Nocturnal hypoglycemia is frequent in patients with

Size: px
Start display at page:

Download "Nocturnal hypoglycemia is frequent in patients with"

Transcription

1 DIABETES TECHNOLOGY & THERAPEUTICS Volume 15, Number 5, 2013 ª Mary Ann Liebert, Inc. DOI: /dia ORIGINAL ARTICLE Nocturnal Continuous Glucose Monitoring: Accuracy and Reliability of Hypoglycemia Detection in Patients with Type 1 Diabetes at High Risk of Severe Hypoglycemia Christiane Bay, 1,2 Peter Lommer Kristensen, MD, PhD, 1,3 Ulrik Pedersen-Bjergaard, MD, DMSc, 1 Lise Tarnow, MD, DMSc, 4,5 and Birger Thorsteinsson, MD, DMSc 1,2 Abstract Background: A reliable method to detect biochemical nocturnal hypoglycemia is highly needed, especially in patients with recurrent severe hypoglycemia. We evaluated reliability of nocturnal continuous glucose monitoring (CGM) in patients with type 1 diabetes at high risk of severe hypoglycemia. Patients and Methods: Seventy-two type 1 diabetes patients with recurrent severe hypoglycemia (two or more events within the last year) participated for 4 nights in blinded CGM recordings (Guardian Ò REAL-Time CGMS and Sof-Sensor Ò ; Medtronic MiniMed, Northridge, CA). Blood was drawn hourly from 23:00 to 07:00 h for plasma glucose (PG) measurements (gold standard). Results: Valid data were obtained in 217 nights. The sensitivity of CGM was 65% (95% confidence interval, 53 77%) below 4 mmol/l, 40% (24 56%) below 3 mmol/l, and 17% (0 47%) below 2.2 mmol/l. PG and CGM readings correlated in the total measurement range (Spearman s q = 0.82; P < 0.001). In the normo- and hyperglycemic ranges CGM underestimated PG by 1.1 mmol/l ( mmol/l) (P < 0.001); in contrast, in the hypoglycemic range (PG < 4 mmol/l) CGM overestimated PG levels by 1.0 mmol/l (P < 0.001). The mean absolute relative differences in the hypo- ( 3.9 mmol/l), normo- (4 9.9 mmol/l), and hyperglycemic ( 10 mmol/l) ranges were 45% (37 53%), 23% (22 25%), and 20% (19 21%), respectively. Continuous glucose error grid analysis indicated a clinical accuracy of 56%, 99%, and 93% in the hypo-, normo-, and hyperglycemic ranges, respectively. Conclusions: The accuracy in the hypoglycemic range of nocturnal CGM data using Sof-Sensor is suboptimal in type 1 diabetes patients at high risk of severe hypoglycemia. To ensure clinical useful sensitivity in detection of nocturnal hypoglycemic episodes, an alarm threshold should not be lower than 4 mmol/l. Introduction Nocturnal hypoglycemia is frequent in patients with insulin-treated diabetes and asymptomatic in most cases. 1,2 It may have negative effects on health and well-being and even, although rarely, cause sudden death. Patients with impaired hypoglycemia awareness and recurrent severe hypoglycemia are at particular high risk of experiencing symptomatic as well as asymptomatic episodes of nocturnal hypoglycemia. 2 Continuous glucose monitoring (CGM) provides a method to obtain continuous glucose profiles and has proven helpful in detecting asymptomatic nocturnal hypoglycemia. 3 However, the accuracy of CGM has been questioned as nighttime hypoglycemia has been reported to be spurious when based on CGM recordings as the accuracy and precision of the sensor are less than expected. 4 In order to alarm the patient during impending nocturnal hypoglycemia, it is crucial that real-time CGM sensor readings reflect the 1 Endocrinology Section, Department of Cardiology, Nephrology and Endocrinology, Hillerød Hospital, Hillerød, Denmark. 2 Faculty of Health and Medical Sciences, University of Copenhagen, Copenhagen, Denmark. 3 Department of Endocrinology, Herlev University Hospital, Herlev, Denmark. 4 Steno Diabetes Center, Gentofte, Denmark. 5 Health, University of Aarhus, Aarhus, Denmark. This study is registered at with clinical trial identification number NCT An abstract of this work was presented at the 48 th Annual Meeting of the European Association for the Study of Diabetes,

2 372 BAY ET AL. actual blood glucose level. Therefore we addressed the issue of the reliability of nocturnal CGM in a cohort of type 1 diabetes patients with recurrent severe hypoglycemia. This is the subgroup of patients with the greatest potential benefit from a real-time CGM-based hypoglycemia alarm. There are no previous studies of hypoglycemic CGM reliability in this patient population. Preliminary results of this study have been presented elsewhere in abstract form. 5 Subjects and Methods Study design The present study is part of an investigator-initiated, prospective, randomized, open, blinded endpoint (PROBE) multicenter, 2-year crossover trial (HypoAna) designed to investigate the effect of basal-bolus therapy with subcutaneous injection of insulin analogs (insulin detemir/insulin aspart) versus human NPH insulin/soluble insulin on the frequency of severe hypoglycemia in high-risk patients (i.e., patients with type 1 diabetes and at least two episodes of severe hypoglycemia during the year before the study). 6 The CGM study comprised two overnight stays in each HypoAna treatment arm (at 6, 12, 18, and 24 months) at the inpatient facilities at Steno Diabetes Center. A blinded subcutaneous CGM system was used (Guardian Ò REAL-Time CGMS using Sof-sensor Ò MMT-7020C; Medtronic MiniMed, Northridge, CA). Written informed consent to participate was obtained from all patients. The study was approved by The Regional Committee on Biomedical Research Ethics (protocol number H-KA ). Subjects All adult (>18 years of age) patients diagnosed with type 1 diabetes mellitus and followed up at six Danish outpatient clinics in were screened to identify candidates for participation in the HypoAna trial. In total, 3,861 subjects completed a given or mailed questionnaire about hypoglycemia as previously reported. 7 The inclusion criteria for the HypoAna trial were type 1 diabetes for more than 5 years, age >18 years, basal-bolus treatment with human insulin or insulin analogs, a negative pregnancy test, and at least two episodes of severe hypoglycemia (defined by need for assistance from another person to treat the episode) in the previous year. 6 The most important exclusion criteria were adrenal insufficiency, growth hormone deficiency, untreated myxedema, and recent history of major cardiovascular events. In total, 159 subjects fulfilling these criteria were recruited to the HypoAna study. 6 A subgroup of 110 of these patients who were followed at Steno Diabetes Center or Hillerød Hospital was invited to participate in the study of CGM performance during nighttime. Seventy-six patients agreed to participate, and 72 patients completed the study. Four patients had to be excluded (three did not return, and one was excluded because of technical failures of the CGM device during the first visit and because of the patient s failing to return at the subsequent three scheduled visits). Data on patients history of diabetes were extracted from medical records. Previous experience of hypoglycemia was assessed in the questionnaire, and the patient s state of hypoglycemia awareness was classified according to a validated method based on the patient s self-reported ability to recognize symptoms during a hypoglycemic episode. 8 Subjects who reported always being able to recognize symptoms of hypoglycemia were classified as aware, those usually recognizing symptoms were classified as having impaired awareness, and patients occasionally or never recognizing symptoms were classified as being unaware of hypoglycemia. Study setting Patients arrived at the outpatient clinic at Steno Diabetes Center at 18:00 h for dinner and were observed continuously for clinical bedside signs of hypoglycemia during the night. In case of confirmed hypoglycemia the event was treated according to standard clinical instructions. The CGM device was mounted after dinner. The device consists of (1) a subcutaneous glucose oxidase sensor that converts interstitial glucose to an electrical current, (2) a radiofrequency transmitter connected to the sensor that wirelessly transmits the electrical information to a monitor every 5 min, and (3) a monitor that stores and displays the glucose data. 9 A glucokinase-based amperometric minimally invasive glucose sensor estimates plasma glucose (PG) based on the interstitial glucose concentration in the subcutaneous tissue. At bedtime and at 2:00 h the CGM device was calibrated with fingerstick meter blood glucose measurements performed according to the recommendations from the manufacturer and supervised by a diabetes specialist nurse. Hypoglycemic glucose values were not used for calibration. In the case of hypoglycemia at the planned time of calibration, hypoglycemia was treated according to clinical recommendations, and calibration was postponed until normalization of the glucose level. The CGM device was from the manufacturer dedicated to the study by blinding the display, and the alarm functions were switched off. Venous blood samples were drawn every hour from 23:00 to 07:00 h from an intravenous line inserted into an antecubital vein. As the reference method (gold standard) we determined PG on a Roche Hitachi 912 chemistry analyzer (Roche Diagnostics GmbH, Mannheim, Germany) the following day in an accredited laboratory. Plasma was stored at - 20 C until analysis. End points Hypoglycemic episodes were classified according to three glucose thresholds: < 4mmol/L(thresholdindicatedbythe American Diabetes Association and the European Medicines Agency), < 3 mmol/l (the commonly used threshold), and < 2.2 mmol/l (detection limit of the CGMS). According to the CGM hypoglycemia definition used by the United Kingdom Hypoglycaemia Study Group, 10 avalidepisodeof nocturnal hypoglycemia lasted at least 20 min, and recovery was considered to have occurred once the CGM glucose level remained above the defined threshold for a further 20 min. Statistics Different methods for detection of agreement between PG and CGM measurements were used: 1. Sensitivity and specificity were calculated to evaluate the ability of detecting hypoglycemic nights by CGM:

3 NOCTURNAL CONTINUOUS GLUCOSE MONITORING 373 sensitivity = true positive/(true positive + false negative) and specificity = true negative/(true negative + false positive) according to the three thresholds for hypoglycemia. 2. Spearman s correlation coefficient q between PG and CGM glucose values was determined by linear regression analysis. 3. Bland Altman analysis was done with PG as the reference. 4. The mean difference (MD) (MD = PG CGM glucose value), mean absolute difference (MAD) (MAD = jmdj), and mean absolute relative difference (MARD) (MARD = MAD/PG) were calculated for all paired measurements and for glucose ranges of hypoglycemia (< 4 mmol/l), normoglycemia (4 9.9 mmol/l), and hyperglycemia ( 10 mmol/l). 5. Continuous glucose error grid analysis (CG-EGA) (software from The Epsilon Group, Charlottesville, VA) was used to assess the clinical accuracy of the CGM sensors. The interdependency of successive data was taken into account by combining point accuracy (P- EGA) with rate accuracy (R-EGA). P-EGA reflects the difference between two paired samples at one point in time, and R-EGA is defined as the difference in glucose level between two consecutive measurements divided by the time interval between these measurements, which indicates the ability of CGM to follow the direction and tempo of the PG changes. In the CG-EGA matrix the CGM readings and corresponding PG values are divided into five zones (A E) according to accuracy regarding PG changes. Zone A represents rates of change in CGM readings similar to PG rates, points in Zones B D represent rates of change with increasing inaccuracy, and points in Zone E represent readings with a rate opposite to the PG rate. CGM readings are considered clinically accurate when falling into Zones A and B of both the P-EGA and R-EGA. Clinically benign errors are defined as those that are unlikely to result in clinically negative therapeutic consequences (i.e., points with acceptable P-EGA) (Zones A or B) but significant errors in R-EGA (Zones C E). Clinically significant rate errors are those in Zones C E on R-EGA, which could lead to serious decision errors in management. Points in these zones could prompt decisions that could result in PG levels outside the target range. Zone C (upper and lower) represents errors that would result in overcorrecting acceptable PG levels, Zone D (upper and lower) represents failure to detect hypoglycaemia, and Zone E (upper and lower) represents erroneous readings (i.e., inverse treatment should be administered 11,12 ). 6. A Kaplan Meier estimation of the cumulative probability of hypoglycemia-free nighttime according to PG was made. Data from the CGM device were obtained from the files generated by the Medtronic MiniMed software. The distribution of data was tested with the Kolmogorov Smirnov test, and data were compared using paired t tests or standard nonparametric tests as appropriate (Wilcoxon signed ranks test). Continuous variables were expressed as mean 1 SD or median (range) if the variables were not normally distributed. One-way analysis of variance and multiple comparison tests applying Bonferroni s correction were conducted in order to test differences in MARD and MAD and calculated on logarithmic values; MD could not be analyzed in this way because of the occurrence of negative values. The level of statistical significance is P < 0.05 (two-sided). Calculations were performed using statistical software (SPSS version for the Macintosh; SPSS Inc., Chicago, IL). Results In total, 217 nights with valid data was obtained in the completing 72 patients. Patient characteristics are shown in Table 1. Eighteen nights were missing because of CGM failure to log data, and 53 nights were missing because of patients failing to return for the scheduled visits. All participants were sleeping with the device during venous blood sampling. No device-related adverse events were reported. Occurrence of nocturnal hypoglycemia Prevalences of episodes of nocturnal hypoglycemia according to simultaneously measured PG and CGM values and the three thresholds are shown in Table 2. An analysis including all CGM glucose values measured 15 min before and after hypoglycemic events detected by PG did not change the CGM hypoglycemia prevalence results (data not shown). Fifty-three percent of the patients experienced at least one hypoglycemic night with PG < 4 mmol/l, 25% were hypoglycemic with PG < 3 mmol/l, and 8% of patients experienced at least one hypoglycemic night with PG < 2.2 mmol/l. Twenty-eight percent of the nights were hypoglycemic (PG < 4 mmol/l). The largest risk of experiencing a hypoglycemic episode existed at bedtime. Thus, patients went to bed at 23:00 h being hypoglycemic in 18 (8%) of the nights, whereas the occurrence of hypoglycemia was almost evenly distributed throughout the rest of the night. The duration of nocturnal hypoglycemia (PG < 4 mmol/l) was skewed, with a median duration of hypoglycemia at 2 h and a mean duration of 2.7 h. Table 1. Characteristics of 72 Adult Patients with Type 1 Diabetes and Recurrent Severe Hypoglycemia Characteristic Value Duration of diabetes (years) [median (range)] 31 (5 64) Gender (% females) 36 BMI (kg/m 2 ) [mean (SD)] 25.0 (3.1) Age (years) [median (range)] 55 (24 77) Age at onset of diabetes (years) 25 (1 54) [median (range)] HbA1c (%) [mean (SD)] 8.1 (1.1) C-peptide (% negative) 82 Severe hypoglycemia, episodes 2 (2 20) per patient per year [median (range)] Hypoglycemia awareness (%) Normal 1 Impaired 39 Unaware 60 BMI, body mass index; HbA1c, glycosylated hemoglobin.

4 374 BAY ET AL. Table 2. Clinical Performance of the Continuous Glucose Monitoring System According to Three Thresholds for Hypoglycemia Hypoglycemia threshold < 4 mmol/l < 3 mmol/l < 2.2 mmol/l Prevalence of hypoglycemia according to (nights, %) PG 60 (28) 35 (16) 6 (3) CGMS 71 (33) 25 (12) 7 (3) Sensitivity (%, 95% CI) 65 (53 77) 40 (24 56) 17 (0 47) Specificity (%, 95% CI) 80 (73 86) 94 (91 97) 97 (95 99) Performance of CGMS versus PG Correct detection of hypoglycemia (% nights) Failure to detect hypoglycemia (% nights) False alert rate (% nights) Correct identification of absence of hypoglycemia (% nights) Plasma glucose (PG) was considered the gold standard. CGMS, continuous glucose monitoring system. Performance of CGM In total, 1,786 PG values were correlated with simultaneous CGM measurements (slope = 0.76, intercept = 1.2 mmol/l; P < 0.001). At 167 hourly time points of simultaneous measurement either a PG or a CGM value was missing (26 PG samples and 141 CGM readings). The Spearman s rank correlation coefficient was 0.82 (P < 0.001) over the total range of measured values but only 0.32 (P < 0.001) in the hypoglycemic range (< 4 mmol/l). The Bland Altman analysis showed that CGM underestimates PG at 67% of the readings in the total measurement range ( mmol/l), corresponding to a mean difference of mmol/l (95% confidence interval, ) (P < 0.001) (Fig. 1). In contrast, at PG < 4 mmol/l CGM overestimates PG levels by mmol/l (P < 0.001). MAD was lower at hypoglycemia than at normoglycemia (P < 0.03) and hyperglycemia (P < 0.001) (Table 3). MARD was 45% in the hypoglycemic range, which was higher than in the normo- or hyperglycemic ranges (both P < 0.001) (Table 3). CG-EGA showed an overall clinical accuracy of 56.4% (Zones A and B) in the hypoglycemic range (PG < 4 mmol/l), 98.6% in the normoglycemic range (PG 4 10 mmol/l), and 92.8% in the hyperglycemic range (PG > 10 mmol/l) (Fig. 2). The sensitivity of CGM according to the three thresholds was 65%, 40%, and 17% with a specificity of 80%, 94%, and 97%, respectively (Table 2). Discussion We assessed the clinical reliability of nocturnal CGM using the REAL-time CGMS with the Sof-Sensor in patients with type 1 diabetes at high risk of severe hypoglycemia and found that in contrast to at normo- or hyperglycemia, the accuracy of CGM was suboptimal in the hypoglycemic range. Only the highest tested threshold of 4 mmol/l resulted in clinically useful sensitivity in detection of nocturnal hypoglycemia. The frequency of nocturnal hypoglycemia in our cohort, whether defined as < 4, < 3, or < 2.2 mmol/l, corresponds well with reported results In these studies blood glucose levels were also measured hourly during the night in patients with type 1 diabetes, but their previous experience with severe hypoglycemia and their state of hypoglycemia awareness are not stated. 13,14 Although the overall regression analysis showed good agreement between PG and CGM, the clinical accuracy of CGM in the hypoglycemic range was less good. CGM overestimated PG by 1 mmol/l in the hypoglycemic range < 4 mmol/l. In accordance with this finding, a recent study reported that CGM overestimates glucose levels by mmol/l in the hypoglycemic range < 5 mmol/l. 16 This difference may have major clinical implications as CGM has become an increasingly common method for evaluation of overnight insulin requirements and an important tool for discovering otherwise undetected nighttime hypoglycemia. In accordance with previous studies, 17 we found that MARD in the hypoglycemic range was higher than in the total glycemic range. Zijlstra et al. 18 recently also reported that CGM was least accurate in the hypoglycemic range. In their study, including 2,317 paired data in 18 type 1 diabetes patients, CGM failed to detect more than half of the true hypoglycemic events (CGM < 3.9 mmol/l). Both MARD and results of the Clarke error grid analysis were also comparable to our results. 18 The clinical accuracy of CGM has been investigated in various clinical settings using various sensors, making comparisons difficult. 4,9,17 22 Because no consensus exists concerning reporting of the reliability of CGM, we choose to report several parameters of CGM performance. It should be emphasized that outcomes of testing CGM are highly dependent on the setting. 23 Results may vary between subjects with small oscillations in blood glucose levels to type 1 diabetes patients with much more fluctuating blood glucose levels. 19 Our patients probably represent the subset with the greatest glycemic instability. Accuracy also depends on the number of days on CGM. Thus, Day 1 of sensor wear, which was the day of interest in this study, has higher MARD than other days. 20,21 Moreover, the number of blood glucose calibrations and the equipment used (i.e., a home blood glucose meter versus a laboratory analyzer) influences accuracy. 23 Finally, but most importantly, accuracy metrics are highly dependent on variation of the blood glucose level, primarily the relative proportion of low glucose concentrations, and consistently demonstrate less reliability in the hypoglycemic range. 17,18,21,22 The major strength of our study is that we challenged the CGM device maximally by using it only single nights, with

5 NOCTURNAL CONTINUOUS GLUCOSE MONITORING 375 FIG. 1. Bland Altman plot enabling detection of a tendency for the variation of continuous glucose monitoring (CGM) glucose to change with the level of plasma glucose. Differences between plasma glucose and CGM glucose are plotted against the mean of both methods. Limits of agreement are indicated, and the horizontal dashed line Mean illustrates the mean difference. (Color graphics available online at two calibrations on a home blood glucose meter, in type 1 diabetes subjects with fluctuating blood glucose levels even at night. This may have underestimated the reliability of the CGM. Nonetheless, our results are comparable to those of the study by Zijlstra et al., 18 even though they excluded type 1 diabetes patients with recurrent severe hypoglycemia (more than one episode within 12 months) and performed their study in-house for 9 2 days with calibration using a laboratory glucose analyzer every 4 h. Furthermore, our study is the first to be conducted on type 1 diabetes patients at particularly high risk of severe hypoglycemia who might be those benefitting the most from sleeping with real-time CGM. There are limitations in our study. First, because of technical difficulties complete CGM glucose readings were lacking in 8% of the nights (calibration errors, lack of signal between monitor caused by either an improper connection between the transmitter and the sensor, lost sensor alert or weak signal alert). Previous studies have reported rates of sensor failure or malfunction up to 18%. 24 Second, noncompliance during the last visits may indicate that the study setting was too inconvenient or time-consuming for some of the patients. Finally, we only had PG samples drawn hourly to minimize disturbance of the patient s sleep, which we compared with the corresponding single CGM glucose value. CGM might have Table 3. Continuous Glucose Monitoring System Sensor Accuracy by Mean and Mean Absolute Relative Differences at Hypo-, Normo-, and Hyperglycemic Ranges MD (SD) 95% CI MAD (SD) 95% CI MARD (SD) 95% CI MARD within 20% Hypoglycemia ( 3.9 mmol/l) (1.7) [ - 1.3; - 0.7] 1.3 (1.5) [1.1; 1.6] 45 (51) [37; 53] 34 Normoglycemia (4 9.9 mmol/l) 0.6 (2.0) [0.4; 0.7] 1.6 (1.4) [1.5; 1.7] 23 (21) [22; 25] 53 Hyperglycemia ( 10 mmol/l) 2.0 (2.9) [1.8; 2.2] 2.7 (2.2) [2.6; 2.9] 20 (15) [19; 21] 59 CI, confidence interval; MAD, mean absolute difference (mmol/l); MARD, mean absolute relative difference (%); MD, mean difference (mmol/l).

6 376 BAY ET AL. FIG. 2. glucose. Continuous glucose error grid analysis combining rate and point accuracy stratified by glucose range. PG, plasma detected hypoglycemia a few minutes before or after this single point of comparison. However, an analysis including all CGM glucose values 15 min before and after hypoglycemic events missed by CGM did not change the hypoglycemia prevalence results. In conclusion, the reliability of CGM in patients with type 1 diabetes at high risk of severe hypoglycemia is comparable to that previously reported in patients without this problem. This implies rather low accuracy and overestimation of glucose levels in the hypoglycemic range. Absolute values of CGM recordings of nighttime hypoglycemia should therefore be interpreted with caution in clinical practice. CGM detected the majority of the nocturnal hypoglycemic episodes < 4 mmol/l as indicated by our data. The point accuracy in the hypoglycemic range was suboptimal, resulting in the need for a relatively high threshold of 4 mmol/l or above to obtain a clinically useful event sensitivity in detection of nocturnal hypoglycemia, at the expense of specificity. Improved predictive algorithms in the newer generations of CGM may compensate for some of the lacking point accuracy and thereby improve event accuracy. CGM may still be useful as a research tool to document nocturnal hypoglycemia in a much more detailed manner than PG. However, firm definitions of hypoglycemia in terms of threshold and duration are important. Continued efforts to refine the CGM technology with more sophisticated predictive algorithms and improved sensor performance as reported for the Veo calibration algorithm 25 and the Enlite glucose sensor 21,22 are needed. Studies specifically addressing the reliability of the new sensor generations in the most likely candidates for chronic sensor use, namely those prone to severe hypoglycemia, are warranted using standardized study designs and metrics to allow comparisons of different CGM systems. Acknowledgments C.B. received research grant support from The Research Foundation of Hillerød Hospital. The staff at the outpatient diabetes clinic, Endocrinology Section, Hillerød Hospital, and at Steno Diabetes Center, Gentofte, are acknowledged for their cooperation during the patient recruitment phase and for careful handling of patients and data. Medtronic Denmark A/ S provided Guardian REAL-Time continuous glucose monitors for the study. Author Disclosure Statement U.P.-B. has received lecture honoraria from Medtronic Denmark A/S but has not received reimbursements, fees, funding, or salary from any organization relating to the content or the preparation of this manuscript. C.B., P.L.K., L.T., and B.T. declare that they have no competing financial or other interests related to the preparation or the content of the manuscript. References 1. Gale EA, Tattersall RB: Unrecognised nocturnal hypoglycaemia in insulin-treated diabetics. Lancet 1979;1: Jauch-Chara K, Schultes B: Sleep and the response to hypoglycaemia. Best Pract Res Clin Endocrinol Metab 2010; 24: Pickup JC, Freeman SC, Sutton AJ: Glycaemic control in type 1 diabetes during real time continuous glucose monitoring compared with self monitoring of blood glucose: metaanalysis of randomised controlled trials using individual patient data. BMJ 2011;343:d McGowan K, Thomas W, Moran A: Spurious reporting of nocturnal hypoglycemia by CGMS in patients with tightly controlled type 1 diabetes. Diabetes Care 2002;25: Bay C, Kristensen PL, Pedersen-Bjergaard U, Tarnow L, Thorsteinsson B: Accuracy of nocturnal Continuous Glucose Monitoring System (CGMS) in patients with type 1 diabetes at high risk of severe hypoglycaemia [abstract]. Diabetologia 2012;55(Suppl 1):S Kristensen PL, Pedersen-Bjergaard U, Beck-Nielsen H, Norgaard K, Perrild H, Christiansen JS, Jensen T, Parving HH,

7 NOCTURNAL CONTINUOUS GLUCOSE MONITORING 377 Thorsteinsson B, Tarnow L: A prospective randomised cross-over study of the effect of insulin analogues and human insulin on the frequency of severe hypoglycaemia in patients with type 1 diabetes and recurrent hypoglycaemia (the HypoAna trial): study rationale and design. BMC Endocr Disord 2012;12: Kristensen PL, Hansen LS, Jespersen MJ, Pedersen-Bjergaard U, Beck-Nielsen H, Christiansen JS, Nørgaard K, Perrild H, Parving HH, Thorsteinsson B, Tarnow L: Insulin analogues and severe hypoglycaemia in type 1 diabetes. Diabetes Res Clin Pract 2012;96: Pedersen-Bjergaard U, Pramming S, Thorsteinsson B: Recall of severe hypoglycaemia and self-estimated state of awareness in type 1 diabetes. Diabetes Metab Res Rev 2003;19: Bode B, Gross K, Rikalo N, Schwartz S, Wahl T, Page C, Gross T, Mastrototaro J: Alarms based on real-time sensor glucose values alert patients to hypo- and hyperglycemia: the Guardian continuous monitoring system. Diabetes Technol Ther 2004;6: UK Hypoglycaemia Study Group: Risk of hypoglycaemia in types 1 and 2 diabetes: effects of treatment modalities and their duration. Diabetologia 2007;50: Kovatchev BP, Gonder-Frederick LA, Cox DJ, Clarke WL: Evaluating the accuracy of continuous glucose-monitoring sensors: continuous glucose-error grid analysis illustrated by TheraSense Freestyle Navigator data. Diabetes Care 2004;27: Clarke WL: The original Clarke Error Grid Analysis (EGA). Diabetes Technol Ther 2005;7: Pramming S, Thorsteinsson B, Bendtson I, Ronn B, Binder C: Nocturnal hypoglycaemia in patients receiving conventional treatment with insulin. BMJ 1985;291: Bendtson I, Kverneland A, Pramming S, Binder C: Incidence of nocturnal hypoglycaemia in insulin-dependent diabetic patients on intensive therapy. Acta Med Scand 1988;223: Wentholt IM, Maran A, Masurel N, Heine RJ, Hoekstra JB, DeVries JH: Nocturnal hypoglycaemia in Type 1 diabetic patients, assessed with continuous glucose monitoring: frequency, duration and associations. Diabet Med 2007;24: Davey RJ, Stevens K, Jones TW, Fournier PA: The effect of short-term use of the Guardian RT continuous glucose monitoring system on fear of hypoglycaemia in patients with type 1 diabetes mellitus. Prim Care Diabetes 2012;6: Mastrototaro J, Shin J, Marcus A, Sulur G: The accuracy and efficacy of real-time continuous glucose monitoring sensor in patients with type 1 diabetes. Diabetes Technol Ther 2008;10: Zijlstra E, Heise T, Nosek L, Heinemann L, Heckermann S: Continuous glucose monitoring: quality of hypoglycaemia detection. Diabetes Obes Metab 2012;53: Gross TM, Bode BW, Einhorn D, Kayne DM, Reed JH, White NH, Mastrototaro JJ: Performance evaluation of the Mini- Med continuous glucose monitoring system during patient home use. Diabetes Technol Ther 2000;2: Peoples T, Bailey T, Ronald B, Zisser HC, Janowski B, Huang S, Talbot C, Yang Q: Accuracy performance of the Medtronic NexSensor for 6 days in an inpatient setting using abdomen and buttocks insertion sites. J Diabetes Sci Technol 2011; 5: Lee S, Bailey T, Brazg R, Christiansen M, Ahmann A, Henry R, Garg S, Watkins E, Kaufman F: Accuracy and acceptability of the 6-day Enlite continuous subcutaneous glucose sensor [abstract]. Diabetologia 2012;55(Suppl 1):S Keenan DB, Mastrototaro JJ, Zisser H, Cooper KA, Raghavendhar G, Lee SW, Yusi J, Bailey TS, Brazg RL, Shah RV: Accuracy of the Enlite 6-day glucose sensor with Guardian and Veo calibration algorithms. Diabetes Technol Ther 2012;14: Simpson P, Peyser T, Price D, Garcia A, Nakamura K: Impact of study design and analytic techniques on reported accuracy of continuous glucose monitoring (CGM) systems. Diabetologia 2012;55(Suppl 1):S431 S Goldberg PA, Siegel MD, Russell RR, Sherwin RS, Halickman JI, Cooper DA, Dziura JD, Inzucchi SE: Experience with the continuous glucose monitoring system in a medical intensive care unit. Diabetes Technol Ther 2004;6: Keenan DB, Cartaya R, Mastrototaro JJ: Accuracy of a new real-time continuous glucose monitoring algorithm. J Diabetes Sci Technol 2010;4: Address correspondence to: Christiane Bay Endocrinology Section Department of Cardiology, Nephrology and Endocrinology Hillerød Hospital Dyrehavevej Hillerød, Denmark kdp938@alumni.ku.dk

Norbert Hermanns, PhD 1,2, Beatrix Schumann, MD 2, Bernhard Kulzer, PhD 1,2, and Thomas Haak, MD 1,2. Original Article

Norbert Hermanns, PhD 1,2, Beatrix Schumann, MD 2, Bernhard Kulzer, PhD 1,2, and Thomas Haak, MD 1,2. Original Article 524105DSTXXX10.1177/1932296814524105Journal of Diabetes Science and TechnologyHermanns et al research-article2014 Original Article The Impact of Continuous Glucose Monitoring on Low Interstitial Glucose

More information

Long-term effects of continuous glucose monitoring on HbA 1c levels: An audit

Long-term effects of continuous glucose monitoring on HbA 1c levels: An audit Long-term effects of continuous glucose monitoring on Julie Brake Continuous glucose monitoring (CGM) has become a common and useful tool in diabetes care. To understand whether a 72-hour glucose profile

More information

Supplementary Appendix

Supplementary Appendix Supplementary Appendix This appendix has been provided by the authors to give readers additional information about their work. Supplement to: Phillip M, Battelino T, Atlas E, et al. Nocturnal glucose control

More information

Accuracy Requirements for a Hypoglycemia Detector: An Analytical Model to Evaluate the Effects of Bias, Precision, and Rate of Glucose Change

Accuracy Requirements for a Hypoglycemia Detector: An Analytical Model to Evaluate the Effects of Bias, Precision, and Rate of Glucose Change Journal of Diabetes Science and Technology Volume 1, Issue 5, September 2007 Diabetes Technology Society SYMPOSIUM Accuracy Requirements for a Hypoglycemia Detector: An Analytical Model to Evaluate the

More information

Artificial Pancreas Device System (APDS)

Artificial Pancreas Device System (APDS) Medical Policy Manual Durable Medical Equipment, Policy No. 77 Artificial Pancreas Device System (APDS) Next Review: October 2019 Last Review: October 2018 Effective: November 1, 2018 IMPORTANT REMINDER

More information

Agreement between Glucose Trends Derived from Three Simultaneously Worn Continuous Glucose Sensors

Agreement between Glucose Trends Derived from Three Simultaneously Worn Continuous Glucose Sensors Journal of Diabetes Science and Technology Volume 2, Issue 5, September 2008 Diabetes Technology Society ORIGINAL ARTICLES Agreement between Glucose Trends Derived from Three Simultaneously Worn Continuous

More information

Performance Evaluations of Continuous Glucose Monitoring Systems: Precision Absolute Relative Deviation Is Part of the Assessment

Performance Evaluations of Continuous Glucose Monitoring Systems: Precision Absolute Relative Deviation Is Part of the Assessment Journal of Diabetes Science and Technology Volume 7, Issue 4, July 2013 Diabetes Technology Society SYMPOSIUM Performance Evaluations of Continuous Glucose Monitoring Systems: Karin, B.S., Günther Schmelzeisen-Redeker,

More information

Comparison of the Numerical and Clinical Accuracy of Four Continuous Glucose Monitors

Comparison of the Numerical and Clinical Accuracy of Four Continuous Glucose Monitors Emerging Treatments and Technologies O R I G I N A L A R T I C L E Comparison of the Numerical and Clinical Accuracy of Four Continuous Glucose Monitors BORIS KOVATCHEV, PHD 1 STACEY ANDERSON, MD 1 2 LUTZ

More information

Subjects and Methods 2017, 64 (8),

Subjects and Methods 2017, 64 (8), 2017, 64 (8), 827-832 Careful readings for a flash glucose monitoring system in nondiabetic Japanese subjects: individual differences and discrepancy in glucose concentrarion after glucose loading Keiko

More information

Today s Goals 10/6/2017. New Frontiers in Diabetes Technology. Disclosures

Today s Goals 10/6/2017. New Frontiers in Diabetes Technology. Disclosures New Frontiers in Diabetes Technology Marie E. McDonnell, MD Director, Brigham and Women's Diabetes Program Division of Endocrinology, Diabetes and Hypertension Brigham and Women s Hospital Today s Goals

More information

Sensor-augmented pump systems provide insulin delivery

Sensor-augmented pump systems provide insulin delivery DIABETES TECHNOLOGY & THERAPEUTICS Volume 18, Number 10, 2016 Mary Ann Liebert, Inc. DOI: 10.1089/dia.2016.0216 ORIGINAL ARTICLE Effectiveness of Automated Insulin Management Features of the MiniMed c

More information

Diabetes II Insulin pumps; Continuous glucose monitoring system (CGMS) Ernest Asamoah, MD FACE FACP FRCP (Lond)

Diabetes II Insulin pumps; Continuous glucose monitoring system (CGMS) Ernest Asamoah, MD FACE FACP FRCP (Lond) Diabetes II Insulin pumps; Continuous glucose monitoring system (CGMS) Ernest Asamoah, MD FACE FACP FRCP (Lond) 9501366-011 20110401 Objectives Understand the need for insulin pumps and CGMS in managing

More information

Research: Complications Asymptomatic hypoglycaemia in Type 1 diabetes: incidence and risk factors

Research: Complications Asymptomatic hypoglycaemia in Type 1 diabetes: incidence and risk factors Research: Complications Asymptomatic hypoglycaemia in Type 1 diabetes: incidence and risk factors M. M. Henriksen 1, H. U. Andersen 2, B. Thorsteinsson 1,3 and U. Pedersen-Bjergaard 1,3 1 Department of

More information

PREVENTION OF NOCTURNAL HYPOGLYCEMIA USING PREDICTIVE LOW GLUCOSE SUSPEND (PLGS)

PREVENTION OF NOCTURNAL HYPOGLYCEMIA USING PREDICTIVE LOW GLUCOSE SUSPEND (PLGS) PREVENTION OF NOCTURNAL HYPOGLYCEMIA USING PREDICTIVE LOW GLUCOSE SUSPEND (PLGS) Pathways for Future Treatment and Management of Diabetes H. Peter Chase, MD Carousel of Hope Symposium Beverly Hilton, Beverly

More information

Control of Glycemic Variability for Reducing Hypoglycemia Jae Hyeon Kim

Control of Glycemic Variability for Reducing Hypoglycemia Jae Hyeon Kim Control of Glycemic Variability for Reducing Hypoglycemia Jae Hyeon Kim Division of Endocrinology and Metabolism, Samsung Medical Center, Sungkyunkwan University School of Medicine Conflict of interest

More information

Use of Continuous Glucose Monitoring in the Detection and Prevention of Hypoglycemia

Use of Continuous Glucose Monitoring in the Detection and Prevention of Hypoglycemia Journal of Diabetes Science and Technology Volume 1, Issue 1, January 2007 Diabetes Technology Society SYMPOSIUM Use of Continuous Glucose Monitoring in the Detection and Prevention of Hypoglycemia Howard

More information

The transition from urinary

The transition from urinary PRACTICAL POINTERS Continuous Glucose Monitoring: A Perspective on Its Past, Present, and Future Applications for Diabetes Management Hanna S. Mariani, 1 Brian T. Layden, 1,2 and Grazia Aleppo 1 1 Division

More information

Continuous Glucose Monitoring (CGM)

Continuous Glucose Monitoring (CGM) Continuous Glucose Monitoring (CGM) Date of Origin: 02/2001 Last Review Date: 07/26/2017 Effective Date: 07/26/2017 Dates Reviewed: 04/2004, 04/2005, 03/2006, 11/2006, 12/2007, 03/2008, 09/2008, 04/2009,

More information

Hypoglycemia Detection and Prediction Using Continuous Glucose Monitoring A Study on Hypoglycemic Clamp Data

Hypoglycemia Detection and Prediction Using Continuous Glucose Monitoring A Study on Hypoglycemic Clamp Data Journal of Diabetes Science and Technology Volume 1, Issue 5, September 2007 Diabetes Technology Society SYMPOSIUM Hypoglycemia Detection and Prediction Using Continuous Glucose Monitoring A Study on Hypoglycemic

More information

Paolo Di Bartolo U.O di Diabetologia Dip. Malattie Digestive & Metaboliche AULS Prov. di Ravenna. Ipoglicemie e Monitoraggio Glicemico

Paolo Di Bartolo U.O di Diabetologia Dip. Malattie Digestive & Metaboliche AULS Prov. di Ravenna. Ipoglicemie e Monitoraggio Glicemico Paolo Di Bartolo U.O di Diabetologia Dip. Malattie Digestive & Metaboliche AULS Prov. di Ravenna Ipoglicemie e Monitoraggio Glicemico Management of Hypoglycaemia.if hypoglycemia is a problem, the principles

More information

Sponsor / Company: Sanofi Drug substance(s): insulin glargine (HOE901) According to template: QSD VERSION N 4.0 (07-JUN-2012) Page 1

Sponsor / Company: Sanofi Drug substance(s): insulin glargine (HOE901) According to template: QSD VERSION N 4.0 (07-JUN-2012) Page 1 These results are supplied for informational purposes only. Prescribing decisions should be made based on the approved package insert in the country of prescription. Sponsor / Company: Sanofi Drug substance(s):

More information

What is a CGM? (Continuous Glucose Monitor) The Bionic Pancreas Is Coming

What is a CGM? (Continuous Glucose Monitor) The Bionic Pancreas Is Coming The Bionic Pancreas Is Coming Montana Diabetes Professional Conference October 23, 2014 H. Peter Chase, MD Professor of Pediatrics University of Colorado Barbara Davis Center Stanford: Bruce Buckingham,

More information

James J. Chamberlain, MD 1, Dana Dopita, RN, MSN, CDE 1, Emily Gilgen, RD, CD, CDE 1, and Annie Neuman, MPA, PA-C 1.

James J. Chamberlain, MD 1, Dana Dopita, RN, MSN, CDE 1, Emily Gilgen, RD, CD, CDE 1, and Annie Neuman, MPA, PA-C 1. 604633DSTXXX10.1177/1932296815604633Journal of Diabetes Science and TechnologyChamberlain et al research-article2015 Original Article Impact of Frequent and Persistent Use of Continuous Glucose Monitoring

More information

King s Research Portal

King s Research Portal King s Research Portal DOI: 10.1089/dia.2015.0324 Document Version Publisher's PDF, also known as Version of record Link to publication record in King's Research Portal Citation for published version (APA):

More information

Pumps & Sensors made easy. OPADA ALZOHAILI MD FACE Endocrinology Assistant Professor Wayne State University

Pumps & Sensors made easy. OPADA ALZOHAILI MD FACE Endocrinology Assistant Professor Wayne State University Pumps & Sensors made easy OPADA ALZOHAILI MD FACE Endocrinology Assistant Professor Wayne State University DeFronzo RA. Diabetes. 2009;58:773-795. Ominous Octet Relationship of b-cell Dysfunction and Development

More information

Continuous glucose monitoring (CGM) systems enable

Continuous glucose monitoring (CGM) systems enable DIA-2014-0375-ver9-Wang_1P.3d 05/05/15 2:45pm Page 1 DIABETES TECHNOLOGY & THERAPEUTICS Volume 17, Number 8, 2015 ª Mary Ann Liebert, Inc. DOI: 10.1089/dia.2014.0375 DIA-2014-0375-ver9-Wang_1P Type: original-article

More information

Evaluation of a Continuous Glucose Monitoring System for Home-Use Conditions

Evaluation of a Continuous Glucose Monitoring System for Home-Use Conditions Evaluation of a Continuous Glucose Monitoring System for Home-Use Conditions Bruce Bode, MD 1, Michael Silver, MD 2, Richard Weiss, MS, 3 Kathryn Martin, PharmD. 3 1 Atlanta Diabetes Associates; 2 Medical

More information

Effect of Continuous Glucose Monitoring Accuracy on Clinicians Retrospective Decision Making in Diabetes: A Pilot Study

Effect of Continuous Glucose Monitoring Accuracy on Clinicians Retrospective Decision Making in Diabetes: A Pilot Study 587935DSTXXX10.1177/1932296815587935Journal of Diabetes Science and TechnologyMahmoudi et al research-article2015 Original Article Effect of Continuous Glucose Monitoring Accuracy on Clinicians Retrospective

More information

Comparison of glucose monitoring between Freestyle Libre Pro and ipro2 in patients with diabetes mellitus

Comparison of glucose monitoring between Freestyle Libre Pro and ipro2 in patients with diabetes mellitus Comparison of glucose monitoring between Freestyle Libre Pro and ipro2 in patients with diabetes mellitus Ryo Kumagai 1, iko Muramatsu 1, Masanao Fujii 1, Yukino Katakura 1,KeiIto 1,KeikoFujie 2, Yoshio

More information

WOULD YOU LIKE MORE CONFIDENCE IN YOUR HYPOGLYCAEMIC CONTROL? SMARTGUARD. PROVIDES PREDICTIVE PROTECTION

WOULD YOU LIKE MORE CONFIDENCE IN YOUR HYPOGLYCAEMIC CONTROL? SMARTGUARD. PROVIDES PREDICTIVE PROTECTION WOULD YOU LIKE MORE CONFIDENCE IN YOUR HYPOGLYCAEMIC CONTROL? SMARTGUARD. PROVIDES PREDICTIVE PROTECTION 1,2 MINIMED 640G WITH SMARTGUARD HAVE YOU CONSIDERED THE REAL IMPACT HYPOGLYCAEMIA HAS ON YOUR LIFE?

More information

WOULD YOU LIKE TO REDUCE THE FEAR OF HYPOGLYCAEMIA FOR YOUR PATIENTS?

WOULD YOU LIKE TO REDUCE THE FEAR OF HYPOGLYCAEMIA FOR YOUR PATIENTS? WOULD YOU LIKE TO REDUCE THE FEAR OF HYPOGLYCAEMIA FOR YOUR PATIENTS? THE ONLY SYSTEM CLINICALLY PROVEN TO REDUCE HYPOGLYCAEMIA MINIMED 640G SYSTEM WITH SMARTGUARD TECHNOLOGY 1 HYPOGLYCAEMIA IS CHALLENGING

More information

Early Patient and Clinician Experiences with Continuous Glucose Monitoring

Early Patient and Clinician Experiences with Continuous Glucose Monitoring Feature Article / Early Experiences with CGM Early Patient and Clinician Experiences with Continuous Glucose Monitoring David K. Bloomgarden, MD, FACE; Janine Freeman, RD, LD, CDE; and Elizabeth DeRobertis,

More information

Limitations of Conventional Methods of Self-Monitoring of Blood Glucose

Limitations of Conventional Methods of Self-Monitoring of Blood Glucose Clinical Care/Education/Nutrition O R I G I N A L A R T I C L E Limitations of Conventional Methods of Self-Monitoring of Blood Glucose Lessons learned from 3 days of continuous glucose sensing in pediatric

More information

FreeStyle Mini Blood Glucose Results Are Accurate and Suitable for Use in Glycemic Clamp Protocols

FreeStyle Mini Blood Glucose Results Are Accurate and Suitable for Use in Glycemic Clamp Protocols Journal of Diabetes Science and Technology Volume 2, Issue 5, September 2008 Diabetes Technology Society CLINICAL APPLICATIONS FreeStyle Mini Blood Glucose Results Are Accurate and Suitable for Use in

More information

A Comparison of Time Delay in Three Continuous Glucose Monitors for Adolescents and Adults

A Comparison of Time Delay in Three Continuous Glucose Monitors for Adolescents and Adults 704443DSTXXX10.1177/1932296817704443Journal of Diabetes Science and TechnologySinha et al research-article2017 Original Article A Comparison of Time Delay in Three Continuous Glucose Monitors for Adolescents

More information

Continuous Glucose Monitoring Considerations for the Development of a Closed-Loop Artificial Pancreas System

Continuous Glucose Monitoring Considerations for the Development of a Closed-Loop Artificial Pancreas System Journal of Diabetes Science and Technology Volume 5, Issue 6, November 2011 Diabetes Technology Society SYMPOSIUM Continuous Glucose Monitoring Considerations for the Development of a Closed-Loop Artificial

More information

Advances in Diabetes Care Technologies

Advances in Diabetes Care Technologies 1979 Advances in Diabetes Care Technologies 2015 Introduction Insulin pump use: ~ 20% - 30% of patients with T1DM < 1% of insulin-treated patients with T2DM 2007 FDA estimates ~375,000 insulin pumps for

More information

Advances in Technology in the Treatment of Diabetes Mellitus 2017 How far have we come-how far are we going? Is there a final frontier?

Advances in Technology in the Treatment of Diabetes Mellitus 2017 How far have we come-how far are we going? Is there a final frontier? Advances in Technology in the Treatment of Diabetes Mellitus 2017 How far have we come-how far are we going? Is there a final frontier? Alan B Schorr DO FAAIM FACE www.sugardoc.com abs@sugardoc.com Disclosures

More information

The MiniMed Continuous Glucose Monitoring System. JOHN J. MASTROTOTARO, Ph.D.

The MiniMed Continuous Glucose Monitoring System. JOHN J. MASTROTOTARO, Ph.D. DIABETES TECHNOLOGY & THERAPEUTICS Volume 2, Supplement 1, 2000 Mary Ann Liebert, Inc. The MiniMed Continuous Glucose Monitoring System JOHN J. MASTROTOTARO, Ph.D. DIABETES is a major source of morbidity,

More information

Artificial Pancreas Device Systems. Populations Interventions Comparators Outcomes Individuals: With type 1 diabetes

Artificial Pancreas Device Systems. Populations Interventions Comparators Outcomes Individuals: With type 1 diabetes Protocol Artificial Pancreas Device Systems Medical Benefit Effective Date: 07/01/18 Next Review Date: 01/20 Preauthorization Yes Review Dates: 03/15, 03/16, 03/17, 01/18, 05/18, 01/19 Preauthorization

More information

Continuous Glucose Monitoring Devices Pharmacy Policy

Continuous Glucose Monitoring Devices Pharmacy Policy Line of Business: All Line of Business Effective date: August 16, 2017 Revision date: August 16, 2017 Continuous Glucose Monitoring Devices Pharmacy Policy This policy has been developed through review

More information

Continuous Glucose Monitors for Diabetes Management

Continuous Glucose Monitors for Diabetes Management Continuous Glucose Monitors for Diabetes Management Ryan Huang, DO PGY II, Sonia Garcia-Jayne, DO PGY II Mandeep Gill, DO PGY I, Justin Leeka, DO, PGY I, Catherine Nguyen OMS IV Family Medicine Residency,

More information

Preventing Hypoglycemia Using Predictive Alarm Algorithms and Insulin Pump Suspension

Preventing Hypoglycemia Using Predictive Alarm Algorithms and Insulin Pump Suspension DIABETES TECHNOLOGY & THERAPEUTICS Volume 11, Number 2, 29 Mary Ann Liebert, Inc. DOI: 1.189/dia.28.32 Preventing Hypoglycemia Using Predictive Alarm Algorithms and Insulin Pump Suspension Bruce Buckingham,

More information

What is the role of insulin pumps in the modern day care of patients with Type 1 diabetes?

What is the role of insulin pumps in the modern day care of patients with Type 1 diabetes? What is the role of insulin pumps in the modern day care of patients with Type 1 diabetes? Dr. Fiona Wotherspoon Consultant in Diabetes and Endocrinology Dorset County Hospital Fiona.Wotherspoon@dchft.nhs.uk

More information

Coverage Guidelines. Continuous Glucose Monitors (CGMs)

Coverage Guidelines. Continuous Glucose Monitors (CGMs) Coverage Guidelines Continuous Glucose Monitors (CGMs) Disclaimer: Please note that Baptist Health Plan Coverage Guidelines may be updated throughout the year. A printed version may not be most up to date

More information

VALIDATION OF CONTINUOUS GLUCOSE MONITORING IN CHILDREN AND ADOLESCENTS WITH CYSTIC FIBROSIS - A PROSPECTIVE COHORT STUDY

VALIDATION OF CONTINUOUS GLUCOSE MONITORING IN CHILDREN AND ADOLESCENTS WITH CYSTIC FIBROSIS - A PROSPECTIVE COHORT STUDY Diabetes Care Publish Ahead of Print, published online March 11, 2009 Validation of CGM in Children with CF VALIDATION OF CONTINUOUS GLUCOSE MONITORING IN CHILDREN AND ADOLESCENTS WITH CYSTIC FIBROSIS

More information

Artificial Pancreas Device Systems. Populations Interventions Comparators Outcomes. pump. pump

Artificial Pancreas Device Systems. Populations Interventions Comparators Outcomes. pump. pump Protocol Artificial Pancreas Device Systems (10130) Medical Benefit Effective Date: 04/01/18 Next Review Date: 01/19 Preauthorization Yes Review Dates: 03/15, 03/16, 03/17, 01/18 Preauthorization is required.

More information

Hypoglycemia Prediction with Subject-Specific Recursive Time-Series Models

Hypoglycemia Prediction with Subject-Specific Recursive Time-Series Models Journal of Diabetes Science and Technology Volume 4, Issue 1, January 2010 Diabetes Technology Society SYMPOSIUM Hypoglycemia Prediction with Subject-Specific Recursive Time-Series Models Meriyan, M.E.,

More information

THE ONLY SYSTEM^ CLINICALLY PROVEN TO REDUCE HYPOGLYCAEMIA

THE ONLY SYSTEM^ CLINICALLY PROVEN TO REDUCE HYPOGLYCAEMIA THE ONLY SYSTEM^ CLINICALLY PROVEN TO REDUCE HYPOGLYCAEMIA MINIMED 640G system^ with SmartGuard technology HYPOGLYCAEMIA IS CHALLENGING FOR GLYCAEMIC CONTROL 1 AND quality of life 2 Achieving good glucose

More information

Accuracy and Performance of Continuous Glucose Monitors in Athletes

Accuracy and Performance of Continuous Glucose Monitors in Athletes Accuracy and Performance of Continuous Glucose Monitors in Athletes Felicity Thomas,* Christopher G. Pretty,* Matthew Signal,* Geoffrey Shaw,* J. Geoffrey Chase* * Department of Mechanical Engineering,

More information

Updates in Diabetes Technology

Updates in Diabetes Technology Updates in Diabetes Technology Jessica Kirk, MSN, RN, CPN, CDE Nurse Manager, Endo ECHO No disclosures Disclosures 1 Objectives Distinguish patients appropriate for continuous glucose monitoring and insulin

More information

Performance-powered. The OneTouch. Ping insulin pump and meter-remote.

Performance-powered. The OneTouch. Ping insulin pump and meter-remote. Performance-powered. The OneTouch Ping insulin pump and meter-remote. I We don t just deliver insulin. We deliver outstanding clinical performance. P36337_OTP_DetAid_OmniPodUpdate_r12.indd 1 OneTouch Ping.

More information

Report Reference Guide

Report Reference Guide Report Reference Guide How to use this guide Each type of CareLink report and its components are described in the following sections. Report data used to generate the sample reports was from sample patient

More information

Medical Policy An independent licensee of the Blue Cross Blue Shield Association

Medical Policy An independent licensee of the Blue Cross Blue Shield Association Continuous or Intermittent Monitoring of Glucose in Interstitial Fluid Page 1 of 26 Medical Policy An independent licensee of the Blue Cross Blue Shield Association Title: Continuous or Intermittent Monitoring

More information

Report Reference Guide. THERAPY MANAGEMENT SOFTWARE FOR DIABETES CareLink Report Reference Guide 1

Report Reference Guide. THERAPY MANAGEMENT SOFTWARE FOR DIABETES CareLink Report Reference Guide 1 Report Reference Guide THERAPY MANAGEMENT SOFTWARE FOR DIABETES CareLink Report Reference Guide 1 How to use this guide Each type of CareLink report and its components are described in the following sections.

More information

Continuous Glucose Monitoring and the Reality of Metabolic Control in Preschool Children With Type 1 Diabetes

Continuous Glucose Monitoring and the Reality of Metabolic Control in Preschool Children With Type 1 Diabetes Emerging Treatments and Technologies O R I G I N A L A R T I C L E Continuous Glucose Monitoring and the Reality of Metabolic Control in Preschool Children With Type 1 Diabetes GEORGE S. JEHA, MD 1,2 LEFKOTHEA

More information

Advances Towards the Bionic Pancreas.

Advances Towards the Bionic Pancreas. Advances Towards the Bionic Pancreas. Ron Brazg MD, FACE Rainier Clinical Research Center Renton, WA DISCLOSURES Rainier Clinical Research Center is an independent research facility not directly affiliated

More information

Hypoglycemia a barrier to normoglycemia Are long acting analogues and pumps the answer to the barrier??

Hypoglycemia a barrier to normoglycemia Are long acting analogues and pumps the answer to the barrier?? Hypoglycemia a barrier to normoglycemia Are long acting analogues and pumps the answer to the barrier?? Moshe Phillip Institute of Endocrinology and Diabetes National Center of Childhood Diabetes Schneider

More information

Continuous Glucose Monitoring (CGM)

Continuous Glucose Monitoring (CGM) Continuous Glucose Monitoring (CGM) Background Info A1c Average glucose levels over previous 2-3 months Currently remains gold standard for determining control Indicator of risk for development of complications

More information

Clinical Value and Evidence of Continuous Glucose Monitoring

Clinical Value and Evidence of Continuous Glucose Monitoring Clinical Value and Evidence of Continuous Glucose Monitoring 9402313-012 Objective To review the clinical value and the recent clinical evidence for Professional and Personal CGM Key Points CGM reveals

More information

Continuous Glucose Monitoring: Changing Diabetes Behavior in Real Time and Retrospectively

Continuous Glucose Monitoring: Changing Diabetes Behavior in Real Time and Retrospectively Journal of Diabetes Science and Technology Volume 2, Issue 3, May 2008 Diabetes Technology Society CONTROVERSIES in Continuous Glucose Monitoring Continuous Glucose Monitoring: Changing Diabetes Behavior

More information

Continuous or Intermittent Monitoring of Glucose in Interstitial Fluid. Original Policy Date

Continuous or Intermittent Monitoring of Glucose in Interstitial Fluid. Original Policy Date MP 1.01.15 Continuous or Intermittent Monitoring of Glucose in Interstitial Fluid Medical Policy Section Durable Medical Equipment Issue 12:2013 Original Policy Date 12:2013 Last Review Status/Date Reviewed

More information

Diabetes Care 34: , with an increased risk of SH. In a

Diabetes Care 34: , with an increased risk of SH. In a Clinical Care/Education/Nutrition/Psychosocial Research O R I G I N A L A R T I C L E Factors Predictive of Severe Hypoglycemia in Type 1 Diabetes Analysis from the Juvenile Diabetes Research Foundation

More information

Description. Section: Durable Medical Equipment Effective Date: July 15, 2014 Subsection: Original Policy Date: December 7, 2011 Subject:

Description. Section: Durable Medical Equipment Effective Date: July 15, 2014 Subsection: Original Policy Date: December 7, 2011 Subject: Last Review Status/Date: June 2014 Page: 1 of 13 Description Tight glucose control in patients with diabetes has been associated with improved outcomes. Several devices are available to measure glucose

More information

Insulin Delivery and Glucose Monitoring Methods for Diabetes Mellitus: Comparative Effectiveness

Insulin Delivery and Glucose Monitoring Methods for Diabetes Mellitus: Comparative Effectiveness Insulin Delivery and Glucose Monitoring Methods for Diabetes Mellitus: Comparative Effectiveness Prepared for: Agency for Healthcare Research and Quality (AHRQ) www.ahrq.gov Outline of Material Introduction

More information

hypoglycaemia unawareness keystone 18 July 2014

hypoglycaemia unawareness keystone 18 July 2014 hypoglycaemia unawareness keystone 18 July 2014 Hypoglycaemia unawareness: ( Impaired awareness of hypoglycaemia ) Philip Home Newcastle University Philip Home Duality of interest Manufacturers of glucose-lowering

More information

Glucose Outcomes with the In-Home Use of a Hybrid Closed-Loop Insulin Delivery System in Adolescents and Adults with Type 1 Diabetes

Glucose Outcomes with the In-Home Use of a Hybrid Closed-Loop Insulin Delivery System in Adolescents and Adults with Type 1 Diabetes DIABETES TECHNOLOGY & THERAPEUTICS Volume 19, Number 3, 2017 Mary Ann Liebert, Inc. DOI: 10.1089/dia.2016.0421 ORIGINAL ARTICLE Glucose Outcomes with the In-Home Use of a Hybrid Closed-Loop Insulin Delivery

More information

Effects of Measurement Frequency on Analytical Quality Required for Glucose Measurements in Intensive Care Units: Assessments by Simulation Models

Effects of Measurement Frequency on Analytical Quality Required for Glucose Measurements in Intensive Care Units: Assessments by Simulation Models Clinical Chemistry 60:4 644 650 (2014) Endocrinology and Metabolism Effects of Measurement Frequency on Analytical Quality Required for Glucose Measurements in Intensive Care Units: Assessments by Simulation

More information

Objectives. Glucose Monitoring Today. Tracking Glucose ISO Standards. Continuous Glucose Monitoring & Diabetes Management 3/7/2015

Objectives. Glucose Monitoring Today. Tracking Glucose ISO Standards. Continuous Glucose Monitoring & Diabetes Management 3/7/2015 Objectives Continuous Glucose Monitoring & Diabetes Management Tomas C. Walker, DNP, APRN, CDE Director, Clinical Projects Dexcom, Inc. San Diego, CA 1. Review current limitations of SMBG 2. Understand

More information

Continuous Glucose Monitoring (CGM)

Continuous Glucose Monitoring (CGM) Continuous Glucose Monitoring (CGM) Date of Origin: 02/2001 Last Review Date: 08/22/2018 Effective Date: 08/22/2018 Dates Reviewed: 04/2004, 04/2005, 03/2006, 11/2006, 12/2007, 03/2008, 09/2008, 04/2009,

More information

Subject Index. Breastfeeding, self-monitoring of blood glucose 56

Subject Index. Breastfeeding, self-monitoring of blood glucose 56 Subject Index Animas Vibe 86, 130 Artificial pancreas clinical studies inpatient studies 175 180 outpatient studies outcome assessment 182, 183 technology 180, 181 telemedicine 182 components glucose sensor

More information

Reconstruction of Glucose in Plasma from Interstitial Fluid Continuous Glucose Monitoring Data: Role of Sensor Calibration

Reconstruction of Glucose in Plasma from Interstitial Fluid Continuous Glucose Monitoring Data: Role of Sensor Calibration Journal of Diabetes Science and Technology Volume 1, Issue 5, September 2007 Diabetes Technology Society SYMPOSIUM Reconstruction of Glucose in Plasma from Interstitial Fluid Continuous Glucose Monitoring

More information

Corporate Medical Policy

Corporate Medical Policy Corporate Medical Policy Continuous Monitoring of Glucose in the Interstitial Fluid File Name: Origination: Last CAP Review: Next CAP Review: Last Review: continuous_monitoring_of_glucose_in_the_interstitial_fluid

More information

Continuous Glucose Monitoring: The Future of Diabetes Management

Continuous Glucose Monitoring: The Future of Diabetes Management In Brief Continuous glucose monitoring (CGM) technology has the potential to revolutionize diabetes care in the near future because of the real-time feedback it provides about therapeutic interventions

More information

Sponsor / Company: Sanofi Drug substance(s): HOE901-U300 (insulin glargine) According to template: QSD VERSION N 4.0 (07-JUN-2012) Page 1

Sponsor / Company: Sanofi Drug substance(s): HOE901-U300 (insulin glargine) According to template: QSD VERSION N 4.0 (07-JUN-2012) Page 1 These results are supplied for informational purposes only. Prescribing decisions should be made based on the approved package insert in the country of prescription. Sponsor / Company: Sanofi Drug substance(s):

More information

Effective Health Care Program

Effective Health Care Program Comparative Effectiveness Review Number 57 Effective Health Care Program Methods for Insulin Delivery and Glucose Monitoring: Comparative Effectiveness Executive Summary Background Diabetes mellitus is

More information

PREVALENCE AND EXTENT OF GLYCEMIC EXCURSIONS IN WELL- CONTROLLED PATIENTS WITH TYPE 2 DIABETES MELLITUS USING CONTINUOUS GLUCOSE-MONITORING SYSTEM

PREVALENCE AND EXTENT OF GLYCEMIC EXCURSIONS IN WELL- CONTROLLED PATIENTS WITH TYPE 2 DIABETES MELLITUS USING CONTINUOUS GLUCOSE-MONITORING SYSTEM 66 PREVALENCE AND EXTENT OF GLYCEMIC EXCURSIONS IN WELL- CONTROLLED PATIENTS WITH TYPE 2 DIABETES MELLITUS USING CONTINUOUS GLUCOSE-MONITORING SYSTEM SHOKOUFEH BONAKDARAN, REZA RAJABIAN ABSTRACT BACKGROUND:

More information

Personal statement on Continuous Subcutaneous Insulin Infusion Professor John Pickup

Personal statement on Continuous Subcutaneous Insulin Infusion Professor John Pickup 1 Personal statement on Continuous Subcutaneous Insulin Infusion Professor John Pickup King s College London School of Medicine, Guy s Hospital, London SE1 9RT Experience of the technology I am the lead

More information

Efficacy/pharmacodynamics: 85 Safety: 89

Efficacy/pharmacodynamics: 85 Safety: 89 These results are supplied for informational purposes only. Prescribing decisions should be made based on the approved package insert in the country of prescription. Sponsor/Company: Sanofi Drug substance:

More information

Closing the loop, squaring the circle: Studies on insulin delivery, glucose monitoring and the artificial pancreas Luijf, Y.M.

Closing the loop, squaring the circle: Studies on insulin delivery, glucose monitoring and the artificial pancreas Luijf, Y.M. UvA-DARE (Digital Academic Repository) Closing the loop, squaring the circle: Studies on insulin delivery, glucose monitoring and the artificial pancreas Luijf, Y.M. Link to publication Citation for published

More information

State of California Health and Human Services Agency Department of Health Care Services

State of California Health and Human Services Agency Department of Health Care Services State of California Health and Human Services Agency Department of Health Care Services JENNIFER KENT Director EDMOND G. BROWN JR Governor DATE: N.L.: 03-0317 Index: Benefits TO: ALL COUNTY CALIFORNIA

More information

Advances in Diabetes Care Technologies

Advances in Diabetes Care Technologies Advances in Diabetes Care Technologies 1979 2015 Introduction Roughly 20% to 30% of patients with T1DM and fewer than 1% of insulin-treated patients with T2DM use an insulin pump In 2007, the U.S. FDA

More information

COPYRIGHTED MATERIAL. Chapter 1 An Introduction to Insulin Pump Therapy WHAT IS INSULIN PUMP THERAPY?

COPYRIGHTED MATERIAL. Chapter 1 An Introduction to Insulin Pump Therapy WHAT IS INSULIN PUMP THERAPY? Chapter 1 An Introduction to Insulin Pump Therapy This chapter will provide information on what insulin pump therapy is, and how insulin pumps have developed from the early models introduced in the 1970s

More information

DISCLAIMER: ECHO Nevada emphasizes patient privacy and asks participants to not share ANY Protected Health Information during ECHO clinics.

DISCLAIMER: ECHO Nevada emphasizes patient privacy and asks participants to not share ANY Protected Health Information during ECHO clinics. DISCLAIMER: Video will be taken at this clinic and potentially used in Project ECHO promotional materials. By attending this clinic, you consent to have your photo taken and allow Project ECHO to use this

More information

A Prospective Evaluation of Insulin Dosing Recommendations in Patients with Type 1 Diabetes at Near Normal Glucose Control: Bolus Dosing

A Prospective Evaluation of Insulin Dosing Recommendations in Patients with Type 1 Diabetes at Near Normal Glucose Control: Bolus Dosing Journal of Diabetes Science and Technology Volume 1, Issue 1, January 2007 Diabetes Technology Society ORIGINAL ARTICLES A Prospective Evaluation of Insulin Dosing Recommendations in Patients with Type

More information

DIABETES TECHNOLOGY: WHERE ARE WE NOW AND WHERE ARE WE GOING? Presented by: Tom Brobson

DIABETES TECHNOLOGY: WHERE ARE WE NOW AND WHERE ARE WE GOING? Presented by: Tom Brobson DIABETES TECHNOLOGY: WHERE ARE WE NOW AND WHERE ARE WE GOING? Presented by: Tom Brobson March 3 rd, 2018 Who is this guy? Accelerating Progress 1 Oh that guy! Accelerating Progress 2 STATE OF T1D CARE

More information

Initiating Insulin in Primary Care for Type 2 Diabetes Mellitus. Dr Manish Khanolkar, Diabetologist, Auckland Diabetes Centre

Initiating Insulin in Primary Care for Type 2 Diabetes Mellitus. Dr Manish Khanolkar, Diabetologist, Auckland Diabetes Centre Initiating Insulin in Primary Care for Type 2 Diabetes Mellitus Dr Manish Khanolkar, Diabetologist, Auckland Diabetes Centre Outline How big is the problem? Natural progression of type 2 diabetes What

More information

REPORT INTERPRETATION

REPORT INTERPRETATION REPORT INTERPRETATION: Interpreting ipro Professional Continuous Glucose Monitoring (CGM) Reports and Making Therapy Adjustments TARGET AUDIENCE The audience for this section is physicians, mid-level practitioners,

More information

Initiation and Titration of Insulin in Diabetes Mellitus Type 2

Initiation and Titration of Insulin in Diabetes Mellitus Type 2 Initiation and Titration of Insulin in Diabetes Mellitus Type 2 Greg Doelle MD, MS April 6, 2016 Disclosure I have no actual or potential conflicts of interest in relation to the content of this lecture.

More information

Accuracy and Performance of Continuous Glucose Monitors in Athletes

Accuracy and Performance of Continuous Glucose Monitors in Athletes Accuracy and Performance of Continuous Glucose Monitors in Athletes Felicity Thomas,* Christopher G. Pretty,* Matthew Signal,* J. Geoffrey Chase* * Department of Mechanical Engineering, University of Canterbury,

More information

1. What s the point of a network the case for research? 2. How to use CSII effectively

1. What s the point of a network the case for research? 2. How to use CSII effectively 1. What s the point of a network the case for research? 2. How to use CSII effectively John Pickup King s College London Faculty of Medicine Guy s Hospital, London What should an insulin pump network do?

More information

Accuracy of the 5-day FreeStyle Navigator Continuous Glucose Monitoring System: comparison with frequent laboratory reference measurements

Accuracy of the 5-day FreeStyle Navigator Continuous Glucose Monitoring System: comparison with frequent laboratory reference measurements Diabetes Care In Press, published online March 2, 2007 Accuracy of the 5-day FreeStyle Navigator Continuous Glucose Monitoring System: comparison with frequent laboratory reference measurements SHORT RUNNING

More information

QUESTION 4. WHAT CLINICAL DATA ARE CURRENTLY AVAILABLE TO SUPPORT EXPANDED CGM COVERAGE BY PAYERS AS PERTAINS TO QUESTIONS 1 AND 3?

QUESTION 4. WHAT CLINICAL DATA ARE CURRENTLY AVAILABLE TO SUPPORT EXPANDED CGM COVERAGE BY PAYERS AS PERTAINS TO QUESTIONS 1 AND 3? 500 1 QUESTION 4. WHAT CLINICAL DATA ARE CURRENTLY AVAILABLE TO SUPPORT EXPANDED CGM COVERAGE BY PAYERS AS PERTAINS TO QUESTIONS 1 AND 3? WHAT ADDITIONAL DATA ARE NEEDED? AACE/ACE CGM Consensus Conference:

More information

Real-Time Continuous Glucose Monitoring: From Application to Evaluation

Real-Time Continuous Glucose Monitoring: From Application to Evaluation Real-Time Continuous Glucose Monitoring: From Application to Evaluation Gary Scheiner MS, CDE Owner/Director, Integrated Diabetes Services 333 E. Lancaster Ave., Suite 24 Wynnewood, PA 1996 (877) 735-3648

More information

Assessing the value of the Ambulatory Glucose Profile in clinical practice

Assessing the value of the Ambulatory Glucose Profile in clinical practice Assessing the value of the Ambulatory Glucose Profile in clinical practice STEPHAN MATTHAEI Abstract Glycated haemoglobin (HbA 1c) is a measure of mean blood glucose levels over time. It is not a good

More information

The In-Clinic Close Loop Experience in the US

The In-Clinic Close Loop Experience in the US The In-Clinic Close Loop Experience in the US Keystone Symposium, Practical Ways to Achieve Targets in Diabetes Care July 14, 2012 Francine Ratner Kaufman, MD Chief Medical Officer, VP Global Medical,

More information

Anneloes Kerssen, a Harold W. de Valk, b Gerard H.A. Visser a. D RCOG 2004 BJOG: an International Journal of Obstetrics and Gynaecology

Anneloes Kerssen, a Harold W. de Valk, b Gerard H.A. Visser a. D RCOG 2004 BJOG: an International Journal of Obstetrics and Gynaecology BJOG: an International Journal of Obstetrics and Gynaecology September 2004, Vol. 111, pp. 919 924 DOI: 10.1111/j.1471-0528.2004.00203.x Day-to-day glucose variability during pregnancy in women with Type

More information

Advances in Diabetes Care Technologies

Advances in Diabetes Care Technologies 1979 Advances in Diabetes Care Technologies 2015 Introduction Roughly 20% - 30% of patients with T1DM and fewer than 1% of insulin-treated patients with T2DM use an insulin pump In 2007, the US FDA estimated

More information

The Artificial Pancreas

The Artificial Pancreas SAGLB.DIA.15.2.87(2). Approved October 215 The Artificial Pancreas For scientific and medical purposes only The pancreas responds to changes in blood glucose by releasing insulin (β cells) to lower blood

More information