Testing the Accuracy of ECG Captured by Cronovo through Comparison of ECG Recording to a Standard 12-Lead ECG Recording Device

Similar documents
ISSN: ISO 9001:2008 Certified International Journal of Engineering and Innovative Technology (IJEIT) Volume 2, Issue 10, April 2013

Basic Dysrhythmia Interpretation

ECG UPDATE COURSE DESCRIPTION

Evaluation of the Accuracy of ECG Captured by CardioChip through Comparison of Lead I Recording to a Standard 12-Lead ECG Recording Device

HST-582J/6.555J/16.456J-Biomedical Signal and Image Processing-Spring Laboratory Project 1 The Electrocardiogram

Panorama. Arrhythmia Analysis Frequently Asked Questions

CHAPTER IV PREPROCESSING & FEATURE EXTRACTION IN ECG SIGNALS

ECG Signal Analysis for Abnormality Detection in the Heart beat

Miscellaneous Stuff Keep reading the Outline

SPECTRAL ANALYSIS OF LIFE-THREATENING CARDIAC ARRHYTHMIAS

Appendix D Output Code and Interpretation of Analysis

DIFFERENCE-BASED PARAMETER SET FOR LOCAL HEARTBEAT CLASSIFICATION: RANKING OF THE PARAMETERS

Speed - Accuracy - Exploration. Pathfinder SL

ECG Rhythm Analysis by Using Neuro-Genetic Algorithms

cvrphone: a Novel Point-of-Care Smartphone Based System for Monitoring the Cardiac and Respiratory Systems

Vital Responder: Real-time Health Monitoring of First- Responders

ECG interpretation basics

2017 BDKA Review. Regularity Rate P waves PRI QRS Interpretation. Regularity Rate P waves PRI QRS Interpretation 1/1/2017

ECG Beat Recognition using Principal Components Analysis and Artificial Neural Network

Signal Processing of Stress Test ECG Using MATLAB

Lab Activity 24 EKG. Portland Community College BI 232

Proceedings of the World Small Animal Veterinary Association Sydney, Australia 2007

Performance Identification of Different Heart Diseases Based On Neural Network Classification

Dr.Binoy Skaria 13/07/15

Reading Assignment (p1-91 in Outline ) Objectives What s in an ECG?

An ECG Beat Classification Using Adaptive Neuro- Fuzzy Inference System

Electrocardiography for Healthcare Professionals

ECG Signal Characterization and Correlation To Heart Abnormalities

ECG ABNORMALITIES D R. T AM A R A AL Q U D AH

ECG Underwriting Puzzler Dr. Regina Rosace AVP & Medical Director

The Electrocardiogram

Removal of Baseline Wander from Ecg Signals Using Cosine Window Based Fir Digital Filter

Temporal Analysis and Remote Monitoring of ECG Signal

ABCs of ECGs. Shelby L. Durler

Rate: The atrial and ventricular rates are equal; heart rate is greater than 100 bpm (usually between bpm).

Electrocardiography I Laboratory

The Efficient and Smart Methods for Diagnosis of SVT 대구파티마병원순환기내과정병천

Biomedical Signal Processing

Study methodology for screening candidates to athletes risk

Interpreting Electrocardiograms (ECG) Physiology Name: Per:

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev

DETECTION OF HEART ABNORMALITIES USING LABVIEW

Electrocardiography for Healthcare Professionals

Automatic Detection of Abnormalities in ECG Signals : A MATLAB Study

Assessment of Reliability of Hamilton-Tompkins Algorithm to ECG Parameter Detection

POWER EFFICIENT PROCESSOR FOR PREDICTING VENTRICULAR ARRHYTHMIA BASED ON ECG

An electrocardiogram (ECG) is a recording of the electricity of the heart. Analysis of ECG

The Electrocardiogram part II. Dr. Adelina Vlad, MD PhD

CORONARY ARTERIES HEART

EKG Intermediate Tips, tricks, tools

Rigel UNI-SiM Lite. The most cost-effective patient simulator on the market. Key Features n Compact and cost-effective

EKG Competency for Agency

Supraventricular Arrhythmias. Reading Assignment. Chapter 5 (p17-30)

Outline. Electrical Activity of the Human Heart. What is the Heart? The Heart as a Pump. Anatomy of the Heart. The Hard Work

NOISE DETECTION ALGORITHM

12-Lead ECG Interpretation. Kathy Kuznar, RN, ANP

LABVIEW based expert system for Detection of heart abnormalities

-RHYTHM PRACTICE- By Dr.moanes Msc.cardiology Assistant Lecturer of Cardiology Al Azhar University. OBHG Education Subcommittee

Electrocardiography Biomedical Engineering Kaj-Åge Henneberg

Comparison of ANN and Fuzzy logic based Bradycardia and Tachycardia Arrhythmia detection using ECG signal

PCA Enhanced Kalman Filter for ECG Denoising

A Review on Arrhythmia Detection Using ECG Signal

ECG INTERPRETATION MANUAL

DR QAZI IMTIAZ RASOOL OBJECTIVES

INTERNATIONAL JOURNAL OF ELECTRONICS AND COMMUNICATION ENGINEERING & TECHNOLOGY (IJECET)

Simulation Based R-peak and QRS complex detection in ECG Signal

UNDERSTANDING YOUR ECG: A REVIEW

Family Medicine for English language students of Medical University of Lodz ECG. Jakub Dorożyński

Assessment of ECG frequency and morphology parameters for automatic classification of life-threatening cardiac arrhythmias

11/18/13 ECG SIGNAL ACQUISITION HARDWARE DESIGN. Origin of Bioelectric Signals

Figure 2. Normal ECG tracing. Table 1.

SSRG International Journal of Medical Science ( SSRG IJMS ) Volume 4 Issue 1 January 2017

Cardiac Telemetry Self Study: Part One Cardiovascular Review 2017 THINGS TO REMEMBER

Study Day 1 Study Days 2 to 9 Sequence 1 Placebo for moxifloxacin Study Days 2 to 8: placebo for pazopanib (placebopaz) 800 mg;

Defibrillator. BiomedGuy

Diploma in Electrocardiography

MARS Ambulatory ECG Analysis The power to assess and predict

5- The normal electrocardiogram (ECG)

Cardiac Arrhythmias in Sleep

ID: BFH_ Start time: :00:00. Examination summary

CORONARY ARTERIES. LAD Anterior wall of the left vent Lateral wall of left vent Anterior 2/3 of interventricluar septum R & L bundle branches

Biomedical Signal Processing

Diagnostic capabilities of the implantable therapeutic systems

TELEMETRY BASICS FOR NURSING STUDENTS

Removal of Baseline wander and detection of QRS complex using wavelets

Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO

Assessing Arrhythmia Performance ST/AR Algorithm

EEG, ECG, EMG. Mitesh Shrestha

TEST BANK FOR ECGS MADE EASY 5TH EDITION BY AEHLERT

Powerline Interference Reduction in ECG Using Combination of MA Method and IIR Notch

SVT Discriminators. Definition of SVT Discrimination. Identify which patient populations might benefit from these features

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing

NHA Certified EKG Technician (CET) Test Plan for the CET Exam

BEDSIDE ECG INTERPRETATION

HRV ventricular response during atrial fibrillation. Valentina Corino

RN-BC, MS, CCRN, FAHA

IJRIM Volume 1, Issue 2 (June, 2011) (ISSN ) ECG FEATURE EXTRACTION FOR CLASSIFICATION OF ARRHYTHMIA. Abstract

Rigel UNI-SiM. The world s smallest integrated NiBP, SpO2 and patient simulator. Key Features n Compact and cost-effective

ECG Interpretation Cat Williams, DVM DACVIM (Cardiology)

If the P wave > 0.12 sec( 3 mm) usually in any lead. Notched P wave usually in lead I,aVl may be lead II Negative terminal portion of P wave in V1, 1

Transcription:

Testing the Accuracy of ECG Captured by through Comparison of ECG Recording to a Standard 12-Lead ECG Recording Device Data Analysis a) R-wave Comparison: The mean and standard deviation of R-wave amplitudes for the two devices were calculated and evaluated for significant difference using a paired t-test and a pairwise linear correlation analysis. b) Validation of ECG Signal Quality: Lead I ECG of each participant from Smart watch and Philips Pagewriter Trim III was also saved for offline analysis. Philips ECG were first resampled from 500 Hz to match the sampling rate of at 512 Hz, then two ECGs were preprocessed to remove noises or artifacts which contaminated the ECG raw signal, such as powerline interference, EMG noise, and baseline wander, using s custom algorithms. After the preprocessing, the ECG data were converted to voltage range. The correlation coefficient, power spectral density, and the magnitude squared spectral coherence were calculated. The Results Of the 60 subjects, two cardiovascular doctors found that the morphology of QRS complex, P wave, and T wave to be the same between two device ECG recordings, though ECG had more baseline noise than Philips ECG as illustrated in the example shown in Figure 1. They thought that ECG was accurate for ECG assessment and ECG arrhythmias could be accurately diagnosed by Lead I ECG tracings. Phillips Fig. 1. Effect of baseline noise on ECG

A variety of ECG arrhythmias were found among 60 subjects, as listed in Table 1. Of these, 27/60 (45%) subjects were detected having abnormal ECG, and 5/60 (8%) subjects were detected more than one ECG arrhythmia. Clinical Cases Table 1. Detected arrhythmia Several exemplary clinical cases demonstrated that Lead I ECG enable to detect significant ECG abnormalities. The first case was sinus bradycardia with right bundle branch block (RBBB), manifesting slow heart rate = 49, broad QRS complex > 120ms, and wide, slurred S wave, shown in Figure 2. The second case was atrial fibrillation associated with a premature ventricular contraction (PVC), characterized by varied heart rate and one premature ventricular beat, shown in figure 3. And figure 4 shows a case of sinus tachycardia, with heart rate > 110. Standard Reading Fig. 2. ECG of sinus bradycardia with right bundle branch block (RBBB).

Standard Fig. 3. ECG of atrial fibrillation associated with premature ventricular contraction (PVC). Standard Fig. 4. ECG of sinus tachycardia.

R-wave comparison For each subject s printed copies of both Lead I and Philips Pagewriter Trim III Lead I ECG, R-peak amplitude was measured to the nearest tenth of a millimeter on the corresponding pairs of QRS complexes. The mean and standard deviation of R-peak amplitudes for and Philips ECG recordings were 0.72/0.24 mv and 0.71/0.24 mv, respectively. Of the 60 subjects, eight subjects R-peak amplitude of were different to that of Philips, six subjects 0.1 mv higher than Philips, one subject 0.2 mv higher than Philips, and one subject 0.1 mv lower than Philips. Paired t test (MatLab R2013b Statistics Toolbox) resulted in p<0.05, indicating statistically non-significant amplitude difference. Pairwise linear correlation coefficient further suggested the similarity in R-peak amplitudes from two devices, r=0.9855, p<0.0001. Validation of ECG Signal Quality Offline analysis was applied to validate the similarity of ECGs from both devices, in both time domain and frequency domain. The middle 8-second data in the 10-second recording were used to calculate correlation coefficient, power spectral, and coherence. Coherence was calculated the mean value between 5 Hz and 25 Hz which contains the major frequency components of ECG signal. Figures 5 8 take one subject s plots of ECG waveforms, power spectral, and coherence as an example.

Fig. 6. Power spectral of ECG (red, solid) and Philips ECG (blue, dashed) Fig. 5. Preprocessed ECG (red, solid) and Philips ECG (blue, dashed), and zoomed view.

Fig. 7. Coherence function between ECG and Philips ECG The mean value of correlation coefficients for all subjects is 0.942, shown in Figure 8. Among total 60 Subjects, 52 subjects (87%) had correlation coefficient over 0.9, 5 subjects (8%) between 0.85 and 0.9, and 3 subjects (5%) less than 0.85 but greater than 0.8. Examining the ECG waveforms of those having low correlation coefficients found that baseline noise contaminating ECG was a major reason. Presumably, dry electrodes of are more sensitive to change of skinelectrode impedance than gel-based electrodes used by medical devices. The mean value of coherence for all 60 subjects was 0.945, 55 subjects (92%) above 0.9, 5 subjects (8%) below 0.9, as shown in Figure 9. Offline analysis indicated the high correlation between the two ECGs. CONCLUSIONS Philips Pagewriter Trim III has been FDA approved and widely used for years in medical care environments, as a valid ECG recording tool for cardiac conditions diagnosis. Comparing ECG recorded by this device and by the s ECG produced no significant difference both clinically and statistically. The high similarity and correlation of the two recordings were further proofed by offline analysis. Therefore, s ECG will accurately measure ECG and allow users to learn about and characterize their heart rate and rhythms. Widespread use of the technology could improve public awareness of heart health and identify unknown cardiac conditions at early stage.

Fig. 8. Correlation coefficients of each subjects Fig.9. Coherence of each subject