MRI Guidelines for the Axonics Sacral Neuromodulation System
|
|
- Nickolas Cain
- 6 years ago
- Views:
Transcription
1 MRI Guidelines for the Axonics Sacral Neuromodulation System For Healthcare Professionals Rx only Note: This document is a supplement to the physician manuals provided with the Axonics SNM System. This document only contains information related to MRI use with the Axonics SNM System. For more information on the Axonics SNM System, consult the Tined Lead Implant Manual and the Neurostimulator Implant Manual. Axonics, Axonics Modulation, Axonics Modulation Technologies and Axonics Sacral Neuromodulation System are trademarks of Axonics Modulation Technologies, Inc., registered or pending registration in the U.S. and other countries. English June,
2 Table of Contents SAFE MRI CONDITIONS... 3 POTENTIAL RISKS OF MRI WITH THE AXONICS SNM SYSTEM... 3 Heating Effects... 3 Unintended Stimulation... 4 Image Distortion and Artifacts... 4 Magnetic Field Interactions... 4 Device Malfunction or Damage... 4 Other Precautions... 4 MRI Unsafe Devices... 5 MRI GUIDELINES... 5 Before Starting MRI Head Scan... 5 During MRI Head Scan... 6 Post MRI Head Scan... 7 GLOSSARY... 8 English June,
3 SAFE MRI CONDITIONS The Axonics Sacral Neuromodulation (SNM) System is an MRI Conditional device. MRI examinations of the head can be safely conducted in patients with the Axonics SNM System under the following conditions: Head transmit/receive coil only (No RF transmit body coil) Static magnetic field of 1.5T or 3T horizontal closed bore Normal operating mode Spatial gradient field of 2,500 Gauss/cm or less Gradient slew rate limited to 200 T/m/s Maximum head-averaged specific absorption rate (SAR) of 3.2 W/kg. Axonics Neurostimulator (Model 1101) and Tined lead (Model 1201) implanted Turn the Axonics Neurostimulator Stimulation off In vitro tests and simulations have shown that the patients with the Axonics SNM System may be safely exposed to MRI environments that follow the guidelines described in this document. When using a head transmit coil, the Axonics SNM System is expected to produce a maximum temperature rise of less than 1 C at a maximum head-averaged specific absorption rate (SAR) of 3.2 W/kg. Always obtain the latest MRI guidelines. Refer to the contact information on the last page of this manual, or go to POTENTIAL RISKS OF MRI WITH THE AXONICS SNM SYSTEM The potential risks of performing MRI on patients with an implanted SNM System include: Heating effects around the SNM System, especially electrodes, from radiofrequency (RF) energy Unintended stimulation due to current induced through the SNM lead wire by the timevarying magnetic gradient field or RF field Image distortion and artifacts Magnetic field interactions Device malfunction or damage Heating Effects Modeling and simulation show that with the RF transmit head coil, heating of the Axonics SNM System is minimal. If the specific MRI conditions are not met, e.g., RF body coil is used, heating at a lead electrode could damage the sacral nerve and/or surrounding structures. MRI-related heating is primarily influenced by location of the patient in the MR system and by lead configuration and length. English June,
4 Unintended Stimulation Non-clinical testing suggests that gradient induced or combined field induced current is small so that unintended stimulation to the surrounding tissue is unlikely. Risk of tissue damage due to current induced by the gradient or combined field is extremely low. If a patient suspects any unintended stimulation while in MRI, he/she should inform the MRI technician immediately and then contact their physician. Image Distortion and Artifacts No artifacts or distortion of the brain imaging should be seen when imaging with a RF Head Coil. Magnetic Field Interactions The Axonics SNM System may experience magnetic field interactions associated with the static magnetic field of the MRI system due to small amounts of material in the Neurostimulator being sensitive to magnetic fields. This may cause the Neurostimulator to shift or move slightly within the implant pocket and/or may place mechanical stress on tissues and/or the lead. Patients may feel a slight tugging sensation at the site of the Neurostimulator. Device Malfunction or Damage Tests in various MRI systems have not shown damage to, or malfunction of, the Axonics SNM System. If device malfunction or damage were to occur, it could cause discomfort, unintended stimulation, painful stimulation or direct current stimulation which may result in nerve damage and other associated problems. If a patient suspects a malfunction, he/she should be instructed to exit the MRI system room and use the patient Remote Control to stop stimulation. The patient should then contact their physician for further evaluation. Other Precautions MRI should not be performed in patients with multiple implanted devices. Safety has not been demonstrated in patients with implanted devices in addition to the Axonics SNM System. Do not use the transmit RF body coil for 1.5 T or 3 T imaging. Not all head RF coils are transmit and receive type. Many are receive only. The use of any local receive coil with the body coil in RF transmit mode presents the same RF heating hazards as the body coil alone with no local coils. Open MRI scanners have not been evaluated and should not be used for scanning SNM patients. Exposure of the Axonics SNM System to any RF transmit coil must be avoided. Surgical removal of the entire Axonics Therapy System may be required if an MRI of the lumbar or pelvic area is needed. External devices are MRI Unsafe in the MRI environment. Refer to MRI Unsafe Device section for details. No testing at field strengths other than 1.5 T and 3 T has been performed to evaluate migration or heating. English June,
5 MRI Unsafe Devices The Axonics SNM System Clinician Programmer, Remote Control and Charger and Dock (Figure 1) are MRI Unsafe. These devices must NOT be brought into the MRI system room. Clinician Programmer Remote Control Charger and Dock Figure 1. MRI Unsafe Devices MRI GUIDELINES Recommendations for Head Coil MRI with the Axonics SNM System are based on phantom tests and numerical simulations of worst-case and with the recommended implant configurations of the standard Axonics SNM Neurostimulator (Model 1101) and 30-cm Tined Lead (Model 1201). Before Starting MRI Head Scan 1. Turn the Axonics SNM Neurostimulator (Model 1101) stimulation off with the patient Remote Control or Clinician Programmer. English June,
6 2. Make sure the MRI System and settings used meet the following conditions: Static magnetic field of 1.5T or 3T horizontal closed bore only Spatial gradient field of 2,500 Gauss/cm or less Gradient slew rate limited to 200 T/m/s Normal operating mode only Head transmit/receive coils used only (No RF transmit body coil) Maximum head-averaged specific absorption rate (SAR) of 3.2 W/kg. 3. The patient should be placed in the MRI system in a supine position as shown in Figure 2. Figure 2. Head MR Imaging During MRI Head Scan 1. Monitor the patient both visually and audibly. Discontinue the MRI examination immediately if the patient reports any problems. 2. During the MRI head scan, the patient may feel slight tugging and/or vibration of the neurostimulator. If the tugging or vibration causes the patient considerable discomfort, stop the MRI head scan. English June,
7 After MRI Head Scan After the MRI procedure, turn the Axonics SNM Neurostimulator stimulation back on with the patient Remote Control or Clinician Programmer. If a patient suspects any unexpected change in stimulation after an MRI, he/she should contact their physician. English June,
8 GLOSSARY Hertz (Hz) a unit of frequency defined as cycles per second. One Megahertz (MHz) is one million cycles per second. MR Conditional an item with demonstrated safety in the MR environment within defined conditions. At a minimum, these address the conditions of the static magnetic field, the switched gradient magnetic field and the radio frequency fields. Additional conditions, including specific configurations of the item, may be required. MRI Magnetic Resonance Imaging. Radio Frequency (RF) high frequency electrical fields whose frequencies are in the range of 10,000 Hz and above. The RF used in the 1.5T MRI Scanner is ~64MHz. The RF used in the 3T MRI Scanner is ~128MHz. Specific Absorption Rate (SAR) radio frequency power absorbed per unit of mass (W/kg). Tesla (T) the unit of measure of magnetic field strength. One T is equal to 10,000 Gauss. MRI Transmit/Receive RF Body coil a coil used to transmit and to receive RF energy that encompasses the entire body region within the MR system bore. MRI Transmit/Receive RF Quadrature Head coil a coil used to transmit and to receive RF energy that is constrained to the head region, and configured to use circular polarization (CP). W/kg Watts per kilogram, a measure of the power that is absorbed per kilogram of tissue. English June,
9 Axonics Modulation Technologies, Inc Irvine Center Dr., Ste. 200 Irvine, CA (USA) Tel Fax HealthLink Europe Services BV De Tweeling MC 's-hertogenbosch The Netherlands All Rights Reserved. Copyright Axonics Modulation Technologies, Inc rD English June,
ImageReady MRI Full Body Guidelines for Precision Montage MRI Spinal Cord Stimulator System
ImageReady MRI Full Body Guidelines for Precision Montage MRI Spinal Cord Stimulator System CAUTION: Federal law restricts this device to sale, distribution and use by or on the order of a physician. 91035972-02
More informationAlgovita. MRI Procedure Guidelines. Spinal Cord Stimulation System. 201x
Algovita Spinal Cord Stimulation System MRI Procedure Guidelines Read this manual before performing an MRI scan on a patient implanted with an Algovita Spinal Cord Stimulation System. ONLY 201x Algovita
More informationImageReady MRI Head Only Guidelines for Spectra WaveWriter Spinal Cord Stimulator System
ImageReady MRI Head Only Guidelines for Spectra WaveWriter Spinal Cord Stimulator System CAUTION: Federal law restricts this device to sale, distribution and use by or on the order of a physician. 91171762-01
More informationRead this manual before performing an MRI scan on a patient implanted with an Algovita Spinal Cord Stimulation System.
Algovita Spinal Cord Stimulation System MRI Procedure Guidelines Read this manual before performing an MRI scan on a patient implanted with an Algovita Spinal Cord Stimulation System. ONLY 0300-000175-001
More informationMRI GUIDELINES FOR INSPIRE THERAPY
MRI GUIDELINES FOR INSPIRE THERAPY Clinician s Manual ONLY The following is a trademark of Inspire Medical Systems, Inc.: Inspire Introduction Read the information in this manual prior to conducting an
More informationVNS Therapy System Physician s Manual
VNS Therapy System Physician s Manual Pulse Generator Model 102 Pulse Duo Generator Model 102R Demipulse Generator Model 103 Demipulse Duo Generator Model 104 AspireHC Generator Model 105 AspireSR Generator
More informationMRI GUIDELINES FOR INSPIRE THERAPY
MRI GUIDELINES FOR INSPIRE THERAPY Clinician s Manual The following is a trademark of Inspire Medical Systems, Inc.: Inspire This product and/or the use of this product in a method may be covered by one
More information1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500)
1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500) ONLY NEVRO CORP. All questions or concerns about Nevro products should be forwarded to: Nevro
More information1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500)
1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System (IPG1000 and IPG1500). ONLY NEVRO CORP. All questions or concerns about Nevro products should be forwarded to: Nevro
More information1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System
1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System ONLY 11096 Rev B 1 NEVRO CORP. All questions or concerns about Nevro products should be forwarded to: Nevro Corp.
More informationMRI guidelines for Medtronic neurostimulation systems for chronic pain
MRI guidelines for Medtronic neurostimulation systems for chronic pain See "START HERE" section before conducting MRI. Instructions for use! USA Rx only Explanation of symbols on product or package labeling
More informationMRI Procedure Information. For St. Jude Medical MR Conditional Neurostimulation Systems. Clinician's Manual
MRI Procedure Information For St. Jude Medical MR Conditional Neurostimulation Systems Clinician's Manual Unless otherwise noted, indicates that the name is a trademark of, or licensed to, St. Jude Medical
More information*smith&nephew. MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG. Knee Implants
Knee Implants MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG Knee Implants *smith&nephew Supporting healthcare professionals for over 150 years Summary All knee implants of Smith
More informationMRI Procedure Information. For St. Jude Medical MR Conditional Neurostimulation Systems. Clinician's Manual
MRI Procedure Information For St. Jude Medical MR Conditional Neurostimulation Systems Clinician's Manual Unless otherwise noted, indicates that the name is a trademark of, or licensed to, St. Jude Medical
More information*smith&nephew. MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG. Shoulder Implants
Shoulder Implants MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG Shoulder Implants *smith&nephew Supporting healthcare professionals for over 150 years Summary All shoulder implants
More informationMicra MC1VR01. MRI procedural information for Micra Model MC1VR01 MR Conditional single chamber transcatheter pacing device (VVIR)
Micra MC1VR01 MRI procedural information for Micra Model MC1VR01 MR Conditional single chamber transcatheter pacing device (VVIR) MRI Technical Manual The following list includes trademarks or registered
More information*smith&nephew. MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG. Hip Implants
Hip Implants MRI Safety Information & Parameters for Smith & Nephew Orthopaedics AG Hip Implants *smith&nephew Supporting healthcare professionals for over 150 years Summary All hip implants of Smith &
More informationCochlear Implants. Medical Procedures. for MED EL CI Systems. AW33290_5.0 (English US)
Cochlear Implants Medical Procedures for MED EL CI Systems AW33290_5.0 (English US) This manual provides important instructions and safety information for MED EL CI System users who have to undergo a medical
More informationAdvisa MRI and Ensura MRI SureScan pacing systems
Advisa MRI and Ensura MRI SureScan pacing systems MR Conditional dual chamber and single chamber SureScan pacing systems with SureScan leads MRI Technical Manual The following list includes trademarks
More informationMRI-Ready Systems Manual. MRI Procedure Information for the St. Jude Medical MR Conditional System
MRI-Ready Systems Manual MRI Procedure Information for the St. Jude Medical MR Conditional System CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician. WARNING: This
More informationCochlear Implants. Medical Procedures. for MED EL Implant Systems. AW33290_2.0 (English US)
Cochlear Implants Medical Procedures for MED EL Implant Systems AW33290_2.0 (English US) This manual provides important instructions and safety information for MED EL Implant System users who have to undergo
More informationLow-Profile Wrist Fixator. For stabilization of fractures of the distal radius.
Low-Profile Wrist Fixator. For stabilization of fractures of the distal radius. Technique Guide Part of the External Fixation System Low-Profile Wrist Fixator Indications Intended for stabilization of
More information2014 AAPM 56 th Annual Meeting
2014 AAPM 56 th Annual Meeting SAM Diagnostic Radiology MR Safety - Deep Brain Stimulator and Other Neurostimulators Yunhong Shu, Ph.D. Mayo Clinic, Rochester, MN Outline Background of Neurostimulator
More informationMRI Safety Information for the HiRes Ultra 3D Cochlear Implant
Advanced Bionics AG Laubisrütistrasse 28 8712 Stäfa, Switzerland +41 58 928 78 00 Manufactured by: Advanced Bionics LLC California, U.S.A. +1 661 362 1400 MRI Safety Information for the HiRes Ultra 3D
More informationMRI for Patients with MRI-Conditional Pacing System: Radiographers Role. Lawrance Yip DM, DR, QMH
MRI for Patients with MRI-Conditional Pacing System: Radiographers Role Lawrance Yip DM, DR, QMH MRI & Pacemaker MRI is a crucial and growing imaging modality Plays an important role in clinical management
More informationWarning: Patients should not be exposed to MRI environments until the surgical site following pump implantation is fully healed.
Prometra and Prometra II Programmable Pumps Magnetic Resonance Imaging (MRI) Safety Information GENERAL MR Conditional WARNING: FAILURE TO EMPTY THE PUMP PRIOR TO EXPOSURE TO MRI ENVIRONMENT COULD RESULT
More informationMRI-Ready Systems Manual. MRI Procedure Information for the St. Jude Medical MR Conditional System
MRI-Ready Systems Manual MRI Procedure Information for the St. Jude Medical MR Conditional System Unless otherwise noted, indicates that the name is a trademark of, or licensed to, St. Jude Medical or
More informationExternal Distal Radius Fixator. Supplement to the 8 mm rod fixator system
External Distal Radius Fixator. Supplement to the 8 mm rod fixator system Surgical technique This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation
More informationMedium External Fixator Humeral Shaft Frame. Modular frame for upper extremity use.
Medium External Fixator Humeral Shaft Frame. Modular frame for upper extremity use. Technique Guide Part of the Medium External Fixation System MRI Information Synthes Medium External Fixation devices
More informationMRI Information for Medtronic Spinal and Biologics Devices
M708348B263E Rev C Release Date: 2013-02 MRI Information for Medtronic Spinal and Biologics Devices Not for Distribution in the U.S or its territories. 1 P age Table of Contents MRI Information for Medtronic
More informationMedium External Fixator Pediatric Femoral Shaft Frame. Using medium multi-pin clamps.
Medium External Fixator Pediatric Femoral Shaft Frame. Using medium multi-pin clamps. Technique Guide Part of the Medium External Fixation System MRI Information Synthes Medium External Fixation devices
More informationWallFlex Biliary RX Partially Covered Stent System Prescriptive Information
Caution/Rx Only: Federal Law (USA) restricts this device to sale by or on the order of a physician. Intended Use/Indications for Use The WallFlex Biliary RX is indicated for use in the palliative treatment
More informationMRI SureScan pacemakers & ICDs what you need to know
MRI SureScan pacemakers & ICDs what you need to know Many variables can affect the device patient s likelihood of risks w/ an MRI Type of imaging sequence Type of scanner MRI scan duration Strength of
More informationSafety Issue of Static Magnetic Field (1)
MRI Safety Safety Issue of Static Magnetic Field (1) Currently, typical MRI utilizes static field with strengths of 0.5 to 3 Tesla (FDA guideline is max 4 T for infants < 1 month and 8 T for others) 1
More informationMRI-Ready Systems Manual. MRI Procedure Information for the St. Jude Medical MR Conditional System
MRI-Ready Systems Manual MRI Procedure Information for the St. Jude Medical MR Conditional System Unless otherwise noted, indicates that the name is a trademark of, or licensed to, St. Jude Medical or
More informationMRI Guidelines for VNS Therapy
I Guidelines for VNS Therapy The new I labeling allows patients with an implanted VNS Therapy System to receive fast and high quality scans also with use of Body Coils*. I can be safely performed on patients
More informationUser Manual: Ives MR Conditional Cup Electrodes J.R. Ives, December 22, 2017
User Manual: Ives MR Conditional Cup Electrodes J.R. Ives, December 22, 2017 Ives EEG Solutions; 25 Storey Ave., #118, Newburyport, MA, 01950; ph: 978-358-8006; fax: 978-358-7825; www.iveseegsolutions.com
More informationMagnetic resonance imaging safety of NucleusR 24 cochlear implants at 3.0 T
International Congress Series 1273 (2004) 394 398 www.ics-elsevier.com Magnetic resonance imaging safety of NucleusR 24 cochlear implants at 3.0 T F. Risi*, A. Saldanha, R. Leigh, P. Gibson Cochlear Limited,
More informationVersion überholt. Nicht verwenden. Version obsolète. Ne pas utiliser. Versione obsoleta. Non utilizzare. Verouderde versie. Niet gebruiken.
MRI TECHNICAL GUIDE IMAGEREADY MR CONDITIONAL PACING SYSTEM REF J065, J066, J067, J175, J176, J177, J275, J276, J277, J279, L110, L111, L131, L210, L211, L231, L310, L311, L331, 4456, 4457, 4458, 4459,
More informationSafety of MRI at 1.5 Tesla in Patients with Implanted Sacral Nerve Neurostimulator
european urology 50 (2006) 311 316 available at www.sciencedirect.com journal homepage: www.europeanurology.com Neuro-urology Safety of MRI at 1.5 Tesla in Patients with Implanted Sacral Nerve Neurostimulator
More informationMRI Checklist for MED EL CI and ABI models
MRI Checklist for MED EL CI and ABI models Mi1200 SYNCHRONY Mi1200 SYNCHRONY PIN Mi1210 SYNCHRONY ST...1 Mi1200 SYNCHRONY ABI Mi1200 SYNCHRONY PIN ABI...2 Mi1000 CONCERTO Mi1000 CONCERTO PIN SONATA...3
More informationMRI Exams in Patients with Non MR Conditional Pacemakers and with MR Conditional Pacemakers
MRI Exams in Patients with Non MR Conditional Pacemakers and with MR Conditional Pacemakers Frank G. Shellock, Ph.D., FACC Adjunct Clinical Professor of Radiology and Medicine Keck School of Medicine,
More informationFastFrame External Fixation System. Damage Control Surgical Technique
FastFrame External Fixation System Damage Control Surgical Technique 1 FastFrame External Fixation System Damage Control Surgical Technique Table of Contents Introduction... 2 Indications and Contradictions...
More informationAspireSR. Essential information about your new device
AspireSR Essential information about your new device FAQs for Patients Your doctor recommends that you receive the latest VNS Therapy device. This brochure will give you more information on the differences
More informationSterile-Packaged Large External Fixator Kits. For treatment of long bone and pelvic fractures that require external fixation.
Sterile-Packaged Large External Fixator Kits. For treatment of long bone and pelvic fractures that require external fixation. Large External Fixator Ankle Frame Kit Large External Fixator Trauma Kit Large
More informationGUIDELINE 2-G-004 (formerly APPENDIX 3) HSREB (March 2005) Page 1
HSREB (March 2005) Page 1 Guidelines for Ethics Approval of Research Protocols Involving Human Exposure to Magnetic Resonance Imaging Required Wording in Informed Consent Documentation; The Food & Drug
More informationVNS Therapy System Glossary September 2018
VNS Therapy System Glossary September 2018 1 26-0010-0000/1 (US) ACLS Advanced Cardiac Life Support AE (adverse event) Any symptom, sign, illness, or experience that develops or worsens in severity and/or
More informationBioness Inc Rye Canyon Loop Valencia, CA USA Telephone: or Website: Rx Only Bioness Inc.
Bioness Inc. 25103 Rye Canyon Loop Valencia, CA 91355 USA Telephone: 800.211.9136 or 661.362.4850 Website: www.bioness.com Rx Only 2017 Bioness Inc. StimRouter, Bioness, the Bioness Logo and LiveOn are
More informationButton Plate. Reinforcement for transosseous fixations.
Button Plate. Reinforcement for transosseous fixations. Product Information This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation. Image intensifier
More informationREFILL KIT For use with Prometra Programmable Infusion Systems
REFILL KIT Caution: Federal (USA) Law restricts this device to sale by or on the order of a physician. Table of Contents Contents... 3 Description... 3 Indications... 3 Contraindications... 3 Warnings...
More informationSTANDARD OPERATING PROCEDURE (SOP)
STANDARD OPERATING PROCEDURE (SOP) MRI Scanning for Patients with Cardiac Implantable Electronic Devices (Pacemakers and Implantable Cardiac Defibrillators) Effective date 05/04/2018 Version number V1.0
More informationPrometra Programmable Pump System Brief Summary
Prometra Programmable Pump System Brief Summary Indications, Safety and Warnings Product Instructions for Use and the Infumorph drug labeling must be reviewed prior to use for detailed disclosure. Caution:
More informationTell the doctor who prescribed your MRI scan that you have an implanted Medtronic neurostimulation system.
Before Your I For patients with Medtronic neurostimulation systems for chronic pain The following materials are intended to be used by implant physicians with patients who are to receive Medtronic neurostimulation
More informationUSING YOUR INTELLIS IMPLANTABLE SPINAL CORD STIMULATION SYSTEM
USING YOUR INTELLIS IMPLANTABLE SPINAL CORD STIMULATION SYSTEM This guide provides basic resources and information to help you use your Medtronic Intellis Spinal Cord Stimulation system. Note that the
More informationReplacement Heart Valve Product Description (Stented Tissue) Models Reference
Dear Imaging Center: This letter is in response to your inquiry concerning the safety of performing magnetic resonance (MR) procedures in patients who have been implanted with Edwards Lifesciences LLC
More informationThe Calcaneal Plate. The Synthes non-locking solution for the Calcaneus.
The Calcaneal Plate. The Synthes non-locking solution for the Calcaneus. Surgical Technique This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation.
More informationMRI imaging for patients with cardiac implantable electronic devices (CIEDs)
MRI imaging for patients with cardiac implantable electronic devices (CIEDs) 13 th annual International Winter Arrhythmia School Collingwood, Ontario, Canada February 6, 2016 Andrew C.T. Ha, MD, MSc, FRCPC
More informationLCP Metaphyseal Plates. For extra-articular fractures.
LCP Metaphyseal Plates. For extra-articular fractures. Surgical Technique This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation. Image intensifier
More informationLarge External Fixator Modular Knee Bridge
Using Multi-pin Clamps Large External Fixator Modular Knee Bridge Surgical Technique Large External Fixator Modular Knee Bridge DePuy Synthes Large External Fixation devices are labeled MR Conditional
More informationBefore Your MRI FOR PATIENTS WITH MEDTRONIC NEUROSTIMULATION SYSTEMS FOR CHRONIC PAIN
Before Your I FOR PATIENTS WITH MEDTRONIC NEUROSTIMULATION SYSTEMS FOR CHRONIC PAIN Before Your I Depending on the type of neurostimulation system you have, you may be eligible for either a full-body or
More informationQ & A: How to Safely Scan Patients with a SureScan MRI Pacemaker
Q & A: How to Safely Scan Patients with a SureScan MRI Pacemaker Patient Scheduling and Screening Q: Is there any special (other than regular MRI consent) consent form that should be signed by the patient?
More informationQ-Sense fmri in the MRI environment
Q-Sense fmri in the MRI environment An evaluation at the Erwin L. Hahn Institute for Magnetic Resonance Imaging, and at the Institute of Diagnostic and Interventional Radiology and Neuroradiology, Essen
More informationMRI and CIEDs. Disclosures
Date Clinical Title MRI and CIEDs C. p. Lao Raul Weiss MD, FAHA, FACC, FHRS and CCDS Director, Electrophysiology Fellowship Program Professor of Medicine The Ohio State University Wexner Medical Center
More informationFor Small-Statured Adults and Pediatric Patients. Medium External Fixator Basic Modular Frame
For Small-Statured Adults and Pediatric Patients Medium External Fixator Basic Modular Frame Surgical Technique MRI Information DePuy Synthes Medium External Fixation devices are labeled MR Conditional
More informationIMAGEREADY MR CONDITIONAL PACING SYSTEM
CAUTION: Federal law (USA) restricts this device to sale by or on the order of a physician trained or experienced in device implant and follow-up procedures. MRI TECHNICAL GUIDE IMAGEREADY MR CONDITIONAL
More informationThe VNS Therapy Magnet
The VNS Therapy Magnet A unique benefit of VNS Therapy For patients with epilepsy treated with VNS Therapy What is the Patient Magnet? Use of the magnet is not necessary to receive VNS Therapy. The magnet
More informationSacral Neuromodulation for Dysfunctional Voiders Workshop 1 Monday, August 23, 09:00 10:30
Sacral Neuromodulation for Dysfunctional Voiders Workshop 1 Monday, August 23, 09:00 10:30 Time Time Topic Speaker 09:00 09:05 Introduction to the aims of the workshop and the Magdy Hassouna participants
More informationPocket Adaptor Kit FOR SPINAL CORD STIMULATION. Tip Cards
Pocket Adaptor Kit FOR SPINAL CORD STIMULATION Tip Cards model 74001 (1 4) model 74002 (2 4) The Pocket Adaptor Tip Cards are to be used for educational purposes only. Please refer to the Pocket Adaptor
More informationCardiac Rhythm Therapy // Advanced Features // ProMRI. ProMRI. Checklist and Quick Reference Guide
Cardiac Rhythm Therapy // Advanced Features // ProMRI ProMRI Checklist and Quick Reference Guide PAGE 2 Conditions for MR scans with a BIOTRONIK ProMRI system Usage of an MR scan on a patient having an
More informationBEFORE AND AFTER YOUR MRI SCAN. For patients with St. Jude Medical neurostimulation systems for chronic pain
BEFORE AND AFTER YOUR MRI SCAN For patients with St. Jude Medical neurostimulation systems for chronic pain Preparing for your MRI Depending on what type of neurostimulation system you have, you may be
More informationDISTAL RADIUS. Instructions for Use
DISTAL RADIUS Instructions for Use CAUTION: FEDERAL LAW RESTRICTS THESE DEVICES TO SALE BY OR ON THE ORDER OF A PHYSICIAN. Table of Contents 1. INTENDED USE... 3 2. DEVICE DESCRIPTION... 3 3. METHOD OF
More informationDistal Radius Plate 2.4/2.7 dorsal and volar
Distal Radius Plate 2.4/2.7 dorsal and volar Surgical Technique This publication is not intended for distribution in the USA. Instruments and implants approved by the AO Foundation. Distal Radius Plate
More informationRevision A Date:
Biomet Trauma 56 East Bell Drive P.O. Box 587 Warsaw, Indiana 46581 USA 01-50-4053 Revision A Date: 2016-06 Biomet Unite3D Bridge with OsseoTi Technology ATTENTION OPERATING SURGEON DESCRIPTION Biomet
More informationManufactured for: Bioness Inc Rye Canyon Loop Valencia, CA USA Telephone: (800) Website:
Clinician s Guide Copyright 2016 by Bioness Inc. All Rights Reserved. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system or translated into any language
More informationReplacement Heart Valve Product Description (Stented Tissue) Models Reference
Dear Imaging Center: This letter is in response to your inquiry concerning the safety of performing magnetic resonance (MR) procedures in patients who have been implanted with Edwards Lifesciences LLC
More informationEasyStep. Operative technique
Operative technique EasyStep - Step staple This publication sets forth detailed recommended procedures for using Stryker Osteosynthesis devices and instruments. It offers guidance that you should heed,
More informationLead Management for MRI
Lead Management for MRI Charles J. Love, MD FACC FAHA FHRS CCDS Professor of Medicine Director, Cardiac Rhythm Device Services The Ohio State University Medical Center Columbus, OH USA President, International
More informationCardiac Rhythm Management // Manual // ProMRI. ProMRI. MR conditional device systems
Cardiac Rhythm Management // Manual // ProMRI ProMRI MR conditional device systems Table of Contents Table of Contents Version Comparison............................................ 3 Change History.....................................
More informationYasargil Permanent Aneurysm Clips
Yasargil Permanent Aneurysm Clips Yasargil aneurysm clips Figure captions 1 Yasargil Phynox aneurysm clip, example of a straight clip 2 Yasargil titanium aneurysm clip, example of a straight clip 3 Yasargil
More informationLarge External Fixator Modular Knee Bridge. Using multi-pin clamps.
Large External Fixator Modular Knee Bridge. Using multi-pin clamps. Technique Guide Part of the Large External Fixation System Large External Fixator Modular Knee Bridge Technique Overview Insert Schanz
More informationPatient Essentials. PATIENT S MANUAL for Heart Failure. June 2017
Patient Essentials PATIENT S MANUAL for Heart Failure June 2017 This Patient s Manual is a supplement to the physician s manuals. It is not intended to substitute for advice from your doctor. For a complete
More informationElectrodes (for TransAeris System)
Surgical Procedures for Implantation of TransLoc Electrodes (for TransAeris System) 1 R L 1 TransAeris System TransLoc Electrode Kit 3 3 TransAeris Patient Kit (FrictionLoc) TransAeris Patient Kit (TransAeris
More informationLarge External Fixator Traveling Traction
Used to Correct Angular Deformity Large External Fixator Traveling Traction Surgical Technique Large External Fixator Traveling Traction DePuy Synthes Large External Fixation devices are labeled MR Conditional
More informationExposure to MRI-Related Magnetic Fields and Vertigo in MRI Workers
Utrecht University Institute for Risk Assessment Sciences (IRAS) Exposure to MRI-Related Magnetic Fields and Vertigo in MRI Workers Kristel Schaap, Yvette Christopher-de Vries, Hans Kromhout This project
More informationMemo on The Safety of MRI Scanning of Coronary Artery Stents
Memo on The Safety of MRI Scanning of Coronary Artery Stents Developed by Tobias Gilk, Gilk Radiology Consultants for The problem: Many coronary artery stents have (a) no manufacturer guidance, or (b)
More informationUnderstanding Peripheral
Patient Information Guide Understanding Peripheral Artery Disease Innova Vascular Self-Expanding Stent System Table of Contents Glossary... 2 What is Peripheral Artery Disease?... 4 Treating Peripheral
More informationMedium External Fixator Delta Frame Ankle Bridge
For Small-Statured Adults Medium External Fixator Delta Frame Ankle Bridge Surgical Technique MRI Information DePuy Synthes Medium External Fixation devices are labeled MR Conditional according to the
More informationThe success of any Inter-
Sacral Nerve Neuromodulation (InterStim ) Part II: Review of Programming Helen Rittenmeyer The success of any Inter- Stim Sacral Nerve Neuromodulation program depends on the positioning of the chronic
More informationPATIENT PROGRAMMER 37642
PATIENT PROGRAMMER 37642 Medtronic DBS Therapy user manual Activa PC Model 37601 Activa RC Model 37612 Activa SC Model 37602 Activa SC Model 37603! USA Rx only 2008 Medtronic, Activa, and SoftStart/Stop
More informationREBEL. Platinum Chromium Coronary Stent System. Patient Information Guide
REBEL Patient Information Guide REBEL PATIENT INFORMATION GUIDE You have recently had a REBEL bare metal stent implanted in the coronary arteries of your heart. The following information is important for
More informationAnnual Associate Safety Module Magnetic Resonance Imaging (MRI)
Annual Associate Safety Module Magnetic Resonance Imaging (MRI) MRI Safety and YOU! Rev 12/16 Magnetic Resonance Imaging (MRI) MRI uses a large magnet with a powerful magnetic field, radiofrequencies (RF
More informationPATIENT CONTROLLER ST. JUDE MEDICAL UNDERSTANDING IMPORTANT FEATURES ON YOUR
UNDERSTANDING IMPORTANT FEATURES ON YOUR ST. JUDE MEDICAL PATIENT CONTROLLER THINGS TO KNOW BEFORE AND AFTER AN MRI OR SURGICAL PROCEDURE WHEN YOU HAVE A ST. JUDE MEDICAL PROCLAIM NEUROSTIMULATION SYSTEM
More informationParkinson s disease, Essential Tremor and primary dystonia
Parkinson s disease, Essential Tremor and primary dystonia What is Deep Brain Stimulation and how does it work? Deep Brain Stimulation (DBS) uses one or two surgically implanted medical devices, similar
More informationOBSOLETED. LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation.
LCP Medial Distal Tibia Plate, without Tab. The Low Profile Anatomic Fixation System with Angular Stability and Optimal Screw Orientation. Surgical Technique LCP Small Fragment System This publication
More informationChildren's (Pediatric) MRI for Appendicitis
Scan for mobile link. Children's (Pediatric) MRI for Appendicitis Children's magnetic resonance imaging (MRI) for appendicitis uses a powerful magnetic field, radio waves and a computer to produce detailed
More informationFor the Treatment of Wrist Fractures. Small External Fixator Nonspanning Wrist Frame
For the Treatment of Wrist Fractures Small External Fixator Nonspanning Wrist Frame Surgical Technique Small External Fixator Nonspanning Wrist Frame DePuy Synthes Small External Fixation devices are labeled
More informationPatient Brochure. Clearstream Technologies, Ltd. Moyne Upper Enniscorthy Co. Wexford, Ireland. PK Rev. 0 05/17
Patient Brochure Clearstream Technologies, Ltd. Moyne Upper Enniscorthy Co. Wexford, Ireland PK1411100 Rev. 0 05/17 LIFESTREAM Patient Brochure If you or a member of your family has been diagnosed with
More informationVNS Therapy Guide. For children and those with special needs with drug-resistant epilepsy
VNS Therapy Guide For children and those with special needs with drug-resistant epilepsy This is the story of little Tony, Buddy and Doc. Teur. Little Tony regularly suffered really bad epileptic seizures
More informationCall Medtronic at 1 (800) to verify the patient s current implanted system
MRI SURESCAN SYSTEMS Patient Scanning Process Transvenous Implantable Cardiac Systems PATIENT PRESCREENING SureScan Pacing, Defibrillation, and CRT (CRT-D and CRT-P) Systems Verification Verify that patient
More informationPATIENT CARE AND MRI SAFETY Module 7
PATIENT CARE AND MRI SAFETY Module 7 1 Biological Considerations There are no reported adverse biological effects of extended exposure to MRI. However, several inconsequential and reversible effects of
More information