Correcting laboratory results for the effects of interferences: an approach incorporating uncertainty of measurement

Size: px
Start display at page:

Download "Correcting laboratory results for the effects of interferences: an approach incorporating uncertainty of measurement"

Transcription

1 Original Article Annals of Clinical Biochemistry 2015, Vol. 52(2) ! The Author(s) 2014 Reprints and permissions: sagepub.co.uk/journalspermissions.nav DOI: / acb.sagepub.com Correcting laboratory results for the effects of interferences: an approach incorporating uncertainty of measurement Graham RD Jones 1 and Robert C Hawkins 2 Abstract Background: Results of numerical pathology tests may be subject to interference and many laboratories identify such interferences and withhold results or issue warnings if clinically erroneous results may be issued. Some laboratories choose to correct for the effect of interferences, with the uncertainty of the correction noted as a limitation in this process. We investigate the effect of correcting for the effect of interferences on the ability to release results within defined error goals using the effect of in-vitro haemolysis on serum potassium measurement as an example. Methods: A model was developed to determine the uncertainty of a result corrected for the effect of an interferent with a linear relationship between concentration and effect. The model was used to assess the effect of correction on the results which could be released within specified accuracy criteria. Results: Using the effects of haemolysis on potassium results as an example, the maximum amount of haemolysis in a sample that would change the result by > 0.5 mmol/l, with a frequency of 5%, was increased from approximately 1100 mg/l (no correction) to 8000 mg/l (with correction). Conclusions: With modelling of the factors related to the uncertainties of results in the presence of interferences, it is possible to release results in the presence of significantly higher concentrations of interferences after correction than without correction. Correction of a result for a known bias and allowance for the uncertainty of the correction can be considered consistent with the guide to the expression of uncertainty in measurement (GUM). Keywords Analytical systems, laboratory methods, quality assurance and control, laboratory methods Accepted: 7th April 2014 Introduction For some numerical laboratory tests the presence of potential interferents is assessed so that the laboratory can provide a consistent response. Examples include the identification of haemolysis or icterus on routine chemistry samples using spectrophotometry. 1 In some cases, these (semi) quantitative measurements closely approximate the concentration of haemoglobin or bilirubin while in other systems, only a qualitative grading is provided. In response to the presence of an interferent laboratories may respond in a number of ways. These responses may include withholding results when the interferent exceeds a specified level; issuing the results 1 St Vincent s Hospital, Sydney, Australia 2 Tan Tock Seng Hospital, Singapore Corresponding author: Graham RD Jones, St Vincent s Hospital, Sydney, New South Wales 2010, Australia. gjones@stvincents.com.au

2 Jones and Hawkins 227 with a generalised warning comment about possible effects of the interferent; or issuing the result with an estimation of the amount of change that may have occurred due to the presence of the interference. An alternative approach is to correct for the effect of the interference with the aim of releasing more results. 2 This is a controversial practice with mixed views on its validity and safety. 1,3 7 Some authors have suggested that correction can be done with the associated uncertainty reported in a qualitative or semi-quantatitive manner. 5,7 10 For any response to an interferent the laboratory must determine several factors. One factor is the maximal allowed change in the result due to the interference. This is based on the likely clinical effect of an error of the allowed magnitude. The other key factor is the amount of interferent that may cause the allowed change. This may be based on information supplied by the kit manufacturer, obtained from the literature, or on local experimental evidence. The experimental data on the effect of interferences from any of these sources will show a range of different responses, provided multiple experiments have been performed, which can be described with an average response and the uncertainty about that average. An additional component in determining a response to interferences is the percentage of results which may be erroneous by more than the maximal allowed. The simplest model is based on determining the average concentration of the interferent which produces the maximal allowed change. This model does not account for any uncertainty in the interferent effect and sets the fraction of results affected by more than the maximal allowable change at 50%. A more sophisticated model will include the uncertainty of the effect of the interferent as well as establishing the allowable fraction of results that may exceed this limit. In this paper, we analyse the effect of correction for interferences on the ability to release results within defined standard of accuracy when the uncertainties of the measurement of the measurand of interest, the measurement of the interferent and the slope of the correction algorithm are considered. The correction for known causes of bias, and including consideration of the uncertainty of that correction, is recommended in standard metrology according the Guide to the Expression of Uncertainty of Measurement (GUM). 11 The study is limited to situations where the concentration of the interferent is measured and there is a linear relationship between the concentration of the interferent and the effect on the measurement of the measurand of interest. Common examples where this approach may be applied are the effect of in-vitro haemolysis on results for serum potassium, creatine kinase, folate or lactate dehydrogenase or bilirubin on some assays for creatinine, magnesium or total protein. 12 In this paper, data for the effect of in-vitro haemolysis on potassium measurement used as an example. Methods A model to quantitate the effect of interference on results was developed using Microsoft Excel. The model predicted the amount of interferent which would produce more than a fixed percentage of results beyond acceptable limits, with and without correction for the effect of the interferent. The following inputs were included in the model: measured measurand concentration in the absence of interferent (A o ); uncertainty of the measurement of the measurand concentration expressed as co-efficient of variation (CV A ); measured interferent concentration (B); uncertainty of the measurement of the interferent expressed as co-efficient of variation (CV B ); slope of the effect of the interferent on measurement of measurand (C); uncertainty of the slope of the effect of the interferent on measurement of measurand (CV C, see Appendix for terminology summary). The model also required a maximal allowable error and maximal percentage erroneous results. The inputs to the model were varied and the maximal allowable interferent concentration determined for corrected and uncorrected results. The study was limited to interferents with a linear effect on the result. The combined uncertainty of measurement of A in the absence of interferent is: U A ¼ðCV A AÞ F ðf ¼ 1:96 for two tailed errors and 1:65 for one tailed errorsþ ð1þ Based on the above description it follows that for a given concentration A in the presence of B the best estimate of the measured concentration AB is: AB ¼ Ao þ (B C). Then U AB is the combined uncertainty of the measurand measurement at the measured concentration and the uncertainty of the correction, where the uncertainty of the correction is the combined uncertainty of measurement of the interferent and the uncertainty of the slope of the interferent effect and is given as: U AB ¼ p ððcv A ðao þðb CÞÞÞ 2 þðbc p ðcv 2 C þ CV2 B ÞÞ2 Þ F ðf ¼ 1:96 for two tailed errors and 1:65 for one tailed errorsþ ð2þ The model predicts, for known input values, the best estimate of AB and the 95% confidence interval for determination of AB. A one-tailed probability is used

3 228 Annals of Clinical Biochemistry 52(2) for interpretation when the result of the interference is to adjust results in one direction. The effect of correction for the interferent effect was modelled by removing the bias predicted by the amount of interferent and calculating the uncertainty with the same equation. The average and uncertainties of the slope of the interference effect and the uncertainty of the measurement of the interferent used to predict the effect of the presence of an interferent, are the same as should be applied to remove the effect of an interferent. By way of example the model was populated with representative data for the effect of haemolysis on potassium measurements. The following parameters were based on previously published data examining the effect of haemolysis on serum potassium measurement on a Roche Diagnostics 917 clinical chemistry analyser. 13 Haemolysis was assessed by H index measurement in mg/dl and converted to mg/l for further analysis. This spectrophotometric measurement correlates well with chemical haemoglobin measurement. 14,15 Input variables for the model were: C (K/Hb): mmol/mg, CV C : 9%, CV A : 2%, CV B : 5%. 13 An allowable error due to the combined effect of haemolysis and analytical imprecision was taken as 5% of results outside 0.5 mmol/l. Further analysis was done using different estimates of CV c with other parameters unchanged to assess the effect of the uncertainty of the interferent effect on the reportability of results. Results The model was tested using data for the effect of haemolysis on potassium results. Figure 1 shows the expected range of values for measured potassium in samples with various amounts of haemolysis for a true value of 5.0 mmol/l, a CV A of 2%, a CV B of 5%, a slope for C of and a CV C of 9%. It can be seen that it is predicted that an amount of haemolysis releasing approximately 1100 mg/l haemoglobin will produce 5% of results outside the selected error limits of 0.5 mmol/l potassium. Figure 2 shows the expected range of values for measured potassium under the same conditions after correction for the average effect of haemolysis. In this example, the use of the corrected results allows release of results up to a haemolysis amount of approximately 8000 mg/l as opposed to an amount of 1100 mg/l for the raw results. Alternatively for results up to an haemolysis indicated by 1100 mg/l the 95% confidence interval of the results error is less than 0.3 mmol/l compared with up to 0.5 mmol/l for the raw results. The model was tested with a wide range of inputs for the uncertainty of the slope of the interference effect with other parameters unchanged (Figure 3). It can be seen that the effect of this uncertainty has different effects on ability to release corrected and uncorrected results, however in all cases more corrected results are able to be released. Changes in all other parameters K (mmol/l) H index (mg/l) Figure 1. Effect of haemolysis on potassium measurement before correction. The solid black line is the average, and the upper and lower black short-dashed lines are the 95% confidence intervals for the range of expected results for a sample with 5.0 mmol/l potassium measured with an assay with a CV of 2% and effect of haemolysis as described in the text. The grey short-dashed lines indicate the 95% one-tailed uncertainty limits. The long-dashed horizontal lines indicate the selected allowable error limits of 0.5 mmol/l. At haemolysis indicated by haemoglobin greater than 1100 mg/l (indicated by the arrow) there is a greater than 5% chance that the measurement result will be outside the allowable error limits.

4 Jones and Hawkins K (mmol/l) H index (mg/l) Figure 2. Effect of haemolysis on potassium measurement after correction. The solid line represents the average and the upper and lower black short-dashed lines are the 95% confidence intervals for the range of expected results for a sample with 5.0 mmol/l potassium measured with an assay with a CV of 2% and effect of haemolysis as described in the text after correction for the effect of the haemolysis. The horizontal long-dashed lines indicate the selected allowable error limits of 0.5 mmol/l. At haemolysis indicated by haemoglobin greater than 8000 mg/l (indicated by the vertical arrow) there is a greater than 5% chance that the measurement result will be outside the allowable error limits. Repor ng Limit for haemmolysis (mg/l) % 10% 15% 20% 25% 30% 35% 40% CVc (%) Figure 3. Effect of uncertainty of the slope of the interference effect (CV c ) on the haemolysis limit for releasing potassium results, With all parameters unchanged from the model as shown in Figures 1 and 2, with the exception of CV c, the upper haemolysis limit for releasing results is shown for uncorrected results (black unbroken line) and for corrected results (grey dashed line). The vertical arrow shows the values obtained in the model with a CV c of 9% as used in Figures 1 and 2.

5 230 Annals of Clinical Biochemistry 52(2) were assessed and in all cases the use of results corrected for the effect of the interferent allowed results to be released at a higher interferent concentration compared with uncorrected results, or for the same amount of interference of the expected error was smaller. Discussion Laboratories have a variety of options when handling analytical interference affecting laboratory measurements. These include the following. 1. Reporting all results uncorrected for interference. 2. Reporting results uncorrected for interference up to a given interferent threshold and suppressing results above this threshold. 3. Reporting semi-quantitative results based on corrected results for a given interferent threshold and suppressing results above this threshold. 4. Reporting all results corrected for interference. 5. Reporting results corrected for interference up to a given interferent threshold and suppressing results above this threshold. The best alternative is that which minimises the unacceptable bias rate in reported results. From inspection of equations (1) and (2), it can be seen that a corrected result inevitably has a greater uncertainty in the reported value compared with a result from a sample without an interferent. However, as shown in Figures 1 and 2, much higher interferent concentrations can be tolerated with the corrected model than the uncorrected results before the bias rate becomes unacceptable. In this study, the definition of bias as 0.5 mmol/l with a maximally acceptable rate of 5% was arbitrary but the superiority of the corrected model over the uncorrected approach would seem universally true in the absence of grossly elevated values for CV b and/or CV c. Thus it would appear that options 4 and 5 above are preferable to 1 and 2 in minimising the rate of unacceptably biased results. This finding remains true for different values for the uncertainty of the slope of the interference correction. With any of the responses above laboratories may choose to add a comment indicating the uncertainty of the result due to the process adopted. Approach 3 above can be considered as an expression of the uncertainty, but without providing the best estimate of the result. Previous authors have indicated that correction for interferences can lead to increased uncertainty of the results. On this basis it has been advised that correction for the effect of interferences is not recommended. We agree that any correction process adds uncertainty to a result but would make three additional observations. Firstly that uncorrected results in the presence of an interference display the same uncertainty in their difference from the true value as is seen with corrected results. Secondly that any evidence-based quantitative response to an interferent requires an assessment of the uncertainty of the effect of that interferent. Thirdly we have demonstrated that correcting for the effect of the interference reduces the expected uncertainty of the issued results. As indicated above the removal of a known source bias, and inclusion of the uncertainty of that removal process in the final estimate of the uncertainty of the result of this is consistent with recommendations in the Guide to the Expression of Uncertainty of Measurement (GUM). We believe that this approach may allow more results to be issues from routine laboratories and the uncertainty of released results can be reduced. A key component of the approach is valid comprehensive studies to understand the effect of an interference. Such studies must accurately replicate the real world setting and take all relevant factors into account. The recently published guidelines from the Clinical and Laboratory Standards Institute on serum indices as indicators of interference provide standardised protocols for use by laboratories and diagnostic manufacturers. 1 The resolution of interference measurements available (semi-quantitative or qualitative) will limit the utility of this approach. Exceptions to the application of the correction, for example intravascular rather than in-vitro haemolysis, must be identified and acted on appropriately. It is appreciated that assessment of free haemoglobin concentration in plasma and serum is a incomplete measure of all preanalytical damage to blood constituents (red cells, white cells, platelets) and truly reflects only erythrolysis. 16 If there are multiple sources of interference, e.g. fragile white cells in leukaemia or thrombocytosis affecting potassium results, these could be addressed with separate correction equations provided the data supported a linear relationship and the uncertainties were small enough to make correction worthwhile. Inadequate experimental data may lead to incorrect corrections, in the same way as it may lead to incorrect decisions about whether to release a result of not in the presence of an interferent when corrections are not performed. This study indicates that issuing of results corrected for the effect of interference is superior to uncorrected results when the uncertainties of each final result are included in the decision. The model described here is limited to linear responses, however the principle may be applicable in different settings with different measurable interferences. Declaration of conflicting interest None declared.

6 Jones and Hawkins 231 Funding This research received no specific grant from any funding agency in the public, commercial, or not-for-profit sectors. Ethical approval Not applicable. Guarantor GJ. Contributorship GJ conceived the study. RH researched literature and provided supporting data. Both authors reviewed and edited the manuscript and approved the final version of the manuscript. References 1. Clinical and Laboratory Standards Institute. Hemolysis, icterus and lipaemia/turbidity indices as indicators of interference in clinical laboratory analysis; approved guideline. 3rd ed. Wayne: PA: Clinical and Laboratory Standards Institute, Vermeer HJ, Steen G, Naus AJ, et al. Correction of patient results for Beckman Coulter LX-20 assays affected by interference due to hemoglobin, bilirubin or lipids: a practical approach. Clin Chem Lab Med 2007; 45: Hawkins RC. Hemolysis interference in the ortho-clinical diagnostics vitros ECi ctni assay. Clin Chem 2003; 49: World Health Organisation. Use of anticogulants in diagnostic investigations. Geneva: World Health Organisation, Owens H, Siparsky G, Bajaj L, et al. Correction of factitious hyperkalemia in hemolyzed specimens. Am J Emerg Med 2005; 23: Mansour MM, Azzazy HM and Kazmierczak SC. Correction factors for estimating potassium concentrations in samples with in vitro hemolysis: a detriment to patient safety. Arch Pathol Lab Med 2009; 133: Jeffery J, Sharma A and Ayling RM. Detection of haemolysis and reporting of potassium results in samples from neonates. Ann Clin Biochem 2009; 46: Dimeski G, Clague AE and Hickman PE. Correction and reporting of potassium results in haemolysed samples. Ann Clin Biochem 2005; 42: Hawkins RC. Correction and reporting of potassium results in haemolysed samples. Ann Clin Biochem 2006; 43: Zou J, Nolan DK, LaFiore AR, et al. Estimating the effects of hemolysis on potassium and LDH laboratory results. Clin Chim Acta 2013; 421: Joint Committee for Guides in Metrology. Evaluation of measurement data Guide to the expression of uncertainty in measurement Joint Committee for Guides in Metrology, Jay DW and Provasek D. Characterization and mathematical correction of hemolysis interference in selected Hitachi 717 assays. Clin Chem 1993; 39: Hawkins R. Variability in potassium/hemoglobin ratios for hemolysis correction. Clin Chem 2002; 48: Sonntag O and Glick MR. Serum index and interferogram a new way of testing and depicting the interference of serum chromogens. J Lab Med 1989; 13: Moon-Massat PF, Tierney JP, Hock KG, et al. Hitachi Hemolytic Index correlates with HBOC-201 concentrations: impact on suppression of analyte results. Clin Biochem 2008; 41: Hawkins R. Hemolysis toward clearer terminology. Arch Pathol Lab Med 2010; 134: Appendix Notation Ao AB B C CV A CV B CV C measurand concentration in absence of interferent measurand concentration in presence of interferent with concentration B interferent concentration average slope of interferent effect CV of assay for measurement of A CV of measurement of interferent CV of slope of interferent effect

Coping with Analytical Interferences

Coping with Analytical Interferences Coping with Analytical Interferences (Handling Icteric, Hemolytic and Lipaemic Samples) Graham Jones Department of Chemical Pathology St Vincent s Hospital, Sydney Surabaya Indonesia 2016 Acknowledgements

More information

The preanalytic phase of laboratory medicine is the

The preanalytic phase of laboratory medicine is the Correction Factors for Estimating Potassium Concentrations in Samples With In Vitro Hemolysis A Detriment to Patient Safety Mai M. H. Mansour, MS; Hassan M. E. Azzazy, PhD, DABCC; Steven C. Kazmierczak,

More information

Guide to Fulfillment of Measurement Uncertainty

Guide to Fulfillment of Measurement Uncertainty DIAGNOSTIC ACCREDITATION PROGRAM College of Physicians and Surgeons of British Columbia 300 669 Howe Street Telephone: 604-733-7758 ext. 2635 Vancouver BC V6C 0B4 Toll Free: 1-800-461-3008 (in BC) www.cpsbc.ca

More information

Laboratory management of haemolysed samples results of an RCPAQAP survey. Penny Petinos

Laboratory management of haemolysed samples results of an RCPAQAP survey. Penny Petinos Laboratory management of haemolysed samples results of an RCPAQAP survey Penny Petinos Agenda Background to KIMMS survey on haemolysis Results of the 2010 and 2014 surveys Outcomes/decisions 2 Background

More information

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats Authors: Kate Cote, Ph.D., Graham Bilbrough, MA, VetMB, CertVA, MRCVS,

More information

What information on measurement uncertainty should be communicated to clinicians, and how? Mario Plebani

What information on measurement uncertainty should be communicated to clinicians, and how? Mario Plebani What information on measurement uncertainty should be communicated to clinicians, and how? Mario Plebani OUTLINE OF TALK Uncertainty in medicine and shared decision making Measurement uncertainty in laboratory

More information

Evaluation of VACUETTE CAT Serum Fast Separator Blood Collection Tube for Routine Chemistry Analytes in Comparison to VACUTAINER RST Tube

Evaluation of VACUETTE CAT Serum Fast Separator Blood Collection Tube for Routine Chemistry Analytes in Comparison to VACUTAINER RST Tube Evaluation of VACUETTE CAT Serum Fast Separator Blood Collection Tube for Routine Chemistry Analytes in Comparison to VACUTAINER RST Tube Background: Greiner-Bio-One, Austria has been selling plastic evacuated

More information

AUSTRALIAN PATHOLOGY UNITS AND TERMINOLOGY (APUTS) Harmonised Reference Intervals Chemical Pathology. (v1.1)

AUSTRALIAN PATHOLOGY UNITS AND TERMINOLOGY (APUTS) Harmonised Reference Intervals Chemical Pathology. (v1.1) AUSTRALIAN PATHOLOGY UNITS AND TERMINOLOGY (APUTS) Harmonised Reference Intervals Chemical Pathology (v1.1) ISBN: Pending 1 State Health Publication Number (SHPN): Pending Online copyright RCPA 2014 This

More information

Effect of lipemia interferences in routine clinical Biochemical Tests (Intralаboratory Study)

Effect of lipemia interferences in routine clinical Biochemical Tests (Intralаboratory Study) Standard Scientific Research and Essays Vol 3(7): 187-191, July 2015 (ISSN: 2310-7502) http://www.standresjournals.org/journals/ssre Research Article Effect of lipemia interferences in routine clinical

More information

Quality Initiative In Pathology. Harmonisation of Laboratory Testing

Quality Initiative In Pathology. Harmonisation of Laboratory Testing Quality Initiative In Pathology Harmonisation of Laboratory Testing Harmonisation: what do we mean? Agreement of test results irrespective of the method used or the testing laboratory Requires: Common

More information

Confirmation - A Practical Approach

Confirmation - A Practical Approach Data Mining i for Reference Interval Confirmation - A Practical Approach Graham Jones Chemical Pathologist St Vincent s Hospital, Sydney AACB Webinar 5 th June 2013 Summary A little (more) background Current

More information

Selecting a Risk-Based SQC Procedure for a HbA1c Total QC Plan

Selecting a Risk-Based SQC Procedure for a HbA1c Total QC Plan 729488DSTXXX10.1177/1932296817729488Journal of Diabetes Science and TechnologyWestgard et al research-article2017 Original Article Selecting a Risk-Based SQC Procedure for a HbA1c Total QC Plan Journal

More information

General Chemistry Scheme Guide

General Chemistry Scheme Guide General Chemistry Scheme Guide Copyright WEQAS. All rights reserved. No part of this document may be reproduced or utilised in any form without permission from WEQAS Contents. Scheme details and repertoire.....

More information

TDM A biochemists approach (Vancomycin)

TDM A biochemists approach (Vancomycin) TDM A biochemists approach (Vancomycin) Graham Jones Chemical Pathologist SydPath, St Vincent s Hospital, Sydney ANZ TDM Workshop Sydney May 2015 Declaration I am a chemical Pathologist (not a clinical

More information

Bias in clinical chemistry. Elvar Theodorsson EFLM and Linköping University

Bias in clinical chemistry. Elvar Theodorsson EFLM and Linköping University Bias in clinical chemistry Elvar Theodorsson EFLM and Linköping University Bias a controversial subject Different perspectives Reseachers Users Regulatory bodies Standardisation organisations Metrologists

More information

Why Traceability Matters to Patients?

Why Traceability Matters to Patients? Why Traceability Matters to Patients? (and you are all patients) Graham Jones Department of Chemical Pathology St Vincent s Hospital, Sydney JCTLM Members and Stakeholders meeting, Paris 2017 Acknowledgements

More information

Quantitative results play an important role in diagnosing

Quantitative results play an important role in diagnosing Assessing the Impact of Analytical Error on Perceived Disease Severity Martin H. Kroll, MD; Carl C. Garber, PhD; Caixia Bi, PhD; Stephen C. Suffin, MD Context. The perception of the severity of disease

More information

Define performance goals in standardization: Is fitness for medical purpose the key?

Define performance goals in standardization: Is fitness for medical purpose the key? Define performance goals in standardization: Is fitness for medical purpose the key? 7th CIRME International Scientific Meeting Metrological traceability and assay standardization Stresa, May 24th, 2013

More information

The impact of change in albumin assay on reference intervals, prevalence of hypoalbuminaemia and albumin prescriptions

The impact of change in albumin assay on reference intervals, prevalence of hypoalbuminaemia and albumin prescriptions Research Article Annals of Clinical Biochemistry 2016, Vol. 53(1) 112 116! The Author(s) 2015 Reprints and permissions: sagepub.co.uk/journalspermissions.nav DOI: 10.1177/0004563215599560 acb.sagepub.com

More information

The Whats and Hows of Reference Intervals. Graham Jones Department of Chemical Pathology St Vincent s Hospital, Sydney

The Whats and Hows of Reference Intervals. Graham Jones Department of Chemical Pathology St Vincent s Hospital, Sydney The Whats and Hows of Reference Intervals Graham Jones Department of Chemical Pathology St Vincent s Hospital, Sydney Surabaya Indonesia 2016 Acknowledgements Reference Intervals - Summary What are they

More information

Patient: 55 y female (ambulatory)

Patient: 55 y female (ambulatory) Disclosures Speaking Honoraria Radiometer (Canada) Nova Biomedical, Draeger Roche Diagnostics (Canada) Research Support (Reagents, Instrumentation, Travel) Nova Biomedical Abbott Laboratories (Canada)

More information

Traceability in External Quality Assessment: How Weqas ensures traceability in EQA and stresses its importance to users. David Ducroq.

Traceability in External Quality Assessment: How Weqas ensures traceability in EQA and stresses its importance to users. David Ducroq. Traceability in External Quality Assessment: How Weqas ensures traceability in EQA and stresses its importance to users David Ducroq Weqas Unit 6, Parc Tŷ Glas Llanishen Cardiff UK www.weqas.com Programme

More information

Reference Intervals. Graham Jones / Gus Koerbin

Reference Intervals. Graham Jones / Gus Koerbin Reference Intervals Graham Jones / Gus Koerbin Adult CRI - Harmonisation Harmonisation 1 2012: (13 tests + 1 calculation) Harmonisation 2 2013: Confirm 2012 recommendations. Discussed: albumin, globulin,

More information

ACCESS hstni SCIENTIFIC LITERATURE

ACCESS hstni SCIENTIFIC LITERATURE ACCESS hstni SCIENTIFIC LITERATURE 2017 2018 Table of contents Performance Evaluation of Access hstni A critical evaluation of the Beckman Coulter Access hstni: Analytical performance, reference interval

More information

BIOLOGICAL VARIABILITY: THE CIRME CONTRIBUTION. F. Braga Centre for Metrological Traceability in Laboratory Medicine (CIRME)

BIOLOGICAL VARIABILITY: THE CIRME CONTRIBUTION. F. Braga Centre for Metrological Traceability in Laboratory Medicine (CIRME) BIOLOGICAL VARIABILITY: THE CIRME CONTRIBUTION F. Braga Centre for Metrological Traceability in Laboratory Medicine (CIRME) TOTAL VARIATION (CV (CV T ) T ) PREANALYTICAL VARIATION ANALYTICAL VARIATION

More information

Week 17 and 21 Comparing two assays and Measurement of Uncertainty Explain tools used to compare the performance of two assays, including

Week 17 and 21 Comparing two assays and Measurement of Uncertainty Explain tools used to compare the performance of two assays, including Week 17 and 21 Comparing two assays and Measurement of Uncertainty 2.4.1.4. Explain tools used to compare the performance of two assays, including 2.4.1.4.1. Linear regression 2.4.1.4.2. Bland-Altman plots

More information

Med Chem 535P ~ Diagnostic Medicinal Chemistry. General Comments

Med Chem 535P ~ Diagnostic Medicinal Chemistry. General Comments Med Chem 535P ~ Diagnostic Medicinal Chemistry General Comments Most blood chemistry and serology assays are performed automatically. Larger clinical laboratories often use sophisticated analyzers that

More information

1 Introduction Imprecision Within-run imprecision, results Day-to-day imprecision, results... 2

1 Introduction Imprecision Within-run imprecision, results Day-to-day imprecision, results... 2 Design Verification liquicolor 1 Introduction... 2 2 Imprecision... 2 2.1 Within-run imprecision, results... 2 2.2 Day-to-day imprecision, results... 2 3 Linearity and Detection Limit... 2 3.1 Linearity...

More information

Delta Check Calculation Guide

Delta Check Calculation Guide Delta Check Calculation Guide National Technology 2017, All Rights Reserved By Senior Scientific Researcher, Asmaa Taher Table of Contents Definition... 2 Purpose... 2 Delta Check Research Studies... 2

More information

Electrolyte-balanced heparin in blood gas syringes can introduce a significant bias in the measurement of positively charged electrolytes

Electrolyte-balanced heparin in blood gas syringes can introduce a significant bias in the measurement of positively charged electrolytes Electrolyte-balanced heparin in blood gas syringes can introduce a significant bias in the measurement of positively charged electrolytes Citation for published version (APA): Berkel, van, M., & Scharnhorst,

More information

Stapleton, Mary; Daly, Niamh; O'Kelly, Ruth; Turner, Michael.

Stapleton, Mary; Daly, Niamh; O'Kelly, Ruth; Turner, Michael. Provided by the author(s) and University College Dublin Library in accordance with publisher policies. Please cite the published version when available. Title Time And temperature affect glycolysis in

More information

Method Comparison Report Semi-Annual 1/5/2018

Method Comparison Report Semi-Annual 1/5/2018 Method Comparison Report Semi-Annual 1/5/2018 Prepared for Carl Commissioner Regularatory Commission 123 Commission Drive Anytown, XX, 12345 Prepared by Dr. Mark Mainstay Clinical Laboratory Kennett Community

More information

The impact of preanalytical factors on glucose concentration measurement

The impact of preanalytical factors on glucose concentration measurement The impact of preanalytical factors on glucose concentration measurement Nora Nikolac, PhD University Department of Chemistry, Medical School University Hospital Sestre Milosrdnice, Zagreb, Croatia Glucose

More information

Comparison of Five Automated Serum and Whole Blood Folate Assays

Comparison of Five Automated Serum and Whole Blood Folate Assays Clinical Chemistry / FIVE AUTOMATED FOLATE ASSAYS Comparison of Five Automated Serum and Whole Blood Folate Assays William E. Owen, MT(ASCP), 1 and William L. Roberts, MD, PhD 2 Key Words: Hemolysate;

More information

ENROLLMENT CONFIRMATION

ENROLLMENT CONFIRMATION Step 1: Please review the Facility/Contact information. If any of the information is incorrect, please make the appropriate changes below: Facility/Contact Phone: (850)474-3660 Fax: (850)474-3659 6431

More information

Performance Characteristics of Eight Estradiol Immunoassays

Performance Characteristics of Eight Estradiol Immunoassays Clinical Chemistry / EVALUATION OF EIGHT ESTRADIOL IMMUNOASSAYS Performance Characteristics of Eight Estradiol Immunoassays David T. Yang, MD, 1 William E. Owen, MT(ASCP), 2 Carol S. Ramsay, 3 Hui Xie,

More information

Stability of common biochemical analytes in serum gel tubes subjected to various storage temperatures and times pre-centrifugation

Stability of common biochemical analytes in serum gel tubes subjected to various storage temperatures and times pre-centrifugation Original Article Stability of common biochemical analytes in serum gel tubes subjected to various storage temperatures and times pre-centrifugation Melissa Tanner 1, Neil Kent 1, Brian Smith 2, Stephen

More information

Seeing is not Believing (13-Nov-2004)

Seeing is not Believing (13-Nov-2004) In: 55th Annual Meeting of the American College of Veterinary Pathologists (ACVP) & 39th Annual Meeting of the American Society of Clinical Pathology (ASVCP), ACVP and ASVCP (Eds.) Publisher: American

More information

METHOD VALIDATION: WHY, HOW AND WHEN?

METHOD VALIDATION: WHY, HOW AND WHEN? METHOD VALIDATION: WHY, HOW AND WHEN? Linear-Data Plotter a,b s y/x,r Regression Statistics mean SD,CV Distribution Statistics Comparison Plot Histogram Plot SD Calculator Bias SD Diff t Paired t-test

More information

CCQM-K6: Key Comparison on the Determination of Cholesterol In Serum* Final Report December 2001

CCQM-K6: Key Comparison on the Determination of Cholesterol In Serum* Final Report December 2001 CCQM-K6: Key Comparison on the Determination of Cholesterol In Serum* INTRODUCTION Final Report December 2001 Michael J. Welch, Reenie M. Parris, Lorna T. Sniegoski, and Willie E. May Analytical Chemistry

More information

Assessment of "Average of Normals" Quality Control Procedures and Guidelines for Implementation

Assessment of Average of Normals Quality Control Procedures and Guidelines for Implementation Assessment of "Average of Normals" Quality Control Procedures and Guidelines for Implementation GEORGE S. CEMBROWSKI, M.D., PH.D., ELLIOT P. CHANDLER, M.D., AND JAMES O. WESTGARD, PH.D. The capabilities

More information

Breakout Session C: Harmonisation of the Alert Table.

Breakout Session C: Harmonisation of the Alert Table. Breakout Session C: Harmonisation of the Alert Table. RCPA-AACB High Risk Results Working Party Andrew Georgiou Craig Campbell Grahame Caldwell Hans Schneider Penelope Coates Que Lam Rita Horvath Robert

More information

For In Vitro Diagnostic Use

For In Vitro Diagnostic Use SYNCHRON System(s) Chemistry Information Sheet Lactate REF A95550 For In Vitro Diagnostic Use ANNUAL REVIEW Refer to Review and Revision Coversheet at the front of each method. PRINCIPLE INTENDED USE reagent,

More information

NQLM Nordic Committee for External Quality Assurance Programmes in Laboratory Medicine

NQLM Nordic Committee for External Quality Assurance Programmes in Laboratory Medicine 1(4) NQLM 2001-11-23 Nordic Committee for External Quality Assurance Programmes in Laboratory Medicine NORDIC INTERFERENCE STUDY, March 2001 Effects of bilirubin on some common serum analysis Enclosed

More information

In Vitro Diagnostic Glucose Test System

In Vitro Diagnostic Glucose Test System CDRH Final Guidance 2-Page Cover Sheet Guidance for Industry In Vitro Diagnostic Glucose Test System Document issued on: July 6, 1998 U.S. Department Of Health and Human Services Food and Drug Administration

More information

Creatinine (serum, plasma)

Creatinine (serum, plasma) Creatinine (serum, plasma) 1 Name and description of analyte 1.1 Name of analyte Creatinine 1.2 Alternative names None 1.3 Description of analyte Creatinine is a heterocyclic nitrogenous compound (IUPAC

More information

Evaluation of haemoglobin A1c measurement by an enzymatic method using an automated analyser that has an on-board haemolysis system

Evaluation of haemoglobin A1c measurement by an enzymatic method using an automated analyser that has an on-board haemolysis system Original Article Annals of Clinical Biochemistry 50(5) 443 449! The Author(s) 2013 Reprints and permissions: sagepub.co.uk/journalspermissions.nav DOI: 10.1177/0004563213476859 acb.sagepub.com Evaluation

More information

It is a requirement of ISO that laboratories shall determine the uncertainty of measurement of results, where relevant and possible. (5.6.

It is a requirement of ISO that laboratories shall determine the uncertainty of measurement of results, where relevant and possible. (5.6. Guideline Subject: Uncertainty of Measurement Approval Date: November 2004, November 2009 Review Date: November 2013 Review By: PPAC Number: 2/2004 Introduction This Guideline has been developed to provide

More information

The Virtues and Pitfalls of Implementing a New Test

The Virtues and Pitfalls of Implementing a New Test The Virtues and Pitfalls of Implementing a New Test James H. Nichols, Ph.D., DABCC, FACB Professor of Clinical Pathology, Microbiology and Immunology Associate Medical Director for Clinical Operations

More information

Original Article Interfering effect of bilirubin on the determination of alkaline phosphatase

Original Article Interfering effect of bilirubin on the determination of alkaline phosphatase Int J Clin Exp Med 2014;7(11):4244-4248 www.ijcem.com /ISSN:1940-5901/IJCEM0002083 Original Article Interfering effect of bilirubin on the determination of alkaline phosphatase Zhong Wang, Hongyan Guo,

More information

State of the Art of HbA1c Measurement

State of the Art of HbA1c Measurement State of the Art of HbA1c Measurement XXI Congreso Latinoamericano de Patologia Clinica Y XLII Congreso Mexicano de Patologia Clinica, Cancun October 2012 Randie R. Little, Ph.D. NGSP Network Laboratory

More information

A statistical evaluation of the effects of gender differences in assessment of acute inhalation toxicity

A statistical evaluation of the effects of gender differences in assessment of acute inhalation toxicity A statistical evaluation of the effects of gender differences in assessment of acute inhalation toxicity Human and Experimental Toxicology 30(3) 217 238 ª The Author(s) 2010 Reprints and permission: sagepub.co.uk/journalspermissions.nav

More information

Evaluation Report. Eurolyser Lactate test kit (VT0220, VT0221) on solo analysers

Evaluation Report. Eurolyser Lactate test kit (VT0220, VT0221) on solo analysers Evaluation Report Eurolyser Lactate test kit (VT0220, VT0221) on solo analysers Location: Operators: Eurolyser Diagnostica GmbH Simone Wieser, Bettina Schönauer, Dr. Jürgen Berlanda Report created on 17

More information

It s a Gas! Issues in the Blood Gas Laboratory. D. Robert Dufour, MD, FCAP Consultant Pathologist VA Medical Center, Washington DC

It s a Gas! Issues in the Blood Gas Laboratory. D. Robert Dufour, MD, FCAP Consultant Pathologist VA Medical Center, Washington DC It s a Gas! Issues in the Blood Gas Laboratory D. Robert Dufour, MD, FCAP Consultant Pathologist VA Medical Center, Washington DC Learning Objectives After participating in this session, you will be able

More information

Analysis of Hemoglobin A1c from Dried Blood Spot Samples with the Tina-quant II Immunoturbidimetric Method

Analysis of Hemoglobin A1c from Dried Blood Spot Samples with the Tina-quant II Immunoturbidimetric Method Journal of Diabetes Science and Technology Volume 4, Issue 2, March 2010 Diabetes Technology Society SYMPOSIUM Analysis of Hemoglobin A1c from Dried Blood Spot Samples with the Tina-quant II Immunoturbidimetric

More information

liquicolor Design Verification

liquicolor Design Verification Design Verification liquicolor 1 Introduction... 2 2 Imprecision... 2 2.1 Imprecision for... 2 2.2 Imprecision for... 2 3 Linearity and Detection Limit... 2 3.1 Linearity... 2 3.2 Detection Limit... 4

More information

The analytical phase

The analytical phase The analytical phase Result interpretation Test request Result Sampling Black box: the lab ANALYTICAL PHASE The CASE Uncle Pete, 67 years old Marked abdominal pain 8 pm, ED Acute abdomen? Assessment (+

More information

CAP Laboratory Improvement Programs. Utility of Repeat Testing of Critical Values. A Q-Probes Analysis of 86 Clinical Laboratories

CAP Laboratory Improvement Programs. Utility of Repeat Testing of Critical Values. A Q-Probes Analysis of 86 Clinical Laboratories CAP Laboratory Improvement Programs Utility of Repeat Testing of Critical Values A Q-Probes Analysis of 86 Clinical Laboratories Christopher M. Lehman, MD; Peter J. Howanitz, MD; Rhona Souers, MS; Donald

More information

Analytical Performance in the The Test Evaluation Cycle

Analytical Performance in the The Test Evaluation Cycle Analytical Performance Specifications based on Outcome Studies: Is it possible? Test Evaluation Working Group of EFLM www.efcclm.eu SEALS Department of Clinical Chemistry, Prince of Wales Hospital, Sydney

More information

INTERNAL QUALITY CONTROL ( Q I C) QC)

INTERNAL QUALITY CONTROL ( Q I C) QC) EXTERNAL QUALITY ASSESSMENT PROGRAM (EQAP) BIOCHEMISTRY DEPARTMENT R. Mohammadi Biochemist (Ph.D.) Faculty member of Medical Faculty TWO COMPLEMENTARY COMPONENETS OF TQM ARE Internal Quality Control (IQC)

More information

Indicators of Quality in the pre-analytical Phase for Determining Cardiac Markers in Acute Myocardial infarction CK, CKMB and TnT

Indicators of Quality in the pre-analytical Phase for Determining Cardiac Markers in Acute Myocardial infarction CK, CKMB and TnT Journal of US-China Medical Science 13 (216) 138-145 doi: 1.17265/1548-6648/216.3.3 D DAVID PUBLISHING Indicators of Quality in the pre-analytical Phase for Determining Cardiac Markers in Acute Myocardial

More information

Patricia C. Fallest-Strobl, 1 Elin Olafsdottir, 2 Donald A. Wiebe, 1 and James O. Westgard 1 * Lipoproteins

Patricia C. Fallest-Strobl, 1 Elin Olafsdottir, 2 Donald A. Wiebe, 1 and James O. Westgard 1 * Lipoproteins Clinical Chemistry 43:11 2164 2168 (1997) Lipids and Lipoproteins Comparison of NCEP performance specifications for triglycerides, HDL-, and LDL-cholesterol with operating specifications based on NCEP

More information

The ABL800 FLEX Setting the standard

The ABL800 FLEX Setting the standard The ABL800 FLEX Setting the standard SIMPLY DEDICATED Setting the standard for fast and reliable measurements A look behind Radiometer s superior sensor technology Since the introduction, Radiometer s

More information

Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry.

Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry. 2013/07/30 A93A01315AUS A11A01641 60 ml 15 ml Pentra C400 Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry. QUAL-QA-TEMP-0846

More information

Applicable To Employees of Gundersen Boscobel Area Hospital Laboratory and Gundersen Palmer Lutheran Hospital and Clinics Laboratory.

Applicable To Employees of Gundersen Boscobel Area Hospital Laboratory and Gundersen Palmer Lutheran Hospital and Clinics Laboratory. Subject Creatinine COBAS C311 Index Number Lab-8814 Section Laboratory Subsection Regional/Affiliates Location Category Departmental Contact Tilleraas, Betty Last Revised 1/17/2018 References Required

More information

LIPASE liquicolor. Design Verification. Multipurpose Reagent

LIPASE liquicolor. Design Verification. Multipurpose Reagent Design Verification LIPASE liquicolor Multipurpose Reagent CONTENTS 1 Introduction... 2 2 Imprecision... 2 3 Linearity and Detection Limit... 2 3.1 Linearity... 2 3.2 Detection Limit... 3 4 Recovery of

More information

POTASSIUM/K. 3.6 mmol/l

POTASSIUM/K. 3.6 mmol/l POTASSIUM/K Potassium is measured by ion-selective electrode potentiometry. In the calculation of results for potassium, concentration is related to potential through the Nernst equation. The i-stat System

More information

Clinical Laboratory Quality Practices When Hemolysis Occurs

Clinical Laboratory Quality Practices When Hemolysis Occurs Clinical Laboratory Quality Practices When Hemolysis Occurs Peter J. Howanitz, MD; Christopher M. Lehman, MD; Bruce A. Jones, MD; Frederick A. Meier, MD; Gary L. Horowitz, MD Context. Hemolyzed specimens

More information

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)

1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803) Comparison Study Summary Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC 29720 (803) 286-1480 February 10, 2016 1 PROTOCOL This evaluation was conducted on February 10, 2016 at Springs Memorial

More information

Table. Analytical characteristics of commercial and research cardiac troponin I and T assays declared by the manufacturer.

Table. Analytical characteristics of commercial and research cardiac troponin I and T assays declared by the manufacturer. Table. Analytical characteristics of commercial and research cardiac troponin I and T assays declared by the manufacturer. Commercially available assays - Company/ platform(s)/ assay LoB a LoD b 99 th

More information

Use of PSI on CS instruments

Use of PSI on CS instruments Johan Claes, Version 01, 1 st September 2016 User day hemostasis 2016 Use of PSI on CS instruments Restricted Siemens Healthcare GmbH, 2016 Hemostasis Lab Challenges Preanalytical problems in Haemostasis

More information

Setting of quality standards

Setting of quality standards Setting of quality standards Graham Jones Department of Chemical Pathology St Vincent s Hospital, Sydney AACB ASM Adelaide October 2014 Setting of Quality Standards - 2013 The 2013 QC workshop revealed

More information

FRCPath Part 1 written examination in Clinical Biochemistry

FRCPath Part 1 written examination in Clinical Biochemistry FRCPath Part 1 written examination in Clinical Biochemistry Paper 2 of the Part 1 written examination in Clinical Biochemistry consists of 20 compulsory Short Answer Questions (SAQs), to be answered in

More information

Evaluation of Calculated Low-Density Lipoprotein Against a Direct Assay

Evaluation of Calculated Low-Density Lipoprotein Against a Direct Assay Evaluation of Calculated Low-Density Lipoprotein Against a Direct Assay Navapun Charuruks MD*, Anna Milintagas BSc** * Department of Laboratory Medicine, Faculty of Medicine, Chulalongkorn University **

More information

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY INFORMATION BULLETIN USING THE ACCESS AMH ASSAY IN YOUR LABORATORY ///////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////

More information

Postanalytical phase

Postanalytical phase Postanalytical phase Test request POSTANALYTICAL Result interpretation PHASE Result Sampling Black box: the lab And the RESULT is created The technician approves the result; it is transferred to the lab

More information

Comparison of VACUETTE Heparin Gel Tubes for Common Chemistry Analytes

Comparison of VACUETTE Heparin Gel Tubes for Common Chemistry Analytes Comparison of VACUETTE Heparin Gel Tubes for Common Chemistry Analytes Background: Greiner-Bio-One, Austria has been selling plastic evacuated tubes (VACUETTE ) for venous blood collection since 9. The

More information

GLUH PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE

GLUH PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE SYNCHRON System(s) Chemistry Information Sheet Copyright 2014 Beckman Coulter, Inc. GLUH Glucose REF B24985 For In Vitro Diagnostic Use Rx Only ANNUAL REVIEW Reviewed by Date Reviewed by Date PRINCIPLE

More information

SLAUGHTER PIG MARKETING MANAGEMENT: UTILIZATION OF HIGHLY BIASED HERD SPECIFIC DATA. Henrik Kure

SLAUGHTER PIG MARKETING MANAGEMENT: UTILIZATION OF HIGHLY BIASED HERD SPECIFIC DATA. Henrik Kure SLAUGHTER PIG MARKETING MANAGEMENT: UTILIZATION OF HIGHLY BIASED HERD SPECIFIC DATA Henrik Kure Dina, The Royal Veterinary and Agricuural University Bülowsvej 48 DK 1870 Frederiksberg C. kure@dina.kvl.dk

More information

Determination of hemoglobin is one of the most commonly

Determination of hemoglobin is one of the most commonly ORIGINAL ARTICLE Multiple-Site Analytic Evaluation of a New Portable Analyzer, HemoCue Hb 201+, for Point-of-Care Testing Sten-Erik Bäck, PhD,* Carl G. M. Magnusson, PhD, Lena K. Norlund, MD, PhD, Henning

More information

Analytical performance specifications two years after Milan conference. Prof Mauro Panteghini CIRME Scientific Coordinator

Analytical performance specifications two years after Milan conference. Prof Mauro Panteghini CIRME Scientific Coordinator Analytical performance specifications two years after Milan conference Prof Mauro Panteghini CIRME Scientific Coordinator Definition Analytical performance specifications: Criteria that specify (in numerical

More information

Performance Guide. icheck Iodine The test kit to measure iodine in salt

Performance Guide. icheck Iodine The test kit to measure iodine in salt Performance Guide icheck Iodine The test kit to measure iodine in salt Performance explained This Performance Guide is intended to explain the validity of our test kit icheck Iodine in a nutshell to facilitate

More information

Sample Quality Assurance

Sample Quality Assurance Sample Quality Assurance Phlebotomy Association of Ireland AGM 2012 Julie Riordan BSc, HDip Arts, MSc, F.A.M.L.S. Talk Outline Quality Assurance (QA) Preanalytical Variables (PAVs) & Errors (PAEs) Effects

More information

Polymer Technology Systems, Inc. CardioChek PA Comparison Study

Polymer Technology Systems, Inc. CardioChek PA Comparison Study Polymer Technology Systems, Inc. CardioChek PA Comparison Study Evaluation Protocol: Accuracy Precision Clinical Correlation PTS Panels Lipid Panel Test Strips For Use in Comparisons to a Reference Laboratory

More information

QUANTIFYING THE EFFECT OF SETTING QUALITY CONTROL STANDARD DEVIATIONS GREATER THAN ACTUAL STANDARD DEVIATIONS ON WESTGARD RULES

QUANTIFYING THE EFFECT OF SETTING QUALITY CONTROL STANDARD DEVIATIONS GREATER THAN ACTUAL STANDARD DEVIATIONS ON WESTGARD RULES AACC24 QUANTIFYING THE EFFECT OF SETTING QUALITY CONTROL STANDARD DEVIATIONS GREATER THAN ACTUAL STANDARD DEVIATIONS ON WESTGARD RULES Graham Jones Department of Chemical Pathology, St Vincent s Hospital,

More information

Harmonization of Clinical Laboratory Results is Essential for Quality Patient Care

Harmonization of Clinical Laboratory Results is Essential for Quality Patient Care Harmonization of Clinical Laboratory Results is Essential for Quality Patient Care Greg Miller, PhD Virginia Commonwealth University Richmond, VA Presented at the National Association of Chronic Disease

More information

G. Snodgrass, K. Ackles, A. Blanco and A. Versaggi Ortho Clinical Diagnostics, Rochester, NY 14626

G. Snodgrass, K. Ackles, A. Blanco and A. Versaggi Ortho Clinical Diagnostics, Rochester, NY 14626 Performance of the EKF Diagnostics, Stanbio β-hydroxybutyrate LiquiColor Assay on the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System. G. Snodgrass, K. Ackles, A. Blanco and A. Versaggi

More information

Analytic Bias of Thyroid Function Tests. Analysis of a College of American Pathologists Fresh Frozen Serum Pool by 3900 Clinical Laboratories

Analytic Bias of Thyroid Function Tests. Analysis of a College of American Pathologists Fresh Frozen Serum Pool by 3900 Clinical Laboratories Analytic Bias of Thyroid Function Tests Analysis of a College of American Pathologists Fresh Frozen Serum Pool by 3900 Clinical Laboratories Bernard W. Steele, MD; Edward Wang, PhD; George G. Klee, MD,

More information

BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS

BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS International Standard ISO 15197:2013 (Second Edition 15/05/2013) In vitro diagnostic test systems Requirements for blood-glucose monitoring systems for self-testing

More information

Biological Variation The logical source for analytical goals

Biological Variation The logical source for analytical goals Biological Variation The logical source for analytical goals Presented by: John Yundt-Pacheco Scientific Fellow Quality Systems Division Bio-Rad Laboratories John_yundt-pacheco@bio-rad.com Acknowledgements

More information

Aina Blood Monitoring System

Aina Blood Monitoring System Aina Blood Monitoring System Analytical Performance Summary The Aina Blood Monitoring System is a high quality and versatile multi-parameter diagnostic platform that is CE-marked and approved for sale

More information

C56-A Hemolysis, Icterus, and Lipemia/Turbidity Indices as Indicators of Interference in Clinical Laboratory Analysis; Approved Guideline

C56-A Hemolysis, Icterus, and Lipemia/Turbidity Indices as Indicators of Interference in Clinical Laboratory Analysis; Approved Guideline C56-A Hemolysis, Icterus, and Lipemia/Turbidity Indices as Indicators of Interference in Clinical Laboratory Analysis; Approved Guideline Presented by: Merle B. Smith, MBA, MS, MT(ASCP) Siemens Healthcare

More information

Keywords: albuminuria; albumin/creatinine ratio (ACR); measurements. Introduction

Keywords: albuminuria; albumin/creatinine ratio (ACR); measurements. Introduction Clin Chem Lab Med 2015; 53(11): 1737 1743 Beryl E. Jacobson, David W. Seccombe*, Alex Katayev and Adeera Levin A study examining the bias of albumin and albumin/creatinine ratio measurements in urine DOI

More information

VetScan UA Urine Analyzer Clinical Performance Larry Lem, PhD & Andrew Rosenfeld, DVM, ABVP

VetScan UA Urine Analyzer Clinical Performance Larry Lem, PhD & Andrew Rosenfeld, DVM, ABVP VetScan UA Urine Analyzer Clinical Performance Larry Lem, PhD & Andrew Rosenfeld, DVM, ABVP Abstract Background: The VetScan UA urine strip reader is an analyzer designed to provide semi-quantitative and

More information

Plasma Hemoglobin Determination: Two Manual Methods Compared. An Honors Thesis (HONRS 499) Anne M. Emerick. Sharlene Strahl. Beech Grove, Indiana

Plasma Hemoglobin Determination: Two Manual Methods Compared. An Honors Thesis (HONRS 499) Anne M. Emerick. Sharlene Strahl. Beech Grove, Indiana Plasma Hemoglobin Determination: Two Manual Methods Compared An Honors Thesis (HONRS 499) by Anne M. Emerick Sharlene Strahl St. Francis Hospital and Health Centers Beech Grove, Indiana Dr. Nancy Behforouz

More information

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date SYNCHRON System(s) Chemistry Information Sheet 2015 Beckman Coulter, Inc. All rights reserved. HDLD HDL Cholesterol 650207 For In Vitro Diagnostic Use ANNUAL REVIEW Rx Only Reviewed by Date Reviewed by

More information

Traditional approaches to Reference Intervals

Traditional approaches to Reference Intervals Establishing reference intervals: Unique challenges for the pediatric population Amy L. Pyle, PhD, DABCC Traditional approaches to Reference Intervals Decision Limits Establish Verify Transfer 1 Decision

More information

For in vitro diagnostic use only. For Rx use only. Reviewed by Date Reviewed by Date

For in vitro diagnostic use only. For Rx use only. Reviewed by Date Reviewed by Date AU US Instructions For Use 2018 Beckman Coulter, Inc. All rights reserved. TRIG TRIGLYCERIDE OSR60118 4 x 20 ml R1, 4 x 5 ml R2 OSR61118 4 x 50 ml R1, 4 x 12.5 ml R2 OSR66118 4 x 173 ml R1, 4 x 48 ml R2

More information

ROTUNDA HOSPITAL DEPARTMENT OF LABORATORY MEDICINE

ROTUNDA HOSPITAL DEPARTMENT OF LABORATORY MEDICINE This active test table informs the user of Biochemistry tests available in house. s referred to other sites are recorded in the Referred Table. Issue date: 4 TH April 2016 Contact Phone Number ext.1345/2522

More information

1. Calibra - H - Store at 2-8 ºC.

1. Calibra - H - Store at 2-8 ºC. CALIBRA H Insert Ref.:80 Lot Expiration Calibrator Attention. It is suggested to verify carefully if the lot number printed in this insert corresponds to the lot on the bottle label. Intended use. is a

More information