods Original Article Assessment of the Confidence of the Adhesive Remnant Index Score With Different Methods A. Alper Öz; 1 * Sabahat Yazicioğlu; 2 Nursel Arici; 3 Berat Serdar Akdeniz; 4 Naci Murat; 5 and

More information

The effect of different force magnitudes for placement of orthodontic brackets on shear bond strength, in three adhesive systems

The effect of different force magnitudes for placement of orthodontic brackets on shear bond strength, in three adhesive systems Journal section: Orthodontics Publication Types: Research doi:10.4317/jced.54733 http://dx.doi.org/10.4317/jced.54733 The effect of different force magnitudes for placement of orthodontic brackets on shear

More information

Bond Strength of Direct and Indirect Bonded Brackets After Thermocycling

Bond Strength of Direct and Indirect Bonded Brackets After Thermocycling Marquette University e-publications@marquette School of Dentistry Faculty Research and Publications Dentistry, School of 3-1-2006 Bond Strength of Direct and Indirect Bonded Brackets After Thermocycling

More information

In Vitro Evaluation of a Moisture-Active Adhesive for Indirect Bonding

In Vitro Evaluation of a Moisture-Active Adhesive for Indirect Bonding Original Article In Vitro Evaluation of a Moisture-Active Adhesive for Indirect Bonding Arndt Klocke, Dr med dent, MS a ; Jianmin Shi, MSc b ;Bärbel Kahl-Nieke, Dr med dent, PhD c ; Ulrich Bismayer, Dr

More information

Early Shear Bond Strength of a One-step Self-adhesive on Orthodontic Brackets

Early Shear Bond Strength of a One-step Self-adhesive on Orthodontic Brackets Original Article Early Shear Bond Strength of a One-step Self-adhesive on Orthodontic Brackets Samir E. Bishara a ; Adam W. Ostby b ; Raed Ajlouni c ; John F. Laffoon d ; John J. Warren e ABSTRACT Objective:

More information

The words adhesionpromoters werefirstusedto

The words adhesionpromoters werefirstusedto ORIGINAL ARTICLE Effects of 3 adhesion promoters on the shear bond strength of orthodontic brackets: An in-vitro study Ascensión Vicente, a Luis A. Bravo, b Martín Romero, c Antonio J. Ortíz, d and Manuel

More information

Shear Bond Strength of Acidic Primer, Light-Cure Glass Ionomer, Light-Cure and Self Cure Composite Adhesive Systems - An In Vitro Study

Shear Bond Strength of Acidic Primer, Light-Cure Glass Ionomer, Light-Cure and Self Cure Composite Adhesive Systems - An In Vitro Study Shear Bond Strength of Acidic Primer, Light-Cure Glass Ionomer, Light-Cure and Self Cure Composite Adhesive Systems - An In Vitro Study Krishnakanth Reddy D 1, Kishore M S V 2, Safeena Safeena 3 1 Professor,

More information

In Vitro Investigation of Indirect Bonding with a Hydrophilic Primer

In Vitro Investigation of Indirect Bonding with a Hydrophilic Primer Original Article In Vitro Investigation of Indirect Bonding with a Hydrophilic Primer Arndt Klocke, Dr med dent, MS a ; Jianmin Shi, MSc b ;Bärbel Kahl-Nieke, Dr med dent, PhD c ; Ulrich Bismayer, Dr rer

More information

ORIGINAL RESEARCH. Sandhya CA Shyam Lohakare, Pushpa Vinay Hazarey ABSTRACT

ORIGINAL RESEARCH. Sandhya CA Shyam Lohakare, Pushpa Vinay Hazarey ABSTRACT ORIGINAL RESEARCH Comparison of shear bond strength of no mix-composite resin, self cure resin reinforced glass ionomer cement and light cure resin reinforced glass ionomer cement: an in-vitro study Sandhya

More information

Shear bond strength of ceramic and metallic orthodontic brackets bonded with self-etching primer and conventional bonding adhesives

Shear bond strength of ceramic and metallic orthodontic brackets bonded with self-etching primer and conventional bonding adhesives Electronic Physician (ISSN: 2008-5842) http://www.ephysician.ir January 2017, Volume: 9, Issue: 1, Pages: 3584-3591, DOI: http://dx.doi.org/10.19082/3584 Shear bond strength of ceramic and metallic orthodontic

More information

The Effect of Different Bracket Base Cleaning Method on Shear Bond Strength of Rebonded Brackets

The Effect of Different Bracket Base Cleaning Method on Shear Bond Strength of Rebonded Brackets 10.5005/jp-journals-10024-1417 Emad Original F Al Maaitah Research et al The Effect of Different Bracket Base Cleaning Method on Shear Bond Strength of Rebonded Brackets Emad F Al Maaitah, Sawsan Alomari,

More information

Three-dimensional measurement of enamel loss caused by bonding and debonding of orthodontic brackets. VAN WAES, H, MATTER, T, KREJCI, Ivo.

Three-dimensional measurement of enamel loss caused by bonding and debonding of orthodontic brackets. VAN WAES, H, MATTER, T, KREJCI, Ivo. Article Three-dimensional measurement of enamel loss caused by bonding and debonding of orthodontic brackets VAN WAES, H, MATTER, T, KREJCI, Ivo Abstract A mechanical computerized three-dimensional scanner

More information

Flowable Composite for Orthodontic Bracket Bonding (in vitro study)

Flowable Composite for Orthodontic Bracket Bonding (in vitro study) Tikrit Journal for Dental Sciences 1(01)44-50 Flowable Composite for Orthodontic Bracket Bonding (in vitro study) Bahn G. Agha BDS, MSc. (1) Key words flowable composite; shear bond strength; orthodontic

More information

Effect of bleaching with hydrogen peroxide into different concentrations on shear strength of brackets bonded with a resin-modified glass ionomer

Effect of bleaching with hydrogen peroxide into different concentrations on shear strength of brackets bonded with a resin-modified glass ionomer Braz J Oral Sci. January/March 2008 - Vol. 7 - Number 24 Effect of bleaching with hydrogen peroxide into different concentrations on shear strength of brackets bonded with a resin-modified glass ionomer

More information

LED vs Halogen Light-Curing of Adhesive-Precoated Brackets

LED vs Halogen Light-Curing of Adhesive-Precoated Brackets Allen Press DTPro System GALLEY 56 File # 01TQ Name /angl/78_201 12/05/2007 02:41PM Plate # 0-Composite pg 56 # 1 Original Article LED vs Halogen Light-Curing of Adhesive-Precoated Brackets?1 Davide Mirabella

More information

Influence of saliva contamination on the shear bond strength of adhesives on enamel

Influence of saliva contamination on the shear bond strength of adhesives on enamel Influence of saliva contamination on the shear bond strength of adhesives on enamel Tatiana Feres Assad-Loss 1, Mônica Tostes 2, José Nelson Mucha 3 Objective: To evaluate shear bond strength of 3 adhesive

More information

Shear bond strength of a new polycarbonate bracketan in vitro study with 14 adhesives

Shear bond strength of a new polycarbonate bracketan in vitro study with 14 adhesives European Journal of Orthodontics 18 (1996) 295-301 O 1996 European Orthodontic Society Shear bond strength of a new polycarbonate bracketan in vitro study with 14 adhesives Nilgun Akin-Nergiz,* Ibrahim

More information

S. Shear bond strength and debonding characteristics of metal and ceramic

S. Shear bond strength and debonding characteristics of metal and ceramic Journal section: Orthodontics Publication Types: Research doi:10.4317/jced.52658 http://dx.doi.org/10.4317/jced.52658 Shear bond strength and debonding characteristics of metal and ceramic brackets bonded

More information

Bond Strength of Disinfected Metal and Ceramic Brackets:

Bond Strength of Disinfected Metal and Ceramic Brackets: Original Article Bond Strength of Disinfected Metal and Ceramic s: An In Vitro Study Cornelia Speer a ; Dorothee Zimny b ; Werner Hopfenmueller c ; Eva Andrea Holtgrave d Abstract: The aim of this in vitro

More information

Bond strengths between composite resin and auto cure glass ionomer cement using the co-cure technique

Bond strengths between composite resin and auto cure glass ionomer cement using the co-cure technique SCIENTIFIC ARTICLE Australian Dental Journal 2006;51:(2):175-179 Bond strengths between composite resin and auto cure glass ionomer cement using the co-cure technique GM Knight,* JM McIntyre,* Mulyani*

More information

Ceramic brackets have been the preferred choice

Ceramic brackets have been the preferred choice ORIGINAL ARTICLE Diode laser debonding of ceramic brackets Paul J. Feldon, a Peter E. Murray, b James G. Burch, c Malcolm Meister, c and Matthew A. Freedman d Boca Raton and Fort Lauderdale, Fla Introduction:

More information

Transbond PLUS Color Change Adhesive Lunch and Learn

Transbond PLUS Color Change Adhesive Lunch and Learn Unitek Bonding System Transbond PLUS Color Change Adhesive Lunch and Learn Transbond PLUS Color Change Adhesive Agenda What is Transbond PLUS Color Change Adhesive? Why Transbond PLUS Color Change Adhesive?

More information

[Downloaded free from on Wednesday, September 28, 2016, IP: ]

[Downloaded free from  on Wednesday, September 28, 2016, IP: ] Original Article Effect of fluoridated casein phospopeptide amorphous calcium phosphate complex, chlorhexidine fluoride mouthwash on shear bond strength of orthodontic brackets: A comparative in vitro

More information

Bonding to molars the effect of etch time (an in vitro study)

Bonding to molars the effect of etch time (an in vitro study) European Journal of Orthodontics 20 (1998) 195 199 1998 European Orthodontic Society Bonding to molars the effect of etch time (an in vitro study) Chris D. Johnston*, Donald J. Burden*, David L. Hussey**

More information

The influence of different bracket base surfaces on tensile and shear bond strength

The influence of different bracket base surfaces on tensile and shear bond strength European Journal of Orthodontics 30 (2008) 490 494 doi:10.1093/ejo/cjn029 Advance cess publication 5 August 2008 The Author 2008. Published by Oxford University Press on behalf of the European Orthodontic

More information

Effects of CO 2 laser debonding of a ceramic bracket on the mechanical properties of enamel

Effects of CO 2 laser debonding of a ceramic bracket on the mechanical properties of enamel Original Article Effects of CO 2 laser debonding of a ceramic bracket on the mechanical properties of enamel Masahiro Iijima a ; Yoshitaka Yasuda b ; Takeshi Muguruma c ; Itaru Mizoguchi d ABSTRACT Objective:

More information

An In-Vivo Comparison Of The Efficacy Between Direct And Indirect Bonding Methods

An In-Vivo Comparison Of The Efficacy Between Direct And Indirect Bonding Methods ISPUB.COM The Internet Journal of Dental Science Volume 13 Number 1 An In-Vivo Comparison Of The Efficacy Between Direct And Indirect Bonding Methods P Premanand, A J Shankar Citation P Premanand, A J

More information

The effects of argon laser curing of a resin adhesive on bracket retention and enamel decalcification: a prospective clinical trial

The effects of argon laser curing of a resin adhesive on bracket retention and enamel decalcification: a prospective clinical trial European Journal of Orthodontics 26 (2004) 553 560 European Journal of Orthodontics vol. 26 no. 5 European Orthodontic Society 2004; all rights reserved. The effects of argon laser curing of a resin adhesive

More information

Scanning Electron Microscopy Evaluation of the Bonding Mechanism of a Self-etching Primer on Enamel

Scanning Electron Microscopy Evaluation of the Bonding Mechanism of a Self-etching Primer on Enamel Original Article Scanning Electron Microscopy Evaluation of the Bonding Mechanism of a Self-etching Primer on Enamel Julio P. Cal-Neto a ; José Augusto M. Miguel b Abstract: The aim of this study was to

More information

Dental Research Journal

Dental Research Journal Dental Research Journal Original Article Shear bond strength of orthodontic buccal tubes to porcelain Kathiravan Purmal 1, Mohammad K. Alam 1, Prema Sukumaran 2 1 Orthodontic Unit, School of Dental Science,

More information

The desire to cure on demand has caused more orthodontic

The desire to cure on demand has caused more orthodontic ORIGINAL ARTICLE In-vitro assessment of temperature rise in the pulp during orthodontic bonding Sıddık Malkoç, a Tancan Uysal, b Serdar Üsxümez, c Eren _ Isxman, d and Aslı Baysal e Malatya, Kayseri, Gaziantep,

More information

The halogen light is the most commonly used

The halogen light is the most commonly used ORIGINAL ARTICLE Efficiency of light-emitting diode and halogen units in reducing residual monomers Felipe de Assis Ribeiro Carvalho, a Rhita C. Almeida, a Marco Antonio Almeida, b Lucia H. S. Cevidanes,

More information

Bond strength assessment of metal brackets bonded to porcelain fused to metal surface using different surface conditioning method

Bond strength assessment of metal brackets bonded to porcelain fused to metal surface using different surface conditioning method ORIGINAL RESEARCH Bond strength assessment of metal brackets bonded to porcelain fused to metal surface using different surface conditioning method Vaishnav PD 1, Philip P 2, Shetty S 3, Mogra S 4, Batra

More information

Even though the acid-etching technique is useful

Even though the acid-etching technique is useful SHORT COMMUNICATION Bond failure rates with a self-etching primer: A randomized controlled trial Julio Pedra e Cal-Neto, a Cátia Abdo Quintão, b Marco Antônio de Oliveira Almeida, c and José Augusto Mendes

More information

Various composite resins are used in orthodontic

Various composite resins are used in orthodontic CLINICIAN S CORNER Four applications of reinforced polyethylene fiber material in orthodontic practice Ali Ihya Karaman, DDS, PhD, a Nihal Kir, DDS, b and Sema Belli, DDS, PhD c Konya, Turkey Four different

More information

Bond Strength of Composite Resin Luting Cements to Fiber-reinforced Composite Root Canal Post

Bond Strength of Composite Resin Luting Cements to Fiber-reinforced Composite Root Canal Post Bond Strength of Composite Resin Luting Cements to Fiber-reinforced Composite Root Canal Post Abstract Aims: The aim of this study was to compare the attachment of different composite resin luting cements

More information

The effect of the blue light diode laser on shear-peel bond strength of orthodontic brackets with varying curing time and distance.

The effect of the blue light diode laser on shear-peel bond strength of orthodontic brackets with varying curing time and distance. University of Louisville ThinkIR: The University of Louisville's Institutional Repository Electronic Theses and Dissertations 8-2010 The effect of the blue light diode laser on shear-peel bond strength

More information

UvA-DARE (Digital Academic Repository) A better understanding of orthodontic bracket bonding Algera, T.J. Link to publication

UvA-DARE (Digital Academic Repository) A better understanding of orthodontic bracket bonding Algera, T.J. Link to publication UvA-DARE (Digital Academic Repository) A better understanding of orthodontic bracket bonding Algera, T.J. Link to publication Citation for published version (APA): Algera, T. J. (2009). A better understanding

More information

Until recently, the most popular light-curing unit. Effects of high-intensity curing lights on microleakage under orthodontic bands ORIGINAL ARTICLE

Until recently, the most popular light-curing unit. Effects of high-intensity curing lights on microleakage under orthodontic bands ORIGINAL ARTICLE ORIGINAL ARTICLE Effects of high-intensity curing lights on microleakage under orthodontic bands Tancan Uysal, a Sabri Ilhan Ramoglu, b Mustafa Ulker, c and Huseyin Ertas d Kayseri, Turkey Introduction:

More information

Effect of Self-etchant ph on Shear Bond Strength of Orthodontic Brackets: An in vitro Study

Effect of Self-etchant ph on Shear Bond Strength of Orthodontic Brackets: An in vitro Study 10.5005/jp-journals-10015-1126 ORIGINAL RESEARCH Effect of Self-etchant ph on Shear Bond Strength of Orthodontic Brackets: An in vitro Study Amit Goyal, H Jyothikiran, BM Shivalinga WJD ABSTRACT Aims:

More information

A comparison of shear bond strength and mean survival time between treated and untreated mesh backed stainless steel brackets, an in-vitro assessment.

A comparison of shear bond strength and mean survival time between treated and untreated mesh backed stainless steel brackets, an in-vitro assessment. (J lnd Orthod Soc 2002; 35:124 128) RESEARCH ARTICLE - A comparison of shear bond strength and mean survival time between treated and untreated mesh backed stainless steel brackets, an in-vitro assessment.

More information

The Effect of Argon Laser Irradiation on Demineralization Resistance of Human Enamel Adjacent to Orthodontic Brackets: An In Vitro Study

The Effect of Argon Laser Irradiation on Demineralization Resistance of Human Enamel Adjacent to Orthodontic Brackets: An In Vitro Study Original Article The Effect of Argon Laser Irradiation on Demineralization Resistance of Human Enamel Adjacent to Orthodontic Brackets: An In Vitro Study Lloyd Noel, DMD, MS a ; Joe Rebellato, DDS b ;

More information

MDJ Evaluation the effect of eugenol containing temporary Vol.:9 No.:2 2012

MDJ Evaluation the effect of eugenol containing temporary Vol.:9 No.:2 2012 MDJ Evaluation the effect of eugenol containing temporary fillings on shear bond strength of composite restoration Dr. Farid G. Numan B.D.S., M.Sc Abstract The purpose of this in-vitro study was to evaluate

More information

UvA-DARE (Digital Academic Repository) A better understanding of orthodontic bracket bonding Algera, T.J. Link to publication

UvA-DARE (Digital Academic Repository) A better understanding of orthodontic bracket bonding Algera, T.J. Link to publication UvA-DARE (Digital Academic Repository) A better understanding of orthodontic bracket bonding Algera, T.J. Link to publication Citation for published version (APA): Algera, T. J. (2009). A better understanding

More information

Glass Ionomer Cement as an Orthodontic Bonding Agent

Glass Ionomer Cement as an Orthodontic Bonding Agent 10.5005/jp-journals-10024-1203 Hashim Ali, Sasi Maroli ORIGINAL RESEARCH Hashim Ali, Sasi Maroli ABSTRACT Aim To find an alternative to popular orthodontic bonding agent composite resin and elimination

More information

Comparison of three different orthodontic wires for bonded lingual retainer fabrication

Comparison of three different orthodontic wires for bonded lingual retainer fabrication Original Article THE KOREAN JOURNAL of ORTHODONTICS pissn 2234-7518 eissn 2005-372X http://dx.doi.org/10.4041/kjod.2012.42.1.39 Comparison of three different orthodontic wires for bonded lingual retainer

More information

WHY PLASMA TECHNOLOGY?

WHY PLASMA TECHNOLOGY? WHY PLASMA TECHNOLOGY? Why Plasma Technology? With the increased number of curing lights in the market today, a dental professional is faced with a number of choices. In past years, the quartz-tungsten-halogen

More information

Improving Tensile Bond Strength of Orthodontic Bracket by Applying Papain Gel as an Deproteinization Agent

Improving Tensile Bond Strength of Orthodontic Bracket by Applying Papain Gel as an  Deproteinization Agent doi: 10.14693/jdi.v24i3.1095 ORIGINAL ARTICLE Improving Tensile Bond Strength of Orthodontic Bracket by Applying Papain Gel as an Email Deproteinization Agent Niswati F. Rosyida 1, Sri Suparwitri 1, Pinandi

More information

Assessing the Performance of a Methyl Methacrylate-Based Resin Cement with Self-Etching Primer for Bonding Orthodontic Brackets

Assessing the Performance of a Methyl Methacrylate-Based Resin Cement with Self-Etching Primer for Bonding Orthodontic Brackets Original Article Assessing the Performance of a Methyl Methacrylate-Based Resin Cement with Self-Etching Primer for Bonding Orthodontic Brackets Mayuko Kawasaki, DDS a ; Tohru Hayakawa, PhD b ; Tsutomu

More information

Clinical evaluation of the failure rates of metallic brackets

Clinical evaluation of the failure rates of metallic brackets Universidade de São Paulo Biblioteca Digital da Produção Intelectual - BDPI Departamento de Clínica Infantil, Odontologia Preventiva e Social - FORP/807 Artigos e Materiais de Revistas Científicas - FORP/807

More information

Shear bond strength of metallic and ceramic brackets using color change adhesives

Shear bond strength of metallic and ceramic brackets using color change adhesives Shear bond strength of metallic and ceramic brackets using color change adhesives Aisha de Souza Gomes Stumpf 1, Carlos Bergmann 2, José Renato Prietsch 3, Juliane Vicenzi 4 Objective: To determine the

More information

Evaluation of Degree of Cure and Shear Bond Strength of a Color Changing Light Cure Adhesive

Evaluation of Degree of Cure and Shear Bond Strength of a Color Changing Light Cure Adhesive 10.5005/jp-journals-10021-1116 ORIGINAL ARTICLE Hanumantha Rao Chidipothu, Shyamala Chandrasekhar Evaluation of Degree of Cure and Shear Bond Strength of a Color Changing Light Cure Adhesive 1 Hanumantha

More information

The Impact of Er,Cr:YSGG Laser Recycling on Shear Bond Strength (SBS) of Stainless Steel Orthodontic Brackets

The Impact of Er,Cr:YSGG Laser Recycling on Shear Bond Strength (SBS) of Stainless Steel Orthodontic Brackets European International Journal of Science and Technology Vol. 3 No. 7 September, 2014 The Impact of Er,Cr:YSGG Laser Recycling on Shear Bond Strength (SBS) of Stainless Steel Orthodontic Brackets Faisal

More information

Victory Series Active Self-Ligating Brackets. Reliable. and. Effective

Victory Series Active Self-Ligating Brackets. Reliable. and. Effective Victory Series Active Self-Ligating Brackets Reliable and Effective Victory Series Active Self-Ligating Brackets Reliability Victory Series SL Brackets are widely accepted as an industry benchmark for

More information

Cementation of crowns, bridges, inlays and onlays. Made of conventional porcelain, ceramic, hybrid ceramics or composite resin Cementation of veneers

Cementation of crowns, bridges, inlays and onlays. Made of conventional porcelain, ceramic, hybrid ceramics or composite resin Cementation of veneers Flow Chart Sheet 1621 Sakazu, Kurashiki, Okayama 710-0801, Japan INDICTIONS Case -1-2 Case Case C Case D Case E Cementation of crowns, bridges, inlays and onlays Made of metal Made of conventional porcelain,

More information

Evaluation of bond strength of orthodontic brackets without enamel etching

Evaluation of bond strength of orthodontic brackets without enamel etching Journal section: Operative Dentistry and Endodontics Publication Types: Research doi:10.4317/jced.52253 http://dx.doi.org/10.4317/jced.52253 Evaluation of bond strength of orthodontic brackets without

More information

Comparison of the Heat Generation of Light Curing Units

Comparison of the Heat Generation of Light Curing Units Comparison of the Heat Generation of Light Curing Units Abstract Aim: The aim of this study was to evaluate the heat generation of three different types of light curing units. Methods and Materials: Temperature

More information

Within the past few years, ceramic brackets have

Within the past few years, ceramic brackets have ORIGINAL ARTICLE Debonding of ceramic brackets by a new scanning laser method Mehmet Oguz Oztoprak, a Didem Nalbantgil, a Aysxe Sine Erdem, b Murat Tozlu, b and Tülin Arun c Istanbul, Turkey Introduction:

More information

Effects of silanation time on shear bond strength between a gold alloy surface and metal bracket

Effects of silanation time on shear bond strength between a gold alloy surface and metal bracket Original Article THE KOREAN JOURNAL of ORTHODONTICS pissn 2234-7518 eissn 2005-372X Effects of silanation time on shear bond strength between a gold alloy surface and metal bracket Min-Ho Jung a Won-Jun

More information

Bond Strength of Amorphous Calcium Phosphate Containing Orthodontic Composite Used as a Lingual Retainer Adhesive

Bond Strength of Amorphous Calcium Phosphate Containing Orthodontic Composite Used as a Lingual Retainer Adhesive Original Article Bond Strength of Amorphous Calcium Phosphate Containing Orthodontic Composite Used as a Lingual Retainer Adhesive Tancan Uysal a ; Mustafa Ulker b ; Gulsen Akdogan c ; Sabri Ilhan Ramoglu

More information

Objective: The aim of this study was to assess the impact of blood contamination on shear

Objective: The aim of this study was to assess the impact of blood contamination on shear www.scielo.br/jaos Blood contamination effect on shear bond strength of an orthodontic hydrophilic resin Taís de Morais Alves da CUNHA 1, Bruna Ariela BEHRENS 2, Denise NASCIMENTO 2, Luciana Borges RETAMOSO

More information

Microleakage under Metallic and Ceramic Brackets Bonded with Orthodontic Self-Etching Primer Systems

Microleakage under Metallic and Ceramic Brackets Bonded with Orthodontic Self-Etching Primer Systems Original Article Microleakage under Metallic and Ceramic Brackets Bonded with Orthodontic Self-Etching Primer Systems Tancan Uysal a ; Mustafa Ulker b ; Sabri Ilhan Ramoglu c ; Huseyin Ertas d ABSTRACT

More information

Using Hemostatic Agents During Orthodontic Bonding: An In Vitro Study

Using Hemostatic Agents During Orthodontic Bonding: An In Vitro Study Original Article Using Hemostatic Agents During Orthodontic Bonding: An In Vitro Study Berat Serdar Akdeniz, DDS, PhD; 1, *Abdullah Alper Oz, DDS, PhD; 2 Nursel Arici, DDS, PhD; 3 Onur Demir, DDS, PhD;

More information

Antimicrobial Effects of Zinc Oxide in an Orthodontic Bonding Agent

Antimicrobial Effects of Zinc Oxide in an Orthodontic Bonding Agent Original Article Antimicrobial Effects of Zinc Oxide in an Orthodontic Bonding Agent Clayton Glen Spencer a ; Phillip M. Campbell b ; Peter H. Buschang c ; John Cai d ; Allen L. Honeyman e ABSTRACT Objective:

More information

In nitro bond strength of cements to treated teeth

In nitro bond strength of cements to treated teeth In nitro bond strength of cements to treated teeth Mohamed M. Barakat, BDS, PhD* John M. Powers, PhDt Abstract Adhesive bond strengths of glass ionomer and polycarboxylate cements to human enamel and dentine

More information

Original citation: Eslamian, Ladan, Borzabadi-Farahani, Ali, Tavakol, Pegah, Tavakol, Ali, Amini, Nazila and Lynch, Edward. (2015) Effect of multiple debonding sequences on shear bond strength of new stainless

More information

Incognito Appliance System Bonding Protocols Step-by-Step Procedures

Incognito Appliance System Bonding Protocols Step-by-Step Procedures Incognito Appliance System Bonding Protocols Step-by-Step Procedures 2 Incognito Appliance System Bonding Protocols Summary Bonding Protocols Step-by-Step Procedures Adhesives 3 Bonding Dual Cure, RelyX

More information

The procedure for bonding orthodontic brackets

The procedure for bonding orthodontic brackets SHORT COMMUNICATION Examination of enamel-adhesive interface with focused ion beam and scanning electron microscopy Rogelio José Scougall Vilchis, a Yasuaki Hotta, b and Kohji Yamamoto c Mizuho, Japan

More information

UvA-DARE (Digital Academic Repository) A better understanding of orthodontic bracket bonding Algera, T.J. Link to publication

UvA-DARE (Digital Academic Repository) A better understanding of orthodontic bracket bonding Algera, T.J. Link to publication UvA-DARE (Digital Academic Repository) A better understanding of orthodontic bracket bonding Algera, T.J. Link to publication Citation for published version (APA): Algera, T. J. (2009). A better understanding

More information

Chemical and morphological analysis of the human dental enamel treated with argon laser during orthodontic bonding

Chemical and morphological analysis of the human dental enamel treated with argon laser during orthodontic bonding O r i g i n a l A r t i c l e Chemical and morphological analysis of the human dental enamel treated with argon laser during orthodontic bonding Glaucio Serra Guimarães*, Liliane Siqueira de Morais**,

More information

Bond strengths of different orthodontic adhesives after enamel conditioning with the same self-etching primer

Bond strengths of different orthodontic adhesives after enamel conditioning with the same self-etching primer Bond strengths of different orthodontic adhesives after enamel conditioning with the same self-etching primer Rogelio J. Scougall-Vilchis, * Chrisel Zárate-Díaz, Shusuke Kusakabe + and Kohji Yamamoto +

More information

Comparative shear bond strength of some orthodontic bonding resins to enamel

Comparative shear bond strength of some orthodontic bonding resins to enamel European Journal of thodontic} 18 (1996) 89-93 O 1996 European thodontic Society Comparative shear bond strength of some orthodontic bonding resins to enamel L M. Trimpeneers*, R. M. H. Verbeeck**, L R.

More information

Evaluation of Shear Bond Strength Orthodontic Bracket and Adhesive Remaining Materials on the Enamel Surface Using ARI (Adhesive Remnant Index)

Evaluation of Shear Bond Strength Orthodontic Bracket and Adhesive Remaining Materials on the Enamel Surface Using ARI (Adhesive Remnant Index) Cronicon OPEN ACCESS EC DENTAL SCIENCE Research Article Evaluation of Shear Bond Strength Orthodontic Bracket and Adhesive Remaining Materials on the Enamel Surface Using ARI Rina Sutjiati 1 * and Rudy

More information

Comparitive Evaluation of Enamel Demineralisation using Conventional Light Cure Composite Resin and RMGIC- An Invivo Study

Comparitive Evaluation of Enamel Demineralisation using Conventional Light Cure Composite Resin and RMGIC- An Invivo Study I J Pre Clin Dent Res 2015;2(7):1-9 July-September All rights reserved International Journal of Preventive & Clinical Dental Research Comparitive Evaluation of Enamel Demineralisation using Conventional

More information

Cytotoxicity and shear bond strength of four orthodontic adhesive systems

Cytotoxicity and shear bond strength of four orthodontic adhesive systems European Journal of Orthodontics 30 (2008) 495 502 doi:10.1093/ejo/cjn042 Advance Access publication 12 September 2008 The Author 2008. Published by Oxford University Press on behalf of the European Orthodontic

More information

An in vitro study of the bond strength of light-cured glass ionomer cement in the bonding of orthodontic brackets

An in vitro study of the bond strength of light-cured glass ionomer cement in the bonding of orthodontic brackets European Journal of Orthodontics 18 (1996) 199-204 1996 European Orthodontic Society An in vitro study of the bond strength of light-cured glass ionomer cement in the bonding of orthodontic brackets P.

More information

AN 8 MONTH CLINICAL TRIAL OF BOND FAILUES WITH FOUR DIFFERENT TYPES OF ORTHODONTIC ADHESIVES

AN 8 MONTH CLINICAL TRIAL OF BOND FAILUES WITH FOUR DIFFERENT TYPES OF ORTHODONTIC ADHESIVES AN 8 MONTH CLINICAL TRIAL OF BOND FAILUES WITH FOUR DIFFERENT TYPES OF ORTHODONTIC ADHESIVES G. Basaran 1, T. Özer 1, O. Hamamci 2 Dicle University, Orthodontics Department of Faculty Dentistry, Diyarbakır,

More information

The Reinforced Removable Retainer

The Reinforced Removable Retainer The Reinforced Removable Retainer Abstract The aim of this paper is to present a new type of orthodontic removable retainer, which is of great help to the orthodontist as well as to the patient. The procedures

More information

Comparison of Shear Bond Strength of Orthodontic Brackets Bonded to Enamel Prepared By Er:YAG Laser and Conventional

Comparison of Shear Bond Strength of Orthodontic Brackets Bonded to Enamel Prepared By Er:YAG Laser and Conventional Original Article Comparison of Shear Bond Strength of Orthodontic Brackets Bonded to Enamel Prepared By Er:YAG Laser and Conventional Acid-Etching M.H. Hosseini 1, F. Namvar 2, J. Chalipa 1, K. Saber 3,

More information

Near-UV light detection

Near-UV light detection C L I N I C A L Near-UV light detection Javier Tapia Guadix 1 Near-UV light induced fluorescence has already proven to be very useful as an alternative to classic caries-detector dyes. However its potential

More information

Anisotropy of Tensile Strengths of Bovine Dentin Regarding Dentinal Tubule Orientation and Location

Anisotropy of Tensile Strengths of Bovine Dentin Regarding Dentinal Tubule Orientation and Location Original paper Dental Materials Journal 21 (1): 32-43, 2002 Anisotropy of Tensile Strengths of Bovine Dentin Regarding Dentinal Tubule Orientation and Location Toshiko INOUE, Hidekazu TAKAHASHI and Fumio

More information

How to place a lower bonded retainer

How to place a lower bonded retainer Journal of Orthodontics, Vol. 32, 2005, 206 210 FEATURES SECTION How to place a lower bonded retainer A. A. Shah Department of Oral Health and Development, School of Clinical Dentistry, University of Sheffield,

More information