Chapter -4 GC-MS analysis of hexane and benzene extracts of P. granatum

Size: px
Start display at page:

Download "Chapter -4 GC-MS analysis of hexane and benzene extracts of P. granatum"

Transcription

1 Chapter -4 GC-MS analysis of hexane and benzene extracts of P. granatum If you wish to reach the highest, begin with the lowest

2 4. GC-MS analysis of Hexane and Benzene extracts of P. granatum A. GC-MS of RP-1 Compounds isolated from the hexane fraction of Ethanol extract Methyl heptanes H 3 C Fig Methyl decane H 3 C Fig Methyldec-1-ene Fig Methyl decane H 3 C Fig methylhexane 138 Fig methylhexane

3 H 3 C Fig Heptyl cyclobutane Fig methylpentane H 3 C Fig methylbutan-1-ol H 3 C OH Fig Propylcyclobutane Fig Methyl octanoate O O Fig dec-7-enoic acid O H 3 C Fig.4.12 OH B. GC-MS of RP-2 Compounds isolated from the hexane: benzene fraction of Ethanol extract.

4 1. 2-methylpentan-3-ol OH C H 3 Fig.4.13 CH methyldecane H 3 C Fig methyloctan-1-ol H 3 C OH Fig methyloctane H 3 C Fig ,3-dimethylpentane H 3 C Fig ,5-dimethylhexane or 2,4-dimethylpentane H 3 C H 3 C Fig Octylcyclobutane

5 Fig Methylhexane H 3 C C H 3 Fig Methylheptane H 3 C Fig Pentylcyclobutane Fig Decanoic acid O H 3 C Fig OH C. GC-MS of RP-3 Compounds isolated from the benzene fraction of Ethanol extract. 1. n- Undecane H 3 C Fig Methyl Undecane H 3 C Fig Butylcyclopropane

6 Fig Methyl Pentanol O H Fig Methyl Nonane Fig Methyl Hexane H 3 C Fig Cyclobutane-3-ene-Pentane Fig Methyl Pentane H 3 C Fig ,4 - Dimethyl Pentane

7 H 3 C Fig Methyl Hexane Fig Methyl Hex-2-ene H 3 C Fig Methyl Heptane H 3 C Fig But-2-ene H 3 C Fig Henicos-5-ene H 3 C Fig GAS CHROMATOGRAPHY Gas chromatography produces highly defined separation in organic chemistry, biochemistry, environmental and biological fields. Besides this it finds extensive application for the detection, identification and quantification of fatty acids

8 methyl esters. Among its uses the most common are drug testing and environmental contaminant identification. GC analysis is a common confirmation test. It separates all of the components in a sample and provides a representative spectral output. The technician injects the sample into the injection port of the GC device. The GC instrument vaporizes the sample and then separates and analyzes the various components. Each component ideally produces a specific spectral peak that may be recorded on a paper chart or electronically. The time elapsed between injection and elution is called the "retention time." The retention time can help to differentiate between some compounds. The size of the peaks is proportional to the quantity of the corresponding substances in the specimen analyzed. The peak is measured from the baseline to the tip of the peak. [1] Gas chromatography and mass spectrometry are, in many ways, highly compatible techniques. In both techniques, the sample is in the vapor phase, and both techniques deal with the same amount of sample (typically less than 1 ng). GC can separate volatile and semi-volatile compounds with great resolution, but it cannot identify them. MS can provide detailed structural information on most compounds such that they can be exactly identified, but it cannot readily separate them. Therefore, it was not surprising that the combination of the two techniques was suggested shortly after the development of GC in the mid-1950s. It was in 1950s that the use of a mass spectrometer was was developed as the detector in gas chromatography by Roland Gohlke and Fred McLafferty. [2] Gas chromatography ("GC") and Mass spectrometry ("MS") make an effective combination for chemical analysis. In GC-MS technique, the gas chromatograph is directly coupled to fast speed mass spectrometer. The Gas chromatography produce highly defined separation of complex mixture and mass spectrometer analyses them. [3][4]. PRINCIPLE Chromatography is a very important analytical tool because it allows the chemist to separate components in a mixture for subsequent use or quantification. Most samples that chemists want to analyze are mixtures. The gas chromatograph makes it possible to separate the

9 volatile components of a very small sample and to determine the amount of each component present. The essentials required for the method are an injection port through which samples are loaded, a column on which the components are separated, a regulated flow of a carrier gas (often helium) which carries the sample through the instrument, a detector, and a data processor. In gas chromatography, the temperature of the injection port, column, and detector are controlled by thermostatic heaters. [5] The method was introduced by James and Martin [6] who used a liquid phase of silicone grease with or without 10% stearic acid to separate normal saturated acids up to C 12, which were detected then Cooper and Heywood [7] expanded the method to acids up to C 22. The method has been used for the separation and identification of geometrical and positional isomers [8] [9]. The pioneering work by Thomson, Aston, Dempster and Nier led to the construction of mass spectrometers for precise measurement of ionic masses using magnetic and electrostatic analyzers. Today mass spectrometers are widely used to identify unknown compounds by way of determining their molecular weight and molecular formula at the expense of negligible amount of sample. The mass spectrum of a compound is given in the form of a bar graph representing the abundance of various ions with respect to their mass (m) to charge (z) ratio (m/z). In addition to the molecular ion, which infers the molecular weight of the sample, the other ions present in the mass spectrum are very characteristic of the compound, thus the mass spectrum of a compound becomes the fingerprint of it, in most of the cases [10]. Today, we all know that a single conventional analytical technique often does not suffice to recognize and quantify all analyses of interest in a sample. During the eighties, the use of multidimensional approaches slowly began to make headway to solve many complex problems. The main driving force behind these developments is the determination of trace level concentration of drugs and their metabolites in biomedical, biological, food, agricultural and environmental samples. Now trace-level determinations have becomes 100 fold stringent. GC-MS Instrumentation

10 GC-MS is an integrated composite instrument combining GC, which is excellent in its ability to separate and quantify components with MS, which is excellent in its ability to identify each component. Thus GC-MS is an integrated composite instrument, it consist of 1. Gas chromatograph 2. Interface (connector) 3. Mass spectrometer (MS) 4. Computer Fig The insides of the GC-MS, with the column of the gas chromatograph in the oven on the right.

11 Outline Diagram of Gas Chromatograph and Mass Analyzer This combination of GC-MS consists of determination of a mass spectrum for each chromatographically separated component of a mixture. Samples need not to be isolated or rigorously purified. The exit pressure of the gas chromatograph is approximately atmospheric. On the other hand, the mass spectrometer operates at high vacuum, 10-6 mm Hg. Therefore, the pressure must be reduced during transfer of effluent from the gas chromatograph to the mass spectrometer and for this purpose an interface is introduced [11]. The GC-MS is composed of two major building blocks: the gas chromatograph and the mass spectrometer. The gas chromatograph utilizes a capillary column which depends on the column's dimensions

12 (length, diameter, film thickness) as well as the phase properties (e.g. 5% phenyl polysiloxane). The difference in the chemical properties between different molecules in a mixture will separate the molecules as the sample travels the length of the column. The molecules take different amount of time (called the retention time) to come out of (elute from) the gas chromatograph, and this allows the mass spectrometer downstream to capture, ionize, accelerate, deflect, and detect the ionized molecules separately. The mass spectrometer does this by breaking each molecule into ionized fragments and detecting these fragments using their mass to charge ratio. These two components, used together, allow a much finer degree of substance identification than either unit used separately. It is not possible to make an accurate identification of a particular molecule by gas chromatography or mass spectrometry alone. The mass spectrometry process normally requires a very pure sample while gas chromatography using a traditional detector (e.g. Flame Ionization Detector) detects multiple molecules that happen to take the same amount of time to travel through the column (i.e. have the same retention time) which results in two or more molecules to co-elute. Sometimes two different molecules can also have a similar pattern of ionized fragments in a mass spectrometer (mass spectrum). Combining the two processes makes it extremely unlikely that two different molecules will behave in the same way in both a gas chromatograph and a mass spectrometer. Therefore when an identifying mass spectrum appears at a characteristic retention time in a GC-MS analysis, it typically lends to increased certainty that the analyte of interest is in the sample. Purge and Trap GC-MS For the analysis of volatile compounds a Purge and Trap (P&T) concentrator system may be used to introduce samples. The target analytes are extracted and mixed with water and introduced into an airtight chamber. An inert gas such as Nitrogen (N 2) is bubbled through the water; this is known as purging. The volatile compounds move into the headspace above the water and are drawn along a pressure gradient (caused by the introduction of the purge gas) out of the chamber. The volatile compounds are drawn along a heated line onto

13 a 'trap'. The trap is a column of adsorbent material at ambient temperature that holds the compounds by returning them to the liquid phase. The trap is then heated and the sample compounds are introduced to the GC-MS column via a volatiles interface, which is a split inlet system. P&T GC-MS is particularly suited to volatile organic compounds (VOCs) and BTEX compounds (aromatic compounds associated with petroleum). [12] Types of Mass Spectrometer Detectors The most common type of mass spectrometer (MS) associated with a gas chromatograph (GC) is the quadrupole mass spectrometer, sometimes referred to by the Hewlett-Packard (now Agilent) trade name "Mass Selective Detector" (MSD). Another relatively common detector is the ion trap mass spectrometer. Other detectors are also there such as time of flight (TOF), tandem quadrupoles (MS-MS), or in the case of an ion trap MS n where n indicates the number mass spectrometry stages. Analysis A mass spectrometer is typically utilized in one of two ways: Full Scan or Selective Ion Monitoring (SIM). The typical GC/MS instrument is capable of performing both functions either individually or concomitantly, depending on the setup of the particular instrument. Full scan MS When collecting data in the full scan mode, a target range of mass fragments is determined and put into the instrument's method. An example of a typical broad range of mass fragments to monitor would be m/z 50 to m/z 400. The determination of what range to use is largely dictated by what one anticipates being in the sample while being cognizant of the solvent and other possible interferences. A MS should not be set to look for mass fragments too low or else one may detect air (found as m/z 28 due to nitrogen), carbon dioxide (m/z 44) or other possible interferences. Additionally if one is to use a large scan range then sensitivity of the instrument is decreased due to performing fewer scans per second since each scan will have to detect a wide range of mass fragments.

14 Full scan is useful in determining unknown compounds in a sample. It provides more information than SIM when it comes to confirming or resolving compounds in a sample. During instrument method development it may be common to first analyze test solutions in full scan mode to determine the retention time and the mass fragment fingerprint before moving to a SIM instrument method. Selected ion monitoring In selected ion monitoring (SIM) certain ion fragments are entered into the instrument method and only those mass fragments are detected by the mass spectrometer. The advantages of SIM are that the detection limit is lower since the instrument is only looking at a small number of fragments (e.g. three fragments) during each scan. More scans can take place each second. Since only a few mass fragments of interest are being monitored, matrix interferences are typically lower. To additionally confirm the likelihood of a potentially positive result, it is relatively important to be sure that the ion ratios of the various mass fragments are comparable to a known reference standard. Types of Ionization After the molecules travel the length of the column, pass through the transfer line and enter into the mass spectrometer they are ionized by various methods with typically only one method being used at any given time. Once the sample is fragmented it will then be detected, usually by an electron multiplier diode, which essentially turns the ionized mass fragment into an electrical signal that is then detected. The ionization technique chosen is independent of using Full Scan or SIM. Electron Ionization By far the most common and perhaps standard form of ionization is electron ionization (EI). The molecules enter into the MS (the source is a quadrupole or the ion trap itself in an ion trap MS) where they are bombarded with free electrons emitted from a filament, not much unlike the filament one would find in a standard light bulb. The electrons bombard the molecules, causing the molecule to fragment in a characteristic and reproducible way. This "hard

15 ionization" technique results in the creation of more fragments of low mass to charge ratio (m/z) and few, if any, molecules approaching the molecular mass unit. Hard ionization is considered by mass spectroscopists as the employ of molecular electron bombardment, whereas "soft ionization" is charge by molecular collision with an introduced gas. The molecular fragmentation pattern is dependent upon the electron energy applied to the system, typically 70eV (electron Volts). The use of 70eV facilitates comparison of generated spectra with National Institute of Standard (NIST-USA) library of spectra applying algorithmic matching programs and the use of methods of analysis written by many method standardization agencies. Chemical Ionization In chemical ionization a reagent gas, typically methane or ammonia is introduced into the mass spectrometer. Depending on the technique (positive CI or negative CI) chosen, this reagent gas will interact with the electrons and analyte and cause a 'soft' ionization of the molecule of interest. A softer ionization fragments the molecule to a lower degree than the hard ionization of EI. One of the main benefits of using chemical ionization is that a mass fragment closely corresponding to the molecular weight of the analyte of interest is produced. Positive Chemical Ionization In Positive Chemical Ionization (PCI) the reagent gas interacts with the target molecule, most often with a proton exchange. This produces the species in relatively high amounts. Negative Chemical Ionization In Negative Chemical Ionization (NCI) the reagent gas decreases the impact of the free electrons on the target analyte. This decreased energy typically leaves the fragment in great supply. Interface (direct coupled) - When the inside diameter of the capillary column is 0.25 mm or 0.32 mm; the column is directly connected with the ion source. If

16 the carrier gas flow is less than 2 ml/min, a turbo molecular pump is satisfactory as the vacuum pump. Jet separator A carrier gas jet separator is generally used in GC-MS. Its function is to remove carrier gas completely from the GC effluent to prevent losses in mass spectral resolution due to pressure broadening from ion - molecule collisions. The principle of the jet separator is the preferential diffusion of lighter carrier gas molecule away from the GC effluent stream as it emerges at high velocity from a narrow aerator. The main functions of molecular jet separator are: 1. It ideally passes the entire sample without decomposition, while the entire carrier gas is removed. 2. It increases the sample to carrier gas ratio, when its input and output concentrations are compared. 3. Reduces the outlet pressure from GC column to a pressure suitable for the ion source. The carrier gas is almost always a small molecule such as helium or hydrogen with a high diffusion coefficient, whereas the organic molecules have much lower diffusion coefficients. In operation, the GC effluent (the carrier gas with the organic analytes) is sprayed through a small nozzle, To avoid the problem thermo stable liquid phase of concentration 0.5 to 5% is used. The use of computer system in combination with GC-MS instrument is very important [13-14]. It facilitates the detection and identification of the compound present in minor quantity. The computer must have large care storage and be able to process the data at high speed. The main advantages in computerization of GC-MS are as follows: 1. The establishment of the automatic mass scale with the help of a reference compound or sensor measuring the magnetic field. 2. Collection of output data at faster rate on the GC-MS instrument. 3. The background spectra can be subtracted from the actual spectra for i.e. sample bleed from sample.

17 4. Presentation of mass spectra in normalized and is plotted as a bar graph just after the sample has been injected into the column. General Uses of GC-MS Identification and quantitation of volatile and semi-volatile organic compounds in complex mixtures Determination of molecular weights and (sometimes) elemental compositions of unknown organic compounds in complex mixtures Structural determination of unknown organic compounds in complex mixtures both by matching their spectra with reference spectra and by a priori spectral interpretation Common Applications Quantitation of pollutants in drinking and wastewater using official U.S. Environmental ProtectionAgency (EPA) methods Quantitation of drugs and their metabolites in blood and urine for both pharmacological and forensic applications Identification of unknown organic compounds in hazardous waste dumps Identification of reaction products by synthetic organic chemists Analysis of industrial products for quality control. [15-17] GC-MS Studies The Hexane, Hexane benzene fractions, were examined in GC-MS for its chemical composition by GC-MS engine model, GC Clarus 500; Perkin Elmer and Computer Mass Library (Wiley 138L) of 80,000 compounds with a GC column Elite 1 (100% Methyl Poly Siloxane). The other conditions were as follows: Injector: GC-Clarus 500; Perkin Elmer; Carrier gas flow Helium 1 ml/min; Split ratio 1:25; Sample injected 1μl; Oven temperature 110 o C 2 min hold; Up to 270 o C at the ratio of 5 deg/min 4 min hold; Injector temperature 250 o C; Total GC- time 38 min; MS inlet line temperature 200 o C; Source temperature 200 o C; Electron energy 70eV; Mass Scan ; MS time 39 min. Identification of components

18 The retention index is calculated for all volatile constituents using a homologous series of n-alkanes. The components of oil were identified by matching their mass spectra with those stored in the computer library such as Wiley, New York mass spectral (MS) library and their retention indices (RI) either with authentic compounds or with published data in the literature.

19 Present work: For the present work we have taken the oily fraction of P.granatum (rind) as isolated from the separation of hexane extract. The hexane eluate of ethanol extract of P.granatum yielded a waxy fraction which was rechromatographed on silica gel column. Its hexane eluates yielded waxy fractions, which were small in amount and due to the small quantity and very less separation seen on the TLC plates they were separated by Gas Chromatography and were analysed by mass spectrophotometer associated with it from Saurashtra University Rajkot. The compounds were identified by comparing their retention time and covet indexes with that of literature and by interpretation of mass spectra. SAMPLE- RP-1 CH3 CH3 CH (CH2)4 CH3

20 CH3 CH3 CH (CH2)7 CH3 CH3 CH3 CH2 CH (CH2)5 CH2=CH2

21 CH3 CH3 CH2 CH (CH2)6 CH3 CH3 CH3 CH2 CH (CH2)2 CH3

22 CH3 CH3 CH (CH2)3 CH3 (CH2)6 CH3

23 CH3 CH3 CH (CH2)2 CH3 CH3 CH3 CH CH2 CH2 OH

24 (CH2)2 CH3 O CH3 CO (CH2)6 CH3

25 O CH3CH2 CH=CH (CH2)5 C OH SAMPLE RP - 2 CH3 CH3 C CH (CH2) CH3 H OH

26

27 CH3 CH3 CH2 CH (CH2)4 CH3 CH3 CH3 CH2 CH (CH2)4 CH2OH

28 CH3 CH3 CH3 CH (CH2)2CH CH3 CH3 CH3CH2 C CH2 CH (CH2)4 CH3 CH3

29 CH3 CH3 CH (CH2)3 CH3 (CH2)7 CH3

30 (CH2)4 CH3 CH3 CH3 CH (CH2)4 CH3

31 O CH3 (CH2)8 C OH SAMPLE RP - 3 CH3 (CH2)9 CH3

32

33 CH3 CH3 CH2 CH CH2 CH2 OH CH2 CH2 CH2 CH3

34 CH3 CH2 CH (CH2)5 CH3 CH3 CH3 CH (CH2)3 CH3

35 -CH2 CH=CH CH2 CH3 CH3 CH3 CH (CH2)2 CH3

36 CH3 CH3 CH2 CH (CH2)2 CH3 CH3 CH3 CH3 CH CH2 CH CH3

37 CH3 CH3 CH (CH2)4 CH3 CH3 CH3 (CH2)2 CH=CH CH3

38 CH3 CH=CH CH3 CH3 (CH2)3 CH=CH(CH2)12 CH3

39 Identification of P.granatum (hexane fraction): RP-1 Peak 1: 2-Methylheptane Mass spectrum showed that the molecule was an aliphatic compound. Base peak obtained at m/z 43, showed the presence of ( CH 2 )2- CH3 moiety. The mass peak at m/z 113 confirmed the compound to be an aliphatic compound having 7 carbons. peak 1 was identified as 2-Methylheptane. Thus the Peak 2: 2-Methyldecane Mass spectrum showed that the molecule was a long chain aliphatic compound. Base peak obtained at m/z 43, showed the presence ( CH 2 )2- CH3 moiety. The mass peak at m/z 131 confirmed the compound to be an aliphatic compound having 10 carbons. peak 2 was identified as 2-Methyldecane. Thus the Peak 3: 8-Methyl, 3 dec-1-ene- Mass spectrum showed that the molecule was an aliphatic compound. The possible fragmentation with peaks at 86,71 and 57,43 Base peak obtained at m/z 57, showed the presence of ( CH 2 )n and the mass peak at m/z 133 confirmed the compound to be an unsaturated compound having 10 carbons. peak 3 was identified as 8-Methyl 3dec-1-ene. Thus the Peak 4: 3-Methyldecane Mass spectrum showed that the molecule was an aliphatic compound. Base peak obtained at m/z 57, showed the presence of ( CH 2 ) n The mass peak at m/z 134 confirmed the compound to be an aliphatic compound having 10 carbons. identified as 3-Methyldecane. Thus the peak 1 was Peak 5: 3-Methylhexane Mass spectrum showed that the molecule was a small aliphatic compound. 100% abundant fragment obtained at m/z 57, showed the presence of CH3 CH2CH2CH2 CH 2. The mass peak at m/z 93 confirmed the compound to be an aliphatic compound having 6 carbons. Thus the peak 1 was identified as 3-Methylhexane. Peak 6: 2-Methylhexane Mass spectrum showed that the molecule was a small aliphatic compound. Base peak obtained at m/z 43, showed the presence of ( CH 2 )n The mass peak at m/z 103 confirmed the compound to be an aliphatic compound having 6 carbons. identified as 2-Methylhexane. Thus the peak 1 was

40 Peak 7: Heptylcyclobutane Mass spectrum showed that the molecule was a small aliphatic compound. Base peak obtained at m/z 56, showed the presence of CH 2. The mass peak at m/z 93 confirmed the compound to be an aliphatic compound having 6 carbons. 1 was identified as Heptylcyclobutane. Thus the peak Peak 8: 2-Methylpentane The mass spectrum identified the molecule as an aliphatic hydrocarbon. 100% abundant fragment obtained at m/z 85, which was due to the presence of CH3-(CH2)n. This confirmed the compound to be 2-Methylpentane. Peak 9: 3-Methylbutan-1-ol Mass spectrum showed that the molecule was a long chain aliphatic alcohol. Abundant fragment obtained at m/z 43 and 147 was due to the position of hydroxyl groups. [18] This confirmed the compound to be an aliphatic alcohol. Thus the peak 4 was identified as 3-Methylbutan-1-ol. Peak 10: Propylcyclobutane The mass spectrum identified the molecule as an aliphatic hydrocarbon. 100% abundant fragment obtained at m/z 100, which was due to the presence of CH3CH2CH2-. [21] This confirmed the compound to be Propylcyclobutane Peak 11: Methyl octanoate- Mass spectrum showed the mass peak at 173. The mass fragments at 85, 60 and 43 confirm the long chain of compound. The peak at 73 was due to the cleavage of a fragment by McLefferty rearrangement. [19,20] Thus the peak 11 was identified as Methyloctanoate. Peak 12: 7 -dec-7-enoic acid- Mass spectrum identified the molecule as organic acid. The position of double bond was confirmed by the fragments at 83 and 69, the peak at 55 indicated the formation of CH3- CH2-CH=CH + ion. 10 Carbons were confirmed by the peaks at 97, 83, 69 and the compound was identified as 7 -dec-7-enoic acid Identification of P.granatum (Hexane: Benzene fraction): RP-2 Peak 1: 2-Methylpentan-3-ol 3-Methylbutan-1-ol Mass spectrum showed that the molecule was a long chain aliphatic alcohol. Abundant fragment obtained at m/z 43 and 147 was due to the position of hydroxyl

41 groups. This confirmed the compound to be an aliphatic alcohol. Thus the peak 4 was identified 2-Methylpentan-3-ol [21] Peak 2: 2-Methyldecane The mass spectrum identified the molecule as an aliphatic long chain hydrocarbon. 100% abundant fragment obtained at m/z 71, 57.43, 41, which were due to the presence of CH3 CH2CH2CH2 CH 2. This confirmed the compound to be 2-Methyldecane. Peak 3: 6-Methyloctane-1-ol Mass spectrum showed that the molecule was a long chain aliphatic alcohol. Abundant fragment obtained at m/z 43 and 147 was due to the position of hydroxyl groups. This confirmed the compound to be an aliphatic alcohol. Thus the peak 4 was identified as 6-Methyloctane-1-ol Peak 4: 3-Methyloctane The mass spectrum identified the molecule as an aliphatic hydrocarbon. 100% abundant fragment obtained at m/z 56, which was due to the presence of CH3CH2CH2CH2 CH 2. This confirmed the compound to be 3-Methyloctanoate. Peak 5: 2, 3-Dimethylpentane The mass spectrum identified the molecule as an aliphatic hydrocarbon. 100% abundant fragment obtained at m/z 56, which was due to the presence of CH3CH2CH2CH2 CH 2. This confirmed the compound to be 2, 3-Dimethylpentane. Peak 6: 2, 5-Dimethylhexane The mass spectrum identified the molecule as an aliphatic hydrocarbon. 100% abundant fragment obtained at m/z 56, which was due to the presence of CH3CH2CH2CH2 CH 2. This confirmed the compound to be 2, 5-Dimethylhexane. Peak 7: Octylcyclobutane Mass spectrum showed that the molecule was a small aliphatic compound. Base peak obtained at m/z 56, showed the presence of CH 2. The mass peak at m/z 93 confirmed the compound to be an aliphatic compound having 6 carbons. 1 was identified as Octylcyclobutane. Thus the peak Peak 8: 2-Methylhexane The mass spectrum identified the molecule as an aliphatic hydrocarbon. 100% abundant fragment obtained at m/z 56, which was due to the presence of CH3CH2CH2CH2 CH 2. This confirmed the compound to be 2-Methylhexane. [22] Peak 9: 2-Methylheptane The mass spectrum identified the molecule as an aliphatic hydrocarbon. 100% abundant fragment obtained at m/z 56,

42 which was due to the presence of CH3CH2CH2CH2 CH 2. This confirmed the compound to be 2-Methylheptane. Peak 10: Pentyl cyclobutane Mass spectrum showed that the molecule was a small aliphatic compound. Base peak obtained at m/z 56, showed the presence of CH 2. The mass peak at m/z 93 confirmed the compound to be an aliphatic compound having 6 carbons. Thus the peak 1 was identified as Pentylcyclobutane. Peak 11: Decanoic acid Mass spectrum showed that the molecule was a long chain aliphatic alcohol. Abundant fragments obtained at m/z 43. And the mass peak was seen on 174. This confirmed the compound to be an aliphatic compound. Thus the peak 2 was identified as 2-Methyl- Undecane. Identification of P.granatum (Benzene fraction): RP-3 Peak 1: n-undecane Mass spectrum showed that the molecule was a long chain aliphatic compound. Abundant fragments obtained at m/z 43, were due to the CH 2. This confirmed the compound to be an aliphatic compound having 11 carbons. Thus the peak 1 was identified as n-undecane. Peak: 2. 2-Methyl-Undecane Mass spectrum showed that the molecule was a long chain aliphatic alcohol. Abundant fragments obtained at m/z 43. And the mass peak was seen on 170 which confirmed the compound to be an aliphatic compound. Thus the peak 2 was identified as 2- Methyl-Undecane. Peak 3: Butyl Cyclopropane : Mass spectrum showed that the molecule was a small aliphatic compound. Base peak obtained at m/z 56, showed the presence of CH 2. The mass peak at m/z 93 confirmed the compound to be an aliphatic compound having 6 carbons. 1 was identified as Butylcyclopropane. Thus the peak Peak 4: 3-Methyl Pentan-1-ol Mass spectrum showed that the molecule was a long chain aliphatic alcohol. Abundant fragment obtained at m/z 43 and 145 was due to the position of hydroxyl groups. [22] This confirmed the compound to be an aliphatic alcohol. Thus the peak 4 was identified as 3-Methyl Pentan-1-ol.

43 Peak 5: 3-Methylnonane The mass spectrum identified the molecule as an aliphatic hydrocarbon. 100% abundant fragment obtained at m/z 56, which was due to the presence of CH3CH2CH2CH2 CH 2. This confirmed the compound to be 3-Methylnonane. Peak 6: 2-Methylhexane The mass spectrum identified the molecule as an aliphatic hydrocarbon. 100% abundant fragment obtained at m/z 56, which was due to the presence of CH3CH2CH2CH2 CH 2. This confirmed the compound to be 2-Methylhexane. Peak 7: 2 pent-2-en-1-ylcyclobutane- Mass spectrum showed that the molecule was a small aliphatic compound. Base peak obtained at m/z 56, showed the presence of CH 2. The mass peak at m/z 93 confirmed the compound to be an aliphatic compound having 6 carbons. peak 1 was identified as 2 pent-2-en-1-ylcyclobutane. Thus the Peak 8: 2-Methylpentane The mass spectrum identified the molecule as an aliphatic hydrocarbon. 100% abundant fragment obtained at m/z 56, which was due to the presence of CH3CH2CH2CH2 CH 2. This confirmed the compound to be 3-Methylhpentane. Peak 9: 2, 4 - Dimethyl Pentane The mass spectrum identified the molecule as an aliphatic hydrocarbon. 100% abundant fragment obtained at m/z 56, which was due to the presence of CH3CH2CH2CH2 CH 2. This confirmed the compound to be 2,4-Dimethylpentane. Peak 10: 3-Methylhexane The mass spectrum identified the molecule as an aliphatic hydrocarbon. 100% abundant fragment obtained at m/z 56, which was due to the presence of CH3CH2CH2CH2 CH 2. This confirmed the compound to be 3-Methylhexane. Peak 11: 2-3-methylhex-2-ene Mass spectrum showed that the molecule was an aliphatic compound. The possible fragmentation with peaks at 86,71 and 57,43 Base peak obtained at m/z 57, showed the presence of CH 2 and the mass peak at m/z 133 confirmed the compound to be an unsaturated compound having 10 carbon. peak 3 was identified as 2-3-methylhex-2-ene ene. Thus the Peak 12: 2-Methylheptane The mass spectrum identified the molecule as an aliphatic hydrocarbon. 100% abundant fragment obtained at m/z 56, which was due to the presence of CH3CH2CH2CH2 CH 2. This confirmed the compound to be 3-Methylhexane.

44 Peak 13: 2 -but-2-ene ene Mass spectrum showed that the molecule was an aliphatic compound. The possible fragmentation with peaks at 86, 71 and 57,43 Base peak obtained at m/z 57, showed the presence of CH 2 and the mass peak at m/z 133 confirmed the compound to be an unsaturated compound identified as 2 -but-2-ene. having 10 carbon. Thus the peak 3 was

45 Peak 14: Henicos-5-ene- Mass spectrum showed that the molecule was an aliphatic compound. The peaks at 98, 83, 56 and 41 indicate the long chain aliphatic compound and the presence of a single unsaturation. Thus the compound was identified as Henicos-5-ene. Experimental Work The powdered fruit rind of P. granatum was extracted and then fractionated with various organic solvents as discussed in chapter 3 Processing of Ethanol extract: The ethanol extract was fractionated by liquid-liquid extraction with different solvents in their increasing order of polarity and various fractions were collected. The non polar solvent hexane yielded a fraction which was coated on silica and was again fractionated by hexane which yielded oily fraction which was very less in quantity. There TLC examination in the system Hexane: Benzene (9:1) showed long spot indicating the mixture of components having Rf values very close to each other and could not be separated by TLC or even column chromatography and it was decided to analyze them by GC-MS analysis. GC-MS analysis revealed the presence of 12, 11 and14 different compounds from the three samples RP-1, RP-2, RP-3 respectively. The compounds were identified by comparing their retention time and covet indexes with that of literature and by interpretation of mass spectra. Mass Spectrum: m/z (rel. int. %) Investigation of hexane fraction RP-1 Peak 1: (R. time: 2.125) (1.23%) Line.1 M + 113, 85, 71,57,43, 41, 27 Peak 2: (R. time: 2.375) (22.79%) Line.2 M + 135, 105, 86, 71,57,43, 41, 27 Peak 3: (R. time: 2.157) (19.03%) Line.3 M + 133, 105, 86, 71,57,43, 41, 26 Peak 4: (R. time: 2.675) (22.38%) Line.4 M + 134, 119, 98, 86, 71, 57, 43, 41, 27 Peak 5: (R. time: 2.917) (2.97%) Line.5 M + 93, 85, 71,57,43, 41, 27 Peak 6: (R. time: 2.967) (3.29%) Line.6

46 M + 103, , 71,57,43, 41, 27 Peak 7: (R. time: 3.083) (19.29%) Line.7 M + 122, 105, 84, 70, 56, 42, 41, 27 Peak 8: (R. time: 3.358) (0.62%) Line.8 M + 106, 85, 71, 57, 43, 41, 27 Peak 9: (R. time: 3.433) (1.16%) Line.9 M + 100, 88, 85, 70, 57, 43, 41, 27 Peak 10: (R. time: 3.550) (6.55%) Line.10 M + 123, 100, 84, 56, 55, 44, 27 Peak 11: (R. time: ) (0.51%) Line.11 M + 256, 239, 227, 213, 199, 185, 157, 143, 129, 115, 98, 97, 83, 69, 55, 41, 27 Peak 12: (R. time: ) (0.19%) Line.12 M + 264, 235, 209, 165, 137, 123, 98, 97, 83, 69, 55, 41, 27 Investigation of hexane: benzene fraction RP-2 Peak 1: (R. time: 2.125) (1.59%) Line.1 M + 120, 85, 71, 57, 43, 41, 27 Peak 2: (R. time: 2.383) (17.22%) Line.2 M + 152, 135, 86, 71, 57, 43, 41, 27 Peak 3: (R. time: 2.525) (19.24%) Line.3 M + 147,119, 98, 86, 71, 57, 43, 41, 27 Peak 4: (R. time: 2.658) (22.14%) Line.4 M + 127, 103, 86, 71, 57, 43, 41, 27 Peak 5: (R. time: 2.925) (3.87%) Line.5 M + 104, 98, 85, 71, 57, 43, 41, 27 Peak 6: (R. time: 2.975) (3.66%) Line.6 M + 117, , 71,57,43, 41, 27 Peak 7: (R. time: 3.083) (20.80%) Line. M + 139, 135, 105, 84, 70, 56, 42, 41, 27 Peak 8: (R. time: 3.358) (0.91%) Line.8 M + 99, 85, 71, 57, 43, 41, 40, 27 Peak 9: (R. time: 3.442) (1.66%) Line.9 M + 108, 100, 85, 57, 43, 41, 27 Peak 10: (R. time: 3.550) (8.53%) Line.10 M + 127, 100, 84, 56, 55, 41, 27

47 Peak 11: (R. time: ) (0.40%) Line.11 M + 256, 239, 227, 213, 199, 185, 174, 157, 143, 129, 115, 98, 85, 73, 60, 43 41, 27 Investigation of benzene fraction RP-3 Peak 1: (R. time: 2.125) (1.81%) Line.1 M + 157, 135, 105, 85, 71,57,43, 41, 27

48 Peak 2: (R. time: 2.383) (15.74%) Line.2 M + 150, 135, 107, 86, 71,57,43, 41, 27 Peak 3: (R. time: 2.417) (2.22%) Line.3 M + 96, 70, 67,42, 41, 27 Peak 4: (R. time: 2.525) (18.06%) Line.4 M + 135, 105, 86, 71, 51, 43, 41, 27 Peak 5: (R. time: 2.683) (20.45%) Line.5 M + 140,135, 105, 86, 71, 57,43, 41, 27 Peak 6: (R. time: 2.975) (10.94%) Line.6 M + 115, 100, 85, 71,57,43, 41, 27 Peak 7: (R. time: 3.092) (18.98%) Line.7 M + 135, 105, 87, 70, 56, 55, 41, 27 Peak 8: (R. time: 3.358) (0.90%) Line.8 M + 125, 85, 71, 51, 43, 41, 27 Peak 9: (R. time: 3.442) (1.77%) Line.9 M + 100, 85, 70, 57, 43, 41, 27 Peak 10: (R. time: 3.550) (8.87%) Line.10 M + 102, 84, 83, 70, 56, 41, 27 Peak 11: (R. time: 3.658) (0.14%) Line.11 M + 135, 98, 84, 83, 69,56, 55, 41, 27 Peak 12: (R. time: 3.783) (0.07%) Line.12 M + 135, 98, 84, 70, 56, 55, 41, 27

49 Peak 13: (R. time: 3.867) (0.02%) Line.13 M + 286, 244, 187, 124, 98, 97, 70, 56, 55, 41, 27 Peak 14: (R. time: 4.442) (0.01%) Line.14 M + 244, 159, 130, 98, 84, 83, 68, 56, 55, 41, 27 Table: 4.1 GC-MS details of sample: RP-1 Peak R.Time Area Area% Name Methylheptane Methyldecane Methyldec-1-ene Methyldecane Methylhexane Methylhexane Heptylcyclobutane Methylpentane Methylbutane-1-ol Proply cyclobutane Methyloctanoate dec-7-enoic acid

50 Table: 4.2 GC-MS details of sample: RP-2 Peak R.Time Area Area% Name Methylpentan-3-ol Methyldecane Methyloctane-1-ol Methyloctane ,3-Dimethylpentane ,5-Dimethylhexane Octylcyclobutane Methylhexane Methylheptane Pentyl cyclobutane Decanoic acid

51 Table: 4.3 GC-MS details of sample RP-3 Peak R.Time Area Area% Name n- Undecane methyl Undecane Butyl Cyclopropane methyl pentane-1-ol methyl nonane methyl hexane pent-2-en-1- ylcyclobutane methyl Pentane ,4- dimethyl pentane methyl hexane methylhex-2-ene methyl heptane But-2-ene

52 References 1. Karasek, Francis W.,and Raye Clement, Basic Gas Chromatography Mass Spectrometry: Principles & Techniques. Amsterdam: Elsevier. (1988) 2. Gohlke, R. S. "Time-of-Flight Mass Spectrometry and Gas-Liquid Partition Chromatography". Analytical Chemistry 31: 535. (1959) 3. Gohlke, R; McLafferty, Fred W. "Early gas chromatography/mass spectrometry". Journal of the American Society for Mass Spectrometry 4: 367. (1993). 4. Litchfield C., Riser R. and Isbell A.L., Ibid. 40, 302, (1963) 5. Zen man T.J., Gas Chrom. 3, 18. (1985) 6. Acumen R.G., J. Am. Oil. Chem. Soc., 40,564. (1963) 7. Cropper F.R. and Heywood A., Nature, 132, 110, (1953) 8. James A.T. and Martin A.R.P., J. Bio. Chem., 144, (1956) 9. Lipsky S.R., Landowone R.A., Lovelock J.E., Anal. Chem., 31, 852, (1959) 10. Rose M.E. and Johnson R.A.W., Mass spectrometry for chemist and biochemist, (Cambridge University press, Cambridge), 112, (1982) 11. Eiceman G. A., Hill H.H. and Hovgen F.W., Anal. Chem.,70 (1993) 12. Levy R.K., Gasser H., Herman T.S. and Hovgen F.W., Anal.Chem. 41, (1969) 13. Hedfgall B., Jonsson P.A., Marde Y., Ryhage R. and Wikstrom S., J. Sci. Ins. 2, 1031, (1969) 14. Reynolds, W.E., Bridges, T.C., Tocket, R.E. and Coburn, T.B., 16 th Annual conference on Mass Spectrometry and allied Topics (ASTM, Committee E-14, Pittsburgh, Dena), 77. (1968) 15. Woolley D.W., J. Bio.Chem., 147, , (1943) 16. Chibnall A.C., and Channon H. I., J. Bio. Chem , (1929) 17. Halliday N., Devel J. J. Tr., Tragerman L. J. and Warel W.E., J. Bio. Chem., 132, , (1940) 18. Bailey A.V., et al, J. Am. Oil Chem. Soc., 48, 775, (1971) 19. McLafferty F. W. Interpretation of mass spectra (W.A Benjamin, Inc), , (1973)

53 20. Malta V., et, al, J.Chem.Soc. (B), 548, (1971) 21. Verma Mahesh N. and Madras, Giridhar, Applied Biochemistry & Biotech April 141, , (2007) 22. Silverstein R.M. and Webster F.X. Spectrometric Identification of Organic Compounds VI, (John Wiley & Sons, Inc, New York.), 48, (2003)

Chromatography Vacuum Ultraviolet Spectroscopy

Chromatography Vacuum Ultraviolet Spectroscopy Application Note Differentiation and Determination Differentiation and Determination of Fatty Acid Methyl of Fatty Esters Acid by Gas Methyl Chromatography Esters by Vacuum Gas Ultraviolet Spectroscopy

More information

Analysis of FAMEs Using Cold EI GC/MS for Enhanced Molecular Ion Selectivity

Analysis of FAMEs Using Cold EI GC/MS for Enhanced Molecular Ion Selectivity APPLICATION NOTE Gas Chromatography/ Mass Spectrometry Author: Adam Patkin PerkinElmer, Inc. Shelton, CT Analysis of FAMEs Using GC/MS for Enhanced Molecular Ion Selectivity Introduction Characterization

More information

Mass Spectrometry Introduction

Mass Spectrometry Introduction Mass Spectrometry Introduction Chem 744 Spring 2013 What MS is and is not MS is NOT a spectroscopic method. Molecules are not absorbing EM radiation MS is the generation, separation and characterization

More information

CHM 424L Organic Laboratory, Dr. Laurie S. Starkey Introduction to Mass Spectrometry

CHM 424L Organic Laboratory, Dr. Laurie S. Starkey Introduction to Mass Spectrometry CM 424L rganic Laboratory, Dr. Laurie S. Starkey Introduction to Mass Spectrometry Mass spectrometry is used to determine a sample's molecular mass and molecular formula. Some structural information can

More information

Methods in Mass Spectrometry. Dr. Noam Tal Laboratory of Mass Spectrometry School of Chemistry, Tel Aviv University

Methods in Mass Spectrometry. Dr. Noam Tal Laboratory of Mass Spectrometry School of Chemistry, Tel Aviv University Methods in Mass Spectrometry Dr. Noam Tal Laboratory of Mass Spectrometry School of Chemistry, Tel Aviv University Sample Engineering Chemistry Biology Life Science Medicine Industry IDF / Police Sample

More information

New GCMS Applications

New GCMS Applications New GCMS Applications Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel, Sample Characterization by GCxGC/FID/MSD, Crude Oil Biomarkers Malgorzata Sierocinska Agilent Technologies Waldbronn

More information

DETERMINATION OF COMPOSITION OF TRIACYLGLYCEROLS AND COMPOSITION AND CONTENT OF DI-ACYLGLYCEROLS BY CAPILLARY GAS CHROMATOGRAPHY, IN VEGETABLE OILS

DETERMINATION OF COMPOSITION OF TRIACYLGLYCEROLS AND COMPOSITION AND CONTENT OF DI-ACYLGLYCEROLS BY CAPILLARY GAS CHROMATOGRAPHY, IN VEGETABLE OILS INTERNATIONAL OLIVE COUNCIL COI/T.20/Doc. No 32 November 2013 ENGLISH Original: ENGLISH Príncipe de Vergara, 154 28002 Madrid España Telef.: +34 915 903 638 Fax: +34 915 631 263 - e-mail: iooc@internationaloliveoil.org

More information

Analysis of Organic Acids and Alcohols Using the Agilent J&W DB-624UI Ultra Inert GC Column

Analysis of Organic Acids and Alcohols Using the Agilent J&W DB-624UI Ultra Inert GC Column Analysis of Organic Acids and Alcohols Using the Agilent J&W DB-624UI Ultra Inert GC Column Application Note Food Testing & Agriculture Authors Pat Sasso and Ken Lynam Agilent Technologies, Inc. Abstract

More information

Components of a Mass Spectrometer

Components of a Mass Spectrometer Components of a Mass Spectrometer Sample Introduction Inlet GC LC Direct Insertion (Syringe/Probe) Ionization Ion Separation Ion Detection Ion Source EI,CI,,, MALDI Mass Analyzer Under vacuum TOF, Quadrupole,

More information

Organic Chemistry Laboratory Fall Lecture 3 Gas Chromatography and Mass Spectrometry June

Organic Chemistry Laboratory Fall Lecture 3 Gas Chromatography and Mass Spectrometry June 344 Organic Chemistry Laboratory Fall 2013 Lecture 3 Gas Chromatography and Mass Spectrometry June 19 2013 Chromatography Chromatography separation of a mixture into individual components Paper, Column,

More information

An Introduction and Overview on Comprehensive Two-Dimensional Gas Chromatography (GCxGC): New Opportunities for Unresolved Complex Mixtures

An Introduction and Overview on Comprehensive Two-Dimensional Gas Chromatography (GCxGC): New Opportunities for Unresolved Complex Mixtures An Introduction and Overview on Comprehensive Two-Dimensional Gas Chromatography (GCxGC): New Opportunities for Unresolved Complex Mixtures Daniela Cavagnino GC Product Manager Thermo Fisher Scientific,

More information

Rapid Analysis of 37 FAMEs with the Agilent 8860 Gas Chromatograph

Rapid Analysis of 37 FAMEs with the Agilent 8860 Gas Chromatograph Application Note Food Rapid Analysis of 37 FAMEs with the Agilent 88 Gas Chromatograph Author Youjuan Zhang Agilent Technologies (Shanghai) Co. Ltd., Shanghai 131 P. R. China Abstract An Agilent 88 GC

More information

DIRECT EXTRACTION OF BENZODIAZEPINE METABOLITE WITH SUPERCRITICAL FLUID FROM WHOLE BLOOD

DIRECT EXTRACTION OF BENZODIAZEPINE METABOLITE WITH SUPERCRITICAL FLUID FROM WHOLE BLOOD DIRECT EXTRACTION OF BENZODIAZEPINE METABOLITE WITH SUPERCRITICAL FLUID FROM WHOLE BLOOD Kenichi TAKAICHI, Shuji SAITOH, Yoshio KUMOOKA, Noriko TSUNODA National Research Institute of Police Science, Chiba,

More information

Type Analysis of Citrus Essential Oils by Multidimensional Supercritical Fluid Chromatography/ Gas Chromatography

Type Analysis of Citrus Essential Oils by Multidimensional Supercritical Fluid Chromatography/ Gas Chromatography 25 Type Analysis of Citrus Essential Oils by Multidimensional Supercritical Fluid Chromatography/ Gas Chromatography Takashi YARITA, Akira NoMURA and Yoshiyuki HoRIMOTo Department of Analytical Chemistry,

More information

Ultra Performance Liquid Chromatography Coupled to Orthogonal Quadrupole TOF MS(MS) for Metabolite Identification

Ultra Performance Liquid Chromatography Coupled to Orthogonal Quadrupole TOF MS(MS) for Metabolite Identification 22 SEPARATION SCIENCE REDEFINED MAY 2005 Ultra Performance Liquid Chromatography Coupled to Orthogonal Quadrupole TOF MS(MS) for Metabolite Identification In the drug discovery process the detection and

More information

Small Scale Preparative Isolation of Corticosteroid Degradation Products Using Mass-Based Fraction Collection Application

Small Scale Preparative Isolation of Corticosteroid Degradation Products Using Mass-Based Fraction Collection Application Small Scale Preparative Isolation of Corticosteroid Degradation Products Using Mass-Based Fraction Collection Application Pharmaceutical Author Cliff Woodward Agilent Technologies, Inc. 285 Centerville

More information

Separation of 37 Fatty Acid Methyl Esters Utilizing a High-Efficiency 10 m Capillary GC Column with Optimization in Three Carrier Gases

Separation of 37 Fatty Acid Methyl Esters Utilizing a High-Efficiency 10 m Capillary GC Column with Optimization in Three Carrier Gases APPLICATION NOTE Separation of 37 Fatty Acid Methyl Esters Utilizing a High-Efficiency 10 m Capillary GC Column with Optimization in Three Carrier Gases No. 21557 Aaron L. Lamb Thermo Fisher Scientific,

More information

Rapid and Robust Detection of THC and Its Metabolites in Blood

Rapid and Robust Detection of THC and Its Metabolites in Blood Rapid and Robust Detection of THC and Its Metabolites in Blood Application Note Forensics/Doping Control Author Stephan Baumann Agilent Technologies, Inc. Santa Clara CA 95051 USA Abstract A robust method

More information

Innovations in Large Volume Injection

Innovations in Large Volume Injection Innovations in Large Volume Injection Applications of a Chromatographic Zone as Inlet System for GC/MS Jeffery S. Hollis APEX Technologies Silicon Valley, Introduction In typical GC/MS analyses, only a

More information

2D-LC as an Automated Desalting Tool for MSD Analysis

2D-LC as an Automated Desalting Tool for MSD Analysis 2D-LC as an Automated Desalting Tool for MSD Analysis Direct Mass Selective Detection of a Pharmaceutical Peptide from an MS-Incompatible USP Method Application Note Biologics and Biosimilars Author Sonja

More information

1. Sample Introduction to MS Systems:

1. Sample Introduction to MS Systems: MS Overview: 9.10.08 1. Sample Introduction to MS Systems:...2 1.1. Chromatography Interfaces:...3 1.2. Electron impact: Used mainly in Protein MS hard ionization source...4 1.3. Electrospray Ioniztion:

More information

Martin Lommatzsch Gottfried-Hagener-Str. 60 D Köln. Report

Martin Lommatzsch Gottfried-Hagener-Str. 60 D Köln. Report Martin Lommatzsch Gottfried-Hagener-Str. 60 D-51105 Köln Report Feasibility study concerning the differentiation of mineral oil hydrocarbons from recycled cardboard and oligomers from resins via multidimensional

More information

Analysis of Omega 3 and Omega 6 FAMEs in Fish Oil and Animal Fat Using an Agilent J&W DB-FATWAX Ultra Inert GC Column

Analysis of Omega 3 and Omega 6 FAMEs in Fish Oil and Animal Fat Using an Agilent J&W DB-FATWAX Ultra Inert GC Column Application Note Food Analysis of Omega 3 and Omega 6 FAMEs in Fish Oil and Animal Fat Using an Agilent J&W DB-FATWAX Ultra Inert GC Column Authors Ingrid Van Der Meer, Yun Zou, and Gustavo Serrano Agilent

More information

Mass Spectral Fragmentation Studies of Coumarin-Type Compounds Using GC High-Resolution MS

Mass Spectral Fragmentation Studies of Coumarin-Type Compounds Using GC High-Resolution MS The Open Analytical Chemistry Journal, 2011, 5, 27-36 27 Open Access Mass Spectral Fragmentation Studies of Coumarin-Type Compounds Using GC High-Resolution MS Viorica Lopez-Avila * and George Yefchak

More information

DETECTION OF BEESWAX ADULTERATION WITH HYDROCARBONS USING GAS CHROMATOGRAPHY WITH MASS DETECTOR (GC-MS)

DETECTION OF BEESWAX ADULTERATION WITH HYDROCARBONS USING GAS CHROMATOGRAPHY WITH MASS DETECTOR (GC-MS) DETECTION OF BEESWAX ADULTERATION WITH HYDROCARBONS USING GAS CHROMATOGRAPHY WITH MASS DETECTOR (GC-MS) Ewa Waś,, Helena Rybak-Chmielewska, Teresa Szczęsna e-mail: ewa.was@man.pulawy.pl Research Institute

More information

Fundamentals of Soft Ionization and MS Instrumentation

Fundamentals of Soft Ionization and MS Instrumentation Fundamentals of Soft Ionization and MS Instrumentation Ana Varela Coelho varela@itqb.unl.pt Mass Spectrometry Lab Analytical Services Unit Index Mass spectrometers and its components Ionization methods:

More information

Automated Sample Preparation for Profiling Fatty Acids in Blood and Plasma using the Agilent 7693 ALS

Automated Sample Preparation for Profiling Fatty Acids in Blood and Plasma using the Agilent 7693 ALS Automated Sample Preparation for Profiling Fatty Acids in Blood and Plasma using the Agilent 7693 ALS Application Note Clinical Research Authors Frank David and Bart Tienpont, Research Institute for Chromatography,

More information

LOCALISATION, IDENTIFICATION AND SEPARATION OF MOLECULES. Gilles Frache Materials Characterization Day October 14 th 2016

LOCALISATION, IDENTIFICATION AND SEPARATION OF MOLECULES. Gilles Frache Materials Characterization Day October 14 th 2016 LOCALISATION, IDENTIFICATION AND SEPARATION OF MOLECULES Gilles Frache Materials Characterization Day October 14 th 2016 1 MOLECULAR ANALYSES Which focus? LOCALIZATION of molecules by Mass Spectrometry

More information

Time (min) Supplementary Figure 1: Gas decomposition products of irradiated DMC.

Time (min) Supplementary Figure 1: Gas decomposition products of irradiated DMC. 200000 C 2 CH 3 CH 3 DMC 180000 160000 140000 Intensity 120000 100000 80000 60000 40000 C 2 H 6 CH 3 CH 2 CH 3 CH 3 CCH 3 EMC DEC 20000 C 3 H 8 HCCH 3 5 10 15 20 25 Time (min) Supplementary Figure 1: Gas

More information

Analysis of the fatty acids from Periploca sepium by GC-MS and GC-FID

Analysis of the fatty acids from Periploca sepium by GC-MS and GC-FID Analysis of the fatty acids from Periploca sepium by GC-MS and GC-FID Ling Tong, Lei Zhang, Shuanghui Yu, Xiaohui Chen, Kaishun Bi * Department of Pharmacy, Shenyang Pharmaceutical University, Wenhua Road

More information

Eszopiclone (Lunesta ): An Analytical Profile

Eszopiclone (Lunesta ): An Analytical Profile Eszopiclone (Lunesta ): An Analytical Profile Roxanne E. Franckowski, M.S.* and Robert A. Thompson, Ph.D. U.S. Department of Justice Drug Enforcement Administration Special Testing and Research Laboratory

More information

Column Chromatographic Isolation of Docosahexaenoic Acid from Fish Oil and its Assessment by Analytical Techniques

Column Chromatographic Isolation of Docosahexaenoic Acid from Fish Oil and its Assessment by Analytical Techniques Human Journals Research Article April 2017 Vol.:9, Issue:1 All rights are reserved by Rasika R. Nikhade et al. Column Chromatographic Isolation of Docosahexaenoic Acid from Fish Oil and its Assessment

More information

MALDI-TOF. Introduction. Schematic and Theory of MALDI

MALDI-TOF. Introduction. Schematic and Theory of MALDI MALDI-TOF Proteins and peptides have been characterized by high pressure liquid chromatography (HPLC) or SDS PAGE by generating peptide maps. These peptide maps have been used as fingerprints of protein

More information

Improving the Analysis of 37 Fatty Acid Methyl Esters

Improving the Analysis of 37 Fatty Acid Methyl Esters Application Note Food Testing Improving the Analysis of 37 Fatty Acid Methyl Esters Using Three Types of Capillary GC columns Authors Yun Zou Agilent Technologies (Shanghai) Co.Ltd, Shanghai 200131 P.R.China

More information

O O H. Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL. LC /MS conditions

O O H. Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION EXPERIMENTAL. LC /MS conditions Simplifying Qual/Quan Analysis in Discovery DMPK using UPLC and Xevo TQ MS Robert S. Plumb and Paul D. Rainville Waters Corporation, Milford, MA, U.S. INTRODUCTION The determination of the drug metabolism

More information

Identification of Aromatic Fatty Acid Ethyl Esters

Identification of Aromatic Fatty Acid Ethyl Esters Chapter 3.2 Identification of Aromatic Fatty Acid Ethyl Esters The only use of gas chromatography is not sufficient to determine which compounds are eluting from the catalytic bed. At the beginning of

More information

Solving practical problems. Maria Kuhtinskaja

Solving practical problems. Maria Kuhtinskaja Solving practical problems Maria Kuhtinskaja What does a mass spectrometer do? It measures mass better than any other technique. It can give information about chemical structures. What are mass measurements

More information

Mass Spectrometry Course Árpád Somogyi Chemistry and Biochemistry MassSpectrometry Facility) University of Debrecen, April 12-23, 2010

Mass Spectrometry Course Árpád Somogyi Chemistry and Biochemistry MassSpectrometry Facility) University of Debrecen, April 12-23, 2010 Mass Spectrometry Course Árpád Somogyi Chemistry and Biochemistry MassSpectrometry Facility) University of Debrecen, April 12-23, 2010 Introduction, Ionization Methods Mass Analyzers, Ion Activation Methods

More information

TO-3 BTEX and TPH. Table 13.1a Summary of QC Criteria for TO-3 BTEX Oxygenates

TO-3 BTEX and TPH. Table 13.1a Summary of QC Criteria for TO-3 BTEX Oxygenates TO-3 BTEX and TPH Method TO-3 uses cryogenic trapping and a gas chromatograph with a flame ionization detector (FID) to measure volatile organic compounds collected in Summa Canisters or Tedlar Bags. This

More information

A comparison study of the analysis of volatile organic acids and fatty acids

A comparison study of the analysis of volatile organic acids and fatty acids Application Note Food Testing A comparison study of the analysis of volatile organic acids and fatty acids Using DB-FATWAX Ultra Inert and other WAX GC columns Author Yun Zou Agilent Technologies (Shanghai)

More information

Mass-Based Purification of Natural Product Impurities Using an Agilent 1260 Infinity II Preparative LC/MSD System

Mass-Based Purification of Natural Product Impurities Using an Agilent 1260 Infinity II Preparative LC/MSD System Application Note Food Testing and Agriculture Mass-Based Purification of Natural Product Impurities Using an Agilent 126 Infinity II Preparative LC/MSD System Authors Florian Rieck and Jörg Hippler Agilent

More information

[application note] Simultaneous detection and quantification of D 9 THC, 11-OH-D 9 T H C and D 9 THC-COOH in whole blood by GC tandem quadrupole MS

[application note] Simultaneous detection and quantification of D 9 THC, 11-OH-D 9 T H C and D 9 THC-COOH in whole blood by GC tandem quadrupole MS Simultaneous detection and quantification of D 9 THC, 11-OH-D 9 T H C and D 9 THC-COOH in whole blood by GC tandem quadrupole MS Marie Bresson, Vincent Cirimele, Pascal Kintz, Marion Villain; Laboratoire

More information

Characterization of sixty alkenes in a cat-cracked gasoline naphtha by gas chromatography

Characterization of sixty alkenes in a cat-cracked gasoline naphtha by gas chromatography REFINERY OLEFINES HEXENES ISOMERS ANALYSIS PETROL PROPORTIONS HEPTENES SEPARATION GC-MS Open access accepted manuscript version of Chromatographia 38 (1994) 222-226 Link to publisher Characterization of

More information

CORESTA Recommended Method No. 84

CORESTA Recommended Method No. 84 Cooperation Centre for Scientific Research Relative to Tobacco E-Vapour Sub-Group CORESTA Recommended Method No. 84 DETERMINATION OF GLYCERIN, PROPYLENE GLYCOL, WATER, AND NICOTINE IN THE AEROSOL OF E-CIGARETTES

More information

Separation of Polyphenols by Comprehensive 2D-LC and Molecular Formula Determination by Coupling to Accurate Mass Measurement

Separation of Polyphenols by Comprehensive 2D-LC and Molecular Formula Determination by Coupling to Accurate Mass Measurement Separation of Polyphenols by Comprehensive 2D-LC and Molecular Formula Determination by Coupling to Accurate Mass Measurement Application Note Food Testing & Agriculture Author Edgar Naegele Agilent Technologies,

More information

#20 Separation of Alcohols Using Gas Chromatography

#20 Separation of Alcohols Using Gas Chromatography #20 Separation of Alcohols Using Gas Chromatography Purpose: The purpose of this experiment is to determine the composition of an unknown alcohol mixture by using gas chromatography. Introduction: A gas

More information

Metabolomics: quantifying the phenotype

Metabolomics: quantifying the phenotype Metabolomics: quantifying the phenotype Metabolomics Promises Quantitative Phenotyping What can happen GENOME What appears to be happening Bioinformatics TRANSCRIPTOME What makes it happen PROTEOME Systems

More information

Measuring Lipid Composition LC-MS/MS

Measuring Lipid Composition LC-MS/MS Project: Measuring Lipid Composition LC-MS/MS Verification of expected lipid composition in nanomedical controlled release systems by liquid chromatography tandem mass spectrometry AUTHORED BY: DATE: Sven

More information

Lecture 10: MS Interpretation Part 4 Postulation of Molecular Structures

Lecture 10: MS Interpretation Part 4 Postulation of Molecular Structures Lecture 10: MS Interpretation Part 4 Postulation of Molecular Structures CU- Boulder CHEM 5181 Mass Spectrometry & Chromatography Prof. Jose-Luis Jimenez Postulation of Molecular Structures There are several

More information

Supplement of Evaluation of NO + reagent ion chemistry for online measurements of atmospheric volatile organic compounds

Supplement of Evaluation of NO + reagent ion chemistry for online measurements of atmospheric volatile organic compounds Supplement of Atmos. Meas. Tech., 9, 2909 2925, 2016 http://www.atmos-meas-tech.net/9/2909/2016/ doi:10.5194/amt-9-2909-2016-supplement Author(s) 2016. CC Attribution 3.0 License. Supplement of Evaluation

More information

Supplementary Information

Supplementary Information Supplementary Information Levulinic esters from the acid-catalysed reactions of sugar and alcohol as part of bio-refinery Xun Hu and Chun-Zhu Li* Fuels and Energy Technology Institute, Curtin University

More information

Toxicology Screening of Whole Blood Extracts Using GC/Triple Quadrupole/MS

Toxicology Screening of Whole Blood Extracts Using GC/Triple Quadrupole/MS Toxicology Screening of Whole Blood Extracts Using GC/Triple Quadrupole/MS Application Note Forensic Toxicology Authors Bruce Quimby and Mike Szelewski Agilent Technologies, Inc. 2850 Centerville Road

More information

Trans Fat Determination in the Industrially Processed Edible Oils By Transmission FT-IR Spectroscopy By

Trans Fat Determination in the Industrially Processed Edible Oils By Transmission FT-IR Spectroscopy By Trans Fat Determination in the Industrially Processed Edible Oils By Transmission FT-IR Spectroscopy By Dr. Syed Tufail Hussain Sherazi E-mail: tufail_sherazi@yahoo.com National Center of Excellence in

More information

Biological Mass spectrometry in Protein Chemistry

Biological Mass spectrometry in Protein Chemistry Biological Mass spectrometry in Protein Chemistry Tuula Nyman Institute of Biotechnology tuula.nyman@helsinki.fi MASS SPECTROMETRY is an analytical technique that identifies the chemical composition of

More information

The Development of Analytical Method for the Determination of Azelaic Acid Content in Cosmetic Cream Products

The Development of Analytical Method for the Determination of Azelaic Acid Content in Cosmetic Cream Products IOP Conference Series: Materials Science and Engineering PAPER OPEN ACCESS The Development of Analytical Method for the Determination of Azelaic Acid Content in Cosmetic Cream Products To cite this article:

More information

Agilent 1260 Infinity Analytical SFC System

Agilent 1260 Infinity Analytical SFC System Agilent 1260 Infinity Analytical SFC System Installation Instructions: Checkout Procedure This technical note describes the checkout procedures for the Agilent 1260 Infinity Analytical SFC System. Contents

More information

Catalog No. : Lot No. : Container Size : 300 µl Autosampler Vial Pkg Amt : 150 µl

Catalog No. : Lot No. : Container Size : 300 µl Autosampler Vial Pkg Amt : 150 µl Catalog No. : 30731 Lot No. : 052764 Description : DHA PONA Standard Container Size : 300 µl Autosampler Vial Pkg Amt : 150 µl Expiration Date : 12 months unopened* Storage : 4 or colder *From date of

More information

A Definitive Lipidomics Workflow for Human Plasma Utilizing Off-line Enrichment and Class Specific Separation of Phospholipids

A Definitive Lipidomics Workflow for Human Plasma Utilizing Off-line Enrichment and Class Specific Separation of Phospholipids A Definitive Lipidomics Workflow for Human Plasma Utilizing Off-line Enrichment and Class Specific Separation of Phospholipids Jeremy Netto, 1 Stephen Wong, 1 Federico Torta, 2 Pradeep Narayanaswamy, 2

More information

Why use Signal-To-Noise as a Measure of MS Performance When it is Often Meaningless?

Why use Signal-To-Noise as a Measure of MS Performance When it is Often Meaningless? Why use Signal-To-Noise as a Measure of MS Performance When it is Often Meaningless? Technical Overview Authors Greg Wells, Harry Prest, and Charles William Russ IV, Agilent Technologies, Inc. 285 Centerville

More information

Rapid Separation of Fatty Acid Methyl Esters

Rapid Separation of Fatty Acid Methyl Esters Application Note Food Testing & Agriculture Rapid Separation of Fatty Acid Methyl Esters Using DB-FastFAME Intuvo GC columns Author Yun Zou Agilent Technologies (Shanghai) Co. Ltd. Shanghai 200131 P. R.

More information

Lecture 3. Tandem MS & Protein Sequencing

Lecture 3. Tandem MS & Protein Sequencing Lecture 3 Tandem MS & Protein Sequencing Nancy Allbritton, M.D., Ph.D. Department of Physiology & Biophysics 824-9137 (office) nlallbri@uci.edu Office- Rm D349 Medical Science D Bldg. Tandem MS Steps:

More information

Rapidly Analyze a Wide Range of Glycol Ethers by GC-MS Using the New Rxi -1301Sil MS Column

Rapidly Analyze a Wide Range of Glycol Ethers by GC-MS Using the New Rxi -1301Sil MS Column Rapidly Analyze a Wide Range of Glycol Ethers by GC-MS Using the New Rxi -101Sil MS Column By Rebecca Stevens and Chris English Abstract Chromatographic conditions were developed for a fast GC-MS glycol

More information

Small Molecule Science: Experimental designs for achieving ultra trace analysis

Small Molecule Science: Experimental designs for achieving ultra trace analysis Small Molecule Science: Experimental designs for achieving ultra trace analysis Michael P. Balogh Principal scientist Waters Corporation 211 Waters Corporation 1 www.cosmoscience.org The Society for Small

More information

FATTY ACID PROFILING BY GAS CHROMATOGRAPHY FOR THE SHERLOCK MIS

FATTY ACID PROFILING BY GAS CHROMATOGRAPHY FOR THE SHERLOCK MIS FATTY ACID PROFILING BY GAS CHROMATOGRAPHY FOR THE SHERLOCK MIS Traditional gas chromatography of complex mixtures of compounds requires precision on the part of the chromatography equipment and considerable

More information

MASS SPECTROMETRY BASED METABOLOMICS. Pavel Aronov. ABRF2010 Metabolomics Research Group March 21, 2010

MASS SPECTROMETRY BASED METABOLOMICS. Pavel Aronov. ABRF2010 Metabolomics Research Group March 21, 2010 MASS SPECTROMETRY BASED METABOLOMICS Pavel Aronov ABRF2010 Metabolomics Research Group March 21, 2010 Types of Experiments in Metabolomics targeted non targeted Number of analyzed metabolites is limited

More information

A New Method for the Early Detection of Edible Oil Oxidation

A New Method for the Early Detection of Edible Oil Oxidation WHITE PAPER Early Detection of Edible Oil Oxidation A New Method for the Early Detection of Edible Oil Oxidation Edible oils are used in a wide range of culinary applications. Oils containing unsaturated

More information

LC/MS Method for Comprehensive Analysis of Plasma Lipids

LC/MS Method for Comprehensive Analysis of Plasma Lipids Application Note omics LC/MS Method for Comprehensive Analysis of Plasma s Authors Tomas Cajka and Oliver Fiehn West Coast Metabolomics Center, University of California Davis, 451 Health Sciences Drive,

More information

Single Quadrupole Compact Mass Spectrometer for Liquid Sample

Single Quadrupole Compact Mass Spectrometer for Liquid Sample Single Quadrupole Compact Mass Spectrometer for Liquid Sample YL INSTRUMENT CO., LTD. YL9900 MS quickly provides essential information and improves the chemist's workflow. Mordern organic labs have become

More information

Lipids Analysis. Lipids

Lipids Analysis. Lipids Lipids Analysis Stephen Barnes 3 5 15 Lipids Lipids are mostly very hydrophobic Most are conjugates of fatty acids of a variety of chain lengths, which have different degrees of unsaturation, cis trans

More information

Agilent 7700x ICP-MS

Agilent 7700x ICP-MS Agilent 7700x ICP-MS Metals analysis Mid ppt to low ppb detection range Option to detect in four different gas modes to gain optimum detection of multiple metals or interferences in a sample Autosampler

More information

DETERMINATION OF FATTY ACIDS IN EDIBLE OILS BY CAPILARY GC

DETERMINATION OF FATTY ACIDS IN EDIBLE OILS BY CAPILARY GC DETERMINATION OF FATTY ACIDS IN EDIBLE OILS BY CAPILARY GC Vesna Kostik 1 University Goce Delcev Stip Faculty of Medicine Department of Pharmacy 1 WHY FATTY ACID (FA) ANALYSIS IN EDIBLE OILS The content

More information

Impact of Chromatography on Lipid Profiling of Liver Tissue Extracts

Impact of Chromatography on Lipid Profiling of Liver Tissue Extracts Impact of Chromatography on Lipid Profiling of Liver Tissue Extracts Application Note Clinical Research Authors Mark Sartain and Theodore Sana Agilent Technologies, Inc. Santa Clara, California, USA Introduction

More information

Quantitative Analysis of -Hydroxybutyrate in Hair Using Target Analyte Finding Processing of Comprehensive GC-HRT Data

Quantitative Analysis of -Hydroxybutyrate in Hair Using Target Analyte Finding Processing of Comprehensive GC-HRT Data Quantitative Analysis of -Hydroxybutyrate in Hair Using Target Analyte Finding Processing of Comprehensive GC-HRT Data LECO Corporation; Saint Joseph, Michigan USA Key Words: Pegasus GC-HRT, GHB, Hair,

More information

Secrets to Successful GC-MS/MS Pesticide Method Development

Secrets to Successful GC-MS/MS Pesticide Method Development Secrets to Successful GC-MS/MS Pesticide Method Development Katerina Mastovska Nutritional Chemistry and Food Safety Covance Laboratories QuEChERS vs. Traditional Methods Quick, Easy, Cheap, Effective,

More information

Application Note # FTMS-46 solarix XR: Analysis of Complex Mixtures

Application Note # FTMS-46 solarix XR: Analysis of Complex Mixtures Application Note # FTMS-46 solarix XR: Analysis of Complex Mixtures Introduction Natural organic matter (NOM) as well as crude oil and crude oil fractions are very complex mixtures of organic compounds

More information

Detection of Cannabinoids in Oral Fluid with the Agilent 7010 GC-MS/MS System

Detection of Cannabinoids in Oral Fluid with the Agilent 7010 GC-MS/MS System Application Note Forensics, Workplace Drug Testing Detection of Cannabinoids in Oral Fluid with the Agilent 7010 GC-MS/MS System Authors Fred Feyerherm and Anthony Macherone Agilent Technologies, Inc.

More information

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine

SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine SPE-LC-MS/MS Method for the Determination of Nicotine, Cotinine, and Trans-3-hydroxycotinine in Urine J. Jones, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 709 Key Words SPE, SOLA

More information

Edgar Naegele. Abstract

Edgar Naegele. Abstract Simultaneous determination of metabolic stability and identification of buspirone metabolites using multiple column fast LC/TOF mass spectrometry Application ote Edgar aegele Abstract A recent trend in

More information

Polymer Additive Analysis by EI and APCI

Polymer Additive Analysis by EI and APCI Polymer Additive Analysis by EI and APCI Kate Yu, Eric Block LC/MS Waters Corporation Introduction to polymer additive analysis by LC/MS. Electron Ionization (EI) Advantages Classical spectra Library searchable

More information

Detection of oxygenated polycyclic aromatic hydrocarbons (oxy-pahs) in APCI mode with a single quadrupole mass spectrometer

Detection of oxygenated polycyclic aromatic hydrocarbons (oxy-pahs) in APCI mode with a single quadrupole mass spectrometer APPLICATION NOTE 72819 Detection of oxygenated polycyclic aromatic hydrocarbons (oxy-pahs) in APCI mode with a single quadrupole mass spectrometer Author Maria Grübner; Thermo Fisher Scientific, Germering,

More information

Fast determination of residual glycerol and glycerides in biodiesel by SFC/MS using the Agilent 1260 Infinity Analytical SFC System

Fast determination of residual glycerol and glycerides in biodiesel by SFC/MS using the Agilent 1260 Infinity Analytical SFC System Fast deteration of residual glycerol and glycerides in biodiesel by SFC/MS using the Agilent 126 Infinity Analytical SFC System Application Note Petrochemical Authors Maria Rambla Alegre, Melissa N. Dunkle,

More information

Separation of Macrocyclic Lactones (Avermectins) on FLARE C18 MM & FLARE C18+ Columns

Separation of Macrocyclic Lactones (Avermectins) on FLARE C18 MM & FLARE C18+ Columns Separation of Macrocyclic Lactones (Avermectins) on FLARE C8 MM & FLARE C8+ Columns Introduction Diamond Analytics Technical Note: T05- Avermectins are a series of 6-membered macrocyclic lactone derivatives

More information

2. Ionization Sources 3. Mass Analyzers 4. Tandem Mass Spectrometry

2. Ionization Sources 3. Mass Analyzers 4. Tandem Mass Spectrometry Dr. Sanjeeva Srivastava 1. Fundamental of Mass Spectrometry Role of MS and basic concepts 2. Ionization Sources 3. Mass Analyzers 4. Tandem Mass Spectrometry 2 1 MS basic concepts Mass spectrometry - technique

More information

Quantitative Analysis of Drugs of Abuse in Urine using UHPLC Coupled to Accurate Mass AxION 2 TOF Mass Spectrometer

Quantitative Analysis of Drugs of Abuse in Urine using UHPLC Coupled to Accurate Mass AxION 2 TOF Mass Spectrometer application Note Liquid Chromatography/ Mass Spectrometry Authors Sharanya Reddy Blas Cerda PerkinElmer, Inc. Shelton, CT USA Quantitative Analysis of Drugs of Abuse in Urine using UHPLC Coupled to Accurate

More information

Simple Method (IS-MRM) to Monitor Lysophospholipids and Phospholipids During LC-MS Method Development via In-Source CID

Simple Method (IS-MRM) to Monitor Lysophospholipids and Phospholipids During LC-MS Method Development via In-Source CID Simple Method (IS-MRM) to Monitor Lysophospholipids and Phospholipids During LC-MS Method Development via In-Source CID James Little, Eastman Chemical Company, Kingsport, TN Overview Phospholipids and

More information

The Raptor HILIC-Si Column

The Raptor HILIC-Si Column The Raptor HILIC-Si Column With Raptor LC columns, Restek chemists became the first to combine the speed of superficially porous particles (also known as SPP or core-shell particles) with the resolution

More information

Benefits and Characteristic Applications of High Resolution GC/MS and LC/MS. Frank David RIC and Ghent University

Benefits and Characteristic Applications of High Resolution GC/MS and LC/MS. Frank David RIC and Ghent University Benefits and Characteristic Applications of High Resolution GC/MS and LC/MS. Frank David RIC and Ghent University Mass Spectrometry Structure Elucidation Selective and Sensitive Detection Identification

More information

CAMAG TLC-MS INTERFACE

CAMAG TLC-MS INTERFACE CAMAG TLC-MS INTERFACE 93.1 249.2 40 30 97.1 20 10 250.2 0 200 400 m/z WORLD LEADER IN PLANAR-CHROMATOGRAPHY Identification and elucidation of unknown substances by hyphenation of TLC / HPTLC and MS The

More information

Mass Spectrometry. Actual Instrumentation

Mass Spectrometry. Actual Instrumentation Mass Spectrometry Actual Instrumentation August 2017 See also http://www.uni-bielefeld.de/chemie/analytik/ms f additional infmation 1. MALDI TOF MASS SPECTROMETRY ON THE ULTRAFLEX 2 2. ESI MASS SPECTROMETRY

More information

The four levels of protein structure are: primary structure, secondary structure, tertiary structure, and quaternary structure.

The four levels of protein structure are: primary structure, secondary structure, tertiary structure, and quaternary structure. Proteins Proteins are organic complex nitrogenous compounds of high molecular weight, formed of C, H, O and N. They are formed of a number of amino acids linked together by peptide linkage [-CO-NH-]. Proteins

More information

Analytical Method for 2, 4, 5-T (Targeted to Agricultural, Animal and Fishery Products)

Analytical Method for 2, 4, 5-T (Targeted to Agricultural, Animal and Fishery Products) Analytical Method for 2, 4, 5-T (Targeted to Agricultural, Animal and Fishery Products) The target compound to be determined is 2, 4, 5-T. 1. Instrument Liquid Chromatograph-tandem mass spectrometer (LC-MS/MS)

More information

Fizzy Extraction of Volatile and Semivolatile Compounds into the Gas Phase

Fizzy Extraction of Volatile and Semivolatile Compounds into the Gas Phase SUPPORTING INFORMATION Fizzy Extraction of Volatile and Semivolatile Compounds into the Gas Phase Cheng-Hao Chang, Pawel L. Urban* Department of Applied Chemistry, National Chiao Tung University 1001 University

More information

(2Z,7Z)-Cycloocta-2,7-dienone 4

(2Z,7Z)-Cycloocta-2,7-dienone 4 (2Z,7Z)-Cycloocta-2,7-dienone 4 (R,Z)-7-Methylcyclooct-2-enone 5 (3R,7S)-Methyl 8-hydroxy-3,7-dimethyloctanoate 3 (3R,7S)-Methyl 8-((tert-butyldiphenylsilyl)oxy)-3,7-dimethyloctanoate 7 (3R,7S)-8-((tert-Butyldiphenylsilyl)oxy)-3,7-dimethyloctan-1-ol

More information

methods Electrospray mass spectrometry of human hair wax esters Mark Fitzgerald and Robert C. Murphy 1

methods Electrospray mass spectrometry of human hair wax esters Mark Fitzgerald and Robert C. Murphy 1 Electrospray mass spectrometry of human hair wax esters methods Mark Fitzgerald and Robert C. Murphy 1 Department of Pharmacology, University of Colorado at Denver and Health Sciences Center, Aurora, CO

More information

Quadrupole and Ion Trap Mass Analysers and an introduction to Resolution

Quadrupole and Ion Trap Mass Analysers and an introduction to Resolution Quadrupole and Ion Trap Mass Analysers and an introduction to Resolution A simple definition of a Mass Spectrometer A Mass Spectrometer is an analytical instrument that can separate charged molecules according

More information

IDENTIFICATION AND CONTROLOFRESIDUALSOLVENTS Identification and control of residual solvents EUROPEAN PHARMACOPOEIA 6.

IDENTIFICATION AND CONTROLOFRESIDUALSOLVENTS Identification and control of residual solvents EUROPEAN PHARMACOPOEIA 6. EUROPEAN PHARMACOPOEIA 6.0 2.4.24. Identification and control of residual solvents paper and wash each filter with 3 quantities, each of 15 ml, of methylenechlorider.placethecombinedfiltrateand washings

More information

Sample Concentration and Analysis of Human Hormones in Drinking Water

Sample Concentration and Analysis of Human Hormones in Drinking Water Sample Concentration and Analysis of Human Hormones in Drinking Water Carl Fisher Applications Chemist Ion Chromatography/Sample Preparation Thermo Fisher Scientific March 1, 215 1 The world leader in

More information

Application of Design of Experiment (DOE) to the Simultaneous GC FID/MS Analysis of Monocarboxylic Acids and Glycerides in Biofuels

Application of Design of Experiment (DOE) to the Simultaneous GC FID/MS Analysis of Monocarboxylic Acids and Glycerides in Biofuels Application of Design of Experiment (DOE) to the Simultaneous GC FID/MS Analysis of Monocarboxylic Acids and Glycerides in Biofuels G. Baglayeva Jana Šťávová, B. Diepp, E. Hellrung, W. Seames, and A. Kubátová

More information

David M. Shelow, Restek Corporation, 110 Benner Circle, Bellefonte, PA 16823

David M. Shelow, Restek Corporation, 110 Benner Circle, Bellefonte, PA 16823 USE OF SILCOCAN CANISTERS FOR STORING LOW-LEVEL (1ppb-ppb) REACTIVE SULFURS IN AIR David M. Shelow, Restek Corporation, 11 Benner Circle, Bellefonte, PA 16823 INTRODUCTION Analysis of sulfur-containing

More information

INTRODUCTION TO MALDI IMAGING

INTRODUCTION TO MALDI IMAGING INTRODUCTION TO MALDI IMAGING Marten F. Snel, Emmanuelle Claude, Thérèse McKenna, and James I. Langridge Waters Corporation, Manchester, UK INT RODUCTION The last few years have seen a rapid increase in

More information