Analytical and Clinical Performance of the epoc Blood Analysis System. Experience at a Large Tertiary Academic Medical Center

Size: px
Start display at page:

Download "Analytical and Clinical Performance of the epoc Blood Analysis System. Experience at a Large Tertiary Academic Medical Center"

Transcription

1 AJCP / Original Article Analytical and Clinical Performance of the epoc Blood Analysis System Experience at a Large Tertiary Academic Medical Center Brie A. Stotler, MD, MPH, 1,2 and Alexander Kratz, MD, PhD 1,2 From the 1 Department of Pathology and Cell Biology, Columbia University College of Physicians and Surgeons, New York, NY, and 2 Clinical Laboratory Services, New York-Presbyterian Hospital, New York, NY. Key Words: Point-of-care testing; Blood gas analysis; Evaluation; epoc ABSTRACT Objectives: To describe validation and performance of epoc, a blood gas analysis point-of-care system, in a live clinical setting. Methods: Data were collected for 156 epoc systems over 12 months. Preimplementation precision and correlation studies and postimplementation quality assurance data were collected, including test card, reader, and personal data assistant (PDA) failure rates. Results: The coefficient of variation was clinically acceptable for all analytes. Correlation studies yielded an R 2 from (for sodium) to (for potassium) with the Nova analyzer and from (sodium) to (glucose) with the i-stat. Average test card failure rate was 13%. Of the PDA/reader units, 55% needed repair within 1 year. Conclusions: The analytical performance showed high precision and good correlation with the Nova and i-stat platforms. Test card and instrument failure rates were higher than that of the i-stat system. Blood gas analysis can be performed in a laboratory or at the point of care (POC). The advantages of central laboratory testing include higher accuracy and precision associated with the use of a few large, well-maintained instruments operated by highly trained laboratory professionals, as well as lower costs due to the cheaper reagents and more efficient use of labor. Transport delays, sample queuing, and the need to look up the results in the electronic medical record have led many hospitals to use POC devices for urgent samples. Although POC instruments provide much faster turnaround times than a central laboratory, the reliability of results can be compromised because of testing by clinical staff, whose primary expertise is not laboratory work. In addition, the cost per test is usually higher for POC assays due to higher reagent/cartridge costs. 1,2 Historically, i-stat instruments (Abbott Point of Care, Princeton, NJ) dominated the POC blood gas and electrolyte testing market in the United States. 3-5 The US Food and Drug Administration approved the epoc blood gas and electrolyte analysis system (Alere, Waltham, MA) in 2006 for testing many of the analytes available on the i-stat platform. The epoc instruments, which use similar technology to the i-stat, are being marketed as a costeffective alternative. In addition to cost, the advantages of epoc include built-in Wi-Fi connectivity and individually barcoded cards that can be stored at room temperature. 6 The analytical performance of the epoc blood analysis system was found to be comparable to the i-stat for ph, CO 2, O 2, sodium, potassium, ionized calcium, and hematocrit. 7 Our institution was among the first large academic medical centers to implement the epoc. This communication describes the performance of the epoc system in a large clinical Downloaded from Am J Clin Pathol 2013;140: Stotler_ indd 715

2 Stotler and Kratz / Analytical and Clinical Performance of epoc setting and its effects on the total use of blood gas analysis. We also extend the prior analytical evaluation to include glucose and lactate, two analytes that were not previously available on the epoc, and correlate the epoc results with central laboratory methods. Materials and Methods epoc Blood Analysis System The epoc system consists of test cards containing the sensors, a wireless card reader, and a personal data assistant (PDA) for data analysis. 7 The test cards use selective electrode potentiometry to measure ph, pco 2, sodium, potassium, and ionized calcium. Glucose, po 2, and lactate are measured amperometrically. Hematocrit is determined conductometrically. Bicarbonate, total carbon dioxide, base excess, oxygen saturation, and hemoglobin are calculated. 8 The epoc card reader is a battery-powered portable device with a barcode scanner. Insertion of a test card into the reader releases calibration fluid over the sensors. After completion of a 190-second calibration cycle, the user has seven minutes to apply a patient sample to the card. The reader then sends the sensor data via Bluetooth to the PDA. The PDA is a handheld computer that calculates analytical values from the data sent by the reader. The host displays the results and sends them via Wi-Fi to the hospital network. Evaluation Site Data were collected at the Columbia University Medical Center (CUMC) campus of New York-Presbyterian Hospital (New York, NY), a tertiary care academic medical center with more than 1,000 adult and pediatric inpatient beds. Active transplant, cardiac, and general surgical services; multiple intensive care units (ICUs); and two emergency departments require fast turnaround times for blood gas results. In 1995, the CUMC introduced i-stat POC devices for measuring blood gases, electrolytes, and hemoglobin in adult and pediatric ICUs, operating rooms, and the pediatric emergency department. Initially, two types of instruments were used: 140 blood analysis modules (BAMs) built into patient monitors in the adult ICUs and 52 portable i-stat clinical analyzers (PCAs) used in the pediatric emergency department, pediatric ICUs, and the operating rooms. Over time, the hospital discontinued using BAMs, and by 2008, only the portable PCAs were in use. Departments that did not use i-stat instrumentation sent samples requiring blood gas analysis to the centralized core laboratory, where they were analyzed with Nova CCX instruments (Nova Biomedical, Waltham, MA). In 2010, the hospital replaced the i-stat instrumentation with epoc blood analysis systems. In addition, epoc instruments were deployed in several locations that were not previously using i-stat devices for blood gas measurements. A total of 156 epoc systems were deployed over approximately three months. Each system consisted of a reader and a PDA. The vendor and the POC service staff trained more than 800 nurses, 300 physicians, 70 respiratory therapists, and 20 perfusionists. Training sessions lasted approximately one hour and included hands-on demonstrations of the instruments. Initial competency was assessed by a written quiz and direct observation of testing using quality control (QC) material. Competency was reassessed six months after initial training using direct observation of patient testing combined with an online review course and quiz. Evaluation Protocol Precision was assessed by analyzing three levels of aqueous QC material (Eurotrol BGEM control fluids [Eurotrol, Burlington, MA] and Mission Blood Gas and Electrolytes Control Calibration Verification Set [Diamond Diagnostics, Holliston, MA]) on 10 consecutive days. The precision data shown in this article are based on 37 readers. Three levels of QC material were run on each reader every day for 10 consecutive days, giving 370 measurements for each level of QC material. The means, standard deviations, and coefficients of variation were then calculated with Excel 2008 (Microsoft, Redmond, WA). Linearity studies were performed on all 156 instruments with five levels of commercially available linearity material (RNA Medical, Devens, MA), with three repetitions of each level. Acceptability criteria were a slope between 0.9 and 1.1 and a recovery between 90% and 110%. Correlation studies were performed on 15 randomly selected epoc systems, representing 10% of the analyzers purchased. Residual sample volume from patient samples collected for clinical analysis on Nova CCX instruments in the core laboratory was used for these studies. Samples collected in lithium heparin syringes were of arterial, mixed venous, and venous types. Approximately 40 patient samples were analyzed on each of the 15 epoc systems and on two i-stat instruments; results from the 15 epoc systems were compared with those from the i-stat analyzers and Nova CCX instruments. Quality Assurance Data After implementation of the epoc blood gas analyzer, the following quality assurance (QA) data were collected with the Electronic Data Manager (Alere) and analyzed in Excel 2008 (Microsoft): number of tests performed, number of tests reported in the electronic medical record, test card failure rate, and error codes associated with test card failures. 716 Am J Clin Pathol 2013;140: Downloaded 716 from Stotler_ indd 716

3 AJCP / Original Article The number of instruments replaced during the first year of use was tracked manually. Comparable QA data for i-stat had been collected previously with an i-stat Central Data Station. Data for the core laboratory were obtained with the laboratory information systems (Mysis, Tucson, AZ, and Cerner Millennium, North Kansas City, MO). Results Analytical Performance Precision studies were performed by analyzing three levels of QC material on 10 consecutive days on each epoc instrument. Summary data from 37 representative instruments are shown in Table 1. Coefficients of variation were generally close to the manufacturer s specifications and met clinical needs (0.1% for ph and up to 0.9% for sodium, 3.3% for pco 2, and 1.4% for potassium). Linearity studies showed recoveries between 90% and 110% on all 156 instruments (data not shown). Fifteen randomly selected instruments were correlated with the Nova CCX and i-stat instruments Table 2. The coefficient of determination for all analytes for both instrument comparisons was greater than The R 2 for all analytes ranged from (sodium) to (potassium) with the Nova instrument and from (sodium) to (glucose) with the i-stat instrument. The bias between the methods was assessed at two clinical decision levels for each analyte and met clinical needs. Test Card Failure Rate The epoc analyzer is an automated self-monitoring system. It monitors internal circuitry and reports problems on the display screen. Messages indicate the cause of the malfunction. Examples of user-generated errors include insufficient amount of sample, sample injected too early, or sample not introduced within seven minutes of calibration. Examples of instrument- or test card generated errors include lack of calibration fluid in the test card or failure of QC of the fluidics channels or sensors. Failure rates of the system were broken down into two main categories: (1) Card Failures, in which the entire test card failed and no assays were performed, and (2) Internal Quality Control (IQC) Errors, in which single or multiple analytes failed to pass the IQC and, thus, individual assays were not reported but other results were available. Table 3 shows the card failure rate of the epoc and i-stat cartridges over a 12-month period. The mean complete card failure rate was 13.0% for epoc and 4.7% for i-stat. IQC errors were less frequent than card failures for both the epoc and the i-stat systems. The mean IQC error rate for epoc was 3.1% during Table 1 Precision Data a Manufacturer s Analyte/Quality Specifications for Control Material CV, CV at Comparable Level Mean (SD) % Level of Analyte, % ph (ph units) CV (0.01) CV (0.01) 0.1 CV (0.01) pco 2, mm Hg CV (2.1) CV (0.9) 2.1 CV (0.7) po 2, mm Hg CV (6.1) CV (6.3) 5.7 CV (8.3) Sodium, meq/l CV (0.8) CV (1.2) 0.9 CV (1.4) Potassium, meq/l CV2 1.8 (0.0) CV3 4.2 (0.1) 1.3 CV4 6.2 (0.1) Ionized calcium, mg/dl CV2 5.6 (0.0) CV3 4.4 (0.0) 1.9 CV4 2.0 (0.0) Glucose, mg/dl CV (1.8) CV3 211 (8.0) 3.8 CV (15.7) Lactate, mg/dl CV2 4.5 (0.1) CV (0.1) 3.6 CV (0.4) Hematocrit, % H (0.4) H (0.5) H (0.6) 0.9 CV, coefficient of variation. a Summary data are presented for 37 epoc analyzers. Each level of quality control (QC) material was run on each analyzer on 10 consecutive days, resulting in 370 measurements per level of QC material. the 12-month study period; thus, for the 106,360 successful test cards used, approximately 3,780 cards had at least one IQC error for an analyte. The mean i-stat IQC error rate was 0.9% in the year before epoc implementation. Table 4 highlights the clinical units that had the highest and lowest overall failure rates using the epoc system. The average failure rate was highest in the adult emergency department (43% total failure rate) and in various ICUs. The lowest failure rates were in the children and adult hospital cardiovascular operating rooms (CVORs). The mean total failure rate in 2011 was 13% in the adult CVOR and 15% in the children s CVOR. In the CVOR locations, only perfusionists perform POC blood gas tests, whereas on all other units these tests are primarily performed by nurses, respiratory therapists, and/or anesthesiologists. The children s Downloaded from Am J Clin Pathol 2013;140: Stotler_ indd 717

4 Stotler and Kratz / Analytical and Clinical Performance of epoc Table 2 Correlation Studies of 15 epoc Systems With Nova CCX and i-stat Instruments No. of Decision Decision Instrument/Analyte Assays a Intercept (Slope) R 2 Level 1 Bias (95% CI) Level 2 Bias (95% CI) epoc vs Nova CCX ph (1.070) ( to 0.021) ( to 0.007) pco (1.073) mm Hg 7.8 (7.4 to 8.1) 50 mm Hg 8.9 (8.6 to 9.1) po (0.950) mm Hg 2.2 ( 3.0 to 1.3) 80 mm Hg 4.6 ( 5.2 to 4.1) Sodium (0.995) meq/l 0.0 ( 0.1 to 0.1) 150 meq/l 0.1 ( 0.3 to 0.2) Potassium (0.928) meq/l 0.14 ( 0.15 to 0.12) 5.8 meq/l 0.34 ( 0.35 to 0.33) Ionized calcium (1.106) mg/dl ( to 0.059) 1.4 mg/dl (.011 to 0.001) Glucose (0.863) mg/dl 1.5 (0.8 to 2.3) 180 mg/dl 17 ( 17.7 to 16.3) Lactate NA NA NA NA NA NA NA Hematocrit (1.289) % 1.86 ( 1.98 to 1.73) 55% 4.36 (4.04 to 4.68) epoc vs i-stat ph (0.949) ( to 0.018) ( to 0.028) pco (1.015) mm Hg 0.4 ( 0.7 to 0.2) 50 mm Hg 0.2 (.04 to 0.0) po (1.002) mm Hg 6.4 (5.7 to 7.1) 80 mm Hg 6.5 (6.0 to 6.9) Sodium (0.938) meq/l 1.4 (1.3 to 1.5) 150 meq/l 0.4 (0.3 to 0.6) Potassium (0.939) meq/l 0.03 (0.02 to 0.04) 5.8 meq/l 0.14 ( 0.15 to 0.13) Ionized calcium (1.015) mg/dl ( to 0.032) 1.4 mg/dl (.029 to 0.024) Glucose (0.959) mg/dl 2.6 (1.9 to 3.4) 180 mg/dl 2.9 ( 3.6 to 2.3) Lactate (0.949) mg/dl 0.2 (0.1 to 0.2) 3.4 mg/dl 0.0 (0.0 to 0.1) Hematocrit (1.087) % 1.41 (1.30 to 1.51) 55% 3.29 (3.02 to 3.55) CI, confidence interval; NA, not available (the Nova CCX does not perform lactate measurements). a Differences in the number of assays performed for each individual analyte are due to internal quality control errors, in which single or multiple analytes fail to pass the internal quality control and no results for the particular analyte is available. Table 3 Failure Rates of epoc vs i-stat Test Cards January- April- July- October- Test Card March % June % September % December % Annual % Rates of complete card failures (ie, no results reported from the card) i-stat epoc Rates of cards with at least one IQC failure (one or more analytes are not reported from a card) i-stat 1.0 NA epoc Sum of complete card failures and IQC result errors i-stat 5.4 NA epoc IQC, internal quality control; NA, not available. Table 4 Units With the Highest and Lowest Test Card Failure Total No. No. of No. of Tests With No. (%) of Combined of Test Successful at Least One IQC No. of Card Card Failures and Location Cards Used Tests Result Error a Failures b IQC Result Errors Adult ED (43) Neurology ICU 1,946 1, (23) Cardiac care unit 4,208 3, (22) Medical ICU 7,592 6, ,307 1,539 (20) Children s hospital respiratory therapists 1,496 1, (20) Adult hospital cardiovascular operating room 22,983 20, ,174 2,923 (13) Children s hospital ICU 21,100 18, ,368 2,993 (14) Children s hospital cardiovascular operating room 5,596 4, (15) ED, emergency department; ICU, intensive care unit; IQC, internal quality control. a IQC error can occur for single or multiple analyte(s) during test performance. No result for one or more individual parameters is given, but other analytes on the test card may be reported. b Card failure is failure of the entire test card in which no results for any analytes are reported. 718 Am J Clin Pathol 2013;140: Downloaded 718 from Stotler_ indd 718

5 AJCP / Original Article hospital ICU also had a low total rate of test card failure, averaging 14% for the year. The epoc card failure rate was significantly higher than the failure rate of the i-stat cartridges in the previous year. Because this was a new system, we instituted retraining programs to units with higher-than-average failure rates within 3 months of initial training; however, no appreciable decline in card failure rates was seen after secondary training. By the end of 2011, all units had used the epoc system for at least six months, yet the test card error rate remained high (only a 1.2% decrease over 12 months from 17% to 15.8%). Reader and Host Failure Rate A total of 156 PDAs and card readers were deployed throughout the institution. Each reader was paired with its own PDA; thus, if one of the two parts was malfunctioning, the entire unit would be returned to the biomedical engineering department. Over the course of 2011, 49 (33%) PDAs and 37 (24%) card readers needed to be repaired and/ or replaced for various reasons. Because our usage model considered the paired readers and PDAs as one unit, 55% of the original inventory of PDA/reader units needed to be repaired and/or replaced within one year of implementation. Common reasons for PDA failures included broken power connectors, damaged reset buttons, and host cradles breaking inside the reader. Common causes for reader malfunction included readers not powering on and failing to read barcodes. Investigation into the mechanical failures discovered multiple potential causes, such as corrosion inside the instruments, blood contamination inside the readers, improper host cradle assembly, and use of an inappropriate power supply leading to battery failure. Usage Data In 2010 before implementation of the epoc system, results of 87,532 blood gas panels were reported using the i-stat system. In 2011, 101,808 blood gas panels were reported with the epoc system, a 16.3% increase in POC blood gas testing. Interestingly, the number of blood gas panels reported from the core laboratory decreased from 103,098 panels in 2010 to 78,519 in 2011, a decrease of 23.8%. Overall, the total number of blood gas panels performed at the CUMC (both at the POC and in the core laboratory) decreased from 190,630 in 2010 to 180,327 in 2011, a decrease of 5.4%. Discussion The epoc blood analysis system showed high precision and good correlation with the i-stat platform for all nine directly measured analytes. This confirms previous findings showing good correlation of the epoc and i-stat systems for seven of these analytes (glucose and lactate were not previously available on the epoc platform). 7 We also found good correlation with the Nova CCX, a central laboratory blood gas instrument. However, there was a clear bias toward lower O 2 levels on the epoc and i-stat instruments relative to the Nova CCX results. Of 122,317 epoc test cards used at the CUMC during the first year after implementation, 15,957 (13%) failed. An additional 3,781 (3.1%) cards yielded only partial results because one or more analytes failed the IQC check (IQC errors). This compared unfavorably with a cartridge failure rate of 4.7% and an IQC error rate of 0.9% for the i-stat instruments during the previous year. In addition, 31% of the epoc PDAs and 24% of the card readers failed and had to be returned to the manufacturer for repair or replacement. The high card failure rate after implementation was expected because the epoc system was new to our institution and was also deployed at several sites within the hospital that did not previously perform POC blood gas testing. During the first year after implementation, the POC department retrained staff in units experiencing the highest card failure rates; however, retraining did not significantly decrease card failure rates. The highest card failure rates (20%-40%) continued to occur in areas that previously did not offer POC blood gas testing, confirming unsatisfactory user competency as a major contributing cause. Units that had a dedicated core group of testing personnel, such as the CVOR perfusionists, had the lowest failure rates (15%). Our usage model viewed each paired epoc PDA and reader as one unit and returned them together to the manufacturer for repair. Therefore, when 31% of our PDAs and 24% of our card readers failed during the first year, the failures affected 55% of our inventory of epoc systems. An alternate usage model could have been implemented whereby any PDA could be used with any reader. With this model, the overall number of damaged inventory at any given time would have been less since only the individual damaged component (ie, PDA or reader), and not both parts, would be returned for service. Had we instituted the latter model, the damaged equipment for the first year would have totaled 86 of 312 PDAs and readers, or 28% of the total inventory. Our hospital replaced 52 i-stat instruments with 156 epoc analyzers. This allowed the deployment of more analyzers to established sites of POC testing, as well as expansion of the POC program to new sites. This resulted in an increase in annual POC blood gas testing by 14,276 panels or 16.3%. The number of blood gas panels performed in the core laboratory decreased by 24,579 or 23.8%. This decrease can be directly attributed to the deployment of epoc analyzers to new POC sites that did not use the i-stat before epoc implementation. Downloaded from Am J Clin Pathol 2013;140: Stotler_ indd 719

6 Stotler and Kratz / Analytical and Clinical Performance of epoc Table 5 Factors That Can Influence the Outcome of a POC System Implementation Factor Clinical reliability of the method Cost of disposables Cost of instruments Availability of infrastructure required Previous experience of clinical staff with similar technology Appropriate training of staff Failure rate of test cards Failure rate of instrumentation Attitude of clinical staff toward the system Practical Examples Accuracy, precision, and correlation with other methods in use at the institution Test cards, reagents, calibrators, QC material Wi-Fi, data cables High test card failure rates lead to delays in diagnosis and care, additional costs for replacement cards, labor cost, and staff frustration High instrument failure rates can cause lack of availability of instrumentation and require additional efforts by POC and IT staff for validation and connection to network Willingness of clinical staff to assume responsibility for testing; willingness to trust the POC results IT, information technology; POC, point of care; QC, quality control. In summary, implementing and expanding a new POC blood gas testing program requires investment from all relevant hospital departments. Our experience provides examples of some factors that can determine the success or failure of such a project Table 5. Challenges faced during this implantation arose from a lack of user familiarity with the test system, a high card failure rate, and inventory supply problems due to the high PDA/reader mechanical failure rate. The POC department had to invest a significant amount of time retraining staff and orchestrating equipment repairs. Address reprint requests to Dr Kratz: New York-Presbyterian Hospital, Core Laboratory, 622 W 168th St, PH3-363, New York, NY 10032; ak2651@columbia.edu. Acknowledgments: We thank Steven Spitalnik, MD, for critical reading of the manuscript and Anthony Sireci, MD, for statistical advice. 2. Cox CJ. Acute care testing: blood gases and electrolytes at the point of care. Clin Lab Med. 2001;21: Schneider J, Dudziak R, Westphal K, et al. The i-stat analyzer: a new, hand-held device for the bedside determination of hematocrit, blood gases, and electrolytes [in German]. Anaesthesist. 1997;46: Dascombe BJ, Reaburn PR, Sirotic AC, et al. The reliability of the i-stat clinical portable analyser. J Sci Med Sport. 2007;10: Erickson KA, Wilding P. Evaluation of a novel point-of-care system, the i-stat portable clinical analyzer. Clin Chem. 1993;39: epoc Blood analysis Accessed November 23, Nichols JH, Rajadhayaska A, Rodriguez M. Evaluation of the enterprise point-of-care (EPOC) system for blood gas and electrolyte analysis. Point Care. 2008;7: Epocal. epoc System Manual. 6th ed. Ottawa, Ontario, Canada: Epocal; References 1. Holloway PAH. Point-of-care testing in intensive care. In: Kost GJ, ed. Principles and Practice of Point-of-Care Testing. Philadelphia, PA: Lippincott Williams & Wilkins; 2002: Am J Clin Pathol 2013;140: Downloaded 720 from Stotler_ indd 720

Comparison of Three Whole Blood Creatinine Methods for Estimation of Glomerular Filtration Rate Before Radiographic Contrast Administration

Comparison of Three Whole Blood Creatinine Methods for Estimation of Glomerular Filtration Rate Before Radiographic Contrast Administration Clinical Chemistry / Whole Blood Creatinine for egfr Comparison of Three Whole Blood Creatinine Methods for Estimation of Glomerular Filtration Rate Before Radiographic Contrast Administration Nichole

More information

Polymer Technology Systems, Inc. CardioChek PA Comparison Study

Polymer Technology Systems, Inc. CardioChek PA Comparison Study Polymer Technology Systems, Inc. CardioChek PA Comparison Study Evaluation Protocol: Accuracy Precision Clinical Correlation PTS Panels Lipid Panel Test Strips For Use in Comparisons to a Reference Laboratory

More information

FreeStyle Lite A Blood Glucose Meter That Requires No Coding

FreeStyle Lite A Blood Glucose Meter That Requires No Coding Journal of Diabetes Science and Technology Volume 2, Issue 4, July 2008 Diabetes Technology Society SYMPOSIUM Shridhara, Ph.D. Abstract Background: Abbott Diabetes Care introduced the FreeStyle Lite blood

More information

ASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE

ASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE ASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE Brooke Ann Spaeth, Anne K Shephard, Mark DS Shephard, Timothy H Mathew ABSTRACT

More information

Patient Safety: A Quality System Approach To POCT QC/QA

Patient Safety: A Quality System Approach To POCT QC/QA Patient Safety: A Quality System Approach To POCT QC/QA Ellis Jacobs, Ph.D., DABCC New York University School of Medicine Coler-Goldwater Specialty Hospital & Nursing Facility New York, New York Point-of-Care

More information

The good, the bad, the ugly A month in the life of a Point Of Care Coordinator

The good, the bad, the ugly A month in the life of a Point Of Care Coordinator The good, the bad, the ugly A month in the life of a Point Of Care Coordinator Stefanie Van Heule Ghent University Hospital 23/10/2014 Ghent University Hospital ± 3000 patients / day ± 62 000 total admissions

More information

Basic knowledge of POCT instuments. The Tide Resort Bangsan Beach 18 June 2009

Basic knowledge of POCT instuments. The Tide Resort Bangsan Beach 18 June 2009 Basic knowledge of POCT instuments The Tide Resort Bangsan Beach 18 June 2009 1 POCT instruments concepts Point of care testing Diagnostic medical devices used at or near the site of patient care Bed site,

More information

Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6

Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 Original: 5/02 Minor Revision: 9/15 Full Review: 1/15 Replaces: 1/15 Responsible Person: Point of Care Laboratory Specialist Approving Committee:

More information

I. Subject: Ionized Calcium (Ca++) Analysis Whole Blood

I. Subject: Ionized Calcium (Ca++) Analysis Whole Blood I. Subject: Ionized Calcium (Ca++) Analysis Whole Blood II. Method: i-stat III. Principle: A. Ca++: is measured by ion-selective electrode potentiometry. Concentrations are calculated from the measured

More information

i-stat Alinity v Utilization Guide

i-stat Alinity v Utilization Guide istat Alinity v Utilization Guide The istat Alinity v delivers blood gas, acidbase, electrolyte, chemistry, and hematology results in a completely portable, handheld package. Accuracy is ensured by extensive

More information

CME/SAM. An Examination of the Usefulness of Repeat Testing Practices in a Large Hospital Clinical Chemistry Laboratory

CME/SAM. An Examination of the Usefulness of Repeat Testing Practices in a Large Hospital Clinical Chemistry Laboratory Clinical Chemistry / Repeat Testing Practices in Clinical Chemistry An Examination of the Usefulness of Repeat Testing Practices in a Large Hospital Clinical Chemistry Laboratory Carl O. Deetz, MD, PhD,

More information

Virtual Lectures Planning Committee Disclosure Summary

Virtual Lectures Planning Committee Disclosure Summary Mayo Medical Laboratories Virtual Lectures 2014 MFMER MFMER Virtual Lectures Planning Committee Disclosure Summary As a provider accredited by ACCME, College of Medicine, Mayo Clinic (Mayo School of CPD)

More information

Clinical Evaluation of the Accuracy and Precision of the CDI 500 In-line Blood Gas Monitor With and Without Gas Calibration

Clinical Evaluation of the Accuracy and Precision of the CDI 500 In-line Blood Gas Monitor With and Without Gas Calibration The Journal of ExtraCorporeal Technology Original Articles Clinical Evaluation of the Accuracy and Precision of the CDI 500 In-line Blood Gas Monitor With and Without Gas Calibration Anne Louise Bellaiche,

More information

Point of Care (POC) Glucose Testing Revised on: December, 2018 Reference: Policy LAB 23-A; Point of Care Nova Stat Strip Glucose Meter

Point of Care (POC) Glucose Testing Revised on: December, 2018 Reference: Policy LAB 23-A; Point of Care Nova Stat Strip Glucose Meter Point of Care (POC) Glucose Testing Revised on: December, 2018 Reference: Policy LAB 23-A; Point of Care Nova Stat Strip Glucose Meter The Nova Stat Strip Blood Glucose meter is UHB s approved Glucometer

More information

Summary of Significant Changes at this Revision

Summary of Significant Changes at this Revision COPY Summary of Significant Changes at this Revision CR5890 Removal of Ketone guidelines - refer to Specialist Diabetes nurse (adult wards); Children's ward has their own protocol Remove all references

More information

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: k100322 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE B. Purpose for Submission: Clearance of a new device C. Measurand: Whole

More information

PROCEDURE NO. POC LBH. Printed copies are for reference only. Please refer to the electronic copy for the latest version.

PROCEDURE NO. POC LBH. Printed copies are for reference only. Please refer to the electronic copy for the latest version. Department Of Pathology POC.514.06- Blood Glucose Monitoring Accu-Chek Inform II Procedure-LBH Version# 6 Department PROCEDURE NO. PAGE NO. Point-of-Care Testing POC.514.05 LBH 1 OF 7 Printed copies are

More information

PROCEDURE. TITLE: Bedside Glucose Monitoring PC Laboratory. Issuing Department: Clinical Director Signature: Departments Involved:

PROCEDURE. TITLE: Bedside Glucose Monitoring PC Laboratory. Issuing Department: Clinical Director Signature: Departments Involved: PROCEDURE TITLE: Bedside Glucose Monitoring Issuing Department: Clinical Director Signature: Departments Involved: Laboratory Nursing Effective Date: 10/97 Review Dates: 09/01, 07/02, 05/13 Revision Dates:

More information

i-stat Alinity v Utilization Guide

i-stat Alinity v Utilization Guide istat Alinity v Utilization Guide The istat Alinity v delivers blood gas, acidbase, electrolyte, chemistry, and hematology measurements in a completely portable, handheld package. Accuracy is ensured by

More information

Abbott PXP Glucose Meters Resource: Lin Gustafson, MT(ASCP) Point of Care Coordinator, Laboratory

Abbott PXP Glucose Meters Resource: Lin Gustafson, MT(ASCP) Point of Care Coordinator, Laboratory Abbott PXP Glucose Meters 2015 Resource: Lin Gustafson, MT(ASCP) Point of Care Coordinator, Laboratory Objectives After you complete this Computer-Based Learning (CBL) module, you should be able to: Explain

More information

i-stat CHEM8 + Cartridge Quick Reference Guide

i-stat CHEM8 + Cartridge Quick Reference Guide i-stat CHEM8 + Cartridge Quick Reference Guide For use with a Certificate of Waiver The CHEM8+ cartridge has tests for sodium, potassium, chloride, total carbon dioxide, ionized calcium, glucose, urea

More information

Management of Central Venous Access Devices. Blood Glucose Monitoring

Management of Central Venous Access Devices. Blood Glucose Monitoring Management of Central Venous Access Devices Blood Glucose Monitoring Purpose To provide education on the standard of care regarding the use and monitoring of the Accu- Chek Blood glucose machine, including

More information

Title: Abbott Optium Xceed Glucose Meter Effective date: 16/05/2011. Summary of Significant Changes at this Revision. SOP updated.

Title: Abbott Optium Xceed Glucose Meter Effective date: 16/05/2011. Summary of Significant Changes at this Revision. SOP updated. COPY Summary of Significant Changes at this Revision SOP updated Purpose and Scope Items Required 1. The Abbott Optium Xceed meter is a battery-powered device designed for the measurement of blood glucose

More information

Written By: Joann O Connor Date: 03/05/2017 Effective Date: 03/28/2017 1

Written By: Joann O Connor Date: 03/05/2017 Effective Date: 03/28/2017 1 Department Of Pathology Point of Care Testing POC.108.07 i-stat 1-NW Version#7 Printed copies are for reference only. Please refer to the electronic copy for the latest version. A. Purpose: The i-stat

More information

Analytical performance of a portable critical care blood gas analyzer

Analytical performance of a portable critical care blood gas analyzer Ž. Clinica Chimica Acta 307 2001 107 112 www.elsevier.comrlocaterclinchim Analytical performance of a portable critical care blood gas analyzer Harald Schlebusch a,), Iris Paffenholz b, Rainer Zerback

More information

POTASSIUM/K. 3.6 mmol/l

POTASSIUM/K. 3.6 mmol/l POTASSIUM/K Potassium is measured by ion-selective electrode potentiometry. In the calculation of results for potassium, concentration is related to potential through the Nernst equation. The i-stat System

More information

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 System Accuracy Evaluation of Project number: Fora092614-01 System accuracy evaluation of Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Date: 13 th

More information

Comparison between Manual and Automated Methods for Determination of Canine and Feline Hematocrit and Hemoglobin Concentration

Comparison between Manual and Automated Methods for Determination of Canine and Feline Hematocrit and Hemoglobin Concentration Kasetsart J. (Nat. Sci.) 4 : 655-659 (8) Comparison between and d Methods for Determination of Canine and Feline Hematocrit and Hemoglobin Concentration Kreangsak Prihirunkit *, Chalermpol Lekcharoensuk

More information

Determination of hemoglobin is one of the most commonly

Determination of hemoglobin is one of the most commonly ORIGINAL ARTICLE Multiple-Site Analytic Evaluation of a New Portable Analyzer, HemoCue Hb 201+, for Point-of-Care Testing Sten-Erik Bäck, PhD,* Carl G. M. Magnusson, PhD, Lena K. Norlund, MD, PhD, Henning

More information

Immediate Pathology Results Now Available for all Remote Northern Territorians

Immediate Pathology Results Now Available for all Remote Northern Territorians Immediate Pathology Results Now Available for all Northern Territorians Brooke Spaeth, BMedSc (Hons) Coordinator - Northern Territory Point-of-Care Testing Program Flinders University International Centre

More information

C30-A2 ISBN ISSN Point-of-Care Blood Glucose Testing in Acute and Chronic Care Facilities; Approved Guideline Second Edition

C30-A2 ISBN ISSN Point-of-Care Blood Glucose Testing in Acute and Chronic Care Facilities; Approved Guideline Second Edition ISBN 1-56238-471-6 ISSN 0273-3099 Point-of-Care Blood Glucose Testing in Acute and Chronic Care Facilities; Approved Guideline Second Edition Volume 22 Number 17 David B. Sacks, M.D., Chairholder Patricia

More information

Patient: 55 y female (ambulatory)

Patient: 55 y female (ambulatory) Disclosures Speaking Honoraria Radiometer (Canada) Nova Biomedical, Draeger Roche Diagnostics (Canada) Research Support (Reagents, Instrumentation, Travel) Nova Biomedical Abbott Laboratories (Canada)

More information

POC Testing: The New Standard. InnovaStar From DiaSys.

POC Testing: The New Standard. InnovaStar From DiaSys. POC Testing: The New Standard. InnovaStar From DiaSys. Easy to use, superior precision, rapid results: InnovaStar from DiaSys brings POC testing to a whole new level. A fully automatic, compact analyzer

More information

Summary of Significant Changes at this Revision. Items Required

Summary of Significant Changes at this Revision. Items Required Update Approver/checker Summary of Significant Changes at this Revision Purpose and Scope Items Required 1. The Abbott FreeStyle Optium meter is a battery-powered device designed for the measurement of

More information

Patient Safety: A Quality System Approach To POCT QC/QA

Patient Safety: A Quality System Approach To POCT QC/QA Patient Safety: A Quality System Approach To POCT QC/QA Ellis Jacobs, Ph.D., DABCC New York University School of Medicine Coler-Goldwater Specialty Hospital & Nursing Facility New York, New York Point-of-Care

More information

Kevin K. Nunnink Extracorporeal Membrane Oxygenation Program

Kevin K. Nunnink Extracorporeal Membrane Oxygenation Program PATIE NT I N FO R M ATI O N Kevin K. Nunnink Extracorporeal Membrane Oxygenation Program A family s guide to understanding this specialized treatment for cardiac and pulmonary complications Saint Luke

More information

The Laboratories Role in Global Health

The Laboratories Role in Global Health The Laboratories Role in Global Health Larry Westerman, Ph.D. International Laboratory Branch Division of Global HIV/AIDS Centers for Disease Control and Prevention Center for Global Health International

More information

Alere Technologies GmbH

Alere Technologies GmbH Alere Technologies GmbH Mobile Smart Instruments for Automated Image Analysis and Point of Care Diagnosis April 2011 Alere Technologies GmbH Alere Technologies GmbH Based in Jena, Thuringia, Germany ~300

More information

Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers

Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D iochemistry Analyzers Life Sciences Data for Life. TM YSI Life Sciences White Paper 91 Authors: Kevin Schlueter, PhD,

More information

Parkland Health & Hospital System Women & Infant Specialty Health

Parkland Health & Hospital System Women & Infant Specialty Health Parkland Health & Hospital System Women & Infant Specialty Health NS 1700.04 Nursery Services Procedure Manual Arterial Puncture Practice Statement Upon the written order of the provider, the credentialled

More information

Title/Description: Point of Care Glucose Testing-Waived Department: Organization Wide Personnel: All Effective Date: November 29, 2016 Revised:

Title/Description: Point of Care Glucose Testing-Waived Department: Organization Wide Personnel: All Effective Date: November 29, 2016 Revised: Title/Description: Point of Care Glucose Testing-Waived Department: Organization Wide Personnel: All Effective Date: November 29, 2016 Revised: PURPOSE The Abbott FreeStyle Precision Pro (FSPP) Blood Glucose

More information

HEMOCHRON. Whole Blood Coagulation Systems

HEMOCHRON. Whole Blood Coagulation Systems HEMOCHRON Whole Blood Coagulation Systems Citrated Activated Partial Thromboplastin Time (APTT) Cuvette Correlation Protocol for HEMOCHRON Microcoagulation Instruments MSIG:131 10/06 Dear Medical Professional:

More information

It s a Gas! Issues in the Blood Gas Laboratory. D. Robert Dufour, MD, FCAP Consultant Pathologist VA Medical Center, Washington DC

It s a Gas! Issues in the Blood Gas Laboratory. D. Robert Dufour, MD, FCAP Consultant Pathologist VA Medical Center, Washington DC It s a Gas! Issues in the Blood Gas Laboratory D. Robert Dufour, MD, FCAP Consultant Pathologist VA Medical Center, Washington DC Learning Objectives After participating in this session, you will be able

More information

NEW TECHNOLOGIES FOR MANAGING DIABETES ANGELA THOMPSON DNP, FNP-C, BC-ADM, CDE, FAANP

NEW TECHNOLOGIES FOR MANAGING DIABETES ANGELA THOMPSON DNP, FNP-C, BC-ADM, CDE, FAANP NEW TECHNOLOGIES FOR MANAGING DIABETES ANGELA THOMPSON DNP, FNP-C, BC-ADM, CDE, FAANP No commercial support or sponsorship was received for this project I have nothing to disclose OBJECTIVES Identify at

More information

cobas h 232 POC system The power to make decisions in cardiac emergencies

cobas h 232 POC system The power to make decisions in cardiac emergencies cobas h 232 POC system The power to make decisions in cardiac emergencies cobas h 232 POC system Rapid results at the Point of Care (POC) Wherever and whenever you face an emergency situation, measurement

More information

Quantitative Measurement of Emergency Biomarkers in Parallel

Quantitative Measurement of Emergency Biomarkers in Parallel Point-of-Care Immunoassay Analyzer Quantitative Measurement of Emergency Biomarkers in Parallel Troponin I, Myoglobin, CK-MB, D-Dimer, NTproBNP, hscrp Results in 17 minutes Direct measurement from whole

More information

Relationship between glucose meter error and glycemic control efficacy

Relationship between glucose meter error and glycemic control efficacy Relationship between glucose meter error and glycemic control efficacy Brad S. Karon, M.D., Ph.D. Professor of Laboratory Medicine and Pathology Department of Laboratory Medicine and Pathology Mayo Clinic

More information

Validity and Reliability of a Glucometer Against Industry Reference Standards

Validity and Reliability of a Glucometer Against Industry Reference Standards 514315DSTXXX10.1177/1932296813514315Journal of Diabetes Science and TechnologySalacinski et al research-article2014 Original Article Validity and Reliability of a Glucometer Against Industry Reference

More information

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY INFORMATION BULLETIN USING THE ACCESS AMH ASSAY IN YOUR LABORATORY ///////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////

More information

Evaluation of the Precision and Accuracy of the INRatio 2 Prothrombin Time (PT) Monitoring System

Evaluation of the Precision and Accuracy of the INRatio 2 Prothrombin Time (PT) Monitoring System Evaluation of the Precision and Accuracy of the INRatio 2 Prothrombin Time (PT) Monitoring System Grace DeSantis, PhD Lisa Croner, PhD Michealle Havenhill, BA Shari Kipp, BSc (Honours) MBA Inverness Medical,

More information

Determination of Delay in :flirn Around Time (TAT) of Stat Tests and its Causes: an AKUH Experience

Determination of Delay in :flirn Around Time (TAT) of Stat Tests and its Causes: an AKUH Experience Determination of Delay in :flirn Around Time (TAT) of Stat Tests and its Causes: an AKUH Experience F. Bilwani,I. Siddiqui,S. Vaqar ( Section of Chemical Pathology, Department of Pathology, Aga Khan University

More information

Multiple studies are available concerning the use of

Multiple studies are available concerning the use of SPECIAL ARTICLE Evaluation of the Biosite Ò Quantitative Whole Blood D-dimer Assay and Comparison With the biomérieux VIDAS Ò D-dimer Exclusion Test Validation and Utility For Use in the Central Laboratory

More information

GLYCEMIC CONTROL SURVEY

GLYCEMIC CONTROL SURVEY GLYCEMIC CONTROL SURVEY Objective: To gain an understanding of the current state of glycemic control (ie, intensive insulin therapy and frequent blood glucose testing) protocol use in hospital inpatients.

More information

GlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data

GlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data GlucoMen LX PLUS: Accuracy Evaluations to ISO 15197:2013 with Technical and Specification Data GlucoMen LX PLUS blood glucose and blood ketone meter Accuracy Evaluation to New ISO 15197:2013, with Technical

More information

The Virtues and Pitfalls of Implementing a New Test

The Virtues and Pitfalls of Implementing a New Test The Virtues and Pitfalls of Implementing a New Test James H. Nichols, Ph.D., DABCC, FACB Professor of Clinical Pathology, Microbiology and Immunology Associate Medical Director for Clinical Operations

More information

Compatibility of Na and K values measured by biochemistry analyser and blood gas analyser in different clinical situations

Compatibility of Na and K values measured by biochemistry analyser and blood gas analyser in different clinical situations ORIGINAL ARTICLE Compatibility of Na and K values measured by analyser and blood gas analyser in different clinical situations Ali Haydar Akça *, Muhammed İkbal Şaşmaz Department Of Emergency Medicine,

More information

Quantitative Measurement of Cardiac Biomarkers

Quantitative Measurement of Cardiac Biomarkers Cardiac Biomarker Analyzer Quantitative Measurement of Cardiac Biomarkers Troponin I, Myoglobin, CK-MB, D-Dimer, NTproBNP, hscrp Results in 17 minutes Direct measurement from whole blood samples 6 parallel

More information

Method Comparison Report Semi-Annual 1/5/2018

Method Comparison Report Semi-Annual 1/5/2018 Method Comparison Report Semi-Annual 1/5/2018 Prepared for Carl Commissioner Regularatory Commission 123 Commission Drive Anytown, XX, 12345 Prepared by Dr. Mark Mainstay Clinical Laboratory Kennett Community

More information

Point of Care Testing (POCT)

Point of Care Testing (POCT) Accounting for Pre-analytical Variables in POCT L.V. Rao, PhD Senior Clinical Laboratory Director Associate Professor, Pathology UMass Memorial Medical Center (Worcester, MA) Senior Scientific Director

More information

Alere Triage System. Knowing now matters when rapid treatment decisions can be made at the point-of-care.

Alere Triage System. Knowing now matters when rapid treatment decisions can be made at the point-of-care. System Knowing now matters when rapid treatment decisions can be made at the point-of-care. The Alere Triage System The MeterPro is utilized in thousands of healthcare facilities worldwide. A comprehensive

More information

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Abstract Objective: The goal of the study was to assess

More information

TITLE: Whole Blood Glucose Monitoring Using the Nova StatStrip Meter

TITLE: Whole Blood Glucose Monitoring Using the Nova StatStrip Meter Page 4 of 14 IV. Performance of Quality Control A. Purpose: To detect any errors due to system failures or operator performance B. Frequency: Levels 1 and 3 controls are run every 24 hours of use. C. Quality

More information

Interference and Point-of-Care Testing Devices

Interference and Point-of-Care Testing Devices Interference and Point-of-Care Testing Devices Nam K. Tran, PhD, HCLD, (ABB), FACB Associate Clinical Professor Director of Clinical Chemistry, Special Chemistry/Toxicology and POCT 1 Learning Objectives

More information

Delta Check Calculation Guide

Delta Check Calculation Guide Delta Check Calculation Guide National Technology 2017, All Rights Reserved By Senior Scientific Researcher, Asmaa Taher Table of Contents Definition... 2 Purpose... 2 Delta Check Research Studies... 2

More information

DIAGNOSTIC ACCREDITATION PROGRAM. Spirometry Quality Control Plan

DIAGNOSTIC ACCREDITATION PROGRAM. Spirometry Quality Control Plan DIAGNOSTIC ACCREDITATION PROGRAM Spirometry Quality Control Plan Table of Contents Introduction...1 Spirometry Definitions and Requirements...2 Spirometry Requirements... 2...4 Daily Quality Control (see

More information

Admission of patient CVICU and hemodynamic monitoring

Admission of patient CVICU and hemodynamic monitoring Admission of patient CVICU and hemodynamic monitoring Prepared by: Rami AL-Khatib King Fahad Medical City Pi Prince Salman Heart tcentre CVICU-RN Admission patient to CVICU Introduction All the patients

More information

Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007

Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007 Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007 Thomas J. Spira, M.D. International Laboratory Branch Global AIDS Program Centers for Disease

More information

METHOD VALIDATION: WHY, HOW AND WHEN?

METHOD VALIDATION: WHY, HOW AND WHEN? METHOD VALIDATION: WHY, HOW AND WHEN? Linear-Data Plotter a,b s y/x,r Regression Statistics mean SD,CV Distribution Statistics Comparison Plot Histogram Plot SD Calculator Bias SD Diff t Paired t-test

More information

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014 Comparison Study Summary Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR 97123 July 30, 2014 1. PROTOCOL This study was conducted on July 24 th, 2014 at Tuality Healthcare, Hillsboro, OR. The

More information

M.D.IPA, M.D.IPA Preferred, Optimum Choice and Optimum Choice Preferred STAT Laboratory List Revised Jan. 5, 2017

M.D.IPA, M.D.IPA Preferred, Optimum Choice and Optimum Choice Preferred STAT Laboratory List Revised Jan. 5, 2017 M.D.IPA, M.D.IPA Preferred, Optimum Choice and Optimum Choice Preferred STAT Laboratory List Revised Jan. 5, 2017 If laboratory results are required on a STAT basis, the designated commercial medical laboratory

More information

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803) Comparison Study Summary Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC 29720 (803) 286-1480 February 10, 2016 1 PROTOCOL This evaluation was conducted on February 10, 2016 at Springs Memorial

More information

SODIUM/NA. See below for information on factors affecting results. Certain substances, such as drugs, may affect analyte levels in vivo.

SODIUM/NA. See below for information on factors affecting results. Certain substances, such as drugs, may affect analyte levels in vivo. SODIUM/NA Sodium is measured by ion-selective electrode potentiometry. In the calculation of results for sodium, concentration is related to potential through the Nernst equation. The i-stat System uses

More information

Hemostasis Test Validation, Performance and Reference Intervals

Hemostasis Test Validation, Performance and Reference Intervals Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma City VA Medical Center University of Oklahoma Health Sciences Center

More information

Point of Care testing refers to all laboratory testing that is done outside of the walls of the clinical laboratory in the proximity of the patient.

Point of Care testing refers to all laboratory testing that is done outside of the walls of the clinical laboratory in the proximity of the patient. 1 2 Point of Care testing refers to all laboratory testing that is done outside of the walls of the clinical laboratory in the proximity of the patient. All such tests are considered lab tests and are

More information

pat hways Medtech innovation briefing Published: 9 September 2015 nice.org.uk/guidance/mib38

pat hways Medtech innovation briefing Published: 9 September 2015 nice.org.uk/guidance/mib38 pat hways i STAT T CG4+ and CHEM8+ cartridges for point- of-care testing in the emergency department Medtech innovation briefing Published: 9 September 2015 nice.org.uk/guidance/mib38 Summary The i-stat

More information

Omnitest : 2013

Omnitest : 2013 ISO 15197: 2013 Omnitest 3 Blood Glucose Monitoring System Omnitest 3 Could it be easier? After strip insertion, Omnitest 3 is ready for the measurement. The code is set automatically. Only 0.3 µl sample

More information

Implementing a Modern NDT Inspection Program Utilizing Connected Instruments and Cloud Software Solutions

Implementing a Modern NDT Inspection Program Utilizing Connected Instruments and Cloud Software Solutions Implementing a Modern NDT Inspection Program Utilizing Connected Instruments and Cloud Software Solutions Dianne Hillhouse Business Development Manager Oil/Gas/Chemical Market Olympus Scientific Solutions

More information

Current Status: Active PolicyStat ID: Accu-Chek Whole Blood Glucose Testing Policy PURPOSE

Current Status: Active PolicyStat ID: Accu-Chek Whole Blood Glucose Testing Policy PURPOSE Current Status: Active PolicyStat ID: 4749849 Origination: 12/2013 Last Approved: 03/2018 Last Revised: 03/2018 Next Review: 03/2020 Owner: Area: References: Karla Gonzalez Anchondo: Point of Care Coordinator

More information

Biosensor and its electrochemical transducer

Biosensor and its electrochemical transducer YINAN TANG Background Biosensor and its electrochemical transducer Simple, inexpensive, accurate and sensitive platform Inherent miniaturization for both the detector and control instrumentation Application

More information

A wireless medical sensor application and the problems it poses

A wireless medical sensor application and the problems it poses A wireless medical sensor application and the problems it poses Odd Nordland SINTEF ICT Trondheim Norway odd.nordland@sintef.no 1 The scenario 2 The scenario Diabetes patient Needs regular medical attention

More information

The challenges of quality assessment for point-of-care testing. Anne Stavelin. EQALM Symposium, Bergen 8 October General challenges

The challenges of quality assessment for point-of-care testing. Anne Stavelin. EQALM Symposium, Bergen 8 October General challenges The challenges of quality assessment for point-of-care testing Anne Stavelin EQALM Symposium, Bergen 8 October 2015 www.noklus.no Norwegian quality improvement of primary care laboratories General challenges

More information

DOELAP On-Site Assessment Requirements Checklist Page 1 of 24 Participant:

DOELAP On-Site Assessment Requirements Checklist Page 1 of 24 Participant: DOELAP On-Site Assessment Requirements Checklist Page 1 of 24 General Requirements Y, N, N/A Demonstration of Conformance G.1 Latest version of protocols or procedures G.2 Latest version of dosimeter specifications

More information

Mass Spectrometry Made Simple for Clinical Diagnostic Labs

Mass Spectrometry Made Simple for Clinical Diagnostic Labs Mass Spectrometry Made Simple for Clinical Diagnostic Labs The SCIEX Topaz System with FDA-Cleared Topaz Vitamin D 200M Assay Kit Mass Spectrometry Made Simple SCIEX Topaz System for Clinical Diagnostics

More information

C RITICAL C ARE 2. Angelo Manzoni, Susanna Zanardi, Pietro Zonca Jay Johnson. Glucose biosensor for whole blood analysis

C RITICAL C ARE 2. Angelo Manzoni, Susanna Zanardi, Pietro Zonca Jay Johnson. Glucose biosensor for whole blood analysis C RITICAL C ARE 2 Angelo Manzoni, Susanna Zanardi, Pietro Zonca Jay Johnson Glucose biosensor for whole blood analysis *Angelo Manzoni, *Susanna Zanardi, *Pietro Zonca **Jay Johnson * Instrumentation Laboratory

More information

Empowering the RT with New Noninvasive Monitoring Capabilities

Empowering the RT with New Noninvasive Monitoring Capabilities Empowering the RT with New Noninvasive Monitoring Capabilities Thomas Lamphere BS, RRT, RPFT Executive Director, Pennsylvania Society for Respiratory Care Adjunct Instructor, Respiratory Care Program,

More information

Quicker Diagnostic Testing of Cardiac Conditions in the ER

Quicker Diagnostic Testing of Cardiac Conditions in the ER Transcript Details This is a transcript of an educational program accessible on the ReachMD network. Details about the program and additional media formats for the program are accessible by visiting: https://reachmd.com/programs/clinicians-roundtable/quicker-diagnostic-testing-of-cardiac-conditions-inthe-er/3630/

More information

Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100)

Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100) Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on CUBE analyser (CA0100) Location Location: Eurolyser Diagnostica GmbH Operators: Simone Wieser; Franz Helminger; Michael Gruber Date: July-November

More information

Medtronic MiniMed Insulin Infusion Pumps

Medtronic MiniMed Insulin Infusion Pumps Medtronic MiniMed Insulin Infusion Pumps Patients should always discuss potential risks and benefits with a physician. Please review the product manual prior to use for detailed instructions and disclosure.

More information

MEDICAL DEVICES. our advice on. automated external. defibrillators

MEDICAL DEVICES. our advice on. automated external. defibrillators MEDICAL DEVICES our advice on automated external defibrillators An Automated External Defibrillator (AED) is a medical device that analyses a person s heart rhythm and, when needed, delivers a shock to

More information

Clinical Evaluation for. Embrace No Code Blood Glucose Monitoring System

Clinical Evaluation for. Embrace No Code Blood Glucose Monitoring System Embrace No Code Report Clinical Evaluation for Embrace No Code Blood Glucose Monitoring System Applicant: Apex Biotechnology Corp. Test Date: 2007/8/28~2007/9/21 Clinical Site: SCHMIDT Group Practice Clinic

More information

NORMAL LABORATORY VALUES FOR CHILDREN

NORMAL LABORATORY VALUES FOR CHILDREN Pediatric Drug Lookup Normal Laboratory Values for NORMAL LABORATORY VALUES FOR CHILDREN CHEMISTRY Normal Values Albumin 0-1 y 2.0-4.0 g/dl 1 y to adult 3.5-5.5 g/dl Ammonia Newborns 90-150 mcg/dl 40-120

More information

Determination of Ethanol in Breath and Estimation of Blood Alcohol Concentration with Alcolmeter S-D2

Determination of Ethanol in Breath and Estimation of Blood Alcohol Concentration with Alcolmeter S-D2 Determination of Ethanol in Breath and Estimation of Blood Alcohol Concentration with Alcolmeter S-D2 A.W. Jones and KÄ. Jönsson Departments of Alcohol Toxicology and Internal Medicine, University Hospital,

More information

Summary of Significant Changes at this Revision. Items Required. Printed copies are uncontrolled unless there is an allocated Copy Number on page 1

Summary of Significant Changes at this Revision. Items Required. Printed copies are uncontrolled unless there is an allocated Copy Number on page 1 COPY Summary of Significant Changes at this Revision CR 5279 Update section 10 as per re-wording by NP Purpose and Scope 1. The Abbott Precision Xceed Pro/FreeStyle Precision Pro meter is a battery-powered

More information

Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System

Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Abstract Objective: Both Fora Premium V10 and V30 blood glucose

More information

Session 15 Improved Outcomes and a Proven ROI Model for Quality Improvement: Transforming Diabetes Care

Session 15 Improved Outcomes and a Proven ROI Model for Quality Improvement: Transforming Diabetes Care Session 15 Improved Outcomes and a Proven ROI Model for Quality Improvement: Transforming Diabetes Care Charles G Macias MD, MPH Chief Clinical Systems Integration Officer Director of Evidence-Based Outcomes

More information

Human Hemoglobin Colorimetric Detection Kit

Human Hemoglobin Colorimetric Detection Kit Human Hemoglobin Colorimetric Detection Kit CATALOG NO: IRAAKT2522 LOT NO: SAMPLE INTENDED USE The Hemoglobin detection kit is designed to quantitatively measure all forms of hemoglobin present in blood

More information

TIP CARD. Model Temporary External Pacemaker

TIP CARD. Model Temporary External Pacemaker TIP CARD Model 5392 Temporary External Pacemaker PRE-USE What to do... What NOT to do... Battery Replace the battery for each new patient. Do NOT reuse battery. Physical Check case for cracks/damage.**

More information

SAGE. Nick Beard Vice President, IDX Systems Corp.

SAGE. Nick Beard Vice President, IDX Systems Corp. SAGE Nick Beard Vice President, IDX Systems Corp. Sharable Active Guideline Environment An R&D consortium to develop the technology infrastructure to enable computable clinical guidelines, that will be

More information