Internal Calibration System of Thermo Scientific Varioskan Flash with Improved Sensitivity, Accuracy and Dynamic Range

Similar documents
FT-Raman Surface Mapping of Remineralized Artificial Dental Caries

Thermo Scientific ART Pipette Tips

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine

High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine

Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS

Comparison of Full Scan MS2 and MS3 Linear Ion Trap Approaches for Quantitation of Vitamin D

GC-MS/MS Analysis of Benzodiazepines Using Analyte Protectants

Nitrogen/Protein Determination in Starch by Flash Combustion using Large Sample Weight as an Alternative to the Kjeldahl Method

PIF: Precursor Ion Fingerprinting Searching for a Structurally Diagnostic Fragment Using Combined Targeted and Data Dependent MS n

The Investigation of Factors Contributing to Immunosuppressant Drugs Response Variability in LC-MS/MS Analysis

Thermo Scientific Finntip Pipette Tips. HANDHELD PIPETTING Finntip Pipette Tips. Details

Detection. personal. Now, you can optimize your personal workflow. Promega instruments and reagents integrate easily.

L-Amino Acid Assay Kit

Impurity Profiling of Carbamazepine by HPLC/UV

Thermo Scientific Finnpipette Manual Pipetting Systems. you can handle. anything

Quantitative Analysis of THC and Main Metabolites in Whole Blood Using Tandem Mass Spectrometry and Automated Online Sample Preparation

ab Factor Xa Activity Assay Kit (Fluorometric)

Free Fatty Acid Uptake Assay Kit (Fluorometric)

ab HIV-1 Protease Inhibitor Screening Kit (Fluorometric)

Integrated Targeted Quantitation Method for Insulin and its Therapeutic Analogs

Dynamic Analysis of HIV-Human Protein-Protein Interactions During Infection

Solving One of Chromatography s Biggest Dilemmas Proper Sealing of Chromatography Autosampler Vials

ab Cathepsin K Inhibitor Screening Kit (Fluorometric)

Product information: Human Apolipoprotein B kit 10,000 tests 1. ASSAY DESCRIPTION AND INTENDED USE 2. PROTOCOL AT GLANCE

Determination of Clinically Relevant Compounds using HPLC and Electrochemical Detection with a Boron-Doped Diamond Electrode

Enhancing Sequence Coverage in Proteomics Studies by Using a Combination of Proteolytic Enzymes

PD1/PD-L1 BINDING ASSAY KITS

ab HRV 3C Protease Inhibitor Screening Kit (Colorimetric)

Dayana Argoti, Kerry Hassell, Sarah J. Fair, and Joseph Herman Thermo Fisher Scientific, Franklin, MA, USA

ApplicationNOTE EXACT MASS MEASUREMENT OF ACTIVE COMPONENTS OF TRADITIONAL HERBAL MEDICINES BY ORTHOGONAL ACCELERATION TIME-OF-FLIGHT.

High-Throughput Quantitative LC-MS/MS Analysis of 6 Opiates and 14 Benzodiazepines in Urine

L-Amino Acid Assay Kit

Phenylalanine Assay Kit

Rapid and Direct Analysis of Free Phytosterols by Reversed Phase HPLC with Electrochemical Detection

Detection and Quantification of Inorganic Arsenic in Fruit Juices by Capillary Ion Chromatography with Suppressed Conductivity Detection

Ascorbic Acid Assay Kit

ab HRP Substrate Instructions for Use

Lipase Detection Kit III (Fluorometric)

Protease Activity Assay Kit (Fluorometric Red)

Glucose 6 Phosphate Dehydrogenase Assay Kit (Fluorometric)

For the accurate measurement of glycogen levels in various biological samples.

For the rapid, sensitive and accurate measurement of Aspartate aminotransferase activity in various samples

INSULIN MOUSE SERUM KITS

ab Membrane fluidity kit Instructions for Use For the detection of membrane fluidity in cells

Glucose 6 Phosphate Assay Kit (Colorimetric)

ab HIV-1 Protease Activity Assay Kit

A Rapid UHPLC Method for the Analysis of Biogenic Amines and Metabolites in Microdialysis Samples

Mercury Speciation Determinations in Asian Dietary Supplements

SERUM GLUCAGON KITS PROTOCOL

CytoPainter LysoGreen Indicator Reagent

Glucose Detection Kit

Evaluation of an LC-MS/MS Research Method for the Analysis of 33 Benzodiazepines and their Metabolites

Glucose-6-phosphate Isomerase Activity Assay Kit (Colorimetric)

For the accurate measurement of glucose in cells and tissues, plasma, serum and other body fluids, growth media and food.

PosterReprint OVERVIEW

Myeloperoxidase (MPO) Activity Assay Kit (Fluorometric)

Essential Lipidomics Experiments using the LTQ Orbitrap Hybrid Mass Spectrometer

Singlet Oxygen Production Photosensitized by Fluorescein in Reversed Micellar Solutions. Norio Miyoshi and Giiti Tomita*

ab HMG-CoA Reductase Activity Assay Kit (Colorimetric)

ab Dipeptidyl peptidase IV (DPP4) Activity Assay Kit (Fluorometric)

Glucose and Sucrose Assay Kit

Mouse serum Insulin 200 tests

6-Phosphofructokinase Activity Assay Kit (Colorimetric)

For the rapid, sensitive and accurate measurement of Lactose levels in various samples

Acid Phosphatase Assay Kit (Fluorometric)

Cathepsin K Activity Assay Kit (Fluorometric)

For the rapid, sensitive and accurate measurement of Sucrose levels in various samples

ab Intracellular O 2 Probe

ab83375 Sialic Acid (NANA) Assay Kit Instructions for Use For the rapid, sensitive and accurate measurement of Sialic Acid (NANA) in various samples.

Thermo Scientific LipidSearch Software for Lipidomics Workflows. Automated Identification and Relative. Quantitation of Lipids by LC/MS

Acetoacetate Assay Kit (Colorimetric)

Providing a Universal, One-step Alternative to Liquid-Liquid Extraction in Bioanalysis

For the rapid, sensitive and accurate measurement of Glycerol in cell cultures.

D-Mannitol Colorimetric Assay Kit

Superior Fluorescent Labeling Dyes Spanning the Full Visible Spectrum...1. Trademarks: HiLyte Fluor (AnaSpec, Inc.)

Cisbio Bioassays MAP-Tau assay is only intended for quantitative measurement of microtubule-associated protein tau (MAP-Tau) using HTRF technology.

LC-MS/MS Method for the Determination of Tenofovir from Plasma

Discrete Industrial Analyzer System Reagents for Food and Other Sample Materials

Dako IT S ABOUT TIME. Interpretation Guide. Agilent Pathology Solutions. ALK, ROS1 and RET IQFISH probes (Dako Omnis) MET IQFISH probe (Dako Omnis)

Lipase Detection Kit II (Colorimetric)

Sialic Acid Determination in Glycoproteins: Comparison of Two Liquid Chromatography Methods. Experimental. Introduction. Method 1.

HRM Series PCB Power Relays

For the rapid, sensitive and accurate measurement of Glucose in various samples

ab LDL Uptake Assay Kit (Fluorometric)

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS

slge112 Molecular Allergology Product Characteristics ImmunoCAP ISAC slge 112

ab Tryptophan Assay Kit (Fluorometric)

For the rapid, sensitive and accurate measurement of Maltose levels in various samples.

Thermo Fisher Scientific, Sunnyvale, CA, USA; 2 Thermo Fisher Scientific, San Jose, CA, USA

For the rapid, sensitive and accurate detection of long-chain Free Fatty Acids in various samples.

Choline/Acetylcholine Assay Kit

Carl Fisher, Terri Christison, Hua Yang, Monika Verma, and Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA, USA

Increased Identification Coverage and Throughput for Complex Lipidomes

Component Analysis of an Excitation-Emission Matrix of Water Samples Using PARAFAC Analysis

HER2 FISH pharmdx TM Interpretation Guide - Breast Cancer

Manual. Transfer step. Figure 1. Workflow comparison of manual phenylalanine assay and automated GSP Neonatal Phenylalanine assay

Transcription:

Internal Calibration System of Thermo Scientific Varioskan Flash with Improved Sensitivity, Accuracy and Dynamic Range Marika Raitio and Jorma Lampinen Thermo Fisher Scientific, Vantaa, Finland Key Words Fluorescence Luminescence PMT Gain Multitechnology Reader Spectral Correction Reference Signal Xenon Lamp Introduction When fluorescence signals are measured the results are normally referred as Relative Fluorescence Units (RFUs) to emphasis the fact that resulting numeric values are heavily dependent on the settings of the instrument electronics, mainly PMT gain voltage. Results are also somewhat dependent on the environmental aspects, like ambient temperature that affects the electronic noise levels. The practical effect of this relative nature of fluorescence measurements is that result values become quite difficult to be compared, especially if the results are from different measurements done with different instrument settings. Most of the available instruments partially overcome these difficulties by preventing the use of different electronic settings inside one assay, for example PMT gain voltage is fixed within an assay. This method causes new difficulties that are seen for example as a narrow dynamic measurement range caused by fixed gain voltage. Specially, when high intensity Xenon flash lamps are used for the excitation, the available dynamic range with one gain voltage is commonly reduced to about 3 4 orders of magnitude. Fluorometric assay chemistries offer chemical concentration ranges around 5 6 orders of magnitude, or even more, so dynamic range of fixed gain voltage is not sufficient. If Xenon flash lamp is used for the excitation it is necessary to be able to use several PMT gain voltages within one assay to get a good assay performance: both high sensitivity and large dynamic range. When an instrument with fixed PMT gain is used to measure such an assay where the concentration range exceeds the available dynamic range, either high concentration samples are saturated or low concentration samples are not separated from the background, depending on the PMT gain value selected. In both cases certain samples will give incorrect result values. In addition, the detection system affects also on spectral scanning measurements. The detector system, mainly PMT and monochromator grating, efficiency varies according to the wavelength and that will have an effect on the peak and shape of the spectra. Thermo Scientific Varioskan Flash multitechnology reader has in-built calibration system for all measurement technologies covering all mentioned problems in the measurement. This paper shows how detection systems for fluorescence intensity, time resolved fluorescence and luminescence are calibrated to ensure the best possible performance in each individual measurement. Figure 1. Basic design of Varioskan Flash spectral scanning reader. Optical Design The basic optical design of the Varioskan Flash reader is shown in Figure 1. The reader has two optical modules. LumiSens module is used for normal luminescence measurements without any wavelength selection and for luminescence assays where filters are used to separate two or more wavelengths. Another module is a monochromator based spectral scanning module that is used for all the other detection technologies. Both modules are designed so that at least one dispensing head of the instrument s injector units can dispense the well simultaneously with the measurement optics. This makes it possible to dispense the triggering reagent and measure the signal simultaneously and therefore gives possibility to measure very fast flash type chemical reactions. Both LumiSens module and spectral scanning module have similar calibration system that makes it possible to use several different PMT gain voltages to measure one set of samples. The spectral scanning module includes two reference chips in the plate carrier that produce the calibration signal, one is used for the calibration of the top reading optics and another for the bottom reading optics. The location of the reference chips is shown in Figure 2. These chips produce stable

emission signal over the whole spectral range and is used to calibrate four predefined PMT gain voltages. The basic design for luminometric measurement module is shown in Figure 3. Luminometric measurement module is calibrated using stabilized 560 nm green LED that can produce stable light signal over large intensity range. Reference Chip (Bottom Reading) Reference Chip (Top Reading) Plate Tray Figure 2. Varioskan Flash plate carrier system with PMT gain calibration chips for spectral scanning module. instrument will measure all possible information about the instrument electronics, mechanics and optics, for example PMT sensitivity curves, lamp intensity and spectral characteristics, monochromator relative sensitivity spectra etc. This factory calibration takes about an hour and after that all this information describing how instrument performs is stored inside the instrument memory. Factory calibration needs to be done only after the instrument has been assembled in the factory, or if the internal software is changed or major service has been done. Start-up calibration is performed every time when instrument is switched on to eliminate possible effect of environmental effects or aging of the electronics. This process is a shortened version from factory calibration where instrument checks that all calibration values are still the same that they were during the factory calibration, all calibration data is still completely valid. Runtime calibration is performed before and during each measurement run. It includes the calibration of PMT gain voltages for those wavelengths that will be used in the measurement with spectral scanning module when LumiSens module for luminometry is always calibrated with 560 nm signal. The calibration has the following steps: Reference Light Source PMT Before the measurement PMT gain voltage calibration with the reference signals. The reference chip (when spectral scanning module will be used in assay) or the 560 nm LED (when LumiSens module will be used) sends couple of calibration signals with certain intensity difference between the signals, covering the whole dynamic range. All those wavelengths that have been selected to be used in the assay are calibrated using the reference signals. These reference signals are measured with predefined (during factory calibration) PMT gain voltages. Based on this calibration data, instrument will form a conversion table where relative differences between the PMT gain voltages are normalized and defines light intensity levels where each PMT gain voltage should be used. This conversion table is then used to normalize measurement results that have been measured with different gain voltages. The effect of this PMT gain calibration is shown in Figure 4. Figure 3. Detailed optical design of Varioskan Flash LumiSens module. Calibration process The calibration process is divided into three phases: Initial factory calibration, start-up calibration and runtime calibration. Factory calibration process is the main system for the instrument calibration and it is fully integrated process that is completely based on these reference samples inside the unit. During this process, the During the measurement Xenon lamp intensity calibration. When Xenon lamp flashes, each flash is an individual phenomenon, every flash has little different intensity, flash time etc. Therefore, a small portion of the flash is directed directly to the reference detector that will analyze flash properties (intensity, length etc.). This information is used to eliminate the excitation intensity variations. The effect of Xenon lamp intensity calibration is shown in Figure 5.

Figure 4. Effect of the PMT gain calibration. A) Fluorescein calibration series was measured with different PMT gain voltages without PMT gain calibration. Each PMT gain selection gives its typical about three decades linear dynamic range and same PMT signal response is obtained with several fluorescein concentrations. B) When several PMT gains are calibrated against each others continuous linear dynamic range over seven order of magnitude can be obtained. Each gain is responsible for measuring certain area of the dynamic range. Figure 5. Effect of Xenon lamp intensity calibration. Figure 6. PMT gain selection process during the measurement. 10 ms trial flash is used prior to each measurement to select the correct PMT gain. The PMT gain that was used for the previous measurement is used for the first trial flash. If needed, PMT gain in changed and new trial flash is performed until a correct gain is found.

When these variations in the reading efficiency of the instrument are reflected to the measured spectra totally incorrect spectra can be seen. The technical (uncorrected) and true chemical (corrected) spectra of tryptophan and 4-methylumbelliferone are shown in Figure 8. The comparison of the spectra clearly shows the effect of the spectral correction on the peak emission wavelengths and the difference between peak wavelengths of technical and chemical spectrum can be up to even 15-20 nm with certain labels, for example tryptophan. Figure 7. Wavelength dependent efficiency of the monochromator based detection system of Thermo Scientific Varioskan Flash. As seen, the instrument has the highest detection sensitivity at 409 nm and the sensitivity is remarkably lower at for example low UV and near IR areas. PMT gain voltage selection for each sample. The optimal gain voltage is selected individually for each sample well using one flash test measurement. Based on the emission signal level from this trial flash measurement, the Varioskan Flash selects the optimal gain voltage and performs the actual measurement. When next sample is measured, similar trial flash measurement is performed and a new gain is selected for the new sample. This trial flash test measurement is performed before each measurement in all measurement types, end-point, kinetic and spectral scanning. The schematic diagram of this process is shown as an example in Figure 6. Spectral Correction Efficiency of any detection system is affected by the electronic components included, specially the PMT and in monochromator systems the monochromator grating. The relative efficiency varies according to the wavelength and has an effect specially on the spectra. If these instrument dependent effects are not eliminated from the spectra then the spectra do not precisely represent the true chemical spectra and are called technical spectra. The difference between technical and true chemical spectra is dependent the case, when spectra is measured in the range where instrument s relative efficiency changed rapidly strong difference between technical and chemical spectra is seen. Figure 7. Shows the wavelength-dependent efficiency of the detection system of the Varioskan Flash reader. This reader has a unique spectral correction feature which can be used to correct the shift of the technical spectra caused by the effects of the instrument's detection optics. Figure 8. Examples of the differences in technical and chemical fluorescence spectra of two common fluorometric labels. A) When tryptophan label is used, a strong spectral shift to the higher wavelength is seen in technical spectra due to electronics higher relative sensitivity to wavelengths over 350 nm. B) With Methyl-umbelliferone an opposite effect is seen: Instrument electronics is more sensitive around 400 430 nm than around 440 470 nm.

Conclusions With this fully automatic gain selection it is possible to use optimal gain voltage to measure each sample individually. When high emission signal sample is measured, a low gain voltage is used to ensure sufficient measurement range without the detector saturation. When low emission signal sample is measured, the high gain voltage is used to guarantee good signal to noise ratio and therefore good assay sensitivity. Automatic gain calibration and selection makes it possible to obtain simultaneously both high assay sensitivity and dynamic range sufficient for any fluorometric application. Spectral correction feature is needed to measure instrument independent chemical spectra and can be valuable in assay optimization when small changes in wavelengths can improve the assay performance. In addition to these offices, Thermo Fisher Scientific maintains a network of representative organizations throughout the world. North America: USA / Canada +1 800 522 7763 Europe: Austria +43 1 801 40 0 Belgium +32 53 73 42 41 France +33 2 2803 2180 Germany national toll free 08001-536 376 Germany international +49 6184 90 6940 Italy +39 02 02 95059 448 Netherlands +31 76 571 4440 Nordic countries +358 9 329 100 Russia/CIS +7 (495) 739 76 41 Spain/Portugal +34 93 223 09 18 Switzerland +41 44 454 12 12 UK/Ireland +44 870 609 9203 Asia: China +86 21 6865 4588 or +86 10 8419 3588 India toll free 1800 22 8374 India +91 22 6716 2200 Japan +81 45 453 9220 Other Asian countries +852 2885 4613 Countries not listed: +49 6184 90 6940 or +33 2 2803 2180 www.thermoscientific.com ANMIVFlash09-0467 2010 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.