Application of international standards to diagnostic radiology dosimetry

Size: px
Start display at page:

Download "Application of international standards to diagnostic radiology dosimetry"

Transcription

1 Application of international standards to diagnostic radiology dosimetry Poster No.: C-780 Congress: ECR 2009 Type: Scientific Exhibit Topic: Physics in Radiology Authors: I. D. McLean, A. Meghzifene, F. Pernicka; Vienna/AT Keywords: Dosimetry, phantoms, Calibration, DAP DOI: /ecr2009/C-780 Any information contained in this pdf file is automatically generated from digital material submitted to EPOS by third parties in the form of scientific presentations. References to any names, marks, products, or services of third parties or hypertext links to thirdparty sites or information are provided solely as a convenience to you and do not in any way constitute or imply ECR's endorsement, sponsorship or recommendation of the third party, information, product or service. ECR is not responsible for the content of these pages and does not make any representations regarding the content or accuracy of material in this file. As per copyright regulations, any unauthorised use of the material or parts thereof as well as commercial reproduction or multiple distribution by any traditional or electronically based reproduction/publication method ist strictly prohibited. You agree to defend, indemnify, and hold ECR harmless from and against any and all claims, damages, costs, and expenses, including attorneys' fees, arising from or related to your use of these pages. Please note: Links to movies, ppt slideshows and any other multimedia files are not available in the pdf version of presentations. Page 1 of 14

2 Purpose Purpose: An international protocol on dosimetry in diagnostic radiology has recently been developed by the IAEA (TRS 457) for implementation in both Secondary Standard Dosimetry Laboratories (SSDL) and the clinical workplace. An analysis of the implementation of TRS 457 has been made under a coordinated research programme (CRP) with 11 participants (see Conclusion section). The work plan for the CRP is below (Table 1) and this communication will give some overview of activities 1, 4 and 5 as defined in the Methods section. Table 1. Activity list Activity 1 Setting-up calibration beam qualities at SSDLs Activity 2 Development of calibration procedures including the uncertainty budget at SSDLs Activity 3 Comparison of calibration of a selected instrument in selected beam qualities at SSDLs Activity 4 Evaluation of measurement procedures in hospitals, including: 1. Research the feasibility of implementing the procedures described in TRS 457 for making dosimetric measurements using phantoms and for patient data collection. 2. Report on the availability of dosimetric instrumentation and recommended phantoms, and the possibility of phantom fabrication if needed. 3. Create uncertainty budget for each type of dosimetric estimation, including the dose estimation from patient data. 4. Compare phantom measurements with patient dose data for each modality. Activity 5 Calibration of KAP meters at the SSDLs and at the clinical centres Activity 6 TLD dosimetry audit for SSDLs and clinical centres Page 2 of 14

3 Activity 7 The implementation of practical peak voltage (PPV) What does a secondary standard's laboratory do and look like? SSDLs have a role to calibrate radiation detectors for diagnostic radiology use that comply with International Electrotechnical Commission (IEC) standards. They are typically equipped with Xray systems with a stationary anode X-ray tube utilising a continuous tube current (e.g ma). In most cases, a shutter is incorporated allowing the timing control for an exposure. IEC-61674, IEC, Geneva (1997). Setup for diagnostic X ray calibration at IAEA laboratory Typical calibration X ray tubes and shutter system Page 3 of 14

4 Methods and Materials Methods and Materials: The coordinated research project, over the last 3 years, had two main directions. First (A), to investigate the suitability of dosimetry standards and protocols used to calibrate dosimetric instruments through SSDLs, including detectors for mammography, CT and kerma area product (KAP) meters. The second (B) was to investigate the clinical application of dosimetry protocols. For this communication the work of establishing reference beam qualities in SSDLs is examined, with special attention to areas of interest such as mammography and KAP meter calibration. The IEC has defined reference beam conditions for use in diagnostic radiology and the beams discussed in TRS 457 are summarised below Table: Radiation qualities for calibrations of diagnostic dosimeters used in TRS 457 Page 4 of 14

5 Radiation quality Radiation origin Material of additional filter an Application RQR Radiation beam no phantom emerging from X ray assembly General radiography, fluoroscopy and dental applications (measurements free in air) RQA Radiation beam with Aluminium an added filter Measurements behind the patient (on the image intensifier) RQT Radiation beam with Copper an added filter CT applications (measurements free in air) RQR-M Radiation beam no phantom emerging from X ray assembly Mammography applications (measurements free in air) RQA-M Radiation beam with Aluminium an added filter Mammography studies IEC-61267, IEC, Geneva (2005). Table: Calibration beams: RQR (IEC 61267) Beam quality Tube voltage HVL1 Homogeneity coefficient, h kv mm Al RQR RQR RQR RQR 5* RQR RQR RQR RQR RQR X ray setup with filter wheel used to create beam qualities. Note monitor chamber needed when comparing the standard and detector under calibration Page 5 of 14

6 Table: Calibration beams: RQR(M) (IEC 61267) Radiation quality X ray tube voltage Nominal first half-value layer Mo target kv mm Al RQR-M RQR-M 2* RQR-M RQR-M SSDL laboratories used TRS 457 to establish the above beam qualities for calibration of general diagnostic dosimeters and mammographic dosimeters. The calibration of KAP meters is also described in TRS 457 and the use of RQR beams for this was investigated. Clinical Applications Page 6 of 14

7 TRS 457 is a complementary publication to ICRU 74 and uses defined terminology and symbols. Quantity ICRU IAEA Incident air kerma Ka.i Ki Entrance-surface air kerma Ka,e Ke Air kerma-area product PKA PKA Air kerma-length product PKL PKL CT air-kerma index free in-air CK (integration -# to +#) Ca,100 (integration -50 to +50 mm) CT air-kerma index in the CK,PMMA (CK,PMMA,100) standard phantom CPMMA,100 These quantities are measured using protocols for two basic situations; (i) with the use of phantoms and (ii) to estimate dose to patients from radiographic and patient data combined with a characterization of relevant tube output. This is relatively straightforward for general radiography and also fluoroscopy where recommended phantoms can be utilised. In this communication the comparison of results for phantom and patient doses are reported. In the case of CT dose estimation TRS 457 utilises the IEC system of CT dosimetry which utilises on two circular PMMA phantoms of defined diameter. The practicality of the data collection methods for patients is reported on. Phantoms used for abdominal and chest dosimetry, also CT below: Page 7 of 14

8 Page 8 of 14

9 Results Results: Establishment of beam qualities: The required set of IEC beam qualities were successfully implemented in the SSDL laboratories in the project. While the calibration protocols developed for most radiographic applications, including plane X-ray, proved suitable and transferable to SSDLs it was noticed that the RQR (M) beams did not cover the full range of clinical use in mammography as shown below for a molybdenum target. Page 9 of 14

10 (courtesy C Hourdakis) In this graph it is noted that the compression paddle (not accounted for in IEC beam qualities creates clinical beams that easily exceed calibration conditions. In addition it is recognised that target materials other than W and Mo are also used and not covered by calibration standards. While well designed ionization chambers show little change in the energy correction factor kq, some solid state detectors may have considerable differences and so the definition of appropriate beam qualities is important clinically. More complexity was encountered with the calibration of KAP meters, where new beam qualities may be required to account for conditions appropriate to the high dose areas of interventional radiology (LIN 2007). Note that for thin objects (patients) that the X ray unit described above will use 60 kv and filtrations of 0.9 mm Cu. This has been confirmed by (Vano et al. 2008) Such beam qualities are not covered by RQR beams. As mentioned above this may not be of great concern for detectors which have little energy dependence such as ionisation chambers with the Page 10 of 14

11 noteable exception of KAP meters which have very thin metal (transparent) electrodes and display considerable energy dependence as seen below. This data comes from STUK and is adapted from Toroi et al (2008) The participants from STUK (above) have further shown that when the clinical beams as described by Lin et al are used that the behaviour of the energy dependant calibration factor cannot be characterised by the simply use of HVL, further emphasising the need for separate new beam calibration conditions. As is clearly seen this has significant implications for accurate dose estimation in interventional radiology. It is also clear that the use of only one energy correction term for all beam qualities will considerably increase the uncertainty of measurement used in patient dose estimation. Page 11 of 14

12 Other work in resulting from TRS457 confirms the need to calibrate field KAP meters in situ with the use of a reference instrument, rather than the independent calibration of KAP meters in standard laboratories. Clinical protocols and measurements Generally the clinical application was successful; however, more work is still required in implementation for effective use in the countries that contributed. Considering patient dose and phantom dose for measurements made on the same X ray units the figures below show that generally patient dose exceeds phantom dose. (courtesy P Homolka) Page 12 of 14

13 The average ratio of these results is also summarised in a single figure. Ratio of patient to phantom doses, dotted line corresponds to mean value, solid line in the box to median. The shaded area contains the 1st to 3rd quartiles. As seen there is a large spread in the ratio of patient to phantom dose. This is hardly surprising given the geographical spread of the data collected, including sites in Asia and Europe, where the average size of patient would be expected to vary significantly. While it might be felt that the collection of patient data is more reliable that the use of phantom data it should be considered that phantom dose values, which can be made rapidly, are valuable for dose intercomparison exercises. The area of CT dosimetry was found to be satisfactory for most contemporary needs; however, continuing developments in equipment design pose ongoing challenges, as does the diversity of acquisition parameters used by different scanners. Page 13 of 14

14 Conclusion Conclusion: Establishment of beam qualities: TRS 457 was found to give sufficient guidance for the successful establishment of IEC beam qualities to allow the calibration of appropriate detectors for use in diagnostic radiology. It was further found however that limitation exist in the IEC beam qualities as they do not cover a number of significant possible radiological setups that are used for mammography and interventional examinations. Clinical protocols and measurements Generally units and protocols useful for patient and phantom dose estimation are able to be applied to the clinical environment. Further it is clear that both patient and phantom dose are important in diagnostic dosimetry. Patient dose, while time consuming and demanding in terms of data collection, should result in accurate estimations. Phantom dose estimations on the other hand can be made rapidly and are useful in intercomparisons. Recommendations from areas covered in this communication. 1. Since existing IEC beam qualities do not fully cover important areas of diagnostic radiology, such as significant conditions for mammography and interventional radiology, particularly for small patients, new appropriate beam conditions should be agreed upon for definition by the IEC and then established at primary standards laboratories. 2. The determination of patient dose as described in TRS457 has significant clinical impact when the results are compared to appropriate dose reference levels. Together with relevant image quality data, this will inform the process of optimization. 3. CT Scanner type specific data sheets should be used to assist in data collection 4. Further work is needed to extend the CT procedures to accommodate the new developments in CT technology. 5. Field KAP meters should be calibrated (or cross calibrated) against a reference KAP meter or air kerma meter in situ in the clinical environment. Page 14 of 14

Direct half value layer measurements in mammography - is near enough good enough?

Direct half value layer measurements in mammography - is near enough good enough? Direct half value layer measurements in mammography - is near enough good enough? Poster No.: R-0127 Congress: Type: Authors: 2014 CSM Scientific Exhibit J. Diffey 1, L. Cartwright 2, J. Crocker 1, J.

More information

Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography

Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography Using Monte Carlo Method for Evaluation of kvp & mas variation effect on Absorbed Dose in Mammography Poster No.: C-2078 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Exhibit F. Salmani Rezaei,

More information

Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA

Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA Examination of dose evaluation method for mammography: comparison of ACR, Euref and IAEA Poster No.: C-0618 Congress: ECR 2011 Type: Authors: Scientific Paper T. Otsuka 1, S. Suzuki 1, M. Ogura 1, T. Igarashi

More information

Multislice CT versus cone beam CT: dosimetric and image quality comparision.

Multislice CT versus cone beam CT: dosimetric and image quality comparision. Multislice CT versus cone beam CT: dosimetric and image quality comparision. Poster No.: C-2589 Congress: ECR 2015 Type: Authors: Scientific Exhibit C. Dionisi 1, A. Crispo 2, E. Stefani 3, A. Lovaglio

More information

64-MDCT imaging of the pancreas: Scan protocol optimisation by different scan delay regimes

64-MDCT imaging of the pancreas: Scan protocol optimisation by different scan delay regimes 64-MDCT imaging of the pancreas: Scan protocol optimisation by different scan delay regimes Poster No.: C-051 Congress: ECR 2009 Type: Scientific Exhibit Topic: Abdominal and Gastrointestinal Authors:

More information

Standardizing mammographic breast compression: Pressure rather than force?

Standardizing mammographic breast compression: Pressure rather than force? Standardizing mammographic breast compression: Pressure rather than force? Poster No.: C-2383 Congress: ECR 2012 Type: Authors: Scientific Paper J. E. de Groot 1, W. Branderhorst 1, M. Broeders 2, G. J.

More information

Lifetime risk of radiation-induced cancer from screening mammography

Lifetime risk of radiation-induced cancer from screening mammography Lifetime risk of radiation-induced cancer from screening mammography Poster No.: C-0558 Congress: ECR 2015 Type: Scientific Exhibit Authors: R. M. K. M.Ali, A. England, P. Hogg; Manchester/UK Keywords:

More information

Evaluation and Optimization of Exposure Index in Lumbar Spine and Pelvis X-Ray

Evaluation and Optimization of Exposure Index in Lumbar Spine and Pelvis X-Ray Evaluation and Optimization of Exposure Index in Lumbar Spine and Pelvis X-Ray Poster No.: C-1492 Congress: ECR 2012 Type: Scientific Exhibit Authors: R. M. Lopes, L. H. Domingos ; Mortágua/PT, Azoia de

More information

AFib is the most common cardiac arrhythmia and its prevalence and incidence increases with age (Fuster V. et al. Circulation 2006).

AFib is the most common cardiac arrhythmia and its prevalence and incidence increases with age (Fuster V. et al. Circulation 2006). Feasibility, image quality and radiation dose of coronary CT angiography (CCTA) in patients with atrial fibrillation using a new generation 256 multi-detector CT (MDCT) Poster No.: C-2378 Congress: ECR

More information

Artifact in Head CT Images Due to Air Bubbles in X-Ray Tube Oil

Artifact in Head CT Images Due to Air Bubbles in X-Ray Tube Oil Artifact in Head CT Images Due to Air Bubbles in X-Ray Tube Oil Poster No.: C-0671 Congress: ECR 2016 Type: Educational Exhibit Authors: H. Patel 1, W. Liu 2, J. DeSanto 2, S. Meagher 2, M. Zagardo 2,

More information

User Guide for Dental and Maxillofacial Cone Beam Computed Tomography (CBCT)

User Guide for Dental and Maxillofacial Cone Beam Computed Tomography (CBCT) User Guide for Dental and Maxillofacial Cone Beam Computed Tomography (CBCT) Poster No.: C-0756 Congress: ECR 2014 Type: Educational Exhibit Authors: J. Ukkonen, J. Asp; Helsinki/FI Keywords: Education

More information

Radiological features of Legionella Pneumophila Pneumonia

Radiological features of Legionella Pneumophila Pneumonia Radiological features of Legionella Pneumophila Pneumonia Poster No.: E-0048 Congress: ESTI 2012 Type: Scientific Exhibit Authors: M. Vinciguerra, L. Stefanetti, E. Teti, G. Argentieri, L. G. 1 1 1 1 1

More information

Scientific Exhibit Authors:

Scientific Exhibit Authors: Clinical Audit on Optimization of Radiation Dose from MDCT: Effect on Diagnostic Reference Levels for Brain, Sinus, Cervical Spine, Chest, Abdomen-Pelvis, and Lumbar Spine Examinations and on Nationwide

More information

Slowly growing malignant nodules and rapidly growing benign nodules: Evaluation of the value of volume doubling time

Slowly growing malignant nodules and rapidly growing benign nodules: Evaluation of the value of volume doubling time Slowly growing malignant nodules and rapidly growing benign nodules: Evaluation of the value of volume doubling time Poster No.: C-208 Congress: ECR 2009 Type: Educational Exhibit Topic: Chest Authors:

More information

Dose reduction in Hologic Selenia FFDM units through AEC optimization, without compromising diagnostic image quality.

Dose reduction in Hologic Selenia FFDM units through AEC optimization, without compromising diagnostic image quality. Dose reduction in Hologic Selenia FFDM units through AEC optimization, without compromising diagnostic image quality. Poster No.: C-1636 Congress: ECR 2015 Type: Authors: Keywords: DOI: Scientific Exhibit

More information

Influence of pulsed fluoroscopy and special radiation risk training on the radiation dose in pneumatic reduction of ileocoecal intussusceptions.

Influence of pulsed fluoroscopy and special radiation risk training on the radiation dose in pneumatic reduction of ileocoecal intussusceptions. Influence of pulsed fluoroscopy and special radiation risk training on the radiation dose in pneumatic reduction of ileocoecal intussusceptions. Poster No.: C-0599 Congress: ECR 2013 Type: Authors: Keywords:

More information

FDG-18 PET/CT - radiation dose and dose-reduction strategy

FDG-18 PET/CT - radiation dose and dose-reduction strategy FDG-18 PET/CT - radiation dose and dose-reduction strategy Poster No.: C-1856 Congress: ECR 2014 Type: Authors: Keywords: DOI: Scientific Exhibit P. Nicholson, S. McSweeney, K. O'Regan; Cork/IE Radiation

More information

Computed tomographic dacryocystography as compared with X-ray dacryocystography in patients with dacryostenosis

Computed tomographic dacryocystography as compared with X-ray dacryocystography in patients with dacryostenosis Computed tomographic dacryocystography as compared with X-ray dacryocystography in patients with dacryostenosis Poster No.: C-1887 Congress: ECR 2016 Type: Authors: Keywords: DOI: Educational Exhibit M.

More information

Quantifying Dual-energy computed tomography (DECT) in patients with renal calculi using a Toshiba Aquilion One Scanner.

Quantifying Dual-energy computed tomography (DECT) in patients with renal calculi using a Toshiba Aquilion One Scanner. Quantifying Dual-energy computed tomography (DECT) in patients with renal calculi using a Toshiba Aquilion One Scanner. Poster No.: C-2613 Congress: ECR 2015 Type: Authors: Keywords: DOI: Scientific Exhibit

More information

Purpose. Methods and Materials. Results

Purpose. Methods and Materials. Results Prevalence and significance of hypoattenuating hepatic lesions deemed too small to characterise: How are we following up these lesions and what are the outcomes? Poster No.: C-014 Congress: ECR 2009 Type:

More information

Digital chest radiography: collimation and dose reduction

Digital chest radiography: collimation and dose reduction Digital chest radiography: collimation and dose reduction Poster No.: C-1939 Congress: ECR 2015 Type: Scientific Exhibit Authors: J. Debess, K. Johnsen, K. Vejle Sørensen, H. Thomsen ; 1 1 2 1 1 2 Aalborg

More information

Scientific Exhibit Authors: V. Moustakas, E. Karallas, K. Koutsopoulos ; Rodos/GR, 2

Scientific Exhibit Authors: V. Moustakas, E. Karallas, K. Koutsopoulos ; Rodos/GR, 2 Diagnosis of Acute Appendicitis: the role of Color Doppler Ultrasound as first-line imaging method and evaluation of the higher diagnostic performances of CT against its disadvantages. Poster No.: C-0708

More information

Analysis of patients and operators radiation doses in transcatheter aortic valve implantation (TAVI) and evaluation of radiation risks

Analysis of patients and operators radiation doses in transcatheter aortic valve implantation (TAVI) and evaluation of radiation risks Analysis of patients and operators radiation doses in transcatheter aortic valve implantation (TAVI) and evaluation of radiation risks Poster No.: C-2993 Congress: ECR 2010 Type: Topic: Scientific Exhibit

More information

Computed tomography for pulmonary embolism: scan assessment of a one-year cohort and estimated cancer risk associated with diagnostic irradiation.

Computed tomography for pulmonary embolism: scan assessment of a one-year cohort and estimated cancer risk associated with diagnostic irradiation. Computed tomography for pulmonary embolism: scan assessment of a one-year cohort and estimated cancer risk associated with diagnostic irradiation. Award: Poster No.: E-0004 Certificate of Merit Congress:

More information

Cavitary lung lesion: Two different diagnosis with similar appearence

Cavitary lung lesion: Two different diagnosis with similar appearence Cavitary lung lesion: Two different diagnosis with similar appearence Poster No.: P-0043 Congress: ESTI 2015 Type: Educational Poster Authors: M. Yesildag, H. Kalkan, K. Ödev; Konya/TR Keywords: Infection,

More information

Computed tomography and Modified RECIST criteria for assessment of response in malignant pleural mesothelioma

Computed tomography and Modified RECIST criteria for assessment of response in malignant pleural mesothelioma Computed tomography and Modified RECIST criteria for assessment of response in malignant pleural mesothelioma Poster No.: C-0729 Congress: ECR 2013 Type: Scientific Exhibit Authors: A. Marin, I. Pozek,

More information

Bolus administration of esmolol allows for safe and effective heart rate control during coronary computed tomography angiography

Bolus administration of esmolol allows for safe and effective heart rate control during coronary computed tomography angiography Bolus administration of esmolol allows for safe and effective heart rate control during coronary computed tomography angiography Poster No.: C-1342 Congress: ECR 2013 Type: Scientific Exhibit Authors:

More information

BI-RADS 3, 4 and 5 lesions on US: Five categories and their diagnostic efficacy and pitfalls in interpretation

BI-RADS 3, 4 and 5 lesions on US: Five categories and their diagnostic efficacy and pitfalls in interpretation BI-RADS 3, 4 and 5 lesions on US: Five categories and their diagnostic efficacy and pitfalls in interpretation e-poster: C-118 Congress: ECR 2008 Type: Educational Exhibit Topic: Breast / Ultrasound Authors:

More information

Single cold nodule in Graves' disease: benign vs malignant

Single cold nodule in Graves' disease: benign vs malignant Single cold nodule in Graves' disease: benign vs malignant Poster No.: C-0073 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Paper L. I. Sonoda, M. Halim, K. Balan; Cambridge/UK Head and neck,

More information

Cognitive target MRI-TRUS fusion biopsies of MRI detected PIRADS 4 and 5 lesions

Cognitive target MRI-TRUS fusion biopsies of MRI detected PIRADS 4 and 5 lesions Cognitive target MRI-TRUS fusion biopsies of MRI detected PIRADS 4 and 5 lesions Poster No.: B-0704 Congress: ECR 2015 Type: Scientific Paper Authors: P. P. van Westerveld, J. Vriesema, J. H. W. van den

More information

Metal Artifact Reduction by Dual Energy CT

Metal Artifact Reduction by Dual Energy CT Metal Artifact Reduction by Dual Energy CT Poster No.: C-0108 Congress: ECR 2011 Type: Authors: Keywords: DOI: Scientific Paper T. Johnson, F. Bamberg, A. Dierks, H.-C. Becker, M. F. Reiser; Munich/DE

More information

Identification and numbering of lumbar vertebrae using various anatomical landmarks on MRI of lumbosacral spine

Identification and numbering of lumbar vertebrae using various anatomical landmarks on MRI of lumbosacral spine Identification and numbering of lumbar vertebrae using various anatomical landmarks on MRI of lumbosacral spine Poster No.: C-2125 Congress: ECR 2015 Type: Authors: Scientific Exhibit S. patil 1, A. M.

More information

Contrast agents, Abdomen, CT, Contrast agent-intravenous, Cancer /ecr2015/C-1760

Contrast agents, Abdomen, CT, Contrast agent-intravenous, Cancer /ecr2015/C-1760 Can Weight-Adapted IV Contrast Media Protocols Reduce Iodine Dose and Still Produce a Diagnostic Contrast Enhancement Level in Abdomino-Pelvic CT Scans? Poster No.: C-1760 Congress: ECR 2015 Type: Authors:

More information

Seemingly isolated greater trochanter fractures do not exist

Seemingly isolated greater trochanter fractures do not exist Seemingly isolated greater trochanter fractures do not exist Poster No.: B-0950 Congress: ECR 2012 Type: Scientific Paper Authors: D. Dunker, J. H. Göthlin, M. Geijer ; Gothenburg/SE, Lund/SE Keywords:

More information

Radiation induced DNA double strand breaks undergoing PET/CT examinations.

Radiation induced DNA double strand breaks undergoing PET/CT examinations. Radiation induced DNA double strand breaks undergoing PET/CT examinations. Poster No.: C-1685 Congress: ECR 2011 Type: Scientific Exhibit Authors: M. S. May, M. Brand, T. Löwe, M. Sommer, C. Engert, U.

More information

Hyperechoic breast lesions can be malignant.

Hyperechoic breast lesions can be malignant. Hyperechoic breast lesions can be malignant. Poster No.: C-0041 Congress: ECR 2015 Type: Educational Exhibit Authors: G. Babu, R. bradley; Edinburgh/UK Keywords: Breast, Ultrasound, Biopsy, Cancer DOI:

More information

Comparison of radiation doses of various approaches of MR arthrograms with fluoroscopic guided contrast injection

Comparison of radiation doses of various approaches of MR arthrograms with fluoroscopic guided contrast injection Comparison of radiation doses of various approaches of MR arthrograms with fluoroscopic guided contrast injection Poster No.: C-1294 Congress: ECR 2014 Type: Scientific Exhibit Authors: H. M. Cheng, C.

More information

3D cine PCA enables rapid and comprehensive hemodynamic assessment of the abdominal aorta

3D cine PCA enables rapid and comprehensive hemodynamic assessment of the abdominal aorta 3D cine PCA enables rapid and comprehensive hemodynamic assessment of the abdominal aorta Poster No.: C-1138 Congress: ECR 2013 Type: Scientific Exhibit Authors: M. Takahashi 1, Y. Takehara 2, H. Isoda

More information

Computed tomography for the detection of thumb base osteoarthritis, comparison with digital radiography.

Computed tomography for the detection of thumb base osteoarthritis, comparison with digital radiography. Computed tomography for the detection of thumb base osteoarthritis, comparison with digital radiography. Poster No.: C-1981 Congress: ECR 2012 Type: Scientific Exhibit Authors: M. S. Saltzherr, J. W. van

More information

CT Fluoroscopy-Guided vs Multislice CT Biopsy ModeGuided Lung Biopies:a preliminary experience

CT Fluoroscopy-Guided vs Multislice CT Biopsy ModeGuided Lung Biopies:a preliminary experience CT Fluoroscopy-Guided vs Multislice CT Biopsy ModeGuided Lung Biopies:a preliminary experience Poster No.: C-0097 Congress: ECR 2016 Type: Scientific Exhibit Authors: A. Casarin, G. Rech, C. Cicero, A.

More information

The Role of Radionuclide Lymphoscintigraphy in the Diagnosis of Lymphedema of the Extremities

The Role of Radionuclide Lymphoscintigraphy in the Diagnosis of Lymphedema of the Extremities The Role of Radionuclide Lymphoscintigraphy in the Diagnosis of Lymphedema of the Extremities Poster No.: C-1229 Congress: ECR 2014 Type: Authors: Keywords: DOI: Scientific Exhibit M. Osher 1, A. Pallas

More information

Audit of CT Pulmonary Angiogram in suspected pulmonary embolism patients

Audit of CT Pulmonary Angiogram in suspected pulmonary embolism patients Audit of CT Pulmonary Angiogram in suspected pulmonary embolism patients Poster No.: C-2511 Congress: ECR 2012 Type: Scientific Exhibit Authors: N. D. Gupta, M. K. Heir, P. Bradding; Leicester/UK Keywords:

More information

Shear Wave Elastography in diagnostics of supraspinatus tendon.

Shear Wave Elastography in diagnostics of supraspinatus tendon. Shear Wave Elastography in diagnostics of supraspinatus tendon. Poster No.: C-2168 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit V. Saltykova; Moscow/RU Musculoskeletal joint, Musculoskeletal

More information

PGMI classification of screening mammograms prior to interval cancer. Comparison with radiologists' consensus classification.

PGMI classification of screening mammograms prior to interval cancer. Comparison with radiologists' consensus classification. PGMI classification of screening mammograms prior to interval cancer. Comparison with radiologists' consensus classification. Poster No.: C-0673 Congress: ECR 2016 Type: Authors: Keywords: DOI: Scientific

More information

Feasibility of magnetic resonance elastography using myofascial phantom model

Feasibility of magnetic resonance elastography using myofascial phantom model Feasibility of magnetic resonance elastography using myofascial phantom model Poster No.: C-0971 Congress: ECR 2013 Type: Scientific Exhibit Authors: H. J. Kang, J.-S. Yoon, S.-J. Hong, C.-H. Oh, S. H.

More information

How to plan a Zenith AAA stent-graft from a CTA: Basic measurements and concepts explained

How to plan a Zenith AAA stent-graft from a CTA: Basic measurements and concepts explained How to plan a Zenith AAA stent-graft from a CTA: Basic measurements and concepts explained Poster No.: C-3077 Congress: ECR 2010 Type: Educational Exhibit Topic: Vascular Authors: D. V. Thomas; Northampton/UK

More information

Quantitative imaging of hepatic cirrhosis on abdominal CT images

Quantitative imaging of hepatic cirrhosis on abdominal CT images Quantitative imaging of hepatic cirrhosis on abdominal CT images Poster No.: C-0556 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit S. Kido, A. Nakamura, Y. Hirano; Ube/JP Cirrhosis,

More information

Low-dose computed tomography (CT) protocol in the screening of patients with social exposure to asbestos

Low-dose computed tomography (CT) protocol in the screening of patients with social exposure to asbestos Low-dose computed tomography (CT) protocol in the screening of patients with social exposure to asbestos Poster No.: C-3032 Congress: ECR 2010 Type: Scientific Exhibit Topic: Radiographers Authors: P.

More information

Diffuse high-attenuation within mediastinal lymph nodes on non-enhanced CT scan: Usefulness in the prediction of benignancy

Diffuse high-attenuation within mediastinal lymph nodes on non-enhanced CT scan: Usefulness in the prediction of benignancy Diffuse high-attenuation within mediastinal lymph nodes on non-enhanced CT scan: Usefulness in the prediction of benignancy Poster No.: C-1785 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific

More information

Scientific Exhibit. Authors: D. Takenaka, Y. Ohno, Y. Onishi, K. Matsumoto, T.

Scientific Exhibit. Authors: D. Takenaka, Y. Ohno, Y. Onishi, K. Matsumoto, T. The feasibility of biphasic contrast-media-injection-protocol for chest imaging on 320-slice volume MDCT: Direct comparison of biphasic and bolus contrast-media injection protocols on 320-slice volume

More information

The effect of CT dose reduction on performance of a diagnostic task

The effect of CT dose reduction on performance of a diagnostic task The effect of CT dose reduction on performance of a diagnostic task Poster No.: B-0856 Congress: ECR 2012 Type: Scientific Paper Authors: D. Zhang, M. Khatonabadi, H. Kim, C. Jude, E. Zaragoza, S. McNitt-Gray,

More information

The effect of CT dose reduction on performance of a diagnostic task

The effect of CT dose reduction on performance of a diagnostic task The effect of CT dose reduction on performance of a diagnostic task Poster No.: B-0856 Congress: ECR 2012 Type: Scientific Paper Authors: D. Zhang, M. Khatonabadi, H. Kim, C. Jude, E. Zaragoza, S. McNitt-Gray,

More information

Reliability of the pronator quadratus fat pad sign to predict the severity of distal radius fractures

Reliability of the pronator quadratus fat pad sign to predict the severity of distal radius fractures Reliability of the pronator quadratus fat pad sign to predict the severity of distal radius fractures Poster No.: C-0669 Congress: ECR 2014 Type: Scientific Exhibit Authors: J. Tonak, I. Wobbe, R. L. Duschka,

More information

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution

Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Testing of the Implementation of the Code of Practice on Dosimetry in X-ray Diagnostic Radiology Hungarian Contribution Ferenc Giczi a*, Sándor Pellet b, Ian Donald McLean c and Ahmed Meghzifene c a Széchenyi

More information

Is the lateral neck radiograph still a viable investigation for an impacted fish bone in the aero-digestive tract: going back to basics.

Is the lateral neck radiograph still a viable investigation for an impacted fish bone in the aero-digestive tract: going back to basics. Is the lateral neck radiograph still a viable investigation for an impacted fish bone in the aero-digestive tract: going back to basics. Poster No.: C-0353 Congress: ECR 2011 Type: Scientific Exhibit Authors:

More information

Audit of CT Pulmonary Angiogram in suspected pulmonary embolism patients

Audit of CT Pulmonary Angiogram in suspected pulmonary embolism patients Audit of CT Pulmonary Angiogram in suspected pulmonary embolism patients Poster No.: C-2511 Congress: ECR 2012 Type: Scientific Exhibit Authors: N. D. Gupta, M. K. Heir, P. Bradding; Leicester/UK Keywords:

More information

Breast ultrasound appearances after Mammotome vacuumassisted

Breast ultrasound appearances after Mammotome vacuumassisted Breast ultrasound appearances after Mammotome vacuumassisted biopsy. Poster No.: C-1924 Congress: ECR 2011 Type: Educational Exhibit Authors: R. Patel 1, G. R. Kaplan 2 ; 1 London/UK, 2 Herts/UK Keywords:

More information

Medical device adverse incident reporting in interventional radiology

Medical device adverse incident reporting in interventional radiology Medical device adverse incident reporting in interventional radiology Poster No.: C-0226 Congress: ECR 2014 Type: Authors: Scientific Exhibit N. PARVIZI 1, I. Robertson 2, R. G. McWilliams 3 ; 1 Oxford/UK,

More information

Correlation Between BIRADS Classification and Ultrasound -guided Tru-Cut Biopsy Results of Breast Lesions: Retrospective Analysis of 285 Patients

Correlation Between BIRADS Classification and Ultrasound -guided Tru-Cut Biopsy Results of Breast Lesions: Retrospective Analysis of 285 Patients Correlation Between BIRADS Classification and Ultrasound -guided Tru-Cut Biopsy Results of Breast Lesions: Retrospective Analysis of 285 Patients Poster No.: C-1433 Congress: ECR 2014 Type: Scientific

More information

Radiological Investigation of Renal Colic in an Emergency Department of a Teaching Hospital

Radiological Investigation of Renal Colic in an Emergency Department of a Teaching Hospital Radiological Investigation of Renal Colic in an Emergency Department of a Teaching Hospital Poster No.: C-0892 Congress: ECR 2014 Type: Authors: Keywords: DOI: Scientific Exhibit A. Koo; Leeds, West Yorkshire/UK

More information

Diffusion-weighted MRI (DWI) "claw sign" is useful in differentiation of infectious from degenerative Modic I signal changes of the spine

Diffusion-weighted MRI (DWI) claw sign is useful in differentiation of infectious from degenerative Modic I signal changes of the spine Diffusion-weighted MRI (DWI) "claw sign" is useful in differentiation of infectious from degenerative Modic I signal changes of the spine Poster No.: C-0894 Congress: ECR 2012 Type: Scientific Exhibit

More information

Emerging Referral Patterns for Whole-Body Diffusion Weighted Imaging (WB-DWI) in an Oncology Center

Emerging Referral Patterns for Whole-Body Diffusion Weighted Imaging (WB-DWI) in an Oncology Center Emerging Referral Patterns for Whole-Body Diffusion Weighted Imaging (WB-DWI) in an Oncology Center Poster No.: C-1296 Congress: ECR 2014 Type: Scientific Exhibit Authors: G. Petralia 1, G. Conte 1, S.

More information

Radiation dose levels assessment in mammography

Radiation dose levels assessment in mammography Radiation dose levels assessment in mammography Poster No.: B-0871 Congress: ECR 2013 Type: Scientific Paper Authors: C. N. Lourenço, P. Sousa, S. Rodrigues, A. F. Abrantes, L. P. V. Ribeiro, K. B. Azevedo,

More information

Monophasic versus biphasic contrast application in CT of patients with head and neck tumour

Monophasic versus biphasic contrast application in CT of patients with head and neck tumour Monophasic versus biphasic contrast application in CT of patients with head and neck tumour Poster No.: C-3331 Congress: ECR 2010 Type: Topic: Authors: Keywords: DOI: Scientific Exhibit Head and Neck G.

More information

Cumulative radiation CT dose in young traumatic patients: a single centre 5 years retrospective review

Cumulative radiation CT dose in young traumatic patients: a single centre 5 years retrospective review Cumulative radiation CT dose in young traumatic patients: a single centre 5 years retrospective review Poster No.: C-1604 Congress: ECR 2014 Type: Scientific Exhibit Authors: S. Salerno, C. Geraci, E.

More information

Extravasation of Contrast Medium during CT Scanning Tracking and Reduction of Rate of Extravasation

Extravasation of Contrast Medium during CT Scanning Tracking and Reduction of Rate of Extravasation Extravasation of Contrast Medium during CT Scanning Tracking and Reduction of Rate of Extravasation Poster No.: C-0072 Congress: ECR 2015 Type: Scientific Exhibit Authors: M. S. K. Teo, C. Ong, A. Ying,

More information

High density thrombi of pulmonary embolism on precontrast CT scan: Is it dangerous?

High density thrombi of pulmonary embolism on precontrast CT scan: Is it dangerous? High density thrombi of pulmonary embolism on precontrast CT scan: Is it dangerous? Poster No.: C-1753 Congress: ECR 2013 Type: Authors: Keywords: DOI: Scientific Exhibit B. Y. Lee, H. R. KIM, J. I. Jung,

More information

Using diffusion-tensor imaging and tractography (DTT) to study biological characteristics of glyoma in brain stem for neurosurgical planning

Using diffusion-tensor imaging and tractography (DTT) to study biological characteristics of glyoma in brain stem for neurosurgical planning Using diffusion-tensor imaging and tractography (DTT) to study biological characteristics of glyoma in brain stem for neurosurgical planning Poster No.: C-0016 Congress: ECR 2014 Type: Authors: Keywords:

More information

DEXA Scores and Bone Density Measured on Routine CT Scans

DEXA Scores and Bone Density Measured on Routine CT Scans DEXA Scores and Bone Density Measured on Routine CT Scans Poster No.: R-0072 Congress: 2015 ASM Type: Scientific Exhibit Authors: C. Annabattula, P. Phadke; Sydney/AU Keywords: Screening, Nuclear medicine

More information

A pictorial review of normal anatomical appearences of Pericardial recesses on multislice Computed Tomography.

A pictorial review of normal anatomical appearences of Pericardial recesses on multislice Computed Tomography. A pictorial review of normal anatomical appearences of Pericardial recesses on multislice Computed Tomography. Poster No.: C-1787 Congress: ECR 2012 Type: Educational Exhibit Authors: N. Ahmed 1, G. Avery

More information

Cierny-Mader classification of chronic osteomyelitis: Preoperative evaluation with cross-sectional imaging

Cierny-Mader classification of chronic osteomyelitis: Preoperative evaluation with cross-sectional imaging Cierny-Mader classification of chronic osteomyelitis: Preoperative evaluation with cross-sectional imaging Poster No.: C-590 Congress: ECR 2009 Type: Topic: Educational Exhibit Musculoskeletal Authors:

More information

A dosimetric analysis of the Aeroform(tm) tissue expander in radiation therapy

A dosimetric analysis of the Aeroform(tm) tissue expander in radiation therapy A dosimetric analysis of the Aeroform(tm) tissue expander in radiation therapy Poster No.: R-0008 Congress: 2014 CSM Type: Scientific Exhibit Authors: T. Tran, W. Ding, L. Melvern, M. Chao, B. Subramanian;

More information

Contrast enhanced spectral mammography: A literature review

Contrast enhanced spectral mammography: A literature review Contrast enhanced spectral mammography: A literature review Poster No.: R-0172 Congress: RANZCR-AOCR 2012 Type: Authors: Keywords: DOI: Educational Exhibit S. Buzynski, D. Taylor; Perth/AU Breast, Digital

More information

"Ultrasound measurements of the lateral ventricles in neonates: A comparison of multiple measurements methods."

Ultrasound measurements of the lateral ventricles in neonates: A comparison of multiple measurements methods. "Ultrasound measurements of the lateral ventricles in neonates: A comparison of multiple measurements methods." Poster No.: C-1557 Congress: ECR 2014 Type: Authors: Keywords: DOI: Educational Exhibit I.

More information

ARDS - a must know. Page 1 of 14

ARDS - a must know. Page 1 of 14 ARDS - a must know Poster No.: C-1683 Congress: ECR 2016 Type: Authors: Keywords: DOI: Educational Exhibit M. Cristian; Turda/RO Education and training, Edema, Acute, Localisation, Education, Digital radiography,

More information

Abdominal fat distribution (subcutaneous vs. visceral abdominal fat compartments): correlation with gender, age, BMI and waist circumference

Abdominal fat distribution (subcutaneous vs. visceral abdominal fat compartments): correlation with gender, age, BMI and waist circumference Abdominal fat distribution (subcutaneous vs. visceral abdominal fat compartments): correlation with gender, age, BMI and waist circumference Poster No.: C-934 Congress: ECR 202 Type: Scientific Exhibit

More information

Delayed Vertebral Augmentation With Spinejack Technique in A3 Type Vertebral Compression Fractures

Delayed Vertebral Augmentation With Spinejack Technique in A3 Type Vertebral Compression Fractures Delayed Vertebral Augmentation With Spinejack Technique in A3 Type Vertebral Compression Fractures Poster No.: C-0358 Congress: ECR 2016 Type: Authors: Keywords: DOI: Scientific Exhibit J. Chiras; Paris

More information

Uozu city/jp, Minatoku, Tokyo/JP Bones, Extremities, CT, Surgery, Physics, Artifacts, Image verification /ecr2014/C-0462

Uozu city/jp, Minatoku, Tokyo/JP Bones, Extremities, CT, Surgery, Physics, Artifacts, Image verification /ecr2014/C-0462 Metal Artifact Reduction Algorithm enablesreduce metal artifacts and improvement of diagnosis in the postoperative Pedicle screws implant for Spinal Fusion: A Phantom Study Poster No.: C-0462 Congress:

More information

Evaluation of BI-RADS 3 lesions in women with a high risk of hereditary breast cancer.

Evaluation of BI-RADS 3 lesions in women with a high risk of hereditary breast cancer. Evaluation of BI-RADS 3 lesions in women with a high risk of hereditary breast cancer. Poster No.: C-0346 Congress: ECR 2014 Type: Scientific Exhibit Authors: A. Thomas 1, R. Dominguez Oronoz 1, S. Roche

More information

The Abdominal plain film: A justified 21st century imaging investigation?

The Abdominal plain film: A justified 21st century imaging investigation? The Abdominal plain film: A justified 21st century imaging investigation? Poster No.: C-0877 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Exhibit Z. J. Hussain 1, H. F. D'Costa 2 ; 1 Oxford/UK,

More information

The "whirl sign". Diagnostic accuracy for intestinal volvulus.

The whirl sign. Diagnostic accuracy for intestinal volvulus. The "whirl sign". Diagnostic accuracy for intestinal volvulus. Poster No.: C-0670 Congress: ECR 2014 Type: Scientific Exhibit Authors: M. Pire, M. Marti, A. Borobia, A. Verón; Madrid/ES Keywords: Abdomen,

More information

Idiopathic dilatation of the pulmonary artery : radiographic and MDCT features in 6 cases

Idiopathic dilatation of the pulmonary artery : radiographic and MDCT features in 6 cases Idiopathic dilatation of the pulmonary artery : radiographic and MDCT features in 6 cases Poster No.: P-0075 Congress: ESTI 2014 Type: Authors: Educational Poster J. J. Woo 1, K. Y. Lee 2, Y. Cho 1, J.

More information

Sonographic and Mammographic Features of Phyllodes Tumours of the Breast: Correlation with Histological Grade

Sonographic and Mammographic Features of Phyllodes Tumours of the Breast: Correlation with Histological Grade Sonographic and Mammographic Features of Phyllodes Tumours of the Breast: Correlation with Histological Grade Poster No.: C-0046 Congress: ECR 2014 Type: Authors: Keywords: DOI: Scientific Exhibit C. Y.

More information

Ultrasonic evaluation of superior mesenteric vein in cancer of the pancreatic head

Ultrasonic evaluation of superior mesenteric vein in cancer of the pancreatic head Ultrasonic evaluation of superior mesenteric vein in cancer of the pancreatic head Poster No.: C-1430 Congress: ECR 2012 Type: Authors: Keywords: DOI: Scientific Exhibit E. Fisenko, N. Vetsheva, E. Pershina;

More information

Long bones manifestations of congenital syphilis

Long bones manifestations of congenital syphilis Long bones manifestations of congenital syphilis Poster No.: C-0139 Congress: ECR 2011 Type: Educational Exhibit Authors: T. F. de Souza 1, P. P. Collier 1, E. J. M. Bronzatto 1, G. L. P. Keywords: DOI:

More information

How to obtain the waist circumference for retrospective studies - a prospective validation of CT images for the evaluation of the abdominal perimeter

How to obtain the waist circumference for retrospective studies - a prospective validation of CT images for the evaluation of the abdominal perimeter How to obtain the waist circumference for retrospective studies - a prospective validation of CT images for the evaluation of the abdominal perimeter Poster No.: C-0281 Congress: ECR 2013 Type: Scientific

More information

The Virtual Lung Nodule Clinic

The Virtual Lung Nodule Clinic The Virtual Lung Nodule Clinic Poster No.: C-1023 Congress: ECR 2016 Type: Educational Exhibit Authors: S. Higgins, F. C. Lyall, J. Taylor, J. goldman, S. Rolin, B. 1 2 1 2 2 3 2 2 3 Soar ; Torbay/UK,

More information

The impact of SID and collimation on backscatter in radiography

The impact of SID and collimation on backscatter in radiography The impact of SID and collimation on backscatter in radiography Poster No.: C-2931 Congress: ECR 2010 Type: Scientific Exhibit Topic: Physics in Radiology Authors: M. Joyce 1, J. T. Ryan 2, M. F. Mc Entee

More information

Comparison of MRI and ultrasound based liver volumetry in iron overload diseases

Comparison of MRI and ultrasound based liver volumetry in iron overload diseases Comparison of MRI and ultrasound based liver volumetry in iron overload diseases Poster No.: C-0913 Congress: ECR 2015 Type: Scientific Exhibit Authors: S. Keller, B. P. Schönnagel, P. Nielsen, M. T. Brehmer,

More information

NICE Guidelines for C-Spine Imaging: Real Life Impact

NICE Guidelines for C-Spine Imaging: Real Life Impact NICE Guidelines for C-Spine Imaging: Real Life Impact Poster No.: C-1367 Congress: ECR 2017 Type: Scientific Exhibit Authors: D. Weinberg, I. Djoukhadar, G. Potter; Salford/UK Keywords: Trauma, Audit and

More information

Role of ultrasound in the evaluation of the ileocecal valve

Role of ultrasound in the evaluation of the ileocecal valve Role of ultrasound in the evaluation of the ileocecal valve Poster No.: C-1581 Congress: ECR 2010 Type: Scientific Exhibit Topic: GI Tract Authors: M. Mohammed, M. Hussain, U. Momin, S. Lakhtakia, N. D.

More information

The imaging evaluation of breast implants

The imaging evaluation of breast implants The imaging evaluation of breast implants Poster No.: C-0654 Congress: ECR 2014 Type: Authors: Keywords: DOI: Educational Exhibit N. Rotaru; Chisinau/MD Neoplasia, Education and training, Cancer, Imaging

More information

Imaging Gorham's disease (vanishing bone)

Imaging Gorham's disease (vanishing bone) Imaging Gorham's disease (vanishing bone) Poster No.: C-2201 Congress: ECR 2010 Type: Scientific Exhibit Topic: Musculoskeletal Authors: D. Vanel, P. Ruggieri, M. Alberghini; Bologna/IT Keywords: Gorham,

More information

Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters

Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters Thoracic examinations with 16, 64, 128 and 256 slices CT: comparison of exposure doses measured with an anthropomorphic phantom and TLD dosimeters Poster No.: C-2584 Congress: ECR 2015 Type: Scientific

More information

CT evaluation of small bowel carcinoid tumors

CT evaluation of small bowel carcinoid tumors CT evaluation of small bowel carcinoid tumors Poster No.: C-0060 Congress: ECR 2015 Type: Educational Exhibit Authors: N. V. V. P. Costa, L. Nascimento, T. Bilhim ; Estoril/PT, PT, 1 2 3 1 2 3 Lisbon/PT

More information

Is ascites a sensible predictive sign of peritoneal involvement in patients with ovarian carcinoma?: our experience with FDG-PET/CT

Is ascites a sensible predictive sign of peritoneal involvement in patients with ovarian carcinoma?: our experience with FDG-PET/CT Is ascites a sensible predictive sign of peritoneal involvement in patients with ovarian carcinoma?: our experience with FDG-PET/CT Poster No.: C-1019 Congress: ECR 2013 Type: Scientific Exhibit Authors:

More information

Assessment of Vasospasm and Delayed Cerebral Ischemia after Subarachnoid Hemorrhage: Current concepts and Value of CT Perfusion and CT Angiography

Assessment of Vasospasm and Delayed Cerebral Ischemia after Subarachnoid Hemorrhage: Current concepts and Value of CT Perfusion and CT Angiography Assessment of Vasospasm and Delayed Cerebral Ischemia after Subarachnoid Hemorrhage: Current concepts and Value of CT Perfusion and CT Angiography Poster No.: C-2563 Congress: ECR 2012 Type: Educational

More information

Basal ganglia and thalamus in Parkinson disease: structural and connectivity associated changes

Basal ganglia and thalamus in Parkinson disease: structural and connectivity associated changes Basal ganglia and thalamus in Parkinson disease: structural and connectivity associated changes Poster No.: C-0851 Congress: ECR 2014 Type: Authors: Keywords: DOI: Scientific Exhibit M. Onu, L. Badea,

More information