FreeStyle Mini Blood Glucose Results Are Accurate and Suitable for Use in Glycemic Clamp Protocols

Size: px
Start display at page:

Download "FreeStyle Mini Blood Glucose Results Are Accurate and Suitable for Use in Glycemic Clamp Protocols"

Transcription

1 Journal of Diabetes Science and Technology Volume 2, Issue 5, September 2008 Diabetes Technology Society CLINICAL APPLICATIONS FreeStyle Mini Blood Glucose Results Are Accurate and Suitable for Use in Glycemic Clamp Protocols Ohad, M.D., 1 Sigal Shaklai, M.D., Ph.D., 1 Ephraim Gabis, B.Sc., 2 and Michael A. Pani, M.D. 3 Abstract Objective: We assessed the accuracy of the FreeStyle Mini (FSM) meter for use in glycemic clamp and meal protocols in comparison with the HemoCue Glucose 201 DM Analyzer (HemoCue) and the YSI 2300 STAT Glucose Oxidase Analyzer (YSI). Methods: Seven volunteers with type 2 diabetes mellitus, years old, underwent a frequently sampled meal test and a graded hyperglycemic test, on two separate days, with one of the volunteers undergoing each test twice. Samples for glucose measurements were obtained from arterialized venous blood. A total of 420 samples (with glucose levels ranging from 63 to 388 mg/dl) were available for comparison. On average, 10 measurements were available for every 5 mg/dl increment in glucose level in the range of mg/dl. Blood glucose measurements were done on each sample with the FSM, HemoCue, and YSI. Results: FreeStyle Mini blood glucose values correlated closely with the YSI readings. Of the FSM measurements, 99.0% were within the Clarke error grid zone A; 51.3%, 84.7%, and 96.2% of the FSM readings were within 5%, 10% and 15% of the YSI values, respectively. The FSM was significantly more accurate than the HemoCue (84.7% vs 76.6% of results within 10% of the YSI results; p =.0038). The mean average relative difference of the FSM (5.8%) was also significantly lower than that of the HemoCue (6.8%; p =.0013) Conclusions: The FSM provides accurate results and constitutes a suitable alternative for bedside blood glucose measurements in experimental procedures, helping to reduce sample size, turnaround time, and cost. J Diabetes Sci Technol 2008;2(5): Author Affiliations: 1 Institute of Endocrinology, Ch. Sheba Medical Center, Sackler School of Medicine, Tel Aviv University, Israel; 2 Orsense, Nes Ziona, Israel; and 3 Abbott Diabetes Care, Wiesbaden, Germany Abbreviations: (FSM) FreeStyle Mini, (HemoCue) HemoCue Glucose 201 DM Analyzer, (ISO) International Organization for Standardization, (MAD) mean absolute difference, (MARD) mean average relative difference, (SD) standard deviation, (T2DM) type 2 diabetes mellitus, (YSI) Yellow Springs Instruments 2300 STAT Glucose Oxidase Analyzer Keywords: glucose monitor, glycemic clamp, diabetes mellitus Corresponding Author: Ohad, M.D., Institute of Endocrinology, Ch. Sheba Medical Center, Tel Hashomer, Israel 52621; address ohad.cohen@sheba.health.gov.il 890

2 Introduction Clamp protocols are commonly used for the study of metabolic dysfunction in type 2 diabetes mellitus (T2DM) and other carbohydrate and lipid disorders. 1 In these protocols, emphasis is put on the accuracy of glucose measurements, as these are critical for the clamping procedure and for the interpretation of data. Glucose measurements are performed at intervals as short as two minutes and usually include measurements per single clamp procedure. The resulting need for a rapid and accurate diagnostic tool requiring only small blood samples to minimize blood loss is challenging. Glucose oxidase-based analyzers, such as YSI 2300 STAT (YSI) (Yellow Springs Instruments Incorporated, Yellow Springs, OH), are considered accurate and provide reference measurements for glucose. The use of the YSI as a bed-side tool rather than as a source of reference measurements is impractical, because of the relatively long turnaround intervals and calibration time outs that necessitate the use of more than one analyzer. Some institutes use the Beckman Glucose Analyzer (Fullerton, CA) 2 as a bed-side tool, though it is still cumbersome and needs trained staff for rapid and accurate operation. The accuracy of hand-held devices for glucose measurements, though sufficient for homebased clinical management, is inadequate for clamp procedures. To fulfill the need for accurate and rapid glucose measurements that use minimal blood samples, studies have been conducted to assess the accuracy of hand-held devices in clamp protocols. Brunner et al. 3 found the analytical accuracy of a hand-held meter, as evaluated by variance components analysis, to be suboptimal. More than 3% of measurements deviated by more than +/-20% from the reference value, rendering the home glucose monitor inadequate for clamp procedures. Similar results from a hyperinsulinemic clamp study were reported by Stork et al. 4 with 4% of samples deviating by more than +/-20%. Consequently, the YSI and Beckman analyzers remain the most frequently used devices in clamp protocols. Unlike most other hand-held meters, the FreeStyle Mini [(FSM) Abbott Diabetes Care, Alameda, CA] uses a glucose dehydrogenase assay based on coulometry and requires only a small sample (0.3 µl) to produce highly accurate results across the mg/dl range, independent of hematocrit, temperature, and many interfering substances. 5 Given the unmet need in clamp studies for an accurate, rapid, and cost-effective glucose monitoring system that requires only small blood samples, we investigated the suitability of using the FSM in glucose clamp protocols. Materials and Methods A total of 420 arterialized venous blood samples were obtained from seven patients with T2DM during eight extended meal tests and eight stepped hyperglycemic clamp tests. These clamp protocols were part of an extended study of the effects of medications on beta-cell function. All protocols performed at our clinical research center were approved by the center s internal review board and informed consent was obtained from every participant. These tests were conducted as follows: Graded hyperglycemic clamp. After an overnight fast, patients were admitted to the clinical research center and, after resting supine, they had a 20-gauge catheter inserted into an antecubital vein (for the infusion of test substances) and another catheter threaded in a retrograde manner into a wrist vein of the same arm, for blood sampling. The hand was kept in a heated box (55 C) for the duration of the study, to achieve and maintain arterialization of venous blood. The study protocol consisted of three consecutive 30-minute squarewave steps of hyperglycemia (each of 50 mg/dl above the previous level), achieved by a variable infusion of 20% dextrose, followed by an intravenous bolus of 5 g arginine. Blood samples for the determination of plasma glucose, insulin, and C-peptide concentrations were obtained every 2 minutes for the first 10 minutes and every 5 minutes thereafter for the final 20 minutes of each step. Meal test. The patients were admitted to the metabolism unit after an overnight (11- to 12-h) fast. An 18-gauge cannula was inserted in a retrograde manner into a dorsal hand vein to obtain arterialized venous blood samples. A mixed meal (10 kcal/kg, 45% carbohydrate, 15% protein, 40% fat) was consumed within 15 minutes and blood samples were obtained at 0, 5, 10, 15, 20, 30, 40, 50, 60, 75, 90, 120, 150, 180, 210, and 240 minutes. Blood samples were retrieved through a 3 way stopcock. When collecting the blood sample, the first 2 ml were discarded, and immediately thereafter a sample of approximately 3 ml was drawn, 2 ml were put into a prechilled vial containing powdered sodium fluoride and potassium oxalate (BD Vacutainer Plus; Becton, Dickinson, and Company, Franklin Lakes, NJ) and immediately 891

3 centrifuged for 15 s at 5800 g, separated, and placed on ice; the venous line was then flushed with saline. Subsequently, the glucose level in the plasma was analyzed with the YSI. From the remainder of the sample, a small amount of whole blood was immediately applied from the tip of the syringe and analyzed with the FSM glucose meter and a HemoCue Glucose 201 DM Analyzer (HemoCue AB, Ängelholm, Sweden). All instruments were maintained according to the manufacturers recommendations. Additionally, the YSI was calibrated at the start of every experimental day. Statistical Analysis Values were used for comparisons without corrections. The following statistical analyses were performed using SAS software (SAS Institute Inc., Cary, NC): Two-tailed Student s t-test on mean values from the HemoCue, FSM, and YSI with levels of significance defined as p <.05. Bias measures: Method of residuals: the differences between the YSI plasma glucose concentrations and the blood glucose meter readings (residuals) were calculated and presented as the mean, median, and standard deviation (SD) of the differences. Percentages were calculated for the HemoCue and FSM values within 5%, 10%, and 15%and within the International Organization for Standardization (ISO) requirements of the YSI value. System accuracy plot. Clarke and consensus grids. 6 Data are presented as means ± SD, unless otherwise indicated. As this trial was not devised exclusively to determine the accuracy of the FSM, no power analysis was performed or is presented. Data were collected from successive patients without excluding patients from the analysis. The use of successive data prevents data bias. Results A total of 420 matched glucose values were obtained from five male and two female patients undergoing eight meal tests and eight stepped hyperglycemic clamps (one female underwent two sets of studies). The mean age was 55.8 years (range: years), and weight 88.3 ± 15.6 kg. The mean duration of diabetes was five years and the mean hemoglobin A1c was 7.4 ± 0.6%. Mean glucose levels as obtained by the YSI, FSM, and HemoCue were ± 63.5 mg/dl, ± 68.3 mg/dl and ± 70.8 mg/dl, respectively. Glucose measurements ranged from 63 to 388 mg/dl; on average, 10 measurements were available for every 5 mg/dl increment in glucose levels in the range of mg/dl. The two protocol procedures differed in glucose levels with means for the meal test of ± 59.9 mg/dl, ± 63.8 mg/dl, and ± 65.1 mg/dl for YSI, FSM, and HemoCue, respectively. For the clamp, mean glucose levels were ± 59.6 mg/dl, ± 64.5 mg/dl, and ± 67.7 mg/dl for the YSI, FSM, and HemoCue, respectively. The results from the three glucose analyzers were highly correlated with an r 2 of 0.98 for the FSM vs YSI, 0.96 for the FSM vs HemoCue, and 0.97 for theysi vs HemoCue (p = nonsignificant). Grid analysis for the FSM (Figure 1) showed that 99.5% and 99.0% of the readings were in zone A on consensus and Clarke error grids, respectively. The results for the HemoCue were 99.5% and 98.6%, respectively (Figure 2). Bias measures for the FSM are presented in Table 1. The mean, median, and percent overall differences between the FSM and YSI were 7.5 mg/dl, 7.0 mg/dl, and 3.0%, respectively. The absolute mean, median, and percent differences between the FSM and YSI were 13.0 mg/dl, 11.0 mg/dl, and 5.8%, respectively. The differences during the hyperglycemic clamp were slightly greater than those during the meal tests. Of the FSM readings, 99.0% were Table 1. Bias Measures (FSM vs YSI) Measure Glucose Level Mean SD Median Absolute difference (mg/dl) All results Absolute percent difference All results Absolute difference (mg/dl) Clamp Absolute percent difference Clamp Absolute difference (mg/dl) Meal Absolute percent difference Meal Within 5% Within 10% Within 15% Within 20% (ISO) 215/419 (51.3%) 355/419 (84.7%) 403/419 (96.2%) 415/419 (99.0%) 892

4 within the ISO specifications. 7 A total of 51.3%, 84.7%, and 96.2% of the FSM glucose values were within 5%, 10%, and 15% of the YSI values, respectively. The absolute mean (15.7 mg/dl), median (13.0 mg/dl) and percent differences (6.8%) were significantly higher for the HemoCue (Table 2). A total of 42.7%, 76.6%, and 94.3% of the HemoCue glucose values fell within 5%, 10%, and 15% of the YSI values, respectively (p <.005 vs FSM) (Table 3). Discussion This study demonstrates the accuracy of the FSM during meal and graded hyperglycemic clamp protocols. These studies are characterized by rapid undulations in glucose levels spanning the range of mg/dl, as depicted in Figure 3, a representative graded hyperglycemic clamp. Consensus and Clarke error grids demonstrated the excellent clinical acceptability of the FSM results (Figure 1). However, these performance criteria, pertaining to the Consensus error grid Clarke error grid Figure 1. YSI vs FSM performance (Consensus and Clarke error grids). Table 2. Comparison of Bias Measures (FSM vs HemoCue) Mean absolute difference Figure 2. YSI vs HemoCue performance (Clarke error grids). FS-YSI HemoCue- YSI Measure N Mean SD Mean SD Absolute difference (mg/dl) Absolute percent difference (%) Median absolute difference FS-YSI HemoCue- YSI Measure N Median Median Absolute difference (mg/dl) Absolute percent difference (%) Paired difference (t-test p value) -2.6 (p <.0001) (p =.0013) Wilcoxon Rank Sum p value

5 Table 3. Percentage of FSM and HemoCue Measurements Within YSI Measurements (glucose 75mg/dl) System Within 5% Within 10% Within 15% Within 20% (ISO) FreeStyle 215/419 (51.3%) 355/419 (84.7%) 403/419 (96.2%) 415/419 (99.0%) HemoCue 179/419 (42.7%) 321/419 (76.6%) 395/419 (94.3%) 414/419 (98.8%) Fisher s Exact Test 2-sided p-value NS:.2565 NS: >.99 NS, nonsignificant Glucose (mg/dl) Time Figure 3. Representative plot of a graded hyperglycemic clamp. home-based clinical use of this device might not be appropriate in a clamp protocol, where high analytical accuracy is needed. To assess the analytical accuracy of the FSM on a level appropriate for clamp studies, bias measures were performed. The results of these studies indicated that the FSM was highly accurate, with a mean absolute difference (MAD) and mean average relative difference (MARD) of 13.0 mg/dl and 5.8%, respectively. The accuracy of the FSM was significantly superior to that of the HemoCue analyzer. Similar results for the HemoCue were presented by Trajanoski et al. 8 using a similar protocol but applying a hyperinsulinemic stepped hypoglycemic clamp. Thus, the study by Trajanoski et al. and the present study reflect a wide range of glucose levels, spanning from hypoglycemia to hyperglycemia. Similarly, arterialized venous blood was used and the reference values were obtained by a Beckman Glucose Analyzer II. The better results of the FSM can be due to assay accuracy and/or the fact that our study was conducted mostly in the hyperglycemic range. Whole blood glucose levels (by HemoCue) were not converted to match plasma levels (by YSI), as the MAD was 13.0 for the FSM and 15.7 (higher) for the HemoCue. In a similar study comparing the HemoCue to the YSI, Stork et al. 4 converted the whole blood glucose obtained by the HemoCue to plasma levels obtained by the YSI, by adding 10% to HemoCue whole blood. The differences between the results of the studies using conversion to plasma are not clear, as both studies used arterialized blood, which might lessen the differences between whole blood levels measured by a hand-held glucometer and serum levels measured by the YSI. 9 Finally, correcting the readings would make them only more accurate, but would not change the relationship between the FSM and HemoCue readings. Again, different glucose ranges, with an emphasis on hypoglycemia in other studies, 3,4,8 versus hyperglycemia in this study, might be a contributing factor to the differences between the results. For the FSM, the MAD and MARD of 13.0 mg/dl and 5.8%, as obtained in this study, indicate that this device is acceptable for use in glucose clamping protocols, particularly when rapid information is needed to correspond to rapid alterations in glycemic levels. Our findings concur with those of Brunner et al. 3 ; we found an individual variance in accuracy such that half of the patients had a MARD less than 5% and four patients had a MARD above 5%. Thus, we calibrate FSM readings during clamp studies with YSI measurements performed at different critical steps, for example, during a steady state interval or a step before increasing the glucose level; this culminates in 6-10 YSI measurements during the clamp. Additional advantages of the FSM are the convenience of use, small sample volume, rapid results, and low cost. The protocol is, as described, for stepped hyperglycemic clamp studies; it has been similarly described and used in several studies of beta-cell function, including studies of responses to different medications. This protocol was not devised specifically for assessing the accuracy of the FSM. In conclusion, we have shown that under specific conditions present in graded hyperglycemic clamp protocols, the measurements of the FSM glucose meter are sufficiently accurate to supply valid and useful glycemic data. Calibration with YSI results during the clamp procedure is recommended, as individual factors might impact the accuracy of the FSM measurements. 894

6 References: DeFronzo RA, Tobin JD, Andres R. Glucose clamp technique: a method for quantifying insulin secretion and resistance. Am J Physiol. 1979;237(3):E Clarke WL, Anderson S, Farhy L, Breton M, Gonder-Frederick L, Cox D, Kovatchev B. Evaluating the clinical accuracy of two continuous glucose sensors using continuous glucose-error grid analysis. Diabetes Care 2005;28(10): Brunner GA, Ellmerer M, Sendlhofer G, Wutte A, Trajanoski Z, Schaupp L, Quehenberger F, Wach P, Krejs GJ, Pieber TR. Validation of home blood glucose meters with respect to clinical and analytical approaches. Diabetes Care. 1998;21(4): Stork AD, Kemperman H, Erkelens DW, Veneman TF. Comparison of the accuracy of the HemoCue glucose analyzer with the Yellow Springs Instrument glucose oxidase analyzer, particularly in hypoglycemia. Eur J Endocrinol. 2005;153(2): Feldman B, McGarraugh G, Heller A, Bohannon N, Skyler J, DeLeeuw E, Clarke D. FreeStyle: a small-volume electrochemical glucose sensor for home blood glucose testing. Diabetes Technol Ther. 2000;2(2): Clarke WL, Cox D, Gonder-Frederick LA, Carter W, Pohl SL. Evaluating clinical accuracy of systems for self-monitoring of blood glucose. Diabetes Care. 1987;10(5): International Organization for Standardization. Requirements for in vitro blood glucose monitoring systems for self-testing in managing diabetes mellitus. ISO/TC 212/WG 3. In: Draft International Standard ISO/DIS Geneva, Switzerland: International Organization for Standardization; Trajanoski Z, Brunner GA, Gfrerer RJ, Wach P, Pieber TR. Accuracy of home blood glucose meters during hypoglycemia. Diabetes Care. 1996;19(12): Liu D, Moberg E, Kollind M, Lins PE, Adamson U, Macdonald IA. Arterial, arterialized venous and capillary blood glucose measurements in normal man during hyperinsulinaemic euglycaemia and hypoglycaemia. Diabetologia. 1992;35(3):

FreeStyle Lite A Blood Glucose Meter That Requires No Coding

FreeStyle Lite A Blood Glucose Meter That Requires No Coding Journal of Diabetes Science and Technology Volume 2, Issue 4, July 2008 Diabetes Technology Society SYMPOSIUM Shridhara, Ph.D. Abstract Background: Abbott Diabetes Care introduced the FreeStyle Lite blood

More information

Accuracy and Reliability of the Nova StatStrip Glucose Meter for Real-Time Blood Glucose Determinations during Glucose Clamp Studies

Accuracy and Reliability of the Nova StatStrip Glucose Meter for Real-Time Blood Glucose Determinations during Glucose Clamp Studies Journal of Diabetes Science and Technology Volume 4, Issue 5, September 2010 Diabetes Technology Society ORIGINAL ARTICLES Accuracy and Reliability of the Nova StatStrip Glucose Meter for Real-Time Blood

More information

Bern Harrison, B.A., Cheryl Leazenby, B.S., and Solveig Halldorsdottir, Ph.D.

Bern Harrison, B.A., Cheryl Leazenby, B.S., and Solveig Halldorsdottir, Ph.D. Journal of Diabetes Science and Technology Volume 5, Issue 4, July 2011 Diabetes Technology Society TECHNOLOGY REPORTS Accuracy of the CONTOUR Blood Glucose Monitoring System Bern, B.A., Cheryl Leazenby,

More information

Comparison of the Numerical and Clinical Accuracy of Four Continuous Glucose Monitors

Comparison of the Numerical and Clinical Accuracy of Four Continuous Glucose Monitors Emerging Treatments and Technologies O R I G I N A L A R T I C L E Comparison of the Numerical and Clinical Accuracy of Four Continuous Glucose Monitors BORIS KOVATCHEV, PHD 1 STACEY ANDERSON, MD 1 2 LUTZ

More information

Antisense Mediated Lowering of Plasma Apolipoprotein C-III by Volanesorsen Improves Dyslipidemia and Insulin Sensitivity in Type 2 Diabetes

Antisense Mediated Lowering of Plasma Apolipoprotein C-III by Volanesorsen Improves Dyslipidemia and Insulin Sensitivity in Type 2 Diabetes Antisense Mediated Lowering of Plasma Apolipoprotein C-III by Volanesorsen Improves Dyslipidemia and Insulin Sensitivity in Type 2 Diabetes Digenio A, et al. Table of Contents Detailed Methods for Clinical

More information

Supplementary Appendix

Supplementary Appendix Supplementary Appendix This appendix has been provided by the authors to give readers additional information about their work. Supplement to: Phillip M, Battelino T, Atlas E, et al. Nocturnal glucose control

More information

Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers

Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D iochemistry Analyzers Life Sciences Data for Life. TM YSI Life Sciences White Paper 91 Authors: Kevin Schlueter, PhD,

More information

Validity and Reliability of a Glucometer Against Industry Reference Standards

Validity and Reliability of a Glucometer Against Industry Reference Standards 514315DSTXXX10.1177/1932296813514315Journal of Diabetes Science and TechnologySalacinski et al research-article2014 Original Article Validity and Reliability of a Glucometer Against Industry Reference

More information

Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System

Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System Summary Objectives: To demonstrate that the TRUE METRIX Blood Glucose Monitoring System, from Trividia Health,

More information

A New Consensus Error Grid to Evaluate the Clinical Significance of Inaccuracies in the Measurement of Blood Glucose

A New Consensus Error Grid to Evaluate the Clinical Significance of Inaccuracies in the Measurement of Blood Glucose Emerging Treatments and O R I G I N A L A R T I C L E Technologies A New Consensus Error Grid to Evaluate the Clinical Significance of Inaccuracies in the Measurement of Blood Glucose JOAN L. PARKES, PHD

More information

Comparison of the accuracy of the hemocue glucose analyzer with the Yellow Springs Instrument glucose oxidase analyzer, particularly in hypoglycemia

Comparison of the accuracy of the hemocue glucose analyzer with the Yellow Springs Instrument glucose oxidase analyzer, particularly in hypoglycemia European Journal of Endocrinology (2005) 153 275 281 ISSN 0804-4643 CLINICAL STUDY Comparison of the accuracy of the hemocue glucose analyzer with the Yellow Springs Instrument glucose oxidase analyzer,

More information

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 System Accuracy Evaluation of Project number: Fora092614-01 System accuracy evaluation of Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Date: 13 th

More information

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Abstract Objective: The goal of the study was to assess

More information

Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System

Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System Summary Objectives: To demonstrate that the TRUE METRIX Self-Monitoring Blood Glucose System*, from Trividia

More information

Relationship between glucose meter error and glycemic control efficacy

Relationship between glucose meter error and glycemic control efficacy Relationship between glucose meter error and glycemic control efficacy Brad S. Karon, M.D., Ph.D. Professor of Laboratory Medicine and Pathology Department of Laboratory Medicine and Pathology Mayo Clinic

More information

GlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data

GlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data GlucoMen LX PLUS: Accuracy Evaluations to ISO 15197:2013 with Technical and Specification Data GlucoMen LX PLUS blood glucose and blood ketone meter Accuracy Evaluation to New ISO 15197:2013, with Technical

More information

Clinical Evaluation for. Embrace No Code Blood Glucose Monitoring System

Clinical Evaluation for. Embrace No Code Blood Glucose Monitoring System Embrace No Code Report Clinical Evaluation for Embrace No Code Blood Glucose Monitoring System Applicant: Apex Biotechnology Corp. Test Date: 2007/8/28~2007/9/21 Clinical Site: SCHMIDT Group Practice Clinic

More information

Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System

Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System Summary Objectives: To demonstrate that the TRUE METRIX AIR Blood Glucose Monitoring System, from Trividia

More information

Use of Continuous Glucose Monitoring in the Detection and Prevention of Hypoglycemia

Use of Continuous Glucose Monitoring in the Detection and Prevention of Hypoglycemia Journal of Diabetes Science and Technology Volume 1, Issue 1, January 2007 Diabetes Technology Society SYMPOSIUM Use of Continuous Glucose Monitoring in the Detection and Prevention of Hypoglycemia Howard

More information

Accuracy of the 5-day FreeStyle Navigator Continuous Glucose Monitoring System: comparison with frequent laboratory reference measurements

Accuracy of the 5-day FreeStyle Navigator Continuous Glucose Monitoring System: comparison with frequent laboratory reference measurements Diabetes Care In Press, published online March 2, 2007 Accuracy of the 5-day FreeStyle Navigator Continuous Glucose Monitoring System: comparison with frequent laboratory reference measurements SHORT RUNNING

More information

Standard Operating Procedures Clinical and Translational Research Center

Standard Operating Procedures Clinical and Translational Research Center Standard Operating Procedures Clinical and Translational Research Center Title: Approved By: Number of Pages: Frequently Sampled Intravenous Glucose Tolerance Test (FSIVGTT) Effective March 19, 2012 Date:

More information

Clinical Impact of Prandial State, Exercise, and Site Preparation on the Equivalence of Alternative-Site Blood Glucose Testing

Clinical Impact of Prandial State, Exercise, and Site Preparation on the Equivalence of Alternative-Site Blood Glucose Testing Clinical Care/Education/Nutrition O R I G I N A L A R T I C L E Clinical Impact of Prandial State, Exercise, and Preparation on the Equivalence of Alternative- Blood Glucose Testing DAWN M. BINA, CMT ROBYN

More information

Ultra Blue. Test Strips UNISTRIP. Charlotte, NC V2-10/25/2018 TECHNOLOGIES,LLC

Ultra Blue. Test Strips UNISTRIP. Charlotte, NC V2-10/25/2018 TECHNOLOGIES,LLC VS UNISTRIP TECHNOLOGIES,LLC Charlotte, NC 28269 1-866-889-6229 V2-10/25/2018 UniStrip Generic Blood Glucose VS OneTouch Ultra Blue 1. Purpose This study is intended to evaluate the difference in values

More information

Performance Evaluations of Continuous Glucose Monitoring Systems: Precision Absolute Relative Deviation Is Part of the Assessment

Performance Evaluations of Continuous Glucose Monitoring Systems: Precision Absolute Relative Deviation Is Part of the Assessment Journal of Diabetes Science and Technology Volume 7, Issue 4, July 2013 Diabetes Technology Society SYMPOSIUM Performance Evaluations of Continuous Glucose Monitoring Systems: Karin, B.S., Günther Schmelzeisen-Redeker,

More information

Accuracy of the 5-Day FreeStyle Navigator Continuous Glucose Monitoring System. Comparison with frequent laboratory reference measurements

Accuracy of the 5-Day FreeStyle Navigator Continuous Glucose Monitoring System. Comparison with frequent laboratory reference measurements Emerging Treatments and Technologies O R I G I N A L A R T I C L E Accuracy of the 5-Day FreeStyle Navigator Continuous Glucose Monitoring System Comparison with frequent laboratory reference measurements

More information

Hypoglycemia Detection and Prediction Using Continuous Glucose Monitoring A Study on Hypoglycemic Clamp Data

Hypoglycemia Detection and Prediction Using Continuous Glucose Monitoring A Study on Hypoglycemic Clamp Data Journal of Diabetes Science and Technology Volume 1, Issue 5, September 2007 Diabetes Technology Society SYMPOSIUM Hypoglycemia Detection and Prediction Using Continuous Glucose Monitoring A Study on Hypoglycemic

More information

Aina Blood Monitoring System

Aina Blood Monitoring System Aina Blood Monitoring System Analytical Performance Summary The Aina Blood Monitoring System is a high quality and versatile multi-parameter diagnostic platform that is CE-marked and approved for sale

More information

White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard

White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard 7500-10032 Rev F Table of Contents Executive Summary 3 Changes introduced

More information

Today s Goals 10/6/2017. New Frontiers in Diabetes Technology. Disclosures

Today s Goals 10/6/2017. New Frontiers in Diabetes Technology. Disclosures New Frontiers in Diabetes Technology Marie E. McDonnell, MD Director, Brigham and Women's Diabetes Program Division of Endocrinology, Diabetes and Hypertension Brigham and Women s Hospital Today s Goals

More information

Agreement between Glucose Trends Derived from Three Simultaneously Worn Continuous Glucose Sensors

Agreement between Glucose Trends Derived from Three Simultaneously Worn Continuous Glucose Sensors Journal of Diabetes Science and Technology Volume 2, Issue 5, September 2008 Diabetes Technology Society ORIGINAL ARTICLES Agreement between Glucose Trends Derived from Three Simultaneously Worn Continuous

More information

Determination of Ethanol in Breath and Estimation of Blood Alcohol Concentration with Alcolmeter S-D2

Determination of Ethanol in Breath and Estimation of Blood Alcohol Concentration with Alcolmeter S-D2 Determination of Ethanol in Breath and Estimation of Blood Alcohol Concentration with Alcolmeter S-D2 A.W. Jones and KÄ. Jönsson Departments of Alcohol Toxicology and Internal Medicine, University Hospital,

More information

Subjects and Methods 2017, 64 (8),

Subjects and Methods 2017, 64 (8), 2017, 64 (8), 827-832 Careful readings for a flash glucose monitoring system in nondiabetic Japanese subjects: individual differences and discrepancy in glucose concentrarion after glucose loading Keiko

More information

Diabetes II Insulin pumps; Continuous glucose monitoring system (CGMS) Ernest Asamoah, MD FACE FACP FRCP (Lond)

Diabetes II Insulin pumps; Continuous glucose monitoring system (CGMS) Ernest Asamoah, MD FACE FACP FRCP (Lond) Diabetes II Insulin pumps; Continuous glucose monitoring system (CGMS) Ernest Asamoah, MD FACE FACP FRCP (Lond) 9501366-011 20110401 Objectives Understand the need for insulin pumps and CGMS in managing

More information

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k103542 B. Purpose for Submission: Alternative blood glucose test strips for use with LifeScan OneTouch

More information

Insulin Tolerance Test Protocol - RNS Endocrinology

Insulin Tolerance Test Protocol - RNS Endocrinology Page 1 of 7 - RNS Endocrinology Test name Insulin tolerance test. Alternate test names. Related Tests. Indication(s) Investigation of the hypothalamic pituitary axis (HPA) with regard to the release of

More information

Diabetologia 9 Springer-Verlag 1984

Diabetologia 9 Springer-Verlag 1984 Diabetologia (1984) 26:203 207 Diabetologia 9 Springer-Verlag 1984 How does glucose regulate the human pancreatic A cell in vivo? C. M. Asplin*, P. M. Hollander** and J. P. Palmer Diabetes Research Center

More information

Early Patient and Clinician Experiences with Continuous Glucose Monitoring

Early Patient and Clinician Experiences with Continuous Glucose Monitoring Feature Article / Early Experiences with CGM Early Patient and Clinician Experiences with Continuous Glucose Monitoring David K. Bloomgarden, MD, FACE; Janine Freeman, RD, LD, CDE; and Elizabeth DeRobertis,

More information

Evolution of Analytical Performance in Portable Glucose Meters in the Last Decade

Evolution of Analytical Performance in Portable Glucose Meters in the Last Decade Emerging Treatments and Technologies O R I G I N A L A R T I C L E Evolution of Analytical Performance in Portable Glucose Meters in the Last Decade PHILIP BÖHME, MD MICHELE FLORIOT, MD MARIE-AUDE SIRVEAUX,

More information

Anomaly Detection in the Artificial Pancreas

Anomaly Detection in the Artificial Pancreas Anomaly Detection in the Artificial Pancreas Avila L., Martínez E. INGAR (CONICET-UTN) Abstract The integration of subcutaneous sensing and insulin delivery technologies with novel control strategies has

More information

Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System

Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Abstract Objective: Both Fora Premium V10 and V30 blood glucose

More information

The Realities of Technology in Type 1 Diabetes

The Realities of Technology in Type 1 Diabetes The Realities of Technology in Type 1 Diabetes May 6, 2017 Rosanna Fiallo-scharer, MD Margaret Frederick, RN Disclosures I have no conflicts of interest to disclose I will discuss some unapproved treatments

More information

Exenatide Treatment for 6 Months Improves Insulin Sensitivity in Adults With Type 1 Diabetes

Exenatide Treatment for 6 Months Improves Insulin Sensitivity in Adults With Type 1 Diabetes 666 Diabetes Care Volume 37, March 2014 CLIN CARE/EDUCATION/NUTRITION/PSYCHOSOCIAL Exenatide Treatment for 6 Months Improves Insulin Sensitivity in Adults With Type 1 Diabetes Gayatri Sarkar, 1 May Alattar,

More information

Simone Del Favero, PhD, Andrea Facchinetti, PhD, Giovanni Sparacino, PhD, and Claudio Cobelli, PhD

Simone Del Favero, PhD, Andrea Facchinetti, PhD, Giovanni Sparacino, PhD, and Claudio Cobelli, PhD DIA-2016-0413-ver9-Favero_3P.3d 03//17 12:40pm Page 1 DIABETES TECHNOLOGY & THERAPEUTICS Volume 19, Number 4, 2017 ª Mary Ann Liebert, Inc. DOI:.89/dia.2016.0413 ORIGINAL ARTICLE Retrofitting Real-Life

More information

Computing the Surveillance Error Grid Analysis: Procedure and Examples

Computing the Surveillance Error Grid Analysis: Procedure and Examples 5959DSTXXX.77/99685959Journal of Diabetes Science and TechnologyKovatchev et al research-article Original Article Computing the Surveillance Error Grid Analysis: Procedure and Examples Journal of Diabetes

More information

Report. Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013

Report. Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013 Institute of Diabetes Gerhardt Katsch Karlsburg (IDK) Report Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013 Sponsor: The study was

More information

Objectives. Glucose Monitoring Today. Tracking Glucose ISO Standards. Continuous Glucose Monitoring & Diabetes Management 3/7/2015

Objectives. Glucose Monitoring Today. Tracking Glucose ISO Standards. Continuous Glucose Monitoring & Diabetes Management 3/7/2015 Objectives Continuous Glucose Monitoring & Diabetes Management Tomas C. Walker, DNP, APRN, CDE Director, Clinical Projects Dexcom, Inc. San Diego, CA 1. Review current limitations of SMBG 2. Understand

More information

Better Diagnostics for Life

Better Diagnostics for Life NASDAQ: NMRD Better Diagnostics for Life Interim Clinical Presentation sugarbeat 12 th September 2018 A Prospective Single Centre Evaluation of the Accuracy and safety of the sugarbeat Non-invasive Continuous

More information

Evaluation of portable blood glucose meters using canine and feline pooled blood samples

Evaluation of portable blood glucose meters using canine and feline pooled blood samples Polish Journal of Veterinary Sciences Vol. 19, No. 4 (2016), 707 713 DOI 10.1515/pjvs-2016-0089 Original article Evaluation of portable blood glucose meters using canine and feline pooled blood samples

More information

Preventing Hypoglycemia Using Predictive Alarm Algorithms and Insulin Pump Suspension

Preventing Hypoglycemia Using Predictive Alarm Algorithms and Insulin Pump Suspension DIABETES TECHNOLOGY & THERAPEUTICS Volume 11, Number 2, 29 Mary Ann Liebert, Inc. DOI: 1.189/dia.28.32 Preventing Hypoglycemia Using Predictive Alarm Algorithms and Insulin Pump Suspension Bruce Buckingham,

More information

Efficacy/pharmacodynamics: 85 Safety: 89

Efficacy/pharmacodynamics: 85 Safety: 89 These results are supplied for informational purposes only. Prescribing decisions should be made based on the approved package insert in the country of prescription. Sponsor/Company: Sanofi Drug substance:

More information

Cross-Matches for Bioequivalence Evaluation Division using Needle Free Jet Injector (Comfort-In) and conventional Pen type syringe.

Cross-Matches for Bioequivalence Evaluation Division using Needle Free Jet Injector (Comfort-In) and conventional Pen type syringe. Cross-Matches for Bioequivalence Evaluation Division using Needle Free Jet Injector (Comfort-In) and conventional Pen type syringe. Dr. EunJig, Lee http://www.yuhs.or.kr/en/hospitals/severance/clinic_dept/endo_dept/phy_directory/docprofile.asp?sno=1734

More information

The Current Environment of CGM Technologies. Barry H. Ginsberg, M.D., Ph.D.

The Current Environment of CGM Technologies. Barry H. Ginsberg, M.D., Ph.D. Journal of Diabetes Science and Technology Volume 1, Issue 1, January 2007 Diabetes Technology Society SYMPOSIUM The Current Environment of CGM Technologies Introduction Barry H., M.D., Ph.D. This presentation

More information

Closing the loop, squaring the circle: Studies on insulin delivery, glucose monitoring and the artificial pancreas Luijf, Y.M.

Closing the loop, squaring the circle: Studies on insulin delivery, glucose monitoring and the artificial pancreas Luijf, Y.M. UvA-DARE (Digital Academic Repository) Closing the loop, squaring the circle: Studies on insulin delivery, glucose monitoring and the artificial pancreas Luijf, Y.M. Link to publication Citation for published

More information

Accuracy of the SEVEN Continuous Glucose Monitoring System: Comparison with Frequently Sampled Venous Glucose Measurements

Accuracy of the SEVEN Continuous Glucose Monitoring System: Comparison with Frequently Sampled Venous Glucose Measurements Journal of Diabetes Science and Technology Volume 3, Issue 5, September 2009 Diabetes Technology Society ORIGINAL ARTICLES Accuracy of the SEVEN Continuous Glucose Monitoring System: Howard C., M.D., 1

More information

Performance of the Medtronic Sentrino continuous glucose management (CGM) system in the cardiac intensive care unit

Performance of the Medtronic Sentrino continuous glucose management (CGM) system in the cardiac intensive care unit Open Access Research Performance of the Medtronic Sentrino continuous glucose management (CGM) system in the cardiac intensive care unit Mikhail Kosiborod, 1,2 Rebecca K Gottlieb, 3 Julie A Sekella, 3

More information

Glucagon secretion in relation to insulin sensitivity in healthy subjects

Glucagon secretion in relation to insulin sensitivity in healthy subjects Diabetologia (2006) 49: 117 122 DOI 10.1007/s00125-005-0056-8 ARTICLE B. Ahrén Glucagon secretion in relation to insulin sensitivity in healthy subjects Received: 4 July 2005 / Accepted: 12 September 2005

More information

Non-invasive blood glucose measurement by near infrared spectroscopy: Machine drift, time drift and physiological effect

Non-invasive blood glucose measurement by near infrared spectroscopy: Machine drift, time drift and physiological effect Spectroscopy 24 (2010) 629 639 629 DOI 10.3233/SPE-2010-0485 IOS Press Non-invasive blood glucose measurement by near infrared spectroscopy: Machine drift, time drift and physiological effect Simon C.H.

More information

Continuous Glucose Monitoring and the Reality of Metabolic Control in Preschool Children With Type 1 Diabetes

Continuous Glucose Monitoring and the Reality of Metabolic Control in Preschool Children With Type 1 Diabetes Emerging Treatments and Technologies O R I G I N A L A R T I C L E Continuous Glucose Monitoring and the Reality of Metabolic Control in Preschool Children With Type 1 Diabetes GEORGE S. JEHA, MD 1,2 LEFKOTHEA

More information

EVALUATION OF THE PERFORMANCE OF THE TRUETRACK BLOOD GLUCOSE MONITORING SYSTEM USING THE ISO 15197:2003 ACCURACY AND USER EVALUATION REQUIREMENTS

EVALUATION OF THE PERFORMANCE OF THE TRUETRACK BLOOD GLUCOSE MONITORING SYSTEM USING THE ISO 15197:2003 ACCURACY AND USER EVALUATION REQUIREMENTS EVALUATION OF THE PERFORMANCE OF THE TRUETRACK BLOOD GLUCOSE MONITORING SYSTEM USING THE ISO 15197:23 ACCURACY AND USER EVALUATION REQUIREMENTS Douglas E. Bell, PhD Teri A. Sasse, RN, MS BACKGROUND: The

More information

Continuous Glucose Monitoring Considerations for the Development of a Closed-Loop Artificial Pancreas System

Continuous Glucose Monitoring Considerations for the Development of a Closed-Loop Artificial Pancreas System Journal of Diabetes Science and Technology Volume 5, Issue 6, November 2011 Diabetes Technology Society SYMPOSIUM Continuous Glucose Monitoring Considerations for the Development of a Closed-Loop Artificial

More information

Screening of Elevated Glucose Levels in Gingival Crevice Blood Using a Novel, Sensitive Self-Monitoring Device

Screening of Elevated Glucose Levels in Gingival Crevice Blood Using a Novel, Sensitive Self-Monitoring Device Original Paper DOI: 10.1159/000080 474 Received: June 23, 2003 Revised: February 28, 2004 Screening of Elevated Glucose Levels in Gingival Crevice Blood Using a Novel, Sensitive Self-Monitoring Device

More information

CareLink. software REPORT REFERENCE GUIDE. Management Software for Diabetes

CareLink. software REPORT REFERENCE GUIDE. Management Software for Diabetes CareLink Management Software for Diabetes software REPORT REFERENCE GUIDE How to use this guide Each type of CareLink report and its components are described in the following sections. Report data used

More information

NICO /14 Nipro Diagnostics, Inc.

NICO /14 Nipro Diagnostics, Inc. Clinical Performance of the TRUEresult twist Blood Glucose Monitoring System () Using ISO 15197:13 Criteria for Accuracy Utilizing GDH-FAD Enzyme NICO-8 1/1 Nipro Diagnostics, Inc. NICO_8_mmol.indd CLINICAL

More information

Advances in Diabetes Care Technologies

Advances in Diabetes Care Technologies 1979 Advances in Diabetes Care Technologies 2015 Introduction Insulin pump use: ~ 20% - 30% of patients with T1DM < 1% of insulin-treated patients with T2DM 2007 FDA estimates ~375,000 insulin pumps for

More information

Clinical Implications and Economic Impact of Accuracy Differences among Commercially Available Blood Glucose Monitoring Systems

Clinical Implications and Economic Impact of Accuracy Differences among Commercially Available Blood Glucose Monitoring Systems Journal of Diabetes Science and Technology Volume 7, Issue 2, March 2013 Diabetes Technology Society ORIGINAL ARTICLE Clinical Implications and Economic Impact of Accuracy Differences Erwin S., Ph.D.,

More information

Dr Sander Zwart MD, FACP, ECNU Arizona-AACE 2016 Annual Meeting

Dr Sander Zwart MD, FACP, ECNU Arizona-AACE 2016 Annual Meeting Dr Sander Zwart MD, FACP, ECNU Arizona-AACE 2016 Annual Meeting Disclaimers None Except products and software discussed are all proprietary, so if I mention a brand product it is because there is no other

More information

Glycemic Variability:

Glycemic Variability: Glycemic Variability: Do the Differences Make a Difference? Kim L Kelly, PharmD, BCPS, FCCP Define Variability VERY LOW GLYCEMIC VARIABILITY LOW GLYCEMIC VARIABILITY HIGH GLYCEMIC VARIABILITY So what s

More information

Statistical Approach of Assessing the Reliability of Glucose Sensors: The GLYCENSIT Procedure

Statistical Approach of Assessing the Reliability of Glucose Sensors: The GLYCENSIT Procedure Journal of Diabetes Science and Technology Volume 2, Issue 6, November 2008 Diabetes Technology Society SYMPOSIUM Statistical Approach of Assessing the Reliability of Glucose Sensors: The GLYCENSIT Procedure

More information

Advances in Diabetes Care Technologies

Advances in Diabetes Care Technologies Advances in Diabetes Care Technologies 1979 2015 Introduction Roughly 20% to 30% of patients with T1DM and fewer than 1% of insulin-treated patients with T2DM use an insulin pump In 2007, the U.S. FDA

More information

GD50. Simple. Accurate. Reliable. Test Simply. Live Better. All FORA Diamond Series Blood Glucose Monitoring Systems use FORA Diamond Test Strips.

GD50. Simple. Accurate. Reliable. Test Simply. Live Better. All FORA Diamond Series Blood Glucose Monitoring Systems use FORA Diamond Test Strips. GD50 Higher Measuring Accuracy Simple. Accurate. Reliable. Test Simply. Live Better. All FORA Diamond Series Blood Glucose Monitoring Systems use FORA Diamond Test Strips. GD50 Simple. Accurate. Reliable.

More information

Report Reference Guide

Report Reference Guide Report Reference Guide How to use this guide Each type of CareLink report and its components are described in the following sections. Report data used to generate the sample reports was from sample patient

More information

Accuracy Requirements for a Hypoglycemia Detector: An Analytical Model to Evaluate the Effects of Bias, Precision, and Rate of Glucose Change

Accuracy Requirements for a Hypoglycemia Detector: An Analytical Model to Evaluate the Effects of Bias, Precision, and Rate of Glucose Change Journal of Diabetes Science and Technology Volume 1, Issue 5, September 2007 Diabetes Technology Society SYMPOSIUM Accuracy Requirements for a Hypoglycemia Detector: An Analytical Model to Evaluate the

More information

NEW TECHNOLOGIES FOR MANAGING DIABETES ANGELA THOMPSON DNP, FNP-C, BC-ADM, CDE, FAANP

NEW TECHNOLOGIES FOR MANAGING DIABETES ANGELA THOMPSON DNP, FNP-C, BC-ADM, CDE, FAANP NEW TECHNOLOGIES FOR MANAGING DIABETES ANGELA THOMPSON DNP, FNP-C, BC-ADM, CDE, FAANP No commercial support or sponsorship was received for this project I have nothing to disclose OBJECTIVES Identify at

More information

Reconstruction of Glucose in Plasma from Interstitial Fluid Continuous Glucose Monitoring Data: Role of Sensor Calibration

Reconstruction of Glucose in Plasma from Interstitial Fluid Continuous Glucose Monitoring Data: Role of Sensor Calibration Journal of Diabetes Science and Technology Volume 1, Issue 5, September 2007 Diabetes Technology Society SYMPOSIUM Reconstruction of Glucose in Plasma from Interstitial Fluid Continuous Glucose Monitoring

More information

Smart & Simple Mobile Solutions.

Smart & Simple Mobile Solutions. Higher Measuring Accuracy Smart & Simple Mobile Solutions. Test Simply. Live Better. FORA Diamond lets you connect all your Bluetooth devices. *Also available with a Micro USB. ifora Diabetes Manager All

More information

INSULIN INITIATION AND INTENSIFICATION WITH A FOCUS ON HYPOGLYCEMIA REDUCTION

INSULIN INITIATION AND INTENSIFICATION WITH A FOCUS ON HYPOGLYCEMIA REDUCTION INSULIN INITIATION AND INTENSIFICATION WITH A FOCUS ON HYPOGLYCEMIA REDUCTION Jaiwant Rangi, MD, FACE Nov 10 th 2018 DISCLOSURES Speaker Novo Nordisk Sanofi-Aventis Boheringer Ingleheim Merck Abbvie Abbott

More information

Subject Index. Breastfeeding, self-monitoring of blood glucose 56

Subject Index. Breastfeeding, self-monitoring of blood glucose 56 Subject Index Animas Vibe 86, 130 Artificial pancreas clinical studies inpatient studies 175 180 outpatient studies outcome assessment 182, 183 technology 180, 181 telemedicine 182 components glucose sensor

More information

Clinical Performance of the TRUE2go Blood Glucose Monitoring System Exceeds Minimum Criteria for Accuracy Using ISO 15197:2013

Clinical Performance of the TRUE2go Blood Glucose Monitoring System Exceeds Minimum Criteria for Accuracy Using ISO 15197:2013 linical Performance of the TRUE2go lood Glucose Monitoring System Exceeds Minimum riteria for Accuracy Using ISO 197:213 LINIAL STUY SUMMARY EVALUATION OJETIVES This clinical evaluation: 1. Evaluates the

More information

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: k100322 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE B. Purpose for Submission: Clearance of a new device C. Measurand: Whole

More information

Performance of three portable blood glucose monitoring devices used in a veterinary application

Performance of three portable blood glucose monitoring devices used in a veterinary application Performance of three portable blood glucose monitoring devices used in a veterinary application Prepared by Michael I. Lindinger, PhD, Vice-President of Research, The Nutraceutical Alliance, Campbellville,

More information

The Benefits of Continuous Glucose Monitoring and a Glucose Monitoring Schedule in Individuals with Type 1 Diabetes during Recreational Diving

The Benefits of Continuous Glucose Monitoring and a Glucose Monitoring Schedule in Individuals with Type 1 Diabetes during Recreational Diving Journal of Diabetes Science and Technology Volume 2, Issue 5, September 2008 Diabetes Technology Society ORIGINAL ARTICLES The Benefits of Continuous Glucose Monitoring and a Glucose Monitoring Schedule

More information

Imaging Core Laboratory Standard Operating Procedure

Imaging Core Laboratory Standard Operating Procedure Imaging Core Laboratory Standard Operating Procedure Patient Preparation and FDG Administration Procedure for ACRIN FDG-PET Scans 1.0 PURPOSE To outline the procedure for patient preparation and for 18

More information

Since the discovery of insulin, the management of type 1 diabetes

Since the discovery of insulin, the management of type 1 diabetes SPECIAL Clinical FEATURE Review Realistic Expectations and Practical Use of Continuous Glucose Monitoring for the Endocrinologist Irl B. Hirsch University of Washington School of Medicine, Seattle, Washington

More information

Learning Objectives. At the conclusion of this module, participants should be better able to:

Learning Objectives. At the conclusion of this module, participants should be better able to: Learning Objectives At the conclusion of this module, participants should be better able to: Treat asymptomatic neonatal hypoglycemia with buccal dextrose gel Develop patient-specific approaches to intravenous

More information

Pumps & Sensors made easy. OPADA ALZOHAILI MD FACE Endocrinology Assistant Professor Wayne State University

Pumps & Sensors made easy. OPADA ALZOHAILI MD FACE Endocrinology Assistant Professor Wayne State University Pumps & Sensors made easy OPADA ALZOHAILI MD FACE Endocrinology Assistant Professor Wayne State University DeFronzo RA. Diabetes. 2009;58:773-795. Ominous Octet Relationship of b-cell Dysfunction and Development

More information

Blood Glucose Monitoring System. Owner s Booklet

Blood Glucose Monitoring System. Owner s Booklet Blood Glucose Monitoring System Owner s Booklet The FreeStyle Freedom Lite Meter and How It Works System Check Screen This display always appears when the meter is turned on. You should check that your

More information

Advances Towards the Bionic Pancreas.

Advances Towards the Bionic Pancreas. Advances Towards the Bionic Pancreas. Ron Brazg MD, FACE Rainier Clinical Research Center Renton, WA DISCLOSURES Rainier Clinical Research Center is an independent research facility not directly affiliated

More information

Diabetes: Definition Pathophysiology Treatment Goals. By Scott Magee, MD, FACE

Diabetes: Definition Pathophysiology Treatment Goals. By Scott Magee, MD, FACE Diabetes: Definition Pathophysiology Treatment Goals By Scott Magee, MD, FACE Disclosures No disclosures to report Definition of Diabetes Mellitus Diabetes Mellitus comprises a group of disorders characterized

More information

Advances in Diabetes Care Technologies

Advances in Diabetes Care Technologies 1979 Advances in Diabetes Care Technologies 2015 Introduction Roughly 20% - 30% of patients with T1DM and fewer than 1% of insulin-treated patients with T2DM use an insulin pump In 2007, the US FDA estimated

More information

Evaluation of the Analytical Performance of the Coulometry-Based Optium Omega Blood Glucose Meter

Evaluation of the Analytical Performance of the Coulometry-Based Optium Omega Blood Glucose Meter Journal of Diabetes Science and Technology Volume 5, Issue 6, November 2011 Diabetes Technology Society TECHNOLOGY REPORT Evaluation of the Analytical Performance of the Coulometry-Based Optium Omega Blood

More information

Norbert Hermanns, PhD 1,2, Beatrix Schumann, MD 2, Bernhard Kulzer, PhD 1,2, and Thomas Haak, MD 1,2. Original Article

Norbert Hermanns, PhD 1,2, Beatrix Schumann, MD 2, Bernhard Kulzer, PhD 1,2, and Thomas Haak, MD 1,2. Original Article 524105DSTXXX10.1177/1932296814524105Journal of Diabetes Science and TechnologyHermanns et al research-article2014 Original Article The Impact of Continuous Glucose Monitoring on Low Interstitial Glucose

More information

Updates in Diabetes Technology

Updates in Diabetes Technology Updates in Diabetes Technology Jessica Kirk, MSN, RN, CPN, CDE Nurse Manager, Endo ECHO No disclosures Disclosures 1 Objectives Distinguish patients appropriate for continuous glucose monitoring and insulin

More information

The Commission on Dental Accreditation s

The Commission on Dental Accreditation s Teaching Casual Random Blood Glucose Screening to Second-Year Dental Students Thomas W. Radmer, D.D.S., M.S.; Moawia M. Kassab, D.D.S., M.S.; Denis P. Lynch, D.D.S., Ph.D.; Martin Walsh, B.S. Abstract:

More information

Somatostatin is an endogenous peptide and neurotransmitter

Somatostatin is an endogenous peptide and neurotransmitter Rapid Publication Somatostatin Impairs Clearance of Exogenous Insulin in Humans ELI IPP, YSEF SINAI, BENJAMIN BAR-Z, RAFAEL NESHER, AND ERL CERASI SUMMARY Somatostatin has been widely used to suppress

More information

Figure 2.1: Glucose meter

Figure 2.1: Glucose meter CHAPTER TWO: MONITORING TECHNOLOGIES 2.1 Introduction Glucose monitoring is a method of self-testing glucose (blood sugar) levels for the management of diabetes. Traditionally, it involves pricking the

More information

Performance Evaluation of B. Braun Omnitest 5 Blood Glucose Monitoring System for Self-Monitoring of Blood Glucose

Performance Evaluation of B. Braun Omnitest 5 Blood Glucose Monitoring System for Self-Monitoring of Blood Glucose ORIGINL RTILE https://doi.org/10.15263/jlmqa.2016.38.4.234 Performance Evaluation of. raun Omnitest 5 lood Glucose Monitoring System for Self-Monitoring of lood Glucose Kyungso Jeon 1 and Miseon Shin 2

More information

Quality Control of Self-Monitoring of Blood Glucose: Why and How?

Quality Control of Self-Monitoring of Blood Glucose: Why and How? Journal of Diabetes Science and Technology Volume 1, Issue 2, March 2007 Diabetes Technology Society SYMPOSIUM Quality Control of Self-Monitoring of Blood Glucose: Why and How? Bogdan, M.D., Ph.D. and

More information

Accuracy Evaluation of Five Blood Glucose Monitoring Systems: The North American Comparator Trial

Accuracy Evaluation of Five Blood Glucose Monitoring Systems: The North American Comparator Trial Journal of Diabetes Science and Technology Volume 7, Issue 5, September 2013 Diabetes Technology Society ORIGINAL ARTICLE Accuracy Evaluation of Five Blood Glucose Monitoring Systems: The North American

More information

Glucose, glucose oxidase/peroxidase method, photometry, diabetes mellitus.

Glucose, glucose oxidase/peroxidase method, photometry, diabetes mellitus. Quantitative Determination TEAS Related topics Glucose, glucose oxidase/peroxidase method, photometry, diabetes mellitus. Principle Blood sugar is the key energy provider for our cells. The primary hormone

More information