Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan

Size: px
Start display at page:

Download "Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan"

Transcription

1 Oximeters Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan

2 Pulse oximeter Masimo pulse CO-oximeter ts/continuous/radical-7/ Oximeter

3 Oxygen transport in blood Hemoglobin Oxygen Hb HbO Deoxygen Hb Hb Hb+O HbO Oximeters 3

4 Arterial saturation of oxygen SaO SaO is affected by partial pressure of arterial oxygen, PO. At low partial pressures, such as tissue capillary pressure of 40 mmhg Millimeter of mercury Oxygen is released from hemoglobin, leading to a low SaO of about 75%. At high partial pressures, such as pulmonary capillary pressure of 95 mmhg Oxygen binds with hemoglobin, leading to high SaO of about 95 mmhg. Oximeters 4

5 Oxyhemoglobin dissociation curve Oximeters 5

6 Shifts of the oxyhemoglobin dissociation curve CO reacts with water to form carbonic acid HCO 3, which dissociates into hydrogen ions H + and bicarbonate ions HCO 3 When blood temperature increases, oxygen delivery to tissues is increased, making more oxygen available for binding to hemoglobin. BPG,3 biphosphoglycerate is a metabolically important phosphate compound. BPG concentration increases during hypoxic conditions of longer than a few hours. Oximeter 6

7 Carbon monoxide-hemoglobin dissociation curve A CO partial pressure of only 0.4 mmhg in alveoli causes CO to compete equally with oxygen for hemoglobin binding 1/50 of normal alveolar oxygen pressure of 100 mmhg. A CO partial pressure of only 0.6 mmhg can be lethal. Oximeter 7

8 Beer-Lambert s law Oximeters 8

9 Light interaction with absorption effect Beer-Lambert s law I x x I x di x ai x dx ax I x I x e 0 ai x x Absorbance of a wavelength of light absorption path length, l concentration, c absorbing species by ελ, molar absorptivity Oximeters 9

10 Oxygen saturation SaO, SpO measurement Relative amount of oxygen carried by the hemoglobin Two specific wavelengths : λ 1 : a red wavelength eg. 660 nm λ : a near infrared wavelength eg. 805 nm SaO C C HbO HbO C Hb Oximeters 10

11 Electromagnetic spectrum Oximeters 11

12 Light absorption signal produce AC output produce DC output Oximeters 1

13 SaO 0 10 d d C C IR Hb Hb IR HbO HbO IR I I 0 10 d d C C R Hb Hb R HbO HbO R I I Hb Hb IR HbO HbO IR Hb Hb R HbO HbO R DC IR IR DC R R C C C C I I I I IR R log log HbO R Hb R HbO IR Hb IR Hb R Hb IR Hb HbO HbO IR R IR R C C C SaO 13

14 Molar absorptivities of hemoglobin deoxyhemoglobin and oxyhemoglobin Oximeters 14

15 Total absorbance in tissue Scattering Variable absorbance not due to arterial blood Nonspecific sources of optical attenuation such as light scattering and geometric factors Oximeter 15

16 Pulse oximeter Takuo Aoyagi recognized that only pulsatile changes in light transmission through living tissues were necessary for estimating hemoglobin concentrations. Aoyagi s pulsatile observation is summarized in ratio calculations Zeros because each absorbance are constant The pulsatile path lengths for Hb and HbO are assumed to be equal Biopotential amplifiers 16

17 Arterial saturation of oxygen SaO Oximeter 17

18 Approximation of R in actual implementation AC refers to the pulsatile component of transmitted light DC refers to the nonpulsatile component of transmitted light Biopotential amplifiers 18

19 Calibration curve Scott Wilber made pulse oximetry accurate by abandoning the Beer-Lambert law R-vs.-SaO calibration curve was converted to a software lookup table for later real-time SpO estimation. Oximeter 19

20 Photodetector circuits /36086/does-this-photodiode-circuit-work

21 Wearable photoplethysmographic biosensors H.H. Asada et al, IEEE EMB, 003 Oximeter 1

22 Adaptive noise canceler Oximeter

23 Rapid desaturation-resaturation occurring during tapping motion Control A healthy volunteer inspired air from a face mask, in which the fraction of inspired oxygen was adjusted to decrease the control SpO values down from 100% to approximately 70% and up to 100% again. Masimo SET uses adaptive noise cancellation to minimize the motion effect Nellcor N 00, Nellcor N 3000 pulse oximeters Oximeter 3

24 Masimo s signal extraction technology SET adaptive noise cancellation Oximeter 4

25 Masimo s SET cont. Masimo s original signal extraction technology SET algorithm for adaptive noise cancellation. Ratios R corresponding to SpO values are input to the reference signal generator to output candidate noise sources. Each candidate noise source is input to the adaptive filter, with the filter output power then plotted A peak power corresponding to a lower SpO value may correspond to the ratio associated with saturation of venous blood rv. Oximeter 5

26 Masimo s SET cont. Because of two wavelengths, there are two input signals The reference noise source is defined as a function of the ratio Oximeter 6

27 Near-infrared spectroscopy NIRS by laser diode Uses the near-infrared region from 800 nm to 500 nm Applications in pharmaceutical and medical diagnostics including blood sugar and oximetry. OxiplexTS, ISS Inc, USA Oximeters 7

28 NIRS sensing by LEDs An sensor with central emitter and eight surrounding, detachable detectors Oximeters 8

29 Evoked cerebral function examination Electroencephalography EEG Near infrared NIR Magnetoencephalography MEG Functional MRI

30 Cerebral functional areas Oximeters 30

31 Neurovascular coupling Neurovascular coupling produces changes in blood flow and oxygen consumption Oximeters 31

32 Continuous-wave NIRS Ratio of reflected to incident light intensity, I/I 0, depends Absorption coefficient Changes of hemoglobin concentration [O Hb] [HHb] Path length L and differential path length factor B A I ln I OHb 0 u L [ O Hb] HHb [ HHb] B L Oximeters 3

33 Continuous-wave NIRS cont. Using two NIRS lights with different wavelengths for computing the change of hemoglobin concentrations ] [ ] [ Hb O HHb Hb O HHb HHb HHb Hb O HHb Hb O HHb Hb O Hb O u u HHb u u O Hb Oximeters 33

34 Application in brain oxygenation With 0% of oxygen consumption occurring in the human brain, any deficiency in oxygen supply may result in injury OxiplexTS, ISS Inc, USA Biomedical Optic 34

35 Cerebral Near-Infrared spectroscopy monitoring Monitoring of cerebral oxygenation Carotid stenosis or occlusion Brain ischemia Sleep apnea Ischemia of rat model Ischemia Cerebral response to carotid balloon inflation Obstructive sleep apnea

36 Related functional area for NIRS Prefrontal cortices PFC Premotor cortices PMC Supplement motor areas SMA Sensorimotor cortices SMC Oximeters 36

37 Brain oxygenation & reorganization in stroke patients Before rehabilitation oxyhb After rehabilitation Gait task TotalHb oxyhb deoxyhb Lesion Lesion

38 Application in peripheral vascular disease PVD PVD is a narrowing of the vessels carrying blood to the muscles in the legs and arms. Most patients report experiences of pain in the extremities due to inadequate blood flow and oxygen delivery to the exercising muscle. OxiplexTS, ISS Inc, USA Oximeters 38

39 Vessel occulsion test

40 NIRS demonstration Pulse Oximetry NEJM Muscular oxygenation under exercise Functional NIRS

41 Reference G. D. Baura, Medical device technologies, Elsevier Inc., 01. JG. Webster, Medical Instrumentation, application and design, 3rd, Houghton Mifflin, 000. JJ.Carr, JM.Brown, Introduction to Biomedical Equipment Technology, 4nd Edition, Prentice-Hall, 000.

Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan

Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan Oximeters Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan chanhl@mail.cgu.edu.tw Oxygen transport in blood Hemoglobin Oxygen O Deoxygen +O O Oximeters Arterial saturation

More information

Biomedical Instrumentation D. The Photoplethysmogram

Biomedical Instrumentation D. The Photoplethysmogram Biomedical Instrumentation D. The Photoplethysmogram Dr Gari Clifford Based on slides by Prof. Lionel Tarassenko The need for real-time oxygen saturation monitoring Respiratory failure & pulmonary disease

More information

Optical Sensor System for Hemoglobin Measurement

Optical Sensor System for Hemoglobin Measurement International Journal of Computational Engineering Research Vol, 03 Issue, 7 Optical Sensor System for Hemoglobin Measurement Rajashree Doshi 1, Anagha Panditrao 2 1 Department of Instrumentation and Control,

More information

Muscle OxyGen monitor. The Science Behind Moxy

Muscle OxyGen monitor. The Science Behind Moxy Muscle OxyGen monitor The Science Behind Moxy How Does Moxy Monitor Work? Moxy Monitor works by shining near-infrared light onto the skin and detecting some of the light after it has travelled into the

More information

Correlation of Ratio of Ratio of Near Infrared Vs Infrared Photoplethysmograph Spectroscopy for the Non-Invasive Measurement of Haemoglobin

Correlation of Ratio of Ratio of Near Infrared Vs Infrared Photoplethysmograph Spectroscopy for the Non-Invasive Measurement of Haemoglobin Correlation of Ratio of Ratio of Near Infrared Vs Infrared Photoplethysmograph Spectroscopy for the Non-Invasive Measurement of Haemoglobin Ashita Diwate 1, Yogesh Kumar 2, Dr. Sanjeev Kumar 3, Prof. T.

More information

How Does Pulse Oximetry Work? SpO2 Sensors Absorption at the Sensor Site Oxyhemoglobin Dissociation Curve

How Does Pulse Oximetry Work? SpO2 Sensors Absorption at the Sensor Site Oxyhemoglobin Dissociation Curve SpO2 Monitoring Contents 1 Introduction 1 What is SpO 2? How Does Pulse Oximetry Work? SpO2 Sensors Absorption at the Sensor Site Oxyhemoglobin Dissociation Curve 5 How Do I Use SpO2? Choosing a Sensor

More information

Chandra Ramamoorthy MBBS; FRCA (UK) Professor of Anesthesiology, Stanford University. Director of Pediatric Cardiac Anesthesiology

Chandra Ramamoorthy MBBS; FRCA (UK) Professor of Anesthesiology, Stanford University. Director of Pediatric Cardiac Anesthesiology Should NIRS be Standard Care for Pediatric CPB Chandra Ramamoorthy MBBS; FRCA (UK) Professor of Anesthesiology, Stanford University Director of Pediatric Cardiac Anesthesiology Stanford Children s Hospital

More information

Research Journal of Pharmaceutical, Biological and Chemical Sciences

Research Journal of Pharmaceutical, Biological and Chemical Sciences Research Journal of Pharmaceutical, Biological and Chemical Sciences Oxygen Glucose and Heart rate Monitor: An non Invasive approach Anne Frank Joe A*, Veeramani Saikrupa, and Prajukti Panda Sathyabama

More information

Oxygen Saturation Monitors & Pulse Oximetry. D. J. McMahon rev cewood

Oxygen Saturation Monitors & Pulse Oximetry. D. J. McMahon rev cewood Oxygen Saturation Monitors & Pulse Oximetry D. J. McMahon 141105 rev cewood 2017-11-14 2 3 Key Points Oxygen Saturation Monitors & Pulse Oximetry : Role of hemoglobin in respiration Basic principle of

More information

Measuring Heart Rate and Blood Oxygen Levels for Portable and Wearable Devices

Measuring Heart Rate and Blood Oxygen Levels for Portable and Wearable Devices Measuring Heart Rate and Blood Oxygen Levels for Portable and Wearable Devices By Zhang Feng, Senior Medical Applications Engineer Marten Smith, Medical Marketing Manager Medical Products Group Microchip

More information

Estimation of Haemoglobin Using Non-Invasive Technique

Estimation of Haemoglobin Using Non-Invasive Technique Estimation of Haemoglobin Using Non-Invasive Technique Tatiparti Padma #1, Pinjala Jahnavi *2 #Department of ECE, GRIET, Hyderabad, Telangana *Department of ECE, GRIET, Hyderabad, Telangana 1tatipartipadma@gmail.com,

More information

Gas Exchange in the Tissues

Gas Exchange in the Tissues Gas Exchange in the Tissues As the systemic arterial blood enters capillaries throughout the body, it is separated from the interstitial fluid by only the thin capillary wall, which is highly permeable

More information

ISSN: [Cortes* et al., 7(3): March, 2018] Impact Factor: 5.164

ISSN: [Cortes* et al., 7(3): March, 2018] Impact Factor: 5.164 IJESRT INTERNATIONAL JOURNAL OF ENGINEERING SCIENCES & RESEARCH TECHNOLOGY DESING OF A HYPOXIA DEVICE DETECTOR WITH WIRELESS ALERT Ing. Teo Cortes *1 & Dra. Fatima Moumtadi *2 *Department of Electronics,

More information

Design and Development of Non-Invasive Prototype to Measure Pulse Rate, Blood Glucose and Oxygen Saturation Level in Arterial Blood

Design and Development of Non-Invasive Prototype to Measure Pulse Rate, Blood Glucose and Oxygen Saturation Level in Arterial Blood Design and Development of Non-Invasive Prototype to Measure Pulse Rate, Blood Glucose and Oxygen Saturation Level in Arterial Blood Nazo Haroon Department of Mechatronics Engineering, CEME National University

More information

Development of a New Rehabilitation System Based on a Brain-Computer Interface Using Near-Infrared Spectroscopy

Development of a New Rehabilitation System Based on a Brain-Computer Interface Using Near-Infrared Spectroscopy Development of a New Rehabilitation System Based on a Brain-Computer Interface Using Near-Infrared Spectroscopy Takafumi Nagaoka, Kaoru Sakatani, Takayuki Awano, Noriaki Yokose, Tatsuya Hoshino, Yoshihiro

More information

Dr. D.McG.Clarkson Determining the Accuracy of SpO2 Values

Dr. D.McG.Clarkson Determining the Accuracy of SpO2 Values Dr. D.McG.Clarkson Determining the Accuracy of SpO2 Values 1 Life saving technology The availability of the LED provided a convenient light which provides advantages of :- Compact size Modest power requirements

More information

Dr. Puntarica Suwanprathes. Version 2007

Dr. Puntarica Suwanprathes. Version 2007 Dr. Puntarica Suwanprathes Version 2007 O 2 and CO 2 transport in blood Oxyhemoglobin dissociation curve O 2 consumption (VO 2 ) CO 2 production (VCO 2 ) O 2 capacity O 2 content: CaO 2 or CvO 2 %saturation

More information

Permanent City Research Online URL:

Permanent City Research Online URL: Kyriacou, P. A., Pal, S. K., Langford, R. & Jones, DP (2006). Electro-optical techniques for the investigation of oesophageal photoplethysmographic signals and blood oxygen saturation in burns. Measurement

More information

Journal of Faculty of Engineering & Technology DESIGN AND IMPLEMENTATION OF A WEARABLE HEALTH DEVICE

Journal of Faculty of Engineering & Technology DESIGN AND IMPLEMENTATION OF A WEARABLE HEALTH DEVICE PAK BULLET TRAIN (PBT) JFET 22(2) (2015) 39-44 Journal of Faculty of Engineering & Technology journal homepage: www.pu.edu.pk/journals/index.php/jfet/index DESIGN AND IMPLEMENTATION OF A WEARABLE HEALTH

More information

Electroencephalogram (EEG) Hsiao-Lung Chan Dept Electrical Engineering Chang Gung University

Electroencephalogram (EEG) Hsiao-Lung Chan Dept Electrical Engineering Chang Gung University Electroencephalogram (EEG) Hsiao-Lung Chan Dept Electrical Engineering Chang Gung University chanhl@mail.cgu.edu.tw Cerebral function examination Electroencephalography (EEG) Near infrared ray spectroscopy

More information

Arterial blood oxygen saturation during blood pressure cuff-induced hypoperfusion

Arterial blood oxygen saturation during blood pressure cuff-induced hypoperfusion Journal of Physics: Conference Series Arterial blood oxygen saturation during blood pressure cuff-induced hypoperfusion To cite this article: P A Kyriacou et al 27 J. Phys.: Conf. Ser. 85 1226 Related

More information

Clinically Available Optical Topography System

Clinically Available Optical Topography System Clinically Available Optical Topography System Clinically Available Optical Topography System 18 Fumio Kawaguchi Noriyoshi Ichikawa Noriyuki Fujiwara Yûichi Yamashita Shingo Kawasaki OVERVIEW: Progress

More information

JAMES R. KNIGHT

JAMES R. KNIGHT JAMES R. KNIGHT knightjr@ah.org James is currently the supervisor of clinical engineering at Sonora Regional Medical Center. He is the chairman of the CMIA Training & Education Committee. He also develops

More information

Design a system of measurement of heart rate, oxygen saturation in blood and body temperature with non-invasive method

Design a system of measurement of heart rate, oxygen saturation in blood and body temperature with non-invasive method IOP Conference Series: Materials Science and Engineering PAPER OPEN ACCESS Design a system of measurement of heart rate, oxygen saturation in blood and body temperature with non-invasive method To cite

More information

Monitoring Tumor Therapy

Monitoring Tumor Therapy Monitoring Tumor Therapy Zenalux Biomedical, Inc. The Zenascope was used to monitor the physiological response to a vascular disrupting agent in mouse tumors. Changes in hemoglobin concentration, hemoglobin

More information

Non-invasive Hemoglobin Measurement for Anemia Diagnosis

Non-invasive Hemoglobin Measurement for Anemia Diagnosis Non-invasive Hemoglobin Measurement for Anemia Diagnosis Raditya Artha Rochmanto 1, Hasballah Zakaria 2 Ratih Devi Alviana 3, Nurhalim Shahib 4 Department of Biomedical Engineering Medical Faculty Institut

More information

PBL SEMINAR. HEMOGLOBIN, O 2 -TRANSPORT and CYANOSIS An Overview

PBL SEMINAR. HEMOGLOBIN, O 2 -TRANSPORT and CYANOSIS An Overview 1 University of Papua New Guinea School of Medicine and Health Sciences Division of Basic Medical Sciences Discipline of Biochemistry and Molecular Biology PBL SEMINAR HEMOGLOBIN, O 2 -TRANSPORT and CYANOSIS

More information

ORIGINAL RESEARCH. Effect of road traffic accident contaminants on pulse oximetry among normoxaemic volunteers

ORIGINAL RESEARCH. Effect of road traffic accident contaminants on pulse oximetry among normoxaemic volunteers ISSN 1447-4999 ORIGINAL RESEARCH Effect of road traffic accident contaminants on pulse oximetry among normoxaemic volunteers Dr Gyorgyi Kamaras MD, FCEM 1,2, Dr Tamas Geller MD 1, Dr Csaba Dioszeghy MD

More information

Pulse oximetry use in small animals

Pulse oximetry use in small animals Vet Times The website for the veterinary profession https://www.vettimes.co.uk Pulse oximetry use in small animals Author : Clara Rigotti, Marieke de Vries Categories : Vets Date : October 10, 2011 Clara

More information

Consumer Near Infrared Spectroscopy as a Training and Recovery Tool. Jon Tarkington

Consumer Near Infrared Spectroscopy as a Training and Recovery Tool. Jon Tarkington Consumer Near Infrared Spectroscopy as a Training and Recovery Tool Jon Tarkington Outline - What is NIRS? - Devices - Data interpretation - Practical Application - Future Impacts - Suffix light Disclaimers

More information

ALCO Regulations. Protocol pg. 47

ALCO Regulations. Protocol pg. 47 For the EMT-Basic Objectives Understand ALCO regulations relative to monitoring pulse oximetry by the EMT-B Review the signs and symptoms of respiratory compromise. Understand the importance of adequate

More information

Pulse oximetry revisited. Dr Liesel Bösenberg Specialist Physician and Fellow in Critical Care Kalafong Hospital University of Pretoria

Pulse oximetry revisited. Dr Liesel Bösenberg Specialist Physician and Fellow in Critical Care Kalafong Hospital University of Pretoria Pulse oximetry revisited Dr Liesel Bösenberg Specialist Physician and Fellow in Critical Care Kalafong Hospital University of Pretoria Topics that will be discussed and dissected: Revisiting physiology

More information

CEREBRAL OXIMETRY IN INFANTS WITH HIE DAPHNA YASOVA BARBEAU, MD FN3 MEETING 2018

CEREBRAL OXIMETRY IN INFANTS WITH HIE DAPHNA YASOVA BARBEAU, MD FN3 MEETING 2018 CEREBRAL OXIMETRY IN INFANTS WITH HIE DAPHNA YASOVA BARBEAU, MD FN3 MEETING 2018 OBJECTIVES 1. Understand how cerebral oximetry works 2. Understand how cerebral oximetry may guide intervention 3. Learn

More information

Development of an Optical Sensor Based System for Detection and Monitoring of Hemoglobin in Blood

Development of an Optical Sensor Based System for Detection and Monitoring of Hemoglobin in Blood International Journal of Electronics Engineering Research. ISSN 0975-6450 Volume 9, Number 4 (2017) pp. 483-489 Research India Publications http://www.ripublication.com Development of an Optical Sensor

More information

Disclosures. Set up audience participation. Test question. Outline. Neuromonitoring What and When? IP for monitoring technology licensed to Medtronic

Disclosures. Set up audience participation. Test question. Outline. Neuromonitoring What and When? IP for monitoring technology licensed to Medtronic Neuromonitoring What and When? Disclosures IP for monitoring technology licensed to Medtronic Ken Brady, MD Pediatrics, Anesthesia, Critical Care Texas Children s Hospital Baylor College of Medicine Set

More information

Oxygen Level Measurement Techniques: Pulse Oximetry

Oxygen Level Measurement Techniques: Pulse Oximetry Oxygen Level Measurement Techniques: Pulse Oximetry Reem Abedalmoniam Ismail 1 and Sharief F. Babiker 2 1 Computer Engineering Dept., ElNeelain University, Khartoum, Sudan 2 Electrical and Electronics

More information

3. Which of the following would be inconsistent with respiratory alkalosis? A. ph = 7.57 B. PaCO = 30 mm Hg C. ph = 7.63 D.

3. Which of the following would be inconsistent with respiratory alkalosis? A. ph = 7.57 B. PaCO = 30 mm Hg C. ph = 7.63 D. Pilbeam: Mechanical Ventilation, 4 th Edition Test Bank Chapter 1: Oxygenation and Acid-Base Evaluation MULTIPLE CHOICE 1. The diffusion of carbon dioxide across the alveolar capillary membrane is. A.

More information

Signal Quality Analysis in Pulse Oximetry: Modelling and Detection of Motion Artifact

Signal Quality Analysis in Pulse Oximetry: Modelling and Detection of Motion Artifact Signal Quality Analysis in Pulse Oximetry: Modelling and Detection of Motion Artifact by Geoffrey Clarke, B.Eng. A thesis submitted to the Faculty of Graduate and Postdoctoral Affairs in partial fulfillment

More information

Appendix D An unresponsive patient with shallow, gasping breaths at a rate of six per minute requires:

Appendix D An unresponsive patient with shallow, gasping breaths at a rate of six per minute requires: Answer Key Appendix D-2 1. An unresponsive patient with shallow, gasping breaths at a rate of six per minute requires: a. oxygen given via nasal cannula b. immediate transport to a medical facility c.

More information

Appendix E Choose the sign or symptom that best indicates severe respiratory distress.

Appendix E Choose the sign or symptom that best indicates severe respiratory distress. Appendix E-2 1. In Kansas EMT-B may monitor pulse oximetry: a. after they complete the EMT-B course b. when the service purchases the state approved pulse oximeters c. when the service director receives

More information

Western Hospital System. PSG in History. SENSORS in the field of SLEEP. PSG in History continued. Remember

Western Hospital System. PSG in History. SENSORS in the field of SLEEP. PSG in History continued. Remember SENSORS in the field of SLEEP Mrs. Gaye Cherry: Scientist in Charge Department of Sleep and Respiratory Medicine Sleep Disorders Unit Western Hospital PSG in History 1875: Discovery of brain-wave activity

More information

BLOOD GLUCOSE LEVEL MONITORING BY NONINVASIVE METHOD USING NEAR INFRA RED SENSOR

BLOOD GLUCOSE LEVEL MONITORING BY NONINVASIVE METHOD USING NEAR INFRA RED SENSOR BLOOD GLUCOSE LEVEL MONITORING BY NONINVASIVE METHOD USING NEAR INFRA RED SENSOR P.Daarani 1 & A.Kavithamani 2 International Journal of Latest Trends in Engineering and Technology pp. 141-147 DOI: http://dx.doi.org/10.21172/1.ires.19

More information

International Journal of Research and Applications July - Sep 2017 Transactions 4(15):

International Journal of Research and Applications July - Sep 2017 Transactions 4(15): DOI: 10.17812/IJRA.4.15(98)2017 International Journal of Research and Applications July - Sep 2017 Transactions 4(15): 585-591 B I O - M E D I C A L R E V I E W R E P O R T eissn : 2349 0020 pissn : 2394

More information

Saturday Physics Honors presents. Professor Enrico Gratton, Looking Into the Brain with Lasers. (With some demonstrations)

Saturday Physics Honors presents. Professor Enrico Gratton, Looking Into the Brain with Lasers. (With some demonstrations) Saturday Physics Honors presents Professor Enrico Gratton, Looking Into the Brain with Lasers (With some demonstrations) Can we see through the body using light? What we expect to see? Startrek Movie clip

More information

Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis

Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis Journal of Physics: Conference Series PAPER OPEN ACCESS Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis To cite this article: Valerie Tan et al 2018 J. Phys.: Conf. Ser. 1049 012062

More information

Quantitative Performance Assessment Studies in Air Transportation

Quantitative Performance Assessment Studies in Air Transportation Quantitative Performance Assessment Studies in Air Transportation Portable Brain Activity Monitoring Technology Kurtulus Izzetoglu, PhD Drexel University Optical Brain Imaging Group Drexel University,

More information

Check a Pulse! When to Question SpO 2, NIBP & EtCO 2 Readings

Check a Pulse! When to Question SpO 2, NIBP & EtCO 2 Readings Check a Pulse! When to Question SpO 2, NIBP & EtCO 2 Readings Mike McEvoy, PhD, RN, CCRN, NRP Professor Emeritus - Critical Care Medicine Albany Medical College Albany, New York Chair Resuscitation Committee

More information

Is Hypoxia common despite oxygen treatment in the Acute Medical ward?

Is Hypoxia common despite oxygen treatment in the Acute Medical ward? Is Hypoxia common despite oxygen treatment in the Acute Medical ward? Dr. R Dhar, Dr. RN Harrison and Dr. PD Snashall University Hospital of North Tees, Stockton on Tees, TS198PE, United Kingdom (Received

More information

A Three-Wavelength Pulse Oximeter for Carboxyhemoglobin Determination

A Three-Wavelength Pulse Oximeter for Carboxyhemoglobin Determination Three-Wavelength Pulse Oximeter for Carboxyhemoglobin Determination R. P. Buinevicius 1 J. G. Webster 2 1 PhD Student, Department of Electrical and Computer Engineering University of Wisconsin 1415 Engineering

More information

Testing Masimo rainbow Technology with vpad-a1 1. Background

Testing Masimo rainbow Technology with vpad-a1 1. Background 1 Testing Masimo rainbow Technology with vpad-a1 1. Background 1.1 Pulse Oximetry Pulse oximetery measures oxygen saturation or the ratio of oxyhemoglobin to the total amount of hemoglobin in the arterial

More information

City, University of London Institutional Repository

City, University of London Institutional Repository City Research Online City, University of London Institutional Repository Citation: Kyriacou, P. A., Shafqat, K. & Pal, S. K. (29). Pilot investigation of photoplethysmographic signals and blood oxygen

More information

Electroencephalograms and Neuro-Rehabilitation. Hsiao-Lung Chan Dept Electrical Engineering Chang Gung University

Electroencephalograms and Neuro-Rehabilitation. Hsiao-Lung Chan Dept Electrical Engineering Chang Gung University Electroencephalograms and Neuro-Rehabilitation Hsiao-Lung Chan Dept Electrical Engineering Chang Gung University chanhl@mail.cgu.edu.tw Brain Cerebrum ( 大腦 ) Receives and processes conscious sensation

More information

Introduction. Device Description. Dual-Band Spectral Output

Introduction. Device Description. Dual-Band Spectral Output Optimum Spectrum and Pulse Shape for Vascular Lesion Treatment: The Science Behind MaxG James Childs, Ph.D.; Andrei Erofeev, Ph.D.; Mikhail Smirnov, Ph.D.; and Gregory Altshuler, Ph.D., Sc.D. Introduction

More information

Light is wearable Health Monitoring and Fitness Tracking Light is OSRAM

Light is wearable Health Monitoring and Fitness Tracking Light is OSRAM www.osram-os.com Light is wearable Health Monitoring and Fitness Tracking Light is OSRAM Sensor products Exactly what you need: high efficiency and great diversity OSRAM Opto Semiconductors offers a great

More information

Biomedical Instrumentation E. Blood Pressure

Biomedical Instrumentation E. Blood Pressure Biomedical Instrumentation E. Blood Pressure Dr Gari Clifford Adapted from slides by Prof. Lionel Tarassenko Blood pressure Blood is pumped around the body by the heart. It makes its way around the body

More information

Optical non-invasive methods for characterization of the human health status

Optical non-invasive methods for characterization of the human health status Optical non-invasive methods for characterization of the human health status Abstract Jens Kraitl and Hartmut Ewald Institute of General Electrical Engineering, University of Rostock, Rostock, Germany

More information

Respiratory System 1. A function of the structure labelled X is to

Respiratory System 1. A function of the structure labelled X is to 1 Respiratory System 1. A function of the structure labelled X is to A. produce sound. B. exchange gases. C. carry air into and out of the lung. D. stimulate the breathing centre in the brain. 2. Identify

More information

Hybrid EEG-HEG based Neurofeedback Device

Hybrid EEG-HEG based Neurofeedback Device APSIPA ASC 2011 Xi an Hybrid EEG-HEG based Neurofeedback Device Supassorn Rodrak *, Supatcha Namtong, and Yodchanan Wongsawat ** Department of Biomedical Engineering, Faculty of Engineering, Mahidol University,

More information

Pulse Oximeter Accuracy Study

Pulse Oximeter Accuracy Study Patient Monitoring Technical Library SpO2 Monitoring Pulse Oximeter Accuracy Study 0 614-902709 Printed : 2008/12/24 This document is owned or controlled by Nihon Kohden and is protected by copyright law.

More information

Near Infrared Spectroscopy in Neonatal Intensive Care

Near Infrared Spectroscopy in Neonatal Intensive Care Near Infrared Spectroscopy in Neonatal Intensive Care Adré J. du Plessis, MBChB Fetal-Neonatal Neurology, Children s Hospital, Boston, MA Children s Hospital and Harvard Medical School Disclosure Statement

More information

Lecture 10. Circulatory systems; flow dynamics, flow regulation in response to environmental and internal conditions.

Lecture 10. Circulatory systems; flow dynamics, flow regulation in response to environmental and internal conditions. Lecture 10 Circulatory systems; flow dynamics, flow regulation in response to environmental and internal conditions Professor Simchon Influence of P O2 on Hemoglobin Saturation Hemoglobin saturation plotted

More information

Dual channel photoplethysmography studies of cardio-vascular response to the body position changes

Dual channel photoplethysmography studies of cardio-vascular response to the body position changes Dual channel photoplethysmography studies of cardio-vascular response to the body position changes R. Erts, I. Kukulis, J. Spigulis, L. Kere University of Latvia, Raina Blvd. 9, Riga, LV-, Latvia Renars.erts@lu.lv

More information

Blood Pressure Estimation Using Photoplethysmography (PPG)

Blood Pressure Estimation Using Photoplethysmography (PPG) Blood Pressure Estimation Using Photoplethysmography (PPG) 1 Siddhi Sham Karande, BE E&TC, VIIT Pune. 2 Kiran Rajendrasingh Thakur, BE E&TC, VIIT Pune. 3 Sneha Dattatraya Waghmare, BE E&TC, VIIT Pune 4

More information

OPTICAL PERFUSION AND OXYGENATION CHARACTERIZATION IN A LIVER PHANTOM. A Thesis TRAVIS J. KING

OPTICAL PERFUSION AND OXYGENATION CHARACTERIZATION IN A LIVER PHANTOM. A Thesis TRAVIS J. KING OPTICAL PERFUSION AND OXYGENATION CHARACTERIZATION IN A LIVER PHANTOM A Thesis by TRAVIS J. KING Submitted to the Office of Graduate Studies of Texas A&M University in partial fulfillment of the requirements

More information

GLUCOSE CONCENTRATION MEASUREMENT USING FIBRE OPTIC SENSOR

GLUCOSE CONCENTRATION MEASUREMENT USING FIBRE OPTIC SENSOR Chapter 4 GLUCOSE CONCENTRATION MEASUREMENT USING FIBRE OPTIC SENSOR Abstract Design and development ofan evanescent wave fibre optic sensor for the determination ofglucose concentration is described in

More information

Noninvasive Detection of Bilirubin in Discrete Vessels

Noninvasive Detection of Bilirubin in Discrete Vessels , July 2-4, 2014, London, U.K. Noninvasive Detection of Bilirubin in Discrete Vessels Mark McEwen, Karen J. Reynolds Abstract Bilirubin, at the levels encountered in hyperbilirubinaemia, makes a similar

More information

Smart Probe for Uric Acid Detection

Smart Probe for Uric Acid Detection Smart Probe for Uric Acid Detection Rahul Pimpare 1, Samyak Ramteke 2, Pauroosh Kaushal 3 1 Student,Department of instrumentation & Control Engineering, D Y Patil College of Engineering Akurdi, Pune,India.

More information

Question Expected Answers Marks Additional Guidance 1 (a) C ; E ; A ; B ; 4. PhysicsAndMathsTutor.com

Question Expected Answers Marks Additional Guidance 1 (a) C ; E ; A ; B ; 4. PhysicsAndMathsTutor.com Question Expected Answers Marks Additional Guidance 1 (a) C ; E ; A ; B ; 4 Question Expected Answers Marks Additional Guidance (b) (i) P wave combined with larger peak before QRS complex ; Note: - look

More information

Use of the Free Electron Laser for the Noninvasive Determination of Retinal Oxyhemoglobin Saturation by Near Infrared Reflectance Spectrophotometry

Use of the Free Electron Laser for the Noninvasive Determination of Retinal Oxyhemoglobin Saturation by Near Infrared Reflectance Spectrophotometry Use of the Free Electron Laser for the Noninvasive Determination of Retinal Oxyhemoglobin Saturation by Near Infrared Reflectance Spectrophotometry Ref: Eye, M.C. Escher, 1946 Ref: Eye, M.C. Escher, 1946

More information

NEAR INFRARED TRANSMISSION SPECTROSCOPY AS APPLIED TO FATS AND OIL

NEAR INFRARED TRANSMISSION SPECTROSCOPY AS APPLIED TO FATS AND OIL NEAR INFRARED TRANSMISSION SPECTROSCOPY AS APPLIED TO FATS AND OIL Phillip J. Clancy, NIR Technology Systems, 56 Kitchener Pde, Bankstown, NSW, Australia. Near Infrared Transmission (NIT) Spectroscopy

More information

Pulse Oximeter Accuracy Study

Pulse Oximeter Accuracy Study Patient Monitoring Technical Library SpO2 Monitoring Pulse Oximeter Accuracy Study 0 614-902709A Printed : 2014/5/26 This document is owned or controlled by Nihon Kohden and is protected by copyright law.

More information

Blood Gases, ph, Acid- Base Balance

Blood Gases, ph, Acid- Base Balance Blood Gases, ph, Acid- Base Balance Blood Gases Acid-Base Physiology Clinical Acid-Base Disturbances Blood Gases Respiratory Gas Exchange Chemical Control of Respiration Dyshemoglobins Oxygen Transport

More information

Current bedside monitors of brain blood flow and oxygen delivery

Current bedside monitors of brain blood flow and oxygen delivery 24. Brain Chemistry Current bedside monitors of brain blood flow and oxygen delivery Global monitors Cannot detect regional abnormalities Local monitors Sample only a small region of the brain and highly

More information

Chapter 43 Monitoring of Filter Patency During Carotid Artery Stenting Using Near-Infrared Spectroscopy with High Time-Resolution

Chapter 43 Monitoring of Filter Patency During Carotid Artery Stenting Using Near-Infrared Spectroscopy with High Time-Resolution Chapter 43 Monitoring of Filter Patency During Carotid Artery Stenting Using Near-Infrared Spectroscopy with High Time-Resolution Takahiro Igarashi, Kaoru Sakatani, Tadashi Shibuya, Teruyasu Hirayama,

More information

Health Tech Symposium Fall, Dan Sommers P.E. EMT-P

Health Tech Symposium Fall, Dan Sommers P.E. EMT-P Health Tech Symposium Fall, 2009 Dan Sommers P.E. EMT-P Human Physiological Signals Simple Explanations for Complicated Systems Ref: Atlas of Human Anatomy, 4 th Edition Simple Schematic RA LA RV LV PCR

More information

Blood Pressure. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan

Blood Pressure. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan Bood Pressure Hsiao-Lung Chan, Ph.D. Dept Eectrica Engineering Chang Gung University, Taiwan chanh@mai.cgu.edu.tw Bood pressure traves from aorta toward smaer vesses 2 Bood pressure 3 Hagen-Poiseuie s

More information

Development of an Optical Brain-Computer Interface Using Dynamic Topographical Pattern Classification

Development of an Optical Brain-Computer Interface Using Dynamic Topographical Pattern Classification Development of an Optical Brain-Computer Interface Using Dynamic Topographical Pattern Classification by Larissa Christina Schudlo A thesis submitted in conformity with the requirements for the degree

More information

Competitive Intelligence Report Technology & Patents

Competitive Intelligence Report Technology & Patents Competitive Intelligence Report Technology & Patents N o n i n v a s i v e d i a b e t e s m o n i t o r i n g 01 November - 15 November, 2018 7 Published Granted Inactive 5 8 0 25 2 1 Start ups Companies

More information

Respiratory Physiology Part II. Bio 219 Napa Valley College Dr. Adam Ross

Respiratory Physiology Part II. Bio 219 Napa Valley College Dr. Adam Ross Respiratory Physiology Part II Bio 219 Napa Valley College Dr. Adam Ross Gas exchange Gas exchange in the lungs (to capillaries) occurs by diffusion across respiratory membrane due to differences in partial

More information

Infinity MCable -Masimo SET

Infinity MCable -Masimo SET Infinity MCable -Masimo SET Bring the advantages of Masimo s Signal Extraction Technology (SET ) to your pulse oximetry monitoring The noninvasive, motion-tolerant Infinity MCable -Masimo SET works with

More information

Near Infrared Spectroscopy in Medical Diagnostics

Near Infrared Spectroscopy in Medical Diagnostics Near Infrared Spectroscopy in Medical Diagnostics Michael D. Wider, PhD Vice President Of Technology Somanetics Corporation Adjunct Assistant Professor in Physiology Wayne State University School of Medicine

More information

This work is licensed under a Creative Commons Attribution-Noncommercial-Share Alike 3.0 License.

This work is licensed under a Creative Commons Attribution-Noncommercial-Share Alike 3.0 License. University of Massachusetts Medical School escholarship@umms UMass Center for Clinical and Translational Science Research Retreat 2017 UMass Center for Clinical and Translational Science Research Retreat

More information

Concurrent near-infrared spectroscopy (NIRS) and functional magnetic resonance imaging (fmri) of the brain

Concurrent near-infrared spectroscopy (NIRS) and functional magnetic resonance imaging (fmri) of the brain Motor cortex activation fmri Near-infrared imaging Concurrent near-infrared spectroscopy (NIRS) and functional magnetic resonance imaging (fmri) of the brain Sergio Fantini s group, Department of Biomedical

More information

Endorsements. Review Board. Pulse Oximetry and CO Oximetry 6/13/2007. Oximetry and CO-Oximetry for EMTs and

Endorsements. Review Board. Pulse Oximetry and CO Oximetry 6/13/2007. Oximetry and CO-Oximetry for EMTs and Oximetry and CO-Oximetry for EMTs and First Responders Bryan E. Bledsoe, DO, FACEP Endorsements This educational module has been endorsed by the following professional organizations: 2 Review Board Roy

More information

Cerebral Hemodynamics Measured by Near-Infrared Spectroscopy at Rest and During Motor Activation

Cerebral Hemodynamics Measured by Near-Infrared Spectroscopy at Rest and During Motor Activation Cerebral Hemodynamics Measured by Near-Infrared Spectroscopy at Rest and During Motor Activation Maria Angela Franceschini, Sergio Fantini Bioengineering Center, Department of Electrical Engineering and

More information

Chapter 62 Monitoring of Hemodynamic Change in Patients with Carotid Artery Stenosis During the Tilt Test Using Wearable Near-Infrared Spectroscopy

Chapter 62 Monitoring of Hemodynamic Change in Patients with Carotid Artery Stenosis During the Tilt Test Using Wearable Near-Infrared Spectroscopy Chapter 62 Monitoring of Hemodynamic Change in Patients with Carotid Artery Stenosis During the Tilt Test Using Wearable Near-Infrared Spectroscopy Takahiro Igarashi, Kaoru Sakatani, Norio Fujiwara, Yoshihiro

More information

Empowering the RT with New Noninvasive Monitoring Capabilities

Empowering the RT with New Noninvasive Monitoring Capabilities Empowering the RT with New Noninvasive Monitoring Capabilities Thomas Lamphere BS, RRT, RPFT Executive Director, Pennsylvania Society for Respiratory Care Adjunct Instructor, Respiratory Care Program,

More information

ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE

ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE Vet Times The website for the veterinary profession https://www.vettimes.co.uk ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE Author : CARL BRADBROOK Categories : Vets Date : October 7, 2013

More information

OpenStax-CNX module: m Transport of Gases. OpenStax College. Abstract

OpenStax-CNX module: m Transport of Gases. OpenStax College. Abstract OpenStax-CNX module: m46545 1 Transport of Gases OpenStax College This work is produced by OpenStax-CNX and licensed under the Creative Commons Attribution License 3.0 By the end of this section, you will

More information

NOTE: CONTENT CONTAINED IN THIS DOCUMENT IS TAKEN FROM ROSEN S EMERGENCY MEDICINE 9th Ed.

NOTE: CONTENT CONTAINED IN THIS DOCUMENT IS TAKEN FROM ROSEN S EMERGENCY MEDICINE 9th Ed. Chapter 5 Monitoring the Emergency Patient NOTE: CONTENT CONTAINED IN THIS DOCUMENT IS TAKEN FROM ROSEN S EMERGENCY MEDICINE 9th Ed. Italicized text is quoted directly from Rosen s. Key Concepts: 1. Monitoring

More information

ANALYSIS OF PHOTOPLETHYSMOGRAPHIC SIGNALS OF CARDIOVASCULAR PATIENTS

ANALYSIS OF PHOTOPLETHYSMOGRAPHIC SIGNALS OF CARDIOVASCULAR PATIENTS ANALYSIS OF PHOTOPLETHYSMOGRAPHIC SIGNALS OF CARDIOVASCULAR PATIENTS V.S. Murthy 1, Sripad Ramamoorthy 1, Narayanan Srinivasan 2, Sriram Rajagopal 2, M. Mukunda Rao 3 1 Department of Physics, Indian Institute

More information

Oximetry. Jan Bakker. Grants Hutchinson Pulsion.

Oximetry. Jan Bakker. Grants Hutchinson Pulsion. Oximetry Jan Bakker jan.bakker@erasmusmc.nl www.erasmusmc.me Grants Hutchinson Pulsion Postgraduate Workshop on Hemodynamic Monitoring ISICEM 2012 - Brussels Oximetry arterial (pulse oximetry) central/mixed

More information

Temporal decoupling of oxy- and deoxy-hemoglobin hemodynamic responses detected by functional near-infrared spectroscopy (fnirs)

Temporal decoupling of oxy- and deoxy-hemoglobin hemodynamic responses detected by functional near-infrared spectroscopy (fnirs) Temporal decoupling of oxy- and deoxy-hemoglobin hemodynamic responses detected by functional near-infrared spectroscopy (fnirs) Nicoladie D. Tam a, and George Zouridakis b a Department of Biological Sciences,

More information

Design of a SpO 2 Pulse Oximeter Prototype

Design of a SpO 2 Pulse Oximeter Prototype VIETNAM NATIONAL UNIVERSITY - HCMC International University Biomedical Engineering Department Design of a SpO Pulse Oximeter Prototype by Le Nguyen Trong Nghia A thesis submitted to the Biomedical Engineering

More information

Time domain diffuse optical imaging and spectroscopy for biomedical diagnostics. Rinaldo Cubeddu

Time domain diffuse optical imaging and spectroscopy for biomedical diagnostics. Rinaldo Cubeddu Time domain diffuse optical imaging and spectroscopy for biomedical diagnostics SIF, Pisa, September 22-26, 2014 Introduction: Time-Resolved Diffuse Optical Spectroscopy Time Resolved Diffusion Equation

More information

Evaluation of oesophageal reflectance pulse oximetry in major burns patients

Evaluation of oesophageal reflectance pulse oximetry in major burns patients Burns 31 (2005) 337 341 www.elsevier.com/locate/burns Evaluation of oesophageal reflectance pulse oximetry in major burns patients S.K. Pal a, *, P.A. Kyriacou b, S. Kumaran c, S. Fadheel c, R. Emamdee

More information

Blood Pressure Laboratory

Blood Pressure Laboratory Introduction The blood that circulates throughout the body maintains a flow and pressure. The nervous system can change the flow and pressure based on the particular needs at a given time. For example,

More information

Choosing Mindray SpO2

Choosing Mindray SpO2 Choosing Mindray SpO2 Aug. 2014 Choosing Mindray SpO2 The culmination of over 20 years of in-house SpO2 technology research. 4 advanced technologies make Mindray a leading SpO2 provider 5 dimensions of

More information

Lecture 5. Dr. Sameh Sarray Hlaoui

Lecture 5. Dr. Sameh Sarray Hlaoui Lecture 5 Myoglobin & Hemoglobin Dr. Sameh Sarray Hlaoui Myoglobin and Hemoglobin Myoglobin - Myoglobin and Hemoglobin are (metalloprotein containing a heme prosthetic group). hemeproteins - Function as

More information

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model Instrumentation and Control Department, JSS Academy of Technical Education, Noida (U.P.), India Abstract- The non-invasive

More information