RESPONSE OF 10 PULSE OXIMETERS TO AN IN VITRO TEST SYSTEM

Size: px
Start display at page:

Download "RESPONSE OF 10 PULSE OXIMETERS TO AN IN VITRO TEST SYSTEM"

Transcription

1 British Journal of Anaesthesia 1992; 68: RESPONSE OF 10 PULSE OXIMETERS TO AN IN VITRO TEST SYSTEM K. J. REYNOLDS, J. T. B. MOYLE, M. K. SYKES AND C. E. W. HAHN SUMMARY Pulse oximeters are often used in situations in which severe hypoxaemia may occur. We have developed an in vitro system to test the accuracy of pulse oximeter calibration. The probe of10 different oximeters was attached to a model finger in an in vitro blood circuit, and pulse oximeter readings (Sp Oi ) were compared with multi-wavelength in vitro oximeter readings (SoJ over a range of So 2 values from to 100%. The oximeters tested varied widely in their accuracy and linearity. We conclude that the system can test the accuracy, reproducibility and linearity of response of pulse oximeter readings at low oxyhaemoglobin saturations. KEY WORDS Equipment: pulse oximeters. Measurement techniques, pulse oximetry Pulse oximeters determine the oxygen saturation of arterial blood by a non-invasive optical technique. They are calibrated using data obtained by inducing hypoxaemia in healthy adult subjects. This is achieved easily at high saturations, but at saturations less than approximately 75 %, the acquisition of data becomes more difficult and possibly unethical. Data are limited, therefore, at these low saturations, and calibration is mainly by extrapolation. As calibration is performed individually by each manufacturer for his own instrument, the data used for calibration have been obtained from several subject population types and sizes. It is likely, therefore, that accuracy differs widely between different instruments, especially at low saturations. This has been confirmed in extensive in vivo studies by Severinghaus and colleagues [1,2]. We have developed an in vitro system for testing the performance of pulse oximeters over a wide range of saturations. This study determined the response of a range of different pulse oximeters to the in vitro test system. circuit by a computer-controlled peristaltic pump. The desired waveform and pulse rate of blood in the circuit are fed into the microcomputer, and appropriate control pulses are applied to a stepper motor driving the pumphead. Blood is pumped in pulsatile fashion through a cuvette acting as a model finger (fig. 2) and the probe of the pulse oximeter being tested is positioned across the finger. As it is impossible to construct a model finger reproducing exactly the anatomical features of a real finger, we aimed to keep the design of the cuvette as simple as possible (fig. 2). Blood enters from one end of the finger and flows in a thin Computer controlled pumphead Pulse oximeter Pulse oximeter probe FtG. 1. The pulse oximeter test system. Gas supply Membrane oxygenator METHODS Test system An in vitro test system was designed capable of producing signals similar to those obtained from pulsatile arterial blood in the finger (fig. 1). Whole blood obtained from adult subjects and anticoagulated with heparin is pumped through the K. J. REYNOLDS*, B.A., M.SC. ; J. T. B. MOYLE, M.B., B.S., I.ENG., M.INST.M.C, M.I.ELEC.I.E., F.C.ANAES.; M. K. SYKES, M.B., B.CHIR., F.C.ANAES., F.F.A.R.A.C.S. (HON.), F.F.A.(S.A-) (HON.); C. E. W. HAHN, M.A., B.SC, M.sc. D.PHiL.; Nuffield Department of Anaesthetics, John Raddiffe Hospital, Oxford. Accepted for Publication: November 2, Present address, for correspondence: c/o Physiology Department, University of Leicester, University Road, Leicester LEI 7RH.

2 366 BRITISH JOURNAL OF ANAESTHESIA Light from pulse oximeter probe TABLB I. Pulse oximeters tested, probe used, data points measured and number of units of blood used Pulse oximeter Type of probe No. data points Blood used (units) Black Clear Perspex Perspex Blood in Pulsating blood channels Silicon rubber membranes Datex Satlite (DA) Nellcor N-200 (NE) Ohmeda Biox 3700 (OH) Physiocontrol Lifestat 1600 (PH) Radiometer OX1 (RA) SiMed S-100 (SI) MiniOX IV (MI) Novametrix 515A (NO) Kontron 7840 (KO) Hewlett-Packard Monitor (HP) Blood out FIG. 2. The model finger used in the test system. Top: side view; below: face view. (1-mm) layer through the cuvette, over the finger tip end and back in a similar fashion along the underside. Both inlet and outlet are tapered to prevent flow separation. The silicone rubber membranes (0.5 mm thick) allow blood pulsations within the model finger. The a.c. :d.c. signal ratio may be altered by adjusting the gate clamps at points A and B in figure 1. This ratio should be in the physiological range (1:100 to 5:100 [3]). This design enables flow within the model finger to be developed fully within a short distance from each end of the cuvette, and the pulsations produced in the finger to be consistent along its length. This is necessary because the position of the light sources relative to the "fingertip " end of the oximeter probe varies between makes of oximeter, and so for a valid comparison to be made between different oximeters, the pulsations must be consistent at all positions along the cuvette. A Datex Satlite Versalite (ear) probe was positioned in turn at several places along the length of the cuvette, and the saturation value recorded by the pulse oximeter was found to be repeatable at all positions. The overall dimensions of the model finger were similar to those of a real finger, to ensure that all oximeter finger probes fitted snugly over the finger. To provide some scattering of the incident light, a piece of translucent draughtsman's paper was inserted between the light sources in the probe and the model finger. This paper, together with the silicon rubber membranes and the red blood cells, causes diffusion of the incident light. The saturation of blood flowing through the circuit is controlled by a small parallel plate membrane oxygenator. The blood flows between two parallel microporous polypropylene membranes, each with an effective area of 6.4 x 16.0 cm 2. Nitrogen, oxygen and carbon dioxide pass by diffusion across the membranes. The composition of the gas mixture passing through the membrane oxygenator is controlled by flowmeters. The percentage of carbon dioxide is maintained at 6 %, while the percentage of oxygen is varied; the balance of the gas mixture is nitrogen. Finally, the blood enters a small reservoir which damps the pulse wave completely before it returns to the pump. The blood in the reservoir is exposed to the same mixture of gases as that passing through the membrane oxygenator in order to maintain equilibrium. Experimental methods Ten pulse oximeters from different manufacturers (table I) were used to test the in vitro system. The probe of each pulse oximeter was placed over the model finger. Blood in the circuit was desaturated to approximately % by passing a mixture of 2 % oxygen, 6% carbon dioxide, balance nitrogen, through the membrane oxygenator. When equilibrium was reached and the pulse oximeter reading became stable, measurements displayed by the pulse oximeter (5p Ot ) were compared with blood samples withdrawn simultaneously from a sampling port immediately next to the outlet of the model finger. The oxygen saturation (So 2 ) of the blood samples was measured immediately using a co-oximeter (OSM3 Hemoximeter, Radiometer Ltd). The Hemoximeter uses six wavelengths of light to measure SOj directly, and is capable, therefore, of distinguishing the abnormal haemoglobins, carboxyhaemoglobin and methaemoglobin. The concentration of oxygen in the gas mixture was increased in 1 % steps, and 5p o, and So t readings recorded when equilibrium was reached at each saturation. A least-squares linear regression analysis was performed on the data from each pulse oximeter and the correlation coefficients calculated. As pulse oximeters are generally calibrated from empirical data for saturations greater than approximately 75 % and by extrapolation for values less than 75%, we decided also to analyse these two data ranges separately. Another useful method of analysis of the data is by calculation of mean bias, (Sp Oj So,,), and precision, which is the SD of the bias. Mean bias and precision were calculated for each oximeter over all three data ranges.

3 PULSE OXIMETER TEST SYSTEM 367 TABLE II. Linear regression analysis of Sp Oj on So,for the oximeters tested over the saturation ranges, and %. (See table I for details of oximeters.) CI = Confidence interval; SEE = standard error of estimate; n = number of readings Pulse oximeter and probe DA NE OH PH RA SI MI NO KO HP So, range (%) n Total haemoglobin concentration (thb) of the blood was adjusted to g %. Data were obtained also using die Datex Satlite for thb of 8 g % and 14 g%, to see if haemoglobin concentration affected the results. The SD of haemoximeter readings was estimated by analysing 10 samples from 1 ml of blood in a 1-ml syringe. The syringe was capped and placed in ice between readings to prevent the So 2 changing. The measurements were repeated on two further 1-ml samples of blood. Several in vivo measurements were made on hypoxaemic patients in an attempt to confirm the in vitro results. The same pulse oximeters and probes as used for the in vitro measurements were placed at random on the fingers of each patient, and Sp Oj values recorded. Simultaneously, an arterial blood sample was obtained from each patient and analysed in die OSM3 Hemoximeter to obtain So 2. RESULTS Table I shows the number of data points and number of different units of blood used for each pulse oximeter. In all blood samples, the carboxyhaemoglobin and methaemoglobin concentrations were < 1.5%. These small concentrations of abnormal haemoglobins should not affect the accuracy of the pulse oximeter readings. Linear regression, correlation coefficient, mean bias and precision are given in table II for the three saturation ranges %, % and % for each pulse oximeter. Slope (±95% CI) 0.80 (0.04) (0.09) 0.71 (0.12) 0.94 (0.09) 0.94 (0.26) 1.07(0.26) 0.84 (0.04) 0.88 (0.09) 0.82(0.13) (0.06) 0.82(0.16) 1.08(0.21) 0.59 (0.03) 0.72 (0.06) 0.47(0.11) 0. (0.04) 0.68 (0.08) 0.35(0.11) 0.59 (0.04) 0.57 (0.07) 0.62(0.13) 0.65 (0.04) 0.66 (0.08) 0.69(0.13) 0.58 (0.04) 0.76(0.10) 0.46(0.12) 0.49 (0.04) 0.69 (0.05) 0.32(0.10) X 2 (0 Q SEE n Intercept r Bias (mean (SD)) 2.7 (3.7) 1.0(2.0) 5.4 (4.2) 0.2(3.1) -0.5 (3.0) 1.4(2.9) 1.8(3.4) -0.2 (2.0) 3.7 (3.3) 3.5 (2.7) 2.9 (2.2) 4.2 (3.0) 7.6 (6.5) 2.3 (2.5) 13.1 (4.5) 10.2 (7.3) 4.2(3.1) 16.1 (5.0) 11.3(5.9) 6.6 (3.7) 16.0 (3.5) 9.7 (5.0) 5.6 (3.0) 13.7 (2.8) 6.0 (6.2) 0.9 (2.9) 10.7 (4.4) 9.5 (7.6) 3.1 (2.4) 15.8 (5.4) Hemoximeter SOj (%) FIG. 3. Comparison between Datex Satlite pulse oximeter readings (Sp Ot ) obtained from the test system, and Hemoximeter readings (So,). Linear regression lines and data points:, data set So, > 75 %; O, So, < 75 %; indicates identity. In vivo data points: A > 75 %, A < 75 %. The data points for three makes of oximeter are plotted in figures 3-5, to show the variation in response between different oximeters. In vivo results obtained from hypoxic patients are plotted on the same figures to allow comparison with the results obtained from the test system. On each figure, regression lines are drawn for test system data from

4 368 BRITISH JOURNAL OF ANAESTHESIA 100 -i o (D 1 70H Hemoximeter So, (%) FIG. 4. Comparison between SiMed S-100 pulse oximeter readings (5p Ot ) obtained rora the test system, and Hemoximeter readings (So,). (See figure 3 for explanation of symbols.) O > o z 100 -i Hemoximeter So, (%) FIG. 5. Comparison between Novametrix 515A pulse oximeter readings (5p o J obtained from the test system, and Hemoximeter readings (So,). (See figure 3 for explanation of symbols.) the two saturation ranges % (dashed line) and % (solid line). These two lines are useful to show the linearity of results between the two data ranges. All pulse oximeters tended to over-read; this effect was greater at low saturations (table II). The most extreme results were produced by the SiMed (fig. 4) and the HP Monitor, both of which produced readings of approximately 75 % for a true saturation of %. The correlation coefficient over the entire range measured ( %) was Ss 0.94 for all instruments. The correlation coefficient over the range % for nine of the 10 instruments was ^. Over this range, the Nellcor produced a correlation coefficient of This smaller value was attributable to the smaller number of points in the analysis, and one outlying point, rather than to any lesser reproducibility of the results. Correlation at lesser saturation was less good, with six pulse oximeters producing correlation coefficients of < 0.90 in the % range. In all cases (except the Nellcor), the residual variance (or scatter) was greater in the lesser saturation range. Each pulse oximeter was tested using at least five different units of blood on different days (table I). The results for each oximeter were consistent in repeated sets, which is a necessary requirement for a performance test system. No significant difference was found between results using the Datex Satlite from blood samples with haemoglobin concentrations of 8 g % and 14 g%. However, scatter of the results can be attributed partly to the SD of 5o 2 measurement, which was 0.4 %. DISCUSSION All the pulse oximeters showed a tendency to over-read at low saturations. This is confirmed by the results of the regression analysis, with regression coefficients for the % So i range varying between 0.49 for the HP Monitor and for the Physiocontrol, and positive values for the mean bias for all pulse oximeters over the range %. This tendency to over-read was noticed also by Sidi and colleagues using an Ohmeda Biox 3700 and a Nellcor N-100 on hypoxic dogs [4]. In contrast, in vivo measurements made by Severinghaus on healthy adults subjected to profound hypoxia, gave values for regression coefficients between 0.86 and 1.42, and negative biases on several occasions [2], It was not possible for us to obtain a complete set of in vivo results to confirm our in vitro results. However, a small number of readings were obtained from hypoxaemic patients with arterial saturations as little as 56 %. These in vivo points are plotted with the in vitro results for the three oximeters in figures 3 5. The results confirm our in vitro results, with all oximeters over-reading to some degree at low saturations. For example, a range of pulse oximeter readings from 69% (Ohmeda) to 84% (SiMed) (fig. 4) was obtained from a patient with an arterial saturation of 68.0 %. As oximeters are calibrated mainly by extrapolation of data at low saturations, calibration in this range is speculative. For this reason, the two ranges greater and less than 75 % were analysed separately. In order to assess the linearity of response of the pulse oximeters, regression coefficients of the entire saturation range (%), the "calibrated" range (%), and the "extrapolated" range ( %) were compared, with 95 % confidence intervals for the coefficients (table II). For six of the 10 pulse oximeters, all three coefficients coincided within 95 % confidence limits. Four pulse oximeters, however, (RA, SI, KO, HP) had a significant difference between the regression coefficient in the % So, range and that in the % range. This suggests non-linearity between the calibrated range and the uncalibrated range obtained by extrapol-

5 PULSE OXIMETER TEST SYSTEM 369 ation. An example of this non-linearity of results is shown in figure 4 for the SiMed pulse oximeter. In contrast, results presented in figures 3 and 5 for two other pulse oximeters showed no significant nonlinearity. The increase in scatter at low saturations is in agreement with in vivo observations [2, 5]. A possible explanation for this increasing scatter at low saturations may be as follows. The relatively large extinction coefficient of reduced haemoglobin at 660 nm implies that, as saturation decreases, less red light is transmitted through blood, and it appears darker. Any errors in the transmitted light signal at the red wavelength are therefore increased relative to the red signal at low saturations, which leads to a greater variance in 5p o, at these saturations. Indeed, we found that, if the saturation of the blood in the circuit was reduced to much less than %, in some cases the pulse oximeter would register a "noisy signal" warning, suggesting that the errors in the red signal were becoming large relative to the magnitude of the signal itself. The use of a system such as this may enable manufacturers to reassess and improve the calibration algorithms for their oximeters. However, the results presented in this paper are intended to be used as an assessment of the test system, and should not be extrapolated to a validation of the pulse oximeters themselves, as only one unit from each manufacturer was used. (Severinghaus [2], for example, found large variations between different sets of results using different instruments from the same manufacturer.) Also, die calibration algorithms are continually being updated by the manufacturers, and die oximeters used in our study are not necessarily the latest models and software versions. Kagle and colleagues demonstrated a significant difference between results obtained from an Ohmeda 3700 Version J software and diose from the same make of oximeter but containing modified Version XJ1 software [6]. These studies investigated die use of an in vitro system to compare pulse oximeter accuracy. The test system was used under laboratory conditions. Pulse oximeters are known to be affected by a large number of external factors, for example response time [1], accuracy at low perfusion [7, 8], interference by external electromagnetic sources [9, 10] and effect of motion [11]. The test system is potentially able to test the response of pulse oximeters to all diese artefacts. ACKNOWLEDGEMENTS We thank the Department of Health for generous support, and the manufacturers for providing the pulse oximeters. REFERENCES 1. Severinghaus JW, Naif eh KH. Accuracy of response of six pulse oximeters to profound hypoxia. Anesthesiology 1987; 67: Severinghaus JW, Naifeh KH, Koh SO. Errors in 14 pulse oximeters during profound hypoxia. Journal of Clinical Monitoring 1989; S: Huch A, Huch R, K6nig V, Neuman MR, Parker D, Yount J, Lubbers D. Limitations of pulse oximetry. Lancet 1988; 1: Sidi A, Rush W, Gravenstcin N, Ruiz B, Paulus DA, Davis RF. Pulse oximetry fails to accurately detect low levels of arterial hemoglobin oxygen saturation in dogs. Journal of Clinical Monitoring 1987; 3: Ridley SA. A comparison of two pulse oximeters. Assessment of accuracy at low arterial saturation in paediatric surgical patients. Anaesthesia 1988; 43: Kagle DM, Alexander CM, Berko RS, Giuffre M, Gross JB. Evaluation of the Ohmeda 3700 pulse oximeter: steady-state and transient response characteristics. Anesthesiology 1987; 66: Clayton DG, Webb RK, Ralston AC, Duthie D, Runciman WB. A comparison of the performance of 20 pulse oximeters under conditions of poor perfusion. Anaesthesia 1991; 46: Falconer RJ, Robinson BJ. Comparison of pulse oximeters: accuracy at low arterial pressure in volunteers. British Journal of Anaesthesia 1990; 65: Costarino AT, Davis DA, Keon TP. Falsely normal saturation reading with the pulse oximeter. Anesthesiology 1987; 67: Block FE. Interference in a pulse oximeter from a fiberoptic light source. Journal of Clinical Monitoring 1987; 3: Langton JA, Hanning CD. Effect of motion artefact on pulse oximeters: evaluation of four instruments and finger probes. British Journal of Anaesthesia 1990; 65:

Pulse Oximeter Accuracy Study

Pulse Oximeter Accuracy Study Patient Monitoring Technical Library SpO2 Monitoring Pulse Oximeter Accuracy Study 0 614-902709 Printed : 2008/12/24 This document is owned or controlled by Nihon Kohden and is protected by copyright law.

More information

Pulse Oximeter Accuracy Study

Pulse Oximeter Accuracy Study Patient Monitoring Technical Library SpO2 Monitoring Pulse Oximeter Accuracy Study 0 614-902709A Printed : 2014/5/26 This document is owned or controlled by Nihon Kohden and is protected by copyright law.

More information

COPD. Ebrahim Razi, Hossein Akbari

COPD. Ebrahim Razi, Hossein Akbari ORIGINAL ARTICLE COPD Turkish Respiratory Journal 2006; 7(2): 43-47 A Comparison of Arterial Oxygen Saturation Measured Both by Pulse Oximeter and Arterial Blood Gas Analyzer in Hypoxemic and Non-hypoxemic

More information

Pulse oximetry in the accident and emergency department

Pulse oximetry in the accident and emergency department Archives of Emergency Medicine, 1989, 6, 137-142 Pulse oximetry in the accident and emergency department C. J. HOLBURN & M. J. ALLEN Accident and Emergency Department, Leicester Royal Infirmary, Leicester,

More information

Permanent City Research Online URL:

Permanent City Research Online URL: Kyriacou, P. A., Pal, S. K., Langford, R. & Jones, DP (2006). Electro-optical techniques for the investigation of oesophageal photoplethysmographic signals and blood oxygen saturation in burns. Measurement

More information

Biomedical Instrumentation D. The Photoplethysmogram

Biomedical Instrumentation D. The Photoplethysmogram Biomedical Instrumentation D. The Photoplethysmogram Dr Gari Clifford Based on slides by Prof. Lionel Tarassenko The need for real-time oxygen saturation monitoring Respiratory failure & pulmonary disease

More information

A comparison of the performance of 20 pulse oximeters under conditions of poor perfusion

A comparison of the performance of 20 pulse oximeters under conditions of poor perfusion Anaesthesia, 1991, Volume 46, pages 3-10 A comparison of the performance of 20 pulse oximeters under conditions of poor perfusion D. G. CLAYTON, R. K. WEBB, A. C. RALSTON, D. DUTHIE AND W. B. RUNCIMAN

More information

Accuracy and precision of fourteen pulse oximeters

Accuracy and precision of fourteen pulse oximeters Eur Respir J 1991, 4, 1-119 TECHNICAL NOTE Accuracy and precision of fourteen pulse oximeters B. Hannhart*, J-P. Haberer**, C. Saunier*, M-C. Laxenaire** Accuracy and precision of fourteen pulse oxi~ters.

More information

Pulse oximetry use in small animals

Pulse oximetry use in small animals Vet Times The website for the veterinary profession https://www.vettimes.co.uk Pulse oximetry use in small animals Author : Clara Rigotti, Marieke de Vries Categories : Vets Date : October 10, 2011 Clara

More information

Dr. D.McG.Clarkson Determining the Accuracy of SpO2 Values

Dr. D.McG.Clarkson Determining the Accuracy of SpO2 Values Dr. D.McG.Clarkson Determining the Accuracy of SpO2 Values 1 Life saving technology The availability of the LED provided a convenient light which provides advantages of :- Compact size Modest power requirements

More information

Open Access. Abstract. Introduction. oximeter oxygen saturation (SpO 2. ) in intensive care patients have reported that the minimum SpO 2

Open Access. Abstract. Introduction. oximeter oxygen saturation (SpO 2. ) in intensive care patients have reported that the minimum SpO 2 Research Do changes in pulse oximeter oxygen saturation predict equivalent changes in arterial oxygen saturation? Gavin D Perkins 1, Daniel F McAuley 2, Simon Giles 3, Helen Routledge 4 and Fang Gao 5

More information

CORRELATION BETWEEN MEASUREMENT OF ARTERIAL SA TURA TION BY PULSE OXIMETRY AND BY HEMOXYMETER IN CHILDREN WITH CONGENITAL HEART DISEASE

CORRELATION BETWEEN MEASUREMENT OF ARTERIAL SA TURA TION BY PULSE OXIMETRY AND BY HEMOXYMETER IN CHILDREN WITH CONGENITAL HEART DISEASE 16 CORRELATION BETWEEN MEASUREMENT OF ARTERIAL SA TURA TION BY PULSE OXIMETRY AND BY HEMOXYMETER IN CHILDREN WITH CONGENITAL HEART DISEASE OMAR GALAL, MO, PhO; NEIL WILSON, MO Pulse oximetry is a noninvasive

More information

Evaluation of oesophageal reflectance pulse oximetry in major burns patients

Evaluation of oesophageal reflectance pulse oximetry in major burns patients Burns 31 (2005) 337 341 www.elsevier.com/locate/burns Evaluation of oesophageal reflectance pulse oximetry in major burns patients S.K. Pal a, *, P.A. Kyriacou b, S. Kumaran c, S. Fadheel c, R. Emamdee

More information

Dual channel photoplethysmography studies of cardio-vascular response to the body position changes

Dual channel photoplethysmography studies of cardio-vascular response to the body position changes Dual channel photoplethysmography studies of cardio-vascular response to the body position changes R. Erts, I. Kukulis, J. Spigulis, L. Kere University of Latvia, Raina Blvd. 9, Riga, LV-, Latvia Renars.erts@lu.lv

More information

Permanent City Research Online URL:

Permanent City Research Online URL: Pal, S. K., Kyriacou, P. A., Kumaran, S., Fadheel, S., Emamdee, R., Langford, R. M. & Jones, D. P. (2005). Evaluation of oesophageal reflectance pulse oximetry in major burns patients. Burns, 31(3), pp.

More information

Assessing Agreement Between Methods Of Clinical Measurement

Assessing Agreement Between Methods Of Clinical Measurement University of York Department of Health Sciences Measuring Health and Disease Assessing Agreement Between Methods Of Clinical Measurement Based on Bland JM, Altman DG. (1986). Statistical methods for assessing

More information

Noninvasive Detection of Bilirubin in Discrete Vessels

Noninvasive Detection of Bilirubin in Discrete Vessels , July 2-4, 2014, London, U.K. Noninvasive Detection of Bilirubin in Discrete Vessels Mark McEwen, Karen J. Reynolds Abstract Bilirubin, at the levels encountered in hyperbilirubinaemia, makes a similar

More information

ORIGINAL RESEARCH. Effect of road traffic accident contaminants on pulse oximetry among normoxaemic volunteers

ORIGINAL RESEARCH. Effect of road traffic accident contaminants on pulse oximetry among normoxaemic volunteers ISSN 1447-4999 ORIGINAL RESEARCH Effect of road traffic accident contaminants on pulse oximetry among normoxaemic volunteers Dr Gyorgyi Kamaras MD, FCEM 1,2, Dr Tamas Geller MD 1, Dr Csaba Dioszeghy MD

More information

The role of pulse oximetry in the accident and emergency department

The role of pulse oximetry in the accident and emergency department Archives of Emergency Medicine, 1989, 6, 211-215 The role of pulse oximetry in the accident and emergency department M. A. LAMBERT & J. CRINNION Accident and Emergency Department, Leeds General Infirmary,

More information

Oxygen Saturation Monitors & Pulse Oximetry. D. J. McMahon rev cewood

Oxygen Saturation Monitors & Pulse Oximetry. D. J. McMahon rev cewood Oxygen Saturation Monitors & Pulse Oximetry D. J. McMahon 141105 rev cewood 2017-11-14 2 3 Key Points Oxygen Saturation Monitors & Pulse Oximetry : Role of hemoglobin in respiration Basic principle of

More information

NITROUS OXIDE SEDATION CAUSES POST-HYPERVENTILATION APNOEAf

NITROUS OXIDE SEDATION CAUSES POST-HYPERVENTILATION APNOEAf British Journal of Anaesthesia 1991; 67: 7-12 NITROUS OXIDE SEDATION CAUSES POST-HYPERVENTILATION APNOEAf D. NORTHWOOD, D. J. SAPSFORD, J. G. JONES, D. GRIFFITHS AND C. WILKINS SUMMARY We have studied,

More information

City, University of London Institutional Repository

City, University of London Institutional Repository City Research Online City, University of London Institutional Repository Citation: Kyriacou, P. A., Shafqat, K. & Pal, S. K. (29). Pilot investigation of photoplethysmographic signals and blood oxygen

More information

NITROUS OXIDE ELIMINATION AND DIFFUSION HYPOXIA DURING NORMO- AND HYPOVENTILATION

NITROUS OXIDE ELIMINATION AND DIFFUSION HYPOXIA DURING NORMO- AND HYPOVENTILATION British Journal of Anaesthesia 1993; 71: 189-193 NITROUS OXIDE ELIMINATION AND DIFFUSION HYPOXIA DURING NORMO- AND HYPOVENTILATION S. EINARSSON, O. STENQVIST, A. BENGTSSON, E. HOULTZ AND J. P. BENGTSON

More information

Arterial blood oxygen saturation during blood pressure cuff-induced hypoperfusion

Arterial blood oxygen saturation during blood pressure cuff-induced hypoperfusion Journal of Physics: Conference Series Arterial blood oxygen saturation during blood pressure cuff-induced hypoperfusion To cite this article: P A Kyriacou et al 27 J. Phys.: Conf. Ser. 85 1226 Related

More information

End Tidal CO2 Not All Its Cracked Up To Be The Limitations of PETCO2 In Sedation Analgesia

End Tidal CO2 Not All Its Cracked Up To Be The Limitations of PETCO2 In Sedation Analgesia End Tidal CO2 Not All Its Cracked Up To Be The Limitations of PETCO2 In Sedation Analgesia Tidal Volume Noninvasive monitoring of ventilation and exhaled carbon dioxide of a patient End Tidal CO2 Produces

More information

The desaturation response time of finger pulse oximeters during mild hypothermia*

The desaturation response time of finger pulse oximeters during mild hypothermia* APPARATUS The desaturation response time of finger pulse oximeters during mild hypothermia* D. B. MacLeod, 1 L. I. Cortinez, 2 J. C. Keifer, 3 D. Cameron, 4 D. R. Wright, 1 W. D. White, 5 E. W. Moretti,

More information

WHEN DOES BLOOD HAEMOLYSE? A Temperature Study

WHEN DOES BLOOD HAEMOLYSE? A Temperature Study Br. J. Anaesth. (1974), 46, 742 WHEN DOES BLOOD HAEMOLYSE? A Temperature Study C. CHALMERS AND W. J. RUSSELL SUMMARY Incubation of blood in vitro for up to 1 hour at temperatures below 45 C C caused no

More information

Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan

Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan Oximeters Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan chanhl@mail.cgu.edu.tw Pulse oximeter Masimo pulse CO-oximeter http://www.masimo.com/produc ts/continuous/radical-7/

More information

JUGULAR BULB CANNULATION: DESCRIPTION OF A CANNULATION TECHNIQUE AND VALIDATION OF A NEW CONTINUOUS MONITOR

JUGULAR BULB CANNULATION: DESCRIPTION OF A CANNULATION TECHNIQUE AND VALIDATION OF A NEW CONTINUOUS MONITOR British Journal of Anaesthesia 1991; 67: 553-558 JUGULAR BULB CANNULATION: DESCRIPTION OF A CANNULATION TECHNIQUE AND VALIDATION OF A NEW CONTINUOUS MONITOR P. J. D. ANDREWS, N. M. DEARDEN AND J. D. MILLER

More information

British Journal of Anaesthesia 104 (6): (2010) doi: /bja/aeq092 Advance Access publication April 23, 2010

British Journal of Anaesthesia 104 (6): (2010) doi: /bja/aeq092 Advance Access publication April 23, 2010 RESPIRATION AND THE AIRWAY Detection of hypoventilation during deep sedation in patients undergoing ambulatory gynaecological hysteroscopy: a comparison between transcutaneous and nasal end-tidal carbon

More information

Is Hypoxia common despite oxygen treatment in the Acute Medical ward?

Is Hypoxia common despite oxygen treatment in the Acute Medical ward? Is Hypoxia common despite oxygen treatment in the Acute Medical ward? Dr. R Dhar, Dr. RN Harrison and Dr. PD Snashall University Hospital of North Tees, Stockton on Tees, TS198PE, United Kingdom (Received

More information

Trust Nonin Oximeters, Even For Your Sickest Patients

Trust Nonin Oximeters, Even For Your Sickest Patients Trust Nonin Oximeters, Even For Your Sickest Patients Trust Nonin Oximeters, Even For Your Sickest Patients An oximeter comparison study of healthy subjects in an independent hypoxia lab simulating the

More information

ExperimentalPhysiology

ExperimentalPhysiology Exp Physiol 97.5 (2012) pp 557 561 557 Editorial ExperimentalPhysiology Categorized or continuous? Strength of an association and linear regression Gordon B. Drummond 1 and Sarah L. Vowler 2 1 Department

More information

This work is licensed under a Creative Commons Attribution-Noncommercial-Share Alike 3.0 License.

This work is licensed under a Creative Commons Attribution-Noncommercial-Share Alike 3.0 License. University of Massachusetts Medical School escholarship@umms UMass Center for Clinical and Translational Science Research Retreat 2017 UMass Center for Clinical and Translational Science Research Retreat

More information

How Does Pulse Oximetry Work? SpO2 Sensors Absorption at the Sensor Site Oxyhemoglobin Dissociation Curve

How Does Pulse Oximetry Work? SpO2 Sensors Absorption at the Sensor Site Oxyhemoglobin Dissociation Curve SpO2 Monitoring Contents 1 Introduction 1 What is SpO 2? How Does Pulse Oximetry Work? SpO2 Sensors Absorption at the Sensor Site Oxyhemoglobin Dissociation Curve 5 How Do I Use SpO2? Choosing a Sensor

More information

Choosing Mindray SpO2

Choosing Mindray SpO2 Choosing Mindray SpO2 Aug. 2014 Choosing Mindray SpO2 The culmination of over 20 years of in-house SpO2 technology research. 4 advanced technologies make Mindray a leading SpO2 provider 5 dimensions of

More information

Design a system of measurement of heart rate, oxygen saturation in blood and body temperature with non-invasive method

Design a system of measurement of heart rate, oxygen saturation in blood and body temperature with non-invasive method IOP Conference Series: Materials Science and Engineering PAPER OPEN ACCESS Design a system of measurement of heart rate, oxygen saturation in blood and body temperature with non-invasive method To cite

More information

Assessing Global Patient Status with the Halo Index

Assessing Global Patient Status with the Halo Index Whitepaper Halo: Assessing Global Patient Status with the Halo Index Summary Physiologic deterioration often occurs long before a patient crisis and manifests through subtle and often undetected changes

More information

Western Hospital System. PSG in History. SENSORS in the field of SLEEP. PSG in History continued. Remember

Western Hospital System. PSG in History. SENSORS in the field of SLEEP. PSG in History continued. Remember SENSORS in the field of SLEEP Mrs. Gaye Cherry: Scientist in Charge Department of Sleep and Respiratory Medicine Sleep Disorders Unit Western Hospital PSG in History 1875: Discovery of brain-wave activity

More information

ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE

ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE Vet Times The website for the veterinary profession https://www.vettimes.co.uk ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE Author : CARL BRADBROOK Categories : Vets Date : October 7, 2013

More information

THE VENTILATORY RESPONSE TO HYPOXIA DURING EXERCISE IN CYANOTIC CONGENITAL HEART DISEASE

THE VENTILATORY RESPONSE TO HYPOXIA DURING EXERCISE IN CYANOTIC CONGENITAL HEART DISEASE Clinical Science and Molecular Medicine (1973) 45,99-5. THE VENTILATORY RESPONSE TO HYPOXIA DURING EXERCISE IN CYANOTIC CONGENITAL HEART DISEASE M. R. H. TAYLOR Department of Paediatrics, Institute of

More information

Investigation of the human oesophagus as a new monitoring site for blood oxygen saturation

Investigation of the human oesophagus as a new monitoring site for blood oxygen saturation INSTITUTE OF PHYSICS PUBLISHING PHYSIOLOGICAL MEASUREMENT Physiol. Meas. 22 (2001) 223 232 www.iop.org/journals/pm PII: S0967-3334(01)15804-1 Investigation of the human oesophagus as a new monitoring site

More information

Chandra Ramamoorthy MBBS; FRCA (UK) Professor of Anesthesiology, Stanford University. Director of Pediatric Cardiac Anesthesiology

Chandra Ramamoorthy MBBS; FRCA (UK) Professor of Anesthesiology, Stanford University. Director of Pediatric Cardiac Anesthesiology Should NIRS be Standard Care for Pediatric CPB Chandra Ramamoorthy MBBS; FRCA (UK) Professor of Anesthesiology, Stanford University Director of Pediatric Cardiac Anesthesiology Stanford Children s Hospital

More information

Intra-Arterial Blood Pressure Monitoring

Intra-Arterial Blood Pressure Monitoring Intra-Arterial Blood Pressure Monitoring Verghese Cherian Associate Professor of Anesthesiology Hershey Medical Center Intra-Arterial Blood Pressure (IABP) measurement has several advantages over a non-invasive

More information

Potential errors in pulse oximetry

Potential errors in pulse oximetry Anaesthesia, 199 1, Volume 46, pages 29 1-295 APPARATUS Potential errors in pulse oximetry 111: Effects of interference, dyes, dyshaemoglobins and other pigments* A. C. RALSTON, R. K. WEBB AND W. B. RUNCIMAN

More information

Research Journal of Pharmaceutical, Biological and Chemical Sciences

Research Journal of Pharmaceutical, Biological and Chemical Sciences Research Journal of Pharmaceutical, Biological and Chemical Sciences Oxygen Glucose and Heart rate Monitor: An non Invasive approach Anne Frank Joe A*, Veeramani Saikrupa, and Prajukti Panda Sathyabama

More information

Evaluation of Masimo signal extraction technology pulse oximetry in anaesthetized pregnant sheep

Evaluation of Masimo signal extraction technology pulse oximetry in anaesthetized pregnant sheep Veterinary Anaesthesia and Analgesia, 13, 40, 149 156 doi:10.1111/j.1467-2995.12.00772.x RESEARCH PAPER Evaluation of Masimo signal extraction technology pulse oximetry in anaesthetized pregnant sheep

More information

2 Modeling the ventilatory response to carbon dioxide in humans after bilateral and unilateral carotid body resection (CBR)

2 Modeling the ventilatory response to carbon dioxide in humans after bilateral and unilateral carotid body resection (CBR) 2 Modeling the ventilatory response to carbon dioxide in humans after bilateral and unilateral carotid body resection (CBR) IT IS AXIOMATIC that the respiratory chemoreceptors sense and respond to changes

More information

ALCO Regulations. Protocol pg. 47

ALCO Regulations. Protocol pg. 47 For the EMT-Basic Objectives Understand ALCO regulations relative to monitoring pulse oximetry by the EMT-B Review the signs and symptoms of respiratory compromise. Understand the importance of adequate

More information

Nonin Medical OEM Pulse Oximetry Solutions. Superior Performance for Challenging Patient Conditions

Nonin Medical OEM Pulse Oximetry Solutions. Superior Performance for Challenging Patient Conditions Nonin Medical OEM Pulse Oximetry Solutions Superior Performance for Challenging Patient Conditions COPD patients often present with complex co-morbidities When treating patients with Chronic Obstructive

More information

" Annex SpO 2 SPIROMETER D A T O S P I R MU2 *USER'S MANUAL* Rev.: 2.02

 Annex SpO 2 SPIROMETER D A T O S P I R MU2 *USER'S MANUAL* Rev.: 2.02 " Annex SpO 2 " SPIROMETER D A T O S P I R - 120 511-890-MU2 *USER'S MANUAL* 2012-11 Rev.: 2.02 SPIROMETER DATOSPIR-120 USER MANUAL "Annex SpO 2 " E. 2.02 1. INSTRUCTIONS FOR USE AND INSTALLATION 1.1.

More information

Clinical Evaluation of the Accuracy and Precision of the CDI 500 In-line Blood Gas Monitor With and Without Gas Calibration

Clinical Evaluation of the Accuracy and Precision of the CDI 500 In-line Blood Gas Monitor With and Without Gas Calibration The Journal of ExtraCorporeal Technology Original Articles Clinical Evaluation of the Accuracy and Precision of the CDI 500 In-line Blood Gas Monitor With and Without Gas Calibration Anne Louise Bellaiche,

More information

1- Cochlear Impedance Telemetry

1- Cochlear Impedance Telemetry INTRA-OPERATIVE COCHLEAR IMPLANT MEASURMENTS SAMIR ASAL M.D 1- Cochlear Impedance Telemetry 1 Cochlear implants used presently permit bi--directional communication between the inner and outer parts of

More information

Journal of Faculty of Engineering & Technology DESIGN AND IMPLEMENTATION OF A WEARABLE HEALTH DEVICE

Journal of Faculty of Engineering & Technology DESIGN AND IMPLEMENTATION OF A WEARABLE HEALTH DEVICE PAK BULLET TRAIN (PBT) JFET 22(2) (2015) 39-44 Journal of Faculty of Engineering & Technology journal homepage: www.pu.edu.pk/journals/index.php/jfet/index DESIGN AND IMPLEMENTATION OF A WEARABLE HEALTH

More information

FreeStyle Lite A Blood Glucose Meter That Requires No Coding

FreeStyle Lite A Blood Glucose Meter That Requires No Coding Journal of Diabetes Science and Technology Volume 2, Issue 4, July 2008 Diabetes Technology Society SYMPOSIUM Shridhara, Ph.D. Abstract Background: Abbott Diabetes Care introduced the FreeStyle Lite blood

More information

Evaluation of Pulse Rate and Arterial Oxygen Saturation (Sa0 ) Levels in Children During Routine Dental Procedures

Evaluation of Pulse Rate and Arterial Oxygen Saturation (Sa0 ) Levels in Children During Routine Dental Procedures Abstract ORIGINAL RESEARCH Evaluation of Pulse Rate and Arterial Oxygen Saturation (Sa0 ) Levels in Children During Routine Dental Procedures 1 1 1 Singla D, Anand A, Dhama P, Sharma A Aim: The aim of

More information

Signal Quality Analysis in Pulse Oximetry: Modelling and Detection of Motion Artifact

Signal Quality Analysis in Pulse Oximetry: Modelling and Detection of Motion Artifact Signal Quality Analysis in Pulse Oximetry: Modelling and Detection of Motion Artifact by Geoffrey Clarke, B.Eng. A thesis submitted to the Faculty of Graduate and Postdoctoral Affairs in partial fulfillment

More information

Zacurate CMS 500DL. Finger Pulse Oximeter. User Manual

Zacurate CMS 500DL. Finger Pulse Oximeter. User Manual Zacurate CMS 500DL Finger Pulse Oximeter Distributed by: Health Connector LLC 4910 Wright Road Suite 120 Stafford, TX, 77477 Phone: +1-832-939-8290 http://www.medshopandbeyond.com orders@medshopandbeyond.com

More information

Unit 1 Exploring and Understanding Data

Unit 1 Exploring and Understanding Data Unit 1 Exploring and Understanding Data Area Principle Bar Chart Boxplot Conditional Distribution Dotplot Empirical Rule Five Number Summary Frequency Distribution Frequency Polygon Histogram Interquartile

More information

Pulse oximetry revisited. Dr Liesel Bösenberg Specialist Physician and Fellow in Critical Care Kalafong Hospital University of Pretoria

Pulse oximetry revisited. Dr Liesel Bösenberg Specialist Physician and Fellow in Critical Care Kalafong Hospital University of Pretoria Pulse oximetry revisited Dr Liesel Bösenberg Specialist Physician and Fellow in Critical Care Kalafong Hospital University of Pretoria Topics that will be discussed and dissected: Revisiting physiology

More information

Knowledge and views of paediatricians about pulse oximetry: a nationwide online survey in Nigeria

Knowledge and views of paediatricians about pulse oximetry: a nationwide online survey in Nigeria Knowledge and views of paediatricians about pulse oximetry: a nationwide online survey in Nigeria A E Orimadegun, B O Ogunbosi, and F O Akinbami Abstract Pulse oximetry is a non-invasive method of measuring

More information

Autotitrator. Technical Information and Specifications. mpt-2. detailed specification sheets from

Autotitrator. Technical Information and Specifications. mpt-2. detailed specification sheets from Technical Information and Specifications detailed specification sheets from www.malvern.co.uk/autotitrator MPT-2 Autotitrator Description The MPT-2 autotitrator is an optional accessory compatible with

More information

Study Of Effects Of Varying Durations Of Pre-Oxygenation. J Khandrani, A Modak, B Pachpande, G Walsinge, A Ghosh

Study Of Effects Of Varying Durations Of Pre-Oxygenation. J Khandrani, A Modak, B Pachpande, G Walsinge, A Ghosh ISPUB.COM The Internet Journal of Anesthesiology Volume 20 Number 1 J Khandrani, A Modak, B Pachpande, G Walsinge, A Ghosh Citation J Khandrani, A Modak, B Pachpande, G Walsinge, A Ghosh.. The Internet

More information

Model 130 Pulse Oximeter

Model 130 Pulse Oximeter Model 130 Pulse Oximeter User s Manual MEDIAID INC. 17517 Fabrica Way Cerritos California 90703 USA Tel.: 714-367-2848 Fax: 714-367-2852 Web: www.mediaidinc.com Email: info@mediaidinc.com POX010-130 Contents

More information

Air or oxygen as driving gas for nebulised salbutamol

Air or oxygen as driving gas for nebulised salbutamol Archives of Disease in Childhood, 1988, 63, 9-94 Air or oxygen as driving gas for nebulised salbutamol J G A GLEESON, S GREEN, AND J F PRICE Paediatric Respiratory Laboratory, Department of Thoracic Medicine,

More information

International Journal of Health Sciences and Research ISSN:

International Journal of Health Sciences and Research   ISSN: International Journal of Health Sciences and Research www.ijhsr.org ISSN: 2249-9571 Review Article Pulse Oximetry - Working Principles in Pulpal Vitality Testing Jayashankar DN 1@, Zakirulla M 2, Sudhakar

More information

Testing Masimo rainbow Technology with vpad-a1 1. Background

Testing Masimo rainbow Technology with vpad-a1 1. Background 1 Testing Masimo rainbow Technology with vpad-a1 1. Background 1.1 Pulse Oximetry Pulse oximetery measures oxygen saturation or the ratio of oxyhemoglobin to the total amount of hemoglobin in the arterial

More information

Prediction of sleep-disordered breathing by unattended overnight oximetry

Prediction of sleep-disordered breathing by unattended overnight oximetry J. Sleep Res. (1999) 8, 51 55 Prediction of sleep-disordered breathing by unattended overnight oximetry L. G. OLSON, A. AMBROGETTI ands. G. GYULAY Discipline of Medicine, University of Newcastle and Sleep

More information

TruSignal SpO2 Sensors

TruSignal SpO2 Sensors Product Overview TruSignal SpO2 Sensors Photo Here Delivering exceptional oximetry performance Page 1 About GE TruSignal SpO2 sensors Page 2 About TruSignal SpO2 sensors Delivering exceptional oximetry

More information

Infinity MCable -Masimo SET

Infinity MCable -Masimo SET Infinity MCable -Masimo SET Bring the advantages of Masimo s Signal Extraction Technology (SET ) to your pulse oximetry monitoring The noninvasive, motion-tolerant Infinity MCable -Masimo SET works with

More information

ABSTRACTS. Section A: Nellcor Clinical Laboratory

ABSTRACTS. Section A: Nellcor Clinical Laboratory ABSTRACTS Section A: Nellcor Clinical Laboratory A1. Effects of Motion on Three Pulse Oximeters Designed for Use in Motion During Stable Normoxia and Hypoxia Michael W. Jopling, M.D, Paul D. Mannheimer,

More information

pulse oximetry sensors 2007

pulse oximetry sensors 2007 pulse oximetry sensors 2007 Instruments and sensors containing Masimo SET technology are identifi ed with the Masimo SET logo. Look for the Masimo SET designation on both the sensors and monitors to ensure

More information

Muscle OxyGen monitor. The Science Behind Moxy

Muscle OxyGen monitor. The Science Behind Moxy Muscle OxyGen monitor The Science Behind Moxy How Does Moxy Monitor Work? Moxy Monitor works by shining near-infrared light onto the skin and detecting some of the light after it has travelled into the

More information

Measuring Heart Rate and Blood Oxygen Levels for Portable and Wearable Devices

Measuring Heart Rate and Blood Oxygen Levels for Portable and Wearable Devices Measuring Heart Rate and Blood Oxygen Levels for Portable and Wearable Devices By Zhang Feng, Senior Medical Applications Engineer Marten Smith, Medical Marketing Manager Medical Products Group Microchip

More information

MedStar Health considers Continuous Home Pulse Oximetry medically necessary for the following indications:

MedStar Health considers Continuous Home Pulse Oximetry medically necessary for the following indications: MedStar Health, Inc. POLICY AND PROCEDURE MANUAL MP.006.MH Continuous Home Pulse Oximetry This policy applies to the following lines of business: MedStar Employee (Select) MedStar MA DSNP CSNP MedStar

More information

Posted: 11/27/2011 on Medscape; Published Br J Anaesth. 2011;107(2): Oxford University Press

Posted: 11/27/2011 on Medscape; Published Br J Anaesth. 2011;107(2): Oxford University Press Posted: 11/27/2011 on Medscape; Published Br J Anaesth. 2011;107(2):209-217. 2011 Oxford University Press Effect of Phenylephrine and Ephedrine Bolus Treatment on Cerebral Oxygenation in Anaesthetized

More information

REGRESSION MODELLING IN PREDICTING MILK PRODUCTION DEPENDING ON DAIRY BOVINE LIVESTOCK

REGRESSION MODELLING IN PREDICTING MILK PRODUCTION DEPENDING ON DAIRY BOVINE LIVESTOCK REGRESSION MODELLING IN PREDICTING MILK PRODUCTION DEPENDING ON DAIRY BOVINE LIVESTOCK Agatha POPESCU University of Agricultural Sciences and Veterinary Medicine Bucharest, 59 Marasti, District 1, 11464,

More information

Cardioplegia Circuit Products { CARDIOPLEGIA ADAPTERS}

Cardioplegia Circuit Products { CARDIOPLEGIA ADAPTERS} Cardioplegia Circuit Products { CARDIOPLEGIA ADAPTERS} Cardioplegia adapters are designed to permit customization of the cardioplegia circuit. Medtronic adapters can be used to connect multiple cardioplegia

More information

An assessment of the accuracy of pulse oximeters

An assessment of the accuracy of pulse oximeters Anaesthesia 212, 67, 396 41 Original Article doi:1.1111/j.1365-244.211.721.x An assessment of the accuracy of pulse oximeters Q. J. W. Milner 1 and G. R. Mathews 2 1 Consultant, Department of Anaesthesia,

More information

Correlation of Ratio of Ratio of Near Infrared Vs Infrared Photoplethysmograph Spectroscopy for the Non-Invasive Measurement of Haemoglobin

Correlation of Ratio of Ratio of Near Infrared Vs Infrared Photoplethysmograph Spectroscopy for the Non-Invasive Measurement of Haemoglobin Correlation of Ratio of Ratio of Near Infrared Vs Infrared Photoplethysmograph Spectroscopy for the Non-Invasive Measurement of Haemoglobin Ashita Diwate 1, Yogesh Kumar 2, Dr. Sanjeev Kumar 3, Prof. T.

More information

European Board of Anaesthesiology (EBA) recommendations for minimal monitoring during Anaesthesia and Recovery

European Board of Anaesthesiology (EBA) recommendations for minimal monitoring during Anaesthesia and Recovery European Board of Anaesthesiology (EBA) recommendations for minimal monitoring during Anaesthesia and Recovery INTRODUCTION The European Board of Anaesthesiology regards it as essential that certain core

More information

Non- invasive vascular testing. Pros and Cons of ABIs and Alternative Physiologic Assessments

Non- invasive vascular testing. Pros and Cons of ABIs and Alternative Physiologic Assessments Non- invasive vascular testing Pros and Cons of ABIs and Alternative Physiologic Assessments Non- Invasive Physiologic Arterial Studies Segmental Systolic Pressure Measurements ABIs, TBIs, and full segmentals

More information

There is convincing evidence in clinical studies

There is convincing evidence in clinical studies AJH 1998;11:1413 1417 Reliability of Reporting Self-Measured Blood Pressure Values by Hypertensive Patients Thomas Mengden, Rosa Maria Hernandez Medina, Belen Beltran, Elena Alvarez, Karin Kraft, and Hans

More information

ACTIVE TRANSPORT OF SALICYLATE BY RAT JEJUNUM

ACTIVE TRANSPORT OF SALICYLATE BY RAT JEJUNUM Quarterly Journal of Experimental Physiology (1981) 66, 91-98 91 Printed in Great Britain ACTIVE TRANSPORT OF SALICYLATE BY RAT JEJUNUM R. B. FISHER University Laboratory of Physiology, Oxford (RECEIVED

More information

Determination of Ethanol in Breath and Estimation of Blood Alcohol Concentration with Alcolmeter S-D2

Determination of Ethanol in Breath and Estimation of Blood Alcohol Concentration with Alcolmeter S-D2 Determination of Ethanol in Breath and Estimation of Blood Alcohol Concentration with Alcolmeter S-D2 A.W. Jones and KÄ. Jönsson Departments of Alcohol Toxicology and Internal Medicine, University Hospital,

More information

All motion is not created equal.

All motion is not created equal. Datex-Ohmeda 3800 and 3900 pulse oximeters TruTrak + technology designed to improve pulse oximetry performance during clinical patient motion. All motion is not created equal. Datex-Ohmeda 3800 and 3900

More information

Edwards FloTrac Sensor & Performance Assessments of the FloTrac Sensor and Vigileo Monitor

Edwards FloTrac Sensor & Performance Assessments of the FloTrac Sensor and Vigileo Monitor Edwards FloTrac Sensor & Edwards Vigileo Monitor Performance Assessments of the FloTrac Sensor and Vigileo Monitor 1 Topics System Configuration Performance and Validation Dr. William T. McGee, Validation

More information

Arterial Peaks in Regional Cerebral Blood Flow 133 Xenon Clearance Curves

Arterial Peaks in Regional Cerebral Blood Flow 133 Xenon Clearance Curves Arterial Peaks in Regional Cerebral Blood Flow 133 Xenon Clearance Curves BY STEPHEN G. ROSENBAUM, B.A., LINNETTE D. ILIFF, B.SC, M. PHIL., PH.D., J. W. D. BULL, M.D., F.R.C.P., F.F.R., G. H. DU BOULAY,

More information

HbA1c (Human) ELISA Kit

HbA1c (Human) ELISA Kit HbA1c (Human) ELISA Kit Cat. No.:DEIA3509 Pkg.Size:96T Intended use GHbA1c (Human) ELISA Kit is a sandwich enzyme immunoassay for the quantitative measurement of human GHbA1c. General Description vhemoglobin,

More information

escco, estimated Continuous Cardiac Output device DESEBBE OLIVIER SAUVEGARDE CLINIC, LYON, FRANCE

escco, estimated Continuous Cardiac Output device DESEBBE OLIVIER SAUVEGARDE CLINIC, LYON, FRANCE escco, estimated Continuous Cardiac Output device DESEBBE OLIVIER SAUVEGARDE CLINIC, LYON, FRANCE No conflict of interest u Technology, based on Pulse Wave Transit Time u Limits of the technology u Precision

More information

DC Unbalance Tolerance Testing in PSE s

DC Unbalance Tolerance Testing in PSE s The PoE Impairment Much like cable attenuation, alien crosstalk, and impedance mismatches present physical layer impairments to packet traffic on 10/100/1000Base-T links, Power-over-Ethernet (PoE) also

More information

Appendix E Choose the sign or symptom that best indicates severe respiratory distress.

Appendix E Choose the sign or symptom that best indicates severe respiratory distress. Appendix E-2 1. In Kansas EMT-B may monitor pulse oximetry: a. after they complete the EMT-B course b. when the service purchases the state approved pulse oximeters c. when the service director receives

More information

1.4 - Linear Regression and MS Excel

1.4 - Linear Regression and MS Excel 1.4 - Linear Regression and MS Excel Regression is an analytic technique for determining the relationship between a dependent variable and an independent variable. When the two variables have a linear

More information

Oxygen targeting in preterm infants using the Masimo SET Radical pulse oximeter

Oxygen targeting in preterm infants using the Masimo SET Radical pulse oximeter 1 Simpson Centre for Reproductive Health, Royal Infi rmary of Edinburgh, Edinburgh, UK 2 National Perinatal Epidemiology Unit, University of Oxford, Oxford, UK Correspondence to Dr B J Stenson, Simpson

More information

Chapter VI Development of ISFET model. simulation

Chapter VI Development of ISFET model. simulation Chapter VI Development of ISFET model using ZnO as gate and its simulation Introduction Theory of ISFET Experimental work Results and discussion Conclusion 6.1 General: The human bodies ph and electrolyte

More information

Design and Development of Non-Invasive Prototype to Measure Pulse Rate, Blood Glucose and Oxygen Saturation Level in Arterial Blood

Design and Development of Non-Invasive Prototype to Measure Pulse Rate, Blood Glucose and Oxygen Saturation Level in Arterial Blood Design and Development of Non-Invasive Prototype to Measure Pulse Rate, Blood Glucose and Oxygen Saturation Level in Arterial Blood Nazo Haroon Department of Mechatronics Engineering, CEME National University

More information

Check a Pulse! When to Question SpO 2, NIBP & EtCO 2 Readings

Check a Pulse! When to Question SpO 2, NIBP & EtCO 2 Readings Check a Pulse! When to Question SpO 2, NIBP & EtCO 2 Readings Mike McEvoy, PhD, RN, CCRN, NRP Professor Emeritus - Critical Care Medicine Albany Medical College Albany, New York Chair Resuscitation Committee

More information

Plasma Hemoglobin Determination: Two Manual Methods Compared. An Honors Thesis (HONRS 499) Anne M. Emerick. Sharlene Strahl. Beech Grove, Indiana

Plasma Hemoglobin Determination: Two Manual Methods Compared. An Honors Thesis (HONRS 499) Anne M. Emerick. Sharlene Strahl. Beech Grove, Indiana Plasma Hemoglobin Determination: Two Manual Methods Compared An Honors Thesis (HONRS 499) by Anne M. Emerick Sharlene Strahl St. Francis Hospital and Health Centers Beech Grove, Indiana Dr. Nancy Behforouz

More information

Edinburgh Research Explorer

Edinburgh Research Explorer Edinburgh Research Explorer Oxygen targeting in preterm infants using the Masimo SET Radical pulse oximeter Citation for published version: Johnston, ED, Boyle, B, Juszczak, E, King, A, Brocklehurst, P

More information

Estimation of Haemoglobin Using Non-Invasive Technique

Estimation of Haemoglobin Using Non-Invasive Technique Estimation of Haemoglobin Using Non-Invasive Technique Tatiparti Padma #1, Pinjala Jahnavi *2 #Department of ECE, GRIET, Hyderabad, Telangana *Department of ECE, GRIET, Hyderabad, Telangana 1tatipartipadma@gmail.com,

More information