Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing
|
|
- Morgan Watts
- 5 years ago
- Views:
Transcription
1 Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing Key Points Defibrillators: - know the definition & electrical value of a joule - monophasic vs biphasic types: - differences in output - know typical voltage and current values - advantages of biphasic - AED s - how they differ from full-function defibrillators - cardioversion - what it is, what it is not, and how a defibrillator is used for it - external pacers in defibrillators: - asynchronous and synchronous D. J. McMahon cewood rev local manufacturers on parade; Museum of History & Industry (MOHAI)
2 joule: Unit of energy or work: 1 Newton through a distance of 1 meter or James P Joule 1 amp through 1 ohm for 1 second Mathematically, the joule can be defined both ways, but the convention is to use joules for energy, and newton-meters for torque. One joule = one watt-second = ~ 0.24 calorie = ~ 0.74 foot-pound Essential monophasic defibrillator with switch A in this position, defibrillation is happening, (not charging) Coil The Lown waveform: Monophasic, damped
3 Basic Lown waveform creation Specific Lown waveform creation Energy in the capacitor : U = ½ CV p 2 Biphasic defibrillation: > 1990 s improvement in defib circuitry. > Achieves defibrillation in fewer attempts, using lower energies in each attempt. > Less time delay in salvaging a cardiac crisis, and less damage to the heart. > became biphasic truncated exponential (BTE) waveform (Physio Control and Philips)
4 Current & Time at Various Patient Impedances Insulated Gate Bipolar Transistor (IGBT): - Combines an isolated gate FET for the input, and a bipolar power transistor as a switch, in a single device. - Has the gate-drive characteristics of a MOSFET and the high-current and low-saturation-voltage capability of a bipolar transistor. - Used in medium- to high-power applications. Test loads are 50Ω
5 Some manufacturers of biphasic defibs provide a maximum energy of 200 joules, while many provide a maximum of 360 joules.
6 Pads have replaced paddles in most applications of defibrillators. Pads assure a better contact with the patient s skin, so resistive losses are minimized. They are also safer for the user. Paddles, internal as well as external, are still used in specific clinical areas (O.R., PACU, etc.) Defib voltages are very brief, but very high and potentially lethal! paddles pads spoons for OR Correct placement of the pads or paddles is important
7 Internal paddles for use in open-heart surgery: 2010 AHA Guidelines for pediatric defibrillation: With a manual defibrillator use a dose of 2 J/kg for the first attempt, and 4 J/kg for subsequent attempts. Placed directly on the heart to re-start the beat. Defibrillator limits energy to them at 50 joules. ECG Input Protection from Defibrillator Pulses: NE2H lamps
8 Automatic External Defibrillators (AED s) > Semi-automated defibs that analyze cardiac status and deliver therapeutic shock as indicated, using a pre-set algorithm and voice commands. > Intended for non-clinical settings, for use by non-clinicians. > Typical biomed maintenance consists of a functional check and a battery check. Conductive gel left on paddles from a previous use. Typical block diagram for an AED
9 Defibs for Cardioversion (this is not defibrillation!) > The use of a small energy pulse from a defibrillator to convert a non-lethal arrhythmia (e.g. atrial flutter) to a normal sinus rhythm > Delivers a pulse about 30ms after the peak of the R-wave, synchronized by the defibrillator > Modern units can discriminate the R-wave from the T-wave or noise spikes Timing of cardioversion pulse Defibs as External Pacemakers Defibrillators can be used as temporary external pacemakers for patients with bradycardia or other issues. Pacing is either synchronous or asynchronous : In Synchronous pacing ( Demand mode): pacer fires only when no complex is sensed for a predetermined amount of time. Do NOT cardiovert across the T-wave! In Asynchronous pacing: pacer fires at a fixed rate at preset intervals, independent of cardiac activity. ( not synchronized means the pacemaker leads and the heart follows)
10 Typical external pacemakers deliver a 40-millisecond impulse for each beat. The average current necessary for external pacing is from ma, depending on the cardiac status. In pacing, the same defibrillator electrodes are used.
11 Major Makers of full-function Defibrillators and AEDs: October, 2014 Test equipment: Defibrillator Analyzers BC Biomedical Fluke Biomedical
Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev
Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing D. J. McMahon 141001 cewood rev 2017-10-04 Key Points Defibrillators: - know the definition & electrical value
More informationDEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION
1 DEFIBRILLATORS The two atria contract together and pump blood through the valves into the two ventricles, when the action potentials spread rapidly across the atria surface. After a critical time delay,
More informationM Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use
DEFIBRILLATOR OPTION General Information Introduction M Series products are available with an advanced electrical design that provides a unique rectilinear biphasic waveform for defibrillation and cardioversion.
More informationDEFIBRILLATORS. Prof. Yasser Mostafa Kadah
DEFIBRILLATORS Prof. Yasser Mostafa Kadah Basics Defibrillation is definitive treatment for life-threatening cardiac arrhythmias such as ventricular fibrillation Defibrillation consists of delivering therapeutic
More informationOBJECTIVE. 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation.
Defibrillators OBJECTIVE 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation. 4. Types and classes of defibrillator 5. Describe
More informationImplantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević
Implantable Cardioverter-Defibril Defibrillators Ratko Magjarević University of Zagreb Faculty of Electrical Engineering and Computing Croatia ratko.magjarevic@fer.hr Ventricular Fibrillation Ventricular
More informationBiphasic Clinical Summaries
Biphasic Clinical Summaries Defibrillation of Ventricular Fibrillation and Ventricular Tachycardia Background Physio-Control conducted a multi-centered, prospective, randomized and blinded clinical trial
More informationCardioversion / Defibrillation / Transcutaneous Pacer Helpful Hints [ZOLL] 2018
Cardioversion / Defibrillation / Transcutaneous Pacer Helpful Hints [ZOLL] 08 4 Universal cable may be inserted to the side Defib Short Block to perform 0 J check Universal Cable must be connected to APEX
More informationEarly Defibrillation. Dr. M. Ravishankar
Early Defibrillation Dr. M. Ravishankar THE CHAIN OF SURVIVAL CONCEPT Early Access Early CPR Early Defibrillation Early Advance Care 1 st Link 2 nd Link 3 rd Link 4 th Link Why early defibrillation VF
More informationDefibrillator. BiomedGuy
Defibrillator BiomedGuy Medtronic Physiocontrol LifePak 10 Introduction This life-support system is used by paramedic, hospital staff, and other trained authorized healthcare providers. Provides, ECG,
More informationThe heart's "natural" pacemaker is called the sinoatrial (SA) node or sinus node.
PACEMAKER Natural pacemaker: The heart's "natural" pacemaker is called the sinoatrial (SA) node or sinus node. Artificial pacemaker: It is a small, battery-operated device that helps the heart beat in
More informationThe Automated Defibrillator: A Biomedical Engineering Success Story. Dr. James A. Smith
The Automated Defibrillator: A Biomedical Engineering Success Story Dr. James A. Smith A Aorta VCS Superior Vena Cava RA Right Atrium RV Right Ventricle LV Left Ventricle The Heart Beating Heart: Video
More informationPediatric pacemakers & ICDs:
Pediatric pacemakers & ICDs: perioperative management Manchula Navaratnam Clinical Assistant Professor LPCH, Stanford SPA 2016 Conflict of interest: none Objectives Indications in pediatrics Components
More information1. Types of Cardiac Arrhythmia
Defibrillators Structure 1. Types of Cardiac Arrhythmia 2. Modes and Waves 3. Evolution 4. external Defibrillator I. Structure & Function II. Sequence of Operation 5. Internal Defibrillator I. Short-Term
More informationCardiac Electrical Therapies. By Omar AL-Rawajfah, PhD, RN
Cardiac Electrical Therapies By Omar AL-Rawajfah, PhD, RN Outlines What are cardiac electrical therapies Ablation Defibrillation Cardioversion What are the nursing considerations for each type of therapy
More informationQUICK SETUP GUIDE SECULIFE DFBASE
QUICK SETUP GUIDE SECULIFE DFBASE SECULIFE DFBASE is a microprocessor-based instrument that is used in the testing of defibrillators. It measures the energy output and provides information about the defibrillator
More informationWhite Paper. High performance resistors are key to meeting the demanding requirements of portable medical electronics designs. Issued in June 2014
White Paper High performance resistors are key to meeting the demanding requirements of portable medical electronics designs Issued in June 2014 The contents of this White Paper are protected by copyright
More informationThe Facts about Biphasic Defibrillation
The Facts about Biphasic Defibrillation Introduction In reference to the SMART Biphasic waveform, Guidelines 2000 for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care, published by the American
More informationPresented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO
Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO Follow me on Twitter! CEE Med Updates@BarbaraFurryRN Like me on Facebook! What is a
More informationThe Nuts and Bolts of ICD Therapy
Electrical Management of Cardiac Rhythm Disorders For Cardiology Fellows December 5-8 Austin, Texas The Nuts and Bolts of ICD Therapy 1 2 Action Potential Localized Differences in Conduction Conduction
More informationRN-BC, MS, CCRN, FAHA
Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO Follow me on Twitter! CEE Med Updates@BarbaraFurryRN Like me on Facebook! 1 A. Atropine
More informationEnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds ERI 10.7 years**** 35cc, 68g Programmable Active Can
EnTrust D154ATG Dual Chamber ICD 35J delivered 8 seconds BOL, 11 seconds ERI 7.7 years** 35cc***, 68g Programmable Active Can EnTrust D154VRC Single Chamber ICD 35J delivered 8 seconds BOL, 11.8 seconds
More informationCardioverting with Confidence! Class Code 142
Cardioverting with Confidence! Class Code 142 Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO Follow me on Twitter! CEE Med Updates@BarbaraFurryRN
More informationCardioverting with Confidence: Isn t It About Time!
Cardioverting with Confidence: Isn t It About Time! Barbara Furry RN-BC, MS, CCRN, FAHA Cardioverting with Style! 1 Definition of Cardioversion Cardioversion is an electrical method to restore a rapid
More informationPuzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC
Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Pacemaker: An electric device implanted in the body to regulate the heart beat. Delivers electrical stimuli over leads with electrodes in contact
More informationFlowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields
Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields Identify patient with an ICD at POAC. Notify Cardiac Physiologist that patient is due to have surgery and
More informationMARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa
MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa INTENSIVE CARE POLICY & PROCEDURES Policy Number: R-25 Subject: Policy: Purpose: Cardioversion - Synchronized An RN may perform a synchronized
More informationElectrical Interventions
3 Electrical Interventions Quick Contents Rationale for Electricity - p. 74 Paddle Placement - p. 77 Defibrillation - p. 82 Using an AED- p. 85 Manual Defibrillation - p. 86 Cardioversion - p. 87 Transcutaneous
More informationModel 5392 EPG Temporary Pacer
Model 5392 EPG Temporary Pacer Compatible Components Reference Card 5392 Surgical Cables 5487 Disposable, short 5487L Disposable, long 5832S Reusable, small clip 5833S 5833SL Disposable, small clip, short
More information1-Epinephrine 2-Atropine 3-Amiodarone 4-Lidocaine 5-Magnesium
١ 1-Epinephrine 2-Atropine 3-Amiodarone 4-Lidocaine 5-Magnesium ٢ When VF/pulseless VT cardiac arrest is associated with torsades de pointes, providers may administer magnesium sulfate at a dose of 1-2
More informationPanorama. Arrhythmia Analysis Frequently Asked Questions
Panorama Arrhythmia Analysis Frequently Asked Questions What ECG vectors are used for Beat Detection? 3-wire lead set 5-wire lead set and 12 lead What ECG vectors are used for Beat Typing? 3-wire lead
More informationTranscutaneous Pacing on Zoll Defibrillator Version 4 Related Documents SOP-Zoll defibrillator ; Phillips MRX SOP Dr Randal McRoberts
Emergency Medical Retrieval Service (EMRS) www.emrs.scot.nhs.uk Standard Operating Procedure Public Distribution Title Transcutaneous Pacing on Zoll Defibrillator Version 4 Related Documents SOP-Zoll defibrillator
More informationICD: Basics, Programming and Trouble-shooting
ICD: Basics, Programming and Trouble-shooting Amir AbdelWahab, MD Electrophysiology and Pacing Service Cardiology Department Cairo University Feb 2013 Evolution of ICD Technology ICD Evolution Indications
More informationSMART Biphasic. A Proven, Fixed, Low-Energy Defibrillation Waveform
SMART Biphasic A Proven, Fixed, Low-Energy Defibrillation Waveform INTRODUCTION The 1990s heralded a new era of transthoracic defibrillation in which the rules of conventional practice no longer apply.
More informationBiomedical Instrumentation
Biomedical Instrumentation Prof. Dr. Nizamettin AYDIN naydin@yildiz.edu.tr naydin@ieee.org http://www.yildiz.edu.tr/~naydin Therapeutic and Prosthetic Devices 1 Figure 13.1 Block diagram of an asynchronous
More informationAutomated External Defibrillation Principle of Early Defibrillation States that all BLS personnel be trained, equipped and allowed to operate a if
1 2 3 4 5 6 Automated External Defibrillation Principle of Early Defibrillation States that all BLS personnel be trained, equipped and allowed to operate a if they are expected to respond to persons in
More informationTherapeutic/Prosthetic Devices Pacemakers & Defibrillators. Biomedical Instrumentation JHU Applied Physics Lab. Dr. Nitish V. Thakor.
Lecture 8 Therapeutic/Prosthetic Devices Pacemakers & Defibrillators Dr. Nitish V. Thakor Biomedical Instrumentation JHU Applied Physics Lab Introduction Major use of medical electronics is as a diagnostic
More informationDefibrillation. Learning outcomes. Introduction. Mechanism of defibrillation. Factors affecting defibrillation. success. Transthoracic impedance
Defibrillation CHAPTER 9 Learning outcomes To understand: The mechanism of defibrillation The factors affecting defibrillation success The importance of minimising interruptions to chest compressions during
More informationINTERNAL CARDIOVERSION. Lancashire & South Cumbria Cardiac Network
INTERNAL CARDIOVERSION Lancashire & South Cumbria Cardiac Network DC Cardioversion AF & AFL safe 1,2 efficacious 3,4 (60-94%) 5 SR - Increases exercise tolerance 6 Maintainence SR unlikely in patients
More informationPacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report
Case Report Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report Taku Nishida MD, Tamio Nakajima MD, PhD, Yutaka Goryo MD, Ken-ichi Ishigami MD, PhD, Hiroyuki Kawata
More informationPower to see the big picture
Power to see the big picture Philips HeartStart MRx Monitor/Defibrillator Product information The capabilities you need and the performance HeartStart MRx unites Philips industry-leading monitoring with
More informationHeartStart. XL+ cardiac resuscitation. Ready. to save lives. Not available for sale in the U.S.
HeartStart XL+ cardiac resuscitation Ready to save lives Not available for sale in the U.S. Ready to respond, revive, improve Hospital cardiac emergencies are typically stressful and chaotic events. The
More informationEssentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH
Essentials of Pacemakers and ICD s Rajesh Banker, MD, MPH Pacemakers have 4 basic functions: Stimulate cardiac depolarization Sense intrinsic cardiac function Respond to increased metabolic demand by providing
More informationPacemaker Concepts and Terminology*
Pacemaker Concepts and Terminology* BAROUH V BERKOVITS, E.E., Ing. Associate in Surgery, Harvard Medical School, Boston, Mass., Associate in Electrophysiology, Miami University, Miami, Fla., Senior Research
More informationElectrocardiography for Healthcare Professionals
Electrocardiography for Healthcare Professionals Kathryn A. Booth Thomas O Brien Chapter 10: Pacemaker Rhythms and Bundle Branch Block Learning Outcomes 10.1 Describe the various pacemaker rhythms. 10.2
More informationAdvanced ICD Concepts
1 4 2 5 7 3 6 8 Advanced ICD Concepts This presentation is provided with the understanding that the slide content must not be altered in any manner as the content is subject to FDA regulations. This presentation
More informationPeri-operative management of pacemakers and implantable cardiac defibrillators
Guideline Title: Peri-operative management of patients fitted with Permanent Pacemakers (PPMs) and Implantable Cardioverter Defibrillators (ICDs). Author(s): Jessica Osman (Chief Pacing Physiologist),
More informationFigure 2. Normal ECG tracing. Table 1.
Figure 2. Normal ECG tracing that navigates through the left ventricle. Following these bundle branches the impulse finally passes to the terminal points called Purkinje fibers. These Purkinje fibers are
More informationPACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I
1 PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I Cynthia Webner DNP, RN, CCNS, CCRN-CMC Karen Marzlin DNP, RN, CCNS, CCRN-CMC 2 PROFESSIONAL NURSING PRACTICE CAN ONLY ADVANCE AS MUCH AS INDIVIDUAL
More informationTesting the Accuracy of ECG Captured by Cronovo through Comparison of ECG Recording to a Standard 12-Lead ECG Recording Device
Testing the Accuracy of ECG Captured by through Comparison of ECG Recording to a Standard 12-Lead ECG Recording Device Data Analysis a) R-wave Comparison: The mean and standard deviation of R-wave amplitudes
More informationPacing Lead Implant Testing. Document Identifier
Pacing Lead Implant Testing 1 Objectives Upon completion of this presentation, the participant should be able to: Name the two primary surgical options for implanting pacing leads Describe three significant
More informationA Study of Automated External Defibrillator with the Help of Monophasic and Biphasic Pulse
A Study of Automated External Defibrillator with the Help of Monophasic and Biphasic Pulse K. Malathy 1, S. Dhaarani 2, N. Bairavi 3, V. Niranjhana 4 Assistant Professor, Dept of ECE, Dr.SJS Paul Memorial
More informationReview of Pacemakers and ICD Therapy: Overview and Patient Management
Review of Pacemakers and ICD Therapy: Overview and Patient Management Pacing Systems Charles J. Love, MD FACC FAHA FHRS CCDS Professor of Medicine Director, Cardiac Rhythm Device Services OSU Division
More informationCardiac Pacing. Learning outcomes. Introduction. The cardiac impulse - its formation and its failure CHAPTER. To understand:
Cardiac Pacing CHAPTER 10 Learning outcomes To understand: The indications for cardiac pacing in the peri-arrest setting How to perform percussion pacing How to apply non-invasive, transcutaneous electrical
More informationPerioperative Management of Cardiac Implantable Devices
Financial Disclosures Breandan Sullivan MD Assistant Professor Co Director Cardiothoracic and Vascular Surgery ICU Department of Anesthesiology and Critical Care Medicine University of Colorado None Perioperative
More informationEE 4BD4 Lecture 20. Therapeutic Stimulation
EE 4BD4 Lecture 20 Therapeutic Stimulation 1 2 Extracellular Stimulation (at cathode) 3 4 Design of FES (cont.): Example stimulus waveform shapes: monophasic, biphasic, chopped, triphasic, and asymmetric,
More informationTranscutaneous Pacing. Approval: Medical Director James Stubblefield, MD. Approval: EMS Director Michael Petrie
Transcutaneous Pacing 6091D Low Frequency/High Risk: Transcutaneous Pacing Applies To: Paramedics Effective Approval: Medical Director James Stubblefield, MD Approval: EMS Director Michael Petrie Expires
More informationProject Title Temporary Pacemaker Training Simulator
Project Title Temporary Pacemaker Training Simulator Project Description Problem: There is no available training device for temporary pacemakers (pacemakers). A training device will have to essentially
More informationImproving quality of resuscitation. Defibrillators TEC-5600 series
Improving quality of resuscitation Defibrillators TEC-5600 series Transitioning back to Life Improving quality of resuscitation Saving Life Shock Efficiency on Demand Fast shocks and continuous observation
More informationEHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs
EHRA Accreditation Exam - Sample MCQs Cardiac Pacing and ICDs Dear EHRA Member, Dear Colleague, As you know, the EHRA Accreditation Process is becoming increasingly recognised as an important step for
More informationMEDTRONIC CARELINK NETWORK FOR PACEMAKERS. Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring
MEDTRONIC CARELINK NETWORK FOR PACEMAKERS Comparison between the Medtronic CareLink Network for Pacemakers and Transtelephonic Monitoring Transtelephonic Monitoring Transmission What can you determine
More informationPERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS
PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS DR SUSAN CORCORAN CARDIOLOGIST ONCE UPON A TIME.. Single chamber pacemakers Programmed at 70/min VVI 70 UNIPOLAR SYSTEMS A Unipolar Pacing
More informationDEFIB «U S E-HEART. CardioStart MEDICAL MEDIO TECHNOLOGIES BIPHASIC DEFIBRILLATORS К-ИЕЛКГ. Wmwie
«U S DEFIB Т^П MEDICAL MEDIO TECHNOLOGIES BIPHASIC DEFIBRILLATORS i E-HEART CardioStart К-ИЕЛКГ Wmwie E-HEART CardioStart The lightest! ONLY 8,82 lbs 4Kg With the innovative flip down display. Multiple
More informationCPR Guidelines - Update
CPR Guidelines - Update This Guidelines Highlights publication summarizes the key issues and changes in the 2010 American Heart Association (AHA) Guidelines for Cardiopulmonary Resuscitation (CPR) and
More informationCardiac Resuscitation. HeartStart FRx. Lead the way. to save a life Philips HeartStart FRx defibrillator with Life Guidance
Cardiac Resuscitation HeartStart FRx Lead the way to save a life Philips HeartStart FRx defibrillator with Life Guidance For those who get there first With access to the right equipment and support, everyone
More informationVENTRICULAR DEFIBRILLATOR
VENTRICULAR DEFIBRILLATOR Group No: B03 Ritesh Agarwal (06004037) ritesh_agarwal@iitb.ac.in Sanket Kabra (06007017) sanketkabra@iitb.ac.in Prateek Mittal (06007021) prateekm@iitb.ac.in Supervisor: Prof.
More informationMARS Ambulatory ECG Analysis The power to assess and predict
GE Healthcare MARS Ambulatory ECG Analysis The power to assess and predict Connecting hearts and minds Prevention starts with knowledge Around the world, heart disease is one of our fastest-growing health
More informationBeneHeart D3. Defibrillator/Monitor. More than a fast defibrillator
BeneHeart D3 Defibrillator/Monitor More than a fast defibrillator Monitoring 4-in-1 design and powerful function for full spectrum applications 4-in-1 design and powerful function for full spectrum applications
More informationNewer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity.
Pacemakers & Defibrillators A pacemaker system consists of a battery, a computerized generator and wires with sensors called electrodes on one end. The battery powers the generator, and both are surrounded
More informationM3535A/M3536A Instructions for Use. HeartStart MRx
M3535A/M3536A Instructions for Use HeartStart MRx 1Table of Contents 1 Introduction 3 Overview 3 Intended Use 5 Defibrillation Therapy 5 Synchronized Cardioversion Therapy 5 Noninvasive Pacing Therapy
More informationSVT Discriminators. Definition of SVT Discrimination. Identify which patient populations might benefit from these features
Definition of SVT Discrimination Identify which patient populations might benefit from these features Understand the 4 types of SVT discriminators used by St Jude Medical Be aware of programmable parameters
More informationDefibrillators TEC-8300K series. Improving quality of resuscitation
Defibrillators TEC-8300K series Improving quality of resuscitation Transitioning back to life For intensive treatment and advanced life support Pre-hospital Out-of-hospital 12-lead ECG acquisition and
More informationShock-induced termination of cardiac arrhythmias
Shock-induced termination of cardiac arrhythmias Group members: Baltazar Chavez-Diaz, Chen Jiang, Sarah Schwenck, Weide Wang, and Jinglei Zhang Abstract: Cardiac arrhythmias occur when blood flow to the
More informationCardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc.
Cardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc. The Miracle of Living February 21, 2018 Matthew Ostrom MD,FACC,FHRS Division of
More informationBEDSIDE ECG INTERPRETATION
BEDSIDE ECG INTERPRETATION Presented by: Ryan Dean, RN, MSN, CCRN, CCNS, CFRN Flight Nurse 2017 Based on presentations originally by Gennifer DePaoli, RN Objectives Hospital policies Electrical conduction
More informationCardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up
Cardiac Rhythm Device Management PBL STOP Your acronym for a standardized follow-up What s in it for you? What do you need to feel comfortable with a pacemaker/icd follow-up? 2 CORE OBJECTIVE Provide a
More informationChain of Survival. Highlights of 2010 American Heart Guidelines CPR
Highlights of 2010 American Heart Guidelines CPR Compressions rate of at least 100/min. allow for complete chest recoil Adult CPR depth of at least 2 inches Child/Infant CPR depth of 1/3 anterior/posterior
More informationAbout atrial fibrillation (AFib) Atrial Fibrillation (AFib) What is AFib? What s the danger? Who gets AFib?
Understanding AFib Atrial Fibrillation (AFib) About AFib 3 How Your Heart Works 4 Types of AFib 5 Symptoms 5 Risk Factors 5 How is AFib Diagnosed? 6 Treatment 6 What to Ask Your Doctor 7 A normal heartbeat
More informationAWAY Model # RTLAGF-900 REV
AWAY Model # RTLAGF-900 REV2.4.9.15 Contents User manual English 2 General Description 2 Intended Use 3 Included in this Package 4 Instructions for Use 7 Application Duration 7 Electrode Placement 8 Tips
More informationDefibrillator Monitor E-HEART. User Manual
User Manual Defibrillator Monitor E-HEART Defibrillator Monitor E-HEART Manufacturer: US Defib Medical Technologies LLC Address : 7831 NW 72 nd Avenue, Medley - Miami Zip Code: 33166 Phone: +1 305 8877552
More informationProSim 3 and 2. Vital Signs Simulators
ProSim 3 and 2 Vital Signs Simulators Technical Data Don t need a comprehensive patient monitor tester? The 6-in-1 ProSim 3 and 4-in-1 ProSim 2 Vital Signs Simulators are clear choices for biomedical engineers
More informationIndications The AED mode is to be used only on patients in cardiopulmonary arrest.
14 October 2008 TO: Dr. Copass FR: Brian W. Chinn B3 A2 1018 RE: SFD LP12 Guidelines: I. AED / Advisory II. Manual A. Defibrillation B. Synchronize Cardioversion C. Noninvasive Pacing 1. Demand 2. Non-Demand
More informationSTUDY REPORT ON THE EFFECT OF RADIO WAVES ON MEDICAL DEVICES
STUDY REPORT ON THE EFFECT OF RADIO WAVES ON MEDICAL DEVICES March, 2007 Ministry of Internal Affairs and Communications (MIC) Introduction We are moving toward the creation of a ubiquitous network society
More informationDiagnostic capabilities of the implantable therapeutic systems
Cardiac pacing 2012 and beyound Monday August 27, 2012 Diagnostic capabilities of the implantable therapeutic systems Pekka Raatikainen Heart Center Co. Tampere University Hospital and University of Tampere
More informationMichigan Pediatric Cardiac Protocols. Date: November 15, 2012 Page 1 of 1 TABLE OF CONTENTS
Date: November 15, 2012 Page 1 of 1 TABLE OF CONTENTS Pediatric Asystole Section 4-1 Pediatric Bradycardia Section 4-2 Pediatric Cardiac Arrest General Section 4-3 Pediatric Narrow Complex Tachycardia
More informationPacemaker Simulator ST2361 Learning Material Ver.1.1 An ISO 9001:2008 company Save paper, save trees, save earth Dear User, We request you to use the Learning material in the CD form provided with this
More information09/05/2018. True/False: All pacemakers have defibrillation capabilities.
Paul G. Tarasi M.D. Staff Anesthesiologist Allegheny Health Network Adjunct Clinical Assistant Professor of Anesthesiology Temple University School of Medicine Describe different types of implanted cardiac
More informationSARASOTA MEMORIAL HOSPITAL
SARASOTA MEMORIAL HOSPITAL TITLE: NURSING PROCEDURE PACEMAKERS AND DETERMINING PACEMAKER DATE: REVIEWED: PAGES: 10/99 11/18 1 of 7 PS1094 ISSUED FOR: Nursing RESPONSIBILITY: *Qualified RN PURPOSE: To establish
More informationDysrhythmias Chapter
Dysrhythmias Chapter 36 1.5.17 Copyright 2014 by Mosby, an imprint of Elsevier Inc. Copyright 2014 by Mosby, an imprint of Elsevier Inc. Properties of Cardiac Cells Automaticity Excitability Conductivity
More informationProSim 3 and 2 Vital Signs Simulators
ProSim 3 and 2 Vital Signs Simulators Technical Data Don t need a comprehensive patient monitor tester? The 6-in-1 ProSim 3 and 4-in-1 ProSim 2 Vital Signs Simulators are clear choices for biomedical engineers
More informationMODEL EPG SINGLE CHAMBER TEMPORARY EXTERNAL PACEMAKER
MODEL 53401 EPG SINGLE CHAMBER TEMPORARY EXTERNAL PACEMAKER In-service Guide 53401 TRAINING LEARNING OBJECTIVES Identify the control knobs and be able to explain how to use them to program device. Distinguish
More informationRigel UNI-SiM Lite. The most cost-effective patient simulator on the market. Key Features n Compact and cost-effective
Rigel UNI-SiM Lite The most cost-effective patient simulator on the market The UNI-SiM Lite is a handheld and battery-operated vital signs simulator, designed to be a cost effective and portable solution
More informationACLS Prep. Preparation is key to a successful ACLS experience. Please complete the ACLS Pretest and Please complete this ACLS Prep.
November, 2013 ACLS Prep Preparation is key to a successful ACLS experience. Please complete the ACLS Pretest and Please complete this ACLS Prep. ACLS Prep Preparation is key to a successful ACLS experience.
More informationReviewed Date: Supersedes (Policy #/description & date): Prepared by: Jacque Callis, RN Approved by: D.R. Johnson, MD
External Pacemaker Set-up and Revised Date: 11/13/2014 Page 1 of 5 POLICY: The transcutaneous external pacemaker will be used as a noninvasive method of delivering electrical stimuli to the myocardium
More informationEngineering involved in Cardiac Arrest Management Sahar Sorek
Engineering involved in Cardiac Arrest Management Sahar Sorek The University of Rhode Island College of Engineering BME 281-0001 Fascinating Facts: 424,000/year out of hospital CA s. More than half of
More informationNon-Invasive Transcutaneous Pacing
Non-Invasive Transcutaneous Pacing Application Note Introduction This Application Note provides an overview of non-invasive transcutaneous pacing. It includes a brief history of pacing, describes the electrical
More informationRigel UNI-SiM. The world s smallest integrated NiBP, SpO2 and patient simulator. Key Features n Compact and cost-effective
Rigel UNI-SiM The world s smallest integrated NiBP, SpO2 and patient simulator. The UNI-SiM is a handheld and battery-operated vital signs simulator capable of undertaking six synchronized vital signs
More informationSupplemental Material
Supplemental Material 1 Table S1. Codes for Patient Selection Cohort Codes Primary PM CPT: 33206 or 33207 or 33208 (without 33225) ICD-9 proc: 37.81, 37.82, 37.83 Primary ICD Replacement PM Replacement