Radiation Protection in BNCT Patients
|
|
- Poppy Dean
- 5 years ago
- Views:
Transcription
1 Radiation Protection in BNCT Patients J. Kessler 2, H. Blaumann 1, D. Feld 2,, E. Scharnichia 1, I. Levanón 1, C. Fernández 1, G. Facchini 1, M. Casal 3, S. González 2, J. Longhino 1, O. Calzetta 1, B. Roth 3, P. Menéndez 3, S. Liberman 2, M. D. Pereira 4 1 Centro Atómico Bariloche, Comisión Nacional de Energía Atómica, Avda Bustillo 9500, S. C. de Bariloche, 8400 Río Negro, Argentina 2 Centro Atómico Constituyentes, Comisión Nacional de Energía Atómica, Avda General Paz 1499, San Martin, B1650KNA Provincia de Buenos Aires, Argentina 3 Instituto de Oncología Ángel Roffo, Avda San Martín 5481, 1417, Ciudad Autónoma de Buenos Aires, Argentina 4 Agencia Nacional de Promoción Científica y Tecnológica, Av. Córdoba 831, Ciudad Autónoma de Buenos Aires, Argentina Abstract. Boron Neutron Capture Therapy (BNCT) is a technique that selectively targets cancer cells while sparing normal tissues by virtue of the differential uptake of a 10 B carrier compound in tumor. The National Atomic Energy Commission (CNEA) and the Oncology Institute Angel H. Roffo (IOAR) began a BNCT programme in 2003 for treating cutaneous skin melanomas in extremities. The neutron beam used is the hyperthermal one developed at the RA-6 Reactor of the Bariloche Atomic Centre (CAB). The prescribed dose is delivered in one fraction and therefore patient positioning and knowledge of the dose received by normal tissue are crucial. 10 irradiations have been done since The dose prescription was determined by the maximum tolerable skin dose. Due to the characteristics of this treatment the patient body might be exposed both to the primary beam and to the secondary photon beam produced by neutron capture at the target itself. A patient radiation-monitoring plan was implemented in order to evaluate the gamma dose delivered to sensible organs of each patient. An acrylic water-filled whole body phantom was used for preliminary gamma dose and thermal neutron flux measurements at positions related to patient s body sensible organs considering tentative patient positions. The beam port shielding was, in this way, optimized. TLD-700 and Manganese foils were used for gamma and thermal neutron detection. The TLD-700 thermal neutron response was previously evaluated with the in-phantom beam dosimetry characterization. In-vivo dosimetry with Thermal Luminescent Dosimeters (TLD) is implemented in order to evaluate gamma dose to sensible organs of each patient. These organs are chosen depending on its distance from the zone to be irradiated and its radio-sensibility. TLD s have been positioned outside the irradiated area. Maximum gamma dose received outside the radiation field in healthy tissues was well below tolerance dose for the compromised organs. KEYWORDS: BNCT, TLD, tolerance dose, positioning 1. Introduction BNCT is a radiation selective technique which uses a neutron beam impinging over cells containing a 10 B compound, boronophenylalanine (BPA) in this case, which is infused intravenously before the beginning of irradiation. The beam contains neutrons as well as photons which have to be taken into account for radiation protection. The BPA distinguishes between normal and tumor cells and mainly selects the latter ones. The reaction 10 B(n, α) 7 Li produces an α (alpha) particle and a 7 Li recoil nucleus of high Linear Energy Transfer (LET) and very short range in tissue, as well as an associated photon with an energy of MeV. The former particles are the responsible of the very high dose absorbed by tumor cells, while normal cells receive a considerably lower dose. The CNEA and the IOAR outlined a protocol for the treatment of metastatic melanomas in extremities. This type of skin melanoma is a highly lethal disease for which most of the adjuvant treatments have failed [1]. BNCT treatments began in Argentina in 2003 using the RA-6 reactor belonging to the CNEA and located in Bariloche. Since then, 7 patients with multiple subcutaneous skin metastases of melanoma have been treated and 10 irradiations have been carried out. As the beam is a mixed one, neutron as well as gamma doses to normal tissue have to be known for protection purposes. The outcome concerning gamma dose to normal tissue measured in patients, is presented in this paper. 1
2 2. Material and Methods The whole dose is delivered in one session for each field and the patient is treated through a unique entrance port of 15 cm diameter (Fig. 1). Dose prescription is done taking into account the maximum tolerable dose in normal skin for a single and unique fraction, and the values ranged from 16.5 to 24 Gy-Eq [2]. Figure 1: Schematic representation of the BNCT facility showing the main components of the beam port 2.1 Beam characteristics The RA-6 is a pool type reactor with 500 kw of nominal power. The original epithermal beam was thermalized for the treatment of superficial tumors, being now a hyperthermal mixed beam containing neutrons of lower energy getting the maximum flux between 0.5 cm and 1.0 cm depth in water. Fig. 1 shows a schematic representation of the port and its main components [3]. Dosimetric distribution of gamma dose has been done in a body-shaped water phantom using TLD 700 distributed in 16 different points of the phantom, simulating several regions of the body. The thermal neutron flux has been estimated with Mn foils [4]. 2.2 Patient positioning The main difference between traditional Radiotherapy and this modality is that the port of the reactor is stationary and cannot be rotated around the patient like the gantry of a typical external beam machine as a linear accelerator or a Co-60 unit. Thus, the positioning of patient is highly complicated, the resulting positions being sometimes a little uncomfortable (Fig. 2). 2
3 Figure 2 a) Positioning scheme b) Irradiation region of patient Nº 6 for patient Nº 6 The patient position is simulated at the Simulation Room (SR) in Buenos Aires using immobilisation devices which are afterwards taken to the Treatment Room (TR) in Bariloche (Fig. 3). The Simulation Room and the Treatment Room both have identical gridded windows. Moreover, the window of the Simulation Room has a beam s-eye view as it is made of transparent acrylic (Fig. 4). All the anatomical marks are photographed at the SR and patient position coordinates registered in a simulation report in order to reproduce the positioning at the TR, where there isn t a beam s eye view. The position is repeated and verified at the TR one day before treatment. Immobilisation is crucial for these treatments as the irradiation times are about 60 minutes. Figure 3: Treatment Room showing the port Figure 4: The beam s eye view at the Simulation Room 3
4 2.3 Patient irradiation and in-vivo dosimetry Patients are infused intravenously over 90 minutes, blood samples are extracted in spans of minutes during this period and until the beginning of irradiation, in order to analyze blood boron concentration. Those values are plotted versus time and fitted with an open two-compartiment pharmaco kinetic model for getting the final irradiation time [5]. After infusion patient is carefully positioned at the TR following the simulation report. Marks are added on the patient for monitoring patient movement from outside through a video camera during the irradiation (Fig. 5). Figure 5: Record of patient position at the beginning and end of treatment. LiF (TLD) is commonly used for in vivo dosimetry in other radiotherapy modalities. Furthermore, the use of TLD-700 in a mixed neutron-photon beam is reported and their accuracy is well determined [6],[7]. For each irradiation field, between 5 and 7 TLD s 700 are distributed over the patient, outside the irradiation area (Fig. 6) in order to estimate the gamma dose received by healthy tissues. The position of TLD s is chosen according to the localisation of each tumor and the type of organs at risk (OR) at the surroundings of the irradiated region. It is worth while to emphasize that staff remains in a controlled area during the irradiation and they don t enter to the TR till the dose rate is below 0.20 msv/h. Dose values read from personal dosimeters have been always below 0.15 msv. 4
5 Figure 6: TLD localisation for patient Nº3 for fields 3 (Fig.6 a), 2 (Fig. 6b) and 1 (Fig. 6c) a) b) TLD c) TLD As an example, we illustrate the particular case of patient Nº 3 whose schedule included 3 irradiation fields. The treated regions (the calf and the heel of right leg and the right foot sole) and the TLD s positions for each field are shown in Fig. 6. For one patient a protective borated polyethylene block was needed in order to avoid high neutron doses to parts of the body which were at the same level of the port (Fig.7). 5
6 Fig. 7: Patient Nº 2 with the shielding block Shielding block 3. Results An excellent reproduction of patient positioning is achieved and the monitoring showed that patient movement provoked a shift of only 2 to 4 mm of the irradiated region. This shift produces an estimated change in thermal neutron flux of 5%. The dose values got from the TLD s for the patient taken as example are shown in Table 1. As it was previously stated, the thermal neutron dose is reported to be 10% [4] to 20% [7], and these values are comparable to the uncertainty in the estimation of gamma dose. The contribution from thermal neutron field to TLD s is included and so, the values in Table 1 give higher dose levels than the actual absorbed gamma dose in inner organs at risk by a factor of 2.3 [4], so they can be considered a conservative estimate of dose. Measurements performed by Blaumann in the anthropomorphic water filled acrylic phantom [4] showed that outside the irradiated area, neutron dose was of the same order of magnitude than the uncertainty in the determination of gamma dose, thus it can be neglected. Based on these results, it was considered that doses in organs at risk outside the irradiated zone were due mainly to gamma rays, and TLD 700 were used to estimate it. The OR for all cases were both lungs and bladder and TLD s have been placed near them. The rest of TLD s locations have been chosen according to their proximity to the port. In patient 3, left knee and abdomen were near the port for all fields, so TLD s have been positioned there. For field 1 (right calf) the left groin was measured because it was near to the ovaries, for field 2 (right heel), the forehead and the eyes because they were facing the port, and for field 3 (right foot sole) again because the forehead was facing the port and the left foot was near the port. For all patients, doses have been well below 1Gy. This result also includes the estimations for the patient who had the protective block. 6
7 Table 1: Gamma dose from TLDs vs. body region for the field Nº 1 of patient Nº 3 Field Localisation Doses [cgy] Left groin Right Calf 2 Right Heel 3 Right Foot Sole Abdomen 8.6 Left knee 16.1 Right lung 7.9 Left lung 3.8 Bladder 11.4 Forehead 7.2 Abdomen 6.9 Left knee 10.7 Right lung 6.3 Left lung 6.2 Bladder 6.2 Forehead 7.7 Abdomen 9.4 Left knee 15.7 Right lung 8.3 Left lung 8.1 Bladder 10.3 Left foot Conclusion In spite of the uncomfortable position, the movement of the irradiated zone during irradiation was only of 2 to 4 mm and it is perfectly acceptable as it is the similar tolerable displacement in others radiotherapy modalities. The beam port shielding proved to be effective for patient radiation protection according to the results of the TLD s doses. Maximum dose received outside the radiation field in healthy tissues was well below tolerance dose for the sites measured. TLD s proved to be a useful tool to estimate doses in OR for this kind of treatments. The video system to monitor patietnts during irradiation is useful to determine their movements, and also provides sufficient data to calculate the retrospective dose distribution when needed. The general BNCT procedure resulted safe from the radiation protection point of view. REFERENCES [1] PAWLIK, T.M., SONDAK, V.K. Malignant melanoma: current state of primary and adjuvant treatment Crit. Rev. Oncol. Hemat. (2003) 45(3), [2] ROTH, B.M, et al. BNCT Clinical trials of Skin Melanoma Patients in Argentina 1st Congress of the Latin American of the Oncological Radiation Therapy Association, Montevideo, Uruguay (2007) [3] BLAUMANN, H.R., et al. Boron neutron capture therapy of skin melanomas at the RA-6 reactor. A procedural approach to beam set up and performance evaluation for upcoming clinical trials, Med. Phys. (2004) 31(1), [4] BLAUMANN, H., Evaluación de la distribución de dosis gamma en un fantoma cuerpo entero en la facilidad de BNCT del reactor RA-6, Informe técnico CNEA-CAB 47/014/02 (in Spanish) 7
8 [5] GONZÁLEZ, S.J., et al. First BNCT treatment of a skin melanoma in Argentina: Dosimetric Analysis and clinical outcome, Applied Radiation and Isotopes, (2004) 61, [6] KRY, S. F. et al: The use of LiF(TLD-100) as an out-of-field dosimeter, J. Appl. Clinical Medical Physics, (2007), 8 (4) [7] CARITA ASCHAN, Applicability of thermoluminiscent dosimeters in X-Ray organ dose determination and in the dosimetry of systemic and boron neutron capture radiotherapy University of Helsinki, 1999, Report Series in Physics, HU-P-D77, ISSN , ISBN
Single Photon Emission Tomography Approach for Online Patient Dose Assessment in Boron Neutron Capture Therapy
Single Photon Emission Tomography Approach for Online Patient Dose Assessment in Boron Neutron Capture Therapy Daniel M Minsky a,b,c*, Alejandro Valda a,b, Andrés J Kreiner a,b,c and Alejandro A Burlon
More informationCOMPARING THE RESPONSES OF TLD 100, TLD 600, TLD 700 AND TLD 400 IN MIXED NEUTRON-GAMMA FIELDS
2015 International Nuclear Atlantic Conference - INAC 2015 São Paulo, SP, Brazil, October 4-9, 2015 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-06-9 COMPARING THE RESPONSES OF TLD
More informationProgress in Reactor and Accelerator Based BNCT at Kyoto University Research Reactor Institute
Progress in Reactor and Accelerator Based BNCT at Kyoto University Research Reactor Institute Yoshinori Sakurai 1 Kyoto University Research Reactor Institute Asashiro-nishi 2-1010, Kumatori-cho, Sennan-gun,
More informationToday, I will present the second of the two lectures on neutron interactions.
Today, I will present the second of the two lectures on neutron interactions. 1 The main goal of this lecture is to tell you a little about clinical neutron therapy, first with fast neutron beams, and
More informationARTICLE IN PRESS. Applied Radiation and Isotopes
Applied Radiation and Isotopes 67 (2009) S54 S58 Contents lists available at ScienceDirect Applied Radiation and Isotopes journal homepage: www.elsevier.com/locate/apradiso Dynamic infrared imaging of
More informationNeutron - a tool in the cancer treatment - C. Paunoiu
Neutron - a tool in the cancer treatment - C. Paunoiu 146 5/17/2009 The neutron -a a tool in the cancer treatment Dr.Constantin PĂUNOIU, constantin.paunoiu@nuclear.ro http://www.nuclear.ro biological material
More informationEpithermal neutron beams from the 7 Li(p,n) reaction near the threshold for neutron capture therapy
IL NUOVO CIMENTO 38 C (2015) 179 DOI 10.1393/ncc/i2015-15179-9 Colloquia: UCANS-V Epithermal neutron beams from the 7 Li(p,n) reaction near the threshold for neutron capture therapy I. Porras( 1 ),J.Praena(
More informationPreliminary study of MAGAT polymer gel dosimetry for boron-neutron capture therapy
Journal of Physics: Conference Series OPEN ACCESS Preliminary study of MAGAT polymer gel dosimetry for boron-neutron capture therapy To cite this article: Shin-ichiro Hayashi et al 2015 J. Phys.: Conf.
More informationTHERMOLUMINESCENT (TL) DOSIMETRY OF SLOW-NEUTRON FIELDS AT RADIOTHERAPY DOSE LEVEL
THERMOLUMINESCENT (TL) DOSIMETRY OF SLOW-NEUTRON FIELDS AT RADIOTHERAPY DOSE LEVEL G. Gambarini Dipartimento di Fisica dell Università, Milano, Italy e-mail grazia.gambarini http://users.unimi.it/~frixy/
More informationNon-target dose from radiotherapy: Magnitude, Evaluation, and Impact. Stephen F. Kry, Ph.D., D.ABR.
Non-target dose from radiotherapy: Magnitude, Evaluation, and Impact Stephen F. Kry, Ph.D., D.ABR. Goals Compare out-of-field doses from various techniques Methods to reduce out-of-field doses Impact of
More informationNeutrons. ρ σ. where. Neutrons act like photons in the sense that they are attenuated as. Unlike photons, neutrons interact via the strong interaction
Neutrons Neutrons act like photons in the sense that they are attenuated as I = I 0 e μx where Unlike photons, neutrons interact via the strong interaction μ = The cross sections are much smaller than
More informationTHE HIGH-CURRENT DEUTERON ACCELERATOR FOR THE NEUTRON THERAPY. S.V. Akulinichev, A. V. Andreev, V.M. Skorkin
THE HIGH-CURRENT DEUTERON ACCELERATOR FOR THE NEUTRON THERAPY S.V. Akulinichev, A. V. Andreev, V.M. Skorkin Institute for Nuclear Research of the RAS, Russia THE PROJECT OF NEUTRON SOURCES FOR THE NEUTRON
More informationUse of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments
Use of Bubble Detectors to Characterize Neutron Dose Distribution in a Radiotherapy Treatment Room used for IMRT treatments Alana Hudson *1 1 Tom Baker Cancer Centre, Department of Medical Physics, 1331
More informationNeutron Interactions Part 2. Neutron shielding. Neutron shielding. George Starkschall, Ph.D. Department of Radiation Physics
Neutron Interactions Part 2 George Starkschall, Ph.D. Department of Radiation Physics Neutron shielding Fast neutrons Slow down rapidly by scatter in hydrogenous materials, e.g., polyethylene, paraffin,
More informationDesign and Performance Of A Thermal Neutron Beam for Boron Neutron Capture Therapy At The University Of Missouri Research Reactor
Design and Performance Of A Thermal Neutron Beam for Boron Neutron Capture Therapy At The University Of Missouri Research Reactor J.D. Brockman J.C. McKibben In situ activation reaction, 10 B(n, a) 7 Li;
More informationPhD értekezés tézisei PHASE I CLINICAL STUDY ON BORON NEUTRON CAPTURE THERAPY. Dr Katalin Hideghéty. A Doktori Iskola vezetıje Prof. Dr.
PhD értekezés tézisei PHASE I CLINICAL STUDY ON BORON NEUTRON CAPTURE THERAPY Dr Katalin Hideghéty A Doktori Iskola vezetıje Prof. Dr. Nagy Judit témavezetı Prof. Dr. Ember István Pécsi Tudományegyetem
More informationChapter Introduction
Chapter 1 Introduction Malignancy is a life-threatening condition in which tumor cells are of a highly invasive character and penetrate normal organs and exhibit an obstinate resistance against cancer
More informationRPI DOSIMETRY. Therefore the description of the work is grouped in two main areas:
RPI DOSIMETRY Introduction The main objective of the RPI Dosimetry is the characterization of the radiation field of the facilities available in the reactor. Most of the use of the reactor centers in irradiation
More informationRadiation Dosimetry at the BNL High Flux Beam Reactor and Medical Research Reactor
BNL-66807 1 1 1 1 2 Holden, N.E., Hu, J-P., Greenberg, D.D., Reciniello, R.N., Farrell, K. and Greenwood, L.R. Radiation Dosimetry at the BNL High Flux Beam Reactor and Medical Research Reactor Reference:Holden,N.E.,
More informationFETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS
FETUS ABSORBED DOSE EVALUATION IN HEAD AND NECK RADIOTHERAPY PROCEDURES OF PREGNANT PATIENTS Etieli Camargo da Costa 1, Luiz Antonio Ribeiro da Rosa 2 Delano Valdivino Santos Batista 3 1 Instituto de Radioproteção
More informationA Feasibility Study of the SLOWPOKE-2 Reactor as a Neutron Source for Boron Neutron Cancer Treatment
A Feasibility Study of the SLOWPOKE- Reactor as a Neutron Source for Boron Neutron Cancer Treatment Introduction M.J. McCall, M. Pierre Royal Military College of Canada, Kingston, Ontario, Canada K7K 7B4
More informationCalibration of two 90 Sr+ 90 Y dermatological applicators
Calibration of two 90 Sr+ 90 Y dermatological applicators Patrícia L. Antonio *, Linda V. E. Caldas Instituto de Pesquisas Energéticas e Nucleares, IPEN-CNEN/SP Av. Prof. Lineu Prestes 2242, 05508-000,
More informationProtection of the contralateral breast during radiation therapy for breast cancer
Protection of the contralateral breast during radiation therapy for breast cancer Edgardo Garrigó a*, Alejandro Germanier b, Silvia Zunino a a Instituto Privado de Radioterapia, Ob Oro 423 (5000) Córdoba,
More information6/29/2012 WHAT IS IN THIS PRESENTATION? MANAGEMENT OF PRIMARY DEVICES INVESTIGATED MAJOR ISSUES WITH CARDIAC DEVICES AND FROM MED PHYS LISTSERVS
6/29/2012 MANAGEMENT OF RADIOTHERAPY PATIENTS WITH IMPLANTED CARDIAC DEVICES Dimitris Mihailidis, PhD., Charleston Radiation Therapy Consultants Charleston, WV 25304 WHAT IS IN THIS PRESENTATION? Types
More informationResponse evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry
Response evaluation of CaSO4:Dy; LiF:Mg,Ti and LiF:Mg,Ti microdosimeters using liquid water phantom for clinical photon beams dosimetry Luciana C. Matsushima 1, Glauco R. Veneziani 1, Roberto K. Sakuraba
More informationROPES eye plaque dosimetry: commissioning and verification of an ophthalmic brachytherapy treatment planning system
University of Wollongong Research Online Faculty of Engineering and Information Sciences - Papers: Part A Faculty of Engineering and Information Sciences 2013 ROPES eye plaque dosimetry: commissioning
More informationOptimization of an accelerator-based epithermal neutron source for neutron capture therapy
Applied Radiation and Isotopes 61 (2004) 1009 1013 Optimization of an accelerator-based epithermal neutron source for neutron capture therapy O.E. Kononov a, *, V.N. Kononov a, M.V. Bokhovko a, V.V. Korobeynikov
More informationX-RAYS INDIVIDUAL DOSE ASSESSMENT USING TLD DOSIMETERS
X-RAYS INDIVIDUAL DOSE ASSESSMENT USING TLD DOSIMETERS Carlos Salas Nucleoeléctrica Argentina SA - Embalse NPP Environmental Control and Dosimetry Division INTRODUCTION: This paper describes the methodology
More informationPeak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam
Peak temperature ratio of TLD glow curves to investigate the spatial variation of LET in a clinical proton beam University of Chicago CDH Proton Center LET study C. Reft 1, H. Ramirez 2 and M. Pankuch
More informationNeutron sources from Nuclear Reactors
Sorgenti di neutroni e loro applicazioni in ambito INFN Neutron sources from Nuclear Reactors Alfonso Santagata alfonso.santagata@enea.it ENEA * Via Anguillarese, 301 00123 Rome Italy * Agezia nazionale
More informationMODELLING A GAMMA IRRADIATION PROCESS USING THE MONTE CARLO METHOD
2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte, MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 MODELLING A GAMMA IRRADIATION
More informationFirst application of hemi-body electron beam irradiation for Kaposi sarcoma at the lower extremities
JBUON 2018; 23(1): 268-272 ISSN: 1107-0625, online ISSN: 2241-6293 www.jbuon.com E-mail: editorial_office@jbuon.com SHORT COMMUNICATION First application of hemi-body electron beam irradiation for Kaposi
More informationThe Therapeutic Ratio in BNCT: Assessment using the Rat 9L Gliosarcoma Brain Tumor and Spinal Cord Models
BNL-63408 The Therapeutic Ratio in BNCT: Assessment using the Rat 9L Gliosarcoma Brain Tumor and Spinal Cord Models J.A. Coderre, G.M. Morris 1, P.L. Micca, M.M. Nawrocky, CD. Fisher, A. Bywaters and J;W.
More informationRPC s Credentialing Programs for Clinical Trials
RPC s Credentialing Programs for Clinical Trials July 19, 2010 Geoffrey S. Ibbott, Ph.D. and RPC Staff Mission The mission of the Radiological Physics Center is to assure NCI and the Cooperative Groups
More informationDesign of a BSA for Producing Epithermal Neutron for BNCT
Siam Physics Congress 2015 Design of a BSA for Producing Epithermal Neutron for BNCT M Asnal *, T Liamsuwan 2 and T Onjun 1 1 Sirindhorn International Institute of Technology, Thammasat University 2 Nuclear
More informationOption D: Medicinal Chemistry
Option D: Medicinal Chemistry Basics - unstable radioactive nuclei emit radiation in the form of smaller particles alpha, beta, positron, proton, neutron, & gamma are all used in nuclear medicine unstable
More informationCALIBRATION OF A TLD ALBEDO INDIVIDUAL NEUTRON MONITOR
2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 CALIBRATION OF A TLD
More informationShielding Design Basis and its Calculation of High Energy Medical Linac Installed in Bangladesh Atomic Energy Commission, Bangladesh
Universal Journal of Medical Science 5(2): 27-31, 2017 DOI: 10.13189/ujmsj.2017.050202 http://www.hrpub.org Shielding Design Basis and its Calculation of High Energy Medical Linac Installed in Bangladesh
More informationNeutron dose evaluation in radiotherapy
Neutron dose evaluation in radiotherapy Francesco d Errico University of Pisa, Italy Yale University, USA Radiation therapy with a linear accelerator (LINAC) Photoneutron production in accelerator head
More informationDiode calibration for dose determination in total body irradiation
Iran. J. Radiat. Res., 2008; 6 (1): 43-50 Diode calibration for dose determination in total body irradiation M.Allahverdi 1*, Gh. Geraily 1, M. Esfehani 3,A. Sharafi 2,A. Shirazi 1 1 Department of Medical
More informationRadiation qualities in carbon-ion radiotherapy at NIRS/HIMAC
Radiation qualities in carbon-ion radiotherapy at NIRS/ Shunsuke YONAI Radiological Protection Section Research Center for Charged Particle Therapy National Institute of Radiological Sciences (NIRS) E-mail:
More informationDistribution of Water Phantom BNCT Kartini Research Reactor Based Using PHITS
Indonesian Journal of Physics and Nuclear Applications Volume 3, Number 2, June 2018, p. 43-48 e-issn 2550-0570, FSM UKSW Publication Distribution of Water Phantom BNCT Kartini Research Reactor Based Using
More informationSHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES
SHIELDING TECHNIQUES FOR CURRENT RADIATION THERAPY MODALITIES MELISSA C. MARTIN, M.S., FACR, FAAPM PRESIDENT AAPM - 2017 PRESIDENT - THERAPY PHYSICS INC., GARDENA, CA MELISSA@THERAPYPHYSICS.COM AAPM Spring
More informationSpatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, Pamela Myers, Ph.D.
Spatially Fractionated Radiation Therapy: GRID Sponsored by.decimal Friday, August 22, 2014 Pamela Myers, Ph.D. Introduction o o o o o Outline GRID compensator Purpose of SFRT/GRID therapy Fractionation
More informationStudy on Microdosimetry for Boron Neutron Capture Therapy
Progress in NUCLEAR SCIENCE and TECHNOLOGY, Vol. 2, pp.242-246 (2011) ARTICLE Study on Microdosimetry for Boron Neutron Capture Therapy Tetsuya MUKAWA 1,*, Tetsuo MATSUMOTO 1 and Koji NIITA 2 1 Tokyo City
More informationRadiologic Units: What You Need to Know
Radiologic Units: What You Need to Know TODD VAN AUKEN M.ED. RT (R)(MR) Agenda Greys, Sieverts, Coulombs per kg, & Becquerel's Conventional Units Other Concepts (LET, Q-Factor, Effective Dose, NCRP Report
More informationBoron Neutron Capture Therapy (BNCT) - Low-Energy Neutron Spectrometer for Neutron Field Characterization - )
Boron Neutron Capture Therapy (BNCT) - Low-Energy Neutron Spectrometer for Neutron Field Characterization - ) Isao MURATA and Tsubasa OBATA Division of Electrical, Electronic and Information Engineering,
More informationDjoko S. Pudjorahardjo, Widarto, Isman Mulyadi T
Indonesian Journal of Physics and NuclearApplications Volume 3, Number 1, February 2018, p. 15-20 e-issn 2550-0570, FSM UKSWPublication 2549-046X, FSM UKSWPublication Detail Engineering Design of Compact
More informationNeutron Radiotherapy: Past, Present, and Future Directions
Neutron Radiotherapy: Past, Present, and Future Directions Theodore L. Phillips Lecture -- 2014 George E. Laramore Ph.D., M.D. NONE Conflicts of Interest Peter Wootton Professor of Radiation Oncology University
More informationOut-of-field dosimetry in radiotherapy for input to epidemiological studies. Roger Harrison
MELODI 7th Workshop, 9 11 November 2015 Helmholtz Zentrum München Next Generation Radiation Protection Research Out-of-field dosimetry in radiotherapy for input to epidemiological studies Roger Harrison
More informationA TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM *
Romanian Reports in Physics, Vol. 66, No. 2, P. 394 400, 2014 A TREATMENT PLANNING STUDY COMPARING VMAT WITH 3D CONFORMAL RADIOTHERAPY FOR PROSTATE CANCER USING PINNACLE PLANNING SYSTEM * D. ADAM 1,2,
More information22.55 Principles of Radiation Interactions Name: Spring 2004 Professor Coderre Exam 2 April 30, 2004
22.55 Principles of Radiation Interactions Name: Spring 2004 Professor Coderre Exam 2 April 30, 2004 You have 2 hours to complete this exam. This exam is closed book. Please show all work on the attached
More informationApplication of the Multi-Model Monte-Carlo Treatment Planning System Combined with PHITS to Proton Radiotherapy
1 1 6 2 1 2 3 5 6 7 9 1 1 1 1 2 1 3 1 1 5 1 6 1 7 1 1 9 2 2 1 Progress in NUCLEAR SCIENCE and TECHNOLOGY, Vol. 2, pp.213-21 (211) DOI: 1.15669/pnst.2.213 ARTICLE Application of the Multi-Model Monte-Carlo
More informationDosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy
Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti M. Bruzzi,M. Bucciolini, L. Marrazzo, D. Menichelli University of
More informationADVANCES IN RADIATION TECHNOLOGIES IN THE TREATMENT OF CANCER
ADVANCES IN RADIATION TECHNOLOGIES IN THE TREATMENT OF CANCER Bro. Dr. Collie Miller IARC/WHO Based on trends in the incidence of cancer, the International Agency for Research on Cancer (IARC) and WHO
More informationV. 4. Design and Benchmark Experiment for Cyclotron-based Neutron Source for BNCT
CYRIC Annual Report 2002 V. 4. Design and Benchmark Experiment for Cyclotron-based Neutron Source for BNCT Yonai S., ItogaT., Nakamura T., Baba M., Yashima, H. Yokobori H. *, and Tahara Y. ** Cyclotron
More informationIntroduction. Measurement of Secondary Radiation for Electron and Proton Accelerators. Introduction - Photons. Introduction - Neutrons.
Measurement of Secondary Radiation for Electron and Proton Accelerators D. Followill, Ph.D. Radiological Physics Center U. T. M. D. Anderson Cancer Center Introduction Patients undergoing radiation therapy
More informationThe determination of timer error and its role in the administration of specified doses
JURNAL FIZIK MALAYSIA VOLUME 29, NUMBER 1&2 2008 The determination of timer error and its role in the administration of specified doses S. B. Samat a *, C. J. Evans b, T. Kadni c and M. T. Dolah c a School
More informationQA for Clinical Dosimetry with Emphasis on Clinical Trials
QA for Clinical Dosimetry with Emphasis on Clinical Trials Geoffrey S. Ibbott, Ph.D. and RPC Staff G. Ibbott, AAPM Summer School, June 24, 2009 1 QA Infrastructure for Clinical Trials Participating Institutions
More informationChapters from Clinical Oncology
Chapters from Clinical Oncology Lecture notes University of Szeged Faculty of Medicine Department of Oncotherapy 2012. 1 RADIOTHERAPY Technical aspects Dr. Elemér Szil Introduction There are three possibilities
More informationCharacterization of OSL response of LiF:Mg,Ti and microlif:mg, Ti to 60 Co gamma source
Characterization of OSL response of LiF:Mg,Ti and microlif:mg, Ti to 60 Co gamma source Amanda Bravim, Leticia Lucente Campos Instituto de Pesquisas Energéticas e Nucleares (IPEN / CNEN - SP) Av. Professor
More informationCHARACTERIZATION OF TLD-100 DOSIMETER FOR MONITORING THE EXTREMITIES OF WORKERS IN NUCLEAR MEDICINE
2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 CHARACTERIZATION OF TLD-100
More informationWhat Can Go Wrong in Radiation Treatment: Data from the RPC. Geoffrey S. Ibbott, Ph.D. and RPC Staff
What Can Go Wrong in Radiation Treatment: Data from the RPC Geoffrey S. Ibbott, Ph.D. and RPC Staff Clinical NCI Trials CALGB 1768 Participating Institutions NCCTG ECOG COG ACOSOG SWOG QARC RPC ATC RTOG
More informationDose Homogeneity of the Total Body Irradiation in vivo and in vitro confirmed with Thermoluminescent Dosimeter
Dose Homogeneity of the Total Body Irradiation in vivo and in vitro confirmed with Thermoluminescent Dosimeter E.K. Chie M.D. 1, S.W. Park M.D. 1, W.S. Kang PhD. 1,2, I.H. Kim M.D. 1,2, S.W. Ha M.D. 1,2,
More informationRadiation Monitoring Instruments
Radiation Monitoring Instruments This set of slides is based on Chapter 4 authored byg. Rajan, J. Izewska of the IAEA publication (ISBN 92-0-107304-6): Radiation Oncology Physics: A Handbook for Teachers
More informationAssessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine
2017 IJSRST Volume 3 Issue 3 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Assessment of Dosimetric Functions of An Equinox 100 Telecobalt Machine Samuel Nii Adu Tagoe
More informationIAEA-CN THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS*
THE ESTRO-EQUAL RESULTS FOR PHOTON AND ELECTRON BEAMS CHECKS IN EUROPEAN RADIOTHERAPY BEAMS* H. Ferreira, A. Dutreix, A. Bridier, D. Marre, J. Chavaudra, H. Svensson ESTRO-EQUAL Measuring Laboratory, Service
More informationMONITORING OF OCCUPATIONAL EXPOSURE AT NUCLEAR FACILITIES
GUIDE YVL 7.10 / 29 JANUARY 2002 MONITORING OF OCCUPATIONAL EXPOSURE AT NUCLEAR FACILITIES 1 GENERAL 3 2 PROVISIONS OF THE RADIATION ACT AND DECREE 3 3 MONITORING OF RADIATION EXPOSURE 4 3.1 General requirements
More informationMedical Use of Radioisotopes
Medical Use of Radioisotopes Therapy Radioisotopes prove to be useful in the application of brachytherapy, the procedure for using temporary irradiation close to the area of disease (i.e. cancer) 10% Medical
More informationStudy of neutron-dna interaction at the IPEN BNCT Research facility ABSTRACT
Study of neutron-dna interaction at the IPEN BNCT Research facility Maritza Rodriguez Gual 1, Oscar Rodriguez Hoyos 1, Fernando Guzman Martinez 1, Airton Deppman 2, J.D.T.Aruda Neto 2, V.P.Likhachev 2,
More informationTLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience
IAEA-CN-96-82 TLD as a tool for remote verification of output for radiotherapy beams: 25 years of experience J. Francisco Aguirre, Ramesh C. Tailor, Geoffrey S. Ibbott, Marilyn Stovall and William F. Hanson
More informationCharacterization and implementation of Pencil Beam Scanning proton therapy techniques: from spot scanning to continuous scanning
Characterization and implementation of Pencil Beam Scanning proton therapy techniques: from spot scanning to continuous scanning Supervisors Prof. V. Patera PhD R. Van Roermund Candidate Annalisa Patriarca
More informationnuclear science and technology
EUROPEAN COMMISSION nuclear science and technology Evaluation of Individual Dosimetry in Mixed Neutron and Photon Radiation Fields (EVIDOS) Editor: Helmut Schuhmacher, Physikalisch-Technische Bundesanstalt
More informationNeutron-Gamma Mixed field Dosimetry on a Child phantom under Therapeutic Proton Irradiation using TL Dosimeters
133 Clinical Radiation Oncology Neutron-Gamma Mixed field Dosimetry on a Child phantom under Therapeutic Proton Irradiation using TL Dosimeters Bhaskar Mukherjee 1,2, Carolina Lina 1 Vladimir Mares 3 1Westdeutsches
More informationPET Guidance of Therapy for BNCT and in vivo B-10 imaging
INFN LNL Legnaro 17-19 Novembre 2009 Principles of Positron Emission Tomography and Radiopharmaceuticals PET Guidance of Therapy for BNCT and in vivo B-10 imaging Luca Menichetti, Ph.D C.N.R. Institute
More informationRadiation physics and radiation protection. University of Szeged Department of Nuclear Medicine
Radiation physics and radiation protection University of Szeged Department of Nuclear Medicine Radiation doses to the population 1 Radiation doses to the population 2 Sources of radiation 1 Radiation we
More informationPhysical Bases : Which Isotopes?
Physical Bases : Which Isotopes? S. Gnesin Institute of Radiation Physics, Lausanne University Hospital, Lausanne, Switzerland 1/53 Theranostic Bruxelles, 2 Octobrer 2017 Theranostic : use of diagnostic
More informationD DAVID PUBLISHING. Uncertainties of in vivo Dosimetry Using Semiconductors. I. Introduction. 2. Methodology
Journal of Life Sciences 9 (2015) 120-126 doi: 10.17265/1934-7391/2015.03.005 D DAVID PUBLISHING Uncertainties of in vivo Dosimetry Using Semiconductors Zeina Al Kattar, Hanna El Balaa and Saeed Zahran
More informationDosimetric characteristics of 137 Cs sources used in after loading Selectron system by Monte Carlo method
Iran. J. Radiat. Res., 2007; 5 (3): 147-152 Dosimetric characteristics of Cs sources used in after loading Selectron system by Monte Carlo method M.B.Tavakoli, D. Shahbazi-Gahrouei *, M. Hosseinpour Department
More informationThe physics of boron neutron capture therapy: an emerging and innovative treatment for glioblastoma and melanoma
Physics and Astronomy Department Physics and Astronomy Comps Papers Carleton College Year 2005 The physics of boron neutron capture therapy: an emerging and innovative treatment for glioblastoma and melanoma
More informationThis article was published in an Elsevier journal. The attached copy is furnished to the author for non-commercial research and education use, including for instruction at the author s institution, sharing
More informationPatient dosimetry for total body irradiation using single-use MOSFET detectors
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 9, NUMBER 4, FALL 2008 Patient dosimetry for total body irradiation using single-use MOSFET detectors Tina Marie Briere, 1 Ramesh Tailor, 1 Naresh Tolani,
More informationTomoTherapy. Michelle Roach CNC Radiation Oncology Liverpool Hospital CNSA. May 2016
TomoTherapy Michelle Roach CNC Radiation Oncology Liverpool Hospital CNSA May 2016 TomoTherapy The Facts Greek Tomo = slice Advanced form of IMRT 3D computerised tomography (CT) imaging immediately prior
More informationTitle. Author(s)Ishikawa, Masayori; Tanaka, Kenichi; Endo, Satrou; H. CitationJournal of Radiation Research, 56(2): Issue Date
Title Application of an ultraminiature thermal neutron mon therapy Author(s)Ishikawa, Masayori; Tanaka, Kenichi; Endo, Satrou; H CitationJournal of Radiation Research, 56(2): 391-396 Issue Date 2015-03
More informationGDR in Radiotherapy Treatment Fields with 18 MV Accelerators
GDR in Radiotherapy Treatment Fields with 18 MV Accelerators R. R. Martín-Landrove 1, J. Dávila 1, H. R. Vega-Carrillo 2, M. T. Barrera 3, A.J. Kreiner 4, F. Pino 3, H. Barros 3, E. D. Greaves 3 and L.
More informationThin Beryllium target for 9 Be(d,n)- driven BNCT
Thin Beryllium target for 9 Be(d,n)- driven BNCT M.E.Capoulat 1-3, D.M.Minsky 1-3, L.Gagetti 1-3, M. Suárez Anzorena 1, M.F.del Grosso 1-2, J.Bergueiro 1, D.Cartelli 1-3, M.Baldo 1, W.Castell 1, J.Gomez
More informationSUMMARY OF PERSONAL DOSIMETRY PRACTICIES IN RCA MEMBER COUNTRIES
A Personal Dosimetry Intercomparison Study in Asian and Pacific Region Hiroyuki MURAKAMI and Fumiaki TAKAHASHI Department of Health Physics, Japan Atomic Energy Research Institute and Richard V. Griffith
More informationRisk of Secondary Fatal Malignancies from Hi-Art Tomotherapy
Risk of Secondary Fatal Malignancies from Hi-Art Tomotherapy Susannah Lazar David Followill, Ph.D. John Gibbons, Ph.D. * Anita Mahajan, M.D. Mohammad Salehpour, Ph.D. Marilyn Stovall, Ph.D. R. Allen White,
More informationI. Equipments for external beam radiotherapy
I. Equipments for external beam radiotherapy 5 linear accelerators (LINACs): Varian TrueBeam 6, 10 & 18 MV photons, 6-18 MeV electrons, image-guided (IGRT) and intensity modulated radiotherapy (IMRT),
More informationBORON NEUTRON CAPTURE THERAPY SETUP FOR A LINEAR ACCELERATOR
BORON NEUTRON CAPTURE THERAPY SETUP FOR A LINEAR ACCELERATOR C. F. CHIOJDEANU, C. PAVEL, F. CONSTANTIN National Institute for Physics and Nuclear Engineering Horia Hulubei, P.O. Box MG-6, RO-077125 Bucharest-Mãgurele,
More informationOutline. Outline 3/30/12. Second Cancers from. Radiotherapy Procedures. Stephen F. Kry, Ph.D., D.ABR.
Second Cancers from Radiotherapy Procedures Stephen F. Kry, Ph.D., D.ABR. Outline Radiation and cancer induction Medically exposed people Estimating risk of second cancers Minimizing the risk Outline Radiation
More informationRadiotherapy. Marta Anguiano Millán. Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada
Departamento de Física Atómica, Molecular y Nuclear Facultad de Ciencias. Universidad de Granada Overview Introduction Overview Introduction Brachytherapy Radioisotopes in contact with the tumor Overview
More informationThe ARN Critical Dosimetry System. Gregori, B.N.; Papadópulos, S.; Cruzate, J.A.; Equillor, H.E. and Kunst, J.J.
The ARN Critical Dosimetry System Gregori, B.N.; Papadópulos, S.; Cruzate, J.A.; Equillor, H.E. and Kunst, J.J. Presentado en: 11 th International Congress on the International Radiation Protection Association.
More informationDosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy
Dosimetric characterization with 62 MeV protons of a silicon segmented detector for 2D dose verifications in radiotherapy C. Talamonti a*, M. Bucciolini a, L. Marrazzo a, D. Menichelli a. a) Department
More informationCOMPARISON OF RADIOBIOLOGICAL EFFECTS OF CARBON IONS TO PROTONS ON A RESISTANT HUMAN MELANOMA CELL LINE
COMPARISON OF RADIOBIOLOGICAL EFFECTS OF CARBON IONS TO PROTONS ON A RESISTANT HUMAN MELANOMA CELL LINE I. Petrovi a, A. Risti -Fira a, L. Kori anac a, J. Požega a, F. Di Rosa b, P. Cirrone b and G. Cuttone
More informationMeasurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients
Measurement of Dose to Implanted Cardiac Devices in Radiotherapy Patients Moyed Miften, PhD Professor and Chief Physicist University of Colorado Chester Reft, PhD Associate Professor University of Chicago
More informationTo Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence
SCIENCE & TECHNOLOGY To Reduce Hot Dose Spots in Craniospinal Irradiation: An IMRT Approach with Matching Beam Divergence Alburuj R. Rahman*, Jian Z. Wang, Dr. Z. Huang, Dr. J. Montebello Department of
More informationDosimetry Optimization System and Integrated Software (DOSIS): a comparison against Fluka code results over a standard phantom
Dosimetry Optimization System and Integrated Software (DOSIS): a comparison against Fluka code results over a standard phantom Institute of Physics E. Gaviola - CONICET & LIIFAMIRX - Laboratorio de Investigación
More informationIntroduction to clinical Radiotherapy
Introduction to clinical Radiotherapy Sarah Brothwood Radiotherapy Practice Educator Rosemere Cancer Centre Sarah.brothwood@lthtr.nhs.uk 01772 522694 Radiotherapy We have been able to see and document
More informationOut-of-field organ doses from therapeutic irradiation during childhood: is there an excess risk for second cancer induction?
Out-of-field organ doses from therapeutic irradiation during childhood: is there an excess risk for second cancer induction? Poster No.: C-0096 Congress: ECR 2012 Type: Scientific Paper Authors: M. Mazonakis,
More information