Solving One of Chromatography s Biggest Dilemmas Proper Sealing of Chromatography Autosampler Vials

Similar documents
Achieving Secure Sealing and Instrument Handling in 9 mm Screw Thread Chromatography Vials

FT-Raman Surface Mapping of Remineralized Artificial Dental Caries

Increasing Extraction Efficiency of Wet Samples Using a Novel New Polymer During Accelerated Solvent Extraction

Measuring Phytosterols in Health Supplements by LC/MS. Marcus Miller and William Schnute Thermo Fisher Scientific, San Jose, CA, USA

The Investigation of Factors Contributing to Immunosuppressant Drugs Response Variability in LC-MS/MS Analysis

Dayana Argoti, Kerry Hassell, Sarah J. Fair, and Joseph Herman Thermo Fisher Scientific, Franklin, MA, USA

LC-MS/MS Method for the Determination of 21 Opiates and Opiate Derivatives in Urine

Results of Using Re-washed Vials and Closures

Enhanced LC-MS Sensitivity of Vitamin D Assay by Selection of Appropriate Mobile Phase

Detection and Quantification of Inorganic Arsenic in Fruit Juices by Capillary Ion Chromatography with Suppressed Conductivity Detection

Determination of Clinically Relevant Compounds using HPLC and Electrochemical Detection with a Boron-Doped Diamond Electrode

GC-MS/MS Analysis of Benzodiazepines Using Analyte Protectants

LC-MS/MS Method for the Determination of Tenofovir from Plasma

High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine

High-Throughput Quantitative LC-MS/MS Analysis of 6 Opiates and 14 Benzodiazepines in Urine

A Rapid UHPLC Method for the Analysis of Biogenic Amines and Metabolites in Microdialysis Samples

Thermo Scientific ART Pipette Tips

Impurity Profiling of Carbamazepine by HPLC/UV

Mercury Speciation Determinations in Asian Dietary Supplements

Quantitative Analysis of THC and Main Metabolites in Whole Blood Using Tandem Mass Spectrometry and Automated Online Sample Preparation

Comparison of Full Scan MS2 and MS3 Linear Ion Trap Approaches for Quantitation of Vitamin D

Nitrogen/Protein Determination in Starch by Flash Combustion using Large Sample Weight as an Alternative to the Kjeldahl Method

Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

Rapid and Direct Analysis of Free Phytosterols by Reversed Phase HPLC with Electrochemical Detection

PIF: Precursor Ion Fingerprinting Searching for a Structurally Diagnostic Fragment Using Combined Targeted and Data Dependent MS n

Gas Chromatography Syringe Selection Guide

A New Era in SFC Column Technology

Internal Calibration System of Thermo Scientific Varioskan Flash with Improved Sensitivity, Accuracy and Dynamic Range

Sialic Acid Determination in Glycoproteins: Comparison of Two Liquid Chromatography Methods. Experimental. Introduction. Method 1.

Carl Fisher, Terri Christison, Hua Yang, Monika Verma, and Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA, USA

Determination of Ethyl Sulfate Impurity in Indinavir Sulfate Drug Using Ion Chromatography

Analysis of Phospholipids in Natural Samples by Normal Phase HPLC and Corona Charged Aerosol Detection

Fast and easy separation of 23 drugs of abuse. including high, stable resolution of isobaric opioids from human urine by UHPLC-MS/MS

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine

Bio-Rad Laboratories. Hemoglobin Testing. VARIANT II Hemoglobin Testing System For HbA 1c

Improved Extraction and Analysis of Hexavalent Chromium from Soil and Water

Evaluation of an LC-MS/MS Research Method for the Analysis of 33 Benzodiazepines and their Metabolites

Marc Plante, Bruce Bailey, and Ian N. Acworth Thermo Fisher Scientific, Chelmsford, MA, USA

Determination of Phenolic Compounds in Apple Orchard Soil

Thermo Fisher Scientific, Sunnyvale, CA, USA; 2 Thermo Fisher Scientific, San Jose, CA, USA

Determination of Polymerized Triglycerides by High Pressure Liquid Chromatography and Corona Veo Charged Aerosol Detector

Thermo Scientific Finntip Pipette Tips. HANDHELD PIPETTING Finntip Pipette Tips. Details

Increased Efficiency of Biomolecule Identification by Optimization of Trypsin Digestion Buffers

HRM Series PCB Power Relays

Infectious Disease Testing. ULTRA Product Line. Safety is not a Matter of Chance

Thermo Scientific Finnpipette Manual Pipetting Systems. you can handle. anything

Enhancing Sequence Coverage in Proteomics Studies by Using a Combination of Proteolytic Enzymes

Ordering Information. ProteOn XPR36 System. Protein Interaction Analysis

Dynamic Analysis of HIV-Human Protein-Protein Interactions During Infection

PyroMark Q24 CpG MGMT Handbook

[ Care and Use Manual ]

Bio-Rad Laboratories HEMOGLOBIN Testing Bio-Rad A1c. VARIANT II TURBO Link System. Fully-Automated HbA 1c Testing

Use of a Tandem Mass Spectrometry Research Method for the Analysis of Amino Acids and Acylcarnitines in Dried Blood Spots

Fast UHPLC Methods for Analysis of Amino Acids

autosamplers ASI-100 and ASI-100T Autosamplers

Results you can trust

Bio-Rad Laboratories HEMOGLOBIN Testing Bio-Rad A1c. VARIANT II TURBO Link. Fully-Automated HbA 1c Testing

ApplicationNOTE EXACT MASS MEASUREMENT OF ACTIVE COMPONENTS OF TRADITIONAL HERBAL MEDICINES BY ORTHOGONAL ACCELERATION TIME-OF-FLIGHT.

Health Care Providers:

Novel Glycan Column Technology for the LC-MS Analysis of Labeled and Native N-Glycans Released from Proteins and Antibodies

Cardiac Assessment Controls

Integrated Targeted Quantitation Method for Insulin and its Therapeutic Analogs

Analysis of Emulsifiers in Foods by High Pressure Liquid Chromatography and Corona Charged Aerosol Detection

Immunohematology. IH-QC Modular System. Select. Combine. Control.

Determination of Carbohydrates in Kombucha Using HPAE-PAD

Chemical compatibility of the Pall QPoint TM Docking Station - Tap Assembly with common surface disinfectants

BioPlex 2200 Infectious Disease Panels

Injection Techniques Questionnaire (ITQ) WorldWide Results Needle Gauge

Providing a Universal, One-step Alternative to Liquid-Liquid Extraction in Bioanalysis

Discrete Industrial Analyzer System Reagents for Food and Other Sample Materials

Improving Food Process Control Using the

P6KE Series DO-15. Mechanical Data

Rapid and sensitive UHPLC screening for water soluble vitamins in sports beverages

Direct Determination of Native N-linked Glycans by UHPLC with Charged Aerosol Detection

PosterREPRINT INTRODUCTION. 2-D PAGE of Mouse Liver Samples. 2-D PAGE of E.coli Samples. Digestion / Cleanup. EXPERIMENTAL 1-D PAGE of BSA Samples

Determination of Inorganic Ions and Organic Acids in Non-Alcoholic Carbonated Beverages

ab HIV-1 Protease Inhibitor Screening Kit (Fluorometric)

NobelProcera Implant Bar Overdenture complete range of fixed and removable solutions.

SS3 Series. Controlled Avalanche Power Diodes. Features: Mechanical Data: SMC/DO-214AB. Page 1 28/03/06 V1.0

Heart Health Exploring the OTC options

Cardiac Assessment Controls A critical element of reliable cardiac testing

For personal use only

slge112 Molecular Allergology Product Characteristics ImmunoCAP ISAC slge 112

Rapid and sensitive UHPLC screening of additives in carbonated beverages with a robust organic acid column

NEW! Thermo Scientific ClipTip Pipetting Systems. sealed in security. feel the difference

A New HILIC/RP Mixed-Mode Column and Its Applications in Surfactant Analysis

PRODUCT INFORMATION. New Reagents for Dako CoverStainer. Choose the H&E staining intensity you want.

Separation of 37 Fatty Acid Methyl Esters Utilizing a High-Efficiency 10 m Capillary GC Column with Optimization in Three Carrier Gases

Characterization of Used Cooking Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Bio-Rad Laboratories. Cardiac Assessment Controls. Value Assigned for Clinical Laboratory and Point of Care Test Systems

Forensic and clinical products and services

Dako IT S ABOUT TIME. Interpretation Guide. Agilent Pathology Solutions. ALK, ROS1 and RET IQFISH probes (Dako Omnis) MET IQFISH probe (Dako Omnis)

Determination of Benzoate in Liquid Food Products by Reagent-Free Ion Chromatography

New Solvent Grade Targeted for Trace Analysis by UHPLC-MS

Determination of Clinically Relevant Compounds using HPLC and Electrochemical Detection with Boron-Doped Diamond Electrode

artus EBV QS-RGQ Kit Performance Characteristics May 2012 Sample & Assay Technologies Analytical sensitivity plasma

Testing robustness: Immunosuppressant drugs in blood with a TSQ Quantis MS for clinical research

Transcription:

Solving One of Chromatography s Biggest Dilemmas Proper Sealing of Chromatography Autosampler Vials D. Edwards 2, D. Lennartz 1, L.Shick 2 1 Thermo Fisher Scientific, Langerwehe, Germany 2 Thermo Fisher Scientific, Suwanee, GA, USA

Abstract Just one more turn. That has been the chromatographers answer to the age old question of how tight should I seal my autosampler chromatography vial closure. Unfortunately for a lot of chromatographers this answer has resulted in a myriad of chromatographic problems. Truth be told, Just one more turn is more than likely the absolute worst thing that one can do to seal an autosampler vial. For all current chromatographic screw thread autosampler vial products, the limit of tightening is reached when the septum can no longer be compressed between the vial rim and the top of the cap. Elastomeric seals are most effective when they are tightened to a point of 50-80% compression. Less than 50% compression risks not completely sealing the vial. Greater than 80% compression can cause septum extrusion and does not allow for the elasticity of the seal required for most effective performance. It is rarely the case that a closure will be tightened to the point where compression is less than 50%. Failures due to over compression of the septum are much more frequent. How tight is tight enough is very subjective and different for each operator. There are no tools to measure torque for a small closure such as the 9 mm cap. The Advance Vial Closure System (AVCS) removes the subjectivity out of achieving the optimal compression when sealing a vial. Designed as a complete system, AVCS allows the closure, septum and vial to work together to prevent compression from exceeding the optimal range. This paper will review the design features of AVCS and provide data to support how AVCS eliminates under/over compression of vial seals while improving robotic autosampler pick-up of vials, assuring more centered vial closure positioning, and greater optical recognition of vials placed in autosamplers. Introduction A study of the effects of the typical operator response to evaporative sample loss and septum dislodging during the use of standard 9 mm chromatography vial and closures was conducted. In this study, sample loss was measured for vials tightened to the perceived optimal degree were compared to vials slightly over-tightened beyond optimal, and to newly introduced vials designed to eliminate the common causes of evaporative sample loss. Methods All glass samples were placed into a 40 C incubator for 24 hours and cooled to room temperature prior to use. Room Temperature Evaporation Weight in grams was recorded for empty vials with caps loosely attached to a resolution of 0.001 g. Approximately 1.3 g of pure methanol was added to each vial and the cap was attached using the normal amount of torque required to achieve optimal sealing. The prepared samples were allowed to sit at room temperature for 1 hour. Initial filled weight was recorded for each vial as measured on an analytical balance to a resolution of 0.001 g The vials were returned to the rack and allowed to sit at room temperature for an additional 72 hours. Any change in appearance over the course of the incubation period was noted. After 72 hours the final room temperature weight was taken and subtracted from the final 40 C temperature weight to yield the room temperature sample loss in g. The cap on each reweighed vial was tightened by approximately one-fourth to one-half of a full turn and allowed to sit at room temperature. After 72 hours the final over torqued room temperature weight was taken and subtracted from the final 40 C temperature weight to yield the room temperature over torque sample loss in grams. High Temperature Evaporation Closures were readjusted on each vial to the optimal amount of torque. Initial high temperature weight was recorded as measured on a four place analytical balance to a resolution of 0.001 g. The vials were loaded into a sample rack and incubated at 40 C for 72 hours Final 40 C temperature weight was recorded and the difference from the initial high temperature weight calculated as sample loss in grams. Changes in vial weight representing evaporative sample loss were plotted for each vial in the normal and over-torqued position. 2 Solving One of Chromatography s Biggest Dilemmas Proper Sealing of Chromatography Autosampler Vials

Results Evaporative losses from standard 9 mm chromatography vials were measured when the vial closures were tightened to a point that would be considered optimal by most experienced chromatographers. In general the results were acceptable when adequate care is exercised during the attachment of the closure. It was found that for some samples in this group the degree of tightening that produced good sealing characteristics also made the product susceptible to septum dislodging. While standard chromatography vials available from a variety of sources showed only minor amounts of solvent evaporation under the optimal degree of tightening, the ability to sense the proper endpoint proved to be a challenge. There was a noticeable tendency on the part of some operators to tighten the vials past the optimal point. Figures 1a & 1b: Evaporative Loss from Standard 9 mm Chromatography Vials Carefully Tightened Thermo Scientifi c Poster Note PSGSC 0910 0610 3

When standard 9 mm chromatography vials were tightened beyond the optimal sealing point, either in an effort to counteract septum dislodging or to achieve a definite feel of closing, evaporative loss was more pronounced. The results of tests of over tightened vials are present in Figures 2a, 2b, 2c and 2d 4 Solving One of Chromatography s Biggest Dilemmas Proper Sealing of Chromatography Autosampler Vials

Figure 3 shows the test results from a new type of 9 mm vial developed using the Thermo Fisher Scientific Advance Vial Closure System (AVCS) Technology. When an AVCS closure is attached to one of the new SureStop 9 mm vials, there is a positive stopping point detected when optimal septum compression and sealing has occurred. Attempts to further tighten the closure does not cause the same deterioration of performance found with standard vials and closures. Septum dislodging was not detected in tests of this vial and closure combination under either the optimal or over tightened state. Thermo Scientifi c Poster Note PSGSC 0910 0610 5

Figure 4 below shows how over tightening a closure can cause septum displacement. The three vials on the left have been tightened approximately one-eigth of a turn past optimal and allowed to stand undisturbed. On the right is a Thermo Scientific SureStop vial with AVCS closure tightened to the maximum limit. It can be seen that the septum in the standard vial has shifted position under the cap to the extend that it is no longer providing an effective seal. The SureStop vial does not allow the septum to shift position maintaining good sealing properties. Discussion Investigations into evaporation of volatile sample components from 9 mm screw thread vials revealed that there is a significant effect resulting from how much the closure is tightened onto the vial. This magnitude of this effect varied from one product to another, but could always be observed on a few samples out of a larger group of 50 vials. For some vials, the point of optimal sealing did not sufficiently retain the septum so that dislodging during injection could be avoided. The lack of a positive stop when the vial is properly closed contributes to doubts about whether a volatile sample could evaporate while waiting for analysis. In addition it was observed that over tightening of the closure can cause the septum to be extruded from its proper position between the vial rim and cap further compromising vial sealing. In the current experiments a new type of vial and closure were evaluated for sealing under normal and attempted over tightened positions. Sealing of a chromatography vial is dependent on maintaining the correct amount of compression of the septum inside of the cap. The standard 9 mm products allow the septum to be compressed far beyond its functional optimum before the cap reaches a stopping point. The new Thermo Scientific SureStop vial with the associated AVCS closure technology assures that the position of optimum compression can be detected and provides effective sealing even when an attempt is made to exceed the proper end point Sample evaporative losses are minimal and extrusion of the septum is not observed under the full range of use for the vial and closure as a unit. Conclusion Evaporative losses from a variety of chromatography vials were found to be heavily dependent on the amount of torque applied to the closure when tightening it onto the vial. When using commonly available 9 mm vials, the end point for optimal sealing is subject to the judgment of the operator. Sample vials and caps employing a definite end point or stop for the optimal amount of torque applied to the samples provides a perceptive end point position for the closure onto the vial. The positive stop delivered when the closure is properly tightened gives the user complete confidence that the vial is closed in such a manner that an optimal seal has been attained and there is no fear of septum push through. Eliminating the effects of over-tightening, including septum push-through creates higher confidence in the quality of analytical results over standard vials. 6 Solving One of Chromatography s Biggest Dilemmas Proper Sealing of Chromatography Autosampler Vials

www.thermoscientific.com/dionex 2013 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Thermo Fisher Scientific, Sunnyvale, CA USA is ISO 9001:2008 Certified. Australia +61 3 9757 4486 Austria +43 1 333 50 34 0 Belgium +32 53 73 42 41 Brazil +55 11 3731 5140 China +852 2428 3282 Denmark +45 70 23 62 60 France +33 1 60 92 48 00 Germany +49 6126 991 0 India +91 22 2764 2735 Italy +39 02 51 62 1267 Japan +81 6 6885 1213 Korea +82 2 3420 8600 Netherlands +31 76 579 55 55 Singapore +65 6289 1190 Sweden +46 8 473 3380 Switzerland +41 62 205 9966 Taiwan +886 2 8751 6655 UK/Ireland +44 1442 233555 USA and Canada +847 295 7500 PSGSC 0910 0610_E 02/13S