ARTICLE. Accuracy and Reproducibility of Blood Lead Testing in Commercial Laboratories

Size: px
Start display at page:

Download "ARTICLE. Accuracy and Reproducibility of Blood Lead Testing in Commercial Laboratories"

Transcription

1 ARTICLE Accuracy and Reproducibility of Blood Lead Testing in Commercial Laboratories Nishith K. Jobanputra, MPH; Robert Jones, PhD; Gail Buckler, RN, COHN, MPH; Ronald P. Cody, EdD; Michael Gochfeld, MD, PhD; Thomas M. Matte, MD, MPH; David Q. Rich, BS; George G. Rhoads, MD, MPH Objective: To assess the proficiency of commercial laboratories in analyzing lead in clinical blood samples from subjects without overt lead exposure. Design: We submitted masked duplicate blood lead specimens to 8 masked laboratories. Each laboratory received blood aliquots immediately following drawing (time 1) and 2 weeks later (time 2) from 7 human subjects and 3 bovine blood samples with known lead levels of 0.26, 0.57, and 0.79 µmol/l (5.4, 11.8, and 16.4 µg/dl). Of the 8 laboratories, 5 were commercial laboratories, 1 was a state laboratory, 1 was a research laboratory, and 1 was the Centers for Disease Control and Prevention reference laboratory. Outcome Measures: Correlation coefficients were calculated, and differences within and between laboratories were assessed by analysis of variance. Results: Results were obtained for all specimens, with all the human subjects overall mean lead levels being less than 0.48 µmol/l ( 10 µg/dl). Each laboratory reported all human blood specimens appropriately, as having lead levels less than 0.48 µmol/l ( 10 µg/dl) and within 0.14 µmol/l (3 µg/dl) of the overall mean for that subject. All internal reproducibilities were very high (range, ) except for one (0.60), possibly lower because of 1 pair of specimens. Mean differences between blood samples analyzed at time 1 and time 2 ranged from 1.4 to 1.2, with only 2 laboratories having significant differences (P.01). Conclusions: Overall, there was strong reproducibility within and among laboratories, with no overall time trend or interlaboratory or intralaboratory variance. The storage conditions did not seem to affect the aggregate results. The data suggest that through implementation of the Centers for Disease Control and Prevention/Wisconsin Blood Lead Proficiency Testing Program, the Centers for Disease Control and Prevention s Blood Lead Laboratory Reference System, and mandated federal and state proficiency programs, laboratories in this geographic region have improved their performance as compared with previous published studies and an unpublished study. Arch Pediatr Adolesc Med. 1998;152: Editor s Note: What a nice study this is! In some matters, it helps to have Big Sister watching. Catherine D. DeAngelis, MD From the New Jersey Graduate Program in Public Health (Mr Jobanputra) and the Environmental and Occupational Health Sciences Institute (Ms Buckler, Drs Cody, Gochfeld, and Rhoads, and Mr Rich), University of Medicine and Dentistry of New Jersey Robert Wood Johnson Medical School, and Rutgers University, Piscataway, NJ; and Divisions of Environmental Health Laboratory Sciences (Dr Jones) and Environmental Hazards and Health Effects (Dr Matte), Centers for Disease Control and Prevention, National Center for Environmental Health, Atlanta, Ga. BECAUSE OF the subclinical symptoms of chronic lead exposure, diagnosis has traditionally relied heavily on laboratory criteria (blood lead levels, protoporphyrin levels, and, more recently, bone lead levels), rather than observable symptoms. Blood lead level is generally accepted as the most valid biomarker of human exposure to lead. 1 In 1965, the normal blood lead range for children in the United States was µmol/l (15-25 µg/dl) 2 ; whereas today it is generally considered that even levels of 0.48 to 0.72 µmol/l (10-15 µg/dl) can produce developmental delays. 3 A population-based study 4 has shown that blood lead levels have decreased over time as the use of leaded gasoline and leadbased paint has decreased. Also, the demand for blood lead testing has increased since the Centers for Disease Control and Prevention (CDC) 5 recommended, in 1991, that children living in communities with high risks for lead exposure should have repeated blood lead determinations every 6 months until 2 or more consecutive blood lead values are less than 0.48 µmol/l ( 10 µg/dl). Consequently, laboratories are now working closer to their limits of detection, and their ability to consistently reproduce valid results at levels less than 0.48 µmol/l ( 10 µg/dl) is under scrutiny. Several standardization and proficiency programs have been implemented to control such error; the most noteworthy of these is the CDC/ Wisconsin Blood Lead Proficiency Testing Program (currently known as the Federal/Wisconsin Blood Lead Proficiency Testing Program), a nonregulatory program that tests participating commercial laboratories periodically by submitting marked samples of known lead concentra- 548

2 PATIENTS AND METHODS SUPPLIES Vacutainerbloodcollectiontubes(BectonDickinsonLabware, Franklin Lakes, NJ), which contain liquid EDTA anticoagulant, and 23-gauge butterfly blood-collection systems were providedbycdc snationalcenterforenvironmentalhealth. All materials were prescreened for lead content, which was not substantial ( µmol/l [ 0.07 µg/dl] for the collection tubes). One milliliter of an aqueous mixture of 1% vol/ vol nitric acid (HNO 3 ) and 0.001% vol/vol Triton X-100 (AldrichChemicalInc, Milwaukee, Wis), anemulsifier, wasdrawn through the butterfly blood-collection systems (capillary collection) into a clean syringe, then dispensed and measured for lead content. The collection tubes were screened with the addition of 1.0 ml of an aqueous mixture of 1% vol/vol nitric acid (HNO 3 ) and 0.001% vol/vol Triton X-100 solvent, followed by soaking for approximately 12 to 14 hours. Half of the tubes were inverted to maximize contact with the stoppers (or caps). The lead level of the solvent was measured after contact with all the containers to be used in the blood collection. The total amount of lead from all of these sources was added and evaluated so that the total amount of lead contamination that would be caused by the collection devices was less than 5% of the expected blood lead mean value (0.24 µmol/l [5µg/dL]) ofthepatientsamples. Alllaboratories, exceptlaboratory 7, used the EDTA-containing tubes with lavender tops. Laboratory 7 only conducts analyses on certified trace metal free tubes with sodium heparin anticoagulant (tan topped). To maintain anonymity, we complied with this requirement. Butterfly blood-collection systems were used to facilitate the drawing of multiple samples with a minimum number of venipunctures. DRAWING OF BLOOD The National Center for Environmental Health, through the CDC Blood Lead Laboratory Reference System, provided a set of in vivo dosed bovine blood samples at 3 target values to be used as quality control specimens. The target values of 0.26, 0.57, and 0.79 µmol/l (5.4, 11.8, and 16.4 µg/dl) were determined using inductively coupled plasma isotopedilution mass spectroscopy. 9 The bovine blood samples were collected with 1.5 mg of the anticoagulant Na 2 EDTA per milliliter of whole blood. The collection tubes (lavender or tan topped) were washed with 18-megaohm deionized water to remove the EDTA that was normally placed in the tubes during tube production, thus eliminating the problems that twice the amount of EDTA had with specific analytical methods. The bovine blood samples from the CDC Blood Lead Laboratory Reference System were then placed into the tubes, and the tops replaced. All of the manipulations of cleaning, drying, and filling were handled in a class-100 clean room ( 100 particles/m 3 ) to eliminate possible contamination of the samples. All reference samples were refrigerated before and during shipment by overnight express carrier. All human blood samples were drawn in the first quarter of 1995 at the Clinical Center for Environmental and Occupational Health at the Environmental and Occupational Health Sciences Institute in Piscataway, NJ. After receiving informed consent, blood to fill sixteen 2-mL vials were drawn from each of 7 healthy adult volunteers with no history of occupational exposure to lead. The subject pool consisted of the 3 bovine blood samples, labeled as human blood, along with the samples from 7 human subjects. Eight laboratories were used: 5 large commercial laboratories, 1 state laboratory, 1 research laboratory, and the CDC reference laboratory. All 5 commercial laboratories were large national laboratories operating in All 5 processed large numbers of specimens and were selected because of their involvement in blood lead level determination. Each laboratory was sent 2 specimens from each subject, 1 immediately following the drawing of the blood (time 1) and 1 specimen 2 weeks later (time 2), for a total of 80 per batch. Each laboratory analyzed 20 samples, 10 per batch. The specimens were assigned fictitious labels to facilitate blinding. Throughout the 2 weeks before the submission of the second batch, the specimens were stored in polyethylene vials at 4 C, a temperature shown to provide stable storage conditions for whole blood Specimens were either picked up by laboratory personnel or shipped overnight in insulated packages. All laboratories used graphite-furnace atomic absorption spectroscopy for analysis of the specimens. The CDC reference laboratory performed all analyses as previously described. 13,14 Unknown blood samples, bovine quality control materials established by inductively coupled plasma isotopedilution mass spectroscopy (4 levels, different from the 3 target values of bovine blood used as subjects H-J), and aqueous standards were diluted 10-fold (1:10) with a matrix modifier, placed into clean autosampler cups, and analyzed. The matrix modifier was prepared with 0.2% vol/vol highpurity nitric acid, 0.5% vol/vol Triton X-100, and 0.2% wt/ vol ammonium phosphate diluted with ultrapure 18-megaohm water. Stock lead standards for instrument calibration were prepared using National Institute of Standards and Technology Standard Reference Material Quality control materials, including 4 levels of bovine blood quality control materials, were analyzed before and after all samples to verify the accuracy of each analytical run. An atomic absorption spectrophotometer(model4100zl, Perkin-ElmerCorp, Norwalk, Conn) with Zeeman-effect background correction was used for all CDC analyses of study blood specimens. DATA ANALYSIS Blood samples were matched by patient code, batch number, and laboratory identification number. Descriptive statistics were generated, and correlation analyses and analyses of variance were performed. Laboratory 1 and laboratory 2 had minimum detection levels of 1.0 mg/dl. Those values that were reported as nondetectable were assigned a value of 0.02 µmol/l (0.5 µg/dl). The minimum detection level for laboratory 6 was 3.0 µg/dl, and, therefore, all values reported from laboratory 6 as nondetectable were assigned a value of 0.07 µmol/l (1.5 µg/dl). None of the other laboratories reported levels below detection limits. Data are reported as mean ± SD. tion for analysis. 6 The analysis of blood lead specimens is regulated by the Clinical Laboratory Improvement Amendments Act of 1988, 7 which states that acceptable error is within 0.19 µmol/l (4 µg/dl) when blood lead values are less than 19.3 µmol/l ( 40 µg/dl). Although these marked samples are intended to be received without extra attention and care, they are easily identified by supervisors and technicians. To our knowledge, the only prior blinded study 549

3 Table 1. Laboratory Reproducibility: Correlation of Time-1 and Time-2 Blood Lead Levels Laboratory Pearson r * Mean (SD) Difference, µmol/l (0.12) (0.14) (0.08) (0.02) (0.09) (0.06) (0.05) (0.03) Overall 0.92 NA *All r values are significant at P.001, except for laboratory 2 with P =.07. Mean difference is (time-1 blood lead level minus time-2 blood lead level). To convert lead values from micromoles per liter to micrograms per deciliter, divide by NA indicates data not available. P.01. Time-1 Blood Lead Value, µmol/l Time-2 Blood Lead Value, µmol/l Figure 1. Reported blood lead values for all laboratories at time 1 and time 2. To convert lead values from micromoles per liter to micrograms per deciliter, divide by of laboratory proficiency in blood lead determination was conducted by Sargent et al 8 in They reported that there was wide disparity in laboratory performance, with some laboratories meeting or exceeding the CDC/Wisconsin Blood Lead Proficiency Testing Program guidelines and others failing to do so. They also believed that further testing to determine if order-of-magnitude errors 8 occurred frequently enough to warrant the use of blinded quality control or quality assurance samples at routine intervals was necessary. The present study provides a second assessment of routine laboratory performance on blood lead determinationandcomparesinterlaboratoryandintralaboratoryprecision by assessing known and unknown lead levels. RESULTS REPRODUCIBILITY Table 1 presents the within-laboratory reproducibility, which is a measure of the correlation between matched pairs of time 1 and time 2 samples. Figure 1 shows this graphically with reported values at time 1 (micromoles per liter) plotted against those at time 2 (micromoles per liter). All laboratories have Pearson correlation coefficients (r) between 0.92 and 1.00 that were significant at P.001, except laboratory 2, which had an r of 0.60 (P=.07). Laboratory 4 and laboratory 8, the reference laboratory at the CDC, ranked highest in reproducibility at r=1.00 (P.001). Laboratory 2 performed poorly on the samples from subject J in both batches (Table 2 and Figure 2), accounting for the low Pearson coefficient and suggested excessive variability within that laboratory. Mean values obtained for the 10 specimens analyzed at time 1 and at time 2 are shown separately for each laboratory in Table 2. At time 1, the mean blood lead level of all subjects by laboratory ranged from 0.15 to 0.30 µmol/l ( µg/dl). Mean levels at time 2 ranged from 0.13 to 0.25 µmol/l ( µg/ dl). The mean difference (time 1 time 2) ranged from 1.4 to 1.2 (Table 1). Only laboratories 4 and 7 showed a significant negative time trend between the second and first specimen batches (P=.001 and P=.003, respectively). All other laboratories did well, as the overall reproducibility is acceptable. Subject mean blood lead values per laboratory are shown in Table 2. Analysis of variance showed that there was a borderline significant difference among laboratories (P=.06) when comparing the reported blood lead values from each laboratory. Further analysis showed that there was a significant difference between 2 laboratories (P.05), most likely caused by a single outlier. Because of the small sample size, this outlier has considerable effect on the aggregate result. Overall laboratory mean levels ranged from 0.32 to 0.54 µmol/l ( µg/dl), with all but 1 mean level between 0.48 and 0.54 µmol/l (10.0 and 11.3 µg/dl). ACCURACY Table 2 shows the mean result for the bovine subjects (subjects H, I, and J) for each laboratory. Each laboratory analyzed 2 samples from subjects H, I, and J. Although the target value for subjects H, I, and J were known (0.26, 0.57, and 0.79 µmol/l [5.4, 11.8, and 16.4 µg/dl], respectively), many of the laboratories underreported values at both time 1 and time 2 (Figure 2, Figure 3, and Figure 4). Laboratory 2 reported the blood lead level for subject J as 0.15 µmol/l (3.1 µg/dl) for time 2. The overall laboratory mean measurements were 0.2 ± 0.04 µmol/l (5.0 ± 0.8 µg/dl) for subject H, 0.5 ± 0.06 µmol/l (11.0 ± 1.2 µg/dl) for subject I, and 0.68 ± 0.14 µmol/l (14.0 ± 3.0 µg/dl) for subject J. Although all mean blood lead levels, except that obtainedbylaboratory2forsubjectj, arewithinthe0.2-µmol/l (4-µg/dL) deviation required by the Clinical Laboratory Improvement Amendments Act of 1988 proficiency testing standard for blood lead levels of less than 1.93 µmol/l ( 40 µg/dl), it is not clear why these laboratory means are consistently less than the given target values. SENSITIVITY AND SPECIFICITY We examined the ability of each laboratory to correctly classify the individual specimens as greater or less than the CDC action level of 0.48 µmol/l (10 µg/dl). To calculate sen- 550

4 Table 2. Laboratory Results and Means by Subject and Time* Laboratory Results, µmol/l Subject Mean (SD), µmol/l A Time (0.07) Time (0.04) Laboratory means NA B Time (0.05) Time (0.04) Laboratory means NA C Time (0.04) Time (0.06) Laboratory means NA D Time (0.05) Time (0.07) Laboratory means NA E Time (0.05) Time (0.03) Laboratory means NA F Time (0.05) Time (0.04) Laboratory means NA G Time (0.04) Time (0.03) Laboratory means NA H (0.26) Time (0.05) Time (0.05) Laboratory means (0.04) I (0.57) Time (0.08) Time (0.08) Laboratory means (0.06) J (0.79) Time (0.12) Time (0.22) Laboratory means (0.14) *To convert values from micromoles per liter to micrograms per deciliter, divide by Known lead levels in parentheses for bovine blood samples. Levels given in micromole per liter. The minimum detection level was 0.05 µmol/l, reported as less than 0.05 µmol/l, and analyzed as 0.02 µmol/l. The minimum detection level for laboratory 6 was 0.14 µmol/l, reported as less than 0.14 µmol/l, and analyzed as 0.07 µmol/l. NA indicates data not available. sitivity and specificity, it is necessary to have a criterion of truth for each specimen. We used the target values for the bovine specimens and assumed, on the basis of our results and the absence of known exposures, that all volunteers had true values less than 0.48 µmol/l ( 10 µg/dl) (Table 2). Since the target values of 2 bovine blood specimens, subjects I and J, are greater than 0.48 µmol/l ( 10 µg/dl), as determined by inductively coupled plasma isotope-dilution mass spectroscopy, laboratory sensitivity for values greater than or equal to 10 µg/dl was 87.5% (28/32 samples). Specificity for values less than 0.48 µmol/l ( 10 µg/dl) was 100% (128/128 samples), although all values less than 0.48 µmol/l ( 10 µg/dl) were well below 10 µg/dl (Table 2). The highest human mean blood level was 0.2 µmol/l (4.4 µg/dl) at time 1 and 0.2 µmol/l (4.1 µg/dl) at time 2, levels that are well below the action threshold of 0.48 µmol/l (10 µg/dl). Laboratory 2 showed excessive variability and produced substantially lower results than the other laboratories. COMMENT Correlations between values at time 1 and time 2 were high in 7 of the 8 laboratories. Only 2 of 8 laboratories had significant differences between a subject s reported blood lead values at time 1 and time 2 (Table 1). Because the laboratories were unaware that these samples were part of a proficiency study and blinding was believed to have been maintained, the results suggest adequate accuracy and precision in the analysis of blood specimens at low blood lead levels. Only the performance of laboratory 2 was unsatisfactory. It consis- 551

5 Reported Blood Lead Value, µmol/l Time 1 Time Laboratory No. Reported Blood Lead Value, µmol/l Time 1 Time Laboratory No. Figure 2. Reported blood lead values by laboratory for subject J at time 1 and time 2. Dotted line represents the blood lead value (0.79 µmol/l) determined by inductively coupled plasma isotope-dilution mass spectroscopy. This was used as the target value. (To convert lead values from micromoles per liter to micrograms per deciliter, divide by ) Reported Blood Lead Value, µmol/l Time 1 Time Laboratory No. Figure 3. Reported blood lead values by laboratory for subject H at time 1 and time 2. Dotted line represents the blood lead value (0.26 µmol/l) determined by inductively coupled plasma isotope-dilution mass spectroscopy. This was used as the target value. (To convert lead values from micromoles per liter to micrograms per deciliter, divide by ) tently reported values less than the overall means for subjects A through G and underestimated the blood lead levels in subjects H through J. No other laboratory had such a substantial underestimation. Although laboratories 4 and 7 had significant differences between reported blood lead values at time 1 and time 2, these differences were not strong enough to affect the overall time trend. Only a larger sample size with more samples per laboratory would permit individual investigation. The variance found between laboratories seems to be driven by 1 laboratory with a markedly different overall mean level for all specimens. That laboratory s overall mean blood lead level seems to be driven by 1 low outlier. Since the sample size is not large enough to minimize the effects of such outliers, the borderline significance of the variance between laboratories must be considered carefully. Another indicator of adequate accuracy and precision was that all laboratory results were within 4 µg/dl of the overall mean Figure 4. Reported blood lead values by laboratory for subject I at time 1 and time 2. Dotted line represents the blood lead value (0.57 µmol/l) determined by inductively coupled plasma isotope-dilution mass spectroscopy. This was used as the target value. (To convert lead values from micromoles per liter to micrograms per deciliter, divide by ) for each subject at each time. Although Figures 2 through 4 showed substantial differences between reported blood lead values at time 1 and time 2 for subjects H through J, only 1 pair had values more than 4 µg/dl greater and 0.2 µmol/l (4 µg/dl) less than the mean of the pair, and only 4 of 48 reported blood lead values (time 1 or time 2) were more than 0.2 µmol/l ( 4 µg/dl) greater or less than the target value. Three of these unacceptable results were on samples from subject J, 2 of them from laboratory 2. Figure 1 showed that subjects A, D, and F had somewhat higher mean reported blood lead values across laboratories than the other human subjects. It also showed that the mean subject values from all laboratories, excluding subject J from laboratory 2, were within 0.2 µmol/l (4 µg/dl), which limited interlaboratory variance. Several previous studies of laboratory proficiency in blood lead determination found that performance was unsatisfactory. In an unpublished study done by our group at the Environmental and Occupational Health Sciences Institute in 1990, laboratories did not perform satisfactorily in analyzing samples with blood lead levels lower than 0.48 µmol/l ( 10 µg/dl). There were substantial differences between replicate samples, whether submitted 1 week apart or on the same day. There was obvious disparity between individual laboratories (P.01). Although some laboratories fared better than others, none performed satisfactorily, and we consequently arranged to have blood lead determinations, for a randomized trial of the effect of home cleaning, done at the CDC. Published studies, with results such as these, coupled with the decrease in children s blood lead levels, lead the CDC to promote proficiency testing programs and initiate the CDC Blood Lead Laboratory Reference System. Sargent et al 8 concluded that there was wide variation in the ability of laboratories to meet the criteria set forth by the CDC/Wisconsin Blood Lead Proficiency Testing Program, even in 1995 after proficiency programs were in place. The authors believed that further study of laboratory performance on clinical specimens was required to determine if such samples should be sent periodically to test labo- 552

6 ratories. This study used such clinical specimens, along with bovine blood samples with known lead levels, with all those involved in the analysis blinded as to the purpose of the blood samples and found that clinical laboratories have improved their performance on the analysis of lead in blood specimens, but still have room to improve. Sargent et al 8 reported that 11% of specimens with a known lead level of 0.3 µmol/l (6.9 µg/dl) were misclassified as having a level of more than 0.48 µmol/l ( 10 µg/dl), and 42% of specimens with a known lead level of 0.43 µmol/l (9.0 µg/ dl) were misclassified. They also reported that 11% of the specimens with levels known to be 0.89 µmol/l (18.4 µg/ dl) were misclassified as having levels less than 0.48 µmol/l ( 10 µg/dl), and 6% of specimens with known lead levels of 1.59 µmol/l (32.9 µg/dl) were misclassified as having levels less than 0.48 µmol/l ( 10 µg/dl). Our study compared favorably, in that all laboratories reported lead levels for subjects A through G as less than 10 µg/dl, and 0% of subject H specimens, known to be 5.4 µg/dl, were misclassified. Also, only 19% of subject I specimens, with a known lead value of 0.57 µmol/l (11.8 µg/dl), were misclassified, and 6% of subject J specimens, with a known value of 0.79 µmol/l (16.4 µg/dl), were misclassified. An initial problem in those early studies of laboratory proficiency was the absence of a definitive method that could be used to assign certified values to reference materials. In this study, the use of a definitive method (inductively coupled plasma isotope-dilution mass spectroscopy) to assign target values to the bovine quality control materials eliminated this problem. Consequently, we were able to assess the accuracy of laboratory performance by using the bovine blood specimens as standards in the interpretation of human blood lead results. Although it is not known why 6 of the 8 laboratories had means less than the target value both at time 1 and time 2, several causes seem possible, including human error and the inability of the analytical method used (graphite-furnace atomic absorption spectroscopy) to accurately reproduce the lead values generated by the inductively coupled plasma isotopedilution mass spectroscopy. Laboratory 2 substantially underreported the blood lead levels of all specimens, both at time 1 and time 2. Further study with sufficient sample size is needed to more fully explore the problems with the analyses of lead in blood specimens by laboratory 2. Overall, however, the laboratories in this study performed well in relation to what we potentially expected judging from our unpublished study and a published study. 8 Other studies 10,12,20 have addressed the issue of storage time and its effects on lead determinations in human and bovine blood. These studies recommended the use of original collection tubes, a storage temperature of 4 C, and EDTA or heparin anticoagulant. Our results support these recommendations since we show no overall time trend, even after 2 weeks of such storage. Although there were individual time trends in 2 of the 8 laboratories, they were not large enough to affect the overall time trend. A limitation to this study is the small sample size. Because only 160 samples were analyzed, we cannot decisively comment on the current status of blood lead testing in individual commercial laboratories; we can, however, suggest that overall the laboratories are performing well in analyzing samples with these low blood lead levels and seem to have improved their ability to provide more accurate and consistent results over time (since the unpublished assessment by our group in 1990). Although the commercial firms used were large national companies, the analyses could have been performed at regional laboratories. This geographic limitation implies the need for caution in generalizing the results to the nation. This limitation, as well as the inconsistent results found for laboratory 2, suggest the need for a more comprehensive study with a greater number of samples per laboratory to assess regional differences and individual laboratory performance, rather than aggregate performance. Accepted for publication December 29, This work was funded by grant from the Robert Wood Johnson Foundation, Princeton, NJ; cooperative agreement CR from the US Environmental Protection Agency, Washington, DC; and interagency agreement from the National Institute of Child Health and Human Development, Bethesda, Md; and by center grant 2P30ES05022 from the National Institute of Environmental Health Sciences, Research Triangle Park, NC. Reprints: George G. Rhoads, MD, MPH, Environmental Health Division, Environmental and Occupational Health Sciences Institute, Room 234A, 681 Frelinghuysen Rd, Piscataway, NJ REFERENCES 1. Parsons PJ. Monitoring human exposure to lead: an assessment of current laboratoryperformanceforthedeterminationofbloodlead.environres.1992;57: Patterson CC. Contaminated and natural lead environments of man. Arch Environ Health. 1965;11: Agency for Toxic Substances and Disease Registry. The Nature and Extent of Lead Poisoning in Children in the United States: A Report to Congress. Atlanta, Ga: Dept of Health and Human Services, Public Health Service; Pirkle JL, Brody DJ, Gunter EW, et al. The decline in blood lead levels in the United States: the National Health and Nutrition Examination Surveys (NHANES). JAMA. 1994;272: Centers for Disease Control and Prevention. Preventing Lead Poisoning in Young Children: A Statement by the Centers for Disease Control. Atlanta, Ga: Centers for Disease Control; Centers for Disease Control and Prevention. Blood Lead Proficiency Testing. Atlanta, Ga: Dept of Health and Human Services, Public Health Service; Clinical Laboratory Improvement Amendments Act of 1988, 42 CFR to end (1997). 8. Sargent JD, Johnson L, Roda S. Disparities in clinical laboratory performance for blood lead analysis. Arch Pediatr Adolesc Med. 1996;150: Cox DH, Duncan CE, Wyatt GP, Meadow JS, Robinson JB. Bovine blood lead reference material. J Anal Toxicol. 1989;13: Stanton N. Blood lead proficiency testing: overview of the federally sponsored program in the United States. J Int Fed Clin Chem. 1993;5: Wang ST, Peter F. The stability of human blood lead in storage. J Anal Toxicol. 1985;9: Cox DH. Bovine blood quality control material for cadmium, mercury, and lead. J Anal Toxicol. 1989;13: Miller DT, Paschal DC, Gunter EW, Stroud PE, D Angelo J. Determination of blood lead with electrothermal atomic absorption using a L vov platform and matrix modifier. Analyst. 1987;112: Parsons PJ, Slavin W. A rapid Zeeman graphite-furnace atomic absorption spectrometric method for determination of lead in blood. Spectrochimica Acta. 1993; 48B: Boone J, Hearn T, Lewis S. Comparison of interlaboratory results for blood lead with results from a definitive method. Clin Chem. 1979;25: Keppler JF, Maxfield ME, Moss WD, Tietjen G, Linch AL. Interlaboratory evaluation of the reliability of blood lead analyses. Am Ind Hyg Assoc J. 1970;31: Donovan DT, Vought VM, Rakow AB. Laboratories which conduct lead analyses on biologic specimens. Arch Environ Health. 1971;23: Bullock DG, Smith NJ, Whitehead TP. External quality assessment of assays of lead in blood. Clin Chem. 1986;32: Lind B, Elinder CG, Friberg L, Nilsson B, Svartengren M, Vahter M. Quality control in theanalysisforleadandcadmiuminblood.freseniusjanalchem.1987;326: Subramanian KS, Meranger JC, Connor J. The effect of container material, storage time, and temperature on cadmium and lead levels in heparinized human whole blood. J Anal Toxicol. 1983;7:

Assessment of Interlaboratory Performance on the Measurement of Blood Lead Levels in Taiwanese Adults

Assessment of Interlaboratory Performance on the Measurement of Blood Lead Levels in Taiwanese Adults , f Industrial Health, 1995, 33, 181-190 181 Assessment of Interlaboratory Performance on the Measurement of Blood Lead Levels in Taiwanese Adults Saou-Hsing LIOU1), Guang-Yang YANG2), Trong-Neng WU3)

More information

Toxicology Blood Lead Comprehensive Testing

Toxicology Blood Lead Comprehensive Testing Clinical Laboratory s of Practice Blood Lead Comprehensive Testing The following specialty sustaining standards of practices shall be incorporated into the laboratory s quality management system, where

More information

Toxicology Blood Lead Comprehensive Testing

Toxicology Blood Lead Comprehensive Testing Clinical Laboratory s of Practice Blood Lead Comprehensive Testing The following specialty sustaining standards of practices shall be incorporated into the laboratory s quality management system, where

More information

Toxicology Blood Lead Comprehensive Testing

Toxicology Blood Lead Comprehensive Testing Blood Lead Comprehensive Testing The following specialty sustaining standards of practices shall be incorporated into the laboratory s quality management system, where applicable to the scope of services

More information

High performance graphite furnace tube for determination of lead in blood

High performance graphite furnace tube for determination of lead in blood High performance graphite furnace tube for determination of lead in blood Application note Clinical research Authors Kai Robinson, John Sanders, James Barker Agilent Technologies, Melbourne, Australia

More information

TRACE ELEMENTS IN URINE. Event #1, 2010

TRACE ELEMENTS IN URINE. Event #1, 2010 TRACE ELEMENTS IN URINE Event #, 00 March 6, 00 Event #, 00 Urine Arsenic The source of the test materials is human urine obtained from donor volunteers with informed consent. Urine was collected into

More information

Polymer Technology Systems, Inc. CardioChek PA Comparison Study

Polymer Technology Systems, Inc. CardioChek PA Comparison Study Polymer Technology Systems, Inc. CardioChek PA Comparison Study Evaluation Protocol: Accuracy Precision Clinical Correlation PTS Panels Lipid Panel Test Strips For Use in Comparisons to a Reference Laboratory

More information

LeadCare BLOOD LEAD ANALYZER. Quick Reference Guide

LeadCare BLOOD LEAD ANALYZER. Quick Reference Guide LeadCare II BLOOD LEAD ANALYZER Quick Reference Guide Precautions Precautions Caution The LeadCare II Blood Lead Analyzer is a CLIA-waived device. Facilities that perform tests with the LeadCare II System

More information

MAGNITUDE OF SAMPLING AND ANALYTICAL VARIATIONS IN BLOOD AND BREATH ALCOHOL MEASUREMENTS

MAGNITUDE OF SAMPLING AND ANALYTICAL VARIATIONS IN BLOOD AND BREATH ALCOHOL MEASUREMENTS MAGNITUDE OF SAMPLING AND ANALYTICAL VARIATIONS IN BLOOD AND BREATH ALCOHOL MEASUREMENTS A. W. Jones, K. Jensson and P.M. Williams Department of alcohol Toxicology,j National Laboratory of Forensic Chemistry

More information

Evaluation of a Portable Blood Lead Analyzer as an Alternative to Graphite Furnace Atomic Absorption Spectrophotometer

Evaluation of a Portable Blood Lead Analyzer as an Alternative to Graphite Furnace Atomic Absorption Spectrophotometer Evaluation of a Portable Blood Lead Analyzer as an Alternative to Graphite Furnace Atomic Absorption Spectrophotometer Tom M. Olewe 1*, Mutuku A. Mwanthi 2, Joseph K. Wang ombe 2 and Jeffrey K. Griffiths

More information

Mouse C-Peptide ELISA Kit

Mouse C-Peptide ELISA Kit Mouse C-Peptide ELISA Kit Cat.No: DEIA4507 Lot. No. (See product label) Size 96T Intended Use The Mouse C-Peptide ELISA kit is for the quantitative determination of c-peptide in mouse serum, plasma, and

More information

Patient Preparation Unique patient preparation requirements are listed under each test in the Test Directory.

Patient Preparation Unique patient preparation requirements are listed under each test in the Test Directory. Specimen Collection Lab results are only as good as the specimen provided. Patient preparation, venipuncture technique, specimen handling, and transportation can all affect the quantity of results. Health

More information

APOB (Human) ELISA Kit

APOB (Human) ELISA Kit APOB (Human) ELISA Kit Catalog Number KA4330 96 assays Version: 01 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle of the Assay...

More information

AAP Lead Testing Webinar Series

AAP Lead Testing Webinar Series AAP Lead Testing Webinar Series This event is the first of a 4-part webinar series developed as part of the American Academy of Pediatrics (AAP) project, Increasing Capacity for Blood Lead Testing and

More information

Lab Director: Russell P. Tracy, Ph.D. Coordinator: Elaine Cornell Lab/Project Manager: Rebekah Boyle, M.S. Repository Manager: Sarah Nightingale

Lab Director: Russell P. Tracy, Ph.D. Coordinator: Elaine Cornell Lab/Project Manager: Rebekah Boyle, M.S. Repository Manager: Sarah Nightingale Presented by: Jessica Rooney Colchester Research Facility Laboratory for Clinical Biochemistry Research (LCBR) University of Vermont Department of Pathology and Laboratory Medicine Lab Director: Russell

More information

METHOD VALIDATION: WHY, HOW AND WHEN?

METHOD VALIDATION: WHY, HOW AND WHEN? METHOD VALIDATION: WHY, HOW AND WHEN? Linear-Data Plotter a,b s y/x,r Regression Statistics mean SD,CV Distribution Statistics Comparison Plot Histogram Plot SD Calculator Bias SD Diff t Paired t-test

More information

Carnitine / Acylcarnitines Dried Blood Spots LC-MS/MS Analysis Kit User Manual

Carnitine / Acylcarnitines Dried Blood Spots LC-MS/MS Analysis Kit User Manual Page 1 / 14 Carnitine / Acylcarnitines Dried Blood Spots LC-MS/MS Analysis Kit User Manual ZV-3051-0200-15 200 Page 2 / 14 Table of Contents 1. INTENDED USE... 3 2. SUMMARY AND EXPLANATION... 3 3. TEST

More information

Step-by-Step Instructions For OraQuick HCV Rapid Antibody Test

Step-by-Step Instructions For OraQuick HCV Rapid Antibody Test Step-by-Step Instructions For OraQuick HCV Rapid Antibody Test Complexity: WAIVED for fingerstick whole blood and venipuncture whole blood. A Certificate of CLIA Waiver is required to perform the test

More information

8.2 Principles of Quantitative Hematologic Determinations (1)

8.2 Principles of Quantitative Hematologic Determinations (1) 3 8. FIELD CENTER HEMATOLOGY SERVICES 8.1 Clinical Significance Quantitation of the formed elements of the blood (erythrocytes -RBCs, leukocytes - WBCs, and platelets) is important in the ARIC study primarily

More information

TRACE ELEMENTS IN SERUM

TRACE ELEMENTS IN SERUM TRACE ELEMENTS IN SERUM Proficiency Test Report Event #, 05 November th, 05 Event #, 05 Serum Aluminum The test materials for serum Al were prepared from human serum obtained from Tennessee Blood Services,

More information

The Virtues and Pitfalls of Implementing a New Test

The Virtues and Pitfalls of Implementing a New Test The Virtues and Pitfalls of Implementing a New Test James H. Nichols, Ph.D., DABCC, FACB Professor of Clinical Pathology, Microbiology and Immunology Associate Medical Director for Clinical Operations

More information

TRACE ELEMENTS IN SERUM

TRACE ELEMENTS IN SERUM TRACE ELEMENTS IN SERUM Proficiency Test Report Event #, 0 March th, 0 Event #, 0 Serum Aluminum The test materials for serum Al were prepared from human serum obtained from Tennessee Blood Services,

More information

Clinical Chemistry (CHE 221)

Clinical Chemistry (CHE 221) Clinical Chemistry (CHE 221) Experiment # 14 Blood Alcohol Determination by Gas Chromatography and by Reaction with Alcohol Dehydrogenase Name Date Performed Date Submitted Partners Name(s) Partners Name(s)

More information

Farouk AL Quadan lecture 3 Implentation of

Farouk AL Quadan lecture 3 Implentation of How to implement a program? Establish written policies and procedures Assign responsibility for monitoring and reviewing Train staff Obtain control materials Collect data Set target values (mean, SD) Establish

More information

A significant dietary source of lead to Nunavik Inuit: Ptarmigan hunting

A significant dietary source of lead to Nunavik Inuit: Ptarmigan hunting A significant dietary source of lead to Nunavik Inuit: Ptarmigan hunting Michael K. H. Kwan, Ph.D. Toxicologist Nunavik Research Centre, Kuujjuaq, Quebec March 2018 Abstract Lead concentration in ptarmigans

More information

Determination of metals in industrial wastewaters by microwave plasmaatomic emission spectrometry

Determination of metals in industrial wastewaters by microwave plasmaatomic emission spectrometry Determination of metals in industrial wastewaters by microwave plasmaatomic emission spectrometry Application note Environmental Author Terrance Hettipathirana Agilent Technologies Melbourne, Australia

More information

TRACE ELEMENTS IN SERUM. Event #2, 2010

TRACE ELEMENTS IN SERUM. Event #2, 2010 TRACE ELEMENTS IN SERUM Event #, 00 June 8, 00 Event #, 00 Serum Aluminum The test materials for serum Al were prepared from human serum obtained from Tennessee Blood Services, Inc. The units were tested

More information

TRACE ELEMENTS IN SERUM. Educational Event. Proficiency Test Report. Event #2, 2013

TRACE ELEMENTS IN SERUM. Educational Event. Proficiency Test Report. Event #2, 2013 TRACE ELEMENTS IN SERUM Educational Event Proficiency Test Report Event #, 0 May 8 th, 0 Event #, 0 Serum Aluminum The test materials for serum Al were prepared from human serum obtained from Tennessee

More information

METHOD 9079 SCREENING TEST METHOD FOR POLYCHLORINATED BIPHENYLS IN TRANSFORMER OIL

METHOD 9079 SCREENING TEST METHOD FOR POLYCHLORINATED BIPHENYLS IN TRANSFORMER OIL METHOD 9079 SCREENING TEST METHOD FOR POLYCHLORINATED BIPHENYLS IN TRANSFORMER OIL 1.0 SCOPE AND APPLICATION 1.1 Method 9079 may be used to screen hydrocarbon based electrical insulating fluids for polychlorinated

More information

Storage of blood and the mean corpuscular volume

Storage of blood and the mean corpuscular volume J. clin. Path., 1975, 28, 345-349 Storage of blood and the mean corpuscular volume A. C. K. LAWRENCE, J. M. BEVINGTON, AND M. YOUNG From the Haematology Department, Northern General Hospital, Sheffield

More information

SALEM HOSPITAL SALEM, OREGON 97309

SALEM HOSPITAL SALEM, OREGON 97309 SALEM HOSPITAL SALEM, OREGON 97309 Department: Phlebotomy TITLE: BLOOD COLLECTIONS, VENIPUNCTURE Area: Phlebotomy Effective Date: 04/01/96 Authored By: Cindy Humphrey, Diane Duncan Revised: 09/26/2000,

More information

PKU (Phenylketonuria) Serum HPLC Analysis Kit User Manual

PKU (Phenylketonuria) Serum HPLC Analysis Kit User Manual Page 1 / 20 PKU (Phenylketonuria) Serum HPLC Analysis Kit User Manual ZV-4003-0200-10 200 2-8 C Page 2 / 20 Table of Contents 1. INTENDED USE... 3 2. SUMMARY AND EXPLANATION... 3 3. TEST PRINCIPLE... 3

More information

Technical Information Guide

Technical Information Guide Technical Information Guide This Guide provides information relating to the selection, utilization, and interpretation of the Quantose IR test. Information provided is based on peer-reviewed publications

More information

From the Departments of Biochemistry and Medicine, Medical-Dental Schools of Alabama, Birmingham, Alabama

From the Departments of Biochemistry and Medicine, Medical-Dental Schools of Alabama, Birmingham, Alabama SERUM SODIUM AND VALUES IN FOUR HUNDRED NORMAL HUMAN SUBJECTS, DETERMINED BY THE BECKMAN FLAME PHOTOMETER* HOWARD C. ELLIOTT, JR., B.S., AND HOWARD L. HOLLEY, M.D. From the Departments of Biochemistry

More information

Worksheet. Worksheet. Worksheet. Worksheet. Student Performance Guide. Student Performance Guide

Worksheet. Worksheet. Worksheet. Worksheet. Student Performance Guide. Student Performance Guide LESSON 6-3 Laboratory Reagent Preparation and Calculations Worksheet LESSON 6-4 Chemistry Instrumentation in the Physician Office Laboratory Worksheet LESSON 6-6 Measuring Blood Glucose Worksheet LESSON

More information

Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers

Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D iochemistry Analyzers Life Sciences Data for Life. TM YSI Life Sciences White Paper 91 Authors: Kevin Schlueter, PhD,

More information

HEALTH CONSULTATION. Tom Lea Park EL PASO COUNTY METAL SURVEY EL PASO, EL PASO COUNTY, TEXAS EPA FACILITY ID: TX

HEALTH CONSULTATION. Tom Lea Park EL PASO COUNTY METAL SURVEY EL PASO, EL PASO COUNTY, TEXAS EPA FACILITY ID: TX HEALTH CONSULTATION Tom Lea Park EL PASO COUNTY METAL SURVEY EL PASO, EL PASO COUNTY, TEXAS EPA FACILITY ID: TX0000605388 September 6, 2002 Prepared by: The Texas Department of Health Under a Cooperative

More information

TRACE ELEMENTS IN SERUM. Event #3, 2011

TRACE ELEMENTS IN SERUM. Event #3, 2011 TRACE ELEMENTS IN SERUM Event #, 0 November 5, 0 Event #, 0 Serum Aluminum The test materials for serum Al were prepared from human serum obtained from Tennessee Blood Services, Inc. The units were tested

More information

Introduction: Table/Figure Descriptions:

Introduction: Table/Figure Descriptions: Introduction: We have completed the analysis of your HIV RNA Validation Study. The validation plan was designed to verify the installation of an unmodified FDA-approved HIV RNA assay into your laboratory.

More information

RESULTS OF THE BLOOD MONITORING PROGRAM AT THE GARFIELD AVENUE CHROMIUM(VI) REMEDIATION SITES

RESULTS OF THE BLOOD MONITORING PROGRAM AT THE GARFIELD AVENUE CHROMIUM(VI) REMEDIATION SITES RESULTS OF THE BLOOD MONITORING PROGRAM AT THE GARFIELD AVENUE CHROMIUM(VI) REMEDIATION SITES Robert J. Laumbach, M.D., M.P.H., C.I.H. Associate Professor Rutgers School of Public Health Environmental

More information

MHC MULTIMER PROFICIENCY PANEL 2017

MHC MULTIMER PROFICIENCY PANEL 2017 MHC MULTIMER PROFICIENCY PANEL 2017 August 2017 CONTACT Charlotte Halgreen ProficiencyPanel@immudex.com FOR MORE INFORMATION www.proficiencypanel.com MHC MULTIMER PROFICIENCY PANEL 2017 This report summarizes

More information

Title 17 California Code of Regulations As of November, 2010 (Cite as 17 CCR )

Title 17 California Code of Regulations As of November, 2010 (Cite as 17 CCR ) Title 17 California Code of Regulations As of November, 2010 (Cite as 17 CCR ) 1215. Authority. Chapter 5 Sections 436.50-436.63 of Part 1 of Division 1 of the Health and Safety Code. Note: Authority cited:

More information

Introduction 3. Childhood Lead Testing Challenges...3. Risk Factors for Childhood Lead Poisoning 4 7. National Recommendations for Lead Screening 8

Introduction 3. Childhood Lead Testing Challenges...3. Risk Factors for Childhood Lead Poisoning 4 7. National Recommendations for Lead Screening 8 TableofContents 2 Introduction 3 ChildhoodLeadTestingChallenges...3 RiskFactorsforChildhoodLeadPoisoning 47 SectionI:FortheHealthCareProvider NationalRecommendationsforLeadScreening 8 WisconsinBloodLeadScreeningRecommendations.9

More information

Morinaga Mouse C-peptide ELISA Kit

Morinaga Mouse C-peptide ELISA Kit Morinaga Mouse C-peptide ELISA Kit For the quantitative determination of C-peptide in mouse serum, plasma, and fluid 96wells For Laboratory Use Only, not for use in diagnostic procedure Please read full

More information

Mouse Leptin ELISA Kit Instructions

Mouse Leptin ELISA Kit Instructions V.6/Mar/2008 CRYSTAL CHEM INC. Mouse Leptin ELISA Kit Instructions For the quantitative determination of leptin in mouse serum or plasma and fluid Catalog #: 90030 96 Assays For Research Use Only. Not

More information

Vitamin A-E Serum HPLC Analysis Kit

Vitamin A-E Serum HPLC Analysis Kit Vitamin A-E Serum HPLC Analysis Kit ZV-4013-0200-10 200 2-8 C ZIVAK TEKNOLOJI SANAYI VE TICARET LIMITED SIRKETI Gursel Mh. Eski Besiktas Cd. No:44/1-2 KAGITHANE-ISTANBUL-TURKEY www.zivak.com info@zivak.com

More information

Multi Analyte Custom Grade Solution

Multi Analyte Custom Grade Solution 1.0 ACCREDITATION / REGISTRATION INORGANIC VENTURES is accredited to ISO Guide 34, "General Requirements for the Competence of Reference Material Producers" and ISO/IEC 17025, "General Requirements for

More information

02006B 1 vial 02006B 1 vial Store at -20 C. Lyophilized recombinant IL-2

02006B 1 vial 02006B 1 vial Store at -20 C. Lyophilized recombinant IL-2 For detection and measurement of human interleukin 2 Catalog #02006 Catalog #02007 2 Plates 10 Plates Product Description The Human Interleukin 2 (IL-2) ELISA Kit is designed for the quantitative detection

More information

Insulin ELISA. For the quantitative determination of insulin in serum and plasma

Insulin ELISA. For the quantitative determination of insulin in serum and plasma Insulin ELISA For the quantitative determination of insulin in serum and plasma For In Vitro Diagnostic use within the United States of America. This product is for Research Use Only outside of the United

More information

EliKine Free Thyroxine (ft4) ELISA Kit

EliKine Free Thyroxine (ft4) ELISA Kit EliKine Free Thyroxine (ft4) ELISA Kit Booklet Item NO. KET0005 Product Name EliKine Free Thyroxine (ft4) ELISA Kit ATTENTION For laboratory research use only. Not for clinical or diagnostic use TABLE

More information

Reference Material Institute for Clinical Chemistry Standards (ReCCS)

Reference Material Institute for Clinical Chemistry Standards (ReCCS) Reference Material Institute for Clinical Chemistry Standards (ReCCS) Certified Reference Material for Measurement of Total Cholesterol and Glycerides in Human Serum Certificate of Analysis Intended use

More information

NEGATIVE POSITIVE. Rev , 06/11. 5 min. Mono Test. For fi ngertip. blood: 1 DROP 1 DROP For serum, whole blood. plasma or. samples.

NEGATIVE POSITIVE. Rev , 06/11. 5 min. Mono Test. For fi ngertip. blood: 1 DROP 1 DROP For serum, whole blood. plasma or. samples. Mono Test 1 2 1 DROP 1 DROP For serum, plasma or whole blood samples in tubes: For fi ngertip blood: 3 4 5 min POSITIVE NEGATIVE Rev. 3078-0, 06/11 Mono Test CLIA Complexity: Waived for Whole Blood Non-Waived

More information

Counting Spermatozoa - Part 1 By B. W. Pickett, Ph.D. Professor Emeritus, Colorado State University

Counting Spermatozoa - Part 1 By B. W. Pickett, Ph.D. Professor Emeritus, Colorado State University Counting Spermatozoa - Part 1 By B. W. Pickett, Ph.D. Professor Emeritus, Colorado State University A hemacytometer, originally designed to count blood cells, has been used for many years to count other

More information

TSH ELISA Kit Medical Device Licence No.: 16419

TSH ELISA Kit Medical Device Licence No.: 16419 TSH ELISA Kit Medical Device Licence No.: 16419 Enzyme immunoassay kit for the quantitative determination of TSH concentration in serum. Catalog Number: SL100305 Catalog Number: SL100306 Catalog Number:

More information

Keywords: albuminuria; albumin/creatinine ratio (ACR); measurements. Introduction

Keywords: albuminuria; albumin/creatinine ratio (ACR); measurements. Introduction Clin Chem Lab Med 2015; 53(11): 1737 1743 Beryl E. Jacobson, David W. Seccombe*, Alex Katayev and Adeera Levin A study examining the bias of albumin and albumin/creatinine ratio measurements in urine DOI

More information

LAP-003CS. Procedure Title: Author(s): Bonnie Hames, Fannie Posey-Eddy, Chris Roth, Ray Ruiz, Amie Sluiter, David Templeton.

LAP-003CS. Procedure Title: Author(s): Bonnie Hames, Fannie Posey-Eddy, Chris Roth, Ray Ruiz, Amie Sluiter, David Templeton. Biofuels Program Biomass Analysis Technology Team Laboratory Analytical Procedure LAP-003CS Procedure Title: Determination of Acid-Insoluble Lignin in Corn Stover Author(s): Bonnie Hames, Fannie Posey-Eddy,

More information

RayBio Human Granzyme B ELISA Kit

RayBio Human Granzyme B ELISA Kit RayBio Human Granzyme B ELISA Kit Catalog #: ELH-GZMB User Manual Last revised April 15, 2016 Caution: Extraordinarily useful information enclosed ISO 13485 Certified 3607 Parkway Lane, Suite 100 Norcross,

More information

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014 Comparison Study Summary Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR 97123 July 30, 2014 1. PROTOCOL This study was conducted on July 24 th, 2014 at Tuality Healthcare, Hillsboro, OR. The

More information

I. Subject: Ionized Calcium (Ca++) Analysis Whole Blood

I. Subject: Ionized Calcium (Ca++) Analysis Whole Blood I. Subject: Ionized Calcium (Ca++) Analysis Whole Blood II. Method: i-stat III. Principle: A. Ca++: is measured by ion-selective electrode potentiometry. Concentrations are calculated from the measured

More information

Documentation, Codebook, and Frequencies

Documentation, Codebook, and Frequencies Documentation, Codebook, and Frequencies Laboratory Component: Urinary Iodine Survey Years: 2005 to 2006 SAS Export File: UIO_D.XPT First Published: July 2008 Last revised: N/A NHANES 2005 2006 Data Documentation

More information

UV Tracer TM Maleimide NHS ester

UV Tracer TM Maleimide NHS ester UV Tracer TM Maleimide HS ester Product o.: 1020 Product ame: UV-Tracer TM Maleimide-HS ester Chemical Structure: Chemical Composition: C 41 H 67 5 18 Molecular Weight: 1014.08 Appearance: Storage: Yellow

More information

METHOD 8316 ACRYLAMIDE, ACRYLONITRILE AND ACROLEIN BY HIGH PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC)

METHOD 8316 ACRYLAMIDE, ACRYLONITRILE AND ACROLEIN BY HIGH PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC) METHOD 8316 ACRYLAMIDE, ACRYLONITRILE AND ACROLEIN BY HIGH PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC) 1.0 SCOPE AND APPLICATION 1.1 The following compounds can be determined by this method: Compound Name

More information

HbA1c (Human) ELISA Kit

HbA1c (Human) ELISA Kit HbA1c (Human) ELISA Kit Cat. No.:DEIA3509 Pkg.Size:96T Intended use GHbA1c (Human) ELISA Kit is a sandwich enzyme immunoassay for the quantitative measurement of human GHbA1c. General Description vhemoglobin,

More information

ASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE

ASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE ASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE Brooke Ann Spaeth, Anne K Shephard, Mark DS Shephard, Timothy H Mathew ABSTRACT

More information

Application Note novaa 800 D. Determination of Macro and Trace Minerals as well as Toxic Trace Metals in Powdered Milk. Challenge.

Application Note novaa 800 D. Determination of Macro and Trace Minerals as well as Toxic Trace Metals in Powdered Milk. Challenge. Application Note novaa 800 D Determination of Macro and Trace Minerals as well as Toxic Trace Metals in Powdered Milk Introduction The popularity of milk and dairy products is growing worldwide, resulting

More information

2014 Continuing Compliance Master Series Best Practices in Alternative Assessment of Performance

2014 Continuing Compliance Master Series Best Practices in Alternative Assessment of Performance 2014 Continuing Compliance Master Series Best Practices in Alternative Assessment of Performance Brad S. Karon, MD, PhD, FCAP November 19, 2014 www.cap.org Today s Presenter Brad S. Karon, MD, PhD, FCAP

More information

TRACE ELEMENTS IN SERUM

TRACE ELEMENTS IN SERUM TRACE ELEMENTS IN SERUM Proficiency Test Report Event #, 0 July 5 th, 0 Event #, 0 Serum Aluminum Serum materials were obtained from the Dutch EQA (External Quality Assessment) scheme SKML (Stichting

More information

IgG Antibodies To Toxoplasma Gondii ELISA Kit Protocol

IgG Antibodies To Toxoplasma Gondii ELISA Kit Protocol IgG Antibodies To Toxoplasma Gondii ELISA Kit Protocol (Cat. No.:EK-310-85) 330 Beach Road, Burlingame CA Tel: 650-558-8898 Fax: 650-558-1686 E-Mail: info@phoenixpeptide.com www.phoenixpeptide.com INTENDED

More information

Determination of Ethanol in Breath and Estimation of Blood Alcohol Concentration with Alcolmeter S-D2

Determination of Ethanol in Breath and Estimation of Blood Alcohol Concentration with Alcolmeter S-D2 Determination of Ethanol in Breath and Estimation of Blood Alcohol Concentration with Alcolmeter S-D2 A.W. Jones and KÄ. Jönsson Departments of Alcohol Toxicology and Internal Medicine, University Hospital,

More information

TRACE ELEMENTS IN SERUM

TRACE ELEMENTS IN SERUM TRACE ELEMENTS IN SERUM Proficiency Test Report Event #, 0 November 6 th, 0 Event #, 0 Serum Aluminum Serum materials were obtained from the Dutch EQA (External Quality Assessment) scheme SKML (Stichting

More information

Center for Advanced Studies in Measurement and Assessment. CASMA Research Report

Center for Advanced Studies in Measurement and Assessment. CASMA Research Report Center for Advanced Studies in Measurement and Assessment CASMA Research Report Number 39 Evaluation of Comparability of Scores and Passing Decisions for Different Item Pools of Computerized Adaptive Examinations

More information

Human Free Thyroxine (ft4) CLIA Kit

Human Free Thyroxine (ft4) CLIA Kit Human Free Thyroxine (ft4) CLIA Kit Cat. No.:DEEL0226 Pkg.Size:96 tests Intended use For the direct quantitative determination of Free Thyroxine in human serum by chemiluminescence immunoassay (LIA). For

More information

Standardization of Insulin Immunoassays: Report of the American. Supplemental Information: Complete Materials and Methods

Standardization of Insulin Immunoassays: Report of the American. Supplemental Information: Complete Materials and Methods Standardization of Insulin Immunoassays: Report of the American Diabetes Association s Workgroup Santica Marcovina 1, Ronald R. Bowsher 2, Greg Miller 3, Myrlene Staten 4, Gary Myers 5, Samuel P. Caudill

More information

WTQA '99-15th Annual Waste Testing & Quality Assurance Symposium LESSONS LEARNED FROM PERFORMANCE EVALUATION STUDIES

WTQA '99-15th Annual Waste Testing & Quality Assurance Symposium LESSONS LEARNED FROM PERFORMANCE EVALUATION STUDIES LESSONS LEARNED FROM PERFORMANCE EVALUATION STUDIES Ruth L. Forman Quality Assurance Specialist/Principal Rock J. Vitale, CPC Technical Director or Chemistry/Principal Environmental Standards, Inc., 1140

More information

Bovine Insulin ELISA

Bovine Insulin ELISA Bovine Insulin ELISA For quantitative determination of insulin in bovine serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 80-INSBO-E01 Size: 96 wells Version:

More information

Determination of Calcium in Serum Samples by AAS Using a Fuel Lean Flame

Determination of Calcium in Serum Samples by AAS Using a Fuel Lean Flame Determination of Calcium in Serum Samples by AAS Using a Fuel Lean Flame Application Note Atomic Absorption Authors Michael J. Sommer Michael G. Rutman Elizabeth Wask-Rotter Howard Wagoner and Edward T.

More information

Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry.

Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry. 2013/07/30 A93A01315AUS A11A01641 60 ml 15 ml Pentra C400 Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry. QUAL-QA-TEMP-0846

More information

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: k100322 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE B. Purpose for Submission: Clearance of a new device C. Measurand: Whole

More information

Disclosures. Relevant Financial Relationship(s): Nothing to Disclose. Off Label Usage: Nothing to Disclose 9/19/2017. Proficiency Testing

Disclosures. Relevant Financial Relationship(s): Nothing to Disclose. Off Label Usage: Nothing to Disclose 9/19/2017. Proficiency Testing September 11, 2017 Teri Ross, Quality Specialist II Doreen Ryan, MA, MT; Quality Operations Supervisor Deborah J. Wells, MPA, MT(ASCP)SH, Quality Management Coordinator Disclosures Relevant Financial Relationship(s):

More information

PROCEDURES FOR COLLECTING SHIPPING SPECIMENS FOR MEASUREMENT OF NICOTINE, COTININE, CAFFEINE AND CATECHOLAMINES 2017

PROCEDURES FOR COLLECTING SHIPPING SPECIMENS FOR MEASUREMENT OF NICOTINE, COTININE, CAFFEINE AND CATECHOLAMINES 2017 PROCEDURES FOR COLLECTING AND SHIPPING SPECIMENS FOR MEASUREMENT OF NICOTINE, COTININE AND OTHER NICOTINE METABOLITES, TOBACCO ALKALOIDS, AND CARCINOGEN BIOMARKERS Clinical Pharmacology Laboratory San

More information

Method Comparison Report Semi-Annual 1/5/2018

Method Comparison Report Semi-Annual 1/5/2018 Method Comparison Report Semi-Annual 1/5/2018 Prepared for Carl Commissioner Regularatory Commission 123 Commission Drive Anytown, XX, 12345 Prepared by Dr. Mark Mainstay Clinical Laboratory Kennett Community

More information

Analysis of trace elements in nutraceuticals in compliance with USP chapter <2232> Elemental Contaminants in Dietary Supplements

Analysis of trace elements in nutraceuticals in compliance with USP chapter <2232> Elemental Contaminants in Dietary Supplements APPLICATION NOTE 43445 Analysis of trace elements in nutraceuticals in compliance with USP chapter Elemental Contaminants in Dietary Supplements Authors Sanja Asendorf, Application Specialist, Thermo

More information

MULTI-COMPONENT ANALYSIS OF HEAVY METALS

MULTI-COMPONENT ANALYSIS OF HEAVY METALS MULTI-COMPONENT ANALYSIS OF HEAVY METALS in Various Foods Using Shimadzu s AA, ICP, ICPMS & EDX HEAVY METALS TESTING HEAVY METALS POSE A SIGNIFICANT THREAT TO HUMAN HEALTH The U.S. Food and Drug Administration

More information

Human TSH ELISA Kit. User Manual

Human TSH ELISA Kit. User Manual Human TSH ELISA Kit User Manual Catalog number: GTX15585 GeneTex Table of Contents A. Product Description... 2 B. Kit Components... 3 C. Additional Required Materials (not included)... 3 D. Reagent Preparation...

More information

Procine sphingomyelin ELISA Kit

Procine sphingomyelin ELISA Kit Procine sphingomyelin ELISA Kit For the quantitative in vitro determination of Procine sphingomyelin concentrations in serum - plasma - celiac fluid - tissue homogenate - body fluid FOR LABORATORY RESEARCH

More information

CHEMILUMINESCENCE ENZYME IMMUNOASSAY (CLIA) Toxoplasma IgG. Cat # (20-25 C Room temp.) Volume

CHEMILUMINESCENCE ENZYME IMMUNOASSAY (CLIA) Toxoplasma IgG. Cat # (20-25 C Room temp.) Volume DIAGNOSTIC AUTOMATION, INC. 23961 Craftsman Road, Suite D/E/F, Calabasas, CA 91302 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com See external

More information

Stability of prothrombin time and activated partial thromboplastin time tests under different storage conditions

Stability of prothrombin time and activated partial thromboplastin time tests under different storage conditions Clinica Chimica Acta 300 (2000) 13 21 www.elsevier.com/ locate/ clinchim Stability of prothrombin time and activated partial thromboplastin time tests under different storage conditions L.V. Rao *, A.O.

More information

CORESTA Recommended Method No. 84

CORESTA Recommended Method No. 84 Cooperation Centre for Scientific Research Relative to Tobacco E-Vapour Sub-Group CORESTA Recommended Method No. 84 DETERMINATION OF GLYCERIN, PROPYLENE GLYCOL, WATER, AND NICOTINE IN THE AEROSOL OF E-CIGARETTES

More information

HIV Viral Load Quality Assessment Program Summary for Panel HIVVL 2018Oct26

HIV Viral Load Quality Assessment Program Summary for Panel HIVVL 2018Oct26 1 National Laboratory for HIV Reference Services National HIV and Retrovirology Laboratories National Microbiology Laboratory Public Health Agency of Canada HIV Viral Load Quality Assessment Program Summary

More information

HemoCue Hb METHOD AND SAMPLE COLLECTION

HemoCue Hb METHOD AND SAMPLE COLLECTION HemoCue Hb 201 + METHOD AND SAMPLE COLLECTION 1. PURPOSE AND SCOPE The purpose of this document is to describe the procedure for performing a Haemoglobin (Hb) test using the HemoCue Hb 201 + analyser.

More information

Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007

Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007 Forum for Collaborative HIV Research External Validation of CD4 and Viral Load Assays Paris, France June 29, 2007 Thomas J. Spira, M.D. International Laboratory Branch Global AIDS Program Centers for Disease

More information

Cryptosporidium Proficiency Test Program for Laboratories using US EPA Method 1623: Cryptosporidium and Giardia

Cryptosporidium Proficiency Test Program for Laboratories using US EPA Method 1623: Cryptosporidium and Giardia Cryptosporidium Proficiency Test Program for Laboratories using US EPA Method 1623: Cryptosporidium and Giardia in Water by Filtration/IMS/FA, December 2005 version Preferable Specifications to Establish

More information

MHC MULTIMER PROFICIENCY PANEL 2015

MHC MULTIMER PROFICIENCY PANEL 2015 MHC MULTIMER PROFICIENCY PANEL 2015 July 2015 CONTACT Charlotte Halgreen ProficiencyPanel@immudex.com FOR MORE INFORMATION www.proficiencypanel.com MHC MULTIMER PROFICIENCY PANEL 2015 This report summarizes

More information

Recent Advances in the Analysis of Ethanol in Saliva: Evaluation of the QED Device

Recent Advances in the Analysis of Ethanol in Saliva: Evaluation of the QED Device Recent Advances in the Analysis of Ethanol in Saliva: Evaluation of the QED Device A.W. Jones and k A. Jünsson Departments of Alcohol Toxicology and Internal Medicine, University Hospital, 581 85 Linköping,

More information

Accuracy and Precision in Point-of-Care Lipid Testing: CardioChek P A Point-of-Care Test System and PTS Panels Test Strips

Accuracy and Precision in Point-of-Care Lipid Testing: CardioChek P A Point-of-Care Test System and PTS Panels Test Strips Accuracy and Precision in Point-of-Care Lipid Testing: CardioChek P A Point-of-Care Test System and PTS Panels Test Strips Sponsored by Arthur Roberts, MD of The Living Heart Foundation The Living Heart

More information

FAO SPECIFICATIONS FAO PLANT PROTECTION PRODUCTS SULPHUR. FOOD AND AGRICULTURE ORGANIZATION OF THE UNITED NATIONS Rome, 1973

FAO SPECIFICATIONS FAO PLANT PROTECTION PRODUCTS SULPHUR. FOOD AND AGRICULTURE ORGANIZATION OF THE UNITED NATIONS Rome, 1973 AGP: CP/58 FAO SPECIFICATIONS FAO PLANT PROTECTION PRODUCTS SULPHUR FOOD AND AGRICULTURE ORGANIZATION OF THE UNITED NATIONS Rome, 1973 DISCLAIMER 1 FAO specifications are developed with the basic objective

More information

LABORATORY SERVICES TECHNICAL PROCEDURE

LABORATORY SERVICES TECHNICAL PROCEDURE LABORATORY SERVICES TECHNICAL PROCEDURE TITLE: BLOOD CULTURE COLLECTION PROCEDURE NUMBER: SC050.4 PERFORMING SITE(s): University Hospital, Drake Center SITE EFFECTIVE DATE 8/8/13 SUPERSEDES: 1/5/2012 RETIRED

More information

Porcine/Canine Insulin ELISA

Porcine/Canine Insulin ELISA Porcine/Canine Insulin ELISA For the quantitative determination of insulin in porcine or canine serum and plasma. Please read carefully due to Critical Changes, e.g., Calculation of Results. For Research

More information

25-Hydroxyvitamin D2-D3 Serum VD-200 UHPLC Analysis Kit User Manual

25-Hydroxyvitamin D2-D3 Serum VD-200 UHPLC Analysis Kit User Manual Page 1 / 12 25-Hydroxyvitamin D2-D3 Serum VD-200 UHPLC Analysis Kit User Manual ZV-4027-0500-10 500 2-8 C Page 2 / 12 Table of Contents 1. INTENDED USE... 3 2. SUMMARY AND EXPLANATION... 3 3. IVD SYMBOLS...

More information

Patient Information Publications Warren Grant Magnuson Clinical Center National Institutes of Health

Patient Information Publications Warren Grant Magnuson Clinical Center National Institutes of Health Warren Grant Magnuson Clinical Center National Institutes of Health What is a subcutaneous injection? A subcutaneous injection is given in the fatty layer of tissue just under the skin. A subcutaneous

More information