Mercury versus Non-Mercury Sphygmomanometers: A Review of the Comparative Clinical Effectiveness and Guidelines

Size: px
Start display at page:

Download "Mercury versus Non-Mercury Sphygmomanometers: A Review of the Comparative Clinical Effectiveness and Guidelines"

Transcription

1 TITLE: Mercury versus Non-Mercury Sphygmomanometers: A Review of the Comparative Clinical Effectiveness and Guidelines DATE: 25 September 2012 CONTEXT AND POLICY ISSUES A mercury manual sphygmomanometer is a device attached to the forearm to record arterial pulsations. Basic components include an inflatable cuff to put pressure on the artery, a pressure-measuring scale, and a bulb to inflate the cuff. As the cuff is inflated, the mercury in the manometer rises through the column, and the pressure is displayed on the scale. A stethoscope is placed over the brachial artery of the elbow to detect the systolic and diastolic pressure during the slow deflation of the pressure of the cuff. Since the instrument relies on the human judgment to detect systolic and diastolic pressure, it is termed an auscultatory device. 1 Accuracy of mercury sphygmomanometers depend on proper validation, calibration, and maintenance. Several factors may affect the accuracy including measurement technique, proper cuff size, proper placement of the cuff, accurate recording of the systolic and diastolic endpoints, patient s level of anxiety, and patient-physician interaction (e.g. white coat response ). 1 Poor maintenance of the mercury sphygmomanometer may result in loss of accuracy due to equipment deterioration such as an obscure mercury column due to dirt or mercury oxidation, faded scales that are hindering the readings of mercury meniscus, and mercury leak, which is a major health concern. 2 Despite its limitations, the mercury sphygmomanometer is still considered gold standard in terms of accuracy. 3 There are two common types of non-mercury sphygmomanometers: aneroid and oscillometric devices. 1 The aneroid devices are also operated manually and require a normal blood pressure cuff and a stethoscope, but are liquid-free and use mechanical parts to relay the blood pressure to the gauge. 1 The oscillometric devices are operated automatically, and the inflation and deflation of the cuff are controlled electronically. 1 Based on the pressure-wave changes, an algorithm in the oscillating device is used to calculate the systolic and diastolic pressures, which are displayed on a digital readout. 1 Accuracy of mercury-free sphygmomanometers also depends on proper maintenance, calibration and validation. 4 The perceive benefits of oscillometric devices are that they are more accurate, less time consuming, easier to operate, and safer in terms of being mercury-free. 5,6 Oscillometric devices may result in greater withinsubject reliability than manual readings, by eliminating digit preference (the preference to round a value to a lower or higher one), observer bias and the white coat effect. 6 However, the Disclaimer: The Rapid Response Service is an information service for those involved in planning and providing health care in Canada. Rapid responses are based on a limited literature search and are not comprehensive, systematic reviews. The intent is to provide a list of sources and a summary of the best evidence on the topic that CADTH could identify using all reasonable efforts within the time allowed. Rapid responses should be considered along with other types of information and health care considerations. The information included in this response is not intended to replace professional medical advice, nor should it be construed as a recommendation for or against the use of a particular health technology. Readers are also cautioned that a lack of good quality evidence does not necessarily mean a lack of effectiveness particularly in the case of new and emerging health technologies, for which little information can be found, but which may in future prove to be effective. While CADTH has taken care in the preparation of the report to ensure that its contents are accurate, complete and up to date, CADTH does not make any guarantee to that effect. CADTH is not liable for any loss or damages resulting from use of the information in the report. Copyright: This report contains CADTH copyright material. It may be copied and used for non-commercial purposes, provided that attribution is given to CADTH. Links: This report may contain links to other information available on the websites of third parties on the Internet. CADTH does not have control over the content of such sites. Use of third party sites is governed by the owners own terms and conditions.

2 accuracy of oscillometric devices remains questionable and the use of oscillometric devices is not clinically appropriate when patient has an arrhythmia. 7 The aim of this report is to review the comparative accuracy, safety and guidelines regarding the use of mercury versus non-mercury sphygmomanometers for the measurement of blood pressure. RESEARCH QUESTIONS 1. What is the evidence regarding the comparative accuracy of mercury versus nonmercury sphygmomanometers for the measurement of blood pressure? 2. What are the evidence-based guidelines regarding the use of mercury versus nonmercury sphygmomanometers in higher risk patient groups such as pregnancy and pediatric specialty clinics? KEY MESSAGE Automated (oscillometric) office blood pressure devices, with proper maintenance, calibration and validation, can be used as a replacement for the mercury sphygmomanometer in office settings for the measurement of blood pressure. In pregnant women, children, patients with arrhythmia, and hypertensive or hypotensive patients, identified guidelines recommend that the measurement readings of automated devices should be validated with a mercury sphygmomanometer. There is no evidence on the comparison of aneroid devices with mercury sphygmomanometer. METHODS Literature Search Strategy A limited literature search was conducted on key resources including PubMed, The Cochrane Library (2012 Issue 8), Canadian and major international health technology agencies, as well as a focused Internet search. Methodological filters were applied to limit retrieval to health technology assessments, systematic reviews, meta-analyses, randomized controlled trials, nonrandomized studies containing safety data, and guidelines. The search was also limited to English language documents published between August 27, 2007 and August 27, Selection Criteria and Methods One reviewer screened the titles and abstracts of the retrieved publications and evaluated the full-text publications for the final article selection, according to selection criteria presented in Table 1. Table 1: Selection Criteria Population Q1: Adult and pediatric patients in hospital or medical clinic Q2: High risk patients Intervention Non-mercury sphygmomanometers Comparator Mercury sphygmomanometers (gold standard) Mercury versus Non-Mercury Sphygmomanometers 2

3 Outcomes Study Designs Accuracy and calibration needs for non-mercury sphygmomanometers Possible harms (acute inhalation or chronic exposure) due to mercury spill with mercury sphygmomanometers Guidelines regarding the use of non-mercury sphygmomanometers Health technology assessments, systematic reviews, meta-analyses, randomized controlled trials, non-randomized controlled trials (for safety only), guidelines Exclusion Criteria Studies were excluded if they did not satisfy the selection criteria in Table 1, if they were published prior to 2007, duplicate publications of the same study, or included in a selected health technology assessment or systematic review. Critical Appraisal of Individual Studies The methodological quality of the systematic reviews were assessed using AMSTAR checklist. 8 The quality of the RCTs and observational studies was assessed using Downs and Black checklist. 9 The Appraisal of Guidelines Research & Evaluation (AGREE) instrument 10 was used to evaluate the quality of the included guidelines. For the critical appraisal of studies, a numeric score was not calculated. Instead, the strength and limitations of the study were described. SUMMARY OF EVIDENCE Quantity of Research Available The literature search yielded 55 citations. Upon screening titles and abstracts, eight potential relevant articles were retrieved for full-text review. Ten additional relevant reports were retrieved from other sources. Of the 18 potentially relevant articles, nine were included in this review. They are two systematic reviews, 11,12 three RCTs, and four guidelines The study selection process is outlined in a PRISMA flowchart (Appendix 1). Summary of Study Characteristics A summary of study characteristics can be found in Appendix 2. Systematic reviews and meta-analyses: The systematic review and meta-analysis by Stergiou et al assessed the accuracy of automated blood pressure measurement in patients with sustained atrial fibrillation. Eight observational studies (12 validations: four ambulatory, five home, and three office devices) published between January 1986 and October 2011 with a total of 566 participants (57% men, mean age 73 years) with sustained atrial fibrillation were included. Sample sizes ranged from 20 to 255 participants. A standard mercury sphygmomanometer was used as reference method in five validations, a Hawksley random-zero method in five, and an aneroid sphygmomanometer in two. For comparison of the two methods, blood pressure was measured either sequential (nine validations) or simultaneous (three validations). Average values pooled by random-effects meta- Mercury versus Non-Mercury Sphygmomanometers 3

4 analysis and accounting for heterogeneity were estimated with 95% confidence intervals for correlation coefficients and for differences between automated and auscultatory blood pressure measurements (systolic and diastolic). The systematic review by Skirton et al compared the accuracy and appropriateness of auscultatory (manual) and oscillometric (automated) devices for measuring blood pressure in clinical settings. Sixteen studies (15 observational studies and 1 RCT) published in English between January 1997 and May 2009 were included. Quality of the evidence was appraised and the results were summarized in a narrative form. The score for quality appraisal ranged from 45 to 95%, covering study design, subject group, means of analysis, sample size and estimates of variance. In nine of the studies, the manual reading was done by trained observers, while in three studies the readings were made by clinical nurses. In other studies, measurements were taken by nurses with or without specific training or by medical residents. The practice settings for the studies included emergency care, hypertension clinics, general wards, or outpatient clinics. Sample sizes ranged from 33 to 997 participants. The way measurements were taken and recorded varied among studies. Randomized controlled trials: The trial by Brown et al was a prospective randomized open study with blinded endpoints design. The objective was to determine whether women diagnosed with hypertension in pregnancy, using traditional mercury sphygmomanometers, have equivalent pregnancy outcomes when blood pressure is measured thereafter by either an automated blood pressure recorder or a mercury sphygmomanometer throughout pregnancy. Pregnant women with hypertension (mean age 31 years, N=220) were randomized into one of the two groups: the mercury sphygmomanometer (n=110) or the automated (Omron HEM-705CP) blood pressure recording group. The primary outcome was the number of women in each group developing any episode of severe hypertension ( 170 mm Hg systolic and/or 110 mm Hg diastolic), which was used as a surrogate marker for detecting the risk of stroke and alerting the cases of severe preeclampsia or gestational hypertension. Secondary endpoints included gestational time at birth, small for gestational age (<10 th centile) rates, caesarean section, and induction of labor rates. The multi-site cluster randomized controlled trial by Myers et al compared the quality and accuracy of manual office blood pressure and automated blood pressure using the awake ambulatory blood pressure as a gold standard. The setting was primary care practices in five cities in eastern Canada. The sites consisted of either one family physician or a group of two or three physicians sharing the same office space. Adult patients (over 45 years, N=555) with systolic hypertension (160 mm Hg / 95 mm Hg for untreated or 140 mm Hg / 90 mm Hg for treated) and no serious comorbidities were randomly assigned to either manual office blood pressure with continuing use of manual sphygmomanometer (control group) or automated office blood pressure (intervention group) using the BpTRU device. The baseline characteristics were balanced between groups in terms of age, gender, duration of hypertension, antihypertensive treatment and cigarette smokers. The BpTRU is a fully automated sphygmomanometer that records blood pressure by an oscillometric method. It was set to take readings at two minute intervals (from the start of one reading to the start of the next one) while the patients were left alone during five readings. For the manual office blood pressure control group, physicians continued recording blood pressure as before, without any additional instructions on proper blood pressure measurement technique. All patients were instructed to have 24 hour ambulatory blood pressure monitoring using a Spacelabs Model unit, which has been validated for accuracy. The device was set to record blood pressure at 15 minute intervals between 6:00 and 22:00 and at 30 minute intervals during the night. The primary outcome was Mercury versus Non-Mercury Sphygmomanometers 4

5 the mean estimated difference between the mean awake ambulatory blood pressure and the automated blood pressure (intervention group) or manual office blood pressure (control group) at the first office visit after enrolment. Lamarre-Cliché et al conducted a randomized, crossover study to compare the blood pressure measurement by mercury sphygmomanometer and other three blood pressure measurement modalities, namely the automated office blood pressure (AOBP) device (BpTRU), the ambulatory blood pressure measurement (ABPM) device (Spacelabs, model ), and self-measurement (Omron HEM-780CAN). Patients (mean age 58 years, N=101) from hypertension clinic with average blood pressure of 134 ± 12.7/83.6 ± 8.8 mm Hg were recruited and randomized to either the office or home group. The office group was then randomized to undergo an AOBP Sphygmomanometer or Sphygmomanometer AOBP sequence. The home group was randomized to undergo either a self-measurement ABPM or ABPM self measurement sequence. The total length of the study was 8 to 14 days. The primary study outcome was achievement of blood pressure goal. Guidelines: The characteristics on the grading of recommendations and levels of evidence used to develop the corresponding guidelines are summarized in Appendix 3. Summary of Critical Appraisal The strengths and limitations of included studies are summarized in Appendix 4. The methodological quality assessment of the systematic review by Stergiou et al met 9 out of 11 items, while that of the systematic review by Skirton et al met 8 out of 11 items of the AMSTAR criteria. In Stergiou, the scientific quality of the include studies was not assessed and considered in the conclusions. In Skirton, the likelihood of publication bias was not assessed, methods of study selection and data extraction were not described, and there was no pooling of the results from the included studies. Overall, all three included RCTs met most items of the Downs and Black checklist for measuring study quality. A major limitation of all RCTs was the lack of identification of potential confounders. One RCT 13 made attempt to blind the observers undertaking data collection and analyzing data. Although the scope and purpose were explicit, the included guidelines suffered from multiple limitations, covering stakeholder involvement, rigour of development applicability, and editorial independence according to AGREE instrument. Methods used to search for the evidence were not reported in the identified guidelines. One guideline 16 reported procedures for updating the recommendations. Clarity and presentation criteria were met in three guidelines. 16,17,19 Summary of Findings The main study findings and authors conclusions from the clinical studies can be found in Appendix 5. Systematic review: In the systematic review and meta-analysis by Stergiou et al., 11 automated measurements were higher than manual blood pressure measurements [pooled average systolic difference 0.5 mm Hg, 95% CI -0.9, 1.9; diastolic 2.5 mm Hg, 95% CI -0.6, 5.7). Pooled correlation coefficients between automated and manual bleed pressure measurements were stronger for systolic than Mercury versus Non-Mercury Sphygmomanometers 5

6 diastolic blood pressure (r=0.89 versus 0.76, P<0.001). Four of eight studies reporting data regarding the within 5 mm Hg ESH-IP criterion have passed for SBP (at least 65% of comparisons within 5 mm Hg) and two for DBP, suggesting better measurement accuracy for systolic blood pressure. Three studies showed no impact of heart rate on the blood pressure differences between automated and manual measurements. It was concluded that automated blood pressure monitors are accurate in measuring systolic but not diastolic blood pressure in elderly with atrial fibrillation. In the systematic review by Skirton et al 2011, 12 oscillometric (automated) devices were found less accurate than the mercury sphygmomanometer in 10 of the included studies. However, in most case the authors of the included studies concluded that the differences were not clinically significant, and the oscillometric devices could be used in the clinical settings, with the exceptions being used with hypertensive patients, patients with arrhythmia and patients with trauma. Randomized controlled trial: In the trial by Brown et al. 2012, 13 blood pressures obtained by six sequential recordings (alternating between mercury and automated) at randomization, according to the allocated group were similar. Both groups had similar maternal and fetal outcomes. The percentage of women with any episode of severe hypertension was similar between groups (39% for mercury vs. 44% for automated, P=0.5). The average number of episodes of severe hypertension per woman affected was also similar between groups (3 ± 2 for mercury vs. 4 ± 5 for automated, P=0.3). Small gestational age (<10 th centile) occurred more frequently in those using automated blood pressure than mercury sphygmomanometer (17% vs. 12%), but the difference was not statistically significant (P=0.3). Birth weight and perinatal mortality were similar among groups. It was concluded that the use of automated blood pressure recording is associated with similar maternal and fetal outcomes compared with mercury sphygmomanometer. In the trial by Myers et al. 2011, 14 both groups showed a drop in mean office blood pressure between the pre-study office visit and blood pressure recorded after enrolment. The reduction in the automated group was numerically higher than that in the manual group, but the difference was not statistically significant. With respect to the primary outcome measure, the mean estimated difference between awake ambulatory blood pressure and office blood pressure after enrolment was -2.3 (95% confidence interval [CI] -0.3 to -4.3) / -3.3 (95% CI -2.2 to -4.4) (P=0.02 / P<0.001) for automated group and -6.5 (95% CI -4.3 to -8.6) / -4.3 (95% CI -2.9 to - 5.8) (P<0.001 / P<0.001). The difference for systolic manual office blood pressure (-6.5) was significantly greater (P=0.006) than that for the systolic automated office blood pressure (-2.3). Automated office blood pressure showed a stronger (P<0.001) within group correlation (r=0.34 / r=0.56) with the awake ambulatory blood pressure compared with manual office blood pressure. The number of individual readings that showed digit preference, with readings being rounded off to the nearest zero, was significantly reduced (P<0.001) when automated office blood pressure readings were taken. It was concluded that the quality and accuracy of the automated blood pressure was better than the manual office blood pressure in relation to the awake ambulatory blood pressure. The automated office blood pressure measurement significantly reduced the white coat response compared with the manual office blood pressure measurement. In the trial by Lamarre-Cliché et al. 2011, 15 average blood pressures taken from all measurement methods were similar with the exception of daytime ambulatory blood pressure measurement (ABPM), which resulted in significantly greater systolic pressure compared with other methods. With the sphygmomanometer, blood pressure decreased by 1.2 ± 5.5 (P=0.029) / 0.4 ± 4.3 (P=0.40) mm Hg between the first and third measurements. With the automated Mercury versus Non-Mercury Sphygmomanometers 6

7 office blood pressure (AOBP) device, blood pressure decreased by 2.8 ± 7.5 (P=0.0004) / 1.1 ± 4.7 (P=0.40) mm Hg between the first and fifth measurements. According to cut-off values, target blood pressure of 140/90 mm Hg was reached in 67% of patients according to office sphygmomanometer and 70% according to the AOBP. Kappa coefficients measuring concordance showed fair agreement between two office methods (AOBP and sphygmomanometer; kappa coefficient 0.653; 95% CI to 0.813). AOBP had the best diagnostic performance when using a sphygmomanometer as the gold standard but had variable results when choosing other standards. It was concluded that AOBP device can be used as a replacement for the sphygmomanometer in office settings for the measurement of blood pressure in hypertensive patients. Guidelines: Three guidelines were identified that provided recommendations on the use of blood pressure measuring devices in hypertensive adults, in children, and in pregnant women. Recommendations stated that extra care should be given to the use of automated device in measuring blood pressure for patients having pulse irregularity, because the device may not measure blood pressure accurately. In pregnant women, mercury sphygmomanometer remains the gold standard for measurement of blood pressure. Automated or aneroid devices can be used to monitor blood pressure in women with preeclampsia. Because automated and aneroid devices are prone to error, it is recommended that a mercury sphygmomanometer should be in place for validation. In children, mercury or aneroid sphygmomanometer is recommended to be used for diagnostic evaluation of blood pressure elevation. However, automated devices may be acceptable for routine surveillance or in newborns and young infants, in whom auscultation may be difficult. Limitations Clinical Several limitations of the two systematic reviews 11,12 included differences in patient populations among studies, differences in protocol design, the lack of information about sample selection, and small sample sizes in some included studies. Many included studies in the systematic reviews were of observational design, which suffered from potential biases in selection and measurement. Due to the nature of the studies, it is difficult to conduct a blinded RCT in blood pressure measurement. All three RCTs used open-labeled design. However, one RCT 13 reported an attempt to blind those measuring the outcomes. The use of one type of automated device (BpTRU in two trials 14,15 and Omron in one trial 13 ) in each trial makes it difficult to extrapolate the findings to other automated devices. The study population in each trial was selective and it is not known for certain if the findings could be applied to the general population. Guidelines Of the four identified guidelines, one was from Canada, 19 which had recommendations for the type of device for blood pressure measurement in pregnant women. There was no specific guideline emphasizing the use of blood pressure monitoring devices, rather recommendations on the measurement of blood pressure were part of the management guidelines of specific individuals with hypertension. There were no tools or algorithms for the use of different type of devices in each situation. Mercury versus Non-Mercury Sphygmomanometers 7

8 CONCLUSIONS AND IMPLICATIONS FOR DECISION OR POLICY MAKING The collective current evidence suggests that automated office blood pressure monitoring devices, which are mercury-free, can be used in substitution of the mercury sphygmomanometer in clinical settings, although there are still some concerns about the accuracy of the devices with respect to specific populations such as those with hypertension, hypotension, arrhythmia, and children. Evidence from one RCT suggests that the maternal and fetal outcomes were comparable between automated and manual devices used to monitor throughout pregnancy. The mercury sphygmomanometer is still considered as the gold standard according to clinical practice guidelines for blood pressure measurement in patients with pulse irregularity, in children, and in pregnant women with hypertension. In those patients, it was recommended that blood pressure readings by automated devices should be verified by a mercury sphygmomanometer or an aneroid device. No evidence on the comparison of aneroid devices with mercury sphygmomanometers was identified. PREPARED BY: Canadian Agency for Drugs and Technologies in Health Tel: Mercury versus Non-Mercury Sphygmomanometers 8

9 REFERENCES 1. World Health Organization. Replacement of mercury thermometers and sphygmomanometers in health care. Technical guidance [Internet]. Geneva: WHO; [cited 2012 Sep 14]. Available from: 2. Markandu ND, Whitcher F, Arnold A, Carney C. The mercury sphygmomanometer should be abandoned before it is proscribed. J Hum Hypertens Jan;14(1): Buchanan S, Orris P, Karliner J. Alternatives to the mercury sphygmomanometer. J Public Health Policy Feb;32(1): Scientific Committee on Emerging and Newly Identified Health Risks. Mercury sphygmomanometers in healthcare and the feasibility of alternatives [Internet]. Brussels: European Commission; [cited 2012 Sep 14]. Available from: 5. Medicines and Healthcare products Regulatory Agency. Minutes of the meeting of the Committee on Blood Pressure Monitoring in Clinical Practice, Tuesday, 4 November London (UK): The Agency; Myers MG, Godwin M. Automated office blood pressure. Can J Cardiol May;28(3): Beevers G, Lip GY, O'Brien E. ABC of hypertension. Blood pressure measurement. Part I- sphygmomanometry: factors common to all techniques. BMJ [Internet] Apr 21 [cited 2012 Sep 14];322(7292): Available from: 8. Shea BJ, Grimshaw JM, Wells GA, Boers M, Andersson N, Hamel C, et al. Development of AMSTAR: a measurement tool to assess the methodological quality of systematic reviews. BMC Med Res Methodol [Internet] Feb 15 [cited 2012 Aug 8];7:10. Available from: 9. Downs SH, Black N. The feasibility of creating a checklist for the assessment of the methodological quality both of randomised and non-randomised studies of health care interventions. J Epidemiol Community Health [Internet] Jun [cited 2012 Aug 8];52(6): Available from: The AGREE Collaboration. Appraisal of guidelines for research and evaluation (AGREE) instrument [Internet]. London: The AGREE Research Trust; 2001 Sep. [cited 2012 Aug 8]. Available from: Stergiou GS, Kollias A, Destounis A, Tzamouranis D. Automated blood pressure measurement in atrial fibrillation: a systematic review and meta-analysis. J Hypertens Aug 21. Epub ahead of print. Mercury versus Non-Mercury Sphygmomanometers 9

10 12. Skirton H, Chamberlain W, Lawson C, Ryan H, Young E. A systematic review of variability and reliability of manual and automated blood pressure readings. J Clin Nurs Mar;20(5-6): Brown MA, Roberts LM, Mackenzie C, Mangos G, Davis GK. A prospective randomized study of automated versus mercury blood pressure recordings in hypertensive pregnancy (PRAM Study). Hypertens Pregnancy. 2012;31(1): Myers MG, Godwin M, Dawes M, Kiss A, Tobe SW, Grant FC, et al. Conventional versus automated measurement of blood pressure in primary care patients with systolic hypertension: randomised parallel design controlled trial. BMJ [Internet] [cited 2012 Sep 4];342:d286. Available from: Lamarre-Cliche M, Cheong NN, Larochelle P. Comparative assessment of four blood pressure measurement methods in hypertensives. Can J Cardiol Jul;27(4): National Institute for Health and Clinical Excellence. Hypertension: clinical management of primary hypertension in adults [Internet]. London: NICE; 2011 Aug. 36 p. [cited 2012 Sep 4]. (Clinical Guidelines CG127). Available from: Cincinnati Children's Hospital Medical Center. Blood pressure measurement in children [Internet]. Cincinnati (OH): The Centre; 2009 Jan 9. 9 p. [cited 2012 Sep 4]. (Best evidence statement (BESt)). Available from: Lowe SA, Brown MA, Dekker G, Gatt S, McLintock C, McMahon L, et al. Guidelines for the management of hypertensive disorders of pregnancy 2008 [Internet]. Sydney: Society of Obstetric Medicine of Australia and New Zealand (SOMANZ); p. [cited 2012 Sep 4]. Available from: Magee LA, Helewa M, Moutquin JM, von DP, Hypertension Guideline Committee, Strategic Training Initiative in Research in the Reproductive Health Sciences (STIRRHS) Scholars. Diagnosis and classification. Diagnosis, evaluation, and management of the hypertensive disorders of pregnancy. J Obstet Gynaecol Can Mar;30(3 Suppl):S9- S15. Mercury versus Non-Mercury Sphygmomanometers 10

11 APPENDIX 1: Selection of Included Studies 55 citations identified from electronic literature search and screened 47 citations excluded 8 potentially relevant articles retrieved for scrutiny (full text, if available) 10 potentially relevant reports retrieved from other sources (grey literature, hand search) 18 potentially relevant reports 9 reports excluded: irrelevant comparator (3) published in language other than English (1) observational study with no safety data (1) other (review articles, editorials)(4) 9 reports included in review 2 SR/MA 3 RCTs 4 guidelines Mercury versus Non-Mercury Sphygmomanometers 11

12 APPENDIX 2: Characteristics of Included Clinical Studies First Author, Publication Year, Country Study Design Systematic reviews and meta-analyses Stergiou et al Greece Skirton et al UK Systematic review and meta-analysis Systematic review Randomized controlled trials Brown et al. RCT open study with blinded end Australia points Source of funding: not reported Myers et al Canada Public funding Lamarre-Cliché et al Canada Source of funding: not reported Follow-up: remainder of pregnancy Multi-site cluster RCT RCT, crossover Length of study: 8-14 days Patient characteristics, sample Size (n) Individuals (57% men, mean age 73 years) with sustained atrial fibrillation from 8 studies, N=566 Participants (inpatients and outpatients) with or without elevated blood pressure from 16 included studies (15 observational studies and 1 RCT), N=4,429 Women (mean age 31 years) with hypertension in pregnancy (outpatients and inpatients), N=220 Patients (>45 years) with systolic hypertension and no serious comorbidities, N=555 Hypertensive patients (mean age 58 years) BP=blood pressure; RCT=randomized controlled trial Intervention Comparators Clinical Outcomes Oscillometric (automated) devices Oscillometric (automated) devices Automated (Omron HEM- 705 CP) devices, n=110 Automated office blood pressure device (BpTRU), n=303 Automated office blood pressure device (BpTRU), n=101 Mercury sphygmomanometer Ausculatory (manual) devices Mercury sphygmomanometer, n=110 Manual blood pressure device, n=252 Office mercury sphygmomanometer 24-hour ambulatory Daytime ambulatory Self-measurement Mean BP Mean BP Number of women having episode of severe hypertension Maternal outcomes Fetal outcomes Mean BP Mean BP Mean BP Target BP Mercury versus Non-Mercury Sphygmomanometers 12

13 APPENDIX 3: Grading of Recommendations and Levels of Evidence Guideline Society or Institute National Institute for Health and Clinical Excellence (NICE), UK, Cincinnati Children s Hospital Medical Center. Best evidence statement, Society of Obstetric Medicine of Australia and New Zealand (SOMANZ) Society of Obstetricians and Gynaecologists of Canada Magee et al Recommendation Level of Evidence The Center established a Guideline Development Group, which reviewed the evidence and updated the recommendations. An independent Guideline Review Panel oversaw the updating of the guideline. For more information, see Strongly recommended : There is consensus the benefits clearly outweigh risks and burdens (or visa-versa for negative recommendations) Recommended : There is consensus that benefits are closely balanced with risks and burdens No recommendation made: There is a lack of consensus to direct development of a recommendation Not indicated A. There is good evidence to recommend the clinical preventive action B. There is fair evidence to recommend the clinical preventive action C. the existing evidence is conflicting and does not allow to make a recommendation for or against use of the clinical preventive action; however, other factors may influence decision-making D. There is fair evidence to recommend against the clinical prevention action E. There is good evidence to recommend against the clinical prevention action I. There is insufficient evidence (in quantity or quality) to make a recommendation; however, other factors may influence decision-making 1a or 1b: Systematic review, meta-analysis, or meta-analysis of multiple studies 2a or 2b: Best study design for domain 3a or 3b: Fair study design for domain 4a: or 4b: Weak study design for domain 5: Other: General review, expert opinion, case report, consensus report, or guideline a = good quality study; b = lesser quality study Not indicated I: Evidence obtained from at least one properly randomized controlled trial II-1: Evidence from well-designed controlled trials without randomization II-2: Evidence from well-designed cohort (prospective or retrospective) or case-control studies, preferably from more than one centre or research group II-3: Evidence obtained from comparisons between times or places with or without the intervention. Dramatic results in uncontrolled experiments (such as results of treatment with penicillin in 1940s) could also be included in this category III: Opinions of respected authorities, based on clinical experience, descriptive studies, or reports of expert committees Mercury versus Non-Mercury Sphygmomanometers 13

14 APPENDIX 4: Summary of Study Strengths and Limitations First Author, Publication Year Strengths Systematic Reviews and Meta-analyses Stergiou et al Scored Yes on 9 out of 11 items of the AMSTAR checklist (1-6, 9-11) Skirton et al Scored Yes on 8 out of 11 items of the AMSTAR checklist (1, 3-8, 11) Limitations Scored No on 2 out of 11 items of the AMSTAR checklist (7, 8). List of excluded studies not provided Scored No on one item (10), Can t tell on one item (2), and Not applicable on one item (9) AMSTAR check list 1. Was an a priori design provided? 2. Was there duplicate study selection and data extraction? 3. Was a comprehensive literature search performed? 4. Was the status of publication (i.e., grey literature) used as an inclusion criteria? 5. Was a list of studies (included and excluded) provided? 6. Were the characteristics of the included studies provided? 7. Was the scientific quality of the included studies assessed and documented? 8. Was the scientific quality of the included studies used appropriately in formulating conclusions? 9. Were the methods used to combine the findings of studies appropriate? 10. Was the likelihood of publication bias assessed? 11. Was the conflict of interest stated? Randomized controlled trials Brown et al Hypothesis/objective, main outcomes to be measured, characteristics of included patients, interventions of interest and the main findings were explicit The study provided mean and standard deviation or percentage for outcomes All important adverse events were reported The actual p-values were reported The recruited subjects represented the entire population within a specified period of time Staff, places and facilities where patients were treated, were representative of the treatment the majority of patients received Attempt was made to blind those measuring the main outcomes of the intervention. Follow-up was the same for all study patients Statistical tests used to assess the main outcomes were appropriate The outcome measures were clearly described Patients in both groups were recruited from the same population, over the same period of time, and were randomized to the interventions groups The trial was powered to detect the difference in primary outcome Myers et al Hypothesis/objective, main outcomes to be measured, characteristics of A list of principal confounders was not provided The characteristics of patients lost to followup were not described Attempt was made to blind study subjects (not applicable in this study) Unable to determine compliance with the interventions Random assignment was not concealed from staff Unable to determine if there was adequate adjustment for confounding in the analyses, or losses of patients to follow-up were taken into account A list of principal confounders was not provided Mercury versus Non-Mercury Sphygmomanometers 14

15 First Author, Publication Year Strengths included patients, interventions of interest and the main findings were explicit The study provided mean differences and 95% confidence intervals for outcomes The actual p-values were reported The recruited subjects represented the entire population within a specified period of time Staff, places and facilities where patients were treated, were representative of the treatment the majority of patients received Follow-up was the same for all study patients Statistical tests used to assess the main outcomes were appropriate The outcome measures were clearly described Patients in both groups were recruited from the same population, over the same period of time, and were randomized to the interventions groups The trial was powered to detect the difference in primary outcome Lamarre-Cliché et al. Hypothesis/objective, main outcomes to be measured, characteristics of included patients, interventions of interest and the main findings were explicit The study provided mean differences and 95% confidence intervals for outcomes The actual p-values were reported The recruited subjects represented the entire population within a specified period of time Staff, places and facilities where patients were treated, were representative of the treatment the majority of patients received Follow-up was the same for all study patients Statistical tests used to assess the main outcomes were appropriate The outcome measures were clearly described Patients in both groups were recruited from the same population, over the same period of time, and were randomized to the interventions groups The trial was powered to detect the Guidelines National Institute for Health and Clinical difference in primary outcome Objectives and target patients population were explicit. Limitations All important adverse events were not reported The characteristics of patients lost to followup were not described No attempt was made to blind study subjects (not applicable in this study) No attempt was made to blind those measuring the main outcomes of the intervention. Unable to determine compliance with the interventions Random assignment was not concealed from staff Unable to determine if there was adequate adjustment for confounding in the analyses, or losses of patients to follow-up were taken into account A list of principal confounders was not provided All important adverse events were not reported The characteristics of patients lost to followup were not described No attempt was made to blind study subjects (not applicable in this study) No attempt was made to blind those measuring the main outcomes of the intervention. Unable to determine compliance with the interventions Random assignment was not concealed from staff Unable to determine if there was adequate adjustment for confounding in the analyses, or losses of patients to follow-up were taken into account The guidelines are not supported with application tools Mercury versus Non-Mercury Sphygmomanometers 15

16 First Author, Publication Year Excellence (NICE), UK, Cincinnati Children s Hospital Medical Center. Best evidence statement, Society of Obstetric Medicine of Australia and New Zealand (SOMANZ) Society of Obstetricians and Gynaecologists of Canada Magee et al Strengths A procedure for updating the guideline was provided The recommendations were specific and unambiguous Key recommendations were easily identified The editorial independence of the guidelines was specified. Objectives and target patients population were explicit. The recommendations were specific and unambiguous Key recommendations were easily identified Objectives and target patients population were explicit. The recommendations were specific and unambiguous Objectives and target patients population were explicit. The recommendations were specific and unambiguous Key recommendations were easily identified Formulation of the recommendations was based on clinical evidence Limitations The cost implication of applying these guidelines was not evaluated and reported. The recommendations were based on consensus of appraisal committee members, who considered evidence from a number of sources including systematic reviews, assessment reports and manufacturers documents. Patient s preferences and views were not explicitly taken into consideration. Methods used for searching the evidence were not reported. The guidelines are not supported with application tools A procedure for updating the guideline was not provided The cost implication of applying these guidelines was not evaluated and reported. Formulation of the recommendations was based on expert consensus. Patient s preferences and views were not explicitly taken into consideration. Methods used for searching the evidence were not reported. The editorial independence of the guidelines was not specified. The guidelines are not supported with application tools A procedure for updating the guideline was not provided The cost implication of applying these guidelines was not evaluated and reported. Formulation of the recommendations was based on literature reviews and expert consensus. Patient s preferences and views were not explicitly taken into consideration. Methods used for searching the evidence were not reported. The editorial independence of the guidelines was not specified. Key recommendations were not easily identified The guidelines are not supported with application tools A procedure for updating the guideline was not provided The cost implication of applying these guidelines was not evaluated and reported. Patient s preferences and views were not explicitly taken into consideration. Methods used for searching the evidence were not reported. The editorial independence of the guidelines was not specified. Financial disclosures and conflict of interest were not stated. Mercury versus Non-Mercury Sphygmomanometers 16

17 APPENDIX 5: Main Study Findings and Authors Conclusions Clinical First Author, Publication Year, Country Main Findings Systematic Review Stergiou et al. Correlation coefficient between automated and manual SBP/DBP (4 studies): For SBP: r=0.89, 95% CI 0.84, 0.94 For DBP: r=0.76, 95% CI 0.70, 0.81, p<0.001 for difference Average blood pressure differences between automated and manual method (6 studies): SBP difference = 0.5 mm Hg, 95% CI -0.9, 1.9 DBP difference = 2.5 mm Hg, 95% CI -0.6, 5.7 Absolute blood pressure differences within 5 mm Hg Four of eight studies reporting data regarding the within 5 mm Hg ESH-IP criterion have passed for SBP (at least 65% of comparisons within 5 mm Hg) and only two for DBP Impact of heart rate Heart rate had no impact on the blood pressure differences between automated and manual measurements Authors Conclusions: this analysis suggests that the currently available automated blood pressure monitors that have been validated in individuals with sinus rhythm appear to be accurate in measuring systolic blood pressure but not diastolic blood pressure in individuals with sustained atrial fibrillation p.8 Skirton et al Study Intervention vs. control Population Main findings UK Crosssectional survey Crosssectional Prospective Crosssectional study Crosssectional Retrospective crosssectional study Prospective cohort study Crosssectional quasi- Automated (BpTRU) vs. manual (mercury) Automated (Dinamap) vs. manual (mercury) Automated (Dinamap) vs. manual (mercury) Automated BP wrist monitor (Omron Rx) vs. manual (mercury) Automated (Dinamap) vs. manual (mercury) Automated (Dinamap) vs. manual (mercury) Automated (Omron R3 and Visomat OZ2) vs. manual (mercury) 50 patients in a hypertension or internal medicine clinic 47 hospital patients 171 patients in emergency department Results suggested that the automated device should replace manual instruments in clinical practice and research Dinamap demonstrably less accurate than mercury sphygmomanometer, especially systolic pressure (p<0.05) The automated device consistently overestimated systolic BP and underestimated diastolic BP 85 participants Wrist device readings are more closely correlated with mercury sphygmomanometer readings using appropriately sized cuffs. 388 trauma patients 85 participants (outpatients or staff of a large teaching hospital) 85 patients attending a medical Manual BP readings were more reflective of patient condition. Automated BP measurements were higher than manual readings, especially in hypotensive patients The automated device is acceptable for clinical use, according to BHS criteria All three types of devices showed good reproducibility. Automated devices performed Mercury versus Non-Mercury Sphygmomanometers 17

18 First Author, Publication Year, Country Main Findings experimental outpatient clinic less well than mercury sphygmomanometer in hypertensive patients. Measuring BP on the wrist using automated devices was not found to be satisfactory Crosssectional Post-hoc analysis of observational Prospective cohort Prospective cohort Prospective crosssectional Prospective crosssectional Prospective cohort aneroid and oscillometric vs. manual (mercury) Automated (OMRON 705 CP) vs. manual (mercury) Automated (Dinamap Pro-care 400) vs. manual (mercury) Automated (TM-2564G) vs. sphygmomanometer aneroid Welch Allyn Tycos 767 vs. manual (mercury) Automated (BpTRU) vs. manual (mercury) Automated (Dinamap Pro-care) vs. manual (mercury) 500 participants from Hypertension Center, Preventive Medicine Research Center and local fairs 313 untreated hypertensive patients 300 patients at one tertiary care teaching hospital 85 inpatients from one hospital 997 participants at high risk of diabetes 238 Ontario residents Oscillometric device slightly overestimated systolic BP and underestimated diastolic BP compared with mercury sphygmomanometer. Oscillometric device was comparable in performance to the aneroid device, and both could be used to replace the mercury manometer There are differences between readings obtained by both devices. Authors state reasonable good correlation between the two devices for systolic and diastolic pressure. Some observer bias noted when mercury manometer was used notably repeated measure and terminal digit bias. Interoperator measurements were more consistent for the automated device than intraoperator measurements using the mercury device Diastolic readings taken with the device did not differ more than 5 mmhg from those taken with the sphygmomanometer. 99.2% of the systolic readings taken with the device were within 5 mmhg of sphygmomanometer There were no clinically significant differences between readings according to the type of sphygmomanometer Mean BP taken with automated device was lower than that taken with mercury sphygmomanometer. Authors conclude the BpTRU able to be used because it reduced white coat effect 33 outpatients 79% of measurements with device were within 5 mm Hg of the systolic reading taken with Mercury versus Non-Mercury Sphygmomanometers 18

19 First Author, Publication Year, Country Main Findings Prospective Randomized controlled trial Automated (OMRON) vs. sphygmomanometer Automated (IVAC 4200) vs. manual (mercury) 907 hypertensive patients 145 stable inpatients mercury, while 77% of diastolic readings were within 5 mmhg. The device is acceptable for clinical use, according to the Advancement of Medical Instrumentation 99% readings by traditional nurses ended in zero. Using a level of plus and minus 5 mm Hg, traditional nurse over estimated systolic BP in 45% cases and underestimated systolic BP in 26.5% cases. Automated device overestimated systolic BP in 51.3% cases and underestimated systolic BP in 12% cases IVAC measurements are less than satisfactory in both systolic and diastolic BP. Arrhythmia or low K5 reduces IVAC accuracy. IVAC should not be used for patients with hypertension Authors Conclusions: There are situation where the substitution of oscillometric for auscultatory devices could have particularly serious repercussions for the patient, such as when the patient is either hypertensive or hypotensive. However, further research is required on the use of aneroid sphygmomanometers as a replacement for mercury devices p.602 obtained by six sequential recordings (alternating between mercury and Randomized controlled trials Brown et al. Blood pressures automated) Mercury (n=110) Automated (n=110) Average BP using mercury manometer (mm Hg) 139 (9) / 90 (7) 137 (9) / 91 (7) Average BP using automated device (m Hg) 136 (12) / 86 (8) 137 (11) / 87 (8) Difference in average BP (mm Hg) 2 (8) / 3 (6) 1 (7) / 5 (5) Average SBP difference between methods (%) 5 mm Hg mm Hg mm Hg Average DBP difference between methods (%) 5 mm Hg mm Hg mm Hg Maternal and fetal outcomes were similar between groups. Mercury Automated p- value Gestation at last antenatal visit (weeks) 37 (2) 37 (2) 0.1 Gestation at birth (weeks) 38 (3) 37 (3) 0.08 Diagnosis at birth (%) Essential hypertension Mercury versus Non-Mercury Sphygmomanometers 19

TITLE: Delivery of Electroconvulsive Therapy in Non-Hospital Settings: A Review of the Safety and Guidelines

TITLE: Delivery of Electroconvulsive Therapy in Non-Hospital Settings: A Review of the Safety and Guidelines TITLE: Delivery of Electroconvulsive Therapy in Non-Hospital Settings: A Review of the Safety and Guidelines DATE: 08 May 2014 CONTEXT AND POLICY ISSUES Electroconvulsive therapy (ECT) is a treatment that

More information

TITLE: Montelukast for Sleep Apnea: A Review of the Clinical Effectiveness, Cost Effectiveness, and Guidelines

TITLE: Montelukast for Sleep Apnea: A Review of the Clinical Effectiveness, Cost Effectiveness, and Guidelines TITLE: Montelukast for Sleep Apnea: A Review of the Clinical Effectiveness, Cost Effectiveness, and Guidelines DATE: 17 January 2014 CONTEXT AND POLICY ISSUES Obstructive sleep apnea (OSA) is a common

More information

DATE: 04 April 2012 CONTEXT AND POLICY ISSUES

DATE: 04 April 2012 CONTEXT AND POLICY ISSUES TITLE: Procedure Site Bleeding Complications Following Percutaneous Coronary Interventions or Angioplasty: A Review of Clinical Evidence and Guidelines DATE: 04 April 2012 CONTEXT AND POLICY ISSUES Percutaneous

More information

DATE: 17 July 2012 CONTEXT AND POLICY ISSUES

DATE: 17 July 2012 CONTEXT AND POLICY ISSUES TITLE: Sterile Pre-filled Saline Syringes for Acute Care Patients: A Review of Clinical Evidence, Cost-effectiveness, Evidence-based Guidelines, and Safety DATE: 17 July 2012 CONTEXT AND POLICY ISSUES

More information

DR JIRAR TOPOUCHIAN PROF PAROUNAK ZELVEIAN PROF ROLAND ASMAR. September 8 th, Principal Investigator and Study Chair:

DR JIRAR TOPOUCHIAN PROF PAROUNAK ZELVEIAN PROF ROLAND ASMAR. September 8 th, Principal Investigator and Study Chair: Page 1 of 10 Validation of the automatic blood pressure measurements device, the OMRON EVOLV (HEM- 7600 T-E) in Pregnancy according to the Modified European Society of Hypertension International Protocol

More information

Cardiopulmonary Resuscitation Feedback Devices for Adult Patients in Cardiac Arrest: A Review of Clinical Effectiveness and Guidelines

Cardiopulmonary Resuscitation Feedback Devices for Adult Patients in Cardiac Arrest: A Review of Clinical Effectiveness and Guidelines TITLE: Cardiopulmonary Resuscitation Feedback Devices for Adult Patients in Cardiac Arrest: A Review of Clinical Effectiveness and Guidelines DATE: 20 April 2015 CONTEXT AND POLICY ISSUES The incidence

More information

This report will provide a review on the comparative clinical effectiveness and safety between intranasal triamcinolone and beclomethasone.

This report will provide a review on the comparative clinical effectiveness and safety between intranasal triamcinolone and beclomethasone. TITLE: Intranasal Triamcinolone versus Intranasal Beclomethasone for Acute and Chronic Sinus Inflammation: A Review of Comparative Clinical Effectiveness and Safety DATE: 29 January 2013 CONTEXT AND POLICY

More information

TITLE: Acetylsalicylic Acid for Venous Thromboembolism Prophylaxis: A Review of Clinical Evidence, Benefits and Harms

TITLE: Acetylsalicylic Acid for Venous Thromboembolism Prophylaxis: A Review of Clinical Evidence, Benefits and Harms TITLE: Acetylsalicylic Acid for Venous Thromboembolism Prophylaxis: A Review of Clinical Evidence, Benefits and Harms DATE: 23 August 2011 CONTEXT AND POLICY ISSUES: Thromboembolism occurs when a blood

More information

DATE: 28 May 2015 CONTEXT AND POLICY ISSUES

DATE: 28 May 2015 CONTEXT AND POLICY ISSUES TITLE: Neonatal Vitamin K Administration for the Prevention of Hemorrhagic Disease: A Review of the Clinical Effectiveness, Comparative Effectiveness, and Guidelines DATE: 28 May 2015 CONTEXT AND POLICY

More information

TITLE: Immediate Osseointegrated Implants for Cancer Patients: A Review of Clinical and Cost-Effectiveness

TITLE: Immediate Osseointegrated Implants for Cancer Patients: A Review of Clinical and Cost-Effectiveness TITLE: Immediate Osseointegrated Implants for Cancer Patients: A Review of Clinical and Cost-Effectiveness DATE: 13 January 2015 CONTEXT AND POLICY ISSUES According to the World health Organization, the

More information

TITLE: Patient-Controlled Analgesia for Acute Injury Transfers: A Review of the Clinical Effectiveness, Safety, and Guidelines

TITLE: Patient-Controlled Analgesia for Acute Injury Transfers: A Review of the Clinical Effectiveness, Safety, and Guidelines TITLE: Patient-Controlled Analgesia for Acute Injury Transfers: A Review of the Clinical Effectiveness, Safety, and Guidelines DATE: 11 August 2014 CONTEXT AND POLICY ISSUES Patient-controlled analgesia

More information

and bias, which are known to be present in self-home and in professional office BP measurements taken using the auscultatory technique [7].

and bias, which are known to be present in self-home and in professional office BP measurements taken using the auscultatory technique [7]. Devices and technology 37 Validation of the A&D UM-11 professional hybrid device for office blood pressure measurement according to the International Protocol George S. Stergiou, Periklis P. Giovas, Charilaos

More information

TITLE: Optimal Oxygen Saturation Range for Adults Suffering from Traumatic Brain Injury: A Review of Patient Benefit, Harms, and Guidelines

TITLE: Optimal Oxygen Saturation Range for Adults Suffering from Traumatic Brain Injury: A Review of Patient Benefit, Harms, and Guidelines TITLE: Optimal Oxygen Saturation Range for Adults Suffering from Traumatic Brain Injury: A Review of Patient Benefit, Harms, and Guidelines DATE: 11 April 2014 CONTEXT AND POLICY ISSUES Traumatic brain

More information

TITLE: Hyperbaric Oxygen Therapy for Adults with Mental Illness: A Review of the Clinical Effectiveness

TITLE: Hyperbaric Oxygen Therapy for Adults with Mental Illness: A Review of the Clinical Effectiveness TITLE: Hyperbaric Oxygen Therapy for Adults with Mental Illness: A Review of the Clinical Effectiveness DATE: 27 August 2014 CONTEXT AND POLICY ISSUES Mental illness, including major depressive episode,

More information

Ambulatory BP Monitoring: Getting the Diagnosis of Hypertension Right. Anthony J. Viera, MD, MPH, FAHA Professor and Chair

Ambulatory BP Monitoring: Getting the Diagnosis of Hypertension Right. Anthony J. Viera, MD, MPH, FAHA Professor and Chair Ambulatory BP Monitoring: Getting the Diagnosis of Hypertension Right Anthony J. Viera, MD, MPH, FAHA Professor and Chair Objectives Review limitations of office BP in making a correct diagnosis of hypertension

More information

TITLE: Fusidic Acid for Ophthalmic Infections: A Review of Clinical and Cost Effectiveness and Safety

TITLE: Fusidic Acid for Ophthalmic Infections: A Review of Clinical and Cost Effectiveness and Safety TITLE: Fusidic Acid for Ophthalmic Infections: A Review of Clinical and Cost Effectiveness and Safety DATE: 22 February 2013 CONTEXT AND POLICY ISSUES The antibiotic fusidic acid is available in a 1% suspension

More information

& Wilkins. a Division of Cardiology, Schulich Heart Centre, b Institute for Clinical and

& Wilkins. a Division of Cardiology, Schulich Heart Centre, b Institute for Clinical and Original article 333 Optimum frequency of office blood pressure measurement using an automated sphygmomanometer Martin G. Myers a, Miguel Valdivieso a and Alexander Kiss b,c Objective To determine the

More information

TITLE: Shilla and MAGEC Systems for Growing Children with Scoliosis: A Review of the Clinical Benefits and Cost-Effectiveness

TITLE: Shilla and MAGEC Systems for Growing Children with Scoliosis: A Review of the Clinical Benefits and Cost-Effectiveness TITLE: Shilla and MAGEC Systems for Growing Children with Scoliosis: A Review of the Clinical Benefits and Cost-Effectiveness DATE: 17 January 2013 CONTEXT AND POLICY ISSUES Early-onset scoliosis (EOS)

More information

CADTH RAPID RESPONSE REPORT: SUMMARY WITH CRITICAL APPRAISAL

CADTH RAPID RESPONSE REPORT: SUMMARY WITH CRITICAL APPRAISAL CADTH RAPID RESPONSE REPORT: SUMMARY WITH CRITICAL APPRAISAL Day Surgery versus Overnight Stay for Laparoscopic Cholecystectomy and Laparoscopic Hernia Repair: A Review of Comparative Clinical Effectiveness

More information

DATE: 22 May 2013 CONTEXT AND POLICY ISSUES

DATE: 22 May 2013 CONTEXT AND POLICY ISSUES TITLE: Intraperitoneal versus Intravenous Chemotherapy for the Treatment of Ovarian Cancer: A Review of the Comparative Clinical Evidence, Safety, Cost- Effectiveness, and Guidelines DATE: 22 May 2013

More information

Comparison of manual versus automated blood pressure measurement in intensive care unit, coronary care unit, and emergency room.

Comparison of manual versus automated blood pressure measurement in intensive care unit, coronary care unit, and emergency room. Comparison of manual versus automated blood pressure measurement in intensive care unit, coronary care unit, and emergency room Abstract Ahmad Mirdamadi (1), Mostafa Etebari (2) Original Article BACKGROUND:

More information

TITLE: Diagnosing, Screening, and Monitoring Depression in the Elderly: A Review of Guidelines

TITLE: Diagnosing, Screening, and Monitoring Depression in the Elderly: A Review of Guidelines TITLE: Diagnosing, Screening, and Monitoring Depression in the Elderly: A Review of Guidelines DATE: 8 September 2015 CONTEXT AND POLICY ISSUES Depression disorders are one the most common and burdensome

More information

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension

Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home. blood pressure monitoring according to the European Society of Hypertension Validation of the SEJOY BP-1307 upper arm blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010 Short title: Validation

More information

Increasing Frequency of Self-Monitoring Blood Glucose Test Strips During Pregnancy: A Review of the Clinical and Cost- Effectiveness and Guidelines

Increasing Frequency of Self-Monitoring Blood Glucose Test Strips During Pregnancy: A Review of the Clinical and Cost- Effectiveness and Guidelines CADTH RAPID RESPONSE REPORT: SUMMARY WITH CRITICAL APPRAISAL Increasing Frequency of Self-Monitoring Blood Glucose Test Strips During Pregnancy: A Review of the Clinical and Cost- Effectiveness and Guidelines

More information

OHTAC Recommendation: Twenty-Four-Hour Ambulatory Blood Pressure Monitoring in Hypertension. Ontario Health Technology Advisory Committee

OHTAC Recommendation: Twenty-Four-Hour Ambulatory Blood Pressure Monitoring in Hypertension. Ontario Health Technology Advisory Committee OHTAC Recommendation: Twenty-Four-Hour Ambulatory Blood Pressure Monitoring in Hypertension Ontario Health Technology Advisory Committee May 2012 Background Hypertension in Canada Hypertension occurs when

More information

TITLE: Periodic Dental Examinations for Oral Health: A Review of Clinical Effectiveness, Cost Effectiveness, and Guidelines

TITLE: Periodic Dental Examinations for Oral Health: A Review of Clinical Effectiveness, Cost Effectiveness, and Guidelines TITLE: Periodic Dental Examinations for Oral Health: A Review of Clinical Effectiveness, Cost Effectiveness, and Guidelines DATE: 21 May 2014 CONTEXT AND POLICY ISSUES While there is no accepted universal

More information

CADTH RAPID RESPONSE REPORT: REFERENCE LIST Side Effect Free Chemotherapy for the Treatment of Cancer: Clinical Effectiveness

CADTH RAPID RESPONSE REPORT: REFERENCE LIST Side Effect Free Chemotherapy for the Treatment of Cancer: Clinical Effectiveness CADTH RAPID RESPONSE REPORT: REFERENCE LIST Side Effect Free Chemotherapy for the Treatment of Cancer: Clinical Effectiveness Service Line: Rapid Response Service Version: 1.0 Publication Date: August

More information

DATE: 22 June 2015 CONTEXT AND POLICY ISSUES

DATE: 22 June 2015 CONTEXT AND POLICY ISSUES TITLE: Non-Invasive Methods for Diagnosis and Monitoring of Liver Fibrosis in Patients with Chronic Hepatitis B and C: A Review of Diagnostic Accuracy, Clinical Effectiveness, Cost-Effectiveness and Guidelines

More information

TITLE: Use of Bupropion in Patients with Depression and the Associated Risk of Seizures: Safety

TITLE: Use of Bupropion in Patients with Depression and the Associated Risk of Seizures: Safety TITLE: Use of Bupropion in Patients with Depression and the Associated Risk of Seizures: Safety DATE: 1 April 2010 CONTEXT AND POLICY ISSUES: The prevalence of epilepsy ranges from 0.5% to 1%, 1,2 and

More information

TITLE: Long-acting Opioids for Chronic Non-cancer Pain: A Review of the Clinical Efficacy and Safety

TITLE: Long-acting Opioids for Chronic Non-cancer Pain: A Review of the Clinical Efficacy and Safety TITLE: Long-acting Opioids for Chronic Non-cancer Pain: A Review of the Clinical Efficacy and Safety DATE: 27 August 2015 CONTEXT AND POLICY ISSUES Chronic non-cancer pain (CNCP) of neuropathic or proprioceptive

More information

CADTH RAPID RESPONSE REPORT: SUMMARY WITH CRITICAL APPRAISAL

CADTH RAPID RESPONSE REPORT: SUMMARY WITH CRITICAL APPRAISAL CADTH RAPID RESPONSE REPORT: SUMMARY WITH CRITICAL APPRAISAL Central Venous Access Devices (CVADs) and Peripherally Inserted Central Catheters (PICCs) for Adult and Pediatric : A Review of Clinical Effectiveness

More information

DATE: 12 January 2017 CONTEXT AND POLICY ISSUES

DATE: 12 January 2017 CONTEXT AND POLICY ISSUES TITLE: Basal-Bolus Versus Sliding-Scale Insulin Therapy in the Acute Care Hospital Setting: A Review of Comparative Clinical Effectiveness and Cost-Effectiveness DATE: 12 January 2017 CONTEXT AND POLICY

More information

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited.

Copyright Lippincott Williams & Wilkins. Unauthorized reproduction of this article is prohibited. Devices and technology 261 Validation of the ROSSMAX blood pressure measuring monitor according to the European Society of Hypertension International Protocol for Validation of Blood Pressure Measuring

More information

TITLE: Hyperbaric Oxygen Therapy for Autism: A Review of the Clinical and Cost- Effectiveness

TITLE: Hyperbaric Oxygen Therapy for Autism: A Review of the Clinical and Cost- Effectiveness TITLE: Hyperbaric Oxygen Therapy for Autism: A Review of the Clinical and Cost- Effectiveness DATE: 19 November 2009 CONTEXT AND POLICY ISSUES: Autism is a pervasive developmental disorder characterized

More information

1 Department of Medical Physics, Royal Infirmary of Edinburgh, 2 Department of. Received 11 June 2004 Accepted 7 September 2004

1 Department of Medical Physics, Royal Infirmary of Edinburgh, 2 Department of. Received 11 June 2004 Accepted 7 September 2004 Devices and technology 1 Do SpaceLabs ambulatory non-invasive blood pressure recorders measure blood pressure consistently over several years use? John N. Amoore 1, David Dewar 2, Kathleen Gough 3 and

More information

DATE: 29 Aug 2012 CONTEXT AND POLICY ISSUES

DATE: 29 Aug 2012 CONTEXT AND POLICY ISSUES TITLE: Tiotropium Compared with Ipratropium for Patients with Moderate to Severe Chronic Obstructive Pulmonary Disease: A Review of the Clinical Effectiveness DATE: 29 Aug 2012 CONTEXT AND POLICY ISSUES

More information

HOW CONSISTENT ARE THE BLOOD PRESSURE AND PULSE RATE MEASUREMENTS OF THE ELECTRONIC BP APPARATUS AND THE MANUAL SPHYGMOMANOMETER

HOW CONSISTENT ARE THE BLOOD PRESSURE AND PULSE RATE MEASUREMENTS OF THE ELECTRONIC BP APPARATUS AND THE MANUAL SPHYGMOMANOMETER HOW CONSISTENT ARE THE BLOOD PRESSURE AND PULSE RATE MEASUREMENTS OF THE ELECTRONIC BP APPARATUS AND THE MANUAL SPHYGMOMANOMETER Naser KA. Teaching Hospital Peradeniya, Peradeniya, Sri Lanka Zawahir S

More information

AFFORDABLE TECHNOLOGY

AFFORDABLE TECHNOLOGY World Health Organization AFFORDABLE TECHNOLOGY BLOOD PRESSURE MEASURING DEVICES FOR LOW RESOURCE SETTINGS CARDIOVASCULAR DISEASES Blood Pressure Measurement in Low Resource Settings Annex 1: Blood Pressure

More information

24-Hour Ambulatory Blood Pressure Monitoring

24-Hour Ambulatory Blood Pressure Monitoring No. 15 Jan 2003 Before CCO decides to undertake a health technology assessment, a pre-assessment of the literature is performed. Pre-assessments are based on a limited literature search; they are not extensive,

More information

TITLE: Nutritional Supplementation for Patients with Cancer: A Review of the Clinical Effectiveness and Guidelines

TITLE: Nutritional Supplementation for Patients with Cancer: A Review of the Clinical Effectiveness and Guidelines TITLE: Nutritional Supplementation for Patients with Cancer: A Review of the Clinical Effectiveness and Guidelines DATE: 12 February 2014 CONTEXT AND POLICY ISSUES People with cancer who are well nourished

More information

TITLE: Crushed Buprenorphine or Buprenorphine-Naloxone for Opioid Dependency: A Review of the Clinical Effectiveness and Guidelines

TITLE: Crushed Buprenorphine or Buprenorphine-Naloxone for Opioid Dependency: A Review of the Clinical Effectiveness and Guidelines TITLE: Crushed Buprenorphine or Buprenorphine-Naloxone for Opioid Dependency: A Review of the Clinical Effectiveness and Guidelines DATE: 18 July 2016 CONTEXT AND POLICY ISSUES Opioid dependence is a chronic

More information

Assessing Blood Pressure for Clinical Research: Pearls & Pitfalls

Assessing Blood Pressure for Clinical Research: Pearls & Pitfalls Assessing Blood Pressure for Clinical Research: Pearls & Pitfalls Anthony J. Viera, MD, MPH, FAHA Department of Family Medicine Hypertension Research Program UNC School of Medicine Objectives Review limitations

More information

DATE: 07 August 2012 CONTEXT AND POLICY ISSUES

DATE: 07 August 2012 CONTEXT AND POLICY ISSUES TITLE: Adaptive Servo Ventilation versus Continuous or Bi-Level Positive Airway Pressure: A Review of the Clinical Effectiveness, Cost-Effectiveness and Guidelines DATE: 07 August 2012 CONTEXT AND POLICY

More information

DATE: 10 February 2016 CONTEXT AND POLICY ISSUES

DATE: 10 February 2016 CONTEXT AND POLICY ISSUES TITLE: Neuromuscular Monitoring for Patients Receiving Continuous Paralytic or Neuromuscular Blocking Agents: A Review of the Clinical Effectiveness and Guidelines DATE: 10 February 2016 CONTEXT AND POLICY

More information

DATE: 26 August 2013 CONTEXT AND POLICY ISSUES

DATE: 26 August 2013 CONTEXT AND POLICY ISSUES TITLE: Nd:YAG Laser for the Treatment of Patients with Hidradenitis Suppurativa: A Review of the Clinical Effectiveness, Cost-Effectiveness, and Safety DATE: 26 August 2013 CONTEXT AND POLICY ISSUES Hidradenitis

More information

TITLE: Discontinuation of Benzodiazepines and Other Sedative-Hypnotic Sleep Medication in Hospitalized Patients: Clinical Evidence and Guidelines

TITLE: Discontinuation of Benzodiazepines and Other Sedative-Hypnotic Sleep Medication in Hospitalized Patients: Clinical Evidence and Guidelines TITLE: Discontinuation of Benzodiazepines and Other Sedative-Hypnotic Sleep Medication in Hospitalized Patients: Clinical Evidence and Guidelines DATE: 25 September 2015 RESEARCH QUESTIONS 1. What is the

More information

TITLE: Mattresses for Chronic Back or Neck Pain: A Review of the Clinical Effectiveness and Guidelines

TITLE: Mattresses for Chronic Back or Neck Pain: A Review of the Clinical Effectiveness and Guidelines TITLE: Mattresses for Chronic Back or Neck Pain: A Review of the Clinical Effectiveness and Guidelines DATE: 14 May 2014 CONTEXT AND POLICY ISSUES Persistent back pain is a common presentation in primary

More information

Service Line: Rapid Response Service Version: 1.0 Publication Date: April 4, 2018 Report Length: 14 Pages

Service Line: Rapid Response Service Version: 1.0 Publication Date: April 4, 2018 Report Length: 14 Pages CADTH RAPID RESPONSE REPORT: SUMMARY WITH CRITICAL APPRAISAL Chest X-Rays Around Placement in Long-Term Care Facilities: A Review of Clinical Utility and Guidelines Service Line: Rapid Response Service

More information

TITLE: Optimal Care of Chronic, Non-Healing, Lower Extremity Wounds: A Review of Clinical Evidence and Guidelines

TITLE: Optimal Care of Chronic, Non-Healing, Lower Extremity Wounds: A Review of Clinical Evidence and Guidelines TITLE: Optimal Care of Chronic, Non-Healing, Lower Extremity Wounds: A Review of Clinical Evidence and Guidelines DATE: 17 December 2013 CONTEXT AND POLICY ISSUES Chronic wounds are those which do not

More information

Service Line: Rapid Response Service Version: 1.0 Publication Date: October 30, 2018 Report Length: 7 Pages

Service Line: Rapid Response Service Version: 1.0 Publication Date: October 30, 2018 Report Length: 7 Pages CADTH RAPID RESPONSE REPORT: SUMMARY WITH CRITICAL APPRAISAL Topical Silver Nitrate for the Management of Hemostasis: A Review of Clinical Effectiveness, Cost- Effectiveness, and Guidelines Service Line:

More information

TITLE: Antimicrobial Ointments for Patients Undergoing Hemodialysis: A Review of Evidence-Based Guidelines

TITLE: Antimicrobial Ointments for Patients Undergoing Hemodialysis: A Review of Evidence-Based Guidelines TITLE: Antimicrobial Ointments for Patients Undergoing Hemodialysis: A Review of Evidence-Based Guidelines DATE: 20 June 2013 CONTEXT AND POLICY ISSUES Vascular access is an important concern in patients

More information

CADTH Therapeutic Review

CADTH Therapeutic Review Canadian Agency for Drugs and Technologies in Health Agence canadienne des médicaments et des technologies de la santé CADTH Therapeutic Review August 2012 Volume 1, Issue 1A Antithrombotic Therapy for

More information

TITLE: Tranexamic Acid for the Management of Bleeding: A Review of the Clinical Effectiveness and Guidelines

TITLE: Tranexamic Acid for the Management of Bleeding: A Review of the Clinical Effectiveness and Guidelines TITLE: Tranexamic Acid for the Management of Bleeding: A Review of the Clinical Effectiveness and Guidelines DATE: 24 June 2013 CONTEXT AND POLICY ISSUES Bleeding or hemorrhage indicates a blood loss from

More information

Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents

Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents Original article http://dx.doi.org/10.3345/kjp.2011.54.11.463 Korean J Pediatr 2011;54(11):463-469 Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents

More information

1. Department of Gynecology and Obstetrics, St. Joseph's Hospital Berlin Tempelhof, Germany

1. Department of Gynecology and Obstetrics, St. Joseph's Hospital Berlin Tempelhof, Germany Page 1 of 9 Validation of the TONOPORT VI ambulatory blood pressure monitor, according to the European Society of Hypertension International Protocol revision 2010 Michael Abou-Dakn 1, Cornelius Döhmen

More information

External Oscillatory Blood Pressure - EOBPTM

External Oscillatory Blood Pressure - EOBPTM External Oscillatory Blood Pressure - EOBPTM Development of Novel Principle To Measure Blood Pressure Mindaugas Pranevicius, M.D., Osvaldas Pranevicius, M.D., Ph.D. Pranevicius Biotech Inc., Forest Hills,

More information

Evidence Supporting the Recommendations. Implementation of the Guideline. Benefits/Harms of Implementing the Guideline Recommendations

Evidence Supporting the Recommendations. Implementation of the Guideline. Benefits/Harms of Implementing the Guideline Recommendations 1 / 6 Surgical safety checklist in obstetrics and gynaecology Guideline Developer(s) Society of Obstetricians and Gynaecologists of Canada Date Released 2013 Jan Full Text Guideline Surgical safety checklist

More information

Received 10 April 2008 Revised 23 June 2008 Accepted 24 June 2008

Received 10 April 2008 Revised 23 June 2008 Accepted 24 June 2008 342 Original article Development of an accurate oscillometric blood pressure device for low resource settings Annemarie de Greeff, Hannah Nathan, Nina Stafford, Bing Liu and Andrew H. Shennan Objective

More information

TITLE: Dose of Electrical Current for Biphasic Defibrillators for Synchronous Cardioversion in Patients with Tachyarrhythmia: Guidelines

TITLE: Dose of Electrical Current for Biphasic Defibrillators for Synchronous Cardioversion in Patients with Tachyarrhythmia: Guidelines TITLE: Dose of Electrical Current for Biphasic Defibrillators for Synchronous Cardioversion in Patients with Tachyarrhythmia: Guidelines DATE: 5 May 2009 RESEARCH QUESTION: What recommendations exist for

More information

TITLE: Type of Thermometer for Inpatients: Clinical-Effectiveness and Guidelines

TITLE: Type of Thermometer for Inpatients: Clinical-Effectiveness and Guidelines TITLE: Type of Thermometer for Inpatients: Clinical-Effectiveness and Guidelines DATE: 28 May 2009 RESEARCH QUESTIONS: 1. What is the clinical-effectiveness of tympanic, temporal, oral, and rectal thermometers

More information

Clinical Review Report (Sample)

Clinical Review Report (Sample) CADTH COMMON DRUG REVIEW Clinical Review Report (Sample) GENERIC DRUG NAME (BRAND NAME) (Manufacturer) Indication: Text Disclaimer: The information in this document is intended to help Canadian health

More information

TITLE: Newborn Eye Prophylaxis: A Review of Clinical Effectiveness and Guidelines

TITLE: Newborn Eye Prophylaxis: A Review of Clinical Effectiveness and Guidelines TITLE: Newborn Eye Prophylaxis: A Review of Clinical Effectiveness and Guidelines DATE: 03 May 2016 CONTEXT AND POLICY ISSUES Ophthalmia neonatorum (ON) is a type of conjunctivitis that occurs in up to

More information

DATE: 09 December 2009 CONTEXT AND POLICY ISSUES:

DATE: 09 December 2009 CONTEXT AND POLICY ISSUES: TITLE: Tiotropium Compared with Ipratropium for Patients with Moderate to Severe Chronic Obstructive Pulmonary Disease: A Review of the Clinical Effectiveness DATE: 09 December 2009 CONTEXT AND POLICY

More information

TITLE: Critical Incident Stress Debriefing for First Responders: A Review of the Clinical Benefit and Harm

TITLE: Critical Incident Stress Debriefing for First Responders: A Review of the Clinical Benefit and Harm TITLE: Critical Incident Stress Debriefing for First Responders: A Review of the Clinical Benefit and Harm DATE: 12 February 2010 CONTEXT AND POLICY ISSUES: Critical incident stress debriefing (CISD) is

More information

KDIGO Controversies Conference on Blood Pressure in CKD

KDIGO Controversies Conference on Blood Pressure in CKD KDIGO Controversies Conference on Blood Pressure in CKD September 7-10, 2017 Edinburgh, Scotland Kidney Disease: Improving Global Outcomes (KDIGO) is an international organization whose mission is to improve

More information

DEVICES AND TECHNOLOGY

DEVICES AND TECHNOLOGY DEVICES AND TECHNOLOGY Validation of a new algorithm for the BPM-100 electronic oscillometric office blood pressure monitor James M. Wright ab, Gurdial S. Mattu a, Thomas L. Perry Jr ab, Mark E. Gelfer

More information

A GUIDE TO OUR NIBP TECHNOLOGY

A GUIDE TO OUR NIBP TECHNOLOGY GE Healthcare THE DINAMAP DIFFERENCE A GUIDE TO OUR NIBP TECHNOLOGY OUR TECHNOLOGICAL ADVANTAGES THE OSCILLOMETRIC METHODOLOGY Oscillometry is the most commonly used means of indirect blood pressure measurement

More information

GRADE. Grading of Recommendations Assessment, Development and Evaluation. British Association of Dermatologists April 2018

GRADE. Grading of Recommendations Assessment, Development and Evaluation. British Association of Dermatologists April 2018 GRADE Grading of Recommendations Assessment, Development and Evaluation British Association of Dermatologists April 2018 Previous grading system Level of evidence Strength of recommendation Level of evidence

More information

TITLE: Australian Sheepskins for the Management of Pressure Ulcers: A Review of the Clinical-Effectiveness, Cost-Effectiveness, and Guidelines

TITLE: Australian Sheepskins for the Management of Pressure Ulcers: A Review of the Clinical-Effectiveness, Cost-Effectiveness, and Guidelines TITLE: Australian Sheepskins for the Management of Pressure Ulcers: A Review of the Clinical-Effectiveness, Cost-Effectiveness, and Guidelines DATE: 21 July 2009 CONTEXT AND POLICY ISSUES: Pressure ulcers,

More information

Systematic reviews and meta-analyses of observational studies (MOOSE): Checklist.

Systematic reviews and meta-analyses of observational studies (MOOSE): Checklist. Systematic reviews and meta-analyses of observational studies (MOOSE): Checklist. MOOSE Checklist Infliximab reduces hospitalizations and surgery interventions in patients with inflammatory bowel disease:

More information

Overview and Comparisons of Risk of Bias and Strength of Evidence Assessment Tools: Opportunities and Challenges of Application in Developing DRIs

Overview and Comparisons of Risk of Bias and Strength of Evidence Assessment Tools: Opportunities and Challenges of Application in Developing DRIs Workshop on Guiding Principles for the Inclusion of Chronic Diseases Endpoints in Future Dietary Reference Intakes (DRIs) Overview and Comparisons of Risk of Bias and Strength of Evidence Assessment Tools:

More information

TITLE: Olopatadine for the Treatment of Allergic Conjunctivitis: A Review of the Clinical Efficacy, Safety, and Cost-Effectiveness

TITLE: Olopatadine for the Treatment of Allergic Conjunctivitis: A Review of the Clinical Efficacy, Safety, and Cost-Effectiveness TITLE: Olopatadine for the Treatment of Allergic Conjunctivitis: A Review of the Clinical Efficacy, Safety, and Cost-Effectiveness DATE: 17 March 2016 CONTEXT AND POLICY ISSUES Allergic conjunctivitis,

More information

STUDIES INVOLVING THE NON-INVASIVE MEASUREMENT OF BLOOD PRESSURE IN THE ARM

STUDIES INVOLVING THE NON-INVASIVE MEASUREMENT OF BLOOD PRESSURE IN THE ARM STUDIES INVOLVING THE NON-INVASIVE MEASUREMENT OF BLOOD PRESSURE IN THE ARM 1. SCOPE Blood pressure is a vital sign that is used in many clinical situations, and assesses the pressure of blood as it flows

More information

TITLE: Gabapentin for HIV-associated Neuropathic Pain: A Review of the Clinical Effectiveness

TITLE: Gabapentin for HIV-associated Neuropathic Pain: A Review of the Clinical Effectiveness TITLE: Gabapentin for HIV-associated Neuropathic Pain: A Review of the Clinical Effectiveness DATE: 22 January 2016 CONTEXT AND POLICY ISSUES Neuropathic pain can be severe and debilitating and results

More information

CADTH RAPID RESPONSE REPORT: SUMMARY WITH CRITICAL APPRAISAL Nabilone for Chronic Pain Management: A Review of Clinical Effectiveness and Guidelines

CADTH RAPID RESPONSE REPORT: SUMMARY WITH CRITICAL APPRAISAL Nabilone for Chronic Pain Management: A Review of Clinical Effectiveness and Guidelines CADTH RAPID RESPONSE REPORT: SUMMARY WITH CRITICAL APPRAISAL Nabilone for Chronic Pain Management: A Review of Clinical Effectiveness and Guidelines Service Line: Rapid Response Service Version: 1.0 Publication

More information

Home blood pressure measurement with oscillometric upper-arm devices

Home blood pressure measurement with oscillometric upper-arm devices REVIEW Home blood pressure measurement with oscillometric upper-arm devices R.L. Braam, Th. Thien *# Department of Internal Medicine, University Medical Centre St Radboud, PO Box 9101, 6500 HB Nijmegen,

More information

Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension

Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension Validation of the OMRON 705 IT blood pressure measuring device according to the International Protocol of the European Society of Hypertension M. EL ASSAAD, J. TOPOUCHIAN, G. LABAKI, R. ASMAR L Institut

More information

Diagnostic accuracy of a home blood pressure monitor to detect atrial fibrillation

Diagnostic accuracy of a home blood pressure monitor to detect atrial fibrillation (2009), 1 5 & 2009 Macmillan Publishers Limited All rights reserved 0950-9240/09 $32.00 www.nature.com/jhh ORIGINAL ARTICLE Diagnostic accuracy of a home blood pressure monitor to detect atrial fibrillation

More information

DATE: 24 January 2014 CONTEXT AND POLICY ISSUES

DATE: 24 January 2014 CONTEXT AND POLICY ISSUES TITLE: Oncotype DX in Women and Men with ER-Positive HER2-Negative Early Stage Breast Cancer Who are Lymph Node-Positive: A Review of Clinical Effectiveness and Guidelines DATE: 24 January 2014 CONTEXT

More information

Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods

Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods CLINICAL STUDIES Blood Pressure Measurement During Pregnancy: Auscultatory Versus Oscillatory Methods Linda A. Green, RN, MS, Robin D. Froman, RN, PhD, FAAN Objective: To determine the equivalence of auscultatory

More information

TITLE: Naltrexone for the Treatment of Alcohol Dependence in Individuals with Co- Dependencies: A Review of the Clinical Effectiveness

TITLE: Naltrexone for the Treatment of Alcohol Dependence in Individuals with Co- Dependencies: A Review of the Clinical Effectiveness TITLE: Naltrexone for the Treatment of Alcohol Dependence in Individuals with Co- Dependencies: A Review of the Clinical Effectiveness DATE: 08 October 2009 CONTEXT AND POLICY ISSUES: Poly-drug abuse is

More information

TITLE: Left Atrial Appendage Occlusion: Economic Impact and Existing HTA Recommendations

TITLE: Left Atrial Appendage Occlusion: Economic Impact and Existing HTA Recommendations TITLE: Left Atrial Appendage Occlusion: Economic Impact and Existing HTA Recommendations DATE: 29 September 2010 CONTEXT AND POLICY ISSUES: Atrial fibrillation (AF) is the irregular, rapid beating of the

More information

a Medical Physics Department, Guy s & St Thomas NHS Foundation Trust and b King s College School of Medicine, St Thomas Campus, London, UK

a Medical Physics Department, Guy s & St Thomas NHS Foundation Trust and b King s College School of Medicine, St Thomas Campus, London, UK Devices and technology 27 Validation of the Omron 705IT (HEM-759-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol Andrew Coleman a, Paul Freeman

More information

Managing Blood Pressure: It Takes a Team

Managing Blood Pressure: It Takes a Team Managing Blood Pressure: It Takes a Team Module Two How to Take Blood Pressure October 2011 MODULE TWO: How to Take Blood Pressure Module Two: HOW TO TAKE BLOOD PRESSURE Objectives Demonstrate proper technique

More information

RESEARCH INTRODUCTION

RESEARCH INTRODUCTION Conventional versus automated measurement of blood pressure in primary care patients with systolic hypertension: randomised parallel design controlled trial Martin G Myers, professor of medicine, 1 Marshall

More information

No. 1 cause of invalidity No. 2 cause of dementia No. 3 cause of death

No. 1 cause of invalidity No. 2 cause of dementia No. 3 cause of death STROKE statistics No. 1 cause of invalidity No. 2 cause of dementia No. 3 cause of death Every 6 seconds a person in the world has a STROKE regardless of age and gender. Worldwide 15 million people suffer

More information

Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation

Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation ORIGINAL RESEARCH Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation Tyler S Lamb BSc Amar Thakrar MD, FRCPC Mahua Ghosh MD, PhD Merne P Wilson MSc Thomas W Wilson

More information

Health Interventions in Ambulatory Cancer Care Centres DRAFT. Objectives. Methods

Health Interventions in Ambulatory Cancer Care Centres DRAFT. Objectives. Methods ENVIRONMENTAL SCAN Health Interventions in Ambulatory Cancer Care Centres Context Cancer, a complex, chronic condition, will affect an estimated two in five Canadians in their lifetime. 1 Cancer requires

More information

Service Line: Rapid Response Service Version: 1.0 Publication Date: January 21, 2019 Report Length: 5 Pages

Service Line: Rapid Response Service Version: 1.0 Publication Date: January 21, 2019 Report Length: 5 Pages CADTH RAPID RESPONSE REPORT: REFERENCE LIST Topical Cantharidin/ Salicylic Acid/ Podophyllin for the Treatment of Warts: Clinical Effectiveness and Guidelines Service Line: Rapid Response Service Version:

More information

Controlled Trials. Spyros Kitsiou, PhD

Controlled Trials. Spyros Kitsiou, PhD Assessing Risk of Bias in Randomized Controlled Trials Spyros Kitsiou, PhD Assistant Professor Department of Biomedical and Health Information Sciences College of Applied Health Sciences University of

More information

The measurement of blood pressure and hypertension. Handout Fenyvesi Tamás III.Department of Medicine

The measurement of blood pressure and hypertension. Handout Fenyvesi Tamás III.Department of Medicine The measurement of blood pressure and hypertension Handout Fenyvesi Tamás III.Department of Medicine 1 2 History of the blood pressure measurement direct Stephen HALES 1726 horse carotid indirect Riva-Rocci

More information

Blood Pressure Competency Training

Blood Pressure Competency Training Blood Pressure Competency Training WORKBOOK Table of Contents 3..... Introduction 4..... Background 6..... Equipment Accurate blood pressure What impacts a patient s blood pressure Strategies for proper

More information

GETTING THE NUMBERS RIGHT: ACCURATE MEASUREMENT OF BLOOD PRESSURE

GETTING THE NUMBERS RIGHT: ACCURATE MEASUREMENT OF BLOOD PRESSURE GETTING THE NUMBERS RIGHT: ACCURATE MEASUREMENT OF BLOOD PRESSURE Pamela B. Morris, MD, FACC, FAHA, FASPC, FNLA Professor Medicine and Cardiovascular Diseases Director, Seinsheimer Cardiovascular Health

More information

Is Traditional Clinic Blood Pressure Dead?

Is Traditional Clinic Blood Pressure Dead? Royal College of Physicans May 16 th 2017 Is Traditional Clinic Blood Pressure Dead? Professor Bryan Williams MD FRCP FAHA FESC Chair of Medicine UCL Director National Institute for Health Research Biomedical

More information

Title: Parenteral Iron Therapy for Anemia: A Clinical and Cost-Effectiveness Review

Title: Parenteral Iron Therapy for Anemia: A Clinical and Cost-Effectiveness Review Title: Parenteral Iron Therapy for Anemia: A Clinical and Cost-Effectiveness Review Date: 14 February 2008 Context and policy issues: Anemia is a complication of chronic diseases and commonly occurs in

More information

Primary hypertension in adults

Primary hypertension in adults Primary hypertension in adults NICE provided the content for this booklet which is independent of any company or product advertised Hypertension Welcome NICE published an updated guideline on the diagnosis

More information

Surveillance report Published: 8 June 2017 nice.org.uk. NICE All rights reserved.

Surveillance report Published: 8 June 2017 nice.org.uk. NICE All rights reserved. Surveillance report 2017 Antenatal and postnatal mental health: clinical management and service guidance (2014) NICE guideline CG192 Surveillance report Published: 8 June 2017 nice.org.uk NICE 2017. All

More information

Drug Class Literature Scan: Pancreatic Enzymes

Drug Class Literature Scan: Pancreatic Enzymes Copyright 2012 Oregon State University. All Rights Reserved Drug Use Research & Management Program Oregon State University, 500 Summer Street NE, E35 Salem, Oregon 97301-1079 Phone 503-947-5220 Fax 503-947-1119

More information

How do we diagnose hypertension today? Presentation Subtitle

How do we diagnose hypertension today? Presentation Subtitle How do we diagnose hypertension today? Presentation Subtitle Renata Cífková Case 1 JM, a 64-year-old lady referred to our center because of undesirable effects of her antihypertensive medication Personal

More information

Research. A comparison of blood pressure measurement over a sleeved arm versus a bare arm. The measurement of blood pressure is one of the

Research. A comparison of blood pressure measurement over a sleeved arm versus a bare arm. The measurement of blood pressure is one of the A comparison of blood pressure measurement over a sleeved arm versus a bare arm Grace Ma MD, Norman Sabin MD, Martin Dawes MBBS MD @ See related article page 591 DOI:10.1503/cmaj.070975 Abstract Background:

More information