Anterior chamber depth measurement: interchangeability between Pentacam and other devices

Size: px
Start display at page:

Download "Anterior chamber depth measurement: interchangeability between Pentacam and other devices"

Transcription

1 UPDATE/REVIEW Anterior chamber depth measurement: interchangeability between Pentacam and other devices Alberto Domínguez-Vicent, MSc 1 ; Daniel Monsálvez-Romín, OD 1 ; Vicent Sanchis, MSc 1 ; Santiago García-Lázaro, OD, PhD 1 ; Robert Montés-Micó, OD, PhD 1. ABSTRACT: We reviewed a series of recently published studies in which the anterior chamber depth (ACD) distance was measured by Pentacam or Pentacam HR and compared with other measuring devices. Since some published studies reach contradictory conclusions regarding device interchangeability and include different ocular conditions, for instance, phakic eyes, pseudophakic eyes, cataractous eye, and so on, this review was carried out to try to clarify which clinical devices can or cannot be considered as interchangeable with Pentacam or Pentacam HR for the measurement of ACD distance in clinical practice. J Emmetropia 2012; 3: Nowadays, the measurement of anterior chamber depth (ACD) distance has become increasingly important in ophthalmic practice, for instance, in the planning of cataract 1-5 and refractive 1-9 surgery, calculation of phakic or pseudophakic IOL power and diameter 4,5,10-14, screening for glaucoma risk factors 2,3,14-16, and so on. Many papers focusing on device interchangeability for the measurement of ACD distance have been published. However, the studies published include different ocular conditions. For this reason, we decided to review all studies published in Medline before August The inclusion criterion was that the study had to evaluate device interchangeability between Pentacam or Pentacam HR and other devices that measure ACD distance. The devices included in this review were Pentacam (Oculus, Wetzlar, Germany), Pentacam HR (Oculus, Wetzlar, Germany), A-Scan, Ultrasound Biomicroscopy (UBM), Orbscan (Bausch&Lomb Surgical Inc., San Dimas, California, USA), Orbscan II (Bausch&Lomb Surgical Inc., San Dimas, California, USA), Galilei (Ziemer, Switzerland), Visante OCT (Carl Zeiss Meditec Inc., Dublin, California, USA), IOLMaster (Carl Zeiss Meditec, Jena, Germany) and Lenstar LS 900/Biograph (Haag-Streit AG, Koeniz, Switzerland/ Alcon Laboratories Inc., Ft Worth, Texas, USA). In this context, the purpose of this paper was to clarify which devices are interchangeable with Pentacam or Pentacam HR in the measurement of ACD distances. We also included a brief description of the measurement principle used by the devices included in this review. 1 Optometry Research Group (GIO), Department of Optics. University of Valencia. Spain. Financial disclosure: This research was supported in part by Research Grant #SAF E# awarded by the Ministry of Science and Innovation (Ministerio de Ciencia e Innovación) and by a BAF research scholarship (Generalitat Valenciana) awarded to Alberto Domínguez Vicent. The authors have no proprietary interest in any of the products or devices mentioned in this article. Corresponding Author: Alberto Domínguez Vicent. Department of Optics - University of Valencia. C/ Dr. Moliner, Burjassot. Spain. Phone: Fax: aldovi@alumni.uv.es MEASUREMENT PRINCIPLES Pentacam The Pentacam is a non-contact device using a rotating Scheimpflug camera. It uses a monochromatic slit light source to measure anterior segment topography. Up to 50 slit images with 500 measurement points on the front and back of the corneal surface are acquired in 2 seconds over a 180-degree rotation. The internal software creates a three-dimensional reconstruction of the anterior segment by using the elevation data of these images, which gives information about the anterior and posterior surface of the cornea, and ACD from endothelium to crystalline lens SECOIR Sociedad Española de Cirugía Ocular Implanto-Refractiva ISSN:

2 216 COMPARISON OF ANTERIOR CHAMBER DEPTH MEASUREMENT DEVICES Figure 1. Pentacam HR (left) and Galilei (right). Figure 2. Orbscan II (left) and Visante OCT (right). Pentacam HR Pentacam HR (Figure 1 left) is one of the latest developments in three-dimensional topographers and is based on the Scheimpflug principle. The rotating camera rotates around the optical axes of the eye to calculate a three-dimensional model of the anterior segment. The system has two different three-dimensional modes. One is a regular three-dimensional scan that takes 50 images in 2 seconds. The other is a special three-dimensional high-resolution corneal mode, in which the camera takes 50 images per second. The internal software creates a three-dimensional reconstruction of the anterior segment by using the elevation data of these images, giving information about the anterior and posterior surface of the cornea, and ACD from endothelium to crystalline lens. A-Scan This ultrasound contact device, which works at 10-MHz, is used to measure eye biometric dimensions, i.e. axial length, anterior chamber depth, posterior chamber depth and corneal and lens thickness. It calculates ACD on the basis of the difference in the time taken for ultrasound waves to reflect back to its receiver from the posterior corneal surface and anterior lens surface. Ultrasound Biomicroscopy This ultrasound contact device, which works at 50 MHz, is used to image ocular tissue depths up to 4 to 5 mm with lateral and axial physical resolution of approximately 50 and 25 µm. ACD is measured with a ruler over the sonogram. Orbscan The Orbscan is a noncontact topography system that is used for anterior segment evaluation. It uses two different techniques: a Placido disk system, which screens anterior corneal topography and creates an elevation map; and a horizontally moving scanning camera, which acquires slit images. After image reconstruction, a mathematical three-dimensional model of the cornea and the anterior segment is calculated. To calculate the ACD, the system software automatically detects the corneal endothelium surface and anterior surface of the crystalline lens on the acquired images. Orbscan II The Orbscan II (Figure 2 left) is a non-invasive topography system that scans the anterior segment, combining a three-dimensional scanning slit beam system with an added Placido attachment for evaluating corneal surfaces. This device calculates ACD in the same manner as Orbscan. Galilei The Galilei (Figure 1 right) is a non-invasive diagnostic system designed for the analysis of the anterior eye segment. The system is based on a rotating dual-scheimpflug camera integrated with a Placido topographer. This device captures slit images from opposite sides of the illuminated slit and averages the elevation data obtained from corresponding opposite slit images. This dual Scheimpflug imaging technique improves the detection of the posterior corneal surface and provides outstanding accuracy in pachymetry across the entire cornea, even when the camera is decentered because of eye movements. Visante OCT Visante OCT (Figure 2 right) is a time-domain non-invasive system that employs high-resolution images. The light source is a superluminescent light-emitting diode with a short wavelength that has

3 COMPARISON OF ANTERIOR CHAMBER DEPTH MEASUREMENT DEVICES 217 Figure 3. Lenstar (left) and IOLMaster (right). a limited penetration depth into the eye. By moving the scanning spot laterally across the eye, this device acquires multiple A-Scans and aligns them to construct two-dimensional images analogous to an ultrasound B-scan. The scanning speed is 4,000 axial scans/second and each image frame has 500 axial scans/image. It can be used to measure central corneal thickness and ACD. IOLMaster The IOLMaster (Figure 3 right) is designed to measure the parameters used in intraocular lens (IOL) calculation, including axial length, corneal curvature, ACD and WTW distance. The IOLMaster emits 780-nm infrared light and uses partial coherence interferometry to measure axial length. Keratometry and ACD are determined automatically using a lateral slit illumination of the corneal and crystalline lens and a white-light-emitting diode of 590 nm as the light source. The lateral slit illumination is 0.7 mm wide and employed at an angle of 30 degrees during ACD measurements. Lenstar The Lenstar (Figure 3 left) optical biometer is based on low coherence reflectometry, with an 820-mm superluminescent diode. The Lenstar detects anterior and posterior corneal and anterior crystalline lens peaks in the optical low coherence reflectometry waveform to measure the ACD and corneal thickness. In addition to ACD and corneal thickness, the Lenstar also measures axial length. ANTERIOR CHAMBER DEPTH Each study included in this study compared Pentacam or Pentacam HR with one of the following devices: A-Scan, UBM, Orbscan, Orbscan II, Galilei, Visante OCT, IOLMaster, Lenstar and MRI. Pentacam vs. A-Scan Németh et al. 7, Elbaz et al. 17 and Su et al. 18 studied the interchangeability of Pentacam and A-Scan. These authors included normal eyes in their sample but reached different conclusions: while Németh et al. 7 obtained comparable ACD values, Elbaz et al. 17 and Su et al. 18 obtained significantly lower ACD values with A-Scan. However, in these two latter studies the mean difference lay within clinical tolerance limits. Németh et al. 7 and Su et al. 18 also included pseudophakic eyes in their sample and in these cases the ACD values provided by A-Scan and Pentacam were not comparable. Therefore, it can be concluded that A-Scan and Pentacam can be used interchangeably in clinical practice for phakic subjects, but not for pseudophakic eyes. Pentacam HR vs. A-Scan Szalai et al. 19 compared ACD measures provided by Pentacam HR and A-Scan. They observed that A-Scan yielded significantly higher ACD values than Pentacam HR, thus concluding that these two devices cannot be assumed to be interchangeable in clinical practice. However, for IOL power calculation these devices can be used interchangeably, since the mean difference between these two devices is within clinical tolerance limits. Pentacam vs. Orbscan Lackner et al. 12, Kiraly et al. 20 and Yazici et al. 21 compared Orbscan with Pentacam, but the results they achieved were rather contradictor. Lackner et al. 11 obtained significantly higher ACD values with Orbscan, while in contrast, in the study of Yazici et al. 21, it was Orbscan that produced significantly lower ACD values. However, the mean difference observed by Lackner et al. 11 and by Yazici et al. 21 was within clinically acceptable levels, which is why these two authors concluded that these devices can be used interchangeably. Pentacam vs. Orbscan II Five studies 8,15,16,23,24 evaluated the interchangeability between Orbscan II and Pentacam. Doors et al. 8, Dinc et al. 16, Hashemi et al. 23 and Kim et al. 24 obtained statistically significant differences between the values produced by these two devices, thus concluding that they should not be used as though they were interchangeable. In contrast, Utine et al. 15 did not observe any statistically significant differences between Orbscan II and Pentacam, thus assuming their interchangeability. It is worth mentioning that the mean difference between Orbscan II and Pentacam was within clinically acceptable tolerance limits in four 8,15,16,24 out of these five studies, which suggests that further studies should be carried out to clarify whether or not these two devices are interchangeable for the measurement of ACD in normal eyes.

4 218 COMPARISON OF ANTERIOR CHAMBER DEPTH MEASUREMENT DEVICES Moreover, Doors et al. 8 measured ACD in phakic eyes following IOL implantation and Hashemi et al. 23 and Kim et al. 24 following laser refractive surgery. While in all three cases the authors observed statistically significant differences between Orbscan II and Pentacam, the mean difference was within clinically acceptance levels, except for the Doors et al. 8 study. Thus, we can conclude that these are interchangeable after refractive surgery but not after phakic IOL implantation. Pentacam HR vs. Orbscan II Salouti et al. 22 compared Pentacam HR with Orbscan II, obtaining significantly higher ACD values with Orbscan II, thus concluding that these devices should not be used interchangeably. Pentacam vs. Galilei Kiraly et al. 20 compared the ACD values obtained with Pentacam and Galilei. They did not find any statistically significant differences between the two devices, thus concluding that Pentacam and Galilei can be employed interchangeably in clinical practice. Pentacam vs. Visante OCT Doors et al. 8, Dinc et al. 16, Kiraly et al. 20, Yazici et al. 21 and O Donnell et al. 25 compared ACD measured with Pentacam and with Visante OCT. No statistically significant differences were found either by Kiraly et al. 20 or by Yazici et al. 21, thus concluding that these two devices can be used interchangeably in clinical practice. On the contrary, Door et al. 8, Dinc et al. 16 and O Donnell et al. 25 obtained significantly higher ACD values with Visante OCT than with Pentacam, thus concluding that these devices cannot be assumed to be interchangeable. However, since the mean difference obtained in these studies was within clinically acceptable limits, we believe that Pentacam and Visante OCT can be used interchangeably in normal eyes. Doors et al. 8 also included in their study a comparison of patients that had undergone phakic IOL implantation. In this group they also found statistically significant differences but the mean difference was again within clinical tolerance limits. Consequently, this study suggests that Pentacam and Visante OCT can be also used interchangeably to measure ACD in eyes that have undergone phakic IOL implantation. Pentacam vs. IOLMaster Eight studies 6,10,11,15-18,20 have so far compared Pentacam with IOLMaster for the measurement of ACD, all reaching similar conclusions. Savant et al. 6, Woodmass et al. 10, and Kiraly et al. 20 did not obtain any statistically significant differences between Pentacam and IOLMaster, and they concluded that these two devices may be used interchangeably in clinical practice. On the other hand, Reuland et al. 11, Utine et al. 15, Dinc et al. 16, Elbaz et al. 17 and Su et al. 18 observed statistically significant differences between Pentacam and IOLMaster, but they concluded that Pentacam and IOLMaster can be considered as interchangeable for use in clinical practice to calculate IOL power since the mean difference between these two devices lay within clinical tolerance levels. However, contradictory results were found for the estimation of IOL vaulting: while Reuland et al. 11, Utine et al. 15, Elbaz et al. 17 and Su et al. 18 obtained mean differences within clinical tolerance limits, Dinc et al. 16 did not. Moreover, Su et al. 18 also compared Pentacam with IOLMaster in pseudophakic subjects. Since the authors did not obtain any statistically significant differences, we may conclude that Pentacam and IOLMaster can be used interchangeably for the measurement of ACD in pseudophakic eyes. Pentacam vs. Lenstar Only two studies so far (O Donnell et al. 25 and Huang et al. 26 ) have compared Pentacam with Lenstar for the measurement of ACD. Although O Donnell et al. 25 found statistically significant differences between these devices, as the mean difference was acceptable from a clinical point of view, it was concluded that they were interchangeable. However, Huang et al. 26 did not find statistically significant differences between Pentacam and Lenstar. These studies, then, suggest that these two devices can be used interchangeably in clinical practice. Pentacam HR vs. Galilei Salouti et al. 22 also studied the interchangeability between Pentacam HR and Galilei. They did not obtain significantly differences and concluded that these devices could be used interchangeably. Pentacam HR vs. IOLMaster Németh et al. 27 studied the interchangeability of Pentacam HR and IOLMaster in phakic eyes. These authors did not find any statistically significant differences between these two devices. This study, then, suggests that Pentacam HR and IOLMaster can be used interchangeably in the clinical practice. DISCUSSION After these reviews, it can be concluded that the following devices can be used interchangeably with Pentacam or Pentacam HR in clinical practice: Pentacam-A-Scan 7,17,18 for phakic patients, Pentacam HR-A-Scan 19 to calculate IOL power, Pentacam- Orbscan 12,20,21 for phakic patients, Pentacam-Orbscan II 24 after laser refractive surgery, Pentacam-Galilei 20 for phakic patients, Pentacam-Visante OCT 8,16,20,21,25 for

5 COMPARISON OF ANTERIOR CHAMBER DEPTH MEASUREMENT DEVICES 219 phakic patients and after phakic IOL implantation, Pentacam-IOLMaster 6,10,11,15-18,20 to calculate IOL power and after pseudophakic IOL implantation, Pentacam-Lenstar 25,26 for phakic patients, Pentacam HR-Galilei 22 for phakic patients and Pentacam HR-IOLMaster 26 for phakic patients. The device pair whose interchangeability is still unclear is Pentacam-Orbscan II 8,15,16,20,23,24 for phakic patients. Moreover, the device pairs that cannot be used interchangeably in clinical practice are Pentacam-A- Scan 7,18 for pseudophakic patients, Pentacam-Orbscan II 8 after phakic IOL implantation and Pentacam HR-Orbscan II 22 for healthy subjects. The present review clarifies the interchangeability of ACD measurements between Pentacam or Pentacam HR and other devices. However, discrepancies between studies may be due to the differences in the measurement method upon which each device is based. For instance, A-Scan and UBM rely on ultrasound; Orbscan and Orbscan II rely on scanning-slit topography; Pentacam, Pentacam HR and Galilei use Scheimpflug photography; Visante OCT uses low coherence interferometry; IOLMaster uses a lateral slit illumination and is based on partial coherence interferometry; whereas Lenstar measures ACD with optical biometry. Other sources of discrepancies may be the age group included in each study and the ability of the researcher to control the accommodation state during measurement in those studies that included young subjects. Finally, further studies should be undertaken, aimed as studying the interchangeability of ACD measurements across different age groups, ocular conditions and interchangeability with other devices. It would also be interesting to define a relationship between the gold standard for ACD and the rest of the devices that are able to measure these distances. This would be useful for surgeons who could then know the ACD that a gold standard would yield without the need for having the device in the ophthalmology clinic. As a result, some postsurgical problems, such as those related to IOL power, could be avoided. To conclude, Pentacam (for phakic eyes) and Pentacam HR (for cataractous eyes) can be used interchangeably with the two gold standards used to measure ACD (IOLMaster and A-Scan). This review also reveals that the interchangeability of the device depends on the patients ocular condition and this must be taken into consideration in clinical practice. REFERENCES 1. Frisch BI, Rabsilber TM, Becker KA, Reuland AJ, Auffarth GU. Comparison of anterior chamber depth measurements using Orbscan II and IOLMaster. Eur J Ophthalmol. 2007; 17: Rabsilber TM, Becker KA, Frisch IB, Auffarth GU. Anterior chamber depth in relation to refractive status measured with the Orbscan II Topography System. J Cataract Refract Surg. 2003; 29: Hashemi H, Yazdani K, Mehravaran S, Fotouhi A. Anterior chamber depth measurement with a-scan ultrasonography, Orbscan II, and IOLMaster. Optom Vis Sci. 2005; 82: Salouti R, Nowroozzadeh MH, Zamani H, Ghoreyshi M, Salouti R. Comparison of horizontal corneal diameter measurements using Galilei, EyeSys and Orbscan II systems. Clin Exp Optom. 2009; 92(5): Baumeister M, Terzi E, Ekici Y, Kohnen T. Comparison of manual and automated methods to determine horizontal corneal diameter. J Cataract Refract Surg. 2004; 30: Savant V, Chavan R, Pushpoth S, Ilango B. Comparability and intra-/interobserver reliability of anterior chamber depth measurements with the Pentacam and IOLMaster. J Refract Surg. 2008; 24: Németh G, Vajas A, Kolozsvari B, Berta A, Modis L. Anterior chamber depth measurements in phakic and pseudophakic eyes: Pentacam versus ultrasound device. J Cataract Refract Surg. 2006; 32: Doors M, Cruysberg LPJ, Berendschot TTJM, Bradander J, Verbakel F, Webers CAB, Nujitis RMMA. Comparison of central corneal thickness and anterior chamber depth measurements using three imaging technologies in normal eyes and after phakic intraocular lens implantation. Graefes Arch Clin Exp Ophthalmol. 2009; 247: Kohnen T, Thomala MC, Cichocki M, Strenger A. Internal anterior chamber diameter using optical coherence tomography compared with white-to-white distances using automated measurements. J Cataract Refract Surg. 2006; 32: Woodmass J, Rocha G. A comparison of Scheimpflug imaging simulated and Holladay equivalent keratometry values with partial coherence interferometry keratometry measurements in phakic eyes. Can J Ophthalmol. 2009; 44: Reuland MS, Reuland AJ, Nishi Y, Auffarth GU. Corneal radii and anterior chamber depth measurements using the IOLMaster versus the Pentacam. J Refract Surg. 2007; 23: Lackner B, Schmidingrer G, Skorpik C. Validity and repeatability of anterior chamber depth measurements with Pentacam and Orbscan. Optom Vis Sci. 2005; 82: Reddy AR, Pande MV, Finn P, El-Gogary H. Comparative estimation of anterior chamber depth by ultrasonography, Orbscan II, and IOLMaster. J Cataract Refract Surg. 2004; 30: Rosa N, Lanza M, Capasso L, Lucci M, Polito B, Romano A. Anterior chamber depth measurement before and after photorefractive keratectomy. Ophthalmology. 2006; 113: Utine C.A, Altin F, Cakir H, Perente I. Comparison of anterior chamber depth measurements taken with the Pentacam, Orbscan IIz and IOLMaster in myopic and emmetropic eyes. Acta Ophthalmol. 2009; 87: Dinc UA, Gorgun E, Oncel B, Yenerel MN, Levent A. Assessment of anterior chamber depth using Visante Optical Coherence Tomography, slitlamp Optical Coherence Tomography, IOLMaster, Pentacam and Orbscan IIz. Ophthalmologica. 2010; 224: Elbaz U, Barkana Y, Gerber Y, Avni I, Zadok D. Comparison of different techniques of anterior chamber

6 220 COMPARISON OF ANTERIOR CHAMBER DEPTH MEASUREMENT DEVICES depth and keratometric measurements. Am J Ophthlmol. 2007; 143: Su PF, Lo A.Y, Hu C.Y, Chang SW. Anterior chamber depth measurement in phakic and pseudophakic eyes. Optom Vis Sci. 2008; 85: Szalai E, Berta A, Németh G, Hassan Z, Módis L. Anterior chamber depth measurements obtained with PentacamHR imaging system and conventional A-scan ultrasound. Ophthalmic Surg Lasers imaging. 2011; 42: Kiraly L, Duncker G. Biometry of the anterior eye segment for implantation of phakic anterior chamber lenses. A comparison of current measurement devices. Ophthalmologe. 2012; 109(3): Yazici AT, Bozkurt E, Alagoz C, et al. Central corneal thickness, anterior chamber depth, and pupil diameter measurements using Visante OCT, Orbscan, and Pentacam. J Refract Surg. 2010; 26: Salouti R, Nowroozzadeh MH, Zamani M, Ghoreyshi M, Salouti R. Comparison of anterior chamber depth measurements using Galilei, HR Pentacam, and Orbscan II. Optometry. 2010; 81: Hashemi H, Mehravaran S. Corneal changes after laser refractive surgery for myopia: comparison of Orbscan II and Pentacam findings. J Cataract Refract Surg. 2007; 33: Kim S.W, Sun J.S, Chang JH, Kim EK. Anterior segment measurement using Pentacam and Orbscan II 1 to 5 years after refractive surgery. J Refract Surg. 2009; 25: O Donnell C, Hartwig A, Radhakrishnan H. Comparison of central corneal thickness and anterior chamber depth measured using LenStar LS900, Pentacam and Visante AC-OCT. Cornea. 2012; 31(9): Huang J, Pesudovs K, Wen D, et al. Comparison of anterior segment measurements with rotating Scheimpflug photography and partial coherence reflectometry. J Cataract Refract Surg. 2011; 37(2): Németh G, Hassan Z, Módis L, Szalai E, Katona K, Berta A. Comparison of anterior chamber deoth measurement conducted with Pentacam HR and IOLMaster. Ophthalmic Surg Lasers imaging. 2011; 42: First author: Alberto Domínguez Vicent, MSc Optometry Research Group (GIO), Valencia (Spain)

Research Article Reproducibility of the Optical Biometer OA-1000 (Tomey)

Research Article Reproducibility of the Optical Biometer OA-1000 (Tomey) BioMed Research International, Article ID 814761, 6 pages http://dx.doi.org/10.1155/2014/814761 Research Article Reproducibility of the Optical Biometer OA-1000 (Tomey) Susanne Christiane Goebels, 1 Berthold

More information

Learn Connect Succeed. JCAHPO Regional Meetings 2017

Learn Connect Succeed. JCAHPO Regional Meetings 2017 Learn Connect Succeed JCAHPO Regional Meetings 2017 Faculty Biometry and IOL Calculations ASCRS and ASOA Symposium and Congress Los Angeles, CA Daniel H. Chang, M.D. - Empire Eye and Laser Center Bakersfield,

More information

Central corneal thickness (CCT) measurement is a critical

Central corneal thickness (CCT) measurement is a critical ARTICLE Topographic Paracentral Corneal Thickness With Pentacam and Orbscan: Effect of Acoustic Factor Javier González-Pérez, O.D., M.Sc., Ph.D., Jose M. González-Méijome, O.D., Ph.D., María T. Rodríguez

More information

INTRODUCTION J. DAWCZYNSKI, E. KOENIGSDOERFFER, R. AUGSTEN, J. STROBEL. Department of Ophthalmology, University Hospital Jena, Jena - Germany

INTRODUCTION J. DAWCZYNSKI, E. KOENIGSDOERFFER, R. AUGSTEN, J. STROBEL. Department of Ophthalmology, University Hospital Jena, Jena - Germany European Journal of Ophthalmology / Vol. 17 no. 3, 2007 / pp. 363-367 Anterior segment optical coherence tomography for evaluation of changes in anterior chamber angle and depth after intraocular lens

More information

Comparison of Central Corneal Thickness Measurements with Pentacam, Orbscan II, and Ultrasound Pachymeter

Comparison of Central Corneal Thickness Measurements with Pentacam, Orbscan II, and Ultrasound Pachymeter Comparison of Central Corneal Thickness Measurements with Pentacam, Orbscan II, and Ultrasound Pachymeter Abbas-Ali Yekta, PhD 1 Hassan Hashemi, MD 2,3 Mehdi KhabazKhoob, MSc 3 Asghar Dostdar, MSc 4 Shiva

More information

OPTOMETRY RESEARCH PAPER

OPTOMETRY RESEARCH PAPER C L I N I C A L A N D E X P E R I M E N T A L OPTOMETRY RESEARCH PAPER Agreement study of keratometric values measured by Biograph/LENSTAR, auto-kerato-refractometer and Pentacam: Decision for IOL calculation

More information

THE PENTACAM AXL. Improving Cataract Surgery Outcomes. Optical biometry and anterior segment tomography in one device

THE PENTACAM AXL. Improving Cataract Surgery Outcomes. Optical biometry and anterior segment tomography in one device Insert to November/December 2016 Sponsored by OCULUS THE PENTACAM AXL Improving Cataract Surgery Outcomes Optical biometry and anterior segment tomography in one device A New Way to Calculate IOL Power

More information

Biometry and intraocular lens power calculation Alexander C. Lee, Mujtaba A. Qazi and Jay S. Pepose

Biometry and intraocular lens power calculation Alexander C. Lee, Mujtaba A. Qazi and Jay S. Pepose Biometry and intraocular lens power calculation Alexander C. Lee, Mujtaba A. Qazi and Jay S. Pepose Pepose Vision Institute, and the Department of Ophthalmology and Visual Sciences, Washington University

More information

Jinhai Huang, 1,2 Weicong Lu, 1,2 Giacomo Savini, 3 Hao Chen, 1 Chengfang Wang, 1 Xinxin Yu, 1 Fangjun Bao, 1,2 and Qinmei Wang 1,2. 1.

Jinhai Huang, 1,2 Weicong Lu, 1,2 Giacomo Savini, 3 Hao Chen, 1 Chengfang Wang, 1 Xinxin Yu, 1 Fangjun Bao, 1,2 and Qinmei Wang 1,2. 1. Ophthalmology Volume 2016, Article ID 6347236, 5 pages http://dx.doi.org/10.1155/2016/6347236 Research Article Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence

More information

Predictability and accuracy of IOL formulas in high myopia

Predictability and accuracy of IOL formulas in high myopia ARTICLE Predictability and accuracy of IOL formulas in high myopia Mohamed Yasser Sayed Saif, MD 1 ; Mohamed Othman Abdel Khalek, MD 1 ; Ahmed Tamer Sayed Saif, MD 2 ; Passant Sayed Saif, MD 3 ; Sherif

More information

Accuracy of Biometry for Intraocular Lens Implantation Using the New Partial Coherence Interferometer, AL-scan

Accuracy of Biometry for Intraocular Lens Implantation Using the New Partial Coherence Interferometer, AL-scan pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2014;28(6):444-450 http://dx.doi.org/10.3341/kjo.2014.28.6.444 Original Article Accuracy of Biometry for Intraocular Lens Implantation Using the New

More information

Comparison of Intraocular Lens Power Calculation Methods Following Myopic Laser Refractive Surgery: New Options Using a Rotating Scheimpflug Camera

Comparison of Intraocular Lens Power Calculation Methods Following Myopic Laser Refractive Surgery: New Options Using a Rotating Scheimpflug Camera pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2018;32(6):497-505 https://doi.org/10.3341/kjo.2018.0008 Original Article Comparison of Intraocular Lens Power Calculation Methods Following Myopic

More information

Comparative Efficacy of the New Optical Biometer on Intraocular Lens Power Calculation (AL-Scan versus IOLMaster)

Comparative Efficacy of the New Optical Biometer on Intraocular Lens Power Calculation (AL-Scan versus IOLMaster) pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2018;32(3):241-248 https://doi.org/10.3341/kjo.2017.0063 Original Article Comparative Efficacy of the New Optical Biometer on Intraocular Lens Power

More information

Measurement of Angle Kappa Using Ultrasound Biomicroscopy and Corneal Topography

Measurement of Angle Kappa Using Ultrasound Biomicroscopy and Corneal Topography pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2017;31(3):257-262 https://doi.org/10.3341/kjo.2016.0021 Original Article Measurement of Angle Kappa Using Ultrasound Biomicroscopy and Corneal Topography

More information

FEP Medical Policy Manual

FEP Medical Policy Manual FEP Medical Manual 9.03.05 Corneal Topography/Computer-Assisted Corneal Topography/ Photokeratoscopy Last Review: September 2016 Next Review: September 2017 Related Policies 9.03.28 Corneal Collagen Cross-linking

More information

Cataract Surgery in the Patient with a History of LASIK or PRK

Cataract Surgery in the Patient with a History of LASIK or PRK Cataract Surgery in the Patient with a History of LASIK or PRK #56996-RS April 2018 Sebastian Lesniak, MD Matossian Eye Associates None Disclosures Bio Matossian Eye Associates, Hopewell NJ, 7/2015 Present

More information

IOL Master. Photorefractive keratectomy

IOL Master. Photorefractive keratectomy 89/11/3 : 89/9/3 : 1389 /119 / IOL Master * Photorefractive keratectomy 3 2 1 2 4 Photorefractive keratectomy(prk) (IOL) :. IoL-Master.. (clinical trial) :. 30 20 35.. PRK. IOL Master ( GM300,china) IOL.

More information

Corporate Medical Policy

Corporate Medical Policy Corporate Medical Policy Optical Coherence Tomography (OCT) Anterior Segment of the Eye File Name: Origination: Last CAP Review: Next CAP Review: Last Review: optical_coherence_tomography_(oct)_anterior_segment_of_the_eye

More information

Structural changes of the anterior chamber following cataract surgery during infancy

Structural changes of the anterior chamber following cataract surgery during infancy Structural changes of the anterior chamber following cataract surgery during infancy Matthew Nguyen, Emory University Marla Shainberg, Emory University Allen Beck, Emory University Scott Lambert, Emory

More information

Author s Affiliation. Original Article. Comparison of Biometry in Phakic and Dense Modes. Muhammad Suhail Sarwar. Unaiza Mariam

Author s Affiliation. Original Article. Comparison of Biometry in Phakic and Dense Modes. Muhammad Suhail Sarwar. Unaiza Mariam Comparison of Biometry in Phakic and Dense Modes Original Article Purpose: To compare biometry readings in phakic and dense modes. Author s Affiliation Muhammad Suhail Sarwar Unaiza Mariam Correspondence

More information

RECENT ADVANCES IN ANTERIOR SEGMENT IMAGING OF THE EYE. by Eszter Szalai, M.D. Supervisor: László Módis, M.D., Ph.D.

RECENT ADVANCES IN ANTERIOR SEGMENT IMAGING OF THE EYE. by Eszter Szalai, M.D. Supervisor: László Módis, M.D., Ph.D. SHORT THESIS FOR THE DEGREE OF DOCTOR OF PHILOSOPHY (Ph.D.) RECENT ADVANCES IN ANTERIOR SEGMENT IMAGING OF THE EYE by Eszter Szalai, M.D. Supervisor: László Módis, M.D., Ph.D. UNIVERSITY OF DEBRECEN DOCTORAL

More information

Comparison between Pentacam HR and Orbscan II after Hyperopic Photorefractive Keratectomy

Comparison between Pentacam HR and Orbscan II after Hyperopic Photorefractive Keratectomy Original Article Comparison between Pentacam HR and Orbscan II after Hyperopic Photorefractive Keratectomy Mahmoud Jabbarvand 1, MD; Farshad Askarizadeh 2, PhD; Mohamad Reza Sedaghat 3, MD Hadi Ghadimi

More information

Patient Selection IOL Power Calculation. Patient Selection. Biometry IOL-Power calculation using Vericalc 2.0. AC-Depth > 3.0 mm (FDA 3.

Patient Selection IOL Power Calculation. Patient Selection. Biometry IOL-Power calculation using Vericalc 2.0. AC-Depth > 3.0 mm (FDA 3. Verisyse and VeriFlex Patient Selection IOL Power Verisyse and VeriFlex Iris Claw Technology for Correcting Refractive Errors in Phakic and Aphakic Eyes ESCRS 2011 Intructional Course 61 Josef Ruckhofer

More information

Esteve-Taboada, JJ. University of Plymouth

Esteve-Taboada, JJ. University of Plymouth University of Plymouth PEARL Faculty of Health and Human Sciences https://pearl.plymouth.ac.uk School of Health Professions 2017-07 Non-invasive measurements of the dynamic changes in the ciliary muscle,

More information

A Comparative Study on the Inter-Session and Inter- Examiner Reliability of Corneal Power Measurement Using Various Keratometry Instruments

A Comparative Study on the Inter-Session and Inter- Examiner Reliability of Corneal Power Measurement Using Various Keratometry Instruments A Comparative Study on the Inter-Session and Inter- Examiner Reliability of Corneal Power Measurement Using Various Keratometry Instruments Md Muziman Syah MM a,b, Mutalib HA a, Sharanjeet Kaur MS a, Khairidzan

More information

Correcting inter-device bias of horizontal white-to-white and sulcus-to-sulcus measures used for implantable collamer lens sizing

Correcting inter-device bias of horizontal white-to-white and sulcus-to-sulcus measures used for implantable collamer lens sizing Correcting inter-device bias of horizontal white-to-white and sulcus-to-sulcus measures used for implantable collamer lens sizing Ivo Guber, MD 1, Ciara Bergin, PhD 1, Sylvie Perritaz, Optom 1, François

More information

EUROTIMES ESCRS SUPPLEMENT FEBRUARY 2007

EUROTIMES ESCRS SUPPLEMENT FEBRUARY 2007 Pentacam opens eyes to new diagnostic possibilities Scheimpflug based anterior segment tomography EUROTIMES ESCRS SUPPLEMENT FEBRUARY 2007 XXIV Congress of the ESCRS London, September 2006 Introduction

More information

A Study to Reach the Proper Measurement of the White to White Diameter Using both the Caliper and IOL Master for ICL Suitable Size

A Study to Reach the Proper Measurement of the White to White Diameter Using both the Caliper and IOL Master for ICL Suitable Size Current Science International Volume : 06 Issue : 01 Jan.- Mar. 2017 Pages: 41-49 A Study to Reach the Proper Measurement of the White to White Diameter Using both the Caliper and IOL Master for ICL Suitable

More information

INTRODUCTION MIKLOS SCHNEIDER 1, GABOR BORGULYA 2, ANDRAS SERES 1, ZOLTAN Z. NAGY 1, JANOS NEMETH 1

INTRODUCTION MIKLOS SCHNEIDER 1, GABOR BORGULYA 2, ANDRAS SERES 1, ZOLTAN Z. NAGY 1, JANOS NEMETH 1 European Journal of Ophthalmology / Vol. 19 no. 2, 2009 / pp. 180-187 Central corneal thickness measurements with optical coherence tomography and ultrasound pachymetry in healthy subjects and in patients

More information

3/23/2016. Diagnostic Services Taylor Pannell CRA, OCT-C. Services Available. Important info for the Tech to know. Visual Fields

3/23/2016. Diagnostic Services Taylor Pannell CRA, OCT-C. Services Available. Important info for the Tech to know. Visual Fields Services Available Diagnostic Services Taylor Pannell CRA, OCT-C Static and Kinetic Visual Fields Pachymetry Anterior and Posterior Segment OCT Fundus Photos FAF,FA,ICG Slit Lamp Photography Confocal HRT

More information

Preoperative anterior segment optical coherence tomography as a predictor of postoperative phakic intraocular lens position

Preoperative anterior segment optical coherence tomography as a predictor of postoperative phakic intraocular lens position ARTICLE Preoperative anterior segment optical coherence tomography as a predictor of postoperative phakic intraocular lens position Mohammad Reza Fallah Tafti, MD, Reza Soltani Moghadam, MD, Amir Houshang

More information

Corneal densitometry using Pentacam based scheimpflug imaging system: Indian rural population

Corneal densitometry using Pentacam based scheimpflug imaging system: Indian rural population Original article: Corneal densitometry using Pentacam based scheimpflug imaging system: Indian rural population Dr Nikhil Mahajan*, Prof. Swati Tomar** **Professor,*Resident Department of Ophthalmology,

More information

Factors Affecting the Accuracy of Intraocular Lens Power Calculation with Lenstar

Factors Affecting the Accuracy of Intraocular Lens Power Calculation with Lenstar Original Article www.cmj.ac.kr Factors Affecting the Accuracy of Intraocular Lens Power Calculation with Lenstar Tae Hee Lee, Mi Sun Sung, Lian Cui, Ying Li and Kyung Chul Yoon* Department of Ophthalmology,

More information

Partial Coherence Interferometry as a Technique to Measure the Axial Length of the Eye Archived Medical Policy

Partial Coherence Interferometry as a Technique to Measure the Axial Length of the Eye Archived Medical Policy Partial Coherence Interferometry as a Technique to Measure the Axial Length of the Eye Applies to all products administered or underwritten by Blue Cross and Blue Shield of Louisiana and its subsidiary,

More information

NIH Public Access Author Manuscript Cornea. Author manuscript; available in PMC 2014 May 29.

NIH Public Access Author Manuscript Cornea. Author manuscript; available in PMC 2014 May 29. NIH Public Access Author Manuscript Published in final edited form as: Cornea. 2013 December ; 32(12): 1544 1548. doi:10.1097/ico.0b013e3182a7f39d. Repeatability of corneal epithelial thickness measurements

More information

Comparison of Corneal and Anterior Chamber Parameters following Myopic laser in situ keratomileusis and photorefractive keratectomy by

Comparison of Corneal and Anterior Chamber Parameters following Myopic laser in situ keratomileusis and photorefractive keratectomy by Comparison of Corneal and Anterior Chamber Parameters following Myopic laser in situ keratomileusis and photorefractive keratectomy by Pentacam as A New Imaging Technique Mohammad Ali Zare, MD 1 Hassan

More information

Interpretation of corneal tomography

Interpretation of corneal tomography Interpretation of corneal tomography Presented by Chameen Samarawickrama - Westmead Hospital - Liverpool Hospital - University of Sydney - University of New South Wales The University of Sydney Page 1

More information

Corneal Power Measurements Using Scheimpflug Imaging in Eyes With Prior Corneal Refractive Surgery

Corneal Power Measurements Using Scheimpflug Imaging in Eyes With Prior Corneal Refractive Surgery Corneal Power Measurements Using Scheimpflug Imaging in Eyes With Prior Corneal Refractive Surgery Jack T. Holladay, MD, MSEE, FACS; Warren E. Hill, MD, FACS; Andreas Steinmueller, MSc ABSTRACT PURPOSE:

More information

ADVANCED PLANNING DEVICES

ADVANCED PLANNING DEVICES ADVANCED PLANNING DEVICES Surgeons discuss the strengths and weaknesses of four systems. The Galilei BY NEEL R. DESAI, MD It is no coincidence that the biometric analysis of potential refractive cataract

More information

Populations Interventions Comparators Outcomes Individuals: Who are being evaluated for angleclosure

Populations Interventions Comparators Outcomes Individuals: Who are being evaluated for angleclosure Protocol Optical Coherence Tomography of the Anterior Eye Segment (90318) Medical Benefit Effective Date: 07/01/14 Next Review Date: 05/18 Preauthorization No Review Dates: 07/11, 07/12, 07/13, 05/14,

More information

Routine OCT and UBM of the anterior segment

Routine OCT and UBM of the anterior segment Reprinted from No. 160 december 2012 Volume 17 Routine OCT and UBM of the anterior segment Michel Puech ISSN : 1274-5243 R e p r i n t e d w i t h t h e s u p p o r t o f t h e L a b o r a t o r y Q u

More information

OPTOMETRY ORIGINAL PAPER

OPTOMETRY ORIGINAL PAPER C L I N I C A L A N D E X P E R I M E N T A L OPTOMETRY ORIGINAL PAPER Repeatability and agreement of two A-scan ultrasonic biometers and IOLMaster in non-orthokeratology subjects and post-orthokeratology

More information

Comparison of Newer IOL Power Calculation Methods for Eyes With Previous Radial Keratotomy PATIENTS AND METHODS. Patients

Comparison of Newer IOL Power Calculation Methods for Eyes With Previous Radial Keratotomy PATIENTS AND METHODS. Patients Special Issue Comparison of Newer IOL Power Calculation Methods for Eyes With Previous Radial Keratotomy Jack X. Ma, 1 Maolong Tang, 2 Li Wang, 3 Mitchell P. Weikert, 3 David Huang, 2 and Douglas D. Koch

More information

Corneal Topography / Computer- Assisted Corneal Topography/Photokeratoscopy

Corneal Topography / Computer- Assisted Corneal Topography/Photokeratoscopy Corneal Topography / Computer- Assisted Corneal Topography/Photokeratoscopy Policy Number: 9.03.05 Last Review: 10/2018 Origination: 10/2000 Next Review: 10/2019 Policy Blue Cross and Blue Shield of Kansas

More information

Estimation accuracy of surgically induced astigmatism on the cornea when neglecting the posterior corneal surface measurement

Estimation accuracy of surgically induced astigmatism on the cornea when neglecting the posterior corneal surface measurement Estimation accuracy of surgically induced astigmatism on the cornea when neglecting the posterior corneal surface measurement Li-Sheng Cheng, 1 Ching-Yao Tsai, 2,3 Ray Jui-Fang Tsai, 4 Shiow-Wen Liou,

More information

CENTRAL CORNEAL THICKNESS; COMPARISON OF CENTRAL CORNEAL THICKNESS MEASUREMENTS BY PENTACAM AND ULTRASOUND PACHYMETRY, IN MYOPIC AND KERATOCONUS EYES.

CENTRAL CORNEAL THICKNESS; COMPARISON OF CENTRAL CORNEAL THICKNESS MEASUREMENTS BY PENTACAM AND ULTRASOUND PACHYMETRY, IN MYOPIC AND KERATOCONUS EYES. The Professional Medical Journal DOI: 10.29309/TPMJ/18.4609 1. FCPS Assistant Professor Allied Hospital, Punjab Medical College, Faisalabad, Pakistan. 2. FCPS Assistant Professor Liaquat University of

More information

Keratometric measurements and IOL calculations in pseudophakic post-dsaek patients

Keratometric measurements and IOL calculations in pseudophakic post-dsaek patients Xu et al. BMC Ophthalmology (2018) 18:268 https://doi.org/10.1186/s12886-018-0931-y RESEARCH ARTICLE Open Access Keratometric measurements and IOL calculations in pseudophakic post-dsaek patients Ke Xu,

More information

Prediction accuracy of intraocular lens power calculation methods after laser refractive surgery

Prediction accuracy of intraocular lens power calculation methods after laser refractive surgery Wu et al. BMC Ophthalmology (2017) 17:44 DOI 10.1186/s12886-017-0439-x RESEARCH ARTICLE Open Access Prediction accuracy of intraocular lens power calculation methods after laser refractive surgery Yubo

More information

Description of iatrogenic corneal ectasia in patients without traditional risk factors

Description of iatrogenic corneal ectasia in patients without traditional risk factors ARTICLE Description of iatrogenic corneal ectasia in patients without traditional risk factors Julio Ortega-Usobiaga, MD, PhD 1 ; Rosario Cobo-Soriano, MD, PhD 1 ; Fernando Llovet-Osuna, MD, PhD 1 ; Stephan

More information

Refractive Dilemma. Challenging Case

Refractive Dilemma. Challenging Case Challenging Case Refractive Dilemma Section Editor: Alireza Baradaran-Rafii, MD Case presentation A 21-year old man was referred to an ophthalmology clinic insisting on getting rid of his glasses which

More information

MEDICAL POLICY SUBJECT: CORNEAL ULTRASOUND PACHYMETRY. POLICY NUMBER: CATEGORY: Technology Assessment

MEDICAL POLICY SUBJECT: CORNEAL ULTRASOUND PACHYMETRY. POLICY NUMBER: CATEGORY: Technology Assessment MEDICAL POLICY SUBJECT: CORNEAL ULTRASOUND,, PAGE: 1 OF: 5 If a product excludes coverage for a service, it is not covered, and medical policy criteria do not apply. If a commercial product, including

More information

Issue 15 The following key clinical peer reviewed journals will be reviewed: MONTHLY RESEARCH UPDATE 151(3) American Journal of Ophthalmology 129(5)

Issue 15 The following key clinical peer reviewed journals will be reviewed: MONTHLY RESEARCH UPDATE 151(3) American Journal of Ophthalmology 129(5) Welcome to Bausch and Lomb s monthly research update. With our background in clinical ophthalmic research, mainly of the anterior eye, Bausch and Lomb have asked us to produce an independent report of

More information

Artiflex Toric Phakic Intraocular Lens Implantation in Congenital Nystagmus

Artiflex Toric Phakic Intraocular Lens Implantation in Congenital Nystagmus 273 This is an Open Access article licensed under the terms of the Creative Commons Attribution- NonCommercial-NoDerivs 3.0 License (www.karger.com/oa-license), applicable to the online version of the

More information

Evaluation of Anterior Segment Parameter Changes Using the Sirius after Uneventful Phacoemulsification

Evaluation of Anterior Segment Parameter Changes Using the Sirius after Uneventful Phacoemulsification pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2016;30(4):251-257 http://dx.doi.org/10.3341/kjo.2016.30.4.251 Original Article Evaluation of Anterior Segment Parameter Changes Using the Sirius after

More information

The World s fastest OCT. As simple as pressing. the start button

The World s fastest OCT. As simple as pressing. the start button The World s fastest OCT As simple as pressing the start button lution continues Optopol engineering team, designers of the first commercially available Spectral Domain OCT in the world, are proud to present

More information

Updated practical intraocular lens power calculation after refractive surgery

Updated practical intraocular lens power calculation after refractive surgery REVIEW C URRENT OPINION Updated practical intraocular lens power calculation after refractive surgery Gowtham Jonna and Prabjot Channa Purpose of review Since its introduction in the 1980s, more than 40

More information

Accuracy of swept-source optical coherence tomography based biometry for intraocular lens power calculation: a retrospective cross sectional study

Accuracy of swept-source optical coherence tomography based biometry for intraocular lens power calculation: a retrospective cross sectional study An et al. BMC Ophthalmology (2019) 19:30 https://doi.org/10.1186/s12886-019-1036-y RESEARCH ARTICLE Open Access Accuracy of swept-source optical coherence tomography based biometry for intraocular lens

More information

Accuracy and Repeatability of a New Portable Ultrasound Pachymeter

Accuracy and Repeatability of a New Portable Ultrasound Pachymeter Accuracy and Repeatability of a New Portable Ultrasound Pachymeter Queirós A 1, González-Méijome JM 1, Fernandes P 1, Jorge J 1, Almeida J B 1, Parafita MA 2 1 Department of Physics (Optometry), School

More information

Bilateral Microphthalmos Associated with Papillomacular Fold, Severe Hyperopia and Steep Cornea

Bilateral Microphthalmos Associated with Papillomacular Fold, Severe Hyperopia and Steep Cornea Bilateral Microphthalmos Associated with Papillomacular Fold, Severe Hyperopia and Steep Cornea Mojtaba Abrishami, MD 1 Alireza Maleki, MD 2 Ali Hamidian-Shoormasti, MD 3 Mostafa Abrishami, MD 4 Abstract

More information

What is the Value of Swept Source oct Technology in Biometry? Experts discussed the IOLMaster 700 at the ESCRS ebook. Content provided by:

What is the Value of Swept Source oct Technology in Biometry? Experts discussed the IOLMaster 700 at the ESCRS ebook. Content provided by: ebook Content provided by: What is the Value of Swept Source oct Technology in Biometry? Experts discussed the IOLMaster 700 at the ESCRS 2016 Participating experts: G. Barrett, MD, Australia; D. Chang,

More information

Comparison of ultrasound and optic biometry with respect to eye refractive errors after phacoemulsification

Comparison of ultrasound and optic biometry with respect to eye refractive errors after phacoemulsification Received: 5.5.2007 Accepted: 22.1.2008 Comparison of ultrasound and optic biometry with respect to eye refractive errors after phacoemulsification Heidarali Moeini*, Fateme Eslami**, Akram Rismanchian*,

More information

Refractive Changes after Removal of Anterior IOLs in Temporary Piggyback IOL Implantation for Congenital Cataracts

Refractive Changes after Removal of Anterior IOLs in Temporary Piggyback IOL Implantation for Congenital Cataracts pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2013;27(2):93-97 http://dx.doi.org/10.3341/kjo.2013.27.2.93 Original Article Refractive Changes after Removal of Anterior IOLs in Temporary Piggyback

More information

CATARACT SURGERY AFTER RADIAL KERATOTOMY

CATARACT SURGERY AFTER RADIAL KERATOTOMY AFTER RADIAL KERATOTOMY How to avoid disappointment. BY BERNARD MATHYS, MD CATARACT SURGERY Radial keratotomy (RK; Figure 1) was a popular refractive surgical procedure to correct myopia in the 1970s and

More information

Inaccuracy of Intraocular Lens Power Prediction for Cataract Surgery in Angle-Closure Glaucoma

Inaccuracy of Intraocular Lens Power Prediction for Cataract Surgery in Angle-Closure Glaucoma Original Article DOI 10.3349/ymj.2009.50.2.206 pissn: 0513-5796, eissn: 1976-2437 Yonsei Med J 50(2):206-210, 2009 Inaccuracy of Intraocular Lens Power Prediction for Cataract Surgery in Angle-Closure

More information

Accuracy of Intraocular Lens Calculation Formulas

Accuracy of Intraocular Lens Calculation Formulas Accuracy of Intraocular Lens Calculation Formulas Ronald B. Melles, MD, 1 Jack T. Holladay, MD, MSEE, 2 William J. Chang, MD 1 Purpose: To compare the accuracy of intraocular lens (IOL) calculation formulas

More information

03/04/2015. LOC Talk Anterior Chamber & Gonioscopy 1st April Methods of Assessing Anterior Chamber Depth (and angle width) Outline

03/04/2015. LOC Talk Anterior Chamber & Gonioscopy 1st April Methods of Assessing Anterior Chamber Depth (and angle width) Outline LOC Talk Anterior & 1st April 2015 Mr. Areeb Moosavi MBBS BSc FRCOphth Glaucoma Consultant Milton Keynes University Hospital NHS Foundation Trust Methods of Assessing Anterior Open Versus Closed angle

More information

Trabeculectomy is an effective method for lowering

Trabeculectomy is an effective method for lowering ORIGINAL STUDY Refractive Outcome of Cataract Surgery in Eyes With Prior Trabeculectomy: Risk Factors for Postoperative Myopia Oliver L. Yeh, MD, Karine D. Bojikian, MD, Mark A. Slabaugh, MD, and Philip

More information

Comparison of the Effects of Bright Light, Phenylephrine 2.5%, Tropicamide 1%, and Pilocarpine 2% on Anterior Chamber Depth and IOL Power

Comparison of the Effects of Bright Light, Phenylephrine 2.5%, Tropicamide 1%, and Pilocarpine 2% on Anterior Chamber Depth and IOL Power Comparison of the Effects of Bright Light, Phenylephrine 2.5%, Tropicamide 1%, and Pilocarpine 2% on Anterior Chamber Depth and IOL Power Mona Sane MD, Vincent Imbrogno DO, Hoon Jung MD The authors have

More information

Advancing Cataract Surgery With Diagnostic Technologies

Advancing Cataract Surgery With Diagnostic Technologies Advancing Cataract Surgery With Diagnostic Technologies Surgeons respond to the questions: What diagnostic tests are essential for the best refractive outcomes in cataract surgery, and why? BY DAVID F.

More information

The Harvard community has made this article openly available. Please share how this access benefits you. Your story matters.

The Harvard community has made this article openly available. Please share how this access benefits you. Your story matters. Central Corneal Thickness in Highly Myopic Eyes: Inter-device Agreement of Ultrasonic Pachymetry, Pentacam and Orbscan II Before and After Photorefractive Keratectomy The Harvard community has made this

More information

Ultrasound Biomicroscopy & Glaucoma Care

Ultrasound Biomicroscopy & Glaucoma Care S P E C IA L S E C T I O N S P O N S O R E D B Y Q UA N T E L M E D I C A L Ultrasound Biomicroscopy & Glaucoma Care Visualizing angle closure and its mechanisms from screening to post-surgical assessment

More information

BioMedical Engineering OnLine. Open Access RESEARCH. Michele Lanza 1*, Robert Koprowski 2 and Mario Bifani Sconocchia 1

BioMedical Engineering OnLine. Open Access RESEARCH. Michele Lanza 1*, Robert Koprowski 2 and Mario Bifani Sconocchia 1 https://doi.org/10.1186/s12938-018-0542-0 BioMedical Engineering OnLine RESEARCH Open Access Improving accuracy of corneal power measurement with partial coherence interferometry after corneal refractive

More information

IOL Calculations in Short, Long, and Postrefractive Eyes

IOL Calculations in Short, Long, and Postrefractive Eyes IOL Calculations in Short, Long, and Postrefractive Eyes Ildamaris Montes de Oca, MD Sabite E. Gökce, MD Katie Hallahan, MD Li Wang, MD, PhD Douglas D. Koch, MD Intraocular lens (IOL) Calculations: Fundamentals

More information

Over the past decade, new anterior chamber (AC) imaging

Over the past decade, new anterior chamber (AC) imaging Glaucoma Biometric Evaluation of Anterior Chamber Changes after Physiologic Pupil Dilation Using Pentacam and Anterior Segment Optical Coherence Tomography Florent Aptel, 1,2 Christophe Chiquet, 1,2 Sylvain

More information

ZEISS presents innovations that support ophthalmologists in their work

ZEISS presents innovations that support ophthalmologists in their work ZEISS presents innovations that support ophthalmologists in their work At the 2015 Annual Meeting of the American Academy of Ophthalmology (AAO) in Las Vegas, ZEISS announces innovations for clinical excellence

More information

Comparison of Corneal Power and Intraocular Lens Power Calculation Methods after LASIK for Myopia

Comparison of Corneal Power and Intraocular Lens Power Calculation Methods after LASIK for Myopia Comparison of Corneal Power and Intraocular Lens Power Calculation Methods after LASIK for Myopia Seyed Mohammad Reza Taheri, MD 1 Azita Kheiltash, MD, MPH 2 Hassan Hashemi, MD 1,3 Abstract Purpose: To

More information

Axial Length Shortening After Cataract Surgery: New Approach to Solve the Question

Axial Length Shortening After Cataract Surgery: New Approach to Solve the Question Article https://doi.org/10.1167/tvst.7.6.34 Axial Length Shortening After Cataract Surgery: New Approach to Solve the Question Maddalena De Bernardo 1, Giulio Salerno 1, Palmiro Cornetta 1, and Nicola

More information

The Effect of Corneal Irregularity on Astigmatism Measurement by Automated versus Ray Tracing Keratometry

The Effect of Corneal Irregularity on Astigmatism Measurement by Automated versus Ray Tracing Keratometry The Effect of Corneal Irregularity on Astigmatism Measurement by Automated versus Ray Tracing Keratometry Hyun Cheol Roh, MD, Roy S. Chuck, MD, PhD, Jimmy K. Lee, MD, and Choul Yong Park, MD, PhD Abstract:

More information

The Comparison of Central and Mean True-Net Power (Pentacam) in Calculating IOL-Power After Refractive Surgery

The Comparison of Central and Mean True-Net Power (Pentacam) in Calculating IOL-Power After Refractive Surgery 접수번호 :08-024 Korean Journal of Ophthalmology 2009;23:1-5 ISSN : 1011-8942 DOI : 10.3341/kjo.2009.23.1.1 The Comparison of Central and Mean True-Net Power (Pentacam) in Calculating IOL-Power After Refractive

More information

Central and Peripheral Changes in Anterior Corneal Topography after Orthokeratology and Laser in situ Keratomileusis

Central and Peripheral Changes in Anterior Corneal Topography after Orthokeratology and Laser in situ Keratomileusis Central and Peripheral Changes in Anterior Corneal Topography after Orthokeratology and Laser in situ Keratomileusis Han-Yin Sun 1, 2, 3, Hsiu-Wan Yang 4, I-Tsung Wu 4, Jung-Kai Tseng 2, 3 1, 5* and Shun-Fa

More information

Immersion Vs Contact Biometery for Axial Length Measurement before Phacoemulsification

Immersion Vs Contact Biometery for Axial Length Measurement before Phacoemulsification Original Article Immersion Vs Contact Biometery for Axial Length Measurement before Phacoemulsification with Foldable IOL Irum Abbas, Atif Mansoor Ahmad, Tahir Mahmood Pak J Ophthalmol 2009, Vol. 25 No.

More information

BY MICHAEL W. BELIN, MD

BY MICHAEL W. BELIN, MD Point/Counterpoint: The Pentacam Ve The Pentacam s rotating imaging provides a more accurate picture of the cornea. BY MICHAEL W. BELIN, MD It was not long ago that discussions concerned whether topography

More information

Change of Capsulotomy Over 1 Year in Femtosecond Laser-Assisted Cataract Surgery and Its Impact on Visual Quality

Change of Capsulotomy Over 1 Year in Femtosecond Laser-Assisted Cataract Surgery and Its Impact on Visual Quality ORIGINAL ARTICLE Change of Capsulotomy Over 1 Year in Femtosecond Laser-Assisted Cataract Surgery and Its Impact on Visual Quality Christophe Panthier, MD; Florent Costantini, MD; Jean Claude Rigal-Sastourné,

More information

The aim of refractive surgery is to modify the refractive

The aim of refractive surgery is to modify the refractive Phakic IOLs: An Overview These lenses are fundamental tools in a successful refractive surgery practice. By António Marinho, MD, PhD The aim of refractive surgery is to modify the refractive power of the

More information

JACK T. HOLLADAY, MD, MSEE, FACS (MODERATOR) MICHAEL W. BELIN, MD, FACS ARTURO S. CHAYET, MD MATTHIAS MAUS, MD PAOLO VINCIGUERRA, MD

JACK T. HOLLADAY, MD, MSEE, FACS (MODERATOR) MICHAEL W. BELIN, MD, FACS ARTURO S. CHAYET, MD MATTHIAS MAUS, MD PAOLO VINCIGUERRA, MD Next-Generation Technology for the Cataract & Refractive Surgeon Produced under an unrestricted educational grant from Oculus, Inc., and Oculus Optikgeraete GmbH. This monograph is based on an educational

More information

Toric IOL implantation in a patient with keratoconus and previous penetrating keratoplasty: a case report and review of literature

Toric IOL implantation in a patient with keratoconus and previous penetrating keratoplasty: a case report and review of literature Allard and Zetterberg BMC Ophthalmology (2018) 18:215 https://doi.org/10.1186/s12886-018-0895-y CASE REPORT Toric IOL implantation in a patient with keratoconus and previous penetrating keratoplasty: a

More information

Medical Policy An independent licensee of the Blue Cross Blue Shield Association

Medical Policy An independent licensee of the Blue Cross Blue Shield Association Optical Coherence Tomography (OCT) of the Anterior Eye Segment Page 1 of 15 Medical Policy An independent licensee of the Blue Cross Blue Shield Association Title: Optical Coherence Tomography (OCT) of

More information

Why Cataract and Refractive Surgeons Need the Pentacam Four surgeons discuss the device s value, including new software additions.

Why Cataract and Refractive Surgeons Need the Pentacam Four surgeons discuss the device s value, including new software additions. Why Cataract and Refractive Surgeons Need the Pentacam Four surgeons discuss the device s value, including new software additions. Produced under an unrestricted educational grant from Oculus, Inc.. Extremely

More information

Biometric Measurements: Average Intra-Ocular Lens Power for Sudanese Patients with Senile Cataract

Biometric Measurements: Average Intra-Ocular Lens Power for Sudanese Patients with Senile Cataract Optics 2015; 4(4): 25-30 Published online September 28, 2015 (http://www.sciencepublishinggroup.com/j/optics) doi: 10.11648/j.optics.20150404.11 ISSN: 2328-7780 (Print); ISSN: 2328-7810 (Online) Biometric

More information

Repeatability and reproducibility of corneal thickness using SOCT Copernicus HR.

Repeatability and reproducibility of corneal thickness using SOCT Copernicus HR. 1 Repeatability and reproducibility of corneal thickness using SOCT Copernicus HR. Silvia Vidal MSc, Valentín Viqueira PhD, David Mas PhD, Begoña Domenech PhD. Dept. Optica, Universidad de Alicante, c/

More information

Myopic Shift after Implantation of a Novel Diffractive Trifocal Intraocular Lens in Korean Eyes

Myopic Shift after Implantation of a Novel Diffractive Trifocal Intraocular Lens in Korean Eyes pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2018;32(1):16-22 https://doi.org/10.3341/kjo.2017.0060 Original Article Myopic Shift after Implantation of a Novel Diffractive Trifocal Intraocular

More information

Medical Policy. MP Optical Coherence Tomography of the Anterior Eye Segment

Medical Policy. MP Optical Coherence Tomography of the Anterior Eye Segment Medical Policy MP 9.03.18 BCBSA Ref. Policy: 9.03.18 Last Review: 03/29/2018 Effective Date: 03/29/2018 Section: Other Related Policies 9.03.05 Corneal Topography/Computer-Assisted Corneal Topography/Photokeratoscopy

More information

IOL Power Calculation for Children

IOL Power Calculation for Children 1 IOL Power Calculation for Children Rupal H. Trivedi, MD MSCR M. Edward Wilson, MD The authors have no financial interest in the subject matter of this presentation. Intraocular lens (IOL) implantation

More information

Anterior segment imaging

Anterior segment imaging CET CONTINUING Sponsored by 1 CET POINT Anterior segment imaging Sundeep Vaswani, BSc (Hons), MCOptom 39 The anterior segment of the eye encompasses all structures from the front surface of the cornea

More information

Effects of Choroidal Thickness on Refractive Outcome Following Cataract Surgery in Primary Angle Closure

Effects of Choroidal Thickness on Refractive Outcome Following Cataract Surgery in Primary Angle Closure pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2018;32(5):382-390 https://doi.org/10.3341/kjo.2017.0129 Original Article Effects of Choroidal Thickness on Refractive Outcome Following Cataract Surgery

More information

Refractive Surgery Dilemma

Refractive Surgery Dilemma Refractive Surgery Dilemma Section Editor: lireza aradaran-rafii, MD CSE PRESENTTION 33-year-old man seeking refractive surgery presented with refractive error of -1.75-4.0 20 in the right and -0.75-2.5

More information

PRIMUS 200 from ZEISS The essential OCT

PRIMUS 200 from ZEISS The essential OCT EN 00_00I The contents of the brochure may differ from the current status of approval of the product in your country. Please contact your regional representative for more information. Subject to change

More information

Anterior segment imaging

Anterior segment imaging Article Date: 11/1/2016 Anterior segment imaging AS OCT vs. UBM vs. endoscope; case based approaches BY BENJAMIN BERT, MD, FACS AND BRIAN FRANCIS, MD, MS Currently, numerous imaging modalities are available

More information

Central Corneal Thickness and its Association with Ocular Parameters

Central Corneal Thickness and its Association with Ocular Parameters Central Corneal Thickness and its Association with Ocular Parameters Authors: An-Fei Li 1,2 Mei-Ju Chen 1,2 Affiliations: 1 Department of Ophthalmology, Taipei Veterans General Hospital 2 National Yang-Ming

More information