Validation and application of a canine-specific automated high-sensitivity C-reactive protein assay

Size: px
Start display at page:

Download "Validation and application of a canine-specific automated high-sensitivity C-reactive protein assay"

Transcription

1 5751VDIXXX / Validation of a canine-specific hscrp assayhillström et al. 015 Full Scientific Report Validation and application of a canine-specific automated high-sensitivity C-reactive protein assay Journal of Veterinary Diagnostic Investigation 2015, Vol. 27(2) The Author(s) Reprints and permissions: sagepub.com/journalspermissions.nav DOI: / jvdi.sagepub.com Anna Hillström, 1 Ragnvi Hagman, Josefin Söder, Jens Häggström, Ingrid Ljungvall, Mads Kjelgaard-Hansen Abstract. Measurement of low concentrations of C-reactive protein (CRP) in dogs has previously been performed with nonautomated assays. The aim of this study was to validate an automated high-sensitivity CRP (hscrp) assay, developed by modifying a routinely used canine-specific immunoturbidimetric CRP test (ccrp). Imprecision, linearity under dilution, limit of blank (LOB), limit of detection (LOD), and limit of quantification (LOQ) were determined for the hscrp test, as well as the presence of prozone effect and interferences. The imprecision, measured as intra-assay variation, was 2.7%. The assay was acceptably linear under dilution. An analytically relevant prozone effect was present for samples with CRP concentration >150 mg/l, and there were mild interferences from hemolysis and lipemia. The LOB, LOD, and LOQ were 0.10 mg/l, 0.22 mg/l, and 0.50 mg/l, respectively. A method comparison study with a canine-specific enzyme-linked immunosorbent assay (ELISA) was performed, showing poor agreement between the hscrp test and the ELISA. An additional aim of the study was to apply the hscrp test to clinical research samples. Serum samples from 7 dogs undergoing ovariohysterectomy were collected pre- and postoperatively, and CRP was measured with both the ccrp and hscrp assay. The expected postoperative increase in CRP was detected earlier with the hscrp test, compared with the ccrp test. The hscrp assay was further applied on samples from 6 lean and 9 overweight dogs. There was no significant difference in CRP concentration between the groups (P = 0.06). In conclusion, the hscrp test had acceptable analytical performance, and the assay was successfully applied to clinical research samples. Key words: C-reactive protein; dogs; method comparison; obesity; validation. Introduction C-reactive protein (CRP) is a major positive acute phase protein (APP) in dogs, produced in the liver in response to increased concentrations of pro-inflammatory cytokines. 4,9 The hallmark of a major APP is a low blood concentration in the healthy dog, and a rapid and marked increase during inflammation. 4,9 This makes major APPs such as CRP valuable as clinical markers for diagnosing and monitoring inflammatory diseases in veterinary medicine. 4,9,12,26 In people, the use of CRP has extended beyond the role as a mere diagnostic test, and the association between this biomarker and lifestyle-related diseases such as obesity, diabetes mellitus, and cardiovascular disease is gaining increasing interest. 8,33 In this context, assays have been developed that measure CRP in the low or ultra-low concentration range, and these tests are then referred to as high-sensitivity CRP (hscrp) assays. The hscrp test is different from the routinely used CRP test regarding methodology applied, purpose of testing, and interpretation of the results. Examples of applications of hscrp testing in human beings are to assess the risk of future cardiovascular disease, 15,34,35,40 and as a prognostic marker in cancer. 1 Similarly, measurements of CRP concentrations in the low range have been performed in dogs in a research setting. It has been investigated if obesity in dogs is associated with increased CRP concentrations, 11,37 as it is in people, 5,32 and the association between cardiac disease and CRP concentrations has been studied. 7,25 In order to investigate minor changes in CRP concentrations in the low range in dogs, routine CRP tests such as the previously validated immunoturbidimetric assays 10,14,20 are not suitable. These tests are developed for assessing moderately to markedly increased CRP concentrations in ill dogs, and they are neither sufficiently accurate nor precise at low CRP concentrations. From the Department of Clinical Sciences, Swedish University of Agricultural Sciences, Uppsala, Sweden (Hillström, Hagman, Häggström, Ljungvall); Department of Anatomy, Physiology and Biochemistry, Swedish University of Agricultural Sciences, Uppsala, Sweden (Söder); and Translational Haemophilia Pharmacology, Novo Nordisk, Maaloev, Denmark (Kjelgaard-Hansen). 1 Corresponding Author: Anna Hillström, Department of Clinical Sciences, Swedish University of Agricultural Sciences, Box 7054, Uppsala, Sweden. anna.hillstrom@slu.se

2 Validation of a canine-specific hscrp assay 183 Instead, tests such as enzyme-linked immunosorbent assay (ELISA) 21 and time-resolved immunofluorometric assay (TR-IFMA) 31 have been used; these assays have low limits of detection (LODs) and can therefore be considered as hscrp methods. However, no automated hscrp test is currently available for use in dogs to our knowledge. The aim of this study was to validate an automated caninespecific hscrp assay. An additional aim was to apply this test for analyzing CRP in serum samples collected for clinical research purposes. We hypothesized that the hscrp test could detect increased CRP concentrations earlier in postoperative samples from dogs undergoing surgery, compared with the routinely used canine-specific CRP test. Second, we hypothesized that overweight dogs would have higher CRP concentrations compared to lean dogs. Material and methods A method validation study was performed to establish the analytical performance of the hscrp assay. After satisfactory completion of the validation study, the hscrp test was applied to serum samples from 2 clinical research studies. Animals and samples Method validation study. Blood samples were collected for diagnostic purposes from canine patients admitted to the University Animal Hospital at the Swedish University of Agricultural Sciences (UDS-SLU; Uppsala, Sweden), in tubes with clot activator. a Sera were prepared by centrifugation at 2,000 g for 5 min, and stored in cryotubes b at 20 C for a maximum of 4 weeks. Samples were thawed at room temperature and thoroughly mixed immediately prior to analysis. Clinical research studies. The current studies were approved by the Local Ethical Committees in Uppsala, Sweden. Informed owner consent was obtained prior to sampling. Blood samples were collected by venipuncture in tubes with clot activator. a Sera were prepared as described above, and were frozen in cryotubes. b C-reactive protein after surgery. The ability of the hscrp test to detect a postoperative increase in CRP concentration was investigated, and then compared with the performance of the routinely used canine-specific immunoturbidimetric CRP (ccrp) assay, c on which the hscrp test was based. Dogs undergoing ovariohysterectomy (OHE) at UDS-SLU were sampled immediately prior to surgery (baseline) and 2, 4, 6, and 8 hr after the start of surgery, defined as the time point of the initial skin incision. Inclusion criteria were that the dog had an elective open approach OHE, and that the preoperative CRP concentration measured with the ccrp assay was <6.8 mg/l, which was the reference interval for CRP at UDS. Samples were stored at 80 C for a maximum of 3 months prior to analysis. C-reactive protein in lean and overweight dogs. To study CRP concentrations in lean and overweight dogs, serum samples were obtained from healthy intact male Labrador Retrievers 3 6 years of age. The dog owners were interviewed concerning the health status of their dogs, and routine hematology and biochemistry profiles and urinalysis were performed in all dogs. All dogs underwent routine clinical examination, which also included visual evaluation of spontaneous lameness in walk and trot on a straight track, as well as examination of oral health status in nonsedated animals. Only dogs reported to be healthy by their owners, without significant abnormal findings on clinical examination and with laboratory test results within the reference intervals, were included in the study. The body condition score (BCS) was determined as previously described 24 by a single examiner, and dogs were allocated to either a lean group (BCS 5) or an overweight group (BCS 6). Samples were stored at 70 C for a maximum of 12 months prior to analysis. C-reactive protein analysis Validation study. C-reactive protein was measured with a hscrp assay, which was based on a previously validated, canine-specific immunoturbidimetric CRP method. c,14 In this assay, polyclonal chicken anti-canine CRP antibodies react with CRP in a serum sample, causing an increased turbidity that is measured spectrophotometrically. In the original assay, 2 μl of serum sample was mixed with 70 μl of reagent containing anti-crp antibodies. To develop the hscrp test, the assay application was modified by increasing the sample volume 10-fold to 20 μl, whereas the reagent volume was kept unchanged. The calibrator kit of the original CRP assay was used to create the calibrator curve for the hscrp test, after manually prediluting the calibrators 1:10 with 0.9% NaCl immediately prior to calibration. Analyses were performed on a fully automated, open-system clinical chemistry/immunoassay analyzer. d Two control pools made of pooled canine sera were analyzed once daily during the experiment. Clinical research studies. Serum samples from dogs undergoing surgery were analyzed with the hscrp and the ccrp assay in duplicates in a single run. The limit of quantification (LOQ) of the ccrp assay was 6.8 mg/l, and samples below this limit were reported as <6.8 mg/l. Samples from the lean and overweight dogs were first analyzed with a previously validated human immunoturbidimetric CRP assay e,20 calibrated with canine CRP. f All samples that had CRP concentrations <30 mg/l were thereafter reanalyzed with the hscrp assay in duplicates in a single run. Quality goals and determination of assay characteristics Quality goals were derived from data of biological variation in healthy dogs, 22 as recommended. 19 The maximal allowable

3 184 Hillström et al. imprecision (CV max ) was set to 12%, the maximal allowable bias (bias max ) to 9.5%, and maximal allowable total error (TE a ) to 29.6%. 22 The maximal optimal imprecision (CV opt ), bias (bias opt ), and total error (TE opt ) were 6.1%, 4.8%, and 14.8%, respectively. 13,22 The within-run imprecision was determined by analyzing 3 serum samples in 15 replicates in 1 run, and the between-run imprecision by measuring 3 samples once daily on 5 consecutive days; the latter samples were frozen in aliquots to avoid repeated freeze thaw episodes. The intra- and interassay variations, expressed as coefficients of variation (CV), were calculated and compared to CV max and CV opt. No recalibration was performed during the imprecision study. In a study of linearity under dilution, a sample with CRP concentration near the upper end of the assay range was used. The sample was manually diluted with 0.9% NaCl to obtain 11 samples with concentrations of 0.9, 0.8, 0.7, 0.6, 0.5, 0.4, 0.3, 0.2, 0.1, 0.05, and times the original concentration. Samples were analyzed in duplicate in a single run. Inaccuracy was further tested in a spike and recovery study. A serum pool with low CRP concentration (base pool) was spiked with 4.7 mg/l and 8.7 mg/l of purified canine CRP. f The base pool and the spiked samples were analyzed in triplicate, and recovery of the spiked amount was calculated. Acceptable recovery after spiking was set to %. Presence of prozone effect was evaluated by analyzing 6 serum samples with both the hscrp and ccrp assay. The CRP concentrations and the absorbances reported by the hscrp method were compared to the CRP concentrations obtained with the ccrp method, which has no analytical relevant prozone effect up to at least 1,200 mg/l. 14 An analytical relevant prozone effect was considered to be present if samples with high CRP concentrations were reported to be false-low by the hscrp assay. The limit of blank (LOB) was determined by analyzing a blank sample, 0.9% NaCl, 20 times in a row. LOB was calculated as follows 28 : LOB = mean SD blank blank The limit of detection (LOD) was calculated according to the following formula 28 : LOD = LOB SD low sample, where the standard deviation (SD) of a low sample was determined by analyzing a sample with an expected CRP concentration of 0.25 mg/l, 10 times in a row. The LOQ, the lowest amount of analyte in a sample that can be quantitatively determined with stated acceptable imprecision and trueness, 28 was established. A start sample with a CRP concentration of approximately 8 mg/l, as determined by the hscrp assay, was manually diluted with 0.9% NaCl to the expected concentrations of 1.0, 0.5, and 0.25 mg/l. The diluted samples were analyzed 10 times in a row on 2 different days, with a recalibration performed between the 2 days. The start sample was analyzed in triplicate on both days. Bias in this experiment was defined as the difference between the observed mean value and the expected value, which was based on the concentration of the start sample and dilution factors. Imprecision was calculated as the CV value of the replicate measurements. The lowest concentration that had TE smaller than TE a (29.6%) was set as the LOQ of the assay. Interference from hemolysis and lipemia was tested as previously described. 14 A hemolytic solution was prepared according to Clinical and Laboratory Standards Institute guidelines, 6 and a commercial fat emulsion g was purchased for testing lipemia. Test pools were created by adding interferents to serum samples, to obtain 3 hemolytic test pools and 3 lipemic test pools. The final concentrations of hemoglobin in the test pools were 5 g/l and 10 g/l, and final triglyceride concentrations were 10 g/l and 25 g/l. Control pools were made with the same serum samples used for creating the test pools, but, instead of an interfering substance, the same volume of 0.9% NaCl was added to create hemolytic control pools, and the same volume of distilled water was added to create lipemic control pools. All samples were measured in triplicate. An interfering effect causing a change in CRP concentration of 10% was considered acceptable. The hscrp method was compared with a previously validated canine high-sensitivity CRP ELISA, h,21 in order to investigate if the methods could be used interchangeably. According to the manufacturer, the reportable range of the ELISA was mg/l when a 1:500 predilution of samples was taken into account. To optimize the assay for samples in the low range, samples in this study were prediluted 1:100, corresponding to a reportable range of mg/l. The exact LOQ of the ELISA was not investigated, and 0.75 mg/l was set as the concentration corresponding to LOQ. To evaluate performance of the ELISA after modification, linearity was assessed by serial diluting a serum sample with CRP concentration of approximately 10 mg/l. The diluted samples were analyzed in triplicate in 1 run, and the ELISA was found to be acceptably linear using the same acceptance criteria as for the hscrp study (data not shown). The imprecision of the ELISA, calculated as CV value of the triplicate measurements, was lower than the CV value for intra-assay variation reported by the manufacturer ( %). With the exception of the modified predilution step, the ELISA testing was performed according to the manufacturer s instructions. In the method comparison study, serum samples were analyzed in duplicate with both methods within 2 hr using a single 96-well ELISA plate. The CVs of the duplicate measurements were calculated, and if it was higher than CV max (12%) with either of the methods, the sample was excluded. Two canine serum control samples were analyzed with both methods.

4 Validation of a canine-specific hscrp assay 185 Table 1. Imprecision for an automated canine-specific high-sensitivity C-reactive protein assay.* Intra-assay variation Interassay variation Sample Mean (mg/l) SD (mg/l) CV (%) Mean (mg/l) SD (mg/l) CV (%) * Intra-assay variation was based on 15 replicate measurements in 1 run. Interassay variation was determined by analyzing 3 samples once daily on 5 consecutive days. SD = standard deviation; CV = coefficient of variation. Statistical analysis Method validation study. Statistical analyses were performed using a statistical software program. i Arithmetic means, SDs, and CVs were calculated using routine descriptive statistics. Linearity under dilution was evaluated by visual inspection, and by first-order and polynomial regression analysis, 27 with a significance level of α = Statistical nonlinearity was present if the second- or third-order model had better fit than the first-order model. Analytical relevant nonlinearity was present if the difference in CRP concentration predicted by the linear model and the best-fitting model exceeded TE a. In the study of interference, samples were measured in triplicate (n = 3). This number of replicates was required for detecting interference with 95% confidence and power, and was calculated by the following formula 6 : ( ) 2 2 s/d 1 α/2 1 β max, n = z + z where α = 0.05, β = 0.05, s = within-run precision of the method, and d max = maximum allowable difference caused by interference (±10%). The difference in concentration between the test pool and control pool (d obs ) was calculated. The 95% confidence intervals (CIs) for d obs were calculated according to the following equation 6 : d ± t s (2/n) obs n 1 The acceptance limits, expressed in mg/l, were ±0.1 CRP control pool. If the 95% CI of d obs was within the acceptance limits, it was concluded with 95% confidence that no interfering effect was present. In the method comparison study, a Bland Altman plot including the 95% limits of agreement was used to evaluate the difference between methods. 2 The mean difference (d mean ) between the methods and the SD of the differences (SD diff ) were calculated. The 95% limits of agreement were calculated by the formula d mean ± 1.96 SD diff, and the area between the upper and lower limit in the plot represented the range in which 95% of the differences between the methods were expected to be. Clinical research studies. Statistical analyses were performed using a commercially available software program. j Data was examined graphically, and normality was assessed using the Anderson Darling test. Log transformation was performed on nonnormal distributed data. To investigate at which time point after surgery an increase in CRP could be detected compared to the baseline, data was log-transformed and evaluated with a paired t-test with Bonferroni correction. A 2-sample t-test was used to compare CRP concentrations in lean and overweight dogs, after log-transformation of data. CRP concentrations <LOQ were assigned a value of 0.5 LOQ for the statistical analyses. The significance level was set at P < Results Method validation study The modified calibrator curve ranged from 0 to 30 mg/l. In the imprecision study, the CV values were lower than CV opt (6.1%) for all tested samples (Table 1). In the study of linearity under dilution, data was evaluated visually by comparing the observed concentration for each diluted sample with the expected concentration (Fig. 1). The third-order polynomial regression model showed best fit, indicating the presence of statistical nonlinearity. The difference in predicted CRP concentration by the first- and third-order model was <TE a for all samples, except for the sample with lowest CRP concentration. For this sample, linear regression predicted a value of 0.42 mg/l, compared to the value of 0.67 mg/l predicted by the third-order model. When the analysis was repeated with the lowest sample excluded, the difference between the first- and third-order model was <TE a for all samples. In the spike and recovery study, the base pool had a CRP concentration of 3.30 mg/l, and the recoveries after spiking were 114% and 117%, respectively. An analytical relevant prozone effect was present for samples with CRP concentrations greater than ~150 mg/l (Fig. 2). Two samples with CRP concentrations of 228 mg/l and 311 mg/l had absorbances that were lower than the absorbance of the upper calibrator of the hscrp assay. The CRP concentrations of these samples were reported to be 20 mg/l and 14 mg/l,

5 186 Hillström et al. measure CRP concentrations between 0.32 and 1.02 times the ELISA concentration. This difference was larger than could be accepted based on the predetermined TE a (29.6%). In method comparison studies, the imprecision of both methods should be considered; when measuring samples in duplicate with both methods, as performed in the current study, this will cause the SD of the difference to be lower than if single measurements are performed. This can be corrected when calculating the limits of agreement 2 ; however, because it had already been concluded that the methods did not agree, further calculations were not performed in this study. Figure 1. Linearity under dilution for an automated caninespecific high-sensitivity C-reactive protein (hscrp) assay. respectively, instead of being correctly reported as >30 mg/l by the assay. LOB and LOD was 0.10 mg/l and 0.22 mg/l, respectively. The LOQ was set to 0.5 mg/l; TE at this concentration was 26.5%, which was lower than TE a of 29.6%. At the expected concentration of 0.25 mg/l, TE was 70.3%. The CV values for the samples with expected CRP concentrations of 1.0 mg/l, 0.5 mg/l, and 0.25 mg/l were 4.7%, 9.0%, and 29.3%, respectively. Interference from hemolysis was present at 10 g/l hemoglobin in a sample with CRP concentration of approximately 3 mg/l, but not in the other tested hemolytic samples (Table 2). In the experiment in which a commercial fat emulsion was added to serum, no interfering effect was detected at a triglyceride concentration of 10 g/l, and only minor interference was present at a triglyceride concentration of 25 g/l (Table 2). In the method comparison study, 40 patient samples were included. No samples were excluded because of unacceptable CV of duplicate measurements (>12%); the median CV values of the duplicate measurements were 3.0% for the ELISA and 0.9% for the hscrp assay. Three samples had CRP concentrations <LOQ with both methods, and 4 samples had concentrations above the upper limit of the ELISA (12 mg/l). For the remaining 33 samples, visual inspection of the difference plot revealed a marked difference between the methods (Fig. 3). Because the differences between the 2 methods increased with increasing CRP concentrations, data was logtransformed before calculating the 95% limits of agreement (Fig. 4). The log-transformed bias was 0.56 and the 95% limits of agreements (95% CI) were 1.14 ( 1.32; 0.96) and 0.02 ( 0.16; 0.20). The anti-logs of the lower and upper limits of agreement were 0.32 and 1.02, respectively. The interpretation was that in 95% of the cases, the hscrp assay will Clinical research studies High-sensitivity CRP after surgery. Seven dogs of 4 different breeds were included in the study: 3 German Shepherd dogs, 2 Labrador Retrievers, 1 Airedale Terrier, and 1 Danish-Swedish Farmdog. The median age was 4 years (range: 1 7 years). The median CRP concentrations measured with the hscrp assay was 0.83 mg/l prior to surgery (baseline). At 2, 4, 6, and 8 hr after surgery, the median CRP concentrations were 0.69 mg/l, 1.44 mg/l, 6.14 mg/l, and 13.9 mg/l, respectively. The difference in CRP concentration compared with baseline was statistically significant at 6 hr (P = 0.01) and 8 hr (P < 0.01), but not at 2 hr (P = 1.00) and 4 hr (P = 0.16). With the ccrp assay, a significant increase in CRP was detected at 8 hr (P < 0.01) but not at 2 hr (P = 1.00), 4 hr (P = 1.00), or 6 hr (P = 0.74). High-sensitivity CRP in lean and overweight dogs. Fifteen dogs were included in the study; 6 dogs had a BCS of 5 and were allocated to the lean group, and 9 dogs had a BCS of 6 7 (median 6) and were allocated to the overweight group. Mean weight in the lean group was 34 kg, and in the overweight group 38 kg. The median (range) age was 5 (3 6) years and 5 (5 6) years in the lean and overweight group, respectively. The median (range) CRP concentration in the lean group was 0.7 (< ) mg/l, compared with 1.1 ( ) mg/l in the overweight group (P = 0.06). Discussion In this study, an automated, canine-specific hscrp assay was validated and applied on samples from clinical research studies. The validation indicated that the hscrp test can be used to accurately measure low CRP concentrations in canine serum samples. The imprecision of the hscrp assay met the requirement for optimal performance, expressed as a CV value of <6.1%, 13,22 and was lower than previously reported for canine-specific hscrp methods. 21,31 The focus in the current study was on within-run imprecision, because the recommendation for using such a test in research is to analyze all samples from 1 study in a single run, to avoid interassay variation. The hscrp assay was acceptably linear under dilution for the intended working range 30 mg/l, with the

6 Validation of a canine-specific hscrp assay 187 Figure 2. Prozone effect with the automated canine-specific high-sensitivity C-reactive protein (hscrp) assay. The absorbance with the hscrp assay was plotted against the CRP concentration as determined by the canine-specific (c)crp assay. The horizontal line illustrates the absorbance of the top calibrator of the hscrp assay with a concentration of 30 mg/l. An analytical relevant prozone effect was present, because in the 2 highest samples the absorbances were lower than the absorbance of the upper calibrator of the hscrp assay. Table 2. Interference from hemolysis and lipemia for an automated canine-specific high-sensitivity C-reactive protein (CRP) assay.* Interferent CRP control pool (mg/l) CRP test pool (mg/l) d obs (mg/l) Acceptance limits of interference (mg/l) Hemoglobin (5 g/l) ( 0.18; 0.22) ( 0.31; 0.31) Hemoglobin (10 g/l) (0.06; 0.39) ( 0.73; 0.73) (0.02; 0.42) ( 0.28; 0.28) Triglycerides (10 g/l) ( 0.24; 0.14) ( 0.77; 0.77) ( 0.19; 0.21) ( 0.30; 0.30) Triglycerides (25 g/l) ( 0.17; 0.22) ( 0.16; 0.16) * The CRP concentrations of 6 control pools were compared to the CRP concentration of 6 corresponding test pools containing interfering substance of different concentrations. The difference between the 2 pools (d obs ) was calculated and compared to the acceptance limits, representing an allowable change in CRP concentration of ±10% by interference. Numbers in parentheses are 95% confidence levels. exception of the most diluted sample with the lowest hscrp concentration. This sample did not meet the quality requirement in the linearity study, but yet the sample had an expected CRP concentration above the LOQ of 0.5 mg/l determined in a separate experiment. The different study designs in the linearity and LOQ experiments can explain the discrepancy. In the linearity study, the whole intended measurement range was evaluated by diluting a sample with CRP concentration of approximately 30 mg/l, whereas in the LOQ experiment, a start sample with CRP concentration of approximately 8 mg/l was diluted to expected CRP concentrations of 1 mg/l, 0.5 mg/l, and 0.25 mg/l. The concentration of the start sample represented the true value in this experiment, and a different LOQ may have been obtained if a start sample with another CRP concentration was chosen. This needs to be acknowledged by researchers performing studies where an exact LOQ is of importance. The original ccrp test, on which the hscrp test was based, was not affected by relevant prozone effect. 14 By increasing the sample volume 10 times in the hscrp test, without changing the reagent volume, a prozone effect was expected to occur. The reason for this is that in a sample with high CRP concentration, the amount of CRP protein will be relatively much higher than the amount of CRPbinding antibodies, preventing successful formation of antigen antibody complexes and leading to erroneous results. In our experiment, a prozone effect was illustrated by the fact that samples with CRP concentrations greater than ~150 mg/l were falsely reported to have low CRP concentrations (<30 mg/l) with the hscrp assay. Therefore, it is crucial

7 188 Hillström et al. Figure 3. Bland Altman difference plot for C-reactive protein (CRP) concentrations in canine sera measured with an automated immunoturbidimetric canine-specific high-sensitivity CRP assay (hscrp) and a canine-specific enzyme-linked immunosorbent assay (ELISA; n = 35). Figure 4. Bland Altman difference plot with bias (solid line) and 95% limits of agreement (dashed lines) for log transformed C-reactive protein (CRP) concentrations measured with an automated immunoturbidimetric canine-specific high-sensitivity CRP assay (hscrp) and a canine-specific enzyme-linked immunosorbent assay (ELISA; n = 35). that only samples with low CRP concentrations are analyzed with the hscrp assay. The results from the interference study showed that, in the hscrp assay, there was no statistically significant interference from hemolysis at a hemoglobin concentration of 5 g/l, which corresponds to moderate hemolysis that is easily detected by visual inspection. At higher hemoglobin concentrations (10 g/l), slight interference was present, and therefore such samples should not be analyzed with the hscrp assay. Interference from hemolysis has previously been shown with the canine CRP ELISA used as comparative method in this study, 21 whereas no interfering effect was present with a TR-IFMA. 30 In samples that were made lipemic by adding a commercial fat emulsion to sera, no interfering effect was present at a triglyceride concentration of 10 g/l, and only a mild interfering effect was present at a triglyceride concentration of 25 g/l, which represents severe hypertriglyceridemia. The presence of lipemia in a sample will cause increased absorbance of the blank reading in the instrument compared with the absorbance of a nonlipemic sample. After the blank reading, when sample and reagent containing antibodies are mixed, the absorbance increases even further. In naturally lipemic samples analyzed with the hscrp assay, the absorbance occasionally was so high that it exceeded the optical limit of the instrument and therefore no result was reported (data not shown). This interference did not occur in the samples to which a commercial fat emulsion was added, despite a higher triglyceride concentration in the experimental samples than in the naturally lipemic samples (data not shown). The commercial fat emulsion does not contain triglycerides of the same sizes that are most commonly present in vivo. 29 Therefore, the results from the experiment with added fat solution cannot be directly transferred to naturally lipemic samples. Another approach for testing the effect of lipemia could be to measure CRP in samples before and after ultracentrifugation, 3 and experiments using this method could be performed to further characterize the effect of natural lipemia on hscrp measurements. There is no gold standard method available for measuring canine CRP, and therefore it was not possible to assess true bias in the method comparison study. Instead, a comparison between the new assay and a previously validated ELISA was performed to investigate relative bias. Because the agreement between the methods was poor, the assays should not be used interchangeably. Most often, the hscrp test measured lower CRP concentrations than the comparative test, but there was a marked variation between individual samples, and the cause of this was not determined. Compared with the ELISA, the hscrp test had the important advantage that it was automated, which diminished the risk of manual errors and made it much less labor intensive to operate. Also, a large number of samples could be analyzed in 1 run with the hscrp assay, thereby avoiding the interassay variation caused by the use of multiple ELISA plates. No method comparison study between the ccrp test and the hscrp test was performed, because of the different measurement ranges of the assays. Instead, the methods were compared when applied on clinical research samples from dogs undergoing surgery. The hypothesis, that the hscrp test would be superior for detecting an expected increase in CRP

8 Validation of a canine-specific hscrp assay 189 concentrations at an early stage after OHE, was confirmed. The hscrp assay is therefore the preferred test for monitoring CRP changes during the first postoperative hours in dogs undergoing surgery. In people, obesity has been associated with increased concentrations of CRP, 5,32 whereas studies of obese dogs have reported both increased 11 and decreased 37,39 CRP concentrations. Different study designs and inclusion and exclusion criteria could explain these divergent results. One possible pitfall when investigating differences in CRP concentrations in apparently healthy dogs is that it can be difficult to exclude with certainty the presence of low-grade inflammatory diseases such as inflammatory bowel disease, rhinitis, periodontitis, and osteoarthritis, which may affect CRP concentrations. 16,18,36,41 In addition, factors such as age, sex, or breed could potentially influence CRP concentrations in dogs. In people, CRP concentrations are positively correlated with increasing age, 17,32 and differences between men and women have been reported. 32 No significant age or sex differences in CRP concentrations were found in a previous study of 144 healthy Beagles aged 3 months to 6 years, 23 but breed differences have been reported. 25,38 In the present study, dogs of a single breed and sex were included to avoid these 2 factors as possible confounders. The hypothesis that overweight dogs had increased CRP concentrations compared with lean dogs was not statistically confirmed. The dogs in the present study were only mildly overweight and a different result may have been obtained if grossly obese dogs had also been included. Larger studies on well-characterized dogs are required to establish the association between BCS and CRP concentrations in dogs. To facilitate additional research about conditions involving CRP within low and ultralow serum concentrations, a hscrp assay should ideally be commercially available and easily accessible. These goals are only partially obtained with this validated hscrp assay as neither the calibrators nor controls, as purchased from the manufacturer, are adapted for a hscrp assay. Therefore, if these commercially available reagents are to be used for hscrp testing on a new instrument, or if a new batch is purchased, a validation study should be performed prior to evaluation of subject samples. To date, 4 different batches have been successfully validated for hscrp testing at the authors laboratory (data not shown), illustrating the assay s suitability to the research laboratory setting. Acknowledgments The authors acknowledge Kathrin Sunde (Gentian AS) and Anne Frösegård (University Animal Hospital, SLU) for advice and laboratory assistance. Sources and manufacturers a. Vacutainer, Becton-Dickinson, Stockholm, Sweden. b. SC Micro tube, Sarstedt AG & Co, Nümbrecht, Germany. c. Gentian ccrp, Gentian AS, Moss, Norway. d. Abbott Architect c4000, Abbott Park, IL. e. Randox Laboratories Ltd, Crumlin, United Kingdom. f. LifeDiagnostics, West Chester, PA. g. Intralipid 200 g/l, Fresenius Kabi AB, Uppsala, Sweden. h. Phase EIA canine CRP Assay, Tridelta Development Ltd, Maynooth, Ireland. i. Analyse-it Software Ltd, Leeds, United Kingdom. j. Minitab Inc., State College, PA. Declaration of conflicting interests Anna Hillström s PhD project at the Swedish University of Agricultural Sciences (SLU) was partly financed by Gentian AS, the company manufacturing the assay investigated in this article. The research was performed independently, with collaborative parties outside the university having no influence over study design, data acquisition, analyses, results, manuscript preparation, or scientific publication. Funding The study was financed by Eurostars project Mammalian CRP through Vinnova project References 1. Allin KH, Nordestgaard BG. Elevated C-reactive protein in the diagnosis, prognosis, and cause of cancer. Crit Rev Clin Lab Sci 2011;48: Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1: Calmarza P, Cordero J. Lipemia interferences in routine clinical biochemical tests. Biochem Med (Zagreb) 2011;21: Ceron JJ, et al. Acute phase proteins in dogs and cats: current knowledge and future perspectives. Vet Clin Pathol 2005;34: Choi J, et al. Obesity and C-reactive protein in various populations: a systematic review and meta-analysis. Obes Rev 2013;14: Clinical and Laboratory Standards Institute (CLSI). Interference testing in clinical chemistry; approved guideline. 2nd ed. Wayne, PA: CLSI, CLSI Document No. EP07-A2. 7. Cunningham SM, et al. Systemic inflammation and endothelial dysfunction in dogs with congestive heart failure. J Vet Intern Med 2012;26: Dali-Youcef N, et al. Metabolic inflammation: connecting obesity and insulin resistance. Ann Med 2013;45: Eckersall PD, Bell R. Acute phase proteins: biomarkers of infection and inflammation in veterinary medicine. Vet J 2010;185: Eckersall PD, et al. An immunoturbidimetric assay for canine C-reactive protein. Vet Res Commun 1991;15: Eirmann LA, et al. Comparison of adipokine concentrations and markers of inflammation in obese versus lean dogs. Int J Appl Res Vet Med 2009;7: Gebhardt C, et al. Use of C-reactive protein to predict outcome in dogs with systemic inflammatory response syndrome or sepsis. J Vet Emerg Crit Care (San Antonio) 2009;19: Harr KE, et al. ASVCP guidelines: allowable total error guidelines for biochemistry. Vet Clin Pathol 2013;42:

9 190 Hillström et al. 14. Hillström A, et al. Validation of a commercially available automated canine-specific immunoturbidimetric method for measuring canine C-reactive protein. Vet Clin Pathol 2014;43: Hingorani AD, et al. C-reactive protein and coronary heart disease: predictive test or therapeutic target? Clin Chem 2009;55: Hurter K, et al. Measurements of C-reactive protein in serum and lactate dehydrogenase in serum and synovial fluid of patients with osteoarthritis. Vet J 2005;169: Imhof A, et al. Distributions of C-reactive protein measured by high-sensitivity assays in apparently healthy men and women from different populations in Europe. Clin Chem 2003;49: Jergens AE, et al. A scoring index for disease activity in canine inflammatory bowel disease. J Vet Intern Med 2003;17: Kenny D, et al. Consensus agreement: conference on strategies to set global quality specifications in laboratory medicine. Scand J Clin Lab Invest 1999;59: Kjelgaard-Hansen M, et al. Evaluation of a commercially available human C-reactive protein (CRP) turbidometric immunoassay for determination of canine serum CRP concentration. Vet Clin Pathol 2003;32: Kjelgaard-Hansen M, et al. Evaluation of a commercially available enzyme-linked immunosorbent assay (ELISA) for the determination of C-reactive protein in canine serum. J Vet Med A Physiol Pathol Clin Med 2003;50: Kjelgaard-Hansen M, et al. Study on biological variability of five acute-phase reactants in dogs. Comp Clin Pathol 2003;12: Kuribayashi T, et al. Determination of serum C-reactive protein (CRP) in healthy beagle dogs of various ages and pregnant beagle dogs. Exp Anim 2003;52: Laflamme D. Development and validation of a body condition score system for dogs. Canine Pract 1997;22: Ljungvall I, et al. Cardiac troponin I is associated with severity of myxomatous mitral valve disease, age, and C-reactive protein in dogs. J Vet Intern Med 2010;24: Nakamura M, et al. C-reactive protein concentration in dogs with various diseases. J Vet Med Sci 2008;70: National Committee for Clinical Laboratory Standards (NCCLS). Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; approved guideline. Wayne, PA: NCCLS, NCCLS Document No. EP6-A. 28. National Committee for Clinical Laboratory Standards (NCCLS). Protocols for Determination of Limits of Detection and Limits of Quantification; approved guideline. Wayne, PA: NCCLS, NCCLS Document No. EP17-A. 29. Nikolac N. Lipemia: causes, interference mechanisms, detection and management. Biochem Med (Zagreb) 2014;24: Parra MD, Cerón JJ. Effects of haemolysis, lipaemia and bilirubinaemia in canine C-reactive protein and haptoglobin determination by time-resolved fluorometry: short communication. Acta Vet Hung 2007;55: Parra MD, et al. Analytical and clinical validation of a timeresolved immunofluorometric assay (TR-IFMA) for canine C-reactive protein in serum. Vet Res Commun 2006;30: Poledne R, et al. Proinflammatory status, genetics and atherosclerosis. Physiol Res 2009;58(Suppl 2):S Ridker PM. Inflammatory biomarkers and risks of myocardial infarction, stroke, diabetes, and total mortality: implications for longevity. Nutr Rev 2007;65:S Ridker PM, et al. Reduction in C-reactive protein and LDL cholesterol and cardiovascular event rates after initiation of rosuvastatin: a prospective study of the JUPITER trial. Lancet 2009;373: Ridker PM, et al. Should C-reactive protein be added to metabolic syndrome and to assessment of global cardiovascular risk? Circulation 2004;109: Sheahan D, et al. Acute phase protein concentrations in dogs with nasal disease. Vet Rec 2010;167: Veiga AP, et al. Association of canine obesity with reduced serum levels of C-reactive protein. J Vet Diagn Invest 2008;20: Wong VM, et al. Serum C-reactive protein concentrations in healthy Miniature Schnauzer dogs. Vet Clin Pathol 2011;40: Yamka RM, et al. Identification of canine markers related to obesity and the effects of weight loss on the markers of interest. Int J Appl Res Vet Med 2006;4: Yousuf O, et al. High-sensitivity C-reactive protein and cardiovascular disease: a resolute belief or an elusive link? J Am Coll Cardiol 2013;62: Yu G, et al. Effect of periodontitis on susceptibility to atrial fibrillation in an animal model. J Electrocardiol 2010;43:

Validation of a commercially available automated canine-specific immunoturbidimetric method for measuring canine C-reactive protein

Validation of a commercially available automated canine-specific immunoturbidimetric method for measuring canine C-reactive protein Veterinary Clinical Pathology ISSN 0275-6382 ORIGINAL RESEARCH Validation of a commercially available automated canine-specific immunoturbidimetric method for measuring canine C-reactive protein Anna Hillstr

More information

Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000

Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000 v2-jan214 Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4 Intended Use The canine CRP immunoassay on Abbott's Architect c4 is an in vitro diagnostic test for quantitative determination

More information

Gentian Canine CRP Immunoassay Application Note for scil VitroVet*

Gentian Canine CRP Immunoassay Application Note for scil VitroVet* v02-jan2014 Gentian Canine CRP Immunoassay Application Note for scil VitroVet* Intended Use The canine CRP immunoassay on scil VitroVet is an in vitro diagnostic test for quantitative determination of

More information

Evaluation of a species-specific C-reactive protein assay for the dog on the ABX Pentra 400 clinical chemistry analyzer

Evaluation of a species-specific C-reactive protein assay for the dog on the ABX Pentra 400 clinical chemistry analyzer Hindenberg et al. BMC Veterinary Research (2017) 13:146 DOI 10.1186/s12917-017-1065-9 RESEARCH ARTICLE Open Access Evaluation of a species-specific C-reactive protein assay for the dog on the ABX Pentra

More information

Canine CRP Detection Kit

Canine CRP Detection Kit Canine CRP Detection Kit Catalog # 6027 For Research Use Only - Not Human or Therapeutic Use Acute phase proteins (APPs) are blood proteins released from hepatocytes as an integral part of acute phase

More information

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats Authors: Kate Cote, Ph.D., Graham Bilbrough, MA, VetMB, CertVA, MRCVS,

More information

Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100)

Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100) Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on CUBE analyser (CA0100) Location Location: Eurolyser Diagnostica GmbH Operators: Simone Wieser; Franz Helminger; Michael Gruber Date: July-November

More information

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY INFORMATION BULLETIN USING THE ACCESS AMH ASSAY IN YOUR LABORATORY ///////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////

More information

Determination of hemoglobin is one of the most commonly

Determination of hemoglobin is one of the most commonly ORIGINAL ARTICLE Multiple-Site Analytic Evaluation of a New Portable Analyzer, HemoCue Hb 201+, for Point-of-Care Testing Sten-Erik Bäck, PhD,* Carl G. M. Magnusson, PhD, Lena K. Norlund, MD, PhD, Henning

More information

Chapter 2. High-sensitivity C-reactive protein methods examined

Chapter 2. High-sensitivity C-reactive protein methods examined Chapter High-sensitivity C-reactive protein methods examined Snježana Rothkrantz-Kos, Maria PJ Schmitz, Otto Bekers, Paul PCA Menheere, Marja P van Dieijen-Visser Clin Chem ;48:359-6 3 Chapter Abstract

More information

GLUH PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE

GLUH PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE SYNCHRON System(s) Chemistry Information Sheet Copyright 2014 Beckman Coulter, Inc. GLUH Glucose REF B24985 For In Vitro Diagnostic Use Rx Only ANNUAL REVIEW Reviewed by Date Reviewed by Date PRINCIPLE

More information

Evaluation of the EUROLYSER-SOLO Total T4 assay. for veterinary samples

Evaluation of the EUROLYSER-SOLO Total T4 assay. for veterinary samples Evaluation of the EUROLYSER-SOLO Total T4 assay for veterinary samples Dr. rer.-nat. Volker Schlüter Zacherlweg 18 D-82061 Neuried Date: 2011 April 27 th Scope of the study Goal of this study was the validation

More information

ACCESS hstni SCIENTIFIC LITERATURE

ACCESS hstni SCIENTIFIC LITERATURE ACCESS hstni SCIENTIFIC LITERATURE 2017 2018 Table of contents Performance Evaluation of Access hstni A critical evaluation of the Beckman Coulter Access hstni: Analytical performance, reference interval

More information

G. Snodgrass, K. Ackles, A. Blanco and A. Versaggi Ortho Clinical Diagnostics, Rochester, NY 14626

G. Snodgrass, K. Ackles, A. Blanco and A. Versaggi Ortho Clinical Diagnostics, Rochester, NY 14626 Performance of the EKF Diagnostics, Stanbio β-hydroxybutyrate LiquiColor Assay on the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System. G. Snodgrass, K. Ackles, A. Blanco and A. Versaggi

More information

SCIEX Vitamin D 200M Assay for the Topaz System

SCIEX Vitamin D 200M Assay for the Topaz System The First FDA-Cleared LC-MS/MS Assay for Vitamin D SCIEX Vitamin D 200M Assay for the Topaz System The first FDA-cleared LC-MS/MS assay for Vitamin D Vitamin D is an important building block for human

More information

Insulin (Porcine/Canine) ELISA

Insulin (Porcine/Canine) ELISA Insulin (Porcine/Canine) ELISA For the quantitative measurement of insulin in Porcine/Canine serum and plasma (EDTA) For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 80-INSPO-E01

More information

BÜHLMANN fcal turbo. Calprotectin turbidimetric assay for professional use. Reagent Kit B-KCAL-RSET. Revision date:

BÜHLMANN fcal turbo. Calprotectin turbidimetric assay for professional use. Reagent Kit B-KCAL-RSET. Revision date: BÜHLMANN fcal turbo Calprotectin turbidimetric assay for professional use Reagent Kit B-KCAL-RSET Revision date: 2017-02-24 BÜHLMANN LABORATORIES AG Baselstrasse 55 4124 Schönenbuch, Switzerland Tel.:

More information

Human Troponin I ELISA Kit

Human Troponin I ELISA Kit Human Troponin I ELISA Kit Cat. No.:DEIA2358 Pkg.Size:96T Intended use This kit is intended for the quantitative determination of cardiac troponin I in human serum. Measurement of troponin I values are

More information

DKK-1 ELISA, Cat.No. BI For the quantitative determination of DKK-1 in human serum

DKK-1 ELISA, Cat.No. BI For the quantitative determination of DKK-1 in human serum DKK-1 ELISA, Cat.No. BI-20413 For the quantitative determination of DKK-1 in human serum ASSAY CHARACTERISTICS Method Standard range Conversion factor Sample volume / well Incubation time, temp. Sensitivity

More information

liquicolor (AMP Buffer, IFCC) Design Verification

liquicolor (AMP Buffer, IFCC) Design Verification Design Verification (AMP Buffer, IFCC) liquicolor 1 Introduction... 2 2 Imprecision... 2 3 arity and Detection Limit... 2 3.1 arity... 2 3.2 Detection Limit... 3 4 Comparison of Methods... 3 5 Stability...

More information

PERIOSTIN ELISA CONTENTS

PERIOSTIN ELISA CONTENTS PERIOSTIN ELISA for the quantitative determination of Periostin in human serum, EDTA plasma, heparin plasma, and citrate plasma Cat. No. BI-20433. 12 x 8 tests CONTENTS ASSAY CHARACTERISTICS Summary...

More information

Measurement of total thyroxine concentration in serum from dogs and cats by use of various methods Objective Sample Population Procedure Results

Measurement of total thyroxine concentration in serum from dogs and cats by use of various methods Objective Sample Population Procedure Results Measurement of total thyroxine concentration in serum from dogs and cats by use of various methods Robert J. Kemppainen, DVM, PhD, and Jeremy R. Birchfield, DVM Objective To compare results obtained from

More information

liquicolor Design Verification

liquicolor Design Verification Design Verification liquicolor 1 Introduction... 2 2 Imprecision... 2 2.1 Imprecision for... 2 2.2 Imprecision for... 2 3 Linearity and Detection Limit... 2 3.1 Linearity... 2 3.2 Detection Limit... 4

More information

The Adiponectin Turbidimetric Immunoassay Reagent Kit

The Adiponectin Turbidimetric Immunoassay Reagent Kit The Adiponectin Turbidimetric Immunoassay Reagent Kit Catalogue number: 51010 For the quantitative determination of Adiponectin in human serum and plasma This package insert must be read in its entirety

More information

METHOD VALIDATION: WHY, HOW AND WHEN?

METHOD VALIDATION: WHY, HOW AND WHEN? METHOD VALIDATION: WHY, HOW AND WHEN? Linear-Data Plotter a,b s y/x,r Regression Statistics mean SD,CV Distribution Statistics Comparison Plot Histogram Plot SD Calculator Bias SD Diff t Paired t-test

More information

HbA1c (Human) ELISA Kit

HbA1c (Human) ELISA Kit HbA1c (Human) ELISA Kit Cat. No.:DEIA3509 Pkg.Size:96T Intended use GHbA1c (Human) ELISA Kit is a sandwich enzyme immunoassay for the quantitative measurement of human GHbA1c. General Description vhemoglobin,

More information

TNF-alpha ELISA. For Research Use Only. Not For Use In Diagnostic Procedures.

TNF-alpha ELISA. For Research Use Only. Not For Use In Diagnostic Procedures. TNF-alpha ELISA For the quantitative determination of TNF-alpha in serum, plasma, buffered solution or cell culture medium. For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number:

More information

Porcine/Canine Insulin ELISA

Porcine/Canine Insulin ELISA Porcine/Canine Insulin ELISA For the quantitative determination of insulin in porcine or canine serum and plasma. Please read carefully due to Critical Changes, e.g., Calculation of Results. For Research

More information

Eileen Crimmins 1 Jianwei Hu 2 Peifeng Hu 3 Wei Huang 2 Jungki Kim 1 Yuhui Shi 2 John Strauss 1 Lu Zhang 1 Xiaohui Zhao 2 Yaohui Zhao 2

Eileen Crimmins 1 Jianwei Hu 2 Peifeng Hu 3 Wei Huang 2 Jungki Kim 1 Yuhui Shi 2 John Strauss 1 Lu Zhang 1 Xiaohui Zhao 2 Yaohui Zhao 2 CHARLS Pilot: Blood-based Biomarker Documentation Eileen Crimmins 1 Jianwei Hu 2 Peifeng Hu 3 Wei Huang 2 Jungki Kim 1 Yuhui Shi 2 John Strauss 1 Lu Zhang 1 Xiaohui Zhao 2 Yaohui Zhao 2 June 2011 1 University

More information

EliKine Free Thyroxine (ft4) ELISA Kit

EliKine Free Thyroxine (ft4) ELISA Kit EliKine Free Thyroxine (ft4) ELISA Kit Booklet Item NO. KET0005 Product Name EliKine Free Thyroxine (ft4) ELISA Kit ATTENTION For laboratory research use only. Not for clinical or diagnostic use TABLE

More information

Total Thyroxine ELISA (T4)

Total Thyroxine ELISA (T4) Design Verification Total Thyroxine ELISA (T4) Contents 1 Assay Principle... 2 2 Imprecision... 2 Within-run Imprecision... 2 Between run Imprecision... 2 3 Comparison of Methods, Accuracy... 2 4 Linearity...

More information

DIHYDROSTREPTOMYCIN and STREPTOMYCIN First draft prepared by Lynn G. Friedlander, Rockville, MD, USA Rainer W. Stephany, Bilthoven, The Netherlands

DIHYDROSTREPTOMYCIN and STREPTOMYCIN First draft prepared by Lynn G. Friedlander, Rockville, MD, USA Rainer W. Stephany, Bilthoven, The Netherlands DIHYDROSTREPTOMYCIN and STREPTOMYCIN First draft prepared by Lynn G. Friedlander, Rockville, MD, USA Rainer W. Stephany, Bilthoven, The Netherlands ADDENDUM To the monograph and addenda prepared by the

More information

IDEXX Catalyst SDMA Test for in-house measurement of SDMA concentration in serum from dogs and cats

IDEXX Catalyst SDMA Test for in-house measurement of SDMA concentration in serum from dogs and cats IDEXX Catalyst SDMA Test for in-house measurement of SDMA in serum from dogs and cats By Graham Bilbrough, Barbara Evert, Karen Hathaway, Gina Panagakos, Jane Robertson, and Maha Yerramilli Introduction

More information

FinTest IgG4 Screen 20 ELISA KIT

FinTest IgG4 Screen 20 ELISA KIT FinTest IgG4 Screen 20 ELISA KIT Cat. No.:DEIA6196 Pkg.Size:96T Intended use Enzyme immunoassay (microtiter strips) for the detection and the quantitative determination of IgG4 antibodies against 20 Food

More information

Bovine Insulin ELISA

Bovine Insulin ELISA Bovine Insulin ELISA For quantitative determination of insulin in bovine serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 80-INSBO-E01 Size: 96 wells Version:

More information

Human LDL ELISA Kit. Innovative Research, Inc.

Human LDL ELISA Kit. Innovative Research, Inc. Human LDL ELISA Kit Catalog No: IRKTAH2582 Lot No: SAMPLE INTRODUCTION Human low-density lipoprotein (LDL) transports cholesterol from the liver to tissues where it is incorporated into cell membranes.

More information

IFLS C-Reactive Protein Data User Guide

IFLS C-Reactive Protein Data User Guide WORKING PAPER IFLS C-Reactive Protein Data User Guide Perry Hu, Elizabeth Henny Herningtyas, John Strauss, Eileen Crimmins, Jung Ki Kim, and Bondan Sikoki RAND Labor & Population WR-675/7 October 2013

More information

Rat cholesterol ELISA Kit

Rat cholesterol ELISA Kit Rat cholesterol ELISA Kit Catalog No. CSB-E11706r (96T) This immunoassay kit allows for the in vitro quantitative determination of rat Cholesterol concentrations in serum, plasma and other biological fluids.

More information

LIPASE liquicolor. Design Verification. Multipurpose Reagent

LIPASE liquicolor. Design Verification. Multipurpose Reagent Design Verification LIPASE liquicolor Multipurpose Reagent CONTENTS 1 Introduction... 2 2 Imprecision... 2 3 Linearity and Detection Limit... 2 3.1 Linearity... 2 3.2 Detection Limit... 3 4 Recovery of

More information

Performance of three portable blood glucose monitoring devices used in a veterinary application

Performance of three portable blood glucose monitoring devices used in a veterinary application Performance of three portable blood glucose monitoring devices used in a veterinary application Prepared by Michael I. Lindinger, PhD, Vice-President of Research, The Nutraceutical Alliance, Campbellville,

More information

CoQ10(Coenzyme Q10) ELISA Kit

CoQ10(Coenzyme Q10) ELISA Kit CoQ10(Coenzyme Q10) ELISA Kit Catalogue No.: EU0196 Size: 48T/96T Reactivity: Universal Detection Range: 0.781-50ng/ml Sensitivity:

More information

Triiodothyronine (T3) ELISA

Triiodothyronine (T3) ELISA For Research Use Only. Not for use in Diagnostic Procedures. INTENDED USE The GenWay, Inc. Triiodothyronine (T3) ELISA Kit is intended for the detection of total T3 in human serum or plasma. For research

More information

Annex 1 Instruction for Use

Annex 1 Instruction for Use Annex 1 Instruction for Use THYROID STIMULATING HORMONE (TSH) ELISA KIT Catalog No. E1003 INTENDED USE The Autobio TSH assay is designed for the quantitative determination of thyroid stimulating hormone

More information

Original Article Interfering effect of bilirubin on the determination of alkaline phosphatase

Original Article Interfering effect of bilirubin on the determination of alkaline phosphatase Int J Clin Exp Med 2014;7(11):4244-4248 www.ijcem.com /ISSN:1940-5901/IJCEM0002083 Original Article Interfering effect of bilirubin on the determination of alkaline phosphatase Zhong Wang, Hongyan Guo,

More information

1 Introduction Imprecision Within-run imprecision, results Day-to-day imprecision, results... 2

1 Introduction Imprecision Within-run imprecision, results Day-to-day imprecision, results... 2 Design Verification liquicolor 1 Introduction... 2 2 Imprecision... 2 2.1 Within-run imprecision, results... 2 2.2 Day-to-day imprecision, results... 2 3 Linearity and Detection Limit... 2 3.1 Linearity...

More information

GSI Canine IL-5 ELISA Kit-2 Plates DataSheet

GSI Canine IL-5 ELISA Kit-2 Plates DataSheet Interleukin5 (IL5) is a secreted glycoprotein that belongs to the α-helical group of cytokines (1, 2, 3). Unlike other family members, it is present as a covalently linked antiparallel dimer (4, 5). IL-5

More information

Influenza A IgG ELISA

Influenza A IgG ELISA Influenza A IgG ELISA For the qualitative determination of IgG-class antibodies against Influenza A virus in human serum or plasma (citrate, heparin). For Research Use Only. Not For Use In Diagnostic Procedures.

More information

Apolipoprotein A-1 ELISA

Apolipoprotein A-1 ELISA Apolipoprotein A-1 ELISA For the quantitative determination of apolipoprotein A1 in serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures. Please read carefully due to Critical

More information

Article in press - uncorrected proof

Article in press - uncorrected proof Clin Chem Lab Med 2006;44(6):768 773 2006 by Walter de Gruyter Berlin New York. DOI 10.1515/CCLM.2006.125 2006/60 Integration between the Tele-Cardiology Unit and the central laboratory: methodological

More information

LDL (Human) ELISA Kit

LDL (Human) ELISA Kit LDL (Human) ELISA Kit Cat. No.:DEIA3864 Pkg.Size:96T Intended use This immunoassay kit allows for the specific measurement of human low density lipoprotein, LDL concentrations in cell culture supernates,

More information

Mouse C3 (Complement Factor 3) ELISA Kit

Mouse C3 (Complement Factor 3) ELISA Kit Mouse C3 (Complement Factor 3) ELISA Kit Cat. No.:DEIA8289 Pkg.Size:96T Intended use The Mouse C3 (Complement Factor 3) ELISA Kit is a highly sensitive two-site enzyme linked immunoassay (ELISA) for measuring

More information

Design Verification. Form: /5 Rev. 007 valid of

Design Verification. Form: /5 Rev. 007 valid of Design Verification g 1 Determination of Diagnostic Sensitivity and Diagnostic Specificity... 2 Crossreactivity... 3 Potentially interfering substances... 3 2 Imprecision... 4 3 Standardisation... 4 4

More information

Abraxis Progesterone (bovine) ELISA Kit

Abraxis Progesterone (bovine) ELISA Kit Abraxis Progesterone (bovine) ELISA Kit Enzyme immunoassay for the quantitative determination of progesterone in bovine milk/serum/plasma samples PN5081M 96 Tests For Research Use Only. Not for use in

More information

Canine Thyroid Stimulating Hormone, TSH ELISA Kit

Canine Thyroid Stimulating Hormone, TSH ELISA Kit Canine Thyroid Stimulating Hormone, TSH ELISA Kit Catalog No: E0463c 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH

More information

Human Thyroid Stimulating Hormone CLIA kit

Human Thyroid Stimulating Hormone CLIA kit Human Thyroid Stimulating Hormone CLIA kit Cat. No.:DEEL0223 Pkg.Size:96 tests Intended use For the direct quantitative determination of Thyroid Stimulating Hormone in human serum by chemiluminescence

More information

Serum Amyloid A ELISA

Serum Amyloid A ELISA Serum Amyloid A ELISA For the quantitative determination of serum amyloid A (SAA) in serum plasma For Research use Only. Not for Use in Diagnostic Procedures Please see Appendix A for Reference Serum information

More information

HumaTex CRP. Design Verification. Contents

HumaTex CRP. Design Verification. Contents Design Verification HumaTex CRP Contents 1 Function... 2 2 Reproducibility... 2 3 Sensitivity and dynamic range... 2 Preparation of serum control panel... 2 Sensitivity test results... 2 Prozone check...

More information

Product Information. Instruction for Use. Troponin I ELISA kit Catalog Number: EA Storage Temperature: 2 8 C

Product Information. Instruction for Use. Troponin I ELISA kit Catalog Number: EA Storage Temperature: 2 8 C 9620 Medical Center Dr., Suite 200, Rockville, MD 20850 Phone: 1.888.267.4436 Fax: 301-340-9254 Email: techsupport@origene.com Web: www.origene.com Product Information Troponin I ELISA kit Catalog Number:

More information

BNP Fragment EIA (Cat.No. BI-20852W) For the Determination of BNP Fragment in Human Samples

BNP Fragment EIA (Cat.No. BI-20852W) For the Determination of BNP Fragment in Human Samples BNP Fragment EIA (Cat.No. BI-20852W) For the Determination of BNP Fragment in Human Samples ASSAY CHARACTERISTICS Method Competitive Enzyme Immunoassay, HRP/TMB. Microtiter plates are coated with a polyclonal

More information

APOB (Human) ELISA Kit

APOB (Human) ELISA Kit APOB (Human) ELISA Kit Catalog Number KA4330 96 assays Version: 01 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle of the Assay...

More information

Human Neurology 3-Plex A

Human Neurology 3-Plex A Human Neurology 3-Plex A SUMMARY AND EXPLANATION OF THE TEST The Human N3PA assay is a digital immunoassay for the quantitative determination of total Tau, Aβ42, and Aβ40 in human plasma and CSF. Determination

More information

Insulin ELISA. For the quantitative determination of insulin in serum and plasma.

Insulin ELISA. For the quantitative determination of insulin in serum and plasma. Insulin ELISA For the quantitative determination of insulin in serum and plasma. For In Vitro Diagnostic use within the United States of America. This product is for Research Use Only outside of the United

More information

Prothrombin (Human) ELISA Kit

Prothrombin (Human) ELISA Kit Prothrombin (Human) ELISA Kit Catalog Number KA0496 96 assays Version: 04 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Background... 3 Principle of the Assay... 3 General

More information

VetScan UA Urine Analyzer Clinical Performance Larry Lem, PhD & Andrew Rosenfeld, DVM, ABVP

VetScan UA Urine Analyzer Clinical Performance Larry Lem, PhD & Andrew Rosenfeld, DVM, ABVP VetScan UA Urine Analyzer Clinical Performance Larry Lem, PhD & Andrew Rosenfeld, DVM, ABVP Abstract Background: The VetScan UA urine strip reader is an analyzer designed to provide semi-quantitative and

More information

human Total Cathepsin B Catalog Number: DY2176

human Total Cathepsin B Catalog Number: DY2176 human Total Cathepsin B Catalog Number: DY2176 This DuoSet ELISA Development kit contains the basic components required for the development of sandwich ELISAs to measure natural and recombinant human Total

More information

MANUAL RBP4 (human) Competitive ELISA Kit Cat. No. AG-45A-0010YEK-KI01 For research use only. Not for diagnostic use.

MANUAL RBP4 (human) Competitive ELISA Kit Cat. No. AG-45A-0010YEK-KI01 For research use only. Not for diagnostic use. MANUAL RBP4 (human) Competitive ELISA Kit For research use only. Not for diagnostic use. Version 3 (04-May-2015) Cat. No. AG-45A-0010YEK-KI01 www.adipogen.com Table of Contents 1. Intended Use 3 2. Introduction

More information

Hyaluronan Quantification Kit

Hyaluronan Quantification Kit Product manual (ELISA-like assay for HA / Hyaluronic Acid) Updated on Jan. 6th, 207 - Background Hyaluronan (HA) is an unbranched glycosaminoglycan composed of repeating disaccharide units of D-glucronic

More information

Cortisol (Sheep) ELISA Kit

Cortisol (Sheep) ELISA Kit Cortisol (Sheep) ELISA Kit Catalog Number KA0919 96 assays Version: 03 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle of the

More information

Multiple studies are available concerning the use of

Multiple studies are available concerning the use of SPECIAL ARTICLE Evaluation of the Biosite Ò Quantitative Whole Blood D-dimer Assay and Comparison With the biomérieux VIDAS Ò D-dimer Exclusion Test Validation and Utility For Use in the Central Laboratory

More information

Adiponectin, TG/HDL-cholesterol index and hs-crp. Predictors of insulin resistance.

Adiponectin, TG/HDL-cholesterol index and hs-crp. Predictors of insulin resistance. ORIGINAL ARTICLE Adiponectin, TG/HDL-cholesterol index and hs-crp. Predictors of insulin resistance. Bonneau GA y Pedrozo WR 1 Ministry of Public Health, Province of Misiones, 2 School of Exact, Chemical

More information

Rubella virus IgG ELISA Kit

Rubella virus IgG ELISA Kit Rubella virus IgG ELISA Kit Catalog Number KA0223 96 assays Version: 03 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle of

More information

Validation Report for the Neogen Fentanyl Kit for ELISA Screening of Whole Blood and Urine Specimens

Validation Report for the Neogen Fentanyl Kit for ELISA Screening of Whole Blood and Urine Specimens Validation Report for the Neogen Fentanyl Kit for ELISA Screening of Whole Blood and Urine Specimens This document describes the validation of a Neogen Fentanyl kit for the semi-quantitative analysis of

More information

Mercodia Porcine C-peptide ELISA

Mercodia Porcine C-peptide ELISA Mercodia Porcine C-peptide ELISA Directions for Use 10-1256-01 REAGENTS FOR 96 DETERMINATIONS Manufactured by Mercodia AB, Sylveniusgatan 8A, SE-754 50 Uppsala, Sweden EXPLANATION OF SYMBOLS USED ON LABELS

More information

CRP (Human) ELISA Kit

CRP (Human) ELISA Kit CRP (Human) ELISA Kit Catalog Number KA0238 96 assays Version: 03 Intended for research use only www.abnova.com Table of Contents Introduction... 3 Intended Use... 3 Background... 3 Principle of the Assay...

More information

Collection of Blood Specimens by Venipuncture for Plasma-Based Coagulation Assays Necessity of a Discard Tube

Collection of Blood Specimens by Venipuncture for Plasma-Based Coagulation Assays Necessity of a Discard Tube Coagulation and Transfusion Medicine / Discard Tube Unnecessary for Coagulation Assays Collection of Blood Specimens by Venipuncture for Plasma-Based Coagulation Assays Necessity of a Discard Tube Maarten

More information

Mercodia Glucagon ELISA

Mercodia Glucagon ELISA Mercodia Glucagon ELISA Directions for Use 10-1271-01 REAGENTS FOR 96 DETERMINATIONS For Research Use Only Manufactured by Mercodia AB, Sylveniusgatan 8A, SE-754 50 Uppsala, Sweden EXPLANATION OF SYMBOLS

More information

Mouse Ultrasensitive Insulin ELISA

Mouse Ultrasensitive Insulin ELISA Mouse Ultrasensitive Insulin ELISA For the quantitative determination of insulin in mouse serum and plasma. Please read carefully due to Critical Changes, e.g., Calculation of Results. For Research Use

More information

Mouse/Rat THYROXINE (T4) ELISA Catalog No (96 Tests)

Mouse/Rat THYROXINE (T4) ELISA Catalog No (96 Tests) For Research Use Only. Not for use in Diagnostic Procedures. INTENDED USE The GenWay, Inc. Mouse/Rat Thryroxine Kit is intended for the detection of total T4 in mouse/rat serum or plasma. SUMMARY AND EXPLANATION

More information

Mouse Adiponectin Immunoassay Kit

Mouse Adiponectin Immunoassay Kit Antibody and Immunoassay Services Li Ka Shing Faculty of Medicine, The University of Hong Kong Mouse Adiponectin Immunoassay Kit Catalogue Number: 32010 For the quantitative determination of mouse adiponectin

More information

STELLUX C-peptide Chemiluminescence ELISA

STELLUX C-peptide Chemiluminescence ELISA STELLUX C-peptide Chemiluminescence ELISA For the quantitative determination of Human C-peptide in Human Serum, Heparin Plasma, EDTA plasma, and Tissue Culture Supernatants. For In Vitro Diagnostic use

More information

Canine C-Reactive Protein ELISA

Canine C-Reactive Protein ELISA Canine C-Reactive Protein ELISA Product Data Sheet Cat. No.: RH931CRP01DCR For Research Use Only Page 1 of 12 VERSION 51 020611 46 CONTENTS 1. INTENDED USE 3 2. PRINCIPLE OF ASSAY 3 3. REAGENTS AND MATERIALS

More information

GUIDE TO THE EVALUATION OF COMMUTABILITY OF CONTROL MATERIALS

GUIDE TO THE EVALUATION OF COMMUTABILITY OF CONTROL MATERIALS GUIDE TO THE EVALUATION OF COMMUTABILITY OF CONTROL MATERIALS Ferruccio Ceriotti Servizio di Medicina di Laboratorio, Ospedale San Raffaele, Milano F. Ceriotti, Milano, 27-11-2015 2 Preamble Most of the

More information

Human Alpha 1 microglobulin ELISA Kit

Human Alpha 1 microglobulin ELISA Kit Human Alpha 1 microglobulin ELISA Kit Catalogue No.: EH4144 Size: 48T/96T Reactivity: Human Range:0.625-40ng/ml Sensitivity:

More information

Instructions for use. TSH rat ELISA. Please use only the valid version of the Instructions for Use provided with the kit AR E-8600

Instructions for use. TSH rat ELISA. Please use only the valid version of the Instructions for Use provided with the kit AR E-8600 Instructions for use TSH rat ELISA AR E8600 TSH rat ELISA INTRODUCTION INTENDED USE The TSH rat ELISA is an enzyme immunoassay for the quantitative measurement of TSH in rat serum. For research use only.

More information

The Virtues and Pitfalls of Implementing a New Test

The Virtues and Pitfalls of Implementing a New Test The Virtues and Pitfalls of Implementing a New Test James H. Nichols, Ph.D., DABCC, FACB Professor of Clinical Pathology, Microbiology and Immunology Associate Medical Director for Clinical Operations

More information

Rat Insulin ELISA. For the quantitative determination of insulin in rat serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures.

Rat Insulin ELISA. For the quantitative determination of insulin in rat serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures. Rat Insulin ELISA For the quantitative determination of insulin in rat serum and plasma For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: Size: 80-INSRT-E01, E10 96 wells, 10

More information

METHOD VALIDATION CASE

METHOD VALIDATION CASE METHOD VALIDATION CASE METHOD VALIDATION PROTOCOL CLIA Regulation 493.1253 (2) 1.Accuracy (closeness to true/comparative method) 2.Precision (reproducibility) 3.Reference Interval 4.Reportable range (linearity,

More information

Mouse C-Peptide ELISA Kit

Mouse C-Peptide ELISA Kit Mouse C-Peptide ELISA Kit Cat.No: DEIA4507 Lot. No. (See product label) Size 96T Intended Use The Mouse C-Peptide ELISA kit is for the quantitative determination of c-peptide in mouse serum, plasma, and

More information

MLT Intact Proinsulin Kit

MLT Intact Proinsulin Kit IV2-002 English MLT Intact Proinsulin Kit For in-vitro diagnostic use only MLT (Molecular Light Technology) Chemiluminescence assay for the measurement of Intact Proinsulin in human samples. 480 Invitron

More information

02006B 1 vial 02006B 1 vial Store at -20 C. Lyophilized recombinant IL-2

02006B 1 vial 02006B 1 vial Store at -20 C. Lyophilized recombinant IL-2 For detection and measurement of human interleukin 2 Catalog #02006 Catalog #02007 2 Plates 10 Plates Product Description The Human Interleukin 2 (IL-2) ELISA Kit is designed for the quantitative detection

More information

We noted a problem with the use of the hematology

We noted a problem with the use of the hematology Standard Article J Vet Intern Med 2017;31:1708 1716 Feline Differential Leukocyte Count with ProCyte Dx: Frequency and Severity of a Neutrophil-Lymphocyte Error and How to Avoid It H.W. Tvedten, V. Andersson,

More information

Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University

Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University Alert or critical values represent those assay results that require prompt, rapid clinical attention to avert significant study-participant

More information

University of Bristol - Explore Bristol Research

University of Bristol - Explore Bristol Research Irvine, K., Burt, K., & Papasouliotis, K. (2016). Evaluation of an in-practice wet chemistry analyser using canine and feline serum samples. Journal of Veterinary Diagnostic Investigation, 28(1), 38-45.

More information

Human Thyroid-Peroxidase Antibody, TPO-Ab ELISA Kit

Human Thyroid-Peroxidase Antibody, TPO-Ab ELISA Kit Human Thyroid-Peroxidase Antibody, TPO-Ab ELISA Kit Catalog No: E0442h 96 Tests Operating instruction www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH

More information

The analytical phase

The analytical phase The analytical phase Result interpretation Test request Result Sampling Black box: the lab ANALYTICAL PHASE The CASE Uncle Pete, 67 years old Marked abdominal pain 8 pm, ED Acute abdomen? Assessment (+

More information

Performance Characteristics of Eight Estradiol Immunoassays

Performance Characteristics of Eight Estradiol Immunoassays Clinical Chemistry / EVALUATION OF EIGHT ESTRADIOL IMMUNOASSAYS Performance Characteristics of Eight Estradiol Immunoassays David T. Yang, MD, 1 William E. Owen, MT(ASCP), 2 Carol S. Ramsay, 3 Hui Xie,

More information

Procine sphingomyelin ELISA Kit

Procine sphingomyelin ELISA Kit Procine sphingomyelin ELISA Kit For the quantitative in vitro determination of Procine sphingomyelin concentrations in serum - plasma - celiac fluid - tissue homogenate - body fluid FOR LABORATORY RESEARCH

More information

Table. Analytical characteristics of commercial and research cardiac troponin I and T assays declared by the manufacturer.

Table. Analytical characteristics of commercial and research cardiac troponin I and T assays declared by the manufacturer. Table. Analytical characteristics of commercial and research cardiac troponin I and T assays declared by the manufacturer. Commercially available assays - Company/ platform(s)/ assay LoB a LoD b 99 th

More information

Evaluation of the Abbott CELL-DYN 4000 Hematology Analyzer

Evaluation of the Abbott CELL-DYN 4000 Hematology Analyzer Hematopathology / EVALUATION OF THE ABBOTT CELL-DYN 4 HEMATOLOGY ANALYZER Evaluation of the Abbott CELL-DYN 4 Hematology Analyzer Ernesto Grimaldi, MD, and Francesco Scopacasa, PhD Key Words: Abbott CD

More information

See external label 2 C-8 C Σ=96 tests Cat # 3171Z. Free Estriol. Cat # 3171Z. Enzyme Linked Immunosorbent Assay

See external label 2 C-8 C Σ=96 tests Cat # 3171Z. Free Estriol. Cat # 3171Z. Enzyme Linked Immunosorbent Assay DIAGNOSTIC AUTOMATION, INC. 23961 Craftsman Road, Suite D/E/F, Calabasas, CA 91302 Tel: (818) 591-3030 Fax: (818) 591-8383 onestep@rapidtest.com technicalsupport@rapidtest.com www.rapidtest.com See external

More information