1 DART MSI of drugs of abuse in hair Wilco Duvivier, Teris van Beek, Michel Nielen October 5, 2017

2 Forensic hair evidence Forensically interesting: Many compounds accumulate in hair Prolonged detectability versus body fluids Non-invasive sampling, easy storage Conventional hair analysis: + Sensitive LC- or GC-MS(/MS) analysis + Segmentation to estimate time of drug intake - Destructive and inefficient sample preparation Digestion and analysis - Only rough estimated time of drug intake (>1 month intervals) Moosmann et al., TIAFT2013 2

3 Hair scan concepts: hair scan What if we....could target intact locks of hair..would not need MALDI matrix or other sample preparation..could make hair analysis fast and easy to perform Ambient ionization mass spectrometry 3

4 DART ionization DART: Direct Analysis in Real Time Metastable species He(2 3 S) ºC Analyte Charged analyte ions He DART glow discharge Mass spectrometer sample Radostina Manova 4

5 DART hair scan method Intact lock of hair attached to stainless steel mesh Scanned at 0.2 mm/s, DART source at 250 C Orbitrap MS: ~ 1 scan per second with high mass resolution Ideal DART temperature setting for THC on hair 8x10 5 T actual = 135 C Average intensity THC signal 7x10 5 6x10 5 5x10 5 4x10 5 3x10 5 2x10 5 1x DART gas beam temperature setting ( C) Duvivier et al., Rapid Commun. Mass Spectrom. 2014, 28,

6 DART-MS scan accuracy Hair spiked with cannabis extract and quinine using TLC-sprayer through 2 mm slit (x-scan at 0.2 mm/s) Spatial resolution: ~ 2.5 mm (~ 1 week history) Improvement over conventional hair analysis 6

7 Drug user hair samples Chronogram of chronic cannabis user hair sample (high THC content: ~5 ng/mg) Duvivier et al., Rapid Commun. Mass Spectrom. 2014, 28,

8 Expansion of method: cocaine Cocaine: different concentrations spiked on blank hair EIC: Blank 0.00 EIC: ng/mg 5.09E2 EIC: ng/mg = SoHT cut-off Cocaine analysis using DART hair scan suitable for forensic applications! 1.95E3 Duvivier et al., Analytical Chemistry 2016, 88,

9 Expansion of method: cocaine Cocaine user hair sample (> 50 ng/mg) Reported use: 2.5 gram daily (!) TIC 7.83E6 Cocaine m/z E6 9

10 Expansion of method: cocaine Cocaine user hair sample (0.5 ng/mg = SoHT cut-off) Reported use: twice a month gram TIC 6.69E6 Cocaine m/z E3 10

11 Confirmation by MS/MS Q-orbitrap instrument (Thermo Q-Exactive) Data dependent MS/MS: m/z of known compounds in inclusion list, when detected: selected for fragmentation in HCD cell Full scan and MS/MS data also retrospective! 11

12 TIC (full scan) Cocaine Cocaine user hair sample Confirmation by MS/MS Continuous cocaine use Amphetamine THC MDMA Recent amphetamine use? _10ppmMix_DART_pos250-DD6 #20 RT: 0.18 AV: 1 NL: 1.05E6 F: FTMS + p NSI d Full ms @hcd35.00 [ ] C 10 H 16 O 2 N = ppm No evidence of cannabis use MDMA use: continuous; recently increased? C 8 H 15 = ppm C 17 H 22 O 4 N = C 5 H 9 = ppm ppm C 9 H 12 O N = ppm 0 Duvivier et al. Analytical Chemistry 2016, 88, m/z Relative Abundance 12

13 Retrospective data analysis Multiple drugs detected with different trends Duvivier et al., Analytical Chemistry 2016, 88,

14 External contamination or incorporated THC? External contamination is a major issue in forensic hair analysis What is measured using the DART-MS hair scan? Incorporated compounds, or External contamination < < 14

15 Metabolite detection Cocaine user hair sample Cocaine m/z m/z E7 Benzoyl- m/z ecgonine 7.38E6 main metabolite BZE at 20% of cocaine intensity shows that detected cocaine is incorporated, not from external contamination

16 Conclusions Direct analysis of intact locks of hair, with sufficient sensitivity and better time resolution Analysis time in the order of minutes Targeted analysis with confirmation by exact mass and data dependent product ion scans Multiple additional hits found using retrospective data analysis Metabolite detection proves measurement of incorporated compounds

17 Acknowledgments Wilco Duvivier Michel Nielen Thijs Meijer Saskia Sterk Ed Pennings

NON TARGETED SEARCHING FOR FOOD

NON TARGETED SEARCHING FOR FOOD NON TARGETED SEARCHING FOR FOOD CONTAMINANTS USING ORBITRAP HIGH RESOLUTION MASS SPECTROMETRY Michal Godula 1, Adrian Charlton 2 and Klaus Mittendorf 1 1 Thermo Fisher Scientific, Dreieich, Germany 2 Food

More information

Don t miss a thing on your peptide mapping journey How to get full coverage peptide maps using high resolution accurate mass spectrometry

Don t miss a thing on your peptide mapping journey How to get full coverage peptide maps using high resolution accurate mass spectrometry Don t miss a thing on your peptide mapping journey How to get full coverage peptide maps using high resolution accurate mass spectrometry Kai Scheffler, PhD BioPharma Support Expert,LSMS Europe The world

More information

Mass Spectrometry Infrastructure

Mass Spectrometry Infrastructure Mass Spectrometry Infrastructure Todd Williams, Ph.D. Director KU Mass Spectrometry and Analytical Proteomics Laboratory Mass Spectrometry Lab B025 Malott Hall Mission The Mass Spectrometry and analytical

More information

The Comparison of High Resolution MS with Triple Quadrupole MS for the Analysis of Oligonucleotides

The Comparison of High Resolution MS with Triple Quadrupole MS for the Analysis of Oligonucleotides The Comparison of High Resolution MS with Triple Quadrupole MS for the Analysis of Oligonucleotides Mohammed Abrar Unilabs York Bioanalytical Solutions Outline Introduction Why LC-MS/MS? Limitations of

More information

Advances in Hybrid Mass Spectrometry

Advances in Hybrid Mass Spectrometry The world leader in serving science Advances in Hybrid Mass Spectrometry ESAC 2008 Claire Dauly Field Marketing Specialist, Proteomics New hybrids instruments LTQ Orbitrap XL with ETD MALDI LTQ Orbitrap

More information

New Instruments and Services

New Instruments and Services New Instruments and Services Liwen Zhang Mass Spectrometry and Proteomics Facility The Ohio State University Summer Workshop 2016 Thermo Orbitrap Fusion http://planetorbitrap.com/orbitrap fusion Thermo

More information

UPLC-HRMS: A tool for multi-residue veterinary drug methods

UPLC-HRMS: A tool for multi-residue veterinary drug methods AOAC, Paris, November 23-24, 2009 UPLC-HRMS: A tool for multi-residue veterinary drug methods Anton Kaufmann Official Food Control Authority of the Canton of Zurich (Kantonales Labor Zürich) The challenge

More information

Amadeo R. Fernández-Alba

Amadeo R. Fernández-Alba % of compounds % of compounds % of compounds % of compounds Amadeo R. Fernández-Alba LC-Orbitrap QExactive Focus Instrumental LOQ 1% 9% 8% 7% 6% 5% 4% 3% 2% 1% %.1 mg/g.2 mg/g Tomato.5 mg/g ddms2 Target

More information

Target Analyses in Parallel Reaction Monitoring Mode (PRM)

Target Analyses in Parallel Reaction Monitoring Mode (PRM) Target Analses in Parallel Reaction Monitoring Mode (PRM) Skline Webinar Januar 13, 2015 Bruno Domon, PhD Head Luxembourg Clinical Proteomics Center Invited Professor Universit of Luxembourg INTRODUCTION

More information

Quantitative Analysis of Drugs of Abuse in Urine using UHPLC Coupled to Accurate Mass AxION 2 TOF Mass Spectrometer

Quantitative Analysis of Drugs of Abuse in Urine using UHPLC Coupled to Accurate Mass AxION 2 TOF Mass Spectrometer application Note Liquid Chromatography/ Mass Spectrometry Authors Sharanya Reddy Blas Cerda PerkinElmer, Inc. Shelton, CT USA Quantitative Analysis of Drugs of Abuse in Urine using UHPLC Coupled to Accurate

More information

Quantification with Proteome Discoverer. Bernard Delanghe

Quantification with Proteome Discoverer. Bernard Delanghe Quantification with Proteome Discoverer Bernard Delanghe Overview: Which approach to use? Proteome Discoverer Quantification Method What When to use Metabolic labeling SILAC Cell culture systems Small

More information

How to Use TOF and Q-TOF Mass Spectrometers

How to Use TOF and Q-TOF Mass Spectrometers How to Use TOF and Q-TOF Mass Spectrometers October 2011 What do TOF and Q-TOF offer? TOF Fast scanning of full spectrum High resolution full scan spectra Accurate mass measurements Q-TOF Fast scanning

More information

Rapid Generation of Fatty Acid Methyl Ester Profiles using (Direct Analysis in Real Time) DART- Mass Spectrometry

Rapid Generation of Fatty Acid Methyl Ester Profiles using (Direct Analysis in Real Time) DART- Mass Spectrometry Rapid Generation of Fatty Acid Methyl Ester Profiles using (Direct Analysis in Real Time) DART- Mass Spectrometry Brian Musselman, Jordan Krechmer, Tim Ng, and Joseph LaPoint IonSense, Inc. 999 Broadway

More information

LOCALISATION, IDENTIFICATION AND SEPARATION OF MOLECULES. Gilles Frache Materials Characterization Day October 14 th 2016

LOCALISATION, IDENTIFICATION AND SEPARATION OF MOLECULES. Gilles Frache Materials Characterization Day October 14 th 2016 LOCALISATION, IDENTIFICATION AND SEPARATION OF MOLECULES Gilles Frache Materials Characterization Day October 14 th 2016 1 MOLECULAR ANALYSES Which focus? LOCALIZATION of molecules by Mass Spectrometry

More information

Michal Godula Thermo Fisher Scientific. The world leader in serving science

Michal Godula Thermo Fisher Scientific. The world leader in serving science Resolving the food authenticity challenges using advanced isotopic ratio and Thermo Scientific Orbitrap high resolution mass spectrometry tools in practice Michal Godula Thermo Fisher Scientific The world

More information

New Instruments and Services

New Instruments and Services New Instruments and Services http://planetorbitrap.com/orbitrap fusion Combining the best of quadrupole, Orbitrap, and ion trap mass analysis in a revolutionary Tribrid architecture, the Orbitrap Fusion

More information

New Mass Spectrometry Tools to Transform Metabolomics and Lipidomics

New Mass Spectrometry Tools to Transform Metabolomics and Lipidomics New Mass Spectrometry Tools to Transform Metabolomics and Lipidomics July.3.13 Ken Miller Vice President of Marketing, Life Sciences Mass Spectrometry 1 The world leader in serving science Omics & the

More information

FOURIER TRANSFORM MASS SPECTROMETRY

FOURIER TRANSFORM MASS SPECTROMETRY FOURIER TRANSFORM MASS SPECTROMETRY https://goo.gl/vx3ogw FT-ICR Theory Ion Cyclotron Motion Inward directed Lorentz force causes ions to move in circular orbits about the magnetic field axis Alan G. Marshall,

More information

Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS

Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS Jenny Chen, Hongxia Wang, Zhiqi Hao, Patrick Bennett, and Greg Kilby Thermo Fisher

More information

Quantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System

Quantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System Quantitative Analysis of Vit D Metabolites in Human Plasma using Exactive System Marta Kozak Clinical Research Applications Group Thermo Fisher Scientific San Jose CA Clinical Research use only, Not for

More information

for new contaminants at ultra trace level by using high resolution mass spectrometry

for new contaminants at ultra trace level by using high resolution mass spectrometry Non-targeted screening for new contaminants at ultra trace level by using high resolution mass spectrometry Dr Igor Fochi LSMS Product Specialist Thermo Fisher Scientific What is Environmental Analysis

More information

Jennifer P. Pascali, PhD. dtolabs, Resana (TV)

Jennifer P. Pascali, PhD. dtolabs, Resana (TV) Jennifer P. Pascali, PhD dtolabs, Resana (TV) Agilent Technologies Users Meeting, 20 May 2014 The project dtolabs dtolabs is an analytical research centre founded by two leading companies in the production

More information

MALDI Imaging Drug Imaging Detlev Suckau Head of R&D MALDI Bruker Daltonik GmbH. December 19,

MALDI Imaging Drug Imaging Detlev Suckau Head of R&D MALDI Bruker Daltonik GmbH. December 19, MALDI Imaging Drug Imaging Detlev Suckau Head of R&D MALDI Bruker Daltonik GmbH December 19, 2014 1 The principle of MALDI imaging Spatially resolved mass spectra are recorded Each mass signal represents

More information

2. Ionization Sources 3. Mass Analyzers 4. Tandem Mass Spectrometry

2. Ionization Sources 3. Mass Analyzers 4. Tandem Mass Spectrometry Dr. Sanjeeva Srivastava 1. Fundamental of Mass Spectrometry Role of MS and basic concepts 2. Ionization Sources 3. Mass Analyzers 4. Tandem Mass Spectrometry 2 1 MS basic concepts Mass spectrometry - technique

More information

Sequence Identification And Spatial Distribution of Rat Brain Tryptic Peptides Using MALDI Mass Spectrometric Imaging

Sequence Identification And Spatial Distribution of Rat Brain Tryptic Peptides Using MALDI Mass Spectrometric Imaging Sequence Identification And Spatial Distribution of Rat Brain Tryptic Peptides Using MALDI Mass Spectrometric Imaging AB SCIEX MALDI TOF/TOF* Systems Patrick Pribil AB SCIEX, Canada MALDI mass spectrometric

More information

The Development of LC/MS Methods for Determination of Polar Drugs of Abuse in Biological Samples

The Development of LC/MS Methods for Determination of Polar Drugs of Abuse in Biological Samples WA20259 The Development of LC/MS Methods for Determination of Polar Drugs of Abuse in Biological Samples Michael S. Young and Kevin M. Jenkins Waters Corporation, 34 Maple Street, Milford, MA 01757 Introduction

More information

Quantitative Analysis of -Hydroxybutyrate in Hair Using Target Analyte Finding Processing of Comprehensive GC-HRT Data

Quantitative Analysis of -Hydroxybutyrate in Hair Using Target Analyte Finding Processing of Comprehensive GC-HRT Data Quantitative Analysis of -Hydroxybutyrate in Hair Using Target Analyte Finding Processing of Comprehensive GC-HRT Data LECO Corporation; Saint Joseph, Michigan USA Key Words: Pegasus GC-HRT, GHB, Hair,

More information

Ten Second DART -MS: Ultra-Fast QC Screening of Foods and Juices with Next Generation DART systems

Ten Second DART -MS: Ultra-Fast QC Screening of Foods and Juices with Next Generation DART systems Ten Second DART -MS: Ultra-Fast QC Screening of Foods and Juices with Next Generation DART systems IonSense, Inc. 999 Broadway Suite 404 Saugus, MA 01906 USA DART-SVP Ionization Source DART -SVP Ion Source

More information

High-sensitivity Orbitrap mass analysis of intact macromolecular assemblies. R. J. Rose, E. Damoc, E. Denisov, A. Makarov, A. J. R.

High-sensitivity Orbitrap mass analysis of intact macromolecular assemblies. R. J. Rose, E. Damoc, E. Denisov, A. Makarov, A. J. R. High-sensitivity Orbitrap mass analysis of intact macromolecular assemblies R. J. Rose, E. Damoc, E. Denisov, A. Makarov, A. J. R. Heck SUPPLEMENTARY INFORMATION HCD multipole C -trap Transport octapole

More information

SCS Mass Spectrometry Laboratory

SCS Mass Spectrometry Laboratory SCS Mass Spectrometry Laboratory Contact Information Staff 31 Noyes Laboratory (8:00-5:00 M-F) 217-333-2545 http://scs.illinois.edu/massspec/ Furong Sun (frs@illinois.edu) Furong Sun Director Training

More information

New Methodologies for POPs Analysis using Orbitrap GC-MS in Official Food Control

New Methodologies for POPs Analysis using Orbitrap GC-MS in Official Food Control New Methodologies for POPs Analysis using Orbitrap GC-MS in Official Food Control Dr. Nuria Cortés-Francisco Servei de Química Laboratori Agència Salut Pública Barcelona Vancouver, August 22 nd, 217 Summary

More information

Small Molecule Science: Experimental designs for achieving ultra trace analysis

Small Molecule Science: Experimental designs for achieving ultra trace analysis Small Molecule Science: Experimental designs for achieving ultra trace analysis Michael P. Balogh Principal scientist Waters Corporation 211 Waters Corporation 1 www.cosmoscience.org The Society for Small

More information

Workflow for Screening and Quantification of the SAMHSA (NIDA) Panel in Urine Using UHPLC-TOF

Workflow for Screening and Quantification of the SAMHSA (NIDA) Panel in Urine Using UHPLC-TOF APPLICATIO OTE Liquid Chromatography/ Mass Spectrometry Authors: Avinash Dalmia Joanne Mather PerkinElmer, Inc. Shelton, CT Workflow for Screening and Quantification of the SAMHSA (IDA) Panel in Urine

More information

High Resolution Accurate Mass Spectrometry in Clinical Toxicology. William Clarke, PhD, MBA, DABCC Johns Hopkins University School of Medicine

High Resolution Accurate Mass Spectrometry in Clinical Toxicology. William Clarke, PhD, MBA, DABCC Johns Hopkins University School of Medicine High Resolution Accurate Mass Spectrometry in Clinical Toxicology Applied to an ED HIV Study William Clarke, PhD, MBA, DABCC Johns Hopkins University School of Medicine Disclosure William Clarke, PhD received

More information

Lipidomic Analysis by UPLC-QTOF MS

Lipidomic Analysis by UPLC-QTOF MS Lipidomic Analysis by UPLC-QTOF MS Version: 1 Edited by: Oliver Fiehn Summary Reagents and Materials Protocol Summary:Lipidomic analysis by UPLC-QTOF mass spectrometry Reagents and Materials: Reagent/Material

More information

Experiences with application of full scan high resolution MS for pesticide residue analysis

Experiences with application of full scan high resolution MS for pesticide residue analysis Experiences with application of full scan high resolution MS for pesticide residue analysis Marc Tienstra, Hans Mol, Paul Zomer? EPRW; Cyprus; 25-May-216 Current routine practice Chromatography with mass

More information

Quantitative Analysis of THC and Main Metabolites in Whole Blood Using Tandem Mass Spectrometry and Automated Online Sample Preparation

Quantitative Analysis of THC and Main Metabolites in Whole Blood Using Tandem Mass Spectrometry and Automated Online Sample Preparation Quantitative Analysis of THC and Main Metabolites in Whole Blood Using Tandem Mass Spectrometry and Automated Online Sample Preparation Valérie Thibert, Bénédicte Duretz Thermo Fisher Scientific, Courtaboeuf,

More information

Advancing your Forensic Toxicology Analyses; Adopting the Latest in Mass Spectrometry Innovations

Advancing your Forensic Toxicology Analyses; Adopting the Latest in Mass Spectrometry Innovations Advancing your Forensic Toxicology Analyses; Adopting the Latest in Mass Spectrometry Innovations For Research Use Only. Not for use in diagnostic procedures. 1 2015 AB Sciex RUO-MKT-11-1018-A For research

More information

Mass Spectrometry at the Laboratory of Food Chemistry. Edwin Bakx Laboratory of Food Chemistry Wageningen University

Mass Spectrometry at the Laboratory of Food Chemistry. Edwin Bakx Laboratory of Food Chemistry Wageningen University Mass Spectrometry at the Wageningen University Mass Spectrometry at the 3 UPLC/CE - ESI - Ion trap MS systems UPLC Thermo Acella with a Velos or VelosPro CE Beckman PA800 with a Thermo VelosPro 1 UPLC-

More information

Rapid Biofluid Analysis using the Prosolia Velox 360 PaperSpray System and Mass Spectrometry

Rapid Biofluid Analysis using the Prosolia Velox 360 PaperSpray System and Mass Spectrometry Rapid Biofluid Analysis using the Prosolia Velox 36 PaperSpray System and Mass Spectrometry Mari Prieto Conaway, PhD Thermo Fisher Scientific Product Manager, Ion Sources The world leader in serving science

More information

Ion Source. Mass Analyzer. Detector. intensity. mass/charge

Ion Source. Mass Analyzer. Detector. intensity. mass/charge Proteomics Informatics Overview of spectrometry (Week 2) Ion Source Analyzer Detector Peptide Fragmentation Ion Source Analyzer 1 Fragmentation Analyzer 2 Detector b y Liquid Chromatography (LC)-MS/MS

More information

Characterization of an Unknown Compound Using the LTQ Orbitrap

Characterization of an Unknown Compound Using the LTQ Orbitrap Characterization of an Unknown Compound Using the LTQ rbitrap Donald Daley, Russell Scammell, Argenta Discovery Limited, 8/9 Spire Green Centre, Flex Meadow, Harlow, Essex, CM19 5TR, UK bjectives unknown

More information

Enhancing Sequence Coverage in Proteomics Studies by Using a Combination of Proteolytic Enzymes

Enhancing Sequence Coverage in Proteomics Studies by Using a Combination of Proteolytic Enzymes Enhancing Sequence Coverage in Proteomics Studies by Using a Combination of Proteolytic Enzymes Dominic Baeumlisberger 2, Christopher Kurz 3, Tabiwang N. Arrey, Marion Rohmer 2, Carola Schiller 3, Thomas

More information

Screening and quantitation of drugs in oral fluids using Q Exactive system

Screening and quantitation of drugs in oral fluids using Q Exactive system Screening and quantitation of drugs in oral fluids using Q Exactive system Marta Kozak November 28,2011 Goal To develop a robust and sensitive quantitative LC-MS/MS analysis method for drugs of abuse in

More information

Toxicology Screening of Whole Blood Extracts Using GC/Triple Quadrupole/MS

Toxicology Screening of Whole Blood Extracts Using GC/Triple Quadrupole/MS Toxicology Screening of Whole Blood Extracts Using GC/Triple Quadrupole/MS Application Note Forensic Toxicology Authors Bruce Quimby and Mike Szelewski Agilent Technologies, Inc. 2850 Centerville Road

More information

Direct Analysis of Folic Acid in Digestive Juices by LC/TOF-MS Application

Direct Analysis of Folic Acid in Digestive Juices by LC/TOF-MS Application Direct Analysis of Folic Acid in Digestive Juices by LC/TOF-MS Application Food Safety Authors M. H. Blokland, P. W. Zoontjes, S. S. Sterk, and L. A. van Ginkel, National Institute of Public Health and

More information

Introduction to the Oligo HTCS Systems. Novatia, LLC

Introduction to the Oligo HTCS Systems. Novatia, LLC Introduction to the Oligo HTCS Systems Novatia, LLC What is the Oligo HTCS and what can it do? Automated ESI/MS system that can analyze up to 14 oligonucleotide samples/day (>3/day with twin autosampler)

More information

Mass Spectrometry Course Árpád Somogyi Chemistry and Biochemistry MassSpectrometry Facility) University of Debrecen, April 12-23, 2010

Mass Spectrometry Course Árpád Somogyi Chemistry and Biochemistry MassSpectrometry Facility) University of Debrecen, April 12-23, 2010 Mass Spectrometry Course Árpád Somogyi Chemistry and Biochemistry MassSpectrometry Facility) University of Debrecen, April 12-23, 2010 Introduction, Ionization Methods Mass Analyzers, Ion Activation Methods

More information

Metabolite identification in metabolomics: Database and interpretation of MSMS spectra

Metabolite identification in metabolomics: Database and interpretation of MSMS spectra Metabolite identification in metabolomics: Database and interpretation of MSMS spectra Jeevan K. Prasain, PhD Department of Pharmacology and Toxicology, UAB jprasain@uab.edu utline Introduction Putative

More information

Introduction to Proteomics 1.0

Introduction to Proteomics 1.0 Introduction to Proteomics 1.0 CMSP Workshop Pratik Jagtap Managing Director, CMSP Objectives Why are we here? For participants: Learn basics of MS-based proteomics Learn what s necessary for success using

More information

Proteomics of body liquids as a source for potential methods for medical diagnostics Prof. Dr. Evgeny Nikolaev

Proteomics of body liquids as a source for potential methods for medical diagnostics Prof. Dr. Evgeny Nikolaev Proteomics of body liquids as a source for potential methods for medical diagnostics Prof. Dr. Evgeny Nikolaev Institute for Biochemical Physics, Rus. Acad. Sci., Moscow, Russia. Institute for Energy Problems

More information

Ten Second DART -MS: Ultra-Fast QC Screening of Dietary Supplements and Pharmaceuticals with a Next Generation DART Source

Ten Second DART -MS: Ultra-Fast QC Screening of Dietary Supplements and Pharmaceuticals with a Next Generation DART Source Ten Second DART -MS: Ultra-Fast QC Screening of Dietary Supplements and Pharmaceuticals with a Next Generation DART Source Elizabeth Crawford, Joseph LaPointe,and Brian Musselman IonSense, Inc. 999 Broadway

More information

Data Independent MALDI Imaging HDMS E for Visualization and Identification of Lipids Directly from a Single Tissue Section

Data Independent MALDI Imaging HDMS E for Visualization and Identification of Lipids Directly from a Single Tissue Section Data Independent MALDI Imaging HDMS E for Visualization and Identification of Lipids Directly from a Single Tissue Section Emmanuelle Claude, Mark Towers, and Kieran Neeson Waters Corporation, Manchester,

More information

NeoSal Oral Fluid Collection System Solutions for Forensic Drug Detection

NeoSal Oral Fluid Collection System Solutions for Forensic Drug Detection PERFORMANCE OF THE NEOSAL ORAL FLUID COLLECTION SYSTEM Neogen develops and manufactures a comprehensive range of ELISA test kits and accessories for forensic drug detection. Neogen offers solutions to

More information

MASS SPECTROMETRY BASED METABOLOMICS. Pavel Aronov. ABRF2010 Metabolomics Research Group March 21, 2010

MASS SPECTROMETRY BASED METABOLOMICS. Pavel Aronov. ABRF2010 Metabolomics Research Group March 21, 2010 MASS SPECTROMETRY BASED METABOLOMICS Pavel Aronov ABRF2010 Metabolomics Research Group March 21, 2010 Types of Experiments in Metabolomics targeted non targeted Number of analyzed metabolites is limited

More information

High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine

High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring and in Urine Xiaolei Xie, Joe DiBussolo, Marta Kozak; Thermo Fisher Scientific, San Jose, CA Application Note 627 Key Words, norbuprenorphine,

More information

«Tit» «Name» «Fam_Nam» «Organization» «Address1» «Address_2» «PO_Box» «City». «Country» Sevilla, February, 2016

«Tit» «Name» «Fam_Nam» «Organization» «Address1» «Address_2» «PO_Box» «City». «Country» Sevilla, February, 2016 «Tit» «Name» «Fam_Nam» «Organization» «Address» «Address_» «PO_Box» «City». «Country» Proficiency Test 5- Laboratory Code: «Lab_code» Sevilla, February, 6 Dear participants, Enclosed you will find the

More information

Agilent 7700x ICP-MS

Agilent 7700x ICP-MS Agilent 7700x ICP-MS Metals analysis Mid ppt to low ppb detection range Option to detect in four different gas modes to gain optimum detection of multiple metals or interferences in a sample Autosampler

More information

An Alternative Approach: Top-Down Bioanalysis of Intact Large Molecules Can this be part of the future? Lecture 8, Page 27

An Alternative Approach: Top-Down Bioanalysis of Intact Large Molecules Can this be part of the future? Lecture 8, Page 27 An Alternative Approach: Top-Down Bioanalysis of Intact Large Molecules Can this be part of the future? Lecture 8, Page 27 Top-down HRAM Bioanalysis of Native Proteins/Molecules Relative Abundance 100

More information

Applications of High Resolution Mass Spectrometry in Forensic Toxicology. Patrick Kyle, PhD.

Applications of High Resolution Mass Spectrometry in Forensic Toxicology. Patrick Kyle, PhD. Applications of High Resolution Mass Spectrometry in Forensic Toxicology Patrick Kyle, PhD pkyle@umc.edu Financial Disclosures Grant/Research Support: None Salary/Consultant Fees: None Board/Committee/Advisory

More information

Analytical Workflow for Extractable and Leachable Impurities

Analytical Workflow for Extractable and Leachable Impurities Analytical Workflow for Extractable and Leachable Impurities AAPS 2015 26 October 2015, Orlando FL Kyle D Silva The world leader in serving science WHAT are EXTRACTABLES and LEACHABLES? 2 3 What is E&L

More information

Biological Mass spectrometry in Protein Chemistry

Biological Mass spectrometry in Protein Chemistry Biological Mass spectrometry in Protein Chemistry Tuula Nyman Institute of Biotechnology tuula.nyman@helsinki.fi MASS SPECTROMETRY is an analytical technique that identifies the chemical composition of

More information

Research to Routine Workflows for Large and Small Molecules using the Q Exactive HR/MS

Research to Routine Workflows for Large and Small Molecules using the Q Exactive HR/MS Research to Routine Workflows for Large and Small Molecules using the Q Exactive HR/MS Sept. 22, 2011 Patrick Bennett Director Pharma Strategic Marketing Thermo Fisher Scientific Agenda Definitions Research

More information

Mass Spectrometry. Actual Instrumentation

Mass Spectrometry. Actual Instrumentation Mass Spectrometry Actual Instrumentation August 2017 See also http://www.uni-bielefeld.de/chemie/analytik/ms f additional infmation 1. MALDI TOF MASS SPECTROMETRY ON THE ULTRAFLEX 2 2. ESI MASS SPECTROMETRY

More information

Identification of Imidacloprid Metabolites in Onions Using High Resolution Mass Spectrometry and Accurate Mass Tools

Identification of Imidacloprid Metabolites in Onions Using High Resolution Mass Spectrometry and Accurate Mass Tools Identification of Imidacloprid Metabolites in Onions Using High Resolution Mass Spectrometry and Accurate Mass Tools Jerry Zweigenbaum 1, Paul Zavitsanos 1, Imma Ferrer 2, and Michael Thurman 2 1 Agilent

More information

Benefits and Characteristic Applications of High Resolution GC/MS and LC/MS. Frank David RIC and Ghent University

Benefits and Characteristic Applications of High Resolution GC/MS and LC/MS. Frank David RIC and Ghent University Benefits and Characteristic Applications of High Resolution GC/MS and LC/MS. Frank David RIC and Ghent University Mass Spectrometry Structure Elucidation Selective and Sensitive Detection Identification

More information

The Development of LC/MS Methods for Determination of MDMA (Ecstasy) and Metabolites in Biological Samples

The Development of LC/MS Methods for Determination of MDMA (Ecstasy) and Metabolites in Biological Samples WA20714 The Development of LC/MS Methods for Determination of MDMA (Ecstasy) and Metabolites in Biological Samples Michael S. Young and Kevin M. Jenkins Waters Corporation, 34 Maple Street, Milford, MA

More information

Comparison of Full Scan MS2 and MS3 Linear Ion Trap Approaches for Quantitation of Vitamin D

Comparison of Full Scan MS2 and MS3 Linear Ion Trap Approaches for Quantitation of Vitamin D Comparison of Full Scan MS2 and MS3 Linear Ion Trap Approaches for Quantitation of Vitamin D Julie A. Horner 1, Marta Kozak 1, Subodh Nimkar 1, and August A. Specht 1 1 Thermo Fisher Scientific, San Jose,

More information

Food Analysis Applications with Triple Quadrupole LC/MS and TOF LC/MS

Food Analysis Applications with Triple Quadrupole LC/MS and TOF LC/MS Food Analysis Applications with Triple Quadrupole LC/MS and TOF LC/MS Linda Côté Application Specialist Page 1 1) Banned Antibiotics substances in food 2) Pesticides Residues screening in Food Page 2 1

More information

Fundamentals of Soft Ionization and MS Instrumentation

Fundamentals of Soft Ionization and MS Instrumentation Fundamentals of Soft Ionization and MS Instrumentation Ana Varela Coelho varela@itqb.unl.pt Mass Spectrometry Lab Analytical Services Unit Index Mass spectrometers and its components Ionization methods:

More information

INTRODUCTION TO MALDI IMAGING

INTRODUCTION TO MALDI IMAGING INTRODUCTION TO MALDI IMAGING Marten F. Snel, Emmanuelle Claude, Thérèse McKenna, and James I. Langridge Waters Corporation, Manchester, UK INT RODUCTION The last few years have seen a rapid increase in

More information

Orbitrap technology update

Orbitrap technology update Orbitrap technology update Users Meeting Somerset, Oct 12, 211 Dr. Thomas Moehring Outline Users Meeting 21 Technology Update 211 - Orbitrap Elite - Q Exactive Summary 2 Your Goal: The ideal mass analyzer

More information

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS New Solvent Grade Targeted for Trace Analysis by UHPLC-MS Subhra Bhattacharya, Deva H. Puranam, and Stephen C. Roemer Thermo Fisher Scientific Fisher Chemical, One Reagent Lane, Fair Lawn, NJ Material

More information

Comparison of mass spectrometers performances

Comparison of mass spectrometers performances Comparison of mass spectrometers performances Instrument Mass Mass Sensitivity resolution accuracy Quadrupole 1 x 10 3 0.1 Da* 0.5-1.0 pmol DE-MALDI 2 x 10 4 20 ppm 1-10 fmol peptide 1-5 pmol protein Ion

More information

Automating Mass Spectrometry-Based Quantitative Glycomics using Tandem Mass Tag (TMT) Reagents with SimGlycan

Automating Mass Spectrometry-Based Quantitative Glycomics using Tandem Mass Tag (TMT) Reagents with SimGlycan PREMIER Biosoft Automating Mass Spectrometry-Based Quantitative Glycomics using Tandem Mass Tag (TMT) Reagents with SimGlycan Ne uaca2-3galb1-4glc NAcb1 6 Gal NAca -Thr 3 Ne uaca2-3galb1 Ningombam Sanjib

More information

FOURIER TRANSFORM MASS SPECTROMETRY

FOURIER TRANSFORM MASS SPECTROMETRY FOURIER TRANSFORM MASS SPECTROMETRY FT-ICR Theory Ion Cyclotron Motion Inward directed Lorentz force causes ions to move in circular orbits about the magnetic field axis Alan G. Marshall, Christopher L.

More information

Mass Spectrometry based metabolomics

Mass Spectrometry based metabolomics Mass Spectrometry based metabolomics Metabolomics- A realm of small molecules (

More information

Determination of Benzodiazepines in Urine by CE-MS/MS

Determination of Benzodiazepines in Urine by CE-MS/MS Determination of Benzodiazepines in Urine by CE-MS/MS Application ote Forensic Toxicology Authors audimir Lucio do Lago Department of Fundamental Chemistry, Institute of Chemistry University of São Paulo,

More information

FOURIER TRANSFORM MASS SPECTROMETRY

FOURIER TRANSFORM MASS SPECTROMETRY FOURIER TRANSFORM MASS SPECTROMETRY FT-ICR Theory Ion Cyclotron Motion Inward directed Lorentz force causes ions to move in circular orbits about the magnetic field axis Alan G. Marshall, Christopher L.

More information

Ultra High Definition Optimizing all Analytical Dimensions

Ultra High Definition Optimizing all Analytical Dimensions Ultra High Definition Optimizing all Analytical Dimensions Sensitivity Dynamic Range Signal Response Linearity Separation Speed Peak Capacity Chromatogram Mass Spectrum Mass Accuracy Resolving Power Acquisition

More information

Join the mass movement towards mass spectrometry

Join the mass movement towards mass spectrometry Join the mass movement towards mass spectrometry Thermo Scientific ISQ EC single quadrupole mass spectrometer Embrace the power of mass spectrometry Achieving a comprehensive understanding of the samples

More information

Workflow for Screening and Quantification of the SAMHSA (NIDA) Panel in Serum Using UHPLC-TOF

Workflow for Screening and Quantification of the SAMHSA (NIDA) Panel in Serum Using UHPLC-TOF APPLICATIO OTE Liquid Chromatography/ Mass Spectrometry Authors: Avinash Dalmia Bonnie Marmor PerkinElmer, Inc. Shelton, CT Workflow for Screening and Quantification of the SAMHSA (IDA) Panel in Serum

More information

Components of a Mass Spectrometer

Components of a Mass Spectrometer Components of a Mass Spectrometer Sample Introduction Inlet GC LC Direct Insertion (Syringe/Probe) Ionization Ion Separation Ion Detection Ion Source EI,CI,,, MALDI Mass Analyzer Under vacuum TOF, Quadrupole,

More information

SIEVE 2.1 Proteomics Example

SIEVE 2.1 Proteomics Example SIEVE 2.1 Proteomics Example Software Overview What is SIEVE? SIEVE is Thermo Scientific s differential software solution. SIEVE will continue to enhance our current product for label-free differential

More information

Methods in Mass Spectrometry. Dr. Noam Tal Laboratory of Mass Spectrometry School of Chemistry, Tel Aviv University

Methods in Mass Spectrometry. Dr. Noam Tal Laboratory of Mass Spectrometry School of Chemistry, Tel Aviv University Methods in Mass Spectrometry Dr. Noam Tal Laboratory of Mass Spectrometry School of Chemistry, Tel Aviv University Sample Engineering Chemistry Biology Life Science Medicine Industry IDF / Police Sample

More information

Metabolite identification in metabolomics: Metlin Database and interpretation of MSMS spectra

Metabolite identification in metabolomics: Metlin Database and interpretation of MSMS spectra Metabolite identification in metabolomics: Metlin Database and interpretation of MSMS spectra Jeevan K. Prasain, PhD Department of Pharmacology and Toxicology, UAB jprasain@uab.edu Outline Introduction

More information

NIH Public Access Author Manuscript J Proteome Res. Author manuscript; available in PMC 2014 July 05.

NIH Public Access Author Manuscript J Proteome Res. Author manuscript; available in PMC 2014 July 05. NIH Public Access Author Manuscript Published in final edited form as: J Proteome Res. 2013 July 5; 12(7): 3071 3086. doi:10.1021/pr3011588. Evaluation and Optimization of Mass Spectrometric Settings during

More information

Development and validation of a screening method for drug residues in fish, shrimp and eel using LC-HRMS

Development and validation of a screening method for drug residues in fish, shrimp and eel using LC-HRMS Development and validation of a screening method for drug residues in fish, shrimp and eel using LC-HRMS Sherri Turnipseed 1, Joseph Storey 1, Robert Burger 2, Aaron Johnson 2, Jack Lohne 1, Wendy Andersen

More information

Designer Fentanyls Drugs that kill and how to detect them. Cyclopropylfentanyl

Designer Fentanyls Drugs that kill and how to detect them. Cyclopropylfentanyl Designer Fentanyls Drugs that kill and how to detect them Cyclopropylfentanyl Science for a safer world The in vitro metabolism of cyclopropylfentanyl Simon Hudson & Charlotte Cutler, Sport and Specialised

More information

SWATH Acquisition Enables the Ultra-Fast and Accurate Determination of Novel Synthetic Opioids

SWATH Acquisition Enables the Ultra-Fast and Accurate Determination of Novel Synthetic Opioids SWATH Acquisition Enables the Ultra-Fast and Accurate Determination of Novel Synthetic Opioids Data Independent Acquisition on TripleTOF and X-Series QTOF Systems for Seized Drug Analysis Oscar G. Cabrices

More information

A NOVEL METHOD OF M/Z DRIFT CORRECTION FOR OA-TOF MASS SPECTROMETERS BASED ON CONSTRUCTION OF LIBRARIES OF MATRIX COMPONENTS.

A NOVEL METHOD OF M/Z DRIFT CORRECTION FOR OA-TOF MASS SPECTROMETERS BASED ON CONSTRUCTION OF LIBRARIES OF MATRIX COMPONENTS. A NOVEL METHOD OF M/Z DRIFT CORRECTION FOR OA-TOF MASS SPECTROMETERS BASED ON CONSTRUCTION OF LIBRARIES OF MATRIX COMPONENTS. Martin R Green*, Keith Richardson, John Chipperfield, Nick Tomczyk, Martin

More information

Strategies for Rapid Determination of Impurities using DART-MS from Surfaces

Strategies for Rapid Determination of Impurities using DART-MS from Surfaces Strategies for Rapid Determination of Impurities using DART-MS from Surfaces Brian Musselman, Elizabeth Crawford, Catherine Stacy and Robert Cody IonSense, Inc. Saugus, MA,. USA DART Ionization Source

More information

1 Seminario di Spettrometria di Massa. Università degli studi di Milano 23/06/2017 The world leader in serving science

1 Seminario di Spettrometria di Massa. Università degli studi di Milano 23/06/2017 The world leader in serving science Innovazioni tecnologiche per l analisi GC-MS in bassa ed alta risoluzione con il nuovo GC-Orbitrap Dott.ssa Cristina Neri, Ph.D Sales Specialist GC-GC&MS 1 Seminario di Spettrometria di Massa. Università

More information

Rapid and Sensitive Screening of Benzodiazepines in Serum Using Liquid Chromatography-APCI-Linear Ion Trap System

Rapid and Sensitive Screening of Benzodiazepines in Serum Using Liquid Chromatography-APCI-Linear Ion Trap System T2007 Seattle, Washington Rapid and Sensitive Screening of Benzodiazepines in Serum Using Liquid hromatography-api-linear Ion Trap System Viktor Vorisek *, Vilma Habrdova, Pavel Zivny and Vladimir Palicka

More information

Towards High Resolution MS in Regulated Bioanalysis

Towards High Resolution MS in Regulated Bioanalysis Towards High Resolution MS in Regulated Bioanalysis Benno Ingelse MSD 3 rd EB focus meeting June 12 th 13 th Brussels, Belgium Contributors Gary Adamson Ken Anderson Kevin Bateman Cynthia Chavez-Eng Inhou

More information

Increased Identification Coverage and Throughput for Complex Lipidomes

Increased Identification Coverage and Throughput for Complex Lipidomes Increased Identification Coverage and Throughput for Complex Lipidomes Reiko Kiyonami, David Peake, Yingying Huang, Thermo Fisher Scientific, San Jose, CA USA Application Note 607 Key Words Q Exactive

More information

Supporting information

Supporting information Supporting information A novel lipidomics workflow for improved human plasma identification and quantification using RPLC-MSn methods and isotope dilution strategies Evelyn Rampler 1,2,3, Angela Criscuolo

More information

Detection of Cannabinoids in Oral Fluid with the Agilent 7010 GC-MS/MS System

Detection of Cannabinoids in Oral Fluid with the Agilent 7010 GC-MS/MS System Application Note Forensics, Workplace Drug Testing Detection of Cannabinoids in Oral Fluid with the Agilent 7010 GC-MS/MS System Authors Fred Feyerherm and Anthony Macherone Agilent Technologies, Inc.

More information

Occupational exposure to alcohol-based hand sanitizers: the diriment role of alcohol biomarkers in hair

Occupational exposure to alcohol-based hand sanitizers: the diriment role of alcohol biomarkers in hair Occupational exposure to alcohol-based hand sanitizers: the diriment role of alcohol biomarkers in hair Submission by Bozzo Andrea mobile: +393401932599 e-mail: andrea2bozzo@gmail.com Introduction The

More information

The use of mass spectrometry in lipidomics. Outlines

The use of mass spectrometry in lipidomics. Outlines The use of mass spectrometry in lipidomics Jeevan Prasain jprasain@uab.edu 6-2612 utlines Brief introduction to lipidomics Analytical methodology: MS/MS structure elucidation of phospholipids Phospholipid

More information