Determination of Ethyl Sulfate Impurity in Indinavir Sulfate Drug Using Ion Chromatography

Similar documents
Detection and Quantification of Inorganic Arsenic in Fruit Juices by Capillary Ion Chromatography with Suppressed Conductivity Detection

Rapid determination of phosphate and citrate in carbonated soft drinks using ion chromatography

Mercury Speciation Determinations in Asian Dietary Supplements

Determination of Benzoate in Liquid Food Products by Reagent-Free Ion Chromatography

Improved Extraction and Analysis of Hexavalent Chromium from Soil and Water

Determination of Inorganic Ions and Organic Acids in Non-Alcoholic Carbonated Beverages

Sialic Acid Determination in Glycoproteins: Comparison of Two Liquid Chromatography Methods. Experimental. Introduction. Method 1.

Rapid Determination of Phosphate and Citrate in Carbonated Soft Drinks Using a Reagent-Free Ion Chromatography System

Determination of Methacholine Chloride and Potential Impurities Using a Reagent-Free Ion Chromatography System

Determination of Organic Acids in Kombucha Using a High-Pressure Ion Chromatography System

Determination of Carbachol In Ophthalmic Solutions Using a Reagent-Free Ion Chromatography System

Identification and Quantification at ppb Levels of Common Cations and Amines by IC-MS

Application Note 178 INTRODUCTION

Carl Fisher, Terri Christison, Hua Yang, Monika Verma, and Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA, USA

Determination of Carbohydrates in Kombucha Using HPAE-PAD

Application Note 178. Brian DeBorba and Jeff Rohrer Thermo Fisher Scientific, Sunnyvale, CA, USA. 10 Typically, this requires the use of sample

Determination of Plant-Derived Neutral Oligo- and Polysaccharides Using the CarboPac PA200

Separation of Chromium (III) and Chromium (VI) by Ion Chromatography

Determination of Water- and Fat-Soluble Vitamins in Nutritional Supplements by HPLC with UV Detection

The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection

A HPAE-PAD Chromatographic Assay for Carbohydrates in Urine as a Measure of Intestinal Permeability

IC-MS Environmental Applications - Water Testing. Application Notebook

A New HILIC/RP Mixed-Mode Column and Its Applications in Surfactant Analysis

Measuring Phytosterols in Health Supplements by LC/MS. Marcus Miller and William Schnute Thermo Fisher Scientific, San Jose, CA, USA

HPAE-PAD Determination of Infant Formula Sialic Acid

Determination of Trace Sodium in Cranberry Powder

Robust and Fast Analysis of Tobacco-Specific Nitrosamines by LC-MS/MS

INTRODUCTION. Application Note 176

autosamplers ASI-100 and ASI-100T Autosamplers

Determination of Phenolic Compounds in Apple Orchard Soil

Rapid and sensitive UHPLC screening of additives in carbonated beverages with a robust organic acid column

Determination of Tigecycline in a Cell Lysate

Determination of Iodide and Iodate in Seawater and Iodized Table Salt by HPLC with UV Detection

Enhanced LC-MS Sensitivity of Vitamin D Assay by Selection of Appropriate Mobile Phase

Ion Chromatography: A Versatile Technique for the Analysis of Beer

High-Throughput, Cost-Efficient LC-MS/MS Forensic Method for Measuring Buprenorphine and Norbuprenorphine in Urine

Determination of Polydextrose in Foods by AOAC Method

The Investigation of Factors Contributing to Immunosuppressant Drugs Response Variability in LC-MS/MS Analysis

Rapid and sensitive UHPLC screening for water soluble vitamins in sports beverages

Analysis of Chloride, Phosphate, Malate, Sulfite, Tartrate, Sulfate, and Oxalate in Red and White Wine

High-Throughput Quantitative LC-MS/MS Analysis of 6 Opiates and 14 Benzodiazepines in Urine

Determination of perchlorate in environmental waters using a compact ion chromatography system coupled with a single quadrupole mass spectrometer

Determination of Lactose in Lactose-Free Milk Products by High-Performance. anion-exchange chromatography with. Pulsed Amperometric Detection

Improving Food Process Control Using the

Providing a Universal, One-step Alternative to Liquid-Liquid Extraction in Bioanalysis

columns CarboPac PA20 Column

A Rapid UHPLC Method for the Analysis of Biogenic Amines and Metabolites in Microdialysis Samples

Increasing Extraction Efficiency of Wet Samples Using a Novel New Polymer During Accelerated Solvent Extraction

Two-dimensional HPLC determination of water-soluble vitamins in a nutritional drink

Solving One of Chromatography s Biggest Dilemmas Proper Sealing of Chromatography Autosampler Vials

Determination of Steviol Glycosides by HPLC with UV and ELS Detections

Direct Determination of Sialic Acids in Glycoprotein Hydrolyzates by HPAE-PAD

GC-MS/MS Analysis of Benzodiazepines Using Analyte Protectants

A Versatile Detector for the Sensitive and Selective Measurement of Numerous Fat-Soluble Vitamins and Antioxidants in Human Plasma and Plant Extracts

Core-Shell Technology for Proteins and Peptides

Two-Dimensional HPLC Determination of Water-Soluble Vitamins in a Nutritional Drink

Determination of Clinically Relevant Compounds using HPLC and Electrochemical Detection with a Boron-Doped Diamond Electrode

Determination of Carbohydrates in Beverages and Milk Products by HPAE-PAD and Capillary HPAE-PAD

Determination of Sucralose Using HPAE-PAD

[ Care and Use Manual ]

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

Fig.1. Denatonium benzoate (DB) chemical structure

Doubling the throughput of long chromatographic methods by using a novel Dual LC workflow

Impurity Profiling of Carbamazepine by HPLC/UV

Bio-Rad Laboratories HEMOGLOBIN Testing Bio-Rad A1c. VARIANT II TURBO Link System. Fully-Automated HbA 1c Testing

Bioanalytical Quantitation of Biotherapeutics Using Intact Protein vs. Proteolytic Peptides by LC-HR/AM on a Q Exactive MS

Comparison of Full Scan MS2 and MS3 Linear Ion Trap Approaches for Quantitation of Vitamin D

Determination of Sucralose in Reduced- Carbohydrate Colas using High-Performance Anion-Exchange Chromatography with Pulsed Amperometric Detection

Evaluation of an LC-MS/MS Research Method for the Analysis of 33 Benzodiazepines and their Metabolites

Separation of 2AA-Labeled N-Linked Glycans from Glycoproteins on a High Resolution Mixed-Mode Column

Glycoprotein monosaccharide analysis using HPAE-PAD with manually prepared eluent

GlycanPac AXR-1 Columns

Two-Dimensional HPLC Combined with On-Line SPE for Determination of Sudan Dyes I IV in Chili Oil

Stress Degradation Studies And Validation Method For Quantification Of Aprepitent In Formulations By Using RP-HPLC

Analysis of Amino Acids Derived Online Using an Agilent AdvanceBio AAA Column

Ensuring High Sensitivity and Consistent Response in UHPLC-MS Analyses

Sugars in Honey Using HPAE-PAD: What Is the Best Column?

Thermo Scientific Dionex CarboPac MA1 Column

Comparison of conventional HPLC with UPLC method for determination of albuterol sulfate and it s impurities in pharmaceutical formulation

Chemical compatibility of the Pall QPoint TM Docking Station - Tap Assembly with common surface disinfectants

Simultaneous Analysis of Active Pharmaceutical Ingredients and Their Counter-Ions Using a Mixed-Mode Column

Rapid Analysis of Water-Soluble Vitamins in Infant Formula by Standard-Addition

APPLICATIONS TN µm Core-Shell Particle. (continued on page 2) Page 1 of 8. For additional technical notes, visit

ApplicationNOTE EXACT MASS MEASUREMENT OF ACTIVE COMPONENTS OF TRADITIONAL HERBAL MEDICINES BY ORTHOGONAL ACCELERATION TIME-OF-FLIGHT.

ab ORAC Assay Kit

Dayana Argoti, Kerry Hassell, Sarah J. Fair, and Joseph Herman Thermo Fisher Scientific, Franklin, MA, USA

APPLICATIONS TN Determination of Sugars in Animal Feed using HPLC-ELSD

Marc Plante, Bruce Bailey, and Ian N. Acworth Thermo Fisher Scientific, Chelmsford, MA, USA

ab HIV-1 Protease Inhibitor Screening Kit (Fluorometric)

ab Homocysteine Assay Kit (Fluorometric) 1

Bio-Rad Laboratories HEMOGLOBIN Testing Bio-Rad A1c. VARIANT II TURBO Link. Fully-Automated HbA 1c Testing

Nitrate/Nitrite Assay Kit Manual Catalog #

Bio-Rad Laboratories. Hemoglobin Testing. VARIANT II Hemoglobin Testing System For HbA 1c

A New Era in SFC Column Technology

Rapid and Direct Analysis of Free Phytosterols by Reversed Phase HPLC with Electrochemical Detection

MS/MS as an LC Detector for the Screening of Drugs and Their Metabolites in Race Horse Urine

Determination of Total Nitrogen and Phosphorus in Wastewaters by Alkaline Persulfate Digestion Followed by IC

LC-MS/MS Method for the Determination of 21 Opiates and Opiate Derivatives in Urine

Transcription:

Application Note 8 Determination of Ethyl Sulfate Impurity in INTRODUCTION Indinavir sulfate is a specific and potent inhibitor of HIV- protease and is widely used in the treatment of AIDS. Indinavir sulfate is prepared by dissolving indinavir base in anhydrous ethanol that is then treated with sulfuric acid. At ambient temperature, a mixture of ethanol and sulfuric acid reacts to yield monoethylsulfate (ethyl sulfate). Therefore, treatment of indinavir base in anhydrous ethanol with sulfuric acid is performed at a controlled temperature of less than C to avoid the formation of ethyl sulfate. However, it is very important for manufacturers of drug products to monitor the level of anticipated process-related and degradation impurities before commercial release to prove the consistency of the manufacturing process. It is also important to monitor drug stability during storage. Ion chromatography (IC) has been successfully used to measure ionic drug degradation products and processrelated impurities. N-methylpyrrolidine, a degradation product of cefipime, has been measured in cefipime and simulated Cefipime for Injection., In addition, IC also has been used determine the process impurity ethylhexanoate in clavulanate. The work here shows an IC method to measure ethyl sulfate in indinavir sulfate. This method uses a Thermo Scientific Dionex IonPac ASA column with a carbonate/bicarbonate eluent to separate ethyl sulfate from other anions, including sulfate. The eluent was produced using an eluent generator to improve ease of use and retention time reproducibility. Reproducible chromatography is demonstrated without the labor or possible error of eluent preparation, and without the need to prepare sulfuric acid solutions for suppressed conductivity detection. EQUIPMENT Thermo Scientific Dionex ICS- or Dionex ICS-5 system including:* DP Dual Pump DC Detector/Chromatography module with dual-temperature zone equipped with 6-port injection valves EG Eluent Generator module EPM Electrolytic ph Modifier Thermo Scientific Dionex Chromeleon Chromatography Data System software, Version 6.8, SR9 *A Dionex ICS- or other Dionex IC system featuring eluent generation also can be used. Alternately, if the eluents are manually prepared, any Dionex IC system can be used. REAGENTS AND STANDARDS Deionized (DI) water, Type I reagent grade, 8 MΩ-cm resistivity or better Dionex Seven Anion Standard II (P/N 5759) Sulfuric acid, 98% (H SO, Merck) Ethanol (Absolute), AR grade (C H 5 OH, RCI Labscan)

PREPARATION OF SOLUTIONS AND REAGENTS Carbonate (.7 mm )/Bicarbonate (. mm) Eluent Generate carbonate/bicarbonate eluent on-line by pumping high-quality DI water (8 MΩ-cm resistivity or better) through the Thermo Scientific Dionex EluGen EGC II K Cartridge and EPM. Use Chromeleon software to control the concentration of carbonate/bicarbonate. Add backpressure tubing to achieve psi system backpressure, which allows the EG degasser to work properly. Refer to the Dionex ICS- operator s manual for how to add backpressure. Alternately, prepare the eluent manually by pipetting 5. ml of.5 M Na (P/N 76) and.6 ml of.5 M NaH (P/N 76) into a L volumetric flask and dilute to volume with DI water. Eluent generation was used to produce the data in this study. Ethyl Sulfate Stock Solution Transfer.5 ml ethanol into a ml volumetric flask. Slowly add.5 ml 98% sulfuric acid to the same volumetric flask. Gently swirl the flask to mix and bring to volume with DI water. Note: this preparation must be performed in a fume hood while wearing goggles and protective clothing. Mixture of Seven Anion Secondary Standard Add.5 ml Dionex Seven Anion Standard Stock Solution to a ml volumetric flask and bring to volume with DI water. Ethyl Sulfate Standard Solution Add. ml ethyl sulfate stock solution to a 5 ml volumetric flask and bring to volume with DI water. Preparation Dissolve a capsule of the indinavir sulfate drug in a ml volumetric flask with DI water. Filter the sample solution with a.5 µm syringe filter. CONDITIONS Column: Guard: Dionex IonPac ASA ( mm) (P/N 6) Dionex IonPac AGA ( 5 mm) (P/N 65) Eluent Source: Dionex EluGen EGC II K cartridge (P/N 589) with EPM Electrolytic ph Modifier and EGC Carbonate Mixer, mm (P/N 6686) Eluent:.7 mm Carbonate/. mm bicarbonate Column Temp.: ºC Pressure: ~ psi Detection: Suppressed conductivity, Thermo Scientific Dionex ASRS, mm (P/N 655), recycle mode, current 5 ma Background: ~ RESULTS AND DISCUSSION Ethyl Sulfate Production Ethyl sulfate is not commercially available, but it can be easily produced in the laboratory by mixing sulfuric acid and ethanol. 5 Ethyl sulfate was prepared as described in the Ethyl Sulfate Stock Solution section, and ultimately used to prepare the Ethyl Sulfate Standard Solution used for the chromatography in this study. IC was used to judge the success of ethyl sulfate production. The Ethyl Sulfate Standard Solution was injected into the IC system and the chromatography was compared to ethanol and sulfuric acid (diluted in the same manner as was the Ethyl Sulfate Stock Solution to prepare the Ethyl Sulfate Standard Solution [Figure ]). The Ethyl Sulfate Standard Solution exhibited two peaks: one at approximately 5 min and another at the retention time of sulfate. No peaks were found in the ethanol solution and only a peak corresponding to sulfate was found in the sulfuric acid solution. Therefore, the peak at approximately 5 min was assumed to be ethyl sulfate. Determination of Ethyl Sulfate Impurity in

Column: Dionex IonPac ASA Analytical, mm Dionex IonPac AGA Guard, 5 mm with EPM and EGC Carbonate Mixer Eluent:.7 mm Carbonate/. mm bicarbonate Column Temp.: C Detection: Suppressed conductivity, Dionex ASRS, mm, recycle mode, current 5 ma : A. Ethanol B. Mixture of ethanol and sulfuric acid C. Sulfuric acid. Ethyl sulfate. Sulfate Column: Dionex IonPac ASA Analytical, mm Dionex IonPac AGA Guard, 5 mm with EPM and EGC Carbonate Mixer Eluent:.7 mm Carbonate/. mm bicarbonate Column Temp.: C Detection: Suppressed conductivity, Dionex ASRS, mm, recycle mode, current 5 ma : A. Dionex Seven Anion Standard B. Prepared ethyl sulfate mg/l. Fluoride. Chloride 5. Nitrite 5. Ethyl sulfate 5. Bromide 5 6. Nitrate 5 7. Phosphate 8. Sulfate 5 C B A 5 6 7 8 A -5 8 6 88 B Figure. Overlay of chromatograms of ethanol, sulfuric acid, and a mixture of ethanol and sulfuric acid. Separation The Dionex IonPac AS, AS, and ASA columns were evaluated to determine the best separation of ethyl sulfate from the other inorganic anions. The Dionex IonPac ASA column provided the best separation. Figure shows chromatograms of a mixture of seven common inorganic anions and the prepared Ethyl Sulfate Standard Solution. Note that ethyl sulfate was well resolved from the other anions. The Ethyl Sulfate Standard Solution was injected five times to evaluate stability. The peak area was stable and there was no noticeable increase or decrease in retention time (Table ). - 6 8 6 8 88 Figure. Chromatograms of a Dionex Seven Anion Standard and the prepared ethyl sulfate. Table. Area of Ethyl Sulfate Standard Solution Injection No. Area (*min).79.76.76.86 5.76 Average.77 % RSD. Application Note 8

Analysis A customer kindly provided the indinavir sulfate final drug product in the form of capsules that had been stored on a shelf for a long period of time prior to the analysis reported here. Five capsules of sample were prepared and each sample was injected five times into the IC system. Figure shows the chromatography of one of the capsule samples. The putative ethyl sulfate peak was well resolved from the other peaks in the sample. Spiking the sample solution with the Ethyl Sulfate Standard Solution confirmed that the peak in the sample had the same retention time as the peak in the standard (Figure ). The position of the peak tentatively identified as ethyl sulfate relative to sulfate and other peaks in the capsule sample was also consistent with a peak identified as ethyl sulfate in a publication that uses a longer IC method. Peak area data from the 5 sample injections are shown in Table. These data show that ethyl sulfate was found in all five capsules and the peak area response was reproducible. There was also a consistent amount of ethyl sulfate in the five capsules. Column: Dionex IonPac ASA Analytical, mm Dionex IonPac AGA Guard, 5 mm with EPM and EGC Carbonate Mixer Eluent:.7 mm Carbonate/. mm bicarbonate Column Temp.: C Detection: Suppressed conductivity, Dionex ASRS, mm, recycle mode, current 5 ma : Indinavir sulfate drug sample and spiked indinavir sulfate drug sample. Chloride. Ethyl sulfate. Nitrate. Sulfate -5 8 6 88 Column: Dionex IonPac ASA Analytical, mm Dionex IonPac AGA Guard, 5 mm with EPM and EGC Carbonate Mixer Eluent:.7 mm Carbonate/. mm bicarbonate Column Temp.: C Detection: Suppressed conductivity, Dionex ASRS, mm, recycle mode, current 5 ma : Indinavir sulfate drug sample. Chloride. Ethyl sulfate. Nitrate. Sulfate Figure. Overlay chromatograms of the indinavir sulfate and spiked indinavir sulfate sample. Injection No. Table. Area of Ethyl Sulfate in s Area (*min) 5.56.589.85.57.988.55.5.8.9.99.99.55.86.9.95.598.58.87.87.99 5.55.587.87.887.895 Average.55.56.8.9.95 RSD.8%.8%.8%.%.87 % -5 8 6 88 Figure. Chromatogram of the indinavir sulfate drug sample. Determination of Ethyl Sulfate Impurity in

CONCLUSION This work shows IC separation of ethyl sulfate in an indinavir sulfate drug sample. This method only requires the addition of DI water to the IC system to deliver reproducible chromatography while avoiding the time, labor, and potential error of manual eluent preparation. This method uses an electrolytic suppressor in recycle mode, and therefore does not require the analyst to prepare sulfuric acid solutions for eluent suppression prior to conductivity detection. REFERENCES. Prasanna, S. J.; Shama, M. K.; Mukkanti, K.; Sivakumaran, M.; Kumar, K.S.R.P.; Kumar, V. J. J. Validation of a Sensitive Ion Chromatography Method for Determination of Monoethylsulfate in Indinavir Sulfate Drug Substance. Pharm. Biomed. Anal. 9, 5, 65 69.. Dionex Corporation, Determination of N-Methylpyrrolidine in Cefepime Using a Reagent- Free Ion Chromatography System. Application Note 99, LPN 5, 8, Sunnyvale, CA.. Dionex Corporation, Determination of N-Methylpyrrolidine in Cefepime with Nonsuppressed Conductivity Detection. Application Note 59, LPN 586, 8, Sunnyvale, CA.. Dionex Corporation, Determination of -Ethylhexanoic Acid Impurity in Clavulanate. Application Note 6, LPN 68,, Sunnyvale, CA. 5. Wikipedia. http://en.wikipedia.org/wiki/ethyl_sulfate (accessed Jun 5, ). IonPac, Chromeleon, and ASRS are registered trademarks of Dionex Corporation. Other trademarks are the property of Thermo Fisher Scientific, Inc. and its subsidiaries. Speed Simplicity Solutions Dionex Products 8 Titan Way P.O. Box 6 Sunnyvale, CA 988-6 (8) 77-7 North America U.S./Canada (87) 95-75 South America Brazil (55) 7 5 Europe Austria () 66 5 5 Benelux () 68 9768 () 5 9 Denmark (5) 6 6 9 9 France () 9 Germany (9) 66 99 Ireland (5) 6 6 Italy (9) 5 6 67 Sweden (6) 8 7 8 Switzerland () 6 5 9966 United Kingdom () 76 697 Asia Pacific Australia (6) 9 5 China (85) 8 8 India (9) 76 75 Japan (8) 6 6885 Korea (8) 65 58 Singapore (65) 689 9 Taiwan (886) 875 6655 LPN 88 PDF 9/6 6 Thermo Fisher Scientific, Inc. www.thermoscientific.com/dionex Application Note 8