Arm position and blood pressure: a risk factor for hypertension?

Size: px
Start display at page:

Download "Arm position and blood pressure: a risk factor for hypertension?"

Transcription

1 (3) 17, & 3 Nature Publishing Group All rights reserved /03 $ ORIGINAL ARTICLE Arm position and blood pressure: a risk factor for hypertension? A Mourad, S Carney, A Gillies, B Jones, R Nanra and P Trevillian Department of Nephrology, John Hunter Hospital, Hunter Region Mail Centre, NSW, Australia The objective of this study was to re-evaluate the effect of arm position on blood pressure (BP) measurement with auscultatory and oscillometric methods including ambulatory blood pressure monitoring (ABPM). The setting was the hospital outpatient department and the subjects chosen were normotensive and hypertensive. The effect of lowering the arm from heart level on indirect systolic BP (SBP) and diastolic BP (DBP) measurement as well as the importance of supporting the horizontal arm were measured. In the sitting position, lowering the supported horizontal arm to the dependent position increased BP measured by a mercury device from / to / mmhg in normotensive subjects, a mean increase of 8/7 mmhg (Po0.01). In hypertensive subjects, a similar manoeuvre increased BP from / to / mmhg, an increase of 23/10 mmhg (Po0.01). Combined results from normotensive and hypertensive subjects demonstrate a direct and proportional association between BP (SBP and DBP) and the increase produced by arm dependency. Similar changes and associations were noted with oscillometric devices in the clinic situation. However, supporting the horizontal arm did not alter BP. Of particular interest, analysis of 13 hypertensive subjects who underwent ABPM on two occasions, once with the arm in the usual position and once with the arm held horizontally for BP measurement during waking hours, demonstrated changes comparable to the other devices. The mean 12-hour BP was / mmhg during the former period and significantly decreased to / mmhg during the latter period (Po0.01). Regression analysis of the change in SBP and DBP with arm position change again demonstrated a close correlation (r 2 ¼ and ; Po0.001) with the artefact being larger with higher systolic and diastolic pressures. In conclusion, arm movements lead to significant artefacts in BP measurement, which are greater, the higher the systolic or diastolic pressure. These systematic errors occur when using both auscultatory and oscillometric (clinic and ABPM) devices and might lead to an erroneous diagnosis of hypertension and unnecessary medication, particularly in individuals with high normal BP levels. Since clinical interpretations of heart level vary, the horizontal arm position should be the unambiguous standard for all sitting and standing BP auscultatory and oscillometric measurements. (3) 17, doi: /sj.jhh.1563 Keywords: blood pressure; indirect measurement; mercury and oscillometric device; ABPM Introduction Blood pressure (BP) measurement is the most common clinical measurement in hospitals, consulting rooms, and more recently home and the workplace. Most clinical practice audits reinforce the importance of the cuffed upper arm being at heart level; 1,2 yet published definitions and clinical interpretations of heart level vary, possibly owing to the heart extending from the sternal angle to the fifth intercostal space. 3 WHO guidelines recommend that the cubital fossa be placed at the fourth intercostal space. The American Heart Association defines Correspondence: Dr SL Carney, Department of Nephrology, John Hunter Hospital, Locked Bag 1, Hunter Region Mail Centre, NSW 2310, Australia. scarney@mail.newcastle.edu.au Received 22 July 2; revised 17 January 3; accepted 14 February 3 heart level as arbitrarily taken to the junction of the fourth intercostal space and the lower left sternal border. As a subsequent practical clinical guide, these guidelines state: When the patient is seated, placing the arm on a nearby tabletop a little above waist level will result in a satisfactory position. 5 The British Hypertension Society recommends that the arm be horizontal at the mid-sternal level. 6 This definition best approximates the right atrium, which is the reference level. 7 While the influence of arm position on BP measurement was first reported almost years ago, 8 Webster et al 9 more recently re-evaluated this issue and documented a mean increase of 11 (SBP)/ 12 (DBP) mmhg when the horizontal arm was made dependent in hypertensive subjects. Similar changes were noted in normotensives, and in addition up to 10/6 mmhg increases were measured if the horizontal arm was unsupported. These

2 390 important observations have been replicated using an oscillometric device 10 and confirmed by intraarterial BP measurements. 11 Despite studies confirming the importance of arm position on BP measurement, it is likely that arm position varies markedly in clinical practice and also in published studies. Resting an arm on a desk or arm of a chair as often recommended could vary the BP significantly, depending not just on the dimensions of the furniture but also on the height of the individual and also their posture. It was the intention of this study to repeat part of the Webster et al 9 protocol to decide if arm position errors are more or less apparent in hypertensive subjects when compared to normotensives, an issue that has not been evaluated. In contrast to that study, the Hawksley random zero sphygmomanometer was not used owing to concerns about its accuracy. 12,13 However, to minimise potential observer bias, an automatic oscillometric device was also used. Confirmation of the effect of arm position on BP with oscillometric devices is important, given their increasing use particularly in hospitals and the home. Evaluating the effect of arm position on ambulatory blood pressure monitoring (ABPM) was an additional study objective, given the increasingly important role of ABPM in hypertension management. It was postulated that the usually dependent arm during daytime ABPM significantly overestimates BP and exaggerates nocturnal dipping. Subjects and methods Arm position and clinic BP A total of 51 subjects participated, 26 being hypertensive, two of whom were not on medication. BP was measured with a standard mercury sphygmomanometer and validated 14,15 oscillometric device (OMRON HEM 705CP, Japan) in random order. The cuff was on the nondominant arm. With both devices, the BP was measured with the patient sitting and standing and with the arm dependent, horizontal unsupported and horizontal supported. These manoeuvres were performed in random order. The manufacturer s instructions for using OMRON oscillometric devices are comparable to the American Heart Association recommendations with the arm being flexed and resting on a desk or table. BP measurement was carried out in a comfortable environment by a trained observer (AM) after the subject had been sitting or standing for 5 min. A minute was allowed between each arm position change. Conversation was avoided during the procedure. Phase 1 Korotkoff sounds represented systolic BP (SBP) and Phase V the diastolic BP (DBP). BP technique followed the British Hypertension Society recommendations 6 including the horizontal position of the arm using oscillometric and mercury devices. Arm position and ABPM In total, 15 subjects participated: 10 were on antihypertensive medications and 5 were untreated while being evaluated following clinic diagnosis of hypertension. They were randomly allocated to usual or horizontal arm position during the first of two monitoring periods. Both monitoring periods were performed on days of similar activity. During the horizontal ABPM period, patients were asked to hold the arm in the horizontal position during each BP measurement when not in bed. They were allowed to rest the arm on nearby furniture or hold it without support. They were not advised on the possible effects of arm position on BP. Successful completion of greater than 90% of BP readings was required for each ABPM period using a spacelabs oscillometric device (Space Labs Inc., Redmond, WA, USA), with the cuff on the nondominant arm. Volunteers in both studies were in sinus rhythm, did not have cardiac failure, autonomic neuropathy or a recent history of postural hypotension, and were not taking tricyclic antidepressants or a-adrenoceptor blockers. Statistical analysis Student s paired t-test was conducted to compare systolic and diastolic pressures. Analyses compared dependent position with horizontal supported, horizontal supported vs horizontal unsupported while sitting and standing with both devices. Simple linear regression was used to compare the relations between horizontal and dependent arm positions and the level of SBP and DBP. Differences between usual and horizontal ABPM periods were examined by analysis of variance for repeated measurements. Differences were tested using Students paired t-test. Linear relations between patient BP were estimated using the least-squares method. Results are expressed as mean 7 s.d. Results Arm position and clinic BP Hypertensive subjects were older than the normotensive group, although mean body weight and arm circumference were similar between the groups (Table 1a). BP measured with a mercury sphygmomanometer was comparable in the sitting and standing positions in both normotensive and hypertensive groups (Table 2a). However, lowering the horizontal supported arm to the dependent position increased mean BP by 8/7 mmhg in sitting normotensive subjects and by 7/5 mmhg when standing (Po0.01 for both SBP and DBP). In hypertensive subjects, the same arm manoeuvre increased mean BP by 23/10 mmhg sitting and 21/10 mmhg standing (Po0.01 for both SBP and DBP). When comparing the supported arm with the unsupported arm,

3 Table 1 Patient characteristics 391 (a) Arm position and clinic BP Normotensive (n=25) Hypertensive (n=26) Mean 7 s.d. (range) Mean 7 s.d. (range) Sex (M/F) 10/15 10/16 Age (years) (17 58) * (18 81) Body weight (kg) ( 131) (38 140) Body height (cm) ( 183) ( ) Arm circumference (cm) (20 40) (20 46) (b) Arm position and ABPM Mean 7 s.d. (range) Sex (M/F) 9/4 (18 80) Age (years) Arm horizontal day 1 7 Antihypertensive treatment 9 *Po0.05. Table 2 Clinic and ABPM results (a) Auscultatory and oscillometric blood pressure (mmhg) measurements in normotensive and hypertensive subjects Posture and arm position Normotensive Hypertensive Systolic Diastolic Systolic Diastolic Auscultatory Sitting Horizontal supported 103 (10) 60 (7) 143 (21) 78 (17) Horizontal unsupported 103 (10) 60 (7) 142 (21) 77 (18) Dependent 111 (14)** 67 (10)** 166 (29)** 88 (20)** Standing Horizontal supported 103 (11) 63 (9) 142 (19) 77 (15) Horizontal unsupported 102 (12) 62 (10) 143 (20) 77 (15) Dependent 110 (13)** 68 (10)** 163 (25)** 87 (17)** Oscillometric Sitting Horizontal supported 103 (11) 62 (7) 145 (24) 79 (18) Horizontal unsupported 102 (11) 63 (8) 144 (25) 81 (18) Dependent 113 (13)** 72 (8)** 163 (29)** 88 (19)** Standing Horizontal supported 102 (12) 63 (9) 139 (24) 77 (13) Horizontal unsupported 101 (10) 63 (7) 139 (25) 78 (13) Dependent 114 (14)** 73 (10)** 157 (29)** 88 (16)** (b) ABPM daytime results Mean 12-h SBP Mean 12-h DBP Arm horizontal Arm usual position ** ** **Po0.01. mean BP was not significantly different. A regression analysis of the differences between sitting and standing BP in normotensive and hypertensive subjects using both mercury and oscillometric devices was performed. Figures 1 and 2 represent the sitting auscultatory SBP (y ¼ x ) and DBP (y ¼ x ) results, and sitting oscillometric and standing auscultatory and oscillometric results were similar as were the regression equations for each analysis. In particular the difference between the supported horizontal and dependent arm increased with the increase in SBP and DBP. For example, an SBP of 155 mmhg in the horizontal arm would become 180 mmhg in the dependent position, an increase of 25 mmhg. A DBP of 85 mmhg could be incorrectly measured as

4 r 2 = P <0.001 dependent (mmhg) unsupported (mmhg) r 2 = P< horizontal (mmhg) Figure 1 Sitting auscultatory SBP: supported horizontal compared to dependent arm position supported (mmhg) Figure 3 Sitting auscultatory SBP: supported compared to unsupported horizontal arm position. 2 r 2 = P< dependent (mmhg) mmhg BPM horizontal (mmhg) Figure 2 Sitting auscultatory DBP: supported horizontal compared to dependent arm position. 96 mmhg, an increase of 11 mmhg following arm dependence. Evaluation of the regression analyses for systolic pressure (horizontal supported compared to dependent) demonstrated a slope that was significantly different from the diastolic pressure relation, regardless of whether sitting or standing. As demonstrated in Figure 3, supporting the arm did not significantly alter sitting SBP or DBP when compared to horizontal unsupported auscultatory readings (y ¼ x ). This observation persisted in standing subjects for SBP and DBP and was also noted with both oscillometric measurements. Arm position and ABPM Adequate measurements were achieved in 13 of 15 subjects, nine of whom were male and their age ranged from 18 to 80 years (Table 1b). Nine patients were taking antihypertensive medications and the period of assessment varied between 2 and 4 days TIME Figure 4 Analysis of the mean hourly SBP, DBP and pulse rate, during the 12-h awake period. Values are expressed as mean 7 s.d: + Po0.05, ++ Po0.01, and +++ Po0.001; (K) horizontal arm; (&) usual arm position. Results were analysed over a 12-h period between and h when all subjects were awake and out of bed. During this period, the mean SBP/DBP was / mmhg with the arm horizontal and / mmhg with the arm in the usual position for ABPM. Both systolic and diastolic pressures were significantly different (Po0.01) when the two 12-hourly evaluation periods were compared (Table 2b). An analysis of the mean hourly SBP (Figure 4) also demonstrated a persistently lower SBP in patients when their arm was horizontal during the 12-h awake period, and these differences were also statistically significant at 09:00 h (Po0.01), h (Po0.01), h (Po0.05), h (Po0.05) and h (Po0.001). While each mean hourly DBP was also consistently lower over this 12-h period, these differences did not reach statistical significance. The pulse rate was

5 a Arm Position (mmhg) b Horizontal (mmhg) r 2 = P< r 2 = P < comparable during each period (Figure 4). A comparison of mean horizontal vs usual arm position over the 12-h awake period in each of the 13 subjects revealed a significant correlation for both SBP (r 2 ¼ ) and DBP (r 2 ¼ ) (Figure 5), the subsequent regression lines being significantly different from zero and unity (Po0.001). Thus, the higher the SBP or DBP, the greater the artefact because of arm position. There was no difference in night-time sleeping SBP or DBP between the two periods. Discussion usual arm position (mmhg) Figure 5 SBP (a) and DBP (b) during 12-h awake period: horizontal compared to usual arm postion. While this study confirms the significant increase in SBP and DBP by making the arm dependent noted in earlier studies, including recent ones using a Hawksley random-zero sphygmomanometer, 9,10 it also demonstrates that oscillometric devices including those used for ABPM are subject to similar artefacts. The absolute mean changes found during arm movement with a mercury device were similar to those found by Webster et al, 4 and Netea et al. 10 In sitting hypertensives, Netea et al 10 noted a mean increase of 10/11 mmhg in subjects from a mean SBP/DBP of 133/78 mmhg. While the mean SBP/DBP of our hypertensive subjects was higher (143/ 78 mmhg), using our calculations we would have predicted an increase of 18/10 mmhg in their patients. In the earlier data from Webster et al, 9 their sitting hypertensive group mean SBP/DBP was 140/90 mmhg, which increased by 18/14 mmhg with dependency. Our prediction of 20/12 mmhg is comparable. Netea et al 10 also measured the effect of arm position using an oscillometric device and demonstrated a similar but slightly lower effect of arm movement. We observed similar effects of arm position with mercury and both oscillometric devices. Despite the widespread belief that the arm should be supported and confirmatory results from Webster et al, 9 we were unable to demonstrate any significant differences in sitting or standing subjects. We suspect that the additional cardiac work required to hold the arm up in healthy people is relatively trivial; however, on balance it is reasonable to maintain this standard during BP measurement. The 13 mmhg SBP and 8 mmhg DBP mean difference between the two ABPM periods is also consistent with our clinic study with mercury and oscillometric devices. Again the higher the SBP or DBP, the greater the effect of arm position, potentially a significant problem in patient management. During this study, we could not ensure that all pressure measurements during the arm horizontal day were in fact horizontal, and while complete compliance was reported, BP differences may have been reduced by impaired compliance. While the two situations are not strictly comparable, it is of interest to compare the ABPM and clinic oscillometric device results. Applying the clinic regression analysis of dependent and horizontal pressures to the 13 ABPM subjects, a BP of 141/74 mmhg with the arm horizontal (mean 12-h BP) should increase to 161/84 mmhg with the arm dependent. This prediction is close to the measured mean of 154/ 82 mmhg. The regression analyses of SBP and DBP in horizontal vs usual arm position in the 13 ABPM subjects are also similar to the clinic OMRON findings. Minor differences between clinic and ABPM studies are probably because of a variation in the arm position during the usual ABPM monitoring period. Although the role of ABPM in the diagnosis and treatment of hypertension now appears well defined, individual reproducibility remains an important issue. ABPM is often repeated as part of the diagnostic or treatment process, and while ABPM reproducibility appears to be superior to clinic BP measurement, particularly in large groups, 19 a significant variation has been detected in some individuals. 20 For example, some studies have demonstrated that BP recorded in hypertensives and normotensives during a work day can be higher than during a nonworking day, 21,22 although a large individual variation was observed. Seasonal variations can also alter BP in normotensive and hypertensive individuals. 23 The ability of arm 393

6 394 position to alter ABPM significantly is another confounding factor in the interpretation of ABPM in individuals and this probably contributes to reported reproducibility problems, particularly when comparing work to nonwork days rather than the implied effects of work exertion and stress. The recent report by van der Steen et al 24 that body and arm positions can significantly influence sleeping ABPM supports this hypothesis. Guidelines for ABPM use recommend only that patients hold their arm still during pressure recordings, 25 and unless the subject is seated or lying, the arm will probably be dependent. The correlation between the level of BP and arm position has not been evaluated before and is an important clinical observation. While changes in arm position in subjects with low normal BP, particularly a normal SBP, would have few clinical implications, in subjects with borderline high pressures lowering the arm even if not to the dependent position could clearly produce a spurious diagnosis of hypertension. Some publications recommend correction factors for arm movement, with the American Heart Association suggesting a 0.8 mmhg correction for every 1 cm displacement of the sensing site with respect to heart level. 5 However, in a borderline or hypertensive population such corrections underestimate the potential error. It is believed that the difference in BP values produced by arm movement is largely accounted for by hydrostatic pressure 5 and this is supported by Netea et al, 11 although the differences demonstrated in regression analysis between SBP and DBP in our study cast doubt on such a simplistic view. While the British Hypertension Society recommends the method of differences as its preferred technique when comparing methods for measuring BP, 26 least-products regression analysis may be superior. 27 Nevertheless, ordinary least-squares regression analysis was used in this study despite its limitations because it is readily comprehensible to clinicians and, in particular, it clearly defines the magnification of the error produced by arm dependency as both SBP and DBP increases. Identifying fixed and proportional bias does not illuminate the important clinical observations of this study. This study confirms the significant effect of arm position on auscultatory BP. Such an effect occurs in the sitting and standing positions as well as with an oscillometric device. In addition, it is now clear that the higher the BP, the greater the error produced, particularly in the measurement of SBP. Given the current focus on the diagnosis and treatment of systolic hypertension, the potential for spurious readings should cause concern. The indirect measurement of BP is prone to numerous errors because of poor technique and even the presence of back support when sitting may influence BP. 28 Nevertheless, the marked effect of arm position on BP has been relatively ignored probably because the accepted heart level position is unclear and open to misinterpretation. Even a relatively small downward arm movement in a patient with borderline or high BP could significantly influence diagnosis and treatment. To avoid significant and probable widespread errors in current BP measurement, we commend all national and international organisations to clearly and unambiguously state that the arm must be supported in the horizontal position when BP is being measured, whether sitting or standing and whether auscultatory or oscillometric devices are being used. Furthermore, clinical research manuscripts should state arm position to allow comparison, although in the recently published comparison between BP and mortality from the seven countries study research group, 29 BP was measured in the supine position, which possibly serendipitously minimised arm position artefacts between several centres. The application of such an unambiguous standard using ABPM is, however, less clear. Many community studies have facilitated the development of useful clinical criteria for hypertension diagnosis and assessment and to discount this information would be difficult, particularly since it is probably convenient for individuals to leave their arm at their side during cuff inflation. However, daytime ABPM is usually much lower than clinic pressures even if it is still artificially high owing to the dependent arm. On the other hand, night-time dipping is in part due to a daytime dependent arm. References 1 Feher M, Harris-StJohn K, Lant A. Blood pressure measurement by junior doctorsfa gap in medical education? Health Trends 1992; 24: Carney S et al. Hospital blood pressure measurement: staff and device assessment. J Qual Clin Pract 1999; 19: Moore KL. Clinically Oriented Anatomy, 3rd edn. Williams & Wilkins: Baltimore, MD, WHO. Guidelines for the management of mild hypertension. Memorandum from a WHO/ISH meeting. Hypertension 1993; 22: American Heart Association. Recommendations for human blood pressure determination by sphygmomanometers: report of a special taskforce appointed by the Steering committee. Hypertension 1988; 11: 210A 222A. 6 Petrie JC, O Brien ET, Littler WA, de Swiet M. British Hypertension Society Recommendations on blood pressure measurement. BMJ 1986; 293: Guyton A. Textbook of Medical Physiology. WB Saunders Company: Philadelphia; 1986, p Hill L. Measurement of systolic blood pressure in men. Heart 1909; 1: Webster J, Newnham D, Petrie JC, Lovell HG. Influence of arm position on measurement of blood pressure. BMJ 1984; 288: Netea RT, Lenders JWM, Smits P, Thien T. Arm position is important for blood pressure measurements. J Hum Hypertens 1999; 13:

7 11 Netea RT et al. Influence of arm position on intraarterial blood pressure measurement. J Hum Hypertens 1999; 13: Parker D et al. A comparison of the random zero and standard mercury sphygmomanometer. Hypertension 1988; 11: O Brien E, Mee F, Atkins N, O Mally K. Inaccuracy of the Hawksley random zero sphygmomanometer. Lancet 1990; 336: Mann S. Inaccuracy of electronic sphygmomanometers. Clin Exp Pharmacol Physiol 1992; 19: O Brien E, Mee F, Atkins N, Thomas M. Evaluation of three devices for self-measurement of blood pressure according to the revised British Hypertension Society Protocol: the Omron HEM-705CP, Phillips HP5332, and Nissei DS-175. Blood Press Monitor 1996; Fotherby MC, Potter JF. Reproducibility of ambulatory and clinic blood pressure measurements in elderly hypertensive subjects. J Hypertens 1993; 11: Mansoor GA, McCabe EJ, White WB. Long-term reproducibility of ambulatory blood pressure. J Hypertens 1994; 12: Palatini P, Mormino P, Caneli C. Factors affecting ambulatory blood pressure reproducibility Results of the HARVEST trial. Hypertension 1994; 23: Engfeldt P et al. 24-Hour ambulatory blood pressure monitoring in elderly normotensive individuals and its reproducibility after one year. J Hum Hypertens 1994; 8: Prisant LM et al. Repeatability of automated ambulatory blood pressure measurements. J Fam Pract 1992; 11: Peiper C, Warren K, Pickering TG. A comparison of ambulatory blood pressure and heart rate at home and work on work day and non-work days. J Hypertens 1993; 11: Enstrom I, Pennert K. Does it matter whether ambulatory blood pressure is recorded duirng a work day or non-work day? J Hypertens 1996; 14: Sega R et al. Seasonal variations in home and ambulatory blood pressure in the PAMELA population. J Hypertens 1998; 16: Van der Steen MS, Pleijers AMLJ, Lenders JCUM, Thein T. Influence of different supine body positions on blood pressure: consequences for night blood pressure/dipper-status. J Hypertens 0; 18: Prasad N, Isles C. Ambulatory blood pressure monitoring: a guide for general practitioners. BMJ 1996; 313: O Brien E et al. The British Hypertension Society Protocol for the evaluation of automated and semiautomated blood pressure measuring devices with special reference to ambulatory systems. J Hypertens 1990; 8: Ludbrook J. Comparing methods of measurement. Clin Exp Pharmacol Physiol 1997; 24: Cushman WC, Cooper KM, Horne RA, Meydrech EF. Effect of back support and stethoscope on seated blood pressure determinations. Am J Hypertens 1990; 3: Van den Hoogen PCW et al. The relation between blood pressure and mortality due to coronary heart disease among men in different parts of the world. N Engl J Med 0; 342:

Does masked hypertension exist in healthy volunteers and apparently well-controlled hypertensive patients?

Does masked hypertension exist in healthy volunteers and apparently well-controlled hypertensive patients? O R I G I N A L A R T I C L E Does masked hypertension exist in healthy volunteers and apparently well-controlled hypertensive patients? I. Aksoy, J. Deinum, J.W.M. Lenders, Th. Thien *# Department of

More information

and bias, which are known to be present in self-home and in professional office BP measurements taken using the auscultatory technique [7].

and bias, which are known to be present in self-home and in professional office BP measurements taken using the auscultatory technique [7]. Devices and technology 37 Validation of the A&D UM-11 professional hybrid device for office blood pressure measurement according to the International Protocol George S. Stergiou, Periklis P. Giovas, Charilaos

More information

AFFORDABLE TECHNOLOGY

AFFORDABLE TECHNOLOGY World Health Organization AFFORDABLE TECHNOLOGY BLOOD PRESSURE MEASURING DEVICES FOR LOW RESOURCE SETTINGS CARDIOVASCULAR DISEASES Blood Pressure Measurement in Low Resource Settings Annex 1: Blood Pressure

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 261 Validation of the ROSSMAX blood pressure measuring monitor according to the European Society of Hypertension International Protocol for Validation of Blood Pressure Measuring

More information

Home blood pressure measurement with oscillometric upper-arm devices

Home blood pressure measurement with oscillometric upper-arm devices REVIEW Home blood pressure measurement with oscillometric upper-arm devices R.L. Braam, Th. Thien *# Department of Internal Medicine, University Medical Centre St Radboud, PO Box 9101, 6500 HB Nijmegen,

More information

ANTIHYPERTENSIVE DRUG THERAPY IN CONSIDERATION OF CIRCADIAN BLOOD PRESSURE VARIATION*

ANTIHYPERTENSIVE DRUG THERAPY IN CONSIDERATION OF CIRCADIAN BLOOD PRESSURE VARIATION* Progress in Clinical Medicine 1 ANTIHYPERTENSIVE DRUG THERAPY IN CONSIDERATION OF CIRCADIAN BLOOD PRESSURE VARIATION* Keishi ABE** Asian Med. J. 44(2): 83 90, 2001 Abstract: J-MUBA was a large-scale clinical

More information

Methods DEVICES AND TECHNOLOGY

Methods DEVICES AND TECHNOLOGY DEVICES AND TECHNOLOGY Validation of the Omron HEM-907 device for blood pressure measurement Mohamed A. El Assaad, Jirar A. Topouchian, Bernadette M. Darné and Roland G. Asmar Background The aim of this

More information

1 Department of Medical Physics, Royal Infirmary of Edinburgh, 2 Department of. Received 11 June 2004 Accepted 7 September 2004

1 Department of Medical Physics, Royal Infirmary of Edinburgh, 2 Department of. Received 11 June 2004 Accepted 7 September 2004 Devices and technology 1 Do SpaceLabs ambulatory non-invasive blood pressure recorders measure blood pressure consistently over several years use? John N. Amoore 1, David Dewar 2, Kathleen Gough 3 and

More information

Ambulatory blood pressure monitoring (ABPM) is. Accuracy of Ambulatory Blood Pressure Monitors in Routine Clinical Practice.

Ambulatory blood pressure monitoring (ABPM) is. Accuracy of Ambulatory Blood Pressure Monitors in Routine Clinical Practice. AJH 26; 19:81 89 BP Measurement Accuracy of Ambulatory Blood Pressure Monitors in Routine Clinical Practice Tony C. Y. Pang and Mark A. Brown Background: To determine the extent of discrepancies between

More information

Copyright: DOI link to article: Date deposited: This work is licensed under a Creative Commons Attribution 4.0 International License

Copyright: DOI link to article: Date deposited: This work is licensed under a Creative Commons Attribution 4.0 International License Pan F, Zheng DC, He PY, Murray A. Does the Position or Contact Pressure of the Stethoscope Make Any Difference to Clinical Blood Pressure Measurements: An Observational Study. Medicine 214, 93(29). Copyright:

More information

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010 Short title: Validation

More information

STATE OF THE ART BP ASSESSMENT

STATE OF THE ART BP ASSESSMENT STATE OF THE ART BP ASSESSMENT PROFESSOR MOLECULAR PHARMACOLOGY CONWAY INSTITUE UNIVERSITY COLLEGE DUBLIN CO-CHAIRMAN BLOOD PRESCSURE MANAGEMENT IN LOW RESOURCE SETTINGS CENTRE FOR INTERNATIONAL HUMANITARIAN

More information

HOW CONSISTENT ARE THE BLOOD PRESSURE AND PULSE RATE MEASUREMENTS OF THE ELECTRONIC BP APPARATUS AND THE MANUAL SPHYGMOMANOMETER

HOW CONSISTENT ARE THE BLOOD PRESSURE AND PULSE RATE MEASUREMENTS OF THE ELECTRONIC BP APPARATUS AND THE MANUAL SPHYGMOMANOMETER HOW CONSISTENT ARE THE BLOOD PRESSURE AND PULSE RATE MEASUREMENTS OF THE ELECTRONIC BP APPARATUS AND THE MANUAL SPHYGMOMANOMETER Naser KA. Teaching Hospital Peradeniya, Peradeniya, Sri Lanka Zawahir S

More information

Todd S. Perlstein, MD FIFTH ANNUAL SYMPOSIUM

Todd S. Perlstein, MD FIFTH ANNUAL SYMPOSIUM Todd S. Perlstein, MD FIFTH ANNUAL SYMPOSIUM Faculty Disclosure I have no financial interest to disclose No off-label use of medications will be discussed FIFTH ANNUAL SYMPOSIUM Recognize changes between

More information

a Medical Physics Department, Guy s & St Thomas NHS Foundation Trust and b King s College School of Medicine, St Thomas Campus, London, UK

a Medical Physics Department, Guy s & St Thomas NHS Foundation Trust and b King s College School of Medicine, St Thomas Campus, London, UK Devices and technology 27 Validation of the Omron 705IT (HEM-759-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol Andrew Coleman a, Paul Freeman

More information

Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension

Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension M. EL ASSAAD, J. TOPOUCHIAN, G. LABAKI, R. ASMAR L Institut

More information

There is convincing evidence in clinical studies

There is convincing evidence in clinical studies AJH 1998;11:1413 1417 Reliability of Reporting Self-Measured Blood Pressure Values by Hypertensive Patients Thomas Mengden, Rosa Maria Hernandez Medina, Belen Beltran, Elena Alvarez, Karin Kraft, and Hans

More information

The magnitude and duration of ambulatory blood pressure reduction following acute exercise

The magnitude and duration of ambulatory blood pressure reduction following acute exercise Journal of Human Hypertension (1999) 13, 361 366 1999 Stockton Press. All rights reserved 0950-9240/99 $12.00 http://www.stockton-press.co.uk/jhh ORIGINAL ARTICLE The magnitude and duration of ambulatory

More information

& Wilkins. a Division of Cardiology, Schulich Heart Centre, b Institute for Clinical and

& Wilkins. a Division of Cardiology, Schulich Heart Centre, b Institute for Clinical and Original article 333 Optimum frequency of office blood pressure measurement using an automated sphygmomanometer Martin G. Myers a, Miguel Valdivieso a and Alexander Kiss b,c Objective To determine the

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 57 Comparison of wrist-type and arm-type 24-h blood pressure monitoring devices for ambulatory use Takahiro Komori a, Kazuo Eguchi a, Satoshi Hoshide a, Bryan Williams b and Kazuomi

More information

DR JIRAR TOPOUCHIAN PROF PAROUNAK ZELVEIAN PROF ROLAND ASMAR. September 8 th, Principal Investigator and Study Chair:

DR JIRAR TOPOUCHIAN PROF PAROUNAK ZELVEIAN PROF ROLAND ASMAR. September 8 th, Principal Investigator and Study Chair: Page 1 of 10 Validation of the automatic blood pressure measurements device, the OMRON EVOLV (HEM- 7600 T-E) in Pregnancy according to the Modified European Society of Hypertension International Protocol

More information

GETTING THE NUMBERS RIGHT: ACCURATE MEASUREMENT OF BLOOD PRESSURE

GETTING THE NUMBERS RIGHT: ACCURATE MEASUREMENT OF BLOOD PRESSURE GETTING THE NUMBERS RIGHT: ACCURATE MEASUREMENT OF BLOOD PRESSURE Pamela B. Morris, MD, FACC, FAHA, FASPC, FNLA Professor Medicine and Cardiovascular Diseases Director, Seinsheimer Cardiovascular Health

More information

Comparison of arbitrary definitions of circadian time periods with those determined by wrist actigraphy in analysis of ABPM data

Comparison of arbitrary definitions of circadian time periods with those determined by wrist actigraphy in analysis of ABPM data Journal of Human Hypertension (1999) 13, 449 453 1999 Stockton Press. All rights reserved 0950-9240/99 $12.00 http://www.stockton-press.co.uk/jhh ORIGINAL ARTICLE Comparison of arbitrary definitions of

More information

BLOOD PRESSURE MEASUREMENT HOME BASED OR OFFICE BP MONITORING WHICH, HOW AND WHEN?

BLOOD PRESSURE MEASUREMENT HOME BASED OR OFFICE BP MONITORING WHICH, HOW AND WHEN? BLOOD PRESSURE MEASUREMENT HOME BASED OR OFFICE BP MONITORING WHICH, HOW AND WHEN? DECLARATION OF INTEREST Medical Director and Chairman, Advisory Board dabl Ltd., Dublin, Ireland. BLOOD PRESSURE MEASUREMENT

More information

a Centre de Médecine Cardiovasculaire, Paris, France, b Lebanese Hospital and Received 7 June 2009 Revised 15 September 2009 Accepted 9 October 2009

a Centre de Médecine Cardiovasculaire, Paris, France, b Lebanese Hospital and Received 7 June 2009 Revised 15 September 2009 Accepted 9 October 2009 42 Original article Validation of three professional devices measuring office blood pressure according to three different methods: the Omron BP, the Omron HBP T and the Pic Indolor Professional Roland

More information

Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard

Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard Evaluation of the Accutracker I1 noninvasive ambulatory blood pressure recorder according to the AAMI Standard EOIN O'BRIENt, NEIL ATKINS, JAMES SHERIDAN, COLM KILLEEN, RICHARD COSTELLOE, SHANE LYDON and

More information

BRIEF COMMUNICATIONS. KEY WORDS: Ambulatory blood pressure monitoring, placebo effect, antihypertensive drug trials.

BRIEF COMMUNICATIONS. KEY WORDS: Ambulatory blood pressure monitoring, placebo effect, antihypertensive drug trials. AJH 1995; 8:311-315 BRIEF COMMUNICATIONS Lack of Placebo Effect on Ambulatory Blood Pressure Giuseppe Mancia, Stefano Omboni, Gianfranco Parati, Antonella Ravogli, Alessandra Villani, and Alberto Zanchetti

More information

Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol

Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol Clinical Evaluation of an Oscillometric NIBP Technology During Hemodialysis According to the British Hypertension Society Protocol Abstract Objectives Existing concerns over the accuracy of automated blood

More information

Received 10 April 2008 Revised 23 June 2008 Accepted 24 June 2008

Received 10 April 2008 Revised 23 June 2008 Accepted 24 June 2008 342 Original article Development of an accurate oscillometric blood pressure device for low resource settings Annemarie de Greeff, Hannah Nathan, Nina Stafford, Bing Liu and Andrew H. Shennan Objective

More information

White coat and masked hypertension

White coat and masked hypertension White coat and masked hypertension Conflict of interest Support from Spacelabs, Microlife. Honoraria from Novartis, Elpen, Boeringer-Ingelheim, CANA, Lilly, MSD, Sanofi, Menarini, Ciezi, Astra-Zeneca.

More information

Data Collection Worksheet

Data Collection Worksheet Data Collection Worksheet 1. Has a doctor or nurse ever said that you have: High blood pressure or hypertension? [ ] No [ ] Yes [ ] Not Sure 1.a. If Yes, then at what age were you first told this? Age

More information

Automatic oscillometric blood pressure devices are frequently

Automatic oscillometric blood pressure devices are frequently Arterial Stiffness as Underlying Mechanism of Disagreement Between an Oscillometric Blood Pressure Monitor and a Sphygmomanometer Nicole M. van Popele, Willem Jan W. Bos, Nicole A.M. de Beer, Deirdre A.M.

More information

Prognostic significance of blood pressure measured in the office, at home and during ambulatory monitoring in older patients in general practice

Prognostic significance of blood pressure measured in the office, at home and during ambulatory monitoring in older patients in general practice (2005) 19, 801 807 & 2005 Nature Publishing Group All rights reserved 0950-9240/05 $30.00 www.nature.com/jhh ORIGINAL ARTICLE Prognostic significance of blood pressure measured in the office, at home and

More information

1. Department of Gynecology and Obstetrics, St. Joseph's Hospital Berlin Tempelhof, Germany

1. Department of Gynecology and Obstetrics, St. Joseph's Hospital Berlin Tempelhof, Germany Page 1 of 9 Validation of the TONOPORT VI ambulatory blood pressure monitor, according to the European Society of Hypertension International Protocol revision 2010 Michael Abou-Dakn 1, Cornelius Döhmen

More information

Ambulatory BP Monitoring: Getting the Diagnosis of Hypertension Right. Anthony J. Viera, MD, MPH, FAHA Professor and Chair

Ambulatory BP Monitoring: Getting the Diagnosis of Hypertension Right. Anthony J. Viera, MD, MPH, FAHA Professor and Chair Ambulatory BP Monitoring: Getting the Diagnosis of Hypertension Right Anthony J. Viera, MD, MPH, FAHA Professor and Chair Objectives Review limitations of office BP in making a correct diagnosis of hypertension

More information

Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods

Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods CLINICAL STUDIES Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods Linda A. Green, RN, MS, Robin D. Froman, RN, PhD, FAAN Objective: To determine the equivalence of auscultatory

More information

Importance Of Taking Manual Blood Pressure In Both Arms

Importance Of Taking Manual Blood Pressure In Both Arms Importance Of Taking Manual Blood Pressure In Both Arms Its importance to health is now well-recognised, in that high blood pressure can lead to Have a look at our slide show on how to take a blood pressure

More information

ORIGINAL ARTICLE AMBULATORY BLOOD PRESSURE IN OBESITY. Introduction. Patients and Methods

ORIGINAL ARTICLE AMBULATORY BLOOD PRESSURE IN OBESITY. Introduction. Patients and Methods Vol. 2, Issue 1, pages 31-36 ORIGINAL ARTICLE AMBULATORY BLOOD PRESSURE IN OBESITY By Alejandro de la Sierra, MD Luis M. Ruilope, MD Hypertension Units, Hospital Clinico, Barcelona & Hospital 12 de Octubre,

More information

DEVICES AND TECHNOLOGY

DEVICES AND TECHNOLOGY DEVICES AND TECHNOLOGY Validation of a new algorithm for the BPM-100 electronic oscillometric office blood pressure monitor James M. Wright ab, Gurdial S. Mattu a, Thomas L. Perry Jr ab, Mark E. Gelfer

More information

Unreliable oscillometric blood pressure measurement: prevalence, repeatability and characteristics of the phenomenon

Unreliable oscillometric blood pressure measurement: prevalence, repeatability and characteristics of the phenomenon (2009) 23, 794 800 & 2009 Macmillan Publishers Limited All rights reserved 0950-9240/09 $32.00 www.nature.com/jhh ORIGINAL ARTICLE Unreliable oscillometric blood pressure measurement: prevalence, repeatability

More information

Indirect ambulatory blood pressure monitoring devices

Indirect ambulatory blood pressure monitoring devices AJH 1992;5:880-886 Comparison of Ambulatory and Clinic Blood Pressure and Heart Rate in Older Persons With Isolated Systolic Hypertension Gale H. Rutan, Robert H. McDonald, and Lewis H. Kuller We compared

More information

University of Padova, Padua, Italy, and HARVEST Study Group, Italy

University of Padova, Padua, Italy, and HARVEST Study Group, Italy University of Padova, Padua, Italy, and HARVEST Study Group, Italy ISOLATED SYSTOLIC HYPERTENSION IN THE YOUNG DOES NOT IMPLY AN INCREASED RISK OF FUTURE HYPERTENSION NEEDING TREATMENT Mos L, Saladini

More information

Prognostic significance of blood pressure measured on rising

Prognostic significance of blood pressure measured on rising (2001) 15, 413 417 2001 Nature Publishing Group All rights reserved 0950-9240/01 $15.00 www.nature.com/jhh ORIGINAL ARTICLE Prognostic significance of blood pressure measured on rising P Gosse, C Cipriano,

More information

Vital Signs. Vital Signs. Pulse. Temperature. Respiration. Blood Pressure

Vital Signs. Vital Signs. Pulse. Temperature. Respiration. Blood Pressure Vital Signs Jarvis, Chapter 9 Vital Signs Classic Vital Signs TPR/BP Temperature Pulse Respirations Blood Pressure Additional Vital Signs Height Weight BMI (Kg/m2) or (702Xlbs/in2) Supine, orthostatic

More information

External Oscillatory Blood Pressure - EOBPTM

External Oscillatory Blood Pressure - EOBPTM External Oscillatory Blood Pressure - EOBPTM Development of Novel Principle To Measure Blood Pressure Mindaugas Pranevicius, M.D., Osvaldas Pranevicius, M.D., Ph.D. Pranevicius Biotech Inc., Forest Hills,

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 16 Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-71-E)

More information

Copyright: DOI link to paper: Date deposited: This work is licensed under a Creative Commons Attribution 2.5 Generic License

Copyright: DOI link to paper: Date deposited: This work is licensed under a Creative Commons Attribution 2.5 Generic License Zheng DC, Liu CY, Amoore J, Mieke S, Murray A. Need for Re-validation of Automated Blood Pressure Devices for use in Unstable Conditions. In: 215 Computing in Cardiology Conference (CinC). 215, Nice, France:

More information

Echocardiographic definition of left ventricular hypertrophy in the hypertensive: which method of indexation of left ventricular mass?

Echocardiographic definition of left ventricular hypertrophy in the hypertensive: which method of indexation of left ventricular mass? Journal of Human Hypertension (1999) 13, 505 509 1999 Stockton Press. All rights reserved 0950-9240/99 $12.00 http://www.stockton-press.co.uk/jhh ORIGINAL ARTICLE Echocardiographic definition of left ventricular

More information

THE ACCURATE ASSESSMENT

THE ACCURATE ASSESSMENT ORIGINAL INVESTIGATION Consistency of Blood Pressure Differences Between the Left and Right s Kazuo Eguchi, MD; Mona Yacoub, MD; Juhee Jhalani; William Gerin, PhD; Joseph E. Schwartz, PhD; Thomas G. Pickering,

More information

Evaluation of the Extent and Duration of the ABPM Effect in Hypertensive Patients

Evaluation of the Extent and Duration of the ABPM Effect in Hypertensive Patients Journal of the American College of Cardiology Vol. 40, No. 4, 2002 2002 by the American College of Cardiology Foundation ISSN 0735-1097/02/$22.00 Published by Elsevier Science Inc. PII S0735-1097(02)02011-9

More information

The measurement of blood pressure and hypertension. Handout Fenyvesi Tamás III.Department of Medicine

The measurement of blood pressure and hypertension. Handout Fenyvesi Tamás III.Department of Medicine The measurement of blood pressure and hypertension Handout Fenyvesi Tamás III.Department of Medicine 1 2 History of the blood pressure measurement direct Stephen HALES 1726 horse carotid indirect Riva-Rocci

More information

Importance of Ambulatory Blood Pressure Monitoring in Adolescents

Importance of Ambulatory Blood Pressure Monitoring in Adolescents Importance of Ambulatory Blood Pressure Monitoring in Adolescents Josep Redon, MD, PhD, FAHA Internal Medicine Hospital Clinico Universitario de Valencia University of Valencia CIBERObn Instituto de Salud

More information

How well do office and exercise blood pressures predict sustained hypertension? A Dundee Step Test Study

How well do office and exercise blood pressures predict sustained hypertension? A Dundee Step Test Study (2000) 14, 429 433 2000 Macmillan Publishers Ltd All rights reserved 0950-9240/00 $15.00 www.nature.com/jhh ORIGINAL ARTICLE How well do office and exercise blood pressures predict sustained hypertension?

More information

Use and Interpretation of Home Blood Pressure Monitoring

Use and Interpretation of Home Blood Pressure Monitoring Use and Interpretation of Home Blood Pressure Monitoring BLOOD PRESSURE in year 2015 PARAMETERS TODAY Gianfranco Parati University of Milano-Bicocca & Ospedale S.Luca, IRCCS, Istituto Auxologico Italiano

More information

Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents

Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents Original article http://dx.doi.org/10.3345/kjp.2011.54.11.463 Korean J Pediatr 2011;54(11):463-469 Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents

More information

The effect of a change in ambient temperature on blood pressure in normotensives

The effect of a change in ambient temperature on blood pressure in normotensives (2001) 15, 113 117 2001 Nature Publishing Group All rights reserved 0950-9240/01 $15.00 www.nature.com/jhh ORIGINAL ARTICLE The effect of a change in ambient temperature on blood pressure in normotensives

More information

Taking and recording blood pressure and pulse. City Gate Training Centre all rights reserved

Taking and recording blood pressure and pulse. City Gate Training Centre all rights reserved Taking and recording blood pressure and pulse City Gate Training Centre all rights reserved Course Content: Definition of Blood Pressure Normal Blood Pressure Readings Systolic and Diastolic Blood Pressure

More information

Assessing Blood Pressure for Clinical Research: Pearls & Pitfalls

Assessing Blood Pressure for Clinical Research: Pearls & Pitfalls Assessing Blood Pressure for Clinical Research: Pearls & Pitfalls Anthony J. Viera, MD, MPH, FAHA Department of Family Medicine Hypertension Research Program UNC School of Medicine Objectives Review limitations

More information

DIFFERENTE RELAZIONE TRA VALORI PRESSORI E MASSA VENTRICOLARE SX NEI DUE SESSI IN PAZIENTI IPERTESI.

DIFFERENTE RELAZIONE TRA VALORI PRESSORI E MASSA VENTRICOLARE SX NEI DUE SESSI IN PAZIENTI IPERTESI. DIFFERENTE RELAZIONE TRA VALORI PRESSORI E MASSA VENTRICOLARE SX NEI DUE SESSI IN PAZIENTI IPERTESI. Franco Cipollini, Carlo Porta, Enrica Arcangeli, Carla Breschi, & Giuseppe Seghieri Azienda USL 3, Ambulatorio

More information

Comparison of manual versus automated blood pressure measurement in intensive care unit, coronary care unit, and emergency room.

Comparison of manual versus automated blood pressure measurement in intensive care unit, coronary care unit, and emergency room. Comparison of manual versus automated blood pressure measurement in intensive care unit, coronary care unit, and emergency room Abstract Ahmad Mirdamadi (1), Mostafa Etebari (2) Original Article BACKGROUND:

More information

Mercury versus Non-Mercury Sphygmomanometers: A Review of the Comparative Clinical Effectiveness and Guidelines

Mercury versus Non-Mercury Sphygmomanometers: A Review of the Comparative Clinical Effectiveness and Guidelines TITLE: Mercury versus Non-Mercury Sphygmomanometers: A Review of the Comparative Clinical Effectiveness and Guidelines DATE: 25 September 2012 CONTEXT AND POLICY ISSUES A mercury manual sphygmomanometer

More information

Calculation of Trough-to-Peak Ratio in the Research Unit Setting

Calculation of Trough-to-Peak Ratio in the Research Unit Setting A]H 1996; 9:71S-75S Calculation of Trough-to-Peak Ratio in the Research Unit Setting Advantages and Disadvantages Henry L. Elliott and Peter A. Meredith The trough-to-peak ratio for the response to an

More information

In clinical practice, blood pressure (BP) treatment

In clinical practice, blood pressure (BP) treatment Original Paper Does Hidden Undercuffing Occur Among Obese Patients? Effect of Arm Sizes and Other Predictors of the Difference Between Wrist and Upper Arm Blood Pressures Hardik Doshi, MD; Alan B. Weder,

More information

K/DOQI Clinical Practice Guidelines on Hypertension and Antihypertensive Agents in Chronic Kidney Disease

K/DOQI Clinical Practice Guidelines on Hypertension and Antihypertensive Agents in Chronic Kidney Disease K/DOQI Clinical Practice Guidelines on Hypertension and Antihypertensive Agents in Chronic Kidney Disease GUIDELINE 1: GOALS OF ANTIHYPERTENSIVE THERAPY IN CKD Hypertension is common in CKD, and is a risk

More information

Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation

Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation ORIGINAL RESEARCH Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation Tyler S Lamb BSc Amar Thakrar MD, FRCPC Mahua Ghosh MD, PhD Merne P Wilson MSc Thomas W Wilson

More information

가정혈압의활용 CARDIOVASCULAR CENTER. Wook Bum Pyun M.D., Ph.D. HOME BLOOD PRESSURE MONITORING. Ewha Womans University, school of Medicine

가정혈압의활용 CARDIOVASCULAR CENTER. Wook Bum Pyun M.D., Ph.D. HOME BLOOD PRESSURE MONITORING. Ewha Womans University, school of Medicine 가정혈압의활용 HOME BLOOD PRESSURE MONITORING CARDIOVASCULAR CENTER Wook Bum Pyun M.D., Ph.D. pwb423@ewha.ac.kr Ewha Womans University, school of Medicine Non-Invasive Blood Pressure Measurement 5-20% Resistant

More information

Please do not hesitate to contact our dedicated service unit for more details regarding this program at

Please do not hesitate to contact our dedicated service unit for more details regarding this program at The Blue Cross and Blue Shield Federal Employee Program (FEP) and the American Medical Association (AMA) have come together in a collaborative effort to provide physicians with resources designed to improve

More information

Hypertension is a common medical disorder, affecting. Is Resistant Hypertension Really Resistant? Mark A. Brown, Megan L. Buddle, and Allison Martin

Hypertension is a common medical disorder, affecting. Is Resistant Hypertension Really Resistant? Mark A. Brown, Megan L. Buddle, and Allison Martin AJH 2001; 14:1263 1269 Is Resistant Hypertension Really Resistant? Mark A. Brown, Megan L. Buddle, and Allison Martin Background: Managing resistant hypertension is difficult and mostly involves expensive

More information

Section 03: Pre Exercise Evaluations and Risk Factor Assessment

Section 03: Pre Exercise Evaluations and Risk Factor Assessment Section 03: Pre Exercise Evaluations and Risk Factor Assessment ACSM Guidelines: Chapter 3 Pre Exercise Evaluations ACSM Manual: Chapter 3 Risk Factor Assessments HPHE 4450 Dr. Cheatham Purpose The extent

More information

Hypertension and Cardiovascular Disease

Hypertension and Cardiovascular Disease Hypertension and Cardiovascular Disease Copyright 2017 by Sea Courses Inc. All rights reserved. No part of this document may be reproduced, copied, stored, or transmitted in any form or by any means graphic,

More information

Autonomic Variation of Blood Pressure in Middle Aged Diabetics: A Prospective Study

Autonomic Variation of Blood Pressure in Middle Aged Diabetics: A Prospective Study Original Article DOI: 10.17354/ijss/2016/16 Autonomic Variation of Blood Pressure in Middle Aged Diabetics: A Prospective Study M Usharani 1, K Chandini 2 1 Professor and Head, Department of Physiology,

More information

Hypertension Update 2016 AREEF ISHANI, MD MS CHIEF OF MEDICINE MINNEAPOLIS VA MEDICAL CENTER PROFESSOR OF MEDICINE UNIVERSITY OF MINNESOTA

Hypertension Update 2016 AREEF ISHANI, MD MS CHIEF OF MEDICINE MINNEAPOLIS VA MEDICAL CENTER PROFESSOR OF MEDICINE UNIVERSITY OF MINNESOTA Hypertension Update 2016 AREEF ISHANI, MD MS CHIEF OF MEDICINE MINNEAPOLIS VA MEDICAL CENTER PROFESSOR OF MEDICINE UNIVERSITY OF MINNESOTA Case 1 What should be your BP goal for an elderly (> 75 yrs of

More information

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE. MEASURING BLOOD PRESSURE - MANUAL (equ04)

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE. MEASURING BLOOD PRESSURE - MANUAL (equ04) SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE TITLE: ISSUED FOR: MEASURING BLOOD PRESSURE - MANUAL Nursing DATE: REVIEWED: PAGES: 2/80 7/17 1 of 5 RESPONSIBILITY: RN, LPN, Patient Care Technician Multi-skilled

More information

Blood Pressure Competency Training

Blood Pressure Competency Training Blood Pressure Competency Training WORKBOOK Table of Contents 3..... Introduction 4..... Background 6..... Equipment Accurate blood pressure What impacts a patient s blood pressure Strategies for proper

More information

Techniques of Vital Signs. John Gazewood, MD, MSPH Department of Family Medicine

Techniques of Vital Signs. John Gazewood, MD, MSPH Department of Family Medicine Techniques of Vital Signs John Gazewood, MD, MSPH Department of Family Medicine Objectives Define normal in several ways. Describe correct technique of measuring vital signs. Describe correct technique

More information

How do we diagnose hypertension today? Presentation Subtitle

How do we diagnose hypertension today? Presentation Subtitle How do we diagnose hypertension today? Presentation Subtitle Renata Cífková Case 1 JM, a 64-year-old lady referred to our center because of undesirable effects of her antihypertensive medication Personal

More information

High-dose monotherapy vs low-dose combination therapy of calcium channel blockers and angiotensin receptor blockers in mild to moderate hypertension

High-dose monotherapy vs low-dose combination therapy of calcium channel blockers and angiotensin receptor blockers in mild to moderate hypertension (2005) 19, 491 496 & 2005 Nature Publishing Group All rights reserved 0950-9240/05 $30.00 www.nature.com/jhh ORIGINAL ARTICLE High-dose monotherapy vs low-dose combination therapy of calcium channel blockers

More information

Received 24 February 2015 Revised 29 April 2015 Accepted 20 May 2015

Received 24 February 2015 Revised 29 April 2015 Accepted 20 May 2015 Original article 1 Clinical practice of ambulatory versus home blood pressure monitoring in hypertensive patients Jorge A. Paolasso, Florencia Crespo, Viviana Arias, Eduardo A. Moreyra, Ariel Volmaro,

More information

Clinical Medical Policy Department Clinical Affairs Division DESCRIPTION

Clinical Medical Policy Department Clinical Affairs Division DESCRIPTION Ambulatory Blood Pressure Monitoring (ABPM) [For the list of services and procedures that need preauthorization, please refer to www.mcs.com.pr. go to Comunicados a Proveedores, and click Cartas Circulares.]

More information

The increasing awareness of hypertension as a serious

The increasing awareness of hypertension as a serious AJH 2002; 15:787 792 Evaluation of the Performance of a Wrist Blood Pressure Measuring Device With a Position Sensor Compared to Ambulatory 24-Hour Blood Pressure Measurements Sakir Uen, Burkard Weisser,

More information

Blood Pressure Monitors Nissei

Blood Pressure Monitors Nissei We have made it easy for you to find a PDF Ebooks without any digging. And by having access to our ebooks online or by storing it on your computer, you have convenient answers with blood pressure monitors

More information

Summary of recommendations

Summary of recommendations Summary of recommendations Measuring blood pressure (BP) Use the recommended technique at every BP reading to ensure accurate measurements and avoid common errs. Pay particular attention to the following:

More information

Does Talking and Different Body (Sitting, Supine, Standing) Positions affect Blood Pressure?

Does Talking and Different Body (Sitting, Supine, Standing) Positions affect Blood Pressure? International Journal of Nursing December 2018, Vol. 5,. 2, pp. 94-99 ISSN 2373-7662 (Print) 2373-7670 (Online) Copyright The Author(s). All Rights Reserved. Published by American Research Institute for

More information

Thesis for doctoral degree (Ph.D.) 2008 ROLE OF AMBULATORY BLOOD PRESSURE MONITORING AFTER PEDIATRIC RENAL TRANSPLANTATION Rafael Tomás Krmar

Thesis for doctoral degree (Ph.D.) 2008 ROLE OF AMBULATORY BLOOD PRESSURE MONITORING AFTER PEDIATRIC RENAL TRANSPLANTATION Rafael Tomás Krmar Thesis for doctoral degree (Ph.D.) 2008 Thesis for doctoral degree (Ph.D.) 2008 ROLE OF AMBULATORY BLOOD PRESSURE MONITORING AFTER PEDIATRIC RENAL TRANSPLANTATION Rafael Tomás Krmar ROLE OF AMBULATORY

More information

Circadian rhythm of blood pressure is transformed from a dipper to a non-dipper pattern in shift workers with hypertension

Circadian rhythm of blood pressure is transformed from a dipper to a non-dipper pattern in shift workers with hypertension (2002) 16, 193 197 2002 Nature Publishing Group All rights reserved 0950-9240/02 $25.00 www.nature.com/jhh ORIGINAL ARTICLE Circadian rhythm of blood pressure is transformed from a dipper to a non-dipper

More information

Primary hypertension in adults

Primary hypertension in adults Primary hypertension in adults NICE provided the content for this booklet which is independent of any company or product advertised Hypertension Welcome NICE published an updated guideline on the diagnosis

More information

Blood Pressure LIMBO How Low To Go?

Blood Pressure LIMBO How Low To Go? Blood Pressure LIMBO How Low To Go? Joseph L. Kummer, MD, FACC Bryan Heart Spring Conference April 21 st, 2018 Hypertension Epidemiology Over a billion people have hypertension Major cause of morbidity

More information

DIURNAL VARIATIONS IN BLOOD PRESSURE AND THEIR RELATION WITH CAROTID ARTERY INTIMA-MEDIA THICKENING

DIURNAL VARIATIONS IN BLOOD PRESSURE AND THEIR RELATION WITH CAROTID ARTERY INTIMA-MEDIA THICKENING DIURNAL VARIATIONS IN BLOOD PRESSURE AND THEIR RELATION WITH CAROTID ARTERY INTIMA-MEDIA THICKENING Sh Narooei (1), B Soroor (2), F Zaker (3) Abstract INTRODUCTION: Hypertension is a very common cardiovascular

More information

2. Measure a subject's blood pressure and heart rate both at rest and during exercise.

2. Measure a subject's blood pressure and heart rate both at rest and during exercise. Lab Activity 11 The Cardiovascular System Student Learning Objectives After completing this lab, you should be able to: 1. Define, explain and correctly use the key terms. 2. Measure a subject's blood

More information

Normal Ambulatory Blood Pressure: A Clinical-Practice- Based Analysis of Recent American Heart Association Recommendations

Normal Ambulatory Blood Pressure: A Clinical-Practice- Based Analysis of Recent American Heart Association Recommendations The American Journal of Medicine (2006) 119, 69.e13-69.e18 CLINICAL RESEARCH STUDY Normal Ambulatory Blood Pressure: A Clinical-Practice- Based Analysis of Recent American Heart Association Recommendations

More information

Comparison of the Assessment of Orthostatic Hypotension Using Peripheral and Central Blood Pressure Measurements

Comparison of the Assessment of Orthostatic Hypotension Using Peripheral and Central Blood Pressure Measurements Original Article J Clin Med Res. 2018;10(4):309-313 Comparison of the Assessment of Orthostatic Hypotension Using Peripheral and Central Blood Pressure Measurements Kannayiram Alagiakrishnan a, d, Ruojin

More information

Hypertension Putting the Guidelines into Practice

Hypertension Putting the Guidelines into Practice Hypertension 2017 Putting the Guidelines into Practice Disclosures Relationships with commercial interests: Grants/Research Support: Speakers Bureau/Honoraria: Consulting Fees: Data Safety and Monitoring:

More information

KDIGO Controversies Conference on Blood Pressure in CKD

KDIGO Controversies Conference on Blood Pressure in CKD KDIGO Controversies Conference on Blood Pressure in CKD September 7-10, 2017 Edinburgh, Scotland Kidney Disease: Improving Global Outcomes (KDIGO) is an international organization whose mission is to improve

More information

Verapamil SR and trandolapril combination therapy in hypertension a clinical trial of factorial design

Verapamil SR and trandolapril combination therapy in hypertension a clinical trial of factorial design Br J Clin Pharmacol 1998; 45: 491 495 Verapamil SR and trandolapril combination therapy in hypertension a clinical trial of factorial design Juergen Scholze, 1 Peter Zilles 2 & Daniele Compagnone 2 on

More information

Chapter-V. Summary, Conclusions and Recommendations

Chapter-V. Summary, Conclusions and Recommendations Summary, Conclusions and Recommendations INTRODUCTION The work included in this thesis entitled, Circadian heart rate and blood pressure variability in apparently healthy subjects using ABPM has been divided

More information

BIOE221. Session 2. Vital Signs. Biosciences Dept.

BIOE221. Session 2. Vital Signs. Biosciences Dept. BIOE221 Session 2 Vital Signs Biosciences Dept. www.endeavour.edu.au Session Objectives cont o Understand the physiology related to the vital signs of temperature, pulse, respiration and blood pressure.

More information

Is Traditional Clinic Blood Pressure Dead?

Is Traditional Clinic Blood Pressure Dead? Royal College of Physicans May 16 th 2017 Is Traditional Clinic Blood Pressure Dead? Professor Bryan Williams MD FRCP FAHA FESC Chair of Medicine UCL Director National Institute for Health Research Biomedical

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 223 Validation of the A&D UA-705 device for self-measurement of blood pressure according to the British Hypertension Society protocol Paolo Verdecchia, Fabio Angeli, Roberto Gattobigio,

More information

SECONDARY HYPERTENSION

SECONDARY HYPERTENSION HYPERTENSION Hypertension is the clinical term used to describe a high blood pressure of 140/90 mmhg or higher (National Institute of Health 1997). It is such a health risk the World Health Organisation

More information

Assessment of techniques for measurement of blood pressure in infants and children

Assessment of techniques for measurement of blood pressure in infants and children Archives of Disease in Childhood, 1973, 48, 932. Assessment of techniques for measurement of blood pressure in infants and children A. M. ELSEED, E. A. SHINEBOURNE, and M. C. JOSEPH From the Paediatric

More information