MRI GUIDELINES FOR INSPIRE THERAPY

Similar documents
MRI GUIDELINES FOR INSPIRE THERAPY

MRI Procedure Information. For St. Jude Medical MR Conditional Neurostimulation Systems. Clinician's Manual

MRI Procedure Information. For St. Jude Medical MR Conditional Neurostimulation Systems. Clinician's Manual

Algovita. MRI Procedure Guidelines. Spinal Cord Stimulation System. 201x

Read this manual before performing an MRI scan on a patient implanted with an Algovita Spinal Cord Stimulation System.

MRI Guidelines for the Axonics Sacral Neuromodulation System

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500)

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500)

ImageReady MRI Head Only Guidelines for Spectra WaveWriter Spinal Cord Stimulator System

ImageReady MRI Full Body Guidelines for Precision Montage MRI Spinal Cord Stimulator System

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System

VNS Therapy System Physician s Manual

MRI guidelines for Medtronic neurostimulation systems for chronic pain

Micra MC1VR01. MRI procedural information for Micra Model MC1VR01 MR Conditional single chamber transcatheter pacing device (VVIR)

Advisa MRI and Ensura MRI SureScan pacing systems

MRI-Ready Systems Manual. MRI Procedure Information for the St. Jude Medical MR Conditional System

Warning: Patients should not be exposed to MRI environments until the surgical site following pump implantation is fully healed.

MRI-Ready Systems Manual. MRI Procedure Information for the St. Jude Medical MR Conditional System

*smith&nephew. MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG. Knee Implants

*smith&nephew. MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG. Shoulder Implants

Cochlear Implants. Medical Procedures. for MED EL CI Systems. AW33290_5.0 (English US)

Cochlear Implants. Medical Procedures. for MED EL Implant Systems. AW33290_2.0 (English US)

MRI Safety Information for the HiRes Ultra 3D Cochlear Implant

MRI-Ready Systems Manual. MRI Procedure Information for the St. Jude Medical MR Conditional System

Version überholt. Nicht verwenden. Version obsolète. Ne pas utiliser. Versione obsoleta. Non utilizzare. Verouderde versie. Niet gebruiken.

Before Your MRI FOR PATIENTS WITH MEDTRONIC NEUROSTIMULATION SYSTEMS FOR CHRONIC PAIN

MRI SureScan pacemakers & ICDs what you need to know

MRI for Patients with MRI-Conditional Pacing System: Radiographers Role. Lawrance Yip DM, DR, QMH

*smith&nephew. MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG. Hip Implants

2014 AAPM 56 th Annual Meeting

Low-Profile Wrist Fixator. For stabilization of fractures of the distal radius.

REFILL KIT For use with Prometra Programmable Infusion Systems

PATIENT MANUAL. Inspire IV Generator Model Rx Only

Tell the doctor who prescribed your MRI scan that you have an implanted Medtronic neurostimulation system.

Q & A: How to Safely Scan Patients with a SureScan MRI Pacemaker

Parkinson s disease, Essential Tremor and primary dystonia

PATIENT PROGRAMMER 37642

MRI Guidelines for VNS Therapy

FastFrame External Fixation System. Damage Control Surgical Technique

BEFORE AND AFTER YOUR MRI SCAN. For patients with St. Jude Medical neurostimulation systems for chronic pain

WallFlex Biliary RX Partially Covered Stent System Prescriptive Information

Safety Issue of Static Magnetic Field (1)

Cardiac Rhythm Management // Manual // ProMRI. ProMRI. MR conditional device systems

AspireSR. Essential information about your new device

Cardiac Rhythm Therapy // Advanced Features // ProMRI. ProMRI. Checklist and Quick Reference Guide

Magnetic resonance imaging safety of NucleusR 24 cochlear implants at 3.0 T

Medium External Fixator Pediatric Femoral Shaft Frame. Using medium multi-pin clamps.

Yasargil Permanent Aneurysm Clips

SJM MRI ACTIVATOR HANDHELD DEVICE WORKFLOW Model: EX4000. SJM-EDTR Item approved for U.S. use only.

External Distal Radius Fixator. Supplement to the 8 mm rod fixator system

USING YOUR INTELLIS IMPLANTABLE SPINAL CORD STIMULATION SYSTEM

PATIENT CONTROLLER ST. JUDE MEDICAL UNDERSTANDING IMPORTANT FEATURES ON YOUR

Call Medtronic at 1 (800) to verify the patient s current implanted system

MRI Information for Medtronic Spinal and Biologics Devices

The Calcaneal Plate. The Synthes non-locking solution for the Calcaneus.

DISTAL RADIUS. Instructions for Use

Medium External Fixator Humeral Shaft Frame. Modular frame for upper extremity use.

LCP Metaphyseal Plates. For extra-articular fractures.

Medtronic, Inc.

Prometra Programmable Pump System Brief Summary

STANDARD OPERATING PROCEDURE (SOP)

MRI SURESCAN SYSTEMS. Patient Scanning Process PATIENT PRESCREENING. SureScan Pacing, Defibrillation, and CRT-D Systems Verification

Electrodes (for TransAeris System)

PARKINSON S SYMPTOM TRACKER

IMAGEREADY MR CONDITIONAL PACING SYSTEM

The VNS Therapy Magnet

Button Plate. Reinforcement for transosseous fixations.

Pocket Adaptor Kit FOR SPINAL CORD STIMULATION. Tip Cards

MRI Checklist for MED EL CI and ABI models

Quick Reference Guide

Bioness Inc Rye Canyon Loop Valencia, CA USA Telephone: or Website: Rx Only Bioness Inc.

THN. Sleep Therapy Study. ImThera. Information for Participants. Caution: Investigational device. Limited by United States law to investigational use.

Revision A Date:

Quick Reference Guide

Quick Reference Guide

Sterile-Packaged Large External Fixator Kits. For treatment of long bone and pelvic fractures that require external fixation.

Department of Radiology

How to Record & Send Information

Inspire Medical Systems. Physician Billing Guide

EasyStep. Operative technique

Annual Associate Safety Module Magnetic Resonance Imaging (MRI)

Out of pain. Into relief.

MRI Exams in Patients with Non MR Conditional Pacemakers and with MR Conditional Pacemakers

CARDIAC DEVICE MR-CONDITIONAL PRODUCT SUMMARY CHART

Prodigy Neurostimulation System Programming and Reference. Models 3855, Clinician's Manual

GUIDELINE 2-G-004 (formerly APPENDIX 3) HSREB (March 2005) Page 1

User Manual: Ives MR Conditional Cup Electrodes J.R. Ives, December 22, 2017

An Introduction to VNS Therapy

TAKE CONTROL OF YOUR FUTURE

An Introduction to VNS Therapy

Replacement Heart Valve Product Description (Stented Tissue) Models Reference

USING YOUR IMPLANTED SPINAL CORD STIMULATION SYSTEM

The table below shows the density and velocity of waves in two different substances. Density / kg m 3 Velocity / m s 1

CATHETER REVISION KIT (REF 11830) For use with Prometra Programmable Infusion Systems

VariAx Compression Plating System. Operative technique

JRI Thompson Hemiarthroplasty

Yes, out of sight out of mind.

Distal Radius Plate 2.4/2.7 dorsal and volar

ProPocket TM. User Guide

PATIENT CARE AND MRI SAFETY Module 7

Transcription:

MRI GUIDELINES FOR INSPIRE THERAPY Clinician s Manual ONLY

The following is a trademark of Inspire Medical Systems, Inc.: Inspire

Introduction Read the information in this manual prior to conducting an MRI scan on a patient with an implanted Inspire Medical Systems upper airway stimulation device. This manual contains information about the components that comprise the MRI Conditional system, applicable warnings and precautions related to the MR Conditional system, and the requirements that you must follow in order for the implanted Inspire system to be conditionally safe for MRI scans. Refer to the System Implant Manual for non-mri related information. If you have any questions, contact Therapy Support. See Therapy Support on page 7. NOTE: Before conducting an MRI scan, always ensure that you are using the most recent version of these MRI procedures. Contact Therapy Support or get the most recent version online at manuals.inspiresleep.com. Symbols and Definitions The following symbols may be used in the document and on some of the products and packaging: Symbol Description Caution, Consult Accompanying Documents Magnetic Resonance (MR) Conditional. Magnetic Resonance (MR) Unsafe. Manufacturer For USA audiences only MRI Guidelines for Inspire Therapy English 1

Models and Implant Locations for MR Conditional Neurostimulation Systems The following table lists all Inspire IPG model numbers and identifies those that can comprise an MR Conditional neurostimulation system. WARNING: For an implanted system to be an MR Conditional system, the IPG must be an approved MR Conditional model. The system components must also be implanted in the approved location; otherwise, the implanted system is considered untested. Table 1. IPG Scan Eligibility Model Description X-Ray ID Eligibility Details 3028 Inspire IV IMS1 Head and extremity scans only 3024 Inspire II NCR Not eligible for MRI The following table lists the approved locations for implanted components. Table 2. Approved Locations for Implanted Components 1 2 3 1 2 3 1. Implant location for the tip of the stimulation lead 2. Implant location for the IPG 3. Implant location for the respiratory sensing lead 2 English MRI Guidelines for Inspire Therapy

MRI Safety Information Non-clinical testing has demonstrated that the Model 3028 Inspire IV IPG system is MR Conditional. A patient with this device can be safely scanned in an MR system meeting the following conditions: Static magnetic field of 1.5T Cylindrical bore MR system Maximum spatial field gradient of 1900 gauss/cm (19T/m) Transmit/receive birdcage head coil (circularly polarized only) centered at the nasion or transmit/ receive extremity coil (circularly polarized only) only (no RF trasmit body coil) Maximum MR system reported, head specific absorption rate (SAR) of 3.2 W/kg Maximum gradient slew rate of 200 T/m/s per axis Maximum active scan time of 30 minutes per session with a minimum of 60 minutes wait between sessions OR: FPO:B Static magnetic field of 1.5T Cylindrical bore MR system Maximum spatial field gradient of 1900 gauss/cm (19T/m) Transmit/receive birdcage head coil (circularly polarized only) centered at the nasion or transmit/ receive extremity coil (circularly polarized only) only (no RF trasmit body coil) Maximum active scan time of 30 minutes per session with a minimum of 60 minutes wait between sessions Under the scan conditions defined above, the Model 3028 Inspire IV IPG system is expected to produce a maximum temperature rise of less than 4 C after 15 minutes of continuous scanning. In non-clinical testing, the image artifact caused by the device extends approximately 3.6mm from the stimulation lead when imaged with a gradient echo pulse sequence and a 1.5T MRI system. Warnings and Precautions Read this section for warnings and precautions related to an MR Conditional neurostimulation system. Warnings No body coil. Do not use the body coil to perform an MRI scan. Do not use a receive-only head or extremity coil with an RF trasmit body coil. It will result in unsafe heating and may cause nerve damage. No receive-only head or extremity coil. Do not use a receive-only head or extremity coil. A receive-only head or extremity coil utilizes the body coil for RF transmit. Use of the body coil for RF transmit will result in unsafe heating and may cause nerve damage. RF coil position. When performing a head scan, center the transmit/receive head coil at the patient's nasion. When performing an extremity scan do not include any implanted components in the RF field. Implant locations can be confirmed with x-ray images. Incorrect RF coil positioning may result in unsafe heating resulting in nerve damage. RF power. Do not conduct MRI scans in first or second level controlled operating mode. These modes allow higher levels of RF energy that may cause unsafe heating resulting in nerve damage. MRI Guidelines for Inspire Therapy English 3

Hydrogen only. Imaging atoms other than hydrogen has not been tested and could result in serious patient injury. Scan time. Exceeding the active scan time limit may cause unsafe heating resulting in nerve damage. Cylindrical bore 1.5T only. Scanning a patient with an MR system other than a cylindrical bore 1.5T scanner has not been tested and may result in serious patient injury. Unapproved components. Do not perform an MRI scan on patients who have any components of a neurostimulation that are unapproved for use in an MR environment. Abandoned devices. Do not perform an MRI scan on patients who have an abandoned IPG or lead. Nonfunctional leads. Do not perform an MRI scan on patients with broken or nonfunctional leads. MRI scans of patients with nonfunctional leads may result in higher than normal heating occurring at the location of the implanted lead electrodes. Skin erosion. Do not perform an MRI scan on patients who have any portion of their implanted system exposed due to skin erosion. The MRI scan may cause heating of the system, which could result in serious patient injury. Other implanted medical devices. Scanning patients who have other MR Conditional devices is acceptable as long all the MR Conditional requirements for each of the implanted devices are met. Do not conduct an MRI scan if any conditions or implants prohibit it. If you are unclear what implants are present, perform an X-ray to determine the implant type and location. Precautions Therapy off. Inspire therapy is intended for use during sleep only. Inspire therapy must be off during the MRI scan. Failure to turn therapy off during an MRI may result in unintended changes to stimulation. External devices. Do not allow external devices into the scanner magnet room (zone 4), such as the sleep remote and Inspire programmer. Because these devices contain ferromagnetic material, they can be affected by the MRI magnet, may present a projectile hazard, and are considered MR Unsafe. Electromagnetic interference (EMI). Some electrical equipment, such as an MRI machine, may generate enough EMI to interfere with the operation of the internal or external electronic components of a neurostimulation system if the equipment and the system component is too close to the system component. To mitigate the effects of possible EMI, increase the distance between the electrical equipment and the system component that is affected, and try performing the operation again. Potential Adverse Events The Inspire Medical System MR Conditional UAS system has been designed to minimize the potential adverse events that may cause patient harm. The following potential adverse events that may occur in the MRI environment: Lead electrode heating resulting in tissue damage or serious patient injury IPG heating resulting in tissue damage in the implant pocket or patient discomfort or both Induced currents on leads resulting in over stimulation or unintended stimulation Damage to the IPG or leads causing the system to fail to deliver stimulation or causing the system to deliver over stimulation Damage to the functionality or mechanical integrity of the IPG resulting in the inability to communicate with the IPG Movement or vibration of the IPG or leads 4 English MRI Guidelines for Inspire Therapy

Preparing a Patient for an MRI Scan Before conducting an MRI scan, you must perform the following steps: 1. Confirm that the implanted Inspire IPG and system are MR Conditional. 2. Confirm that no adverse conditions to MR scanning are present. 3. Select the MRI parameters according to the scanning requirements. 4. Notify the patient of potential interactions. The following sections provide more information about each of these steps. You can also use the form in the appendix of this manual as a quick-reference checklist to help you determine a patient's eligibility for an MRI scan. NOTE: Before the day of the MRI procedure, inform patients to bring their patient ID card and their sleep remote with them. Step 1: Confirm the Implanted System Contains Only MR Conditional Components To confirm that the patient's implanted Inspire System is MR Conditional, review the patient's identification card for his or her system. When an Inspire system is implanted, a patient receives an identification card that identifies the model numbers of the implanted components to help you identify them as MR Conditional. 1. Request the identification card from the patient. 2. Cross-reference the model numbers on the card with the model numbers of the MR Conditional components identified in Table 1. NOTE: If a patient does not have his or her system identification card, consider other means of confirming the MR Conditional system, such as referencing the patient's medical history, x-ray, or contacting Therapy Support. CAUTION: Do not bring the sleep remote or Inspire programmer into the scanner magnet room (Zone 4). These devices contain ferromagnetic material that can be affected by the MRI magnet and may present a projectile hazard. The sleep remote and programmer are MR Unsafe. Step 2: Confirm No Adverse Conditions to Scanning Are Present If any conditions exist that could make an MRI scan unsafe, do not scan the patient. Such conditions include: The presence of implanted neurostimulation components that are not listed as MR Conditional as outlined in Table 1. The location of MR Conditional components in an area other than what is noted in Table 2. The presence of broken or non-functional MR Conditional leads. The presence of abandoned devices, such as an IPG or lead. Any exposed portions of MR Conditional neurostimulation system components due to skin erosion. The presence of any other implanted devices (active or passive implanted devices) that prohibit safe scanning. The presence of a fever in the patient the day of the scan. MRI Guidelines for Inspire Therapy English 5

Using the Sleep Remote to Confirm Normal Operation Within 30 minutes prior to each scan, use the patient's sleep remote to confirm normal system operation. To confirm the patient's Inspire system is operating normally, follow these steps. 1. Using the patient's sleep remote, find the Therapy On/Pause button as shown in Figure 1. Therapy On/Pause Status ring Decrease (stimulation strength) Increase (stimulation strength) Therapy Off Figure 1. Sleep remote front view 2. Activate therapy by pressing the Therapy On/Pause button and holding the remote over the Inspire IPG as shown in Figure 2. IPG Radio signals Sleep remote Figure 2. Communication between sleep remote and IPG 3. Confirm that stimulation occurs and clearly moves the patient's tongue. 4. Turn the patient's therapy off by pressing the Therapy Off button as shown in Figure 1 and holding the remote over the Inspire IPG as shown in Figure 2. 5. Turn the remote over and confirm the generator light as shown in Figure 3 is NOT illuminated. If the generator light is illuminated, DO NOT scan the patient. Refer the patient to their sleep physician to resolve the issue with their implant. 6 English MRI Guidelines for Inspire Therapy

6. Confirm therapy has been turned off by confirming that the remote therapy status ring and stimulation strength gauge lights are illuminated white and are not flashing. Stimulation strength gauge Generator Upper limit Sleep remote Figure 3. Sleep remote back view Step 3: Select the MRI Parameters According to the Scanning Requirements Set up the MRI equipment per MRI Safety Information on page 3. Step 4: Notify the Patient of Potential Interactions Notify the patient that they may notice the following interactions with their implanted system during the scan: Vibration of the IPG IPG or stimulation electrode heating. Mild stimulation of the tongue, similar to therapy. Instruct the patient to notify the scan operator if either condition becomes uncomfortable. Performing the Scan and Monitoring the Patient While performing the scan, follow these guidelines: Leave any external control devices, such as a programmer or sleep remote, out of the scanner magnet room (Zone 4). If not performing a head scan, ensure that the transmit-receive coil does not cover any part of the implanted MR Conditional system. Keep the duration of the total active scanning time to 30 minutes or less per session. Wait at least 60 minutes between scanning sessions. During the MRI scan, visually and audibly monitor the patient, including verbal communication. When selecting the field of view and imaging parameters, consider that minimal image distortion may occur around an implanted lead or IPG. Consider these factors also when interpreting the MRI images. Stop the scan if the patient feels uncomfortable stimulation or heating. Therapy Support For technical questions and support please contact Inspire therapy support at 1-844-OSA-HELP (844-672-4357). For additional assistance, call your local Inspire Medical Systems representative. MRI Guidelines for Inspire Therapy English 7

Appendix A: Patient Eligibility Form for MRI Head or Extremity Scans Complete this form to help you determine the eligibility of a patient with an implanted neurostimulation system for an MRI head or extremity scan using the RF transmit-receive coil only. If the answers to all of the following questions are Yes, consult the MRI procedures in this manual for complete information on conducting an MRI scan. If the answer to any of the questions is No, do not perform the scan. If the answer to any of the questions is Unsure, contact the patient's physician or Inspire Therapy Support for help. WARNING: Scanning patients who have other MR Conditional devices is acceptable as long as all the MR Conditional requirements for each of the implanted devices are met. Do not conduct an MRI scan if any conditions or implants prohibit it. If you are unclear what implants are present, perform an X-ray to determine the implant type and location. NOTE: Before conducting an MRI scan, always ensure that you using the most recent version of these MRI procedures. Contact Therapy Support or retrieve the most recent version online at manuals.inspiresleep.com Patient's Name Physician's name and contact information (office name, address, phone number) Date of eligibility assessment IPG Model: IPG Location: Stimulation Lead Model: Stimulation Lead Location: Sense Lead Model: Sense Lead Location: Eligibility Question Yes No Unsure Approved Implant Locations 1. Does the patient have an IPG model that is MR Conditional? (See the patient's ID card for IPG model) 2. Are the MR Conditional components the only Inspire components implanted? 3. Identify the location of the implanted components and mark them on the diagram to the right. Is the IPG within zone 1, and is the stimulation lead tip within zone 2? 4. Is the coil to be used a local transmit-receive coil? ZONE 2 ZONE 1 Zone 1, IPG location: sub-clavicle region 5. If a head coil is being used, will it be centered on the nasion? 6. Is the patient free of broken or abandoned devices? 7. Is the MRI for an approved scan region? Zone 2, Stimulation lead tip location: tongue base. 8 English MRI Guidelines for Inspire Therapy

a

Manufacturer Inspire Medical Systems, Inc. 9700 63rd Avenue North Maple Grove, MN 55369 USA Tel: +1-844-672-4357 +1-763-205-7970 Fax: +1-763-537-4310 www.inspiresleep.com Inspire Medical Systems 2017 All Rights Reserved 200-261-101 Rev B