System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013
|
|
- Justina Andrews
- 6 years ago
- Views:
Transcription
1 System accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013
2 Abstract Objective: The goal of the study was to assess the accuracy and reliability of Fora Test N Go blood glucose monitoring system (BGMS) in the clinical studies according to the international standard ISO 15197: 2013(E) criteria and more stringent guidelines. Methods: The capillary blood specimens was collected via fingertip stick from participants at AMCR Institute (700 West El Norte Parkway, Suite 201, Escondido, CA 92026) and tested on three different lots of test strips. Results: The Fora Test N Go BGMS met the requirements of ISO 15197:2013 from all three lots of test strips tested. Within ± 5 mg/dl & ± 5 % (Within ± 0.28 mmol/l & ± 5 %) & ± 10 % (Within ± 0.56 mmol/l & ± 10 %) Within ± 15 mg/dl & ± 15 % (Within ± 0.83 mmol/l & ± 15 %) 425 / 600 (71.0 %) 579 / 600 (96.5 %) 592 / 600 (98.6 %) This BGMS is also compliant with the requirements of DIN EN ISO 15197:2003 with all 3 reagent system lot(s). Within ± 5 mg/dl & ± 5 % (Within ± 0.28 mmol/l & ± 5 %) & ± 10 % (Within ± 0.56 mmol/l & ± 10 %) & ± 15 % (Within ± 0.56 mmol/l & ± 15 %) Within ± 15 mg/dl & ± 20 % (Within ± 0.83 mmol/l & ± 20 %) 336 / 600 (55%) 518 / 600 (86.0 %) 587 / 600 (98.0 %) 598 / 600 (99.6 %) Conclusion: The study suggest that the Fora Test N Go blood glucose monitoring system not only met the current accuracy standard (ISO15197:2003), but also met with the minimum acceptable accuracy standard published in the newer 2013 guideline (ISO15197:2013) when compared the test results with a laboratory reference. 2
3 1. Introduction This study was to evaluate the system accuracy following ISO 15197: protocol and if possible, the newly published ISO 19197: guideline. EN ISO 15197:2003 is an international standard published by International Standard Origination (ISO) specifying requirements for in vitro glucose monitoring systems on self-testing in managing diabetes mellitus. The minimum acceptable accuracy for results produced by a glucose monitoring system shall be as follows: Ninety-five percent (95 %) of the individual glucose results shall fall within ± 15 mg/dl (± 0.83 mmol/l) of the results of the reference measurement at glucose concentrations < 75 mg/dl (< 4.2 mmol/l) and within ± 20 % at glucose concentrations 75 mg/dl ( 4.2 mmol/l). In 2013, ISO has published a newer standard ISO 15197:2013 (E) which tightens the accuracy acceptance criteria. The minimum acceptable accuracy for results produced by a glucose monitoring system has been changed to the followings: Ninety-five percent (95 %) of the individual glucose results shall fall within ± 15 mg/dl (± 0.83 mmol/l) of the results of the reference measurement at glucose concentrations < 100 mg/dl (< 5.55 mmol/l) and within ± 15 % at glucose concentrations 100 mg/dl ( 5.55 mmol/l). EN ISO 15197: 2003 ISO 15197: 2013 (E) Maximum deviation from glucose reference value At glucose levels < 75 mg/dl (< 4.2 mmol/l) At glucose levels 75 mg/dl ( 4.2 mmol/l) ± 15mg/dl (0.83 mmol/l) ± 20% At glucose levels < 100 mg/dl (< 5.5 mmol/l) ± 15mg/dl (0.83 mmol/l) ± 15% Parkes Error Grid Not Required At glucose levels 100 mg/dl ( 4.2 mmol/l) 99% of results within Zones A and B 2. Material and Methods 2.1 Reference standard The reference used in this study was a Yellow Springs 2300 (YSI 2300) manufactured by YSI incorporated (Yellow Springs, OH) which detected glucose concentration based on a glucose oxidase method The readings of the YSI analyzer was validated by running the analyzer with the National Institute of Standards and Technology (NIST) secondary reference material SRM 965a which consists of four levels of standard glucose concentrations. Glucose reference was determined as the average of at least duplicated measurements from the YSI. 3
4 2.2 Preparation of Samples Capillary blood samples were collected based on the ISO 15197:2013 guideline to have certain percentage of subject samples in the following concentration categories: Concentration category Percentage of samples [%] Glucose concentration [mg/dl (mmol/l)] Number of test subjects based on ( 2.77) > (> ) > (> ) > (> ) > (> ) > (> ) > 400 (> 22.20) Test of Blood Glucose Sample capillary blood was obtained by a finger stick using a 25G safety lancet. The blood sample was immediately collected onto a Lithium-Heparin Microvette (~ 200 μl) The capillary blood was then placed on a parafilm and tested on Fora Test N Go blood glucose monitoring systems (6 devices per station with test strip lot 1, 2, and 3). The remaining blood in the lithium-heparin microvette was then immediately placed on a centrifuge to separate plasma. The glucose concentration of the plasma was then determined by the YSI 2300 using one electrode. The measurement was repeated once and the true glucose concentration of the subject glucose was determined by the average of the YSI values. 2.4 Statistical analysis In this study, the following statistical analyses were performed: Bias analysis according to Bland and Altman 5. Regression analysis according to Passing and Bablok Results 3.1 Test results Accuracy results following ISO 15197:2013 (E) Summary: Within Consensus Error Grid zones A and B 600 / 600 (100 %) Conclusion: The Fora Test N Go BGMS met the requirements of new ISO standard (ISO 15197:2013). 4
5 Detail results: System accuracy results for glucose concentrations < 100 mg/dl (5.55 mmol/l) Within ± 5 mg/dl (Within ± 0.28 mmol/l) (Within ± 0.56 mmol/l) Within ± 15 mg/dl (Within ± 0.83 mmol/l) 47 / 64 (73.4 %) 61 / 64 (95.3 %) 64 / 64 (100 %) 50 / 64 (78.1 %) 61 / 64 (95.3 %) 63 / 64 (98.4 %) 49 / 64 (76.6 %) 64 / 64 (100 %) 64 / 64 (100 %) System accuracy results for glucose concentrations 100 mg/dl (5.55 mmol/l) Within ± 5 % Within ± 10 % Within ± 15 % 72 / 136 (52.9 %) 118 / 136 (86.8 %) 134 / 136 (98.5 %) 59 / 136 (43.4 %) 109 / 136 (80.2 %) 136 / 136 (100 %) 64 / 136 (47.1 %) 108 / 136 (79.4 %) 130 / 136 (95.6 %) Accuracy results following ISO 15197:2003 Summary: Within Consensus Error Grid zones A and B 600 / 600 (100 %) Conclusion: Fora Test N Go BGMS also is in compliance with the requirements of DIN EN ISO 15197:2003 for all 3 reagent system lot(s). Detail results: System accuracy results for glucose concentrations < 75 mg/dl (4.2 mmol/l) Within ± 5 mg/dl (Within ± 0.28 mmol/l) (Within ± 0.56 mmol/l) Within ± 15 mg/dl (Within ± 0.83 mmol/l) 29 / 38 (76.3 %) 36 / 38 (94.7 %) 38 / 38 (100 %) 32 / 38 (84.2 %) 35 / 38 (92.1 %) 37 / 38 (97.4 %) 31 / 38 (81.6 %) 38 / 38 (100 %) 38 / 38 (100 %) System accuracy results for glucose concentrations 75 mg/dl (4.2 mmol/l) Within ± 5 % Within ± 10 % Within ± 15 % Within ± 20 % 88 / 162 (54.3 %) 142 / 162 (87.7 %) 160 / 162 (98.8 %) 161 / 162 (99.4 %) 75 / 162 (46.3 %) 134 / 162 (82.7 %) 162 / 162 (100 %) 162 / 162 (100 %) 81 / 162 (50.0 %) 133 / 162 (82.1 %) 156 / 162 (96.3 %) 162 / 162 (100 %) 5
6 3.1.3 Consensus Error Grid for FORA Test N Go with three reagent system lots Region Number Percent A % B % C % D % E
7 Summary of regression analysis (Passing and Bablok) n=200 Regression Reagent system lot 1: TD14D228-C0E Reagent system lot 2: TD14 D828-C0E Reagent system lot 3: TD14 D928-C0E Y= X Y= X Y= X 7
8 3.2.4 Absolute differences between FORA Test N Go and YSI 2300 STAT PLUS 8
9 4. Conclusion and discussion: This study showed that the Fora Test N Go blood glucose monitoring system not only met the current accuracy standard (ISO15197:2003), but also exceeded the minimum acceptable accuracy standard published in the newer 2013 guideline (ISO15197:2013) when compared the test results with a laboratory reference. In addition, 99% of the data points fall within Zone A & B of Consensus Error Grid, which means that when user obtained a value using Fora Test N Go monitor, it should not result in a clinical decision that could be deemed inappropriate. There are several glucose monitoring systems out in the market. Since FDA is in preparation of tightening the accuracy standard to ± 15% by adopting the new ISO15197:2013 standard. Based on the results, it will be reasonable assumption that Fora Test N Go met the new ISO standard, and therefore would be a reasonable selection for users if they like to use meter with a higher accuracy. Reference a International Diabetes Federation, the global burden. b. In vitro diagnostic test systems - Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. EN ISO 15197:2003. c. In vitro diagnostic test systems - Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. ISO 15197:2013. d. Parkes JL, Slatin SL, Pardo S, Ginsberg BH: A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose. Diabetes Care 2000; 23: e. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 1: f. Passing H, Bablok. A new biometrical procedure for testing the equality of measurements from two different analytical methods. Application of linear regression procedures for method comparison studies in clinical chemistry, Part I. J Clin Chem Clin Biochem 1983; 21:
Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System
Equivalent Accuracy Evaluation of FORA Premium V10 Blood Glucose Monitoring System as Compared to Fora V30 Blood Glucose Monitoring System Abstract Objective: Both Fora Premium V10 and V30 blood glucose
More informationSystem accuracy evaluation of FORA Test N Go Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013
System Accuracy Evaluation of Project number: Fora092614-01 System accuracy evaluation of Blood Glucose Monitoring System versus YSI 2300 STAT Plus glucose analyzer following ISO 15197:2013 Date: 13 th
More informationClinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System
Clinical Accuracy and User Performance of the TRUE METRIX AIR Blood Glucose Monitoring System Summary Objectives: To demonstrate that the TRUE METRIX AIR Blood Glucose Monitoring System, from Trividia
More informationEvaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System
Evaluation of Accuracy and User Performance of the TRUE METRIX Blood Glucose Monitoring System Summary Objectives: To demonstrate that the TRUE METRIX Blood Glucose Monitoring System, from Trividia Health,
More informationReport. Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013
Institute of Diabetes Gerhardt Katsch Karlsburg (IDK) Report Evaluation of system accuracy of the blood glucose monitoring system VivaChek Ino according to DIN EN ISO 15197: 2013 Sponsor: The study was
More informationEvaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System
Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System Summary Objectives: To demonstrate that the TRUE METRIX Self-Monitoring Blood Glucose System*, from Trividia
More informationBern Harrison, B.A., Cheryl Leazenby, B.S., and Solveig Halldorsdottir, Ph.D.
Journal of Diabetes Science and Technology Volume 5, Issue 4, July 2011 Diabetes Technology Society TECHNOLOGY REPORTS Accuracy of the CONTOUR Blood Glucose Monitoring System Bern, B.A., Cheryl Leazenby,
More informationWhite Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard
White Paper, Evaluation of WaveSense JAZZ Blood Glucose Monitoring System Analytical Performance to EN ISO 15197:2015 Standard 7500-10032 Rev F Table of Contents Executive Summary 3 Changes introduced
More informationFreeStyle Lite A Blood Glucose Meter That Requires No Coding
Journal of Diabetes Science and Technology Volume 2, Issue 4, July 2008 Diabetes Technology Society SYMPOSIUM Shridhara, Ph.D. Abstract Background: Abbott Diabetes Care introduced the FreeStyle Lite blood
More informationNICO /14 Nipro Diagnostics, Inc.
Clinical Performance of the TRUEresult twist Blood Glucose Monitoring System () Using ISO 15197:13 Criteria for Accuracy Utilizing GDH-FAD Enzyme NICO-8 1/1 Nipro Diagnostics, Inc. NICO_8_mmol.indd CLINICAL
More informationGD50. Simple. Accurate. Reliable. Test Simply. Live Better. All FORA Diamond Series Blood Glucose Monitoring Systems use FORA Diamond Test Strips.
GD50 Higher Measuring Accuracy Simple. Accurate. Reliable. Test Simply. Live Better. All FORA Diamond Series Blood Glucose Monitoring Systems use FORA Diamond Test Strips. GD50 Simple. Accurate. Reliable.
More informationEVALUATION OF THE PERFORMANCE OF THE TRUETRACK BLOOD GLUCOSE MONITORING SYSTEM USING THE ISO 15197:2003 ACCURACY AND USER EVALUATION REQUIREMENTS
EVALUATION OF THE PERFORMANCE OF THE TRUETRACK BLOOD GLUCOSE MONITORING SYSTEM USING THE ISO 15197:23 ACCURACY AND USER EVALUATION REQUIREMENTS Douglas E. Bell, PhD Teri A. Sasse, RN, MS BACKGROUND: The
More informationSmart & Simple Mobile Solutions.
Higher Measuring Accuracy Smart & Simple Mobile Solutions. Test Simply. Live Better. FORA Diamond lets you connect all your Bluetooth devices. *Also available with a Micro USB. ifora Diabetes Manager All
More informationGlucoMen LX PLUS blood glucose and blood ketone meter. Accuracy Evaluation to New ISO 15197:2013, with Technical and Specification Data
GlucoMen LX PLUS: Accuracy Evaluations to ISO 15197:2013 with Technical and Specification Data GlucoMen LX PLUS blood glucose and blood ketone meter Accuracy Evaluation to New ISO 15197:2013, with Technical
More informationClinical Performance of the TRUE2go Blood Glucose Monitoring System Exceeds Minimum Criteria for Accuracy Using ISO 15197:2013
linical Performance of the TRUE2go lood Glucose Monitoring System Exceeds Minimum riteria for Accuracy Using ISO 197:213 LINIAL STUY SUMMARY EVALUATION OJETIVES This clinical evaluation: 1. Evaluates the
More informationGlucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D Biochemistry Analyzers
Glucose Analytical Comparability Evaluation of the YSI 2300 STAT Plus and YSI 2900D iochemistry Analyzers Life Sciences Data for Life. TM YSI Life Sciences White Paper 91 Authors: Kevin Schlueter, PhD,
More informationAina Blood Monitoring System
Aina Blood Monitoring System Analytical Performance Summary The Aina Blood Monitoring System is a high quality and versatile multi-parameter diagnostic platform that is CE-marked and approved for sale
More informationClinical Evaluation for. Embrace No Code Blood Glucose Monitoring System
Embrace No Code Report Clinical Evaluation for Embrace No Code Blood Glucose Monitoring System Applicant: Apex Biotechnology Corp. Test Date: 2007/8/28~2007/9/21 Clinical Site: SCHMIDT Group Practice Clinic
More informationA New, Wireless-enabled Blood Glucose Monitoring System That Links to a Smart Mobile Device: Accuracy and User Performance Evaluation
691301DSTXXX10.1177/1932296817691301Journal of Diabetes Science and TechnologyChristiansen et al research-article2017 Original Article A New, Wireless-enabled Blood Glucose Monitoring System That Links
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number: k100322 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE B. Purpose for Submission: Clearance of a new device C. Measurand: Whole
More informationAccuracy and Precision Performance of the Accu-Chek Mobile System. Introduction. Method
Accuracy and Precision Performance of the Accu-Chek Mobile System I. ACCURACY The accuracy of the system was assessed according to ISO 15197. Introduction The purpose of this study was to determine the
More informationValidity and Reliability of a Glucometer Against Industry Reference Standards
514315DSTXXX10.1177/1932296813514315Journal of Diabetes Science and TechnologySalacinski et al research-article2014 Original Article Validity and Reliability of a Glucometer Against Industry Reference
More informationPerformance Evaluation of B. Braun Omnitest 5 Blood Glucose Monitoring System for Self-Monitoring of Blood Glucose
ORIGINL RTILE https://doi.org/10.15263/jlmqa.2016.38.4.234 Performance Evaluation of. raun Omnitest 5 lood Glucose Monitoring System for Self-Monitoring of lood Glucose Kyungso Jeon 1 and Miseon Shin 2
More information(a) y = 1.0x + 0.0; r = ; N = 60 (b) y = 1.0x + 0.0; r = ; N = Lot 1, Li-heparin whole blood, HbA1c (%)
cobas b system - performance evaluation Study report from a multicenter evaluation of the new cobas b system for the measurement of HbAc and lipid panel Introduction The new cobas b system provides a point-of-care
More informationBLOOD GLUCOSE METERS ACCURACY REQUIREMENTS
BLOOD GLUCOSE METERS ACCURACY REQUIREMENTS International Standard ISO 15197:2013 (Second Edition 15/05/2013) In vitro diagnostic test systems Requirements for blood-glucose monitoring systems for self-testing
More informationPerformance of three portable blood glucose monitoring devices used in a veterinary application
Performance of three portable blood glucose monitoring devices used in a veterinary application Prepared by Michael I. Lindinger, PhD, Vice-President of Research, The Nutraceutical Alliance, Campbellville,
More information3. System Accuracy Evaluation LUX Hgb test 1. Samples
3. System Accuracy Evaluation LUX Hgb test 1. Samples - Type of sample: Capillary Blood System accuracy evaluation was performed with 200 fresh capillary blood samples, each with sufficient volume to be
More informationUltra Blue. Test Strips UNISTRIP. Charlotte, NC V2-10/25/2018 TECHNOLOGIES,LLC
VS UNISTRIP TECHNOLOGIES,LLC Charlotte, NC 28269 1-866-889-6229 V2-10/25/2018 UniStrip Generic Blood Glucose VS OneTouch Ultra Blue 1. Purpose This study is intended to evaluate the difference in values
More information510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k103542 B. Purpose for Submission: Alternative blood glucose test strips for use with LifeScan OneTouch
More informationAccuracy Evaluation of Five Blood Glucose Monitoring Systems: The North American Comparator Trial
Journal of Diabetes Science and Technology Volume 7, Issue 5, September 2013 Diabetes Technology Society ORIGINAL ARTICLE Accuracy Evaluation of Five Blood Glucose Monitoring Systems: The North American
More informationA meter designed for glucose self-measurement manufactured by LifeScan, Johnson & Johnson. Report from an evaluation organised by SKUP
SKUP Scandinavian evaluation of laboratory equipment for primary health care ONETOUCH Ultra A meter designed for glucose self-measurement manufactured by LifeScan, Johnson & Johnson Report from an evaluation
More informationCOBAS, ACCU-CHECK INFORM, MODULAR and LIFE NEEDS ANSWERS are trademarks of Roche Roche
Accu-Chek Inform II test strip Results of the evaluation study on newborns, conducted by the Flevo Hospital (FVZH) Almere and the Maxima Medical Center (MMC) Veldhoven in the Netherlands COBAS, ACCU-CHECK
More informationPoint-of-Care Blood Glucose Meter Accuracy in the Hospital Setting
In Brief Point-of-care (POC) tests provide analytical information that can be used to make decisions at patients bedside, as opposed to laboratory tests that must be run at a central laboratory. POC testing
More information1 PROTOCOL. Comparison Study Summary. Springs Memorial Hospital. Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC (803)
Comparison Study Summary Springs Memorial Hospital 800 W. Meeting St. Lancaster, SC 29720 (803) 286-1480 February 10, 2016 1 PROTOCOL This evaluation was conducted on February 10, 2016 at Springs Memorial
More informationISO 15197:2013 / EN ISO 15197:2015 Clinical Validation Study List for Blood Glucose Monitoring System
List No. STUDY TITLE DATE / PLACE SCIENTIST BRIEF RESULTS TYPE OF STUDY 01 Intermeiate measurement precision Jan 2018 The FORA 6 Connect fulfille the requirements of evaluation of FORA 6 Connect Multi-functional
More informationA meter designed for glucose self-measurement manufactured by Roche Diagnostics. Report from an evaluation organised by SKUP
SKUP Scandinavian evaluation of laboratory equipment for primary health care ACCU-CHEK Compact Plus A meter designed for glucose self-measurement manufactured by Roche Diagnostics Report from an evaluation
More informationA model for managing and monitoring the quality of glucometers used in a high-volume clinical setting
Original papers A model for managing and monitoring the quality of glucometers used in a high-volume clinical setting Güzin Aykal*, Ayşenur Yegin, Özgür Tekeli, Necat Yilmaz Central Laboratories, Antalya
More informationOGCARE is a blood glucose monitoring device for checking glucose concentration in whole blood from fingertip pricking.
ITALIAN PRODUCT S E L F T E S T I N G Personal medical devices for metabolic disorders MONITORING and PREVENTION, taking advantage of reflectometric and electrochemical technologies. Blood glucose monitoring
More information1. PROTOCOL. Comparison Study Summary. Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR July 30, 2014
Comparison Study Summary Tuality Healthcare 324 SE 9 th Ave. Suite E Hillsboro, OR 97123 July 30, 2014 1. PROTOCOL This study was conducted on July 24 th, 2014 at Tuality Healthcare, Hillsboro, OR. The
More informationPolymer Technology Systems, Inc. CardioChek PA Comparison Study
Polymer Technology Systems, Inc. CardioChek PA Comparison Study Evaluation Protocol: Accuracy Precision Clinical Correlation PTS Panels Lipid Panel Test Strips For Use in Comparisons to a Reference Laboratory
More informationSUMMARY... 2 PLANNING OF THE EVALUATION... 4 ANALYTICAL QUALITY SPECIFICATIONS... 5
USER-EVALUATION ASCENSIA CONTOUR Blood Glucose Meter System Table of contents SUMMARY... 2 PLANNING OF THE EVALUATION... 4 ANALYTICAL QUALITY SPECIFICATIONS... 5 MATERIALS AND METHODS... 6 ASCENSIA CONTOUR...
More informationMeter and test strips designed for glucose self-measurement manufactured by Arkray, Inc. Report from an evaluation organised by SKUP
SKUP Scandinavian evaluation of laboratory equipment for primary health care GLUCOCARD X-METER GLUCOCARD X-SENSOR Meter and test strips designed for glucose self-measurement manufactured by Arkray, Inc.
More informationASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE
ASSESSMENT OF A POINT-OF-CARE DEVICE FOR MEASURING CREATININE IN A COMMUNITY SCREENING PROGRAM FOR CHRONIC KIDNEY DISEASE Brooke Ann Spaeth, Anne K Shephard, Mark DS Shephard, Timothy H Mathew ABSTRACT
More informationGB/T Translated English of Chinese Standard: GB/T
Translated English of Chinese Standard: GB/T 19634-2005 www.chinesestandard.net Sales@ChineseStandard.net GB ICS 11.100 C 44 National Standard of the People s Republic of China GB/T 19634-2005 In vitro
More informationMulti User Model. Product Overview
Multi User Model Product Overview Features and Benefits NO CODING EASY TO USE o o o Large LCD display. Large buttons with simple functions Easy Setup. Pre-Setup before shipping. Easy strip insertion. SMALL
More informationWHO Prequalification of In Vitro Diagnostics PUBLIC REPORT. Product: OraQuick HCV Rapid Antibody Test Kit WHO reference number: PQDx
WHO Prequalification of In Vitro Diagnostics PUBLIC REPORT Product: OraQuick HCV Rapid Antibody Test Kit WHO reference number: PQDx 0244-055-00 OraQuick HCV Rapid Antibody Test Kit with product codes 1001-0270
More informationRelationship between glucose meter error and glycemic control efficacy
Relationship between glucose meter error and glycemic control efficacy Brad S. Karon, M.D., Ph.D. Professor of Laboratory Medicine and Pathology Department of Laboratory Medicine and Pathology Mayo Clinic
More informationAnalytical error of home glucose monitors: a comparison of 18 systems
Original Article Ann Clin Biochem 1999; 36: 72-79 Analytical error of home glucose monitors: a comparison of 18 systems Roger N Johnson and John R Baker From the Department of Clinical Biochemistry, Green
More informationTheraSense FreeStyle blood glucose meter
May 2002 Evaluation Report MDA 02049 TheraSense FreeStyle blood glucose meter Crown Copyright 60 WHAT YOU CAN EXPECT FROM MDA EVALUATION REPORTS The Device Evaluation Service (DES) aims to provide independent
More informationMQ Comparison of glucometers using heparin whole blood
MQ 2014-1 Comparison of glucometers using heparin whole blood Note: The instrument comparison is structured as a survey. This is a sample test, not a complete evaluation. Introduction Glucose self-testing
More informationComputing the Surveillance Error Grid Analysis: Procedure and Examples
5959DSTXXX.77/99685959Journal of Diabetes Science and TechnologyKovatchev et al research-article Original Article Computing the Surveillance Error Grid Analysis: Procedure and Examples Journal of Diabetes
More informationEVALUATING THE UNCERTAINTY OF MEASUREMENT ON BLOOD'S GLUCOSE LEVEL
BIOMEDICAL ENGINEERING- APPLICATIONS, BASIS & COMMUNICATIONS EVALUATING THE UNCERTAINTY OF MEASUREMENT ON BLOOD'S GLUCOSE LEVEL 91 HUNG-CHUN HUANG 1,2, CHIA-HUNG CHIEN 1,2, CHENG-YI WANG 1, FOK-CHING CHONG
More informationSKUP Scandinavian evaluation of laboratory equipment for primary health care
S K U P i n D e n m a r k, D E K S, R i g s h o s p i t a l e t - G l o s t r u p, DK- 2 6 0 0 G l o s t r u p, P h o n e + 4 5 3 8 6 3 4 4 0 0, w w w. s k u p. n u S K U P i n N o r w a y, N o k l u s,
More informationSAMPLE. Glucose Monitoring in Settings Without Laboratory Support
POCT13c Glucose Monitoring in Settings Without Laboratory Support This guideline focuses on performance of point-of-care glucose monitoring systems, with an emphasis on safety practices, quality control,
More informationCustomer Bulletin. Questions/Comments? Please call Customer Service at (Direct) or (Toll-free inside USA).
Customer Bulletin Assignment of Ranges for PTS Panels Multi-Chemistry Controls (Cat. No./REF 721) and PTS Panels Controls (Cat. No./REF 722) Background: PTS Panels Multi-Chemistry Controls and Controls
More informationEvaluation of the Analytical Performance of the Coulometry-Based Optium Omega Blood Glucose Meter
Journal of Diabetes Science and Technology Volume 5, Issue 6, November 2011 Diabetes Technology Society TECHNOLOGY REPORT Evaluation of the Analytical Performance of the Coulometry-Based Optium Omega Blood
More informationComparison of Three Whole Blood Creatinine Methods for Estimation of Glomerular Filtration Rate Before Radiographic Contrast Administration
Clinical Chemistry / Whole Blood Creatinine for egfr Comparison of Three Whole Blood Creatinine Methods for Estimation of Glomerular Filtration Rate Before Radiographic Contrast Administration Nichole
More informationOmnitest : 2013
ISO 15197: 2013 Omnitest 3 Blood Glucose Monitoring System Omnitest 3 Could it be easier? After strip insertion, Omnitest 3 is ready for the measurement. The code is set automatically. Only 0.3 µl sample
More informationWELL CARD HEALTH TOOL BOX
WELL CARD HEALTH TOOL BOX 1 ADVOCATE Brand Advantage Well Card Health Members Will Save 52-70% On Diabetic Supplies Free Talking Meter Or Dash Meter Programs To Well Card Health Members 3 to 5 day UPS
More informationPrecision, Accuracy, and User Acceptance of the OneTouch SelectSimple Blood Glucose Monitoring System
Journal of Diabetes Science and Technology Volume 5, Issue 6, November 2011 Diabetes Technology Society TECHNOLOGY REPORT Precision, Accuracy, and User Acceptance of the OneTouch SelectSimple Blood Glucose
More informationRETAIL BULK DIABETIC SUPPLY TOOL BOX
RETAIL BULK DIABETIC SUPPLY TOOL BOX 1 ADVOCATE Brand Advantage Employers Can Save 52-72% On Diabetic Supplies Free Talking Meter - No Coding Required One of the Most Accurate Blood Glucose Meters on the
More informationSystem Accuracy Evaluation LUX Chol HDL test
System Accuracy Evaluation LUX Chol HDL test 1. Minimum Acceptable System Accuracy 1.1 System Accuracy Requirement The minimum acceptable accuracy for results produced by a Lipidmonitoring systemshall
More informationSCIEX Vitamin D 200M Assay for the Topaz System
The First FDA-Cleared LC-MS/MS Assay for Vitamin D SCIEX Vitamin D 200M Assay for the Topaz System The first FDA-cleared LC-MS/MS assay for Vitamin D Vitamin D is an important building block for human
More informationMQ Comparison of glucometers using whole blood
MQ 2017-1 Comparison of glucometers using whole blood Note: The instrument comparison is structured as a survey. This is a sample test, not a complete evaluation. Introduction Glucose self-testing devices
More informationClinical Impact of Prandial State, Exercise, and Site Preparation on the Equivalence of Alternative-Site Blood Glucose Testing
Clinical Care/Education/Nutrition O R I G I N A L A R T I C L E Clinical Impact of Prandial State, Exercise, and Preparation on the Equivalence of Alternative- Blood Glucose Testing DAWN M. BINA, CMT ROBYN
More informationStatistical Considerations: Study Designs and Challenges in the Development and Validation of Cancer Biomarkers
MD-TIP Workshop at UVA Campus, 2011 Statistical Considerations: Study Designs and Challenges in the Development and Validation of Cancer Biomarkers Meijuan Li, PhD Acting Team Leader Diagnostic Devices
More informationBLOOD GLUCOSE MONITORING: CLINICAL NEED AND TECHNOLOGY
BLOOD GLUCOSE MONITORING: CLINICAL NEED AND TECHNOLOGY DAVID C KLONOFF, MD Medical Director Diabetes Research Institute of Mills-Peninsula Health Services San Mateo, California Clinical Professor of Medicine,
More informationAccuracy and Reliability of the Nova StatStrip Glucose Meter for Real-Time Blood Glucose Determinations during Glucose Clamp Studies
Journal of Diabetes Science and Technology Volume 4, Issue 5, September 2010 Diabetes Technology Society ORIGINAL ARTICLES Accuracy and Reliability of the Nova StatStrip Glucose Meter for Real-Time Blood
More informationMQ Comparison of glucometers using whole blood
MQ 2018-2 Comparison of glucometers using whole blood Note: The instrument comparison is structured as a survey. This is a sample test, not a complete evaluation. Introduction Glucose self-testing devices
More informationAccuracy and Precision in Point-of-Care Lipid Testing: CardioChek P A Point-of-Care Test System and PTS Panels Test Strips
Accuracy and Precision in Point-of-Care Lipid Testing: CardioChek P A Point-of-Care Test System and PTS Panels Test Strips Sponsored by Arthur Roberts, MD of The Living Heart Foundation The Living Heart
More informationPROCEDURE. TITLE: Bedside Glucose Monitoring PC Laboratory. Issuing Department: Clinical Director Signature: Departments Involved:
PROCEDURE TITLE: Bedside Glucose Monitoring Issuing Department: Clinical Director Signature: Departments Involved: Laboratory Nursing Effective Date: 10/97 Review Dates: 09/01, 07/02, 05/13 Revision Dates:
More informationMARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa
MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa CARE OF PATIENT POLICY & PROCEDURE Policy Number: 4:10 Subject: Policy: Glucose Monitoring (Accuchek) Nursing department staff and laboratory staff
More informationEvaluation of the Precision and Accuracy of the INRatio 2 Prothrombin Time (PT) Monitoring System
Evaluation of the Precision and Accuracy of the INRatio 2 Prothrombin Time (PT) Monitoring System Grace DeSantis, PhD Lisa Croner, PhD Michealle Havenhill, BA Shari Kipp, BSc (Honours) MBA Inverness Medical,
More informationABSTRACT ORIGINAL RESEARCH. Guido Freckmann. Nina Jendrike. Annette Baumstark. Stefan Pleus. Christina Liebing. Cornelia Haug
Diabetes Ther (2018) 9:683 697 https://doi.org/10.1007/s13300-018-0392-6 ORIGINAL RESEARCH User Performance Evaluation of Four Blood Glucose Monitoring Systems Applying ISO 15197:2013 Accuracy Criteria
More informationPackage ega. March 21, 2017
Title Error Grid Analysis Version 2.0.0 Package ega March 21, 2017 Maintainer Daniel Schmolze Functions for assigning Clarke or Parkes (Consensus) error grid zones to blood glucose values,
More informationSKUP Scandinavian evaluation of laboratory equipment for primary health care
SKUP Scandinavian evaluation of laboratory equipment for primary health care Accu-Chek Mobile Meter and test cassettes designed for glucose self-testing manufactured by Roche Diagnostics GmbH Report from
More informationAccuracy and Usability Evaluation of Six Commercially Available Blood Glucose Monitoring Systems
Original Article Accuracy and Usability Evaluation of Six Commercially Available Blood Glucose Monitoring Systems DOI: 10.7860/JCDR/2018/34055.11671 Internal Medicine Section Anastasios Tentolouris 1,
More informationIt s Just a Waived Glucose, Isn t It?
It s Just a Waived Glucose, Isn t It? What Is the Next Step? Becky Damiani, MT (ASCP), Senior Inspection Specialist Laboratory Accreditation Program Objectives Understand the CLIA requirements surrounding
More informationWhy do we need SD LipidoCare?
SD Biosensor, Inc. Innovative Global Leader of In Vitro Diagnostics Total supplier of Blood Glucose Monitoring System and Lipid Analyzer System in vitro diagnostics test kits for Point of Care Testing
More informationFreeStyle Mini Blood Glucose Results Are Accurate and Suitable for Use in Glycemic Clamp Protocols
Journal of Diabetes Science and Technology Volume 2, Issue 5, September 2008 Diabetes Technology Society CLINICAL APPLICATIONS FreeStyle Mini Blood Glucose Results Are Accurate and Suitable for Use in
More informationISO INTERNATIONAL STANDARD
INTERNATIONAL STANDARD ISO 15197 First edition 2003-05-01 In vitro diagnostic test systems Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus Systèmes d'essais
More informationEffects of Simulated Altitude on Blood Glucose Meter Performance: Implications for In-Flight Blood Glucose Monitoring
Journal of Diabetes Science and Technology Volume 6, Issue 4, July 2012 Diabetes Technology Society ORIGINAL ARTICLE Effects of Simulated Altitude on Blood Glucose Meter Performance: Tolu, M.B.B.S., M.R.C.P.,
More informationOneTouch Blood Glucose Monitoring Systems Impact of New Technologies on the Efficacy of Self-monitoring Blood Glucose
Blood Glucose Monitoring Systems Impact of New Technologies on the Efficacy of Self-monitoring Blood Glucose Laurence B Katz, Mike Grady, Steven J Setford, and Brian L Levy US Endocrinology SUPPLEMENT
More informationUSING THE ACCESS AMH ASSAY IN YOUR LABORATORY
INFORMATION BULLETIN USING THE ACCESS AMH ASSAY IN YOUR LABORATORY ///////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////
More informationBMJ Open. A method-comparison study regarding the validity and reliability of the Lactate Plus hand-held lactate meter
BMJ Open A method-comparison study regarding the validity and reliability of the Lactate Plus hand-held lactate meter Journal: BMJ Open Manuscript ID: bmjopen-0-00 Article Type: Research Date Submitted
More information1 PROTOCOL. Comparison Study Summary
Comparison Study Summary OnSite Health 210 S. Michigan Street, Suite 110 South Bend, IN 46601 574-243-5108 ext. 203 July 27, 2015 1 PROTOCOL This internal evaluation was conducted on July 17 & 20, 2015
More informationBlood Glucose Monitoring System USER GUIDE. For Use With Ascensia CONTOUR Blood Glucose Meter and Ascensia MICROFILL Test Strips
Blood Glucose Monitoring System USER GUIDE For Use With Ascensia CONTOUR Blood Glucose Meter and Ascensia MICROFILL Test Strips Thank you for choosing the Ascensia CONTOUR Blood Glucose Monitoring System!
More informationManagement of Central Venous Access Devices. Blood Glucose Monitoring
Management of Central Venous Access Devices Blood Glucose Monitoring Purpose To provide education on the standard of care regarding the use and monitoring of the Accu- Chek Blood glucose machine, including
More informationPoint-of-Care glucose analysis in neonates using modified quinoprotein glucose dehydrogenase
Washington University School of Medicine Digital Commons@Becker Open Access Publications 2013 Point-of-Care glucose analysis in neonates using modified quinoprotein glucose dehydrogenase Dennis J. Dietzen
More informationImportance of Cell Stock Concentration for Accurate Target Cell Recovery
TECHNICAL NOTE Importance of Cell Stock Concentration for Accurate Target Cell Recovery INTRODUCTION 10x Genomics Single Cell Protocols require suspensions of viable, single cells (Single Cell Protocols
More informationStatistical techniques to evaluate the agreement degree of medicine measurements
Statistical techniques to evaluate the agreement degree of medicine measurements Luís M. Grilo 1, Helena L. Grilo 2, António de Oliveira 3 1 lgrilo@ipt.pt, Mathematics Department, Polytechnic Institute
More informationThe Effect of Alcohol Prep Pads and Blood Drop Number On Capillary Blood Glucose Values
University of New Hampshire University of New Hampshire Scholars' Repository Honors Theses and Capstones Student Scholarship Spring 2017 The Effect of Alcohol Prep Pads and Blood Drop Number On Capillary
More informationWorksheet. Worksheet. Worksheet. Worksheet. Student Performance Guide. Student Performance Guide
LESSON 6-3 Laboratory Reagent Preparation and Calculations Worksheet LESSON 6-4 Chemistry Instrumentation in the Physician Office Laboratory Worksheet LESSON 6-6 Measuring Blood Glucose Worksheet LESSON
More informationliquicolor Design Verification
Design Verification liquicolor 1 Introduction... 2 2 Imprecision... 2 2.1 Imprecision for... 2 2.2 Imprecision for... 2 3 Linearity and Detection Limit... 2 3.1 Linearity... 2 3.2 Detection Limit... 4
More informationCONTOUR NEXT and CONTOUR XT
SKUP Scandinavian evaluation of laboratory equipment for primary health care CONTOUR NEXT and CONTOUR XT Test strips and meter designed for self-monitoring of blood glucose Report from the evaluation The
More informationGentian Canine CRP Immunoassay Application Note for Abbott Architect * c4000
v2-jan214 Gentian Canine CRP Immunoassay Application Note for Abbott Architect * c4 Intended Use The canine CRP immunoassay on Abbott's Architect c4 is an in vitro diagnostic test for quantitative determination
More information13. POINT-OF-CARE CREATININE TESTING IN HIGH- RISK PATIENTS
How to Cite this article: Point-of-Care Creatinine Testing in High-Risk Patients - ejifcc 20/01 2009 http://www.ifcc.org 13. POINT-OF-CARE CREATININE TESTING IN HIGH- RISK PATIENTS Yolanda B. de Rijke
More informationToxicology Blood Lead Comprehensive Testing
Clinical Laboratory s of Practice Blood Lead Comprehensive Testing The following specialty sustaining standards of practices shall be incorporated into the laboratory s quality management system, where
More informationJournal of Diabetes and Metabolic Disorders; 2011; Vol 10, pp 1-6. Keywords: Glycated hemoglobin, Paper filter, Diabetes, Dried blood spot.
Journal of Diabetes and Metabolic Disorders; 2011; Vol 10, pp 1-6 Glycated hemoglobin measurements from dried blood spots: reliability and relation to results obtained from whole blood samples Ozra Tabatabaei-Malazy
More information