A NEW APPROACH TO BEARING SURFACES FOR TOTAL HIP ARTHROPLASTY
|
|
- Hilda Lamb
- 5 years ago
- Views:
Transcription
1 A NEW APPROACH TO BEARING SURFACES FOR TOTAL HIP ARTHROPLASTY by Timothy McTighe, Exec. Dir., JISRF, Chagrin Falls, Ohio Ying Ko, Ph.D., Cincinnati, Ohio Russell B. Bennett, Ph.D., Cincinnati, Ohio James Adams, P.E., Cincinnati, Ohio A POSTER EXHIBIT AT THE 1994 AAOS MEETING NEW ORLEANS, LOUISIANA 1
2 INTRODUCTION The most common cause of proximal femoral bone loss is due to osteolysis. Although the specific cause of lysis in THA is not known, it has been attributed to a variety of factors, including motion of the implant and foreign body reaction to particulate debris, in particular to polymeric debris. It has been almost two decades since Willert first described the problem of polyethylene wear leading to Osteolysis periprosthetic inflammation, granuloma, bone resorption, and implant loosening. Since then, many studies have documented the finding of particulate bone cement and polyethylene in periprosthetic tissues. 3,4 In normal wearing artificial joints, linear wear rates of mm per year result in the generation of about mm 3(25 to 100 mg) of polyethylene debris annually. 5,1 On a basis of Excessive Wear known dimensions of polyethylene particles found in tissues around hip prostheses, this equated to the annual production of tens to hundreds of billions of particles. Variations of polyethylene wear rates probably relate to acetabular implant design, femoral head size, 8 femoral head material, and at least in part to the quality of the polyethylene used.. Wide variations are known to exist between i batches of polyethylene and between different polyethylene suppliers. Examples of Design Flaws Examples of Failures Examples of Poor Contact of Liner With Metal Cups Fiber Mesh Cup Constrained Socket Increased Wear of Poly Cups 2
3 REVIEW Based on favorable clinical trails in Europe during the past decade, improved ceramicon-ceramic and metal-on-metal bearing Ceramic Articulation Mittlemeier Design Metal Articulation Cup Design combinations have been renewed as possible solutions to the problem of polyethylene wear. Ideally, the bearing surfaces for most sliding, rotating, or articulating bearing surfaces systems will be made from material having relatively high strength, high wear, and corrosion resistance, a high resistance to creep, and low frictional moments. However, in reality no one material presently exhibits all of these characteristics. Ceramics have characteristics which are very desirable for use in sliding, rotating, and articulating bearing systems., In addition to high compressive strength, they exhibit high wear and corrosion resistance with relatively low frictional moments. However, use of such ceramic materials in bearing systems has been inhibited because such materials are susceptible to fracture due to their relatively low tensile and shear strengths. This weakness of ceramic materials is one reason Why metal and/or polymeric materials have been used for many bearing surfaces. Compared to bearing ceramics, bearing metals and polymers typically have lower wear and corrosion resistance or resistance to creep and higher frictional moments. In bearing -systems where ceramics have been used, their low tensile and shear strengths often force the adoption of costly design compromises. Thus, one design compromise has been to make the entire bearing component rather than just a portion thereof out of solid ceramic, thereby Surface Roughness: CoCr Surface Roughness: Ceramic Ceramic Metal Wear Rates Surface Wettability 3
4 increasing the amount of ceramic used and, therefore, effectively increasing the structural strength of the bearing surface.12 Such a solid ceramic bearing component can be larger and bulkier than its metal and/or polymeric counterpart. INTRINSIC SEGMENTED CERAMIC CUP DESIGN This paper will review one such concept of ceramic-on-ceramic articulation for use in total hip arthroplasty In an attempt to address these real life problems, a segmented ceramic bearing system has been developed. This segmented bearing system provides ceramic surfaces for mechanical bearings that would apply loads over a greater bearing surface area, resulting in i reduced bearing stresses and, in turn, reduce creep, wear, and likelihood of fracture of the bearing surfaces. Ceramic Truncated Cone Design Making an entire bearing component, like the acetabular cup, out of solid ceramic helps to compensate for the relatively poor tensile and shear strength typically found with ceramics. Also, because bearing ceramics are typically inflexible, additional manufacturing quality control of the geometry of both articular surfaces must be maintained in order to maximize the contact area between the two surfaces. If tight control is not maintained, point contacts may develop between the bearing surfaces. As the contact area between two bearing surfaces decreases, the stress that is transmitted between the surfaces increases. This can result in greater wear and increased possibility of fracture of one or both surfaces. In the past one solution to this problem has been to manufacture prosthesis with matching pairs of heads and cups. However, this solution is not only costly due to maintaining the quality levels required, but are additional inventory costs while making surgical intervention more difficult. The acetabular component is designed with several ceramic articular segments that are backed and held in a pre-determined pattern and configuration by either polyetheretherketone or polyethylene. Both of these materials, have a lower elastic modulus than the segmented ceramic material. In addition, the polymeric material is reduced in height so that only the segmented ceramic material articulates with a ceramic femoral head. lntrinisic Segmented Ceramic Cup Because of its resilience and lower elastic modulus, the polymeric material flexes as loads are transmitted between bearing surfaces while the shape of the surfaces of the segments remain relatively unchanged. This freedom of movement of the segments, under an applied load, allows for greater contact area between bearing surfaces because the segments as a group are able to conform to the geometry of the opposing bearing surface. Thus, rather than having highly localized stress concentrations 4
5 typically occurring in bearing systems, any applied load is shared by a number of segments which result in lower stress being applied to the bearing surfaces and each segment. An additional feature of this design is the formation of channels generated by locating the polymeric material slightly below the surface of the ceramic segments for lubrication and for allowing debris that finds Lubrication Channels its way into the bearing to either pass between the segments or be trapped in the polymeric material. Segmented Evolution This concept has evolved over the past five years from the ceramic in a hex shape imbedded in polysulfone to a current design that is circular imbedded in either polyetheretherketone or polyethylene. TESTING Post-fatigue testing (10 million cycles) has demonstrated no significant mechanical failures of the grout material (polysulfone) or of the ceramic bearing. SEM evaluations did demonstrate a small micro fracture within the grout and a polishing effect on the ceramic bearing surface. Hex Design Pre-Test Hex Design Post-Test 500X Pre-Test 500X Post-Test This test suggests that the bearing surface might benefit from pre-cycling to reduce initial ceramic debris. 5
6 Ongoing wear testing comparing different grout materials (peek and poly) on a P.M. state-of-the-art wear tester in conjunction with contact area and finite element analysis studies will help to determine the value of this design. To date we are optimistically encouraged by the preliminary work concerning this unique approach. However, only additional solid basic science results can justify in-vivo clinical evaluation. P.M. Wear Tester Cup Being Tested SUMMARY Intrinsic Segmental Ceramic Bearing Surface Hemispherical Design High Wear Resistance Low Friction High Compression Strength Greater Bearing Surface Area Self-Adjusting Design (Lower Surface Stress) Lubrication Channels Cost Effective Note: This device is not available for commercial use. 6
7 REFERENCES 1. McTighe, T.; Introduction Update News, JISRF, April, Willert, H.G. and Semlitsch, M.; Reactions of the Articular Capsule to Wear Products of Artificial joint Prostheses. J. Biomed. Mater. Res., 11:157, Howie, D.W.; Tissue Response in Relation to Type of Wear Particles Around Failed Hip Arthroplasties, J. Arthrop., 5:377, Schmalzied, T.P., Justy, M., Harris, W.H., Periprosthetic Bone Loss in THA. Polyethylene Wear Debris and the Concept of the Effective Joint Space. J. Bone Joint Surg. 74A:849, Wroblewski, B.M.; Direction and Rate of Socket Wear in Charnley Low-friction Arthroplasty. J. Bone Surg., 67B:757, Wright, T.M., and Rimnac, C.M.; Ultra-High-Molecular-Weight Polyethylene In; joint Replacement Arthroplasty (Morrey, B.F., ed). Churchill Livingstone, NTY, 1991, pp Tradonsky, S., Postak, P.D., Froimson, A.I., Greenwald, A.S.; Performance Characteristics of Two-Piece Acetabular Cups, Scientific Exhibit, AAOS, 1991, p Livermore, J., Ilstrup, D., Morrey, B.; Effects of Femoral Head Size on Wear of the Polyethylene Acetabular Component. J. Bone Joint Surg., 72A;518, Agins, H.J., Alcock, N.W., Bannsal, M. et al.; Metallic Wear in Failed Titanium-Alloy Total Hip Replacements: A Histological and Quantitative Analysis, J. Bone Joint Surg., 70A:347, Doerre, E., Beutler, H. and Geduldig, D.; Anforderungen un Oxederamische Werkstaffe als Biomaterial fur Kenstliche Gelenke, Arch. Orthop. Unfallchir, 1975, 83, p Geduldig, D., Doerre, E., Happel, M. and Lade, R.; Med. Orthop. Techn, 1975,95, p Mittelmeier, H.; Total Hip Replacement with the Autophor Cement-Free Ceramic Prosthesis in E. Morscher (Ed.), The Cementless Fixation of Hip Endoprosthesis, Springer-Veriag, Berlin, 1984, pp Graham, John W.; Internal Report on File, Orthopaedic Development Corporation, Reprint request to: T. McTighe, 8183 Stoneybrook Drive, Chagrin Falls, Ohio
Wear Properties of UHMWPE in CHARITE Artificial Inter-vertebral Disc
Mechanics of Contact and Lubrication, MTM G230 Department of Mechanical & Industrial Enineering Northeastern University Spring 2006 Wear Properties of UHMWPE in CHARITE Artificial Inter-vertebral Disc
More informationCement Polished Tapered Stems of 12/14 Taper. 96 mm 98 mm 104 mm 110 mm 116 mm 122 mm 128 mm. Ceramic Femoral Head. Outer Diameter
Design Philosophy Cementless Stems of 12/14 Taper Stem Length Neck shaft Angle STEM-N3 STEM-N4 STEM-N5 STEM-N6 STEM-N7 STEM-N8 STEM-N9 123 mm 125 mm 130 mm 135 mm 140 mm 145 mm 150 mm 132 Cement Polished
More informationExamination of Porous-Coated Patellar Components and Analysis of the Reasons for Their Retrieval
Examination of Porous-Coated Patellar Components and Analysis of the Reasons for Their Retrieval J. P. Collier,* M. B. Mayor,+ V. A. Surprenant,* H. P. Surprenant,* and R.E. Jensen* *Dartmouth Biomedical
More informationStability without compromise. Epsilon Durasul Constrained Insert
Stability without compromise Epsilon Durasul Constrained Insert History of Dislocation and Constrained Acetabular Inserts The Need for a Better Solution The Problem Dislocation is the second most common
More informationThe information contained in this document is intended for healthcare professionals only.
The information contained in this document is intended for healthcare professionals only. Trident Hemispherical Acetabular Shell Trident PSL HA Acetabular Shell Featuring: The Trident Acetabular System
More informationUHMWPE LONGEVITY: INFLUENCES OF THE ORTHOPAEDIC TRIAD
UHMWPE LONGEVITY: INFLUENCES OF THE ORTHOPAEDIC TRIAD AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS 66th Annual Meeting February 4-9, 1999 Anaheim, California COMMITTEE ON BIOMEDICAL ENGINEERING COMMITTEE ON
More informationRM Philosophy Innovation based on clinical evidence.
RM Philosophy Innovation based on clinical evidence RM Philosophy Innovation based on clinical evidence Design 1973 RM Classic Cup First uncoated RM Classic Cup out of Polyacetal was inserted by E. Morscher
More informationThe information contained in this document is intended for healthcare professionals only.
The information contained in this document is intended for healthcare professionals only. Trident Poly Acetabular System Trident PSL HA Acetabular Shell with X3 Poly insert Trident Hemispherical Acetabular
More informationENCYCLOPEDIC HANDBOOK OF BIOMATERIALS AND BIOENGINEERING
ENCYCLOPEDIC HANDBOOK OF BIOMATERIALS AND BIOENGINEERING PartB: Applications Volume I edited by Donald L. Wise Northeastern University Boston, Massachusetts Debra J. Trantolo Cambridge Scientific. Inc.
More informationCase report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis?
Case report: Pain L THR [ post THR 2 years; with history of trivial fall] Your Diagnosis? Diagnosis: Ceramic head fracture In the 1970 s, Boutin implemented ceramic in modern total hip arthroplasty (THA).
More informationRegenerex Porous Titanium Construct. Knees Hips Extremities Cement and Accessories PMI Trauma Technology
Regenerex Porous Titanium Construct Knees Hips Extremities Cement and Accessories PMI Trauma Technology One Surgeon. One Patient. Over 1 million times per year, Biomet helps one surgeon provide personalized
More informationESC. Enhanced Stability Liners. Design Rationale & Surgical Technique
ESC Enhanced Stability Liners Design Rationale & Surgical Technique Choice Without Compromise DePuy Synthes PINNACLE Hip Solutions are designed with a wide range of acetabular cup options, biological and
More informationThe Spatial Location of Impingement in Total Hip Arthroplasty
The Journal of Arthroplasty Vol. 15 No. 3 2000 The Spatial Location of Impingement in Total Hip Arthroplasty Motoi Yamaguchi, MD, PhD,* Toshihiro Akisue, MD, PhD,* Thomas W. Bauer, MD, PhD,* and Yasushi
More informationIndex. Gainslen prosthesis 125. Debonding 22-3 Degradation products 40-1 Diffusion theory of adhesive wear 137 Dual energy X-ray absorption 95-6, 98
Index Abrasion 40-1 mechanism of 137 Acetabular components 59-63 advantages 59-60 disadvantages 60-3 Harris Galante Porous cup 59, 60 Porous Coated Anatomic cup 60 Adhesins 155-6 Adhesive wear 137 Alloys
More informationInternational Journal of Engineering Research and Development e A Mathematical Model to Calculate Contact Stresses In Artificial Human Hip Joint
International Journal of Engineering Research and Development e-issn: 2278-067X, p-issn: 2278-800X, www.ijerd.com Volume 6, Issue 12 (May 2013), PP. 119-123 A Mathematical Model to Calculate Contact Stresses
More informationARTICLE IN PRESS. Surface Oxidized Zirconium Total Hip Arthroplasty Head Damage Due to Closed Reduction. Effects on Polyethylene Wear
The Journal of Arthroplasty Vol. 00 No. 0 2008 Surface Oxidized Zirconium Total Hip Arthroplasty Head Damage Due to Closed Reduction Effects on Polyethylene Wear William L. Jaffe, MD,* Eric J. Strauss,
More informationEnhancing stability and increasing range of motion. Metasul LDH Large Diameter Head
Enhancing stability and increasing range of motion Metasul LDH Large Diameter Head The Metasul large diameter head technology is the result of in-depth research, development and clinical experience that
More informationEnhanced Stability Constrained Liners. Design Rationale Surgical Technique
Enhanced Stability Constrained Liners Design Rationale Surgical Technique The Pinnacle Acetabular Cup System was designed to maximize the number of options available to the surgeon, and provide those options
More informationPyroCarbon Implants Durability + Biocompatibility
PyroCarbon Implants Durability + Biocompatibility What is PyroCarbon? PyroCarbon is a specific form of carbon that has been tailored for durability and biocompatibility. PyroCarbon exhibits favorable wear
More informationDePuy Orthopaedics, Inc. 700 Orthopaedic Drive Warsaw, IN USA Tel: +1 (800) Fax: +1 (574)
References 1. Sanchez-Sotelo J, Haidukewych GJ, Boberg CJ. Hospital Cost of Dislocation After Primary Total Hip Arthroplasty. J Bone Joint Surg. 2006;88A:290-294. 2. Brodner W, Grübl A, Jankovsky R, Meisinger
More information1.1 Ceramic on Ceramic Bearings Used with Proximal Modular Stems in THA K. J. Keggi, J. M. Keggi, R. E. Kennon and T. McTighe
1.1 Ceramic on Ceramic Bearings Used with Proximal Modular Stems in THA K. J. Keggi, J. M. Keggi, R. E. Kennon and T. McTighe Abstract Introduction: Osteolysis generated by wear debris remains a problem
More informationM-SERIES. Modular Femoral Stem AN ACCURATE MATCH EVERY TIME
M-SERIES Modular Femoral Stem AN ACCURATE MATCH EVERY TIME 54 Extension 114 Flexion High How do you ensure optimal post-operative joint stability? MECHANICAL RANGE OF MOTION Cup Position: 45 degrees abduction,
More informationZimmer Segmental System
Zimmer Segmental System Simple solutions for solving complex salvage cases A Step Forward The Zimmer Segmental System is designed to address patients with severe bone loss associated with disease, trauma
More informationSystematic Review of Clinical Outcome using Titanium Fiber Mesh Porous Ingrowth Surfaces
Systematic Review of Clinical Outcome using Titanium Fiber Mesh Porous Ingrowth Surfaces Timothy L. Tan, BS 1, Brenna C. Moeljadi, BS 2, Edward Ebramzadeh, PhD 3, Patricia Campbell, PhD 3, Sophia Nicole
More informationRadiographic Wear Measurements in a Cementless Metal-Backed Modular Cobalt-Chromium Acetabular Component
The Journal of Arthroplasty Vol. 16 No. 7 2001 Radiographic Wear Measurements in a Cementless Metal-Backed Modular Cobalt-Chromium Acetabular Component Robert L. Barrack, MD,* Carlos Lavernia, MD, Edward
More informationDesign Rationale and Clinical Review of a New Compliant Bearing Material for Acetabular Reconstruction
Joint Implant Surgery & Research Foundation Chagrin Falls, Ohio, USA Design Rationale and Clinical Review of a New Compliant Bearing Material for Acetabular Reconstruction Timothy McTighe, Dr. H.S. (hc)*;
More informationUnderstanding Hip Implant Options
Understanding Hip Implant Options Cup (Socket) Bearing (Liner) Head Stem (Femur) Modern Hip Implants (Ball) Metal Femoral Head on Marathon Cross-linked Polyethylene Liner Proven materials Polyethylene
More informationCAUTION: Ceramic liners are not approved for use in the United States.
Total Hip Prostheses, Self-Centering Hip Prostheses and Hemi-Hip Prostheses IMPORTANT: This essential product information sheet does not include all of the information necessary for selection and use of
More informationSURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS
SURGICAL TECHNIQUE CEMENTED & PRESS-FIT UNIFIED INSTRUMENTATION INTRAOPERATIVE FLEXIBILITY PROVEN BIOMECHANICS INTRODUCTION The Summit Tapered Hip System s comprehensive set of implants and instruments
More informationProduct Information. *smith&nephew POLARCUP Dual Mobility System
APPLYING SCIENCE TO MOBILITY Combined with VERILAST Technology for unmatched performance Product Information *smith&nephew POLARCUP Dual Mobility System APPLYING SCIENCE TO MOBILITY Combined with VERILAST
More informationOPTIMIZATION OF SKELETAL HIP IMPLANT CROSS-SECTIONS
OPTIMIZATION OF SKELETAL HIP IMPLANT CROSS-SECTIONS Sudesh Sivarasu, Pearline Beulah, Lazar Mathew School of Bio Sciences & Technology, VIT University, Vellore sudeshsivarasu@gmail.com Abstract: Aseptic
More information[ICESTM-2018] ISSN Impact Factor
GLOBAL JOURNAL OF ENGINEERING SCIENCE AND RESEARCHES MODELING AND FINITE ELEMENT ANALYSIS OF KNEE JOINT PROSTHESIS U.D.S.Prathap varma *1,S.Rajesh 2, B.Suresh Kumar 3 & P.Rama Murthy Raju 4 *1 M.TechScholar,
More informationReaching new heights. Comprehensive. Efficient. Simple.
Reaching new heights Comprehensive. Efficient. Simple. Various acetabular cup choices Compatible with the different head and liner options including VERILAST Technology Reach for proven OR efficient Instrumentation
More information2.79J/3.96J/BE.441/HST522J BIOMATERIALS FOR JOINT REPLACEMENT
Massachusetts Institute of Technology Harvard Medical School Brigham and Women s/massachusetts General Hosp. VA Boston Healthcare System 2.79J/3.96J/BE.441/HST522J BIOMATERIALS FOR JOINT REPLACEMENT M.
More informationPoor mid-term results of total hip arthroplasty with use of a Hylamer liner
Acta Orthop. Belg., 2008, 74, 337-342 ORIGINAL STUDY Poor mid-term results of total hip arthroplasty with use of a Hylamer liner Adrian SKWARA, Simone STRACKE, Carsten O. TIBESKU, Susanne FUCHS-WINKELMANN
More informationBone Bangalore
Dr Suresh Annamalai MBBS, MRCS(Edn), FRCS( Tr & Orth)(Edn), FEBOT(European Board), Young Hip and Knee Fellowship(Harrogate, UK) Consultant Arthroplasty and Arthroscopic Surgeon Manipal Hospital, Whitefield,
More informationVersatility in a proven design. Trilogy Acetabular System
Versatility in a proven design Trilogy Acetabular System Wide-ranging options. Clear-cut solutions. Offering a broad selection of shell and liner configurations, the Trilogy Acetabular System is designed
More informationArcos Modular Femoral Revision System
Arcos Modular Femoral Revision System Arcos System Simplify the Complex The Arcos Modular Femoral Revision System meets the demands of complex hip revision surgery by offering surgeons and OR staff the
More informationThe bi-metric XR series of Hip implants blends biomet s proven philosophy and tradition with Advanced technology. porous primary hip series
The bi-metric XR series of Hip implants blends biomet s proven philosophy and tradition with Advanced technology porous primary hip series BI-METRIC XR SERIES implant technology The Bi-Metric XR Hip Series
More informationOptimum implant geometry
Design Rationale Optimum implant geometry Extending the proven Tri-Lock heritage The original Tri-Lock was introduced in 1981. This implant was the first proximally coated tapered-wedge hip stem available
More informationBIOTRIBOLOGICAL AND MECHANICAL ANALYSIS OF TOTAL HIP ARTHROPLASTIY FAILURE: How to break the code
BIOTRIBOLOGICAL AND MECHANICAL ANALYSIS OF TOTAL HIP ARTHROPLASTIY FAILURE: How to break the code Patrick Chaboche, MD patrick.chaboche@wanadoo.fr Former Head of Department of Orthopaedic Surgery. Centre
More informationEarly failure of total hip replacements implanted at distant hospitals to reduce waiting lists
: 31 35 doi 10.1308/1478708051450 Audit Early failure of total hip replacements implanted at distant hospitals to reduce waiting lists Jac Ciampolini, Matthew JW Hubble Princess Elizabeth Orthopaedic Centre,
More informationCeramTec Medical Products. BIOLOX DUO Ceramic Bipolar System. Bone-sparing joint reconstruction with a maximum range of motion
CeramTec Medical Products BIOLOX DUO Ceramic Bipolar System Bone-sparing joint reconstruction with a maximum range of motion 2 BIOLOX DUO Ceramic Bipolar System Less is more BIOLOX DUO is an alternative
More informationTOTAL HIP REPLACEMENT:
THR Prosthesis Design TOTAL HIP REPLACEMENT: PROSTHESIS DESIGN FEATURES JESS JOHNSTON & MELINDA ZIETH History of Hip Prosthesis Joint Replacement Registry Implant Design Technology & Future History and
More informationKnee arthroplasty: What radiologists should know.
Knee arthroplasty: What radiologists should know. Poster No.: C-2174 Congress: ECR 2014 Type: Educational Exhibit Authors: J. Araújo 1, J. Pires 2, J. Oliveira 3, C. Macedo 1, S. Magalhães 1, I. Ferreira
More informationStability of Modular Acetabular Components Depends on the Lacking Mechanism Design, Temperature, and Axial Load
University of Tennessee, Knoxville Trace: Tennessee Research and Creative Exchange University of Tennessee Honors Thesis Projects University of Tennessee Honors Program Spring 5-1997 Stability of Modular
More informationHedrocel trabecular metal monoblock acetabular cups : mid-term results
Acta Orthop. Belg., 2006, 72, 326-331 ORIGINAL STUDY Hedrocel trabecular metal monoblock acetabular cups : mid-term results Michiel MULIER, Bart RYS, Lieven MOKE From the University Hospital Pellenberg,
More informationEffect of alumina femoral heads on polyethylene wear in cemented total hip arthroplasty
Effect of alumina femoral heads on polyethylene wear in cemented total hip arthroplasty OLD VERSUS CURRENT ALUMINA K. Tanaka, J. Tamura, K. Kawanabe, M. Shimizu, T. Nakamura From Kyoto University, Japan
More informationTotal Hip Arthroplasty Using Metal Head on a Highly Cross-linked Polyethylene Liner
ORIGINAL ARTICLE Hip Pelvis 27(4): 216-222, 2015 http://dx.doi.org/10.5371/hp.2015.27.4.216 Print ISSN 2287-3260 Online ISSN 2287-3279 Total Hip Arthroplasty Using Metal Head on a Highly Cross-linked Polyethylene
More informationTOTAL KNEE ARTHROPLASTY (TKA)
TOTAL KNEE ARTHROPLASTY (TKA) 1 Anatomy, Biomechanics, and Design 2 Femur Medial and lateral condyles Convex, asymmetric Medial larger than lateral 3 Tibia Tibial plateau Medial tibial condyle: concave
More informationTrident Ceramic Acetabular System Technical Monograph
Trident Ceramic Acetabular System Technical Monograph Introduction Clinical research and laboratory testing have demonstrated the benefits of ceramic-on-ceramic bearings in THA. Among these benefits are
More informationTotal Elbow Arthroplasty: an Update
Total Elbow Arthroplasty: an Update Emilie Cheung, M.D. Associate Professor Chief, Shoulder and Elbow Service Stanford University Department of Orthopedic Surgery Procedure volumes and rates increased
More informationSurface evaluation of orthopedic hip implants marketed in Brazil
Surface evaluation of orthopedic hip implants marketed in Brazil M.M.Souza 1, R.M. Trommer 2, M.M.Maru 2, C.R.M.Roesler 3, W.S.Barros 1, M.S. Dutra 3 1 Dimensional Metrology Laboratory, Inmetro. Brazil;
More informationHard Implants. Lecture #13
Hard Implants Lecture #13 Refs: 1. Park, J. B. and Bronzino, J. D., eds., Biomaterials: Principles and Applications. CRC Press, Boca Raton, FL, 2003. 2. Park, J. B., Biomaterials Science and Engineering.
More informationMedical Products Division. BIOLOX DUO Ceramic Bipolar System. Bone-sparing joint reconstruction with a maximum range of motion
Medical Products Division BIOLOX DUO Ceramic Bipolar System Bone-sparing joint reconstruction with a maximum range of motion 2 BIOLOX DUO Keramisches Bipolarsystem BIOLOX DUO Ceramic Bipolar System Less
More informationEffect of Femoral Head Size to Contact Stress at Pure Titanium Femoral Ball Head Outer Surface of Hip Joint Implant
Effect of Femoral Head Size to Contact Stress at Pure Titanium Femoral Ball Head Outer Surface of Hip Joint Implant Agung Setyo Darmawan, Waluyo Adi Siswanto, Haryanti Samekto Faculty of Mechanical and
More informationContinuing the Tradition. VerSys Heritage Hip System
Continuing the Tradition VerSys Heritage Hip System Heritage Following the Tradition The low-friction hip prosthesis developed by Sir John Charnley has more than a 20-year history of outstanding results.
More informationActive Articulation E1
Active Articulation E1 Dual Mobility Hip System Design Rationale Knees Hips Extremities Cement and Accessories PMI Trauma Technology Active Articulation E1 Dual Mobility Hip System Dual Mobility: Not a
More informationOptimum implant geometry
Surgical Technique Optimum implant geometry Extending proven Tri-Lock heritage The original Tri-Lock was introduced in 1981. This implant was the first proximally coated tapered-wedge hip stem available
More informationOptimum implant geometry
Design Rationale Optimum implant geometry Extending proven Tri-Lock heritage The original Tri-Lock was introduced in 1981. This implant was the first proximally coated tapered-wedge hip stem available
More informationG7 ACETABULAR SYSTEM. beautifully. efficient.
G7 ACETABULAR SYSTEM beautifully efficient. HERITAGE MEETS MODERN DESIGN SIMPLICITY, EFFICIENCY AND PERFORMANCE The G7 System is a multi-bearing acetabular platform delivering simplicity, efficiency and
More informationThe power of simplicity
The power of simplicity The power of simplicity Consider the facts, and take a look at the future: Revision TKA procedures are projected to double between now and 2015. 1 Restoration of a normal joint
More informationZimmer TM. Continuum Acetabular System. Bringing together proven technologies.
Zimmer TM Continuum Acetabular System Bringing together proven technologies. The power to meet individual patient needs. Zimmer Continuum Acetabular System provides highly flexible solutions for orthopaedic
More informationPrinciples of acetabular fixation in primary and revision hip arthroplasty Piotr Wojciechowski, Damian Kusz, Anna WAGNER
Principles of acetabular fixation in primary and revision hip arthroplasty Piotr Wojciechowski, Damian Kusz, Anna WAGNER Department of Orthopedics and Traumatology Medical University of Silesia Head of
More informationFIXATION FIRST WEAR REDUCTION OXIDATIVE STABILITY MECHANICAL INTEGRITY PROVEN TECHNOLOGY
FIXATION FIRST WEAR REDUCTION OXIDATIVE STABILITY MECHANICAL INTEGRITY PROVEN TECHNOLOGY Fixation First FIXATION FIRST 1958 Charnley Hip Over 1,000,000 implanted 1 96.2 percent survivorship at 32 years
More informationMARATHON Cross-Linked Polyethylene
MARATHON Cross-Linked Polyethylene Fixation First... 1 Wear Reduction... 2 Oxidative Stability... 3 Mechanical Integrity... 4 Proven Technology...5-6 References... 7 Fixation First FIXATION FIRST Prodigy
More informationDUAL MOBILITY CUP, CEMENTED OR CEMENTLESS
HIP Documentation DUAL MOBILITY CUP, CEMENTED OR CEMENTLESS >>DUAL MOBILITY ACETABULAR CUPS Dual mobility acetabular cup P r Gilles Bousquet invented the concept of dual mobility more than thirty years
More informationRetrieval Analysis of a Polycarbonate-Urethane Acetabular Cup: A Case Report
Journal of Long-Term Effects of Medical Implants, 18(1): 69 74 (2008) Retrieval Analysis of a Polycarbonate-Urethane Acetabular Cup: A Case Report W. Siebert, 1 S. Mai, 1 & S. Kurtz 2 1 Orthopädische Klinik
More informationEvolution. Medial-Pivot Knee System The Bi-Cruciate-Substituting Knee. Key Aspects
Evolution Medial-Pivot Knee System The Bi-Cruciate-Substituting Knee Key Aspects MicroPort s EVOLUTION Medial-Pivot Knee System was designed to recreate the natural anatomy that is lost during a total
More informationSession 3: Controversies for Bearing Surfaces in THR
: Controversies for Bearing Surfaces in THR Learning Objectives Upon completion of this activity, participants should be able to: 1. Understand the indications and known methods for the use of each of
More informationProgeny Hip Stem. Surgical Protocol and Product Specifications
Progeny Hip Stem Surgical Protocol and Product Specifications Progeny Hip Stem Introduction With emphasis on maximum stability and ease of use, the StelKast ProgenyTM Hip System provides the surgeon with
More informationCERAMIC-ON-METAL. References: Cat No: version 2
CERAMIC-ON-METAL References: 1. Beaule, P. E., et al. Jumbo femoral head for the treatment of recurrent dislocation following total hip replacement. J.Bone Joint Surg (2002). 84-A: 256-63. 2. Jolles, B.
More informationNational Joint Replacement Registry. Lay Summary 2015 Annual Report Hip and Knee Replacement
National Joint Replacement Registry Lay Summary 2015 Annual Report Hip and Knee Replacement SUPPLEMENTARY REPORT 2015 TABLE OF CONTENTS Introduction... 1 A brief history of the Registry origins... 1 The
More informationCLASSIFICATION OF MOBILE BEARING KNEE DESIGN: MOBILITY AND CONSTRAINT
CLASSIFICATION OF MOBILE BEAING KNEE DESIGN: MOBILITY AND CONSTAINT Orthopaedic esearch Laboratories Lutheran Hospital Cleveland Clinic Health System Christine S. Heim, B.Sc. Paul. D. Postak, B.Sc. Nicholas
More informationPolyethylene wear in total hip prostheses: the influence of direction of linear wear on volumetric wear determined from radiographic data
Skeletal Radiol (2003) 32:679 686 DOI 10.1007/s00256-003-0685-2 ARTICLE R. Košak V. Antolič V. Pavlovčič V. Kralj-Iglič I. Milošev G. Vidmar A. Iglič Polyethylene wear in total hip prostheses: the influence
More informationSurface Roughness of Femoral Head Prostheses After Dislocation
An Original Study Surface Roughness of Femoral Head Prostheses After Dislocation Kenny T. Mai, MD, Christopher Verioti, DO, Darryl D Lima, MD, PhD, Clifford W. Colwell, Jr., MD, and Kace A. Ezzet, MD Abstract
More informationHOSPITAL "A" (2007) ORTHOPEDIC JOINT IMPLANT COMPONENT MATRIX
HOSPITAL "A" (2007) ORTHOPEDIC JOINT IMPLANT COMPONENT MATRIX HIP IMPLANTS TOTAL HIPS Ultra High Demand (metal on metal) Femur, ingrowth Femoral head, metal Acetabular shell, ingrowth Acetabular liner,
More informationOrthopedic Surgery (Adult Hip Service) (Seoul National University Hospital)
Orthopedic Surgery (Adult Hip Service) (Seoul National University Hospital) 1. Introduction Total hip replacement arthroplasty (THRA) using a metal-on-polyethylene articulation has become one of the most
More informationThe Leader in Orthopaedic Innovation
The Leader in Orthopaedic Innovation Wright is a leading international manufacturer and distributor of superior, easy to use, and innovative orthopaedic implants and instrumentation. For over 50 years,
More informationCAUTION Federal law (USA) restricts this device to sale, by or on the order of a physician.
CAUTION Federal law (USA) restricts this device to sale, by or on the order of a physician. ENGLISH Mpact 3D Metal Implants and Augments 3D Metal INSTRUCTION FOR USE Important notice: the device(s) can
More information*smith&nephew SL-PLUS Cementless Femoral Hip System. Product Information
Product Information *smith&nephew SL-PLUS Cementless Femoral Hip System First Came the Philosophy to develop a universal hip system that could be used in almost every indication, immaterial to the patient
More informationFemoral Head Diameter, Neck And Taper Length Influence The Mechanical Load And Micromotion In The Taper Connection Of Total Hip Replacements
Femoral Head Diameter, Neck And Taper Length Influence The Mechanical Load And Micromotion In The Taper Connection Of Total Hip Replacements Daniel Kluess, Sarah Bierbaum, Wolfram Mittelmeier, Rainer Bader,
More informationDevelopment and Analysis of Composite Material for In-Situ Hip Prosthesis
Development and Analysis of Composite Material for In-Situ Hip Prosthesis R.Gowthaman P.G. Student, Manufacturing Engineering, Department of Mechanical Engineering, Thanthai Periyar Govt. Institute of
More informationDoes the self-centering mechanism of bipolar hip endoprosthesis really work in vivo?
Journal of Orthopaedic Surgery 2005;13(1):46-51 Does the self-centering mechanism of bipolar hip endoprosthesis really work in vivo? H Tsumura, N Kaku, T Torisu Department of Orthopedic Surgery, Oita University,
More informationFeaturing. Technology. Product Rationale
Featuring Technology Product Rationale 2 Optimum implant geometry Extending proven TRI-LOCK heritage The original TRI-LOCK Stem was introduced in 1981. This implant was the first proximally coated tapered-wedge
More informationTen-year in vivo wear measurement of a fully congruent mobile bearing unicompartmental knee arthroplasty
Ten-year in vivo wear measurement of a fully congruent mobile bearing unicompartmental knee arthroplasty A. J. Price, A. Short, C. Kellett, D. Beard, H. Gill, H. Pandit, C. A. F. Dodd, D. W. Murray From
More informationPeggers Super Summaries: THR and Revision Concepts
Total Hip replacement: ARTICUALTIONS Ceramic on polyethylene Ceramic on ceramic for young Metal on metal (resurfacing) Metal (cobalt chrome) on polyethylene WEAR Wear debris similar to bacteria o < 10
More informationBIOLOX delta Option Ceramic Femoral Head System. Product Features and Instructions for Use
BIOLOX delta Option Ceramic Femoral Head System Product Features and Instructions for Use Product Features BIOLOX delta ceramic heads are composed of 75% aluminum oxide and approximately 25% zirconia.
More informationR3think your options. For US distribution only.
R3think your options For US distribution only. R3think acetabular surgery It is widely accepted that total hip replacement surgery is one of the most successful procedures done today. Typically more thought
More informationSection of total knee replacement. Total Knee Replacement System. Knieendoprothesen System. Système de prothèse totale de genou
Section of total knee replacement Total Knee Replacement System Knieendoprothesen System Système de prothèse totale de genou Introduction: This knee system features great versality with its modular component
More informationThickness effects on maximum von-mises stress of a cement mantle in total hip replacement - a finite element study
Journal of Applied Biomaterials & Biomechanics 2009; Vol. 7 no. 2, pp. 111-115 Società Italiana Biomateriali Thickness effects on maximum von-mises stress of a cement mantle in total hip replacement -
More informationTrabecular Metal Technology The Best Thing Next to Bone
Trabecular Metal Technology The Best Thing Next to Bone Trabecular Metal Material resembles trabecular bone in its cellular structure and weight-bearing characteristics. By approximating the mechanical
More informationPainful Metal-on-Metal Total Hip Arthroplasty
The Journal of Arthroplasty Vol. 21 No. 2 2006 Painful Metal-on-Metal Total Hip Arthroplasty Cambize Shahrdar, MD,* Pat Campbell, PhD,y Joseph Mirra, MD,y and Lawrence D. Dorr, MD* Abstract: Two patients
More informationBIOLOX delta Option Ceramic Femoral Head System
BIOLOX delta Option Ceramic Femoral Head System Product Features and Instructions for Use Knees Hips Extremities Cement and Accessories PMI Technology Product Features BIOLOX delta ceramic heads are composed
More informationAGGRESSIVE GRANULOMATOUS LESIONS IN CEMENTLESS TOTAL HIP ARTHROPLASTY
AGGRESSIVE GRANULOMATOUS LESIONS IN CEMENTLESS TOTAL HIP ARTHROPLASTY SEPPO SANTAVIRTA, VEIJO HOIKKA, ANTTI ESKOLA, YRJO T. KONTTINEN, TIMO PAAVILAINEN, KAJ TALLROTH From the Orthopaedic Hospitalofthe
More information(c) B. Ravi, IIT Bombay 1
Collaborative Engineering Evaluation, Testing TKP Safety, TKP System Mechanism FEA Physical testing Fatigue Wear What How Evaluate Testing Judgement Criteria's Testing Conditions Parameters Measured OrthoCAD
More informationThe Role of Stem Modularity for THA in a Community Based Practice
Chagrin Falls, Ohio, USA The Role of Stem Modularity for THA in a Community Based Practice Louis Keppler, MD* and Timothy McTighe, Dr. HS (hc)** Abstract: Every few years there are concerns raised regarding
More informationCurrent Concepts of Using Large Femoral Heads in Total Hip Arthroplasty
REVIEW ARTICLE Hip Pelvis 28(3): 134-141, 2016 http://dx.doi.org/10.5371/hp.2016.28.3.134 Print ISSN 2287-3260 Online ISSN 2287-3279 Current Concepts of Using Large Femoral Heads in Total Hip Arthroplasty
More informationImpingement in THA. Georgi I. Wassilew. Orthopaedic Department, Centrum für Muskuloskeletale Chirurgie Univ.-Prof. Dr. med. C.
Georgi I. Wassilew Orthopaedic Department, Centrum für Muskuloskeletale Chirurgie Univ.-Prof. Dr. med. C. Perka Charité - Universitätsmedizin Berlin Orthoload club, Berlin Reasons for revision in THA Reason
More informationADDRESSING CLINICAL ISSUES OF CEMENTLESS HIP ARTHROPLASTY
E C H E L O N P R I M A R Y H I P S Y S T E M P R O D U C T R A T I O N A L E ADDRESSING CLINICAL ISSUES OF CEMENTLESS HIP ARTHROPLASTY Echelon Primary Total Hip System HIGH OFFSET STANDARD OFFSET Cementless
More information