Re irradiation volumetric modulated arc therapy optimization based on cumulative biologically effective dose objectives

Similar documents
Reirradiazione. La radioterapia stereotassica ablativa: torace. Pierluigi Bonomo Firenze

The dependence of optimal fractionation schemes on the spatial dose distribution

Dosimetric Analysis of 3DCRT or IMRT with Vaginal-cuff Brachytherapy (VCB) for Gynaecological Cancer

Efficient SIB-IMRT planning of head & neck patients with Pinnacle 3 -DMPO

Evaluation of Monaco treatment planning system for hypofractionated stereotactic volumetric arc radiotherapy of multiple brain metastases

Original Article. Teyyiba Kanwal, Muhammad Khalid, Syed Ijaz Hussain Shah, Khawar Nadeem

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *

SBRT TREATMENT PLANNING: TIPS + TRICKS. Rachel A. Hackett CMD, RT(T)

Treatment Planning & IGRT Credentialing for NRG SBRT Trials

Automated Plan Quality Check with Scripting. Rajesh Gutti, Ph.D. Clinical Medical Physicist

Radiation Damage Comparison between Intensity Modulated Radiotherapy (IMRT) and Field-in-field (FIF) Technique In Breast Cancer Treatments

Knowledge-Based IMRT Treatment Planning for Prostate Cancer: Experience with 101. Cases from Duke Clinic. Deon Martina Dick

and Strength of Recommendations

A Dosimetric Comparison of Whole-Lung Treatment Techniques. in the Pediatric Population

The Evolution of SBRT and Hypofractionation in Thoracic Radiation Oncology

Silvia Pella, PhD, DABR Brian Doozan, MS South Florida Radiation Oncology Florida Atlantic University Advanced Radiation Physics Boca Raton, Florida

Evaluation of Whole-Field and Split-Field Intensity Modulation Radiation Therapy (IMRT) Techniques in Head and Neck Cancer

Treatment Planning for Lung. Kristi Hendrickson, PhD, DABR University of Washington Dept. of Radiation Oncology

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 17, NUMBER 6, 2016

20. Background. Oligometastases. Oligometastases: bone (including spine) and lymph nodes

Radiation Planning Index for dose distribution evaluation in stereotactic radiotherapy

Feasibility of the partial-single arc technique in RapidArc planning for prostate cancer treatment

A new homogeneity index based on statistical analysis of the dose volume histogram

Ashley Pyfferoen, MS, CMD. Gundersen Health Systems La Crosse, WI

Optimising Radiotherapy Using NTCP Models: 17 Years in Ann Arbor

Defining Target Volumes and Organs at Risk: a common language

Comparison of IMRT and VMAT Plan for Advanced Stage Non-Small Cell Lung Cancer Treatment

Overview. Proton Therapy in lung cancer 8/3/2016 IMPLEMENTATION OF PBS PROTON THERAPY TREATMENT FOR FREE BREATHING LUNG CANCER PATIENTS

Radiotherapy Planning (Contouring Lung Cancer for Radiotherapy dose prescription) Dr Raj K Shrimali

CURRICULUM OUTLINE FOR TRANSITIONING FROM 2-D RT TO 3-D CRT AND IMRT

A dosimetric evaluation of VMAT for the treatment of non-small cell lung cancer

The Physics of Oesophageal Cancer Radiotherapy

Treatment Planning Evaluation of Volumetric Modulated Arc Therapy (VMAT) for Craniospinal Irradiation (CSI)

Comparing conformal, arc radiotherapy and helical tomotherapy in craniospinal irradiation planning

IMRT - the physician s eye-view. Cinzia Iotti Department of Radiation Oncology S.Maria Nuova Hospital Reggio Emilia

Tecniche Radioterapiche U. Ricardi

RTTs role in lung SABR

Stereotactic body radiotherapy (SBRT) has been increasingly

The Effects of DIBH on Liver Dose during Right-Breast Treatments: A Case Study Abstract: Introduction: Case Description: Conclusion: Introduction

IMRT Planning Basics AAMD Student Webinar

The Effects of DIBH on Liver Dose during Right-Breast Treatments Introduction

8/3/2016. Implementation of Pencil Beam Scanning (PBS) Proton Therapy Treatment for Liver Patients. Acknowledgement. Overview

Disclosure. Paul Medin teaches radiosurgery courses sponsored by BrainLAB Many animals (and humans) were harmed to make this presentation possible!

biij Initial experience in treating lung cancer with helical tomotherapy

Corporate Medical Policy

Flattening Filter Free beam

New Thinking on Fractionation in Radiotherapy

Lung Spine Phantom. Guidelines for Planning and Irradiating the IROC Spine Phantom. MARCH 2014

A treatment planning study comparing Elekta VMAT and fixed field IMRT using the varian treatment planning system eclipse

SBRT fundamentals. Outline 8/2/2012. Stereotactic Body Radiation Therapy Quality Assurance Educational Session

IGRT Protocol Design and Informed Margins. Conflict of Interest. Outline 7/7/2017. DJ Vile, PhD. I have no conflict of interest to disclose

Leila E. A. Nichol Royal Surrey County Hospital

IMAT: intensity-modulated arc therapy

Corporate Medical Policy

Radiobiological Impact of Planning Techniques for Prostate Cancer in Terms of Tumor Control Probability and Normal Tissue Complication Probability

Ranking radiotherapy treatment plans: physical or biological objectives?

IGRT Solution for the Living Patient and the Dynamic Treatment Problem

RADIOBIOLOIGCALLY BASED TREATMENT PLANNING: THE NEXT FRONTIER. Teddy LaMaster, MS

Derivation of mean dose tolerances for new fractionation schemes and treatment modalities

WHOLE-BRAIN RADIOTHERAPY WITH SIMULTANEOUS INTEGRATED BOOST TO MULTIPLE BRAIN METASTASES USING VOLUMETRIC MODULATED ARC THERAPY

THE TRANSITION FROM 2D TO 3D AND TO IMRT - RATIONALE AND CRITICAL ELEMENTS

Clinical Usage of Knowledge Based Treatment Planning

IMPACT OF FOUR-DIMENSIONAL COMPUTED TOMOGRAPHY PULMONARY VENTILATION IMAGING-BASED FUNCTIONAL AVOIDANCE FOR LUNG CANCER RADIOTHERAPY

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 6, 2015

Does the IMRT technique allow improvement of treatment plans (e.g. lung sparing) for lung cancer patients with small lung volume: a planning study

Protocol of Radiotherapy for Small Cell Lung Cancer

Isoeffective Dose Specification of Normal Liver in Yttrium-90 Microsphere Radioembolization*

Guidelines for the use of inversely planned treatment techniques in Clinical Trials: IMRT, VMAT, TomoTherapy

Linac or Non-Linac Demystifying And Decoding The Physics Of SBRT/SABR

Corporate Medical Policy

To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence

A Comparison of IMRT and VMAT Technique for the Treatment of Rectal Cancer

Reena Phurailatpam. Intensity Modulated Radiation Therapy of Medulloblastoma using Helical TomoTherapy: Initial Experience from planning to delivery

Baotian Huang 1, Lili Wu 1, Peixian Lin 2 and Chuangzhen Chen 1*

Vaginal Sparing with Volumetric Modulated Arc Therapy (VMAT) for Rectal Cancer. Scott Boulet BSc, RT(T)

PMP. Original Article. Introduction. Hee Jung Kim, Sung Yong Park, Young Hee Park, Ah Ram Chang

Cell survival following high dose rate flattening filter free (FFF) and conventional dose rate irradiation

A dosimetric comparison between volumetric-modulated arc therapy and dynamic conformal arc therapy in SBRT

Outline. Contour quality control. Dosimetric impact of contouring errors and variability in Intensity Modulated Radiation Therapy (IMRT)

A SIMPLE METHOD OF OBTAINING EQUIVALENT DOSES FOR USE IN HDR BRACHYTHERAPY

Response Evaluation after Stereotactic Ablative Radiotherapy for Lung Cancer

Impact of Contouring Variability on Dose- Volume Metrics used in Treatment Plan Optimization of Prostate IMRT

Dosimetric Effects of Using Generalized Equivalent Uniform Dose (geud) in Plan Optimization

The performance of the progressive resolution optimizer (PRO) for RapidArc planning in targets with low-density media

8/3/2016. Background. CT-Ventilation. Validation, Clinical Endpoints and Opportunities for CT Ventilation. 4DCT-Ventilation Imaging.

Radiotherapy physics & Equipments

Trajectory Modulated Arc Therapy: Application to Partial Breast Irradiation. Research and development to advance radiotherapy

Elekta - a partner and world-leading supplier

New Radiation Treatment Modalities in the Treatment of Lung Cancer

Evaluation of Three-dimensional Conformal Radiotherapy and Intensity Modulated Radiotherapy Techniques in High-Grade Gliomas

3D Pre-treatment Dose Verification for Stereotactic Body Radiation Therapy Patients

Clinical Implications Of Dose Summation And Adaptation

Development of an Expert Consensus Target Delineation Atlas for Thymic Cancers: Initial Quantitative Analysis of an Expert Survey.

Radiotherapy What are our options and what is on the horizon. Dr Kevin So Specialist Radiation Oncologist Epworth Radiation Oncology

A Novel Fluence Modified Base Plan IMRT for Second-irradiation Planning Technique

Stereotactic body radiotherapy for central lung tumors, yes we can!

A guide to using multi-criteria optimization (MCO) for IMRT planning in RayStation

Local control rates exceeding 90% have been reported using

REVISITING ICRU VOLUME DEFINITIONS. Eduardo Rosenblatt Vienna, Austria

Outline. Chapter 12 Treatment Planning Combination of Beams. Opposing pairs of beams. Combination of beams. Opposing pairs of beams

Transcription:

Received: 29 June 2018 Revised: 24 August 2018 Accepted: 27 August 2018 DOI: 10.1002/acm2.12481 TECHNICAL NOTE Re irradiation volumetric modulated arc therapy optimization based on cumulative biologically effective dose objectives Nevin McVicar 1 Steven Thomas 1 Mitchell Liu 2 Hannah Carolan 2 Alanah Bergman 1 1 Department of Medical Physics, BC Cancer Vancouver Cancer Centre, Vancouver, BC, Canada 2 Department of Radiation Oncology, BC Cancer Vancouver Cancer Centre, Vancouver, BC, Canada Author to whom correspondence should be addressed. Nevin McVicar E-mail: mcvicarn@rvh.on.ca Abstract The objective of this note is to introduce a clinical tool that generates ideal base plan dose distributions to enable re irradiation volumetric modulated arc therapy (VMAT) optimization based on cumulative biological effective dose objectives for specific organs at risk (OARs). The tool is demonstrated with a lung cancer case that required re irradiation at our clinic. First, previous treatment dose is deformed onto the retreatment computed tomography (CT) using commercial software. Then, the in house Matlab tool alters the deformed previous dose using radiobiological concepts on a voxel by voxel manner to generate an ideal base plan dose distribution. Ideal base plans that were generated using the in house Matlab tool were compatible with the Varian Eclipse treatment planning system. The tool enabled optimization of VMAT re irradiation plans using cumulative dose limits for OARs and all OAR cumulative dose objectives were met on the first optimization for the recurrent lung cancer case tested. PACS 87 KEY WORDS adaptive replanning, linear-quadratic model, radiobiology, VMAT dose optimization algorithms 1 INTRODUCTION High dose re irradiation has emerged as a feasible treatment option in lung cancer patients with locoregional relapse and few other treatment options. Re irradiation of locoregional relapse is becoming more common as lung cancer patients continue to live longer. Modern radiotherapy inverse plan optimization and delivery technologies such as intensity modulated radiotherapy (IMRT) and volumetric Abbreviations: BED, biologically effective dose; CT, computed tomography; DICOM, digital imaging and communications in medicine; DVH, dose-volume histogram; LQM, linear-quadratic model; OAR, organ at risk; PTV, planning target volume; SABR, stereotactic ablative radiotherapy; TPS, treatment planning system; VMAT, volumetric modulated arc therapy modulated arc therapy (VMAT) improve sparing of organs at risk (OARs) and make high dose re irradiation more feasible. Still, re irradiation introduces several complexities and plan optimization is cumbersome. For example, patient anatomy changes and dose accumulation to OARs must be accounted for. The challenge of changing patient anatomy has largely been met through rigid and deformable image registration (IR) algorithms that enable IR guided dose transfer between planning computed tomography (CT) scans. 1 In addition, computer software capable of converting voxel doses using the linear quadratic model (LQM) 2 to calculate biologically effective dose (BED) distributions is now commercially available. 3,4 Converting previously delivered physical dose to BED allows ---------------------------------------------------------------------------------------------------------------------------------------------------------------------- This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. 2018 BC Cancer Agency Vancouver Cancer Centre. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. J Appl Clin Med Phys 2018; 19:6:341 345 wileyonlinelibrary.com/journal/jacmp 341

342 MCVICAR ET AL. radiation oncologists to account for nonlinear biological response to differing dose per fraction. 4 Moreover, BED distributions are theoretically additive according to the LQM meaning distributions from separate courses of treatment can be summed to quantify cumulative BED for each voxel. 3 6 The mechanistic LQM is often considered to be over simplistic 7 ; however, it is almost universally used to adjust for fraction size because it is practical, biologically based and acceptably accurate for dose per fractions below 15 Gy. 2,8 Based on recent reports, a growing number of radiation oncologists are using cumulative BED distributions to guide their re irradiation plan evaluation. 3,6 Several institutions have reported OAR toxicity along with cumulative BED metrics in lung cancer patients receiving re irradiation, 9,10 with some instances of severe toxicities after delivering high cumulative dose to the aorta, 11,12 esophagus, 9,13 and lungs. 14 16 This clinical evidence has motivated radiation oncologists to prescribe cumulative BED limits for specific OARs. 3,6 While IR and dose accumulation techniques improve plan evaluation and clinical decision making, re irradiation plan optimization remains difficult because commercial treatment planning systems (TPSs) are not designed to optimize based on cumulative BED objectives as outlined below. Throughout VMAT optimization, TPSs display dose volume histograms (DVHs) for OARs and planning target volumes (PTVs). Visualization of evolving DVHs during optimization enables dosimetrists to influence the optimization by interactively adjusting dose objectives. The ability to monitor and adjust dose objectives throughout optimization is essential to the quality and efficiency of VMAT planning. 17 For initial irradiation plans, the evolving DVHs display total physical dose delivered over all planned fractions. For re irradiation optimization, TPSs allow for VMAT optimization using previously delivered dose as a base plan. Throughout optimization with a base plan, displayed DVHs result from the sum of base plan and optimizing plan dose distributions and are henceforth referred to as sum DVHs. Re irradiation plan optimization using previously delivered physical dose as a base plan may be useful when initial and re irradiation plans deliver the same dose per fraction. However, when significantly different dose per fractions are used, sum DVHs become nonsense since nonlinear radiobiological effects are not accounted for. Thus, plan optimization is cumbersome because dosimetrists lose the ability to adjust objectives based on evolving DVHs. Furthermore, after each optimization cycle, dosimetrists or physicists must convert dose to BED to determine whether cumulative BED objectives are satisfied. 3,6 Application of base plan dose is further limited in re irradiation scenarios because optimization of uniform target dose is impossible due to the presence of heterogenous base plan dose within the new PTV, which is henceforth referred to as PTV 2. In this technical note, we introduce a clinical tool that enables direct VMAT optimization of cumulative dose objectives using an ideal base plan dose that is compatible with a commercial TPS. Our proposed tool manipulates previously delivered dose to circumvent conventional base plan limitations and help guide VMAT optimization of re irradiation. Specifically, ideal base plans restore the ability to monitor whether cumulative dose objectives are achieved using the evolving sum DVHs throughout optimization. The tool is demonstrated retrospectively for a lung cancer case where re irradiation was prescribed. 2 METHODS AND RESULTS 2.A Patient data A patient with recurrent lung cancer previously treated and retreated at our clinic was used in this study. An initial course of VMAT was used to deliver a prescription of 60 Gy in 30 fractions to a primary lung PTV (PTV 1 ). Approximately 3 yr later, a second primary lesion was discovered and a course of stereotactic ablative radiotherapy (SABR) was prescribed to deliver 48 Gy in four fractions to the PTV 2. The clinically delivered initial course dose, referred to as D 1, was used in this study along with the initial and retreatment CTs, referred to as CT 1 and CT 2, respectively. To guide the optimization of the re irradiation dose distribution (D 2 ), a radiation oncologist retrospectively provided a list of OAR specific α/β ratios and prioritized cumulative dose limits which are listed in both physical dose (Gy per 4 fractions) and BED (Gy α/β ) in Table 1. 18 T ABLE 1 Summary of biological effective dose to organs at risk. Cumulative dose limits Priority Organ α/β BED metric BED 1 BED 1_onCT2 BED 2 BED sum BED (Gy α/β ) Physical (Gy/4 fx) 1 Spinal canal 2 D max (Gy 2 ) 47.2 46.9 12.1 47.1 D max < 110.5 D max < 26 2 PBT a 3 D max (Gy 3 ) 21.8 21.5 32.2 34.7 D max < 135.7 D max < 34.8 3 Esophagus 3 D max (Gy 3 ) 100.5 104.3 12.7 104.4 D max < 105 D max < 30 4 Lungs PTV b 3 V 53.3Gy3 (%) 3.5 1.8 1.3 3.3 V 53.3 < 10% V 20 < 10% 5 Heart 3 D max (Gy 3 ) 104.6 107.8 40.5 107.9 D max < 130.3 D max < 34 6 Skin 3 D max (Gy 3 ) 47.3 48.5 99.5 105.0 D max < 144 D max < 36 BED 1 doses are reported using CT 1 contours. BED 1_onCT2 doses are reported using CT 2 contours after warping BED 1 onto CT 2. BED 2 and BED sum doses are reported using CT 2 contours. a PBT, proximal bronchial tree. b Lungs PTV contours are different in CT 1 and CT 2 due to different PTVs.

MCVICAR ET AL. 343 F IG.1. Workflow for retreatment VMAT planning and evaluation: (a) Initial treatment dose (D 1 ) calculated on initial treatment planning CT 1. (b) D 1 warped onto re irradiation CT 2 to produce D 1_onCT2. (c) Dose within OARs is altered using radiobiological concepts to generate an ideal base plan. (d) Dose uniformity within retreat PTV 2 is achieved by setting dose to 0 Gy to produce D 1_baseplan. (e, f) Re irradiation VMAT plan is optimized using cumulative OAR and PTV dose objectives that account for base plan, D 1_baseplan. During optimization, sum DVHs accurately reflect whether dose objectives are achieved. (g) PTV 2 and OAR dose assessment using retreatment dose (D 2 ). (h) Finally, D 1_onCT2 and D 2 are converted to BED and summed to assess cumulative BED to OARs. In (b d) and (g h), PTV 2 is outlined in red. 2.B CT 1 to CT 2 deformable image registration An overview of the re irradiation planning workflow that incorporates the proposed tool is shown in Fig. 1. First, CT 1 to CT 2 deformable IR was performed and D 1 was warped onto CT 2 to generate D 1_onCT2 using a commercial intensity based IR algorithm (MIM Software Inc., Cleveland, OH, USA). As per clinical protocol, deformable IR was performed by a medical physicist and verified independently by a second medical physicist and a radiation oncologist. Deformable IR errors are not explicitly quantified or accounted for in this workflow. 2.C Generation of base plan The tool is an in house Matlab (The Mathworks Inc., Natick, MA, USA) program that manipulates D 1_onCT2 to generate an ideal base plan dose for re irradiation optimization, referred to as D 1_baseplan. The tool requires user specified initial course prescription (R x1 ) in Gy, the number of fractions for initial (n 1 ) and re irradiation (n 2 ) courses, and OARspecific priorities and α/β ratios. In addition, the tool requires a cumulative dose limit (D L ) specified in Gy per n 2 fractions (D L = d L n 2 ) for each OAR. The tool allows for one D L per OAR and D L must represent a maximum point dose limit (e.g., D max < D L ) or a maximum dose to a specified volume limit (e.g., V DL % OAR volume). The tool requires the digital imaging and communications in medicine (DICOM) structure file associated with CT 2 to label each voxel in D 1_onCT2 as a specific OAR or PTV 2. In cases of overlapping contours, voxels are labeled as the highest priority structure. Since TPSs display the sum DVH (D 1_baseplan + optimizing D 2 ) throughout optimization in Gy per n 2 fractions, D 1_baseplan voxel dose is also specified in Gy per n 2 fractions. 2.C.1 OAR base plan dose For each OAR, D 1_onCT2 is converted voxel by voxel to a physical dose in Gy per n 2 fractions equal to the radiobiological fraction of D L previously delivered by initial treatment. To do this, D L is converted to BED (BED L ) using the LQM formalism 2,19 : BED L ¼ d L n 2 1 þ d L : Similarly, D 1_onCT2 is converted to BED (BED 1 ) for all voxels. Then, for each voxel with BED 1 < BED L, the allowed re irradiation voxel dose is defined as D allowed = d allowed n 2, with d allowed calculated using the equality: BED L BED 1 BED allowed ¼ d allowed n 2 1 þ dallowed ; and then, D 1_baseplan voxel dose is set to D L D allowed. Here, we highlight two important points regarding voxels with BED 1 < BED L that receive exactly D allowed during re irradiation: (a) D 1_baseplan + D allowed = D L and therefore, these voxels are easily monitored on the sum DVH throughout optimization and (b) BED 1 + BED allowed = BED L ; hence, the sum DVH value at D L accurately portrays whether D L is satisfied or not. Furthermore, voxels with BED 1 < BED L that receive under D allowed will appropriately appear below D L on the sum DVH throughout optimization, whereas voxels with BED 1 < BED L that receive over D allowed appear above

344 MCVICAR ET AL. D L. Hence, the advantage to the dosimetrist is that evolving sum DVHs reliably illustrate whether cumulative dose objectives are met throughout optimization. For OARs where D L represents a maximum dose to a specified volume, some voxels will have BED 1 BED L. For these voxels, D 1_baseplan voxel dose is set to the isoeffective dose delivered in n 2 fractions, referred to as D 1_n2 where D 1_n2 = d 1_n2 n 2, with d 1_n2 calculated using the equality: 1 þ d 1 BED 1 ¼d 1 n 1 ¼d 1 n2 n 2 1 þ d 1 n2 The isoeffective dose conversion essentially scales D 1_onCT2 dose to the re irradiation fractionation such that it will correctly appear above D L on the sum DVH throughout the entire optimization. It must be recognized that none of the OAR sum DVH values represent true physical or biological dose when D 1_baseplan is used. However, each sum DVH may be regarded as a LQM scaled approximation that accurately reports whether D L is satisfied. 2.C.2 PTV 2 base plan dose All voxels labeled as PTV 2 are set to an arbitrary uniform dose in D 1_baseplan. For practical purposes, PTV 2 voxel doses are set to 0 Gy in this case. The uniform base plan dose in PTV 2 is required to enable optimization of a uniform re irradiation dose to PTV 2. Finally, D 1_baseplan is saved as a DICOM dose file, using the original dose files DICOM header information. The Matlab generated DICOM base plan dose files are then imported to the version 13.0 Eclipse TPS. 2.D Re irradiation VMAT optimization The re irradiation plan consisted of two full coplanar arcs placed at the center of PTV 2. For the recurrent lung case, D 1_baseplan was generated using tool inputs: R x1 /n 1 =60/30, n 2 = 4 along with all OARspecific priorities, α/β, and D L listed in Table 1. Optimization dose objectives were set according to prioritized cumulative dose limits in Gy per n 2 fractions listed in Table 1. Since PTV 2 voxel dose is 0 Gy in D 1_baseplan, PTV 2 lower dose objective is set to R x2 (48 Gy). 2.E. Re irradiation plan evaluation After optimization, D 2 and D 1_onCT2 were converted to BED using MIM software and the cumulative BED was calculated. Cumulative dose limits were satisfied and OAR specific sum BED metrics are reported in Table 1. For the re irradiation plan only, the percent volume of PTV 2 that received 100% of the R x2 (V 100% ) was 99.9% and the D max was 128% (61.6 Gy), which is acceptable for SABR treatment at our clinic. 3 DISCUSSION To our knowledge, this is the first strategy to use the previously delivered dose and radiobiological concepts to generate an ideal threedimensional base plan for re irradiation optimization. The impact of the ideal base plan on the final optimized dose distribution is outlined further here. In serial OARs where D L represents a maximum point dose limit, voxels that previously received higher doses will be preferentially spared compared to voxels that received lower doses in order to achieve cumulative D max < D L in each voxel. This effect is in agreement with other published re irradiation studies that limited cumulative maximum point dose for serial OARs such as aorta, esophagus, and spinal cord. Conversely, in parallel OARs where D L represents a maximum dose to a specified volume limit, voxels that previously received doses > D L will be preferentially irradiated compared to voxels that received doses < D L in order to minimize cumulative V DL. For example, when the Lung PTV dose objective is V 20Gy 10%, voxels in the lung that previously received >20 Gy are not spared during re irradiation optimization. Again, this effect is in agreement with previous findings that patients retreated with infield relapses experienced lower rates of pneumonitis compared with those retreated with out of field relapses, 14,16 which suggests that previously irradiated lung tissue may have fibrosis and be less susceptible to radiation pneumonitis. 20 Hence, our tool enables serial and parallel OARs to be optimized in different manners based on the type of dose objective, D L, used. Key limitations of our algorithm are that it only supports one objective per OAR and it cannot support cumulative mean dose objectives. As standard fractionation schemes (1.8 2 Gy per fraction) are replaced with hypofractionated stereotactic ablative radiotherapy (>6 Gy per fraction), oncologists increasingly rely on the LQM to derive optimal re irradiation fractionation schemes and OAR dose objectives. The use of LQM derived cumulative BED dose constraints is controversial because the additivity of BED distributions is theoretical and very difficult to validate clinically. In this work, the inherent uncertainty of LQM derived BED values is not explicitly quantified and cumulative BED distributions should only serve to compliment careful judgment of the treating physician. Moreover, the basic LQM formalism used in this technical note does not explicitly account for the impact of ablation and/or regeneration of surrounding vasculature and stromal tissue, which is especially relevant in re irradiation scenarios. It may be appropriate to adapt the proposed tool to employ more complex models (e.g., to account for tissue recovery between treatment courses) if it is deemed necessary by the treating physician. 7,21 Finally, planned re irradiation dose distributions must always be evaluated alone in addition to the cumulative BED distribution prior to re irradiation commencement. In this work, a tool was shown to facilitate optimization of VMAT re irradiation plans using cumulative dose limits for OARs. Ideal base plans generated using our tool were compatible with the Varian Eclipse TPS and similar tools could be developed for other TPSs that accommodate base plans. The presented tool may also be used for inverse planning of IMRT plans as long as base plans can be used during optimization. Although the underlying algorithm used to create the

MCVICAR ET AL. 345 ideal base plan is complex, the tool is clinically practical because it only requires simple OAR specific inputs to generate an ideal base plan for each patient. Furthermore, use of the ideal base plan eliminates timeconsuming iterations of plan optimization, dose conversion to BED and BED accumulation to verify whether cumulative dose objectives are achieved or not. Future work will aim to incorporate representation of dosimetric uncertainties associated with deformable IR errors during cumulative dose evaluation. 22 Furthermore, a script will be developed using Eclipse scripting Application Programming Interface to streamline clinical use of this tool. Finally, we will investigate potential improvements in planning time required and overall plan quality introduced by the proposed planning approach. CONFLICT OF INTEREST The authors declare no conflict of interest. REFERENCES 1. Pukala J, Johnson PB, Shah AP, et al. Benchmarking of five commercial deformable image registration algorithms for head and neck patients. J Appl Clin Med Phys. 2016;17:25 40. 2. Fowler JF. The linear quadratic formula and progress in fractionated radiotherapy. Br J Radiol. 1989;62:679 694. 3. Boman E, Kapanen M, Pickup L, Lahtela SL. Importance of deformable image registration and biological dose summation in planning of radiotherapy retreatments. Med Dosim. 2017;42:296 303. 4. Lee SP, Leu MY, Smathers JB, McBride WH, Parker RG, Withers HR. Biologically effective dose distribution based on the linear quadratic model and its clinical relevance. Int J Rad Oncol Biol Phys. 1995;33:375 389. 5. Jones B, Dale RG, Deehan C, Hopkins KI, Morgan DA. The role of biologically effective dose (BED) in clinical oncology. Clin Oncol. 2001;13:71 81. 6. Jumeau R, Péguret N, Zulliger C, Moeckli R, Bourhis J, Ozsahin EM. Optimization of re irradiation using deformable registration: a case study. BJR Case Rep. 2016;2:20150412. 7. Park C, Papiez L, Zhang S, Story M, Timmerman RD. Universal survival curve and single fraction equivalent dose: useful tools in understanding potency of ablative radiotherapy. Int J Radiat Oncol Biol Phys. 2008;70:847 852. 8. Brenner DJ. The linear quadratic model is an appropriate methodology for determining isoeffective doses at large doses per fraction. Sem Radiat Oncol. 2008;18:234 239. 9. Binkley MS, Hiniker SM, Chaudhuri A, et al. Dosimetric factors and toxicity in highly conformal thoracic reirradiation. Int J Radiat Oncol Biol Phys. 2016;94:808 815. 10. De Ruysscher D, Faivre-Finn C, Le Pechoux C, Peeters S, Belderbos J. High dose re irradiation following radical radiotherapy for nonsmall cell lung cancer. Lancet Oncol. 2014;15:e620 e624. 11. Evans JD, Gomez DR, Amini A, et al. Aortic dose constraints when reirradiating thoracic tumors. Radiother Oncol. 2013;106: 327 332. 12. Kilburn JM, Kuremsky JG, Blackstock AW, et al. Thoracic re irradiation using stereotactic body radiotherapy (SBRT) techniques as first or second course of treatment. Radiother Oncol. 2014;110: 505 510. 13. Griffioen GH, Dahele M, de Haan PF, van de Ven PM, Slotman BJ, Senan S. High dose, conventionally fractionated thoracic reirradiation for lung tumors. Lung Cancer. 2014;83:356 362. 14. Liu H, Zhang X, Vinogradskiy YY, Swisher SG, Komaki R, Chang JY. Predicting radiation pneumonitis after stereotactic ablative radiation therapy in patients previously treated with conventional thoracic radiation therapy. Int J Radiat Oncol Biol Phys. 2012;84: 1017 1023. 15. McAvoy S, Ciura K, Wei C, et al. Definitive reirradiation for locoregionally recurrent non small cell lung cancer with proton beam therapy or intensity modulated radiation therapy: predictors of highgrade toxicity and survival outcomes. Int J Radiat Oncol Biol Phys. 2014;90:819 827. 16. Ren C, Ji T, Liu T, Dang J, Li G. The risk and predictors for severe radiation pneumonitis in lung cancer patients treated with thoracic reirradiation. Radiat Oncol. 2018;13:69. 17. Otto K. Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys. 2008;35:310 317. 18. Videtic GM, Hu C, Singh AK, et al. NRG Oncology RTOG 0915 (NCCTG N0927): a randomized phase II study comparing 2 stereotactic body radiation therapy (SBRT) schedules for medically inoperable patients with stage I peripheral non small cell lung cancer. Int J Radiat Oncol Biol Phys. 2015;93:757. 19. Douglas BG, Fowler JF. The effect of multiple small doses of x rays on skin reactions in the mouse and a basic interpretation. Radiat Res. 1976;66:401 426. 20. Kelly P, Balter PA, Rebueno N, et al. Stereotactic body radiation therapy for patients with lung cancer previously treated with thoracic radiation. Int J Radiat Oncol Biol Phys. 2010;78:1387 1393. 21. Jones B, Hopewell JW. Alternative models for estimating the radiotherapy retreatment dose for the spinal cord. Int J Radiat Biol. 2014;90:731 741. 22. Vickress J, Battista J, Barnett R, Yartsev S. Representing the dosimetric impact of deformable image registration errors. Phys Med Biol. 2017;62:N391 N403.