on Venous and Capillary Blood Samples

Size: px
Start display at page:

Download "on Venous and Capillary Blood Samples"

Transcription

1 Evaluation of the Sebia CAPILLARYS Flex Piercing for the Measurement of on Venous and Capillary Blood Samples Olivier Heylen, MSc, Stefaan Van Neyghem, MLT, Sandra Exterbille, MLT, Charline Wehlou, MD, Frans Gorus, MD, PhD, and Ilse Weets, MD, PhD From the Department of Clinical Chemistry and Radio-immunology, Universitair Ziekenhuis Brussel, Brussels, Belgium. Key Words: Analytical performance; Capillary zone electrophoresis; Hemoglobin A c ; HPLC Am J Clin Pathol June ;:8-8 DOI:.9/AJCPRUQCJBANSV ABSTRACT Objectives: We evaluated the Sebia CAPILLARYS Flex Piercing (Cap FP; Sebia, Lisses, France) for measurement of hemoglobin A c ( ) on venous and capillary blood samples. Methods: We analyzed whole-blood samples and control materials with the Cap FP and Tosoh G8 (Tosoh Corporation, Tokyo, Japan). Capillary blood samples were analyzed on the Cap FP on different storage conditions and were compared with venous samples. Results: Both instruments achieved total imprecision of less than.% (International Federation of Clinical Chemistry units). Bias was mmol/mol or less and mmol/mol or less for the Cap FP and Tosoh G8, respectively. The Cap FP was not prone to common interferences. The Tosoh G8 showed significant bias only for carbamylated hemoglobin and did not completely separate hemoglobin D and hemoglobin E. On the Cap FP, storage of capillary blood at room temperature showed no significant bias. There was good agreement with venous blood. Conclusions: The Cap FP and Tosoh G8 perform excellently for determination. Capillary blood can be analyzed on the Cap FP as an acceptable alternative to venous blood and point-of-care testing. Home collection and central analysis of capillary blood could contribute to a reduction of health care costs without reducing quality of determination. Hemoglobin A c ( ) measurement is widely accepted as the most important marker for glycemic control and assessment of the risk of chronic complications both in type and type diabetes mellitus. - Since the American Diabetes Association has introduced as one of the parameters that can be used for the diagnosis of diabetes, has gained importance as a biological marker. In view of the diabetes epidemic worldwide, the number of determinations will increase. Over the past years, major efforts have been made to standardize the determination. In a first phase, values were aligned with the Diabetes Control and Complications Trial Research Group (DCCT), and currently, it is recommended that methods are standardized according to the International Federation of Clinical Chemistry (IFCC) reference method. In parallel with the evolution toward standardization of the determination, criteria for precision and accuracy as well as for diabetes control became more stringent. The methods most commonly used to quantify are immunoassays, ion-exchange high-performance liquid chromatography (HPLC), boronate affinity HPLC, and enzymatic assays. More recently, capillary zone electrophoresis (CZE) has been introduced for the determination. Both the HPLC and CZE methods have the advantage of showing the presence of hemoglobin variants, which is mainly of importance for the interpretation of results if there is a high frequency of hemoglobin variants in the background population.,8 Potential devices for home samples comprise capillary sampling and dried blood spot on filter paper. Samples can be taken at home, are easy to send by regular mail, and can be Downloaded from Am J Clin Pathol ;: DOI:.9/AJCPRUQCJBANSV 8

2 Heylen et al / Evaluation of CAPILLARYS Flex Piercing analyzed before going to the physician for treatment evaluation. 9 Besides saving extra travel for the patient to the clinician for a phlebotomy, intensive follow-up by home sampling improves glycemic control and therefore is especially interesting for patients with poor metabolic control (eg, children, adolescents, and the elderly). Recent studies have been published regarding the performance of the recently commercially released Sebia CAPIL- LARYS Flex Piercing analyzer (Sebia Cap FP; Sebia, Lisses, France).,, In this evaluation, we studied the Sebia Sebia Cap FP and the Tosoh G8 for the determination of in comparison with the Tosoh G instrument (Tosoh Corporation, Tokyo, Japan). In addition, we focused on the determination of on the Sebia Cap FP in a conventional and capillary blood sample in terms of bias between both sample types and stability in different conditions for the capillary sample. glycidyl methacrylate-ethylene glycol methacrylate copolymer cation-exchange column (TSKgel Glyco His, Tokyo, Japan) maintained at C in an aluminum block oven. Each hemoglobin fraction is eluted from the column by a step gradient of three buffers with increasing ionic strength. Approximately µl of whole blood is automatically diluted (/) with hemolysis and wash dilution in the dilution port. Next, the diluted sample is injected onto the column, and the separated fractions are continuously monitored by the diode detector system at and nm. Comparison Instrument The HLC-G (Tosoh G; Tosoh Corporation) analyzer, which is routinely used in our laboratory, was used as a comparator. The Tosoh G instrument was used according to the manufacturer s instructions with the Tosoh calibration materials. Specimens Materials and Methods Analyzers Sebia Cap FP The Sebia Cap FP instrument is an automated analyzer for the quantitative analysis of proteins (serum and urine), immunotyping (serum and urine), hemoglobin variants, carbohydrate-deficient transferrin, and, more recently,. The CAPILLARYS kit is based on the CZE principle, where hemoglobin fractions are separated by their charge in an alkaline buffer (ph 9.). By using this buffer, hemoglobins are detected in the following order, from cathode to anode: A /C, E, S, D, F, A, other hemoglobin A (HbA), and finally A c. The Sebia Cap FP analyzer has eight silica capillaries functioning in parallel, thus allowing eight simultaneous analyses for. Before each run, the capillaries are washed with wash solution and prepared for the next analysis with buffer. Then, 8 µl of the whole-blood sample is automatically diluted (/) with a hemolyzing solution and is injected at the anodic end of the capillary by hydrodynamic injection. A high-voltage protein separation is then performed and direct detection of the hemoglobins is made at the cathodic end of the capillary at nm. Tosoh G8 Analyzer The Tosoh HLC-G8 (Tosoh G8; Tosoh Corporation) instrument is a fully automatic HPLC benchtop analyzer for the determination of and hemoglobin variants. The Tosoh G8 separates the various components of hemoglobin into six fractions. Separation is achieved by a nonporous Control Samples Quality control materials used throughout the evaluation were Lyphochek Diabetes Control Levels and (Bio-Rad Laboratories, Irvine, CA) and CAPILLARYS Control and (Sebia). Both are human lyophilized whole-blood assayed controls. External quality control (EQC) samples supplied by the Wetenschappelijk Instituut Volksgezondheid Institut Scientifique de Santé Publique (WIV-ISP; Belgian Scientific Institute for Public Health, Brussels, Belgium) were used for bias evaluation. The samples contain lyophilized hemolysate of human origin, and the targets were set by an IFCC reference method. Patient Samples Venous blood samples were collected into S-Monovette potassium EDTA tubes (Sarstedt, Nürnbrecht, Germany). Capillary blood samples were collected by finger prick into -μl EDTA End-to-End capillaries (Sarstedt), which were subsequently stored into screw-cap microtubes (Sarstedt) either with or without hemolysis reagent (Sebia). This study was a prospective study of samples submitted for routine analysis of. Residual material of whole-blood samples from diabetic and nondiabetic patients attending the UZ Brussel (Brussels, Belgium) was included. The samples in this study were used according to procedures supervised by local authorities responsible for ethical research. All results were expressed in IFCC units. The relationship between the IFCC units and DCCT/National Glycohemoglobin Standardization Program (NGSP) results is defined by the following master equation: NGSP = (.9 IFCC) Am J Clin Pathol ;:8-8 Downloaded 88 from DOI:.9/AJCPRUQCJBANSV

3 Evaluation Protocol Bias Estimation Trueness was tested according to Clinical and Laboratory Standards Institute (CLSI) EP9-A guidelines using EQC samples of the national health care survey supplied by the WIV-ISP (; n = ). Linear regression analysis was performed and bias was estimated at levels of,, and 9 mmol/mol. Imprecision Within-run imprecision was determined by replicates at two concentrations, using blood from a diabetic and a nondiabetic participant. Total imprecision was estimated according to CLSI EP-A guidelines. This was done with single lots of reagents and columns by processing two runs per day during days, using the bilevel Lyphochek controls. Performance data were first evaluated by the Grubbs outlier test, followed by calculation of the mean, standard deviation, and coefficient of variation (CV). Method Comparison Each test instrument was individually compared with the same comparative in-house instrument. Samples over a broad range from diabetic and nondiabetic participants attending the hospital were analyzed over at least operating days. Assessment of between-method outliers was performed by visual inspection of the plots and by statistical outlier testing according to CLSI EP9-A guidelines. Data were analyzed by Passing-Bablok regression analysis and by absolute difference plot. Linearity Linearity was assessed following the CLSI EP-A guidelines by mixing a high and a low sample. The wholeblood samples were hemolyzed first and then proportionally mixed. Linear regression analysis of expected vs observed values obtained in duplicates was then performed, after visual inspection for outliers. Carryover Carryover between blood samples with a low (L) and a high (H) concentration was checked by analyzing them in the following order: L, H, L, H, L, H, L, H, L, H, and L. Results were calculated according to EP-Evaluator Release 9 (David G Rhoads Associates, Kennett Square, PA). Assessment of Interfering Substances As described in CLSI EP-A guidelines, the potential error caused by potentially interfering substances (labile, carbamylated hemoglobin, bilirubin, triglycerides, and hemoglobin content) was compared with an allowable error that was set prior to conducting the interference experiments. According to the criteria of the national health care survey of the WIV-ISP, the allowable error was set at mmol/mol. The influence of hemoglobin variants was also investigated. Labile A c Influence of labile A c was examined by incubating erythrocytes of three patients (one with normal and two with elevated values) with glucose solutions up to 8 mmol/l to increase the labile fraction. Erythrocytes were incubated at C for hours with mixing every minutes. Carbamylated Hemoglobin Erythrocytes from three different patients (one with normal and two with elevated values) were incubated with various concentrations of potassium cyanate (up to mmol/l) for hours at C to achieve in vitro carbamylation of the hemoglobin. Lipemia and Bilirubin Interference of triglycerides and bilirubin was assessed by mixing erythrocytes of two patients (one with normal and one with elevated values) with lipemic plasma and with icteric plasma to achieve final concentrations of mmol/l triglycerides and 8 µmol/l bilirubin, respectively. Hemoglobin Content The effect of the hemoglobin concentration on the value was evaluated by manually diluting one whole-blood sample with a normal value with saline, thus obtaining hemoglobin values ranging from to. g/l. Hemoglobin Variants The capacity of demonstrating the presence of hemoglobin variants and correctly reporting the percentage was investigated using samples containing hemoglobin AS, hemoglobin AC, hemoglobin AD (HbAD), and hemoglobin AE (HbAE). Capillary on Sebia Cap FP Stability of on capillaries during storage and transport in different weather conditions was evaluated by preparing capillary samples from venous blood samples (one normal and two elevated values) and storing them at C to 8 C, room temperature (RT), and C for up to 8 days, with and without Sebia hemolyzing solution. Agreement of venous blood samples and simultaneously obtained capillary blood from the fingertip of healthy volunteers and 8 diabetic patients was also evaluated. Capillary blood samples were analyzed after days of storage at RT in Sebia hemolyzing Downloaded from Am J Clin Pathol ;: DOI:.9/AJCPRUQCJBANSV 89

4 Heylen et al / Evaluation of CAPILLARYS Flex Piercing solution. The allowable bias was set at mmol/mol, according to the criteria of the national health care survey of the WIV-ISP. 8 Statistical Analysis Statistical and graphical analysis was performed with Analyse-it software, version. (Analyse-it Software, Leeds, England) and Microsoft Excel (Microsoft, Redmond, WA). Results Bias Estimation Bias values are shown in Table. Linear regression analysis showed a slope of. and an intercept of.8 for the Tosoh G8 and a slope of. and an intercept of.8 for the Sebia Cap FP. Bias at levels of,, and 9 mmol/mol are shown in Table. The bias increases at higher levels for the Tosoh G8 but still remains acceptable, while Sebia Cap FP shows overall excellent trueness according to the analytical goals from the national health care survey of the WIV-ISP. 8 Imprecision Results for within-run and total imprecision are shown in Table. CVs below.% for both analyzers show excellent precision based on the analytical goals from the national health care survey of the WIV-ISP. 8 Method Comparison Passing-Bablok regression analysis of the data from the comparison between the Sebia analyzer and the Tosoh G Figure A showed a slope (9% confidence interval [CI]) of.9 (.9-.9) and an intercept of. (.8-.). Absolute difference plot Figure B showed a mean difference of. mmol/mol. Passing-Bablok regression analysis for the Tosoh G8 vs Tosoh G Figure C yielded a slope (9% CI) of.98 (.9 to.99) and an intercept of. (.89 to.). Data from the absolute difference plot Figure D showed a mean difference of. mmol/mol when comparing the Tosoh G8 with the Tosoh G. Linearity Linear regression analysis of the data obtained by proportional mixing of hemolysates containing low and high concentration is shown in Table. Linearity was demonstrated between 9 and mmol/mol for both analyzers, with the 9% CI for the intercept and slope comprising and, respectively. Carryover With a carryover of. and. for the Sebia Cap FP and Tosoh G8, respectively, results were below the error Table Overall Performance Characteristics for the Measurement of on the Tosoh G8 and Sebia Cap FP Parameter Tosoh G8 Sebia Cap FP Trueness (EP9), mmol/mol Bias at mmol/mol Bias at mmol/mol Bias at 9 mmol/mol Imprecision (EP) Within-run imprecision, % CV at mmol/mol..8 CV at mmol/mol.. Total imprecision, % CV at mmol/mol.. CV at 8 mmol/mol..8 Linearity (EP) Intercept (9% CI). (. to.). (. to.8) Slope (9% CI). (.98 to.).99 (.9 to.) Carryover Not significant Not significant Interferences (EP) Labile A c (8 mmol/l glucose) Bias < mmol/mol Bias < mmol/mol Carbamylated hemoglobin ( mmol/l KCNO) Bias up to mmol/mol Bias < mmol/mol Bilirubin (8 µmol/l) Bias < mmol/mol Bias < mmol/mol Lipemia (triglycerides mmol/l) Bias < mmol/mol Bias < mmol/mol Hemoglobin content (hemoglobin - g/l) Bias < mmol/mol Bias < mmol/mol Hemoglobin AS HbS separated from other fractions HbS separated from other fractions Hemoglobin AC HbC separated from other fractions HbC separated from other fractions Hemoglobin AD HbD elutes just after HbA HbD separated from other fractions Hemoglobin AE HbE coelutes with HbA HbE separated from other fractions CI, confidence interval; CV, coefficient of variation; HbA, hemoglobin A; HbC, hemoglobin C; HbD, hemoglobin D; HbE, hemoglobin E; HbS, hemoglobin S; KCNO, potassium cyanate. 8 Am J Clin Pathol ;:8-8 Downloaded 8 from DOI:.9/AJCPRUQCJBANSV

5 limit (. for the Sebia Cap FP and. for the Tosoh G8), which indicates the absence of carryover between high and low samples. Assessment of Interferences Assessment of interferences is summarized in Table. Labile A c On the Tosoh G8, peaks identified as labile A c increased as a function of glucose concentration. All samples showed no significant bias at glucose concentrations up to 8 mmol/l. On the Sebia analyzer, values were within ± mmol/ mol of the baseline values for all three patient samples, which is significantly below the total allowable error. Carbamylated Hemoglobin On the Tosoh G8, the in vitro experiments with potassium cyanate showed that the labile fraction increased for all three investigated samples, while the observed bias for the normal and one elevated sample did not exceed the total allowable bias with potassium cyanate concentrations up to mmol/l. However, for the highest sample with a baseline value of mmol/mol, a negative bias of mmol/mol was observed at a potassium cyanate concentration A B Sebia Cap FP ( mmol/mol) Difference (Sebia Cap FP Tosoh G) Tosoh G ( mmol/mol) Mean C D Tosoh G8 ( mmol/mol) Difference (Tosoh G8 Tosoh G) Tosoh G ( mmol/mol) Mean Figure Passing-Bablok regression analysis and absolute difference plot of hemoglobin A c ( ) results, using whole-blood patient samples, obtained with the Sebia Cap FP (Sebia, Lisses, France) (A, B) and the Tosoh G8 (Tosoh Corporation, Tokyo, Japan) (C, D) compared with the Tosoh G (Tosoh Corporation). Passing-Bablok: blue solid line (with 9% confidence interval [CI]) represents the linear curve. Absolute difference plot: blue solid line (with 9% CI) represents mean absolute difference, while light blue dashed lines (with 9% CI) represent 9% limits of agreement. Downloaded from Am J Clin Pathol ;: DOI:.9/AJCPRUQCJBANSV 8

6 Heylen et al / Evaluation of CAPILLARYS Flex Piercing of mmol/l. On Sebia Cap FP, the other HbA fractions increased with increasing potassium cyanate concentrations. However, all values were within ± mmol/mol of the baseline values of all three patient samples. Lipemia and Bilirubin Since measured values were within ± mmol/mol of the baseline values for both analyzers, no interference was observed on a sample with a normal and an elevated value with up to mmol/l triglycerides and 8 µmol/l bilirubin. Hemoglobin Content Dilutions of a patient sample with 9 mmol/mol, generating hemoglobin concentrations from g/l (undiluted) down to. g/l, were assessed for separation profiles, total chromatographic area, and results on the Tosoh G8. This range of dilutions resulted in areas of,9 to units, and all provided good resolution and satisfactory integration without significant bias. The same evaluation was done for the Sebia Cap FP by judging results and total optical density values. At a theoretical hemoglobin concentration of. g/l, there is a significant A Percent AA.8%. min AB.%. min.. F.%.9 min SAC 8.%/ mmol LAC+.%.9 min....8 A 8.%.9 min P.%. min P.%.9 min H-V 8.9%.8 min P.9%.9 min HbS B Percent AA.%. min AB.8%. min.. F.%.8 min SAC.8%/ mmol LAC+.%.9 min....8 A 8.%.9 min P.% min P.8%.8 min H-V.%. min HbC Figure Hemoglobin A c ( ) chromatograms on the Tosoh G8 (Tosoh Corporation, Tokyo, Japan) for some variant hemoglobins (labeled peaks), showing resolution of the separation. Hemoglobin S (HbS) (A), hemoglobin C (HbC) (B). 8 Am J Clin Pathol ;:8-8 Downloaded 8 from DOI:.9/AJCPRUQCJBANSV

7 negative bias of mmol/mol, but adequate resolution and integration without significant bias was obtained for concentrations greater than g/l. Hemoglobin Variants The chromatograms of the Tosoh G8 for the most common hemoglobin variants in our population are shown in Figure. These illustrate that the investigated variant hemoglobins were separated from other fractions, except for hemoglobin D, which elutes just after HbA, and hemoglobin E, which elutes together with the HbA fraction and thus generates a falsely decreased value. On the Sebia Cap FP, all the variants investigated were clearly separated and was quantified reliably Figure. Capillary on the Sebia Cap FP Stability of the result on capillaries with different storage conditions is shown in Figure. 8 The normal and two elevated samples did not exceed ± mmol/ mol after 8 days of storage at C to 8 C and RT. At a normal value of mmol/mol, the bias remained within the allowable range after 8 days of storage at C. C Percent AB.9%. min FP.%. min AA.9%. min.. SAC.9%/ mmol F.8%. min LAC+.%. min....8 A 88.%.9 min H-V.%.98 min HbD P.%. min P.%. min D Percent AA.%. min FP.%. min AB.99%. min.. F.%.9 min SAC.9%/ mmol LAC+.8%.9 min....8 A 9.8%.9 min HbA + HbE Figure (cont) Hemoglobin D (HbD) C), and unresolved hemoglobin E (HbE) coeluting with hemoglobin A (A). D). The red-shaded peak is. AA, hemoglobin A a ; AB, hemoglobin A b ; F, hemoglobin F; LAC, labile. Downloaded from Am J Clin Pathol ;: DOI:.9/AJCPRUQCJBANSV 8

8 Heylen et al / Evaluation of CAPILLARYS Flex Piercing In the sample with an elevated level of mmol/ mol stored at C, however, the positive bias exceeded mmol/mol after days and increased to mmol/mol after 8 days. The sample with the highest concentration (88 mmol/mol) stored at C showed a positive bias of mmol/mol after days, which increased further to mmol/ mol after 8 days. Storage with or without Sebia hemolyzing solution yielded similar results. However, in the latter A B HbA HbA HbC HbS HbF or variant HbA HbA Other c Hb Other Hb Electrophoresis Electrophoresis Fractions % Ref % mmol/mol Ref mmol/mol Fractions % Ref % mmol/mol Ref mmol/mol Other Hb Hb A HbF or variant HbA Other Hb HbA C HbA D HbA HbD HbE HbA Other Hb HbA Other Hb Electrophoresis Electrophoresis Fractions Other Hb HbA HbA % Ref %. mmol/mol Ref mmol/mol. Fractions Other Hb HbA HbA Figure Hemoglobin A c ( ) electropherograms on the Sebia Cap FP (Sebia, Lisses, France) for some variant hemoglobins (labeled peaks), showing the resolution of the separation. Hemoglobin S (HbS) (A), hemoglobin C (HbC) (B), hemoglobin D (HbD) (C), and hemoglobin E (HbE) (D). % Ref %. mmol/mol Ref mmol/mol. 8 Am J Clin Pathol ;:8-8 Downloaded 8 from DOI:.9/AJCPRUQCJBANSV

9 condition, results could no longer be obtained for all three samples after 8 days of storage at C. In these samples, calculation was not possible because of pronounced degradation of the profile with distortion of and/or HbA peaks. Passing-Bablok regression analysis comparing the capillary samples without hemolyzing solution and the venous samples Figure A showed a borderline significant slope and intercept of. (9% CI,. to.) and.8 (9% CI,. to.), respectively. 8 Passing-Bablok regression analysis comparing the capillary samples with hemolyzing solution and the venous samples showed similar results Figure C. The slope and intercept were. (9% CI,. to.) and.9 (9% CI,. to.), respectively. Absolute difference plots Figure B and Figure D showed a good agreement between the venous and capillary finger prick samples. After storage days at RT, none of the analyzed capillary samples stored with the Sebia hemolyzing solution showed a bias greater than mmol/mol while one stored without Sebia hemolyzing solution showed a negative bias greater than mmol/mol as compared with the venous sample. The mean bias observed for capillary samples against venous samples was mmol/mol in both conditions (with or without Sebia hemolyzing solution). Discussion For diagnosis and monitoring of diabetes mellitus, it is crucial to measure with an analyzer that provides precise and accurate results. 9, Reproducibility studies showed that CVs met the stringent requirements of the national health care survey. One should keep in mind that imprecision is expressed in IFCC units and thus should not be compared with imprecision results of other analyzers expressed in DCCT units, since CVs are smaller when using the latter. Estimation of trueness was excellent for the Sebia Cap FP. For the Tosoh G8, the bias tended to increase at higher concentrations but was still acceptable. Linearity was demonstrated over a broad clinical range of values, and no carryover was found for both analyzers. In our experiments, we have shown that the Sebia Cap FP was not subjected to the most common interferences. The Tosoh G8 yielded equally good results for most interferants with a few exceptions. First, there was a significant negative bias at the highest concentration of potassium cyanate, due to the appearance of carbamylated hemoglobin in the peak of labile A c. Since this elevated fraction was not seen in samples of patients undergoing uremic dialysis (data not shown), our experiment was probably conducted with concentrations of potassium cyanate that generate carbamylated hemoglobin A B 8 Days of Storage 8 Days of Storage C 9 8 Figure Stability of capillary hemoglobin A c ( ) during storage with (solid lines) and without (dashed lines) hemolyzing solution on a low (A), elevated (B), and highly elevated (C) sample on the Sebia Cap FP (Sebia, Lisses, France). Blue lines represent storage at C to 8 C, purple lines show storage at room temperature, and green lines show storage at C. Orange dashed lines represent the total allowable error, set at ± mmol/mol. 8 8 Days of Storage Downloaded from Am J Clin Pathol ;: DOI:.9/AJCPRUQCJBANSV 8

10 Heylen et al / Evaluation of CAPILLARYS Flex Piercing concentrations that are not even seen in individuals with endstage renal disease. Nevertheless, we suggest paying attention to samples showing an increase in the labile A c fraction for the Tosoh G8. Second, falsely decreased values could be obtained because HbAD is not fully separated from HbA and because of coelution with HbA of HbAE and glycated hemoglobin S., Diabetologists stress the importance of having an result at the time of the patient s visit to the doctor to improve patient management. This can be performed through point-ofcare testing (POCT) or through home sampling. Analytical performance of POCT instruments has been largely disappointing so far., Moreover, shipment of capillary blood to the central laboratory has the advantage over POCT in that it is measured by the same measurement method as for the venous samples. 9 Storage of capillary samples during several days showed an increasing positive bias at higher temperatures. Although this situation could possibly occur during summertime, it is rather unlikely that these high temperatures arise continuously during the entire mailing process. Furthermore, it is shown that capillary results did not significantly differ from each other when mailed during winter or summer. Our study demonstrates that capillary blood samples can reside at room temperature for at least days before being assayed. In support of the use of capillary blood, there was good agreement A B H Capillary 8 Difference ( H Capillary Venous ) 8 Venous 8 Venous C D +H Capillary 8 Difference (+H Capillary Venous ) 8 8 Venous Venous Figure Passing-Bablok regression analysis and absolute difference plot of capillary hemoglobin A c ( ) without (A, B) and with (C, D) hemolyzing solution compared with venous on the Sebia Cap FP (Sebia, Lisses, France). Green dots represent nondiabetic patient samples, while red dots are the diabetic patient samples. Passing-Bablok: blue solid line (with 9% confidence interval [CI]) represents the linear regression curve. Absolute difference plot: blue solid line (with 9% CI) represents mean absolute difference, while orange dashed lines represent the total allowable error, set at ± mmol/mol. 8 8 Am J Clin Pathol ;:8-8 Downloaded 8 from DOI:.9/AJCPRUQCJBANSV

11 between the venous and capillary samples. In fact, only one of 8 samples exceeded the allowable bias. The collection of this capillary blood sample from a healthy volunteer went more difficult than usual, and the sample was stored without hemolyzing solution. We conclude that capillary samples taken at home for measurement on the Sebia Cap FP are an acceptable, easy, and inexpensive alternative for venous blood samples when preanalytical factors such as mailing time and temperature are well controlled. This can provide the clinician with timely and important information at the moment of the patient s consultation without the need for an extra visit for a phlebotomy and with state-of-the-art allowable error. This strategy should be particularly beneficial for diabetic children and the elderly. Address reprint requests to Dr Weets: Dept of Clinical Chemistry, Universitair Ziekenhuis Brussel, Laarbeeklaan, Brussels, Belgium; ilse.weets@uzbrussel.be. Tosoh Bioscience Belgium and Sebia Benelux provided the Tosoh G8 and CAPILLARYS Flex Piercing analyzers and reagents used in this evaluation. The evaluation of capillary sampling was part of a Sebia R&D development project. References. UK Prospective Diabetes Study (UKPDS) Group. Effect of intensive blood-glucose control with metformin on complications in overweight patients with type diabetes (UKPDS ). Lancet. 998;:8-8.. Diabetes Control and Complications Trial Research Group. The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. N Engl J Med. 99;: Gorus F, Mathieu C, Gerlo E. How should HbAc measurements be reported? Diabetologia. ;9:-.. American Diabetes Association. Standards of medical care in diabetes. Diabetes Care. ;(suppl):s-s.. Jeppsson JO, Kobold U, Barr J et al. Approved IFCC reference method for the measurement of HbAc in human blood. Clin Chem Lab Med. ;: Little RR, Rohlfing CL, Sacks DB; National Glycohemoglobin Standardization Program (NGSP) Steering Committee. Status of hemoglobin A c measurement and goals for improvement: from chaos to order for improving diabetes care. Clin Chem. ;:-.. Weykamp C, Waenink-Wieggers H, Kemna E, et al. HbAc: performance of the Sebia CAPILLARYS Flex Piercing. Clin Chem Lab Med. ;:e9-e. 8. Chapelle JP, Teixeira J, Maisin D, et al. Multicentre evaluation of the Tosoh HbAc G8 analyser. Clin Chem Lab Med. ;8:-. 9. Fokkema MR, Bakker AJ, de Boer F, et al. HbAc measurements from dried blood spots: validation and patient satisfaction. Clin Chem Lab Med. 9;:9-.. Bognetti ER, Zoja A, Ciralli F, et al. HbAc determination on capillary blood sample: validity, stability, and potential usefulness. Diabetes Care. 99;8:-.. Jaisson S, Leroy N, Meurice J, et al. First evaluation of Capillarys Flex Piercing (Sebia) as a new analyzer for HbAc assay by capillary electrophoresis. Clin Chem Lab Med. ;:9-.. Marinova M, Altinier S, Caldini A, et al. Multicenter evaluation of hemoglobin Ac assay on capillary electrophoresis. Clin Chim Acta. ;:-.. Clinical and Laboratory Standards Institute (CLSI). Method Comparison and Bias Estimation Using Patient Samples: Approved Guideline Second Edition (Interim Revision). Wayne, PA: CLSI;. CLSI document EP9-A-IR.. National Committee for Clinical Laboratory Standards (NCCLS). Evaluation of Precision Performance of Quantitative Measurement Methods: Approved Guideline. Wayne, PA: NCCLS;. NCCLS document EP-A.. Grubbs FE. Procedures for detecting outlying observations in samples. Technometrics. 99;:-.. National Committee for Clinical Laboratory Standards (NCCLS). Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach: Approved Guideline. Wayne, PA: NCCLS;. NCCLS document EP-A.. Clinical and Laboratory Standards Institute (CLSI). Interference Testing in Clinical Chemistry: Approved Guideline Second Edition. Wayne, PA: CLSI;. CLSI document EP-A. 8. Wetenschappelijk Instituut Volksgezondheid. WIV information for analytical goals for measurement of HbAc. quality/rapports/_down/hbac//hbac-jaarrapport-. pdf. Accessed July,. 9. American Diabetes Association. Implications of the Diabetes Control and Complications Trial. Diabetes Care. ;(suppl):s-s.. Schindhelm RK, Lenters-Westra E, Slingerland RJ. Glycated haemoglobin A(c) (HbA(c)) in the diagnosis of diabetes mellitus: don t forget the performance of the HbA(c) assay. Diabetes Med. ;:-.. Weykamp CW, Mosca A, Gillery P, et al. The analytical goals for hemoglobin A(c) measurement in IFCC units and National Glycohemoglobin Standardization Program Units are different. Clin Chem. ;:-.. Weykamp CW, Penders TJ, Muskiet FA, et al. Influence of hemoglobin variants and derivatives on glycohemoglobin determinations, as investigated by laboratories using methods. Clin Chem. 99;9:-.. Aleyassine H. Low proportions of glycosylated hemoglobin associated with hemoglobin S and hemoglobin C. Clin Chem. 99;:8-8.. Lenters-Westra E, Slingerland RJ. Six of eight hemoglobin Ac point-of-care instruments do not meet the general accepted analytical performance criteria. Clin Chem. ;:-.. Shephard M, Whiting M. Assessment of the practicability and analytical performance of a point-of-care affinity chromatography haemoglobin Ac analyser for use in the non-laboratory setting. Ann Clin Biochem. ;:-.. Bognetti ER, Zoja A, Cogliardi A, et al. HbAc assay on capillary blood sample collected at home: a reliable method. Diabetes Care. 99;9:-8. Downloaded from Am J Clin Pathol ;: DOI:.9/AJCPRUQCJBANSV 8

assay Introduction Conclusions: The D-100 TM system proved to be a robust and reliable method for HbA 1c

assay Introduction Conclusions: The D-100 TM system proved to be a robust and reliable method for HbA 1c Clin Chem Lab Med 215; aop Stéphane Jaisson*, Nathalie Leroy, Emmanuelle Guillard, Aurore Desmons and Philippe Gillery Analytical performances of the D- TM hemoglobin testing system (Bio-Rad) for assay

More information

Tosoh Automated Glycohemoglobin Analyzer HLC-723GX

Tosoh Automated Glycohemoglobin Analyzer HLC-723GX Tosoh Automated Glycohemoglobin Analyzer HLC-723GX The Diabetes Epidemic and the role of HbA 1c Diabetes is recognised worldwide as a disease that is reaching epidemic proportions. (1) IDF region Adult

More information

Hemoglobin A1c: Comparison of Pointe Scientific s 2-Part Direct Hemoglobin A1c with the Bio-Rad Variant II and the Tosoh G8

Hemoglobin A1c: Comparison of Pointe Scientific s 2-Part Direct Hemoglobin A1c with the Bio-Rad Variant II and the Tosoh G8 Hemoglobin A1c: Comparison of Pointe Scientific s 2-Part Direct Hemoglobin A1c with the Bio-Rad Variant II and the Tosoh G8 Joseph D. Artiss, Ph.D., FACB, Associate Professor of Pathology, Wayne State

More information

HbA1c enzymatic assay evaluation at AN academic

HbA1c enzymatic assay evaluation at AN academic Peer reviewed ORIGINAL ARTICLE HbA1c enzymatic assay evaluation at AN academic laboratory B Phiri-Ramongane Mmed Chem Path, FCP Path(SA)-Chem Department of Chemical Pathology, Sefako Makgatho Health Sciences

More information

HPLC ANALYZER. Gold Standard Accuracy by Ion-Exchange HbA1c. First HPLC Analyzer to Aid in Diagnosis of Diabetes

HPLC ANALYZER. Gold Standard Accuracy by Ion-Exchange HbA1c. First HPLC Analyzer to Aid in Diagnosis of Diabetes HPLC ANALYZER Gold Standard Accuracy by Ion-Exchange HbA1c First HPLC Analyzer to Aid in Diagnosis of Diabetes TOSOH BIOSCIENCE www.tosohbioscience.us HPLC ANALYZER HbA1c analysis in 1.6 minutes Direct

More information

Effects of hemoglobin C, D, E and S traits on measurements of hemoglobin A1c by twelve methods

Effects of hemoglobin C, D, E and S traits on measurements of hemoglobin A1c by twelve methods Effects of hemoglobin C, D, E and S traits on measurements of hemoglobin A1c by twelve methods Curt Rohlfing, University of Missouri Steven Hanson, University of Missouri Cas Weykamp, Location Queen Beatrix

More information

Evaluation Report. Eurolyser HbA1c Test Kit on Smart and Cube Analyser

Evaluation Report. Eurolyser HbA1c Test Kit on Smart and Cube Analyser Evaluation Report Eurolyser HbA1c Test Kit on Smart and Cube Analyser Locations Location 1: Klinisk kemi University Hospital MAS Malmö Operator: Dr. Benny Larsson Date: June/July 2007 Location 2: Eurolyser

More information

determination Department of Laboratory Science, Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangdong, China (LA 1c

determination Department of Laboratory Science, Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangdong, China (LA 1c Original papers A comparative evaluation of the analytical performances of Capillarys 2 Flex Piercing, Tosoh HLC-723 G8, Premier Hb9210, and Roche Cobas c501 Tina-quant Gen 2 analyzers for determination

More information

State of the Art of HbA1c Measurement

State of the Art of HbA1c Measurement State of the Art of HbA1c Measurement XXI Congreso Latinoamericano de Patologia Clinica Y XLII Congreso Mexicano de Patologia Clinica, Cancun October 2012 Randie R. Little, Ph.D. NGSP Network Laboratory

More information

Analysis of Hemoglobin A1c from Dried Blood Spot Samples with the Tina-quant II Immunoturbidimetric Method

Analysis of Hemoglobin A1c from Dried Blood Spot Samples with the Tina-quant II Immunoturbidimetric Method Journal of Diabetes Science and Technology Volume 4, Issue 2, March 2010 Diabetes Technology Society SYMPOSIUM Analysis of Hemoglobin A1c from Dried Blood Spot Samples with the Tina-quant II Immunoturbidimetric

More information

Evaluation of haemoglobin A1c measurement by an enzymatic method using an automated analyser that has an on-board haemolysis system

Evaluation of haemoglobin A1c measurement by an enzymatic method using an automated analyser that has an on-board haemolysis system Original Article Annals of Clinical Biochemistry 50(5) 443 449! The Author(s) 2013 Reprints and permissions: sagepub.co.uk/journalspermissions.nav DOI: 10.1177/0004563213476859 acb.sagepub.com Evaluation

More information

Evaluation of a new hemoglobin A1c analyzer for point- of- care testing

Evaluation of a new hemoglobin A1c analyzer for point- of- care testing Received: 19 September 2016 Accepted: 17 January 2017 DOI: 10.1002/jcla.22172 RESEARCH ARTICLE Evaluation of a new hemoglobin A1c analyzer for point- of- care testing Rui Zhou 1 Wei Wang 2 Zhi-Xin Song

More information

Bio-Rad Laboratories. Hemoglobin Testing. VARIANT II Hemoglobin Testing System For HbA 1c

Bio-Rad Laboratories. Hemoglobin Testing. VARIANT II Hemoglobin Testing System For HbA 1c Bio-Rad Laboratories Hemoglobin Testing VARIANT II Hemoglobin Testing System For HbA 1c Bio-Rad CDM System CDM 5.1 VII Instrument Patient Report V2_A1c_NU Patient Data Sample ID: AD Patient ID: Name: Physician:

More information

Evidence Based HbA1c Accuracy. Sabrina Koetsier Medical Scientist RCPAQAP Chemical Pathology

Evidence Based HbA1c Accuracy. Sabrina Koetsier Medical Scientist RCPAQAP Chemical Pathology Evidence Based HbA1c Accuracy Sabrina Koetsier Medical Scientist RCPAQAP Chemical Pathology Fresh lycohaemoglobin This program is designed to assist laboratories to assess the accuracy of their hihba1c

More information

META -ANALYSIS. Hb A1c by Sebia Capillary Electrophoresis. Analytical performances meta-analysis Second Edition

META -ANALYSIS. Hb A1c by Sebia Capillary Electrophoresis. Analytical performances meta-analysis Second Edition META -ANALYSIS Hb A1c by Sebia Capillary Electrophoresis Analytical performances meta-analysis Second Edition INTRODUCTION [Analytical performances of Hb A1c by Sebia capillary electrophoresis - Meta-analysis]

More information

(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%)

(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%) cobas b system - performance evaluation Study report from a multicenter evaluation of the new cobas b system for the measurement of HbAc and lipid panel Introduction The new cobas b system provides a point-of-care

More information

Calibration of ion-exchange HPLC measurements of glycohemoglobin: effect on interassay precision

Calibration of ion-exchange HPLC measurements of glycohemoglobin: effect on interassay precision Clinical Chemistry 43:12 2353 2357 (1997) Endocrinology and Metabolism Calibration of ion-exchange HPLC measurements of glycohemoglobin: effect on interassay precision Erik Gerlo* and Frans Gorus To investigate

More information

REVIEW Global standardisation of HbA 1c

REVIEW Global standardisation of HbA 1c Malaysian J Pathol 2008; 30(2) : 67 71 REVIEW Global standardisation of Leslie C LAI, FRCP, FRCPath Gleneagles Intan Medical Centre, Kuala Lumpur, Malaysia Abstract is used for assessing glycaemic control

More information

Detection and quantitation of hemoglobin (Hb) fractions

Detection and quantitation of hemoglobin (Hb) fractions Evaluation of a Dual Hemoglobin A 2 /A 1c Quantitation Kit on the Bio-Rad Variant II Automated Hemoglobin Analyzer John D. Lafferty, MLT, ART; Andrew G. McFarlane, MLT, ART; David H. K. Chui, MD, FRCP

More information

Defining HbA1c: the indispensable decision to approach measurement standardization

Defining HbA1c: the indispensable decision to approach measurement standardization Defining HbA1c: the indispensable decision to approach measurement standardization Jan-Olof Jeppsson IFCC WG for HbA1c standardization Clinical Chemistry, Malmö University Hospital Sweden 6 November, 2007

More information

The influence of sample freezing at 80 C for 2 12 weeks on glycated haemoglobin (HbA 1c

The influence of sample freezing at 80 C for 2 12 weeks on glycated haemoglobin (HbA 1c Original papers The influence of sample freezing at 80 C for 2 12 weeks on glycated haemoglobin ( ) method on Bio-Rad D-10 auto analyzer Katarzyna Bergmann*, Grazyna Sypniewska Department of Laboratory

More information

Research Article Hb A1c Separation by High Performance Liquid Chromatography in Hemoglobinopathies

Research Article Hb A1c Separation by High Performance Liquid Chromatography in Hemoglobinopathies Scientifica Volume 2016, Article ID 2698362, 4 pages http://dx.doi.org/10.1155/2016/2698362 Research Article Hb A1c Separation by High Performance Liquid Chromatography in Hemoglobinopathies Vani Chandrashekar

More information

Sandra Božičević dipl. ing. med. biochem, Marijana Vučić Lovrenčić, PhD Vuk Vrhovac University Clinic, Zagreb, Croatia

Sandra Božičević dipl. ing. med. biochem, Marijana Vučić Lovrenčić, PhD Vuk Vrhovac University Clinic, Zagreb, Croatia 16. PRE-ANALYTICAL, ANALYTICAL AND POST-ANALYTICAL FACTORS INFLU- ENCING SPECIFIC TESTS FOR DIAGNOSIS AND MONITORING OF DM-National network in quality assessment Prof. Dubravka Juretić, Ph.D. University

More information

HbA1c3 PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE

HbA1c3 PRINCIPLE REF B ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE SYNCHRON System(s) Chemistry Information Sheet Copyright 2014 Beckman Coulter, Inc. HbA1c3 Hemoglobin A1c3 REF B36415 For In Vitro Diagnostic Use Rx Only ANNUAL REVIEW Reviewed by Date Reviewed by Date

More information

COMPARISON OF CATION EXCHAGE HPLC AND IMMUNOTURBIDIMETRIC METHOD FOR DETERMINATION OF HbA1c

COMPARISON OF CATION EXCHAGE HPLC AND IMMUNOTURBIDIMETRIC METHOD FOR DETERMINATION OF HbA1c D:\Biomedica Vol.27, Jul. Dec. 2011\Bio-1.Doc P. 161 165 (KC) IV COMPARISON OF CATION EXCHAGE HPLC AND IMMUNOTURBIDIMETRIC METHOD FOR DETERMINATION OF HbA1c FARZANA YASMEEN, ASIM MUMTAZ, SALEEM-UZ-ZAMAN

More information

Aina Blood Monitoring System

Aina Blood Monitoring System Aina Blood Monitoring System Analytical Performance Summary The Aina Blood Monitoring System is a high quality and versatile multi-parameter diagnostic platform that is CE-marked and approved for sale

More information

Improving the Accuracy of Hemoglobin A 1c. : Your Help Is Needed

Improving the Accuracy of Hemoglobin A 1c. : Your Help Is Needed Improving the Accuracy of Hemoglobin A 1c : Your Help Is Needed Kristina Jackson Behan, PhD, MT(ASCP) (Clinical Laboratory Sciences Program, University of West Florida, Pensacola, FL) DOI: 10.1309/RTVTWPYPTPA5YAGU

More information

Auto HbA1C ANALYZER HbA1LC HPLC 3660

Auto HbA1C ANALYZER HbA1LC HPLC 3660 Turnkey Laboratories Solutions Auto HbA1C ANALYZER HbA1LC HPLC 3660 Viewable Accuracy Real Time Summarized Chromatogram Authentic Results EPC / PRODUCTS / APPLICATION / SOFTWARE / ACCESSORIES / CONSUMABLES

More information

Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000

Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000 v2-jan214 Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4 Intended Use The canine CRP immunoassay on Abbott's Architect c4 is an in vitro diagnostic test for quantitative determination

More information

THE HEALTH AND RETIREMENT STUDY: INVESTIGATION OF DRIED BLOOD SPOT HBA1C MEASUREMENT ERRORS USING TIPTEMP STAMPS

THE HEALTH AND RETIREMENT STUDY: INVESTIGATION OF DRIED BLOOD SPOT HBA1C MEASUREMENT ERRORS USING TIPTEMP STAMPS : INVESTIGATION OF DRIED BLOOD SPOT HBA1C MEASUREMENT ERRORS USING TIPTEMP STAMPS April 29, 2015 PAA Biomarker Network Meeting Jessica Faul, Ph.D., M.P.H. Health and Retirement Study Survey Research Center

More information

Maintaining Quality in Laboratory Medicine. Glycated Haemoglobin Scheme Guide

Maintaining Quality in Laboratory Medicine. Glycated Haemoglobin Scheme Guide Maintaining Quality in Laboratory Medicine Glycated Haemoglobin Scheme Guide Contents 1. Scheme Details and repertoire 1.1 Source Material and Serum Integrity 3 1.2 List of Analytes and Frequency of Distribution

More information

HbA1c and Diabetes. Steven Weier Senior Lecturer School of Biomedical Sciences Faculty of Health QUT. CRICOS No J

HbA1c and Diabetes. Steven Weier Senior Lecturer School of Biomedical Sciences Faculty of Health QUT. CRICOS No J HbA1c and Diabetes Steven Weier Senior Lecturer School of Biomedical Sciences Faculty of Health QUT Queensland University of Technology History of HbA1c 1966: Holmquist and Schroeder identify five subtypes

More information

MALDI-TOF analysis of whole blood: its usefulness and potential in the assessment of HbA1c levels

MALDI-TOF analysis of whole blood: its usefulness and potential in the assessment of HbA1c levels MALDI-TOF analysis of whole blood: its usefulness and potential in the assessment of HbA1c levels Jane Y. Yang, David A. Herold Department of Pathology, University of California San Diego, 9500 Gilman

More information

HbA1c PRINCIPLE REF ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE

HbA1c PRINCIPLE REF ANNUAL REVIEW Reviewed by. Date. Date INTENDED USE SYNCHRON System(s) Chemistry Information Sheet Copyright 2013 Beckman Coulter, Inc. Hemoglobin A1c REF 650262 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by Date Reviewed by Date PRINCIPLE INTENDED

More information

References Required document for Laboratory Accreditation by the College of American Pathologists.

References Required document for Laboratory Accreditation by the College of American Pathologists. Subject Hemoglobin A1c Cobas c501 Index Number Lab-4004 Section Laboratory Subsection Chemistry Category Departmental Contact Benjamin Michel Last Revised 9/12/2016 References Required document for Laboratory

More information

Commentary by R. Little, Ph.D., NGSP Network Coordinator for the NGSP Steering Committee

Commentary by R. Little, Ph.D., NGSP Network Coordinator for the NGSP Steering Committee College of American Pathologists (CAP) GH5 Survey Data: (updated 8/17) The American Diabetes Association (ADA) recommends that The A1C test should be performed using a method that is certified by the NGSP.

More information

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY

USING THE ACCESS AMH ASSAY IN YOUR LABORATORY INFORMATION BULLETIN USING THE ACCESS AMH ASSAY IN YOUR LABORATORY ///////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////

More information

Commentary by R. Little, Ph.D., NGSP Network Coordinator for the NGSP Steering Committee

Commentary by R. Little, Ph.D., NGSP Network Coordinator for the NGSP Steering Committee College of American Pathologists (CAP) GH5 Survey Data: (updated 12/17) The American Diabetes Association (ADA) recommends that The A1C test should be performed using a method that is certified by the

More information

Journal of Diabetes and Metabolic Disorders; 2011; Vol 10, pp 1-6. Keywords: Glycated hemoglobin, Paper filter, Diabetes, Dried blood spot.

Journal of Diabetes and Metabolic Disorders; 2011; Vol 10, pp 1-6. Keywords: Glycated hemoglobin, Paper filter, Diabetes, Dried blood spot. Journal of Diabetes and Metabolic Disorders; 2011; Vol 10, pp 1-6 Glycated hemoglobin measurements from dried blood spots: reliability and relation to results obtained from whole blood samples Ozra Tabatabaei-Malazy

More information

Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry.

Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry. 2013/07/30 A93A01315AUS A11A01641 60 ml 15 ml Pentra C400 Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry. QUAL-QA-TEMP-0846

More information

Selecting a Hemoglobin A1C Point of Care Instrument. Objectives. Disclosure / Conflict of Interest 3/11/2015

Selecting a Hemoglobin A1C Point of Care Instrument. Objectives. Disclosure / Conflict of Interest 3/11/2015 Selecting a Hemoglobin A1C Point of Care Instrument Heather P. Whitley, PharmD, BCPS, CDE Associate Clinical Professor of Pharmacy Practice Auburn University Harrison School of Pharmacy Objectives Compare

More information

Glycosylated Hemoglobin A1c (HbA1c)

Glycosylated Hemoglobin A1c (HbA1c) APPROVED BY:... DR WAN MOHD IZANI WAN MOHAMED ENDOCRINE LABORATORY DIRECTOR CONTROLLED COPY NO: 0 REGISTERED HOLDER ENDOCRINOLOGY LABORATORY Glycosylated Hemoglobin A1c (HbA1c) PREPARED BY DESIGNATION

More information

Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100)

Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on. CUBE analyser (CA0100) Evaluation Report: Eurolyser CRP test (ST0100 and ST0102) on CUBE analyser (CA0100) Location Location: Eurolyser Diagnostica GmbH Operators: Simone Wieser; Franz Helminger; Michael Gruber Date: July-November

More information

Biomedical Research 2017; 28 (4):

Biomedical Research 2017; 28 (4): Biomedical Research 2017; 28 (4): 1765-1769 ISSN 0970-938X www.biomedres.info Analytical calibre of high performance liquid chromatography and ion exchange chromatography resin methods in estimation of

More information

Rapid Quantitation of Hemoglobin S in Sickle Cell Patients Using the Variant II Turbo Analyzer

Rapid Quantitation of Hemoglobin S in Sickle Cell Patients Using the Variant II Turbo Analyzer 32 Available online at www.annclinlabsci.org Annals of Clinical & Laboratory Science, vol. 39, no. 1, 2009 Rapid Quantitation of Hemoglobin S in Sickle Cell Patients Using the Variant II Turbo Analyzer

More information

Clinical Chemistry 54: (2008)

Clinical Chemistry 54: (2008) Clinical Chemistry 54:8 1379 1385 (2008) Informatics and Statistics Statistical Methods for Monitoring the Relationship between the IFCC Reference Measurement Procedure for Hemoglobin A 1c and the Designated

More information

In Vitro Diagnostic Medical Device. Store at 2-8 C. Store at C. CAUTION: Consult accompanying documents

In Vitro Diagnostic Medical Device. Store at 2-8 C. Store at C. CAUTION: Consult accompanying documents E AxSYM HbA1c 3L93-20 ABOL051/R2 Read Highlighted Changes Revised March, 2008 AxSYM HbA1c Customer Service United States: 1-877-4ABBOTT International: Call your Abbott Representative This package insert

More information

The Integrated Cardiovascular Clinical Network CHSA

The Integrated Cardiovascular Clinical Network CHSA The Integrated Cardiovascular Clinical Network CHSA Roche cobas b 101 HbA1c and Lipids Evaluation July 2013 Introduction Cardiovascular disease (CVD) is the leading cause of mortality in Australia, accounting

More information

HbA1c: A Review of Analytical and Clinical Aspects

HbA1c: A Review of Analytical and Clinical Aspects Review Article Clinical Chemistry Ann Lab Med 2013;33:1-8 http://dx.doi.org/10.3343/alm.2013.33.6.1 ISSN 2234-3806 eissn 2234-3814 HbA1c: A Review of Analytical and Clinical Aspects Cas Weykamp, Ph.D.

More information

Gentian Canine CRP Immunoassay Application Note for scil VitroVet*

Gentian Canine CRP Immunoassay Application Note for scil VitroVet* v02-jan2014 Gentian Canine CRP Immunoassay Application Note for scil VitroVet* Intended Use The canine CRP immunoassay on scil VitroVet is an in vitro diagnostic test for quantitative determination of

More information

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats

IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats IDEXX Catalyst One Chemistry Analyzer for In-house Measurement of Total Thyroxine (TT 4 ) Concentration in Serum from Dogs and Cats Authors: Kate Cote, Ph.D., Graham Bilbrough, MA, VetMB, CertVA, MRCVS,

More information

Direct HbA1c testing capabilities on the RX modena

Direct HbA1c testing capabilities on the RX modena Direct testing capabilities on the RX modena INTRODUCTION What is? The term refers to glycated haemoglobin. It develops when haemoglobin, a protein within red blood cells that carries oxygen throughout

More information

Rat Hemoglobin A1c (HbA1c) Kit Instructions

Rat Hemoglobin A1c (HbA1c) Kit Instructions V.3 Crystal Chem Rat Hemoglobin A1c (HbA1c) Kit Instructions For the quantitative determination of hemoglobin A1c (HbA1c) in rat whole blood Catalog #80300 96 Assays For research use only. Not for use

More information

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014

1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014 Comparison Study Summary Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR 97123 July 30, 2014 1. PROTOCOL This study was conducted on July 24 th, 2014 at Tuality Healthcare, Hillsboro, OR. The

More information

Determination of hemoglobin is one of the most commonly

Determination of hemoglobin is one of the most commonly ORIGINAL ARTICLE Multiple-Site Analytic Evaluation of a New Portable Analyzer, HemoCue Hb 201+, for Point-of-Care Testing Sten-Erik Bäck, PhD,* Carl G. M. Magnusson, PhD, Lena K. Norlund, MD, PhD, Henning

More information

College of American Pathologists (CAP) GH2 Survey Data: (updated 12/09) 2009 GH2-B (fresh pooled samples) * = NGSP certified at the time of the survey

College of American Pathologists (CAP) GH2 Survey Data: (updated 12/09) 2009 GH2-B (fresh pooled samples) * = NGSP certified at the time of the survey College of American Pathologists (CAP) GH2 Survey Data: (updated 12/09) The American Diabetes Association (ADA) recommends that laboratories use only HbA1c assay methods that have been NGSP certified and

More information

HbA1c: what the GP needs to know

HbA1c: what the GP needs to know North West London Pathology HbA1c: what the GP needs to know Dr Shivani Misra MRCP, FRCPath, PhD Consultant in Metabolic Medicine & Chemical Pathology s.misra@nhs.net Twitter: @ShivaniM_KC Learning Objectives

More information

Professional Procedure Guide (90821B 04/2008)

Professional Procedure Guide (90821B 04/2008) Professional Procedure Guide (90821B 04/2008) This test is WAIVED under the Clinical Laboratory Improvement Amendments of 1988 (CLIA). If a laboratory modifies the test instructions, the test will no longer

More information

liquicolor (AMP Buffer, IFCC) Design Verification

liquicolor (AMP Buffer, IFCC) Design Verification Design Verification (AMP Buffer, IFCC) liquicolor 1 Introduction... 2 2 Imprecision... 2 3 arity and Detection Limit... 2 3.1 arity... 2 3.2 Detection Limit... 3 4 Comparison of Methods... 3 5 Stability...

More information

Quantitative Measurement of HbA lc by an Immunoturbidimetric Assay Compared to a Standard HPLC Method

Quantitative Measurement of HbA lc by an Immunoturbidimetric Assay Compared to a Standard HPLC Method CLINICAL CHEMISTRY Original Article, Quantitative Measurement of HbA lc by an Immunoturbidimetric Assay Compared to a Standard HPLC Method AHMAD HAMWI, MD, CHRISTIAN R. SCHWEIGER, MD, MARIO VEITL, MD,

More information

CAPILLARYS HEMOGLOBIN

CAPILLARYS HEMOGLOBIN CAPILLARYS HEMOGLOBIN TRAINING MANUAL Sebia, Inc. 1-800-835-6497 Copyright 2005 Sebia, Inc. All rights reserved. The content of this manual may change without notice. This manual may not be reprinted or

More information

LIPASE liquicolor. Design Verification. Multipurpose Reagent

LIPASE liquicolor. Design Verification. Multipurpose Reagent Design Verification LIPASE liquicolor Multipurpose Reagent CONTENTS 1 Introduction... 2 2 Imprecision... 2 3 Linearity and Detection Limit... 2 3.1 Linearity... 2 3.2 Detection Limit... 3 4 Recovery of

More information

Dubrovnik 2014 Sverre Sandberg, Norwegian Quality Improvement of Primary Care Laboratories Noklus, Bergen, Norway

Dubrovnik 2014 Sverre Sandberg, Norwegian Quality Improvement of Primary Care Laboratories Noklus, Bergen, Norway POC testing instruments for diagnosing and monitoring diabetes in clinical settings Dubrovnik 2014 Sverre Sandberg, Norwegian Quality Improvement of Primary Care Laboratories Noklus, Bergen, Norway Does

More information

Hemoglobinopathy and thalassemia detection

Hemoglobinopathy and thalassemia detection Hemoglobinopathy and thalassemia detection Traditional methods and a novel method capillary electrophoresis technology By Aigars Brants, PhD Published October 2011 In humans, two pairs of unlike globin

More information

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013

System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Abstract Objective: The goal of the study was to assess

More information

In normal human erythrocytes,

In normal human erythrocytes, Journal of the association of physicians of india JANUARY 2014 VOL. 62 9 Editorial HbA1c Result, Does it Depend Upon the Testing Methods? Samar Banerjee * * Dept of Medicine, Vivekananda Institute of Medical

More information

Comparison of Sebia Capillarys Capillary Electrophoresis With the Primus High-Pressure Liquid Chromatography in the Evaluation of Hemoglobinopathies

Comparison of Sebia Capillarys Capillary Electrophoresis With the Primus High-Pressure Liquid Chromatography in the Evaluation of Hemoglobinopathies Hematopathology / Evaluation of Hemoglobinopathies Comparison of Sebia Capillarys Capillary Electrophoresis With the Primus High-Pressure Liquid Chromatography in the Evaluation of Hemoglobinopathies David

More information

BÜHLMANN fcal turbo. Calprotectin turbidimetric assay for professional use. Reagent Kit B-KCAL-RSET. Revision date:

BÜHLMANN fcal turbo. Calprotectin turbidimetric assay for professional use. Reagent Kit B-KCAL-RSET. Revision date: BÜHLMANN fcal turbo Calprotectin turbidimetric assay for professional use Reagent Kit B-KCAL-RSET Revision date: 2017-02-24 BÜHLMANN LABORATORIES AG Baselstrasse 55 4124 Schönenbuch, Switzerland Tel.:

More information

EXTERNAL QUALITY ASSESSMENT FOR HAEMOGLOBIN A 2

EXTERNAL QUALITY ASSESSMENT FOR HAEMOGLOBIN A 2 EXTERNAL QUALITY ASSESSMENT FOR HAEMOGLOBIN A 2 Dr Barbara abaawild The importance of Hb A 2 measurement Accurate and reliable measurement of Hb A 2 is essential for the diagnosis of beta thalassaemia

More information

Unraveling Hemoglobinopathies with Capillary Electrophoresis

Unraveling Hemoglobinopathies with Capillary Electrophoresis Session Number 2002 Unraveling Hemoglobinopathies with Capillary Electrophoresis David F. Keren, M.D. Professor of Pathology Division Director, Clinical Pathology The University of Michigan dkeren@med.umich.edu

More information

A validation study of after reconstitution stability of diabetes: level 1 and diabetes level 2 controls

A validation study of after reconstitution stability of diabetes: level 1 and diabetes level 2 controls A validation study of after reconstitution stability of diabetes: level 1 and diabetes level 2 controls Shyamali Pal R B Diagnostic Private Limited, Lake Town, Kolkata, India INFO ABSTRACT Corresponding

More information

DIRECT ENZYMATIC HbA1c ASSAY

DIRECT ENZYMATIC HbA1c ASSAY DIRECT ENZYMATIC HbA1c ASSAY ACCURATE AND SIMPLE DIAZYME INNOVATIONS IN CLINICAL DIAGNOSTICS About Diazyme Diazyme Laboratories is a Division of General Atomics located in Poway, California. Diazyme uses

More information

Selecting a Risk-Based SQC Procedure for a HbA1c Total QC Plan

Selecting a Risk-Based SQC Procedure for a HbA1c Total QC Plan 729488DSTXXX10.1177/1932296817729488Journal of Diabetes Science and TechnologyWestgard et al research-article2017 Original Article Selecting a Risk-Based SQC Procedure for a HbA1c Total QC Plan Journal

More information

SCIEX Vitamin D 200M Assay for the Topaz System

SCIEX Vitamin D 200M Assay for the Topaz System The First FDA-Cleared LC-MS/MS Assay for Vitamin D SCIEX Vitamin D 200M Assay for the Topaz System The first FDA-cleared LC-MS/MS assay for Vitamin D Vitamin D is an important building block for human

More information

METHOD VALIDATION: WHY, HOW AND WHEN?

METHOD VALIDATION: WHY, HOW AND WHEN? METHOD VALIDATION: WHY, HOW AND WHEN? Linear-Data Plotter a,b s y/x,r Regression Statistics mean SD,CV Distribution Statistics Comparison Plot Histogram Plot SD Calculator Bias SD Diff t Paired t-test

More information

Performance Characteristics of a New Single-Sample Freezing Point Depression Osmometer

Performance Characteristics of a New Single-Sample Freezing Point Depression Osmometer AACC, July 2007 Performance Characteristics of a New Single-Sample Freezing Point Depression Osmometer E. Garry, M. Pesta, N. Zampa Advanced Instruments, Inc., Norwood, MA 02062 AbstracT The Advanced Model

More information

Use of Fructosyl Peptide Oxidase for HbA1c Assay

Use of Fructosyl Peptide Oxidase for HbA1c Assay 569573DSTXXX10.1177/1932296815569573Journal of Diabetes Science and TechnologyYonehara et al research-article2015 Symposium Use of Fructosyl Peptide Oxidase for HbA1c Assay Journal of Diabetes Science

More information

ABX Pentra. HbA1c WB. Reagent for quantitative in-vitro determination of Hemoglobin A1c (HbA1c). A11A01702

ABX Pentra. HbA1c WB. Reagent for quantitative in-vitro determination of Hemoglobin A1c (HbA1c). A11A01702 (Whole blood) Réf.: A11A01702 Volume R1: 1 x 23 ml Volume R2: 1 x 23 ml Volume R3: 1 x 110 ml Volume R4: 2 x 21 ml Volume R5: 1 x 25 ml ABX Pentra Whole blood Reagent for quantitative in-vitro determination

More information

Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University

Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University Presented by Marcelo Cardona, MT(ASCP) Johns Hopkins University Alert or critical values represent those assay results that require prompt, rapid clinical attention to avert significant study-participant

More information

Original Article Interfering effect of bilirubin on the determination of alkaline phosphatase

Original Article Interfering effect of bilirubin on the determination of alkaline phosphatase Int J Clin Exp Med 2014;7(11):4244-4248 www.ijcem.com /ISSN:1940-5901/IJCEM0002083 Original Article Interfering effect of bilirubin on the determination of alkaline phosphatase Zhong Wang, Hongyan Guo,

More information

Journal of Chemical and Pharmaceutical Research

Journal of Chemical and Pharmaceutical Research Available on line www.jocpr.com Journal of Chemical and Pharmaceutical Research ISSN No: 0975-7384 CODEN(USA): JCPRC5 J. Chem. Pharm. Res., 2011, 3(2):770-775 Validation of Rapid Liquid Chromatographic

More information

Should HbA1C measured by POC instruments be used for diagnosis of diabetes? Sverre Sandberg, Norwegian Quality Improvement of Primary Care

Should HbA1C measured by POC instruments be used for diagnosis of diabetes? Sverre Sandberg, Norwegian Quality Improvement of Primary Care Should HbA1C measured by POC instruments be used for diagnosis of diabetes? Sverre Sandberg, Norwegian Quality Improvement of Primary Care Laboratories Noklus, Bergen, Norway NOKLUS a POC organisation

More information

Manual. Transfer step. Figure 1. Workflow comparison of manual phenylalanine assay and automated GSP Neonatal Phenylalanine assay

Manual. Transfer step. Figure 1. Workflow comparison of manual phenylalanine assay and automated GSP Neonatal Phenylalanine assay TECHNICAL NOTE GSP Neonatal Phenylalanine kit 3308-0010/3308-001B GSP Neonatal Phenylalanine kit (3308-0010/3308-001B) is a fully Introduction automated enzymatic assay for the GSP system. Optimized for

More information

for simultaneous management of diabetes and screening for thalassemia HbA 1c . We used Capillarys 2 Flex Piercing (Capillarys 2FP) HbA 1c

for simultaneous management of diabetes and screening for thalassemia HbA 1c . We used Capillarys 2 Flex Piercing (Capillarys 2FP) HbA 1c Original papers by Capillarys 2 Flex Piercing programme for simultaneous management of diabetes and screening for thalassemia Peifeng Ke 1, Jiawei Liu 1,2, Yan Chao 1, Xiaobin Wu 1, Yujuan Xiong 1, Li

More information

READ HIGHLIGHTED CHANGES

READ HIGHLIGHTED CHANGES BIO-FLASH anti-hbs 3000-8581 100 tests The BIO-FLASH anti-hbs is a fully automated chemiluminescent simultaneous immunoassay for quantitative measurement of antibodies to Hepatitis B surface antigen (anti-hbs)

More information

BIOLOGICAL VARIABILITY: THE CIRME CONTRIBUTION. F. Braga Centre for Metrological Traceability in Laboratory Medicine (CIRME)

BIOLOGICAL VARIABILITY: THE CIRME CONTRIBUTION. F. Braga Centre for Metrological Traceability in Laboratory Medicine (CIRME) BIOLOGICAL VARIABILITY: THE CIRME CONTRIBUTION F. Braga Centre for Metrological Traceability in Laboratory Medicine (CIRME) TOTAL VARIATION (CV (CV T ) T ) PREANALYTICAL VARIATION ANALYTICAL VARIATION

More information

HbA1c (Human) ELISA Kit

HbA1c (Human) ELISA Kit HbA1c (Human) ELISA Kit Cat. No.:DEIA3509 Pkg.Size:96T Intended use GHbA1c (Human) ELISA Kit is a sandwich enzyme immunoassay for the quantitative measurement of human GHbA1c. General Description vhemoglobin,

More information

Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System

Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Abstract Objective: Both Fora Premium V10 and V30 blood glucose

More information

FreeStyle Lite A Blood Glucose Meter That Requires No Coding

FreeStyle Lite A Blood Glucose Meter That Requires No Coding Journal of Diabetes Science and Technology Volume 2, Issue 4, July 2008 Diabetes Technology Society SYMPOSIUM Shridhara, Ph.D. Abstract Background: Abbott Diabetes Care introduced the FreeStyle Lite blood

More information

Accuracy of Filter Paper Method for Measuring Glycated Hemoglobin

Accuracy of Filter Paper Method for Measuring Glycated Hemoglobin Original Article Accuracy of Filter Paper Method for Measuring Glycated Hemoglobin Anjali*, FS Geethanjali+, R Selva Kumar+, MS Seshadri* Abstract Background and Objectives: Glycated hemoglobin (HbA1c)

More information

Insulin (Porcine/Canine) ELISA

Insulin (Porcine/Canine) ELISA Insulin (Porcine/Canine) ELISA For the quantitative measurement of insulin in Porcine/Canine serum and plasma (EDTA) For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 80-INSPO-E01

More information

Method Comparison Report Semi-Annual 1/5/2018

Method Comparison Report Semi-Annual 1/5/2018 Method Comparison Report Semi-Annual 1/5/2018 Prepared for Carl Commissioner Regularatory Commission 123 Commission Drive Anytown, XX, 12345 Prepared by Dr. Mark Mainstay Clinical Laboratory Kennett Community

More information

Implementation of HbA1c as a Diagnostic Test in New Zealand

Implementation of HbA1c as a Diagnostic Test in New Zealand Implementation of HbA1c as a Diagnostic Test in New Zealand Chris Florkowski MD MRCP(UK) FRACP FRCPA FFSc Consultant in Chemical Pathology, Canterbury Health Laboratories Diabetes Physician, Christchurch

More information

Bio-Rad Laboratories HEMOGLOBIN Testing Bio-Rad A1c. VARIANT II TURBO Link. Fully-Automated HbA 1c Testing

Bio-Rad Laboratories HEMOGLOBIN Testing Bio-Rad A1c. VARIANT II TURBO Link. Fully-Automated HbA 1c Testing Bio-Rad Laboratories HEMOGLOBIN Testing Bio-Rad A1c VARIANT II TURBO Link Fully-Automated HbA 1c Testing Bio-Rad Laboratories HEMOGLOBIN Testing VARIANT II TURBO Link Efficient streamlined process Powerful

More information

Can glycated albumin assist in management of diabetes mellitus? Jasna Lenicek Krleza, PhD. Children s Hospital Zagreb, Croatia

Can glycated albumin assist in management of diabetes mellitus? Jasna Lenicek Krleza, PhD. Children s Hospital Zagreb, Croatia Can glycated albumin assist in management of diabetes mellitus? Jasna Lenicek Krleza, PhD. Children s Hospital Zagreb, Croatia DIABETES MELLITUS COURSE SUBJECT 2001 2014 WHEN WE PREPARED PROGRAMME of THIS

More information

Correlation between eag Values Calculated from HbA1C Values and Fasting Plasma Glucose Levels for Diagnosis and Prognosis of Diabetes

Correlation between eag Values Calculated from HbA1C Values and Fasting Plasma Glucose Levels for Diagnosis and Prognosis of Diabetes Correlation between eag Values Calculated from HbA1C Values and Fasting Plasma Glucose Levels for Diagnosis and Prognosis of Diabetes Omkar Kadhane 1, Usha Mukundan 2, Shirish Malvankar 3, Nikhil Ranade

More information

Bovine Insulin ELISA

Bovine Insulin ELISA Bovine Insulin ELISA For quantitative determination of insulin in bovine serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures. Catalog Number: 80-INSBO-E01 Size: 96 wells Version:

More information

Expression of Hemoglobin Variant Migration by Capillary Electrophoresis Relative to Hemoglobin A 2. Improves Precision

Expression of Hemoglobin Variant Migration by Capillary Electrophoresis Relative to Hemoglobin A 2. Improves Precision Hematopathology / Hb Variant Migration by CE Relative to Hb Expression of Hemoglobin Variant Migration by Capillary Electrophoresis Relative to Hemoglobin Improves Precision David F. Keren, MD,, Renee

More information

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: k100322 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE B. Purpose for Submission: Clearance of a new device C. Measurand: Whole

More information