Reduction of Radiation Exposure in Percutaneous Pedicle Screw Placement by Using a New Extension Instrument

Similar documents
THE UNIVERSITY OF OKLAHOMA HEALTH SCIENCES CENTER GRADUATE COLLEGE RADIATION DOSE ESTIMATION FOR DIAGNOSTIC MODALITIES

Radiation Safety for New Medical Physics Graduate Students

Exposure of the thyroid to radiation during routine orthopaedic procedures

The Fluoroscopic Technique for Monitoring Distraction of a Non-Invasive Lengthening Device in Early Onset Scoliosis

Accuracy of Percutaneous Pedicle Screw Placement in Thoracolumbar Spines: A Comparison Study Between Experienced. Spine Surgeons and Residents

IMAGE-GUIDED SPINE SURGERY

Why radiation protection matters?

3/5/2015. Don t Electrocute Me!: Common Misconceptions in Imaging and Radiation Safety (and What to Do About Them)

Biological Effects of Ionizing Radiation Module 8 - AAPM/RSNA Curriculum. Basic Radiation Biology

WHICH FACTORS PREDICT SHOULDER ASYMMETRY IN PATIENTS WITH LENKE TYPE 1 AND 3 CURVES FOLLOWING PEDICLE SCREW INSTRUMENTATION?

The x-rays produced penetrate the body which absorbs, refracts, or reflects the x-ray beam energy depending on the tissue. Bone

Three-dimensional imaging guided vertebral pedicle screw placement: accuracy analysis doi: /j.issn

Risk of ionising radiation to trainee orthopaedic surgeons

Managing the imaging dose during Image-guided Radiotherapy. Martin J Murphy PhD Department of Radiation Oncology Virginia Commonwealth University

5/19/2017. Interspinous Process Fixation with the Minuteman G3. What is the Minuteman G3. How Does it Work?

Ionizing Radiation Exposure of the Population of the United States. David A. Schauer Executive Director

Pre-operative Computed Tomographic(CT) Scan Templating in Adolescent Idiopathic Scoliosis(AIS): Is it Really Necessary

Pedicle screw insertion : robotic assistance versus conventional C-arm fluoroscopy

Modern orthopedic and spine surgeons strive to minimize

Percutaneous Posterior Fixation: A Unique Entity to minimize Further Damage to Patient with Traumatic Spine

Intraoperative case studies. Portable full body 32-slice CT scanner

Topic: Percutaneous Axial Anterior Lumbar Fusion Date of Origin: June Section: Surgery Last Reviewed Date: June 2013

Krueger-Gilbert Health Physics, Inc.

VIPER PRIME System Cadaver Time Study

Article Excerpts: Radiation protection concerns among staff performing Fluoroscopic procedures.

Introduction of FIREFLY Technology

Radiation Exposure of the Hand and Chest during C-arm Fluoroscopy-Guided Procedures

Ionizing Radiation. Michael J. Vala, CHP. Bristol-Myers Squibb

Disclosures. Sacroiliac Joint (SIJ) Pain. Modified Triplanar Fluoroscopic Approach in Percutaneous Fixation of Sacroiliac Joint 5/12/2017

MISS in Thoracolumbar Fractures

Evaluation of Personal Shields Used in Selected Radiology Departments

SCALING Radiographic Technique

Krueger Gilbert Health Physics, Inc.

Advantages of MISS. Disclosures. Thoracolumbar Trauma: Minimally Invasive Techniques. Minimal Invasive Spine Surgery 11/8/2013.

Adolescent Idiopathic Scoliosis

Utilize radiation safety principles to reduce the amount of radiation used to achieve desired clinical result.

Departement of Neurosurgery A.O.R.N A. Cardarelli- Naples.

An official journal of Neuro Spinal Surgeons Association, India

Chapter 16 Worksheet Code It

Radiation Safety & Determining Need for Radiographs

The print quality of this copy is not an accurate representation of the original.

Replacement Code for Interbody Cage for Disc

Single-Thread Screw Available in 25-45mm lengths (5mm Increments) Dual-Thread Screw Available in 30-45mm lengths (5mm Increments)

Patient Dosimetry in Mammography and Tomosynthesis:

Preliminary Experience With CT/Fluoro Matched Navigation In The Anterior Approach To The Thoracolumbar Spine

TRANSLAMINAR FACET SCREW FIXATION

Invivo Dosimetry for Mammography with and without Lead Apron Using the Glass Dosimeters

RPR 48B. APPLICATION TO USE DIAGNOSTIC EXTERNAL RADIATION IN HUMAN RESEARCH STUDIES

Radiation Protection- Cath lab

CT Radiation Risks and Dose Reduction

Bone Densitometry Radiation dose: what you need to know

Mapping the ASRT Objectives for Radiation Protection (47 Objectives) and Radiation Biology (21 Objectives) to this Text

Upper Limb Imaging Requirements

DETERMINATION OF ENTRANCE SKIN DOSE FROM DIAGNOSTIC X-RAY OF HUMAN CHEST AT FEDERAL MEDICAL CENTRE KEFFI, NIGERIA

Current trends in spinal surgery are directed toward

Radiation protection for patient and staff during routine EVAR and TEVAR procedures Miltos Matsagkas, MD, PhD, FEBVS

Robotic Spine Surgery [TASE]: MZOR

PEDICLE SCREW IMPACTATION IN THE AORTA AFTER DIRECT VERTEBRAL ROTATION

Out-of-field doses of CyberKnife in stereotactic radiotherapy of prostate cancer patients

TECHNICAL BROCHURE. Capture Facet Fixation System

Topic: Percutaneous Axial Anterior Lumbar Fusion Date of Origin: June Section: Surgery Last Reviewed Date: May 2014

2013 Charleston Swallowing Conference

FEE RULES RADIATION ONCOLOGY FEE SCHEDULE CONTENTS

Adolescent Idiopathic Scoliosis

EXAMINATION CONTENT SPECIFICATIONS ARRT BOARD APPROVED: JANUARY 2017 IMPLEMENTATION DATE: JULY 1, 2017

1 Course Syllabus + Study Guide for Lecture and Laboratory

STRUCTURED EDUCATION REQUIREMENTS IMPLEMENTATION DATE: JULY 1, 2017

Radiation Safety Characteristics of the NOMAD Portable X-ray System

Outcomes of an accelerated discharge pathway after spinal fusion

Radiation Exposure Intra-Operatively In Treating Humerus Shaft Fractures A Comparative Study between Plating And Nailing Group

Cost Analysis of Magnetically Controlled Growing Rods Compared with Traditional

Module Rhodes

in developing institutional policies, procedures, and /or protocols. The Canadian Society of

1105 two (2) vertebrae... 1, add on per additional vertebra

The Bone Densitometry Examination

Degenerative spondylolisthesis at the L4 L5 in a 32-year-old female with previous fusion for idiopathic scoliosis: A case report

Medical Errors in Radiation Therapy 2014

PediLoc 3.5mm and 4.5mm Contour Femur Plate Surgical Technique

Royal Oak IBFD System Surgical Technique Posterior Lumbar Interbody Fusion (PLIF)

Brachytherapy. What is brachytherapy and how is it used?

PHYS 383: Applications of physics in medicine (offered at the University of Waterloo from Jan 2015)

Vascular & Interventional Education Days Thomas M Griglock, Ph.D., DABR Chief Diagnostic Imaging Physicist, OHSU

Multiple Choice Identify the letter of the choice that best completes the statement or answers the question.

Clinical Reference Guide

Ballista Percutaneous Screw Placement System. Surgical Technique

KEY WORDS ultra-low radiation imaging; pedicle screw accuracy; minimally invasive; transforaminal lumbar interbody fusion

Case Study: Jordan. Conditions Treated Cleidocranial Dysostosis. Age Range During Treatment 13 Years 14 Years

The Most Reliable Spinal Solution

Bone Densitometry. Total 30 Maximum CE 14. DXA Scanning (10) 7

Pedicle screw placement accuracy in thoracic and lumbar spinal surgery with a patient-matched targeting guide: A cadaveric study

Navigated placement of iliac bolts: description of a new technique

ApiFix New minimal invasive method to treat Adolescent Idiopathic Scoliosis Short fixation followed by Specific Physiotherapy Program

SURGICAL TECHNIQUE. SECURIS Pedicle Screw System for Minimally Invasive Surgery. 2 I SECURIS Pedicle Screw System

Radiation Safety Characteristics of the NOMAD Portable X-ray System

A Patient s Guide to Intraoperative Monitoring

Measurement of Entrance Skin Dose and Calculation of Effective Dose for Common Diagnostic X-Ray Examinations in Kashan, Iran

created by high-voltage devices Examples include medical and dental x-rays, light, microwaves and nuclear energy

WHAT IS SMART SPINAL FUSION?

Transcription:

Reduction of Radiation Exposure in Percutaneous Pedicle Screw Placement by Using a New Extension Instrument Franquiz M. Juan, Myers J.Behnam, Thomas A. Michael, Greenhalgh J. Travis, Shoup J. Andrew Study Design: Experimental assessment of a method to reduce surgeon s radiation exposure by using an extension instrument during lumbar pedicle screw placement. Objective: To assess the reduction of radiation exposure to surgeons when using a new extension instrument (SurGenTec s ALARA radiation exposure reduction instrument). Summary of Background Data: Pedicle screw fixation is the gold standard for stabilization in spine surgery. Pedicle screw instrumentation may be performed using either a minimally invasive or open approach. Both techniques generally require the use of extensive fluoroscopy to help ensure the proper trajectory and placement of pedicle screws. There are many ways to dampen the radiation exposure including intraoperative navigation, intraoperative CT scan, and ultralow radiation imaging. These methods reduce the overall radiation exposure but are not cost effective and there is a learning curve to become proficient with these techniques. In addition, many of these expensive devices and machines are constantly being improved, so technology can expire quickly and become obsolete. Many surgeons attempt to reduce radiation exposure by using a Kocher clamp to keep further distance from the ionizing scatter radiation. Radiation exposure to medical staff has been a long-term concern when performing these procedures. Physicians can exceed their lifetime occupational radiation limit within the first decade of their career [1]. The main source of radiation exposure is the scatter radiation produced by the patient. Scatter radiation is higher in the proximity to the region of the patient getting the primary radiation beam. The surgeon s hands which are closer to the screw placement area generally receive the highest radiation exposure. High-dose or cumulative fluoroscopic x-ray radiation can cause DNA damage, hair loss, burns, and necrosis potentially causing the need for skin grafting [2]. While other sensitive tissues and organs such as the eyes, breasts, thyroid may also be affected, if in close proximity to the scatter radiation. Nearly half of all orthopedic procedures expose the physician to enough radiation to increase the risk of thyroid cancer [3]. By following ALARA (As low As Reasonably Achievable) recommendation, dose exposure should be reduced as much as possible. One simple

method is to extend the distance between surgeon s hands/body to the area exposed to the radiation beam. The purpose of this study is to assess the reduction of dose exposure to surgeons at different proximities to the patient. This will be performed using the ALARA extension instrument developed by SurGenTec (Boca Raton, FL) and compared to an often used Kocher clamp. Both instruments may be used to increase the distance between the surgeon and radiation source. The data reduction of radiation exposure is compared for long term exposures of 1, 3, 5 and 10 years. Methods: Radiation exposure was produced by a C-arm GEMS OEC 9800. Source to imaging distance (SID) is 100 cm. The procedure was simulated by using a kvp of 85 kv and 3.6 ma. No magnification was used. The procedure was performed in posterior-anterior (PA) and lateral (LAT) positions. In position PA, the source to skin distance (SSD) was 42 cm and the x-ray tube was below the table. In the LAT position the SSD was 42 cm. Radiation scattering was simulated with 4 acrylic sheets of 15 cm x 15 cm and total thickness of 8 cm. Acrylic sheets were placed in the center of the table and field of view of the x-ray tube. Radiation exposure was measured by using a calibrated RadCal ionization chamber model 10X6-180 (180 cc). Scattered radiation exposure was measured at four positions: A) Center of acrylic sheets. (This is the position of the needle during the procedure). B) Distance of 18 cm from the center of acrylic sheets. (This is the position of the Surgeon s hand at the Kocher clamp). C) Distance of 46 cm from the center of acrylic sheets. (This is the position of the Surgeon s hand at the end of SurGenTec's radiation reduction instrument). D) distance of 98 cm from the center of the acrylic sheets. (This is the position of the Surgeon's shoulder while grasping the end of SurGenTec's radiation reduction instrument. Radiation dose at this position could be considered as an approximation of the dose received by the thyroid gland and the lens of the eye). A bilateral/single level lumbar spine fusion procedure requires 4 pedicle screws. According to three separate studies, the average fluoroscopy (radiation) time it would take to access one pedicle and insert one screw was 35.5 seconds [4-6]. Assuming half of that time would be used to access the pedicle and half would be used to insert the screw. It would take 17.75 seconds of radiation exposure time to access one pedicle. Thus, the radiation exposure time would be a total of 71 seconds to access four pedicles. Assuming each surgeon performs three bilateral/single level lumbar spine instrumented fusions procedures weekly at an average of 144 per year. The total radiation exposure at surgeon s hand position at 1, 3, 5 and 10 years was calculated.

Table I. Experimental setup Radiation Machine C-arm GEMS OEC 9800 kvp 85 ma 3.6 Fluoroscopy time (per procedure) 71 s Scatter phantom Acrylic block of 15 x 15 x 8 cm Dosimetry results expressed in msv are valid only for this Experimental setup. Results Radiation dose to surgeon s hand is higher in the LAT position X-ray tube-image receptor. When using the SurGenTec extension instrument, the radiation dose is significantly reduced in both the PA and LAT positions. In the PA position, SurGenTec extension instrument(c=46cm) reduces the dose of a surgeon s hand by approximately 15 times compared to using a Kocher clamp (b=18 cm). The difference of dose between the position of the needle (d = 0) and the surgeon s hand in the SurGenTec extension instrument(c=46cm) is 293 times less radiation exposure.

Table II. Radiation dose. X ray tube-image receptor in PA position Dose per Position procedure Dose 1 year Dose 3 years Dose 5 years Dose 10 years A (d = 0) 0.778 112 336 560.0 1120.0 B (d=18 cm) 0.04 5.8 17.62 29.2 58.6 C (d = 46 cm) 0.002 0.38 1.14 1.92 3.82 D (d = 98 cm) 0.00048 0.070 0.210 0.35 0.70 A: Needle position; B: Surgeon s hand when using Kocher clamp; C: Surgeon s hand when using SurGenTec extension clamp; D: Surgeons shoulder while grasping the end of SurGenTec s radiation reduction instrument (this position is representative of dose to lens of the eye and thyroid gland). Table III. Radiation dose. X -ray tube-image receptor in LAT position Dose per Position procedure Dose 1 year Dose 3 years Dose 5 years Dose 10 years A (d = 0) 0.072 10.5 31.4 52.4 105 B (d=18 cm) 0.044 6.42 19.42 32.2 64.2 C (d = 46 cm) 0.010 1.50 4.48 7.46 14.94 D (d = 98 cm) 0.004 0.62 1.84 3.1 6.2 A: Needle position; B: Surgeon s hand when using Kocher clamp; C: Surgeon s hand when using SurGenTec extension clamp; D: Surgeon s shoulder while grasping the end of SurGenTec s radiation reduction instrument (this position is representative of dose to lens of the eye and thyroid gland).

Conclusions There is a significant reduction of ionizing radiation exposure when using the SurGenTec extension instrument versus tradition methods such as using a Kocher clamp for pedicle screw placement. The use of SurGenTec extension instrument will significantly decrease the radiation exposure to the surgeon s hands, eyes, thyroid, and the rest of their body when used properly. When using the SurGenTec extension instrument in the PA position, 15 times more radiation exposure procedures can be performed to get the same hand s dose received when using the Kocher clamp. When using the SurGentec extension instrument there would be minimal radiation exposure to the body(thyroid gland, lens of eye, etc) due to the distance (98cm) between the radiation source and the surgeons body. This SurGenTec extension instrument is a simple and cost effective option to reduce short term and long term radiation exposure during pedicle screw placement. References 1. Ul Haque, M & Shufflebarger, HL & O'Brien, M & Macagno, A. Radiation exposure during pedicle screw placement in adolescent idiopathic scoliosis: is fluoroscopy safe?. Spine 2006; 1; 31(21):2516-20. doi: 10.1097/01.brs.0000238675.91612.2f 2. Balter, S & Hopewell, JW & Miller, DL & Wagner, LK & Zelefsky, MJ. Fluoroscopically Guided Interventional Procedures: A Review of Radiation Effects on Patients' Skin and Hair. Radiology 2010; 254:2. 3. Devalia, KL & Peter, VK & Madanur, MA & Braithwaite, IJ. Exposure of the thyroid to radiation during routine orthopaedic procedures. Acta Orthop Belg. 2006 Oct; 72(5), 615-620. 4. Mroz, Thomas E. & Abdullah, Kalil G. & Steinmetz, Michael P. & Klineberg, Eric O. & Lieberman, Isador H. Radiation exposure to the surgeon during percutaneous pedicle screw placement. J Spinal Disord Tech. 2011 Jun; 24(4): 264 267. doi: 10.1097/BSD.0b013e3181eed618 5. Gwynne-Jones, David & A. Robertson, Peter & Lunt, Brian & Jackson, Suzanne. (2000). Radiation Exposure During Fluoroscopically Assisted Pedicle Screw Insertion in the Lumbar Spine. Spine. 25. 1538-41. doi: 10.1097/00007632-200006150-00013. 6. David C. Briski, Bradford S. Waddell, Brandon W. Cook, and Joseph M. Zavatsky. Reduction in Fluoroscopy Time, Radiation Dosage, and Complications Using an Innovative Guide Wire for Percutaneous Pedicle Screws. Global Spine Journal 2015; 05 A201. doi: 10.1055/s-0035-1554305