Direct half value layer measurements in mammography - is near enough good enough?
|
|
- Sheena Bruce
- 5 years ago
- Views:
Transcription
1 Direct half value layer measurements in mammography - is near enough good enough? Poster No.: R-0127 Congress: Type: Authors: 2014 CSM Scientific Exhibit J. Diffey 1, L. Cartwright 2, J. Crocker 1, J. Heggie 3 ; 1 Newcastle/AU, 2 SYDNEY/AU, 3 MELBOURNE/AU Keywords: DOI: Breast, Mammography, Physics, Dosimetry, Dosimetric comparison, Quality assurance /ranzcr2014/R-0127 Any information contained in this pdf file is automatically generated from digital material submitted to EPOS by third parties in the form of scientific presentations. References to any names, marks, products, or services of third parties or hypertext links to thirdparty sites or information are provided solely as a convenience to you and do not in any way constitute or imply RANZCR/AIR/ACPSEM's endorsement, sponsorship or recommendation of the third party, information, product or service. RANZCR/AIR/ ACPSEM is not responsible for the content of these pages and does not make any representations regarding the content or accuracy of material in this file. As per copyright regulations, any unauthorised use of the material or parts thereof as well as commercial reproduction or multiple distribution by any traditional or electronically based reproduction/publication method ist strictly prohibited. You agree to defend, indemnify, and hold RANZCR/AIR/ACPSEM harmless from and against any and all claims, damages, costs, and expenses, including attorneys' fees, arising from or related to your use of these pages. Please note: Links to movies,.ppt slideshows,.doc documents and any other multimedia files are not available in the pdf version of presentations. Page 1 of 11
2 Aim The accuracy of mean glandular dose (MGD) measurement in mammography depends upon several factors, one of which is half value layer (HVL). This can be calculated from measurements of air kerma with and without aluminium filters in the beam, or measured directly using a solid-state dosimetry system. The aims of this work were to: Compare direct measurements of HVL to calculated values, Assess the agreement within and between dosimetry systems from two vendors, Assess the variation in direct measurements with mas, field size and dosimeter setting, and Examine the effect of these variations in HVL on MGD. Methods and materials HVL measurements were made on two full field digital mammography systems with a range of target/filter combinations: GE Senographe DS with Mo/Mo, Mo/Rh and Rh/Rh Hologic Selenia Dimensions with W/Rh and W/Ag For all exposures, lead rubber was placed on the breast support platform to protect the digital detector. The dosimeter was placed 6 cm back from the chest wall edge in the midline and the compression paddle was always in the beam. The following investigations were carried out: Comparison of Calculated HVL to Direct Measurement For each target/filter combination, HVL was calculated using high purity aluminium (Al) filters (Fig. 1 on page 3) and measured directly using 4 dosimeters (3 from Vendor A; 1 from Vendor B). Measurement of HVL using Al filters requires raising the compression paddle to reduce scatter (Fig. 2 on page 4) and collimating the X-ray beam to the sensitive detector area to achieve narrow beam geometry (Fig. 3 on page 6). A measurement of air kerma is made with no attenuating material in the beam, then Al filters are added in 0.1 Page 2 of 11
3 mm increments (Fig. 4 on page 6) until less than half of the initial unattenuated dose is recorded. All measurements were made using 20 mas. Direct measurements were made using broad beam geometry (no collimation of the X- ray field), the "Paddle" dosimeter setting and 10 mas. Three measurements were made for each target/filter combination to assess repeatability. Agreement within and between dosimetry systems from two vendors The direct measurements from the 4 dosimeters were compared for each target/filter combination to assess consistency within and between dosimetry systems from two vendors. Direct measurement variation with mas Direct measurements were obtained using 10, 20 and 40 mas. Direct measurement variation with field size Direct measurements were made using narrow beam geometry (Fig. 3 on page 6) and broad beam geometry. Direct measurement variation with dosimeter setting When multiple possible dosimetry settings were available, including whether the paddle was in beam or not, these were investigated Mean glandular dose (MGD) was calculated for each measured HVL, assuming typical kv and mas values from mammography quality control tests. The effect of HVL variation on MGD was examined. Images for this section: Page 3 of 11
4 Fig. 1: High purity Al filters of varying thicknesses are used to determine HVL Page 4 of 11
5 Page 5 of 11
6 Fig. 2: Experimental set-up to determine HVL using Al filters. The compression paddle is raised to reduce scatter. Fig. 3: Narrow beam geometry. A manual collimator was placed on the raised paddle to restrict the field of view to the sensitive detector area (illuminated by the light field). Page 6 of 11
7 Fig. 4: Al filters were added in 0.1 mm increments until less than half of the unattenuated dose was recorded. Page 7 of 11
8 Results Comparison of Calculated HVL to Direct Measurement Fig. 5 on page shows the direct HVL measurements and their variation from the calculated values for Mo/Mo, Mo/Rh and Rh/Rh target/filter combinations. All HVL values were within the acceptable range, as defined in the RANZCR spreadsheet. Fig. 5: Comparison of calculated and directly measured HVL on the GE Senograph DS with Mo/Mo, Mo/Rh and Rh/Rh beam qualities References: Hunter New England Imaging - Newcastle/AU The maximum deviations from the calculated values were exhibited by Vendor A for Mo/ Mo (8.5%) and Mo/Rh (10.9%) and by Vendor B for Rh/Rh (-10.8%). The corresponding differences in MGD were 7.6%, 11.9% and -8.7% respectively. For the W target, our initial direct HVL measurements differed from the calculated values by a maximum of 3.9% (W/Rh) and -2.4% (W/Ag) for Vendor A and -22.7% (W/Rh) and -32.0% (W/Ag) for Vendor B. These results suggest that the dosimeter from Vendor B had not been correctly calibrated, although it is interesting to note that the HVL measurements were within the RANZCR-defined acceptable range. Page 8 of 11
9 Our results were shared with Vendor B, who supplied a file to correct the problem. Fortunately, there was no need to send the dosimeter to them for re-calibration. Measurements were repeated and results are shown in Fig. 6 on page for W/Rh and W/Ag. Fig. 6: Comparison of calculated and directly measured HVL on the Hologic Selenia Dimensions with W/Rh and W/Ag beam qualities References: Hunter New England Imaging - Newcastle/AU Despite the improvements, the maximum deviations from the calculated values were still exhibited by Vendor B, but were reduced to -5.6% for W/Rh and -10.2% for W/Ag. The corresponding differences in MGD were -4.2% and -8.3% respectively. Agreement within and between dosimetry systems from two vendors All dosimeters exhibited good repeatability. The coefficient of variation (COV) was dependent upon target/filter combination for Vendor A and ranged from ; COV for Vendor B was for all target/filters. The 3 units from Vendor A gave consistent results for all target/filter combinations, with a typical COV of Page 9 of 11
10 The direct HVL measurements from Vendor A were higher than those for Vendor B, as shown in Fig 7, with the extent of agreement being dependent upon target/filter combination. Fig. 7: Percentage difference in direct HVL measurements between Vendor A and Vendor B References: Hunter New England Imaging - Newcastle/AU It is interesting to note that when compared to the calculated HVL for Rh/Rh and W/Rh, Vendor A over-estimated the value and Vendor B under-estimated the value. Direct measurement variation with mas, field size and dosimeter setting Direct measurements were not significantly affected by mas used (<1%), beam geometry (<2%) or dosimeter setting used (Paddle/No Paddle #2% and W/Rh / W/Rh* <1%). Conclusion Direct HVL measurements exhibit vendor variation and differ from the calculated HVL. However, the variation is within vendor-specification; Vendor A guarantees that their Page 10 of 11
11 direct HVL measurement is accurate to within ±5%, while Vendor B guarantees that their direct HVL measurement is accurate to within ±10%. The effect on resulting MGD calculations is generally not significant considering the other errors and assumptions involved in determining MGD. A 10% increase in the calculated MGD would be very unlikely to cause MGD to exceed the acceptable level, especially for mammography equipment with integrated digital detectors. In Australia, there is no tolerance level for MGD variation with time, although in the UK, MGD must remain within 25% of the baseline value. So near enough is good enough (if dosimeters have been correctly calibrated) and direct HVL measurement saves time during mammography quality control testing. However, we recommend that the accuracy of direct HVL measurements is verified for each target/ filter combination for new solid-state dosimetry systems and after each calibration. Personal information References Sobol WT and Wu X, Parametrization of mammography normalized average glandular dose tables, Med Phys 24 (1997) Boone JM, Fewell TR and Jennings RJ, Molybdenum, rhodium, and tungsten anode spectral models using interpolating polynomials with application to mammography, Med. Phys. 24 (1997) Boone JM, Normalized glandular dose.dgn. coefficients for arbitrary x-ray spectra in mammography: Computer-fit values of Monte Carlo derived data Med. Phys. 29 (2002) Page 11 of 11
Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA
Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA Poster No.: C-0618 Congress: ECR 2011 Type: Authors: Scientific Paper T. Otsuka 1, S. Suzuki 1, M. Ogura 1, T. Igarashi
More informationUsing Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography
Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography Poster No.: C-2078 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Exhibit F. Salmani Rezaei,
More informationAn audit of radiation dose of 4D CT in a radiotherapy department
An audit of radiation dose of 4D CT in a radiotherapy department Poster No.: R-0097 Congress: Type: Authors: Keywords: DOI: 2014 CSM Scientific Exhibit T. Hubbard, J. Callahan, J. Cramb, R. Budd, T. Kron;
More informationApplication of international standards to diagnostic radiology dosimetry
Application of international standards to diagnostic radiology dosimetry Poster No.: C-780 Congress: ECR 2009 Type: Scientific Exhibit Topic: Physics in Radiology Authors: I. D. McLean, A. Meghzifene,
More informationA dosimetric analysis of the Aeroform(tm) tissue expander in radiation therapy
A dosimetric analysis of the Aeroform(tm) tissue expander in radiation therapy Poster No.: R-0008 Congress: 2014 CSM Type: Scientific Exhibit Authors: T. Tran, W. Ding, L. Melvern, M. Chao, B. Subramanian;
More informationImaging of gastrointestinal perforation: Is there a place for plain radiography?
Imaging of gastrointestinal perforation: Is there a place for plain radiography? Poster No.: R-0243 Congress: Type: Authors: 2014 CSM Scientific Exhibit M. L. Chan 1, A. Steward 2, M. Schneider 3 ; 1 BOX
More informationAlfred Health's CT quality control program
Alfred Health's CT quality control program Poster No.: R-0037 Congress: Type: Authors: 2014 CSM Scientific Exhibit A. Perdomo, Z. Brady, N. Tran, L. Hudson, K. Provis; PRAHRAN/ AU Keywords: Quality assurance,
More informationImpact of objective volumetric breast density estimates on mean glandular dose calculations in digital mammography
Impact of objective volumetric breast density estimates on mean glandular dose calculations in digital mammography Poster No.: C-1576 Congress: ECR 2015 Type: Scientific Exhibit Authors: G. Gennaro, I.
More informationDose reduction in Hologic Selenia FFDM units through AEC optimization, without compromising diagnostic image quality.
Dose reduction in Hologic Selenia FFDM units through AEC optimization, without compromising diagnostic image quality. Poster No.: C-1636 Congress: ECR 2015 Type: Authors: Keywords: DOI: Scientific Exhibit
More informationCan the compact economical Nano-X linear accelerator system be a solution for improving access to radiotherapy globally?
Can the compact economical Nano-X linear accelerator system be a solution for improving access to globally? Poster No.: R-0142 Congress: Type: Authors: 2014 CSM Scientific Exhibit P. Lazarakis, E. Eslick,
More informationLifetime risk of radiation-induced cancer from screening mammography
Lifetime risk of radiation-induced cancer from screening mammography Poster No.: C-0558 Congress: ECR 2015 Type: Scientific Exhibit Authors: R. M. K. M.Ali, A. England, P. Hogg; Manchester/UK Keywords:
More informationStandardizing mammographic breast compression: Pressure rather than force?
Standardizing mammographic breast compression: Pressure rather than force? Poster No.: C-2383 Congress: ECR 2012 Type: Authors: Scientific Paper J. E. de Groot 1, W. Branderhorst 1, M. Broeders 2, G. J.
More informationUtilization patterns and yield of Computed Tomography Pulmonary Angiogram (CTPA) at a tertiary teaching hospital - Liverpool Hospital
Utilization patterns and yield of Computed Tomography Pulmonary Angiogram (CTPA) at a tertiary teaching hospital - Liverpool Hospital Poster No.: R-0017 Congress: Type: Authors: Keywords: DOI: 2014 CSM
More informationRadiation dose levels assessment in mammography
Radiation dose levels assessment in mammography Poster No.: B-0871 Congress: ECR 2013 Type: Scientific Paper Authors: C. N. Lourenço, P. Sousa, S. Rodrigues, A. F. Abrantes, L. P. V. Ribeiro, K. B. Azevedo,
More informationMeasuring cell density in prostate cancer imaging as an input for radiotherapy treatment planning
Measuring cell density in prostate cancer imaging as an input for radiotherapy treatment planning Poster No.: R-0262 Congress: 2014 CSM Type: Scientific Exhibit Authors: H. Reynolds, S. Williams, A. Zhang,
More informationMultislice CT versus cone beam CT: dosimetric and image quality comparision.
Multislice CT versus cone beam CT: dosimetric and image quality comparision. Poster No.: C-2589 Congress: ECR 2015 Type: Authors: Scientific Exhibit C. Dionisi 1, A. Crispo 2, E. Stefani 3, A. Lovaglio
More informationThe use of SpaceOAR hydrogel in dose-escalated prostate cancer radiotherapy and its impact on rectal dosimetry
The use of SpaceOAR hydrogel in dose-escalated prostate cancer radiotherapy and its impact on rectal dosimetry Poster No.: R-0296 Congress: Type: Authors: Keywords: DOI: 2014 CSM Scientific Exhibit F.
More information64-MDCT imaging of the pancreas: Scan protocol optimisation by different scan delay regimes
64-MDCT imaging of the pancreas: Scan protocol optimisation by different scan delay regimes Poster No.: C-051 Congress: ECR 2009 Type: Scientific Exhibit Topic: Abdominal and Gastrointestinal Authors:
More informationDosimetry in digital mammography
Dosimetry in digital mammography Professor David Dance NCCPM, Royal Surrey County Hospital, Guildford, United kingdom Outline Why do dosimetry? History Essentials of European breast dosimetry protocol
More informationArtifact in Head CT Images Due to Air Bubbles in X-Ray Tube Oil
Artifact in Head CT Images Due to Air Bubbles in X-Ray Tube Oil Poster No.: C-0671 Congress: ECR 2016 Type: Educational Exhibit Authors: H. Patel 1, W. Liu 2, J. DeSanto 2, S. Meagher 2, M. Zagardo 2,
More informationFDG-18 PET/CT - radiation dose and dose-reduction strategy
FDG-18 PET/CT - radiation dose and dose-reduction strategy Poster No.: C-1856 Congress: ECR 2014 Type: Authors: Keywords: DOI: Scientific Exhibit P. Nicholson, S. McSweeney, K. O'Regan; Cork/IE Radiation
More informationSlowly growing malignant nodules and rapidly growing benign nodules: Evaluation of the value of volume doubling time
Slowly growing malignant nodules and rapidly growing benign nodules: Evaluation of the value of volume doubling time Poster No.: C-208 Congress: ECR 2009 Type: Educational Exhibit Topic: Chest Authors:
More informationContrast enhanced spectral mammography: A literature review
Contrast enhanced spectral mammography: A literature review Poster No.: R-0172 Congress: RANZCR-AOCR 2012 Type: Authors: Keywords: DOI: Educational Exhibit S. Buzynski, D. Taylor; Perth/AU Breast, Digital
More informationRadiotherapy Utilisation in BRAF-mutation Tested Metastatic Melanoma in the Targeted Therapy Era
Radiotherapy Utilisation in BRAF-mutation Tested Metastatic Melanoma in the Targeted Therapy Era Award: Wiley-Blackwell Best Exhibit Award (Radiation Oncology) Poster No.: R-0073 Congress: RANZCR ASM 2013
More informationBI-RADS 3, 4 and 5 lesions on US: Five categories and their diagnostic efficacy and pitfalls in interpretation
BI-RADS 3, 4 and 5 lesions on US: Five categories and their diagnostic efficacy and pitfalls in interpretation e-poster: C-118 Congress: ECR 2008 Type: Educational Exhibit Topic: Breast / Ultrasound Authors:
More informationProstate gold seed fiducial implantation by radiation oncologists: A report on feasibility
Prostate gold seed fiducial implantation by radiation oncologists: A report on feasibility Poster No.: R-0006 Congress: Type: Authors: Keywords: DOI: 2014 CSM Scientific Exhibit C. Jayaratne, A. Brown,
More informationIdentification and numbering of lumbar vertebrae using various anatomical landmarks on MRI of lumbosacral spine
Identification and numbering of lumbar vertebrae using various anatomical landmarks on MRI of lumbosacral spine Poster No.: C-2125 Congress: ECR 2015 Type: Authors: Scientific Exhibit S. patil 1, A. M.
More informationRadiological features of Legionella Pneumophila Pneumonia
Radiological features of Legionella Pneumophila Pneumonia Poster No.: E-0048 Congress: ESTI 2012 Type: Scientific Exhibit Authors: M. Vinciguerra, L. Stefanetti, E. Teti, G. Argentieri, L. G. 1 1 1 1 1
More informationAFib is the most common cardiac arrhythmia and its prevalence and incidence increases with age (Fuster V. et al. Circulation 2006).
Feasibility, image quality and radiation dose of coronary CT angiography (CCTA) in patients with atrial fibrillation using a new generation 256 multi-detector CT (MDCT) Poster No.: C-2378 Congress: ECR
More informationSingle cold nodule in Graves' disease: benign vs malignant
Single cold nodule in Graves' disease: benign vs malignant Poster No.: C-0073 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Paper L. I. Sonoda, M. Halim, K. Balan; Cambridge/UK Head and neck,
More informationEndorectal Balloon during Image Guided Radiation Therapy for Prostate Carcinoma Reduces Radiation Proctitis at 2 Years
Endorectal Balloon during Image Guided Radiation Therapy for Prostate Carcinoma Reduces Radiation Proctitis at 2 Years Poster No.: R-0041 Congress: Type: 2015 ASM Scientific Exhibit Authors: R. Botten,
More informationEmerging techniques in management of post-catheterisation femoral artery psuedoaneurysms
Emerging techniques in management of post-catheterisation femoral artery psuedoaneurysms Poster No.: R-0072 Congress: RANZCR ASM 2013 Type: Educational Exhibit Authors: S. Davis, J. Huang, G. Pavilion,
More informationPGMI classification of screening mammograms prior to interval cancer. Comparison with radiologists' consensus classification.
PGMI classification of screening mammograms prior to interval cancer. Comparison with radiologists' consensus classification. Poster No.: C-0673 Congress: ECR 2016 Type: Authors: Keywords: DOI: Scientific
More informationReliability of the pronator quadratus fat pad sign to predict the severity of distal radius fractures
Reliability of the pronator quadratus fat pad sign to predict the severity of distal radius fractures Poster No.: C-0669 Congress: ECR 2014 Type: Scientific Exhibit Authors: J. Tonak, I. Wobbe, R. L. Duschka,
More informationComputed tomography for pulmonary embolism: scan assessment of a one-year cohort and estimated cancer risk associated with diagnostic irradiation.
Computed tomography for pulmonary embolism: scan assessment of a one-year cohort and estimated cancer risk associated with diagnostic irradiation. Award: Poster No.: E-0004 Certificate of Merit Congress:
More informationCavitary lung lesion: Two different diagnosis with similar appearence
Cavitary lung lesion: Two different diagnosis with similar appearence Poster No.: P-0043 Congress: ESTI 2015 Type: Educational Poster Authors: M. Yesildag, H. Kalkan, K. Ödev; Konya/TR Keywords: Infection,
More informationThe Role of Radionuclide Lymphoscintigraphy in the Diagnosis of Lymphedema of the Extremities
The Role of Radionuclide Lymphoscintigraphy in the Diagnosis of Lymphedema of the Extremities Poster No.: C-1229 Congress: ECR 2014 Type: Authors: Keywords: DOI: Scientific Exhibit M. Osher 1, A. Pallas
More informationEvaluation and Optimization of Exposure Index in Lumbar Spine and Pelvis X-Ray
Evaluation and Optimization of Exposure Index in Lumbar Spine and Pelvis X-Ray Poster No.: C-1492 Congress: ECR 2012 Type: Scientific Exhibit Authors: R. M. Lopes, L. H. Domingos ; Mortágua/PT, Azoia de
More informationDEXA Scores and Bone Density Measured on Routine CT Scans
DEXA Scores and Bone Density Measured on Routine CT Scans Poster No.: R-0072 Congress: 2015 ASM Type: Scientific Exhibit Authors: C. Annabattula, P. Phadke; Sydney/AU Keywords: Screening, Nuclear medicine
More informationCognitive target MRI-TRUS fusion biopsies of MRI detected PIRADS 4 and 5 lesions
Cognitive target MRI-TRUS fusion biopsies of MRI detected PIRADS 4 and 5 lesions Poster No.: B-0704 Congress: ECR 2015 Type: Scientific Paper Authors: P. P. van Westerveld, J. Vriesema, J. H. W. van den
More informationComputed tomographic dacryocystography as compared with X-ray dacryocystography in patients with dacryostenosis
Computed tomographic dacryocystography as compared with X-ray dacryocystography in patients with dacryostenosis Poster No.: C-1887 Congress: ECR 2016 Type: Authors: Keywords: DOI: Educational Exhibit M.
More informationUsing diffusion-tensor imaging and tractography (DTT) to study biological characteristics of glyoma in brain stem for neurosurgical planning
Using diffusion-tensor imaging and tractography (DTT) to study biological characteristics of glyoma in brain stem for neurosurgical planning Poster No.: C-0016 Congress: ECR 2014 Type: Authors: Keywords:
More informationComputed tomography and Modified RECIST criteria for assessment of response in malignant pleural mesothelioma
Computed tomography and Modified RECIST criteria for assessment of response in malignant pleural mesothelioma Poster No.: C-0729 Congress: ECR 2013 Type: Scientific Exhibit Authors: A. Marin, I. Pozek,
More informationDiffusion-weighted MRI (DWI) "claw sign" is useful in differentiation of infectious from degenerative Modic I signal changes of the spine
Diffusion-weighted MRI (DWI) "claw sign" is useful in differentiation of infectious from degenerative Modic I signal changes of the spine Poster No.: C-0894 Congress: ECR 2012 Type: Scientific Exhibit
More informationMonitoring neo-adjuvant chemotherapy: comparison of contrast-enhanced spectral mammography (CESM) and MRI versus breast cancer characteristics
Monitoring neo-adjuvant chemotherapy: comparison of contrast-enhanced spectral mammography (CESM) and MRI versus breast cancer characteristics Poster No.: B-1062 Congress: ECR 2016 Type: Scientific Paper
More informationRadiotherapy and Conservative Surgery For Merkel Cell Carcinoma - The British Columbia Cancer Agency Experience
Radiotherapy and Conservative Surgery For Merkel Cell Carcinoma - The British Columbia Cancer Agency Experience Poster No.: RO-0003 Congress: RANZCR FRO 2012 Type: Scientific Exhibit Authors: C. Harrington,
More informationComparison of rectal and bladder ICRU point doses to the GEC ESTRO volumetric doses in Cervix cancer
Comparison of rectal and bladder ICRU point doses to the GEC ESTRO volumetric doses in Cervix cancer Poster No.: RO-0049 Congress: RANZCR FRO 202 Type: Authors: Scientific Exhibit G. Govindarajulu, A.
More informationScientific Exhibit Authors:
Clinical Audit on Optimization of Radiation Dose from MDCT: Effect on Diagnostic Reference Levels for Brain, Sinus, Cervical Spine, Chest, Abdomen-Pelvis, and Lumbar Spine Examinations and on Nationwide
More informationScientific Exhibit. Authors: D. Takenaka, Y. Ohno, Y. Onishi, K. Matsumoto, T.
The feasibility of biphasic contrast-media-injection-protocol for chest imaging on 320-slice volume MDCT: Direct comparison of biphasic and bolus contrast-media injection protocols on 320-slice volume
More informationDual energy CT in diagnosis of Gout
Dual energy CT in diagnosis of Gout Poster No.: R-0060 Congress: RANZCR ASM 2013 Type: Educational Exhibit Authors: F. Tabatabaie Moghadam, A. Moghaddam, F. Ghazanfari ; 1 1 1 2 2 Brisbane/AU, Melbourne/AU
More informationRole of positron emission mammography (PEM) for assessment of axillary lymph node status in patients with breast cancer
Role of positron emission mammography (PEM) for assessment of axillary lymph node status in patients with breast cancer Poster No.: C-1260 Congress: ECR 2011 Type: Scientific Paper Authors: K. M. Kulkarni,
More informationScientific Exhibit Authors: V. Moustakas, E. Karallas, K. Koutsopoulos ; Rodos/GR, 2
Diagnosis of Acute Appendicitis: the role of Color Doppler Ultrasound as first-line imaging method and evaluation of the higher diagnostic performances of CT against its disadvantages. Poster No.: C-0708
More informationComparison of Image quality in temporal bone MRI at 3T using 2D selective RF excitation versus a routine SPACE sequence
Comparison of Image quality in temporal bone MRI at 3T using 2D selective RF excitation versus a routine SPACE sequence Poster No.: C-1065 Congress: ECR 2015 Type: Authors: Keywords: DOI: Scientific Exhibit
More informationHyperechoic breast lesions can be malignant.
Hyperechoic breast lesions can be malignant. Poster No.: C-0041 Congress: ECR 2015 Type: Educational Exhibit Authors: G. Babu, R. bradley; Edinburgh/UK Keywords: Breast, Ultrasound, Biopsy, Cancer DOI:
More informationDiffuse high-attenuation within mediastinal lymph nodes on non-enhanced CT scan: Usefulness in the prediction of benignancy
Diffuse high-attenuation within mediastinal lymph nodes on non-enhanced CT scan: Usefulness in the prediction of benignancy Poster No.: C-1785 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific
More informationHigh density thrombi of pulmonary embolism on precontrast CT scan: Is it dangerous?
High density thrombi of pulmonary embolism on precontrast CT scan: Is it dangerous? Poster No.: C-1753 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit B. Y. Lee, H. R. KIM, J. I. Jung,
More informationComputed tomography for the detection of thumb base osteoarthritis, comparison with digital radiography.
Computed tomography for the detection of thumb base osteoarthritis, comparison with digital radiography. Poster No.: C-1981 Congress: ECR 2012 Type: Scientific Exhibit Authors: M. S. Saltzherr, J. W. van
More informationPurpose. Methods and Materials. Results
Prevalence and significance of hypoattenuating hepatic lesions deemed too small to characterise: How are we following up these lesions and what are the outcomes? Poster No.: C-014 Congress: ECR 2009 Type:
More informationTissue characterisation, Cancer, Quality assurance /ecr2015/B-0553
Role of DWI at prostatic lesions at 3T-MRI in the discrimination of grading: correlation of imaging, quantitative analysis and pathology at 189 MR-guided prostate biopsies Poster No.: B-0553 Congress:
More informationDigital chest radiography: collimation and dose reduction
Digital chest radiography: collimation and dose reduction Poster No.: C-1939 Congress: ECR 2015 Type: Scientific Exhibit Authors: J. Debess, K. Johnsen, K. Vejle Sørensen, H. Thomsen ; 1 1 2 1 1 2 Aalborg
More informationSignificance of MRI in diagnostics, outcome prognosis and definition the therapeutic tactics for cases of aseptic necrosis of the femoral head
Significance of MRI in diagnostics, outcome prognosis and definition the therapeutic tactics for cases of aseptic necrosis of the femoral head Abstract: 539 Congress: ESMRMB 2013 Type: Scientific Poster
More informationOut-of-field organ doses from therapeutic irradiation during childhood: is there an excess risk for second cancer induction?
Out-of-field organ doses from therapeutic irradiation during childhood: is there an excess risk for second cancer induction? Poster No.: C-0096 Congress: ECR 2012 Type: Scientific Paper Authors: M. Mazonakis,
More informationSeemingly isolated greater trochanter fractures do not exist
Seemingly isolated greater trochanter fractures do not exist Poster No.: B-0950 Congress: ECR 2012 Type: Scientific Paper Authors: D. Dunker, J. H. Göthlin, M. Geijer ; Gothenburg/SE, Lund/SE Keywords:
More informationSonographic and Mammographic Features of Phyllodes Tumours of the Breast: Correlation with Histological Grade
Sonographic and Mammographic Features of Phyllodes Tumours of the Breast: Correlation with Histological Grade Poster No.: C-0046 Congress: ECR 2014 Type: Authors: Keywords: DOI: Scientific Exhibit C. Y.
More informationShear Wave Elastography in diagnostics of supraspinatus tendon.
Shear Wave Elastography in diagnostics of supraspinatus tendon. Poster No.: C-2168 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit V. Saltykova; Moscow/RU Musculoskeletal joint, Musculoskeletal
More informationMonophasic versus biphasic contrast application in CT of patients with head and neck tumour
Monophasic versus biphasic contrast application in CT of patients with head and neck tumour Poster No.: C-3331 Congress: ECR 2010 Type: Topic: Authors: Keywords: DOI: Scientific Exhibit Head and Neck G.
More informationQuantifying Dual-energy computed tomography (DECT) in patients with renal calculi using a Toshiba Aquilion One Scanner.
Quantifying Dual-energy computed tomography (DECT) in patients with renal calculi using a Toshiba Aquilion One Scanner. Poster No.: C-2613 Congress: ECR 2015 Type: Authors: Keywords: DOI: Scientific Exhibit
More informationDoes the synthesised digital mammography (3D-DM) change the ACR density pattern?
Does the synthesised digital mammography (3D-DM) change the ACR density pattern? Poster No.: B-0211 Congress: ECR 2015 Type: Scientific Paper Authors: P. MARTÍNEZ MIRAVETE, M. Millor Muruzábal, P. García-
More informationUltrasonic evaluation of superior mesenteric vein in cancer of the pancreatic head
Ultrasonic evaluation of superior mesenteric vein in cancer of the pancreatic head Poster No.: C-1430 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Exhibit E. Fisenko, N. Vetsheva, E. Pershina;
More informationAnatomical Variations of the Levator Scapulae Muscle - an MR Imaging Study
Anatomical Variations of the Levator Scapulae Muscle - an MR Imaging Study Poster No.: R-0016 Congress: 2015 ASM Type: Scientific Exhibit Authors: J. Au, A. Webb, G. Buirski, P. Smith, M. Pickering, D.
More informationUser Guide for Dental and Maxillofacial Cone Beam Computed Tomography (CBCT)
User Guide for Dental and Maxillofacial Cone Beam Computed Tomography (CBCT) Poster No.: C-0756 Congress: ECR 2014 Type: Educational Exhibit Authors: J. Ukkonen, J. Asp; Helsinki/FI Keywords: Education
More informationMR-guided prostatic biopsy at 3T: the role of PI-RADS-score: a histopahologic-radiologic correlation
MR-guided prostatic biopsy at 3T: the role of PI-RADS-score: a histopahologic-radiologic correlation Poster No.: B-0969 Congress: ECR 2015 Type: Scientific Paper Authors: A. Malich; Nordhausen/DE Keywords:
More informationMedical device adverse incident reporting in interventional radiology
Medical device adverse incident reporting in interventional radiology Poster No.: C-0226 Congress: ECR 2014 Type: Authors: Scientific Exhibit N. PARVIZI 1, I. Robertson 2, R. G. McWilliams 3 ; 1 Oxford/UK,
More informationThoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters
Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters Poster No.: C-2584 Congress: ECR 2015 Type: Scientific
More information"Ultrasound measurements of the lateral ventricles in neonates: A comparison of multiple measurements methods."
"Ultrasound measurements of the lateral ventricles in neonates: A comparison of multiple measurements methods." Poster No.: C-1557 Congress: ECR 2014 Type: Authors: Keywords: DOI: Educational Exhibit I.
More informationComparison of MRI and ultrasound based liver volumetry in iron overload diseases
Comparison of MRI and ultrasound based liver volumetry in iron overload diseases Poster No.: C-0913 Congress: ECR 2015 Type: Scientific Exhibit Authors: S. Keller, B. P. Schönnagel, P. Nielsen, M. T. Brehmer,
More informationBolus administration of esmolol allows for safe and effective heart rate control during coronary computed tomography angiography
Bolus administration of esmolol allows for safe and effective heart rate control during coronary computed tomography angiography Poster No.: C-1342 Congress: ECR 2013 Type: Scientific Exhibit Authors:
More informationInfluence of pulsed fluoroscopy and special radiation risk training on the radiation dose in pneumatic reduction of ileocoecal intussusceptions.
Influence of pulsed fluoroscopy and special radiation risk training on the radiation dose in pneumatic reduction of ileocoecal intussusceptions. Poster No.: C-0599 Congress: ECR 2013 Type: Authors: Keywords:
More informationThe effect of CT dose reduction on performance of a diagnostic task
The effect of CT dose reduction on performance of a diagnostic task Poster No.: B-0856 Congress: ECR 2012 Type: Scientific Paper Authors: D. Zhang, M. Khatonabadi, H. Kim, C. Jude, E. Zaragoza, S. McNitt-Gray,
More informationThe effect of CT dose reduction on performance of a diagnostic task
The effect of CT dose reduction on performance of a diagnostic task Poster No.: B-0856 Congress: ECR 2012 Type: Scientific Paper Authors: D. Zhang, M. Khatonabadi, H. Kim, C. Jude, E. Zaragoza, S. McNitt-Gray,
More informationBasic low - field MR imaging of meniscal injuries in children.
Basic low - field MR imaging of meniscal injuries in children. Poster No.: C-2365 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Exhibit A. Yakimov, M. Nikonova, E. Prokhorova, D. Vybornov,
More informationQuantitative imaging of hepatic cirrhosis on abdominal CT images
Quantitative imaging of hepatic cirrhosis on abdominal CT images Poster No.: C-0556 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit S. Kido, A. Nakamura, Y. Hirano; Ube/JP Cirrhosis,
More informationA pictorial review of normal anatomical appearences of Pericardial recesses on multislice Computed Tomography.
A pictorial review of normal anatomical appearences of Pericardial recesses on multislice Computed Tomography. Poster No.: C-1787 Congress: ECR 2012 Type: Educational Exhibit Authors: N. Ahmed 1, G. Avery
More informationRole of ultrasound in the evaluation of the ileocecal valve
Role of ultrasound in the evaluation of the ileocecal valve Poster No.: C-1581 Congress: ECR 2010 Type: Scientific Exhibit Topic: GI Tract Authors: M. Mohammed, M. Hussain, U. Momin, S. Lakhtakia, N. D.
More informationFeasibility of magnetic resonance elastography using myofascial phantom model
Feasibility of magnetic resonance elastography using myofascial phantom model Poster No.: C-0971 Congress: ECR 2013 Type: Scientific Exhibit Authors: H. J. Kang, J.-S. Yoon, S.-J. Hong, C.-H. Oh, S. H.
More informationNormal Variations and Artifacts in MR Venography that may cause Pitfalls in the Diagnosis of Cerebral Venous Sinus Thrombosis.
Normal Variations and Artifacts in MR Venography that may cause Pitfalls in the Diagnosis of Cerebral Venous Sinus Thrombosis. Poster No.: R-0005 Congress: 2015 ASM Type: Scientific Exhibit Authors: A.
More informationCT-guided core needle biopsy of pulmonary lesions: a better tool for radiologist and clinician
CT-guided core needle biopsy of pulmonary lesions: a better tool for radiologist and clinician Poster No.: C-2476 Congress: ECR 2018 Type: Educational Exhibit Authors: A. Kosolapov, I. Blokhin, P. Gelezhe;
More informationSatisfaction and quality of life: a survey-based assessment in patients with a totally implantable venous port system
Satisfaction and quality of life: a survey-based assessment in patients with a totally implantable venous port system Poster No.: C-0417 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit
More informationTHI-RADS. US differentiation of thyroid lesions.
THI-RADS. US differentiation of thyroid lesions. Poster No.: C-0864 Congress: ECR 2015 Type: Scientific Exhibit Authors: A. N. Sencha, Y. Patrunov, M. S. Mogutov, E. Penyaeva, A. 1 1 1 2 1 1 1 2 Gruzdev,
More informationTHI-RADS. US differentiation of thyroid lesions.
THI-RADS. US differentiation of thyroid lesions. Poster No.: C-0864 Congress: ECR 2015 Type: Scientific Exhibit Authors: A. N. Sencha, Y. Patrunov, M. S. Mogutov, E. Penyaeva, A. 1 1 1 2 1 1 1 2 Gruzdev,
More informationEthanol ablation of benign thyroid cysts and predominantly cystic thyroid nodules: factors that predict outcome.
Ethanol ablation of benign thyroid cysts and predominantly cystic thyroid nodules: factors that predict outcome. Poster No.: C-0322 Congress: ECR 2014 Type: Authors: Keywords: DOI: Scientific Exhibit J.
More informationBreast ultrasound appearances after Mammotome vacuumassisted
Breast ultrasound appearances after Mammotome vacuumassisted biopsy. Poster No.: C-1924 Congress: ECR 2011 Type: Educational Exhibit Authors: R. Patel 1, G. R. Kaplan 2 ; 1 London/UK, 2 Herts/UK Keywords:
More informationCorrelation Between BIRADS Classification and Ultrasound -guided Tru-Cut Biopsy Results of Breast Lesions: Retrospective Analysis of 285 Patients
Correlation Between BIRADS Classification and Ultrasound -guided Tru-Cut Biopsy Results of Breast Lesions: Retrospective Analysis of 285 Patients Poster No.: C-1433 Congress: ECR 2014 Type: Scientific
More informationPopliteal pterygium syndrome
Popliteal pterygium syndrome Poster No.: C-1816 Congress: ECR 2011 Type: Educational Exhibit Authors: L. B. S. Santos, J. L. D. O. Schiavon, O. O. Guimaraes Neto, 1 1 2 3 1 1 C. A. P. Braga, R. S. LEMOS,
More information3D ultrasound applied to abdominal aortic aneurysm: preliminary evaluation of diameter measurement accuracy
3D ultrasound applied to abdominal aortic aneurysm: preliminary evaluation of diameter measurement accuracy Poster No.: C-0493 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Paper A. LONG
More informationInfluence of X-ray contrast media on blood viscoelasticity
Influence of X-ray contrast media on blood viscoelasticity Poster No.: B-630 Congress: ECR 2009 Type: Topic: Authors: Keywords: DOI: Scientific Paper Contrast Media V. Ribitsch, C. Well; Graz/AT Blood
More informationRadiation induced DNA double strand breaks undergoing PET/CT examinations.
Radiation induced DNA double strand breaks undergoing PET/CT examinations. Poster No.: C-1685 Congress: ECR 2011 Type: Scientific Exhibit Authors: M. S. May, M. Brand, T. Löwe, M. Sommer, C. Engert, U.
More informationDigital tomosynthesis in diagnosis of occult hip fractures
Digital tomosynthesis in diagnosis of occult hip fractures Poster No.: B-0781 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Paper M. Geijer 1, D. Collin 2, J. H. Göthlin 2 ; 1 Lund/SE, 2
More informationMetal Artifact Reduction by Dual Energy CT
Metal Artifact Reduction by Dual Energy CT Poster No.: C-0108 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Paper T. Johnson, F. Bamberg, A. Dierks, H.-C. Becker, M. F. Reiser; Munich/DE
More information