The estimated prevalence of sleep apnea is 2% to 4% in

Size: px
Start display at page:

Download "The estimated prevalence of sleep apnea is 2% to 4% in"

Transcription

1 Arrhythmia/Electrophysiology Diagnosis of Sleep-Related Breathing Disorders by Visual Analysis of Transthoracic Impedance Signals in Pacemakers Christoph Scharf, MD; Yong K. Cho, PhD; Konrad E. Bloch, MD; Corinna Brunckhorst, MD; Firat Duru, MD; Kryzstof Balaban, MD; Nancy Foldvary, DO; Lynn Liu, MD; Richard C. Burgess, PhD, MD; Reto Candinas, MD; Bruce L. Wilkoff, MD Background Minute ventilation sensors of cardiac pacemakers measure ventilation by means of transthoracic impedance changes between the pacemaker case and the electrode tip. We investigated whether this technique might detect sleep-related breathing disorders. Methods and Results In 22 patients, analog waveforms of the transthoracic impedance signal measured by the pacemaker minute ventilation sensor over the course of a night were visualized, scored for apnea/hypopnea events, and compared with simultaneous polysomnography. Analysis of transthoracic impedance signals correctly identified the presence or absence of moderate to severe sleep apnea (apnea/hypopnea index, AHI 20 h 1 ) in all patients (receiver operating characteristics, ROC 1.0). The ROC for AHI scores of 5 h 1 and 10 h 1 showed an area under the curve of 0.95, P 0.005, and 0.97, P , respectively. Accuracy over time assessed by comparing events per 5-minute epochs was high (Cronbach reliability coefficient, 0.85; intraclass correlation, 0.73). Event-by-event comparison within 15 seconds revealed agreement in 81% (, 0.77; P 0.001). Conclusions Detection of apnea/hypopnea events by pacemaker minute ventilation sensors is feasible and accurate compared with laboratory polysomnography. This technique might be useful to screen and monitor sleep-related breathing disorders in pacemaker patients. (Circulation. 2004;110: ) Key Words: pacemakers diagnosis sleep apnea syndromes The estimated prevalence of sleep apnea is 2% to 4% in the general population 1 and up to 50% in patients with congestive heart failure. 2 Because sleep-disordered breathing (SDB) is associated with a number of cardiovascular diseases, 3 including hypertension, 4 heart failure, 5,6 and coronary artery disease, 7 it may be frequent in patients with implanted pacemakers or intracardiac defibrillators. However, many patients with SDB remain undiagnosed. Diagnosis of SDB requires multichannel recordings in the sleep laboratory. One method to measure respiration is impedance pneumography. 8,9 After injection of an electrical transthoracic current, the changes in electric conductance (transthoracic impedance) are measured, which depend on the ratio of gas/fluid volumes. This principle is used in minute ventilation sensors of cardiac pacemakers. In a previous study, we have shown that impedance minute ventilation measured by pacemaker sensors is closely correlated with oral-nasal airflow. 10 Consequently, the sensor might also detect disturbances of ventilation during sleep. This study was designed to test the feasibility of SDB detection on the basis of a visual analysis of transthoracic impedance waveforms measured by the pacemaker minute ventilation sensor. Analysis of SDB by pacemakers might be an ideal tool to screen for sleep apnea and monitor treatment effects of cardiac pacing. 11,12 As reference, we used simultaneous clinical standard multichannel polysomnography. 13 Methods Study Design This study included 22 patients at 2 institutions who provided informed written consent approved by the local medical ethics committees and had had a rate-responsive pacemaker with a minute ventilation sensor (Medtronic Kappa 400) implanted at least 6 months earlier (Table 1). Sleep apnea had been diagnosed previously in 2 patients. The rate-response function was programmed on to activate the minute ventilation sensor. Patients reported to the sleep laboratory at 8 PM and were instrumented for a standard overnight sleep study. The lights were turned off between 10 and 11 PM. Each center scored the clinical polysomnography, and 1 investigator analyzed the transthoracic impedance tracing data in a blinded manner. Measurements Polysomnographic recording systems (Institution B Alice 4, Respironics; Vangard, Vangard Inc) included derivations of electroen- Received July 8, 2003; de novo received April 8, 2004; revision received June 16, 2004; accepted June 17, From University Hospital of Zurich, Department of Internal Medicine, Division of Cardiology and Pneumology, Zurich, Switzerland (C.S., K.E.B., C.B., F.D., R.C.); the Cleveland Clinic Foundation, Departments of Cardiovascular Medicine and Pulmonary Medicine, Cleveland, Ohio (K.B., N.F., L.L., R.C.B., B.L.W.); and Medtronic Inc, Minneapolis, Minn (Y.K.C.). Correspondence to Christoph Scharf, MD, Cardiovascular Center, University Hospital Zurich, Rämistrasse 100, CH-8091 Zürich, Switzerland. christoph.scharf@usz.ch 2004 American Heart Association, Inc. Circulation is available at DOI: /01.CIR EB 2562

2 Scharf et al Diagnosis of Sleep-Related Breathing Disorders 2563 TABLE 1. Patient Characteristics (n 22) Sex, M/F 14/8 Mean age, y Body mass index, kg/m 2 (range) (20 41) Ejection fraction, % (range) (15 60) Structural heart disease, n (%) 18 (81) Valvular 7 CAD 5 Hypertension 2 HOCM 1 Heart transplant 1 AV node ablation NICM 2 Sternotomy because of previous open-heart 7 (32) surgery, n (%) Pacemaker lower rate 50/min 1 60/min 8 70/min 11 75/min 1 80/min 1 Impedance of the MV-sensor electrode Location of MV-sensor electrode, 20/2 atrial/ventricular Epworth sleepiness scale (range) (0 16) Previous diagnosis of sleep apnea, n (%) 2 (9) CAD indicates coronary artery disease; HOCM, hypertrophic obstructive cardiomyopathy; and MV, minute ventilation. cephalogram (C3A2, C4A1), electro-oculogram (LEA2, REA1), submental electromyogram, ECG, pulse oximetry, rib cage and abdominal excursions by strain gauges, nasal pressure recording for measurement of ventilation, 14 microphones, body position by accelerometer, and audiovisual monitoring by a low-light camera. 15 An electrical signal for time synchronization of the recording systems was set manually at the beginning of the recording. The transthoracic impedance of the minute ventilation sensor is obtained by injecting a biphasic current pulse (1 ma, 15 s, 16 Hz) into the chest cavity via the proximal electrode of the pacemaker lead and measuring the resulting voltage change between the device case and the distal electrode at a 16-Hz sampling rate. The raw signal was telemetered by use of a modified DR-180 Holter monitor (Rozinn Medical, model RZ152, modified by Northeast Monitoring), filtered with a bandpass from 0.05 to 0.8 Hz, processed, and visualized by use of a commercially available software (MatLab, MathWorks). The processed transthoracic impedance was visualized as an analog waveform and scored manually for apnea/hypopnea events and compared with polysomnographic recordings (Figure 1). Data Analysis Different blinded investigators scored polysomnographic and transthoracic impedance tracings. A hypopnea/apnea event on the transthoracic impedance signal was defined by an amplitude decrease 50% from baseline for at least 10 seconds duration and was compared with polysomnographic recordings. According to previously published guidelines, a hypopnea/apnea event had to fulfill 1 of the 2 following criteria for at least 10 seconds: (1) a clear decrease of 50% from baseline in the amplitude of a valid measure of breathing, ie, nasal pressure swings (or excursions of the rib cage and abdominal strain-gauge signals if nasal pressure swings were absent because of oral breathing), and (2) an amplitude reduction of a valid measure of breathing during sleep that did not reach 50% but was associated with either an oxygen desaturation of 3% or an arousal. 10 The apnea/hypopnea index (AHI) was calculated as the number of such events per hour. Agreement of the 2 methods was assessed by comparing AHI indices according to Bland and Altman. 16 Because 2 different recording systems were used (Vangard or Alice 4 and Holter-DR), time correlation was not precise within seconds. Therefore, the reliability of the method over the entire sleep study duration was assessed by comparing 5-minute epochs with any events as well as the number of apnea/hypopnea events per 5-minute epoch. In addition, an event-by-event analysis was performed to compare the transthoracic impedance signal to a single-channel reference signal (nasal pressure only) without use of the oxygen saturation criterion. Only tracings from 1 institution experienced in validation of nasal pressure recordings were of sufficient quality for this single-channel analysis. 14 For both methods, events were defined by a decrease in amplitude of at least 50% and were considered identical when occurring within a 15-second window. Statistical Analysis Continuous variables are expressed as mean SD. Differences on AHI scorings between the 2 methods were assessed according to Bland and Altman. 16 Agreement of the 2 methods was assessed by Cohen s statistics. Epoch-by-epoch apnea/hypopnea scores from these 2 different methods were compared by intraclass correlation coefficients. Receiver operating characteristic (ROC) curves were used to test the diagnostic accuracy of impedance signal analysis for the detection of sleep apnea syndrome at various AHI cutoffs. 14 Statistical significance was assumed at a value of P All analyses were performed by use of a commercially available statistical package (SPSS for Windows 9.0). Results The mean polysomnographic recording duration was hours. The mean sleep time was hours. Visual analysis of impedance signal waveforms was feasible in all patients (Figure 1) for hours. However, in 2 patients, telemetry errors led to technical loss of maximal 40% of total time. The patients had a wide range of AHI, from 0 to 79.2 h 1 (mean, h 1 ) and reported an Epworth sleepiness score of (Table 1). A significant number of apnea/ hypopnea events (AHI 5/min) was found in 17 of 22 (80%) patients and was correctly diagnosed in 16 of 17 (94%) on impedance signal scoring (Table 2). The diagnosis of moderate to severe SDB (AHI 20 h 1 ) was made in 12 patients (45%) and was identified with 100% sensitivity and specificity on impedance signal scoring (Figure 2). Among these 12 patients, 10 were newly diagnosed with SDB and 4 had not reported typical symptoms. The severity of SDB was not related to the degree of left ventricular dysfunction (Table 2). Comparison of the Two Methods Compared with clinical polysomnographic scoring, the analysis of the transthoracic impedance signal waveform showed a significant correlation of AHI (Pearson s r 0.869, P 0.001, Figure 2). The difference of AHI between the 2 methods (ie, the bias) was on average h 1 (P 0.05) and was increasing with higher AHI (Figure 3). The limits of agreement of AHI (2 SD of the mean difference) were 20.8 h 1, but when the 3 patients with the highest AHI by polysomnography (79, 57, and 46 h 1 ) were omitted from the analysis, the limits of agreement were h 1. Diagnostic accuracy at various cutoff levels was evaluated by ROC. The area under the curve for detection of an AHI of

3 2564 Circulation October 26, 2004 Figure 1. Visualization of first derivative of transthoracic impedance signal measured by pacemaker minute ventilation sensor (top row). This analog waveform is synchronized to simultaneous polysomnographic recordings. Note decrease in pacemaker transthoracic impedance signal waveform in both events (central top, mixed with preservation of thoracic movements to some degree, bottom). Impedance signal. SaO 2 indicates oxygen saturation; HR, heart rate. 5 h 1 was , P The corresponding areas for an AHI of 10 h 1 and 15 h 1 were both , P For the cutoff level of AHI 20 h 1, all patients were correctly classified by the pacemaker (ROC 1.0, 100% specificity and sensitivity). A total of 1637 five-minute epochs were analyzed in all patients over the entire recording time. Of those, 803 epochs included at least 1 apnea/hypopnea event on PSG. The transthoracic impedance signal correctly identified 724 epochs (90%), accounting for a sensitivity of 82 21%, a specificity of 74 18%, and a positive and negative predictive value of 76%. The measure of agreement on Cohen s statistics was (P 0.001). The intraclass correlation coefficient of AHI scores per 5-minute epoch was 0.73 (P 0.001). This suggests that 73% of the AHI variation per 5-minute epoch on impedance scoring was related to variation in the reference polysomnographic scoring. On the basis of polysomnographic findings, the differentiation between predominantly obstructive versus central type of SDB was made in 15 patients (Table 2). Identification of apnea/hypopnea events on transthoracic impedance analysis was feasible for both central and obstructive events; however, the differentiation between the 2 types of events was not possible. When matching apnea/hypopnea events per 5-minute epoch, the transthoracic impedance analysis was more sensitive to central than obstructive epochs (93% versus 83%, P NS) but with a lower specificity (67% versus 71%, P NS). Events were more likely to be overscored in central

4 Scharf et al Diagnosis of Sleep-Related Breathing Disorders 2565 TABLE 2. Detection of Hypopnea/Apnea on Pacemaker Impedance Signal Analysis Compared With Polysomnographic Recordings No. Institution Duration, h AHI PSG AHI PM Difference in AHI Predominant Pattern BMI, kg/m 2 EF Indication for PM Thoracotomy 1 B Obstructive NA NA* SND 2 B Obstructive AVB 3 A Central SND AVB 4 B Obstructive SND 5 A Central AVB 6 B Obstructive SND 7 B Obstructive SND 8 B Obstructive AVB 9 B Obstructive AVB 10 B Obstructive AVB 11 A Central SND 12 A Obstructive AVB 13 A Obstructive AVB 14 A SND AVB 15 B NA 0.25 SND 16 A Central AVB 17 B SND 18 B SND 19 A SND 20 B SND 21 B SND 22 A AVB Average SD PSG indicates polysomnography; BMI, body mass index; PM, pacemaker; AHI, apnea/hypopnea index; SND, sinus node dysfunction; AVB, atrioventricular block; and EF, ejection fraction. *No EF available; no history of heart disease other than SND. SDB and underscored in obstructive SDB (mean AHI difference, versus , P 0.05). The lowest correlation was found in patient 19. This patient had emphysema with a predominantly abdominal breathing pattern while the atrial lead was used for minute ventilation sensing. Another problem was abrupt changes in the signal-to-noise ratio during changes of sleeping position to a certain posture, such as the lateral decubitus position on the side of the device implant. In 2 patients, the impedance signal was often interrupted because of telemetry errors caused by body movements, and 40% of their sleep study had to be excluded. The accuracy of the transthoracic impedance signal analysis was not related to the body mass index of the patients or the electrode impedance of the sensing pacemaker lead. To compare impedance tracings to a single-channel respiration tracing without use of the oxygen saturation criterion, the identification and matching of single events was performed referencing to nasal airflow only. Because of technical reasons, this comparison could be performed only at 1 institution. Of 1276 events on nasal pressure tracings, 1033 (81%) were identified within 15 seconds on impedance tracings (223 were false-positive and 5775 were true-negative 30-second intervals; sensitivity, 81%; specificity, 52%; positive predictive value, 82%; negative predictive value, 95%), leading to an agreement of 0.77 on statistics. Discussion Main Findings This study demonstrates the usefulness of the transthoracic impedance signal recorded by pacemaker minute ventilation sensors to detect pauses and transient reductions in ventilation during sleep. The apnea/hypopnea events identified by analysis of the pacemaker signal led to the identification of 94% of patients with an AHI 5 and to the correct diagnosis of moderate to severe sleep apnea (AHI 20 h 1 ) with 100% sensitivity and specificity. The correlation between AHI derived from impedance signal analysis and polysomnography was 0.87 (P 0.001, Figure 2), and precision decreased only in 3 patients with AHI 40. Excluding these patients, the bias of AHI was 0 12 h 1. The measure of agreement between the 2 methods was 0.7 on statistics, which is similar to other home-based PSG recording systems and satisfactory for clinical screening purposes and for monitoring treatment effects. The intraclass correlation coefficient of 0.73 indicates that 73% variation in apnea/hypopnea scores per 5-minute epoch during the entire sleep study was related to variation in reference PSG scores.

5 2566 Circulation October 26, 2004 Figure 3. Bland-Altman plot of difference between polysomnographic scorings and impedance signal scorings, mean, /h, P 0.05; limits of agreement (2 SD), 20.8/h. Discrepancy was related primarily to a lower precision of pacemaker for 3 patients with very high AHI 40/h. Figure 2. Correlation plot of AHI scores on both methods. Pearson s correlation coefficient, 0.896, P 0.001; slope of theoretical curve, 0.82, P 0.005; 95% CI indicated. Technical Aspects Analog waveforms of transthoracic impedance signals measured by pacemaker minute ventilation sensors reflect changes in thoracic geometry and fluid/gas volume relationship similar to impedance pneumography. 8 The technical problems of impedance pneumography include changes in skin electrical conductivity. 9 The pacemaker has the advantage of using stable internal electrodes with a constant surface, and its correlation to measured ventilation has been excellent. 10 A sampling rate of 16 Hz allows adequate tracking of respiration on an analog waveform. The raw signal reflecting changes in intrathoracic gas volume is a linear dimension, whereas the first derivative would be a surrogate of a flow signal, ie, of changes in volume over time reflecting ventilation, which was preferable for visual analysis. Transthoracic impedance signals are not a surrogate of respiratory inductive plethysmography by thoracic or abdominal sensors but rather of ventilation itself. Therefore, the differentiation between central and obstructive events cannot be made, because the ventilation is decreased in both. For the same reason, however, detection of SDB was possible in both patients with predominantly central and obstructive SDB. The results of an event-by-event comparison referencing to nasal pressure tracings showed that more than 80% of episodes identified by both methods were identical. Thus, a declining amplitude in impedance signal waveforms actually corresponds to a simultaneous decrease in ventilation visualized on nasal airflow. Clinical Implications This method could be used for screening and monitoring sleep-related breathing disorders in patients having a rateresponsive pacemaker with a minute ventilation sensor. The ROC curves starting at 0.95 for AHI 5 h 1 and reaching 1.0 for AHI 20 h 1 indicate sufficient accuracy in detecting and excluding SDB. The development of automatic detection algorithms will allow screening of pacemaker patients, which needs to be tested in larger patient series. Because the transthoracic impedance signal is measured constantly when the minute ventilation sensor is programmed on, SDB might be detected incidentally during routine pacemaker follow-up. When monitoring of sleep-related breathing disorders becomes feasible in pacemaker patients, the therapeutic effect of different atrial pacing algorithms can be assessed. 11 Finally, in patients with congestive heart failure and intracardiac defibrillators or pacemakers, periodic respiration during daytime (Cheyne-Stokes respiration) could be detected, providing important independent prognostic information. 20 This monitoring could be used to guide heart failure therapy either by continuous positive airway pressure 21 or, more elegantly, by atrial and biventricular pacing algorithms. Thus, the present study suggests the potential for novel applications of pacemaker technology in screening, monitoring, and treatment of sleep-related breathing disorders. Limitations The main limitation of this feasibility study is the small number of patients with a high prevalence of SDB. Larger studies are under way to confirm these results by use of automatic download and detection algorithms. We did not differentiate between hypopnea and apnea events, because the physiological consequences are similar. Impedance waveforms were scored false-positive in central events or periodic respiration without desaturation, when the patient was lying on the side of pacemaker implant, or in abdominal breathing. These effects, however, were discrete, and major discrepancies between the 2 methods were found only at AHI scores 40 h 1. In patients with predominantly abdominal breathing pattern, such as emphysema (patient 19) or previous sternotomy, an impedance signal obtained from the ventricular instead of the atrial lead might be preferable, because the location is closer to the diaphragm. Abrupt changes in signal-to-noise ratio were found when changes in sleeping position directed ventilation to the lung opposite to the pacemaker implant side. We also encountered this problem

6 Scharf et al Diagnosis of Sleep-Related Breathing Disorders 2567 when calibrating the minute ventilation sensor in different postures. 22 An autocalibration algorithm guided by the activity sensor might overcome this issue. In this study, we analyzed only effective sleep time diagnosed on polysomnography. In clinical application, sleep time might not be directly available from pacemaker information and might have to be retrieved from the patient s diary. Alternatively, we have proposed an algorithm using mechanical activity variance for the detection of sleep, which could be used for automatic detection of resting periods. 23 Technical Limitations Because we used 2 different recording systems, the polysomnograph and the DR 180 Holter, time synchronization was performed manually. Major time correlation errors were generated by data filtering and processing in 2 different systems at 1 institution. However, matching the number of episodes per 5-minute epoch over the entire registration period, which obtained similar results from both institutions, compensated for this issue. The cardiac activity on transthoracic impedance signals can be seen in Figure 1 and needs to be filtered for automatic detection. In 2 patients, major parts of the recordings were excluded because of Holter telemetry errors caused by body movements. This problem will be resolved by future internal recording storage and download software. Conclusions Transthoracic impedance signals from pacemaker minute ventilation sensors can be used for detection and quantification of hypopnea/apnea events. The agreement with reference polysomnographic scorings is promising for future application in screening and monitoring of sleep-related breathing disorders. This study opens a wide field of applications for future pacemaker generations in various groups of patients. Disclosure Christoph Scharf has received research support from Medtronic; Yong K. Cho is an employee of Medtronic; and Bruce L. Wilkoff has received honoraria for speaking engagements and research for and is an advisor to Medtronic. References 1. Young T, Palta M, Dempsey J, et al. The occurrence of sleep-disordered breathing among middle-aged adults. N Engl J Med. 1993;328: Javaheri S, Parker TJ, Liming JD, et al. Sleep apnea in 81 ambulatory male patients with stable heart failure: types and their prevalences, consequences, and presentations. Circulation. 1998;97: Quan SF, Gersh BJ. Cardiovascular consequences of sleep-disordered breathing: past, present and future: report of a workshop from the National Center on Sleep Disorders Research and the National Heart, Lung, and Blood Institute. Circulation. 2004;109: Peppard PE, Young T, Palta M, et al. Prospective study of the association between sleep-disordered breathing and hypertension. N Engl J Med. 2000;342: Mooe T, Franklin KA, Holmstrom K, et al. Sleep-disordered breathing and coronary artery disease: long-term prognosis. Am J Respir Crit Care Med. 2001;164: Schafer H, Koehler U, Ewig S, et al. Obstructive sleep apnea as a risk marker in coronary artery disease. Cardiology. 1999;92: Peker Y, Kraiczi H, Hedner J, et al. An independent association between obstructive sleep apnoea and coronary artery disease. Eur Respir J. 1999;14: Cooley WL, Longini RL. A new design for an impedance pneumograph. J Appl Physiol. 1968;25: Grenvik A, Ballou S, Mc Ginely E, et al. Impedance pneumography. Chest. 1972;62: Cole CR, Jensen DN, Cho Y, et al. Correlation of impedance minute ventilation with measured minute ventilation in a rate responsive pacemaker. Pacing Clin Electrophysiol. 2001;24: Garrigue S, Bordier P, Jais P, et al. Benefit of atrial pacing in sleep apnea syndrome. N Engl J Med. 2002;346: Garrigue S, Barold SS, Clementy J. Sleep apnea: a new indication for cardiac pacing? Pacing Clin Electrophysiol. 2004;27: Sleep-related breathing disorders in adults: recommendations for syndrome definition and measurement techniques in clinical research. Report of an American Academy of Sleep Medicine Task Force. Sleep. 1999;22: Thurnheer R, Xie X, Bloch KE. Accuracy of nasal cannula pressure recordings for assessment of ventilation during sleep. Am J Respir Crit Care Med. 2001;164: Bloch KE. Polysomnography: a systematic review. Technol Health Care. 1997;5: Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1: Ballester E, Solans M, Vila X, et al. Evaluation of a portable respiratory recording device for detecting apnoeas and hypopnoeas in subjects from a general population. Eur Respir J. 2000;16: Bar A, Pillar G, Dvir I, et al. Evaluation of a portable device based on peripheral arterial tone for unattended home sleep studies. Chest. 2003; 123: Ficker JH, Wiest GH, Wilpert J, et al. Evaluation of a portable recording device (Somnocheck) for use in patients with suspected obstructive sleep apnoea. Respiration. 2001;68: Leite JJ, Mansur AJ, de Freitas HF, et al. Periodic breathing during incremental exercise predicts mortality in patients with chronic heart failure evaluated for cardiac transplantation. J Am Coll Cardiol. 2003; 41: Sin DD, Logan AG, Fitzgerald FS, et al. Effects of continuous positive airway pressure on cardiovascular outcomes in heart failure patients with and without Cheyne-Stokes respiration. Circulation. 2000;102: Duru F, Radicke D, Wilkoff BL, et al. Influence of posture, breathing pattern, and type of exercise on minute ventilation estimation by a pacemaker transthoracic impedance sensor. Pacing Clin Electrophysiol. 2000;23: Duru F, Bloch KE, Weilenmann D, et al. Clinical evaluation of a pacemaker algorithm that adjusts the pacing rate during sleep using activity variance. Pacing Clin Electrophysiol. 2000;23:

Σύνδρομο σπνικής άπνοιας. Ποιός o ρόλος ηοσ ζηη γένεζη και ανηιμεηώπιζη ηων αρρσθμιών;

Σύνδρομο σπνικής άπνοιας. Ποιός o ρόλος ηοσ ζηη γένεζη και ανηιμεηώπιζη ηων αρρσθμιών; Σύνδρομο σπνικής άπνοιας. Ποιός o ρόλος ηοσ ζηη γένεζη και ανηιμεηώπιζη ηων αρρσθμιών; E.N. Σημανηηράκης MD, FESC Επίκ. Καθηγηηής Καρδιολογίας Πανεπιζηημιακό Νοζοκομείο Ηρακλείοσ Epidemiology 4% 2% 24%

More information

ACCURACY OF NASAL CANNULA PRESSURE RECORDINGS FOR ASSESSMENT OF VENTILATION DURING SLEEP

ACCURACY OF NASAL CANNULA PRESSURE RECORDINGS FOR ASSESSMENT OF VENTILATION DURING SLEEP Online Supplement for: ACCURACY OF NASAL CANNULA PRESSURE RECORDINGS FOR ASSESSMENT OF VENTILATION DURING SLEEP METHODS Evaluation of Impaired Nasal Ventilation Impaired nasal breathing, as perceived subjectively,

More information

In-Patient Sleep Testing/Management Boaz Markewitz, MD

In-Patient Sleep Testing/Management Boaz Markewitz, MD In-Patient Sleep Testing/Management Boaz Markewitz, MD Objectives: Discuss inpatient sleep programs and if they provide a benefit to patients and sleep centers Identify things needed to be considered when

More information

Assessment of a wrist-worn device in the detection of obstructive sleep apnea

Assessment of a wrist-worn device in the detection of obstructive sleep apnea Sleep Medicine 4 (2003) 435 442 Original article Assessment of a wrist-worn device in the detection of obstructive sleep apnea Najib T. Ayas a,b,c, Stephen Pittman a,c, Mary MacDonald c, David P. White

More information

Obstructive sleep apnoea How to identify?

Obstructive sleep apnoea How to identify? Obstructive sleep apnoea How to identify? Walter McNicholas MD Newman Professor in Medicine, St. Vincent s University Hospital, University College Dublin, Ireland. Potential conflict of interest None Obstructive

More information

DECLARATION OF CONFLICT OF INTEREST

DECLARATION OF CONFLICT OF INTEREST DECLARATION OF CONFLICT OF INTEREST Obstructive sleep apnoea How to identify? Walter McNicholas MD Newman Professor in Medicine, St. Vincent s University Hospital, University College Dublin, Ireland. Potential

More information

A Deadly Combination: Central Sleep Apnea & Heart Failure

A Deadly Combination: Central Sleep Apnea & Heart Failure A Deadly Combination: Central Sleep Apnea & Heart Failure Sanjaya Gupta, MD FACC FHRS Ohio State University Symposium May 10 th, 2018 Disclosures Boston Scientific: fellowship support, speaking honoraria

More information

A 74-year-old man with severe ischemic cardiomyopathy and atrial fibrillation

A 74-year-old man with severe ischemic cardiomyopathy and atrial fibrillation 1 A 74-year-old man with severe ischemic cardiomyopathy and atrial fibrillation The following 3 minute polysomnogram (PSG) tracing was recorded in a 74-year-old man with severe ischemic cardiomyopathy

More information

Diagnostic Accuracy of the Multivariable Apnea Prediction (MAP) Index as a Screening Tool for Obstructive Sleep Apnea

Diagnostic Accuracy of the Multivariable Apnea Prediction (MAP) Index as a Screening Tool for Obstructive Sleep Apnea Original Article Diagnostic Accuracy of the Multivariable Apnea Prediction (MAP) Index as a Screening Tool for Obstructive Sleep Apnea Ahmad Khajeh-Mehrizi 1,2 and Omid Aminian 1 1. Occupational Sleep

More information

Web-Based Home Sleep Testing

Web-Based Home Sleep Testing Editorial Web-Based Home Sleep Testing Authors: Matthew Tarler, Ph.D., Sarah Weimer, Craig Frederick, Michael Papsidero M.D., Hani Kayyali Abstract: Study Objective: To assess the feasibility and accuracy

More information

Basics of Polysomnography. Chitra Lal, MD, FCCP, FAASM Assistant professor of Medicine, Pulmonary, Critical Care and Sleep, MUSC, Charleston, SC

Basics of Polysomnography. Chitra Lal, MD, FCCP, FAASM Assistant professor of Medicine, Pulmonary, Critical Care and Sleep, MUSC, Charleston, SC Basics of Polysomnography Chitra Lal, MD, FCCP, FAASM Assistant professor of Medicine, Pulmonary, Critical Care and Sleep, MUSC, Charleston, SC Basics of Polysomnography Continuous and simultaneous recording

More information

The most accurate predictors of arterial hypertension in patients with Obstructive Sleep Apnea Syndrome

The most accurate predictors of arterial hypertension in patients with Obstructive Sleep Apnea Syndrome The most accurate predictors of arterial hypertension in patients with Obstructive Sleep Apnea Syndrome Natsios Georgios University Hospital of Larissa, Greece Definitions Obstructive Sleep Apnea (OSA)

More information

Nasal pressure recording in the diagnosis of sleep apnoea hypopnoea syndrome

Nasal pressure recording in the diagnosis of sleep apnoea hypopnoea syndrome 56 Unité de Recherche, Centre de Pneumologie de l Hôpital Laval, Université Laval, Québec, Canada F Sériès I Marc Correspondence to: Dr F Sériès, Centre de Pneumologie, 2725 Chemin Sainte Foy, Sainte Foy

More information

Sleep and the Heart. Physiologic Changes in Cardiovascular Parameters during Sleep

Sleep and the Heart. Physiologic Changes in Cardiovascular Parameters during Sleep Sleep and the Heart Rami N. Khayat, MD Professor of Internal Medicine Medical Director, Department of Respiratory Therapy Division of Pulmonary, Critical Care and Sleep Medicine The Ohio State University

More information

Sleep and the Heart. Rami N. Khayat, MD

Sleep and the Heart. Rami N. Khayat, MD Sleep and the Heart Rami N. Khayat, MD Professor of Internal Medicine Medical Director, Department of Respiratory Therapy Division of Pulmonary, Critical Care and Sleep Medicine The Ohio State University

More information

Prediction of sleep-disordered breathing by unattended overnight oximetry

Prediction of sleep-disordered breathing by unattended overnight oximetry J. Sleep Res. (1999) 8, 51 55 Prediction of sleep-disordered breathing by unattended overnight oximetry L. G. OLSON, A. AMBROGETTI ands. G. GYULAY Discipline of Medicine, University of Newcastle and Sleep

More information

Non-contact Screening System with Two Microwave Radars in the Diagnosis of Sleep Apnea-Hypopnea Syndrome

Non-contact Screening System with Two Microwave Radars in the Diagnosis of Sleep Apnea-Hypopnea Syndrome Medinfo2013 Decision Support Systems and Technologies - II Non-contact Screening System with Two Microwave Radars in the Diagnosis of Sleep Apnea-Hypopnea Syndrome 21 August 2013 M. Kagawa 1, K. Ueki 1,

More information

Sleep Disordered Breathing: Beware Snoring! Dr T A McDonagh Consultant Cardiologist Royal Brompton Hospital London. UK

Sleep Disordered Breathing: Beware Snoring! Dr T A McDonagh Consultant Cardiologist Royal Brompton Hospital London. UK Sleep Disordered Breathing: Beware Snoring! Dr T A McDonagh Consultant Cardiologist Royal Brompton Hospital London. UK Sleep Disordered Breathing in CHF Erratic breathing during sleep known for years e.g.

More information

(To be filled by the treating physician)

(To be filled by the treating physician) CERTIFICATE OF MEDICAL NECESSITY TO BE ISSUED TO CGHS BENEFICIAREIS BEING PRESCRIBED BILEVEL CONTINUOUS POSITIVE AIRWAY PRESSURE (BI-LEVEL CPAP) / BI-LEVEL VENTILATORY SUPPORT SYSTEM Certification Type

More information

Simple diagnostic tools for the Screening of Sleep Apnea in subjects with high risk of cardiovascular disease

Simple diagnostic tools for the Screening of Sleep Apnea in subjects with high risk of cardiovascular disease Cardiovascular diseases remain the number one cause of death worldwide Simple diagnostic tools for the Screening of Sleep Apnea in subjects with high risk of cardiovascular disease Shaoguang Huang MD Department

More information

Chronic NIV in heart failure patients: ASV, NIV and CPAP

Chronic NIV in heart failure patients: ASV, NIV and CPAP Chronic NIV in heart failure patients: ASV, NIV and CPAP João C. Winck, Marta Drummond, Miguel Gonçalves and Tiago Pinto Sleep disordered breathing (SDB), including OSA and central sleep apnoea (CSA),

More information

Symposium 27. Protagonist. sleep disordered breathing ,3 MV TV RR TV. Ela Talent 3 3 CPAP. Therapeutic Research vol. 28 no.

Symposium 27. Protagonist. sleep disordered breathing ,3 MV TV RR TV. Ela Talent 3 3 CPAP. Therapeutic Research vol. 28 no. Symposium 27 Protagonist sleep disordered breathing SDB 1 5 4 2 1 50 80 2,3 4 CPAP 1 5 1 MV TV RR TV Ela Talent 3 3 1 119 Symposium 27 2 10 60 2 8 10 60 50 AHI apnea hypopnea index AHI r 0.869 6 Garrigue

More information

sleepview by midmark Home Sleep Test

sleepview by midmark Home Sleep Test sleepview by midmark HST Home Sleep Test Introducing SleepView. Better for your patients, designed for your practice. Home sleep testing (HST) brings the diagnosis and management of OSA to the front lines

More information

Contact-free Monitoring Technology for Screening of sleep

Contact-free Monitoring Technology for Screening of sleep Contact-free Monitoring Technology for Screening of sleep Tal A, Goldbart A, Yizraeli-Davidovich M, Shinar Z Soroka University Medical Center Faculty of Health Sciences, Ben-Gurion University of the Negev

More information

The AASM Manual for the Scoring of Sleep and Associated Events

The AASM Manual for the Scoring of Sleep and Associated Events The AASM Manual for the Scoring of Sleep and Associated Events The 2007 AASM Scoring Manual vs. the AASM Scoring Manual v2.0 October 2012 The American Academy of Sleep Medicine (AASM) is committed to ensuring

More information

NATIONAL COMPETENCY SKILL STANDARDS FOR PERFORMING POLYSOMNOGRAPHY/SLEEP TECHNOLOGY

NATIONAL COMPETENCY SKILL STANDARDS FOR PERFORMING POLYSOMNOGRAPHY/SLEEP TECHNOLOGY NATIONAL COMPETENCY SKILL STANDARDS FOR PERFORMING POLYSOMNOGRAPHY/SLEEP TECHNOLOGY Polysomnography/Sleep Technology providers practice in accordance with the facility policy and procedure manual which

More information

SLEEP DISORDERED BREATHING The Clinical Conditions

SLEEP DISORDERED BREATHING The Clinical Conditions SLEEP DISORDERED BREATHING The Clinical Conditions Robert G. Hooper, M.D. In the previous portion of this paper, the definitions of the respiratory events that are the hallmarks of problems with breathing

More information

In 1994, the American Sleep Disorders Association

In 1994, the American Sleep Disorders Association Unreliability of Automatic Scoring of MESAM 4 in Assessing Patients With Complicated Obstructive Sleep Apnea Syndrome* Fabio Cirignotta, MD; Susanna Mondini, MD; Roberto Gerardi, MD Barbara Mostacci, MD;

More information

Portable Computerized Polysomnography in Attended and Unattended Settings*

Portable Computerized Polysomnography in Attended and Unattended Settings* Portable Computerized Polysomnography in Attended and Unattended Settings* Ivanka J. Mykytyn; Dimitar Sajkov, MBBS, PhD; Alister M. Neill, MBBS, FRACP; and R. Douglas McEvoy, MBBS, MD, FRACP Study objective:

More information

Novel pathophysiological concepts for the development and impact of sleep apnea in CHF.

Novel pathophysiological concepts for the development and impact of sleep apnea in CHF. Olaf Oldenburg Novel pathophysiological concepts for the development and impact of sleep apnea in CHF. Sleep apnea the need to synchronize the heart, the lung and the brain. Heart Failure 2011 Gothenburg,

More information

Opioids Cause Central and Complex Sleep Apnea in Humans and Reversal With Discontinuation: A Plea for Detoxification

Opioids Cause Central and Complex Sleep Apnea in Humans and Reversal With Discontinuation: A Plea for Detoxification pii: jc-16-00020 http://dx.doi.org/10.5664/jcsm.6628 CASE REPORTS Opioids Cause Central and Complex Sleep Apnea in Humans and Reversal With Discontinuation: A Plea for Detoxification Shahrokh Javaheri,

More information

Heart Failure and Sleep Disordered Breathing (SDB) Unhappy Bedfellows

Heart Failure and Sleep Disordered Breathing (SDB) Unhappy Bedfellows Question Heart Failure and Sleep Disordered Breathing (SDB) Unhappy Bedfellows 1 ResMed 2012 07 2 ResMed 2012 07 Open Airway 3 ResMed 2012 07 Flow Limitation Snore 4 ResMed 2012 07 Apnoea 5 ResMed 2012

More information

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms

Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms Automatic Identification of Implantable Cardioverter-Defibrillator Lead Problems Using Intracardiac Electrograms BD Gunderson, AS Patel, CA Bounds Medtronic, Inc., Minneapolis, USA Abstract Implantable

More information

Evaluation of a 2-Channel Portable Device and a Predictive Model to Screen for Obstructive Sleep Apnea in a Laboratory Environment

Evaluation of a 2-Channel Portable Device and a Predictive Model to Screen for Obstructive Sleep Apnea in a Laboratory Environment Evaluation of a 2-Channel Portable Device and a Predictive Model to Screen for Obstructive Sleep Apnea in a Laboratory Environment Jianyin Zou, Lili Meng MD, Yupu Liu MSc, Xiaoxi Xu, Suru Liu PhD, Jian

More information

Mario Kinsella MD FAASM 10/5/2016

Mario Kinsella MD FAASM 10/5/2016 Mario Kinsella MD FAASM 10/5/2016 Repetitive episodes of apnea or reduced airflow Due to upper airway obstruction during sleep Patients often obese Often have hypertension or DM 1 Obstructive apneas, hypopneas,

More information

Automated analysis of digital oximetry in the diagnosis of obstructive sleep apnoea

Automated analysis of digital oximetry in the diagnosis of obstructive sleep apnoea 302 Division of Respiratory Medicine, Department of Medicine, University of Calgary, Calgary, Alberta, Canada T2N 4N1 J-C Vázquez W H Tsai W W Flemons A Masuda R Brant E Hajduk W A Whitelaw J E Remmers

More information

Introducing the WatchPAT 200 # 1 Home Sleep Study Device

Introducing the WatchPAT 200 # 1 Home Sleep Study Device Introducing the WatchPAT 200 # 1 Home Sleep Study Device Top 10 Medical Innovation for 2010 Cleveland Clinic Fidelis Diagnostics & Itamar Medical Fidelis Diagnostics founded in 2004, is a privately-held

More information

The recommended method for diagnosing sleep

The recommended method for diagnosing sleep reviews Measuring Agreement Between Diagnostic Devices* W. Ward Flemons, MD; and Michael R. Littner, MD, FCCP There is growing interest in using portable monitoring for investigating patients with suspected

More information

Overdrive atrial pacing does not improve obstructive sleep apnoea syndrome

Overdrive atrial pacing does not improve obstructive sleep apnoea syndrome Eur Respir J 2005; 25: 343 347 DOI: 10.1183/09031936.05.00132703 CopyrightßERS Journals Ltd 2005 Overdrive atrial pacing does not improve obstructive sleep apnoea syndrome J-L. Pépin*,#, P. Defaye ", S.

More information

Biomedical Instrumentation

Biomedical Instrumentation Biomedical Instrumentation Prof. Dr. Nizamettin AYDIN naydin@yildiz.edu.tr naydin@ieee.org http://www.yildiz.edu.tr/~naydin Therapeutic and Prosthetic Devices 1 Figure 13.1 Block diagram of an asynchronous

More information

Prevalence of Sleep Disordered Breathing in Congestive Heart Failure as Determined by ApneaLink, a Simplified Screening Device

Prevalence of Sleep Disordered Breathing in Congestive Heart Failure as Determined by ApneaLink, a Simplified Screening Device Prevalence of Sleep Disordered Breathing in Congestive Heart Failure as Determined by ApneaLink, a Simplified Screening Device Susan R. Isakson, BS, 1 Jennifer Beede, BS, 1 Kevin Jiang, BS, 1 Nancy J.

More information

Polysomnography (PSG) (Sleep Studies), Sleep Center

Polysomnography (PSG) (Sleep Studies), Sleep Center Policy Number: 1036 Policy History Approve Date: 07/09/2015 Effective Date: 07/09/2015 Preauthorization All Plans Benefit plans vary in coverage and some plans may not provide coverage for certain service(s)

More information

During the therapeutic titration of nasal continuous

During the therapeutic titration of nasal continuous Cardiogenic Oscillations on the Airflow Signal During Continuous Positive Airway Pressure as a Marker of Central Apnea* Indu Ayappa, PhD; Robert G. Norman, MS, RRT; and David M. Rapoport, MD, FCCP Therapeutic

More information

Respiratory Inductance Plethysmography Improved Diagnostic Sensitivity and Specificity of Obstructive Sleep Apnea

Respiratory Inductance Plethysmography Improved Diagnostic Sensitivity and Specificity of Obstructive Sleep Apnea Respiratory Inductance Plethysmography Improved Diagnostic Sensitivity and Specificity of Obstructive Sleep Apnea Dmitriy Kogan MD, Arad Jain, Shawn Kimbro RPSGT, Guillermo Gutierrez MD PhD, and Vivek

More information

Sleep and the Heart Reversing the Effects of Sleep Apnea to Better Manage Heart Disease

Sleep and the Heart Reversing the Effects of Sleep Apnea to Better Manage Heart Disease 1 Sleep and the Heart Reversing the Effects of Sleep Apnea to Better Manage Heart Disease Rami Khayat, MD Professor of Internal Medicine Director, OSU Sleep Heart Program Medical Director, Department of

More information

Κλινικό Φροντιστήριο Αναγνώριση και καταγραφή αναπνευστικών επεισοδίων Λυκούργος Κολιλέκας Επιμελητής A ΕΣΥ 7η Πνευμονολογική Κλινική ΝΝΘΑ Η ΣΩΤΗΡΙΑ

Κλινικό Φροντιστήριο Αναγνώριση και καταγραφή αναπνευστικών επεισοδίων Λυκούργος Κολιλέκας Επιμελητής A ΕΣΥ 7η Πνευμονολογική Κλινική ΝΝΘΑ Η ΣΩΤΗΡΙΑ Κλινικό Φροντιστήριο Αναγνώριση και καταγραφή αναπνευστικών επεισοδίων Λυκούργος Κολιλέκας Επιμελητής A ΕΣΥ 7 η Πνευμονολογική Κλινική ΝΝΘΑ Η ΣΩΤΗΡΙΑ SCORING SLEEP -Rechtschaffen and Kales (1968) - AASM

More information

Validation of a Self-Applied Unattended Monitor for Sleep Disordered Breathing

Validation of a Self-Applied Unattended Monitor for Sleep Disordered Breathing Scientific investigations Validation of a Self-Applied Unattended Monitor for Sleep Disordered Breathing Indu Ayappa, Ph.D.; Robert G. Norman, Ph.D.; Vijay Seelall, M.D.; David M. Rapoport, M.D. Division

More information

Edoardo Gronda UO cardiologia e Ricerca Dipartimento Cardiovascolare IRCCS MultiMedica

Edoardo Gronda UO cardiologia e Ricerca Dipartimento Cardiovascolare IRCCS MultiMedica Convegno Pneumologia 2016 Milano 16-18 giugno 2016 Centro Congressi Palazzo delle Stelline Edoardo Gronda UO cardiologia e Ricerca Dipartimento Cardiovascolare IRCCS MultiMedica Central apnea 10 second

More information

Key words: Medicare; obstructive sleep apnea; oximetry; sleep apnea syndromes

Key words: Medicare; obstructive sleep apnea; oximetry; sleep apnea syndromes Choice of Oximeter Affects Apnea- Hypopnea Index* Subooha Zafar, MD; Indu Ayappa, PhD; Robert G. Norman, PhD; Ana C. Krieger, MD, FCCP; Joyce A. Walsleben, PhD; and David M. Rapoport, MD, FCCP Study objectives:

More information

Martin R Cowie Professor of Cardiology, National Heart & Lung Institute Imperial College London (Royal Brompton Hospital)

Martin R Cowie Professor of Cardiology, National Heart & Lung Institute Imperial College London (Royal Brompton Hospital) Treatment of Sleep-Disordered Breathing With Predominant Central Sleep Apnoea by Adaptive Servo Ventilation in Patients With Heart Failure and Reduced Ejection Fraction (SERVE-HF) Martin R Cowie Professor

More information

Diabetes & Obstructive Sleep Apnoea risk. Jaynie Pateraki MSc RGN

Diabetes & Obstructive Sleep Apnoea risk. Jaynie Pateraki MSc RGN Diabetes & Obstructive Sleep Apnoea risk Jaynie Pateraki MSc RGN Non-REM - REM - Both - Unrelated - Common disorders of Sleep Sleep Walking Night terrors Periodic leg movements Sleep automatism Nightmares

More information

PVDOMICS. Sleep Core. Cleveland Clinic Cleveland, Ohio

PVDOMICS. Sleep Core. Cleveland Clinic Cleveland, Ohio PVDOMICS Sleep Core Rawan Nawabit, Research Coordinator and Polysomnologist Joan Aylor, Research Coordinator Dr. Reena Mehra, Co-Investigator, Sleep Core Lead Cleveland Clinic Cleveland, Ohio 1 Obstructive

More information

SOMNOcheck micro CARDIO. Simple Assessment of Cardiovascular Risk During Sleep. In sleep medicine In internal medicine In preventive medicine

SOMNOcheck micro CARDIO. Simple Assessment of Cardiovascular Risk During Sleep. In sleep medicine In internal medicine In preventive medicine SOMNOcheck micro CARDIO Simple Assessment of Cardiovascular Risk During Sleep In sleep medicine In internal medicine In preventive medicine SOMNOcheck micro CAR In sleep diagnostics we provide classic

More information

Coding for Sleep Disorders Jennifer Rose V. Molano, MD

Coding for Sleep Disorders Jennifer Rose V. Molano, MD Practice Coding for Sleep Disorders Jennifer Rose V. Molano, MD Accurate coding is an important function of neurologic practice. This section of is part of an ongoing series that presents helpful coding

More information

DRS ed Aritmie Cardiache Iper ed Ipocinetiche: la clinica

DRS ed Aritmie Cardiache Iper ed Ipocinetiche: la clinica Corso Multidisciplinare di Aggiornamento La Sindrome delle Apnee Notturne: una sfida diagnostico terapeutica DRS ed Aritmie Cardiache Iper ed Ipocinetiche: la clinica FRANCESCO PERNA, MD, PhD Laboratorio

More information

New Government O2 Criteria and Expert Panel. Jennifer Despain, RPSGT, RST, AS

New Government O2 Criteria and Expert Panel. Jennifer Despain, RPSGT, RST, AS New Government O2 Criteria and Expert Panel Jennifer Despain, RPSGT, RST, AS Lead Sleep Technologist, Central Utah Clinic Sleep Disorders Center; Provo, Utah Objectives: Review new government O2 criteria

More information

Validating the Watch-PAT for Diagnosing Obstructive Sleep Apnea in Adolescents

Validating the Watch-PAT for Diagnosing Obstructive Sleep Apnea in Adolescents pii: jc-18-00029 http://dx.doi.org/10.5664/jcsm.7386 SCIENTIFIC INVESTIGATIONS Validating the Watch-PAT for Diagnosing Obstructive Sleep Apnea in Adolescents Ji Ho Choi, MD, PhD 1 ; Bora Lee 2 ; Jae Yong

More information

The Latest Technology from CareFusion

The Latest Technology from CareFusion The Latest Technology from CareFusion Contents 1 Introduction... 2 1.1 Overview... 2 1.2 Scope... 2 2.1 Input Recordings... 2 2.2 Automatic Analysis... 3 2.3 Data Mining... 3 3 Results... 4 3.1 AHI comparison...

More information

Overnight fluid shifts in subjects with and without obstructive sleep apnea

Overnight fluid shifts in subjects with and without obstructive sleep apnea Original Article Overnight fluid shifts in subjects with and without obstructive sleep apnea Ning Ding 1 *, Wei Lin 2 *, Xi-Long Zhang 1, Wen-Xiao Ding 1, Bing Gu 3, Bu-Qing Ni 4, Wei Zhang 4, Shi-Jiang

More information

Internet Journal of Medical Update

Internet Journal of Medical Update Internet Journal of Medical Update 2009 July;4(2):24-28 Internet Journal of Medical Update Journal home page: http://www.akspublication.com/ijmu Original Work EEG arousal prediction via hypoxemia indicator

More information

Obstructive sleep apnea (OSA) is common but underdiagnosed

Obstructive sleep apnea (OSA) is common but underdiagnosed The Role of Single-Channel Nasal Airflow Pressure Transducer in the Diagnosis of OSA in the Sleep Laboratory Lydia Makarie Rofail, M.B.B.S., Ph.D. 1,2,3 ; Keith K. H. Wong, M.B.B.S., Ph.D. 1,2,4 ; Gunnar

More information

About VirtuOx. Was marketed exclusively by Phillips Healthcare division, Respironics for 3 years

About VirtuOx. Was marketed exclusively by Phillips Healthcare division, Respironics for 3 years About VirtuOx VirtuOx, Inc. assists physicians and Durable Medical Equipment (DME)( companies diagnose respiratory diseases and qualify patients for home respiratory equipment under the guidelines of CMS

More information

GOALS. Obstructive Sleep Apnea and Cardiovascular Disease (OVERVIEW) FINANCIAL DISCLOSURE 2/1/2017

GOALS. Obstructive Sleep Apnea and Cardiovascular Disease (OVERVIEW) FINANCIAL DISCLOSURE 2/1/2017 Obstructive Sleep Apnea and Cardiovascular Disease (OVERVIEW) 19th Annual Topics in Cardiovascular Care Steven Khov, DO, FAAP Pulmonary Associates of Lancaster, Ltd February 3, 2017 skhov2@lghealth.org

More information

Epidemiology and diagnosis of sleep apnea

Epidemiology and diagnosis of sleep apnea Epidemiology and diagnosis of sleep apnea Dr Raphael Heinzer, MD MPH Center for Investigation and research in Sleep Lausanne University Hospital (CHUV) Switzerland Joint annual meeting SSC/SSCS-SSP 2016

More information

Validating the Watch-PAT for Diagnosing Obstructive Sleep. Apnea in Adolescents

Validating the Watch-PAT for Diagnosing Obstructive Sleep. Apnea in Adolescents Validating the Watch-PAT for Diagnosing Obstructive Sleep Apnea in Adolescents Ji Ho Choi, MD, PhD 1 ; Bora Lee 2 ; Jae Yong Lee, MD, PhD 1 ; Hyun Jun Kim, MD, PhD 3 1 Department of Otorhinolaryngology-Head

More information

Sleep Bruxism and Sleep-Disordered Breathing

Sleep Bruxism and Sleep-Disordered Breathing Sleep Bruxism and Sleep-Disordered Breathing Author STEVEN D BENDER, DDS*, Associate Editor EDWARD J. SWIFT JR., DMD, MS Sleep bruxism (SB) is a repetitive jaw muscle activity with clenching or grinding

More information

Sleep Apnea: Vascular and Metabolic Complications

Sleep Apnea: Vascular and Metabolic Complications Sleep Apnea: Vascular and Metabolic Complications Vahid Mohsenin, M.D. Professor of Medicine Yale University School of Medicine Director, Yale Center for Sleep Medicine Definitions Apnea: Cessation of

More information

The Epworth Sleepiness Scale (ESS) was developed by Johns

The Epworth Sleepiness Scale (ESS) was developed by Johns Clinical Reproducibility of the Epworth Sleepiness Scale Anh Tu Duy Nguyen, M.D. 1 ; Marc A. Baltzan, M.D., M.Sc. 1,2 ; David Small, M.D. 1 ; Norman Wolkove, M.D. 1 ; Simone Guillon, M.D. 3 ; Mark Palayew,

More information

Frequency-domain Index of Oxyhemoglobin Saturation from Pulse Oximetry for Obstructive Sleep Apnea Syndrome

Frequency-domain Index of Oxyhemoglobin Saturation from Pulse Oximetry for Obstructive Sleep Apnea Syndrome Journal of Medical and Biological Engineering, 32(5): 343-348 343 Frequency-domain Index of Oxyhemoglobin Saturation from Pulse Oximetry for Obstructive Sleep Apnea Syndrome Liang-Wen Hang 1,2 Chen-Wen

More information

Sleep Apnea Syndrome in Patients with Atrial Fibrillation 2 Cases Whose Atrial Fibrillation Was Controlled by Treatment for Sleep Apnea Syndrome

Sleep Apnea Syndrome in Patients with Atrial Fibrillation 2 Cases Whose Atrial Fibrillation Was Controlled by Treatment for Sleep Apnea Syndrome Case Report Sleep Apnea Syndrome in Patients with Atrial Fibrillation 2 Cases Whose Atrial Fibrillation Was Controlled by Treatment for Sleep Apnea Syndrome Manabu Fujimoto MD, Yamamoto Masakazu MD Kouseiren

More information

Atrial Overdrive Pacing for the Obstructive Sleep Apnea Hypopnea Syndrome

Atrial Overdrive Pacing for the Obstructive Sleep Apnea Hypopnea Syndrome The new england journal of medicine original article Atrial Overdrive Pacing for the Obstructive Sleep Apnea Hypopnea Syndrome Emmanuel N. Simantirakis, M.D., Sophia E. Schiza, M.D., Stavros I. Chrysostomakis,

More information

TABLE OF CONTENTS. Physician's Manual. I. Device Description 2. II. Intended Use 5. III. Contraindications 5. IV. Warnings and Precautions 5

TABLE OF CONTENTS. Physician's Manual. I. Device Description 2. II. Intended Use 5. III. Contraindications 5. IV. Warnings and Precautions 5 TABLE OF CONTENTS SECTION Page I. Device Description 2 II. Intended Use 5 III. Contraindications 5 IV. Warnings and Precautions 5 V. Adverse Events 5 VI. Clinical Trials 5 VII. Patient Information 6 VIII.

More information

Effectiveness of Portable Monitoring Devices for Diagnosing Obstructive Sleep Apnea: Update of a Systematic Review

Effectiveness of Portable Monitoring Devices for Diagnosing Obstructive Sleep Apnea: Update of a Systematic Review Effectiveness of Portable Monitoring Devices for Diagnosing Obstructive Sleep Apnea: Update of a Systematic Review Submitted to: Agency for Healthcare Research and Quality 540 Gaither Road Rockville, Maryland

More information

The Familial Occurrence of Obstructive Sleep Apnoea Syndrome (OSAS)

The Familial Occurrence of Obstructive Sleep Apnoea Syndrome (OSAS) Global Journal of Respiratory Care, 2014, 1, 17-21 17 The Familial Occurrence of Obstructive Sleep Apnoea Syndrome (OSAS) Piotr Bielicki, Tadeusz Przybylowski, Ryszarda Chazan * Department of Internal

More information

Screening method for severe sleep-disordered breathing in hypertensive patients without daytime sleepiness

Screening method for severe sleep-disordered breathing in hypertensive patients without daytime sleepiness Journal of Cardiology (2009) 53, 79 85 ORIGINAL ARTICLE Screening method for severe sleep-disordered breathing in hypertensive patients without daytime sleepiness Taiji Furukawa (PhD) a,, Masaaki Suzuki

More information

Diagnostic capabilities of the implantable therapeutic systems

Diagnostic capabilities of the implantable therapeutic systems Cardiac pacing 2012 and beyound Monday August 27, 2012 Diagnostic capabilities of the implantable therapeutic systems Pekka Raatikainen Heart Center Co. Tampere University Hospital and University of Tampere

More information

Obstructive sleep apnea (OSA) is characterized by. Quality of Life in Patients with Obstructive Sleep Apnea*

Obstructive sleep apnea (OSA) is characterized by. Quality of Life in Patients with Obstructive Sleep Apnea* Quality of Life in Patients with Obstructive Sleep Apnea* Effect of Nasal Continuous Positive Airway Pressure A Prospective Study Carolyn D Ambrosio, MD; Teri Bowman, MD; and Vahid Mohsenin, MD Background:

More information

Image: Scott Leight, istockphoto

Image: Scott Leight, istockphoto Image: Scott Leight, istockphoto B.D. Kent 1,2 W.T. McNicholas 1,2 1 Pulmonary and Sleep Disorders Unit, St Vincent s University Hospital 2 School of Medicine and Medical Science, University College Dublin,

More information

FA et Apnée du Sommeil

FA et Apnée du Sommeil FA et Apnée du Sommeil La Réunion Octobre 2017 Pascal Defaye CHU Grenoble-Alpes Obstructive Sleep Apnea and AF Incidence of atrial fibrillation (AF), based on presence or absence of OSA. Cumulative frequency

More information

Index SLEEP MEDICINE CLINICS. Note: Page numbers of article titles are in boldface type.

Index SLEEP MEDICINE CLINICS. Note: Page numbers of article titles are in boldface type. 549 SLEEP MEDICINE CLINICS Sleep Med Clin 1 (2007) 549 553 Note: Page numbers of article titles are in boldface type. A Abdominal motion, in assessment of sleep-related breathing disorders, 452 454 Adherence,

More information

FAQ CODING & REIMBURSEMENT. WatchPAT TM Home Sleep Test

FAQ CODING & REIMBURSEMENT. WatchPAT TM Home Sleep Test FAQ CODING & REIMBURSEMENT WatchPAT TM Home Sleep Test TABLE OF CONTENTS PATIENT SELECTION CRITERIA 3 CODING & MODIFIERS 4-6 PLACE OF SERVICE 6 FREQUENCY 7 ACCREDITATION 7 SLEEP MEDICINE GLOSSARY AND ACRONYMS

More information

O bstructive sleep apnoea-hypopnoea (OSAH) is a highly

O bstructive sleep apnoea-hypopnoea (OSAH) is a highly 422 SLEEP-DISORDERED BREATHING Predictive value of automated oxygen saturation analysis for the diagnosis and treatment of obstructive sleep apnoea in a home-based setting V Jobin, P Mayer, F Bellemare...

More information

Split Night Protocols for Adult Patients - Updated July 2012

Split Night Protocols for Adult Patients - Updated July 2012 Split Night Protocols for Adult Patients - Updated July 2012 SUMMARY: Sleep technologists are team members who work under the direction of a physician practicing sleep disorders medicine. Sleep technologists

More information

The Effect of a Mandibular Advancement Device on Apneas and Sleep in Patients With Obstructive Sleep Apnea*

The Effect of a Mandibular Advancement Device on Apneas and Sleep in Patients With Obstructive Sleep Apnea* The Effect of a Mandibular Advancement Device on Apneas and Sleep in Patients With Obstructive Sleep Apnea* Marie Marklund, DDS; Karl A. Franklin, MD, PhD, FCCP; Carin Sahlin, RTA; and Rune Lundgren, MD,

More information

The AASM Manual for the Scoring of Sleep and Associated Events

The AASM Manual for the Scoring of Sleep and Associated Events The AASM Manual for the Scoring of Sleep and Associated Events Summary of Updates in Version 2.1 July 1, 2014 The American Academy of Sleep Medicine (AASM) is committed to ensuring that The AASM Manual

More information

Equivalence of Autoadjusted and Constant Continuous Positive Airway Pressure in Home Treatment of Sleep Apnea*

Equivalence of Autoadjusted and Constant Continuous Positive Airway Pressure in Home Treatment of Sleep Apnea* Original Research SLEEP MEDICINE Equivalence of Autoadjusted and Constant Continuous Positive Airway Pressure in Home Treatment of Sleep Apnea* Yvonne Nussbaumer, MD; Konrad E. Bloch, MD, FCCP; Therese

More information

Apnoeic and obstructive nonapnoeic sleep respiratory events

Apnoeic and obstructive nonapnoeic sleep respiratory events Eur Respir J 29; 34: 156 161 DOI: 1.1183/931936.1628 CopyrightßERS Journals Ltd 29 Apnoeic and obstructive nonapnoeic sleep respiratory events J.F. Masa*, **, J. Corral*, **, J. Teran #, **, M.J. Martin*,

More information

The STOP-Bang Equivalent Model and Prediction of Severity

The STOP-Bang Equivalent Model and Prediction of Severity DOI:.5664/JCSM.36 The STOP-Bang Equivalent Model and Prediction of Severity of Obstructive Sleep Apnea: Relation to Polysomnographic Measurements of the Apnea/Hypopnea Index Robert J. Farney, M.D. ; Brandon

More information

There have been many traffic accidents in our country, and the

There have been many traffic accidents in our country, and the http://dx.doi.org/1.5664/jcsm.2584 Reliability of the Watch-PAT 2 in Detecting Sleep Apnea in Highway bus Drivers Melike Yuceege, M.D. 1 ; Hikmet Firat, M.D. 1 ; Ahmet Demir, M.D. 2 ; Sadik Ardic, M.D.

More information

ONLINE DATA SUPPLEMENT. Impact of Obstructive Sleep Apnea on Left Ventricular Mass and. Diastolic Function

ONLINE DATA SUPPLEMENT. Impact of Obstructive Sleep Apnea on Left Ventricular Mass and. Diastolic Function ONLINE DATA SUPPLEMENT Impact of Obstructive Sleep Apnea on Left Ventricular Mass and Diastolic Function Mitra Niroumand Raffael Kuperstein Zion Sasson Patrick J. Hanly St. Michael s Hospital University

More information

Sleep Apnea and Heart Failure

Sleep Apnea and Heart Failure Sleep Apnea and Heart Failure Micha T. Maeder, MD Cardiology Division Kantonsspital St. Gallen Switzerland micha.maeder@kssg.ch Sleep Disordered Breathing (SDB) in HFrEF 700 HFrEF patients (LVEF

More information

SleepView. SleepView. Monitor + SleepViewSM. Portal Clinical Validation Summary. CliniCal validation

SleepView. SleepView. Monitor + SleepViewSM. Portal Clinical Validation Summary. CliniCal validation CliniCal validation Monitor + SM Portal Clinical Validation Summary 2007 AASM established guidelines for proper Portable Monitoring (PM) methodology to be used in home sleep apnea diagnosis Clinical Guidelines

More information

Sleep and the Heart. Sleep Stages. Sleep and the Heart: non REM 8/31/2016

Sleep and the Heart. Sleep Stages. Sleep and the Heart: non REM 8/31/2016 Sleep and the Heart Overview of sleep Hypertension Arrhythmias Ischemic events CHF Pulmonary Hypertension Cardiac Meds and Sleep Sleep Stages Non-REM sleep(75-80%) Stage 1(5%) Stage 2(50%) Stage 3-4*(15-20%)

More information

International Journal of Scientific & Engineering Research Volume 9, Issue 1, January ISSN

International Journal of Scientific & Engineering Research Volume 9, Issue 1, January ISSN International Journal of Scientific & Engineering Research Volume 9, Issue 1, January-2018 342 The difference of sleep quality between 2-channel ambulatory monitor and diagnostic polysomnography Tengchin

More information

Obstructive sleep apnea (OSA) is the periodic reduction

Obstructive sleep apnea (OSA) is the periodic reduction Obstructive Sleep Apnea and Oxygen Therapy: A Systematic Review of the Literature and Meta-Analysis 1 Department of Anesthesiology, Toronto Western Hospital, University Health Network, University of Toronto,

More information

Auto Servo Ventilation Indications, Basics of Algorithm, and Titration

Auto Servo Ventilation Indications, Basics of Algorithm, and Titration Auto Servo Ventilation Indications, Basics of Algorithm, and Titration 1 ASV Learning Objectives Understand the indications for Auto Servo Ventilation Differentiate obstructive versus central hypopneas

More information

The amount of nightly variability that occurs in sleepdisordered. Nightly variability of sleep-disordered breathing measured over 3 nights

The amount of nightly variability that occurs in sleepdisordered. Nightly variability of sleep-disordered breathing measured over 3 nights Nightly variability of sleep-disordered breathing measured over 3 nights CARL J. STEPNOWSKY, Jr, PHD, WILLIAM C. ORR, PHD, and TERENCE M. DAVIDSON, MD, San Diego, California, and Oklahoma City, Oklahoma

More information

CHANGING SHAPE OF SLEEP STUDIES

CHANGING SHAPE OF SLEEP STUDIES CHANGING SHAPE OF SLEEP STUDIES JAMES C. O BRIEN MD FCCP, FAASM MEDICAL DIRECTOR, PROHEALTH PHYSICIANS SLEEP CENTERS 10/19/2107 GOALS OF MY TALK-- Review the types of sleep studies Provide clinical information

More information

Effect of body mass index on overnight oximetry for the diagnosis of sleep apnea

Effect of body mass index on overnight oximetry for the diagnosis of sleep apnea Respiratory Medicine (2004) 98, 421 427 Effect of body mass index on overnight oximetry for the diagnosis of sleep apnea Hiroshi Nakano*, Togo Ikeda, Makito Hayashi, Etsuko Ohshima, Michiko Itoh, Nahoko

More information

Circadian Variations Influential in Circulatory & Vascular Phenomena

Circadian Variations Influential in Circulatory & Vascular Phenomena SLEEP & STROKE 1 Circadian Variations Influential in Circulatory & Vascular Phenomena Endocrine secretions Thermo regulations Renal Functions Respiratory control Heart Rhythm Hematologic parameters Immune

More information