Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan
|
|
- Darlene Lloyd
- 5 years ago
- Views:
Transcription
1 Oximeters Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan
2 Pulse oximeter Masimo pulse CO-oximeter ts/continuous/radical-7/ Oximeter
3 Oxygen transport in blood Hemoglobin Oxygen Hb HbO Deoxygen Hb Hb Hb+O HbO Oximeters 3
4 Arterial saturation of oxygen SaO SaO is affected by partial pressure of arterial oxygen, PO. At low partial pressures, such as tissue capillary pressure of 40 mmhg Millimeter of mercury Oxygen is released from hemoglobin, leading to a low SaO of about 75%. At high partial pressures, such as pulmonary capillary pressure of 95 mmhg Oxygen binds with hemoglobin, leading to high SaO of about 95 mmhg. Oximeters 4
5 Oxyhemoglobin dissociation curve Oximeters 5
6 Shifts of the oxyhemoglobin dissociation curve CO reacts with water to form carbonic acid HCO 3, which dissociates into hydrogen ions H + and bicarbonate ions HCO 3 When blood temperature increases, oxygen delivery to tissues is increased, making more oxygen available for binding to hemoglobin. BPG,3 biphosphoglycerate is a metabolically important phosphate compound. BPG concentration increases during hypoxic conditions of longer than a few hours. Oximeter 6
7 Carbon monoxide-hemoglobin dissociation curve A CO partial pressure of only 0.4 mmhg in alveoli causes CO to compete equally with oxygen for hemoglobin binding 1/50 of normal alveolar oxygen pressure of 100 mmhg. A CO partial pressure of only 0.6 mmhg can be lethal. Oximeter 7
8 Beer-Lambert s law Oximeters 8
9 Light interaction with absorption effect Beer-Lambert s law I x x I x di x ai x dx ax I x I x e 0 ai x x Absorbance of a wavelength of light absorption path length, l concentration, c absorbing species by ελ, molar absorptivity Oximeters 9
10 Oxygen saturation SaO, SpO measurement Relative amount of oxygen carried by the hemoglobin Two specific wavelengths : λ 1 : a red wavelength eg. 660 nm λ : a near infrared wavelength eg. 805 nm SaO C C HbO HbO C Hb Oximeters 10
11 Electromagnetic spectrum Oximeters 11
12 Light absorption signal produce AC output produce DC output Oximeters 1
13 SaO 0 10 d d C C IR Hb Hb IR HbO HbO IR I I 0 10 d d C C R Hb Hb R HbO HbO R I I Hb Hb IR HbO HbO IR Hb Hb R HbO HbO R DC IR IR DC R R C C C C I I I I IR R log log HbO R Hb R HbO IR Hb IR Hb R Hb IR Hb HbO HbO IR R IR R C C C SaO 13
14 Molar absorptivities of hemoglobin deoxyhemoglobin and oxyhemoglobin Oximeters 14
15 Total absorbance in tissue Scattering Variable absorbance not due to arterial blood Nonspecific sources of optical attenuation such as light scattering and geometric factors Oximeter 15
16 Pulse oximeter Takuo Aoyagi recognized that only pulsatile changes in light transmission through living tissues were necessary for estimating hemoglobin concentrations. Aoyagi s pulsatile observation is summarized in ratio calculations Zeros because each absorbance are constant The pulsatile path lengths for Hb and HbO are assumed to be equal Biopotential amplifiers 16
17 Arterial saturation of oxygen SaO Oximeter 17
18 Approximation of R in actual implementation AC refers to the pulsatile component of transmitted light DC refers to the nonpulsatile component of transmitted light Biopotential amplifiers 18
19 Calibration curve Scott Wilber made pulse oximetry accurate by abandoning the Beer-Lambert law R-vs.-SaO calibration curve was converted to a software lookup table for later real-time SpO estimation. Oximeter 19
20 Photodetector circuits /36086/does-this-photodiode-circuit-work
21 Wearable photoplethysmographic biosensors H.H. Asada et al, IEEE EMB, 003 Oximeter 1
22 Adaptive noise canceler Oximeter
23 Rapid desaturation-resaturation occurring during tapping motion Control A healthy volunteer inspired air from a face mask, in which the fraction of inspired oxygen was adjusted to decrease the control SpO values down from 100% to approximately 70% and up to 100% again. Masimo SET uses adaptive noise cancellation to minimize the motion effect Nellcor N 00, Nellcor N 3000 pulse oximeters Oximeter 3
24 Masimo s signal extraction technology SET adaptive noise cancellation Oximeter 4
25 Masimo s SET cont. Masimo s original signal extraction technology SET algorithm for adaptive noise cancellation. Ratios R corresponding to SpO values are input to the reference signal generator to output candidate noise sources. Each candidate noise source is input to the adaptive filter, with the filter output power then plotted A peak power corresponding to a lower SpO value may correspond to the ratio associated with saturation of venous blood rv. Oximeter 5
26 Masimo s SET cont. Because of two wavelengths, there are two input signals The reference noise source is defined as a function of the ratio Oximeter 6
27 Near-infrared spectroscopy NIRS by laser diode Uses the near-infrared region from 800 nm to 500 nm Applications in pharmaceutical and medical diagnostics including blood sugar and oximetry. OxiplexTS, ISS Inc, USA Oximeters 7
28 NIRS sensing by LEDs An sensor with central emitter and eight surrounding, detachable detectors Oximeters 8
29 Evoked cerebral function examination Electroencephalography EEG Near infrared NIR Magnetoencephalography MEG Functional MRI
30 Cerebral functional areas Oximeters 30
31 Neurovascular coupling Neurovascular coupling produces changes in blood flow and oxygen consumption Oximeters 31
32 Continuous-wave NIRS Ratio of reflected to incident light intensity, I/I 0, depends Absorption coefficient Changes of hemoglobin concentration [O Hb] [HHb] Path length L and differential path length factor B A I ln I OHb 0 u L [ O Hb] HHb [ HHb] B L Oximeters 3
33 Continuous-wave NIRS cont. Using two NIRS lights with different wavelengths for computing the change of hemoglobin concentrations ] [ ] [ Hb O HHb Hb O HHb HHb HHb Hb O HHb Hb O HHb Hb O Hb O u u HHb u u O Hb Oximeters 33
34 Application in brain oxygenation With 0% of oxygen consumption occurring in the human brain, any deficiency in oxygen supply may result in injury OxiplexTS, ISS Inc, USA Biomedical Optic 34
35 Cerebral Near-Infrared spectroscopy monitoring Monitoring of cerebral oxygenation Carotid stenosis or occlusion Brain ischemia Sleep apnea Ischemia of rat model Ischemia Cerebral response to carotid balloon inflation Obstructive sleep apnea
36 Related functional area for NIRS Prefrontal cortices PFC Premotor cortices PMC Supplement motor areas SMA Sensorimotor cortices SMC Oximeters 36
37 Brain oxygenation & reorganization in stroke patients Before rehabilitation oxyhb After rehabilitation Gait task TotalHb oxyhb deoxyhb Lesion Lesion
38 Application in peripheral vascular disease PVD PVD is a narrowing of the vessels carrying blood to the muscles in the legs and arms. Most patients report experiences of pain in the extremities due to inadequate blood flow and oxygen delivery to the exercising muscle. OxiplexTS, ISS Inc, USA Oximeters 38
39 Vessel occulsion test
40 NIRS demonstration Pulse Oximetry NEJM Muscular oxygenation under exercise Functional NIRS
41 Reference G. D. Baura, Medical device technologies, Elsevier Inc., 01. JG. Webster, Medical Instrumentation, application and design, 3rd, Houghton Mifflin, 000. JJ.Carr, JM.Brown, Introduction to Biomedical Equipment Technology, 4nd Edition, Prentice-Hall, 000.
Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan
Oximeters Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan chanhl@mail.cgu.edu.tw Oxygen transport in blood Hemoglobin Oxygen O Deoxygen +O O Oximeters Arterial saturation
More informationBiomedical Instrumentation D. The Photoplethysmogram
Biomedical Instrumentation D. The Photoplethysmogram Dr Gari Clifford Based on slides by Prof. Lionel Tarassenko The need for real-time oxygen saturation monitoring Respiratory failure & pulmonary disease
More informationOptical Sensor System for Hemoglobin Measurement
International Journal of Computational Engineering Research Vol, 03 Issue, 7 Optical Sensor System for Hemoglobin Measurement Rajashree Doshi 1, Anagha Panditrao 2 1 Department of Instrumentation and Control,
More informationMuscle OxyGen monitor. The Science Behind Moxy
Muscle OxyGen monitor The Science Behind Moxy How Does Moxy Monitor Work? Moxy Monitor works by shining near-infrared light onto the skin and detecting some of the light after it has travelled into the
More informationCorrelation of Ratio of Ratio of Near Infrared Vs Infrared Photoplethysmograph Spectroscopy for the Non-Invasive Measurement of Haemoglobin
Correlation of Ratio of Ratio of Near Infrared Vs Infrared Photoplethysmograph Spectroscopy for the Non-Invasive Measurement of Haemoglobin Ashita Diwate 1, Yogesh Kumar 2, Dr. Sanjeev Kumar 3, Prof. T.
More informationHow Does Pulse Oximetry Work? SpO2 Sensors Absorption at the Sensor Site Oxyhemoglobin Dissociation Curve
SpO2 Monitoring Contents 1 Introduction 1 What is SpO 2? How Does Pulse Oximetry Work? SpO2 Sensors Absorption at the Sensor Site Oxyhemoglobin Dissociation Curve 5 How Do I Use SpO2? Choosing a Sensor
More informationChandra Ramamoorthy MBBS; FRCA (UK) Professor of Anesthesiology, Stanford University. Director of Pediatric Cardiac Anesthesiology
Should NIRS be Standard Care for Pediatric CPB Chandra Ramamoorthy MBBS; FRCA (UK) Professor of Anesthesiology, Stanford University Director of Pediatric Cardiac Anesthesiology Stanford Children s Hospital
More informationResearch Journal of Pharmaceutical, Biological and Chemical Sciences
Research Journal of Pharmaceutical, Biological and Chemical Sciences Oxygen Glucose and Heart rate Monitor: An non Invasive approach Anne Frank Joe A*, Veeramani Saikrupa, and Prajukti Panda Sathyabama
More informationOxygen Saturation Monitors & Pulse Oximetry. D. J. McMahon rev cewood
Oxygen Saturation Monitors & Pulse Oximetry D. J. McMahon 141105 rev cewood 2017-11-14 2 3 Key Points Oxygen Saturation Monitors & Pulse Oximetry : Role of hemoglobin in respiration Basic principle of
More informationMeasuring Heart Rate and Blood Oxygen Levels for Portable and Wearable Devices
Measuring Heart Rate and Blood Oxygen Levels for Portable and Wearable Devices By Zhang Feng, Senior Medical Applications Engineer Marten Smith, Medical Marketing Manager Medical Products Group Microchip
More informationEstimation of Haemoglobin Using Non-Invasive Technique
Estimation of Haemoglobin Using Non-Invasive Technique Tatiparti Padma #1, Pinjala Jahnavi *2 #Department of ECE, GRIET, Hyderabad, Telangana *Department of ECE, GRIET, Hyderabad, Telangana 1tatipartipadma@gmail.com,
More informationGas Exchange in the Tissues
Gas Exchange in the Tissues As the systemic arterial blood enters capillaries throughout the body, it is separated from the interstitial fluid by only the thin capillary wall, which is highly permeable
More informationISSN: [Cortes* et al., 7(3): March, 2018] Impact Factor: 5.164
IJESRT INTERNATIONAL JOURNAL OF ENGINEERING SCIENCES & RESEARCH TECHNOLOGY DESING OF A HYPOXIA DEVICE DETECTOR WITH WIRELESS ALERT Ing. Teo Cortes *1 & Dra. Fatima Moumtadi *2 *Department of Electronics,
More informationDesign and Development of Non-Invasive Prototype to Measure Pulse Rate, Blood Glucose and Oxygen Saturation Level in Arterial Blood
Design and Development of Non-Invasive Prototype to Measure Pulse Rate, Blood Glucose and Oxygen Saturation Level in Arterial Blood Nazo Haroon Department of Mechatronics Engineering, CEME National University
More informationDevelopment of a New Rehabilitation System Based on a Brain-Computer Interface Using Near-Infrared Spectroscopy
Development of a New Rehabilitation System Based on a Brain-Computer Interface Using Near-Infrared Spectroscopy Takafumi Nagaoka, Kaoru Sakatani, Takayuki Awano, Noriaki Yokose, Tatsuya Hoshino, Yoshihiro
More informationDr. D.McG.Clarkson Determining the Accuracy of SpO2 Values
Dr. D.McG.Clarkson Determining the Accuracy of SpO2 Values 1 Life saving technology The availability of the LED provided a convenient light which provides advantages of :- Compact size Modest power requirements
More informationDr. Puntarica Suwanprathes. Version 2007
Dr. Puntarica Suwanprathes Version 2007 O 2 and CO 2 transport in blood Oxyhemoglobin dissociation curve O 2 consumption (VO 2 ) CO 2 production (VCO 2 ) O 2 capacity O 2 content: CaO 2 or CvO 2 %saturation
More informationPermanent City Research Online URL:
Kyriacou, P. A., Pal, S. K., Langford, R. & Jones, DP (2006). Electro-optical techniques for the investigation of oesophageal photoplethysmographic signals and blood oxygen saturation in burns. Measurement
More informationJournal of Faculty of Engineering & Technology DESIGN AND IMPLEMENTATION OF A WEARABLE HEALTH DEVICE
PAK BULLET TRAIN (PBT) JFET 22(2) (2015) 39-44 Journal of Faculty of Engineering & Technology journal homepage: www.pu.edu.pk/journals/index.php/jfet/index DESIGN AND IMPLEMENTATION OF A WEARABLE HEALTH
More informationElectroencephalogram (EEG) Hsiao-Lung Chan Dept Electrical Engineering Chang Gung University
Electroencephalogram (EEG) Hsiao-Lung Chan Dept Electrical Engineering Chang Gung University chanhl@mail.cgu.edu.tw Cerebral function examination Electroencephalography (EEG) Near infrared ray spectroscopy
More informationArterial blood oxygen saturation during blood pressure cuff-induced hypoperfusion
Journal of Physics: Conference Series Arterial blood oxygen saturation during blood pressure cuff-induced hypoperfusion To cite this article: P A Kyriacou et al 27 J. Phys.: Conf. Ser. 85 1226 Related
More informationClinically Available Optical Topography System
Clinically Available Optical Topography System Clinically Available Optical Topography System 18 Fumio Kawaguchi Noriyoshi Ichikawa Noriyuki Fujiwara Yûichi Yamashita Shingo Kawasaki OVERVIEW: Progress
More informationJAMES R. KNIGHT
JAMES R. KNIGHT knightjr@ah.org James is currently the supervisor of clinical engineering at Sonora Regional Medical Center. He is the chairman of the CMIA Training & Education Committee. He also develops
More informationDesign a system of measurement of heart rate, oxygen saturation in blood and body temperature with non-invasive method
IOP Conference Series: Materials Science and Engineering PAPER OPEN ACCESS Design a system of measurement of heart rate, oxygen saturation in blood and body temperature with non-invasive method To cite
More informationMonitoring Tumor Therapy
Monitoring Tumor Therapy Zenalux Biomedical, Inc. The Zenascope was used to monitor the physiological response to a vascular disrupting agent in mouse tumors. Changes in hemoglobin concentration, hemoglobin
More informationNon-invasive Hemoglobin Measurement for Anemia Diagnosis
Non-invasive Hemoglobin Measurement for Anemia Diagnosis Raditya Artha Rochmanto 1, Hasballah Zakaria 2 Ratih Devi Alviana 3, Nurhalim Shahib 4 Department of Biomedical Engineering Medical Faculty Institut
More informationPBL SEMINAR. HEMOGLOBIN, O 2 -TRANSPORT and CYANOSIS An Overview
1 University of Papua New Guinea School of Medicine and Health Sciences Division of Basic Medical Sciences Discipline of Biochemistry and Molecular Biology PBL SEMINAR HEMOGLOBIN, O 2 -TRANSPORT and CYANOSIS
More informationORIGINAL RESEARCH. Effect of road traffic accident contaminants on pulse oximetry among normoxaemic volunteers
ISSN 1447-4999 ORIGINAL RESEARCH Effect of road traffic accident contaminants on pulse oximetry among normoxaemic volunteers Dr Gyorgyi Kamaras MD, FCEM 1,2, Dr Tamas Geller MD 1, Dr Csaba Dioszeghy MD
More informationPulse oximetry use in small animals
Vet Times The website for the veterinary profession https://www.vettimes.co.uk Pulse oximetry use in small animals Author : Clara Rigotti, Marieke de Vries Categories : Vets Date : October 10, 2011 Clara
More informationConsumer Near Infrared Spectroscopy as a Training and Recovery Tool. Jon Tarkington
Consumer Near Infrared Spectroscopy as a Training and Recovery Tool Jon Tarkington Outline - What is NIRS? - Devices - Data interpretation - Practical Application - Future Impacts - Suffix light Disclaimers
More informationALCO Regulations. Protocol pg. 47
For the EMT-Basic Objectives Understand ALCO regulations relative to monitoring pulse oximetry by the EMT-B Review the signs and symptoms of respiratory compromise. Understand the importance of adequate
More informationPulse oximetry revisited. Dr Liesel Bösenberg Specialist Physician and Fellow in Critical Care Kalafong Hospital University of Pretoria
Pulse oximetry revisited Dr Liesel Bösenberg Specialist Physician and Fellow in Critical Care Kalafong Hospital University of Pretoria Topics that will be discussed and dissected: Revisiting physiology
More informationCEREBRAL OXIMETRY IN INFANTS WITH HIE DAPHNA YASOVA BARBEAU, MD FN3 MEETING 2018
CEREBRAL OXIMETRY IN INFANTS WITH HIE DAPHNA YASOVA BARBEAU, MD FN3 MEETING 2018 OBJECTIVES 1. Understand how cerebral oximetry works 2. Understand how cerebral oximetry may guide intervention 3. Learn
More informationDevelopment of an Optical Sensor Based System for Detection and Monitoring of Hemoglobin in Blood
International Journal of Electronics Engineering Research. ISSN 0975-6450 Volume 9, Number 4 (2017) pp. 483-489 Research India Publications http://www.ripublication.com Development of an Optical Sensor
More informationDisclosures. Set up audience participation. Test question. Outline. Neuromonitoring What and When? IP for monitoring technology licensed to Medtronic
Neuromonitoring What and When? Disclosures IP for monitoring technology licensed to Medtronic Ken Brady, MD Pediatrics, Anesthesia, Critical Care Texas Children s Hospital Baylor College of Medicine Set
More informationOxygen Level Measurement Techniques: Pulse Oximetry
Oxygen Level Measurement Techniques: Pulse Oximetry Reem Abedalmoniam Ismail 1 and Sharief F. Babiker 2 1 Computer Engineering Dept., ElNeelain University, Khartoum, Sudan 2 Electrical and Electronics
More information3. Which of the following would be inconsistent with respiratory alkalosis? A. ph = 7.57 B. PaCO = 30 mm Hg C. ph = 7.63 D.
Pilbeam: Mechanical Ventilation, 4 th Edition Test Bank Chapter 1: Oxygenation and Acid-Base Evaluation MULTIPLE CHOICE 1. The diffusion of carbon dioxide across the alveolar capillary membrane is. A.
More informationSignal Quality Analysis in Pulse Oximetry: Modelling and Detection of Motion Artifact
Signal Quality Analysis in Pulse Oximetry: Modelling and Detection of Motion Artifact by Geoffrey Clarke, B.Eng. A thesis submitted to the Faculty of Graduate and Postdoctoral Affairs in partial fulfillment
More informationAppendix D An unresponsive patient with shallow, gasping breaths at a rate of six per minute requires:
Answer Key Appendix D-2 1. An unresponsive patient with shallow, gasping breaths at a rate of six per minute requires: a. oxygen given via nasal cannula b. immediate transport to a medical facility c.
More informationAppendix E Choose the sign or symptom that best indicates severe respiratory distress.
Appendix E-2 1. In Kansas EMT-B may monitor pulse oximetry: a. after they complete the EMT-B course b. when the service purchases the state approved pulse oximeters c. when the service director receives
More informationWestern Hospital System. PSG in History. SENSORS in the field of SLEEP. PSG in History continued. Remember
SENSORS in the field of SLEEP Mrs. Gaye Cherry: Scientist in Charge Department of Sleep and Respiratory Medicine Sleep Disorders Unit Western Hospital PSG in History 1875: Discovery of brain-wave activity
More informationBLOOD GLUCOSE LEVEL MONITORING BY NONINVASIVE METHOD USING NEAR INFRA RED SENSOR
BLOOD GLUCOSE LEVEL MONITORING BY NONINVASIVE METHOD USING NEAR INFRA RED SENSOR P.Daarani 1 & A.Kavithamani 2 International Journal of Latest Trends in Engineering and Technology pp. 141-147 DOI: http://dx.doi.org/10.21172/1.ires.19
More informationInternational Journal of Research and Applications July - Sep 2017 Transactions 4(15):
DOI: 10.17812/IJRA.4.15(98)2017 International Journal of Research and Applications July - Sep 2017 Transactions 4(15): 585-591 B I O - M E D I C A L R E V I E W R E P O R T eissn : 2349 0020 pissn : 2394
More informationSaturday Physics Honors presents. Professor Enrico Gratton, Looking Into the Brain with Lasers. (With some demonstrations)
Saturday Physics Honors presents Professor Enrico Gratton, Looking Into the Brain with Lasers (With some demonstrations) Can we see through the body using light? What we expect to see? Startrek Movie clip
More informationContinuous Monitoring of Blood Pressure Based on Heart Pulse Analysis
Journal of Physics: Conference Series PAPER OPEN ACCESS Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis To cite this article: Valerie Tan et al 2018 J. Phys.: Conf. Ser. 1049 012062
More informationQuantitative Performance Assessment Studies in Air Transportation
Quantitative Performance Assessment Studies in Air Transportation Portable Brain Activity Monitoring Technology Kurtulus Izzetoglu, PhD Drexel University Optical Brain Imaging Group Drexel University,
More informationCheck a Pulse! When to Question SpO 2, NIBP & EtCO 2 Readings
Check a Pulse! When to Question SpO 2, NIBP & EtCO 2 Readings Mike McEvoy, PhD, RN, CCRN, NRP Professor Emeritus - Critical Care Medicine Albany Medical College Albany, New York Chair Resuscitation Committee
More informationIs Hypoxia common despite oxygen treatment in the Acute Medical ward?
Is Hypoxia common despite oxygen treatment in the Acute Medical ward? Dr. R Dhar, Dr. RN Harrison and Dr. PD Snashall University Hospital of North Tees, Stockton on Tees, TS198PE, United Kingdom (Received
More informationA Three-Wavelength Pulse Oximeter for Carboxyhemoglobin Determination
Three-Wavelength Pulse Oximeter for Carboxyhemoglobin Determination R. P. Buinevicius 1 J. G. Webster 2 1 PhD Student, Department of Electrical and Computer Engineering University of Wisconsin 1415 Engineering
More informationTesting Masimo rainbow Technology with vpad-a1 1. Background
1 Testing Masimo rainbow Technology with vpad-a1 1. Background 1.1 Pulse Oximetry Pulse oximetery measures oxygen saturation or the ratio of oxyhemoglobin to the total amount of hemoglobin in the arterial
More informationCity, University of London Institutional Repository
City Research Online City, University of London Institutional Repository Citation: Kyriacou, P. A., Shafqat, K. & Pal, S. K. (29). Pilot investigation of photoplethysmographic signals and blood oxygen
More informationElectroencephalograms and Neuro-Rehabilitation. Hsiao-Lung Chan Dept Electrical Engineering Chang Gung University
Electroencephalograms and Neuro-Rehabilitation Hsiao-Lung Chan Dept Electrical Engineering Chang Gung University chanhl@mail.cgu.edu.tw Brain Cerebrum ( 大腦 ) Receives and processes conscious sensation
More informationIntroduction. Device Description. Dual-Band Spectral Output
Optimum Spectrum and Pulse Shape for Vascular Lesion Treatment: The Science Behind MaxG James Childs, Ph.D.; Andrei Erofeev, Ph.D.; Mikhail Smirnov, Ph.D.; and Gregory Altshuler, Ph.D., Sc.D. Introduction
More informationLight is wearable Health Monitoring and Fitness Tracking Light is OSRAM
www.osram-os.com Light is wearable Health Monitoring and Fitness Tracking Light is OSRAM Sensor products Exactly what you need: high efficiency and great diversity OSRAM Opto Semiconductors offers a great
More informationBiomedical Instrumentation E. Blood Pressure
Biomedical Instrumentation E. Blood Pressure Dr Gari Clifford Adapted from slides by Prof. Lionel Tarassenko Blood pressure Blood is pumped around the body by the heart. It makes its way around the body
More informationOptical non-invasive methods for characterization of the human health status
Optical non-invasive methods for characterization of the human health status Abstract Jens Kraitl and Hartmut Ewald Institute of General Electrical Engineering, University of Rostock, Rostock, Germany
More informationRespiratory System 1. A function of the structure labelled X is to
1 Respiratory System 1. A function of the structure labelled X is to A. produce sound. B. exchange gases. C. carry air into and out of the lung. D. stimulate the breathing centre in the brain. 2. Identify
More informationHybrid EEG-HEG based Neurofeedback Device
APSIPA ASC 2011 Xi an Hybrid EEG-HEG based Neurofeedback Device Supassorn Rodrak *, Supatcha Namtong, and Yodchanan Wongsawat ** Department of Biomedical Engineering, Faculty of Engineering, Mahidol University,
More informationPulse Oximeter Accuracy Study
Patient Monitoring Technical Library SpO2 Monitoring Pulse Oximeter Accuracy Study 0 614-902709 Printed : 2008/12/24 This document is owned or controlled by Nihon Kohden and is protected by copyright law.
More informationNear Infrared Spectroscopy in Neonatal Intensive Care
Near Infrared Spectroscopy in Neonatal Intensive Care Adré J. du Plessis, MBChB Fetal-Neonatal Neurology, Children s Hospital, Boston, MA Children s Hospital and Harvard Medical School Disclosure Statement
More informationLecture 10. Circulatory systems; flow dynamics, flow regulation in response to environmental and internal conditions.
Lecture 10 Circulatory systems; flow dynamics, flow regulation in response to environmental and internal conditions Professor Simchon Influence of P O2 on Hemoglobin Saturation Hemoglobin saturation plotted
More informationDual channel photoplethysmography studies of cardio-vascular response to the body position changes
Dual channel photoplethysmography studies of cardio-vascular response to the body position changes R. Erts, I. Kukulis, J. Spigulis, L. Kere University of Latvia, Raina Blvd. 9, Riga, LV-, Latvia Renars.erts@lu.lv
More informationBlood Pressure Estimation Using Photoplethysmography (PPG)
Blood Pressure Estimation Using Photoplethysmography (PPG) 1 Siddhi Sham Karande, BE E&TC, VIIT Pune. 2 Kiran Rajendrasingh Thakur, BE E&TC, VIIT Pune. 3 Sneha Dattatraya Waghmare, BE E&TC, VIIT Pune 4
More informationOPTICAL PERFUSION AND OXYGENATION CHARACTERIZATION IN A LIVER PHANTOM. A Thesis TRAVIS J. KING
OPTICAL PERFUSION AND OXYGENATION CHARACTERIZATION IN A LIVER PHANTOM A Thesis by TRAVIS J. KING Submitted to the Office of Graduate Studies of Texas A&M University in partial fulfillment of the requirements
More informationGLUCOSE CONCENTRATION MEASUREMENT USING FIBRE OPTIC SENSOR
Chapter 4 GLUCOSE CONCENTRATION MEASUREMENT USING FIBRE OPTIC SENSOR Abstract Design and development ofan evanescent wave fibre optic sensor for the determination ofglucose concentration is described in
More informationNoninvasive Detection of Bilirubin in Discrete Vessels
, July 2-4, 2014, London, U.K. Noninvasive Detection of Bilirubin in Discrete Vessels Mark McEwen, Karen J. Reynolds Abstract Bilirubin, at the levels encountered in hyperbilirubinaemia, makes a similar
More informationSmart Probe for Uric Acid Detection
Smart Probe for Uric Acid Detection Rahul Pimpare 1, Samyak Ramteke 2, Pauroosh Kaushal 3 1 Student,Department of instrumentation & Control Engineering, D Y Patil College of Engineering Akurdi, Pune,India.
More informationQuestion Expected Answers Marks Additional Guidance 1 (a) C ; E ; A ; B ; 4. PhysicsAndMathsTutor.com
Question Expected Answers Marks Additional Guidance 1 (a) C ; E ; A ; B ; 4 Question Expected Answers Marks Additional Guidance (b) (i) P wave combined with larger peak before QRS complex ; Note: - look
More informationUse of the Free Electron Laser for the Noninvasive Determination of Retinal Oxyhemoglobin Saturation by Near Infrared Reflectance Spectrophotometry
Use of the Free Electron Laser for the Noninvasive Determination of Retinal Oxyhemoglobin Saturation by Near Infrared Reflectance Spectrophotometry Ref: Eye, M.C. Escher, 1946 Ref: Eye, M.C. Escher, 1946
More informationNEAR INFRARED TRANSMISSION SPECTROSCOPY AS APPLIED TO FATS AND OIL
NEAR INFRARED TRANSMISSION SPECTROSCOPY AS APPLIED TO FATS AND OIL Phillip J. Clancy, NIR Technology Systems, 56 Kitchener Pde, Bankstown, NSW, Australia. Near Infrared Transmission (NIT) Spectroscopy
More informationPulse Oximeter Accuracy Study
Patient Monitoring Technical Library SpO2 Monitoring Pulse Oximeter Accuracy Study 0 614-902709A Printed : 2014/5/26 This document is owned or controlled by Nihon Kohden and is protected by copyright law.
More informationBlood Gases, ph, Acid- Base Balance
Blood Gases, ph, Acid- Base Balance Blood Gases Acid-Base Physiology Clinical Acid-Base Disturbances Blood Gases Respiratory Gas Exchange Chemical Control of Respiration Dyshemoglobins Oxygen Transport
More informationCurrent bedside monitors of brain blood flow and oxygen delivery
24. Brain Chemistry Current bedside monitors of brain blood flow and oxygen delivery Global monitors Cannot detect regional abnormalities Local monitors Sample only a small region of the brain and highly
More informationChapter 43 Monitoring of Filter Patency During Carotid Artery Stenting Using Near-Infrared Spectroscopy with High Time-Resolution
Chapter 43 Monitoring of Filter Patency During Carotid Artery Stenting Using Near-Infrared Spectroscopy with High Time-Resolution Takahiro Igarashi, Kaoru Sakatani, Tadashi Shibuya, Teruyasu Hirayama,
More informationHealth Tech Symposium Fall, Dan Sommers P.E. EMT-P
Health Tech Symposium Fall, 2009 Dan Sommers P.E. EMT-P Human Physiological Signals Simple Explanations for Complicated Systems Ref: Atlas of Human Anatomy, 4 th Edition Simple Schematic RA LA RV LV PCR
More informationBlood Pressure. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan
Bood Pressure Hsiao-Lung Chan, Ph.D. Dept Eectrica Engineering Chang Gung University, Taiwan chanh@mai.cgu.edu.tw Bood pressure traves from aorta toward smaer vesses 2 Bood pressure 3 Hagen-Poiseuie s
More informationDevelopment of an Optical Brain-Computer Interface Using Dynamic Topographical Pattern Classification
Development of an Optical Brain-Computer Interface Using Dynamic Topographical Pattern Classification by Larissa Christina Schudlo A thesis submitted in conformity with the requirements for the degree
More informationCompetitive Intelligence Report Technology & Patents
Competitive Intelligence Report Technology & Patents N o n i n v a s i v e d i a b e t e s m o n i t o r i n g 01 November - 15 November, 2018 7 Published Granted Inactive 5 8 0 25 2 1 Start ups Companies
More informationRespiratory Physiology Part II. Bio 219 Napa Valley College Dr. Adam Ross
Respiratory Physiology Part II Bio 219 Napa Valley College Dr. Adam Ross Gas exchange Gas exchange in the lungs (to capillaries) occurs by diffusion across respiratory membrane due to differences in partial
More informationInfinity MCable -Masimo SET
Infinity MCable -Masimo SET Bring the advantages of Masimo s Signal Extraction Technology (SET ) to your pulse oximetry monitoring The noninvasive, motion-tolerant Infinity MCable -Masimo SET works with
More informationNear Infrared Spectroscopy in Medical Diagnostics
Near Infrared Spectroscopy in Medical Diagnostics Michael D. Wider, PhD Vice President Of Technology Somanetics Corporation Adjunct Assistant Professor in Physiology Wayne State University School of Medicine
More informationThis work is licensed under a Creative Commons Attribution-Noncommercial-Share Alike 3.0 License.
University of Massachusetts Medical School escholarship@umms UMass Center for Clinical and Translational Science Research Retreat 2017 UMass Center for Clinical and Translational Science Research Retreat
More informationConcurrent near-infrared spectroscopy (NIRS) and functional magnetic resonance imaging (fmri) of the brain
Motor cortex activation fmri Near-infrared imaging Concurrent near-infrared spectroscopy (NIRS) and functional magnetic resonance imaging (fmri) of the brain Sergio Fantini s group, Department of Biomedical
More informationEndorsements. Review Board. Pulse Oximetry and CO Oximetry 6/13/2007. Oximetry and CO-Oximetry for EMTs and
Oximetry and CO-Oximetry for EMTs and First Responders Bryan E. Bledsoe, DO, FACEP Endorsements This educational module has been endorsed by the following professional organizations: 2 Review Board Roy
More informationCerebral Hemodynamics Measured by Near-Infrared Spectroscopy at Rest and During Motor Activation
Cerebral Hemodynamics Measured by Near-Infrared Spectroscopy at Rest and During Motor Activation Maria Angela Franceschini, Sergio Fantini Bioengineering Center, Department of Electrical Engineering and
More informationChapter 62 Monitoring of Hemodynamic Change in Patients with Carotid Artery Stenosis During the Tilt Test Using Wearable Near-Infrared Spectroscopy
Chapter 62 Monitoring of Hemodynamic Change in Patients with Carotid Artery Stenosis During the Tilt Test Using Wearable Near-Infrared Spectroscopy Takahiro Igarashi, Kaoru Sakatani, Norio Fujiwara, Yoshihiro
More informationEmpowering the RT with New Noninvasive Monitoring Capabilities
Empowering the RT with New Noninvasive Monitoring Capabilities Thomas Lamphere BS, RRT, RPFT Executive Director, Pennsylvania Society for Respiratory Care Adjunct Instructor, Respiratory Care Program,
More informationADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE
Vet Times The website for the veterinary profession https://www.vettimes.co.uk ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART ONE Author : CARL BRADBROOK Categories : Vets Date : October 7, 2013
More informationOpenStax-CNX module: m Transport of Gases. OpenStax College. Abstract
OpenStax-CNX module: m46545 1 Transport of Gases OpenStax College This work is produced by OpenStax-CNX and licensed under the Creative Commons Attribution License 3.0 By the end of this section, you will
More informationNOTE: CONTENT CONTAINED IN THIS DOCUMENT IS TAKEN FROM ROSEN S EMERGENCY MEDICINE 9th Ed.
Chapter 5 Monitoring the Emergency Patient NOTE: CONTENT CONTAINED IN THIS DOCUMENT IS TAKEN FROM ROSEN S EMERGENCY MEDICINE 9th Ed. Italicized text is quoted directly from Rosen s. Key Concepts: 1. Monitoring
More informationANALYSIS OF PHOTOPLETHYSMOGRAPHIC SIGNALS OF CARDIOVASCULAR PATIENTS
ANALYSIS OF PHOTOPLETHYSMOGRAPHIC SIGNALS OF CARDIOVASCULAR PATIENTS V.S. Murthy 1, Sripad Ramamoorthy 1, Narayanan Srinivasan 2, Sriram Rajagopal 2, M. Mukunda Rao 3 1 Department of Physics, Indian Institute
More informationOximetry. Jan Bakker. Grants Hutchinson Pulsion.
Oximetry Jan Bakker jan.bakker@erasmusmc.nl www.erasmusmc.me Grants Hutchinson Pulsion Postgraduate Workshop on Hemodynamic Monitoring ISICEM 2012 - Brussels Oximetry arterial (pulse oximetry) central/mixed
More informationTemporal decoupling of oxy- and deoxy-hemoglobin hemodynamic responses detected by functional near-infrared spectroscopy (fnirs)
Temporal decoupling of oxy- and deoxy-hemoglobin hemodynamic responses detected by functional near-infrared spectroscopy (fnirs) Nicoladie D. Tam a, and George Zouridakis b a Department of Biological Sciences,
More informationDesign of a SpO 2 Pulse Oximeter Prototype
VIETNAM NATIONAL UNIVERSITY - HCMC International University Biomedical Engineering Department Design of a SpO Pulse Oximeter Prototype by Le Nguyen Trong Nghia A thesis submitted to the Biomedical Engineering
More informationTime domain diffuse optical imaging and spectroscopy for biomedical diagnostics. Rinaldo Cubeddu
Time domain diffuse optical imaging and spectroscopy for biomedical diagnostics SIF, Pisa, September 22-26, 2014 Introduction: Time-Resolved Diffuse Optical Spectroscopy Time Resolved Diffusion Equation
More informationEvaluation of oesophageal reflectance pulse oximetry in major burns patients
Burns 31 (2005) 337 341 www.elsevier.com/locate/burns Evaluation of oesophageal reflectance pulse oximetry in major burns patients S.K. Pal a, *, P.A. Kyriacou b, S. Kumaran c, S. Fadheel c, R. Emamdee
More informationBlood Pressure Laboratory
Introduction The blood that circulates throughout the body maintains a flow and pressure. The nervous system can change the flow and pressure based on the particular needs at a given time. For example,
More informationChoosing Mindray SpO2
Choosing Mindray SpO2 Aug. 2014 Choosing Mindray SpO2 The culmination of over 20 years of in-house SpO2 technology research. 4 advanced technologies make Mindray a leading SpO2 provider 5 dimensions of
More informationLecture 5. Dr. Sameh Sarray Hlaoui
Lecture 5 Myoglobin & Hemoglobin Dr. Sameh Sarray Hlaoui Myoglobin and Hemoglobin Myoglobin - Myoglobin and Hemoglobin are (metalloprotein containing a heme prosthetic group). hemeproteins - Function as
More informationNon-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model
Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model Instrumentation and Control Department, JSS Academy of Technical Education, Noida (U.P.), India Abstract- The non-invasive
More information