Use of Continuous Glucose Monitoring in the Detection and Prevention of Hypoglycemia

Size: px
Start display at page:

Download "Use of Continuous Glucose Monitoring in the Detection and Prevention of Hypoglycemia"

Transcription

1 Journal of Diabetes Science and Technology Volume 1, Issue 1, January 2007 Diabetes Technology Society SYMPOSIUM Use of Continuous Glucose Monitoring in the Detection and Prevention of Hypoglycemia Howard A., M.D. Introduction Intensive diabetes therapy is complicated by an increased rate of hypoglycemia and the development of hypoglycemia unawareness. This presentation focuses on the potential role of real-time continuous glucose monitoring (CGM) in preventing hypoglycemia, and in overcoming this barrier to the implementation of intensive therapy. Issues to be discussed include the differences between capillary blood and interstitial glucose levels, the impact of lag times on sensor calibration and accuracy, and the importance of patient education in minimizing the risk for hypoglycemia from uncontrolled postprandial bolusing. Hypoglycemia: The barrier to intensive insulin therapy Impact of DCCT/EDIC Findings from the Diabetes Control and Complications Trial (DCCT) showed that patients in the intensively-treated group achieved A1c levels approximately 2 points below those achieved in the conventionally-treated group. The Epidemiology of Diabetes Interventions and Complications (EDIC) study, an epidemiologic follow-up of the DCCT subjects, showed a convergence of A1c levels at 4 years post- DCCT; patients in the conventional group improved their A1c somewhat, whereas glycemic control deteriorated in subjects from the experimental group. There have been no systematic evaluations to tease out the factors underlying this lapse in self-care and deterioration in control; however, published data indicate that this deterioration is not related to lack of healthcare insurance. A study by Thompson and colleagues, which looked at how patients have reacted to the findings reported by the DCCT, revealed that approximately 70% of those who wanted to improve their glycemic control cited fears of hypoglycemia as a significant barrier to intensifying their diabetes management. 1 In essence, there is a trade-off between reducing complications and increasing the risk for hypoglycemia. Hypoglycemia is the clinical risk barrier one faces when trying to initiate more intensive diabetes management. Hypoglycemia Unawareness As shown in Figure 1, adapted from work by Zammitt and Frier, declining glucose levels initiate a number of responses: insulin secretion is inhibited, followed by the release of counterregulatory hormones (glucagon and epinephrine) when glucose levels reach approximately 70 mg/dl; this is the basis for the consensus statement that proposes 70 mg/dl as the diagnostic criteria for hypoglycemia. 2 Continued decline in glucose eventually leads to neurologic dysfunction, widespread electroencephalogram (EEG) changes, cognitive dysfunction and severe neuroglycopenia. Counterregulatory abnormalities such as absent suppression of insulin secretion, blunted or absent glucagon response, and blunted adrenaline response further exacerbate the hypoglycemic condition in Type 1 and advanced Type 2 diabetes. Patients who experience frequent episodes of hypoglycemia often develop these abnormalities and lose their ability to detect hypoglycemia; this is referred to as hypoglycemia unawareness and it perpetrates a vicious cycle of recurrent hypoglycemia. Author Affiliation: Joslin Diabetes Center, Harvard Medical School, Boston, Massachusetts Abbreviations: (CGM) continuous glucose monitoring; (DCCT) Diabetes Control and Complications Trial; (EDIC) Epidemiology of Diabetes Interventions and Complications; (EEG) electroencephalogram; (FDA) Food and Drug Administration Key Words: continuous, diabetes, glucose, hypoglycemia, monitor, prevent Corresponding Author: Howard A., M.D., Joslin Diabetes Center, One Joslin Place, Boston, Massachusetts 02215, Howard.Wolper@joslin.harvard.edu J Diabetes Sci Technol 2007; 1:

2 Figure 1. Hierarchy of Responses to Hypoglycemia There is strong evidence, however, that minimizing hypoglycemic episodes through more meticulous glucose control can actually reverse many of these counterregulatory defects. An early study by Fanelli and colleagues looked at patients with Type 1 diabetes who were being treated with multiple daily insulin injections. 3 Patients were converted to a more intensive regimen and closely followed for three months. The goal was to tighten glucose controls in order to minimize hypoglycemic episodes. Subjects underwent a standard hypoglycemic clamp at baseline, 2 weeks, and 3 months duration to measure threshold levels for hypoglycemic symptoms and neuroendocrine responses. As shown in Figure 2, the glycemic thresholds for symptoms and responses normalized after 3 months of meticulous glucose control and hypoglycemia prevention; even some of the glucagon responses recovered. The researchers concluded that hypoglycemia unawareness in Type 1 diabetes is largely reversible. Potential Utility of CGM to Prevent Hypoglycemia CGM can play a significant role in preventing hypoglycemia. Unlike traditional blood glucose monitoring, which looks at only a few points in time, CGM provides comprehensive data that track glucose levels 24 hours a day. This information allows patients to optimize their glycemic control and thus minimize the frequency and severity of hypoglycemic events. A recent study by Garg and colleagues looked at the accuracy, safety and clinical effectiveness of CGM in 91 subjects with Type 1 diabetes. 4 Researchers found that subjects who wore a real-time CGM sensor spent 21% less time hypoglycemic, 23% less time hyperglycemic and 24% more time within the target range of mg/dl than subjects who used traditional blood glucose monitoring to guide treatment. It is important to note that there are several different sensors both on the market and awaiting approval from the Food and Drug Administration (FDA). These sensors vary in their performance characteristics and accuracy. Figure 3 presents data from a study by Clark and colleagues which compares the accuracy of the Medtronic CGMS device (FDA-approved) and Abbott FreeStyle Navigator (awaiting FDA approval) in terms of hypoglycemia detection

3 Figure 3. Accuracy assessment of CGM versus reference blood glucose. Interstitial Fluid vs Capillary Blood Glucose Lag Time A key issue when comparing CGM technology with traditional blood glucose monitoring is that CGM sensors measure interstitial fluid, not capillary glucose. Glucose diffuses from capillary blood into interstitial fluid, resulting in a physiologic lag between glucose levels. The lag time can range from 5 to 15 minutes, depending on the rate of glucose change. Increasing glucose levels are reflected first in capillary blood. However, detecting decreases in glucose (after insulin injection or exercise) may be confounded by the placement of the CGM sensor. For example, if the CGM sensor is placed next to insulin-sensitive tissue, glucose level may decrease first in the interstitial fluid before it is reflected in capillary blood glucose. Because blood glucose concentration is regarded as the reference gold standard, these differences are an important issue in regard to assessment of the accuracy of CGM devices. Clearly, the physiologic lag adds complexity to the development of a reliable hypoglycemic warning alarm. Interstitial Glucose and Cognitive Performance Accurate hypoglycemia measurements are clinically important because they are a marker for patient risk of cognitive impairment and impending severe neuroglycopenia. For CGM devices, the question is whether and/or how well interstitial glucose correlates with cognitive function. Figure 2. Changes in symptoms and neuroendocrine responses resulting from meticulous glycemic control. Evans and colleagues looked at the time course for the onset of and recovery from acute hypoglycemia in healthy subjects. 6 Using standard glucose clamp technique, plasma glucose was allowed to fall rapidly to approximately

4 mg/dl for a period of 90 minutes; Cognitive function was assessed through a battery of sensitive tests throughout the hypoglycemic and recovery periods. The findings showed that while cognitive function became impaired immediately at the onset of hypoglycemia, counterregulatory hormone responses and symptomatic awareness of hypoglycemia were relatively delayed. During the 20-minute recovery period, cognitive function continued to be abnormal even after resolution of symptomatic awareness. (See Figure 4) Data from a smaller study by Cheyne and colleagues, which also used a hypoglycemic clamp, showed that recovery from hypoglycemia was delayed by an average of 26 minutes as measured by interstitial glucose. 7 A second issue rising from use of CGM technology is the tendency to over-bolus based on the data. Clearly, realtime data generated by CGM facilitates optimal glucose control; however, if patients overcompensate for rising glucose levels using frequent bolusing, they have an increased risk of becoming hypoglycemic. In essence, there is trade-off when using CGM with regard to hypoglycemia. On the one hand, there is an increased risk for hypoglycemia if patients are over-bolusing; on the other hand is the ability of the CGM device to detect hypoglycemic events. Just as some patients over-bolus to control hyperglycemia, we also see patients who reduce or discontinue their basal insulin in order to avoid hyperglycemia. This often occurs because patients do not understand the lag time between capillary and interstitial glucose, thereby assessing their glucose to be much lower than it actually is. As a result, they may end up overtreating the hypoglycemia, which, in turn, will lead to a cycle of significant glycemic variability. Figure 4. Delayed restoration of cognitive performance during recovery from hypoglycemia. To date, there have been no studies that examined the correlation between interstitial glucose and cognitive function in individuals with diabetes. However, an animal study by Nielsen and colleagues clearly demonstrated that measurements in the subcutaneous interstitial fat corresponded very closely with measurements in the central nervous system (CNS). 8 To summarize, interstitial fluid glucose concentration may be a better measure of functional status during periods of low glucose than capillary glucose concentration. Practical considerations with use of CGM An important consideration when using CGM technology is that sensor calibration needs to be performed when glucose is in a steady state, which relates back to the lag time issue discussed earlier. For example, if the patient performs the calibration when glucose is rising at 3 mg/dl/ min and there is a 10-minute lag, then there will be a discrepancy of approximately 30 mg/dl between the capillary glucose measurement and the interstitial glucose measurement. This will shift the curve upwards and make the sensor less accurate. Another issue that must be considered is the trade-off between sensitivity and specificity in terms of alarm thresholds. Some patients may feel that they are getting too many alarms. They respond by turning off their alarms and thus derive no benefit from them. This can be addressed by lowering the threshold when patients start using the device. Other patients who fear hypoglycemia, however, may be more tolerant of false alarms. These patients may benefit from starting off with higher thresholds, which lead to more false alarms but also detect more hypoglycemic episodes. Steps Ahead to Facilitate the Adoption of CGM Patient Education An important component of education will be to teach patients how to minimize their risk for hypoglycemia by avoiding over-bolusing. This will require thorough instruction in insulin pharmacodynamics (versus pharmacokinetics) as well as an understanding of the factors that affect postprandial glucose patterns. Patients must also understand when additional bolusing is indicated. At the same time, patients must learn how to minimize their risk for exaggerated rebound. As discussed earlier, this is often the result of overtreating hypoglycemia because they did not consider the lag time between interstitial and capillary glucose. Patients must be cautioned against inappropriate reduction or discontinuation of their basal 149

5 rates. They must also be taught not to rely on interstitial glucose data alone to assess their response to hypoglycemia. Glucose levels should always be confirmed with a capillary glucose test before initiating treatment. Technology Assessment and Improvement Outcome studies are critically important for obtaining insurance coverage for CGM technology. However, if the goal is to strike a balance between accuracy and prevention of hypoglycemia, we will have to assess each device on its own merit; we cannot generalize findings from any one study or device and apply it to all CGM technology. As far as improvements in CGM technology, a key focus should be on minimizing the risk for over-bolusing. It would be helpful to link the insulin-onboard feature on insulin pumps with the CGM screen to help temper patients taking extra boluses postprandially. Another feature would be to have a hypothetical insulin action profile superimposed on the sensor so that patients can more easily conceptualize how much insulin they have onboard. Incorporating a locking system into the insulin pump may also be helpful as a safety measure to prevent patients who are cognitively impaired (due to hypoglycemia) from inadvertently taking additional insulin. 3. Fanelli CG, Epifano L, Rambotti AM, Pampanelli S, Di Vincenzo A, Modarelli F, Lepore M, Annibale B, Ciofetta M, Bottini P, and al. et: Meticulous prevention of hypoglycemia normalizes the glycemic thresholds and magnitude of most of neuroendocrine responses to, symptoms of, and cognitive function during hypoglycemia in intensively treated patients with short-term IDDM. Diabetes 1993; 42: Garg S, Zisser H, Schwartz S, Bailey T, Kaplan R, Ellis S, Jovanovic L: Improvement in Glycemic Excursions With a Transcutaneous, Real- Time Continuous Glucose Sensor: A randomized controlled trial. Diabetes Care 2006; 29: Clarke WL, Anderson S, Farhy L, Breton M, Gonder-Frederick L, Cox D, Kovatchev B: Evaluating the Clinical Accuracy of Two Continuous Glucose Sensors Using Continuous Glucose Error Grid Analysis. Diabetes Care 2005; 28: Evans ML, Pernet A, Lomas J, Jones J, Amiel SA: Delay in onset of awareness of acute hypoglycemia and of restoration of cognitive performance during recovery. Diabetes Care 2000; 23: Cheyne EH, Cavan DA, Kerr D: Performance of a Continuous Glucose Monitoring System During Controlled Hypoglycaemia in Healthy Volunteers. Diabetes Technol Ther 2002; 4: Nielsen JK, Djurhuus CB, Gravholt CH, Carus AC, Granild-Jensen J, Ørskov H, Christiansen JS: Continuous Glucose Monitoring in Interstitial Subcutaneous Adipose Tissue and Skeletal Muscle Reflects Excursions in Cerebral Cortex. Diabetes 2005; 54: Conclusions Fear of hypoglycemia poses a significant barrier to intensifying diabetes control. Frequent hypoglycemic episodes can impair normal counterregulatory responses and lead to hypoglycemia unawareness. CGM has the potential to help patients avoid hypoglycemia, improve their glycemic control and even reverse hypoglycemia unawareness. However, it is important that patients have reasonable expectations of what CGM technology can actually provide. They must receive adequate instruction to use CGM effectively. Disclosures: Member, Medical Advisory Board, Abbott Diabetes Care Member, Medical Advisory Board, J&J/Lifescan Grant support, Medtronic Diabetes References: 1. Thompson CJ, Cummings JF, Chalmers J, Gould C, Newton RW: How have patients reacted to the implications of the DCCT? Diabetes Care 1996; 19: Zammitt NN, Frier BM:Hypoglycemia in Type 2 diabetes: pathophysiology, frequency, and effects of different treatment modalities. Diabetes Care 2005; 28:

Early Patient and Clinician Experiences with Continuous Glucose Monitoring

Early Patient and Clinician Experiences with Continuous Glucose Monitoring Feature Article / Early Experiences with CGM Early Patient and Clinician Experiences with Continuous Glucose Monitoring David K. Bloomgarden, MD, FACE; Janine Freeman, RD, LD, CDE; and Elizabeth DeRobertis,

More information

Continuous Glucose Monitoring: Changing Diabetes Behavior in Real Time and Retrospectively

Continuous Glucose Monitoring: Changing Diabetes Behavior in Real Time and Retrospectively Journal of Diabetes Science and Technology Volume 2, Issue 3, May 2008 Diabetes Technology Society CONTROVERSIES in Continuous Glucose Monitoring Continuous Glucose Monitoring: Changing Diabetes Behavior

More information

The MiniMed Continuous Glucose Monitoring System. JOHN J. MASTROTOTARO, Ph.D.

The MiniMed Continuous Glucose Monitoring System. JOHN J. MASTROTOTARO, Ph.D. DIABETES TECHNOLOGY & THERAPEUTICS Volume 2, Supplement 1, 2000 Mary Ann Liebert, Inc. The MiniMed Continuous Glucose Monitoring System JOHN J. MASTROTOTARO, Ph.D. DIABETES is a major source of morbidity,

More information

Updates in Diabetes Technology

Updates in Diabetes Technology Updates in Diabetes Technology Jessica Kirk, MSN, RN, CPN, CDE Nurse Manager, Endo ECHO No disclosures Disclosures 1 Objectives Distinguish patients appropriate for continuous glucose monitoring and insulin

More information

Hypoglycemia Detection and Prediction Using Continuous Glucose Monitoring A Study on Hypoglycemic Clamp Data

Hypoglycemia Detection and Prediction Using Continuous Glucose Monitoring A Study on Hypoglycemic Clamp Data Journal of Diabetes Science and Technology Volume 1, Issue 5, September 2007 Diabetes Technology Society SYMPOSIUM Hypoglycemia Detection and Prediction Using Continuous Glucose Monitoring A Study on Hypoglycemic

More information

Advances in Diabetes Care Technologies

Advances in Diabetes Care Technologies 1979 Advances in Diabetes Care Technologies 2015 Introduction Insulin pump use: ~ 20% - 30% of patients with T1DM < 1% of insulin-treated patients with T2DM 2007 FDA estimates ~375,000 insulin pumps for

More information

Diabetes II Insulin pumps; Continuous glucose monitoring system (CGMS) Ernest Asamoah, MD FACE FACP FRCP (Lond)

Diabetes II Insulin pumps; Continuous glucose monitoring system (CGMS) Ernest Asamoah, MD FACE FACP FRCP (Lond) Diabetes II Insulin pumps; Continuous glucose monitoring system (CGMS) Ernest Asamoah, MD FACE FACP FRCP (Lond) 9501366-011 20110401 Objectives Understand the need for insulin pumps and CGMS in managing

More information

Continuous Glucose Monitoring (CGM)

Continuous Glucose Monitoring (CGM) Continuous Glucose Monitoring (CGM) Date of Origin: 02/2001 Last Review Date: 07/26/2017 Effective Date: 07/26/2017 Dates Reviewed: 04/2004, 04/2005, 03/2006, 11/2006, 12/2007, 03/2008, 09/2008, 04/2009,

More information

Diabetes Care 34: , with an increased risk of SH. In a

Diabetes Care 34: , with an increased risk of SH. In a Clinical Care/Education/Nutrition/Psychosocial Research O R I G I N A L A R T I C L E Factors Predictive of Severe Hypoglycemia in Type 1 Diabetes Analysis from the Juvenile Diabetes Research Foundation

More information

Comparison of the Numerical and Clinical Accuracy of Four Continuous Glucose Monitors

Comparison of the Numerical and Clinical Accuracy of Four Continuous Glucose Monitors Emerging Treatments and Technologies O R I G I N A L A R T I C L E Comparison of the Numerical and Clinical Accuracy of Four Continuous Glucose Monitors BORIS KOVATCHEV, PHD 1 STACEY ANDERSON, MD 1 2 LUTZ

More information

Clinical Value and Evidence of Continuous Glucose Monitoring

Clinical Value and Evidence of Continuous Glucose Monitoring Clinical Value and Evidence of Continuous Glucose Monitoring 9402313-012 Objective To review the clinical value and the recent clinical evidence for Professional and Personal CGM Key Points CGM reveals

More information

Long-term effects of continuous glucose monitoring on HbA 1c levels: An audit

Long-term effects of continuous glucose monitoring on HbA 1c levels: An audit Long-term effects of continuous glucose monitoring on Julie Brake Continuous glucose monitoring (CGM) has become a common and useful tool in diabetes care. To understand whether a 72-hour glucose profile

More information

Continuous Glucose Monitoring (CGM)

Continuous Glucose Monitoring (CGM) Continuous Glucose Monitoring (CGM) Date of Origin: 02/2001 Last Review Date: 08/22/2018 Effective Date: 08/22/2018 Dates Reviewed: 04/2004, 04/2005, 03/2006, 11/2006, 12/2007, 03/2008, 09/2008, 04/2009,

More information

Advances in Diabetes Care Technologies

Advances in Diabetes Care Technologies 1979 Advances in Diabetes Care Technologies 2015 Introduction Roughly 20% - 30% of patients with T1DM and fewer than 1% of insulin-treated patients with T2DM use an insulin pump In 2007, the US FDA estimated

More information

QUESTION 4. WHAT CLINICAL DATA ARE CURRENTLY AVAILABLE TO SUPPORT EXPANDED CGM COVERAGE BY PAYERS AS PERTAINS TO QUESTIONS 1 AND 3?

QUESTION 4. WHAT CLINICAL DATA ARE CURRENTLY AVAILABLE TO SUPPORT EXPANDED CGM COVERAGE BY PAYERS AS PERTAINS TO QUESTIONS 1 AND 3? 500 1 QUESTION 4. WHAT CLINICAL DATA ARE CURRENTLY AVAILABLE TO SUPPORT EXPANDED CGM COVERAGE BY PAYERS AS PERTAINS TO QUESTIONS 1 AND 3? WHAT ADDITIONAL DATA ARE NEEDED? AACE/ACE CGM Consensus Conference:

More information

The recovery of cognitive function following hypoglycemia

The recovery of cognitive function following hypoglycemia RIEF REPORT Delayed Recovery of Cognitive Function Following Hypoglycemia in Adults With Type 1 Diabetes Effect of Impaired Awareness of Hypoglycemia Nicola N. Zammitt, 1 Roderick E. Warren, 1 Ian J. Deary,

More information

hypoglycaemia unawareness keystone 18 July 2014

hypoglycaemia unawareness keystone 18 July 2014 hypoglycaemia unawareness keystone 18 July 2014 Hypoglycaemia unawareness: ( Impaired awareness of hypoglycaemia ) Philip Home Newcastle University Philip Home Duality of interest Manufacturers of glucose-lowering

More information

Artificial Pancreas Device System (APDS)

Artificial Pancreas Device System (APDS) Medical Policy Manual Durable Medical Equipment, Policy No. 77 Artificial Pancreas Device System (APDS) Next Review: October 2019 Last Review: October 2018 Effective: November 1, 2018 IMPORTANT REMINDER

More information

The In-Clinic Close Loop Experience in the US

The In-Clinic Close Loop Experience in the US The In-Clinic Close Loop Experience in the US Keystone Symposium, Practical Ways to Achieve Targets in Diabetes Care July 14, 2012 Francine Ratner Kaufman, MD Chief Medical Officer, VP Global Medical,

More information

CareLink. software REPORT REFERENCE GUIDE. Management Software for Diabetes

CareLink. software REPORT REFERENCE GUIDE. Management Software for Diabetes CareLink Management Software for Diabetes software REPORT REFERENCE GUIDE How to use this guide Each type of CareLink report and its components are described in the following sections. Report data used

More information

Agreement between Glucose Trends Derived from Three Simultaneously Worn Continuous Glucose Sensors

Agreement between Glucose Trends Derived from Three Simultaneously Worn Continuous Glucose Sensors Journal of Diabetes Science and Technology Volume 2, Issue 5, September 2008 Diabetes Technology Society ORIGINAL ARTICLES Agreement between Glucose Trends Derived from Three Simultaneously Worn Continuous

More information

Supplementary Appendix

Supplementary Appendix Supplementary Appendix This appendix has been provided by the authors to give readers additional information about their work. Supplement to: Phillip M, Battelino T, Atlas E, et al. Nocturnal glucose control

More information

CONTINUOUS OR INTERMITTENT GLUCOSE MONITORING IN INTERSTITIAL FLUID

CONTINUOUS OR INTERMITTENT GLUCOSE MONITORING IN INTERSTITIAL FLUID FLUID Non-Discrimination Statement and Multi-Language Interpreter Services information are located at the end of this document. Coverage for services, procedures, medical devices and drugs are dependent

More information

Real-Time Continuous Glucose Monitoring: From Application to Evaluation

Real-Time Continuous Glucose Monitoring: From Application to Evaluation Real-Time Continuous Glucose Monitoring: From Application to Evaluation Gary Scheiner MS, CDE Owner/Director, Integrated Diabetes Services 333 E. Lancaster Ave., Suite 24 Wynnewood, PA 1996 (877) 735-3648

More information

Reduced Awareness of Hypoglycemia in Adults With IPPM

Reduced Awareness of Hypoglycemia in Adults With IPPM O R I G I N A L A R T I C L E Reduced Awareness of Hypoglycemia in Adults With IPPM A prospective study of hypoglycemic frequency and associated symptoms WILLIAM L. CLARKE, MD DANIEL J. Cox, PHD LINDA

More information

Today s Goals 10/6/2017. New Frontiers in Diabetes Technology. Disclosures

Today s Goals 10/6/2017. New Frontiers in Diabetes Technology. Disclosures New Frontiers in Diabetes Technology Marie E. McDonnell, MD Director, Brigham and Women's Diabetes Program Division of Endocrinology, Diabetes and Hypertension Brigham and Women s Hospital Today s Goals

More information

Cognitive function testing in studies of acute hypoglycaemia: rights and wrongs?

Cognitive function testing in studies of acute hypoglycaemia: rights and wrongs? Diabetologia (998) : 7±79 Ó Springer-Verlag 998 For debates Cognitive function testing in studies of acute hypoglycaemia: rights and wrongs? S. A. Amiel King's College School of Medicine and Dentistry,

More information

Insulin Delivery and Glucose Monitoring Methods for Diabetes Mellitus: Comparative Effectiveness

Insulin Delivery and Glucose Monitoring Methods for Diabetes Mellitus: Comparative Effectiveness Insulin Delivery and Glucose Monitoring Methods for Diabetes Mellitus: Comparative Effectiveness Prepared for: Agency for Healthcare Research and Quality (AHRQ) www.ahrq.gov Outline of Material Introduction

More information

The Current Environment of CGM Technologies. Barry H. Ginsberg, M.D., Ph.D.

The Current Environment of CGM Technologies. Barry H. Ginsberg, M.D., Ph.D. Journal of Diabetes Science and Technology Volume 1, Issue 1, January 2007 Diabetes Technology Society SYMPOSIUM The Current Environment of CGM Technologies Introduction Barry H., M.D., Ph.D. This presentation

More information

Assessing the value of the Ambulatory Glucose Profile in clinical practice

Assessing the value of the Ambulatory Glucose Profile in clinical practice Assessing the value of the Ambulatory Glucose Profile in clinical practice STEPHAN MATTHAEI Abstract Glycated haemoglobin (HbA 1c) is a measure of mean blood glucose levels over time. It is not a good

More information

Making the Most of Continuous Glucose Monitoring

Making the Most of Continuous Glucose Monitoring Making the Most of Continuous Glucose Monitoring Gary Scheiner MS, CDE Owner & Clinical Director Integrated Diabetes Services LLC Wynnewood, PA AADE 2014 Diabetes Educator of the Year gary@integrateddiabetes.com

More information

Control of Glycemic Variability for Reducing Hypoglycemia Jae Hyeon Kim

Control of Glycemic Variability for Reducing Hypoglycemia Jae Hyeon Kim Control of Glycemic Variability for Reducing Hypoglycemia Jae Hyeon Kim Division of Endocrinology and Metabolism, Samsung Medical Center, Sungkyunkwan University School of Medicine Conflict of interest

More information

Continuous Glucose Monitoring

Continuous Glucose Monitoring Continuous Glucose Monitoring What is Continuous Glucose Monitoring? Blood glucose meters measure glucose in your blood and glucose sensors measure glucose levels in the fluid around the cells They are

More information

Preventing Hypoglycemia Using Predictive Alarm Algorithms and Insulin Pump Suspension

Preventing Hypoglycemia Using Predictive Alarm Algorithms and Insulin Pump Suspension DIABETES TECHNOLOGY & THERAPEUTICS Volume 11, Number 2, 29 Mary Ann Liebert, Inc. DOI: 1.189/dia.28.32 Preventing Hypoglycemia Using Predictive Alarm Algorithms and Insulin Pump Suspension Bruce Buckingham,

More information

The Realities of Technology in Type 1 Diabetes

The Realities of Technology in Type 1 Diabetes The Realities of Technology in Type 1 Diabetes May 6, 2017 Rosanna Fiallo-scharer, MD Margaret Frederick, RN Disclosures I have no conflicts of interest to disclose I will discuss some unapproved treatments

More information

The transition from urinary

The transition from urinary PRACTICAL POINTERS Continuous Glucose Monitoring: A Perspective on Its Past, Present, and Future Applications for Diabetes Management Hanna S. Mariani, 1 Brian T. Layden, 1,2 and Grazia Aleppo 1 1 Division

More information

Accuracy Requirements for a Hypoglycemia Detector: An Analytical Model to Evaluate the Effects of Bias, Precision, and Rate of Glucose Change

Accuracy Requirements for a Hypoglycemia Detector: An Analytical Model to Evaluate the Effects of Bias, Precision, and Rate of Glucose Change Journal of Diabetes Science and Technology Volume 1, Issue 5, September 2007 Diabetes Technology Society SYMPOSIUM Accuracy Requirements for a Hypoglycemia Detector: An Analytical Model to Evaluate the

More information

NEW TECHNOLOGIES FOR MANAGING DIABETES ANGELA THOMPSON DNP, FNP-C, BC-ADM, CDE, FAANP

NEW TECHNOLOGIES FOR MANAGING DIABETES ANGELA THOMPSON DNP, FNP-C, BC-ADM, CDE, FAANP NEW TECHNOLOGIES FOR MANAGING DIABETES ANGELA THOMPSON DNP, FNP-C, BC-ADM, CDE, FAANP No commercial support or sponsorship was received for this project I have nothing to disclose OBJECTIVES Identify at

More information

Performance Evaluations of Continuous Glucose Monitoring Systems: Precision Absolute Relative Deviation Is Part of the Assessment

Performance Evaluations of Continuous Glucose Monitoring Systems: Precision Absolute Relative Deviation Is Part of the Assessment Journal of Diabetes Science and Technology Volume 7, Issue 4, July 2013 Diabetes Technology Society SYMPOSIUM Performance Evaluations of Continuous Glucose Monitoring Systems: Karin, B.S., Günther Schmelzeisen-Redeker,

More information

Prediction of severe hypoglycemia. Additional information for this article can be found in an online appendix at

Prediction of severe hypoglycemia. Additional information for this article can be found in an online appendix at Diabetes Care In Press, published online March 15, 2007 Prediction of severe hypoglycemia Received for publication 3 July 2006 and accepted in revised form 7 March 2007. Additional information for this

More information

Advances in Diabetes Care Technologies

Advances in Diabetes Care Technologies Advances in Diabetes Care Technologies 1979 2015 Introduction Roughly 20% to 30% of patients with T1DM and fewer than 1% of insulin-treated patients with T2DM use an insulin pump In 2007, the U.S. FDA

More information

Report Reference Guide

Report Reference Guide Report Reference Guide How to use this guide Each type of CareLink report and its components are described in the following sections. Report data used to generate the sample reports was from sample patient

More information

DISCLAIMER: ECHO Nevada emphasizes patient privacy and asks participants to not share ANY Protected Health Information during ECHO clinics.

DISCLAIMER: ECHO Nevada emphasizes patient privacy and asks participants to not share ANY Protected Health Information during ECHO clinics. DISCLAIMER: Video will be taken at this clinic and potentially used in Project ECHO promotional materials. By attending this clinic, you consent to have your photo taken and allow Project ECHO to use this

More information

THE SHEFFIELD AREA PRESCRIBING GROUP. Position Statement for Prescribing of Freestyle Libre In Type 1 Diabetes. Date: March 2018.

THE SHEFFIELD AREA PRESCRIBING GROUP. Position Statement for Prescribing of Freestyle Libre In Type 1 Diabetes. Date: March 2018. THE SHEFFIELD AREA PRESCRIBING GROUP Position Statement for Prescribing of Freestyle Libre In Type 1 Diabetes Date: March 2018 Overview Freestyle Libre is a flash glucose sensor device that measures *interstitial

More information

Subjects and Methods 2017, 64 (8),

Subjects and Methods 2017, 64 (8), 2017, 64 (8), 827-832 Careful readings for a flash glucose monitoring system in nondiabetic Japanese subjects: individual differences and discrepancy in glucose concentrarion after glucose loading Keiko

More information

Coverage Guidelines. Continuous Glucose Monitors (CGMs)

Coverage Guidelines. Continuous Glucose Monitors (CGMs) Coverage Guidelines Continuous Glucose Monitors (CGMs) Disclaimer: Please note that Baptist Health Plan Coverage Guidelines may be updated throughout the year. A printed version may not be most up to date

More information

Advances in Technology in the Treatment of Diabetes Mellitus 2017 How far have we come-how far are we going? Is there a final frontier?

Advances in Technology in the Treatment of Diabetes Mellitus 2017 How far have we come-how far are we going? Is there a final frontier? Advances in Technology in the Treatment of Diabetes Mellitus 2017 How far have we come-how far are we going? Is there a final frontier? Alan B Schorr DO FAAIM FACE www.sugardoc.com abs@sugardoc.com Disclosures

More information

First steps for success.

First steps for success. First steps for success. Getting to know continuous glucose monitoring (CGM). The Animas Vibe System is approved for persons age 2 and older. Important Safety Information The Animas Vibe Insulin Pump and

More information

Norbert Hermanns, PhD 1,2, Beatrix Schumann, MD 2, Bernhard Kulzer, PhD 1,2, and Thomas Haak, MD 1,2. Original Article

Norbert Hermanns, PhD 1,2, Beatrix Schumann, MD 2, Bernhard Kulzer, PhD 1,2, and Thomas Haak, MD 1,2. Original Article 524105DSTXXX10.1177/1932296814524105Journal of Diabetes Science and TechnologyHermanns et al research-article2014 Original Article The Impact of Continuous Glucose Monitoring on Low Interstitial Glucose

More information

Corporate Medical Policy

Corporate Medical Policy Corporate Medical Policy Continuous Monitoring of Glucose in the Interstitial Fluid File Name: Origination: Last CAP Review: Next CAP Review: Last Review: continuous_monitoring_of_glucose_in_the_interstitial_fluid

More information

FreeStyle Mini Blood Glucose Results Are Accurate and Suitable for Use in Glycemic Clamp Protocols

FreeStyle Mini Blood Glucose Results Are Accurate and Suitable for Use in Glycemic Clamp Protocols Journal of Diabetes Science and Technology Volume 2, Issue 5, September 2008 Diabetes Technology Society CLINICAL APPLICATIONS FreeStyle Mini Blood Glucose Results Are Accurate and Suitable for Use in

More information

Abstract CLINICAL APPLICATIONS. Objectives: Methods: Results: Conclusion: Journal of Diabetes Science and Technology

Abstract CLINICAL APPLICATIONS. Objectives: Methods: Results: Conclusion: Journal of Diabetes Science and Technology Journal of Diabetes Science and Technology Volume 1, Issue 3, May 2007 Diabetes Technology Society CLINICAL APPLICATIONS Combined Insulin Pump Therapy with Real-Time Continuous Glucose Monitoring Significantly

More information

Learning Objectives. At the conclusion of this module, participants should be better able to:

Learning Objectives. At the conclusion of this module, participants should be better able to: Learning Objectives At the conclusion of this module, participants should be better able to: Treat asymptomatic neonatal hypoglycemia with buccal dextrose gel Develop patient-specific approaches to intravenous

More information

Continuous or Intermittent Monitoring of Glucose in Interstitial Fluid. Original Policy Date

Continuous or Intermittent Monitoring of Glucose in Interstitial Fluid. Original Policy Date MP 1.01.15 Continuous or Intermittent Monitoring of Glucose in Interstitial Fluid Medical Policy Section Durable Medical Equipment Issue 12:2013 Original Policy Date 12:2013 Last Review Status/Date Reviewed

More information

Medical Policy An independent licensee of the Blue Cross Blue Shield Association

Medical Policy An independent licensee of the Blue Cross Blue Shield Association Continuous or Intermittent Monitoring of Glucose in Interstitial Fluid Page 1 of 26 Medical Policy An independent licensee of the Blue Cross Blue Shield Association Title: Continuous or Intermittent Monitoring

More information

Acute Glycemic Changes in Brain and Subcutaneous Tissue Measured by Continuous Glucose Monitoring System in Hereditary Hypertriglyceridemic Rat

Acute Glycemic Changes in Brain and Subcutaneous Tissue Measured by Continuous Glucose Monitoring System in Hereditary Hypertriglyceridemic Rat Physiol. Res. 67: 127-131, 2018 Acute Glycemic Changes in Brain and Subcutaneous Tissue Measured by Continuous Glucose Monitoring System in Hereditary Hypertriglyceridemic Rat M. ŽOUREK 1, P. KYSELOVÁ

More information

Diabetes Technology Update. Sarah Konigsberg, MD Diabetes & Endocrine Assoc. April 7, 2018

Diabetes Technology Update. Sarah Konigsberg, MD Diabetes & Endocrine Assoc. April 7, 2018 Diabetes Technology Update Sarah Konigsberg, MD Diabetes & Endocrine Assoc. April 7, 2018 Disclosures None No future technologies are FDA approved Continuous Glucose Monitors Continuous Glucose Monitors

More information

JDRF Perspective on Closed Loop

JDRF Perspective on Closed Loop JDRF Perspective on Closed Loop Aaron J. Kowalski, Ph.D. Assistant Vice President Treatment Therapies Juvenile Diabetes Research Foundation International 1 Presenter Disclosure Aaron Kowalski Disclosed

More information

PREVENTION OF NOCTURNAL HYPOGLYCEMIA USING PREDICTIVE LOW GLUCOSE SUSPEND (PLGS)

PREVENTION OF NOCTURNAL HYPOGLYCEMIA USING PREDICTIVE LOW GLUCOSE SUSPEND (PLGS) PREVENTION OF NOCTURNAL HYPOGLYCEMIA USING PREDICTIVE LOW GLUCOSE SUSPEND (PLGS) Pathways for Future Treatment and Management of Diabetes H. Peter Chase, MD Carousel of Hope Symposium Beverly Hilton, Beverly

More information

death if the deficit is prolonged. Children and adults exposed to hypoglycaemia can develop long-term impairment of cognitive function (Blattner,

death if the deficit is prolonged. Children and adults exposed to hypoglycaemia can develop long-term impairment of cognitive function (Blattner, Introduction Glucose is the primary source of fuel for the cells of the brain. Brain is dependent on a continuous supply of glucose diffusing from the blood into the interstitial tissue within the central

More information

REPORT INTERPRETATION

REPORT INTERPRETATION REPORT INTERPRETATION: Interpreting ipro Professional Continuous Glucose Monitoring (CGM) Reports and Making Therapy Adjustments TARGET AUDIENCE The audience for this section is physicians, mid-level practitioners,

More information

UvA-DARE (Digital Academic Repository) Hypoglycaemia in diabetes Schopman, J.E. Link to publication

UvA-DARE (Digital Academic Repository) Hypoglycaemia in diabetes Schopman, J.E. Link to publication UvA-DARE (Digital Academic Repository) Hypoglycaemia in diabetes Schopman, J.E. Link to publication Citation for published version (APA): Schopman, J. E. (013). Hypoglycaemia in diabetes General rights

More information

Medical Policy An independent licensee of the Blue Cross Blue Shield Association

Medical Policy An independent licensee of the Blue Cross Blue Shield Association Continuous or Intermittent Monitoring of Glucose in Interstitial Fluid Page 1 of 39 Medical Policy An independent licensee of the Blue Cross Blue Shield Association Title: Continuous or Intermittent Monitoring

More information

Report Reference Guide. THERAPY MANAGEMENT SOFTWARE FOR DIABETES CareLink Report Reference Guide 1

Report Reference Guide. THERAPY MANAGEMENT SOFTWARE FOR DIABETES CareLink Report Reference Guide 1 Report Reference Guide THERAPY MANAGEMENT SOFTWARE FOR DIABETES CareLink Report Reference Guide 1 How to use this guide Each type of CareLink report and its components are described in the following sections.

More information

Medical Policy An independent licensee of the Blue Cross Blue Shield Association

Medical Policy An independent licensee of the Blue Cross Blue Shield Association Continuous or Intermittent Monitoring of Glucose in Interstitial Fluid Page 1 of 42 Medical Policy An independent licensee of the Blue Cross Blue Shield Association Title: Continuous or Intermittent Monitoring

More information

Diabetes Management with Continuous Glucose Monitoring & Multiple Daily Injections. Aaron Michels MD

Diabetes Management with Continuous Glucose Monitoring & Multiple Daily Injections. Aaron Michels MD Diabetes Management with Continuous Glucose Monitoring & Multiple Daily Injections Aaron Michels MD Outline SMBG & CGM by age group JDRF CGM Trial Sensor Augmented Insulin Pump Therapy for A1c Reduction

More information

Continuous Glucose Monitoring: The Future of Diabetes Management

Continuous Glucose Monitoring: The Future of Diabetes Management In Brief Continuous glucose monitoring (CGM) technology has the potential to revolutionize diabetes care in the near future because of the real-time feedback it provides about therapeutic interventions

More information

Artificial Pancreas Device Systems. Populations Interventions Comparators Outcomes Individuals: With type 1 diabetes

Artificial Pancreas Device Systems. Populations Interventions Comparators Outcomes Individuals: With type 1 diabetes Protocol Artificial Pancreas Device Systems Medical Benefit Effective Date: 07/01/18 Next Review Date: 01/20 Preauthorization Yes Review Dates: 03/15, 03/16, 03/17, 01/18, 05/18, 01/19 Preauthorization

More information

Continuous Glucose Monitoring System

Continuous Glucose Monitoring System Continuous Glucose Monitoring System Policy Number: Original Effective Date: MM.02.003 03/13/2001 Line(s) of Business: Current Effective Date: HMO; PPO; QUEST Integration 4/1/2018 Section: DME Place(s)

More information

Continuous Glucose Monitoring System

Continuous Glucose Monitoring System Continuous Glucose Monitoring System Policy Number: Original Effective Date: MM.02.003 03/13/2001 Line(s) of Business: Current Effective Date: HMO; PPO; QUEST Integration 04/01/2017 Section: DME Place(s)

More information

Accuracy of the 5-day FreeStyle Navigator Continuous Glucose Monitoring System: comparison with frequent laboratory reference measurements

Accuracy of the 5-day FreeStyle Navigator Continuous Glucose Monitoring System: comparison with frequent laboratory reference measurements Diabetes Care In Press, published online March 2, 2007 Accuracy of the 5-day FreeStyle Navigator Continuous Glucose Monitoring System: comparison with frequent laboratory reference measurements SHORT RUNNING

More information

Glycemic Variability:

Glycemic Variability: Glycemic Variability: Do the Differences Make a Difference? Kim L Kelly, PharmD, BCPS, FCCP Define Variability VERY LOW GLYCEMIC VARIABILITY LOW GLYCEMIC VARIABILITY HIGH GLYCEMIC VARIABILITY So what s

More information

Continuous Glucose Monitoring System

Continuous Glucose Monitoring System Continuous Glucose Monitoring System Policy Number: Original Effective Date: MM.02.003 03/13/2001 Line(s) of Business: Current Effective Date: HMO; PPO; QUEST Integration 04/01/2016 Section: DME Place(s)

More information

1. Continuous Glucose Monitoring

1. Continuous Glucose Monitoring 1. Continuous Glucose Monitoring 1. Physiology of interstitial fluid glucose 2. Comparison of CGM and self-monitored blood glucose (SMBG) data 3. Insulin dosing indication in BGM vs. CGM & the FDA 4. Protection

More information

James J. Chamberlain, MD 1, Dana Dopita, RN, MSN, CDE 1, Emily Gilgen, RD, CD, CDE 1, and Annie Neuman, MPA, PA-C 1.

James J. Chamberlain, MD 1, Dana Dopita, RN, MSN, CDE 1, Emily Gilgen, RD, CD, CDE 1, and Annie Neuman, MPA, PA-C 1. 604633DSTXXX10.1177/1932296815604633Journal of Diabetes Science and TechnologyChamberlain et al research-article2015 Original Article Impact of Frequent and Persistent Use of Continuous Glucose Monitoring

More information

Inpatient Studies of a Kalman-Filter-Based Predictive Pump Shutoff Algorithm

Inpatient Studies of a Kalman-Filter-Based Predictive Pump Shutoff Algorithm Journal of Diabetes Science and Technology Volume 6, Issue 5, September 2012 Diabetes Technology Society ORIGINAL ARTICLE Inpatient Studies of a Kalman-Filter-Based Predictive Pump Shutoff Algorithm Fraser,

More information

Hypoglycemia Prediction with Subject-Specific Recursive Time-Series Models

Hypoglycemia Prediction with Subject-Specific Recursive Time-Series Models Journal of Diabetes Science and Technology Volume 4, Issue 1, January 2010 Diabetes Technology Society SYMPOSIUM Hypoglycemia Prediction with Subject-Specific Recursive Time-Series Models Meriyan, M.E.,

More information

Evaluation of a Continuous Glucose Monitoring System for Home-Use Conditions

Evaluation of a Continuous Glucose Monitoring System for Home-Use Conditions Evaluation of a Continuous Glucose Monitoring System for Home-Use Conditions Bruce Bode, MD 1, Michael Silver, MD 2, Richard Weiss, MS, 3 Kathryn Martin, PharmD. 3 1 Atlanta Diabetes Associates; 2 Medical

More information

Continuous Glucose Monitoring Devices Pharmacy Policy

Continuous Glucose Monitoring Devices Pharmacy Policy Line of Business: All Line of Business Effective date: August 16, 2017 Revision date: August 16, 2017 Continuous Glucose Monitoring Devices Pharmacy Policy This policy has been developed through review

More information

Continuous Glucose Monitors for Diabetes Management

Continuous Glucose Monitors for Diabetes Management Continuous Glucose Monitors for Diabetes Management Ryan Huang, DO PGY II, Sonia Garcia-Jayne, DO PGY II Mandeep Gill, DO PGY I, Justin Leeka, DO, PGY I, Catherine Nguyen OMS IV Family Medicine Residency,

More information

Presented by Dr. Bruce Perkins, MD MPH Dr. Michael Riddell, PhD

Presented by Dr. Bruce Perkins, MD MPH Dr. Michael Riddell, PhD Type 1 Diabetes and Exercise: Optimizing the Medtronic MiniMed Veo Insulin Pump and Continuous Glucose Monitoring (CGM) for Better Glucose Control 1,2 for Healthcare Professionals Presented by Dr. Bruce

More information

Continuous Glucose Monitoring and the Reality of Metabolic Control in Preschool Children With Type 1 Diabetes

Continuous Glucose Monitoring and the Reality of Metabolic Control in Preschool Children With Type 1 Diabetes Emerging Treatments and Technologies O R I G I N A L A R T I C L E Continuous Glucose Monitoring and the Reality of Metabolic Control in Preschool Children With Type 1 Diabetes GEORGE S. JEHA, MD 1,2 LEFKOTHEA

More information

Artificial Pancreas Device Systems. Populations Interventions Comparators Outcomes. pump. pump

Artificial Pancreas Device Systems. Populations Interventions Comparators Outcomes. pump. pump Protocol Artificial Pancreas Device Systems (10130) Medical Benefit Effective Date: 04/01/18 Next Review Date: 01/19 Preauthorization Yes Review Dates: 03/15, 03/16, 03/17, 01/18 Preauthorization is required.

More information

1. What s the point of a network the case for research? 2. How to use CSII effectively

1. What s the point of a network the case for research? 2. How to use CSII effectively 1. What s the point of a network the case for research? 2. How to use CSII effectively John Pickup King s College London Faculty of Medicine Guy s Hospital, London What should an insulin pump network do?

More information

WOULD YOU LIKE TO REDUCE THE FEAR OF HYPOGLYCAEMIA FOR YOUR PATIENTS?

WOULD YOU LIKE TO REDUCE THE FEAR OF HYPOGLYCAEMIA FOR YOUR PATIENTS? WOULD YOU LIKE TO REDUCE THE FEAR OF HYPOGLYCAEMIA FOR YOUR PATIENTS? THE ONLY SYSTEM CLINICALLY PROVEN TO REDUCE HYPOGLYCAEMIA MINIMED 640G SYSTEM WITH SMARTGUARD TECHNOLOGY 1 HYPOGLYCAEMIA IS CHALLENGING

More information

A Comparison of Time Delay in Three Continuous Glucose Monitors for Adolescents and Adults

A Comparison of Time Delay in Three Continuous Glucose Monitors for Adolescents and Adults 704443DSTXXX10.1177/1932296817704443Journal of Diabetes Science and TechnologySinha et al research-article2017 Original Article A Comparison of Time Delay in Three Continuous Glucose Monitors for Adolescents

More information

Practical Strategies for the Clinical Use of Incretin Mimetics CME/CE. CME/CE Released: 09/15/2009; Valid for credit through 09/15/2010

Practical Strategies for the Clinical Use of Incretin Mimetics CME/CE. CME/CE Released: 09/15/2009; Valid for credit through 09/15/2010 Practical Strategies for the Clinical Use of Incretin Mimetics CME/CE Robert R. Henry, MD Authors and Disclosures CME/CE Released: 09/15/2009; Valid for credit through 09/15/2010 Introduction Type 2 diabetes

More information

THE ONLY SYSTEM^ CLINICALLY PROVEN TO REDUCE HYPOGLYCAEMIA

THE ONLY SYSTEM^ CLINICALLY PROVEN TO REDUCE HYPOGLYCAEMIA THE ONLY SYSTEM^ CLINICALLY PROVEN TO REDUCE HYPOGLYCAEMIA MINIMED 640G system^ with SmartGuard technology HYPOGLYCAEMIA IS CHALLENGING FOR GLYCAEMIC CONTROL 1 AND quality of life 2 Achieving good glucose

More information

Welcome to CareLink Pro

Welcome to CareLink Pro Reference Guide Welcome to CareLink Pro This guide was developed to serve as a reference for obtaining patient data and reviewing CareLink Pro reports. Getting Started with CareLink Pro Adding New Patients

More information

In general: Hypoglycemia is common in insulin treated diabetes, but may also occur in people on oral medications, especially sulfonylureas/glinides.

In general: Hypoglycemia is common in insulin treated diabetes, but may also occur in people on oral medications, especially sulfonylureas/glinides. 1 2 3 In general: Hypoglycemia is common in insulin treated diabetes, but may also occur in people on oral medications, especially sulfonylureas/glinides. 4 Answer: b and c Many alcohol containing drinks

More information

INFUSYSTEMS USA. INFUSYSTEM

INFUSYSTEMS USA. INFUSYSTEM INFUSYSTEMS USA. INFUSYSTEM INFUSYSTEMS USA - INFUSYSTEMS USA - INFUSYSTEMS USA - INFUSYSTEMS USA - INFUSYSTEMS USA - INFUSYSTEMS USA - INFUSYSTE Role of Continuous Glucose monitoring in Patients With

More information

Pumps & Sensors made easy. OPADA ALZOHAILI MD FACE Endocrinology Assistant Professor Wayne State University

Pumps & Sensors made easy. OPADA ALZOHAILI MD FACE Endocrinology Assistant Professor Wayne State University Pumps & Sensors made easy OPADA ALZOHAILI MD FACE Endocrinology Assistant Professor Wayne State University DeFronzo RA. Diabetes. 2009;58:773-795. Ominous Octet Relationship of b-cell Dysfunction and Development

More information

INSULIN INITIATION AND INTENSIFICATION WITH A FOCUS ON HYPOGLYCEMIA REDUCTION

INSULIN INITIATION AND INTENSIFICATION WITH A FOCUS ON HYPOGLYCEMIA REDUCTION INSULIN INITIATION AND INTENSIFICATION WITH A FOCUS ON HYPOGLYCEMIA REDUCTION Jaiwant Rangi, MD, FACE Nov 10 th 2018 DISCLOSURES Speaker Novo Nordisk Sanofi-Aventis Boheringer Ingleheim Merck Abbvie Abbott

More information

Continuous or Intermittent Glucose Monitoring in Interstitial Fluid

Continuous or Intermittent Glucose Monitoring in Interstitial Fluid Continuous or Intermittent Glucose Monitoring in Interstitial Fluid Policy Number: 1.01.20 Last Review: 9/2018 Origination: 11/2001 Next Review: 9/2019 Policy Blue Cross and Blue Shield of Kansas City

More information

Insulin Pumps and Glucose Sensors in Diabetes Management

Insulin Pumps and Glucose Sensors in Diabetes Management Diabetes Update+ 2014 Congress Whistler, British Columbia Friday March 21, 2014ǀ 8:15 8:45 am Insulin Pumps and Glucose Sensors in Diabetes Management Bruce A Perkins MD MPH Division of Endocrinology Associate

More information

Objectives. The Problem. We ve come a LONG way, Innovations in Diabetes Care and Management. Barbara Walz, RN, BSN, CDE April 26, 2018

Objectives. The Problem. We ve come a LONG way, Innovations in Diabetes Care and Management. Barbara Walz, RN, BSN, CDE April 26, 2018 Objectives Innovations in Diabetes Care and Management Barbara Walz, RN, BSN, CDE April 26, 2018 1. Briefly discuss the history of diabetes management over the past 100 years 2. Describe the impact of

More information

Objectives. Glucose Monitoring Today. Tracking Glucose ISO Standards. Continuous Glucose Monitoring & Diabetes Management 3/7/2015

Objectives. Glucose Monitoring Today. Tracking Glucose ISO Standards. Continuous Glucose Monitoring & Diabetes Management 3/7/2015 Objectives Continuous Glucose Monitoring & Diabetes Management Tomas C. Walker, DNP, APRN, CDE Director, Clinical Projects Dexcom, Inc. San Diego, CA 1. Review current limitations of SMBG 2. Understand

More information

State of California Health and Human Services Agency Department of Health Care Services

State of California Health and Human Services Agency Department of Health Care Services State of California Health and Human Services Agency Department of Health Care Services JENNIFER KENT Director EDMOND G. BROWN JR Governor DATE: N.L.: 03-0317 Index: Benefits TO: ALL COUNTY CALIFORNIA

More information

Diabetes Management: Current High Tech Innovations

Diabetes Management: Current High Tech Innovations Diabetes Management: Current High Tech Innovations How Far We ve Come in the Last 40 Years William V. Tamborlane, MD Department of Pediatrics Yale School of Medicine Disclosures I am a consultant for:

More information