British Journal ofanaesthesia 85 (3): (2000) EQUIPMENT. The Reflector: a new method for saving anaesthetic vapours
|
|
- Jessie Moody
- 5 years ago
- Views:
Transcription
1 British Journal ofanaesthesia 85 (3): (2000) EQUIPMENT BjA The Reflector: a new method for saving anaesthetic vapours L. Perhag'", P. Reinstrup', R. Thomassoni and O. Werner! 1Department ofanaesthesia, University Hospital of Lund, S Lund, Sweden. 2Department of Inorganic Chemistry 2, University of Lund, PO Box 124, S Lund, Sweden *Corresponding author. Present address: Department ofanaesthesia, Ystad Hospital, S Ystad, Sweden Anaesthesia systems that minimize the use of volatile anaesthetics to reduce cost and pollution are of interest. Closed circuit anaesthesia is the ideal solution, but requires continuous adjustment of fresh gas flow and composition and thus is demanding in routine practice. We describe an alternative system, the Reflector system, which is open in regard to oxygen, nitrogen and N 20, and semiclosed in regard to volatile anaesthetics. The Reflector system is a circle system with a carbon dioxide absorber and an automatic vapour delivery device placed in the inspiratory limb of the circle. A zeolite filter, the Reflector, is placed between the ventilator and the circle. The Reflector functions as a molecular sieve, preventing the volatile anaesthetic from leaving the circle. Isoflurane consumption using the Reflector system in bench tests and an animal study was compared with that of an open system. In bench tests consumption was reduced by 79% and 82%, at a respiratory frequency of 10 and 20 min-i, respectively. The corresponding mean figures from the animal experiment were 65% and 77%. BrJ Anaesth 2000; 85: Keywords: anaesthetic techniques, closed circuit; anaesthetic techniques, inhalation; anaesthetics volatile, isoflurane Accepted for publication: April II, 2000 An anaesthesia circuit with carbon dioxide absorption was described in 1915 by Jackson.' In 1926, Waters 2 developed the to-and-fro absorption method, and the introduction of cyclopropane in 1933 led to widespread use of closed circuit anaesthesia. With the clinical introduction of halothane in 1956 by Raventos.' low-flow and closed systems were used, but halothane vaporizers were imprecise at low gas flows, and lack of knowledge about the pharmacokinetics of halothane resulted in continued use of open/semiclosed systems. The introduction of more ideal," but also more expensive, inhalational anaesthetics and an increasing awareness of atmospheric pollution.' have increased the interest in lowflow and closed circuit methods. These methods preserve humidity and body heat", save money" and reduce atmospheric pollution. The greater use of gas monitors to measure inspired oxygen concentration'' and other gases of the circuit has increased the safety of closed circuit anaesthesia. However, because they require continuous adjustments of gas delivery to the circuit," these systems have not gained much popularity. One solution could be a system that was open in regard to oxygen, nitrogen and N 20 delivery, but closed in regard to volatile anaesthetics. Such a system has been created by introducing a molecular sieve (Reflectorj.l" a semipermeable filter that keeps the volatile anaesthetics on the patient side of the anaesthetic circuit but lets most other gases pass through. We evaluated such a Reflector system in bench tests and in an animal study. Materials and methods Set-up and experimental procedure An anaesthesia circuit with one-way valves and a carbon dioxide absorber (Monosorb; Siemens Elema AB, Solna, Sweden) was used. The tubing was made of polyethylene (Hytrel). The Reflector was placed between a ventilator (Servo 900 C; Siemens Elema AB) and the circuit. A heatmoisture exchanger (Siemens Humidifier 152; Siemens Elema AB) was placed at the tracheal tube connector. A computerized automatic vapour delivery device (AVDD) 11 was placed in the inspiratory limb of the circuit, The Board of Management and Trustees of the British Journal of Anaesthesia 2000
2 The reflector: a method for saving anaesthetic vapours structure, denoted by a three-letter code. Ultras table zeolite Y is of the Faujasite (FAU) structural type, characterized by channels in three dimensions, with an aperture of 7.4 A, connecting to microvoid cages (a(ii) cages) within the structure. Ultrastable zeolites have high thermal stability (up to 1200 C) and are chemically inert; for example, they are insoluble in acids (except HF) and stable to alkali below ph 10. Fig 1 Set-up of the Reflector system. AVDD, automatic vapour device. HME, heat-moisture exchanger. The 'subject' is the connection to model lung or animal. Numerals indicate gas sampling points. distal to the absorber. The AVDD delivered isoflurane at a partial pressure close to 40 kpa, the carrier gas being sucked from the circuit. During each inspiration, the AVDD continuously adjusted the addition of isoflurane vapour into the circuit to keep a set inspiratory isoflurane concentration of 1.5%. To allow continuous assessment of the isoflurane consumption, the reservoir of the AVDD was placed on an electronic balance (G1200; Ohaus, NJ, USA). The ventilator delivered a gas mixture of 40% oxygen in nitrogen with a volume-controlled ventilatory pattern and constant inspiratory flow. The inspiratory time of the ventilatory cycle was 33% and the set end-expiratory pressure was set to 5 em H 20. Side-stream gas analysers (Siemens 120; Siemens Elema AB), recirculating the sampled gas to the circuit, were used. Inspired, end-tidal and reflected isoflurane concentrations and inspired oxygen concentration were measured. Measurements were also taken at the expiratory port of the ventilator to detect 'isoflurane breakthrough' (lower detection limit=o.l %), i.e. leakage of isoflurane on the ventilator side of the Reflector. The set-up of the Reflector system is shown in Figure 1. The reflector The Reflector consisted of a cylindrical plastic container, with a diameter of 70 mm, in which a 30 mm (RS30) or 50 mm (RS50) deep adsorption bed was arranged between two steel grids. A spreader was placed on each side to create laminar flow through the adsorption bed. The pressure gradient over the RS30 was 1.2 and 4.6 em H 20 at 0.5 and 1.0 litres s:', respectively, and that over the RS50 was 3.0 and 11 em H 20 at 0.5 and 1.0 litres s-\ respectively. The dead space of the plastic container, with the zeolite volume subtracted, was identical in both filters. Pellets (110 or 185 ml) of ultras table zeolite Y, a synthetic substance, were used as adsorption material. The pellets were about 1 mm in diamater and were prepared using about 15% by mass of a binder material (A1 203 ). Ultras table zeolites are crystalline microporous silicates built up from corner-sharing [Si0 4t--tetrahedral forming an indefinite framework structure, characterized by the presence of cavities and channels of well-defined shape and size (3-10 A). Each type of zeolite has a unique framework Bench tests Experiments were undertaken to compare isoflurane consumption by RS30 and RS50 with that by a non-rebreathing system. The different components of isoflurane consumption using the two Reflector systems were further studied. The reduction in isoflurane consumption by the added deadspace caused by the geometry of the Reflectors was evaluated using a dummy filter. This filter had a deadspace identical to that of RS30 and RS50 but was filled with glass spheres of 1 mm diameter. With connections all filters had a deadspace of 125 ml. To assess the amountofisoflurane lost due to 'isoflurane breakthrough', a Douglas bag, made of aluminium foil, was placed at the expiratory port of the ventilator. The amount of isoflurane trapped in the Reflector was also assessed by weighing the filter before and after each experiment In bench tests the anaesthesia circle was connected to a lung model consisting of a U-shaped vessel filled with water. The end-expiratory volume of the model lung was 2.5 litres. The ventilator delivered 10 litres min- 1 with a respiratory frequency of 10 and 20 min-i. The inspiratory concentration of isoflurane was set to 1.5%. In control experiments, the Y-piece of the ventilator tubing was connected directly to the model lung, creating an open system. The AVDD was placed in the inspiratory limb. Animal experiments After approval from the local Animal Studies Committee, two pigs (34 and 28 kg; one female and one male) were studied. The pigs were premedicated with 15 mg midazolam i.m. 30 min before anaesthesia. Anaesthesia was induced with ketamine i.v. 1.5 mg kg- 1 and maintained with a continuous infusion of 7 mg kg" h- 1. Continuous infusion of pancuronium mg kg" h- 1 preserved muscle relaxation. A tracheotomy was performed; control of the airway was achieved by a cuffed tube. A pressure transducer attached to a catheter placed in the carotid artery monitored arterial pressure. A central venous cannula was placed in the jugular vein and an i.v. infusion (glucose 25 g 1-1 with 70 mmol Na+, 45 mmol Cl" and 25 mmol acetate per litre) was given at a rate of 6 ml kg- 1 h- 1. In the animal experiments only the RS30 was tested. The pigs were ventilated at a respiratory rate of 10 and 20 min- 1 and tidal volume (VJ was adjusted to keep end-tidal carbon 483
3 Perhag Table 1 The gravimetrically measured consumption at 1.5% inspired concentration of isoflurane. RR=respiratory rate Table 2 Isoflurane consumption displayed in its different components using the Reflector system. RR=respiratory rate System Isoflurane consumption (g/90 min) System Isoflurane consumption (g/90 min) RR=10 RR=20 RS30 RS50 Open Dummy RS30 RS RR=10 RR=20 RR=10 RR=20 Dissolved in circuit and model lung Uptake in filter Loss through filter (breakthrough) dioxide within kpa. End-tidal carbon dioxide was measured with an in-line carbon dioxide analyser (Novametrix 7000, Wallingford, USA). A heat-moisture exchanger was placed at the tracheal tube. Inspiratory concentration of isoflurane was set to 1.5% throughout each experiment. Each phase of the experiments, using the Reflector or the open system in a varied order, lasted 90 min. The zeolite material in the Reflector, when used, was renewed before a new experiment was started. Between each phase there was a 90 min washout period without a filter in the system. During the washout periods the end-expiratory isoflurane concentration decreased rapidly and no isoflurane was detectable after 30 min. 2.0 cf!, 1.5 ~ <.i c Ql c ~ ::> ~ o Inspired conc. Amount supported Endtidal conc. by AVDD Reflected conc Time (min) Calculation of measurement error To verify the validity of isoflurane consumption in the open system, isoflurane expenditure was calculated (Tables 1 and 3) using X 10 3 g litre- 1 as isoflurane density at 25 C and as molecular weight. 12 The quantity of isoflurane used may differ from the calculated quantity because of measurement errors in the weight, the gas analyser and inaccuracy of the delivered volume from the ventilator. The measurement errors of the balance, gas analyser and ventilator are :::':::2%, :::':::5% and :::':::5% respectively, giving a combined measurement error of 7%. Fig 2 Tracing of mean inspired, end-tidal and reflected isoflurane concentrations in animal experiments using the Reflector system (RS30). when using the RS50 were 80% and 82%. The absolute consumption is listed in Table 1. Using the RS30, 'breakthrough', or tracings ~ O %. of l isoflurane on the ventilator side of the Reflector, was detected after 30 min at a respiratory frequency of 10 S-l (V t 1000 ml) and after 44 min at a respiratory frequency of 20 S-l (V t 500 ml). No such breakthrough was detected with the RS50. The amount of isoflurane retrieved from the Douglas bag after 90 min sampling of scavenger gas and the gravimetrically measured weight gain of the filters are listed in Table 2. Results Bench tests The calculated isoflurane consumption over 90 min using a ventilator flow of 10 litres min- 1 and an inspiratory isoflurane concentration of 1.5% is g. The gravimetrically measured isoflurane consumption values for the open system at a respiratory frequency of 10 S-l and 20 S-l (Vt 1000 and 500 ml) were and g, respectively. Compared with the open system, the isoflurane consumption was reduced by 12% and 25% at a respiratory frequency of 10 and 20 min", respectively, when using the Reflector system with the dummy filter. Compared with the open system, the RS30 reduced the consumption of isoflurane by 79% and 82%, respectively. The corresponding figures Animal tests In the reflector experiments (RS30), the inspired oxygen concentration was 37-40%. The preset inspired isoflurane concentration of 1.5 % was reached after 4 :::'::: 2 min (mean :::'::: SD) and kept at this level (:::':::0.1 %) by the AVDD. Within 10 min an end-tidal isoflurane concentration of at least 1.2% was reached. Reflection of isoflurane was detected within 1 min and rapidly reached a plateau of 1.1% within 15 min (Figure 2). Using the RS30, at a respiratory frequency of 10 or 20 min", isoflurane consumption was reduced by 61 78% of that of an open system (Table 3). In one pig, at a respiratory frequency of 10 min", 'isoflurane breakthrough' was detected after 75 min. 484
4 The reflector: a method for saving anaesthetic vapours Table 3 Calculated and gravimetrically measured isoflurane consumption in animal studies using the Reflector (RS30) and open systems after 30, 60 and 90 min. Vt=tidal volume (meant snj: RR=respiratory rate. * indicates breakthrough of isoflurane through the filter which happened after 75 min (pig 1 weighed 34 kg and pig 2 28 kg) Pig # System RR (min-i) V t (litre) Measured isoflurane consumption (g) Calculated isoflurane consumption (g) 30 min 60 min 90 min 30 min 60 min 90 min 1 Reflector ± * 2 Reflector ± open ± open ± Reflector ± Reflector ± open ± open ± Discussion High gas flow administered to an anaesthesia circuit with carbon dioxide absorption is used as a standard technique by many anaesthetists. The inspired concentration of gases is close to the concentrations in the fresh gas delivered into the circuit. The major disadvantages of this approach are its expense and the loss of heat and humidity. The Reflector system was designed as a simple high-flow system which is closed with regard to volatile anaesthetics, in this case isoflurane. This is achieved by inserting a zeolite filter at the interface between the anaesthetic circuit and the Y-piece from the ventilator (Figure 1). The ventilator provides the circle with fresh gas with each tidal volume. Because about 25% of carbon dioxide is reflected by the molecular sieve,13 an anaesthesia circuit with a carbon dioxide absorber was used. We compared the amount of isoflurane consumed using the open system with calculated values. The discrepancy between the calculated consumption of isoflurane and the amount used in the open system can be explained by measurement errors in the components, as consumption of an open system depends only on the delivered volume, if inspired volatile concentration is constant. Thus, the combined error of isoflurane consumption depends on the accuracy of the volume delivered by the ventilator, the inspired isoflurane concentration measured by the gas analyser and the accuracy of the balance. This error was calculated as 7%, and all the measured values were within this level. Assessment of the Reflector system in bench tests was done by comparing a zeolite filter with a dummy filter, using an open system as control. The reduction of isoflurane consumption when changing from the open system to the Reflector system, but using a dummy filter filled with glass spheres, was caused by the added deadspace of this filter when situated between the ventilator and the circle and was more pronounced at high ventilatory frequencies equal to low tidal volumes. Zeolite can reflect the expired isoflurane and thereby retain the volatile anaesthetic in the circle system. The microporous arrangement of the zeolite gives rise to a large internal surface area (1000 m 2 g-l) and microvoid space (0.5 cm' g-l). Because of the connectivity of the tetrahedral building units, all oxygen atoms in the three-dimensional framework are interconnected and shared between two silicon atoms: there are no terminal hydroxyl groups. This makes the (internal and external) surfaces of the zeolite hydrophobic. Together with the size of the micropores, this gives ultras table zeolite Y a high affinity and a high adsorption capacity for non-polar substances, e.g. volatile anaesthetics, and a low affinity for polar substances, e.g. water. The size and shape of the molecules entering the pore system, i.e. the aperture of the channels, restrict the adsorption. The amount adsorbed depends on several factors, including the partial pressure, molecular weight, polarity and polarizability of the adsorbate and on the temperature. We found that both RS30 and RS50 reduced isoflurane consumption by about 80% compared with the open system. Our results support the findings of Thomasson, Luttrop and Werner 10 using a similar filter but a different set-up. However, in a truly closed system the amount of isoflurane used should only be equivalent to the volume of the circle and testlung, saturated to 1.5%, in this set-up approximately 0.6 g of isoflurane. Isoflurane use during Reflector experiments was greater than that in a closed system because of 'isoflurane breakthrough', the passage of volatile anaesthetics through the Reflector and absorption of vapour in the molecular seive of the Reflector. Breakthrough was more pronounced at a large tidal volume while absorption was independent of respiratory frequency. Increasing the amount of zeolite in the filter, from the RS30 to the RS50, did not increase efficacy because each millilitre of zeolite added to the absorption bed traps about 0.1 g of isoflurane when saturated. Therefore, increasing the amount of zeolite to prevent loss through the filter, was in this case, offset by the amount of isoflurane trapped in the filter. Lengthy anaesthesia will be more cost effective, as will the reuse of a loaded filter. The difference between isoflurane use in the open pig experiments (Table 3) and that in the open bench tests 485
5 Perhag (Table I) was caused by differences in tidal volume, which were necessary to keep end-tidal carbon dioxide within normal limits. In the animal experiment, isoflurane consumption by the open system and the RS30 was not quite comparable because the tidal volumes used were higher in the RS30. One reason was that the volume of this system was 2 litres greater, which increased the compressible volume. Futhermore, end-tidal carbon dioxide was only kept within the limit of kpa. The calculated uptake of isoflurane by the pigs in our experiments was 8-9 g.14the reduction in concentration of isoflurane in the expired air, caused by uptake of isoflurane by the pig, may ease the burden on the zeolite filter (RS30) as suggested by the fact that 'isoflurane breakthrough' was always found in the bench tests but only detected in one phase of the animal experiments. Uptake by the pigs was similar to the amount of isoflurane lost through the filter (RS30) in bench tests, thus explaining the small difference between consumption in the bench tests and animal experiments. In conclusion, by using a ventilator attached to the circle system with a zeolite reflector placed in between, isoflurane consumption is reduced by approximately 80% in a bench test and approximately 70% in an animal study. References 1 jackson DE. A new method for the production of general analgesia and anaesthesia with a description of the apparatus used.) Lab Clin Med 1915; I: Waters RM. Advantages and technique of carbon dioxide filtration with inhalation anaesthesia. Anesth Analg 1926; 5: Raventos j. The action of fluothane-a new volatile anaesthetic. Br) Pharmacol 1956; 1I: jones, RM. Desflurane and sevoflurane: inhalation anaesthetic for this decade. Br) Anaesth 1990; 65: Rowland FS. Chlorofluorocarbons and the depletion of stratospheric ozone. Am Sci 1989; 77: Bengtson jp, Sonander H, Stenqvist O. Preservation of humidity and heat of respiratory gases during Anaesthesia. Acta Anaesthesiol Scand 1987; 31: Cotter SM, Petros Aj, Dore C], Barber ND, White DC. Low flow anaesthesia-practice, cost implications and acceptability. Anaesthesia 199I; 46: I Wilson RS, Laver Mj. Oxygen analysis: advances in methodology. Anesthesiology 1972; 37: I Virtue RW. Minimal-flow nitrous oxide anesthesia. Anesthesiology 1974; 40: Thomasson R, Luttropp HH, Werner O. A reflection filter for isoflurane and other anaesthetic vapours. Eur) Anaesthesiol 1989; 6: Perhag L, Larsson A, Luttropp HH, Persson j, Werner O. Method to deliver a preset inspired concentration of anesthetic regardless of degree of rebreathing. Anesthesiology 1991; 75: AIOIO 12 Stevens WC, Kingston HGG. Inhalation anesthesia. In: Barasch PG, Cullen BF, Stoelting RK, eds. Clinical Anesthesia, 2nd edn. Philadelphia: jb Lippincott, 1992; Thomasson R. Zeolites, Curvature and Anesthesia. Thesis, University of Lund, Sweden. 14 Lowe Hj, Ernst EA. The Quantitative Practice ofanesthesia-use of Closed Circuit. Baltimore; Williams and Wilkins, 1981; 16,
NITROUS OXIDE ELIMINATION AND DIFFUSION HYPOXIA DURING NORMO- AND HYPOVENTILATION
British Journal of Anaesthesia 1993; 71: 189-193 NITROUS OXIDE ELIMINATION AND DIFFUSION HYPOXIA DURING NORMO- AND HYPOVENTILATION S. EINARSSON, O. STENQVIST, A. BENGTSSON, E. HOULTZ AND J. P. BENGTSON
More informationSoda lime temperatures during low-flow sevoflurane anaesthesia and differences in dead-space.
Soda lime temperatures during low-flow sevoflurane anaesthesia and differences in dead-space. Luttropp, Hans-Henrik; Johansson, Anders Published in: Acta Anaesthesiologica Scandinavica DOI: 10.1034/j.1399-6576.2002.460505.x
More informationPharmacokinetics. Inhalational Agents. Uptake and Distribution
Pharmacokinetics Inhalational Agents The pharmacokinetics of inhalational agents is divided into four phases Absorption Distribution (to the CNS Metabolism (minimal Excretion (minimal The ultimate goal
More informationComparison of patient spirometry and ventilator spirometry
GE Healthcare Comparison of patient spirometry and ventilator spirometry Test results are based on the Master s thesis, Comparison between patient spirometry and ventilator spirometry by Saana Jenu, 2011
More informationEnd-tidal Control. Introduction. GE Healthcare
GE Healthcare End-tidal Control Introduction End-tidal Control* provides automated target control for volatile anaesthetic agent and patient oxygen during general anaesthesia. End-tidal Control (EtC) is
More informationG. MORIWAKI, H. BITO AND K. IKEDA. Summary. Patients and methods. British Journal of Anaesthesia 1997; 79:
British Journal of Anaesthesia 1997; 79: 782 786 Partly exhausted soda lime or soda lime with water added, inhibits the increase in compound A concentration in the circle system during low-flow sevoflurane
More informationEffect of Charcoal Filter on the Emergence from Sevoflurane Anesthesia in a Semi-Closed Rebreathing Circuit
Original Article DOI 10.3349/ymj.2011.52.4.668 pissn: 0513-5796, eissn: 1976-2437 Yonsei Med J 52(4):668-672, 2011 Effect of Charcoal Filter on the Emergence from Sevoflurane Anesthesia in a Semi-Closed
More informationChapter 25. General Anesthetics
Chapter 25 1. Introduction General anesthetics: 1. Analgesia 2. Amnesia 3. Loss of consciousness 4. Inhibition of sensory and autonomic reflexes 5. Skeletal muscle relaxation An ideal anesthetic: 1. A
More informationKOALA. Adult Bacterial Viral Filters. Medical Pty Ltd
Specialising in Anaesthetic Devices & Airway Management Laproscopic Surgery Maternity & Infant Care Adult Bacterial Viral Filters High efficiency up to 99.9 % @ 30L/Min Low resistance < 2.5 cmh2o @ 50
More informationNomoLine No-moisture sampling lines for intubated and non-intubated patients in low- and high-humidity applications
NomoLine No-moisture sampling lines for intubated and non-intubated patients in low- and high-humidity applications Advanced technology enables cost-effective, hassle-free sidestream capnography and gas
More informationNomoLine No-moisture sampling lines for intubated and non-intubated patients in low- and high-humidity applications
NomoLine No-moisture sampling lines for intubated and non-intubated patients in low- and high-humidity applications Advanced technology enables cost-effective, hassle-free sidestream capnography and gas
More informationISPUB.COM. Review Of Currently Used Inhalation Anesthetics: Part II. O Wenker SIDE EFFECTS OF INHALED ANESTHETICS CARDIOVASCULAR SYSTEM
ISPUB.COM The Internet Journal of Anesthesiology Volume 3 Number 3 O Wenker Citation O Wenker.. The Internet Journal of Anesthesiology. 1998 Volume 3 Number 3. Abstract SIDE EFFECTS OF INHALED ANESTHETICS
More informationBronchoscopes: Occurrence and Management
ORIGIAL ARTICLES Res tk iratory Acidosis wi the Small Ston-Hopkins Bronchoscopes: Occurrence and Management Kang H. Rah, M.D., Arnold M. Salzberg, M.D., C. Paul Boyan, M.D., and Lazar J. Greenfield, M.D.
More informationEnvironmental Health & Safety Policy Manual
Environmental Health & Safety Policy Manual Issue Date: 5/31/2017 Policy # EHS-400.17 Isoflurane Use and Exposure Control Procedures 1.0 PURPOSE: LSUHSC is committed to keeping all exposures to hazardous
More informationAbstract. Med. J. Cairo Univ., Vol. 77, No. 3, June: , MAGDA S. AZER, M.D.
Med. J. Cairo Univ., Vol. 77, No. 3, June: 71-77, 009 www.medicaljournalofcairouniversity.com Evaluation of Target-Controlled Inhalational Anesthesia in Pediatrics Using the Newly Introduced Zeus Anesthesia
More informationGas exchange with a reflecting system for inhalational anaesthesia
Gas exchange with a reflecting system for inhalational anaesthesia Walther Sturesson, Louise Published: 2013-01-01 Link to publication Citation for published version (APA): Walther Sturesson, L. (2013).
More informationGE Healthcare Life Sciences. Quality matters. Whatman TM filters for air monitoring
GE Healthcare Life Sciences Quality matters Whatman TM filters for air monitoring Quality matters Why does quality matter? Particulate testing Examples include PM 10 & PM 2.5 particulate monitoring. Quality
More informationThe effect of a pediatric heat and moisture exchanger on dead space in healthy pediatric anesthesia
Clinical Research Article Korean J Anesthesiol 2012 May 62(5): 418-422 http://dx.doi.org/10.4097/kjae.2012.62.5.418 The effect of a pediatric heat and moisture exchanger on dead space in healthy pediatric
More informationSINGLE BREATH INDUCTION OF ANAESTHESIA WITH ISOFLURANE
Br. J. Anaesth. (987), 59, 24-28 SINGLE BREATH INDUCTION OF ANAESTHESIA WITH ISOFLURANE J. M. LAMBERTY AND I. H. WILSON Two studies have demonstrated that the induction of anaesthesia using a single breath
More informationInhalational Anesthesia. Munir Gharaibeh, MD, PhD, MHPE School of Medicine The University of Jordan February, 2018
Inhalational Anesthesia School of Medicine The University of Jordan February, 2018 mgharaib@ju.edu.jo Inhalational Anesthesia n Gases or volatile liquids n Administration and Elimination is by the lungs
More informationDr. Gavin Parker PINCHER CREEK AB 140
Society of Rural Physicians of Canada 26TH ANNUAL RURAL AND REMOTE MEDICINE COURSE ST. JOHN'S NEWFOUNDLAND AND LABRADOR APRIL 12-14, 2018 Dr. Gavin Parker PINCHER CREEK AB 140 MALIGNANT HYPERTHERMIA Malignant
More informationPharmacokinetics of propofol when given by intravenous
Br. J. clin. Pharmac. (199), 3, 144-148 Pharmacokinetics of propofol when given by intravenous infusion DENIS J. MORGAN', GWEN A. CAMPBELL2,* & DAVID P. CRANKSHAW2 'Victorian College of Pharmacy, 381 Royal
More informationGE Healthcare Life Sciences. Quality matters. Whatman TM filters for air monitoring
GE Healthcare Life Sciences Quality matters Whatman TM filters for air monitoring Quality matters Why does quality matter? Particulate testing Examples include PM 10 & PM 2.5 particulate monitoring. Quality
More informationComparison of isoflurane and sevoflurane for short-term anesthesia in piglets
Veterinary Anaesthesia and Analgesia, 2007, 34, 117 124 doi:10.1111/j.1467-2995.2006.00309.x RESEARCH PAPER Comparison of isoflurane and sevoflurane for short-term anesthesia in piglets David S Hodgson
More informationSurgical Care at the District Hospital. EMERGENCY & ESSENTIAL SURGICAL CARE
Surgical Care at the District Hospital 1 14 Practical Anesthesia Key Points 2 14.1 General Anesthesia Have a clear plan before starting anesthesia Never use an unfamiliar anesthetic technique in an emergency
More information1. When a patient fails to ventilate or oxygenate adequately, the problem is caused by pathophysiological factors such as hyperventilation.
Chapter 1: Principles of Mechanical Ventilation TRUE/FALSE 1. When a patient fails to ventilate or oxygenate adequately, the problem is caused by pathophysiological factors such as hyperventilation. F
More informationVolatile Anaesthetic Agents (Basic Principles)
Volatile Anaesthetic Agents (Basic Principles) KSS School of Anaesthesia Basic Science Course South Coast Training Group Dr S M Walton Consultant Anaesthetist Eastbourne What do you need to know about
More informationCuffed Tracheal Tubes in Children - Myths and Facts. PD Dr. Markus Weiss Department of Anaesthesia University Children s Hospital Zurich Switzerland
Cuffed Tracheal Tubes in Children - Myths and Department of Anaesthesia University Children s Hospital Zurich Switzerland PRO Reduced gas leak, low fresh gas flow Decreased atmospheric pollution Constant
More informationCapnography Connections Guide
Capnography Connections Guide Patient Monitoring Contents I Section 1: Capnography Introduction...1 I Section 2: Capnography & PCA...3 I Section 3: Capnography & Critical Care...7 I Section 4: Capnography
More informationInhalation anesthesia. Somchai Wongpunkamol,MD Anes., CMU
Inhalation anesthesia Somchai Wongpunkamol,MD nes., CMU Inhalation anesthetics is agent that possess anaesthetic qualities that are administered by breathing through an anaesthesia mask or ET tube connected
More informationStudy Of Effects Of Varying Durations Of Pre-Oxygenation. J Khandrani, A Modak, B Pachpande, G Walsinge, A Ghosh
ISPUB.COM The Internet Journal of Anesthesiology Volume 20 Number 1 J Khandrani, A Modak, B Pachpande, G Walsinge, A Ghosh Citation J Khandrani, A Modak, B Pachpande, G Walsinge, A Ghosh.. The Internet
More informationCRITICAL CARE AquaVENT
CRITICAL CARE AquaVENT Heated Breathing Circuits Why Humidification? Normal Respiration During normal respiration, inhaled gases are warmed and moistened as they pass from the upper respiratory tract to
More informationRespiratory Physiology Part II. Bio 219 Napa Valley College Dr. Adam Ross
Respiratory Physiology Part II Bio 219 Napa Valley College Dr. Adam Ross Gas exchange Gas exchange in the lungs (to capillaries) occurs by diffusion across respiratory membrane due to differences in partial
More informationBPK 312 Nutrition for Fitness & Sport. Lecture 4 - Part 2. Measurement of Energy in Food & During Physical Activity
BPK 312 Nutrition for Fitness & Sport Lecture 4 - Part 2 Measurement of Energy in Food & During Physical Activity 1. Heat of Combustion & Energy Value of Foods 2. Measurement of Human Energy Expenditure
More informationW. J. RUSSELL*, M. F. JAMES
Anaesth Intensive Care 2004; 32: 644-648 The Effects on Arterial Haemoglobin Oxygen Saturation and on Shunt of Increasing Cardiac Output with Dopamine or Dobutamine During One-lung Ventilation W. J. RUSSELL*,
More informationEuropean Board of Anaesthesiology (EBA) recommendations for minimal monitoring during Anaesthesia and Recovery
European Board of Anaesthesiology (EBA) recommendations for minimal monitoring during Anaesthesia and Recovery INTRODUCTION The European Board of Anaesthesiology regards it as essential that certain core
More informationKinetics and Monitoring of Inhaled Anesthetics. Copyright , James H Philip, all rights reserved
Kinetics and Monitoring of Inhaled Anesthetics Copyright 1980-2014, James H Philip, all rights reserved Ready Kinetics and Monitoring of Inhaled Anesthetics James H. Philip, M.E.(E.), M.D. Anesthesiologist
More informationPharmacology: Inhalation Anesthetics
Pharmacology: Inhalation Anesthetics This is an edited and abridged version of: Pharmacology: Inhalation Anesthetics by Jch Ko, DVM, MS, DACVA Oklahoma State University - Veterinary Medicine, February
More informationMulti-center (5 centers); United States and Canada. September 10, 1992 to April 9, 1993
vi STUDY SYNOPSIS Study Number: Title: Investigator: GHBA-534 A Phase III, Randomized, Open-Label Study To Compare The Safety, Tolerability And Recovery Characteristics of Sevoflurane Versus Halothane
More informationSevoflurane Output in the Isoflurane/Halothane Diamedica Draw-over Vaporiser
Original Article 252 Sevoflurane Output in the Isoflurane/Halothane Diamedica Draw-over Vaporiser Woo Chiao Tay, 1 MBBS, Wah Tze Tan, 2 MBBS, M Med (Anaes) Abstract Introduction: Draw-over anaesthesia
More informationI. Subject: Continuous Positive Airway Pressure CPAP by Continuous Flow Device
I. Subject: Continuous Positive Airway Pressure CPAP by Continuous Flow Device II. Policy: Continuous Positive Airway Pressure CPAP by the Down's system will be instituted by Respiratory Therapy personnel
More informationTissue Hypoxia and Oxygen Therapy
Tissue Hypoxia and Oxygen Therapy ก ก ก ก ก ก 1. ก ก 2. ก ก 3. tissue hypoxia 4. ก ก ก 5. ก ก ก 6. ก กก ก 7. ก ก tissue hypoxia ก ก ก ก 1. Pathway of oxygen transport 2. Causes of tissue hypoxia 3. Effect
More informationAQA B3.1 Movement of molecules in and out of cells LEVEL 1 Q
AQA B3.1 Movement of molecules in and out of cells LEVEL 1 Q 154 minutes 154 marks Page 1 of 44 Q1. The table shows the percentage of some gases in the air a boy breathed in and out. Gases Air breathed
More informationPULMONARY FUNCTION TESTING. Purposes of Pulmonary Tests. General Categories of Lung Diseases. Types of PF Tests
PULMONARY FUNCTION TESTING Wyka Chapter 13 Various AARC Clinical Practice Guidelines Purposes of Pulmonary Tests Is lung disease present? If so, is it reversible? If so, what type of lung disease is present?
More informationBritish Journal of Anaesthesia 104 (5): (2010) doi: /bja/aeq080 Advance Access publication March 30, 2010
RESPIRATION AND THE AIRWAY Lung recruitment and positive airway pressure before extubation does not improve oxygenation in the post-anaesthesia care unit: a randomized clinical trial A. B. Lumb 1, S. J.
More informationGas Exchange in the Tissues
Gas Exchange in the Tissues As the systemic arterial blood enters capillaries throughout the body, it is separated from the interstitial fluid by only the thin capillary wall, which is highly permeable
More information8/13/11. RSPT 1410 Humidity & Aerosol Therapy Part 3. Humidification Equipment. Aerosol Therapy
1 RSPT 1410 Humidity & Aerosol Therapy Part 3 Wilkins: Chapter 35, p. 775-799 Cairo: Chapter 4, p. 88-143 2 Humidification Equipment A humidifier is a device that adds molecular liquid (e.g. water vapor)
More informationThe diagram below shows the parts of the body that digest and absorb food.
The diagram below shows the parts of the body that digest and absorb food. It also shows some details about the structure of the stomach. (a) Complete the table to show whether each structure is an organ,
More informationJanuary 27, 1992 to November 26, A total of 120 patients (60/site) were enrolled in the study as follows: PATIENT ENROLLMENT
STUDY SYNOPSIS Study Number: Title: GHBA-533 A Phase II, Randomized, Open-Label Study to Compare the Safety and Efficacy of Sevoflurane Versus Halothane Administered with Nitrous Oxide and Oxygen in ASA
More informationNITROUS OXIDE SEDATION CAUSES POST-HYPERVENTILATION APNOEAf
British Journal of Anaesthesia 1991; 67: 7-12 NITROUS OXIDE SEDATION CAUSES POST-HYPERVENTILATION APNOEAf D. NORTHWOOD, D. J. SAPSFORD, J. G. JONES, D. GRIFFITHS AND C. WILKINS SUMMARY We have studied,
More informationLow flow anaesthesia: economic, eco friendly and effective
Original Research Article DOI: 1.18231/2394-4994.217.77 VC Shelgaonkar 1, Annu Choudhary 2,* 1 Associate Professor, 2 Senior Resident, Dept. of Anaesthesia, Maharashtra University of Health Sciences, Nashik,
More informationEffect of low-dose enflurane on the ventilatory response to hypoxia in humans
British Journal of Anaesthesia 1994; 72: 59-514 CLINICAL INVESTIGATIONS Effect of low-dose enflurane on the ventilatory response to hypoxia in humans B. NAGYOVA, K. L. DORRINGTON AND P. A. ROBBINS SUMMARY
More informationCAPNOGRAPHY DR JOHN ROOS
CAPNOGRAPHY DR JOHN ROOS Abraham Lincoln If you want me to speak for an hour give me a moment s notice if you want me to speak for five minutes give me a week. Missed oesophageal intubation Many studies
More informationMinimal & Moderate Sedation. Focus on British Columbia
Minimal & Moderate Sedation Focus on British Columbia Continuum of Sedation in BC Single Oral Sedative Nitrous Oxide & Oxygen Single Oral Sedative and Nitrous Oxide & Oxygen Moderate Sedation Minimal Sedation
More informationADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART TWO
Vet Times The website for the veterinary profession https://www.vettimes.co.uk ADVANCED PATIENT MONITORING DURING ANAESTHESIA: PART TWO Author : CARL BRADBROOK Categories : Vets Date : October 14, 2013
More informationI. Subject: Medication Delivery by Metered Dose Inhaler (MDI)
I. Subject: Medication Delivery by Metered Dose Inhaler (MDI) II. Policy: Aerosol medication administration by metered dose inhaler will be performed upon a physician's order by Respiratory Therapy personnel.
More informationScience Class 8 Topic: Respiration And Circulation Reinforcement Worksheet
Science Class 8 Topic: Respiration And Circulation Reinforcement Worksheet Name: Sec: Date: Q1. MCQs. 1) Which type of blood vessels carries blood away from the heart? a) Veins b) Arteries c) Capillaries
More informationComputational Fluid Dynamics Modeling of Amsino OneMask Oxygen Mask
Computational Fluid Dynamics Modeling of Amsino OneMask Oxygen Mask Abstract This study s objective was to model the Amsino OneMask Oxygen Mask using Computational Fluid Dynamics (CFD). A three-dimensional
More informationAnatomy & Physiology 2 Canale. Respiratory System: Exchange of Gases
Anatomy & Physiology 2 Canale Respiratory System: Exchange of Gases Why is it so hard to hold your breath for Discuss! : ) a long time? Every year carbon monoxide poisoning kills 500 people and sends another
More informationAccumulation of foreign gases during closed-system anaesthesia
British Journal of Anaesthesia 1996; 76: 668 672 Accumulation of foreign gases during closed-system anaesthesia L. VERSICHELEN, G. ROLLY AND H. VERMEULEN Summary In a previous study, accumulation of methane
More informationBLY3H (JAN09BLY3H01) General Certifi cate of Secondary Education January Unit Biology B3. Higher Tier. Time allowed: 45 minutes
Surname Other Names For Examiner s Use Centre Number Candidate Number Candidate Signature General Certifi cate of Secondary Education January 2009 BIOLOGY Unit Biology B3 Higher Tier BLY3H H Monday 12
More informationTransnasal Humidified Rapid Insufflation Ventilatory Exchange (THRIVE): An Optimal Method of Preoxygenation for General Anaesthesia in Obstetrics
Transnasal Humidified Rapid Insufflation Ventilatory Exchange (THRIVE): An Optimal Method of Preoxygenation for General Anaesthesia in Obstetrics Dr E McMaster, Dr E Gent, Dr T Mahendrayogam, Dr A Surendran
More informationPhysiological normal values (Depends on age, strain, health status, type of anesthesia, etc.)
UBC Animal Care Guidelines This SOP will be used for: Inhalational anesthesia SOP: ACC-01-2017 Rodent Anesthesia Injectable anesthesia Submitted by: Shelly McErlane both inhalational and injectable anesthesia
More informationFENTANYL BY CONSTANT RATE I.V. INFUSION FOR POSTOPERATIVE ANALGESIA
Br. J. Anaesth. (1985), 5, 250-254 FENTANYL BY CONSTANT RATE I.V. INFUSION FOR POSTOPERATIVE ANALGESIA W. S. NIMMO AND J. G. TODD is a synthetic opioid analgesic 50 times more potent than morphine, with
More informationUse of the Intubating Laryngeal Mask Airway
340 Anesthesiology 2000; 93:340 5 2000 American Society of Anesthesiologists, Inc. Lippincott Williams & Wilkins, Inc. Use of the Intubating Laryngeal Mask Airway Are Muscle Relaxants Necessary? Janet
More informationPHYSIOEX 3.0 EXERCISE 33B: CARDIOVASCULAR DYNAMICS
PHYSIOEX 3.0 EXERCISE 33B: CARDIOVASCULAR DYNAMICS Objectives 1. To define the following: blood flow; viscosity; peripheral resistance; systole; diastole; end diastolic volume; end systolic volume; stroke
More informationPart 1: Imaging Spectroscopy: From air quality investigations to healthcare applications
Part 1: Imaging Spectroscopy: From air quality investigations to healthcare applications Part 2: Chemical Ionisation Reaction Mass Spectrometry (CIR-MS) for Breath Analysis Research 12 th March 2009 Roland
More informationCORESTA RECOMMENDED METHOD N 8
CORESTA RECOMMENDED METHOD N 8 DETERMINATION OF WATER IN THE MAINSTREAM SMOKE OF CIGARETTES BY GAS CHROMATOGRAPHIC ANALYSIS (August 1991) 1. FIELD OF APPLICATION The method is applicable to the particulate
More informationPCTH 400 Systematic Pharmacology
Objectives At the end of this session, you will be able to: 1. Define the components of general anesthesia; PCTH 400 Systematic Pharmacology Inhaled s and Amnestic Agents Dr. Peter Choi (peter.choi@ubc.ca)
More informationVancouver Coastal Health Guidelines for the use of Respiratory Equipment for Patients on Airborne Precautions in Acute Care Facilities
Vancouver Coastal Health Guidelines for the use of Respiratory Equipment for Patients on Airborne Precautions in Acute Care Facilities Goals 1. To meet respiratory care needs in patients who are on airborne
More informationCh 16 A and P Lecture Notes.notebook May 03, 2017
Table of Contents # Date Title Page # 1. 01/30/17 Ch 8: Muscular System 1 2. 3. 4. 5. 6. 7. 02/14/17 Ch 9: Nervous System 12 03/13/17 Ch 10: Somatic and Special Senses 53 03/27/17 Ch 11: Endocrine System
More informationa. Describe the physiological consequences of intermittent positive pressure ventilation and positive end-expiratory pressure.
B. 10 Applied Respiratory Physiology a. Describe the physiological consequences of intermittent positive pressure ventilation and positive end-expiratory pressure. Intermittent positive pressure ventilation
More information1.40 Prevention of Nosocomial Pneumonia
1.40 Prevention of Nosocomial Pneumonia Purpose Audience Policy Statement: The guideline is designed to reduce the incidence of pneumonia and other acute lower respiratory tract infections. All UTMB healthcare
More informationCapnography. Capnography. Oxygenation. Pulmonary Physiology 4/15/2018. non invasive monitor for ventilation. Edward C. Adlesic, DMD.
Capnography Edward C. Adlesic, DMD University of Pittsburgh School of Dental Medicine 2018 North Carolina Program Capnography non invasive monitor for ventilation measures end tidal CO2 early detection
More informationCough assist T70 for the Tracheostomy Child
Patient and Family Education Cough assist 70 for the racheostomy Child with or without a ventilator What is a Cough assist device? he Cough assist 70 device removes mucus (secretions) from your child s
More informationPORTO 2 VENT CPAP OS. Operator s Manual. PORTO 2VENT CPAP OS System Operator s Manual Part Number Rev I
PORTO 2 VENT CPAP OS Operator s Manual 1 2 TABLE of CONTENTS 1. Introduction 3. Operating Instructions 1a. Definitions 3a. Setting the CPAP Level 1b. General Description 3b. Applying the Breathing Circuit
More informationBi-Level Therapy: Boosting Comfort & Compliance in Apnea Patients
Bi-Level Therapy: Boosting Comfort & Compliance in Apnea Patients Objectives Describe nocturnal ventilation characteristics that may indicate underlying conditions and benefits of bilevel therapy for specific
More informationBritish Journal of Anaesthesia 96 (6): (2006) doi: /bja/ael092 Advance Access publication April 13, 2006 Emergence and recovery in childr
British Journal of Anaesthesia 96 (6): 779 85 (6) doi:.93/bja/ael92 Advance Access publication April 13, 6 Emergence and recovery in children after desflurane and isoflurane anaesthesia: effect of anaesthetic
More informationIssues in Cost Containment
Issues in Cost Containment Impact of a Cost Containment Program on the Use of Volatile Anesthetics and Neuromuscular Blocking Drugs James E Szocik, MD,* David W. Learned, MI) t Department of Anesthesiology,
More informationA randomised comparison between Cobra PLA and classic laryngeal mask airway and laryngeal tube during mechanical ventilation for general anaesthesia
ORIGINAL AND CLINICAL PAPERS Anaesthesiology Intensive Therapy 2013, vol. 45, no 1, 20 24 ISSN 1642 5758 DOI: 10.5603/AIT.2013.0004 www.ait.viamedica.pl A randomised comparison between Cobra PLA and classic
More informationAnaesthetic considerations for laparoscopic surgery in canines
Vet Times The website for the veterinary profession https://www.vettimes.co.uk Anaesthetic considerations for laparoscopic surgery in canines Author : Chris Miller Categories : Canine, Companion animal,
More informationArticles. The Advantages of Nebulization in the Treatment of Mechanically Ventilated Neonates. Kristin Smith, RRT-NPS
Articles The Advantages of Nebulization in the Treatment of Mechanically Ventilated Neonates Kristin Smith, RRT-NPS A major goal in the care of premature babies is growth, and so all therapies are applied
More informationWhy is diffusion insufficient to meet the oxygen requirements of multi-cellular organisms like
Page 95»Question» Question 1: Why is diffusion insufficient to meet the oxygen requirements of multi-cellular organisms like humans? Multicellular organisms such as humans possess complex body designs.
More informationAnesthesia for removal of inhaled foreign bodies in children
Pediatric Anesthesia 2004 14: 947 952 doi:10.1111/j.1460-9592.2004.01309.x Anesthesia for removal of inhaled foreign bodies in children AMIT SOODAN MD, DILIP PAWAR MD AND RAJESHWARI SUBRAMANIUM MD Department
More informationPercutaneous transtracheal ventilation in an obstructed airway model in post-apnoeic sheep
British Journal of Anaesthesia 113 (6): 139 4 (214) Advance Access publication 3 June 214. doi:1.193/bja/aeu188 RESPIRATION AND THE AIRWAY Percutaneous transtracheal ventilation in an obstructed airway
More informationThe use of laryngeal mask airway in dental treatment during sevoflurane deep sedation
Original Article pissn 2383-9309 eissn 2383-9317 J Dent Anesth Pain Med 2016;16(1):49-53 http://dx.doi.org/10.17245/jdapm.2016.16.1.49 The use of laryngeal mask airway in dental treatment during sevoflurane
More informationAdaptation to the waste anesthesia gas system: Gaps in knowledge and opportunities for positive environmental impact
OPEN ACCESS DOI: 10.5455/javar.2015.b106 ORIGINAL ARTICLE J. Adv. Vet. Anim. Res., 2(4): 388-395. Available at- http://bdvets.org/javar Volume 2 Issue 4 (December 2015) Adaptation to the waste anesthesia
More informationTHE BENCHMARK IN CAR AIR PURIFICATION TECHNOLOGY SHEPROS. Safety, Health & Environmental Product Solutions
THE BENCHMARK IN CAR AIR PURIFICATION TECHNOLOGY SHEPROS Safety, Health & Environmental Product Solutions Indoor Air Quality Hazards of Cars Results of recent studies indicate that levels of airborne chemicals
More informationpossibility of a secretion of adrenaline from the suprarenal glands resulting
355 J Physiol. (I942) IOI, 355-36I 6i2.014.465:577 I74.5 THE EFFECT OF ANAESTHESIA ON THE ADRENALINE CONTENT OF THE SUPRARENAL GLANDS BY P. C. ELMES AND A. A. JEFFERSON From the Department of Pharmacology,
More informationNI 60. Non-invasive ventilation without compromise. Homecare Pneumology Neonatology Anaesthesia. Sleep Diagnostics Service Patient Support
NI 60 Non-invasive ventilation without compromise Homecare Pneumology Neonatology Anaesthesia INTENSIVE CARE VENTILATION Sleep Diagnostics Service Patient Support NI 60 Non-invasive ventilation without
More informationDay-to-day management of Tracheostomies & Laryngectomies
Humidification It is mandatory that a method of artificial humidification is utilised when a tracheostomy tube is in situ, for people requiring oxygen therapy dry oxygen should never be given to someone
More informationAgency 71. Kansas Dental Board (Authorized by K.S.A and (Authorized by K.S.A and
Agency 71 Kansas Dental Board Articles 71-4. CONTINUING EDUCATION REQUIREMENTS. 71-5. SEDATIVE AND GENERAL ANAESTHESIA. 71-11. MISCELLANEOUS PROVISIONS. Article 4. CONTINUING EDUCATION REQUIREMENTS 71-4-1.
More informationPMT. Contains ribosomes attached to the endoplasmic reticulum. Genetic material consists of linear chromosomes. Diameter of the cell is 1 µm
1. (a) Complete each box in the table, which compares a prokaryotic and a eukaryotic cell, with a tick if the statement is correct or a cross if it is incorrect. Prokaryotic cell Eukaryotic cell Contains
More informationCARDIO-RESPIRATORY RESPONSE TO EXERCISE IN NORMAL CHILDREN
Clinical Science (1971) 40, 419431. CARDIORESPIRATORY RESPONSE TO EXERCISE IN NORMAL CHILDREN S. GODFREY, C. T. M. DAVIES, E. WOZNIAK AND CAROLYN A. BARNES Institute of Diseases of the Chest, London, and
More informationCardiorespiratory Physiotherapy Tutoring Services 2017
VENTILATOR HYPERINFLATION ***This document is intended to be used as an information resource only it is not intended to be used as a policy document/practice guideline. Before incorporating the use of
More information1. Influence of isoflurane exposure in pregnant rats on the learning and memory of offspring
1. Influence of isoflurane exposure in pregnant rats on the learning and memory of offspring Huang W, Dong Y, Zhao G, et al. BMC Anesthesiology 2018 18:5 Concerns remain about possible effects of general
More informationPilot Of Spontaneous Breathing Vs. Ventilated Model For Hemorrhage And Resuscitation In The Rabbit
Article ID: WMC001137 Pilot Of Spontaneous Breathing Vs. Ventilated Model For Hemorrhage And Resuscitation In The Rabbit Author(s):Dr. Jonathan S. Jahr, MD, Dr. Robert A. Gunther, PhD, Dr. Bernd Driessen,
More informationFOLLOW-UP MEDICAL CARE OF SERVICE MEMBERS AND VETERANS CARDIOPULMONARY EXERCISE TESTING
Cardiopulmonary Exercise Testing Chapter 13 FOLLOW-UP MEDICAL CARE OF SERVICE MEMBERS AND VETERANS CARDIOPULMONARY EXERCISE TESTING WILLIAM ESCHENBACHER, MD* INTRODUCTION AEROBIC METABOLISM ANAEROBIC METABOLISM
More informationPackaging technology and preservation of foods Fátima Poças Biotechnology Faculty Universidade Católica Portuguesa 2017
Packaging technology and preservation of foods Fátima Poças Biotechnology Faculty Universidade Católica Portuguesa 2017 Agenda Packaging systems for different food processing technologies: Frozen foods
More informationNON INVASIVE LIFE SAVERS. Non Invasive Ventilation (NIV)
Table 1. NIV: Mechanisms Of Action Decreases work of breathing Increases functional residual capacity Recruits collapsed alveoli Improves respiratory gas exchange Reverses hypoventilation Maintains upper
More information