Radiation Protection in BNCT Patients

Size: px
Start display at page:

Download "Radiation Protection in BNCT Patients"

Transcription

1 Radiation Protection in BNCT Patients J. Kessler 2, H. Blaumann 1, D. Feld 2,, E. Scharnichia 1, I. Levanón 1, C. Fernández 1, G. Facchini 1, M. Casal 3, S. González 2, J. Longhino 1, O. Calzetta 1, B. Roth 3, P. Menéndez 3, S. Liberman 2, M. D. Pereira 4 1 Centro Atómico Bariloche, Comisión Nacional de Energía Atómica, Avda Bustillo 9500, S. C. de Bariloche, 8400 Río Negro, Argentina 2 Centro Atómico Constituyentes, Comisión Nacional de Energía Atómica, Avda General Paz 1499, San Martin, B1650KNA Provincia de Buenos Aires, Argentina 3 Instituto de Oncología Ángel Roffo, Avda San Martín 5481, 1417, Ciudad Autónoma de Buenos Aires, Argentina 4 Agencia Nacional de Promoción Científica y Tecnológica, Av. Córdoba 831, Ciudad Autónoma de Buenos Aires, Argentina Abstract. Boron Neutron Capture Therapy (BNCT) is a technique that selectively targets cancer cells while sparing normal tissues by virtue of the differential uptake of a 10 B carrier compound in tumor. The National Atomic Energy Commission (CNEA) and the Oncology Institute Angel H. Roffo (IOAR) began a BNCT programme in 2003 for treating cutaneous skin melanomas in extremities. The neutron beam used is the hyperthermal one developed at the RA-6 Reactor of the Bariloche Atomic Centre (CAB). The prescribed dose is delivered in one fraction and therefore patient positioning and knowledge of the dose received by normal tissue are crucial. 10 irradiations have been done since The dose prescription was determined by the maximum tolerable skin dose. Due to the characteristics of this treatment the patient body might be exposed both to the primary beam and to the secondary photon beam produced by neutron capture at the target itself. A patient radiation-monitoring plan was implemented in order to evaluate the gamma dose delivered to sensible organs of each patient. An acrylic water-filled whole body phantom was used for preliminary gamma dose and thermal neutron flux measurements at positions related to patient s body sensible organs considering tentative patient positions. The beam port shielding was, in this way, optimized. TLD-700 and Manganese foils were used for gamma and thermal neutron detection. The TLD-700 thermal neutron response was previously evaluated with the in-phantom beam dosimetry characterization. In-vivo dosimetry with Thermal Luminescent Dosimeters (TLD) is implemented in order to evaluate gamma dose to sensible organs of each patient. These organs are chosen depending on its distance from the zone to be irradiated and its radio-sensibility. TLD s have been positioned outside the irradiated area. Maximum gamma dose received outside the radiation field in healthy tissues was well below tolerance dose for the compromised organs. KEYWORDS: BNCT, TLD, tolerance dose, positioning 1. Introduction BNCT is a radiation selective technique which uses a neutron beam impinging over cells containing a 10 B compound, boronophenylalanine (BPA) in this case, which is infused intravenously before the beginning of irradiation. The beam contains neutrons as well as photons which have to be taken into account for radiation protection. The BPA distinguishes between normal and tumor cells and mainly selects the latter ones. The reaction 10 B(n, α) 7 Li produces an α (alpha) particle and a 7 Li recoil nucleus of high Linear Energy Transfer (LET) and very short range in tissue, as well as an associated photon with an energy of MeV. The former particles are the responsible of the very high dose absorbed by tumor cells, while normal cells receive a considerably lower dose. The CNEA and the IOAR outlined a protocol for the treatment of metastatic melanomas in extremities. This type of skin melanoma is a highly lethal disease for which most of the adjuvant treatments have failed [1]. BNCT treatments began in Argentina in 2003 using the RA-6 reactor belonging to the CNEA and located in Bariloche. Since then, 7 patients with multiple subcutaneous skin metastases of melanoma have been treated and 10 irradiations have been carried out. As the beam is a mixed one, neutron as well as gamma doses to normal tissue have to be known for protection purposes. The outcome concerning gamma dose to normal tissue measured in patients, is presented in this paper. 1

2 2. Material and Methods The whole dose is delivered in one session for each field and the patient is treated through a unique entrance port of 15 cm diameter (Fig. 1). Dose prescription is done taking into account the maximum tolerable dose in normal skin for a single and unique fraction, and the values ranged from 16.5 to 24 Gy-Eq [2]. Figure 1: Schematic representation of the BNCT facility showing the main components of the beam port 2.1 Beam characteristics The RA-6 is a pool type reactor with 500 kw of nominal power. The original epithermal beam was thermalized for the treatment of superficial tumors, being now a hyperthermal mixed beam containing neutrons of lower energy getting the maximum flux between 0.5 cm and 1.0 cm depth in water. Fig. 1 shows a schematic representation of the port and its main components [3]. Dosimetric distribution of gamma dose has been done in a body-shaped water phantom using TLD 700 distributed in 16 different points of the phantom, simulating several regions of the body. The thermal neutron flux has been estimated with Mn foils [4]. 2.2 Patient positioning The main difference between traditional Radiotherapy and this modality is that the port of the reactor is stationary and cannot be rotated around the patient like the gantry of a typical external beam machine as a linear accelerator or a Co-60 unit. Thus, the positioning of patient is highly complicated, the resulting positions being sometimes a little uncomfortable (Fig. 2). 2

3 Figure 2 a) Positioning scheme b) Irradiation region of patient Nº 6 for patient Nº 6 The patient position is simulated at the Simulation Room (SR) in Buenos Aires using immobilisation devices which are afterwards taken to the Treatment Room (TR) in Bariloche (Fig. 3). The Simulation Room and the Treatment Room both have identical gridded windows. Moreover, the window of the Simulation Room has a beam s-eye view as it is made of transparent acrylic (Fig. 4). All the anatomical marks are photographed at the SR and patient position coordinates registered in a simulation report in order to reproduce the positioning at the TR, where there isn t a beam s eye view. The position is repeated and verified at the TR one day before treatment. Immobilisation is crucial for these treatments as the irradiation times are about 60 minutes. Figure 3: Treatment Room showing the port Figure 4: The beam s eye view at the Simulation Room 3

4 2.3 Patient irradiation and in-vivo dosimetry Patients are infused intravenously over 90 minutes, blood samples are extracted in spans of minutes during this period and until the beginning of irradiation, in order to analyze blood boron concentration. Those values are plotted versus time and fitted with an open two-compartiment pharmaco kinetic model for getting the final irradiation time [5]. After infusion patient is carefully positioned at the TR following the simulation report. Marks are added on the patient for monitoring patient movement from outside through a video camera during the irradiation (Fig. 5). Figure 5: Record of patient position at the beginning and end of treatment. LiF (TLD) is commonly used for in vivo dosimetry in other radiotherapy modalities. Furthermore, the use of TLD-700 in a mixed neutron-photon beam is reported and their accuracy is well determined [6],[7]. For each irradiation field, between 5 and 7 TLD s 700 are distributed over the patient, outside the irradiation area (Fig. 6) in order to estimate the gamma dose received by healthy tissues. The position of TLD s is chosen according to the localisation of each tumor and the type of organs at risk (OR) at the surroundings of the irradiated region. It is worth while to emphasize that staff remains in a controlled area during the irradiation and they don t enter to the TR till the dose rate is below 0.20 msv/h. Dose values read from personal dosimeters have been always below 0.15 msv. 4

5 Figure 6: TLD localisation for patient Nº3 for fields 3 (Fig.6 a), 2 (Fig. 6b) and 1 (Fig. 6c) a) b) TLD c) TLD As an example, we illustrate the particular case of patient Nº 3 whose schedule included 3 irradiation fields. The treated regions (the calf and the heel of right leg and the right foot sole) and the TLD s positions for each field are shown in Fig. 6. For one patient a protective borated polyethylene block was needed in order to avoid high neutron doses to parts of the body which were at the same level of the port (Fig.7). 5

6 Fig. 7: Patient Nº 2 with the shielding block Shielding block 3. Results An excellent reproduction of patient positioning is achieved and the monitoring showed that patient movement provoked a shift of only 2 to 4 mm of the irradiated region. This shift produces an estimated change in thermal neutron flux of 5%. The dose values got from the TLD s for the patient taken as example are shown in Table 1. As it was previously stated, the thermal neutron dose is reported to be 10% [4] to 20% [7], and these values are comparable to the uncertainty in the estimation of gamma dose. The contribution from thermal neutron field to TLD s is included and so, the values in Table 1 give higher dose levels than the actual absorbed gamma dose in inner organs at risk by a factor of 2.3 [4], so they can be considered a conservative estimate of dose. Measurements performed by Blaumann in the anthropomorphic water filled acrylic phantom [4] showed that outside the irradiated area, neutron dose was of the same order of magnitude than the uncertainty in the determination of gamma dose, thus it can be neglected. Based on these results, it was considered that doses in organs at risk outside the irradiated zone were due mainly to gamma rays, and TLD 700 were used to estimate it. The OR for all cases were both lungs and bladder and TLD s have been placed near them. The rest of TLD s locations have been chosen according to their proximity to the port. In patient 3, left knee and abdomen were near the port for all fields, so TLD s have been positioned there. For field 1 (right calf) the left groin was measured because it was near to the ovaries, for field 2 (right heel), the forehead and the eyes because they were facing the port, and for field 3 (right foot sole) again because the forehead was facing the port and the left foot was near the port. For all patients, doses have been well below 1Gy. This result also includes the estimations for the patient who had the protective block. 6

7 Table 1: Gamma dose from TLDs vs. body region for the field Nº 1 of patient Nº 3 Field Localisation Doses [cgy] Left groin Right Calf 2 Right Heel 3 Right Foot Sole Abdomen 8.6 Left knee 16.1 Right lung 7.9 Left lung 3.8 Bladder 11.4 Forehead 7.2 Abdomen 6.9 Left knee 10.7 Right lung 6.3 Left lung 6.2 Bladder 6.2 Forehead 7.7 Abdomen 9.4 Left knee 15.7 Right lung 8.3 Left lung 8.1 Bladder 10.3 Left foot Conclusion In spite of the uncomfortable position, the movement of the irradiated zone during irradiation was only of 2 to 4 mm and it is perfectly acceptable as it is the similar tolerable displacement in others radiotherapy modalities. The beam port shielding proved to be effective for patient radiation protection according to the results of the TLD s doses. Maximum dose received outside the radiation field in healthy tissues was well below tolerance dose for the sites measured. TLD s proved to be a useful tool to estimate doses in OR for this kind of treatments. The video system to monitor patietnts during irradiation is useful to determine their movements, and also provides sufficient data to calculate the retrospective dose distribution when needed. The general BNCT procedure resulted safe from the radiation protection point of view. REFERENCES [1] PAWLIK, T.M., SONDAK, V.K. Malignant melanoma: current state of primary and adjuvant treatment Crit. Rev. Oncol. Hemat. (2003) 45(3), [2] ROTH, B.M, et al. BNCT Clinical trials of Skin Melanoma Patients in Argentina 1st Congress of the Latin American of the Oncological Radiation Therapy Association, Montevideo, Uruguay (2007) [3] BLAUMANN, H.R., et al. Boron neutron capture therapy of skin melanomas at the RA-6 reactor. A procedural approach to beam set up and performance evaluation for upcoming clinical trials, Med. Phys. (2004) 31(1), [4] BLAUMANN, H., Evaluación de la distribución de dosis gamma en un fantoma cuerpo entero en la facilidad de BNCT del reactor RA-6, Informe técnico CNEA-CAB 47/014/02 (in Spanish) 7

8 [5] GONZÁLEZ, S.J., et al. First BNCT treatment of a skin melanoma in Argentina: Dosimetric Analysis and clinical outcome, Applied Radiation and Isotopes, (2004) 61, [6] KRY, S. F. et al: The use of LiF(TLD-100) as an out-of-field dosimeter, J. Appl. Clinical Medical Physics, (2007), 8 (4) [7] CARITA ASCHAN, Applicability of thermoluminiscent dosimeters in X-Ray organ dose determination and in the dosimetry of systemic and boron neutron capture radiotherapy University of Helsinki, 1999, Report Series in Physics, HU-P-D77, ISSN , ISBN

Single Photon Emission Tomography Approach for Online Patient Dose Assessment in Boron Neutron Capture Therapy

Single Photon Emission Tomography Approach for Online Patient Dose Assessment in Boron Neutron Capture Therapy Single Photon Emission Tomography Approach for Online Patient Dose Assessment in Boron Neutron Capture Therapy Daniel M Minsky a,b,c*, Alejandro Valda a,b, Andrés J Kreiner a,b,c and Alejandro A Burlon

More information

COMPARING THE RESPONSES OF TLD 100, TLD 600, TLD 700 AND TLD 400 IN MIXED NEUTRON-GAMMA FIELDS

COMPARING THE RESPONSES OF TLD 100, TLD 600, TLD 700 AND TLD 400 IN MIXED NEUTRON-GAMMA FIELDS 2015 International Nuclear Atlantic Conference - INAC 2015 São Paulo, SP, Brazil, October 4-9, 2015 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-06-9 COMPARING THE RESPONSES OF TLD

More information

Progress in Reactor and Accelerator Based BNCT at Kyoto University Research Reactor Institute

Progress in Reactor and Accelerator Based BNCT at Kyoto University Research Reactor Institute Progress in Reactor and Accelerator Based BNCT at Kyoto University Research Reactor Institute Yoshinori Sakurai 1 Kyoto University Research Reactor Institute Asashiro-nishi 2-1010, Kumatori-cho, Sennan-gun,

More information

Today, I will present the second of the two lectures on neutron interactions.

Today, I will present the second of the two lectures on neutron interactions. Today, I will present the second of the two lectures on neutron interactions. 1 The main goal of this lecture is to tell you a little about clinical neutron therapy, first with fast neutron beams, and

More information

ARTICLE IN PRESS. Applied Radiation and Isotopes

ARTICLE IN PRESS. Applied Radiation and Isotopes Applied Radiation and Isotopes 67 (2009) S54 S58 Contents lists available at ScienceDirect Applied Radiation and Isotopes journal homepage: www.elsevier.com/locate/apradiso Dynamic infrared imaging of

More information

Neutron - a tool in the cancer treatment - C. Paunoiu

Neutron - a tool in the cancer treatment - C. Paunoiu Neutron - a tool in the cancer treatment - C. Paunoiu 146 5/17/2009 The neutron -a a tool in the cancer treatment Dr.Constantin PĂUNOIU, constantin.paunoiu@nuclear.ro http://www.nuclear.ro biological material

More information

Epithermal neutron beams from the 7 Li(p,n) reaction near the threshold for neutron capture therapy

Epithermal neutron beams from the 7 Li(p,n) reaction near the threshold for neutron capture therapy IL NUOVO CIMENTO 38 C (2015) 179 DOI 10.1393/ncc/i2015-15179-9 Colloquia: UCANS-V Epithermal neutron beams from the 7 Li(p,n) reaction near the threshold for neutron capture therapy I. Porras( 1 ),J.Praena(

More information

Preliminary study of MAGAT polymer gel dosimetry for boron-neutron capture therapy

Preliminary study of MAGAT polymer gel dosimetry for boron-neutron capture therapy Journal of Physics: Conference Series OPEN ACCESS Preliminary study of MAGAT polymer gel dosimetry for boron-neutron capture therapy To cite this article: Shin-ichiro Hayashi et al 2015 J. Phys.: Conf.

More information

THERMOLUMINESCENT (TL) DOSIMETRY OF SLOW-NEUTRON FIELDS AT RADIOTHERAPY DOSE LEVEL

THERMOLUMINESCENT (TL) DOSIMETRY OF SLOW-NEUTRON FIELDS AT RADIOTHERAPY DOSE LEVEL THERMOLUMINESCENT (TL) DOSIMETRY OF SLOW-NEUTRON FIELDS AT RADIOTHERAPY DOSE LEVEL G. Gambarini Dipartimento di Fisica dell Università, Milano, Italy e-mail grazia.gambarini http://users.unimi.it/~frixy/

More information

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.

Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR. Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact Stephen F. Kry, Ph.D., D.ABR. Goals Compare out-of-field doses from various techniques Methods to reduce out-of-field doses Impact of

More information

Neutrons. ρ σ. where. Neutrons act like photons in the sense that they are attenuated as. Unlike photons, neutrons interact via the strong interaction

Neutrons. ρ σ. where. Neutrons act like photons in the sense that they are attenuated as. Unlike photons, neutrons interact via the strong interaction Neutrons Neutrons act like photons in the sense that they are attenuated as I = I 0 e μx where Unlike photons, neutrons interact via the strong interaction μ = The cross sections are much smaller than

More information

THE HIGH-CURRENT DEUTERON ACCELERATOR FOR THE NEUTRON THERAPY. S.V. Akulinichev, A. V. Andreev, V.M. Skorkin

THE HIGH-CURRENT DEUTERON ACCELERATOR FOR THE NEUTRON THERAPY. S.V. Akulinichev, A. V. Andreev, V.M. Skorkin THE HIGH-CURRENT DEUTERON ACCELERATOR FOR THE NEUTRON THERAPY S.V. Akulinichev, A. V. Andreev, V.M. Skorkin Institute for Nuclear Research of the RAS, Russia THE PROJECT OF NEUTRON SOURCES FOR THE NEUTRON

More information

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments

Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments Alana Hudson *1 1 Tom Baker Cancer Centre, Department of Medical Physics, 1331

More information

Neutron Interactions Part 2. Neutron shielding. Neutron shielding. George Starkschall, Ph.D. Department of Radiation Physics

Neutron Interactions Part 2. Neutron shielding. Neutron shielding. George Starkschall, Ph.D. Department of Radiation Physics Neutron Interactions Part 2 George Starkschall, Ph.D. Department of Radiation Physics Neutron shielding Fast neutrons Slow down rapidly by scatter in hydrogenous materials, e.g., polyethylene, paraffin,

More information

Design and Performance Of A Thermal Neutron Beam for Boron Neutron Capture Therapy At The University Of Missouri Research Reactor

Design and Performance Of A Thermal Neutron Beam for Boron Neutron Capture Therapy At The University Of Missouri Research Reactor Design and Performance Of A Thermal Neutron Beam for Boron Neutron Capture Therapy At The University Of Missouri Research Reactor J.D. Brockman J.C. McKibben In situ activation reaction, 10 B(n, a) 7 Li;

More information

PhD értekezés tézisei PHASE I CLINICAL STUDY ON BORON NEUTRON CAPTURE THERAPY. Dr Katalin Hideghéty. A Doktori Iskola vezetıje Prof. Dr.

PhD értekezés tézisei PHASE I CLINICAL STUDY ON BORON NEUTRON CAPTURE THERAPY. Dr Katalin Hideghéty. A Doktori Iskola vezetıje Prof. Dr. PhD értekezés tézisei PHASE I CLINICAL STUDY ON BORON NEUTRON CAPTURE THERAPY Dr Katalin Hideghéty A Doktori Iskola vezetıje Prof. Dr. Nagy Judit témavezetı Prof. Dr. Ember István Pécsi Tudományegyetem

More information

Chapter Introduction

Chapter Introduction Chapter 1 Introduction Malignancy is a life-threatening condition in which tumor cells are of a highly invasive character and penetrate normal organs and exhibit an obstinate resistance against cancer

More information

RPI DOSIMETRY. Therefore the description of the work is grouped in two main areas:

RPI DOSIMETRY. Therefore the description of the work is grouped in two main areas: RPI DOSIMETRY Introduction The main objective of the RPI Dosimetry is the characterization of the radiation field of the facilities available in the reactor. Most of the use of the reactor centers in irradiation

More information

Radiation Dosimetry at the BNL High Flux Beam Reactor and Medical Research Reactor

Radiation Dosimetry at the BNL High Flux Beam Reactor and Medical Research Reactor BNL-66807 1 1 1 1 2 Holden, N.E., Hu, J-P., Greenberg, D.D., Reciniello, R.N., Farrell, K. and Greenwood, L.R. Radiation Dosimetry at the BNL High Flux Beam Reactor and Medical Research Reactor Reference:Holden,N.E.,

More information

FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS

FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS Etieli Camargo da Costa 1, Luiz Antonio Ribeiro da Rosa 2 Delano Valdivino Santos Batista 3 1 Instituto de Radioproteção

More information

A Feasibility Study of the SLOWPOKE-2 Reactor as a Neutron Source for Boron Neutron Cancer Treatment

A Feasibility Study of the SLOWPOKE-2 Reactor as a Neutron Source for Boron Neutron Cancer Treatment A Feasibility Study of the SLOWPOKE- Reactor as a Neutron Source for Boron Neutron Cancer Treatment Introduction M.J. McCall, M. Pierre Royal Military College of Canada, Kingston, Ontario, Canada K7K 7B4

More information

Calibration of two 90 Sr+ 90 Y dermatological applicators

Calibration of two 90 Sr+ 90 Y dermatological applicators Calibration of two 90 Sr+ 90 Y dermatological applicators Patrícia L. Antonio *, Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares, IPEN-CNEN/SP Av. Prof. Lineu Prestes 2242, 05508-000,

More information

Protection of the contralateral breast during radiation therapy for breast cancer

Protection of the contralateral breast during radiation therapy for breast cancer Protection of the contralateral breast during radiation therapy for breast cancer Edgardo Garrigó a*, Alejandro Germanier b, Silvia Zunino a a Instituto Privado de Radioterapia, Ob Oro 423 (5000) Córdoba,

More information

6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS

6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS 6/29/2012 MANAGEMENT OF RADIOTHERAPY PATIENTS WITH IMPLANTED CARDIAC DEVICES Dimitris Mihailidis, PhD., Charleston Radiation Therapy Consultants Charleston, WV 25304 WHAT IS IN THIS PRESENTATION? Types

More information

Response evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry

Response evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry Response evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry Luciana C. Matsushima 1, Glauco R. Veneziani 1, Roberto K. Sakuraba

More information

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system

ROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2013 ROPES eye plaque dosimetry: commissioning

More information

Optimization of an accelerator-based epithermal neutron source for neutron capture therapy

Optimization of an accelerator-based epithermal neutron source for neutron capture therapy Applied Radiation and Isotopes 61 (2004) 1009 1013 Optimization of an accelerator-based epithermal neutron source for neutron capture therapy O.E. Kononov a, *, V.N. Kononov a, M.V. Bokhovko a, V.V. Korobeynikov

More information

X-RAYS INDIVIDUAL DOSE ASSESSMENT USING TLD DOSIMETERS

X-RAYS INDIVIDUAL DOSE ASSESSMENT USING TLD DOSIMETERS X-RAYS INDIVIDUAL DOSE ASSESSMENT USING TLD DOSIMETERS Carlos Salas Nucleoeléctrica Argentina SA - Embalse NPP Environmental Control and Dosimetry Division INTRODUCTION: This paper describes the methodology

More information

Peak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam

Peak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam Peak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam University of Chicago CDH Proton Center LET study C. Reft 1, H. Ramirez 2 and M. Pankuch

More information

Neutron sources from Nuclear Reactors

Neutron sources from Nuclear Reactors Sorgenti di neutroni e loro applicazioni in ambito INFN Neutron sources from Nuclear Reactors Alfonso Santagata alfonso.santagata@enea.it ENEA * Via Anguillarese, 301 00123 Rome Italy * Agezia nazionale

More information

MODELLING A GAMMA IRRADIATION PROCESS USING THE MONTE CARLO METHOD

MODELLING A GAMMA IRRADIATION PROCESS USING THE MONTE CARLO METHOD 2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte, MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 MODELLING A GAMMA IRRADIATION

More information

First application of hemi-body electron beam irradiation for Kaposi sarcoma at the lower extremities

First application of hemi-body electron beam irradiation for Kaposi sarcoma at the lower extremities JBUON 2018; 23(1): 268-272 ISSN: 1107-0625, online ISSN: 2241-6293 www.jbuon.com E-mail: editorial_office@jbuon.com SHORT COMMUNICATION First application of hemi-body electron beam irradiation for Kaposi

More information

The Therapeutic Ratio in BNCT: Assessment using the Rat 9L Gliosarcoma Brain Tumor and Spinal Cord Models

The Therapeutic Ratio in BNCT: Assessment using the Rat 9L Gliosarcoma Brain Tumor and Spinal Cord Models BNL-63408 The Therapeutic Ratio in BNCT: Assessment using the Rat 9L Gliosarcoma Brain Tumor and Spinal Cord Models J.A. Coderre, G.M. Morris 1, P.L. Micca, M.M. Nawrocky, CD. Fisher, A. Bywaters and J;W.

More information

RPC s Credentialing Programs for Clinical Trials

RPC s Credentialing Programs for Clinical Trials RPC s Credentialing Programs for Clinical Trials July 19, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff Mission The mission of the Radiological Physics Center is to assure NCI and the Cooperative Groups

More information

Design of a BSA for Producing Epithermal Neutron for BNCT

Design of a BSA for Producing Epithermal Neutron for BNCT Siam Physics Congress 2015 Design of a BSA for Producing Epithermal Neutron for BNCT M Asnal *, T Liamsuwan 2 and T Onjun 1 1 Sirindhorn International Institute of Technology, Thammasat University 2 Nuclear

More information

Option D: Medicinal Chemistry

Option D: Medicinal Chemistry Option D: Medicinal Chemistry Basics - unstable radioactive nuclei emit radiation in the form of smaller particles alpha, beta, positron, proton, neutron, & gamma are all used in nuclear medicine unstable

More information

CALIBRATION OF A TLD ALBEDO INDIVIDUAL NEUTRON MONITOR

CALIBRATION OF A TLD ALBEDO INDIVIDUAL NEUTRON MONITOR 2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 CALIBRATION OF A TLD

More information

Shielding Design Basis and its Calculation of High Energy Medical Linac Installed in Bangladesh Atomic Energy Commission, Bangladesh

Shielding Design Basis and its Calculation of High Energy Medical Linac Installed in Bangladesh Atomic Energy Commission, Bangladesh Universal Journal of Medical Science 5(2): 27-31, 2017 DOI: 10.13189/ujmsj.2017.050202 http://www.hrpub.org Shielding Design Basis and its Calculation of High Energy Medical Linac Installed in Bangladesh

More information

Neutron dose evaluation in radiotherapy

Neutron dose evaluation in radiotherapy Neutron dose evaluation in radiotherapy Francesco d Errico University of Pisa, Italy Yale University, USA Radiation therapy with a linear accelerator (LINAC) Photoneutron production in accelerator head

More information

Diode calibration for dose determination in total body irradiation

Diode calibration for dose determination in total body irradiation Iran. J. Radiat. Res., 2008; 6 (1): 43-50 Diode calibration for dose determination in total body irradiation M.Allahverdi 1*, Gh. Geraily 1, M. Esfehani 3,A. Sharafi 2,A. Shirazi 1 1 Department of Medical

More information

Radiation qualities in carbon-ion radiotherapy at NIRS/HIMAC

Radiation qualities in carbon-ion radiotherapy at NIRS/HIMAC Radiation qualities in carbon-ion radiotherapy at NIRS/ Shunsuke YONAI Radiological Protection Section Research Center for Charged Particle Therapy National Institute of Radiological Sciences (NIRS) E-mail:

More information

Distribution of Water Phantom BNCT Kartini Research Reactor Based Using PHITS

Distribution of Water Phantom BNCT Kartini Research Reactor Based Using PHITS Indonesian Journal of Physics and Nuclear Applications Volume 3, Number 2, June 2018, p. 43-48 e-issn 2550-0570, FSM UKSW Publication Distribution of Water Phantom BNCT Kartini Research Reactor Based Using

More information

SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES

SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES MELISSA C. MARTIN, M.S., FACR, FAAPM PRESIDENT AAPM - 2017 PRESIDENT - THERAPY PHYSICS INC., GARDENA, CA MELISSA@THERAPYPHYSICS.COM AAPM Spring

More information

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.

Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D. Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, 2014 Pamela Myers, Ph.D. Introduction o o o o o Outline GRID compensator Purpose of SFRT/GRID therapy Fractionation

More information

Study on Microdosimetry for Boron Neutron Capture Therapy

Study on Microdosimetry for Boron Neutron Capture Therapy Progress in NUCLEAR SCIENCE and TECHNOLOGY, Vol. 2, pp.242-246 (2011) ARTICLE Study on Microdosimetry for Boron Neutron Capture Therapy Tetsuya MUKAWA 1,*, Tetsuo MATSUMOTO 1 and Koji NIITA 2 1 Tokyo City

More information

Radiologic Units: What You Need to Know

Radiologic Units: What You Need to Know Radiologic Units: What You Need to Know TODD VAN AUKEN M.ED. RT (R)(MR) Agenda Greys, Sieverts, Coulombs per kg, & Becquerel's Conventional Units Other Concepts (LET, Q-Factor, Effective Dose, NCRP Report

More information

Boron Neutron Capture Therapy (BNCT) - Low-Energy Neutron Spectrometer for Neutron Field Characterization - )

Boron Neutron Capture Therapy (BNCT) - Low-Energy Neutron Spectrometer for Neutron Field Characterization - ) Boron Neutron Capture Therapy (BNCT) - Low-Energy Neutron Spectrometer for Neutron Field Characterization - ) Isao MURATA and Tsubasa OBATA Division of Electrical, Electronic and Information Engineering,

More information

Djoko S. Pudjorahardjo, Widarto, Isman Mulyadi T

Djoko S. Pudjorahardjo, Widarto, Isman Mulyadi T Indonesian Journal of Physics and NuclearApplications Volume 3, Number 1, February 2018, p. 15-20 e-issn 2550-0570, FSM UKSWPublication 2549-046X, FSM UKSWPublication Detail Engineering Design of Compact

More information

Neutron Radiotherapy: Past, Present, and Future Directions

Neutron Radiotherapy: Past, Present, and Future Directions Neutron Radiotherapy: Past, Present, and Future Directions Theodore L. Phillips Lecture -- 2014 George E. Laramore Ph.D., M.D. NONE Conflicts of Interest Peter Wootton Professor of Radiation Oncology University

More information

Out-of-field dosimetry in radiotherapy for input to epidemiological studies. Roger Harrison

Out-of-field dosimetry in radiotherapy for input to epidemiological studies. Roger Harrison MELODI 7th Workshop, 9 11 November 2015 Helmholtz Zentrum München Next Generation Radiation Protection Research Out-of-field dosimetry in radiotherapy for input to epidemiological studies Roger Harrison

More information

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *

A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM * Romanian Reports in Physics, Vol. 66, No. 2, P. 394 400, 2014 A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM * D. ADAM 1,2,

More information

22.55 Principles of Radiation Interactions Name: Spring 2004 Professor Coderre Exam 2 April 30, 2004

22.55 Principles of Radiation Interactions Name: Spring 2004 Professor Coderre Exam 2 April 30, 2004 22.55 Principles of Radiation Interactions Name: Spring 2004 Professor Coderre Exam 2 April 30, 2004 You have 2 hours to complete this exam. This exam is closed book. Please show all work on the attached

More information

Application of the Multi-Model Monte-Carlo Treatment Planning System Combined with PHITS to Proton Radiotherapy

Application of the Multi-Model Monte-Carlo Treatment Planning System Combined with PHITS to Proton Radiotherapy 1 1 6 2 1 2 3 5 6 7 9 1 1 1 1 2 1 3 1 1 5 1 6 1 7 1 1 9 2 2 1 Progress in NUCLEAR SCIENCE and TECHNOLOGY, Vol. 2, pp.213-21 (211) DOI: 1.15669/pnst.2.213 ARTICLE Application of the Multi-Model Monte-Carlo

More information

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti M. Bruzzi,M. Bucciolini, L. Marrazzo, D. Menichelli University of

More information

ADVANCES IN RADIATION TECHNOLOGIES IN THE TREATMENT OF CANCER

ADVANCES IN RADIATION TECHNOLOGIES IN THE TREATMENT OF CANCER ADVANCES IN RADIATION TECHNOLOGIES IN THE TREATMENT OF CANCER Bro. Dr. Collie Miller IARC/WHO Based on trends in the incidence of cancer, the International Agency for Research on Cancer (IARC) and WHO

More information

V. 4. Design and Benchmark Experiment for Cyclotron-based Neutron Source for BNCT

V. 4. Design and Benchmark Experiment for Cyclotron-based Neutron Source for BNCT CYRIC Annual Report 2002 V. 4. Design and Benchmark Experiment for Cyclotron-based Neutron Source for BNCT Yonai S., ItogaT., Nakamura T., Baba M., Yashima, H. Yokobori H. *, and Tahara Y. ** Cyclotron

More information

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.

Introduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons. Measurement of Secondary Radiation for Electron and Proton Accelerators D. Followill, Ph.D. Radiological Physics Center U. T. M. D. Anderson Cancer Center Introduction Patients undergoing radiation therapy

More information

The determination of timer error and its role in the administration of specified doses

The determination of timer error and its role in the administration of specified doses JURNAL FIZIK MALAYSIA VOLUME 29, NUMBER 1&2 2008 The determination of timer error and its role in the administration of specified doses S. B. Samat a *, C. J. Evans b, T. Kadni c and M. T. Dolah c a School

More information

QA for Clinical Dosimetry with Emphasis on Clinical Trials

QA for Clinical Dosimetry with Emphasis on Clinical Trials QA for Clinical Dosimetry with Emphasis on Clinical Trials Geoffrey S. Ibbott, Ph.D. and RPC Staff G. Ibbott, AAPM Summer School, June 24, 2009 1 QA Infrastructure for Clinical Trials Participating Institutions

More information

Chapters from Clinical Oncology

Chapters from Clinical Oncology Chapters from Clinical Oncology Lecture notes University of Szeged Faculty of Medicine Department of Oncotherapy 2012. 1 RADIOTHERAPY Technical aspects Dr. Elemér Szil Introduction There are three possibilities

More information

Characterization of OSL response of LiF:Mg,Ti and microlif:mg, Ti to 60 Co gamma source

Characterization of OSL response of LiF:Mg,Ti and microlif:mg, Ti to 60 Co gamma source Characterization of OSL response of LiF:Mg,Ti and microlif:mg, Ti to 60 Co gamma source Amanda Bravim, Leticia Lucente Campos Instituto de Pesquisas Energéticas e Nucleares (IPEN / CNEN - SP) Av. Professor

More information

CHARACTERIZATION OF TLD-100 DOSIMETER FOR MONITORING THE EXTREMITIES OF WORKERS IN NUCLEAR MEDICINE

CHARACTERIZATION OF TLD-100 DOSIMETER FOR MONITORING THE EXTREMITIES OF WORKERS IN NUCLEAR MEDICINE 2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 CHARACTERIZATION OF TLD-100

More information

What Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff

What Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff What Can Go Wrong in Radiation Treatment: Data from the RPC Geoffrey S. Ibbott, Ph.D. and RPC Staff Clinical NCI Trials CALGB 1768 Participating Institutions NCCTG ECOG COG ACOSOG SWOG QARC RPC ATC RTOG

More information

Dose Homogeneity of the Total Body Irradiation in vivo and in vitro confirmed with Thermoluminescent Dosimeter

Dose Homogeneity of the Total Body Irradiation in vivo and in vitro confirmed with Thermoluminescent Dosimeter Dose Homogeneity of the Total Body Irradiation in vivo and in vitro confirmed with Thermoluminescent Dosimeter E.K. Chie M.D. 1, S.W. Park M.D. 1, W.S. Kang PhD. 1,2, I.H. Kim M.D. 1,2, S.W. Ha M.D. 1,2,

More information

Radiation Monitoring Instruments

Radiation Monitoring Instruments Radiation Monitoring Instruments This set of slides is based on Chapter 4 authored byg. Rajan, J. Izewska of the IAEA publication (ISBN 92-0-107304-6): Radiation Oncology Physics: A Handbook for Teachers

More information

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine

Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine 2017 IJSRST Volume 3 Issue 3 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine Samuel Nii Adu Tagoe

More information

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*

IAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* H. Ferreira, A. Dutreix, A. Bridier, D. Marre, J. Chavaudra, H. Svensson ESTRO-EQUAL Measuring Laboratory, Service

More information

MONITORING OF OCCUPATIONAL EXPOSURE AT NUCLEAR FACILITIES

MONITORING OF OCCUPATIONAL EXPOSURE AT NUCLEAR FACILITIES GUIDE YVL 7.10 / 29 JANUARY 2002 MONITORING OF OCCUPATIONAL EXPOSURE AT NUCLEAR FACILITIES 1 GENERAL 3 2 PROVISIONS OF THE RADIATION ACT AND DECREE 3 3 MONITORING OF RADIATION EXPOSURE 4 3.1 General requirements

More information

Medical Use of Radioisotopes

Medical Use of Radioisotopes Medical Use of Radioisotopes Therapy Radioisotopes prove to be useful in the application of brachytherapy, the procedure for using temporary irradiation close to the area of disease (i.e. cancer) 10% Medical

More information

Study of neutron-dna interaction at the IPEN BNCT Research facility ABSTRACT

Study of neutron-dna interaction at the IPEN BNCT Research facility ABSTRACT Study of neutron-dna interaction at the IPEN BNCT Research facility Maritza Rodriguez Gual 1, Oscar Rodriguez Hoyos 1, Fernando Guzman Martinez 1, Airton Deppman 2, J.D.T.Aruda Neto 2, V.P.Likhachev 2,

More information

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience

TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience IAEA-CN-96-82 TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience J. Francisco Aguirre, Ramesh C. Tailor, Geoffrey S. Ibbott, Marilyn Stovall and William F. Hanson

More information

Characterization and implementation of Pencil Beam Scanning proton therapy techniques: from spot scanning to continuous scanning

Characterization and implementation of Pencil Beam Scanning proton therapy techniques: from spot scanning to continuous scanning Characterization and implementation of Pencil Beam Scanning proton therapy techniques: from spot scanning to continuous scanning Supervisors Prof. V. Patera PhD R. Van Roermund Candidate Annalisa Patriarca

More information

nuclear science and technology

nuclear science and technology EUROPEAN COMMISSION nuclear science and technology Evaluation of Individual Dosimetry in Mixed Neutron and Photon Radiation Fields (EVIDOS) Editor: Helmut Schuhmacher, Physikalisch-Technische Bundesanstalt

More information

Neutron-Gamma Mixed field Dosimetry on a Child phantom under Therapeutic Proton Irradiation using TL Dosimeters

Neutron-Gamma Mixed field Dosimetry on a Child phantom under Therapeutic Proton Irradiation using TL Dosimeters 133 Clinical Radiation Oncology Neutron-Gamma Mixed field Dosimetry on a Child phantom under Therapeutic Proton Irradiation using TL Dosimeters Bhaskar Mukherjee 1,2, Carolina Lina 1 Vladimir Mares 3 1Westdeutsches

More information

PET Guidance of Therapy for BNCT and in vivo B-10 imaging

PET Guidance of Therapy for BNCT and in vivo B-10 imaging INFN LNL Legnaro 17-19 Novembre 2009 Principles of Positron Emission Tomography and Radiopharmaceuticals PET Guidance of Therapy for BNCT and in vivo B-10 imaging Luca Menichetti, Ph.D C.N.R. Institute

More information

Radiation physics and radiation protection. University of Szeged Department of Nuclear Medicine

Radiation physics and radiation protection. University of Szeged Department of Nuclear Medicine Radiation physics and radiation protection University of Szeged Department of Nuclear Medicine Radiation doses to the population 1 Radiation doses to the population 2 Sources of radiation 1 Radiation we

More information

Physical Bases : Which Isotopes?

Physical Bases : Which Isotopes? Physical Bases : Which Isotopes? S. Gnesin Institute of Radiation Physics, Lausanne University Hospital, Lausanne, Switzerland 1/53 Theranostic Bruxelles, 2 Octobrer 2017 Theranostic : use of diagnostic

More information

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology

D DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology Journal of Life Sciences 9 (2015) 120-126 doi: 10.17265/1934-7391/2015.03.005 D DAVID PUBLISHING Uncertainties of in vivo Dosimetry Using Semiconductors Zeina Al Kattar, Hanna El Balaa and Saeed Zahran

More information

Dosimetric characteristics of 137 Cs sources used in after loading Selectron system by Monte Carlo method

Dosimetric characteristics of 137 Cs sources used in after loading Selectron system by Monte Carlo method Iran. J. Radiat. Res., 2007; 5 (3): 147-152 Dosimetric characteristics of Cs sources used in after loading Selectron system by Monte Carlo method M.B.Tavakoli, D. Shahbazi-Gahrouei *, M. Hosseinpour Department

More information

The physics of boron neutron capture therapy: an emerging and innovative treatment for glioblastoma and melanoma

The physics of boron neutron capture therapy: an emerging and innovative treatment for glioblastoma and melanoma Physics and Astronomy Department Physics and Astronomy Comps Papers Carleton College Year 2005 The physics of boron neutron capture therapy: an emerging and innovative treatment for glioblastoma and melanoma

More information

This article was published in an Elsevier journal. The attached copy is furnished to the author for non-commercial research and education use, including for instruction at the author s institution, sharing

More information

Patient dosimetry for total body irradiation using single-use MOSFET detectors

Patient dosimetry for total body irradiation using single-use MOSFET detectors JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 4, FALL 2008 Patient dosimetry for total body irradiation using single-use MOSFET detectors Tina Marie Briere, 1 Ramesh Tailor, 1 Naresh Tolani,

More information

TomoTherapy. Michelle Roach CNC Radiation Oncology Liverpool Hospital CNSA. May 2016

TomoTherapy. Michelle Roach CNC Radiation Oncology Liverpool Hospital CNSA. May 2016 TomoTherapy Michelle Roach CNC Radiation Oncology Liverpool Hospital CNSA May 2016 TomoTherapy The Facts Greek Tomo = slice Advanced form of IMRT 3D computerised tomography (CT) imaging immediately prior

More information

Title. Author(s)Ishikawa, Masayori; Tanaka, Kenichi; Endo, Satrou; H. CitationJournal of Radiation Research, 56(2): Issue Date

Title. Author(s)Ishikawa, Masayori; Tanaka, Kenichi; Endo, Satrou; H. CitationJournal of Radiation Research, 56(2): Issue Date Title Application of an ultraminiature thermal neutron mon therapy Author(s)Ishikawa, Masayori; Tanaka, Kenichi; Endo, Satrou; H CitationJournal of Radiation Research, 56(2): 391-396 Issue Date 2015-03

More information

GDR in Radiotherapy Treatment Fields with 18 MV Accelerators

GDR in Radiotherapy Treatment Fields with 18 MV Accelerators GDR in Radiotherapy Treatment Fields with 18 MV Accelerators R. R. Martín-Landrove 1, J. Dávila 1, H. R. Vega-Carrillo 2, M. T. Barrera 3, A.J. Kreiner 4, F. Pino 3, H. Barros 3, E. D. Greaves 3 and L.

More information

Thin Beryllium target for 9 Be(d,n)- driven BNCT

Thin Beryllium target for 9 Be(d,n)- driven BNCT Thin Beryllium target for 9 Be(d,n)- driven BNCT M.E.Capoulat 1-3, D.M.Minsky 1-3, L.Gagetti 1-3, M. Suárez Anzorena 1, M.F.del Grosso 1-2, J.Bergueiro 1, D.Cartelli 1-3, M.Baldo 1, W.Castell 1, J.Gomez

More information

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES

SUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES A Personal Dosimetry Intercomparison Study in Asian and Pacific Region Hiroyuki MURAKAMI and Fumiaki TAKAHASHI Department of Health Physics, Japan Atomic Energy Research Institute and Richard V. Griffith

More information

Risk of Secondary Fatal Malignancies from Hi-Art Tomotherapy

Risk of Secondary Fatal Malignancies from Hi-Art Tomotherapy Risk of Secondary Fatal Malignancies from Hi-Art Tomotherapy Susannah Lazar David Followill, Ph.D. John Gibbons, Ph.D. * Anita Mahajan, M.D. Mohammad Salehpour, Ph.D. Marilyn Stovall, Ph.D. R. Allen White,

More information

I. Equipments for external beam radiotherapy

I. Equipments for external beam radiotherapy I. Equipments for external beam radiotherapy 5 linear accelerators (LINACs): Varian TrueBeam 6, 10 & 18 MV photons, 6-18 MeV electrons, image-guided (IGRT) and intensity modulated radiotherapy (IMRT),

More information

BORON NEUTRON CAPTURE THERAPY SETUP FOR A LINEAR ACCELERATOR

BORON NEUTRON CAPTURE THERAPY SETUP FOR A LINEAR ACCELERATOR BORON NEUTRON CAPTURE THERAPY SETUP FOR A LINEAR ACCELERATOR C. F. CHIOJDEANU, C. PAVEL, F. CONSTANTIN National Institute for Physics and Nuclear Engineering Horia Hulubei, P.O. Box MG-6, RO-077125 Bucharest-Mãgurele,

More information

Outline. Outline 3/30/12. Second Cancers from. Radiotherapy Procedures. Stephen F. Kry, Ph.D., D.ABR.

Outline. Outline 3/30/12. Second Cancers from. Radiotherapy Procedures. Stephen F. Kry, Ph.D., D.ABR. Second Cancers from Radiotherapy Procedures Stephen F. Kry, Ph.D., D.ABR. Outline Radiation and cancer induction Medically exposed people Estimating risk of second cancers Minimizing the risk Outline Radiation

More information

Radiotherapy. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada

Radiotherapy. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Overview Introduction Overview Introduction Brachytherapy Radioisotopes in contact with the tumor Overview

More information

The ARN Critical Dosimetry System. Gregori, B.N.; Papadópulos, S.; Cruzate, J.A.; Equillor, H.E. and Kunst, J.J.

The ARN Critical Dosimetry System. Gregori, B.N.; Papadópulos, S.; Cruzate, J.A.; Equillor, H.E. and Kunst, J.J. The ARN Critical Dosimetry System Gregori, B.N.; Papadópulos, S.; Cruzate, J.A.; Equillor, H.E. and Kunst, J.J. Presentado en: 11 th International Congress on the International Radiation Protection Association.

More information

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy

Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti a*, M. Bucciolini a, L. Marrazzo a, D. Menichelli a. a) Department

More information

COMPARISON OF RADIOBIOLOGICAL EFFECTS OF CARBON IONS TO PROTONS ON A RESISTANT HUMAN MELANOMA CELL LINE

COMPARISON OF RADIOBIOLOGICAL EFFECTS OF CARBON IONS TO PROTONS ON A RESISTANT HUMAN MELANOMA CELL LINE COMPARISON OF RADIOBIOLOGICAL EFFECTS OF CARBON IONS TO PROTONS ON A RESISTANT HUMAN MELANOMA CELL LINE I. Petrovi a, A. Risti -Fira a, L. Kori anac a, J. Požega a, F. Di Rosa b, P. Cirrone b and G. Cuttone

More information

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients

Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Moyed Miften, PhD Professor and Chief Physicist University of Colorado Chester Reft, PhD Associate Professor University of Chicago

More information

To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence

To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence SCIENCE & TECHNOLOGY To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence Alburuj R. Rahman*, Jian Z. Wang, Dr. Z. Huang, Dr. J. Montebello Department of

More information

Dosimetry Optimization System and Integrated Software (DOSIS): a comparison against Fluka code results over a standard phantom

Dosimetry Optimization System and Integrated Software (DOSIS): a comparison against Fluka code results over a standard phantom Dosimetry Optimization System and Integrated Software (DOSIS): a comparison against Fluka code results over a standard phantom Institute of Physics E. Gaviola - CONICET & LIIFAMIRX - Laboratorio de Investigación

More information

Introduction to clinical Radiotherapy

Introduction to clinical Radiotherapy Introduction to clinical Radiotherapy Sarah Brothwood Radiotherapy Practice Educator Rosemere Cancer Centre Sarah.brothwood@lthtr.nhs.uk 01772 522694 Radiotherapy We have been able to see and document

More information

Out-of-field organ doses from therapeutic irradiation during childhood: is there an excess risk for second cancer induction?

Out-of-field organ doses from therapeutic irradiation during childhood: is there an excess risk for second cancer induction? Out-of-field organ doses from therapeutic irradiation during childhood: is there an excess risk for second cancer induction? Poster No.: C-0096 Congress: ECR 2012 Type: Scientific Paper Authors: M. Mazonakis,

More information