Monoclonal immunoglobulins, also known as paraproteins,

Size: px
Start display at page:

Download "Monoclonal immunoglobulins, also known as paraproteins,"

Transcription

1 Paraproteins Are a Common Cause of Interferences With Automated Chemistry Methods Yusong Yang, MD, PhD; Peter J. Howanitz, MD; Joan H. Howanitz, MD; Helen Gorfajn, MS; Keung Wong, MS Context. Previous studies have shown that paraproteins caused spurious results on individual analytes including total bilirubin (TBIL), direct bilirubin (DBIL), or HDL-cholesterol (HDL-C). Studies demonstrating paraprotein interferences with multiple analytes measured by different analyzers have not been reported. Objective. To systemically investigate interferences of paraproteins on TBIL, DBIL, and HDL-C measured by the Roche MODULAR and the Olympus AU2700. Design. Eighty-eight serum specimens with monoclonal gammopathies were analyzed using the Roche MOD- ULAR and the Olympus AU2700. Paraprotein interferences with the MODULAR and AU2700 were identified by abnormal absorbance curves and confirmed by results from the Ortho Vitros 950 or inconsistent laboratory information. Results. Spurious results occurred in 89 of 528 measurements; 29 specimens did not demonstrate any interferences whereas 26 specimens gave spurious results in 2 to 4 of the 6 assays. Paraprotein interferences caused spuriously high levels of TBIL in 4 sera measured by the MOD- ULAR. In contrast, paraprotein interferences on DBIL were observed by at least 1 method in 44% (39/88) of sera assayed, occurring almost exclusively with the AU2700. Paraprotein interferences with HDL-C results were present in 35% of specimens assayed with the MODULAR and 16% of specimens assayed with the AU2700. In specimens with interferences, spuriously low AU2700 DBIL, MODULAR HDL-C, and AU2700 HDL-C results occurred with 28%, 90%, and 91% of specimens, respectively. Conclusions. We demonstrated that paraprotein interferences with TBIL, DBIL, and HDL-C are relatively common and provided explanations why these interferences occurred. Although it is difficult to predict which specimens cause interferences, spurious results appeared method and concentration dependent. (Arch Pathol Lab Med. 2008;132: ) Monoclonal immunoglobulins, also known as paraproteins, have been reported to interfere with a number of routine chemistry methods including analyses for creatinine, 1,2 C-reactive protein, 3 glucose, 4 iron, 5,6 inorganic phosphate, 7 9 and uric acid. 10 Recent studies have shown that paraproteins caused false-positive results for total bilirubin (TBIL) and direct bilirubin (DBIL) 14 and falsenegative results for HDL-cholesterol (HDL-C). 12,15 19 The frequency of paraprotein interferences with these analytes on automated chemistry analyzers was reported to be very rare, and comparative studies of paraprotein interferences with TBIL, DBIL, and HDL-C have not been reported with major automated chemistry analyzers. These interferences may cause incorrect diagnoses, inappropriate treatments, longer hospital stays, and increased morbidity in patients who already may be gravely ill. We systemically investigated the potential interference of paraproteins on measurements of TBIL, DBIL, and HDL-C using Accepted for publication October 12, From the Department of Pathology, State University of New York Downstate Medical Center, Brooklyn, NY (Drs Yang, P. J. Howanitz, and J. H. Howanitz, and Ms Gorfajn); and Institute of Pathology, Clinical Chemistry Laboratory, Kings County Hospital Center, Brooklyn, NY (Mr Wong). The authors have no relevant financial interest in the products or companies described in this article. Reprints: Yusong Yang, MD, PhD, Department of Pathology, Box 24, State University of New York Downstate Medical Center, 450 Clarkson Ave, Brooklyn, NY ( yusong.yang@downstate.edu). 88 serum specimens collected from patients with monoclonal gammopathies and measured using 2 commonly used automated chemistry analyzers. MATERIALS AND METHODS This study was conducted as part of a laboratory performance improvement program for patient result reporting. Serum specimens were collected from patients with monoclonal gammopathies determined by serum protein electrophoretic and immunofixation studies during a 1-year period. The sera were free of hemolysis, icterus, and lipemia by routine index checks on automated analyzers. None of the specimens were duplicates or triplicates; if they were from the same patients, they were obtained at different times. Serum protein electrophoretic and immunofixation studies were performed using the SPIFE 3000 (Helena Laboratories, Beaumont, Tex), and serum total protein concentrations were determined with the Roche P800 MODULAR (Roche Diagnostics, Indianapolis, Ind). Paraprotein concentrations were quantitated by the Quickscan 2000 (Helena Laboratories). Serum levels of TBIL, DBIL, and HDL-C were measured independently using the Roche P800 MODULAR, Olympus AU2700 (Olympus Diagnostics, Melville, NY), and Vitros 950 (Ortho-Clinical Diagnostics, Rochester, NY) automated chemistry analyzers. Paraprotein interferences with the Roche MODULAR and Olympus AU2700 were identified by an abnormal reaction absorbance curve obtained directly from the reaction monitor integrated into the automated analyzers and confirmed by at least 1 additional finding. Absorbance curve data from the Olympus AU2700 printouts were plotted manually, whereas similar data Arch Pathol Lab Med Vol 132, February 2008 Paraprotein Interferences With Automated Methods Yang et al 217

2 Figure 1. Scatter plot of total bilirubin (TBIL) concentrations measured by 2 automated chemistry analyzers., Specimens without interferences by paraproteins;, specimens displaying interferences by paraproteins measured by the Roche MODULAR. * To convert to SI units in micromoles per liter, multiply values in milligrams per deciliter by from the Roche MODULAR were used directly. Because it had been shown that the Vitros 950 did not provide spurious results for DBIL and TBIL, it was used for confirmation of interferences. 12,14,16 The Ektachem slide method used in the Vitros 950 was interference free because the specimens had to penetrate through several layers before reaching the reaction layer, and the interfering compounds were probably filtered out. 20 Therefore, when potential interferences were identified and an adequate specimen remained, confirmation of the spurious results was made by additional measurements using the Vitros 950, except for HDL-C, as our instrument did not have this method included in its test menu. Paraprotein interferences also were confirmed by specimens with negative DBIL or negative HDL-C results, specimens with abnormally high TBIL levels from nonicteric patients without evidence of liver diseases and hemolysis, or specimens with DBIL levels higher than TBIL levels. The Roche MODULAR TBIL assay is a chromogenic 2-point end point assay modified from the diazo method 21 that is performed in 1 cuvette. First the patient s serum is incubated with reagent 1 (R1), which contains surfactant to accelerate the reaction and to solubilize protein. Then, reagent 2 (R2) containing a diazonium salt and HCl is added to produce azobilirubin, which is red under acidic conditions. Before R1 is added, the serum specimen should not be turbid. In contrast, the Olympus AU2700 TBIL assay is a 2-cuvette, 1-point end point assay that uses similar reagents. Both the reaction cuvette and the blank cuvette contain surfactant, whereas the chromogen is not added to the blank cuvette. The Roche MODULAR and Olympus AU2700 DBIL assays resemble their manufacturers TBIL assays, other than that a solubilizing agent such as surfactant is not added. When paraprotein interferences with DBIL initially were suspected in the specimens with either negative DBIL values or DBIL concentrations higher than TBIL concentrations, they subsequently were confirmed by abnormal reaction absorbance curves. The Roche HDL-C Plus second-generation assay and the Olympus HDL-C assay are 2-reagent homogenous systems for the selective measurement of serum HDL-C in the presence of other lipoprotein particles. Reagent 1 (R1) selectively forms colorless water-soluble complexes with non-hdl lipoprotein, which is protected from subsequent enzymatic reactions. The HDL-C is then determined enzymatically by the addition of R2 containing cholesterol esterase, cholesterol oxidase, and a chromogen system into the reaction mixture. The HDL-C concentration is determined by the absorbance after the addition of R2 subtracted from the baseline absorbance in the presence of R1. RESULTS The present study included 88 serum specimens from 52 patients with monoclonal gammopathies treated at a major municipal hospital or a university hospital. The paraprotein concentrations of these specimens ranged from 0.5 to 8.9 g/dl with a mean of 3.67 g/dl. The majority (60%) of the paraproteins were immunoglobulin (Ig) G, 18% were IgG, 11% were IgA, and 7% were IgM. Specimens with paraprotein concentrations less than 3.0 g/dl had spurious results in 7.2% of assays; when paraprotein concentrations were 3.0 to 5.0 g/dl, spurious results were seen in 22.4% of assays, and when paraprotein concentrations were greater than 5.0 g/dl, spurious results were seen in 31.3% of assays. Spurious results occurred in 89 of 528 measurements; 29 specimens did not demonstrate any interferences, whereas 26 specimens gave spurious results in 2 to 4 of the 6 assays. Of 88 serum specimens, 4 specimens (4.3%) from 3 patients with monoclonal gammopathies displayed spuriously high Roche MODULAR TBIL concentrations. The interferences in 3 specimens initially were suggested by nonicteric sera with abnormally high TBIL values of greater than 25 mg/dl (428 mol/l), (Figure 1) and subsequently were identified and confirmed along with a fourth specimen by abnormal reaction absorbance curves (Figure 2, C and D) compared with those obtained from normal control serum (Figure 2, A), measurements by the Vitros 950, and icteric serum without measurable paraproteins (Figure 2, B). In our study, the interferences were found exclusively in specimens from patients with IgG or IgG monoclonal gammopathies, and in these 4 specimens, the paraprotein concentrations (mean, 5.45 g/dl) were higher than the overall mean of the 88 monoclonal proteins studied. When we manually performed the Roche MODULAR TBIL assay in test tubes according to manufacturer s suggested protocol using a specimen with the suspected in- 218 Arch Pathol Lab Med Vol 132, February 2008 Paraprotein Interferences With Automated Methods Yang et al

3 Figure 2. Reaction absorbance curves of the Roche MODULAR total bilirubin (TBIL) assay monitored. A, Normal control serum. B, An icteric patient serum without paraprotein. C and D, Nonicteric, nonhemolytic serum specimens containing paraproteins. The gray vertical bars indicate the points at which measurements are taken to calculate results. IgG indicates immunoglobulin G. terference, the specimen was slightly turbid visually before adding R1 but became clear after the addition of R1. However, the addition of R2 caused the reaction mixture to become clear instead of the red translucent appearance observed with the control patient s specimens. When the TBIL concentrations of these sera were measured independently by the Olympus AU2700 and Vitros 950 automated chemistry analyzers, no interferences were identified with these 2 instruments. Using the criteria we established for interference, 39 (44%) of 88 specimens were found to display interferences with DBIL on the Olympus AU2700 (Figure 3). However, only 1 specimen was found to have interferences with DBIL when measured by the Roche MODULAR, and this result was confirmed as spurious by the Vitros 950. We found that these interferences occurred when the reaction absorbance curve of the blank cuvette (DBIL-B) did not parallel the absorbance curve of the reaction cuvette (DBIL-C) (Figure 4). The Olympus instrument was programmed to subtract the absorbance of the blank cuvette from the absorbance of the reaction cuvette at the end point, then tally the difference and convert this difference into a numerical result. If the absorbance of the blank cuvette was lower than the end point absorbance of the reaction cuvette, the instrument reported a positive value (Figure 4, A); in contrast, if the absorbance of the Arch Pathol Lab Med Vol 132, February 2008 Paraprotein Interferences With Automated Methods Yang et al 219

4 Figure 3. Scatter plot of direct bilirubin (DBIL) measurements made by 2 automated chemistry analyzers., Specimens without evidence of paraprotein interferences;, specimens displaying interferences by paraproteins measured by the Olympus AU2700;, specimens displaying interferences by paraproteins measured by the Olympus AU2700 and the Roche MODULAR. * For SI units in micromoles per liter, multiply results in milligrams per deciliter by blank cuvette was higher than the end point absorbance of the reaction cuvette, the instrument reported a negative value (Figure 4, B). Among the 39 specimens with Olympus AU2700 DBIL interferences, 14 specimens had spuriously high values and 25 specimens had spuriously low values. However, the specimens displaying interferences with the Olympus AU2700 DBIL method tended to give irreproducible DBIL values, as demonstrated by a specimen that varied from falsely positive (13.0 mg/dl [222 mol/l]) to falsely negative ( 17.3 mg/dl [296 mol/l]) values when measured repeatedly (data not shown). Figure 5 depicts 31 (35%) of 88 specimens that demonstrated HDL-C interferences with the Roche MODULAR, and of these, 90% were spuriously low. In contrast, 14 (16%) of 88 specimens demonstrated HDL-C interferences when measured by the Olympus AU2700. Of these interfering sera, 7 demonstrated interferences with both methods. Examples of absorbance curves demonstrating interferences with HDL-C as identified by abnormal reaction absorbance curves compared with those from control specimens and confirmed by negative HDL-C values are shown in Figure 6. In Figure 6, B and C, the Roche MOD- ULAR absorbance values of the blank obtained just before the addition of R2 were higher than after the addition of R2, thereby giving spurious results. There was a much higher baseline absorbance before the addition of R2 in the specimens displaying interferences with the Olympus AU2700 HDL-C method compared Figure 4. Absorbance curves from the Olympus AU2700 direct bilirubin (DBIL) assay. A and B, Abnormal reaction absorbance curves of the Olympus AU2700 DBIL assay from 2 patient specimens containing paraproteins. The DBIL assay is a 2-cuvette, end point assay utilizing a sample blank; the sample blank cuvette (DBIL-B) absorbance is subtracted from the color cuvette (DBIL-C) absorbance to determine the net reaction absorbance. The arrows indicate reaction end point. 220 Arch Pathol Lab Med Vol 132, February 2008 Paraprotein Interferences With Automated Methods Yang et al

5 Figure 5. Scatter plot of high-density lipoprotein cholesterol (HDL-C) measurements made by 2 automated chemistry analyzers., Specimens without evidence of paraprotein interferences;, specimens displaying interferences by paraproteins measured by the Roche MODULAR;, specimens displaying interferences by paraproteins measured by the Olympus AU2700; #, specimens displaying interferences by paraproteins measured by the Roche MODULAR and the Olympus AU2700. * For SI units in millimoles per liter, multiply results in milligrams per deciliter by with that in the control specimens; after the addition of R2, absorbance values did not increase as much as expected (Figure 6, D), thereby accounting for the spuriously low values of HDL-C. Visually checking cuvettes containing specimens with the interference also demonstrated white precipitates that were formed after the addition of R1. The paraprotein interferences appear to be concentration dependent, as dilution of patient sera and treatment of the patient by therapeutic plasmapheresis or chemotherapeutic agents causing decreasing monoclonal proteins minimized paraprotein interferences (data not shown). COMMENT Our studies demonstrate that the interferences with TBIL, DBIL, and HDL-C are relatively common in the presence of paraproteins. Comparing results among different analyzers and monitoring absorbance curves led us to conclude that paraprotein interferences are more prevalent than previously reported. Pantanowitz and coworkers 11 found that artifactual TBIL occurred in 2 (1.2%) of 115 patient sera, whereas Nauti et al 14 reported that paraprotein interferences with the Olympus AU2700 DBIL were observed in only 3 (1.5%) of 200 serum specimens. Spuriously low levels of HDL-C on the Roche 917 analyzer caused by paraprotein were found in 2 (13%) of 15 specimens at a major medical center. 12 To our knowledge, paraprotein interferences with HDL-C measured by the Olympus automated chemistry analyzers have not been reported. Our study revealed that the incidence of paraprotein interference with the Olympus AU2700 and Roche MOD- ULAR DBIL assays was 16% and 35%, respectively, in 88 serum specimens. We also found that paraproteins interfere with Olympus AU2700 HDL-C, although the frequency (16%) was not as high as that on the Roche MODULAR (35%). Specimens that gave spurious results in one assay had a high likelihood of yielding a spurious result in other assays. A recent study showed that paraprotein caused only spuriously high levels with Olympus AU2700 DBIL measurements. 14 However, in contrast, our study demonstrated that most DBIL interferences (64%) were spuriously low when measured by the Olympus AU2700 DBIL. Because the Olympus DBIL assay gave exceedingly variable repetitive results ranging from a markedly positive to markedly negative interferences in specimens containing a monoclonal protein, we believe that describing an interference as negative or positive is unsatisfactory. The paraprotein interferences with TBIL, DBIL, and HDL-C appeared to be method dependent. The interferences with TBIL and HDL-C mainly were observed with the Roche MODULAR, while DBIL interferences occurred most frequently with the Olympus AU2700. The Roche TBIL assay is a 1-cuvette, 2-point end point assay. Because the addition of R2 to the reaction mixture changed the assay conditions to more acidic (ph 1 2), paraproteins could form insoluble complexes with R2, which cause the interferences. On the other hand, the Olympus TBIL assay is a 2-cuvette end point assay. Both cuvettes, the sample blank cuvette and the reaction cuvette, contain the same reagents, except the chromogen is only in the reaction cuvette. Therefore, the condition in the assay is not changed as there is no addition of R2. The chromogenic reaction of the DBIL assay occurs at a verylowph( 1). A protein stabilizing agent is included to avoid protein precipitation. It appears that the solubilization capacity of the Olympus AU2700 protein stabilizing agent is not sufficient to prevent the proteins in solution from precipitating at such an acidic ph, which may cause more interferences with DBIL on the Olympus AU2700. As mentioned in the Results, spuriously low HDL-C results observed in the patients with monoclonal gammopathies on the Roche MODULAR and the Olympus AU2700 may be due to the high baseline absorbance in the presence of R1 reagent. The R1 reagent contains a buffer system to selectively form a water-soluble complex Arch Pathol Lab Med Vol 132, February 2008 Paraprotein Interferences With Automated Methods Yang et al 221

6 Figure 6. Comparison of high-density lipoprotein cholesterol (HDL-C) measurements made by 2 automated chemistry analyzers; the Roche MODULAR (A through C) and the Olympus AU2700 (D) HDL-C assay absorbance curves. A, A control patient serum. B and C, Serum specimens containing paraproteins. The gray vertical bars indicate the points at which measurements were taken to calculate results. D, Reaction absorbance curve of the Olympus AU2700 HDL-C assay. The arrows indicate photometric points at which R1 (point 0) and R2 (point 10) reagents were added. The reaction stops at photometric point 27. IgG indicates immunoglobulin G. with low-density lipoprotein, very low-density lipoprotein, and chylomicrons. It may be that the differences of the components or concentrations of the components of the buffer system in the Roche MODULAR and Olympus AU2700 HDL-C assays caused more interferences with HDL-C to be observed on the Roche MODULAR. Our studies demonstrate that paraprotein interferences with DBIL and HDL-C are relatively common, and spurious results of TBIL, DBIL, and HDL-C appear to be methodology and concentration dependent. In a few patients followed for a period of time, treatment that decreased paraprotein concentrations caused interferences to disappear, while disease progression and increasing monoclonal protein concentrations caused interferences to increase. It is important to appreciate the occurrence of these interferences to avoid misinterpretation of the results, especially because low HDL-C ( 40 mg/dl [684 mol/l]) is considered to be an independent risk factor for coronary heart disease. 22 To automatically detect the interference, Smogorzewska et al 12 suggested manually programming the analyzer to trigger a flag when the difference of absorbance between certain points was above a 222 Arch Pathol Lab Med Vol 132, February 2008 Paraprotein Interferences With Automated Methods Yang et al

7 preset value for the Roche MODULAR TBIL assay or the absorbance immediately before the addition of the R2 agent was above a designated value for the Roche AU2700 HDL-C assay. This may help when the abnormal absorbance reaction curves follow a similar kinetic pattern. However, our studies showed that the abnormal reaction curves of the Roche MODULAR TBIL assay and the Olympus AU2700 DBIL assay have different patterns (Figures 2 and 4), which may lead to missing some interferences if a programmable rule is used. An approach that we favor for bilirubin assays is to program the analyzer to automatically flag results for confirmation when specimens have TBIL results greater than 20 mg/dl (342 mol/l) and the icteric flag is not markedly positive. 13 To prevent results preceded by a negative sign from being reported, we recommend programming the laboratory information system to reject patient DBIL and HDL-C results that are preceded by a minus sign. In conclusion, we found that paraprotein interferences with clinical laboratory measurements are far more frequent than previously reported. If an interference occurs with one assay, it is likely that other assays also will be affected. These types of interferences appear in part to be concentration dependent. Specimens with paraproteins cause spurious results by causing precipitation in TBIL, DBIL, and HDL-C assays. References 1. Datta P, Graham GA, Schoen I. Interference by IgG paraproteins in the Jaffe method for creatinine determination. Am J Clin Pathol. 1986;85: Hummel KM, von Ahsen N, Kuhn RB, et al. Pseudohypercreatininemia due to positive interference in enzymatic creatinine measurements caused by monoclonal IgM in patients with Waldenstrom s macroglobulinemia. Nephron. 2000; 86: Yu A, Pira U. False increase in serum C-reactive protein caused by monoclonal IgM-lambda: a case report. Clin Chem Lab Med. 2001;39: Dimeski G, Carter A. Rare IgM interference with Roche/Hitachi modular glucose and gamma-glutamyltransferase methods in heparin samples. Clin Chem. 2005;51: Dorizzi R, Battaglia P, Lora A. Iron measurement in patients with monoclonal immunoglobulin: a further caution. Clin Chem. 1991;37: Bakker AJ. Influence of monoclonal immunoglobulins in direct determinations of iron in serum. Clin Chem. 1991;37: Sinclair D, Smith H, Woodhead P. Spurious hyperphosphataemia caused by an IgA paraprotein: a topic revisited. Ann Clin Biochem. 2004;41: Barutcuoglu B, Parildar Z, Mutaf I, Habif S, Bayindir O. Spuriously elevated inorganic phosphate level in a multiple myeloma patient. Clin Lab Haematol. 2003;25: Savory DJ, Pearce CJ. Paraprotein interference causing pseudohyperphosphataemia: evaluation of an improved methodology. Ann Clin Biochem. 1995; 32: Langman LJ, Allen LC, Romaschin AD. Interference of IgM paraproteins in the Olympus AU800 uric acid assay. Clin Biochem. 1998;31: Pantanowitz L, Horowitz GL, Upalakalin JN, Beckwith BA. Artifactual hyperbilirubinemia due to paraprotein interference. Arch Pathol Lab Med. 2003; 127: Smogorzewska A, Flood JG, Long WH, Dighe AS. Paraprotein interference in automated chemistry analyzers. Clin Chem. 2004;50: Sheppard CA, Allen RC, Austin GE, et al. Paraprotein interference in automated chemistry analyzers. Clin Chem. 2005;51: Nauti A, Barassi A, Merlini G, d Eril GV. Paraprotein interference in an assay of conjugated bilirubin. Clin Chem. 2005;51: Bakker AJ, Zijlstra A, Leemhuis MP. False negative results in the Roche assay for HDL-cholesterol. Ann Clin Biochem. 2003;40: Kadri N, Douville P, Lachance P. Monoclonal paraprotein may interfere with the Roche Direct HDL-C Plus assay. Clin Chem. 2002;48: Tsai LY, Tsai SM, Lee SC, Liu SF. Falsely low LDL-cholesterol concentrations and artifactual undetectable HDL-cholesterol measured by direct methods in a patient with monoclonal paraprotein. Clin Chim Acta. 2005;358: Berth M, Delangehe J. Protein precipitation as a possible important pitfall in the clinical chemistry analysis of blood samples containing monoclonal immunoglobins: 2 case reports and a review of the literature. Acta Clin Belg. 2004; 59: Baca A, Haber RJ, Susishi K, Frost PH, Ng VN. Artifactual undetectable HDL-cholesterol with the Beckman Synchron LX and Vitros 950 assays temporally associated with a paraprotein. Clin Chem. 2004;50: Zaman Z, Sneyers L, Van Orshoven A, Blanckaert N, Marien G. Elimination of paraprotein interference in determination of plasma inorganic phosphate by ammonium molybdate method. Clin Chem. 1995;41: Westwood A. The analysis of bilirubin in serum. Ann Clin Biochem. 1991; 28: Expert Panel on Detection, Evaluation, and Treatment of High Blood Cholesterol in Adults. Executive Summary of The Third Report of The National Cholesterol Education Program (NCEP) Expert Panel on Detection, Evaluation, and Treatment of High Blood Cholesterol in Adults (Adult Treatment Panel III). JAMA. 2001;285: Arch Pathol Lab Med Vol 132, February 2008 Paraprotein Interferences With Automated Methods Yang et al 223

Monoclonal immunoglobulins (Igs) may interfere with

Monoclonal immunoglobulins (Igs) may interfere with Artifactual Hyperbilirubinemia Due to Paraprotein Interference Liron Pantanowitz, MD; Gary L. Horowitz, MD; Jan N. Upalakalin, MD; Bruce A. Beckwith, MD Context. Paraprotein interference in automated chemistry

More information

Seeing is not Believing (13-Nov-2004)

Seeing is not Believing (13-Nov-2004) In: 55th Annual Meeting of the American College of Veterinary Pathologists (ACVP) & 39th Annual Meeting of the American Society of Clinical Pathology (ASVCP), ACVP and ASVCP (Eds.) Publisher: American

More information

Applicable To Employees of Gundersen Boscobel Area Hospital Laboratory and Gundersen Palmer Lutheran Hospital and Clinics Laboratory.

Applicable To Employees of Gundersen Boscobel Area Hospital Laboratory and Gundersen Palmer Lutheran Hospital and Clinics Laboratory. Subject Creatinine COBAS C311 Index Number Lab-8814 Section Laboratory Subsection Regional/Affiliates Location Category Departmental Contact Tilleraas, Betty Last Revised 1/17/2018 References Required

More information

Evaluation of VACUETTE CAT Serum Fast Separator Blood Collection Tube for Routine Chemistry Analytes in Comparison to VACUTAINER RST Tube

Evaluation of VACUETTE CAT Serum Fast Separator Blood Collection Tube for Routine Chemistry Analytes in Comparison to VACUTAINER RST Tube Evaluation of VACUETTE CAT Serum Fast Separator Blood Collection Tube for Routine Chemistry Analytes in Comparison to VACUTAINER RST Tube Background: Greiner-Bio-One, Austria has been selling plastic evacuated

More information

Applicable To Employees of the Gundersen Boscobel Area Hospital Laboratory and Gundersen Palmer Lutheran Hospital and Clinics laboratories.

Applicable To Employees of the Gundersen Boscobel Area Hospital Laboratory and Gundersen Palmer Lutheran Hospital and Clinics laboratories. Subject Calcium C311 Index Number Lab-8810 Section Laboratory Subsection Regional/Affiliates Category Departmental Contact Judy Dayton Last Revised 4/11/2017 References Required document for Laboratory

More information

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date SYNCHRON System(s) Chemistry Information Sheet 2015 Beckman Coulter, Inc. All rights reserved. HDLD HDL Cholesterol 650207 For In Vitro Diagnostic Use ANNUAL REVIEW Rx Only Reviewed by Date Reviewed by

More information

NQLM Nordic Committee for External Quality Assurance Programmes in Laboratory Medicine

NQLM Nordic Committee for External Quality Assurance Programmes in Laboratory Medicine 1(4) NQLM 2001-11-23 Nordic Committee for External Quality Assurance Programmes in Laboratory Medicine NORDIC INTERFERENCE STUDY, March 2001 Effects of bilirubin on some common serum analysis Enclosed

More information

Clinical Chemistry Specific Proteins Presentation

Clinical Chemistry Specific Proteins Presentation Clinical Chemistry Specific Proteins Presentation November 2004 AEROSET and c8000 are registered trademarks of Abbott Laboratories. All other trademarks, brands, trade names and product names are the property

More information

BIOO LIFE SCIENCE PRODUCTS

BIOO LIFE SCIENCE PRODUCTS BIOO LIFE SCIENCE PRODUCTS FOR REFERENCE PURPOSES This manual is for Reference Purposes Only. DO NOT use this protocol to run your assays. Periodically, optimizations and revisions are made to the kit

More information

The Adiponectin Turbidimetric Immunoassay Reagent Kit

The Adiponectin Turbidimetric Immunoassay Reagent Kit The Adiponectin Turbidimetric Immunoassay Reagent Kit Catalogue number: 51010 For the quantitative determination of Adiponectin in human serum and plasma This package insert must be read in its entirety

More information

Estimation of glucose in blood serum

Estimation of glucose in blood serum Estimation of glucose in blood serum Enzymatic estimation of glucose uses a reagent containing two enzymes and a chromogen. Glucose oxidase catalyses the oxidation of glucose to gluconolactone with the

More information

Standard Operating Procedure

Standard Operating Procedure Subject Free Thyroxine (FT4-II) cobas e411 Index Number Lab-1594 Section Regional/Affiliate Subsection Laboratory Category Departmental Contact Goplin, Darcy Last Revised 2/2/2017 References Required document

More information

Hexagon PSA. Design Verification. Contents

Hexagon PSA. Design Verification. Contents Design Verification Hexagon PSA Contents 1. Function...2 2. Sensitivity, Dynamic Range and Traceability...2 Description of Control Materials...2 Analytical Sensitivity...2 3. Clinical Evaluation...3 Evaluation

More information

1 Introduction Imprecision Within-run imprecision, results Day-to-day imprecision, results... 2

1 Introduction Imprecision Within-run imprecision, results Day-to-day imprecision, results... 2 Design Verification liquicolor 1 Introduction... 2 2 Imprecision... 2 2.1 Within-run imprecision, results... 2 2.2 Day-to-day imprecision, results... 2 3 Linearity and Detection Limit... 2 3.1 Linearity...

More information

TABLE OF CONTENTS GENERAL INFORMATION... 1

TABLE OF CONTENTS GENERAL INFORMATION... 1 BIOO RESEARCH PRODUCTS Glucose Assay Kit Manual Catalog #: 5611-01 BIOO Scientific Corp. 2011 TABLE OF CONTENTS GENERAL INFORMATION... 1 Product Description... 1 Procedure Overview... 1 Required Materials

More information

UREA CE Insert. 01 English - Ref.: 27. Ref.:27

UREA CE Insert. 01 English - Ref.: 27. Ref.:27 UREA CE Insert Ref.:27 Intended use. Colorimetric and enzymatic system for urea determination in samples of blood and urine, by end point reaction. Professional use. [Only for in vitro diagnostic use.]

More information

Patient: 55 y female (ambulatory)

Patient: 55 y female (ambulatory) Disclosures Speaking Honoraria Radiometer (Canada) Nova Biomedical, Draeger Roche Diagnostics (Canada) Research Support (Reagents, Instrumentation, Travel) Nova Biomedical Abbott Laboratories (Canada)

More information

J. Nutr. Sci. Vitaminol., 38, , Note. in Tissues

J. Nutr. Sci. Vitaminol., 38, , Note. in Tissues J. Nutr. Sci. Vitaminol., 38, 517-521, 1992 Note A Simple Enzymatic Quantitative in Tissues Analysis of Triglycerides Hiroshi DANNO, Yuu JINCHO, Slamet BUDIYANTO, Yuji FURUKAWA, and Shuichi KIMURA Laboratory

More information

Free Glycerol Assay Kit (Colorimetric)

Free Glycerol Assay Kit (Colorimetric) Product Manual Free Glycerol Assay Kit (Colorimetric) Catalog Number STA-398 100 assays FOR RESEARCH USE ONLY Not for use in diagnostic procedures Introduction Glycerol is the backbone of Triglycerides

More information

References Required document for Laboratory Accreditation by the College of American Pathologists.

References Required document for Laboratory Accreditation by the College of American Pathologists. Subject NT-proBNP Cobas e601 Index Number Lab-4011 Section Laboratory Subsection Chemistry Category Departmental Contact Benjamin Michel Last Revised 10/19/2016 References Required document for Laboratory

More information

Determination of hemoglobin is one of the most commonly

Determination of hemoglobin is one of the most commonly ORIGINAL ARTICLE Multiple-Site Analytic Evaluation of a New Portable Analyzer, HemoCue Hb 201+, for Point-of-Care Testing Sten-Erik Bäck, PhD,* Carl G. M. Magnusson, PhD, Lena K. Norlund, MD, PhD, Henning

More information

Roche/Hitachi - PreciControl ClinChem Multi 2

Roche/Hitachi - PreciControl ClinChem Multi 2 ATRYP Antitrypsin alpha 1 ERM-DA470k/IFCC 9 136 1.36 1.84 0.0184 GPROT Acid glycoprotein alpha 1 CRM 470 2 90.1 0.901 0.890 0.00890 ACP Acid phosphatase total 1-naphthyl phosphate 0.720 43.1 0.00703 0.421

More information

This package insert contains information to run the Cholesterol assay on the ARCHITECT c Systems.

This package insert contains information to run the Cholesterol assay on the ARCHITECT c Systems. en CHOLESTEROL 7D62 304796/R02 CHOLESTEROL This package insert contains information to run the Cholesterol assay on the ARCHITECT c Systems. Read Highlighted Changes Revised October 2012 Package insert

More information

PLASMA LIPOPROTEINS AND LIPIDS DETERMINATION OF PLASMA CHOLESTEROL AND TRIGLICERIDE LEVEL

PLASMA LIPOPROTEINS AND LIPIDS DETERMINATION OF PLASMA CHOLESTEROL AND TRIGLICERIDE LEVEL PLASMA LIPOPROTEINS AND LIPIDS DETERMINATION OF PLASMA CHOLESTEROL AND TRIGLICERIDE LEVEL Lipids are characterized by low polarity and limited solubility in water. Their plasma concentration is about 500-600

More information

PHOS Precipath U plus (10 x 3 ml, for USA) Code Precipath U plus (10 x 3 ml) Code 301

PHOS Precipath U plus (10 x 3 ml, for USA) Code Precipath U plus (10 x 3 ml) Code 301 Order information Analyzer(s) on which cobas c pack(s) can be used 03183793 122 250 tests System ID 07 6614 3 Roche/Hitachi cobas c 311, cobas c 501/502 10759350 190 Calibrator f.a.s. (12 x 3 ml) Code

More information

Experiment 3: Activity Determination

Experiment 3: Activity Determination Experiment 3: Activity Determination Introduction: Specific activity is a method for measuring enzymatic activity and the enzyme purity in a mixture. In order to determine the specific activity of an enzyme,

More information

1. Reagent Store at 2-8 ºC.

1. Reagent Store at 2-8 ºC. LDL Insert Ref.:146 Intended use. System for direct quantitative determination of low density lipoprotein (LDL) in serum or plasma samples. Professional use. [For in vitro diagnostic use.] Test principle.

More information

Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University

Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University Alert or critical values represent those assay results that require prompt, rapid clinical attention to avert significant study-participant

More information

Standard Operating Procedure

Standard Operating Procedure Subject Alkaline Phosphatase C111 Index Number Lab-1504 Section Laboratory Subsection Regional Clinic/Affiliate Hospital Laboratories Category Departmental Contact Kamprud, Elizabeth Last Revised 4/12/2017

More information

Separation of Plasma and Serum and Their Proteins from Whole Blood

Separation of Plasma and Serum and Their Proteins from Whole Blood Separation of Plasma and Serum and Their Proteins from Whole Blood BCH 471 [Practical] BLOOD COMPOSITION Other names to blood cells Red blood cells (erythrocytes) White blood cells (leukocytes) Platelets

More information

Evaluation of a Particle-enhanced Turbidimetric Immunoassay for the Measurement of Immunoglobulin E in an ILab 900 Analyzer

Evaluation of a Particle-enhanced Turbidimetric Immunoassay for the Measurement of Immunoglobulin E in an ILab 900 Analyzer Clinical Chemistry 45:9 1557 1561 (1999) Clinical Immunology Evaluation of a Particle-enhanced Turbidimetric Immunoassay for the Measurement of Immunoglobulin E in an ILab 900 Analyzer Sandra Guaita, Josep

More information

Elevated Urine Zinc Concentration Reduces the Detection of Methamphetamine, Cocaine, THC and Opiates in Urine by EMIT

Elevated Urine Zinc Concentration Reduces the Detection of Methamphetamine, Cocaine, THC and Opiates in Urine by EMIT Journal of Analytical Toxicology 2013;37:665 669 doi:10.1093/jat/bkt056 Advance Access publication July 10, 2013 Article Elevated Urine Zinc Concentration Reduces the Detection of Methamphetamine, Cocaine,

More information

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats Authors: Kate Cote, Ph.D., Graham Bilbrough, MA, VetMB, CertVA, MRCVS,

More information

Β-FRUCTOFURANOSIDASE ENZYME

Β-FRUCTOFURANOSIDASE ENZYME KINETICS ANALYSIS OF Β-FRUCTOFURANOSIDASE ENZYME 2-The effects of enzyme concentration on the rate of an enzyme catalyzed reaction. Systematic names and numbers β-fructofuranosidase (EC 3.2.1.26) Reactions

More information

The analytical phase

The analytical phase The analytical phase Result interpretation Test request Result Sampling Black box: the lab ANALYTICAL PHASE The CASE Uncle Pete, 67 years old Marked abdominal pain 8 pm, ED Acute abdomen? Assessment (+

More information

High-density Lipoprotein Cholesterol (HDL-C) Assay Kit

High-density Lipoprotein Cholesterol (HDL-C) Assay Kit (FOR RESEARCH USE ONLY. DO NOT USE IT IN CLINICAL DIAGNOSIS!) High-density Lipoprotein Cholesterol (HDL-C) Assay Kit (Double reagents) Catalog No: E-BC-K221 Method: Colorimetric method Specification: 96T

More information

C56-A Hemolysis, Icterus, and Lipemia/Turbidity Indices as Indicators of Interference in Clinical Laboratory Analysis; Approved Guideline

C56-A Hemolysis, Icterus, and Lipemia/Turbidity Indices as Indicators of Interference in Clinical Laboratory Analysis; Approved Guideline C56-A Hemolysis, Icterus, and Lipemia/Turbidity Indices as Indicators of Interference in Clinical Laboratory Analysis; Approved Guideline Presented by: Merle B. Smith, MBA, MS, MT(ASCP) Siemens Healthcare

More information

HDL CHOLESTEROL. 01 English - Ref.: 13. Ref.:13. Insert

HDL CHOLESTEROL. 01 English - Ref.: 13. Ref.:13. Insert HDL CHOLESTEROL Insert Ref.:13 Intended use. System for selective precipitation of Low and Very Low Density Lipoproteins (LDL and VLDL) and HDL cholesterol measurement in the supernatant, by an end point

More information

Coping with Analytical Interferences

Coping with Analytical Interferences Coping with Analytical Interferences (Handling Icteric, Hemolytic and Lipaemic Samples) Graham Jones Department of Chemical Pathology St Vincent s Hospital, Sydney Surabaya Indonesia 2016 Acknowledgements

More information

Human LDL Receptor / LDLR ELISA Pair Set

Human LDL Receptor / LDLR ELISA Pair Set Human LDL Receptor / LDLR ELISA Pair Set Catalog Number : SEK10231 To achieve the best assay results, this manual must be read carefully before using this product and the assay is run as summarized in

More information

Urinary 8-Isoprostane ELISA kit

Urinary 8-Isoprostane ELISA kit Urinary 8-Isoprostane ELISA kit Catalog Number: 8isoU Store at -20 C. FOR RESEARCH USE ONLY V.020420 Introduction This competitive ELISA kit is for determination of free and glucronidated 8-isoprostane

More information

ARTICLE. Utility of Direct Measurement of Low-Density Lipoprotein Cholesterol in Dyslipidemic Pediatric Patients

ARTICLE. Utility of Direct Measurement of Low-Density Lipoprotein Cholesterol in Dyslipidemic Pediatric Patients ARTICLE Utility of Direct Measurement of Low-Density Lipoprotein Cholesterol in Dyslipidemic Pediatric Patients Baruch S. Ticho, MD, PhD; Ellis J. Neufeld, MD, PhD; Jane W. Newburger, MD, MPH; Neil Harris,

More information

Human Creatinine Urinary Detection Kit

Human Creatinine Urinary Detection Kit Human Creatinine Urinary CATALOG NO: IRAAKT2509 Detection Kit LOT NO: SAMPLE INTENDED USE The Urinary Creatinine kit is designed to quantitatively measure creatinine present in urine samples. BACKGROUND

More information

S ERUM ALKALINE PHOSPHATASE ACTIVITY is the enzyme analysis most

S ERUM ALKALINE PHOSPHATASE ACTIVITY is the enzyme analysis most Rapid Method for the Quantitative Determination of Serum Alkaline Phosphatase Bernard Klein, Prunella A. Read, and Arthur L. Babson S ERUM ALKALINE PHOSPHATASE ACTIVITY is the enzyme analysis most frequently

More information

QUALITROL 2H Insert. 01 Inglês - Ref.: 72. Ref.:72. Precautions and warnings

QUALITROL 2H Insert. 01 Inglês - Ref.: 72. Ref.:72. Precautions and warnings QUALITROL 2H Insert Ref.:72 Lot 6002 Expiration 2019/08/23 Precautions and warnings Qualitrol 2H is for in vitro diagnostic use only. Attention. It is suggested to verify carefully if the lot number pressed

More information

EXAM COVER SHEET. Course Code: CLS 432. Course Description: Clinical Biochemistry. Final Exam. Duration: 2 hour. 1st semester 1432/1433.

EXAM COVER SHEET. Course Code: CLS 432. Course Description: Clinical Biochemistry. Final Exam. Duration: 2 hour. 1st semester 1432/1433. EXAM COVER SHEET Course Code: CLS 432 Course Description: Clinical Biochemistry Final Exam Duration: 2 hour 1st semester 1432/1433 Student Name: Student Uni No: Part 1 Multiple choice questions Answer

More information

References Required document for Laboratory Accreditation by the College of American Pathologists.

References Required document for Laboratory Accreditation by the College of American Pathologists. Subject Hemoglobin A1c Cobas c501 Index Number Lab-4004 Section Laboratory Subsection Chemistry Category Departmental Contact Benjamin Michel Last Revised 9/12/2016 References Required document for Laboratory

More information

AF HDL and LDL/VLDL Assay Kit

AF HDL and LDL/VLDL Assay Kit A HDL and LDL/VLDL Assay Kit Catalog Number KA1656 100 assays Version: 04 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Principle of the Assay... 3

More information

Human LDL ELISA Kit. Innovative Research, Inc.

Human LDL ELISA Kit. Innovative Research, Inc. Human LDL ELISA Kit Catalog No: IRKTAH2582 Lot No: SAMPLE INTRODUCTION Human low-density lipoprotein (LDL) transports cholesterol from the liver to tissues where it is incorporated into cell membranes.

More information

LDL LD. 01 English - Ref.: 129. Ref.:129. Insert. Intended use. Methodology. Reagents. Test principle. Summary. Precautions and warnings

LDL LD. 01 English - Ref.: 129. Ref.:129. Insert. Intended use. Methodology. Reagents. Test principle. Summary. Precautions and warnings LDL LD Insert Ref.:129 Intended use. System for direct quantitative determination of low density lipoprotein (LDL) in serum or plasma samples. [For in vitro diagnostic use.] Test principle. The method

More information

Enzymatic Assay of ß-GLUCOSIDASE (EC )

Enzymatic Assay of ß-GLUCOSIDASE (EC ) PRINCIPLE: ß-D-Glucoside + H 2 O ß-Glucosidase > D-Glucose + an Alcohol CONDITIONS: T = 37 C, ph = 5.0, A 540nm, Light path = 1 cm METHOD: Colorimetric 1 REAGENTS: A. 100 mm Sodium Acetate Buffer, ph 5.0

More information

Herpes Simplex Virus 2 IgM HSV 2 IgM

Herpes Simplex Virus 2 IgM HSV 2 IgM DIAGNOSTIC AUTOMATION, INC. 21250 Califa Street, Suite 102 and 116, Woodland Hills, CA 91367 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com

More information

Effect of lipemia interferences in routine clinical Biochemical Tests (Intralаboratory Study)

Effect of lipemia interferences in routine clinical Biochemical Tests (Intralаboratory Study) Standard Scientific Research and Essays Vol 3(7): 187-191, July 2015 (ISSN: 2310-7502) http://www.standresjournals.org/journals/ssre Research Article Effect of lipemia interferences in routine clinical

More information

For in vitro diagnostic use only. For Rx use only. Reviewed by Date Reviewed by Date

For in vitro diagnostic use only. For Rx use only. Reviewed by Date Reviewed by Date AU US Instructions For Use 2018 Beckman Coulter, Inc. All rights reserved. TRIG TRIGLYCERIDE OSR60118 4 x 20 ml R1, 4 x 5 ml R2 OSR61118 4 x 50 ml R1, 4 x 12.5 ml R2 OSR66118 4 x 173 ml R1, 4 x 48 ml R2

More information

IGG Diluent NaCl 9 % (50 ml) System-ID English System information For cobas c 311/501 analyzers:

IGG Diluent NaCl 9 % (50 ml) System-ID English System information For cobas c 311/501 analyzers: Order information Analyzer(s) on which cobas c pack(s) can be used 03507432 190 150 tests System-ID 07 6787 5 cobas c 311, cobas c 501/502 11355279 216 Calibrator f.a.s. Proteins (5 1 ml) Code 656 03121305

More information

Influenza A IgG ELISA

Influenza A IgG ELISA Influenza A IgG ELISA For the qualitative determination of IgG-class antibodies against Influenza A virus in human serum or plasma (citrate, heparin). For Research Use Only. Not For Use In Diagnostic Procedures.

More information

Comparison of Manual, Automated, and "Kit" Methods for the Measurement of Glucose. and Urea in Plasma

Comparison of Manual, Automated, and Kit Methods for the Measurement of Glucose. and Urea in Plasma Comparison of Manual, Automated, and "Kit" Methods for the Measurement of Glucose and Urea in Plasma ROBERT C. ELSER, B.S., AND JOHN SAVORY, PH.D. I'ulhology Department, University of Florida College of

More information

Mouse Serum Anti-HDM IgE Antibody Assay Kit

Mouse Serum Anti-HDM IgE Antibody Assay Kit Mouse Serum Anti-HDM IgE Antibody Assay Kit Catalog # 3037 For Research Use Only - Not Human or Therapeutic Use INTRODUCTION Asthma is a common chronic inflammatory disease that affects 300 million people

More information

Comparison of a Homogeneous Assay With a Precipitation Method for the Measurement of HDL Cholesterol in Diabetic Patients

Comparison of a Homogeneous Assay With a Precipitation Method for the Measurement of HDL Cholesterol in Diabetic Patients Clinical Care/Education/Nutrition O R I G I N A L A R T I C L E Comparison of a Homogeneous Assay With a Precipitation Method for the Measurement of HDL Cholesterol in Diabetic Patients TONNY JENSEN, MD

More information

Free Fatty Acid Assay Kit (Fluorometric)

Free Fatty Acid Assay Kit (Fluorometric) Product Manual Free Fatty Acid Assay Kit (Fluorometric) Catalog Number STA-619 100 assays FOR RESEARCH USE ONLY Not for use in diagnostic procedures Introduction Triglycerides (TAG) are a type of lipid

More information

Fructose-6-Phosphate Colorimetric Assay Kit

Fructose-6-Phosphate Colorimetric Assay Kit Fructose-6-Phosphate Colorimetric Assay Kit Catalog No. KM0060 Detection and Quantification of Fructose-6-Phosphate Concentrations in Biological Samples. Research Purposes Only. Not Intended for Diagnostic

More information

B. 1% (w/v) Salicin Substrate Solution (Salicin) (Prepare 50 ml in Reagent A using Salicin, Sigma Prod. No. S-0625.)

B. 1% (w/v) Salicin Substrate Solution (Salicin) (Prepare 50 ml in Reagent A using Salicin, Sigma Prod. No. S-0625.) SIGMA QUALITY CONTROL TEST PROCEDURE (Q]\PDWLFÃ$VVD\ÃRIÃ */8&26,'$6( PRINCIPLE: 'Glucoside + H 2 O Glucosidase > D-Glucose + an Alcohol CONDITIONS: T = 37 C, ph = 5.0, A 540nm, Light path = 1 cm METHOD:

More information

Kinetic assay of serum and urine for urea with use of urease and leucine dehydrogenase

Kinetic assay of serum and urine for urea with use of urease and leucine dehydrogenase Clinical Chemistry 43:10 1932 1936 (1997) Automation and Analytical Techniques Kinetic assay of serum and urine for urea with use of urease and leucine dehydrogenase Yoshitaka Morishita, 1 * Kiyoshi Nakane,

More information

Creatinine Jaffé Gen.2 Indicates cobas c systems on which reagents can be used Order information

Creatinine Jaffé Gen.2 Indicates cobas c systems on which reagents can be used Order information Indicates cobas c systems on which reagents can be used Order information Roche/Hitachi cobas c systems cobas c 311 cobas c 501 700 tests Cat. No. 04810716 190 System-ID 07 6928 2 Calibrator f.a.s. (12

More information

See external label 96 tests HSV 2 IgA. Cat #

See external label 96 tests HSV 2 IgA. Cat # DIAGNOSTIC AUTOMATION, INC. 23961 Craftsman Road, Suite D/E/F, Calabasas, CA 91302 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com See external

More information

Human Hemoglobin Colorimetric Detection Kit

Human Hemoglobin Colorimetric Detection Kit Human Hemoglobin Colorimetric Detection Kit CATALOG NO: IRAAKT2522 LOT NO: SAMPLE INTENDED USE The Hemoglobin detection kit is designed to quantitatively measure all forms of hemoglobin present in blood

More information

Mechanisms of dopamine and dobutamine interference in biochemical tests that use peroxide and peroxidase to generate chromophore

Mechanisms of dopamine and dobutamine interference in biochemical tests that use peroxide and peroxidase to generate chromophore Clinical Chemistry 44:1 155 160 (1998) General Clinical Chemistry Mechanisms of dopamine and dobutamine interference in biochemical tests that use peroxide and peroxidase to generate chromophore Brad S.

More information

APOB (Human) ELISA Kit

APOB (Human) ELISA Kit APOB (Human) ELISA Kit Catalog Number KA4330 96 assays Version: 01 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle of the Assay...

More information

Aina Blood Monitoring System

Aina Blood Monitoring System Aina Blood Monitoring System Analytical Performance Summary The Aina Blood Monitoring System is a high quality and versatile multi-parameter diagnostic platform that is CE-marked and approved for sale

More information

Exendin-4 (Exenatide) ELISA Kit

Exendin-4 (Exenatide) ELISA Kit Exendin-4 (Exenatide) ELISA Kit Catalog: DEIABL227 For the quantitative determination of Exendin-4 in serum or plasma using competitive ELISA method For Research Use Only. Protocol Provided for Informational

More information

Albumin Quantitation by Dye Binding. and Salt FractionationTechniques. Materials and Methods

Albumin Quantitation by Dye Binding. and Salt FractionationTechniques. Materials and Methods Albumin Quantitation by Dye Binding and Salt FractionationTechniques Don S. Miyada, Vernon Baysinger, Solomon Notrica, and R. M. Nakamura Two dye-binding reagents, -(.hyd roxybenzeneazo)benzoic acid (HABA)

More information

Lipid Peroxidation Assay

Lipid Peroxidation Assay Package Insert Lipid Peroxidation Assay 96 Wells For Research Use Only v. 1.0 Eagle Biosciences, Inc. 82 Broad Street, Suite 383, Boston, MA 02110 Phone: 866-419-2019 Fax: 617-419-1110 INTRODUCTION Lipid

More information

LPO-CC (Lipid Peroxides) Kit

LPO-CC (Lipid Peroxides) Kit K-ASSAY LPO-CC (Lipid Peroxides) Kit For the Quantitative Colorimetric Determination of High Levels of Lipid Peroxides in Serum, Plasma and Other Biological Samples. Cat. No. CC-004 For Research Use Only.

More information

Correcting laboratory results for the effects of interferences: an approach incorporating uncertainty of measurement

Correcting laboratory results for the effects of interferences: an approach incorporating uncertainty of measurement Original Article Annals of Clinical Biochemistry 2015, Vol. 52(2) 226 231! The Author(s) 2014 Reprints and permissions: sagepub.co.uk/journalspermissions.nav DOI: 10.1177/0004563214533516 acb.sagepub.com

More information

ab65390 HDL and LDL/VLDL Cholesterol Assay kit

ab65390 HDL and LDL/VLDL Cholesterol Assay kit ab65390 HDL and LDL/VLDL Cholesterol Assay kit (Colorimetric/Fluorometric) Instructions for Use For the rapid, sensitive and accurate measurement of HDL and LDL/VLDL Cholesterol in various samples. This

More information

Original Article Interfering effect of bilirubin on the determination of alkaline phosphatase

Original Article Interfering effect of bilirubin on the determination of alkaline phosphatase Int J Clin Exp Med 2014;7(11):4244-4248 www.ijcem.com /ISSN:1940-5901/IJCEM0002083 Original Article Interfering effect of bilirubin on the determination of alkaline phosphatase Zhong Wang, Hongyan Guo,

More information

INSTRUCTIONS FOR USE

INSTRUCTIONS FOR USE VITROS Chemistry Products Reagent 680 1728 Rx ONLY Intended Use For in vitro diagnostic use only. VITROS Chemistry Products Reagent is used on the VITROS 5,1 FS Chemistry System, the VITROS 4600 Chemistry

More information

Triglyceride determination

Triglyceride determination Triglyceride determination Introduction: - Triglycerides are esters of fatty acids and are hydrolyzed to glycerol and free fatty acids (by lipase) - Triglyceride determinations when performed in conjunction

More information

Nitrate/Nitrite Assay Kit Manual Catalog #

Nitrate/Nitrite Assay Kit Manual Catalog # BIOO RESEARCH PRODUCTS Nitrate/Nitrite Assay Kit Manual Catalog # 1305-01 This kit is manufactured to the international quality standard ISO 9001:2008. ISO CI#: SARA-2009-CA-0114-01-B BIOO Scientific Corp.2011

More information

VetScan UA Urine Analyzer Clinical Performance Larry Lem, PhD & Andrew Rosenfeld, DVM, ABVP

VetScan UA Urine Analyzer Clinical Performance Larry Lem, PhD & Andrew Rosenfeld, DVM, ABVP VetScan UA Urine Analyzer Clinical Performance Larry Lem, PhD & Andrew Rosenfeld, DVM, ABVP Abstract Background: The VetScan UA urine strip reader is an analyzer designed to provide semi-quantitative and

More information

Total Bilirubin Assay Kit Manual Catalog #:

Total Bilirubin Assay Kit Manual Catalog #: Total Bilirubin Assay Kit Manual Catalog #: 3460-10 TABLE OF CONTENTS GENERAL INFORMATION... 2 Product Description... 2 Procedure Overview... 2 Kit Contents, Storage and Shelf Life... 2 Required Materials

More information

Evaluation of VACUETTE Urine CCM tube for Clinical Chemistry

Evaluation of VACUETTE Urine CCM tube for Clinical Chemistry Evaluation of VACUETTE Urine CCM tube for Clinical Chemistry Background The VACUETTE Urine CCM tube is for the collection, transport and storage of urine samples for urine culture and urinalysis in the

More information

CoQ10(Coenzyme Q10) ELISA Kit

CoQ10(Coenzyme Q10) ELISA Kit CoQ10(Coenzyme Q10) ELISA Kit Catalogue No.: EU0196 Size: 48T/96T Reactivity: Universal Detection Range: 0.781-50ng/ml Sensitivity:

More information

VITROS MicroSlide Assay Summary

VITROS MicroSlide Assay Summary ACET Acetaminophen ALB Albumin EDTA 10 9 TDM PV Specialty 5.5 4 PV Isotonic saline or 10 200 μg/ml 66 1323 μmol/l (μmol/l = μg/ml x 6.616) 1.00 6.00 g/dl 10.0-60.0 g/l (g/l = g/dl x 10) Therapeutic: 670

More information

Mouse Anti-HDM IgG Antibody Assay Kit

Mouse Anti-HDM IgG Antibody Assay Kit Mouse Anti-HDM IgG Antibody Assay Kit Catalog # 3030 For Research Use Only - Not Human or Therapeutic Use INTRODUCTION Asthma is a common chronic inflammatory disease that affects 300 million people of

More information

LDLC3. English System information For cobas c 311/501 analyzers:

LDLC3. English System information For cobas c 311/501 analyzers: Order information Analyzer(s) on which cobas c pack(s) can be used 07005717 190 LDL Cholesterol Gen.3 (200 tests) System ID 07 7565 7 Roche/Hitachi cobas c 311, cobas c 501/502 12172623 122 Calibrator

More information

Lipid Concentration And The Extent Of Their Peroxidation In Nigerian Hypertensives

Lipid Concentration And The Extent Of Their Peroxidation In Nigerian Hypertensives ISPUB.COM The Internet Journal of Health Volume 8 Number 1 Lipid Concentration And The Extent Of Their Peroxidation In Nigerian Hypertensives P Igbinaduwa, B Igbinaduwa, I Oforofuo Citation P Igbinaduwa,

More information

3 Linearity and Detection Limit Linearity Detection limit... 4

3 Linearity and Detection Limit Linearity Detection limit... 4 Design Verification liquicolor 1 Introduction... 2 2 Imprecision, Reproducibility... 2 2.1 Data from internal study... 2 2.2 Date from external study (Cholesterol Reference Method Laboratory Network)...

More information

Apolipoprotein A1 (Apo A1) ELISA

Apolipoprotein A1 (Apo A1) ELISA Package Insert Apolipoprotein A1 (Apo A1) ELISA 1 x 96 Wells For Research Use Only (RUO). Not for use in clinical, diagnostic or therapeutic procedures. v. 1.0 Eagle Biosciences, Inc. 20A Northwest Blvd.,

More information

CLINICAL CHEMISTRY REAGENTS. Product Profile

CLINICAL CHEMISTRY REAGENTS. Product Profile Product Profile Why Clinical Chemistry Reagents? Quantitative determination of specific analytes associated with various types of disease. Diagnosis Identifying a disease already present Prognosis Forecasting

More information

EFFECT OF HIGH SALT CONCENTRATIONS ON COLOR PRODUCTION OF THE BIURET REACTION FOR PROTEIN ANALYSIS

EFFECT OF HIGH SALT CONCENTRATIONS ON COLOR PRODUCTION OF THE BIURET REACTION FOR PROTEIN ANALYSIS EFFECT OF HIGH SALT CONCENTRATIONS ON COLOR PRODUCTION OF THE BIURET REACTION FOR PROTEIN ANALYSIS HAROLD L. ROSENTHAL, PH.D., AND TOYOKO KAWAKAMI, M.T. (ASC1>) Division of Biochemistry, Department of

More information

The total protein test is a rough measure of all of the proteins in the plasma. Total protein measurements can reflect:

The total protein test is a rough measure of all of the proteins in the plasma. Total protein measurements can reflect: Proteins Part 2 Introduction The total protein test is a rough measure of all of the proteins in the plasma. Total protein measurements can reflect: nutritional status, kidney disease, liver disease, and

More information

Glutathione Assay Kit

Glutathione Assay Kit Glutathione Assay Kit Catalog Number KA1649 250 assays Version: 02 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle of the Assay...

More information

DIAGNOSTIC AUTOMATION, INC.

DIAGNOSTIC AUTOMATION, INC. DIAGNOSTIC AUTOMATION, INC. 23961 Craftsman Road, Suite E/F, Calabasas, CA 91302 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com See external

More information

Evaluation of the BiliChek Noninvasive Bilirubin Analyzer for Prediction of Serum Bilirubin and Risk of Hyperbilirubinemia

Evaluation of the BiliChek Noninvasive Bilirubin Analyzer for Prediction of Serum Bilirubin and Risk of Hyperbilirubinemia Clinical Chemistry / Evaluation of Transcutaneous Bilirubin Evaluation of the BiliChek Noninvasive Bilirubin Analyzer for Prediction of Serum Bilirubin and Risk of Hyperbilirubinemia Brad S. Karon, MD,

More information

Ascorbic Acid Assay Kit

Ascorbic Acid Assay Kit Ascorbic Acid Assay Kit Catalog Number KA1660 100 assays Version: 04 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Principle of the Assay... 3 General

More information

QMS TACROLIMUS APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800

QMS TACROLIMUS APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800 QMS TACROLIMUS APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800 Beckman Coulter Reagent REF A53727 The QMS Tacrolimus Immunoassay is intended for the quantitative determination of tacrolimus in human whole

More information

Date... Name... Group... Urine sample (Tube No 2)

Date... Name... Group... Urine sample (Tube No 2) Date... Name... Group... Instructions for the practical lesson on biochemistry Topic: Non-protein nitrogen compounds Task 1: Estimation of creatinine in serum and urine 1. Trichloroacetic acid 1.22 mol/l

More information