1 Blood Pressure Measurement Can Isik Electrical Engineering and Computer Science Department Syracuse University Syracuse, NY Abstract Blood pressure is one of the fundamental vital signs, and its measurement is of great importance to medical professionals and the general public alike. This chapter describes direct and indirect methods of measuring blood pressure. There are many techniques of blood pressure measurement in the market, and more are being investigated in the labs, and tested in the field. A number of these techniques are covered here, giving particular emphasis to the oscillatory techniques, extensively employed in devices targeted for clinical use as well as those targeted for home use. Also included are limitations of different techniques, possible remedies to those limitations, and an analysis of the accuracy of a variety of blood pressure measurement devices. Introduction Blood pressure is an important signal in determining the functional integrity of the cardiovascular system. Scientists and physicians have been interested in blood pressure measurement for a long time. The first blood pressure measurement is attributed to Reverend Stephen Hales, who in the early 18 th century connected water-filled glass tubes in the arteries of animals, and correlated their blood pressures to the height of the column of fluid in the tubes. It wasn t until the early 20 th century when the blood pressure measurement was introduced into clinical medicine, albeit with many limitations. Blood pressure measurement techniques are generally put in two broad classes, direct and indirect. Direct techniques of blood pressure measurement, which are also known as invasive techniques, involve a catheter to be inserted in the vascular system. The indirect techniques are non-invasive, with improved patient comfort and safety, but at the expense of accuracy. The accuracy gap between the invasive and non-invasive methods, however, has been narrowing with the increasing computational power available in portable units, which can crunch elaborate signal processing algorithms in a fraction of a second. During a cardiac cycle, blood pressure goes through changes, which correspond to the contraction and relaxation of the cardiac muscle, with terminology that identify different aspects of the cycle. The maximum and minimum pressures over a cardiac cycle are 1

2 called the systolic and diastolic pressures, respectively. The time average of the cardiac pressure over a cycle is called the mean pressure, and the difference between the systolic and diastolic pressures is called the pulse pressure. Table 1. Classification of blood pressure for adults Category Systolic - mmhg Diastolic - mmhg Normal Less than 120 and Less than 80 Prehypertension or Stage 1 Hypertension or Stage 2 Hypertension 160 or higher or 100 or higher Normal blood pressure varies with age, state of health, and other individual conditions. An infant s typical blood pressure is 80/50 mmhg (Systolic/Diastolic). The normal blood pressure increases gradually, and reaches 120/80 for a young adult. Blood pressure is lower during sleep and during pregnancy. Many people experience higher blood pressures in the medical clinic, a phenomenon called the white coat effect. Therefore the ranges given in Table 1 are used as guidelines, rather than diagnostic facts. Direct Techniques The operation of direct measurement techniques can be summarized in very simple terms: they all utilize a pressure transducer that is coupled to the vascular system through a catheter or cannula that is inserted to a blood vessel, followed by a microcontroller unit with electronics and algorithms for signal conditioning, signal processing and decision making. There are many advantages of this set of techniques, including: the pressure is measured very rapidly, usually within one cardiac cycle; the measurement is done to a very high level of accuracy and repeatability; the measurement is continuous, resulting in a graph of pressure against time; the measurement is motion tolerant. Therefore, the direct techniques are utilized when it is necessary to accurately monitor patients vital signs, for example during critical care and in the operating room. Although direct techniques have a lot in common, there are differences in the details of various approaches. Extravascular transducers: the catheter in this type of devices is filled with a saline solution, which transmits the pressure to a chamber that houses the transducer assembly. As a minor disadvantage, this structure affects the measured pressure through the dynamic behavior of the catheter. Since the catheter has a known behavior, this effect can be minimized to insignificant levels through computational compensation (Gibbs). 2

3 Intravascular transducers: The transducer is at the tip of the catheter in this type of devices. Then, the measured signal is not affected by the hydraulics of the fluid in the catheter. The catheter diameter is larger in this class of transducers. Transducer technology: There is a wide spectrum of transducer technologies available to build either kind of transducer. They include metallic or semiconductor strain gages, piezoelectric, variable capacitance, variable inductance, and optical fibers. Appropriate driver and interface circuitry accompanies each technology (Webster, pp ). Other applications of direct pressure measurement: Another advantage of direct measurement techniques is that, they are not limited to measuring the simple arterial pressure. They can be used to obtain central venous, pulmonary arterial, left atrial, right atrial, femoral arterial, umbilical venous, umbilical arterial, and intracranial pressures, by inserting the catheter in the desired site (Hambly). Figure 1. Blood pressure waveform, and systolic, diastolic, and mean pressures, from an invasive monitor screen. (Philips). Sources of errors: Direct blood pressure measurement systems have the flexibility of working with a variety of transducers/probes. It is important that the probes are matched with the appropriate compensation algorithm. Most of the modern equipment does this matching automatically, eliminating the possibility of operator error. An additional source of error occurs when air bubbles get trapped in the catheter. This changes the fluid dynamics of the catheter, causing an unintended mismatch between the catheter and its signal processing algorithm. This may cause distortions in the waveforms and errors in the numeric pressure values extracted from them. It is difficult to recognize this artifact from the waveforms, so it is best to avoid air bubbles in the catheter. Non-Invasive (Indirect) Techniques An overwhelming majority of blood pressure measurements do not require continuous monitoring or extreme accuracy. Therefore non-invasive techniques are used in most cases, maximizing patient comfort and safety. Currently available devices for noninvasive measurement are Manual devices: these devices use the auscultatory technique; Semi-automatic devices: these devices use oscillatory techniques; Automatic devices: while a majority of these devices use oscillatory techniques, there are some that use pulse-wave velocity or plethysmographic methods. 3

4 The Auscultatory Technique In the traditional, manual, indirect measurement system, an occluding cuff is inflated, and a stethoscope is used to listen to the sounds made by the blood flow in the arteries, called Korotkov sounds. When the cuff pressure is above the systolic pressure, blood cannot flow, and no sound is heard. When the cuff pressure is below the diastolic pressure, again, no sound is heard. A manometer connected to the cuff is used to identify the pressures where the transitions from silence to sound to silence are made. This combination of a cuff, an inflating bulb with a release valve, and a manometer is called a sphygmomanometer, and the method, auscultatory technique. Usually, the cuff is placed right above the elbow, elevated to the approximate height of the heart, and the stethoscope over the brachial artery. It is possible to palpate the presence of pulse under the cuff, rather than to use a stethoscope to listen to the sounds. The latter approach works especially well in noisy places where it is hard to hear the heart sounds. This method has various sources of potential error. Most of these sources are due to misplacement of the cuff, problems with hearing soft sounds, and using the wrong cuff size. Using a small cuff on a large size arm would result in overestimating the blood pressure, and vice versa. Nevertheless, an auscultatory measurement performed by an expert health care professional using a clinical grade sphygmomanometer is considered to be the gold standard in non-invasive measurements. Oscillatory Techniques Most automatic devices base their blood pressure estimations on the variations in the pressure of the occluding cuff, as the cuff is inflated or deflated. These variations are due to the combination of two effects: the controlled inflation or deflation of the cuff, and the effect of the arterial pressure changes under the cuff. The Korotkov sounds are not utilized in the oscillatory techniques. The cuff pressure variation data may be collected while the cuff is being inflated, or deflated. Furthermore, the inflation or deflation during the data collection may be controlled in a continuous fashion, or in a step-wise fashion. This variability gives four different strategies in data collection. Their differences may seem insignificant at first, but have significant effects on the way a variety of algorithms are designed. 4

5 Figure 2. Cuff pressure trajectory when data are collected during stepwise deflation of the cuff (Colak2). Pulse Wave Amplitude (mmhg) Baseline Pressure (mmhg) Figure 3. Pulse wave amplitude profile (Colak1) Data in Figure 2 were collected using an experimental system. The cuff is first rapidly inflated to a value higher than the anticipated systolic pressure, an approximate pressure of 170 mmhg in this case. Then it is deflated in small steps until the cuff pressure is below the anticipated diastolic pressure, around 50 mmhg. Please note that when the cuff pressure is very high or very low, the arterial blood pressure variations contribute 5

6 very little to the cuff pressure trajectory. As a matter of fact, the height of those pulses above the cuff pressure baseline is at their maximum when the baseline pressure is equal to the mean arterial pressure (MAP). We demonstrate this in Figure 3, with a plot of pulses relative to their baseline pressure (pulse wave amplitude), against their respective baseline cuff pressures. Please note that only a few of the pulses seen in Figure 2 are transferred to Figure 3 to maintain clarity mmhg Samples Figure 4. Cuff pressure trajectory and pulse wave amplitude when data are collected during continuous inflation of the cuff. Figure 4 shows a cycle of data collected during the continuous inflation of the cuff, as well as the pulse wave amplitude. The pulse wave amplitude is obtained by subtracting the baseline cuff pressure from the raw pressure data. Next, we will return to the example developed in Figures 2 and 3, and continue with the estimation of blood pressure values. It seems trivial to pick the pulse with the tallest height above baseline, and select its baseline pressure to be the MAP. So, for the example at hand, MAP would be just under100 mmhg, as shown in Figure 5. The systolic and diastolic pressures are then estimated from the MAP using a variety of heuristic rules. A common class of these heuristic rules works as follows. First, the peak values (heights) of the pulse wave amplitudes are connected to form an envelope. Again, the baseline pressure at the peak of this envelope is the MAP value. Then, the height of the MAP pulse is reduced by a predetermined systolic ratio, and the intersection of this systolic height with the envelope to the right of the MAP pulse is selected as the systolic location. The baseline pressure at this location is assigned as the estimate of the systolic pressure, as depicted in 6

7 Figure 5. The diastolic pressure is estimated in a similar fashion, by using a ratio of its own to arrive at the diastolic height, and then finding the corresponding intersection with the envelope to the left of the MAP pulse. Pulse Wave Amplitude (mmhg) Peak MAP Height Diastolic Height Systolic Height (mmhg) Diastolic Pressure MAP Systolic Pressure Figure 5. Blood pressure estimation from pulse wave amplitude profile In the example shown in Figure 5, the systolic ratio and diastolic ratio were arbitrarily selected as 0.5 and 0.7, respectively. In a realistic system, those ratios would be found statistically (using methods such as regression, fuzzy rule based systems, neural networks, or evolutionary algorithms) so as to minimize deviations between estimated and actual blood pressure values. Algorithmic Components of Blood Pressure Measurement In the earlier measurement units it was a combination of hardware and software that controlled the various aspects of the automated measurement (or estimation) of blood pressure. With the ever increasing computational power of microcontrollers, all decision making and control are now implemented in software, and with more elaborate algorithms. Here are some of the functions that are included in a measurement system. Please refer to Figure 6 for a typical organization of such algorithms in a blood pressure measurement system. Inflation/deflation control: Whether data collection is done during inflation or deflation, continuously or in steps, there are many challenges to appropriately controlling the air pump. They include maintaining a smooth baseline cuff 7

8 pressure without filtering out the arterial variations, adjusting the pump speed to variations arising from different cuff sizes, arm sizes, and cuff tightness, and selecting the range of cuff pressures for which data will be collected. Pulse detection: This is a fundamental part of extracting features from raw cuff pressure data. It becomes especially challenging when conditions such as arrhythmia, or tremors, affect the regularity of pulses. Pattern recognition techniques with features found in time, frequency, or wavelet domains are used to deal with difficult situations. Blood pressure estimation: The indirect method of measurement is a process of estimating pressures with the use of features extracted from cuff-pressures or other transducer data. This algorithm used to be limited to linear interpolation, as described in the example of Figure 5. Recently, more elaborate decision making and modeling tools such as nonlinear regression, neural networks and fuzzy logic also are being used for this purpose. Algorithm Management/ Cycle Control Inflation/ Deflation Control Cuff Blood Pressure Estimation/ Display Pulse Profile/ Feature Extraction Cuff Pressure Measurement Figure 6. A typical organization of algorithmic components of (oscillatory) blood pressure measurement. Sources of Inaccuracy There are many factors that contribute to the inaccuracies in the automated measurement of blood pressure. The following are some of the more significant sources of error. Sparseness of data: An important design criterion of a blood pressure monitor is to go through a cycle as quickly as possible. However, the faster a device functions, the fewer pulses it will have in a cycle. A cycle time of one minute would yield about pulses, while a 20-minute cycle would have only pulses. The oscillatory techniques are based on collecting cuff pressure due to pulses at baseline pressures that change from above systolic to below diastolic. If we divide a cuff pressure range of about mmhg by the number of pulses in a cycle, we can see that the baseline increment between successive pulses 8

9 varies from 2-3 mmhg in a one-minute cycle to 6-9 mmhg in a 20-second cycle. This quantization error affects accuracy in the estimate of the mean arterial pressure, as well as the shape of the pulse envelope, hence the accuracy of the systolic and diastolic values. Various curve-fitting and interpolation techniques are used to remedy this problem. Pulse extraction uncertainty: Whether the baseline cuff pressure is varied continuously or in steps, figuring out where one pulse ends and another one starts is not a trivial matter. An inspection of Figure 2 will show that there are many artifacts in the data stream that may confuse a pulse extraction algorithm and cause errors in the pulse wave amplitude profile in Figure 3. In addition, common factors such as an irregularity in the pulses as in arrhythmia, small wrinkles or folds in the cuff changing its volume suddenly during data collection, or small movements of the patient may amplify those artifacts. A variety of pattern recognition techniques are employed to improve the accuracy of pulse detection (Osowski). Motion artifacts: the performance of the oscillatory techniques depends on all measurements during a cycle. Therefore any error caused by a motion of the patient may affect the accuracy of the blood pressure estimations. A comparative study of six non-invasive devices has found that average percent errors due to motion artifacts may be as high as 39% (Revision). Remedies to this source of error may be a combination of three strategies: (1) to identify and compensate for minor artifacts, (2) to identify and discard data that include significant artifacts or to repeat the entire cycle if the estimates are deemed unreliable, (3) to incorporate features from additional sensors or monitors such as EKG to help identify motion artifacts (Dowling), (Revision). Other Blood Pressure Measurement Techniques Oscillometry is by far the most common technique in automatic non-invasive blood pressure measurement. However, there are other methods that are found in commercial units or in units that are being developed. In this section few of these methods are summarized, and references are given for further information. It should be noted that algorithmic components and sources of inaccuracy presented within the context of oscillatory technique may apply to other automated measurement methods. Arterial Tonometry: this relatively new technique in blood pressure measurement is inspired by the tonometry devices that were made in the mid 1950s to measure intraocular pressure. The arterial tonometry device is based on a pressure sensor and pneumatic actuator combination, which is placed on the wrist, above the radial artery. When the pressure applied on the artery is adjusted to the appropriate level (called the hold-down pressure), the portion of the artery wall which is facing the actuator is partially flattened. This configuration maximizes the energy transfer between the artery and the sensor, yielding pulses with the highest amplitude. The relative amplitudes of the tonometry pulses are calibrated to the systolic and diastolic pressures. Tonometry is suitable for continuous monitoring applications. Sensor placement sensitivity, calibration difficulties, and motion sensitivity are problems that need improvement (Sato) (Matthys). 9

10 Pulse Wave Velocity: Pulse wave is generated by the heart as it pumps blood, and travels ahead of the pumped blood. By solving analytical equations of fluid dynamics, it has been shown that changes in blood pressure heavily depend on changes in pulse wave velocity. Blood pressure can be continuously calculated from pulse wave velocity, which in turn is calculated from EKG parameters and peripheral pulse wave measured by an SpO 2 probe on the finger or toe. This method is suitable for continuous monitoring, as well as for detecting sudden changes in blood pressure to trigger an oscillometric cycle (Williams). Plethysmographic methods: In this method, changes in the blood volume during a cardiac cycle are sensed using a light emitter and receiver at the finger. Tissue and blood have different infrared light absorbance characteristics. That is, the tissue is practically transparent to the infrared light, while blood is opaque to it. A prototype of a ring-like sensor/signal processor/transmitter combination has been reported (Yang), (Rhee). Different Forms of Blood Pressure Measurement Devices The techniques, algorithms, and transducers discussed in the previous sections have led to a variety of forms of devices, differentiated by where in the body the measurements are taken, or for what purpose the device is used. Ambulatory Blood Pressure Monitoring: These are portable and wearable devices that monitor the patient s blood pressure over a long period, say for 24 hours. While the patient is following her daily routine, the device periodically takes measurements and saves the results. These measurements are later downloaded for analysis by a physician. First ambulatory devices, introduced in the early 1960s were quite rudimentary and used tape recorders to capture the Korotkoff sounds with an occluding cuff. Most current ambulatory devices utilize the oscillatory technique. Since the patient is subjected to repeated blood pressure measurements with an ambulatory device, it is essential to improve motion tolerance, patient comfort, measurement time, and of course overall accuracy of measurement algorithms that are employed in ambulatory monitors. (McGrath) Ambulatory devices have been instrumental in clinical research and practice. Through their use, there have been significant improvements in our understanding of blood pressure dynamics in a variety of physiological and psychological conditions, and concepts such as white-coat hypertension, episodic hypertension, circadian rhythm of blood pressure (e.g., daytime/nighttime variations of blood pressure) have been investigated and added to the medical lexicon. (National) Wrist Blood Pressure Monitoring: These monitors have smaller cuffs than their upperarm-attached counterparts. Hence they are more compact and more conducive to selfmeasurement. It is important that the monitors are held at the heart-level for correct 10

11 measurement. They are popular with the home users, but typically less accurate then the full-size arm monitors. Finger Blood Pressure Monitoring: Finger monitors are not nearly as common as the arm or wrist monitors. The approaches used are auscultatory and plethysmographic. Semi-Automatic Blood Pressure Monitoring: The semi-automatic devices have cuffs that are inflated manually by an attached bulb, like a sphygmomanometer. Once the cuff is inflated, the monitor functions in the same manner as an automatic device, taking cuffpressure measurements while releasing the pressure in a controlled way. These devices are more economical and have longer battery lives than their fully automated counterparts. Accuracy of Blood Pressure Measurement Devices Blood pressure measurement devices play an important role in medicine, as they measure one of the fundamental vital signs. In addition to this traditional use, non-invasive blood pressure devices, especially the automated ones, have become ubiquitous in the homes, regularly used by lay people. Two widely used protocols for testing the accuracy of these devices are those set by the Association for the Advancement of Medical Instrumentation (AAMI), a pass/fail system published in 1987 and revised in 1993, and the protocols of the British Hypertension Society (BHS), an A-D graded system, established in 1990 and revised in These protocols describe in detail the process manufacturers should follow in validating the accuracy of their devices. Their numeric accuracy thresholds can be summarized as follows. A device would pass the AAMI protocols if its measurement error has a mean of no more than 5 mmhg, and a standard deviation of no more than 8 mmhg. The BHS protocol would grant a grade of A to a device if in its measurements 60% of the errors are within 5 mmhg, 85% of the errors are within 10 mmhg, and 95% within 15 mmhg. BHS has progressively less stringent criteria for the grades of B and C, and assigns a grade D if a device performs worse than C. The European Society of Hypertension introduced in 2002 the International Protocol for validation of blood pressure measuring devices in adults (O Brien 2002). The working group that developed this protocol had the benefit of analyzing a large number of studies performed according to the AAMI and BHS standards. One of their motivations was to make the validation process simpler, without compromising its ability to assess the quality of a device. They achieved it by simplifying the rules for selecting subjects for the study. Another change was to devise a multi-stage process that recognized devices with poor accuracy early on. This is a pass/fail process, using performance requirements with multiple error bands. Whether blood pressure measurement devices are used by professionals or lay people, their accuracy is important. Yet, most devices in the market have not been evaluated for accuracy independently, using the established protocols (O Brien 2001). In their study, O Brien et al. surveyed published independent evaluations of manual 11

12 sphygmomanometers, automated devices for clinical use, and automated devices for personal use. If a device was found acceptable by AAMI standards, and received a grade of A or B by BHS standards, for both systolic and diastolic measurements, then it was recommended. Otherwise it was not recommended. Few studies they surveyed had issues such as specificity, so devices reported in those studies were questionably recommended. Table 2. Summary of accuracy of blood pressure measurement devices Device Type Number Surveyed Recommended? Yes Questionable No Manual, Clinical Auto, Clinical Auto, Home, Arm Auto, Home, Wrist Ambulatory Total Table 2 summarizes the result of their survey. It is interesting to note that of the four clinical grade sphygmomanometers, a kind that is highly regarded by health care providers, only one was recommended. Overall, the number of devices not recommended is more than the number of recommended devices. What one should take away from this analysis is that at every level of quality, price, and target market, it is essential to research the accuracy of a device before investing in it and relying on it. References Colak, S., Isik, C., Fuzzy Pulse Qualifier, 23rd International Conference of the North American Fuzzy Information Processing Society (NAFIPS 2004) Proceedings, Banff, June Colak, S., Isik, C., Blood Pressure Estimation Using Neural Networks, IEEE International Conference on Computational Intelligence for Measurement Systems and Applications, Boston, MA, July Dowling, Jr., N.B., Measuring Blood Pressure in Noisy Environments, U.S. Patent No. 6,258,037 B1, July 10, Gibbs, N.C., Gardner, R.M., Dynamics of invasive pressure monitoring systems: clinical and laboratory evaluation, Heart Lung, Vol. 17, pp ,

13 Hambly, P., Measuring The Blood Pressure, Update in Anaesthesia, Issue 11, Article 6, Matthys K., Verdonck P., Development and modelling of arterial applanation tonometry: a review, Technology and Health Care, 2002, Vol.10, No. 1, pp McGrath, B.P., Ambulatory blood pressure monitoring, The Medical Journal of Australia, 2002; Vol. 176, no. 12, pp National High Blood Pressure Education Program (NHBPEP) Working Group Report On Ambulatory Blood Pressure Monitoring, NIH Publication No , Reprinted February Available at: O Brien, E, Atkins, N., Staessen, J., State of the market: a review of ambulatory blood pressure monitoring devices, Hypertension. 1995, No. 26, pp O'Brien E, Waeber B, Parati G, Staessen J, Myers M.G., Blood pressure measuring devices: recommendations of the European Society of Hypertension, BMJ (British Medical Journal), Aug 2001, No. 323, Vol.18, No. 323 (7309):398. Available at O Brien, E, Pickering, T, Asmar, R, Myers, M, Parati, G, Staessen, J, Mengden, T, Imai, Y, Waeber, B, and Palatini, P, Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults, Blood Pressure Monitoring, 2002, Vol 7, pp Available at Osowski, S, Linh, T.H., ECG beat recognition using fuzzy hybrid neural network, IEEE Trans. Biomedical Engineering, Vol. 48, No. 11, Nov. 2001, pp Philips Invasive Monitoring literature at Revision Labs, Beaverton, OR, Noninvasive Blood Pressure Measurement and Motion Artifact: A Comparative Study, December 3, Available at Rhee, S., Yang, B.H., Asada, H.H., Artifact-resistant power-efficient design of fingerring plethysmographic sensors, IEEE Trans. Biomedical Engineering, Vol. 48, No. 7, July 2001, pp Sato, T., Nishinaga, M., Kawamoto, A., Ozawa, T., Takatsuji, H., Accuracy of a continuous blood pressure monitor based on arterial tonometry, Hypertension, Vol. 21, ,

14 U.S. Food And Drug Administration, Non-Invasive Blood Pressure (NIBP) Monitor Guidance. March 10, Available at Webster, J.G., Editor, Medical Instrumentation: Application and Design, 3 rd Ed., Wiley, Williams, B., Pulse wave analysis and hypertension: evangelism versus skepticism, Journal of Hypertension, 2004, Vol. 22, pp Yang, B.H., Asada, H.H., Zhang, Y., Cuff-less continuous monitoring of blood pressure, d Arbeloff Laboratory of Information Systems and Technology, MIT, Progress Report No.2-5, March 31, Available at 14

External Oscillatory Blood Pressure - EOBPTM

External Oscillatory Blood Pressure - EOBPTM External Oscillatory Blood Pressure - EOBPTM Development of Novel Principle To Measure Blood Pressure Mindaugas Pranevicius, M.D., Osvaldas Pranevicius, M.D., Ph.D. Pranevicius Biotech Inc., Forest Hills,

More information

A GUIDE TO OUR NIBP TECHNOLOGY

A GUIDE TO OUR NIBP TECHNOLOGY GE Healthcare THE DINAMAP DIFFERENCE A GUIDE TO OUR NIBP TECHNOLOGY OUR TECHNOLOGICAL ADVANTAGES THE OSCILLOMETRIC METHODOLOGY Oscillometry is the most commonly used means of indirect blood pressure measurement

More information

Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 ( The story so far.

Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 (  The story so far. Medical Electronics Dr Neil Townsend Michaelmas Term 2001 (wwwrobotsoxacuk/~neil/teaching/lectures/med_elec) Blood Pressure has been measured (in some way) for centuries However, clinical interpretation

More information

CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME

CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME 64 CHAPTER 4 ESTIMATION OF BLOOD PRESSURE USING PULSE TRANSIT TIME 4.1 GENERAL This chapter presents the methodologies that are usually adopted for the measurement of blood pressure, heart rate and pulse

More information

1 Non-invasive measurement of arterial pressure

1 Non-invasive measurement of arterial pressure Non-invasive measurement of arterial pressure I. Background A. Circulatory systems Human circulation takes place in a closed system that consists of two subsystems, pulmonary circulation and systemic circulation,

More information

1 Department of Medical Physics, Royal Infirmary of Edinburgh, 2 Department of. Received 11 June 2004 Accepted 7 September 2004

1 Department of Medical Physics, Royal Infirmary of Edinburgh, 2 Department of. Received 11 June 2004 Accepted 7 September 2004 Devices and technology 1 Do SpaceLabs ambulatory non-invasive blood pressure recorders measure blood pressure consistently over several years use? John N. Amoore 1, David Dewar 2, Kathleen Gough 3 and

More information

Blood Pressure Estimation Using Photoplethysmography (PPG)

Blood Pressure Estimation Using Photoplethysmography (PPG) Blood Pressure Estimation Using Photoplethysmography (PPG) 1 Siddhi Sham Karande, BE E&TC, VIIT Pune. 2 Kiran Rajendrasingh Thakur, BE E&TC, VIIT Pune. 3 Sneha Dattatraya Waghmare, BE E&TC, VIIT Pune 4

More information

Biomedical Instrumentation E. Blood Pressure

Biomedical Instrumentation E. Blood Pressure Biomedical Instrumentation E. Blood Pressure Dr Gari Clifford Adapted from slides by Prof. Lionel Tarassenko Blood pressure Blood is pumped around the body by the heart. It makes its way around the body

More information

AFFORDABLE TECHNOLOGY

AFFORDABLE TECHNOLOGY World Health Organization AFFORDABLE TECHNOLOGY BLOOD PRESSURE MEASURING DEVICES FOR LOW RESOURCE SETTINGS CARDIOVASCULAR DISEASES Blood Pressure Measurement in Low Resource Settings Annex 1: Blood Pressure

More information

Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension

Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension M. EL ASSAAD, J. TOPOUCHIAN, G. LABAKI, R. ASMAR L Institut

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 261 Validation of the ROSSMAX blood pressure measuring monitor according to the European Society of Hypertension International Protocol for Validation of Blood Pressure Measuring

More information

DR JIRAR TOPOUCHIAN PROF PAROUNAK ZELVEIAN PROF ROLAND ASMAR. September 8 th, Principal Investigator and Study Chair:

DR JIRAR TOPOUCHIAN PROF PAROUNAK ZELVEIAN PROF ROLAND ASMAR. September 8 th, Principal Investigator and Study Chair: Page 1 of 10 Validation of the automatic blood pressure measurements device, the OMRON EVOLV (HEM- 7600 T-E) in Pregnancy according to the Modified European Society of Hypertension International Protocol

More information

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model Instrumentation and Control Department, JSS Academy of Technical Education, Noida (U.P.), India Abstract- The non-invasive

More information

Clinical Research Coordinator Skills Program

Clinical Research Coordinator Skills Program Clinical Research Coordinator Skills Program Vital Signs This Skill Requires Provider direction Standard Precautions Using purell before and after contact with the patient or the patient s environment

More information

Taking and recording blood pressure and pulse. City Gate Training Centre all rights reserved

Taking and recording blood pressure and pulse. City Gate Training Centre all rights reserved Taking and recording blood pressure and pulse City Gate Training Centre all rights reserved Course Content: Definition of Blood Pressure Normal Blood Pressure Readings Systolic and Diastolic Blood Pressure

More information

#6 - Cardiovascular III Heart Sounds, Pulse Rate, Hemoglobin Saturation, and Blood Pressure

#6 - Cardiovascular III Heart Sounds, Pulse Rate, Hemoglobin Saturation, and Blood Pressure #6 - Cardiovascular III Heart Sounds, Pulse Rate, Hemoglobin Saturation, and Blood Pressure Objectives: Observe slide of artery and vein cross-section Auscultate heart sounds using a stethoscope Measure

More information

Vital Signs. Vital Signs. Pulse. Temperature. Respiration. Blood Pressure

Vital Signs. Vital Signs. Pulse. Temperature. Respiration. Blood Pressure Vital Signs Jarvis, Chapter 9 Vital Signs Classic Vital Signs TPR/BP Temperature Pulse Respirations Blood Pressure Additional Vital Signs Height Weight BMI (Kg/m2) or (702Xlbs/in2) Supine, orthostatic

More information

How To Measure Vital Signs

How To Measure Vital Signs How To Measure Vital Signs How to Use This Tutorial This tutorial is intended for healthcare providers or students to teach basic vital signs skills Use the navigation buttons below to move through the

More information

CAMOSUN COLLEGE BIOLOGY 144 (2010) LABS

CAMOSUN COLLEGE BIOLOGY 144 (2010) LABS LAB 8: CARDIOVASCULAR PHYSIOLOGY PART 1. HEART SOUNDS AND PULSE DETERMINATIONS Introduction Two distinct sounds can be heard during each cardiac cycle. These sounds are commonly described as lub and dup

More information

and bias, which are known to be present in self-home and in professional office BP measurements taken using the auscultatory technique [7].

and bias, which are known to be present in self-home and in professional office BP measurements taken using the auscultatory technique [7]. Devices and technology 37 Validation of the A&D UM-11 professional hybrid device for office blood pressure measurement according to the International Protocol George S. Stergiou, Periklis P. Giovas, Charilaos

More information

Ambulatory Services Orientation & Skill Review Skill Performance Checklist: Vital Signs. Name: Date:

Ambulatory Services Orientation & Skill Review Skill Performance Checklist: Vital Signs. Name: Date: Evaluation Levels Behavior Level 3 Met Performed all skills competently. No behaviors omitted. Level 2 Satisfactory Omitted no more than three (3) non-critical behaviors. Level 1 Unsatisfactory Omitted

More information

ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE

ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE Vet Times The website for the veterinary profession https://www.vettimes.co.uk ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE Author : CARL BRADBROOK Categories : Vets Date : October 7, 2013

More information

Estimation of Systolic and Diastolic Pressure using the Pulse Transit Time

Estimation of Systolic and Diastolic Pressure using the Pulse Transit Time Estimation of Systolic and Diastolic Pressure using the Pulse Transit Time Soo-young Ye, Gi-Ryon Kim, Dong-Keun Jung, Seong-wan Baik, and Gye-rok Jeon Abstract In this paper, algorithm estimating the blood

More information

Received 10 April 2008 Revised 23 June 2008 Accepted 24 June 2008

Received 10 April 2008 Revised 23 June 2008 Accepted 24 June 2008 342 Original article Development of an accurate oscillometric blood pressure device for low resource settings Annemarie de Greeff, Hannah Nathan, Nina Stafford, Bing Liu and Andrew H. Shennan Objective

More information

a Medical Physics Department, Guy s & St Thomas NHS Foundation Trust and b King s College School of Medicine, St Thomas Campus, London, UK

a Medical Physics Department, Guy s & St Thomas NHS Foundation Trust and b King s College School of Medicine, St Thomas Campus, London, UK Devices and technology 27 Validation of the Omron 705IT (HEM-759-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol Andrew Coleman a, Paul Freeman

More information

PHONOCARDIOGRAPHY (PCG)

PHONOCARDIOGRAPHY (PCG) PHONOCARDIOGRAPHY (PCG) The technique of listening to sounds produced by the organs and vessels of the body is called auscultation. The areas at which the heart sounds are heard better are called auscultation

More information

Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination

Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination Dual-Cuff Method Improves Accuracy of Blood Pressure and Hemodynamics Determination JIRI JILEK 1, MILAN STORK 2 1 Carditech, Culver City, California, USA jilekj@usa.net 2 Dept. of Applied Electronics and

More information

Protocol 4: Measuring Blood Pressure

Protocol 4: Measuring Blood Pressure Zool 430L Protocols Page 8 of 12 Protocol 4: Measuring Blood Pressure 4.1 Via Auscultation (listening through stethoscope) The estimate of blood pressure is obtained by cutting off blood flow with a known

More information

HUMAN ANATOMY AND PHYSIOLOGY

HUMAN ANATOMY AND PHYSIOLOGY HUMAN ANATOMY AND PHYSIOLOGY NAME Detection of heart sounds. Clean the ear pieces of the stethoscope before using. The ear pieces should be pointing slightly forward when inserted into the ears because

More information

2. Measure a subject's blood pressure and heart rate both at rest and during exercise.

2. Measure a subject's blood pressure and heart rate both at rest and during exercise. Lab Activity 11 The Cardiovascular System Student Learning Objectives After completing this lab, you should be able to: 1. Define, explain and correctly use the key terms. 2. Measure a subject's blood

More information

High Blood Pressure Blood Pressure Monitoring at Home -- F...

High Blood Pressure Blood Pressure Monitoring at Home -- F... Return to Web version High Blood Pressure Blood Pressure Monitoring at Home Why should I measure my blood pressure at home and keep a record of it? Measuring your blood pressure at home and keeping a record

More information

Blood Pressure Laboratory

Blood Pressure Laboratory Introduction The blood that circulates throughout the body maintains a flow and pressure. The nervous system can change the flow and pressure based on the particular needs at a given time. For example,

More information

Heart Abnormality Detection Technique using PPG Signal

Heart Abnormality Detection Technique using PPG Signal Heart Abnormality Detection Technique using PPG Signal L.F. Umadi, S.N.A.M. Azam and K.A. Sidek Department of Electrical and Computer Engineering, Faculty of Engineering, International Islamic University

More information

science-u.org What affects blood Blood Pressure Directions You Will Need BEST FOR GRADES 5-8 ESTIMATED TIME Minutes

science-u.org What affects blood Blood Pressure Directions You Will Need BEST FOR GRADES 5-8 ESTIMATED TIME Minutes What affects blood pressure? BEST FOR GRADES 5-8 ESTIMATED TIME 25-60 Minutes You Will Need Sphygmomanometer (blood pressure cuff) Stethoscope Blood pressure cuffs and stethoscopes can be found online

More information

JAMES R. KNIGHT

JAMES R. KNIGHT JAMES R. KNIGHT knightjr@ah.org James is currently the supervisor of clinical engineering at Sonora Regional Medical Center. He is the chairman of the CMIA Training & Education Committee. He also develops

More information

Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis

Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis Journal of Physics: Conference Series PAPER OPEN ACCESS Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis To cite this article: Valerie Tan et al 2018 J. Phys.: Conf. Ser. 1049 012062

More information

Lab #3: Electrocardiogram (ECG / EKG)

Lab #3: Electrocardiogram (ECG / EKG) Lab #3: Electrocardiogram (ECG / EKG) An introduction to the recording and analysis of cardiac activity Introduction The beating of the heart is triggered by an electrical signal from the pacemaker. The

More information

LAB: Blood Pressure Measurable Indicator of the Health of the Circulatory System!

LAB: Blood Pressure Measurable Indicator of the Health of the Circulatory System! LAB: Blood Measurable Indicator of the Health of the Circulatory System! Lab Objectives. At the completion of the lab, you should be able to: measure pulse or heart rate (HR) and respiratory rate (RR);

More information

5.2. MEASUREMENT OF BLOOD PRESSURE

5.2. MEASUREMENT OF BLOOD PRESSURE 5.2. MEASUREMENT OF BLOOD PRESSURE Purpose of experiment To examine a person s blood pressure. Tasks of experiment To record the pressure changes in the vascular system: using the traditional method measure

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 57 Comparison of wrist-type and arm-type 24-h blood pressure monitoring devices for ambulatory use Takahiro Komori a, Kazuo Eguchi a, Satoshi Hoshide a, Bryan Williams b and Kazuomi

More information

Chapter 15: Measuring Height, Weight, and Vital Signs. Copyright 2012 Wolters Kluwer Health Lippincott Williams & Wilkins

Chapter 15: Measuring Height, Weight, and Vital Signs. Copyright 2012 Wolters Kluwer Health Lippincott Williams & Wilkins Chapter 15: Measuring Height, Weight, and Vital Signs Height and Weight Weight Baseline measurement at patient s first visit Measured in kg or lbs Common types of scales Balance beam Dial Digital Height

More information

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010 Short title: Validation

More information

37 1 The Circulatory System

37 1 The Circulatory System H T H E E A R T 37 1 The Circulatory System The circulatory system and respiratory system work together to supply cells with the nutrients and oxygen they need to stay alive. a) The respiratory system:

More information

Chapter 29 1/8/2018. Vital Signs. Measuring and Reporting Vital Signs. Key Terms

Chapter 29 1/8/2018. Vital Signs. Measuring and Reporting Vital Signs. Key Terms Chapter 29 Vital Signs Key Terms Blood pressure Body temperature Diastolic pressure Hypertension Hypotension Pulse Pulse rate Respiration Sphygmomanometer Stethoscope Systolic pressure Vital signs Measuring

More information

Blood Pressure Competency Training

Blood Pressure Competency Training Blood Pressure Competency Training WORKBOOK Table of Contents 3..... Introduction 4..... Background 6..... Equipment Accurate blood pressure What impacts a patient s blood pressure Strategies for proper

More information

Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol

Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol Abstract Objectives Existing concerns over the accuracy of automated blood

More information

Lecture #3 - Blood Pressure Recording Procedure. Equipment

Lecture #3 - Blood Pressure Recording Procedure. Equipment Lecture #3 - Blood Pressure Recording Procedure CKiD Blood Pressure Training Equipment Mabis Medic-Kit aneroid sphygmomanometer Mabis Medic-Kit assortment of cuff sizes Stethoscope with bell (pediatric

More information

Vital Signs. Vital Signs. Vital Signs

Vital Signs. Vital Signs. Vital Signs Vital Signs Vital Signs Why do vital signs? Determine relative status of vital organs Establish baseline Monitor response to Rx, meds Observe trends Determine need for further evaluation, Rx, intervention

More information

A Simulation for Estimation of the Blood Pressure using Arterial Pressure-volume Model

A Simulation for Estimation of the Blood Pressure using Arterial Pressure-volume Model A Simulation for Estimation of the Blood Pressure using Arterial Pressure-volume Model Gye-rok Jeon, Jae-hee Jung, In-cheol Kim, Ah-young Jeon, Sang-hwa Yoon, Jung-man Son, Jae-hyung Kim, Soo-young Ye,

More information

Cardiovascular Physiology

Cardiovascular Physiology Cardiovascular Physiology The mammalian heart is a pump that pushes blood around the body and is made of four chambers: right and left atria and right and left ventricles. The two atria act as collecting

More information

IB TOPIC 6.2 THE BLOOD SYSTEM

IB TOPIC 6.2 THE BLOOD SYSTEM IB TOPIC 6.2 THE BLOOD SYSTEM TERMS TO KNOW circulation ventricle artery vein THE BLOOD SYSTEM 6.2.U1 - Arteries convey blood at high pressure from the ventricles to the tissues of the body Circulation

More information

CARDIOVASCULAR MONITORING. Prof. Yasser Mostafa Kadah

CARDIOVASCULAR MONITORING. Prof. Yasser Mostafa Kadah CARDIOVASCULAR MONITORING Prof. Yasser Mostafa Kadah Introduction Cardiovascular monitoring covers monitoring of heart and circulatory functions It makes it possible to commence interventions quickly in

More information

MANUAL BLOOD PRESSURE MONITOR BPM 168B Contents Blood Pressure Monitor Intended Use What is blood pressure?

MANUAL BLOOD PRESSURE MONITOR BPM 168B Contents Blood Pressure Monitor Intended Use What is blood pressure? MANUAL BLOOD PRESSURE MONITOR BPM 168B Contents 01... Blood Pressure Monitor Intended Use 02... 1. What is blood pressure? 02... 2. Why is it useful measure blood pressure at home?... A. WHO blood pressure

More information

Home blood pressure measurement with oscillometric upper-arm devices

Home blood pressure measurement with oscillometric upper-arm devices REVIEW Home blood pressure measurement with oscillometric upper-arm devices R.L. Braam, Th. Thien *# Department of Internal Medicine, University Medical Centre St Radboud, PO Box 9101, 6500 HB Nijmegen,

More information

The measurement of blood pressure and hypertension. Handout Fenyvesi Tamás III.Department of Medicine

The measurement of blood pressure and hypertension. Handout Fenyvesi Tamás III.Department of Medicine The measurement of blood pressure and hypertension Handout Fenyvesi Tamás III.Department of Medicine 1 2 History of the blood pressure measurement direct Stephen HALES 1726 horse carotid indirect Riva-Rocci

More information

Section 03: Pre Exercise Evaluations and Risk Factor Assessment

Section 03: Pre Exercise Evaluations and Risk Factor Assessment Section 03: Pre Exercise Evaluations and Risk Factor Assessment ACSM Guidelines: Chapter 3 Pre Exercise Evaluations ACSM Manual: Chapter 3 Risk Factor Assessments HPHE 4450 Dr. Cheatham Purpose The extent

More information

Characterization of the Oscillometric Method for Measuring Indirect Blood Pressure

Characterization of the Oscillometric Method for Measuring Indirect Blood Pressure Purdue University Purdue e-pubs Weldon School of Biomedical Engineering Faculty Publications Weldon School of Biomedical Engineering 1982 Characterization of the Oscillometric Method for Measuring Indirect

More information

Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard

Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard EOIN O'BRIENt, NEIL ATKINS, JAMES SHERIDAN, COLM KILLEEN, RICHARD COSTELLOE, SHANE LYDON and

More information

STUDIES INVOLVING THE NON-INVASIVE MEASUREMENT OF BLOOD PRESSURE IN THE ARM

STUDIES INVOLVING THE NON-INVASIVE MEASUREMENT OF BLOOD PRESSURE IN THE ARM STUDIES INVOLVING THE NON-INVASIVE MEASUREMENT OF BLOOD PRESSURE IN THE ARM 1. SCOPE Blood pressure is a vital sign that is used in many clinical situations, and assesses the pressure of blood as it flows

More information

LEARNING OUTCOME The students will be able to elicit vital signs correctly on human volunteers/patients

LEARNING OUTCOME The students will be able to elicit vital signs correctly on human volunteers/patients Vital signs (pulse, blood pressure, temperature, respiratory rate, pain) are physiological parameters that a healthcare professional requires when dealing with patients. Accurate measurement of vital signs

More information

1. Department of Gynecology and Obstetrics, St. Joseph's Hospital Berlin Tempelhof, Germany

1. Department of Gynecology and Obstetrics, St. Joseph's Hospital Berlin Tempelhof, Germany Page 1 of 9 Validation of the TONOPORT VI ambulatory blood pressure monitor, according to the European Society of Hypertension International Protocol revision 2010 Michael Abou-Dakn 1, Cornelius Döhmen

More information

SENSATION and CS300 Smaller meets faster. IABP System

SENSATION and CS300 Smaller meets faster. IABP System SENSATION and CS300 Smaller meets faster IABP System Cardiac Assist SENSATION and CS300 3 Cardiac Assist SENSATION and CS300 5 Smaller The SENSATION 7Fr. Fiber-Optic IAB Catheter Faster The CS300 with

More information

Heart Rate and Blood Pressure as Vital Signs

Heart Rate and Blood Pressure as Vital Signs Heart Rate and Blood Pressure as Vital Signs Computer 10 Since the earliest days of medicine heart rate has been recognized as a vital sign an indicator of health, disease, excitement, and stress. Medical

More information

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE. MEASURING BLOOD PRESSURE - MANUAL (equ04)

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE. MEASURING BLOOD PRESSURE - MANUAL (equ04) SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE TITLE: ISSUED FOR: MEASURING BLOOD PRESSURE - MANUAL Nursing DATE: REVIEWED: PAGES: 2/80 7/17 1 of 5 RESPONSIBILITY: RN, LPN, Patient Care Technician Multi-skilled

More information

AN INTRODUCTION TO DOPPLER. Sarah Gardner, Clinical lead, Tissue viability service. Oxford Health NHS Foundation Trust.

AN INTRODUCTION TO DOPPLER. Sarah Gardner, Clinical lead, Tissue viability service. Oxford Health NHS Foundation Trust. AN INTRODUCTION TO DOPPLER Sarah Gardner, Clinical lead, Tissue viability service. Oxford Health NHS Foundation Trust. THE DOPPLER EFFECT The Doppler Principle was described by Physicist and mathematician

More information

Chapter 7. Blood Pressure and Sound

Chapter 7. Blood Pressure and Sound Chapter 7. Blood Pressure and Sound Dr. Kurtuluş Erinç Akdoğan kurtuluserinc@cankaya.edu.tr Blood-pressure values in the various chambers of the heart and in the peripheral vascular system help the physician

More information

EXPERIMENTS WITH BLOOD PRESSURE MONITORING USING ECG AND PPG

EXPERIMENTS WITH BLOOD PRESSURE MONITORING USING ECG AND PPG EXPERIMENTS WITH BLOOD PRESSURE MONITORING USING ECG AND D. Špulák 1, R. Čmejla 1, V. Fabián 2 Czech Technical University in Prague, Faculty of Electrical Engineering, 1 Department of Circuit Theory 2

More information

BLOOD PRESSURE ASSESSMENT

BLOOD PRESSURE ASSESSMENT BLOOD PRESSURE ASSESSMENT Course Principles of Health Science Unit X Vital Signs Course Health Science Unit VI CPR / AED & First Aid Essential Question How do health care skills help to promote health

More information

PART I: HEART ANATOMY

PART I: HEART ANATOMY Lab 7: Heart Sounds and Blood Pressure PART I: HEART ANATOMY a) You should be able to identify the following structures on an adult human heart diagram. the 4 chambers the bicuspid (mitral) and tricuspid

More information

Blood Pressure Determination Using Analysis of Biosignals

Blood Pressure Determination Using Analysis of Biosignals Blood Pressure Determination Using Analysis of Biosignals RADIM ČÍŽ 1, MILAN CHMELAŘ 2 1 Department of Telecommunications Brno University of Technology Purkyňova 118, 612 00 Brno CZECH REPUBLIC cizr@feec.vutbr.cz,

More information

Physiological Measurements Training for Care/Nursing & LD Small Group Home Staff

Physiological Measurements Training for Care/Nursing & LD Small Group Home Staff Physiological Measurements Training for Care/Nursing & LD Small Group Home Staff CONTENTS 1. The National Early Warning Score 2. Respiratory Rate 3. Oxygen Saturations 4. Temperature 5. Blood Pressure

More information

a Centre de Médecine Cardiovasculaire, Paris, France, b Lebanese Hospital and Received 7 June 2009 Revised 15 September 2009 Accepted 9 October 2009

a Centre de Médecine Cardiovasculaire, Paris, France, b Lebanese Hospital and Received 7 June 2009 Revised 15 September 2009 Accepted 9 October 2009 42 Original article Validation of three professional devices measuring office blood pressure according to three different methods: the Omron BP, the Omron HBP T and the Pic Indolor Professional Roland

More information

Managing Blood Pressure: It Takes a Team

Managing Blood Pressure: It Takes a Team Managing Blood Pressure: It Takes a Team Module Two How to Take Blood Pressure October 2011 MODULE TWO: How to Take Blood Pressure Module Two: HOW TO TAKE BLOOD PRESSURE Objectives Demonstrate proper technique

More information

Measuring body temperature, blood pressure, pulse, respiratory rate and oxygen saturation

Measuring body temperature, blood pressure, pulse, respiratory rate and oxygen saturation Vital Signs Measuring body temperature, blood pressure, pulse, respiratory rate and oxygen saturation Aims To ensure that students are able to demonstrate the safe and correct technique for setting up

More information

AP Biology Lab 10 PHYSIOLOGY OF THE CIRCULATORY SYSTEM

AP Biology Lab 10 PHYSIOLOGY OF THE CIRCULATORY SYSTEM AP Biology Laboratory Date: Name and Period: AP Biology Lab 10 PHYSIOLOGY OF THE CIRCULATORY SYSTEM OVERVIEW In this lab you will: 1. in Exercise 10A you will learn how to measure blood pressure. 2. in

More information

DEVICES AND TECHNOLOGY

DEVICES AND TECHNOLOGY DEVICES AND TECHNOLOGY Validation of a new algorithm for the BPM-100 electronic oscillometric office blood pressure monitor James M. Wright ab, Gurdial S. Mattu a, Thomas L. Perry Jr ab, Mark E. Gelfer

More information

Physiology of the Circulatory System modified from

Physiology of the Circulatory System modified from Physiology of the Circulatory System modified from http://www.ekcsk12.org/science/aplabreview/aplab10.htm Introduction The circulatory system functions to deliver oxygen and nutrients to tissues for growth

More information

Hands on Sports Therapy KNOWLEDGE REVIEW QUESTIONS 2004 Thomson Learning It can help to shape a basic fitness training programme

Hands on Sports Therapy KNOWLEDGE REVIEW QUESTIONS 2004 Thomson Learning It can help to shape a basic fitness training programme Hands on Sports Therapy KNOWLEDGE REVIEW QUESTIONS 2004 Thomson Learning 1 CHAPTER 13 Knowledge Review Q1: Why is fitness testing useful? A1: Fitness testing is useful for various reasons: 1. It can help

More information

Chapter 24 Vital Signs. Copyright 2011 Wolters Kluwer Health Lippincott Williams & Wilkins

Chapter 24 Vital Signs. Copyright 2011 Wolters Kluwer Health Lippincott Williams & Wilkins Chapter 24 Vital Signs Vital Signs Temperature Pulse Respiration Blood pressure When to Assess Vital Signs Upon admission to any healthcare agency Based on agency institutional policy and procedures Anytime

More information

GETTING THE NUMBERS RIGHT: ACCURATE MEASUREMENT OF BLOOD PRESSURE

GETTING THE NUMBERS RIGHT: ACCURATE MEASUREMENT OF BLOOD PRESSURE GETTING THE NUMBERS RIGHT: ACCURATE MEASUREMENT OF BLOOD PRESSURE Pamela B. Morris, MD, FACC, FAHA, FASPC, FNLA Professor Medicine and Cardiovascular Diseases Director, Seinsheimer Cardiovascular Health

More information

CATCH A WAVE.. INTRODUCTION NONINVASIVE HEMODYNAMIC MONITORING 4/12/2018

CATCH A WAVE.. INTRODUCTION NONINVASIVE HEMODYNAMIC MONITORING 4/12/2018 WAVES CATCH A WAVE.. W I S C O N S I N P A R A M E D I C S E M I N A R A P R I L 2 0 1 8 K E R I W Y D N E R K R A U S E R N, C C R N, E M T - P Have you considered that if you don't make waves, nobody

More information

Human Cardiovascular Physiology: Blood Pressure and Pulse Determinations

Human Cardiovascular Physiology: Blood Pressure and Pulse Determinations ighapmlre33apg269_274 5/12/04 3:10 PM Page 269 impos03 302:bjighapmL:ighapmLrevshts:layouts: NAME Human Cardiovascular Physiology: Blood Pressure and Pulse Determinations LAB TIME/DATE REVIEW SHEET exercise

More information

The Circulatory System. Lesson Overview. Lesson Overview The Circulatory System

The Circulatory System. Lesson Overview. Lesson Overview The Circulatory System 33.1 THINK ABOUT IT More than one-third of the 1.2 million Americans who suffer a heart attack each year die. This grim evidence shows that the heart and the circulatory system it powers are vital to life.

More information

LAB 9: Metabolic Rates

LAB 9: Metabolic Rates LAB 9: Metabolic Rates Introduction: The cardiovascular (circulatory) system functions to deliver oxygen and nutrients to tissues for growth and metabolism, and to remove metabolic wastes. The heart pumps

More information

AMERICAN ASSOCIATION OF SLEEP TECHNOLOGISTS TECHNICAL GUIDELINE FOR PATIENT ASSESSMENT AND VITAL SIGNS MEASUREMENT AND DOCUMENTATION

AMERICAN ASSOCIATION OF SLEEP TECHNOLOGISTS TECHNICAL GUIDELINE FOR PATIENT ASSESSMENT AND VITAL SIGNS MEASUREMENT AND DOCUMENTATION AMERICAN ASSOCIATION OF SLEEP TECHNOLOGISTS TECHNICAL GUIDELINE FOR PATIENT ASSESSMENT AND VITAL SIGNS MEASUREMENT AND DOCUMENTATION SUMMARY: Sleep technologists observe and monitor physical signs and

More information

Literature Review Aarthi Sivasankar

Literature Review Aarthi Sivasankar Literature Review Aarthi Sivasankar Table of Contents Importance of Blood Pressure How Blood Pressure Monitoring Devices Work Measurements Involved in Taking Blood Pressure Materials Needed for Taking

More information

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists 3,900 116,000 120M Open access books available International authors and editors Downloads Our

More information

Welch Allyn ABPM 6100S. Comprehensive blood pressure monitoring that brings patient comfort home.

Welch Allyn ABPM 6100S. Comprehensive blood pressure monitoring that brings patient comfort home. Welch Allyn ABPM 6100S Comprehensive blood pressure monitoring that brings patient comfort home. Features and Benefits of the Welch Allyn ABPM 6100S Pictorial application instructions, artery markers and

More information

Oscillometric pressure pulse waveforms: their current and prospective applications in biomedical instrumentation

Oscillometric pressure pulse waveforms: their current and prospective applications in biomedical instrumentation Oscillometric pressure pulse waveforms: their current and prospective applications in biomedical instrumentation J. JILEK*, M. STORK** *Carditech, Culver City, California, USA **Department of Applied Electronics

More information

IB TOPIC 6.2 THE BLOOD SYSTEM

IB TOPIC 6.2 THE BLOOD SYSTEM IB TOPIC 6.2 THE BLOOD SYSTEM THE BLOOD SYSTEM TERMS TO KNOW circulation ventricle artery vein 6.2.U1 - Arteries convey blood at high pressure from the ventricles to the tissues of the body Circulation

More information

!"#$ Blood Pressure Monitor. Instruction manual page 1 Gebrauchsanweisung Seite 17 Gebruiksaanwijzing pagina 33

!#$ Blood Pressure Monitor. Instruction manual page 1 Gebrauchsanweisung Seite 17 Gebruiksaanwijzing pagina 33 !"#$ Blood Pressure Monitor Instruction manual page 1 Gebrauchsanweisung Seite 17 Gebruiksaanwijzing pagina 33 Mode d emploi page 49 Manuale de instructone pagina 65 Manual de instrucciones pagina 81 Introduction

More information

The Heart 1 of 34 Boardworks Ltd 2012

The Heart 1 of 34 Boardworks Ltd 2012 The Heart 1 of 34 Boardworks Ltd 2012 2 of 34 Boardworks Ltd 2012 What does the heart do? 3 of 34 Boardworks Ltd 2012 The heart is a muscular organ located in the thorax. It pumps blood continuously around

More information

Objectives of the Heart

Objectives of the Heart Objectives of the Heart Electrical activity of the heart Action potential EKG Cardiac cycle Heart sounds Heart Rate The heart s beat separated into 2 phases Relaxed phase diastole (filling of the chambers)

More information

Chapter 12 - Vital_Signs_and_Monitoring_Devices

Chapter 12 - Vital_Signs_and_Monitoring_Devices Introduction to Emergency Medical Care 1 OBJECTIVES 12.1 Define key terms introduced in this chapter. Slides 13 15, 17, 21 22, 26, 28, 30, 32 33, 35, 44, 47 48, 50, 55, 60 12.2 Identify the vital signs

More information

BP-600 Noninvasive Blood Pressure Sensor

BP-600 Noninvasive Blood Pressure Sensor Technical Note BP-600 Overview A person's cardiac output, peripheral vascular resistance, blood pressure, and other cardiovascular parameters change in response to the activities and events taking place

More information

Assessment of Reliability of Hamilton-Tompkins Algorithm to ECG Parameter Detection

Assessment of Reliability of Hamilton-Tompkins Algorithm to ECG Parameter Detection Proceedings of the 2012 International Conference on Industrial Engineering and Operations Management Istanbul, Turkey, July 3 6, 2012 Assessment of Reliability of Hamilton-Tompkins Algorithm to ECG Parameter

More information

Techniques of Vital Signs. John Gazewood, MD, MSPH Department of Family Medicine

Techniques of Vital Signs. John Gazewood, MD, MSPH Department of Family Medicine Techniques of Vital Signs John Gazewood, MD, MSPH Department of Family Medicine Objectives Define normal in several ways. Describe correct technique of measuring vital signs. Describe correct technique

More information

The Cardiac Cycle Clive M. Baumgarten, Ph.D.

The Cardiac Cycle Clive M. Baumgarten, Ph.D. The Cardiac Cycle Clive M. Baumgarten, Ph.D. OBJECTIVES: 1. Describe periods comprising cardiac cycle and events within each period 2. Describe the temporal relationships between pressure, blood flow,

More information