An Introduction. 1 Introduction MALDI-TOF/TOF_version100623

Similar documents
Biological Mass Spectrometry. April 30, 2014

2. Ionization Sources 3. Mass Analyzers 4. Tandem Mass Spectrometry

Technical Note # TN-31 Redefining MALDI-TOF/TOF Performance

MALDI-TOF. Introduction. Schematic and Theory of MALDI

Matrix Assisted Laser Desorption Ionization Time-of-flight Mass Spectrometry

Ionization Methods. Neutral species Charged species. Removal/addition of electron(s) Removal/addition of proton(s)

Fundamentals of Soft Ionization and MS Instrumentation

Biological Mass spectrometry in Protein Chemistry

Mass Spectrometry. Mass spectrometer MALDI-TOF ESI/MS/MS. Basic components. Ionization source Mass analyzer Detector

Comparison of mass spectrometers performances

Bruker Daltonics. autoflex III smartbeam. The Standard in MALDI-TOF Performance MALDI-TOF/TOF. think forward

Advances in Hybrid Mass Spectrometry

Mass Spectrometry at the Laboratory of Food Chemistry. Edwin Bakx Laboratory of Food Chemistry Wageningen University

1. Sample Introduction to MS Systems:

Methods in Mass Spectrometry. Dr. Noam Tal Laboratory of Mass Spectrometry School of Chemistry, Tel Aviv University

Application Note # ET-17 / MT-99 Characterization of the N-glycosylation Pattern of Antibodies by ESI - and MALDI mass spectrometry

MALDI-TOF analysis of whole blood: its usefulness and potential in the assessment of HbA1c levels

Solving practical problems. Maria Kuhtinskaja

Fundamental Studies of Matrix-Assisted Laser Desorption/Ionization, Using Time-of-Flight Mass Spectrometry to Identify Biological Molecules

Lecture 3. Tandem MS & Protein Sequencing

(III) MALDI instrumentation

SYNAPT G2-S High Definition MS (HDMS) System

Mass Spectrometry Course Árpád Somogyi Chemistry and Biochemistry MassSpectrometry Facility) University of Debrecen, April 12-23, 2010

METHODS AND REVIEWS MATRIX ASSISTED LASER DESORPTION IONIZATION TIME-OF-FLIGHT MASS SPECTROMETRY

Matrix-assisted laser desorption/ionization

REDOX PROTEOMICS. Roman Zubarev.

Structural Elucidation of N-glycans Originating From Ovarian Cancer Cells Using High-Vacuum MALDI Mass Spectrometry

Analysis of Triglycerides in Cooking Oils Using MALDI-TOF Mass Spectrometry and Principal Component Analysis

MALDI Imaging Mass Spectrometry

Biomolecular Mass Spectrometry

N-terminal charge-driven de novo sequencing by using ASDF-incorporated Curved Field Reflectron

New Instruments and Services

Quadrupole and Ion Trap Mass Analysers and an introduction to Resolution

Ionization Methods. Ionization Methods

Introduction to Proteomics 1.0

Introduction to Peptide Sequencing

Tandem MS in Microbiology. Chris Doern, PhD D(ABMM) UT Southwestern Medical Center Dallas, TX

Sequence Identification And Spatial Distribution of Rat Brain Tryptic Peptides Using MALDI Mass Spectrometric Imaging

8/31/2016 APPLICATIONS OF MATRIX ASSISTED LASER DESORPTION/IONIZATION TIME OF FLIGHT MASS SPECTROMETRY. MALDI-TOF Microflex applications

Peptide sequencing using chemically assisted fragmentation (CAF) and Ettan MALDI-ToF Pro mass spectrometry

Mass Spectrometry and Proteomics - Lecture 1 - Matthias Trost Newcastle University

Advancing your Forensic Toxicology Analyses; Adopting the Latest in Mass Spectrometry Innovations

MALDI Imaging Drug Imaging Detlev Suckau Head of R&D MALDI Bruker Daltonik GmbH. December 19,

LOCALISATION, IDENTIFICATION AND SEPARATION OF MOLECULES. Gilles Frache Materials Characterization Day October 14 th 2016

Mass Spectrometry. - Introduction - Ion sources & sample introduction - Mass analyzers - Basics of biomolecule MS - Applications

Mass spectrometry Technologies in Lipid chemistry

PosterREPRINT AN AUTOMATED METHOD TO SELF-CALIBRATE AND REJECT NOISE FROM MALDI PEPTIDE MASS FINGERPRINT SPECTRA

Metabolomics: quantifying the phenotype

Impurity Identification using a Quadrupole - Time of Flight Mass Spectrometer QTOF

PROTEOMICS (Mass spectrometry in Biochemistry)

Advantages of Ion Mobility Q-TOF for Characterization of Diverse Biological Molecules

Chapter 12: Mass Spectrometry: molecular weight of the sample

Mass Spectrometry. Actual Instrumentation

Don t miss a thing on your peptide mapping journey How to get full coverage peptide maps using high resolution accurate mass spectrometry

New Instruments and Services

Mass Spectrometry Introduction

[ Care and Use Manual ]

Three Dimensional Mapping and Imaging of Neuropeptides and Lipids in Crustacean Brain

Brooke Dilmetz. Quantitation of PFAS using MALDI-TOF MS

MALDI-TOF Mass Spectrometry: A New Rapid ID Method in Clinical Microbiology

Automating Mass Spectrometry-Based Quantitative Glycomics using Tandem Mass Tag (TMT) Reagents with SimGlycan

PROTEOMICS (Mass spectrometry in Biochemistry)

MS/MS Scan Modes. Eötvös University, Budapest April 16, MS/MS Scan Modes. Árpád Somogyi. Product Ion Scan Select. Scan. Precursor Ion Scan Scan

Double charge of 33kD peak A1 A2 B1 B2 M2+ M/z. ABRF Proteomics Research Group - Qualitative Proteomics Study Identifier Number 14146

Small Molecule Drug Imaging of Mouse Tissue by MALDI-TOF/TOF Mass Spectrometry and FTMS

PTM Discovery Method for Automated Identification and Sequencing of Phosphopeptides Using the Q TRAP LC/MS/MS System

Analysis of Peptides via Capillary HPLC and Fraction Collection Directly onto a MALDI Plate for Off-line Analysis by MALDI-TOF

Supporting Information. Lysine Propionylation to Boost Proteome Sequence. Coverage and Enable a Silent SILAC Strategy for

Shotgun Proteomics MS/MS. Protein Mixture. proteolysis. Peptide Mixture. Time. Abundance. Abundance. m/z. Abundance. m/z 2. Abundance.

Ion Source. Mass Analyzer. Detector. intensity. mass/charge

Proteomics of body liquids as a source for potential methods for medical diagnostics Prof. Dr. Evgeny Nikolaev

O O H. Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL. LC /MS conditions

Mass spectra of peptides and proteins - and LC analysis of proteomes Stephen Barnes, PhD

PROTEOMICS (Mass spectrometry in Biochemistry)

Applying a Novel Glycan Tagging Reagent, RapiFluor-MS, and an Integrated UPLC-FLR/QTof MS System for Low Abundant N-Glycan Analysis

Characterization of Noncovalent Complexes of Polyelectrolytes by Mass Spectrometry

Multiplex Protein Quantitation using itraq Reagents in a Gel-Based Workflow

amazon speed Innovation with Integrity Ion Trap Performance Beyond Imagination Ion Trap MS

Improve Protein Analysis with the New, Mass Spectrometry- Compatible ProteasMAX Surfactant

More structural information with MS n

Supporting Information. Development of N, S-doped carbon dots as a novel matrix for the. analysis of small molecules by negative ion MALDI-TOF MS

Metabolite identification in metabolomics: Database and interpretation of MSMS spectra

NIH Public Access Author Manuscript J Proteome Res. Author manuscript; available in PMC 2014 July 05.

Protein Analysis using Electrospray Ionization Mass Spectroscopy *

LC/MS/MS SOLUTIONS FOR LIPIDOMICS. Biomarker and Omics Solutions FOR DISCOVERY AND TARGETED LIPIDOMICS

Mass Spectrometry Infrastructure

Targeted and untargeted metabolic profiling by incorporating scanning FAIMS into LC-MS. Kayleigh Arthur

Agilent 7700x ICP-MS

Applications for MALDI-TOF Imaging Present and Future

Automated Sample Preparation/Concentration of Biological Samples Prior to Analysis via MALDI-TOF Mass Spectroscopy Application Note 222

SUPPORTING INFORMATION. Lysine Carbonylation is a Previously Unrecognized Contributor. to Peroxidase Activation of Cytochrome c by Chloramine-T

The 1997 ABRF Mass Spectrometry Committee Collaborative Study: Identification of Phosphopeptides in a Tryptic Digest of Apomyoglobin

SCS Mass Spectrometry Laboratory

Supplementary Information

amazon Turning Speed into Solutions

Multiple Reaction Monitoring for Direct Quantitation of Intact Proteins using a Triple Quadrupole Mass Spectrometer

Protein Identification and Phosphorylation Site Determination by de novo sequencing using PepFrag TM MALDI-Sequencing kit

Proteins: Proteomics & Protein-Protein Interactions Part I

The J105 SIMS. A New Instrument for 3-Dimensional Imaging and Analysis. Paul Blenkinsopp, Ionoptika Ltd

QSTAR Operation. May 3, Bob Seward

Transcription:

MALDI-TOF/TOF mass spectrometry An Introduction 1 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF(/TOF) mass spectrometry MALDI ion source Time-of of-flight (TOF) mass analyzer detector 2 Introduction MALDI-TOF/TOF_version100623

3 Introduction MALDI-TOF/TOF_version100623 MALDI

MALDI stands for... Matrix Assisted Laser Desorption / Ionization invented by Tanaka (Nobel prize awarded 2002), Hillenkamp and Karas in the mid 80s 4 Introduction MALDI-TOF/TOF_version100623

How does MALDI work? Required items: -a sample to be analyzed - a matrix substance -alaser MALDI preparation / MS analysis: 1.) Mix sample and matrix solutions in suitable ratio. 2.) Put a sub µl aliquot of this mixture on the target plate. 3.) Let the mixture co-crystallize. 4.) Insert the target plate in the MALDI mass spec (running under high vacuum). 5.) Shoot with the laser on the MALDI preparation to generate ions (singly charged predominantly). 6.) Collect, analyze and detect the resulting ions. 5 Introduction MALDI-TOF/TOF_version100623

How does MALDI work? Fundamental papers on the principle of MALDI: M. Karas, D. Bachmann, F. Hillenkamp Analytical Chemistry, 57, 2935-2939 (1985) K. Tanaka, H. Waiki, Y. Ido, S. Akita, Y. Yoshida, T. Yoshida Rapid Communications in Mass Spectrometry, 2, 151-153 (1988) R. C. Beavis, B. Chait, K.G. Standing Rapid Communications in Mass Spectrometry, 3, 233-237 (1989) M. Karas, M. Glückmann, J. Schäfer Journal of Mass Spectrometry, 35, 1-12, (2000) R. Zenobi and R. Knochenmuss Mass Spectrometry Reviews, 17, 337-366 (1998) 6 Introduction MALDI-TOF/TOF_version100623

What role does the MALDI matrix play? The matrix transfers the energy needed for ionization from the laser light to the sample molecules. Positive ionization mode: Sample embedded in light-absorbing matrix Sample molecule Matrix molecule Laser Excitation of matrix molecules by laser light Desorption/protonation of sample molecules MH + M Laser energy [M+H] + Matrix Formation of alternative adducts depends on the presence of respective cations (either being ubiquitary present or actively added depending on type of sample): [M+Na] + ; [M+K] + ; [M+Cu] + ; [M+Li] + ; [M+Ag] + 7 Introduction MALDI-TOF/TOF_version100623

What role does the MALDI matrix play? The matrix transfers the energy needed for ionization from the laser light to the sample molecules. Negative ionization mode: Sample embedded in light-absorbing matrix Sample molecule Matrix molecule Laser Excitation of matrix molecules by laser light Desorption/deprotonation of sample molecules M-H - M Laser energy [M-H] - Matrix 8 Introduction MALDI-TOF/TOF_version100623

Types of laser commonly used in MALDI: Nitrogen laser: pro: well structured energy profile contra: slow (maximum 50Hz) Nd:YAG laser: pro: fast (up to 1000Hz) contra: Gaussian energy profile (non-structured) Smartbeam/Smartbeam II (modified Nd:YAG laser): pro: fast (up to 1000Hz) pro: well structured energy profile Reference: A. Holle, A. Haase, M. Kayser, J. Höhndorf, Journal of Mass Spectrometry, 41, 705-716 (2006) 9 Introduction MALDI-TOF/TOF_version100623

Commonly used MALDI matrix substances: Peptides: Proteins: Glycans: Nucleic acids: 4-Hydroxy-α-cyanocinnamic acid (HCCA) 2,5-Dihydroxyacetophenone (DHAP) Sinapinic acid (SA) 2,5-Dihydroxybenzoic acid (DHB) 2,5-Dihydroxybenzoic acid (DHB) 3-Hydroxypicolinic acid (HPA) 2,4,6-Trihydroxyacetophenone (THAP) Why different matrices for different types of sample? It s all about - the amount of energy needed to ionize a particular sample compound (individual matrices show specific energy threshold ) - the stability of a particular sample compound (too hot matrix may lead to non-desired fragmentation of sample compounds) 10 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF (Time of Flight) 11 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF Basic principle of MALDI-Time-of-Flight based mass analysis: High vacuum 10-7 mbar High vacuum 10-7 mbar Intensity ++ + + + + + ++ + m/z IS1 Acceleration MALDI Ion Source Field free TOF analyzer region (drift tube) Detector Time Of Flight (depending on m/z) 12 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF E pot = zeu E kin = 1/2mv 2 zeu = 1/2mv 2 v = 2zeU/m t = L m/2zeu This defines the potential energy at which all ions start from the MALDI target. This equation defines the kinetic energy of ions after acceleration into the flight tube. In the process of ion acceleration, energy is preserved but turns from potential into kinetic energy. Transforming this equation shows the dependency of velocity of moving ions on their m/z value. With the linear velocity v being defined as L drift tube /t flight, the dependency of the ions flight time t flight from m/z becomes obvious. To put the basic MALDI-TOF separation principle into simple words: The larger its m/z, the slower an ion will fly, the longer the measured flight time will be. 13 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF: Resolution Resolution Rdefines, how well two peaks are separated from each other: R=m/ m Δm is usually derived from a mass peak s full width at half maximum (FWHM), as shown below for the exemplaric analysis of a mixture of two compounds X and Y: Mass spectrum of single compound X (yielding a signal @m/z X ) Δm Mass spectrum of single compound Y (yielding a signal @ m/z Y ) Δm: Full Width @ Half Maximum (FWHM) Δm Mass spectrum obtained from mixture [X+Y]: The resolution (i.e. FWHM) provided by the mass spectrometer used here does not allow baseline separation of X and Y. To achieve this, narrower peaks (i.e. smaller FWHM = higher resolution) would be required. 14 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF: Resolution Example spectrum containing 3 singly charged [M+H] + signals at m/z 922Da, 923Da and 924Da: 100% 922.00852 FWHM=0.07264 50% 923.01287 Resolution 15 Introduction MALDI-TOF/TOF_version100623 924.01630 921.0 922.0 923.0 924.0 m/z = m/δm = 922.00852/0.07264 = 12693 The resolution here is obviously more than sufficient to achieve baseline separation of the three compounds contained in the mixture.

MALDI-TOF: Linear mode Intensity ++ + + + + + ++ + m/z IS1 Acceleration MALDI Ion Source 0kV Field free TOF analyzer region (drift tube) Linear detector Time Of Flight (depending on m/z) 16 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF: Linear mode Observation: Ions of identical mass arrive at the detector at slightly different time points, causing peak broadening (i.e. limiting the resolution) Peak broadening effect, limiting the resolution in resulting mass spectrum. Δt + ++ + + + U accel 0kV MALDI Ion Source Field free TOF analyzer region (drift tube) 1040 1045 1050 m/z Linear detector 17 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF: Linear mode Limited in resolution due to spatial spread and energy spread Spatial spread: - initial movement of ions towards different directions - ions are desorbed from different z- coordinates due to heterogeneity in size of matrix crystals Int. Initial energy (=speed) spread: - heterogeneous secondary reactions (ion-ion; ionneutral) For all m/z! 500 1000 v[m/s] Reference: W. Ens, Y. Mao, F. Mayer, K.G. Standing, Rapid Communications In Mass Spectrometry, 5, 117-123 (1991) 18 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF How to improve resolution??? Optimized matrix preparation homogeneity (minimizing spatial spread) Pulsed ion extraction for efficient ion focusing (minimizing initial energy spread) Further ion focusing by means of a reflector TOF setup (minimizing remaining energy spread) 19 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF How to improve resolution??? r Optimized matrix preparation homogeneity Spatial spread Resolution 20 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF: How to improve resolution??? r Pulsed ion extraction for efficient ion focusing in the MALDI ion source Ion source Field-free drift region + + + + Detector IS 1 IS 2 t>t delay Pulsed Ion Extraction E pot + (PIE) + t<t delay 1040 1045 1050 m/z x[cm] 1041 1046 1051 m/z 21 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF How to improve resolution??? r Further ion focusing by means of a reflector TOF setup Ion source Field-free drift region + + + + + + + + Linear detector + + IS 1 IS 2 Reflector detector R2 Reflector (ion mirror) U refl 22 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF How to improve resolution??? MALDI-TOF, Ion source linear Field-free MALDI-TOF, drift region linear + PIE MALDI-TOF, reflector + PIE M 1 + + + + + + + + Linear detector + + IS 1 IS 2 Reflectron detector R2 Reflectron (ion mirror) U refl 1040 1045 1050 m/z 1041 1046 1051 m/z PIE compensates - initial energy spread - initial spatial spread 1046 1050 m/z Reflector compensates - remaining energy spread 23 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF Linear vs. reflector mode FAQ: If MALDI-TOF performed in reflector mode gives so much better resolution - why then use linear mode at all??? Answer: Linear mode is used whenever analytes are not stable enough to survive the energetic stress which is inherent to passing the reflector (ions are deccelerated/re-accelerated in the reflector by a high kv electric field within nanoseconds!!!). Especially larger sized molecules, e.g. intact proteins, show limited stability when passing the reflector field, and may undergo serious fragmentation, which results in either badly resolved spectra (peak fronting due to non-resolved fragments) and/or drastic loss in sensitivity (low mass fragments will miss the reflector detector). 24 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF Linear vs. reflector mode: Cytochrome C (MW avg =12360Da) Linear mode: Low resolution R=1,500 Reflector mode: High resolution R=30,000 Intens. [a.u.] 1500 12360.9 Intens. [a.u.] 2000 12361.0 1250 1500 12358.0 12364.0 Where do actually these isotope peaks originate from? 1000 12356.9 750 1000 12355.9 12366.0 500 500 12368.1 250 0 12340 12350 12360 12370 12380 m/z Spectrum shows one broad peak representing the envelope of the non-resolved isotope peaks. 12340 12350 12360 12370 12380 m/z Spectrum shows all the isotope peaks well separated from each other. 25 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF Where do these isotope peaks originate from? Most elements are found in nature in form of different so called isotopes. Isotopes of an element have nuclei with the same number of protons (the same atomic number) but different numbers of neutrons. However, both, protons and neutrons contribute to the weight of an atom, which explains the difference in weight of isotopes. 26 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF Where do these isotope peaks originate from? Isotope Mass [%] Abundance 1-H 1.007825 99.985 2-H (Deuterium) 2.014000 0.015 12-C 12.00000 98.90 13-C 13.00336 1.10 14-N 14.00307 99.63 15-N 15.00011 0.37 16-O 15.99491 99.76 18-O 17.99916 0.20 19-F 18.99840 100 23-Na 22.98977 100 31-P 30.97376 100 32-S 31.97207 95.03 34-S 33.96787 4.22 35-Cl 34.96885 76.77 37-Cl 36.96590 31.98 39-K 38.96371 93.26 79-Br 78.91834 50.69 81-Br 80.91629 49.31 Elements that are found in nature in form of only one single isotope, are called monoisotopic elements. 27 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF Where do these isotope peaks originate from? For a given molecule, the individual isotopes of all the elements contained in it finally yield a characteristic intensity distribution of isotopic masses. This is shown below by means of the isotopically resolved mass spectra of 3 compounds being different in size: Element composition: C 41 H 69 N 13 O 14 S Monoisotopic mass [M+H] + : 1000.4880 Average mass [M+H] + : 1001.1409 Element composition: C 112 H 164 N 29 O 34 S 2 Monoisotopic mass [M+H] + : 2524.1510 Average mass [M+H] + : 2525.8196 Element composition: C 253 H 363 N 55 O 75 S Monoisotopic mass [M+H] + : 5404.6075 Average mass [M+H] + : 5407.9984 1000.5 1001.5 1002.5 1003.5 2525.2 2526.2 2524.2 2527.2 2528.2 2529.2 5404.6 5405.6 5406.6 5407.6 5408.6 5409.6 5410.6 5411.6 5412.6 998 1000 1002 1004 1006m/z 2522 2524 2526 2528 2530 2532 m/z 5400 5404 5408 5412 5416 m/z The monoisotopic mass is the sum of the masses of all the atoms present in a molecule using the mass of the most abundant isotope for each element. The average mass of a molecule is the sum of elemental masses using the average weighted over all stable isotopes of each element contained in the molecule. 28 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF: Calibration Precision: Variation of values obtained from repetitive measurements performed under identical conditions (random error) Accuracy: Deviation of a measured value from the reference value (systematic error) 29 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF: Calibration U acc d M 1 + + E Ion source M 2 + + L M + 1 + Field-free drift region Detector t of = t delay + t acc + t drift (1) t acc can be calculated as followed. The force to an charge q in an electric field E is F = E q = M a (2) 30 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF: Calibration Where a is the mass dependent acceleration. It is related to the time t acc and the distance d between target and grounded acceleration plate. d = a/2 t acc 2 (3) Replacing in (2) the electric field E = U/d, the charge q = z e, the number of charges z times the elementary charge e, and from (3) a=2d/t acc2 than the time-of-flight through the acceleration region is t acc = d (2m/Uze) (4) 31 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF: Calibration For the calculation of t drift we have to do the following considerations. All ions start from the target at the same potential energy E pot = qu = zeu (5) Where q is the charge of an ion which z number of charges e. e is elementary charge and U is the high voltage applied to the target. Because of the conservation of energy, after acceleration the potential energy is tranferred into kinetic energy E kin =1/2 m v2 (6) From the constant movement in an field-free region we know that the speed v is the distance L divided by the time t drift. Replacing v in (6) by s/t drift we get (as already shown earlier) t drift = L (m/2zeu) (7) 32 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF: Calibration Combining of equation (4) (7) the total time-of-flight is t of =t delay + t acc + t drift = t delay + d (2m/Uze) + L (m/2zeu) t = C 0 +C 1 m TOF equation! The determination of C 0 and C 1 is carried out by means of two signals of known standard masses m 1 and m 2. 33 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF: Calibration By measuring the time-of-flight t 1 and t 2 of two known masses m 1 and m 2 in a mass spectrum, C 0 and C 1 can be determined: t 1 = C 0 +C 1 m 1 C 0, C 1 t 2 = C 0 +C 1 m 2 t [ns] Linear approximation t 2 + t 1 + Calibration mass range m 1 m 2 m/z [Da] For a sufficiently narrow mass range, a linear approximation of t of may be sufficient in terms of calibration accuracy (2 known calibrant masses are required here). However, for the calibration of an extended mass range, a quadratic polynom provides a much better description of the relationship between flight time and m/z. Minimum 4 known calibrant masses are required when applying a quadratic calibration function! 34 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF: Calibration strategies Step 1) External calibration Intens. [a.u.] x10 4 6 757.397 Step 1.1 1672.921 Intens. [a.u.] x10 4 5 4 3 2 1 0 870.530 944.562 Step 1.2 1181.557 1291.708 1373.788 1419.807 1517.871 1565.869 1645.805 Sample spectrum 1797.888 1959.958 750 1000 1250 1500 1750 2000 2250 2500 2750 3000 3250 m /z - apply calibration fit to sample spectra obtained from near neighbour spots 2097.090 Calibrant spectrum 2211.066 2363.150 3208.436 Intens. [a.u.] x10 4 5 4 3 2 1 0 Step 2) Internal re-calibration 870.530 944.562 (optionally) * * 1181.557 1291.708 1373.788 1419.807 1517.871 1565.869 1645.805 1797.888 1959.958 2097.090 2211.066 2363.150 Sample spectrum 750 1000 1250 1500 1750 2000 2250 2500 2750 3000 3250 m/z 3208.436 4 2 1046.542 1296.685 1758.941 2093.079 2465.196 2932.578 3657.937 * denotes compounds of known identity/mass 842.509 Da (trypsin artefact) 2211.104 Da (trypsin artefact) 0 1000 1500 2000 2500 3000 3500 - calibrants of known mass cover mass range of interest - m/z vs. flight time is fitted using a polynom of varying order (depending on size of mass range to be calibrated and number of available calibrant signals, resp.) m/z Internal re-calibration allows for - optimum mass accuracy due to compensation of spot-to-spot heterogeneities that typically cause mass errors after external calibration 35 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF: Calibration strategies When to use which calibration polynom? Compass for Flex series 1.1 and 1.2: Min. number of calibrant peaks Calibration polynom to be used external calibration: 2 linear 4 quadratic internal re-calibration: 1 quadratic (external pre-calibration: quadratic) Compass for Flex series 1.3 or higher: Min. number of calibrant peaks Calibration polynom to be used external calibration: 2 linear 4 quadratic 6 cubic enhanced internal re-calibration: 1 linear correction (external pre-calibration: quadratic) 4 quadratic 6 cubic enhanced internal re-calibration: 1 linear correction (external pre-calibration: cubic enhanced) 4 cubic enhanced Note: For optimum mass accuracy, calibrants in general have to cover the entire mass range that is to be calibrated. Extrapolation of calibration functions will always have a negative effect on the resulting mass accuracy. 36 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF: Typical applications Gel-based proteomics: Protein ID by peptide mass fingerprinting (PMF) - reflectron MALDI-TOF - mass range: 700... 4000Da Biological state A 1) 2D gel separation of proteins 2) Comparative image analysis 3) Excision of regulated spots 4) Enzymatic digestion 5) MALDI-TOF-PMF 6) Protein ID by database search (MASCOT, Phenyx etc.) 1 Database 2 3 4 Intens. [a.u.] x10 4 5 870.530 4 3 Biological state B 2 1 0 944.562 1181.557 1291.708 1373.788 1419.807 1517.871 1565.869 1645.805 1797.888 1959.958 2097.090 2211.066 2363.150 3208.436 750 1000 1250 1500 1750 2000 2250 2500 2750 3000 3250 m/z 37 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF: Typical applications Gel-based proteomics: Protein ID by peptide mass fingerprinting (PMF) 38 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF: Typical applications Gel-based proteomics: Protein ID by peptide mass fingerprinting (PMF) Why is MALDI-TOF the method of choice here? + speed (1 gel spot = 3-15 seconds of measurement) + fully automatized measurement + REALLY high throughput + sensitivity (low amol range) + good mass accuracy and sufficiently high sequence coverage to yield highly significant results 39 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF: Typical applications MALDI-TOFMS profiling of microorganisms: - linear MALDI-TOF - mass range: 2000... 20000Da Identified species Bacillus globigii Intens. [a.u.] 600 Fingerprint search against proprietary spectra library 400 200 Generate MALDI-TOF profile spectrum 0 2000 4000 6000 8000 10000 12000 14000 16000 18000 20000 m/z Mix with MALDI matrix (HCCA), prepare onto a MALDI target plate Unknown microrganism? Select a colony 40 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF: Typical applications MALDI-TOFMS profiling of microorganisms: + analysis speed (5...15s per sample) + minimum sample prep + fully automated + highly significant results 41 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF/TOF 42 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF/TOF: Principal scheme: LIFT Ion source 1 PCIS Ion source 2 PLMS Reflector CID cell Reflector detector Ion path in TOF1 region (linear TOF) Ion path in TOF2 region (reflector TOF) Ion source 1 = MALDI ion source Ion source 2 = LIFT re-acceleration cell PCIS = Timed ion gate PLMS = Post LIFT metastable suppressor 43 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF/TOF: Analysis of a mixture containing 3 compounds (green,red, blue) being different in mass: Source 1 TOF1 region PCIS CID cell v 1 < v 2 < v 3 Molecular ions are separated in TOF1 according to their mass. Part of the ions undergo fragmentation in TOF1 region induced by: - metastable laser induced decay (LID) - collision induced decay (CID) Most important: Fragments and their precursor ions travel at the same velocity. 44 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF/TOF tandem MS: LID vs. CID LID: Laser-Induced Dissociation Most straightforward way to peptide backbone fragmentation (b,y-type ions). Used for protein identification by means of peptide sequencing. CID: Collision-Induced Dissociation (high energy) Additional side chain cleavages. Higher relative intensity of internal fragments. Overall shift of average fragment size towards lower mass. Used as an option in special applications, e.g.: - denovo sequencing (enhanced immonium ions) - differentiation of isobaric aminoacids L and I by respective side chain cleavages - detailed glycan analysis (monomer linkage positions determined by products of specific crossring cleavages) 45 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF/TOF tandem MS: Analysis of a mixture containing 3 compounds (green,red, blue) being different in mass: Selecting the red precursor ion for fragment analysis by MS/MS: Source 1 PCIS Rejected PCIS = Timed Ion Gate (calibrated for time/mass correlation) 46 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF/TOF tandem MS: PCIS = Timed Ion Gate Bradboury-Niellson - Source 1 PCIS + small fringing fields + - + - + - + 47 Introduction MALDI-TOF/TOF_version100623 eff

MALDI-TOF/TOF tandem MS: PCIS = Timed Ion Gate - + - + - Ions to be rejected Ions to be selected Ions to be rejected + - + 48 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF/TOF tandem MS: PCIS = Timed Ion Gate - + - Ions to be rejected Ions to be selected Ions to be rejected + - + - + 49 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF/TOF tandem MS: PCIS = Timed Ion Selector Ions to be rejected Ions to be selected Ions to be rejected 50 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF/TOF tandem MS: PCIS = Timed Ion Selector + - + - + - Ions to be rejected Ions to be selected Ions to be rejected + - 51 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF/TOF tandem MS: Analysis of a mixture containing 3 compounds (green,red, blue) being different in mass: Red precursor ion selected for fragment analysis by MS/MS: Source 1 Ion gate LIFT Ion Source 2 Rejected Ion Source 2 (LIFT): Re-acceleration of fragments and remaining molecular ions. Re-focusing (resolution!!!) of the ions. 52 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF/TOF tandem MS: Reacceleration and focusing potentials applied in the Ion Source 2 (LIFT): Lift 1 Lift 2 Ground 19 kv 17 kv Ground 53 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF/TOF tandem MS: Reacceleration and focusing potentials applied in the Ion Source 2 (LIFT): Lift 1 Lift 2 Ground 19 kv 17 kv Ground 54 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF/TOF tandem MS: Reacceleration and focusing potentials applied in the Ion Source 2 (LIFT): Lift 1 Lift 2 Ground 19 kv 17 kv Ground 55 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF/TOF tandem MS: Reacceleration and focusing potentials applied in the Ion Source 2 (LIFT): Lift 1 Lift 2 Ground 19 kv 17 kv Ground 56 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF/TOF tandem MS: Reacceleration and focusing potentials applied in the Ion Source 2 (LIFT): Lift 1 Lift 2 Ground 19 kv 17 kv Ground 57 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF/TOF tandem MS: FAQ: Why reacceleration of ions in Ion Source 2 (LIFT)? Answer: To minimize the energy spread of the metastable fragments and, thus, to capture on the detector the complete fragment ion spectrum at one fixed reflector voltage. +19 kev 19-27 kev 0-8 kev LIFT Source 1 Source 2 A simple calculation example considering a precursor ion of [M+H] + = 1000Da: Initial acceleration in source 1 = 8kV Precursor [M+H] + = 1000Da E kin precursor = 8keV (100%) Fragment f1 [M+H] + = 500Da E kin f1 = 4keV (50%) Fragment f2 [M+H] + = 100Da E kin f2 = 0.8kV (10%) Reacceleration by 19kV narrows down this energy spread: After reacceleration, even the smallest fragment ions have more than 70% of the energy of the precursor and, thus, will be captured on the detector. E kin precursor = (8+19)keV = 27keV (100%) E kin f1 = (4+19)keV = 23keV (85%) E kin f2 = (0.8+19)kV = 19.8keV(73%) 58 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF/TOF tandem MS: Analysis of a mixture containing 3 compounds (green,red, blue) being different in mass: PCIS LIFT Source 1 Source 2 PLMS Reflector PLMS: Rejected Post-LIFT Metastable Suppressor deflects remaining non-fragmented precursor ions to avoid undesired metastable decay in TOF2 region, which would yield ghost fragment peaks detected at wrong mass. Reflector detector 59 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF/TOF tandem MS: Analysis of a mixture containing 3 compounds (green,red, blue) being different in mass: PCIS LIFT Source 1 Source 2 Rejected PLMS Reflector Rejected Reflector detector 60 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF/TOF tandem MS: Example MS/MS spectrum obtained from a peptide: H y12 y11 y10 y9 y8 y7 y6 y5 y4 Asp-Ala-Phe-Leu-Gly-Ser-Phe-Leu-Tyr-Glu-Tyr-Ser-Arg + b1 b2 b3 b4 b5 b6 b7 b8 b9 b10 b11 b12 y3 y2 y1 H + 61 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF/TOF: Typical applications Gel-based proteomics: Protein ID based on PMF + datadependent MS/MS 1) Acquire PMF. 2) Run PMF database search. 3) Data dependent MS/MS acquisition Intens. [a.u.] x10 5 1.50 1.25 1.00 0.75 0.50 0.25 0.00 #2 #1 1439.821 1479.811 1567.762 #3 #4 803.902 927.501 1006.994 1138.503 1249.633 1399.701 1639.950 1740.855 1805.168 1880.946 1927.827 2045.061 2113.902 2247.949 2344.307 2492.274 2570.243 2649.235 2905.286 3079.242 750 1000 1250 1500 1750 2000 2250 2500 2750 3000 m/z Successful protein ID by PMF Protein ID by PMF failed #1 #2 - acquire MS/MS of matching peptides for confirmation of protein ID - acquire MS/MS of non-matching peptides (identify further proteins, unexpected modifications etc.) PFF database search (combined MS/MS data) - acquire MS/MS of most abundant peptides DeNovo sequencing with subsequent MS BLAST homology search (in case of non- or partially sequenced taxa) #3 #4 62 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF/TOF: Typical applications Gel-based proteomics: Protein ID based on PMF + datadependent MS/MS Advantages of MALDI-TOF/TOF in gel-based proteomics: + speed (PMF + PFF analysis typically < 2min measuring time per gel spot) + amol sensitivity + automated, data dependent MS/MS workflow + improved significancy of combined (PMF+PFF) datasets compared to PMF only data 63 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF/TOF: Typical applications Analysis of posttranslational modifications: Phosphorylation -98Da Neutral loss of?? FQSEEQQQTEDELQDK H 3 PO 4 64 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF/TOF: Typical applications Analysis of posttranslational modifications: Phosphorylation -98Da Neutral loss of FQSEEQQQTEDELQDK H 3 PO 4 Modified threonin is obviously a wrong guess, as all the aminoacid residues following the threonin do not match the detected fragment signal pattern!!! 65 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF/TOF: Typical applications Analysis of posttranslational modifications: Phosphorylation -98Da?? FQSEEQQQTEDELQDK Neutral loss of H 3 PO 4 66 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF/TOF: Typical applications LC-based proteomics: LC-MALDI analysis of complex samples 1) Optional pre-fractionation on protein level (gel; LC; IP etc.) 2) Enzymatic digestion 3) nanolc of resulting peptide mixture 4) LC fraction collection on MALDI plate 5) offline MALDI-TOF/TOF analysis Inte ns. [a.u.] x105 1.50 1.25 1.00 0.75 0.50 0.25 0.00 80 3.90 2 92 7.50 1 1 00 6.99 4 1 13 8.50 3 1 24 9.63 3 139 9.70 1 14 39.8 21 147 9.81 1 15 67.7 62 16 39.9 50 1 740.855 18 05.1 68 1 88 0.94 6 192 7.82 7 2 04 5.06 1 2 113.902 22 47.9 49 23 44.3 07 2 49 2.27 4 257 0.2 43 2 64 9.23 5 2 90 5.28 6 3 07 9.24 2 750 1000 1250 1500 1750 2000 2250 2500 2750 3000 m /z 67 Introduction MALDI-TOF/TOF_version100623

MALDI-TOF/TOF: Typical applications LC-based proteomics: LC-MALDI analysis of complex samples + reduced ion suppression due to separation upfront to the mass spectrometer + no time constraints during data acquisition (no time pressure caused by LC speed) + increased depth of analysis (more protein IDs from complex mixtures; increased sequence coverage for individual proteins) + data dependent, non-redundant MS/MS precursor selection (esp. useful for non-isobaric SILE experiments) + LC separation is frozen on the MALDI plate and, thus, allows archiving of LC separated samples for later re-analysis - a single LC-MS/MS experiment, when compared to online-esi, is rather time consuming 68 Introduction MALDI-TOF/TOF_version100623

MALDI-TDS: Top-down sequencing of intact proteins by means of MALDI-TOF/TOF 69 Introduction MALDI-TOF/TOF_version100623

MALDI-TDS: Top-down sequencing of intact proteins N T E R M S E Q U E N C E C T E R M 70 Introduction MALDI-TOF/TOF_version100623

MALDI-TDS: Top-down sequencing of intact proteins N T E R M S E Q U E N C E C T E R M + MALDI matrix (1,5-DAN, sdhb, SA) 71 Introduction MALDI-TOF/TOF_version100623

MALDI-TDS: Top-down sequencing of intact proteins N T E R M S E Q U E N C E C T E R M + MALDI matrix (1,5-DAN, sdhb, SA) Bruker MALDI-TOF 72 Introduction MALDI-TOF/TOF_version100623

MALDI-TDS: Top-down sequencing of intact proteins N T E R M S E Q U E N C E C T E R M + MALDI matrix (1,5-DAN, sdhb, SA) LASER Bruker MALDI-TOF 73 Introduction MALDI-TOF/TOF_version100623

MALDI-TDS: Top-down sequencing of intact proteins N T E R M S E Q U E N C E C T E R M + MALDI matrix (1,5-DAN, sdhb, SA) LASER In-Source Decay (ISD) upon MALDI process: Intens. [a.u.] 1500 1777.766 1250 2006.893 1000 750 500 250 1804.798 1878.804 1933.827 2034.870 2134.974 2162.955 2263.060 2291.041 2376.140 2433.158 2504.188 2532.212 2589.216 2651.262 2764.341 2807.335 2874.129 3004.160 Bruker MALDI-TOF 0 1800 2000 2200 2400 2600 2800 3000 m/z 74 Introduction MALDI-TOF/TOF_version100623

MALDI-TDS: Top-down sequencing of intact proteins Intact protein N T E R M S E Q U E N C E C T E R M ISD fragments N N T N T E N T E R N T E R M N T E R M N T E R M S E.. S MALDI-ISD.. R M M E R M T E R M C T E R M E C T E R M C E C T E R M 75 Introduction MALDI-TOF/TOF_version100623

MALDI-TDS: Top-down sequencing of intact proteins N T E R M S E Q U E N C E C T E R M N N N T E R M S E Q U E N M R E T C E C N E U Q M N-terminal fragments (c-type ions) C-terminal fragments (z- and/or y-type ions) T N T E N T E R N T E R M N T E R M N T E R M S E.. S MALDI-ISD.. R M E R M T E R M C T E R M E C T E R M C E C T E R M 76 Introduction MALDI-TOF/TOF_version100623

MALDI-TDS: MALDI-ISD-TOF spectrum of Cytochrome C (horse) Heme Acetyl 49 N-term AA (N-terminus modified) 44 C-term AA Spectrum acquired on Bruker Ultraflex III TOF/TOF 77 Introduction MALDI-TOF/TOF_version100623

MALDI-TDS applications: Monoclonal antibodies IgG heavy chain: Aim of analysis: - N-terminal pyroglu??? - C-terminal lysine deleted??? PyroGlu K??? Matching N-terminal fragments confirm N-terminal pyroglu modification. Matching C-terminal fragments confirm absence of C-terminal K. Spectrum acquired on Bruker Ultraflex III TOF/TOF 78 Introduction MALDI-TOF/TOF_version100623

MALDI-TDS applications: PEGylated therapeutic peptides Aim of analysis: Determination of PEGylation site Peptide sequence: Acetyl-WVTHR*LAGLLSR*SGGVVK*KNFVPTDVGPXAF PEG polymer: ~500mer (approx. 22kDa) Linear MALDI-TOF spectrum showing MW distribution of the PEGylated peptide: Yoo et al., 2008 JASMS 79 Introduction MALDI-TOF/TOF_version100623

MALDI-TDS applications: PEGylated therapeutic peptides Aim of analysis: Determination of PEGylation site Peptide sequence: Acetyl-WVTHR*LAGLLSR*SGGVVK*KNFVPTDVGPXAF PEG polymer: ~500mer (approx. 22kDa) MALDI-ISD-TOF spectrum of the PEGylated peptide: Ac-WVTHR*LAGLLSR*SG GVVK*KNFVPTDVGPXAF PEG ~500mer Breakdown of N-terminal fragment series clearly identifies PEGylation site Yoo et al., 2008 JASMS 80 Introduction MALDI-TOF/TOF_version100623

Advanced MALDI-TDS: T 3 sequencing: Further fragmentation (MS 3 ) of ISD fragments by LID ISD LID LIFT Source 1 PCIS Source 2 Rejected PLMS Reflector ISD fragments undergo further metastable decay in TOF1 region Rejected Reflector detector 81 Introduction MALDI-TOF/TOF_version100623

MALDI-T 3 -TOF/TOF spectrum of ISD fragment c 13 (m/z=1516.9da) from Cytochrome C (horse) Spectrum acquired on Bruker ultraflextreme Abs. Int. * 1000 a G b E K G K K I F V Q K y K K K 4.5 4.0 3.5 3.0 2.5 2.0 1.5 1.0 0.5 72.0549 a 1 G V 146.1239 y 1 314.1500 b 3 286.1074 a 3 443.1678 b 4 571.2834 b 5 543.2209 a 5 628.2450 b 6 633.3813 y 5 761.4441 y 6 756.3251 b 7 946.5142 y 8 884.3361 b 8 997.4887 b 9 1116.5572 a 10 1074.6167 y 9 1144.6240 b 10 1243.7509 b 11 1343.7556 a 12 1371.8814 b 12 0.0 200 400 600 800 1000 1200 1400 m/z T 3 sequencing provides access to the very terminal aminoacid sequence, which is commonly not accessible by ISD (due to occurence of matrix clusters and other components occupying the low mass region in MS mode). 82 Introduction MALDI-TOF/TOF_version100623

83 Introduction MALDI-TOF/TOF_version100623 www.bdal.com