1 Emergency Medical Retrieval Service (EMRS) Standard Operating Procedure Public Distribution Title Transcutaneous Pacing on Zoll Defibrillator Version 4 Related Documents SOP-Zoll defibrillator ; Phillips MRX SOP Author Dr Randal McRoberts Reviewer Dr Stephen Hearns THE PHILLIPS MRX DEVICE (DEFAULT MONITOR FOR EMRS) HAS A SEPARATE SOP WHICH INCLUDES A PACING SECTION. THIS SOP PURELY REFERS TO THE ZOLL DEVICE. Aims This document discusses the transcutaneous pacing facilities using the Zoll defibrillator currently in use with EMRS Background Patients in complete heart block may require temporary pacing transcutaneously prior to undergoing formal permanent pacemaker (PPM) insertion at a definitive care centre. Unless a functional transvenous pacing wire has been inserted at the referring site, the patient may require transcutaneous pacing (TCP) by EMRS during transport to definitive care. The E series Zoll (manual) defibrillator/pacing devices currently used as a backup monitor by EMRS have been cleared for use on Military aircraft and SAS-A & SAS-B helicopters. Application EMRS Team Members SAS Paramedics SOP- Transcutaneous Pacing 1

2 Patients appropriate for retrieval team activation Patients in complete heart block or those patients whose bradycardia demands measures beyond that of drug intervention to achieve adequate cardiac output. Advice to GP/Hospital referrer prior to team arrival: Arrange for transvenous pacing if equipment & expertise available Support heart rate and blood pressure according to the current ALS algorithm for bradycardia Continuous ECG/SpO2/Heart Rate/BP monitoring 12 lead ECG & repeat at 30min intervals Consider brady-arrhythmia may be part of ACS and treat accordingly Look for electrolyte abnormalities and treat accordingly If transcutaneous pacing is available and is required, institute its use with sedation for patient comfort as required Medical management on scene The patient should be fully assessed on arrival. The treatment instituted to this point by the referring centre should be noted. 1. Resuscitate as necessary according to ALS guidelines 2. If the patient has had transvenous or transcutaneous pacing, the following should be checked: The effectiveness of this intervention i.e. that the pacing is achieving both electrical and mechanical capture and that the patient s clinical condition reflects this 12 lead ECG Any bloods/investigations if done already (eg: Troponin; U&E) If transvenous: The position of the pacing wire on the CXR. Ensure no immediate complication of the procedure (eg: Pneumothorax) If transcutaneous: The condition/length of use of the pads (replace if have been used for 8hours of continuous pacing) Position of pads Front & back positioning of pads is optimal for pacing Patient comfort check for need for sedation/analgesia to achieve effective pacing SOP- Transcutaneous Pacing 2

3 Having decided clinically on the need for transcutaneous pacing, the following steps should be taken when setting up the Zoll E series defibrillator/pacer device: a) Dry and prepare patient s chest for application of Stat-Padz AND 3 (or 5) Lead ECG. Both are needed to pace in either Demand or Standby mode. On the rare occasion that asynchronous pacing is used, the ECG will not be applied. b) Explain to the patient that it feels uncomfortable during the pacing period c) Check stat padz multi-function electrodes (MFE) expiry date d) Connect the MFE cable to the Zoll device multi-function cable (white opaque connector on MFE to red connector on Zoll) e) Apply the MFE to the patient as per the instructions on the packaging. The pads should be placed in the anterior-posterior orientation for pacing. If AP positioning not possible then consider apex-sternum positioning (less effective for pacing). Ensure adhesion to skin. f) If the MFE pads are not making good contact with the patient, the messages CHECK PADS and POOR PAD CONTACT are alternately displayed and the energy will not be delivered. g) If a short circuit exists between the MFE pads, the message DEFIB PAD SHORT is displayed. h) The following steps can be applied to optimise the set-up of the device for pacing: SOP- Transcutaneous Pacing 3

4 Step 1 The Pacer (Green) mode is accessed by turning the Selector Switch counterclockwise Milliamps (ma) are the unit of pacing energy No AED capability or ANALYZE button can be used in pacing mode Step 3 Turn the Pacer Output dial to adjust the level of discharged milliamps. If capture is achieved, the ECG will have a wide complex reflecting ventricular contraction following the thin Pacing spike marker Step 2 Pacer markers (PM) indicate the rate set to attempt to capture the ventricle. Default settings of 70 Paces Per Minute (PPM) and 0 ma are displayed upon access of Pacer Mode To increase or decrease pacer marker rate, (PM) turn the Pacer Rate Dial i) Once electrical and mechanical capture has been confirmed, dial the ma up 10% from capture threshold as a safety margin The 4:1 button may be pushed at anytime to allow for review of the underlying rhythm. The Zoll device can perform 3 different types of pacing: 1) Demand Pacing: Most frequent form of ventricular pacing. The Pacing rate is set above patient s rate (or lack thereof) and the Pacer Output dial is turned to increase the ma in attempt to obtain capture and pace the ventricles. 2) Stand-by Pacing: Setting the Pacing rate and Pacer output at a back-up rate less than a patient s intrinsic heart rate. The Pacing rate will initially be set above the patient s heart rate and pacer output (ma) is increased to achieve 100% capture. The Pacing rate is then decreased to desired rate below the patient s intrinsic heart rate. Should the HR drop, the stand-by pacer will initiate impulses and begin to pace. 3) Asynchronised Pacing: Rarely used. This form of pacing is performed when no ECG electrodes can be placed due to burns, trauma or interference. The async on/off softkey button is pressed and async mode is displayed. No Pacing spike or electrical capture will be seen on the screen. Mechanical capture will only be proven, by palpating a pulse if one is achieved SOP- Transcutaneous Pacing 4

5 Triage The normal referral pathways for the referring site should be followed where possible and the local cardiology unit capable of permanent pacemaker insertion should be the destination of choice. In the situation where a patient may have had an acute coronary syndrome, that patient may require PCI and as such, must be triaged to a centre capable of providing this. Bibliography 1. ALSG website algorithm for brady-arrhythmia: 2. Zoll website for instruction manual for E series: SOP- Transcutaneous Pacing 5

Reviewed Date: Supersedes (Policy #/description & date): Prepared by: Jacque Callis, RN Approved by: D.R. Johnson, MD

Reviewed Date: Supersedes (Policy #/description & date): Prepared by: Jacque Callis, RN Approved by: D.R. Johnson, MD External Pacemaker Set-up and Revised Date: 11/13/2014 Page 1 of 5 POLICY: The transcutaneous external pacemaker will be used as a noninvasive method of delivering electrical stimuli to the myocardium

More information

Title and contents page 1 Who should read this document 2 Scope of the Guideline 2 Background 2 What is new in this version 2

Title and contents page 1 Who should read this document 2 Scope of the Guideline 2 Background 2 What is new in this version 2 Temporary Transvenous Pacing Guideline Classification: Clinical Guideline Lead Author: Dr Peter Woolfson Additional author(s): Dr Alan Fitchet Sister Joanne Hughes, Matron Julie Winstanley Authors Division:

More information

Figure 2. Normal ECG tracing. Table 1.

Figure 2. Normal ECG tracing. Table 1. Figure 2. Normal ECG tracing that navigates through the left ventricle. Following these bundle branches the impulse finally passes to the terminal points called Purkinje fibers. These Purkinje fibers are

More information

Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO

Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO Follow me on Twitter! CEE Med Updates@BarbaraFurryRN Like me on Facebook! What is a

More information

RN-BC, MS, CCRN, FAHA

RN-BC, MS, CCRN, FAHA Presented By: Barbara Furry, RN-BC, MS, CCRN, FAHA Director The Center of Excellence in Education Director of HERO Follow me on Twitter! CEE Med Updates@BarbaraFurryRN Like me on Facebook! 1 A. Atropine

More information

Cardioversion / Defibrillation / Transcutaneous Pacer Helpful Hints [ZOLL] 2018

Cardioversion / Defibrillation / Transcutaneous Pacer Helpful Hints [ZOLL] 2018 Cardioversion / Defibrillation / Transcutaneous Pacer Helpful Hints [ZOLL] 08 4 Universal cable may be inserted to the side Defib Short Block to perform 0 J check Universal Cable must be connected to APEX

More information

Transcutaneous Pacing. Approval: Medical Director James Stubblefield, MD. Approval: EMS Director Michael Petrie

Transcutaneous Pacing. Approval: Medical Director James Stubblefield, MD. Approval: EMS Director Michael Petrie Transcutaneous Pacing 6091D Low Frequency/High Risk: Transcutaneous Pacing Applies To: Paramedics Effective Approval: Medical Director James Stubblefield, MD Approval: EMS Director Michael Petrie Expires

More information

Early Defibrillation. Dr. M. Ravishankar

Early Defibrillation. Dr. M. Ravishankar Early Defibrillation Dr. M. Ravishankar THE CHAIN OF SURVIVAL CONCEPT Early Access Early CPR Early Defibrillation Early Advance Care 1 st Link 2 nd Link 3 rd Link 4 th Link Why early defibrillation VF

More information

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC

Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Puzzling Pacemakers Cheryl Herrmann, APN, CCRN, CCNS-CSC-CMC Pacemaker: An electric device implanted in the body to regulate the heart beat. Delivers electrical stimuli over leads with electrodes in contact

More information

M Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use

M Series with Rectilinear Biphasic Waveform Defibrillator Option Indications for Use DEFIBRILLATOR OPTION General Information Introduction M Series products are available with an advanced electrical design that provides a unique rectilinear biphasic waveform for defibrillation and cardioversion.

More information

Temporary pacemaker 삼성서울병원 심장혈관센터심장검사실 박정왜 RN, CCDS

Temporary pacemaker 삼성서울병원 심장혈관센터심장검사실 박정왜 RN, CCDS Temporary pacemaker 삼성서울병원 심장혈관센터심장검사실 박정왜 RN, CCDS NBG Codes 1st Letter 2nd Letter 3rd Letter A V D Chamber(s) Paced = atrium = ventricle = dual (both atrium and ventricle) Chamber(s) Sensed A = atrium

More information

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing. D. J. McMahon cewood rev Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing D. J. McMahon 141001 cewood rev 2017-10-04 Key Points Defibrillators: - know the definition & electrical value

More information

Cardiac Pacing. Learning outcomes. Introduction. The cardiac impulse - its formation and its failure CHAPTER. To understand:

Cardiac Pacing. Learning outcomes. Introduction. The cardiac impulse - its formation and its failure CHAPTER. To understand: Cardiac Pacing CHAPTER 10 Learning outcomes To understand: The indications for cardiac pacing in the peri-arrest setting How to perform percussion pacing How to apply non-invasive, transcutaneous electrical

More information

OBJECTIVE. 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation.

OBJECTIVE. 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation. Defibrillators OBJECTIVE 1. Define defibrillation. 2. Describe Need and history of defibrillation. 3. Describe the principle and mechanism of defibrillation. 4. Types and classes of defibrillator 5. Describe

More information

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I

PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I 1 PACEMAKER INTERPRETATION AND DEVICE MANAGEMENT PART I Cynthia Webner DNP, RN, CCNS, CCRN-CMC Karen Marzlin DNP, RN, CCNS, CCRN-CMC 2 PROFESSIONAL NURSING PRACTICE CAN ONLY ADVANCE AS MUCH AS INDIVIDUAL

More information

SARASOTA MEMORIAL HOSPITAL

SARASOTA MEMORIAL HOSPITAL SARASOTA MEMORIAL HOSPITAL TITLE: NURSING PROCEDURE PACEMAKERS AND DETERMINING PACEMAKER DATE: REVIEWED: PAGES: 10/99 11/18 1 of 7 PS1094 ISSUED FOR: Nursing RESPONSIBILITY: *Qualified RN PURPOSE: To establish

More information

Model 5392 EPG Temporary Pacer

Model 5392 EPG Temporary Pacer Model 5392 EPG Temporary Pacer Compatible Components Reference Card 5392 Surgical Cables 5487 Disposable, short 5487L Disposable, long 5832S Reusable, small clip 5833S 5833SL Disposable, small clip, short

More information

MODEL EPG SINGLE CHAMBER TEMPORARY EXTERNAL PACEMAKER

MODEL EPG SINGLE CHAMBER TEMPORARY EXTERNAL PACEMAKER MODEL 53401 EPG SINGLE CHAMBER TEMPORARY EXTERNAL PACEMAKER In-service Guide 53401 TRAINING LEARNING OBJECTIVES Identify the control knobs and be able to explain how to use them to program device. Distinguish

More information

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing

Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing Manual Defibrillators, Automatic External Defibrillators, Cardioversion, and External Pacing Key Points Defibrillators: - know the definition & electrical value of a joule - monophasic vs biphasic types:

More information

Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields

Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields Flowchart for ICD patients undergoing Surgery or procedures involving diathermy/magnetic fields Identify patient with an ICD at POAC. Notify Cardiac Physiologist that patient is due to have surgery and

More information

TIP CARD. Model Temporary External Pacemaker

TIP CARD. Model Temporary External Pacemaker TIP CARD Model 5392 Temporary External Pacemaker PRE-USE What to do... What NOT to do... Battery Replace the battery for each new patient. Do NOT reuse battery. Physical Check case for cracks/damage.**

More information

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa

MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa MARSHALLTOWN MEDICAL & SURGICAL CENTER Marshalltown, Iowa INTENSIVE CARE POLICY & PROCEDURES Policy Number: R-25 Subject: Policy: Purpose: Cardioversion - Synchronized An RN may perform a synchronized

More information

Pacemakers 12/04. Pacemakers. 1. What is a pacemaker?

Pacemakers 12/04. Pacemakers. 1. What is a pacemaker? Pacemakers 12/04 1-What is a pacemaker? 2- What does intrinsic mean? 3- How exactly do pacemakers work on the heart? 4- What are the parts of a pacemaker? 5- Are there different kinds of pacemakers? 6-

More information

Pediatric pacemakers & ICDs:

Pediatric pacemakers & ICDs: Pediatric pacemakers & ICDs: perioperative management Manchula Navaratnam Clinical Assistant Professor LPCH, Stanford SPA 2016 Conflict of interest: none Objectives Indications in pediatrics Components

More information

Peri-operative management of pacemakers and implantable cardiac defibrillators

Peri-operative management of pacemakers and implantable cardiac defibrillators Guideline Title: Peri-operative management of patients fitted with Permanent Pacemakers (PPMs) and Implantable Cardioverter Defibrillators (ICDs). Author(s): Jessica Osman (Chief Pacing Physiologist),

More information

Ass. Prof. Tomon Thongsri, MD Buddhachinaraj Phitsanuloke Hospital

Ass. Prof. Tomon Thongsri, MD Buddhachinaraj Phitsanuloke Hospital Treatment Bradyarrhythmia Ass. Prof. Tomon Thongsri, MD Buddhachinaraj Phitsanuloke Hospital 1 What is the rhythm? Sinus Bradycardia What s rhythm Sinus Bradycardia Treatment Asymptomatic No treatment

More information

The heart's "natural" pacemaker is called the sinoatrial (SA) node or sinus node.

The heart's natural pacemaker is called the sinoatrial (SA) node or sinus node. PACEMAKER Natural pacemaker: The heart's "natural" pacemaker is called the sinoatrial (SA) node or sinus node. Artificial pacemaker: It is a small, battery-operated device that helps the heart beat in

More information

1. What additional information needs to be collected to properly treat this client?

1. What additional information needs to be collected to properly treat this client? CASE 1 A 45-year-old male presents to the emergency department with a complaint of chest pain for the past two hours. 1. What additional information needs to be collected to properly treat this client?

More information

Panorama. Arrhythmia Analysis Frequently Asked Questions

Panorama. Arrhythmia Analysis Frequently Asked Questions Panorama Arrhythmia Analysis Frequently Asked Questions What ECG vectors are used for Beat Detection? 3-wire lead set 5-wire lead set and 12 lead What ECG vectors are used for Beat Typing? 3-wire lead

More information

Electrical Interventions

Electrical Interventions 3 Electrical Interventions Quick Contents Rationale for Electricity - p. 74 Paddle Placement - p. 77 Defibrillation - p. 82 Using an AED- p. 85 Manual Defibrillation - p. 86 Cardioversion - p. 87 Transcutaneous

More information

Non-Invasive Transcutaneous Pacing

Non-Invasive Transcutaneous Pacing Non-Invasive Transcutaneous Pacing Application Note Introduction This Application Note provides an overview of non-invasive transcutaneous pacing. It includes a brief history of pacing, describes the electrical

More information

Indications The AED mode is to be used only on patients in cardiopulmonary arrest.

Indications The AED mode is to be used only on patients in cardiopulmonary arrest. 14 October 2008 TO: Dr. Copass FR: Brian W. Chinn B3 A2 1018 RE: SFD LP12 Guidelines: I. AED / Advisory II. Manual A. Defibrillation B. Synchronize Cardioversion C. Noninvasive Pacing 1. Demand 2. Non-Demand

More information

June 2009 CE. Site code # E-1209

June 2009 CE. Site code # E-1209 1 June 2009 CE Site code # 107200E-1209 Region X SOP Review ACS, Asystole/PEA, Bradycardia, VF/Pulseless VT, Conscious Sedation, Region X Field Triage Criteria Prepared by: Sharon Hopkins, RN, BSN, EMT-P

More information

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS

PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS PERIOPERATIVE MANAGEMENT: CARDIAC PACEMAKERS AND DEFIBRILLATORS DR SUSAN CORCORAN CARDIOLOGIST ONCE UPON A TIME.. Single chamber pacemakers Programmed at 70/min VVI 70 UNIPOLAR SYSTEMS A Unipolar Pacing

More information

Project Title Temporary Pacemaker Training Simulator

Project Title Temporary Pacemaker Training Simulator Project Title Temporary Pacemaker Training Simulator Project Description Problem: There is no available training device for temporary pacemakers (pacemakers). A training device will have to essentially

More information

Automated External Defibrillation Principle of Early Defibrillation States that all BLS personnel be trained, equipped and allowed to operate a if

Automated External Defibrillation Principle of Early Defibrillation States that all BLS personnel be trained, equipped and allowed to operate a if 1 2 3 4 5 6 Automated External Defibrillation Principle of Early Defibrillation States that all BLS personnel be trained, equipped and allowed to operate a if they are expected to respond to persons in

More information

1 Cardiology Acute Care Day 22 April 2013 Arrhythmia Tutorial Course Material

1 Cardiology Acute Care Day 22 April 2013 Arrhythmia Tutorial Course Material 1 Cardiology Acute Care Day 22 April 2013 Arrhythmia Tutorial Course Material Arrhythmia recognition This tutorial builds on the ECG lecture and provides a framework for approaching any ECG to allow the

More information

1-Epinephrine 2-Atropine 3-Amiodarone 4-Lidocaine 5-Magnesium

1-Epinephrine 2-Atropine 3-Amiodarone 4-Lidocaine 5-Magnesium ١ 1-Epinephrine 2-Atropine 3-Amiodarone 4-Lidocaine 5-Magnesium ٢ When VF/pulseless VT cardiac arrest is associated with torsades de pointes, providers may administer magnesium sulfate at a dose of 1-2

More information

MINI MASSAGER. Relaxation,Anywhere, Anytime

MINI MASSAGER. Relaxation,Anywhere, Anytime MINI MASSAGER Relaxation,Anywhere, Anytime Inclu ded in this standard package: Device x 1pc Large pad x 2pcs Small pad x 2pcs Electric wire x 2pcs USB cable & AC adapter x 1set Acupuncture point chart

More information

COALINGA STATE HOSPITAL. NURSING POLICY AND PROCEDURE MANUAL SECTION Emergency Procedures POLICY NUMBER: 716. Effective Date: March 3, 2007

COALINGA STATE HOSPITAL. NURSING POLICY AND PROCEDURE MANUAL SECTION Emergency Procedures POLICY NUMBER: 716. Effective Date: March 3, 2007 COALINGA STATE HOSPITAL NURSING POLICY AND PROCEDURE MANUAL SECTION Emergency Procedures POLICY NUMBER: 716 Effective Date: March 3, 2007 SUBJECT: MONITOR/DEFIBRILLATOR 1. PURPOSE: This policy and procedure

More information

T.E.N.S (Transcutaneous Electrical Nerve Stimulation)

T.E.N.S (Transcutaneous Electrical Nerve Stimulation) You have been prescribed a T.E.N.S machine to help in the management of your pain. You should use it only for the condition for which it is prescribed. Not all pain will respond to the use of T.E.N.S and

More information

Chapter 80 Implantable Cardiac Devices

Chapter 80 Implantable Cardiac Devices Chapter 80 Implantable Cardiac Devices Episode Overview: 1. List 5 Indications for permanent pacing 2. Pacemaker nomenclature - what do the 5 letters mean a. What are common Pacemaker types? 3. List the

More information

NORTH OF ENGLAND CARDIOVASCULAR NETWORK (NECVN)

NORTH OF ENGLAND CARDIOVASCULAR NETWORK (NECVN) NORTH OF ENGLAND CARDIOVASCULAR NETWORK (NECVN) Operational policy for optimal management of: Cardiac pacemakers, Implantable cardioverter defibrillators (ICD), Cardiac resynchronisation devices, (CRT-D,

More information

Michigan Pediatric Cardiac Protocols. Date: November 15, 2012 Page 1 of 1 TABLE OF CONTENTS

Michigan Pediatric Cardiac Protocols. Date: November 15, 2012 Page 1 of 1 TABLE OF CONTENTS Date: November 15, 2012 Page 1 of 1 TABLE OF CONTENTS Pediatric Asystole Section 4-1 Pediatric Bradycardia Section 4-2 Pediatric Cardiac Arrest General Section 4-3 Pediatric Narrow Complex Tachycardia

More information

Implantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević

Implantable Cardioverter-Defibril. Defibrillators. Ratko Magjarević Implantable Cardioverter-Defibril Defibrillators Ratko Magjarević University of Zagreb Faculty of Electrical Engineering and Computing Croatia ratko.magjarevic@fer.hr Ventricular Fibrillation Ventricular

More information

STANDARD OPERATING PROCEDURE (SOP)

STANDARD OPERATING PROCEDURE (SOP) STANDARD OPERATING PROCEDURE (SOP) MRI Scanning for Patients with Cardiac Implantable Electronic Devices (Pacemakers and Implantable Cardiac Defibrillators) Effective date 05/04/2018 Version number V1.0

More information

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity.

Newer pacemakers also can monitor your blood temperature, breathing, and other factors and adjust your heart rate to changes in your activity. Pacemakers & Defibrillators A pacemaker system consists of a battery, a computerized generator and wires with sensors called electrodes on one end. The battery powers the generator, and both are surrounded

More information

Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale

Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale Permanent Pacemaker Implantation Post Cardiac Surgery: A Cautionary Tale Jason Collinson & Stuart Tan Essex Cardiothoracic Centre, Basildon and Thurrock University Hospital. Contact: jason.collinson@nhs.net

More information

Michigan Pediatric Cardiac Protocols. Date: November 15, 2012 Page 1 of 1 TABLE OF CONTENTS

Michigan Pediatric Cardiac Protocols. Date: November 15, 2012 Page 1 of 1 TABLE OF CONTENTS Date: November 15, 2012 Page 1 of 1 TABLE OF CONTENTS Pediatric Asystole Section 4-1 Pediatric Bradycardia Section 4-2 Pediatric Cardiac Arrest General Section 4-3 Pediatric Narrow Complex Tachycardia

More information

MONITORING PATIENTS WITH A TEMPORARY EXTERNAL EPICARDIAL PACEMAKER PURPOSE POLICY STATEMENTS SITE APPLICABILITY PRACTICE LEVEL/COMPETENCIES

MONITORING PATIENTS WITH A TEMPORARY EXTERNAL EPICARDIAL PACEMAKER PURPOSE POLICY STATEMENTS SITE APPLICABILITY PRACTICE LEVEL/COMPETENCIES PURPOSE Guidelines for caring for and monitoring patients with a temporary external epicardial pacemaker. Epicardial pacemaker monitoring is performed on a continuous basis by a nurse to evaluate the effectiveness

More information

ECG Monitoring, Defibrillation and Synchronized Cardioversion with Hands-Free Electrodes

ECG Monitoring, Defibrillation and Synchronized Cardioversion with Hands-Free Electrodes Suggestions for Performance Evaluation with the LIFEPAK 20 Defibrillator/Monitor ECG Monitoring, Defibrillation and Synchronized Cardioversion with Hands-Free Electrodes Name: Unit: Reviewer: Date: This

More information

Defibrillator. BiomedGuy

Defibrillator. BiomedGuy Defibrillator BiomedGuy Medtronic Physiocontrol LifePak 10 Introduction This life-support system is used by paramedic, hospital staff, and other trained authorized healthcare providers. Provides, ECG,

More information

DEFIBRILLATORS. Prof. Yasser Mostafa Kadah

DEFIBRILLATORS. Prof. Yasser Mostafa Kadah DEFIBRILLATORS Prof. Yasser Mostafa Kadah Basics Defibrillation is definitive treatment for life-threatening cardiac arrhythmias such as ventricular fibrillation Defibrillation consists of delivering therapeutic

More information

CORONARY ARTERIES HEART

CORONARY ARTERIES HEART CARDIAC/ECG MODULE THE HEART CORONARY ARTERIES FIBRILLATING HEART CORONARY ARTERIES HEART PRACTICE RHYTHMS PRACTICE RHYTHMS ELECTRICAL CONDUCTION SA Node (60 100) Primary pacemaker AV Node (40 60) ***Creates

More information

EXTERNAL TEMPORARY PACEMAKERS

EXTERNAL TEMPORARY PACEMAKERS Czech Republic EXTERNAL TEMPORARY PACEMAKERS Lukas Peter Engineering and Business Coordinator Pavel Taborsky Chief Executive Officer MEDIATRADE s.r.o. PRAGUE Valasske Mezirici MEDIATRADE s.r.o. More than

More information

Addendum: User Guide AutoPulse Resuscitation System Model 100 with Defibrillator Interface

Addendum: User Guide AutoPulse Resuscitation System Model 100 with Defibrillator Interface Addendum: User Guide AutoPulse Resuscitation System Model 100 with Defibrillator Interface 3.5 Periodic Electrocardiogram (ECG) Monitoring and/or Defibrillation When the AutoPulse is used in conjunction

More information

Defibrillators TEC-8300K series. Improving quality of resuscitation

Defibrillators TEC-8300K series. Improving quality of resuscitation Defibrillators TEC-8300K series Improving quality of resuscitation Transitioning back to life For intensive treatment and advanced life support Pre-hospital Out-of-hospital 12-lead ECG acquisition and

More information

STATE OF OKLAHOMA 2014 EMERGENCY MEDICAL SERVICES PROTOCOLS

STATE OF OKLAHOMA 2014 EMERGENCY MEDICAL SERVICES PROTOCOLS 5M VENTRICULAR ASSIST DEVICE (VAD) MANAGEMENT ADULT EMERGENCY MEDICAL DISPATCHER EMERGENCY MEDICAL RESPONDER EMT EMT-INTERMEDIATE 85 ADVANCED EMT PARAMEDIC A Ventricular Assist Device, or VAD, is a mechanical

More information

Dual Chamber Temporary External Pacemaker

Dual Chamber Temporary External Pacemaker 5392 Dual Chamber Temporary External Pacemaker Technical Manual Caution: Federal law (USA) restricts this device to sale by or on the order of a physician. The following list includes trademarks or registered

More information

Subcutaneous implantable cardioverter defibrillator insertion for preventing sudden cardiac death

Subcutaneous implantable cardioverter defibrillator insertion for preventing sudden cardiac death NATIONAL INSTITUTE FOR HEALTH AND CARE EXCELLENCE Interventional procedure consultation document Subcutaneous implantable cardioverter defibrillator insertion for preventing sudden cardiac death A subcutaneous

More information

Mini Pulse Electronic Stimulator

Mini Pulse Electronic Stimulator Mini Pulse Electronic Stimulator Model: PM-180 Operating Manual IMPORTANT: Please read all instructions before using this product. Retain this manual for future reference. www.santamedical.com IMPORTANT

More information

ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ

ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ ΤΟ ΗΚΓ ΣΤΟΝ ΒΗΜΑΤΟΔΟΤΟΥΜΕΝΟ ΑΣΘΕΝΗ ΤΖΩΡΤΖ ΔΑΔΟΥΣ ΕΠΙΚΟΥΡΟΣ ΚΑΘΗΓΗΤΗΣ Α.Π.Θ. ΜΑΡΙΑ ΚΑΡΑΛΙΟΛΙΟΥ ΕΙΔΙΚΕΥΟΜΕΝΗ ΙΑΤΡΟΣ ΚΑΡΔΙΟΛΟΓΟΣ Β ΚΑΡΔΙΟΛΟΓΙΚΗ ΚΛΙΝΙΚΗ Α.Π.Θ. ΙΠΠΟΚΡΑΤΕΙΟ Γ.Ν.Θ. 5 Why a Pacemaker is Implanted

More information

WHY IS FIRST AID IMPORTANT?

WHY IS FIRST AID IMPORTANT? Chapter 1 -BACKGROUND INFORMATION WHY IS FIRST AID IMPORTANT? 1. 2. SOME THINGS I SHOULD HAVE IN MY FIRST AID KIT 1. 2. 3. 4. 5. VOCABULARY Place the key points of the first aid term in the box provided

More information

TEMPORARY PACEMAKERS HINTS ( DUAL CHAMBER 5392 PACER)

TEMPORARY PACEMAKERS HINTS ( DUAL CHAMBER 5392 PACER) TEMPORARY PACEMAKERS HINTS ( DUAL CHAMBER 5392 PACER) TYPES OF PACERS 2018 1 TRANSCUTANEOUS: ( VENTRICULAR pacing only) Through the skin via gel electrode pads placed on the chest and back. Disadvantage:

More information

Essentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH

Essentials of Pacemakers and ICD s. Rajesh Banker, MD, MPH Essentials of Pacemakers and ICD s Rajesh Banker, MD, MPH Pacemakers have 4 basic functions: Stimulate cardiac depolarization Sense intrinsic cardiac function Respond to increased metabolic demand by providing

More information

Electrocardiography for Healthcare Professionals

Electrocardiography for Healthcare Professionals Electrocardiography for Healthcare Professionals Kathryn A. Booth Thomas O Brien Chapter 10: Pacemaker Rhythms and Bundle Branch Block Learning Outcomes 10.1 Describe the various pacemaker rhythms. 10.2

More information

Icd 10 for pacemaker pocket infection

Icd 10 for pacemaker pocket infection Search Search pages & people Search Search Search pages & people Search Icd 10 for pacemaker pocket infection Oct 29, 2014. 9.1.2 In generator pocket infection, ICED-LI and ICED-IE? 9.1.3 What is the.

More information

Emergency Department Management of Patients with Implantable Cardioverter Defibrillators

Emergency Department Management of Patients with Implantable Cardioverter Defibrillators IAEM Clinical Guideline Emergency Department Management of Patients with Implantable Cardioverter Defibrillators Version 1 June 2014 DISCLAIMER IAEM recognises that patients, their situations, Emergency

More information

AWAY Model # RTLAGF-900 REV

AWAY Model # RTLAGF-900 REV AWAY Model # RTLAGF-900 REV2.4.9.15 Contents User manual English 2 General Description 2 Intended Use 3 Included in this Package 4 Instructions for Use 7 Application Duration 7 Electrode Placement 8 Tips

More information

SJM MRI ACTIVATOR HANDHELD DEVICE WORKFLOW Model: EX4000. SJM-EDTR Item approved for U.S. use only.

SJM MRI ACTIVATOR HANDHELD DEVICE WORKFLOW Model: EX4000. SJM-EDTR Item approved for U.S. use only. SJM MRI ACTIVATOR HANDHELD DEVICE WORKFLOW Model: EX4000 1 APPROVED MRI SCAN CONDITIONS 1.5 TESLA FULL BODY 2 SETUP SJM MRI ACTIVATOR HANDHELD DEVICE The SJM MRI Activator handheld device must be enabled

More information

Adult Advanced Cardiovascular Life Support 2015 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular

Adult Advanced Cardiovascular Life Support 2015 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Adult Advanced Cardiovascular Life Support 2015 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care 1 DR. Alireza Abootalebi Assistant Professor Of

More information

ICD Implantation Patient Information

ICD Implantation Patient Information Melbourne Heart Rhythm ICD Implantation Patient Information The Heart The heart is a pump responsible for maintaining blood supply to the body. It has four chambers. The two upper chambers (the right atrium

More information

QUICK SETUP GUIDE SECULIFE DFBASE

QUICK SETUP GUIDE SECULIFE DFBASE QUICK SETUP GUIDE SECULIFE DFBASE SECULIFE DFBASE is a microprocessor-based instrument that is used in the testing of defibrillators. It measures the energy output and provides information about the defibrillator

More information

UnitedHealthcare Medicare Advantage Cardiology Prior Authorization Program

UnitedHealthcare Medicare Advantage Cardiology Prior Authorization Program Electrophysiology Implant Classification Table The table below contains the codes that apply to our UnitedHealthcare Medicare Advantage cardiology prior Description Includes Generator Placement Includes

More information

Perioperative Management of Cardiac Implantable Devices

Perioperative Management of Cardiac Implantable Devices Financial Disclosures Breandan Sullivan MD Assistant Professor Co Director Cardiothoracic and Vascular Surgery ICU Department of Anesthesiology and Critical Care Medicine University of Colorado None Perioperative

More information

Cardiac Resynchronisation Therapy Patient Information

Cardiac Resynchronisation Therapy Patient Information Melbourne Heart Rhythm Cardiac Resynchronisation Therapy Patient Information Normal Heart Function The heart is a pump responsible for maintaining blood supply to the body. It has four chambers. The two

More information

NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS

NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS NEIL CISPER TECHNICAL FIELD ENGINEER ICD/CRTD BASICS OBJECTIVES Discuss history of ICDs Review the indications for ICD and CRT therapy Describe basic lead and device technology Discuss different therapies

More information

DEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION

DEFIBRILLATORS ATRIAL AND VENTRICULAR FIBRILLATION 1 DEFIBRILLATORS The two atria contract together and pump blood through the valves into the two ventricles, when the action potentials spread rapidly across the atria surface. After a critical time delay,

More information

Pediatric Cardiac Arrest General

Pediatric Cardiac Arrest General Date: November 15, 2012 Page 1 of 5 Pediatric Cardiac Arrest General This protocol should be followed for all pediatric cardiac arrests. If an arrest is of a known traumatic origin refer to the Dead on

More information

Pacing Lead Implant Testing. Document Identifier

Pacing Lead Implant Testing. Document Identifier Pacing Lead Implant Testing 1 Objectives Upon completion of this presentation, the participant should be able to: Name the two primary surgical options for implanting pacing leads Describe three significant

More information

New and Future Trends in EMS. Ron Brown, MD, FACEP Paramedic Lecture Series 2018

New and Future Trends in EMS. Ron Brown, MD, FACEP Paramedic Lecture Series 2018 New and Future Trends in EMS Ron Brown, MD, FACEP Paramedic Lecture Series 2018 New technologies and protocols DSD Mechanical Compression ITD BiPAP Ultrasound Double Sequential Defibrillation Two defibrillators

More information

7 Cedar Street, Suite 202 Sudbury, ON P3E 1A

7 Cedar Street, Suite 202 Sudbury, ON P3E 1A 7 Cedar Street, Suite 202 Sudbury, ON P3E 1A2 705.675.4783 To: Northeast EMS Providers CC: J. Cruickshank, J. Bradbury From: Dr. Jason Prpic, Regional Medical Director Re: Deployment and/or Redeployment

More information

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists

PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists PERMANENT PACEMAKERS AND IMPLANTABLE DEFIBRILLATORS Considerations for intensivists Craig A. McPherson, MD, FACC Associate Professor of Medicine Constantine Manthous, MD, FACP, FCCP Associate Clinical

More information

Supplemental Material

Supplemental Material Supplemental Material 1 Table S1. Codes for Patient Selection Cohort Codes Primary PM CPT: 33206 or 33207 or 33208 (without 33225) ICD-9 proc: 37.81, 37.82, 37.83 Primary ICD Replacement PM Replacement

More information

Cardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc.

Cardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc. Cardiac Implanted Electronic Devices Pacemakers, Defibrillators, Cardiac Resynchronization Devices, Loop Recorders, etc. The Miracle of Living February 21, 2018 Matthew Ostrom MD,FACC,FHRS Division of

More information

Improving quality of resuscitation. Defibrillators TEC-5600 series

Improving quality of resuscitation. Defibrillators TEC-5600 series Improving quality of resuscitation Defibrillators TEC-5600 series Transitioning back to Life Improving quality of resuscitation Saving Life Shock Efficiency on Demand Fast shocks and continuous observation

More information

SIX PACK ABS Item No INSTRUCTION MANUAL. Read entire manual before operating this product. Use only as directed.

SIX PACK ABS Item No INSTRUCTION MANUAL. Read entire manual before operating this product. Use only as directed. SIX PACK ABS Item No. 206098 INSTRUCTION MANUAL Read entire manual before operating this product. Use only as directed. WARNINGS If you are in the care of a physician, consult your physician before using

More information

BEDSIDE ECG INTERPRETATION

BEDSIDE ECG INTERPRETATION BEDSIDE ECG INTERPRETATION Presented by: Ryan Dean, RN, MSN, CCRN, CCNS, CFRN Flight Nurse 2017 Based on presentations originally by Gennifer DePaoli, RN Objectives Hospital policies Electrical conduction

More information

Arrhythmias. Pulmonary Artery

Arrhythmias. Pulmonary Artery Arrhythmias Introduction Cardiac arrhythmia is an irregularity of the heart beat that causes the heart to beat too slowly, too fast, or irregularly. There are different types of arrhythmias. Most arrhythmias

More information

INTERNAL CARDIOVERSION. Lancashire & South Cumbria Cardiac Network

INTERNAL CARDIOVERSION. Lancashire & South Cumbria Cardiac Network INTERNAL CARDIOVERSION Lancashire & South Cumbria Cardiac Network DC Cardioversion AF & AFL safe 1,2 efficacious 3,4 (60-94%) 5 SR - Increases exercise tolerance 6 Maintainence SR unlikely in patients

More information

1. Normal sinus rhythm 2. SINUS BRADYCARDIA

1. Normal sinus rhythm 2. SINUS BRADYCARDIA 1. Normal sinus rhythm 2. SINUS BRADYCARDIA No signs and symptoms observe There are severe signs or symptoms o What are the signs and symptom Hypotension

More information

Transcutaneous Electrical Nerve Stimulation Device GF-3 / GF-3T

Transcutaneous Electrical Nerve Stimulation Device GF-3 / GF-3T Transcutaneous Electrical Nerve Stimulation Device GF-3 / GF-3T Operation Manual Read Before Using GF-3-INS-LAB-RevA08 TABLE OF CONTENTS INTRODUCTION TO TENS INDICATIONS AND CONTRAINDICATIONS WARNINGS

More information

Pacing Codes and Modes Concepts

Pacing Codes and Modes Concepts Pacing Codes and Modes Concepts Pacing codes and modes concepts Objectives Upon completion of this program the participant will be able to: State what the first four positions of the NBG code represent.

More information

Kadlec Regional Medical Center Cardiac Electrophysiology

Kadlec Regional Medical Center Cardiac Electrophysiology Definition of electrophysiology study and ablation Kadlec Regional Medical Center Cardiac Electrophysiology Electrophysiology Study and Ablation An electrophysiology, or EP, study is a test of the heart

More information

Revised 3/3/2008 References provided on last slide

Revised 3/3/2008 References provided on last slide / Medics 12 Lead ECG and STEMI Triage Protocols 5.1 & 4.27 Revised 3/3/2008 References provided on last slide Objectives Purpose of Pre-Hospital 12-Lead ECGs Pre-Hospital 12-Lead ECG Background 12-Lead

More information

TASER and In-Custody Deaths

TASER and In-Custody Deaths TASER and In-Custody Deaths Cardiac Effects of Conducted Electrical Weapons Patrick Tchou, MD Heart and Vascular Institute What is a stun gun? A device that delivers rapid electrical impulses that stimulates

More information

Finally, a pacemaker may be either permanent or temporary, which will also factor into your code selections.

Finally, a pacemaker may be either permanent or temporary, which will also factor into your code selections. 2015 Cardiology Survival Guide Chapter 15: Pacemakers With more than 30 codes and many variables to choose from, you must weigh your options carefully when reporting insertion, revision, or removal of

More information

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up

Cardiac Rhythm Device Management. PBL STOP Your acronym for a standardized follow-up Cardiac Rhythm Device Management PBL STOP Your acronym for a standardized follow-up What s in it for you? What do you need to feel comfortable with a pacemaker/icd follow-up? 2 CORE OBJECTIVE Provide a

More information

Temporary Pacemakers

Temporary Pacemakers 07 Part I Anaesthesia Refresher Course 2018 University of Cape Town Dr Owen Porrill Private Practice Honorary lecturer- University of Cape Town Temporary pacemakers are usually inserted as part of the

More information