Evaluation of the LIAISON 25 OH Vitamin D TOTAL Assay on the new LIAISON XL analyser
|
|
- Junior Carr
- 5 years ago
- Views:
Transcription
1 Evaluation of the LIAISON 25 OH Vitamin D TOTAL Assay on the new LIAISON XL analyser Prof. Etienne Cavalier Department of Clinical Chemistry University of Liège, CHU Sart-Tilman Liège, Belgium etienne.cavalier@chu.ulg.ac.be
2 Presenter Disclosure Information Evaluation of the LIAISON 25 OH Vitamin D TOTAL Assay on the new LIAISON XL analyser I will not discuss off label use or investigational use in my presentation. I have a financial relationship to disclose: Receiving an honoraria payment from DiaSorin to provide this presentation.
3 Workload on the Liaison XL at CHU de Liège 3 rd generation PTH 25(OH)-vitamin D Calcitonin Bone-ALP 22. tests/year 45. tests/year 5. tests/year 5. tests/year All tests accredited vs. ISO
4 Validation of the 25(OH)D on the Liaison XL
5 Validation procedure Precision (modified protocol based on CLSI EP-5A2: 6 serum pools were assayed in three replicates per day on five different days) Functional sensitivity Accuracy profile Lot to lot variability Method comparison: - with Liaison - with LCMS/MS (Perkin Elmer) - with Architect, Centaur, isys and Elecsys
6 Results
7 Precision Sample No. Mean, ng/ml (nmol/l) Repeatability Intermediate precision (CV %) (CV %) 1 1,7 (27) 6,5 7,1 2 21,1 (53) 3,7 3,8 3 31,9 (8) 4, 4,1 4 38,9 (97) 2,7 3,4 5 47,2 (118) 2,4 2,4 6 63,9 (16) 3,8 3,8
8 Limit of quantification 2.5 ng/ml 6.25 nmol/l
9 Accuracy profile
10 Comparison with the Liaison Liaison XL =.92x Liaison + 1.2
11 Comparison with the Liaison
12 Lot-to-lot variation
13 Comparison with other methods
14 3 DiaSorin RIA Liaison XL - DiaSorin RIA SD 11,1 Mean -3, Liaison XL =,96 x DiaSorin RIA - 2, SD -17, Average of Liaison XL and DiaSorin RIA 1 Liaison XL Intercept A -2,6 95% CI -4,5 to -1,6 Proportional differences Slope B,96 95% CI,92 to 1, DiaSorin RIA Significant systematic difference No significant proportional difference
15 LC-MS/MS with PerkinElmer kit Liaison XL =,85 x LCMS/MS PE - 1,24 Liaison XL - LC-MS/MS PERKIN ELMER SD 8,4 Mean -8, SD -25, Average of Liaison XL and LC-MS/MS PERKIN ELMER 12 Liaison XL Intercept A -1,24 95% CI -2,5 to,9 Proportional differences Slope B,85 95% CI,8 to, LC-MS/MS PERKIN ELMER No significant systematic difference Significant proportional difference
16 15 LC-MS/MS with Chromsystems kit Liaison XL =,86 x LCMS/MS CS -,9 Liaison XL - LC-MS/MS Chromsystems SD 6,5 Mean -7, SD -2, Average of Liaison XL and LC-MS/MS Chromsystems 1 Liaison XL Intercept A -,9 95% CI -3,14 to,19 Proportional differences Slope B,86 95% CI,81 to, LC-MS/MS Chromsystems No significant systematic difference Significant proportional difference
17 SD 16,4 Abbott Architect Liaison XL - Architect Mean -7, SD -31,7-5 Liaison XL =,77 x Architect + 3, Average of Liaison XL and Architect 14 Liaison XL Intercept A 3,9 95% CI 2,7 to 5,4 Proportional differences Slope B,77 95% CI,74 to, Architect Significant systematic difference Significant proportional difference
18 SD 9,7 Siemens Centaur Liaison XL - Centaur Mean -2, SD -15,3-2 Liaison XL =,91 x Centaur +, Average of Liaison XL and Centaur 1 Liaison XL Intercept A,7 95% CI -1,15 to 2,2 Proportional differences Slope B,91 95% CI,87 to, Centaur No significant systematic difference Significant proportional difference
19 3 Roche Elecsys Liaison XL - Elecsys SD 14,5 Mean -2, SD -19,6-3 Liaison XL =,94 x Elecsys -, Average of Liaison XL and Elecsys 1 Liaison XL Intercept A -,16 95% CI -1,89 to 1,28 Proportional differences Slope B,94 95% CI,89 to 1, Elecsys No significant systematic difference No significant proportional difference
20 SD 1,1 IDS isys Liaison XL - Isys Mean -11, SD -32,5-5 Liaison XL =,79 x isys -, Average of Liaison XL and Isys Intercept A -,53 Liaison XL % CI -2,16 to,88 Proportional differences Slope B,79 95% CI,75 to, Isys No significant systematic difference Significant proportional difference
21 Conclusions DiaSorin Liaison XL is a friendly and easy-to-use instrument that possesses a lot of interesting new features compared to the former Liaison. From an analytical point of view, the performances of this new instrument are really improved. Nevertheless, a better lot-to-lot consistency could be expected (especially in the lower range) Comparison studies show the good agreement with the DiaSorin RIA, Roche Elecsys and Siemens Centaur. LCMS-MS assays and ids isys are calibrated differently.
Introduction. Etienne Cavalier*, Olivier Rousselle, Nunzio Ferrante, Agnes Carlisi, Caroline Le Goff and Jean-Claude Souberbielle.
DOI 1.1515/cclm-13-138 Clin Chem Lab Med 13; aop Etienne Cavalier*, Olivier Rousselle, Nunzio Ferrante, Agnes Carlisi, Caroline Le Goff and Jean-Claude Souberbielle Technical and clinical evaluation of
More information25-Hydroxyvitamin D assay performance. Graham Carter Imperial College
25-Hydroxyvitamin D assay performance Graham Carter Imperial College www.deqas.org Outline of Talk 1. DEQAS: Organization and Objectives 2. Mean Assay Bias and Precision 3. Bias of individual samples
More information(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%)
cobas b system - performance evaluation Study report from a multicenter evaluation of the new cobas b system for the measurement of HbAc and lipid panel Introduction The new cobas b system provides a point-of-care
More informationAnnual Review Contents. Page 1. Introduction, scheme-wide trends, future developments Insulin C-Peptide 8-10
UK NEQAS Guildford Peptide Hormones Annual Review 212 Contents Page 1. Introduction, scheme-wide trends, future developments 1-3 2. Insulin 3-7 3. C-Peptide 8-1 4. Insulin interpretative exercise 1-12
More informationIntroduction. Nephrol Dial Transplant (2012) 27: doi: /ndt/gfr535 Advance Access publication 22 September 2011
Nephrol Dial Transplant (2012) 27: 1950 1956 doi: 10.1093/ndt/gfr535 Advance Access publication 22 September 2011 Interpretation of serum PTH concentrations with different kits in dialysis according to
More information1,25 Dihydroxyvitamin D
First Unique Automated 1,25 ihydroxyvitamin First fully automated, extraction free immunoassay for the accurate detection of 1,25 ihydroxyvitamin FOR OUTSIE US & CANAA ONLY Clinical background of 1,25
More informationSCIEX Vitamin D 200M Assay for the Topaz System
The First FDA-Cleared LC-MS/MS Assay for Vitamin D SCIEX Vitamin D 200M Assay for the Topaz System The first FDA-cleared LC-MS/MS assay for Vitamin D Vitamin D is an important building block for human
More informationStandardization of DiaSorin and Roche automated third generation PTH assays with an International Standard: impact on clinical populations
DOI 1.1515/cclm-213-127 Clin Chem Lab Med 214; aop Etienne Cavalier *, Pierre Delanaye, Pierre Lukas, Agnes Carlisi, Romy Gadisseur and Jean-Claude Souberbielle Standardization of DiaSorin and Roche automated
More informationDisclosure: No conflict of interests to disclose
Mei Chung, PhD MPH Assistant Professor Nutrition/Infection Unit, Department of Public Health and Community Medicine, School of Medicine, Tufts University 0 1 Disclosure: No conflict of interests to disclose
More informationUSING THE ACCESS AMH ASSAY IN YOUR LABORATORY
INFORMATION BULLETIN USING THE ACCESS AMH ASSAY IN YOUR LABORATORY ///////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////
More informationVitamin D Biomarker Analytical
Vitamin D Biomarker Analytical Issues innhanes Clifford Johnson, Director Centers for Disease Control and Prevention National Center for Health Statistics Presentation Objectives Provide brief background
More informationAccuracy and Precision Performance of the Accu-Chek Mobile System. Introduction. Method
Accuracy and Precision Performance of the Accu-Chek Mobile System I. ACCURACY The accuracy of the system was assessed according to ISO 15197. Introduction The purpose of this study was to determine the
More information25-Hydroxy Vitamin D TOTAL (25HD) on Liaison XL
25-Hydroxy Vitamin D TOTAL (25HD) on Liaison XL I. PRINCIPLE The role of vitamin D in bone and mineral metabolism was recognized from its first identification as a factor that could cure rickets. However,
More informationRecommendations from the Working Group of Senior Scottish Clinical Biochemists on Parathyroid Hormone (PTH) Targets in the Management of Renal Failure
Recommendations from the Working Group of Senior Scottish Clinical Biochemists on Parathyroid Hormone (PTH) Targets in the Management of Renal Failure Chairman: Simon Walker (SW) Group membership: Jim
More informationBone and Mineral. Comprehensive Menu for the Management of Bone and Mineral Related Diseases
Bone and Mineral Comprehensive Menu for the Management of Bone and Mineral Related Diseases Innovation to Assist in Clinical Diagnosis and Treatment DiaSorin offers a specialty line of Bone and Mineral
More informationDiscrepancy between Vitamin D Total Immunoassays due to Various Cross-reactivities
J Bone Metab ;22:7-112 http://dx.doi.org/.105/jbm..22.3.7 pissn 2287-6375 eissn 2287-7029 Original Article Discrepancy between Vitamin D Total Immunoassays due to Various Cross-reactivities Jun Hyung Lee,
More informationDBC 25-Hydroxyvitamin D
D I A G N O S T I C S B I O C H E M C A N A D A DBC 25-Hydroxyvitamin D ELISA OVERVIEW The worldwide aging and chronically ill population is increasing rapidly. It is forecasted that the Global Vitamin
More informationThis supplementary material has been provided by the authors to give readers
Supplementary Online Content LI D, Radulescu A, Shrestha RT, et al. Association of biotin ingestion with performance of hormone and nonhormone assays in healthy adults. JAMA. doi:1.11/jama.217.1375 efigure
More informationAccuSet HCV Performance Panel
Signal to cutoff (s/co) DATA SHEET OVERVIEW AccuSet HCV (0810-0204) is a 25-member validation panel of undiluted, naturally occurring plasma samples (1 vial per member, 1.0 ml per vial). Panel members
More informationAnalytical Variability Among Methods for the Measurement of 25-Hydroxyvitamin D. Still Adding to the Noise
Analytical Variability Among Methods for the Measurement of 25-Hydroxyvitamin D Still Adding to the Noise Earle W. Holmes, PhD, 1 Jean Garbincius, 2 and Kathleen M. McKenna, MBA, MT 2 From the 1 Departments
More informationCommentary by R. Little, Ph.D., NGSP Network Coordinator for the NGSP Steering Committee
College of American Pathologists (CAP) GH5 Survey Data: (updated 8/17) The American Diabetes Association (ADA) recommends that The A1C test should be performed using a method that is certified by the NGSP.
More informationCommentary by R. Little, Ph.D., NGSP Network Coordinator for the NGSP Steering Committee
College of American Pathologists (CAP) GH5 Survey Data: (updated 12/17) The American Diabetes Association (ADA) recommends that The A1C test should be performed using a method that is certified by the
More informationReview. Current Methods for Routine Clinical Laboratory Testing of Vitamin D Levels. Clinical Significance
Current Methods for Routine Clinical Laboratory Testing of Vitamin D Levels Wendy L Arneson, MS, MLS(ASCP) CM * 1 and Dean L Arneson, PharmD, PhD 2 ABSTRACT Requests for vitamin D testing have increased
More informationHarmonisation of Growth Hormone Assays: why, issues. MJ Ellis, Endolab, Canterbury Health Labs, May 2015
Harmonisation of Growth Hormone Assays: why, issues MJ Ellis, Endolab, Canterbury Health Labs, May 2015 Why measure Growth Hormone? diagnose and monitor disorders of growth usually in conjunction with
More informationIMMUNOASSAY PREMIUM - LEVEL 3 (IA PREMIUM 3)
0086 IMMUNOASSAY PREMIUM - LEVEL 3 (IA PREMIUM 3) CAT. NO. IA2640 GTIN: 05055273203868 SIZE: 12 x 5 ml CAT. NO. IA2633 GTIN: 05055273203837 SIZE: 4 x 5 ml LOT NO. 1627EC EXPIRY: 2019-04-28 INTENDED USE
More informationCollege of American Pathologists (CAP) GH2 Survey Data: (updated 12/09) 2009 GH2-B (fresh pooled samples) * = NGSP certified at the time of the survey
College of American Pathologists (CAP) GH2 Survey Data: (updated 12/09) The American Diabetes Association (ADA) recommends that laboratories use only HbA1c assay methods that have been NGSP certified and
More informationNutrition: obtaining reliable biomarker data to study the health status of populations
Nutrition: obtaining reliable biomarker data to study the health status of populations Christine M. Pfeiffer, Ph.D. Chief, Nutritional Biomarkers Branch, DLS AAAS 2011 Annual Meeting Science Without Borders
More information1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014
Comparison Study Summary Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR 97123 July 30, 2014 1. PROTOCOL This study was conducted on July 24 th, 2014 at Tuality Healthcare, Hillsboro, OR. The
More informationPerformance Characteristics of Eight Estradiol Immunoassays
Clinical Chemistry / EVALUATION OF EIGHT ESTRADIOL IMMUNOASSAYS Performance Characteristics of Eight Estradiol Immunoassays David T. Yang, MD, 1 William E. Owen, MT(ASCP), 2 Carol S. Ramsay, 3 Hui Xie,
More informationPrice of High-Throughput 25-Hydroxyvitamin D Immunoassays: Frequency of Inaccurate Results
Price of High-Throughput 25-Hydroxyvitamin D Immunoassays: Frequency of Inaccurate Results Nicole V. Tolan, 1, * Edward J. Yoon, 1, Ashley R. Brady, 1 and Gary L. Horowitz 1, Background: With a 10-year
More informationFree hormone estimates. Never ending story.
Free hormone estimates Never ending story. Roger Ekins (1926-2016) - saturation analysis - comp. immunoassay - microarray - RIA (T4, B12) - Free T4 (never really succeeded) Free hormone hypothesis Law
More informationCommercial insulin immunoassays fail to detect commonly prescribed insulin analogues
Commercial insulin immunoassays fail to detect commonly prescribed insulin analogues Ceri Parfitt Trainee Clinical Scientist Blood Sciences Department Royal Devon and Exeter NHS Foundation Trust November
More informationComparison of Five Automated Serum and Whole Blood Folate Assays
Clinical Chemistry / FIVE AUTOMATED FOLATE ASSAYS Comparison of Five Automated Serum and Whole Blood Folate Assays William E. Owen, MT(ASCP), 1 and William L. Roberts, MD, PhD 2 Key Words: Hemolysate;
More informationStandardization and Certification of the ADVIA Centaur Vitamin D Total Assay
Standardization and Certification of the ADVIA Centaur Vitamin D Total Assay Authors: James Freeman Paul Sibley, PhD Neil Parker, MSc Ryan Spears Kimberly S. Wilson H. Roma Levy, MS Affiliations: Siemens
More informationVitamin D Assay in Human Serum Samples: A Review of Analysis Methods
Johns Hopkins University From the SelectedWorks of Rui Cao Spring February 6, 2017 Vitamin D Assay in Human Serum Samples: A Review of Analysis Methods Rui Cao, Johns Hopkins University Available at: https://works.bepress.com/rui-cao/2/
More informationIMMUNOASSAY PREMIUM PLUS - LEVEL 2 (IA PREMIUM PLUS 2)
0086 IMMUNOSSY PREMIUM PLUS - LEVEL 2 (I PREMIUM PLUS 2) CT. NO. I3110 GTIN: 05055273207262 SIZE: 12 x 5 ml CT. NO. I3112 GTIN: 05055273207286 SIZE: 4 x 5 ml LOT NO. 1727EC EXPIRY: 2020-04-28 INTENDED
More informationSerology and International units
Serology and International units L. Grangeot-Keros, National Reference Laboratory for Rubella, Virology Department, A. Béclère Hospital, Clamart, France Detection of rubella-specific IgG antibody Assays
More informationGentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000
v2-jan214 Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4 Intended Use The canine CRP immunoassay on Abbott's Architect c4 is an in vitro diagnostic test for quantitative determination
More informationMultichem Infectious Disease QC
Infectious Disease QC Contents Introduction....3 Quality Control in the Laboratory...4 Infectious Disease (ID) QC....5 Ready-to-use Concept....6 NRL...7 QConnect...8 IAMQC in Collaboration with QConnect....
More informationZIVAK TECHNOLOGIES. Multitasker-LC Walk Away UHPLC Analyzer For LC-MS/MS
ZIVAK TECHNOLOGIES Multitasker-LC Walk Away UHPLC Analyzer For LC-MS/MS Clinical Diagnostic Market Roche, Siemens,Midray Abbott, Beckman Currrent Clinical Diagnostic Market Immunoassay, ELISA, RIA,FIA
More informationIntegration of steroids analysis in serum using LC-MS/MS with full-automated sample preparation
PO-CON69E Integration of steroids analysis in serum using LC-MS/MS with full-automated sample preparation MSACL 6 EU Stéphane Moreau, Daisuke Kawakami, Toshikazu Minohata Shimadzu Europe GmbH, Duisburg,
More informationHIV-1 AccuVert Seroconversion Panel (PRB966) / Batch #
s/co c/ml DATA SHEET OVERVIEW 0600-0248 (PRB966) is a 10-member panel of undiluted, naturally occurring plasma samples. Panel members represent serial bleeds collected from a single patient over the course
More information25 OH Vitamin D Rapid test
INSTRUCTION FOR USE REF:GDB 7120-25T 25 OH Vitamin D Rapid test Rapid Whole Blood Vitamin D Test A Rapid Sandwich Immunochromatographic Test for the Quantitative Detection of total 25-OH Vitamin D in human
More informationMass Spectrometry Made Simple for Clinical Diagnostic Labs
Mass Spectrometry Made Simple for Clinical Diagnostic Labs The SCIEX Topaz System with FDA-Cleared Topaz Vitamin D 200M Assay Kit Mass Spectrometry Made Simple SCIEX Topaz System for Clinical Diagnostics
More informationVitamin D A game changer for Clinical Laboratories
Vitamin D A game changer for Clinical Laboratories Michael Wright Clinical Chemistry and Endocrinology Department, SEALS Prince of Wales Hospital Sydney, Australia What is 25OH Vitamin D? 25OH Vitamin
More informationACCESS hstni SCIENTIFIC LITERATURE
ACCESS hstni SCIENTIFIC LITERATURE 2017 2018 Table of contents Performance Evaluation of Access hstni A critical evaluation of the Beckman Coulter Access hstni: Analytical performance, reference interval
More informationCortisol Assays. The Good, The Bad and The Indifferent. David Ducroq. Cardiff and Vale ulhb WEQAS. Unit 6, Parc Tŷ Glas. Llanishen.
Cortisol Assays The Good, The Bad and The Indifferent David Ducroq Cardiff and Vale ulhb WEQAS Unit 6, Parc Tŷ Glas Llanishen Cardiff www.weqas.com Summary Brief overview of current methods and challenges
More informationCommutability studies undertaken by the LNE : the case of lipid and lipoprotein testing. Vincent Delatour, PhD
Commutability studies undertaken by the LNE : the case of lipid and lipoprotein testing Vincent Delatour, PhD 1 Traceability in laboratory medicine : regulatory drivers Reform of medical biology in France
More informationChallenges of implementing JCTLM methods in the routine Clinical Laboratories. R.J. Singh, Ph.D. Mayo Clinic
Challenges of implementing JCTLM methods in the routine Clinical Laboratories R.J. Singh, Ph.D. Mayo Clinic 100s of Millions of 25-OH Vitamin D test have been run. Total 25-OH Vitamin D test is not a useful
More informationTable. Analytical characteristics of commercial and research cardiac troponin I and T assays declared by the manufacturer.
Table. Analytical characteristics of commercial and research cardiac troponin I and T assays declared by the manufacturer. Commercially available assays - Company/ platform(s)/ assay LoB a LoD b 99 th
More informationKeywords: albuminuria; albumin/creatinine ratio (ACR); measurements. Introduction
Clin Chem Lab Med 2015; 53(11): 1737 1743 Beryl E. Jacobson, David W. Seccombe*, Alex Katayev and Adeera Levin A study examining the bias of albumin and albumin/creatinine ratio measurements in urine DOI
More informationQMS TACROLIMUS APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800
QMS TACROLIMUS APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800 Beckman Coulter Reagent REF A53727 The QMS Tacrolimus Immunoassay is intended for the quantitative determination of tacrolimus in human whole
More informationMultiple studies are available concerning the use of
SPECIAL ARTICLE Evaluation of the Biosite Ò Quantitative Whole Blood D-dimer Assay and Comparison With the biomérieux VIDAS Ò D-dimer Exclusion Test Validation and Utility For Use in the Central Laboratory
More informationBone Markers and Vascular Calcification in CKD-MBD
Bone Markers and Vascular Calcification in CKD-MBD Pierre Delanaye, MD, PhD Department of Nephrology, Dialysis, Transplantation CHU Sart Tilman University of Liège BELGIUM Bone Markers and Vascular Calcification
More informationMETHOD VALIDATION: WHY, HOW AND WHEN?
METHOD VALIDATION: WHY, HOW AND WHEN? Linear-Data Plotter a,b s y/x,r Regression Statistics mean SD,CV Distribution Statistics Comparison Plot Histogram Plot SD Calculator Bias SD Diff t Paired t-test
More informationCASE STUDY 3: PSA-STANDARDISATION & VARIABILITY COMPARATIVE STUDY IN HOSPITALS
CASE STUDY 3: PSA-STANDARDISATION & VARIABILITY COMPARATIVE STUDY IN HOSPITALS Dr Ophelia Blake FRCPath University Hospital Limerick, IRELAND JCTLM Meeting BIPM, Sevres,France 4 th December 2013 Prostate
More informationBringing quantitative HBsAg to the US provider, drug development and patient network
Bringing quantitative HBsAg to the US provider, drug development and patient network Presenter Robert G Gish MD Adjunct Professor, Stanford University Medical Director HBV Foundation HBV Forum 3 October
More informationSummary of Analytical Method for Quantitative Estimation of Fingolimod and Fingolimod Phosphate from Human Whole Blood Samples
Fingolimod Whole Blood Analysis Summary of Analytical Method for Quantitative Estimation of Fingolimod and Fingolimod Phosphate from Human Whole Blood Samples Study Detail: - Determination of Fingolimod
More informationBio-Rad Laboratories. Cardiac Assessment Controls. Value Assigned for Clinical Laboratory and Point of Care Test Systems
Bio-Rad Laboratories Cardiac Assessment Controls Cardiac Assessment Controls Value Assigned for Clinical Laboratory and Point of Care Test Systems Solving the Challenges of Troponin I Testing for the Central
More informationliquicolor (AMP Buffer, IFCC) Design Verification
Design Verification (AMP Buffer, IFCC) liquicolor 1 Introduction... 2 2 Imprecision... 2 3 arity and Detection Limit... 2 3.1 arity... 2 3.2 Detection Limit... 3 4 Comparison of Methods... 3 5 Stability...
More informationUnderstanding Quality Control for Infectious Disease Testing. Wayne Dimech ASLM Quality Control Workshop Abuja, Nigeria 9 th December 2018
Understanding Quality Control for Infectious Disease Testing Wayne Dimech ASLM Quality Control Workshop Abuja, Nigeria 9 th December 2018 Disclosure Attendance at ASLM part-sponsored by CDC No personal
More informationHIV Serology Quality Assessment Program Revised Summary for Panel HIVSER 2017Oct27
National Laboratory for HIV Reference Services National HIV and Retrovirology Laboratories National Microbiology Laboratory Public Health Agency of Canada HIV Serology Quality Assessment Program Revised
More information1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)
Comparison Study Summary Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC 29720 (803) 286-1480 February 10, 2016 1 PROTOCOL This evaluation was conducted on February 10, 2016 at Springs Memorial
More informationHIV Serology Quality Assessment Program Summary for Panel HIVSER 2017Apr19
National Laboratory for HIV Reference Services National HIV and Retrovirology Laboratories National Microbiology Laboratory Public Health Agency of Canada HIV Serology Quality Assessment Program Summary
More informationProduct Correction Urgent Immediate Action Required
Abbott Ireland Diagnostics Division Lisnamuck, Longford Co. Longford, Ireland DIAGNOSTICS Urgent Field Safety Notice Product Correction Urgent Immediate Action Required Date Issued October 12, 2018 Product
More informationAnalysis of 25-hydroxyvitamin D2 and 25-hydroxyvitamin D3 in Serum Using Ultra-performance Liquid Chromatography with Tandem Mass Spectrometry
Analysis of 25-hydroxyvitamin D2 and 25-hydroxyvitamin D3 in Serum Using Ultra-performance Liquid Chromatography with Tandem Mass Spectrometry Yong-Wha Lee Department of Laboratory Medicine and Genetics,
More informationHemostasis Test Validation, Performance and Reference Intervals
Hemostasis Test Validation, Performance and Reference Intervals Richard A. Marlar, Ph.D. Pathology and Laboratory Medicine Oklahoma City VA Medical Center University of Oklahoma Health Sciences Center
More informationGentian Canine CRP Immunoassay Application Note for scil VitroVet*
v02-jan2014 Gentian Canine CRP Immunoassay Application Note for scil VitroVet* Intended Use The canine CRP immunoassay on scil VitroVet is an in vitro diagnostic test for quantitative determination of
More informationDayana Argoti, Kerry Hassell, Sarah J. Fair, and Joseph Herman Thermo Fisher Scientific, Franklin, MA, USA
The Utilization of Novel Platform in a LC-MS/MS Workflow for the Analysis of Vitamin D, Testosterone, Immunosuppressants, Chemotherapeutics and Cortisol Dayana Argoti, Kerry Hassell, Sarah J. Fair, and
More informationEvaluation of the analytical performances of six measurands for thyroid functions of Mindray CL-2000i system
Original Article Page 1 of 7 Evaluation of the analytical performances of six measurands for thyroid functions of Mindray CL-2i system Andrea Padoan, Chiara Cosma, Mario Plebani Department of Laboratory
More informationComparison between two automated chemiluminescence immunoassays for quantifying 25 (OH) vitamin D
7 ORIGINALS / Rev Osteoporos Metab Miner. 216;8(2):7-74 Torrubia B 1, Alonso I 1, López-Ramiro E 1, Mahillo I 2, De la Piedra C 1 1 Laboratorio de Bioquímica 2 Servicio de Epidemiología Fundación Jiménez
More informationWhite Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard
White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard 7500-10032 Rev F Table of Contents Executive Summary 3 Changes introduced
More informationVitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid
Vitamin D Metabolite Analysis in Biological Samples Using Agilent Captiva EMR Lipid Application Note Clinical Research Authors Derick Lucas and Limian Zhao Agilent Technologies, Inc. Abstract Lipids from
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number: k100322 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE B. Purpose for Submission: Clearance of a new device C. Measurand: Whole
More informationChemical Metrology for Human Health Assessment
Chemical Metrology for Human Health Assessment Metrology and Physical Constants International School of Physics Enrico Fermi Stephen A. Wise Analytical Chemistry Division Material Measurement Laboratory
More information25 OH Vitamin D Rapid Test
INSTRUCTION FOR USE REF:GDB 7120-25T 25 OH Vitamin D Rapid Test A Rapid Sandwich Immunochromatographic Test for Quantitative Detection of total 25-OH Vitamin D in human finger-prick blood For In Vitro
More informationHuman Neurology 3-Plex A
Human Neurology 3-Plex A SUMMARY AND EXPLANATION OF THE TEST The Human N3PA assay is a digital immunoassay for the quantitative determination of total Tau, Aβ42, and Aβ40 in human plasma and CSF. Determination
More informationElecsys bone marker panel. Optimal patient management starts in the laboratory
bone marker panel Optimal patient management starts in the laboratory Complete solution for osteoporosis The most complete bone metabolism panel on a single platform bone marker assays are important diagnostic
More informationManual. Transfer step. Figure 1. Workflow comparison of manual phenylalanine assay and automated GSP Neonatal Phenylalanine assay
TECHNICAL NOTE GSP Neonatal Phenylalanine kit 3308-0010/3308-001B GSP Neonatal Phenylalanine kit (3308-0010/3308-001B) is a fully Introduction automated enzymatic assay for the GSP system. Optimized for
More informationAccuracy and clinical implications of seven 25-hydroxyvitamin D methods compared with liquid chromatography tandem mass spectrometry as a reference
Original Article Accuracy and clinical implications of seven 25-hydroxyvitamin D methods compared with liquid chromatography tandem mass spectrometry as a reference Heinz Jürgen Roth 1, Heinrich Schmidt-Gayk
More informationVitamin D and its metabolites: from now and beyond
In this issue: Laboratory investigation of vitamin D and bone metabolism markers Vitamin D and its metabolites: from now and beyond Etienne Cavalier 1, Jean-Claude Souberbielle 2 1 Department of Clinical
More informationPerformance Characteristics of a New Single-Sample Freezing Point Depression Osmometer
AACC, July 2007 Performance Characteristics of a New Single-Sample Freezing Point Depression Osmometer E. Garry, M. Pesta, N. Zampa Advanced Instruments, Inc., Norwood, MA 02062 AbstracT The Advanced Model
More informationCorrespondence should be addressed to Etienne Cavalier;
Hindawi Publishing Corporation International Journal of Endocrinology Volume 2013, Article ID 327265, 8 pages http://dx.doi.org/10.1155/2013/327265 Clinical Study A Randomised, Double-Blinded, Placebo-Controlled,
More informationHIV-1 Seroconversion Panel
OVERVIEW This Data Sheet contains test results specific for HIV-1. s are undiluted aliquots from plasma units collected from a single donor. No preservatives were added. CAUTION: Potentially infectious
More informationQuantitative Measurement of Cardiac Biomarkers
Cardiac Biomarker Analyzer Quantitative Measurement of Cardiac Biomarkers Troponin I, Myoglobin, CK-MB, D-Dimer, NTproBNP, hscrp Results in 17 minutes Direct measurement from whole blood samples 6 parallel
More informationExpression of HIV-1 Markers During Progression of Infection E E E Days Since 1st Bleed
Seroconversion OVERVIEW HIV-1 AccuVert TM Seroconversion PRB970 is a 4 member, 1.0 ml per vial panel of undiluted, naturally-occurring plasma samples. members represent serial bleeds collected from a single
More informationAldosterone on Liaison XL In use date: June 6, 2017
Aldosterone on Liaison XL In use date: June 6, 2017 I. PRINCIPLE Aldosterone is a steroid hormone of molecular weight 360.4 daltons, the major mineralcorticoid secreted by the adrenal cortex. The role
More informationEvaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100)
Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on CUBE analyser (CA0100) Location Location: Eurolyser Diagnostica GmbH Operators: Simone Wieser; Franz Helminger; Michael Gruber Date: July-November
More information1 PROTOCOL. Comparison Study Summary
Comparison Study Summary OnSite Health 210 S. Michigan Street, Suite 110 South Bend, IN 46601 574-243-5108 ext. 203 July 27, 2015 1 PROTOCOL This internal evaluation was conducted on July 17 & 20, 2015
More informationVDSP Standardisation Issues with 25-hydroxyvitamin D Christopher Sempos, Ph.D. Adjunct Professor, University of Wisconsin-Madison
VDSP Standardisation Issues with 25-hydroxyvitamin D Christopher Sempos, Ph.D. Adjunct Professor, University of Wisconsin-Madison For the Vitamin D Standardization Program (VDSP) Vitamin D: Analytical
More informationHIV-1 Seroconversion Panel PRB964
A SERACARE PANEL PRODUCT Seroconversion INTENDED USE The is a group of serial bleeds from an individual plasma donor during seroconversion. This is intended for use by diagnostics manufacturers and clinical
More informationValidation Report for the Neogen Fentanyl Kit for ELISA Screening of Whole Blood and Urine Specimens
Validation Report for the Neogen Fentanyl Kit for ELISA Screening of Whole Blood and Urine Specimens This document describes the validation of a Neogen Fentanyl kit for the semi-quantitative analysis of
More informationDr. Erin E. Chambers Waters Corporation. Presented by Dr. Diego Rodriguez Cabaleiro Waters Europe Waters Corporation 1
Development of an SPE-LC/MS/MS Assay for the Simultaneous Quantification of Amyloid Beta Peptides in Cerebrospinal Fluid in Support of Alzheimer s Research Dr. Erin E. Chambers Waters Corporation Presented
More informationFOR USE WITH SOFIA. For in vitro diagnostic use.
FOR USE WITH SOFIA For in vitro diagnostic use. INTENDED USE The Sofia Quantitative Vitamin D FIA is an immunofluorescence-based lateral flow assay designed for the quantitative determination of total
More informationHbA1c enzymatic assay evaluation at AN academic
Peer reviewed ORIGINAL ARTICLE HbA1c enzymatic assay evaluation at AN academic laboratory B Phiri-Ramongane Mmed Chem Path, FCP Path(SA)-Chem Department of Chemical Pathology, Sefako Makgatho Health Sciences
More informationIDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats
IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats Authors: Kate Cote, Ph.D., Graham Bilbrough, MA, VetMB, CertVA, MRCVS,
More informationRun 1 Run 2 TDP PLP TDP PLP. 3.0e5 2.8e5 2.2e5. 2.6e5 2.4e5. 2.2e5 2.0e5. Intensity, cps. 1.8e5 1.6e5 1.4e5 1.2e5 1.0e5 4.0e4. 2.0e e5 1.
APPLICATIONS A Simple and Effective High ph LC/MS/MS Method for Determination of Underivatized Vitamin B1 and B6 in Human Whole Blood Ramkumar Dhandapani Technical Manager Ramkumar loves to write poems,
More informationChapter 3.1: Diagnosis of CKD MBD: biochemical abnormalities Kidney International (2009) 76 (Suppl 113), S22 S49. doi: /ki.2009.
chapter 3.1 http://www.kidney-international.org & 2009 KDIGO Chapter 3.1: Diagnosis of CKD MBD: biochemical abnormalities. doi:10.1038/ki.2009.191 Grade for strength of recommendation a Strength Wording
More informationTroponin Assessment. Does it Carry Clinical Message? Stefan Blankenberg. University Heart Center Hamburg
Biomarkers for Optimal Management of Heart Failure Troponin Assessment Does it Carry Clinical Message? Stefan Blankenberg University Heart Center Hamburg Congress of the European Society of Cardiology
More information