Non-invasive measurement of blood and tissue parameters based on VIS-NIR spectroscopy

Similar documents
Optical non-invasive methods for characterization of the human health status

Optical Sensor System for Hemoglobin Measurement

Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan

Biomedical Instrumentation D. The Photoplethysmogram

Muscle OxyGen monitor. The Science Behind Moxy

Non-invasive Hemoglobin Measurement for Anemia Diagnosis

Dual channel photoplethysmography studies of cardio-vascular response to the body position changes

Assessment of Photoplethysmography Method in Extraction of Hemoglobin Concentration

Development of an Optical Sensor Based System for Detection and Monitoring of Hemoglobin in Blood

Oxygen Saturation Monitors & Pulse Oximetry. D. J. McMahon rev cewood

Correlation of Ratio of Ratio of Near Infrared Vs Infrared Photoplethysmograph Spectroscopy for the Non-Invasive Measurement of Haemoglobin

Estimation of Haemoglobin Using Non-Invasive Technique

Noninvasive Blood Glucose Analysis using Near Infrared Absorption Spectroscopy. Abstract

How Does Pulse Oximetry Work? SpO2 Sensors Absorption at the Sensor Site Oxyhemoglobin Dissociation Curve

Monitoring Tumor Therapy

Noninvasive Detection of Bilirubin in Discrete Vessels

Permanent City Research Online URL:

Design a system of measurement of heart rate, oxygen saturation in blood and body temperature with non-invasive method

Introduction. Device Description. Dual-Band Spectral Output

Appendix E Choose the sign or symptom that best indicates severe respiratory distress.

CVS Hemodynamics. Faisal I. Mohammed, MD,PhD.

Oximeters. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan

Pulse oximetry revisited. Dr Liesel Bösenberg Specialist Physician and Fellow in Critical Care Kalafong Hospital University of Pretoria

Light is wearable Health Monitoring and Fitness Tracking Light is OSRAM

Asimplealgorithm for in vivo ocular fundus oximetry compensating for non-haemoglobin absorption and scattering

BLOOD GLUCOSE LEVEL MONITORING BY NONINVASIVE METHOD USING NEAR INFRA RED SENSOR

Continuous Monitoring of Blood Pressure Based on Heart Pulse Analysis

Journal of Faculty of Engineering & Technology DESIGN AND IMPLEMENTATION OF A WEARABLE HEALTH DEVICE

Dr. D.McG.Clarkson Determining the Accuracy of SpO2 Values

CVS Hemodynamics. Change in blood pressure:

Physics of the Cardiovascular System

Blood Flow, Blood Pressure, Cardiac Output. Blood Vessels

Use of the Free Electron Laser for the Noninvasive Determination of Retinal Oxyhemoglobin Saturation by Near Infrared Reflectance Spectrophotometry

Special Lecture 10/28/2012

Design and Development of Non-Invasive Prototype to Measure Pulse Rate, Blood Glucose and Oxygen Saturation Level in Arterial Blood

12.1 The Function of Circulation

Collin County Community College

Heart Abnormality Detection Technique using PPG Signal

Research Journal of Pharmaceutical, Biological and Chemical Sciences

Non-Invasive Method of Blood Pressure Measurement Validated in a Mathematical Model

Empowering the RT with New Noninvasive Monitoring Capabilities

History of Vascular Modelling. William Harvey discovery of the circulation 1628

Non-invasive blood glucose measurement by near infrared spectroscopy: Machine drift, time drift and physiological effect

A Three-Wavelength Pulse Oximeter for Carboxyhemoglobin Determination

Blood Pressure Estimation Using Photoplethysmography (PPG)

The circulatory system transports blood to deliver important substances, such as oxygen, to cells and to remove wastes, such as carbon dioxide.

Cold Laser Program ML830

Blood flows away from the heart in arteries, to the capillaries and back to the heart in the veins

Circulatory, Respiratory, and Excretory Systems

Chapter 12. Capillaries. Circulation. The circulatory system connects with all body tissues

Circulatory System Review

ORIGINAL RESEARCH. Effect of road traffic accident contaminants on pulse oximetry among normoxaemic volunteers

Clinically Available Optical Topography System

International Journal of Research and Applications July - Sep 2017 Transactions 4(15):

Breast Cancer Detection by Determination of Optical Properties of Non- Malignant and Malignant Breast Tissues

PCA Enhanced Kalman Filter for ECG Denoising

Non Invasive methods of Hemoglobin

The Heart 1 of 34 Boardworks Ltd 2012

In vivo optical imaging : revealing endogeneous optical contrast at depth

Non Invasive Estimation of Blood Glucose using Near Infra red Spectroscopy and Double Regression Analysis

Arterial blood oxygen saturation during blood pressure cuff-induced hypoperfusion

2.01 Remember the structures of the circulatory system

CIRCULATORY SYSTEM: BLOOD + BLOOD VESSELS. October 21, 2016

Check a Pulse! When to Question SpO 2, NIBP & EtCO 2 Readings

CIE Biology GCSE. 9: Transport in animals. Notes.

Chapter 27 The Heart and Blood Vessels

Testing Masimo rainbow Technology with vpad-a1 1. Background

Time domain diffuse optical imaging and spectroscopy for biomedical diagnostics. Rinaldo Cubeddu

A. Incorrect! The left ventricle receives oxygenated blood from the lungs via the left atrium.

A. Incorrect! The left ventricle receives oxygenated blood from the lungs via the left atrium.

TEPZZ _ 849 A_T EP A1 (19) (11) EP A1. (12) EUROPEAN PATENT APPLICATION published in accordance with Art.

Human Circulation and Respiration Chapter 38

Measuring Heart Rate and Blood Oxygen Levels for Portable and Wearable Devices

7.L.1.4 Circulatory System Guided Study Notes. Circulation

The Cardiovascular System home study course

1 Non-invasive measurement of arterial pressure

Competitive Intelligence Report Technology & Patents

In vivo Infrared Spectroscopy

Trans Fat Determination in the Industrially Processed Edible Oils By Transmission FT-IR Spectroscopy By

CARDIOVASCULAR SYSTEM Worksheet

Chapter 10 The Circulatory & Lymphatic Systems

Blood Pressure Determination Using Analysis of Biosignals

An in vitro hemodynamic tissue model to study the variations in flow using near infrared spectroscopy

Lecture 8. Heart and Circulatory System. Lecture 8

Circulatory System Objective sheet 3

Blood Functions. Blood and the Cardiovascular System. Blood. Plasma. Erythrocytes (RBCs) Erythrocytes (RBCs) 4/7/2017

Scrub In: Red blood cells are called: Which component of blood is necessary for the initiation of the blood clotting process:

P215 SPRING 2019: CIRCULATORY SYSTEM Chaps 13, 14 & 15: pp , , , I. Major Functions of the Circulatory System

Circulation.notebook March 07, Maintaining Dynamic Equilibrium. Oct 11 3:27 PM. Circulatory System p

1. Which of the following blood vessels has a thin elastic layer? A. Aorta. B. Pulmonary artery. C. Posterior vena cava. D. Mesenteric capillary.

Chapter 1 Methods Describing the Interaction of Laser Radiation with Biological Tissues

Evaluation of oesophageal reflectance pulse oximetry in major burns patients

Outline. Electrical Activity of the Human Heart. What is the Heart? The Heart as a Pump. Anatomy of the Heart. The Hard Work

The Cardiovascular and Lymphatic Systems Cardiovascular System Blood Vessels Blood Vessels Arteries Arteries Arteries

Health Science 20 Circulatory System Notes

Diffuse Reflectance Spectroscopy Characterization for Extraction of Tissue Physiological. Parameters. Janelle Elise Phelps

Determination of hemoglobin is one of the most commonly

What is the mechanism of the audible carotid bruit? How does one calculate the velocity of blood flow?

BLOOD RUNS THROUGH YOUR BODY

Topic 6: Human Physiology

Transcription:

Non-invasive measurement of blood and tissue parameters based on VIS-NIR spectroscopy Jens Kraitl* a, Ulrich Timm a, Hartmut Ewald a a Institute of General Electrical Engineering, University of Rostock, lbert-einstein-str. 2, D-18059 Rostock, Germany STRCT Currently, invasive methods are used to measure the hemoglobin concentration and the most hemoglobin-derivatives, whereby blood is taken from the patient and subsequently analyzed. The noninvasive method presented here allows pain free continuous on-line patient monitoring with minimum risk of infection and facilitates real time data monitoring allowing immediate clinical reaction to the measured data. Visible and near infrared (VIS-NIR) spectroscopy in combination with photo-plethysmography (PPG) is used for a detection of human tissue properties and the measurement of hemoglobin concentration in whole blood and hemoglobin derivatives. The absorption, scattering and the anisotropy of blood and tissue is a function of the irradiated wavelengths. This fact is used to calculate the optical absorbability characteristic of blood and tissue which is yielding information about blood components like hemoglobin-concentration (chb), carboxyhemoglobin (COHb) and arterial oxygen saturation (SaO2). The ratio between the PPG peak to peak pulse amplitudes for each wavelength is used in combination with a dynamic spectrum extraction. The prediction of the bloodand tissue-parameters is based on a Principal Component Regression (PCR) method. The non-invasive sensor system is calibrated with a lab based artificial blood circulatory system and with data from clinical studies. Keywords: hemoglobin, noninvasive, carboxyhemoglobin, spectroscopy, photo-plethysmography 1. INTRODUCTION This paper reports about fundamentals, simulations and measurements of optical absorption characteristics of whole blood. The human circulatory system is a fast regulatory transport system. This closed circuit consists of parallel and serially connected blood vessels for which the heart acts as a circulation pump. The mandatory blood circulation is maintained by the contraction of the muscular cardiac wall. The heart is made up of a left and right cardiac half, each of them being a muscular hollow organ. In functional and morphologic terms the circulatory system consists of two serial sections. The heart also can be considered as two serial pumps. The right cardiac side absorbs the deoxygenated blood from the body and transports it to the lungs (lung circuit), where it is re-oxygenated. The oxygenated blood arrives at the left cardiac side, wherefrom the dispensation to various organs takes place (body circuit). The oxygenated blood in the body circuit is pumped from the left ventricle to the aorta and main arteries whose sub branches lead to the tissue and organs. Following the next stage of branching into arteriole and capillary, the oxygen and nutrients and the ingestion of carbon monoxide and intermediate catabolic products are distributed to the tissue and organs of the body. Finally the capillaries terminate into venules and veins which transport the deoxygenated blood back to the right atrium. sensor system to measure blood components like hemoglobin is presented and corresponding results of in-vitro and in-vivo measurements. s basic technology NIR-spectroscopy and Photoplethysmography (PPG) is used for these non-invasive optical measurements. The characteristic absorption coefficient of blood in the visible and NIR region is well known and is mainly influenced by the different hemoglobin derivates [1]. This fact is used to calculate the optical absorbability characteristics of blood which is yielding information about blood components as arterial oxygen saturation (SaO2), hemoglobin concentration (chb), carboxyhemoglobin (COHb) and met-hemoglobin (MetHb). The measured spectra signals and the ratio between the peak to peak pulse amplitudes are used for a calculation of these parameters. Hemoglobin is the main component of red blood cells. The primary function of Hemoglobin is the transport of oxygen from the lungs to the tissue and carbon dioxide back to the lungs. The Hemoglobin concentration in human blood is an important parameter in evaluating the physiological status of an individual and an essential parameter in every blood count. *j.kraitl@uni-rostock.de; phone +49 381 498-7069; fax +49 381 498-7081; www.iae.uni-rostock.de Optical Diagnostics and Sensing XIII: Toward Point-of-Care Diagnostics, edited by Gerard L. Coté, Proc. of SPIE Vol. 8591, 859105 2013 SPIE CCC code: 1605-742/13/$18 doi: 10.1117/12.2001274 Proc. of SPIE Vol. 8591 859105-1

In currently standards, invasive methods are used to measure the hemoglobin concentration and the most hemoglobin derivates, whereby blood is taken from the patient and subsequently analyzed. part from the discomfort of drawing blood samples, an added disadvantage of this method is the delay between the blood collection and its analysis, which does not allow real time patient monitoring in critical situations. non-invasive method allows pain free continuous online patient monitoring with minimum risk of infection and facilitates real time data monitoring allowing immediate clinical reaction to the measured data. The non-invasive multi-spectral measurement method was tested and analyzed with spectrometers and prototype-devices based on radiation of monochromatic light emitted by laser diodes and by using light emitting diodes (LED) through an area of skin on the finger. The sensors assembled in this investigation are fully integrated into wearable finger clips. Since the near infrared light was found to penetrate a great depth into biological tissues, near infrared spectroscopy has been developed into a non-invasive method for biomedical sensing and clinical diagnosis [2]. Oximetry is well known as typical example of a near-infrared application in clinic and can be used to non-invasive measure the functional arterial oxygen saturation (SpO2) of human blood in-vivo [3]. The hemoglobin concentration in human blood is also an important parameter to evaluate the physiological condition and the capability of oxygen transportation in blood. With this information anemia (a low hemoglobin level) and polycythemia vera (a high hemoglobin level) can be diagnosed and monitored. It is also possible to observe imminent postoperative bleedings and autologous retransfusions. It is well known that pulsatile changes of blood volume in tissue can be observed by measuring the transmission or reflection of light through it [4]. This diagnostic method is called Photoplethysmography (PPG). The newly developed multiwavesensors system operates in the range of 500nm to 1400nm for the measurement of the hemoglobin concentration and hemoglobin derivates. For in-vitro tests and measurements a hemodynamic blood flow model (FM) has been developed which allows spectrometer measurements and the validation of the new sensors. The absorption and scattering of blood in the visible and near infrared range is dominated by the red blood cells, the different hemoglobin derivates and the blood plasma that consists mainly of water. From on optical point of view, the human tissue is a turbid media. The radiation transport equation (1) can be regarded as the mathematical basis of tissue optics for the description of lighttissue interaction and the related phenomena [5]. 1 ( +sñ)l(r,s,t)=-(µ a+µ s)l(r,s,t)+µ s P(s,s')L(r,s',t)dΩ'+Q(r,s,t) (1) cn t 4π Where Lrst (,,) is the radiance intensity of light at position r in direction s at time t, c0 cn = is the velocity of light in tissue with index-of-refraction n, n ( ss, ') is the unit vector direction s and scattered direction, dω ' = sin υ' υ' ϕ' is the element of the solid angle, Srt (,) is the energy intensity of the light source, Srt (,) Qrst (,,) = is a source term. Ω Description of light-tissue interactions using the transport theory requires first to solve Eq.1. ecause of difficulties to obtain exact solutions for biological tissues, several approximations have been made for the representations of Pss (, ') and Lrst (,,). Virtual modeling of turbid media, like biological tissue, and the description of photon transport in this media can lead to a better understanding of photonic processes and enables an effective development and optimization of optical systems. In contrast to the analytical description of light-tissue-interaction with the radiation transport equation a numerical method like Monte Carlo simulation or Monte Carlo ray tracing, provides considerable advantages like greater flexibility during model development and higher accuracy of the results [6]. With the Monte-Carlo simulation (MC) it is possible to trace photons step by step through a turbid medium. large quantity of the traced photons with their resulting paths in the medium results in a realistic mapping of the propagation of light in the target tissue [7]. The medium itself is described by the optical parameters scattering coefficient μ s, absorption coefficient μ a, the anisotropy coefficient g and the index of refraction n. The Henyey-Greenstein phase function describes the probability of the polar deflection angle after each scattering event [8]. The simulations of light-tissue interactions were performed with the optic software SP (reault Research). SP includes functions for simulations of geometric- and physical-optical properties and delivers complete 3D-models of optical and mechanical systems. In addition the Monte-Carlo simulation is already s ' Proc. of SPIE Vol. 8591 859105-2

implemented in the software. esides the simulation of multi-layered tissue models, SP was used to simulate cuvettes and blood transporting tubing s of a hemodynamic blood flow model which has been developed. The results of the simulations provide important information such as optical path lengths, penetration depths and internal distributions of the energy in the target tissues. Furthermore the simulations supplement the analysis of real measurements. Heater 2. HEMODYNMIC LOOD FLOW MODEL hemodynamic blood flow model (FM) based on the human circulatory system was developed to allow a controlled variation of the blood parameters like hemoglobin concentration (chb), oxyhemoglobin (O2Hb resp. SaO2), carboxyhemoglobin (COHb) and methemoglobin (MetHb) [9]. For this reason the optical properties of the blood were observed continuously by spectrometer measurements to determine the absorption, transmission and scattering properties of human whole blood in a wavelength range from 500 to 1000 nm and to test new noninvasive measurement systems. Heated water Erythrocytes 1 lood plasma 1 Figure 1. Structure of the hemodynamic blood flow model (FM) and photos during system operation. The following blood parameters were varied by using the blood flow model: hemoglobin concentration (chb) oxyhemoglobin (O2Hb) carboxyhemoglobin (COHb) deoxyhemoglobin ((HHb) temperature (T) flowrate (Q) kind of fumigation (N 2 -CO, N 2, compressed air ) flowrate of fumigation The model enables a defined circulation of app. 250 ml of human blood. The peristaltic blood pump has a maximum rotation speed of 250 rpm (Q: 0 to 200 ml/min). To determine the optical properties of the blood components two grid spectrometer were used (vantes vaspec 2048x14, Jeti VS140). With these spectrometers it is feasible to cover an overall range of 400nm to 1700 nm to obtain the absorption spectra of the blood. oth spectrometers are connected with optical cuvettes integrated in the FM. ll the measurements taken noninvasively were compared with sample results obtained by a lood Gas nalyzer (G device) to validate the measurements. Fresh erythrocyte concentrate from human donors were centrifuged three times and washed in phosphate buffer solution (300 mmol/l, ph 7.4). y doing this the blood samples contained no plasma, proteins, leukocytes or thrombocytes. Light microscopy control of the samples ensured that they were neither hemolyzed nor were the cells deformed. Concentration of erythrocytes as well as hemoglobin was changed by adding plasma with different intervals. lood circulation and predetermined oxygen states were adjusted with an extra-corporal circulation unit (Stöckert Instruments GmbH, Germany). The blood was gently stirred and kept flowing through the model blood tubes and the specially designed turbulence free cuvettes (for the Proc. of SPIE Vol. 8591 859105-3

spectrometers) with a laminar flow. Oxygen saturation was adjusted by continuous flow of O 2, air and N 2 through a neonatal hollow fiber membrane oxygenator. Important parameters as Temp ( o C), chb (g/l), ph, pco2 (mmhg), po2 (mmhg), HbO2 (%) were also measured by lood Gas nalyzers. ll investigations were derived with different oxygenation and hemoglobin levels. The blood temperature was kept constant at 37 o C via a water heating mechanism. OFM - Y mm cuwlto, rariatòn OQfO. (8e02.100%._ 5%) 3. IN-VITRO LOOD SPECTR NLYSIS The blood absorption spectra are obtained continuously for various oxygenation levels, hemoglobin contents and for variations of carboxyhemoglobin. Figure 2 shows the typical blood spectra (650nm-1000nm) from our measurements where the hemoglobin concentration remains constant (80g/l) and the oxygen saturation (HbO2 resp. SaO2) in the blood varies from 100% to 5%. 2'b Denveuve lnorphon: FM - 2 mm cuvelle, vanalm. O2Hb. (SaO2= 100%.. S`ó) wsr41,n215 I..1 Figure 2. Typical blood absorption spectra () and the Second-Derivative-Spectra () with different SaO2. The principle of measurement is based on the fact of a substantial absorption/transmission difference of light in red and near infrared region between oxygenated hemoglobin (HbO2), deoxygenated hemoglobin (HHb) and blood plasma (optical mainly water). s shown in Figure 2, deoxygenated hemoglobin (HbO2=5%) is optically much denser to the red light (650 nm to 750 nm) than oxygenated hemoglobin (HbO2=100%), whereas the reverse is true in the near infrared region (900 nm to 1000 nm), even to a lesser degree. Note that there is an isosbestic wavelength band in the range of 800 nm to 810 nm within which the transmission of oxyhemoglobin and deoxyhemoglobin is nearly equal. The water absorption is very distinct in the blood near infrared spectrum for the wavelength of above 960 nm. Water has a weak absorption peak around 965nm. If the hemoglobin/hematocrit concentration of blood decreases, the plasma concentration in blood increases, and subsequently the water absorption should be relatively advanced. In figure 2 an absorption peak of deoxyhemoglobin around 760nm is apparently shown for the hypoxia blood but the water absorption at 965nm is not distinguishable. FM -2 mm cuvette, variation hemoglobin, (chb= 469/I...129911, Sa02=98%) 0.9 0.8 0.7 c06 00.5 4 0.4 0.3 0.2 0.1 ---chb=460 ---chb=610 ---chb=740 -chb=83411 ---chb=940 -chb=115g4 ---chb=1 850 700 750 800 850 Wavelength trim] 900 950 1000 Figure 3. lood absorption spectra with different hemoglobin concentrations and identical HbO2 of 98%. Proc. of SPIE Vol. 8591 859105-4

Figure 2 shows the spectra of figure 2 after applying a second derivative transformation to the data of the original spectra. Second derivatives may swing with greater amplitude than the primary spectra. The more distinguishable second derivatives are especially useful for separating out weak peaks of overlapping absorption bands. In some cases derivative spectra can be a good noise filter since changes in base line have negligible effect on derivatives. In figure 2 the absorption peak of deoxyhemoglobin at 760nm and the water absorption peak at 965nm is clearly visible. Figure 3 shows the absorption spectra of blood with different hemoglobin concentrations and identical HbO2 of 98%. The water absorption is very distinct in the blood near infrared spectrum for the wavelength of above 960 nm. If the hemoglobin concentration (resp. the hematocrit) of blood decreases, the plasma concentration in blood increases, and subsequently the water absorption should be relatively advanced. In the measurement shown in Figure 3 it is obvious to see that the water absorption at 965nm becomes more pronounced for a decrease of the hemoglobin concentration in blood (conspicuous at chb=46g/l). The variation of the blood hemoglobin (resp. hematocrit) induces the changes of both the absorption and the scattering coefficients of blood. With the FM and the experimental results it was possible to perform the quantification of chromophore components in scattering media like blood by means of spectral multicomponent analysis which is based upon spectral correlation. The optical spectra of blood are analyzed by combination of this multilinear regression method with a Partial Least Squares Regression (PLSR) and Principal Components Regression (PCR). This both methods are used to model a response variable from the large number of variables (spectra points). oth methods construct new variables, known as components, as linear combinations of the original variables, but they construct those components in different ways. PCR creates components to explain the observed variability in the variables, without considering the response variable at all. PLSR takes the response variable into account and therefore often leads to models that are able to fit the response variable with fewer components. Generally the context can be expressed by the following matrix equation [10] (2) =εcl+e where =( j ) expressed the measured absorption as a function of wavelength for j=1 to n wavelength points. C=(C i ) is the concentration to be estimated by the multilinear regression, where the element C i is the concentration of the i-th chromophore in the blood. E is the residual error in the multicomponent analysis in which multiple scattering effects are involved. The estimation in this form depends on the optical path length L and the term ε= (ε ij ) which expresses the specific absorption coefficients of the chromophores within the blood. During the estimation, a least squares routine is used to fit the measured data over the used spectral region. 200 180 $ 160 140 h en ogl*h n 94 (l4 90 Spectra) Fear Hemoglobin rtoterminaton. N =90 Spectra 't. w 1 3, 10 _- 61g1(í SS S0) f3tand.ltntan.plo1. N.90 Spectra 1 120 100 66 60. z se 20 40 60 60 100 120 140 160 160 200 O lood Oaa Maylor imabeo roloronao - hemoglobin ;gal 40 80 80 100 120 140 160 180 Mean O -Ref (invasive) Predicted (non-invasive} Hemoglobin (g1) Figure 4. Invasive reference chb measured by G vs. predicted noninvasive chb () and the derived scatter plot (). The figure 4 shows the result with the algorithm for the example of hemoglobin concentration in blood. number of n=90 in-vitro measurements were made with a systematical variation of hemoglobin concentration and oxygen saturation separately or for both parameters together. The calibration model for chb is computed with measured blood absorption spectra over the region of 650nm to 1000nm with a number of 351 spectra points. The developed calibration models are applicable for oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin, methemoglobin, hemoglobin concentration and hematocrit. The interaction of all these parameters is eliminated by a recursive algorithm. second example in figure 5 shows the application for a measurement of carboxyhemoglobin in blood. ecause of the spectral characteristics of carboxyhemoglobin the calibration model for COHb is based on absorption spectra in the region of 460nm to 590nm. The figure 5 shows a data set of spectra measured with the FM with a variation of carboxyhemoglobin and oxyhemoglobin (functional). The Figure 5 shows a FM measurement (2 hours) with a variation of carboxyhemoglobin in the blood. The figure shows the good correlation of the computed and predicted noninvasive Proc. of SPIE Vol. 8591 859105-5

values for carboxyhemoglobin (COHb) and the fractional oxygen saturation (HbO2) versus the invasive G reference values. 1 0,99 Wig wwg0 'opadq Iuow01ß9O0W MVOs 01403 WpDU1- PVOv.wuou 01100 r. Woewv09 %Q0 xroah+vo*+wua+wuwb- fwo!naw 0 009 9099.9 %200H : bsorption normalized 0,98 O2Hb: 96 % - COHb: 2% O2Hb: 90% - COHb: 7% 0,97 O2Hb: 84% - COHb: 13% O2Hb: 78% - COHb: 19% O2Hb: 72% - COHb: 24% O2Hb: 62% - COHb: 34% 0,96 460 480 500 520 540 560 580 Wavelength [nm] OZ or 09 auq ul aevww O coi ori Figure 5. lood absorption spectra with different levels of carboxyhemoglobin COHb and fractional oxygen saturation O2Hb () and measurement of COHb and HbO 2 through a 0.5mm cuvette of the FM (). 4. IN-VIVO LOOD SPECTR NLYSIS fter a good correspondence between in-vitro experimental results with cuvettes of the blood tube system (FM) and our mathematic models the applicability of the method for in-vivo measurements was investigated. Figure 6 shows a measurement of the absorption spectra (650nm-1000nm) during a hypoxia study through the fingertip (index finger, dexter) of one subject. The arterial oxygen saturation during this breath down study was reduced to about 75%. Thereby the recorded data of the spectrometer was compared with the data of the blood-gas-analysis G from the. radialis (arterial oxygenic saturation - SaO 2 in %). In figure 6 the in-vivo absorption spectra of the finger with an arterial blood oxygenation SaO2=98%, SaO2=88% and SaO2=75% is represented. The hemoglobin concentration of the subject during the measurement was constant at 145 g/l. n, reo-asap,on spectra aew phalanx or ringer (Sa(Y2-96ro.88 :ane 7e t., anenai ciooa).99.49.999,9099.92.0.999.99,19.19.1994.940990.er9991799.,..119291 98%1 -SO2 08% - S,02,69 nn 14, 999 9, I WI YeavoYaNdi Figure 6. Finger absorption spectra () and the Second-Derivative-Spectra () with different SaO2. t a first appearance the absorption spectra of a finger is comparable with the absorption spectra of whole blood (figure 2). The arterial and venous blood layer in the finger causes the main percentage of absorption. The deoxyhemoglobin absorption peak is distinct in the spectra (caused by venous blood). The water absorption peak at 965nm is much stronger than in whole blood (compare figure 2). This is caused by the high percentage of water in the bloodless tissue. Figure 6 shows the spectra of figure 6 after applying a second derivative transformation to the data of the original spectra. The figure 6 shows besides the absorption peaks of the blood chromophores some weak absorption bands of bloodless tissue components like fat and collagen. Deduced from the spectrometer measurements we developed a new sensor device which is based on a multi-wavelength measurement and allows a continuous non-invasive in-vivo detection of hemoglobin concentration, oxygen saturation and pulse [11]. The arteries contain more blood during the systolic phase of the heart than during the diastolic phase, due to an increased diameter of the arteries during the systolic phase. This effect occurs only in arteries but not in veins. For this reason the absorbance of light in tissues with arteries increases during Proc. of SPIE Vol. 8591 859105-6

systole because the amount of hemoglobin (absorber) is higher and the light passes through a longer optical path length in the arteries. These intensity changes are the so called Photoplethysmography-waves (PPG). The time varying part allows the differentiation between the absorbance due to venous blood (DC part) and that due to the pulsatile component of the total absorbance (C part). This effect is used to separate the arterial blood absorption from absorption of the other tissue components in our application. The in-vivo measurement technique requires a pulse signal for the calculation of the relative attenuation coefficients. Vasoconstriction at the extremities can be a problem, as it decreases the signal amplitude, and therefore the signal to noise ratio. Small signal amplitudes tend to give inaccurate results. This may be a limitation when using the system on various patient groups. One of the newly developed sensor devices emits light in the range of wavelengths between 600 1000 nm (670, 808, 905 and 980 nm). This is the therapeutic window region, in which the blood absorption is dominated by the hemoglobin derivatives. dditionally a wavelength of 1300 nm is integrated. t this wavelength the absorption of water is dominant. The frontpanel of the online software is shown in figure 7 [12]. bicacn, bh0104jelki2che 2E1,120KIK bm, - L25 linear Subjects female Subjects male PMD L25= 23.6(L25) +66,7 N =131 measurements (21 female; 110 male) Rz= 0.9177 PMD -coefficient L25 ([C /DC] /[C /DC]- Ratios) non -invasive measurement Figure 7. Frontpanel online-software for chb measurement () and non-invasive chb vs. invasive chb ( HemoCue TM ) (). pre-study with n=43 adult healthy male and female volunteers (range 19-60 years, 15 women and 28 men) was already implemented to test the ability to measure the hemoglobin content non-invasive and in-vivo. The third or fourth finger of the left hand was connected with the new sensor. HemoCue TM hemoglobin device (HemoCue, Sweden) was used for invasive reference measurements. This system provides reliable quantitative hemoglobin results with the same performance as a large hematology analyzer. drop of capillary blood was taken from the same fingertip used for the photometric measurements for each volunteer, and analyzed with the HemoCue TM device. The figure 7 shows the results of the non-invasive photometric measurements vs. the invasive measured hemoglobin values (131 measurements). In this study, we achieved good results with our application for a non-invasive hemoglobin determination. The example is too small to represent the variability in the population. Therefore the whole data set of 43 volunteers is not sufficient by itself to further evaluate the effects of patient-to-patient variation on the measurement method. The potential of this photometric method to measure the non-invasive hemoglobin and other hemoglobin derivates in-vivo were proved with good results. Future work will involve further clinical studies, which allows an extensive statistical analysis of these measurements. 5. CONCLUSIONS In this paper a multi-wavelength photometric measurement method that provides non-invasive in-vitro and in-vivo spectral measurements in human blood and tissue has been described. The fact that the absorption-coefficients µ a and scattering-coefficients µ s for blood differ at difference wavelengths has been exploited and is used for calculation of the optical absorbability characteristics of human blood yielding information on the blood composition. The results for non-invasive in-vitro determinations of hemoglobin and carboxyhemoglobin of whole blood are shown in the paper. For these measurements a hemodynamic bloodstream model FM has been used which allows spectrometer measurements and the validation of sensor devices. The potential to measure the hemoglobin content of blood in-vivo were proved and demonstrated. What could not be determined so fare in a first approach is a measurement with a large Proc. of SPIE Vol. 8591 859105-7

group of patients and volunteers. Future work will involve further clinical studies with this measurement method. main aspect is the test of an optimized hardware system and a further evaluation of suitable statistical on-line analysis algorithms for the blood components. CKNOWLEDGMENTS This work was conducted within the framework of the project PHOTOSENS and has been funded with financial support from the Ministry of Economics, Labor and Tourism of the State Mecklenburg-Western Pomerania. REFERENCES [1] Roggan,., Friebel, M., Dörschel, K., Hahn,. and Müller, G., Optical properties of circulating human blood in the wavelength range 400-2500 nm, J. iomed. Opt. 4, pp 36-46, (1999). [2] Niemz, M.H., Laser-Tissue Interactions Fundamentals and pplications, Springer-Verlag, erlin New York, 1717-1722 (1996). [3] Wukitsch, M.W., Petterson, M.T., Tobler, D.R., and Prologe, J.., pulse oximetry: analysis of theory, technology, and practice, J. Clin. Monit. 4, 290-301 (1988). [4] Roberts, V.C., Photoplethysmography Fundamental aspects of the optical properties of blood in motion, Trans. Inst. Meas. Control l 4, 101-106 (1982). [5] Martelli, F., Del ianco, S., Ismaelli,., Zaccanti, G., Light Propagation through iological Tissue, SPIE Press, ellingham, Washington (2010). [6] Prahl, S.., Keijzer, M., Jacques, S.L., Welch,.J., Monte Carlo model of light propagation in tissue, Proc. SPIE IS (6th edn.), 5 (1989). [7] Prahl, S.., Inverse adding-doubling program, Oregon Medical Laser Center, St. Vincent Hospital (2011) [8] Tuchin, V., Tissue Optics - Light Scattering Methods and Instruments for Medical Diagnosis (2nd Edition) 145-191, SPIE Press, ellingham, Washington (207). [9] Fricke, D., Koroll, H., Kraitl, J. and Ewald, H., lood flow model for noninvasive diagnostics. Proc. IEEE GCC Conference, Dubai, 343-346 (2011). [10] Cope, M., van der Zee, P., Essenpreis, M., rridge, S.R., and Delpy, D.T, Data analysis methods for near infrared spectroscopy of tissue: problems in determination the relative cytochrome aa 3 concentration, Proc. SPIE, 1431, 251-262 (1990). [11] Timm U., Lewis E., McGrath D., Kraitl J., Ewald H., LED ased Sensor System for Non-Invasive Measurement of the Hemoglobin Concentration in Human lood, IFME Proc. Vol. 23, 825-28 (2008). [12] Kraitl, J., Timm, U., Lewis, E., Ewald, H., Non-invasive sensor for an in vivo hemoglobin measurement. Proc. 10 th IEEE Sensors Conference, Limerick, 276-279 (2011). Proc. of SPIE Vol. 8591 859105-8